Manual Indoor Test Stand
Manual Indoor Test Stand
Contact address:
PSE AG
Emmy-Noether Str. 2
79110 Freiburg
Germany
www.pse.de
February 2011
1. GENERAL INFORMATION .................................................................... 1
1.6. Maintenance................................................................................. 5
I
2.4.3. Removing the glass modules of the artificial sky .................. 18
2.4.4. Disassembling and cleaning of the glass cassettes .............. 19
2.4.5. Assembling and disassembling of the lamps........................ 20
2.4.6. Maintenance....................................................................... 21
2.4.7. Technical specifications ...................................................... 21
3. ELECTRONICS ................................................................................... 47
3.4. Maintenance............................................................................... 58
4. SENSORS ......................................................................................... 59
III
6. GRAPHICAL USER INTERFACE (GUI) .................................................. 71
V
6.13. Web interface and working with it.............................................. 114
1
1.3 Security warnings
To avoid risk of injuries and damage, please read these instructions, especially the security
warnings, carefully!
PSE will not be liable for injuries, defects or damage caused by incorrect operation.
General warning!
Danger général!
Danger of electricity!
Danger électrique!
Please make sure that no persons or parts are in the moving space
of the collector test platform.
Attention installation de convoyage sur rail! T'assure, que ne
personne et pas d'objets se trouve dans l'étendue d'mouvement de
la PEC (plate-forme d'essayer des collecteurs), s.v.p.
Bruising Hazard!
Danger d'écrasement!
2
1 General information
Warning: UV hazard!
Attention: Risque de rayonnement ultraviolet!
Danger of blow-out!
Attention: agent lumineux peut éclater!
Prohibition: No access!
Interdiction: Entrée interdite aux personnes non autorisées!
3
1.3 Security warnings
Mandatory: Wear eye protection! While operating the lamp field, all
staff must wear eye protection against ultraviolet irradiation.
Protection obligatoire de la vue! Pendant le panneau de luminaires
est en service, tout le personnel doit porter des lunettes
protectrices contre le rayonnement ultraviolet.
4
1 General information
1.6. Maintenance
In case of maintenance the system must be disconnected from the grid. Only when the
system is disconnected from the grid staff is allowed to work. In case of maintenance: All
staff members under the lamp field and under the collector test surface must wear helmets.
In case of maintenance: All staff members under the lamp field and
under the collector test surface must wear helmets.
5
1.6 Maintenance
2.2 Vertical Because of wear between spindle gear and screw Every 100 km
carriage nut the nut has to be replaced after 100 km. This
equates to about 39000 times carry up and
down.
Check the lubrication of the rack and the gears of Refer to the manual of
the movement! the manufacturer.
6
1 General information
2.7 Cold water Control the pressure of the cold water circuit! Regularly, best daily
grid
For further information according to the maintenance of the components please refer to the
corresponding subchapters in this manual.
7
1.8 Conformance to engineering standards
1.9. Maxim
8
2. Indoor test stand
11
7 6
4
5
8 9 10
No. Description
1 Lamp field with 8 metal halide lamps for the simulation of the irradiation. Each single
lamp and each lamp row can be moved electrically. Each lamp is controlled by an
Electronic Power Supply (EPS). The complete lamp field, equipped with the artificial
sky, can be moved in vertical direction. Moreover the lamp field may be inclined
within an angle of 0 to 90°.
9
2.1 General description
2 Artificial sky to avoid errors during the test of collector caused by the heat of the
lamps. It consists of 4 glass modules and the air circuit, including air ventilation and
air cooling. The artificial sky is directly attached to the lamp field.
3 Vertical carriage for the height adjustment of the lamp rig (lamp field and artificial
sky).
4 Vertical support, which mainly consists of vertical rails and a spindle gear; it is the
interface of the solar simulator to the building.
5 Collector test platform; it consists of the collector support, the railway carriage and
the thermostat as well as the ventilation unit, the X-Y-scanner and the data
acquisition unit.
6 Collector support, where the test collector is mounted on; the distance to the lamp
field as well as the inclination angle is variable.
7 X-Y-scanner; two axis sensor head with the following sensors: pyranometer for
accurate measurements of the irradiation homogeneity, anemometer for wind speed
measurements, and a solar cell for fast irradiation measurements.
8 Ventilation unit; provides a controlled wind speed on the collector surface. The height
of the ventilation unit can be changed to correspond with different thicknesses of test
collectors.
