0% found this document useful (0 votes)
28 views

Ecourses - Ndri.res - in Moodle File - PHP 92 Lesson 6

Static characteristics of measuring instruments include accuracy, sensitivity, reproducibility, drift, static error, and dead zone. Accuracy refers to how close an instrument reading is to the true value, while precision refers to the clarity and consistency of readings. Static error is the difference between the best measured value and the true value of the quantity being measured. An instrument's performance characteristics are determined through static calibration, which involves comparing the instrument to a standard of known accuracy.

Uploaded by

Amit Kr Godara
Copyright
© © All Rights Reserved
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
28 views

Ecourses - Ndri.res - in Moodle File - PHP 92 Lesson 6

Static characteristics of measuring instruments include accuracy, sensitivity, reproducibility, drift, static error, and dead zone. Accuracy refers to how close an instrument reading is to the true value, while precision refers to the clarity and consistency of readings. Static error is the difference between the best measured value and the true value of the quantity being measured. An instrument's performance characteristics are determined through static calibration, which involves comparing the instrument to a standard of known accuracy.

Uploaded by

Amit Kr Godara
Copyright
© © All Rights Reserved
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 5

12/22/2014 ecourses.ndri.res.in/moodle/file.php/92/Lesson_6.

htm

Module 3. Characteristics of instruments and measurement systems
Lesson 6
STATIC CHARACTERISTICS OF MEASURING INSTRUMENTS ­ I
6.1  Introduction
6.2  Static Calibration
6.3  Static Characteristics
 6.3.1  Scale range and scale span
 6.3.2  True value
 6.3.3  Accuracy          
 6.3.4  Static error
6.4  Numericals
6.1  Introduction

When considering a measurement instrument, it is important to have a clear understanding of all the
parameters  involved  in  defining  the  characteristics  of  the  measurement  device.  By  knowing  the
accuracy  and  resolution  requirements  for  your  application,  you  can  compute  the  total  error  in  the
measurement device you are considering and verify that it satisfies your needs. There are number of
important performance parameters discussed in the following text, but they should not be considered
the ultimate or only parameters to take into account. It is worthwhile to ask the instruments supplier to
clarify  the  meaning  of  the  specifications  in  the  instrument  data  sheet.  Not  knowing  the  true
performance of your instrument could lead you to incorrect readings, and the cost of this error could
be very high.

The  performance  characteristics  of  instruments  and  measurement  systems  can  be  divided  into  two
distinct categories, viz., the Static characteristics, and the Dynamic characteristics. Some applications
involve the measurement of quantities that are either constant or vary very slowly with time. Under
these  circumstances,  it  is  possible  to  define  a  set  of  criteria  that  gives  a  meaningful  description  of
quality  of  measurement  without  interfering  with  dynamic  descriptions  that  involve  the  use  of
differential equations. The characteristics in this set of criteria are called Static Characteristics. Thus
the  static  characteristics  of  a  measurement  system  are  those  which  must  be  considered  when  the
system or instrument is used under a condition not varying with time.

However, many measurements are concerned with rapidly varying quantities. In such cases we must
examine the dynamic relations which exist between the output and the input. This is normally done
with  the  help  of  differential  equations  or  other  methods.  Performance  criteria  based  upon  dynamic
relations constitute the Dynamic Characteristics.

6.2  Static Calibration
All the static performance characteristics are obtained in one form or another by a process called static
calibration.  The  calibration  procedures  involve  a  comparison  of  the  particular  characteristic  with
either  a  primary  standard,  a  secondary  standard  with  a  higher  accuracy  than  the  instrument  to  be
calibrated, or an instrument of known accuracy. It checks the instrument against a known standard and
subsequently  to  errors  in  accuracy.  Actually  all  measuring  instruments  must  be  calibrated  against
some reference instruments which have a higher accuracy. Thus reference instruments in turn must be
calibrated against instrument of still higher grade of accuracy, or against primary standard, or against
http://ecourses.ndri.res.in/moodle/file.php/92/Lesson_6.htm 1/5
12/22/2014 ecourses.ndri.res.in/moodle/file.php/92/Lesson_6.htm

other  standards  of  known  accuracy.  It  is  essential  that  any  measurement  made  must  ultimately  be
traceable to the relevant primary standards.
6.3  Static Characteristics

