Care Vision OM 100 - Service Manual
Care Vision OM 100 - Service Manual
Service Manual
Manual
Pulse Oximeter
Table of Contents
2 Factory Mode 10
2.1. Overview 10
3 Firmware Download 12
3.1. Overview 12
4 Maintenance 18
5 Operation Verification 21
5.1. Overview 21
-1-
6 Disassembly 32
6.1. Overview 32
7 Troubleshooting 35
7.1. Overview 35
7.2. Troubleshooting 36
-2-
For your information
Thank you for purchasing Care Vision OM-100. This product is a medical device that
is designed to check the physiological condition of a patient by measuring the oxyg
This document is provided for service person, it contains device system configuration,
Warranty Period
l This device is manufactured under the strict quality management and insp
l Compensation criteria for the repair and exchange shall follow “Customer
l Any damages incurred while using the device under normal operating condi
tion shall be repaired in the repair center of the manufacture free of cha
l When the device is damaged during the warranty period, send us the prod
uct with the model number, serial number, purchase date and description
of the damage.
l The manufacturer or sales agency is not liable for any damages or loss inc
-3-
Contact Information
Medical Supply Co. Ltd gladly provides our customers with various services and
information of the product. If you have any questions, contact following number or
TEL : +82-70-7438-6034
Technical Support
FAX : +82-33-742-4385
Website http://www.medicalsupply.co.kr/
Note
copyright.
Medical Supply Co., Ltd is not liable for any problems that are incurred due to
Revision History
This document contains its revision history, which is updated when the contents of
the document are changed. The contents of this document, which do not seriously
affect the performance of the device, are subject to change without prior notice.
l Revision : 1.0
-4-
1 System Block Diagram
(1) Power
- AC power
Care Vision OM-100/Care Vision OM-100 contains SMPS (Switch Mode Power Supply)
60Hz. There are two fuses for AC input (line and neutral) to protect the equipment
from overcurrent.
- AC-DC converter
transistor and inductor etc. AC input and DC output is isolated for protect
equipment and users from electric shock.
-5-
- Battery
Care Vision OM-100/Care Vision OM-100 includes an internal power source, Ni-MH
battery. The capacity is 7.2V/2000mAh and the battery can operate the device for
recharged connecting power code. This device includes a battery charge circuit that
controls charging state. If battery level increases over the threshold, battery
- 3.3Vdc
3.3Vdc is generated from the battery or adapter power. It is used for operating MCU,
EEPROM, Sound driver etc. It is also used for operating RTC (Real Time Clock) in
- 5Vdc
5Vdc is used for operating LCD and SpO2 isolation power circuit.
In order to prevent undesired effects from power and main module, SpO2 module
(2) Main
- Main MCU
MCU is used for controlling the user interface and saving trend data etc. It displays
the SpO2 and pulse rate results to 7-segment and SpO2 graph to LCD. Alarm and
beep sounds are generated through speaker. In addition, it receives the input of
- EEPROM
EEPROM is a non-volatile memory used for storing trend data, user settings etc. It
can maintain the data stored during power off. So, the trend data and user settings
-6-
- RTC(Real-time clock)
After setting time, RTC counts and maintains time and date.
The power of RTC is provided continuously so, time is moving on during power off.
- Sound Driver
Sound driver is used for making the auditory signals such as alarm sound, beep
sound etc.
It consists of DAC (digital to analog converter), digital potential meter and audio
power amplifier. It converts from digital signals to analog sound, and amplifies the
(3) SpO2
To protect the patients from electric shock, SpO2 module, the applied part should
- LED Driver
LED driver circuit is used for controlling the electrical current of Red and IR
(infrared) LED in light emitting part of SpO2 sensor. Red and IR LED repeat to turn
on and off alternately and the detector of SpO2 sensor, photodiode, receives the
Photodiode converts from the light signal to the current signal. Analog signal
processing circuit contains ‘IV converter’, ‘sample and hold circuit’, and ‘low pass
filter’.
‘IV converter’ converts from the current signal to the voltage signal and ‘sample
-7-
and hold circuit’ separates Red and IR signal from a photodiode. Low pass filter
- ADC
ADC (analog to digital converter) is used for converting analog signal to digital signal.
