CB (E2) Service Manual State: 01/2002 Created: 03/2002 / Jochen Tussinger
CB (E2) Service Manual State: 01/2002 Created: 03/2002 / Jochen Tussinger
SERIAL-NO. 01-27954 à
Order-No. 9040-0012 / CB 150
Order-No. 9040-0013 / CB 210
CB (E2) Service Manual
Contents
1 Modification levels ...................................................................................................................3
2 Unit overview ..........................................................................................................................4
2.1 The Controller MB1...........................................................................................................5
2.2 Short description of the MB1 Controller ..............................................................................6
3 Function ............................................................................................................................... 12
3.1 The CO2-measuring principle........................................................................................... 12
3.2 Function of the Heating System ....................................................................................... 12
3.3 Flow-Chart of the heating function (basis CB 150 wiring diagram)...................................... 13
3.4 Controller MB1 PIN description (Input / Output) ................................................................ 14
3.5 Function of the CO2 System ............................................................................................ 15
3.6 Flow Chart of the CO2 System (basis CB wiring diagram).................................................. 15
3.7 Function of the Permadry® system .................................................................................. 16
3.8 Flow Chart of the Permadry® system (basis CB wiring diagram) ........................................ 17
3.9 Description of the Function of the Fan Control .................................................................. 17
3.10 Sterilization Mode........................................................................................................ 18
3.11 Hot-air sterilization....................................................................................................... 18
3.12 Hot-air sterilization with inner chamber contaminated with highly infective material .......... 20
4 Trouble Shooting................................................................................................................... 22
5 Most common service work .................................................................................................... 25
5.1 Changeing of the fan....................................................................................................... 26
5.2 Take out of the electronic component board ..................................................................... 28
5.3 Opening of the rear service lid to achieve the area of the air jacket .................................... 29
5.4 Setting of the door heating............................................................................................... 31
5.5 CO2-Reference Measurement .......................................................................................... 32
5.5.1 Measuring of CO2 indirectly via the pH of the cell medium .......................................... 33
5.5.2 Measuring of CO2 directly via chemical indicator tubes ............................................... 34
5.5.3 Measuring of CO2 directly via a electronic measuring device....................................... 35
6 Calibration ............................................................................................................................ 36
6.1 Definition of calibration .................................................................................................... 36
6.2 References for calibration................................................................................................ 36
6.3 Tolerance of the adjustment ............................................................................................ 36
Calibration instructions for CO2 incubator CB with screen controller MB1 ........................................ 37
Temperature / CO2 / O2 controller ................................................................................................ 37
Temperature calibration ............................................................................................................ 37
Calibration (alignment) of the temperature controller ................................................................... 38
Reading out of the actual values: ............................................................................................... 39
Entries: .................................................................................................................................... 39
Result Calibration (alignment) of the temperature controller ......................................................... 39
CO2 calibration (alignment)........................................................................................................ 40
O2 calibration (alignment) .......................................................................................................... 41
7 Maintenance ......................................................................................................................... 48
8 Explosion Drawings of the CB /E2.......................................................................................... 51
8.1 CB Component Board..................................................................................................... 51
8.2 CB Kettle / Sensors / Heaters / Permadry ......................................................................... 52
8.3 CB Door / Sealing / Inner parts ........................................................................................ 53
2
CB (E2) Service Manual
1 Modification levels
The CB CO2 Incubator E2 is a further development of CB CO2 Incubator E1. Especially the Display
controller MB1 inside the red triangle is conspicuous.
The whole electronic is placed inside the lower part of the CB CO2 Incubator (Front Access
Maintenance). Theres no I-box as at the CB CO2 Incubator E1 at the top of the chamber.
08:43:55 20.11.01
W X
TEMP
37.0 36.8 °C EXIT
CO2
5.0 4.9 %
O2
25.0 25.2 %
3
CB (E2) Service Manual
2 Unit overview
(A)
(B)
(C)
(D)
(E)
(F)
(G)
(H)
(I)
(J)
FRONT
(M)
(L)
(K)
K
A) Display controller MB1 for temperature and CO2 as well as O2 (option)
B) Connection socket for low tension supply (option)
C) CO2 sensor
D) Gas mixing head
E) PT 100 temperature probe
F) O2 sensor (option)
G) Internal socket 230V (max. 3 A) (option)
H) Shelf holder bar
I) Shelves
J) Shelf holder
K) Lower housing cover
L) Permadry® water basins
M) Inner glass door
N) Measuring access port
4
CB (E2) Service Manual
Operating Light
08:43:55 20.11.01
W X EXIT-Button
37.0 36.8 °C EXIT
TEMP
C2
O 5.0 4.9 %
(to leave a
O2 25.0 25.2 % Menue)
CONFIG W RESET MENUE VIEW->
ON / OFF Button
(Stand-by)
ENTER-Button
((to confirm a
selection)
Navigation buttons
(corresponding function
will be assigned)
Mainswitch
For the controller MB1, it is possible to show all setpoints and actual values at the same time.
It is also possible to display the actual values portrary.
5
CB (E2) Service Manual
6
CB (E2) Service Manual
7
CB (E2) Service Manual
8
CB (E2) Service Manual
9
CB (E2) Service Manual
10
CB (E2) Service Manual
,,
11
CB (E2) Service Manual
3 Function
Incubators series CB are suited for cultivation of mammal cells under typical conditions of about 37°C.
The incubator allows the set defined pH conditions by justifying NaHCO3 buffer systems of commercial
cell media by keeping exact CO2-atmosphere inside. CB incubators care for high humidity inside to
avoid increasing of the osmolarity caused by evaporation of the cell media. To reach this, the different
functions heating, sterilization and CO2 injection have to work as a perfect team. This know-how is as
Firmware inside the controller type MB1.
The CO2-measuring procedure of the incubator series CB is characterized by fast reaction times, as
well as the highest accuracy and selectivity. The accuracy of the CO2 measuring system bases on a
single-beam infrared measuring cell, which measures in differential mode, with permanently
alternating transmission characteristic of its semi-conductor filter. Due to this highly developed single-
beam principle with Fabry-Perot interferometer (FPI), disturbance variables and aging phenomena in
the measuring system are almost completely eliminated, so that this measuring system, in contrast to
other measuring procedures, remains practically drift-free between calibrations and is absolutely
selective for CO2.
The CO2–measuring cell contains a measuring section inside in which the absorption of infrared light
depends on the number of CO2 -molecules in the beam path. This number of CO2 –molecules changes
with the ambient pressure in relation to a constant volume. The distances between the molecules are
consequently pressure-dependent. The collision frequency of the IR-beam with CO2-molecules
increases therefore by increasing pressure.