9 Thermostat; it controls flow and temperature of the heat transfer medium of the
collector circuit.
The description and the operation instructions of further engines such as mobile scanner,
mobile lamp field, mobile climate chamber and solar air collector test stand are given in
separate manuals.
10
2 Indoor test stand
The vertical carriage is a steel structure, which carries the lamp rig in vertical position and is
driven by the main spindle. It also carries a switch cabinet and the cylinders for adjusting
the tilt angle of the lamp rig. Those cylinders each have a spindle inside.
The spindle gear for the vertical movement must not be used more
than 6 minutes per hour (10 % on-time).
The lamp rig consists of a steel structure which supports the lamp field and the artificial sky.
The artificial sky includes the glass cassettes, two ventilation modules and several air ducts
for air circulation as well as the two water-air-coolers. The whole lamp field is adjustable
between 0 and 90 degrees.
Figure 2.2:
Protection boot
Spindle gear of
the vertical
support
Travelling nut
Safety nut x
Main spindle
Electric motor
11
2.2 Vertical carriage
2.2.2. Maintenance
When working on parts of the main spindle (e.g. replacing of the
drive, maintenance of the brake of the drive) the vertical trolley has
to be supported mechanically. The reason is the spindle is not
locked against turning by drive or brake.
Because of friction between spindle gear and screw nut the nut has
to be replaced after 100 km. This equates to about 39000 times
carry up and down.
Please check the distance between the travelling nut and the safety
nut of the spindle gear (denoted by a red x in Figure 2.2). The
starting distance can be found in the inspection sheet in the annex.
The distance must be checked annually; the measured data have to
be documented in the inspection sheet. For more information
according to maintenance of the worm gear screw jack please refer
to the manual of PFAFF-SILBERBLAU® in the annex.
Figure 2.3:
Left: Grease
nipple for the
bearing of the
vertical support
spindle.
Right:
Main string pot
encoder for the
height
measurement
of the vertical
carriage.
12
2 Indoor test stand
Figure 2.4:
Degrees of
freedom of
lamps,
lamp rows
and lamp field.
With these additional degrees of freedom it is possible to adjust the position of the lamps in
the way that other configurations concerning homogeneity (for example a better
1
MHG stands for “metal halide global”.
13
2.3 Lamp field
homogeneity on a smaller area) are possible. How to move the lamps and the lamp rows is
explained in the chapter software.
Very rarely lamps may burst. Damaged or old lamps may burst more
likely. As rare as this is, the system should never be operated
without or with damaged filter glass (i.e. the spectral selective
glass plates of the MHG-lamps).
To fulfil the requirements of solar simulation, the arc of the lamp
has extreme high luminance and emits a high proportion of
ultraviolet radiation. Irreparable eye damage can occur by looking
directly at the arc. Severe skin damage can occur due to an
exposure of just some minutes.
Figure 2.5:
Recommended
protective
sunglasses
14
2 Indoor test stand
2.3.3. Maintenance
Please check lubrication of the gear rack and the gears of the
lamp row movement. For more information according to
maintenance of the movement please refer to the manual of
LEANTECHNIK® in the .
Figure 2.6:
MHG-lamp in
operation seen
through
welder's
goggles.
15
2.3 Lamp field
Effective degrees of freedom of lamp rig 3 linear, 1 rotational (0° to 90°), actuators
and lamps electrically controlled, height and inclination
sensors
Number of lamps 8
For further information please refer to the attached instructions of ATLAS MTT®!
16
2 Indoor test stand
Figure 2.7:
Condensate
Storage
17
2.4 Artificial sky
Work in pairs!
6. With (minimum) two persons only: Slowly pull the glass modules at the knurled
screws, until both people can carry the module at the shorter edge.
Attention: the two knurled screws are only for pulling the glass
module, NOT for carrying it!
18
2 Indoor test stand
7. Store the glass modules in vertical direction in a clean place and avoid damages
during the storage.
Figure 2.9:
Disassemble
the glass
modules:
step 1.
19
2.4 Artificial sky
Figure 2.10 shows, how to push out the rigid PVC mounting of the glass. Work in pairs!
Figure 2.10
Disassemble
the glass
modules:
step 2.
Now you may clean the glass panes with warm, dematerialized water without chemical
cleaners. Hard materials such as steel wool, scrub sponges with a rough side, scrapers or
stuff like that may scratch the surface. Most suitable is soft cotton cloth or chamois leather.
Work in pairs!
Read the attached instructions of ATLAS MTT®!