Digital signal is easy to calculate than analog signal. So, for SpO2 algorithm
- SpO2 MCU
It is used for controlling SpO2 circuit and running SpO2 algorithm. It calculates SpO2
and pulse rate from ADC output and determines alarm status. SpO2 MCU transmits
(4) Key
Keys are used for user input function. According to push key button, user can operate
the device such as power on/off, system setting, alarm audio pause etc. This device
contains 7 keys.
(5) LCD
LCD (liquid crystal display) is used for displaying SpO2/PR result, graph, trend data,
user setting, battery status, alarm status, etc. The display size is 240 X 64 dots, LCD
(6) 7-segment
3-digits 7-segment is used to display SpO2 and pulse rate with numeric character.
SpO2 result is displayed in red color, pulse rate result is displayed in green color. In
-8-
(7) Speaker
Sound driver circuit makes the sound signal and speaker generates sound according to
(8) Sensor
SpO2 sensor consists of two LEDs(Red and IR) and a photodiode. It contacts to
-9-
2 Factory Mode
2.1. Overview
This chapter mentions factory mode which enables for authorized persons to review and
check information of Care Vision OM-100 including checking the F/W version, EEPROM
etc.
PASSWORD
1111
- 10 -
Table 1. Service mode Items and Functions
Items Functions
In the demo mode, Care Vision OM-100 displays the pre-set value.
1) In power off condition, press ‘power button’ hold on ‘alarm button’ to start
2) When you restart the device, the system is returned to the initial screen.
Note : Don’t Use This Function Except Special Purposes (ex : Demo or Testing)
- 11 -
3 Firmware Download
charge of the service distributed by Medical Supply Co., Ltd. The problems
CAUTION: You must connect power cord to the equipment when download the
firmware.
3.1. Overview
This chapter is to load firmware distributed for Care Vision OM-100 when a previous
firmware may have error or the firmware is updated with a new system. Request the
Equipment Description
- 12 -
3.3. How to Download Firmware
4) Check whether the PC succeeds in searching for the USB connected to the
equipment.
5) In the window OS, Select Start – All programs – Misc program – Telecommunication –
Hyper Terminal (if using window version above XP, search trial program of
6) Run HyperTerminal. Input the proper name on the New Connection window like the
figure below.
- 13 -
7) If the Connection to… screen appears, select COM Port where Care Vision OM-100
- Bit/sec: 115200
- Data bit: 8
- Parity: No
- Stop bit: 1
- Flow control: No
- 14 -
Figure 5. Hyper Terminal Window
10) Return to the device, push the power button holding ▲ button and alarm button.
***********************
Wait Firmware Update
***********************
When the window displays like Figure 6, press [▶] button. Then, display changes to
below Figure 7.
- 15 -
***********************
Firmware Update Mode
***********************
On the display of Figure 8, please press the Menu button on the front case and
choose [PRG3]. Then the down preparation window on the hyper terminal screen
11) Type ‘1’ button on the keyboard. ‘Waiting for the file to be sent (press ‘a’ to
abort)’ message and ‘CCC…’ keep appearing.
- 16 -
12) Click send files ( ) on the menu bar and the ‘file send’ window appears. Select
the protocol ‘Ymodem’.
13) Select the file to send and click Send. Then, the file is downloaded like the
Figure below.
14) When the file transfer is completed, type the ‘2’ on the keyboard or push the
power button on Care Vision OM-100 to reboot. And then, check the normal operation.
- 17 -
4 Maintenance
WARNING: Take care not to spray or spill water on Care Vision OM-100
WARNING: Care Vision OM-100 complies with the waterproof regulation (IPX0)
in IEC 60529.
CAUTION: Disconnect the power code from the equipment before cleaning it.
CAUTION: Take care for liquids or detergents not to penetrate into the
equipment through connector or duct while cleaning the equipment
CAUTION: If the battery shows signs of damage, leakage or crack, let the
authorized one.
CAUTION: When you scrap or recycle parts used for replacement, this may
lead explosion while burning up. Therefore, recycle them based on regulations
on scrapping/recycling.
- Check whether labels are clearly visible. Otherwise, contact the service center of
Medical Supply Co., Ltd.