For this reason, the ambient pressure must be compensated in order to correct the display reading of
the CO2 -concentration in VOL.-%. This is achieved by entering the altitude of the site above the sea
which is described in this manual.
12
CB (E2) Service Manual
The Pt100 temperature probe is equipped with 3 cables, two red and one white cable.
The white cable is connected to pin 1 at E1 – Pt100, the two red cables at pin’s 2 and 3 at E1 – Pt100.
To measure the resistance disconnect all three cables from the controller an measure between the
white cable and one of the red cables, do not measure between both red cables.
Ω
Measuring of the Pt100 resistance between the white
cable PIN 1 and one of the two red cables PIN 2 or 3
13
CB (E2) Service Manual
T (°C) 0 1 2 3 4 5 6 7 8 9 10
-10 96,086 96,478 96,870 97,262 97,653 98,045 98,436 98,827 99,218 99,609 100,000
0 100,000 100,391 100,781 101,172 101,562 101,953 102,343 102,733 103,123 103,513 103,902
10 103,902 104,292 104,681 105,071 105,460 105,849 106,238 106,627 107,016 107,404 107,793
20 107,793 108,181 108,570 108,958 109,346 109,734 110,122 110,509 110,897 111,284 111,672
30 111,672 112,059 112,446 112,833 113,220 113,607 113,994 114,380 114,767 115,153 115,539
40 115,539 115,925 116,311 116,697 117,083 117,469 117,854 118,240 118,625 119,010 119,395
50 119,395 119,780 120,165 120,550 120,934 121,319 121,703 122,087 122,471 122,855 123,239
60 123,239 123,623 124,007 124,390 124,774 125,157 125,540 125,923 126,306 126,689 127,072
70 127,072 127,454 127,837 128,219 128,602 128,984 129,366 129,748 130,130 130,511 130,893
80 130,893 131,274 131,656 132,037 132,418 132,799 133,180 133,561 133,941 134,322 134,702
90 134,702 135,083 135,463 135,843 136,223 136,603 136,982 137,362 137,741 138,121 138,500
100 138,500 138,879 139,258 139,637 140,016 140,395 140,773 141,152 141,530 141,908 142,286
For Example: Your resistance measurement system shows you 114,380 Ω this corresponds to 37°C.
Ground TE
Output Phase-controll
E1 – Pt100 3A / 115-230V AC
Heater air-duct
BA2 – 1A / 115-230V AC
E2 – Pt100
Door Heater
BA3 – 1A / 115-230V AC
E3 – Pt100
Collection Alarm
BA1 – 3A / 115-230V AC
E4 – 4 - 20mA
Heater Kettle
BA4 – 60mA / 5V DC
E2 – Pt100 CO2 Solenoid valve
BA5 – 60mA / 5V DC
Volatge output
45mA / 24VDC
14
CB (E2) Service Manual
1 – white
2 – grey To A1+ / A2-
3 – green
4 – yellow
5 – blue
6 – red
7 – brown To Power supply at the Transformer
8 - pink
H1+
1
-5Y1
2
H2-
15
CB (E2) Service Manual
Cold air, produced by the Permadry® -fan, streams through the air-channel and cools the round cold
water basin. The condensation point is fixed by this.
The Permadry® water basin must be placed correctly. There is a description „FRONT“ impressed.
Lettering
„FRONT“
16
CB (E2) Service Manual
The rotation of the Permadry®-Fan is reverse proportional to the rotation of the Mainfan.
The only difference is: The Mainfan is controlled constant with the help of a phase control level, the
Permadry®-Fan is in interval mode which is given by a solid state relay.
17
CB (E2) Service Manual
The hot-air sterilization gets activated in the CB Menu. The sterilization temperature of
187,5° C is adjusted by the manufacturer. It may not be changed.
The CO2-sensor head is especially adjusted for a specific chamber that it belongs to. It cannot
be operated in another chamber. There is an adhesive label with a number on the sensor head
to avoid any mixing up.
18
CB (E2) Service Manual
CB-Menue
Enter Altitude
Sterilization
Enforced steriliz.
Emergency steriliz.
Sterilization end
Humidity Normal
Sollwert Art
NO
Yes
TEMP 187.5
Display during the
CO 2 0.0 0.0 %
sterilization process
O2 0.0 0.0 %
AUTO
HAND
CONFIG W RESET MENUE VIEW->
Caution: Never open the two unit doors during the sterilization process, as the temperature
reaction time will be interrupted and the sterilization will become ineffective.
Due to the sterilization, units in copper version change their colour. This does not have any negative
effect on the function and quality of the unit.
19
CB (E2) Service Manual
If it is not possible to discharge the unit and to remove all necessary parts for safety reason, the hot-air
sterilization can exceptionally be done nonetheless.
Attention! Note that this procedure destroys the CO2 and the O2 sensor (option)!
CB-Menue
Enter Altitude
Sterilization
Enforced steriliz.
Emergency steriliz.
Sterilization end
Humidity Normal
User-Code
Sollwert Art ?
Grenzwert
Offset +00000
User-Code
Sollwert Art?
Grenzwert
Offset +00001
TEMP 187.5
CO2 0.0 0.0 %
Display during the
emergency sterilization
O2 0.0 0.0 %
process
AUTO
HAND
CONFIG W RESET MENUE VIEW->
20
CB (E2) Service Manual
Caution: Never open the two unit doors during the emergency sterilization process, as the
temperature reaction time will be interrupted and the sterilization will become ineffective.
21
CB (E2) Service Manual
4 Trouble Shooting
Safety hints: Do never unplug the CO2 sensor head while the chamber is switched on.
This could result in sensor destruction or initiation problems.
Hint: The tradename of the controller is MB1
Binder offers a calibration kit for this purpose with analyzed test gas
with a CO2 -concentration of 5 Vol.-%. Because the atmospheric
pressure influences the measurement result of the CO2 sensor system,
the altitude of the site has to be taken into consideration.
• Replace the FPI sensor head and check the sensor current against
the expected value again during exposing the sensor head to test
gas.
If the deviation is now within the tolerance of +/- 0,24mA you can
proceed a re-calibration at the MB1 controller as described in chap.
re-calibration to achieve maximum accuracy.
22
CB (E2) Service Manual
MB1 controller display shows • This means the new plugged CO2 sensor head is not initialized
CO2 readings of 19 to 20 Vol.- successfully.
% CO2 also when there is no • Try a restart of the chamber by switching OFF and after 10 s ON an
CO2 inside the chamber e.g. the main switch.
with open door. • If this worse reading still remains the sensor board has to be
replaced and a re-calibration must be carried out.
FPI sensor board: • This means no connection between sensor head and the FPI
Red and green error LED flash sensor board.
alternately • The sensor head is not plugged correctly – try to plug it again.