Before disassembling the lamps, please first remove the artificial sky glass modules as
described above. Then act as described in the following:
1. Make sure the lamps are disconnected from load.
2. Remove the glass pane which covers the lamp. Use clean cotton gloves.
3. Remove the lamp by loosing all fixture screws.
To mount the lamps, please work in reverse order.
20
2 Indoor test stand
2.4.6. Maintenance
Dimensions of the artificial sky 4 glass modules with circa 2200 x 1100 mm
each
For more information please refer to the datasheets and manuals of glass and ventilation
modules in the annex.
21
2.5 Collector test platform
No. Description
3 Electrically swiveling collector support with integrated electronic tilt angle control.
4 X-Y-scanner with pyranometer, solar cell and anemometer. The first two instruments
are suitable for measurement of irradiance. The latter one detects the wind velocity
on the test surface.
5 Ventilation unit mounted on the lower edge of the test surface; electrically
adjustable to different collector thicknesses.
22
2 Indoor test stand
8 Electrical cabinet "DA2" with Data acquisition unit aka data logger.
9 Box containing pipes, pump and valve for flow control and flow regulation of the
collector circuit. Also contained: heat exchanger between Thermostat and collector
circuit.
12 Trench with rails and energy chains (containing power supply and cold water circuit
pipes).
23
2.5 Collector test platform
Figure 2.12:
The
horizontally
drivable
collector test
platform. This
picture was
taken at a
former indoor
test stand
similar to your
one, but
without a
trench.
Automatically positioning is possible via centralized controls. The horizontal axis for the
collector distance movement is the rail slide in the trench. The movement is done by an
electrical gear-wheel drive which is in contact with a gear rack at the rail slide (manual here).
A second gear-wheel is connected with a sensor and communicates the distances to the
controller. On the basis of this the distance of the horizontal carriage to the solar simulator
can be varied and monitored by the controller.
Figure 2.13:
Electrical gear-
wheel drive c,
sensor of
distance d
and gear rail
e. 2 1
24
2 Indoor test stand
Please make sure that no persons or parts are in the moving space
of the collector test platform, particularly in the trench!.
25
2.5 Collector test platform
This is only allowed, if the collector test platform has been driven in
a horizontal position!
Wear Hard hat while every lifting operation, thus also at-the
collector test platform separation!
The collector test platform and the undercarriage may be separated, as shown in Figure
2.14.
Figure 2.14:
collector test
platform and
undercarriage.
26
2 Indoor test stand
Figure 2.15:
Power supply
connectors of
the collector
test platform.
Then shut the four red stopcocks visible in Figure 2.16. Afterwards separate the collector
test platform from the main cooling circuit by loosing the fitting between the two stopcocks
of supply and return flow.
27
2.5 Collector test platform
Afterwards loose the gear wheel. But never touch the positioning sensor otherwise the
platform will loose the position and the collision detection will not work.
Figure 2. 17:
Electrical gear
wheel.
Now loose the four arrest screws as visible in Figure 2.18 and remove the yellow clamping
plates.
Figure 2.18:
Loosing the
arrest screws
for the collector
test platform.
After this fasten round slings at the four lifting points as shown in Figure 2.19. Fix the two
green slings (length = 2 m) at the therefore destined eyes and the violet ones (length =
1.5 m) at the tilt bearings. Use only the provided polyester round slings. Don't use steel
chains; they could damage the surface of the steel structure. If you do it in another way, the
platform may tilt and crash when lifting …
Conduct the round slings through the superstructures in a way, so that they don’t lie against
any parts (see Figure 2.20). The latter could deform irreparable while lifting.
28
2 Indoor test stand
Figure 2.19:
Fastening of
the round
slings for
collector test
platform lifting.
Now you may lift up the collector test platform. Don’t bother about, if it hangs a bit lopsided.
The sum of the four lifting points doesn't coincide exactly with the center of mass.
Figure 2.20:
Round sling
conduction.
29
2.5 Collector test platform
Well! Now mount the four BLICKLE® swivel caster with brake (foot operated) at the destined
flange plates as shown in Figure 2.21.
Figure 2.21:
Example
mounting
position of the
swivel caster
with brake (foot
operated).
After mounting the heavy duty rollers you can lift the collector test platform down and move
to any other suitable place. Don't forget to lock the foot parking brake after each relocation.
If you want to set back the collector platform onto the undercarriage work in reverse order,
as described above.
