Dynamic Imaging of The Fetal Heart Using Metric Optimized Gating
Dynamic Imaging of The Fetal Heart Using Metric Optimized Gating
Optimized Gating
Christopher W. Roy,1 Mike Seed,2,3 Joshua F. P. van Amerom,1,3 Bahiyah Al Nafisi,2
Lars Grosse-Wortmann,2,3 Shi-Joon Yoo,2,3 and Christopher K. Macgowan1,2,3*
Purpose: Advances in fetal cardiovascular magnetic resonance
imaging have been limited by the absence of a reliable
cardiac gating signal. The purpose of this work was to develop
and validate metric-optimized gating (MOG) for cine imaging
of the fetal heart.
Theory and Methods: Cine MR and electrocardiogram data
were acquired in healthy adult volunteers for validation of the
MOG method. Comparison of MOG and electrocardiogram
reconstructions was performed based on the image quality for
each method, and the difference between MOG and electrocardiogram
trigger times. Fetal images were also acquired,
their quality evaluated by experienced radiologists, and the
theoretical error in the MOG trigger times were calculated.
Results: Excellent agreement between electrocardiogram and
MOG reconstructions was observed. The experimental errors
in adult MOG trigger times for all five volunteers were
6 (7, 25, 17, 8, and 13) ms. Fetal images captured normal and
diseased cardiac dynamics.
Conclusion: MOG for cine imaging of the fetal myocardium
was developed and validated in adults. Using MOG, the first
gated MR images of the human fetal myocardium were
obtained. Small moving structures were visualized during radial
contraction, thus capturing normal fetal cardiac wall motion and
permitting assessment of cardiac function. Magn Reson Med
70:1598–1607, 2013. VC 2013 Wiley Periodicals, Inc.
Key words: fetal magnetic resonance imaging; image metric;
gating; post-processing
Since its inception over 30 years ago, the field of fetal
magnetic resonance imaging (MRI) has grown considerably
and now provides valuable tools for assessing the
fetal anatomy including the brain, lungs, liver, and gastrointestinal
tract (1–4). However, fetal cardiovascular
MRI (CMR) remains relatively underdeveloped. CMR is
an important tool for the noninvasive assessment of cardiac
function, anatomy, and pathology in postnatal subjects.
Advances in fetal CMR may provide useful tools
for the in utero assessment of normal fetal cardiovascular
physiology and diseases that affect the fetal cardiovascular
system. Potential applications of fetal CMR include
congenital heart disease, placental insufficiency, and
other causes of fetal distress. Unfortunately, the development
of fetal CMR has been hampered by the difficulty
in measuring a fetal electrocardiogram (ECG) signal for
cardiac gating (5). The fetal ECG signal is typically an
order of magnitude weaker than that of the mother, and
can be difficult to detect in the presence of electrical
noise from the MRI (6). Current alternative strategies for
fetal gating include ultrasound based techniques and
self-gating. These strategies are described below followed
by the technique presented in this work: metric-optimized
gating (MOG).
Nonimaging ultrasound (fetal cardiotocography) is an
established technique for monitoring the fetal heart rate.
MR-compatible units show promise for fetal cardiac gating,
although they can be limited by interference between
the unit and MR system (7). Furthermore, the likelihood
of fetal movement during the scan may challenge the reliability
of the gating signal. This method has been used to
image the heart of anaesthetized fetal sheep but no direct
human application has been published to date (8).
Another approach to fetal cardiac gating is MRI “selfgating,”
where a periodic gating signal is extracted from
the MR data itself and is used to sort the data retrospectively
(9). Modern implementations exist for both radial
and Cartesian trajectories, and use a variety of signal
detection algorithms (10–12). Self-gating is attractive as
it requires no additional hardware, although the reliability
of the gating signal depends on slice prescription and
temporal resolution, and therefore may not accommodate
all k-space sampling trajectories. Previous publications
have used self-gating for cardiac imaging in the fetal
mouse, chick, and lamb (13–15). These techniques may
be applied to human fetal CMR, but as with ultrasound
gating, no direct application has been published to date.