• The red silicone wire between the Lemo socket for the sensor head
and the FPI sensor board is faulty.
• The soldered contacts of the Lemo socket might be unfixed.
FPI sensor board: • Normal function, no defect.
Green LED flashes and lights
alternately.
FPI sensor board: • Sensor system is faulty. The sensor head and may be the sensor
Red LED is continuously on or board must be replaced. The system must be re-calibrated (see
flashes (green remains off) chap. re-calibration)
during normal operation
CO2 too low inside the The controller MB1 shows also a to low CO2 concentration:
chamber • Check the pressure of the CO2 bottle (max.1.5 bar, min. 1.0 bar)
• Sterilization process has to be off
par example cell-medium is • Check gas inlet valve. When the door is closed the valve has to
colored violet = CO2 too low open with a clicking noise.
• The door switch has to be in closed position when the door is
closed. If it is not closed no CO2 can flow inside the chamber. Press
the door switch by hand and check on the inlet tube whether CO2 is
flowing in
• Check CO2 connection (rear), and the 1mm drill hole of the gas
mixing head. May be there is something jammed.
Binder offers a calibration kit for this purpose with analyzed test gas
with a CO2 -concentration of 5 Vol.-%. Because the atmospheric
pressure influences the measurement result of the CO2 sensor system,
the altitude of the site has to be taken into consideration.
• Change the FPI Sensor Head and check the Sensorcurrent against
the table above during exposing with 5% CO2 Testgas
• If the deviation is within the tolerance of +/- 0,24mA, please make a
re-calibration.
• If the value with the tolerance of +/- 0,24mA at the signal output of
the FPI Sensor board is reached but the controller doesn’t show a
value between 4,8 and 5,2 Vol.-% CO2, there was a wrong
calibration done. Check the correct setting of the altitude and make
a re-calibration.
• Make sure, that you have reset the old calibration.
The CO2 concentration doesn’t • Check the altitude setting at the MB1 controller
go down after a door opening • The Sensor system is defective or wrongley calibrated
of approx. 20s to 0 Vol.-% CO2 • Make a re-calibration and check if the value is now ok.
• If not, you have to change the sensor system. Change first only the
FPI sensor head and make a re-calibration.
The CO2 concentration drifts • If the signal doesn’t reach a stable value, you have to change the
during exposing with 5%- sensor system, first only the FPI sensor head, after that you have to
Testgas make a re-calibration
CO2 Controller shows „-1999“. Means measurement value underrange:
There’s also a red alarm-bell • The Sensor signal between the Sensor Board and the Sensorinput
blinking E4 at the controller MB1 is low. (under 4mA)
• Check if the sensor head puts correctly
• Check the cable between the Sensor Board and the controller MB1
looking for a break or a faulty contact.
• Check the cable between the Sensor Board and the sensor
• If the controller gets the correct signal, but shows still –1999, the
input at the controller is defect. The controller must be changed.
• There’s no current from the transformer (24VDC). Check the power
supply for the FPI sensor. Possibly, the fuse is blown (T500mA).
CO2 controller shows ”9999” Means measurement value exceeding
• The sensor signal between the FPI sensor board and the sensor
input at the controller MB1 at Pin E4 is to high (over 20mA) or
there’s a initialization error – try a restart.
• If the red LED of the sensor board is lit permanent, you have to
change the sensor board.
• If not, check the sensor current and the sensor output of the sensor
board.
• Try a re-calibration with analyzed testgas
• If the error reshows, you have to change the controller
To low humidity inside the • The stpoint of the door heating is to high. (This is a internal setting
chamber of the controller. Please see description in chapter 4.4)
Humidity to high inside the • The Permadry® fan doesn’t run. This fan is for blowing cold air
chamber throgh the air chanel to the cold water basin
Condensation inside
Condensation at the glass • Check the outer door and the inner door fits not tightly
door • Check the two door sealing.
• Adjust the door keeping
• The door heating element is faulty – Check the resistance (normally
278 Ω).
24
CB (E2) Service Manual
Please note:
25
CB (E2) Service Manual
Note:
The fan motor turns when the chamber is
switched on.
26
CB (E2) Service Manual
27
CB (E2) Service Manual
28
CB (E2) Service Manual
5.3 Opening of the rear service lid to achieve the area of the air jacket
This is necessary for following service works.
Note:
The fan motor turns when the chamber is
switched on.
29
CB (E2) Service Manual
4
Inner view:
♦ left: double temperature probe for inner
temperature and safety device
♦ middle: gas mixing head
♦ socket of the FPI-sensor head
30
CB (E2) Service Manual
CAUTION:
08:43:57 15.05.00
The System-Code is only for Service-
Benutzer-Daten EXIT
Konfiguration 2
Not-Sterilisation
Engineers. Don’t give the System-Code
Konfiguration
Steri Abbruch
1
Not-Sterilisation
Parametrierung
Feuchte
Steri Abbruch
Not-Sterilisation
to the customer, it is possibl eto
3. Feuchte
Steri Abbruch change also the parameters, so that the
Feuchte
System-Code ?
4. +00000
System-Code ?
5. +02802
31
CB (E2) Service Manual
The CO2-measuring procedure of the incubator series CB is characterised by fast reaction times, as
well as the highest accuracy and selectivity. The accuracy of the CO2 measuring system bases on a
single-beam infrared measuring cell, which measures in differential mode, with permanently
alternating transmission characteristic of its semi-conductor filter. Due to this highly developed single-
beam principle with Fabry-Perot interferometer (FPI), disturbance variables and aging phenomena in
the measuring system are almost completely eliminated, so that this measuring system, in contrast to
other measuring procedures, remains practically drift-free between calibrations and is absolutely
selective for CO2.
The CO2–measuring cell contains a measuring section inside in which the absorption of infrared light
depends on the number of CO2 -molecules in the beam path. This number of CO2 –molecules changes
with the ambient pressure in relation to a constant volume. The distances between the molecules are
consequently pressure-dependent. The collision frequency of the IR-beam with CO2-molecules
increases therefore by increasing pressure.
For this reason, the ambient pressure must be compensated in order to correct the display reading of
the CO2 -concentration in VOL.-%. This is achieved by entering the altitude of the site above the sea
which is described in the CB operating manual.
In principal there are 3 possibilities to test CO2-concentration inside a incubator which are described
later on.
32
CB (E2) Service Manual
By use of the indirect determination of CO2 concentration via the pH-value of the nutrient it is possible
to check the CO2 -concentration inside the chamber. This method can not be used for re-calibration of
the sensor system but it is a simple method to test for the correct CO2 concentration without any
special CO2 -measuring equipment. Only a accurate pH indicator or a pH measuring electrode is
necessary, but this is standard equipment in cell culture laboratories.
