PROC-TC-012 Procedure For Calibration Flask, Cylinder SOP (PROC - C)
PROC-TC-012 Procedure For Calibration Flask, Cylinder SOP (PROC - C)
PROC-TC-012
Procedure for calibration
Central Organization for Standards and Revision: 1
Quality Control
(COSQC)- Metrology Department Valid after:
PROC - C
WORK PROCEDURE
FOR
Page 1 of 19
Identification:
PROC-TC-012
Procedure for calibration
Central Organization for Standards and Revision: 1
Quality Control
(COSQC)- Metrology Department Valid after:
Table of contents
1. Purpose 3
2. Scope 3
3. Normative References 3
4. Terms And Definition 3
5. General Requirements 5
6. Weighing Equipment Used 6
7. Gravimetric Procedure for Calibration 7
8. Calculations 8
9. Calculations for Measurement Uncertainty 9
10. Associated Documents 14
11. APPENDIX A 18
Page 2 of 19
Identification:
PROC-TC-012
Procedure for calibration
Central Organization for Standards and Revision: 1
Quality Control
(COSQC)- Metrology Department Valid after:
1.0 Purpose
Purpose of this Lab Calibration Procedure is to provide guideline for the calibration of Flask
and Cylinder to Contain.
2.0 Scope
For the purpose of this Lab Calibration Procedure following definitions shall apply:
4.1 Calibration
The set of operations which establishes under specific conditions the relationship between
value indicated by a measuring instrument or measuring system or values represented by a
material measure and the corresponding known value of the measurand.
4.2 Traceability
The property of a measurement result whereby it can be related to appropriate standards
generally national or international standards through an unbroken chain of comparisons.
4.3 Measurement Standards
A Material measure, measuring instrument or system intended to define, realize, conserve or
reproduce a unit by one or more known value of a quantity, in order to transmit them to other
measuring instruments by comparison.
Note
In this method a measurement standard is the “Unit under Calibration”.
4.4 UUC (Unit under Cal)
Unit under Cal is the instrument, which is being calibrated by this method statement.
4.5 Reference Standard
A standard generally of a highest metrological quality available from which measurements
made at a given location are derived.
Page 3 of 19
Identification:
PROC-TC-012
Procedure for calibration
Central Organization for Standards and Revision: 1
Quality Control
(COSQC)- Metrology Department Valid after:
Page 4 of 19
Identification:
PROC-TC-012
Procedure for calibration
Central Organization for Standards and Revision: 1
Quality Control
(COSQC)- Metrology Department Valid after:
6 Weighing Equipment
Page 5 of 19
Identification:
PROC-TC-012
Procedure for calibration
Central Organization for Standards and Revision: 1
Quality Control
(COSQC)- Metrology Department Valid after:
In case, where standard deviation of liquid measurement is specified, than it is possible to use
it as acceptance criterion instead of repeatability and linearity parameters of a weighing
instrument. However, it needs to be assumed, that standard uncertainty is not bigger than
twofold or threefold of reading unit d.
1- After cleaning and drying, weigh the empty flask including the stopper.
2- Place an appropriate sized funnel in the flask to discharge the water below the stopper.
3- Fill from beaker or supply line, maneuvering the funnel So as to wet the entire neck
below the stopper. Let stand for about 2 min to allow the walls to drain.
4- Place the flask under a burette and complete filling and setting the meniscus, taking care
not to splash water on the walls, after which place stopper in neck to lessen evaporation
and weigh.
NOTE: A pipette or dropper with a finely drawn tip may be used to adjust the meniscus instead of
the burette.
Page 6 of 19
Identification:
PROC-TC-012
Procedure for calibration
Central Organization for Standards and Revision: 1
Quality Control
(COSQC)- Metrology Department Valid after:
9.0 Calculations
a) Calculate the mass of water Contain for each cycle by subtracting the balance reading
recorded at the end of the previous cycle from the reading recorded at the end of the cycle. If
appropriate, add the mass of water lost by evaporation as determined in the previous
paragraph.
b) To convert each mass value to a volume at 20 °C (the standard reference temperature for the
calibration of Flask) it must be divided by the density of water corrected to 20 °C. A
correction for air buoyancy, which varies with the air density, must also be applied. Both
these factors are taken into account by multiplying the each mass value by a correction factor
Z. A table specifying the value of Z for a temperatures range of 15 °C and 30 °C and air
pressure range of 80 kPa to 105 kPa. The correction factor is also given as an equation in
ISO/TR 20461:2000.