- Please refer to the ‘Operation Verification’ chapter for accurate verification on the
- 18 -
equipment.
- Please refer to the ‘Operation Verification’ chapter in the service manual for further
electricity safety test. If you fail the electric stability test, do not try to repair it and
Certified personnel should check based on the performance verification method for
every 2 years.
- If the equipment has visible damage or gets mechanical shock (ex, falling), the
personnel should check based on the ‘Operation Verification’ chapter in the service
manual.
- Please refer to the ‘Safety Test’ chapter in the service manual for electric safety
tests. If the equipment fails to pass the electric safety test, please refer to the
- Apply detergents in markets or non-polished detergents to the smooth cloth and wipe
the top, the bottom and the front of the equipment to clean Care Vision OM-100.
Note: If a liquid like water or cleaning agent is spilt, please wipe it well and dry
- 19 -
4.3. Battery Maintenance
ARGUS OXM plus uses Ni-MH battery only. If you do not use the equipment for more than
2 months, please charge the battery before using the equipment. Please refer to the
‘Operation Verification’ chapter in the service manual or the ‘Battery Operation’ chapter
in the user manual to charge the battery through connecting the power cord.
Medical Supply Co., Ltd. recommends to replacing the battery for every 2 years. If you
need to store your ‘ARGUS OXM plus’ for more than 2 months, remove the battery before
storage. Please refer to the Disassemble chapter for detail about replacing and removing
the battery.
Note: If you store your ‘ARGUS OXM plus’ for a long time without removing the battery,
recover its normal capacity. Fully charge the battery for about 4 hours while turning off
the equipment. Medical Supply Co., Ltd. recommends to replacing the battery for every
You may reuse the battery. Do not dispose of it into the trash can.
- 20 -
5 Operation Verification
5.1. Overview
This chapter mentions the operation verification test when the user repairs or maintains
the equipment. All the tests except charging battery and battery test should be done
before the equipment returns to the user. If Care Vision OM-100 fails to pass the tests in
the manual, the equipment should be repaired through proper methods before it returns
to the user.
Charging battery and its operation test should be done before repairing the equipment
whenever battery seems to cause troubles. All the other tests follow repair or check
methods. Batteries should be fully charged before performing battery operation test.
- 21 -
This chapter mentions the configuration values with factory default. If the user
- Use proper power code for Care Vision OM-100 and connect the power cord to power
source.
- Push the power button for 1 second and check whether the equipment is turned on.
- Disconnect the AC power cord from the equipment after it works normally.
- Check the equipment power is turned off after pushing the power switch for more
than 3 seconds.
- If the battery is fully charged, the indicator just shows the status of stopping. If the
battery is charging, the indicator shows the status with its 4 lights.
- Disconnect the AC power cord when the batter is fully charged. Then, enter the
Note: Keeping normal capacity of the battery depends on user’s habits. Therefore,
- 22 -
(3) Battery Discharge
- When the Ni-MH battery is fully charged, disconnect the power cord from the power
source.
- When the battery power remains for about 20 minutes after operating the equipment
for about 4 hours, check that there is an alarm showing the need to charge the battery
- If the battery power is still low and when turned it on, the power is automatically
2. Set the lower limit of oxygen saturation as 90% in alarm limit menu.
- Check whether oxygen saturation and pulse on the LED segment of the
- Check whether alarm sounds, alarm message and SpO2 value is blinking.
- The lower limit of oxygen saturation and oxygen saturation alarm limit are still
blinking.
11. Check if the oxygen saturation on the equipment is shown the same with the
14. Check whether the ‘No finger’ message appears on the LCD screen of the
15. Press the alarm audio paused button and check below.
- Alarm stops.
1. Connect the power cord to the equipment and turn on the equipment.
2. Set the lower limit of oxygen saturation as 90% in alarm limit menu.
6. Push the ▼ button to adjust the beep sound and check the volume of button
6. Push the ▲ button to adjust the alarm sound and check the volume of button
- Time Test
1. Connect the equipment to the power cord and turn on the equipment.
2. Check if the time goes through the right top of the LCD screen on the
equipment.