This method bases on the acid base equilibrium of the buffer system in the nutrient.
The most common media are buffered by NaHCO3. With the help of the pH-value in the medium, a
conclusion can thus be drawn about the concentration of CO2.
This conclusion is directly possible using the diagram below, which expresses the interrelationship
between CO2 concentration in vol. % and the pH of different NaHCO3 buffered media.
We recommend the incubation of an empty sample with medium for 1/2 day under the same
conditions as the cells. The incubation can be performed in a cell culture bottle or in a 50 ml Falcon
tube with open lid.
After gassing, remove the empty sample from the incubator and within 5 minutes measure the pH1-
value with a glass electrode. During the measurement the medium should have the least possible
surface contact with the ambient air, so that the CO2 can only diffuse out slightly. Therefore,
transfusing should be avoided. A significant downward movement will only be observed after 5
minutes, allowing sufficient time for measurement. Naturally, pH-test strips can also be used (pH6-8
range not bleeding).
Following graph shows the pH-value of NaHCO3 buffered media as function of the CO2-concentration:
Vol.% CO2
18 Trade names of common
3,70 media:
16
NaHCO3
14 [g/l]
12
DMEM 3,70
2,20 BME 2,20
10
MEM 2,20
8 Medium 199 2,20
Mc Coy 2,20
1,20
6 F10 1,20
F12 1,20
4 0,85
2 0,35
Example:
If a pH of 7,2 is measured in a medium which is buffered with 2,20 g NaHC03 per liter there must be 8
Vol.-% CO2 in the surrounding of this medium.
33
CB (E2) Service Manual
This is a common do it yourself test of the users. A chemical color reaction in a glass tube shows the
CO2-concentration. A standardized volume of air from the inside of the incubator has to be sucked
through this glass tube to get a quantitative test result. Therefor a special hand pump must be used
with a standardized suction volume.
Notes:
34
CB (E2) Service Manual
The easiest way to measure CO2 concentration are electronic sensor systems. BINDER offers the
portable measuring device model CTM 01 which were specially designed to measure CO2
concentration and temperature inside of CO2 incubators. The CTM 01 can be used both for reference
measurements in certified laboratories, as well as for service purposes.
Description of CTM01:
• The measuring system consists of 2 parts: The measuring station, which remains outside the CO2
incubator, and the sensor housing, which is put in the center of the usable volume of the
incubator. Both parts are connected by a ribbon cable, which can be laid across the door gasket of
the incubator without causing leakage.
• The measuring cell is placed directly in the atmosphere to be measured. This means that there is
no requirement for gas pumps, whose inconstant flow rates can cause inaccuracies. When
converting the measured variable, partial pressure CO2, to the display value, VOL.-% CO2, the
respective altitude above sea level is entered and taken into account, so that the display value
always corresponds immediately to the real concentration in VOL.-% CO2 , irrespective of the
place of use.
• The temperature is determined via a PT100 temperature sensor and is displayed on the large LED
display in 10 seconds alternation with the CO2 - concentration.
• The measured data of temperature and CO2 -concentration are output to an RS 422 interface.
Thus the measured data can be simple recorded, stored and graphically represented via PC using
the APT-COM communication software developed by BINDER.
• For monitoring, the nominal values set in the incubator can be entered into the mobile measuring
system. If these nominal values are exceeded or fallen below by permanently set tolerances, this
is reported by the unit optically and acoustically, as well as via a potential-free alarm contact. The
alarm function only becomes active if both measured values were previously within the tolerance
limits.
35
CB (E2) Service Manual
6 Calibration
When is it worth to correct a deviation between the value measured and the value displayed on the
MB1 controller?
The following table gives you the conditions for an ideal adjustment of a temperature deviation,
Deviations which are smaller than this deviations do not require an adjustment.
Important:
The values on the above chart can only be used if the value of measurement uncertainty on the
calibration certificate of the measuring instrument is smaller than the values of the chart.
When this is not the case, then you must take as limit value for an adjustment the measurement
uncertainty value of the calibration certificate.
36
CB (E2) Service Manual
Type:
Identification No.:
Traceability:
Date of calibration:
Measuring uncertainty
of the measuring device:
Temperature calibration
The temperature calibration is effected in one single procedure at the most common working
temperature (mostly used 37°C) in thermal stationary condition. The unit is checked in empty
condition with one central shelf.
The reference sensor for temperature is led into the empty inner chamber through the door
sealing and is positioned in the middle of the usable volume on one central shelf. Both unit
doors remain closed during the calibration. The incubator has to be pre-heated at the
calibration temperature for at least one hour.
The adjustment of the temperature controller is only necessary in case there is a deviation between
the temperature, measured in centre of the useful volume and the reading of the actual temperature
displayed on the controller MB1 or its safety controller of more than +/-0.2 K. In case the reference
instrument has a measuring uncertainty larger than +/-0.2 K (see its calibration certificate), this is the
confidence criteria.
Temperature set-point W
______________°C
Display reading of the controller
______________°C
Display reading X of the safety controller in menu “user-settings”
______________°C
Display reading of the reference instrument
______________°C
Divergence actual temperature - Reading reference instrument
______________°C
Divergence actual temperature of the safety controller -
Reading reference instrument ______________°C
37
CB (E2) Service Manual
Yes No
Calibration (adjustment) of the temperature controller o o
necessary
Overview :
First of all the former calibration is reset (see chap. reset of the calibration). The input value
correction is deactivated afterwards (set the state of the input val. correction to Off). Then
return to the normal operation mode and let adjust the incubator for 1 hour to the calibration
temperature.
The display reading values of the inner chamber temperature (Analogue input 1), of the preheating
chamber temperature (Analogue input 2) and of the safety controller are set to the display reading
value of the reference instrument. This is done by entering the display value of the reference
instrument to the menu “Input value correction” of analogue input 1, analogue input 2, and of the
safety controller.
Finally the input value correction is activated (set the state of the input val. correction to On).
38
CB (E2) Service Manual
Starting Situation:
During 1 hour without door opening the incubator is adjusted stable to the calibration
temperature (e.g. X = W = 37°C).
The reference instrument shows e.g. 36.5°C.
Example:
Entries:
After the adjustment of the controller the temperature display reading of the reference
instrument is compared with the display reading of the controller MB1 again:
Temperature set point of the Actual value shown on the test Confidence criteria fulfilled
controller equipment
Yes No
______________°C ______________°C o o
39
CB (E2) Service Manual
The incubator has to be pre-heated at the working temperature (common value 37°C) for at
least half an hour. The altitude of the site above the sea level has to be entered to the
temperature/CO2-controller. The procedure is described in chapter 7.3 of the operation manual.