The main static characteristics include:
            (i) Accuracy,                (ii) Sensitivity,                              (iii) Reproducibility,
            (iv) Drift,                     (v) Static error, and                                (vi) Dead zone
The qualities (i), (ii) and (iii) are desirable, while qualities (iv), (v) and (vi) are undesirable. The above
characteristics have been defined in several different ways and the generally accepted definitions are
presented here. Some more quantities have to be defined here which are essential in understanding the
above characteristics.

6.3.1  Scale range and scale span

In  an  analogue  indicating  instrument,  the  measured  value  of  a  variable  is  indicated  on  a  scale  by  a
pointer. The choice of proper range of instruments is important in measurement. The region between
the limits with in which an instrument is designed to operate for measuring, indicating or recording a
physical  quantity  is  called  the  range  of  the  instruments.    The  Scale  Range  of  an  instrument  is  thus
defined as the difference between the largest and the smallest reading of the instrument. Supposing the
highest  point  of  calibration  is  Xmax  units  while  the  lowest  is  Xmin  units  and  the  calibration  is
continuous between the two points, then the instrument range is between Xmin and Xmax . Many times
it is also said that the instrument range is Xmax. The instrument span is the difference between highest
and the lowest point of calibtaration. Thus
                                                    Span = Xmax – Xmin

For example for a thermometer calibrated between 100°C to 400°C, the range is 100°C to 400°C (or
400°C) but the span is 400 – 100 = 300°C.
The  same  is  true  of  digital  instruments.  There  is  another  factor  that  must  be  considered  while
determining the range of the instrument. This is the Frequency Range, which is defined as frequencies
over  which  measurements  can  be  performed  with  a  specified  degree  of  accuracy.  For  example  a
moving iron instrument may have a 0­250 V range and 0­135 Hz frequency range.
6.3.2  True value
The  true  value  of  variable  quantity  being  measured  may  be  defined  as  the  average  of  an  infinite
number of measured values when the average deviation due to the various contributing factors tends
to  zero.  Such  an  ideal  situation  is  impossible  to  realize  the  practice  and  hence  it  is  not  possible  to
determine  the  true  value  of  a  quantity  by  experimental  means.  The  reason  for  this  is  that  there  are
several  factors  such  as  lags,  loading  effects,  wear  or  noise  pick­up  etc.  Normally  an  experimenter
would never know that the value or quantity being measured by experimental means is the ‘true value’
of the quantity or not.
6.3.3  Accuracy         

http://ecourses.ndri.res.in/moodle/file.php/92/Lesson_6.htm 2/5
12/22/2014 ecourses.ndri.res.in/moodle/file.php/92/Lesson_6.htm

Accuracy is the closeness with which an instrument reading approaches the true value of the quantity
being  measured.  Thus  accuracy  of  a  measurement  means  conformity  to  truth.  The  accuracy  of  an
instrument  may  be  expressed  in  many  ways.  The  accuracy  may  be  expressed  as  point  accuracy,
percent of true value or percent of scale range. Point accuracy is stated for one or more points in the
range,  for  example,  the  scale  of  length  may  be  read  with  in  ±  0.2  mm.  Another  common  way  is  to
specify that the instrument is ‘accurate to within ±x percent of instrument span’ at all points on the
scale. Another way of expressing accuracy is based upon instrument range.
 Accuracy is many a time confused with Precision. There is difference in these two terms. The term
‘Precise’  means  clearly  or  sharply  defined.  For  example  an  ammeter  will  possesses  high  degree  of
precision  by  virtue  of  its  clearly  legible,  finely  divided,  distinct  scale  and  a  knife  edge  pointer  with
mirror arrangements to remove parallax. As an example of the difference in meaning of the two terms,
suppose above ammeter can read up to 1/100 of an ampere. Now if its zero adjustment is wrong, every
time  we  take  a  reading,  the  readings  taken  with  this  ammeter  are  not  accurate,  since  they  do  not
confirm to truth on account of its faulty zero adjustment. Though the ammeter is as precise as ever
and  readings  are  consistent  and  clearly  defined  and  can  be  down  to  1/100  of  an  ampere.  The
instrument  can  be  calibrated  remove  the  zero  error.  Thus  the  accuracy  of  the  instrument  can  be
improved upon by calibration but not the precision.
6.3.4  Static error