1. Connect the equipment to the power cord and turn on the equipment.
3. Turn off the equipment by pushing the power button for more than 3 seconds.
4. Disconnect the SpO2 sensor from the equipment and check if the SpO2 sensor
Note : the SpO2 simulator cannot be used to assess the accuracy of a pulse
ARGUS OXM plus complies with the safety regulations in the Class I of the IEC 60601-1
1. Protection Earth
This test checks the protection of the power code earth from the earth of the AC plug.
(1) Connect the AC power cord to the equipment and connect the power code to the
(2) Connect the resistance input lead of the hazard analyzer to the contactable earth
- 26 -
2. Electric Leakage
This test complies with the earth leakage current regulation in the IEC 60601-1. The
applied voltage based on the IEC 60601-1 is 264 V/50Hz – 60Hz AC. All the
(1) Connect the AC power code based on the instruction of the electric hazard
analyzer.
Note: The earth leakage current measures AC main power and the conductor in the
protection earth. The values from the earth leakage current test should not be over the
- 27 -
- External Leakage Current
This test is based on the regulations on the external leakage current in the IEC 60601-
1. This test measures external leakage current not earthed between the external part
and the earth part. The applied voltage based on the IEC 60601-1 is 264 V/50Hz – 60Hz
AC voltage.
Note: The values from the external leakage current test should not be over the
- 28 -
- Patient Leakage Current
This test complies with the patient leakage current standards in the Article 19 for the
Class I, Type BF in the IEC 60601-1. This test measures the patient leakage current
(1) Connect the AC power code to the hazard analyzer based on the recommendations.
(3) Measure the patient leakage current of the hazard analyzer based on the
recommendations.
(4) Measure the patient leakage current of the oxygen saturation (SpO2) with the same
method.
Note: The earth leakage current measures AC main power and the conductor in the
protection earth. The values from the earth leakage current test should not be over the
- 29 -
- Patient Leakage Current (Main voltage on the applied parts)
WARNING: The AC main voltage appears while the test is applied. Take care of
The test is based on the patient leakage current in the Article 19 for the Class I, Type BF
in the IEC 60601-1. The test is applied to the part where 110% of the main voltage is
connected to the patient and the earth. The patient leakage current is measured from
Note: Keep the length of the patient test cable as short as possible in the leakage test.
Note: This test requires the connector connected to the patient as explained in the
(1) Connect the AC power code based on the recommendations of the electric hazard
analyzer.
(3) Measure the patient leakage current based on the recommendations of the hazard
analyzer.
(4) Measure the patient leakage current of the oxygen saturation (SpO2) with the same
way.
Note: The patient leakage current measures forward and backward directions of the
main electrodes. The values of the patient leakage current test should not be over the
- 30 -
Table 7. Patient Leakage Current Value – Main voltage application
- Patient-assistance Current
The test complies with the patient-assistance current in the Article 19 of the Class I,
Type BF in the IEC 60601-1. The applied voltage to the IEC 60601-1 is 264 V/50Hz –
(1) Connect the AC power code based on the recommendations of the electric hazard
analyzer.
(2) Form the SpO2 connector and connect to the input terminals of the hazard analyzer.
(4) The values of the hazard analyzer should not be over the values in the Table 8.
- 31 -
6 Disassembly
6.1. Overview
- 32 -
Figure 11. Disassembling
19 Top Cover
- 33 -
Note: The rear speaker is connected to the power board on the bracket. Therefore,
disassemble the case completely after disconnecting the connector for disassembling the
case.
Note: After disassembling the case, disconnect the internal battery connector first.
- 34 -
7 Troubleshooting
WARNING: When you judge that the measurement of the equipment is not
accurate, verify the state of the patients with other verified methods. Also, be
WARNING: Only the authorized personnel for repairing the equipment may
disassemble and inspect Care Vision OM-100. Users may not repair the
7.1. Overview
The equipment shows error codes when it does not work properly due to error of the
System Error 1 Error of communication data between the Main BD and SpO2 BD
- 35 -
7.2. Troubleshooting
- 36 -
Figure 13. Power Problem –Not working by Battery
- 37 -
2. LCD or LED display is dim and hard to read
- 38 -
Figure 15. Display Problem – LCD Problems
- 39 -
3. Measuring Problem
- 40 -
4. No response to the button or no sound
- 41 -
Figure 18. Other Problem – Key Button Problem
- 42 -