Recent advances in non-gated acquisitions using real
time imaging may be applicable to fetal CMR; however,
these techniques require a combination of aggressive parallel
imaging with non-Cartesian k-space sampling, constraining
the available spatial resolution, temporal
resolution, and signal-to-noise ratio. Even in late gestation,
the diameter of a normal human left ventricle is only
15 mm and the diameters of major vessels are less than
8 mm (16). These small dimensions coupled with high
1Departments of Medical Biophysics and Medical Imaging, University of Toronto,
Toronto, Canada.
2Division of Cardiology, Department of Paediatrics, Labatt Family Heart
Centre, The Hospital for Sick Children, University of Toronto, Toronto,
Canada.
3Department of Diagnostic Imaging, The Hospital for Sick Children, University
are:
@ei;t
@Si;t
¼
1
C
_ ln
Si:t
Bt
__
þ1
__
;
@ei;t
@Bt
¼_
Si;t
Bt _ C
;
@ei;t
@C
¼_
Si:t
C2 _ ln
Si:t
Bt
__
:
Equation [A1] then becomes:
sðei;tÞ ¼
1
C
_ ln
Si;t
Bt
__
þ1
__
_ sðSi;tÞ
_ _2
"
þ
Si;t
Bt _ C
_ _2
_
X
i
_
Si;t _ sðSi;tÞ
_2
X
i
Si;t
2
þ
Si;t
C2 ln
Si;t
Bt
_ _ __2
_
X
t
X
i
_
Si;t _ sðSi;tÞ
_2
X
i
Si;t
2
3
75
12
: [A4]
To explain the dependency of Eq. [A4] on relevant
imaging parameters, several simplifying assumptions can
be made. First, assuming the image noise is approximately
constant over an ROI [i.e., sðSi;tÞ _ sðSÞ] for all
spatial and time indices, sðEÞ simplifies to give:
sðEÞ _
sðSÞ
C
ffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffif
fiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiff
iffi
X
t
X
i
ln
Si;t
Bt
__
þ1
_ _2
þ
Si;t
Bt
_ _2
þ
Si;t _ ln Si;t
Bt
__
C
0
@
1
A
2
_T
0
B@
1
CA
vuuuut
;
where T is the total number of time indices. Furthermore,
the second and third terms of the above expression
are experimentally small relative to the first, and can be
removed:
sðEÞ _
sðSÞ
C
ffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffi
X
t
X
i
ln
Si;t
Bt
__
þ1
_ _2
vuut
:
Finally, ln Si;t
Bt
__
<< 0 and so,
sðEÞ _
sðSÞ
C
ffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffi
X
t
X
i
ln
Si;t
Bt
_ _ __2
vuut
:
The error in the image entropy is then inversely proportional
to the signal-to-noise ratio, scaled by a second
term. The scaling term is minimized when an image consists
of a single bright pixel against a dark background. If
the image is blurred, the scaling term grows, amplifying
the uncertainty in the image entropy. Thus, the uncertainty
in the image entropy is governed by the image
noise but for images that are inherently blurry (poor resolution,
misgating, etc.) the contribution of the image
noise to the uncertainty in the image entropy increases.
The uncertainty in the image entropy is transformed into
an uncertainty in each MOG parameter according to the
following: first, MOG parameters are perturbed about their
optimum values creating one-dimensional entropy landscapes
(Fig. 8). Second, the entropy landscapes are fit with
a parabola. Finally, the uncertainty in each parameter is
given by the change in parameter value required to raise
the entropy an amount r(E) above its optimal value. The
theoretical error is then dependent on the uncertainty in
the image entropy and the curvature of the entropy landscape
which, in turn, is dependent on the underlying
motion and sensitivity of the image entropy to that motion.
REFERENCES
1. Kim D-H, Chung S, Vigneron DB, Barkovich AJ, Glenn OA. Diffusionweighted
imaging of the fetal brain in vivo. Magn Reson Med
2008;59:216–220.
FIG. 8. Entropy as a function of parameter offset. The theoretical
error is given by the timing offset required to raise the entropy an
amount r(E) above its minimum value.