(4)
(5)
(6)
Page 7 of 19
Identification:
PROC-TC-012
Procedure for calibration
Central Organization for Standards and Revision: 1
Quality Control
(COSQC)- Metrology Department Valid after:
Where,
V = volume in ml
W2 = weighed value of the vessel with calibration liquid in g
W1 = weighed value of the empty vessel in g
ρL = density of the air in g/cm³
ρG = density of the weights in g/cm³ (standardized value for steel weights: 8000 kg/m³
= 8 g/cm³)
ρw = density of the calibration liquid in g/cm³
αc = cubic expansion coefficient in 1/K
td = temperature of the volumetric apparatus in °C (for simplification equated with the
temperature of the calibration liquid)
tb = reference temperature in °C
The water density ρw and the air density ρL can be calculated with the formulas stated in
appendix A
c) Add together each of the 10 volumes calculated in the previous paragraph and divide by 10 to
give the mean volume at 20 °C. This can be expressed in microliters or milliliters.
(7)
1. Random error is expressed in volumetric units: microliters [ul] and percent’s [%]. Random
error is marked with CV symbol. The measure of random error is standard deviation
calculated by a formula
(10)
(11)
U
u(Uw ) (16)
3
U – Standard uncertainty of a balance
Page 9 of 19
Identification:
PROC-TC-012
Procedure for calibration
Central Organization for Standards and Revision: 1
Quality Control
(COSQC)- Metrology Department Valid after:
Note: According to the above formula you need to determine uncertainty of particular
measurements in a series, and then determine standard uncertainty for a series of 10
measurements:
d 2
u (m) 10 m c w
2
(21)
2 3
Page 10 of 19
Identification:
PROC-TC-012
Procedure for calibration
Central Organization for Standards and Revision: 1
Quality Control
(COSQC)- Metrology Department Valid after:
d tw
u (t w ) (24)
3
dtw –Thermometer resolution of water temperature measuring
2.9 Air temperature ta:
d
u(t a ) ta (25)
3
dta –Thermometer resolution of air temperature measuring
F V20
cw
m m (29)
3.2 A coefficient related to volume determination V20 is calculated in the following formula:
Page 11 of 19
Identification:
PROC-TC-012
Procedure for calibration
Central Organization for Standards and Revision: 1
Quality Control
(COSQC)- Metrology Department Valid after:
V20
cV 20 1 (30)
V20
3.3 A coefficient related to water temperature determination is calculated in the following formula:
c tw 2,1 10 7 m (31)
3.4 A coefficient related to air temperature determination is calculated in the following formula:
m ( k 2 t a 0 k 3 ) k1 p a
c ta (32)
w2 (t a t a 0 )2
3.5 A coefficient related to relative air humidity determination is calculated in the following
formula:
m k2 ta k3
c (33)
w2 t a t a0
3.6 A coefficient related to air pressure determination is calculated in the following formula:
m k1
c pa (34)
w t a t a0
2
U k u(V20 )
(37)
Page 12 of 19
Identification:
PROC-TC-012
Procedure for calibration
Central Organization for Standards and Revision: 1
Quality Control
(COSQC)- Metrology Department Valid after:
Page 13 of 19
Identification:
PROC-TC-012
Procedure for calibration
Central Organization for Standards and Revision: 1
Quality Control
(COSQC)- Metrology Department Valid after:
Page 14 of 19
Identification:
PROC-TC-012
Procedure for calibration
Central Organization for Standards and Revision: 1
Quality Control
(COSQC)- Metrology Department Valid after:
APPENDIX A
The density of pure water is normally provided from formulations given in the literature.