Measurements done with an instrument always involve errors. No measurement is free from errors. If
the precision of the equipment is adequate, no matter what its accuracy is, a discrepancy will always
be observed between two measured results. Since the accuracy of an instrument is measured in terms
of its error, an understanding and evaluation of the errors is thus essential.
Static  error  is  defined  as  the  difference  between  the  best  measured  value  and  the  true  value  of  the
quantity. Then:
                Es = Am – At                           
Where,     Es = error,
    Am = measured value of quantity, and
    At = true value of quantity.

Es is also called the absolute static error of quantity A. The absolute value of error does not indicate
precisely the accuracy of measurement. For example, an error of ±2 A is negligible when the current
being­measured is of the order of 1000 A while the same error highly significant if the current under
measurement is 10 A. Thus another term relative static error is introduced.  The relative static error is
the ratio of absolute static error to the true value of the quantity under measurement. Thus the relative
static error Er is given by:

               
            Percentage static error % Er = Er x 100                      

http://ecourses.ndri.res.in/moodle/file.php/92/Lesson_6.htm 3/5
12/22/2014 ecourses.ndri.res.in/moodle/file.php/92/Lesson_6.htm

Static Correction
It is the difference between the true value and the measured value of the quantity, or
                                                δC = At ­ Am  

6.4  Numericals                                                                     
1.  A meter reads 115.50 V and the true value of the voltage is 115.44 V. Determine the static error,
and the static correction for this instrument.
Solution:
           The error is:   Es = Am – At = 115.50 – 115.44 = +0.06 V
            Static correction  δC = At ­ Am   = ­0.06 V.

2.  A thermometer reads 71.5 °C and the static correction given is +0.5°C. Determine the true value of
the temperature.
Solution: 
                           True value of the temperature
                            At = Am + δC = 71.5+ 0.5 = 72.0°C.

3.  A  thermometer  is  calibrated  for  the  range  of  100°C  to  150°C.  The  accuracy  is  specified  within
±0.25 percent. What is the maximum static error?
Solution:         
                 Span of thermometer = 150 – 100 = 50°C 

                Maximum static error =  
4.   An analogue indicating instrument with a scale range of 0 – 2.50 V shows a voltage of 1.46 V. A
voltage has a true value of 1.50 V. What are the values of absolute error and correction? Express the
error as a fraction of the true value and the full scale deflection.
Solution: 

             Absolute error = Am – At
                                   = 1.46 – 1.50 = ­0.04 V
            Absolute correction δC = δA = +0.04 V

            Relative error    
           
Relative error expressed as a percentage of full scale division

           =
5.  A pressure indicator showed a reading as 22 bar on a scale range of 0­25 bar. If the true value was
21.4 bar, determine:

http://ecourses.ndri.res.in/moodle/file.php/92/Lesson_6.htm 4/5
12/22/2014 ecourses.ndri.res.in/moodle/file.php/92/Lesson_6.htm

i)      Static error
ii)    Static correction
iii)  Relative static error
Solution:
i)    Static error =  22 – 21.4 = + 0.6 bar
ii)   Static correction = ­ (+0.6) = ­ 0.6 bar
iii)  Relative error = 0.6 / 21.4 = 0.028 or 2.8 %

http://ecourses.ndri.res.in/moodle/file.php/92/Lesson_6.htm 5/5

You might also like