Do the "reunion operation" only in pairs: one person controlling the overhead crane, the
other one controlling the position of the collector test platform. First fix the four polyester
slings again, lift up and remove the rollers.
Let the platform down very slowly onto the undercarriage. During that pay attention that the
dowel pin hits the corresponding conic bushing.
Finally mount the arrest screws an clamping plates again.
Figure 2.22:
Setting
platform back
on the
undercarriage.
30
2 Indoor test stand
Figure 2.23:
Picture of
electrically
swiveling test
platform with
lamps shining
bright. In the
foreground you
see some parts
of the solar air
collector test
stand.
Maintenance, collector fixing and all other works at the platform must only be done in
horizontal end position of the collector support. Please make sure that the collector is fixed
and that there are no parts, e.g. tools, which may fall down when swiveling the tilt angle.
For the connection of the test specimen to the collector circuit, use hoses and fittings
appropriate for very hot and cold water.
31
2.5 Collector test platform
Attention! Make sure, that all locking lever of the bracket retainers
are fixed, before swiveling the test plane!
Figure 2.24:
Support for
semitransparent
PV-modules:
c Holding
Bracket;
d Retainer for
the Brackets;
e Guide rail;
f Example PV- 4
Module;
g Module arrest 5
2
3
To measure the transparency or more precisely the (mean) transmittance T , you may use the
pyranometer installed on the scanner — or alternatively you use an custom irradiance sensor
such as a luxmeter or another sensor with a higher sensitivity.
32
2 Indoor test stand
In either case you first have to carry out a scan to detect the irradiance distribution without
shading. Optimize the lamp positions if desired and repeat the first scan. You'll receive a
discrete matrix of irradiance values E ( xi , y j ) and a mean value E .
Now you may put the module specimen onto the brackets and clamp them with the destined
star-handle arrest screws.
Finally do a second irradiance-scan with the same parameters and the installed shading PV-
modules. This gives you a second data matrix E ' ( xi , y j ) . The mean E ' is.
Finally the mean transmittance/transparency of the PV-module is
E
T= .
E'
Please notice that the simulated sunlight of the simulator isn't as parallel or collimated as
natural sunlight.
2.5.6. X-Y-Scanner
The irradiance and the wind speed in the collector plane have to be measured.
For this purpose a pyranometer is moved automatically over the test surface, with the help of
an X-Y-scanner, and measures the irradiance distribution on the test surface. An additionally
solar cell is installed which allows fast but not that much precise scans of the irradiance. So
this cell cannot be used for the computing of the efficiency factor of the collector but allows
a good overview of the irradiance-homogeneity in the collector plane.
Figure 2.25:
X-Y-Scanner
with
pyranometer,
solar cell and
anemometer.
33
2.5 Collector test platform
Figure 2.26:
X-Y-scanner
with
pyranometer,
solar cell and
anemometer
(CAD drawing).
The third sensor on the X-Y-scanner is an anemometer to measure the wind velocity in the
collector plane according to the EN 12975:2006. It is fixed by a magnet and can be easily set
in protection position (see "brown position" in Figure 2.1). The height of the test probe
above the collector surface has to be set manually.
More and detailed information concerning the sensors you'll find in chapter 4.
Figure 2.27:
X-Y-scanner
with controlled
axes.
Due to the installation height of the pyranometer and the X-Y-scanner, it is not possible to
take irradiance measurements exactly at the level of the collector cover. Therefore, the
measurement data must be taken on a parallel level above the collector cover. After the
34
2 Indoor test stand
irradiance distribution determination, the collector surface must be shifted into the test
level.
You can do this by means of the central control touch screen. The required movement of
collector platform and lamp field is calculated based on the positions and angles of lamp
field and test surface in combination with the thickness of the collector. This repositioning
must be carried out by the test stand user via the centralized controls (see chapter 6
software).
The software developed especially for this purpose ensures an automatic scan process as
well as recording and graphic representation of the measurement results.
Some parts (the brakes) of the linear motion step motors may get
very hot!
When the contact plate sensor of the carriage slide is nearby an inductive sensor the light
goes out and the sensor sends a signal to the control unit. That's no problem if it happens at
the reference sensor. This one serves for orientation that is as benchmark for the movement.
The other two "end"-switches shouldn't be approached in normal operation. Anyhow that
can happen e.g. during a software breakdown. Then the scanner-motion will stop
immediately and can't be activated via the GUI.
In this case you've got to release the brake. There's a button to do this in the GUI (see
section 6.10.2).