Afterwards the incubator is switched off at the main switch (1) and the CO2 sensor is pulled out
of the incubator. The unit doors can now remain open during the calibration. The white filter of
the CO2 sensor is removed and the original BINDER calibration cap is pushed on the sensor
tip. The sensor with the calibration cup is put into the socket of the incubator again and the
incubator is switch on at the main switch (1).
Note: The BINDER service uses larger pressure gas bottles with analysed test gases and an other
flow meter. But the procedure remains as described.
After 5 minutes exposing of the sensor to each test gas the CO2-concentration of the gas is
entered to the MB1 controller according to the following description.
40
CB (E2) Service Manual
The corrected value 5 Vol.-% appears on the controller display after 5 seconds.
Afterwards the exposure with test gas is stopped.
O2 calibration (alignment)
The incubator has to be pre-heated at the working temperature (common value 37°C) for at
least half an hour.
The altitude of the site above the sea level has to be entered to the temperature/ CO2 /O2-
controller. The procedure is described in chapter 7.3 of the operation manual.
The chamber doors can be opened during the calibration procedure. The original BINDER
calibration cap is pushed on the O2-sensor tip only when exposed to the test gas, not when
exposed to ambient air.
41
CB (E2) Service Manual
Note: The BINDER service uses a larger pressure gas bottle with analysed test gas and an other flow
meter. But the procedure remains as described.
After 5 minutes exposing of the sensor to ambient air or to the test gas the entrance signal of
the O2 sensor is corrected according to the table below and is entered to the MB1 controller
according to the following description.
The connection between the oxygen concentration of the test gas or the ambient air and the expected
entrance signal to the MB1 controller is presented in the table below.
If other test gases are used, see table in appendix for according tension values.
The measuring value correcting in the following description is carried out according this table by
entering the expected voltage value (W) and the voltage value (X) that corresponds to the actual
controller display.
42
CB (E2) Service Manual
Sensor without calibration cap, unit door open, unit in normal operation.
Exposition to ambient air (21 vol.-% O2 ) for 5 minutes
Setting:
Normal Display → Config → Configuration 2 → System-Code 2802 → Service Level →
Inp. Val. correct. → Analogue input 5
For Start Value X enter the voltage value that corresponds to the controller display (e.g.,
202.2).
For Start Value W enter the reference voltage value of the O2 concentration of the
ambient air (210).
→ leave menu with 4 x EXIT.
Sensor with pushed-on calibration cap, unit door open, unit in normal operation.
Sensor exposition to test gas 80 vol.-% O2 for 5 minutes:
With sensor exposition to test gas (80 Vol.-% O2 ) the MB1 controller display should display 80 Vol.-%.
According to the above table this corresponds to a voltage of 1434mV.
The corrected value 80 vol.-% appears on the controller display after 5 seconds.
Afterwards the exposure with test gas is stopped.
43
CB (E2) Service Manual
O2 [Vol.-%] U [mV] O2 [Vol.-%] U [mV] O2 [Vol.-%] U [mV] O2 [Vol.-%] U [mV] O2 [Vol.-%] U [mV]
0 0.0 5 45.7 10 93.9 15 144.8 20 198.8
0.1 0.9 5.1 46.6 10.1 94.9 15.1 145.9 20.1 199.9
0.2 1.8 5.2 47.6 10.2 95.9 15.2 146.9 20.2 201.1
0.3 2.7 5.3 48.5 10.3 96.9 15.3 148.0 20.3 202.2
0.4 3.6 5.4 49.5 10.4 97.8 15.4 149.0 20.4 203.3
0.5 4.5 5.5 50.4 10.5 98.8 15.5 150.1 20.5 204.4
0.6 5.4 5.6 51.3 10.6 99.8 15.6 151.1 20.6 205.5
0.7 6.3 5.7 52.3 10.7 100.8 15.7 152.2 20.7 206.7
0.8 7.2 5.8 53.2 10.8 101.8 15.8 153.2 20.8 207.8
0.9 8.1 5.9 54.2 10.9 102.8 15.9 154.3 20.9 208.9
1 9.0 6 55.1 11 103.8 16 155.3 21 210.0
1.1 9.9 6.1 56.1 11.1 104.8 16.1 156.4 21.1 211.2
1.2 10.8 6.2 57.0 11.2 105.8 16.2 157.5 21.2 212.3
1.3 11.7 6.3 58.0 11.3 106.8 16.3 158.5 21.3 213.4
1.4 12.6 6.4 58.9 11.4 107.8 16.4 159.6 21.4 214.6
1.5 13.5 6.5 59.9 11.5 108.9 16.5 160.7 21.5 215.7
1.6 14.4 6.6 60.8 11.6 109.9 16.6 161.7 21.6 216.8
1.7 15.3 6.7 61.8 11.7 110.9 16.7 162.8 21.7 218.0
1.8 16.2 6.8 62.7 11.8 111.9 16.8 163.9 21.8 219.1
1.9 17.1 6.9 63.7 11.9 112.9 16.9 164.9 21.9 220.2
2 18.0 7 64.7 12 113.9 17 166.0 22 221.4