Batista and Paton provide an overview of common formulations used in practice. However it
is generally accepted that the formulation given by Tanaka provides a good basis for
standardization.
g/ml A.1
Where:
t = water temperature, in °C
a1 = -3,983035 ºC
a2 = 301,797 ºC
a3 = 522528,9 (ºC)2
a4 = 69,34881 ºC
a5 = 0,999974950 g/mL
Note: a5 is the density of SMOW water under one atmosphere at this temperature. Many
users of water rely on tap water instead of SMOW. Thus a5 must be changed accordingly to
reflect the density of the water used. The full equation of state for water provided by the
International Association for Properties of the Water Substance (IAPWS) can also be used to
determine the density of the used water and a formulation based on this equation is given in
ref. This provides an alternative to Tanaka and should be used if temperatures above 30 °C
are used.
Where pure water is not available, the relative density of the water may be determined and the
chosen formulation used to determine the temperature expansion factors with insignificant
loss of accuracy.
The air density can be determined according to the formula for the density of moist air -
CIPM-2007
A.2
p =ambient pressure, in Pa
T =thermodynamic temperature = 273,15 + t / ºC, in K
Xv= mole fraction of water vapour
xCO2= mole fraction of carbon dioxide in laboratory air
Z =compressibility factor
If the laboratory ambient conditions are within the limits given then Spieweck’s equation can
be used. The limits are: pressure between 940 hPa and 1080 hPa; temperature between 18 ºC
and 30 ºC; humidity less than 80 %,
Page 15 of 19
Identification:
PROC-TC-012
Procedure for calibration
Central Organization for Standards and Revision: 1
Quality Control
(COSQC)- Metrology Department Valid after:
A.3
Where :
tA= ambient temperature, in °C
pA= ambient pressure, in hPa
hr= relative air humidity, in %
k1 = 3,4844 x 10-4 ºC/hPa
k2 = -2,52 x 10-6 g/mL
k3 = 2,0582 x 10-5 ºC
APPENDIX B
PHV-1
PHV-2
Page 16 of 19
Identification:
PROC-TC-012
Procedure for calibration
Central Organization for Standards and Revision: 1
Quality Control
(COSQC)- Metrology Department Valid after:
APPENDIX C
Page 17 of 19
Identification:
PROC-TC-012
Procedure for calibration
Central Organization for Standards and Revision: 1
Quality Control
(COSQC)- Metrology Department Valid after:
Calibration Certificate
Calibration Certificate
Central Organization for Standardization and Quality Control (COSQC)
Metrology Department - PHYSICS SECTION
P.O. Box13032 Algadria street, Baghdad ,Tel:7785180 - E-Mail : [email protected]
Certificate No.: PHV
Date of issue :
Customer
Name:
Address:
Calibration information
Date of calibration: Due to:
Place of calibration: Physics measurement lab
Method(s) of
calibration:
Calibrated quantity:
Results of calibration: Attached a complete result in Annex 1 of this certificate
Measurement There reported expanded uncertainty is based on GUM Standard and the standard
uncertainty: Uncertainty multiplied by coverage factor k=2 to give confidence level of 95%
Metrological The traceability of measurement results to the SI units is assured by the Nationals tandard maintained at
traceability: Central Organization for standardization and Quality Control through calibration at :
Environmental
conditions of Temp. ( C): ±1ºC R. H.( %) ±5% Pressure ( mbar) ±1mbar
calibration:
Observations, opinions
The results and corrections in Annex 1 should be taken into consideration
or recommendations:
Approved by:
Page 18 of 19
Identification:
PROC-TC-012
Procedure for calibration
Central Organization for Standards and Revision: 1
Quality Control
(COSQC)- Metrology Department Valid after:
Calibration Certificate
Central Organization for Standardization and Quality Control (COSQC)
Metrology Department - PHYSICS SECTION
P.O. Box13032 Algadria street, Baghdad ,Tel:7785180 - E-Mail : [email protected]
Certificate No.: PHV
Date of issue :
Annex 1 ( Results )
Page 19 of 19