1606 Roy et al.
2. Kim D-hyun, Vahidi K, Caughey AB, Coakley FV, Vigneron DB,
Kurhanewicz J, Mow B, Joe BN. In vivo (1) H magnetic resonance
spectroscopy of amniotic fluid and fetal lung at 1.5 T: technical challenges.
JMRI 2008;28:1033–1038.
3. S_rensen A, Holm D, Pedersen M, Tietze A, Stausb_l-Gr_n B, Duus
L, Uldbjerg N. Left-right difference in fetal liver oxygenation during
hypoxia estimated by BOLD MRI in a fetal sheep model. Ultrasound
Obstet Gynecol 2011;38:665–672.
4. Cannie M, Jani J, Dymarkowski S, Deprest J. Fetal magnetic resonance
imaging: luxury or necessity? Ultrasound Obstet Gynecol
2006;27:471–476.
5. Brugger PC. In: Prayer D, editor. Fetal MRI. Berlin, Heidelberg:
Springer Berlin Heidelberg; 2011 pp. 247–257.
6. Peters M, Crowe J, Pi_eri JF, Quartero H, Hayes-Gill B, James D, Stinstra
J, Shakespeare S. Monitoring the fetal heart non-invasively: a
review of methods. J Perinat Med 2001;29:408–416.
7. Feinberg Da, Giese D, Bongers DA, Ramanna S, Zaitsev M, Markl M,
G€unther M. Hybrid ultrasound MRI for improved cardiac imaging
and real-time respiration control. Magn Reson Med 2010;63:290–296.
8. Yamamura J, Kopp I, Frisch M, Fischer R, Valett K, Hecher K, Adam
G, Wedeg€artner U. Cardiac MRI of the fetal heart using a novel triggering
method: initial results in an animal model. JMRI
2012;35:1071–1076.
9. Spraggins TA. Wireless retrospective gating: application to cine cardiac
imaging. Magn Reson Imaging 1990;8:675–681.
10. Crowe ME, Larson AC, Zhang Q, Carr J, White RD, Li D, Simonetti
OP. Automated rectilinear self-gated cardiac cine imaging. Magn
Reson Med 2004;52:782–788.
11. Thompson RB, McVeigh ER. Flow-gated phase-contrast MRI using radial
acquisitions. Magn Reson Med 2004;52:598–604.
12. Larson AC, White RD, Laub G, McVeigh ER, Li D, Simonetti OP. Selfgated
cardiac cine MRI. Magn Reson Med 2004;51:93–102.
13. Hiba B, Richard N, Thibault H, Janier M. Cardiac and respiratory
self-gated cine MRI in the mouse: comparison between radial and
rectilinear techniques at 7T. Magn Reson Med 2007;58:745–753.
14. Holmes WM, McCabe C, Mullin JM, Condon B, Bain MM. Images in
cardiovascular medicine. Noninvasive self-gated magnetic resonance
cardiac imaging of developing chick embryos in ovo. Circulation
2008;117:e346–e347.
15. Yamamura J, Frisch M, Ecker H, Graessner J, Hecher K, Adam G,
Wedeg€artner U. Self-gating MR imaging of the fetal heart: comparison
with real cardiac triggering. Eur Radiol 2011;21:142–149.
16. DeVore GR. Assessing fetal cardiac ventricular function. Semin Fetal
Neonatal Med 2005;10:515–541.
17. Sandmire HF, DeMott RK. Electronic fetal heart rate monitoring:
research guidelines for interpretation. Am J Obstet Gynecol
1998;179:276–277.
18. Jansz MS, Seed M, van Amerom JFP, Wong D, Grosse-Wortmann L,
Yoo S-J, Macgowan CK. Metric optimized gating for fetal cardiac
MRI. Magn Reson Med 2010;64:1304–1314.
19. Lenz GW, Haacke EM, White RD. Retrospective cardiac gating: a
review of technical aspects and future directions. Magn Reson Imaging
1989;7:445–455.
20. Haacke EM, Lenz GW, Nelson AD. Pseudo-gating: elimination of periodic
motion artifacts in magnetic resonance imaging without gating.
Magn Reson Med 1987;4:162–174.
21. McGee KP, Manduca A, Felmlee JP, Riederer SJ, Ehman RL. Image
metric-based correction (autocorrection) of motion effects: analysis of
image metrics. JMRI 2000;11:174–181.