2.1 18.9 7.1 65.6 12.1 114.9 17.1 167.1 22.1 222.5
2.2 19.8 7.2 66.6 12.2 115.9 17.2 168.2 22.2 223.7
2.3 20.7 7.3 67.5 12.3 116.9 17.3 169.2 22.3 224.8
2.4 21.6 7.4 68.5 12.4 118.0 17.4 170.3 22.4 226.0
2.5 22.6 7.5 69.5 12.5 119.0 17.5 171.4 22.5 227.1
2.6 23.5 7.6 70.4 12.6 120.0 17.6 172.5 22.6 228.3
2.7 24.4 7.7 71.4 12.7 121.0 17.7 173.6 22.7 229.4
2.8 25.3 7.8 72.4 12.8 122.0 17.8 174.6 22.8 230.6
2.9 26.2 7.9 73.3 12.9 123.1 17.9 175.7 22.9 231.7
3 27.1 8 74.3 13 124.1 18 176.8 23 232.9
3.1 28.1 8.1 75.3 13.1 125.1 18.1 177.9 23.1 234.0
3.2 29.0 8.2 76.2 13.2 126.1 18.2 179.0 23.2 235.2
3.3 29.9 8.3 77.2 13.3 127.2 18.3 180.1 23.3 236.4
3.4 30.8 8.4 78.2 13.4 128.2 18.4 181.2 23.4 237.5
3.5 31.7 8.5 79.1 13.5 129.2 18.5 182.3 23.5 238.7
3.6 32.7 8.6 80.1 13.6 130.2 18.6 183.4 23.6 239.8
3.7 33.6 8.7 81.1 13.7 131.3 18.7 184.5 23.7 241.0
3.8 34.5 8.8 82.1 13.8 132.3 18.8 185.6 23.8 242.2
3.9 35.4 8.9 83.1 13.9 133.3 18.9 186.7 23.9 243.4
4 36.4 9 84.0 14 134.4 19 187.8 24 244.5
4.1 37.3 9.1 85.0 14.1 135.4 19.1 188.9 24.1 245.7
4.2 38.2 9.2 86.0 14.2 136.5 19.2 190.0 24.2 246.9
4.3 39.2 9.3 87.0 14.3 137.5 19.3 191.1 24.3 248.0
4.4 40.1 9.4 88.0 14.4 138.5 19.4 192.2 24.4 249.2
4.5 41.0 9.5 88.9 14.5 139.6 19.5 193.3 24.5 250.4
4.6 42.0 9.6 89.9 14.6 140.6 19.6 194.4 24.6 251.6
4.7 42.9 9.7 90.9 14.7 141.7 19.7 195.5 24.7 252.8
4.8 43.8 9.8 91.9 14.8 142.7 19.8 196.6 24.8 254.0
4.9 44.8 9.9 92.9 14.9 143.8 19.9 197.7 24.9 255.1
44
CB (E2) Service Manual
O2 [Vol.-%] U [mV] O2 [Vol.-%] U [mV] O2 [Vol.-%] U [mV] O2 [Vol.-%] U [mV] O2 [Vol.-%] U [mV]
25 256.3 30 317.8 35 383.8 40 455.1 45 532.7
25.1 257.5 30.1 319.1 35.1 385.2 40.1 456.6 45.1 534.3
25.2 258.7 30.2 320.3 35.2 386.6 40.2 458.1 45.2 535.9
25.3 259.9 30.3 321.6 35.3 387.9 40.3 459.6 45.3 537.5
25.4 261.1 30.4 322.9 35.4 389.3 40.4 461.1 45.4 539.2
25.5 262.3 30.5 324.2 35.5 390.7 40.5 462.6 45.5 540.8
25.6 263.5 30.6 325.5 35.6 392.1 40.6 464.1 45.6 542.4
25.7 264.7 30.7 326.8 35.7 393.5 40.7 465.6 45.7 544.1
25.8 265.9 30.8 328.0 35.8 394.9 40.8 467.1 45.8 545.7
25.9 267.1 30.9 329.3 35.9 396.3 40.9 468.6 45.9 547.4
26 268.3 31 330.6 36 397.6 41 470.1 46 549.0
26.1 269.5 31.1 331.9 36.1 399.0 41.1 471.6 46.1 550.7
26.2 270.7 31.2 333.2 36.2 400.4 41.2 473.1 46.2 552.3
26.3 271.9 31.3 334.5 36.3 401.8 41.3 474.7 46.3 554.0
26.4 273.1 31.4 335.8 36.4 403.2 41.4 476.2 46.4 555.6
26.5 274.3 31.5 337.1 36.5 404.6 41.5 477.7 46.5 557.3
26.6 275.5 31.6 338.4 36.6 406.0 41.6 479.2 46.6 559.0
26.7 276.8 31.7 339.7 36.7 407.4 41.7 480.8 46.7 560.6
26.8 278.0 31.8 341.0 36.8 408.8 41.8 482.3 46.8 562.3
26.9 279.2 31.9 342.3 36.9 410.3 41.9 483.8 46.9 564.0
27 280.4 32 343.6 37 411.7 42 485.4 47 565.7
27.1 281.6 32.1 344.9 37.1 413.1 42.1 486.9 47.1 567.4
27.2 282.9 32.2 346.2 37.2 414.5 42.2 488.4 47.2 569.0
27.3 284.1 32.3 347.6 37.3 415.9 42.3 490.0 47.3 570.7
27.4 285.3 32.4 348.9 37.4 417.3 42.4 491.5 47.4 572.4
27.5 286.5 32.5 350.2 37.5 418.8 42.5 493.1 47.5 574.1
27.6 287.8 32.6 351.5 37.6 420.2 42.6 494.6 47.6 575.8
27.7 289.0 32.7 352.8 37.7 421.6 42.7 496.2 47.7 577.5
27.8 290.2 32.8 354.2 37.8 423.1 42.8 497.7 47.8 579.2
27.9 291.5 32.9 355.5 37.9 424.5 42.9 499.3 47.9 580.9
28 292.7 33 356.8 38 425.9 43 500.8 48 582.6
28.1 293.9 33.1 358.2 38.1 427.4 43.1 502.4 48.1 584.4
28.2 295.2 33.2 359.5 38.2 428.8 43.2 504.0 48.2 586.1
28.3 296.4 33.3 360.8 38.3 430.3 43.3 505.5 48.3 587.8
28.4 297.7 33.4 362.2 38.4 431.7 43.4 507.1 48.4 589.5
28.5 298.9 33.5 363.5 38.5 433.1 43.5 508.7 48.5 591.3
28.6 300.2 33.6 364.8 38.6 434.6 43.6 510.3 48.6 593.0
28.7 301.4 33.7 366.2 38.7 436.0 43.7 511.9 48.7 594.7
28.8 302.7 33.8 367.5 38.8 437.5 43.8 513.4 48.8 596.5
28.9 303.9 33.9 368.9 38.9 439.0 43.9 515.0 48.9 598.2
29 305.2 34 370.2 39 440.4 44 516.6 49 599.9
29.1 306.4 34.1 371.6 39.1 441.9 44.1 518.2 49.1 601.7
29.2 307.7 34.2 372.9 39.2 443.3 44.2 519.8 49.2 603.5
29.3 308.9 34.3 374.3 39.3 444.8 44.3 521.4 49.3 605.2
29.4 310.2 34.4 375.6 39.4 446.3 44.4 523.0 49.4 607.0
29.5 311.5 34.5 377.0 39.5 447.8 44.5 524.6 49.5 608.7
29.6 312.7 34.6 378.4 39.6 449.2 44.6 526.2 49.6 610.5
29.7 314.0 34.7 379.7 39.7 450.7 44.7 527.8 49.7 612.3
29.8 315.3 34.8 381.1 39.8 452.2 44.8 529.4 49.8 614.0
29.9 316.5 34.9 382.5 39.9 453.7 44.9 531.1 49.9 615.8
45
CB (E2) Service Manual
O2 [Vol.-%] U [mV] O2 [Vol.-%] U [mV] O2 [Vol.-%] U [mV] O2 [Vol.-%] U [mV] O2 [Vol.-%] U [mV]
50 617.6 55 711.5 60 816.4 65 935.4 70 1072.7
50.1 619.4 55.1 713.5 60.1 818.6 65.1 937.9 70.1 1075.7
50.2 621.2 55.2 715.4 60.2 820.9 65.2 940.5 70.2 1078.7
50.3 623.0 55.3 717.4 60.3 823.1 65.3 943.1 70.3 1081.7
50.4 624.8 55.4 719.4 60.4 825.4 65.4 945.6 70.4 1084.7
50.5 626.5 55.5 721.4 60.5 827.6 65.5 948.2 70.5 1087.7
50.6 628.4 55.6 723.4 60.6 829.9 65.6 950.8 70.6 1090.7
50.7 630.2 55.7 725.4 60.7 832.1 65.7 953.4 70.7 1093.8
50.8 632.0 55.8 727.5 60.8 834.4 65.8 956.0 70.8 1096.8
50.9 633.8 55.9 729.5 60.9 836.7 65.9 958.6 70.9 1099.9
51 635.6 56 731.5 61 839.0 66 961.2 71 1102.9
51.1 637.4 56.1 733.5 61.1 841.3 66.1 963.8 71.1 1106.0
51.2 639.2 56.2 735.6 61.2 843.6 66.2 966.5 71.2 1109.1
51.3 641.1 56.3 737.6 61.3 845.9 66.3 969.1 71.3 1112.2
51.4 642.9 56.4 739.6 61.4 848.2 66.4 971.8 71.4 1115.3
51.5 644.7 56.5 741.7 61.5 850.5 66.5 974.4 71.5 1118.4
51.6 646.6 56.6 743.7 61.6 852.8 66.6 977.1 71.6 1121.6
51.7 648.4 56.7 745.8 61.7 855.1 66.7 979.8 71.7 1124.7
51.8 650.3 56.8 747.8 61.8 857.4 66.8 982.4 71.8 1127.9
51.9 652.1 56.9 749.9 61.9 859.8 66.9 985.1 71.9 1131.0
52 654.0 57 752.0 62 862.1 67 987.8 72 1134.2
52.1 655.8 57.1 754.1 62.1 864.5 67.1 990.5 72.1 1137.4
52.2 657.7 57.2 756.1 62.2 866.8 67.2 993.2 72.2 1140.6
52.3 659.6 57.3 758.2 62.3 869.2 67.3 996.0 72.3 1143.8
52.4 661.4 57.4 760.3 62.4 871.5 67.4 998.7 72.4 1147.0
52.5 663.3 57.5 762.4 62.5 873.9 67.5 1001.4 72.5 1150.3
52.6 665.2 57.6 764.5 62.6 876.3 67.6 1004.2 72.6 1153.5
52.7 667.1 57.7 766.6 62.7 878.7 67.7 1006.9 72.7 1156.8
52.8 668.9 57.8 768.7 62.8 881.1 67.8 1009.7 72.8 1160.0
52.9 670.8 57.9 770.8 62.9 883.5 67.9 1012.5 72.9 1163.3
53 672.7 58 772.9 63 885.9 68 1015.2 73 1166.6
53.1 674.6 58.1 775.1 63.1 888.3 68.1 1018.0 73.1 1169.9
53.2 676.5 58.2 777.2 63.2 890.7 68.2 1020.8 73.2 1173.2
53.3 678.4 58.3 779.3 63.3 893.1 68.3 1023.6 73.3 1176.6
53.4 680.3 58.4 781.5 63.4 895.6 68.4 1026.4 73.4 1179.9
53.5 682.3 58.5 783.6 63.5 898.0 68.5 1029.3 73.5 1183.3
53.6 684.2 58.6 785.8 63.6 900.4 68.6 1032.1 73.6 1186.6
53.7 686.1 58.7 787.9 63.7 902.9 68.7 1034.9 73.7 1190.0
53.8 688.0 58.8 790.1 63.8 905.4 68.8 1037.8 73.8 1193.4
53.9 690.0 58.9 792.2 63.9 907.8 68.9 1040.7 73.9 1196.8
54 691.9 59 794.4 64 910.3 69 1043.5 74 1200.2
54.1 693.8 59.1 796.6 64.1 912.8 69.1 1046.4 74.1 1203.7
54.2 695.8 59.2 798.8 64.2 915.3 69.2 1049.3 74.2 1207.1
54.3 697.7 59.3 801.0 64.3 917.7 69.3 1052.2 74.3 1210.6
54.4 699.7 59.4 803.1 64.4 920.2 69.4 1055.1 74.4 1214.1
54.5 701.6 59.5 805.3 64.5 922.8 69.5 1058.0 74.5 1217.5
54.6 703.6 59.6 807.5 64.6 925.3 69.6 1060.9 74.6 1221.0
54.7 705.6 59.7 809.8 64.7 927.8 69.7 1063.9 74.7 1224.6
54.8 707.5 59.8 812.0 64.8 930.3 69.8 1066.8 74.8 1228.1
54.9 709.5 59.9 814.2 64.9 932.8 69.9 1069.8 74.9 1231.6
46
CB (E2) Service Manual
O2 [Vol.-%] U [mV] O2 [Vol.-%] U [mV] O2 [Vol.-%] U [mV] O2 [Vol.-%] U [mV] O2 [Vol.-%] U [mV]
75 1235.2 80 1434.0 85 1690.3 90 2051.6 95 2669.2
75.1 1238.8 80.1 1438.5 85.1 1696.3 90.1 2060.6 95.1 2687.2
75.2 1242.3 80.2 1443.0 85.2 1702.3 90.2 2069.6 95.2 2705.6
75.3 1245.9 80.3 1447.5 85.3 1708.3 90.3 2078.7 95.3 2724.3
75.4 1249.6 80.4 1452.0 85.4 1714.4 90.4 2088.0 95.4 2743.5
75.5 1253.2 80.5 1456.6 85.5 1720.5 90.5 2097.3 95.5 2763.1
75.6 1256.8 80.6 1461.1 85.6 1726.7 90.6 2106.7 95.6 2783.1
75.7 1260.5 80.7 1465.8 85.7 1732.9 90.7 2116.3 95.7 2803.6
75.8 1264.2 80.8 1470.4 85.8 1739.2 90.8 2125.9 95.8 2824.5
75.9 1267.9 80.9 1475.0 85.9 1745.5 90.9 2135.6 95.9 2846.0
76 1271.6 81 1479.7 86 1751.8 91 2145.5 96 2868.0
76.1 1275.3 81.1 1484.4 86.1 1758.2 91.1 2155.4 96.1 2890.6
76.2 1279.0 81.2 1489.1 86.2 1764.6 91.2 2165.5 96.2 2913.7
76.3 1282.8 81.3 1493.9 86.3 1771.1 91.3 2175.7 96.3 2937.5
76.4 1286.5 81.4 1498.7 86.4 1777.6 91.4 2186.0 96.4 2961.9
76.5 1290.3 81.5 1503.5 86.5 1784.2 91.5 2196.4 96.5 2987.0
76.6 1294.1 81.6 1508.3 86.6 1790.8 91.6 2207.0 96.6 3012.8
76.7 1297.9 81.7 1513.2 86.7 1797.5 91.7 2217.6 96.7 3039.4
76.8 1301.8 81.8 1518.0 86.8 1804.2 91.8 2228.4 96.8 3066.8
76.9 1305.6 81.9 1522.9 86.9 1811.0 91.9 2239.4 96.9 3095.1
77 1309.5 82 1527.9 87 1817.8 92 2250.4 97 3124.3
77.1 1313.4 82.1 1532.8 87.1 1824.7 92.1 2261.6 97.1 3154.5
77.2 1317.3 82.2 1537.8 87.2 1831.7 92.2 2273.0 97.2 3185.8
77.3 1321.2 82.3 1542.9 87.3 1838.6 92.3 2284.5 97.3 3218.2
77.4 1325.1 82.4 1547.9 87.4 1845.7 92.4 2296.1 97.4 3251.8