22. Atkinson D, Hill DL, Stoyle PN, Summers PE, Keevil SF. Automatic
correction of motion artifacts in magnetic resonance images using an
entropy focus criterion. IEEE Trans Med Imaging 1997;16:903–910.
23. Bevington P, Robinson DK. Data reduction and error analysis for the
physical sciences, 2nd ed. New York: McGraw-Hill; 1994.
24. Griswold MA, Jakob PM, Heidemann RM, Nittka M, Jellus V, Wang J,
Kiefer B, Haase A. Generalized autocalibrating partially parallel
acquisitions (GRAPPA). Magn Reson Med 2002;47:1202–1210.
25. Lagarias J, Reeds J, Wright M. Convergence properties of the Nelder-
Mead simplexmethod in low dimensions. SiamJ Optim 1998;9:112–147.
26. Foo TK, Bernstein MA, Aisen AM, Hernandez RJ, Collick BD, Bernstein
T. Improved ejection fraction and flow velocity estimates with
use of view sharing and uniform repetition time excitation with fast
cardiac techniques. Radiology 1995;195:471–478.
27. Holdsworth DW, Norley CJ, Frayne R, Steinman DA, Rutt BK. Characterization
of common carotid artery blood-flow waveforms in normal
human subjects. Physiol Meas 1999;20:219–240.
28. Kramer CM, Barkhausen J, Flamm SD, Kim RJ, Nagel E. Standardized
cardiovascular magnetic resonance imaging (CMR) protocols, society
for cardiovascular magnetic resonance: board of trustees task force on
standardized protocols. J Cardiovasc Magn Reson 2008;10:35–45.
29. Tsyvian P, Malkin K, Wladimiroff JW. Assessment of fetal left cardiac
isovolumic relaxation time in appropriate and small-for-gestationalage
fetuses. Ultrasound Med Biol 1995;21:739–743.
30. Manduca A, McGee K, Welch E, Felmlee J, Ehman RL. Autofocusing
of clinical shoulder MR images for correction of motion artifacts.
MICCAI 1998:598–605.
31. McGee KP, Felmlee JP, Jack CR, Manduca A, Riederer SJ, Ehman RL.
Autocorrection of three-dimensional time-of-flight MR angiography of
the Circle of Willis. AJR Am J Roentgenol 2001;176:513–518.
32. McGee K, Felmlee J, Ehman R. Autocorrection of MR images using
multiple metrics. US Patent 6,265,874. 2001.
33. Godfrey ME, Messing B, Cohen SM, Valsky DV, Yagel S. Functional
assessment of the fetal heart: a review. Ultrasound Obstet Gynecol
2012;39:131–144.
34. Loomba RS, Chandrasekar S, Shah PH, Sanan P. The developing role of
fetal magnetic resonance imaging in the diagnosis of congenital cardiac
anomalies: a systematic review. Ann Pediatr Card 2011;4:172–176.
35. Lotz J, Meier C, Leppert A, Galanski M. Cardiovascular flow measurement
with phase-contrast MR imaging: basic facts and implementation.
Radiographics 2002;22:651–671.
36. Thiele H, Nagel E, Paetsch I, Schnackenburg B, Bornstedt A, Kouwenhoven
M, Wahl A, Schuler G, Fleck E. Functional cardiac MR imaging
with steady-state free precession (SSFP) significantly improves endocardial
border delineation without contrast agents. JMRI 2001;14:362–367.
37. Rathi VK, Biederman RWW. Imaging of ventricular function by cardiovascular
magnetic resonance. Curr Cardiol Rep 2004;6:55–61.
38. Alfakih K, Reid S, Jones T, Sivananthan M. Assessment of ventricular
function and mass by cardiac Magn Reson Imaging Eur Radiol
2004;14:1813–1822.
39. Peshock RM, Willett DL, Sayad DE, Hundley WG, Chwialkowski MC,
Clarke GD, Parkey RW. Quantitative MR imaging of the heart. Magn
Reson Imaging Clin N Am 1996;4:287–305.
Dynamic Imaging of the Fetal Heart Using MOG 1607