77.5 1329.1 82.5 1553.0 87.5 1852.8 92.5 2307.9 97.5 3286.8
77.6 1333.0 82.6 1558.1 87.6 1859.9 92.6 2319.9 97.6 3323.2
77.7 1337.0 82.7 1563.2 87.7 1867.2 92.7 2332.0 97.7 3361.1
77.8 1341.0 82.8 1568.4 87.8 1874.4 92.8 2344.3 97.8 3400.7
77.9 1345.0 82.9 1573.6 87.9 1881.8 92.9 2356.8 97.9 3442.1
78 1349.1 83 1578.8 88 1889.2 93 2369.4 98 3485.6
78.1 1353.1 83.1 1584.1 88.1 1896.6 93.1 2382.2 98.1 3531.3
78.2 1357.2 83.2 1589.4 88.2 1904.1 93.2 2395.2 98.2 3579.5
78.3 1361.3 83.3 1594.7 88.3 1911.7 93.3 2408.4 98.3 3630.4
78.4 1365.4 83.4 1600.0 88.4 1919.4 93.4 2421.8 98.4 3684.4
78.5 1369.6 83.5 1605.4 88.5 1927.1 93.5 2435.4 98.5 3741.9
78.6 1373.7 83.6 1610.8 88.6 1934.9 93.6 2449.2 98.6 3803.4
78.7 1377.9 83.7 1616.3 88.7 1942.7 93.7 2463.3 98.7 3869.4
78.8 1382.1 83.8 1621.8 88.8 1950.6 93.8 2477.5 98.8 3940.8
78.9 1386.3 83.9 1627.3 88.9 1958.6 93.9 2492.0 98.9 4018.3
79 1390.5 84 1632.8 89 1966.7 94 2506.7 99 4103.2
79.1 1394.8 84.1 1638.4 89.1 1974.8 94.1 2521.7 99.1 4197.1
79.2 1399.1 84.2 1644.0 89.2 1983.0 94.2 2537.0 99.2 4302.0
79.3 1403.4 84.3 1649.7 89.3 1991.3 94.3 2552.5 99.3 4421.0
79.4 1407.7 84.4 1655.4 89.4 1999.7 94.4 2568.2 99.4 4558.4
79.5 1412.0 84.5 1661.1 89.5 2008.1 94.5 2584.3 99.5 4720.8
79.6 1416.4 84.6 1666.9 89.6 2016.7 94.6 2600.6 99.6 4919.6
79.7 1420.7 84.7 1672.7 89.7 2025.3 94.7 2617.3 99.7 5175.9
79.8 1425.1 84.8 1678.5 89.8 2034.0 94.8 2634.3 99.8 5537.2
79.9 1429.6 84.9 1684.4 89.9 2042.7 94.9 2651.6 99.9 6154.8
100 30157.2
47
CB (E2) Service Manual
7 Maintenance
The BINDER GmbH determines a annual service period to ensure the high quality and reproducibility
of the results elaborated by help of this incubator.
Following tests and maintenance works must be carried out for a comprehensive maintenance.
1 After the door has been open for 30 sec. the temperature (target 37°C) and the CO2 content
(target 5%) must have recovered to 98% of their set value after
10 min..
2 A visual check is performed of the chamber for general condition, wear and corrosion.
3 The doors are checked for serviceability, corrosion, functionality and condition. The door sealing
are examined for wear, seat and damage.
4 The serviceability and functionality of the hinges are tested.
5 The closing mechanism is checked for serviceability and functionality.
6 The seal of the glass doors is checked. The glass doors are examined for damage.
7 The temperature controller MB1 is operated, set values input and all keys checked for
functionality.
8 The safety device 3.1 is tested for functionality. (set temp lower at safety device 3.1)
9 The CO2 controller is operated, set values input and all keys checked for functionality. The CO²
cut-off via the door contact is tested for functionality.
10 The warning lamp is tested.
11 The latest control parameters are tested and input where necessary.
12 The steri filter is checked for contamination and replaced if necessary.
13 The FPI-sensor head is subjected to a visual inspection.
14 The Permadry® water basin is checked for level, tightness, evenness and deposits.
15 The power consumption of the door and chamber heating elements are measured.
16 A complete visual inspection of the gas system is performed.
17 The functionality of the magnetic valves is checked.
18 The pressure switch is checked for functionality.
19 The sterilization function is checked.
20 A visual inspection of the relays is performed.
21 The main fan is checked for functionality.
22 The Permadry® fan is checked for functionality.
23 The availability of the operating manual (e.g. for safety information) is checked.
24 The VDE test including determination of leakage current and insulation resistance is performed
(for Germany – other safety standards may be considered in other countries)
25 The temperature is re-calibrated.
26 The CO2 is re-calibrated.
Following service report should filled out during every maintenance and a copy should be handed over
to the customer.
48
CB (E2) Service Manual
25 Temperature calibration
26 CO2 calibration
Implementation features:
Maintenance issued by: Date: signature :
49
CB (E2) Service Manual
Symbol Meaning
X Defects
OK without error
/ not existence
50
CB (E2) Service Manual
8
9 10
7
6 11
5 12
13
4
3
2
1
14
15
20 16
17
22 18
21 19
51
CB (E2) Service Manual
31
30
29
28
27
26
25
24
23
32
33
34
35
36
52
CB (E2) Service Manual
37 41 43 44
39
38 42
49 48 47 46 45
54
53
52
51
50
53