35
2.5 Collector test platform
Attention!
The collector test platform has to be driven in 0°-position, i.e.
horizontal. Otherwise the cross axis may crash down uncontrollable
and generate serious damage at person and machinery.
If the brake is released you may push the scanner axis manually away from the limit switch.
Possibly you still get an error message on the GUI-screen. Then it's advisable to restart the
GUI.
The ventilation unit is mounted on the lower edge of the test surface. You can vary the height
of the ventilation unit electrically to adapt the position of the air flow to the collector
surface. The air flow speed and the position is set via the test stand’s centralized control
(see section 6.3.7).
36
2 Indoor test stand
2.5.9. Maintenance
Figure 2.30:
Grease fittings
at the lifting
electric
cylinders.
37
2.5 Collector test platform
Horizontal carriage
Railway 9460 mm
Precision ± 1°
38
2 Indoor test stand
Ventilation Unit
X-Y-Scanner
Precision ± 1 mm
39
2.6 Thermostat and collector circuit
Figure 2.31:
Hydraulic plan
of the
thermostat and
collector
circuit.
Strictly speaking there are three circuits: on the left side you see the collector circuit. It's
interconnected with the second circuit of the thermostat via a heat exchanger. Purpose of
the thermostat is heating and cooling of the fluid in the collector circuit.
The thermostat is a "HUBER® Unistat 510w FB spez" and its operation manual is attached in
the annex. Please look there for further information and operation instructions. Switch on
the thermostat before testing. Control happens by means of LabVIEW.
The chiller inside the thermostat is cooled by water out of the cold water grid (see section
2.7) – and that's the third circuit. The internal cooling and heating circuit of the Thermostat
requires special silicon Oil type M40.165.10 (see datasheet here).
Take care that the level of the silicon oil in the sight glass is above
the minimum mark!
The volume flow is recorded by an electromagnetic flow sensor (EFS) and regulated with the
help of a pump controlled via a frequency inverter and a special motorized control valve. In
40
2 Indoor test stand
addition, a temperature sensor on the volume flow meter measures the temperature of the
fluid, so the mass flow can be calculated (see more in chapter 4 and 1).
A drainage tap with a short piece of hose on it allows a complete discharge of the collector
circuit. On the left and right side of the EFS you find to valves with a hose connection facility
for volumetric dosing of the EFS.
Figure 2.32:
Mobile filling
station.
To start the deaeration, the filling station has to be connected to the filling valves on the
collector test platform (c and d.in Figure 2.33).
Figure 2.33:
Filling valves
2
on the collector
platform.
5
41
2.6 Thermostat and collector circuit
42
2 Indoor test stand
There are much more technical specifications and you'll find them in the attached data sheet
which can be found in the end of the manual. For even more operation information please
refer to the manual in the annex.
43
2.7 Connection to the cold water grid
Figure 2.34:
Drawing of the
cooling system
and its 5
6
connection to
the cold water
grid.
3
4 1
No. Description
3 Inlet thermostat
4 Outlet thermostat
44
2 Indoor test stand
2.7.2. Maintenance
45
3. Electronics
The electric components are installed in different cabinets. In the following you will find a
short description of the signals and switches installed at the electric cabinets. For more
technical details please refer the wiring diagrams.
47
3.1 Electrical Cabinets
The switches and buttons at the front side are described in the following picture and chart.
48
3 Electronics
Figure 3.1
Electrical
Cabinet SS 1.
6
9
10 5
11
4
12
3
4 Power Supply Shining, if power supply of the drives for changing the
Lamp position single lamp positions is established.
49
3.1 Electrical Cabinets
9 Reset Emergency Shut Off Reset power supply for the main drives after
activating the emergency shut down.
10 Power Supply Shining, if power supply of the drive for changing the
Lamp Field Height height position of the lamp field is established.
11 Power Supply Shining, if the power supply of the drive for changing
Lamp Field Inclination the inclination of the lamp field is established.
50
3 Electronics
51
3.1 Electrical Cabinets
1 3
1 Main Switch Turns the power supply of the collector test platform
on/off.
52
3 Electronics
Figure 3.3:
Electrical
Cabinet: “Data
acquisition”
2 3 4
5 6 7 8
9 10 11 12
2 Connector 1 • dp_V_50_air
• dp_V_250_air
3 Connector 2 • dp_V_2500_air
• dp_room/air
4 Connector 3 • p_room_air
5 Connector 4 • T_in_air
6 Connector 5 • T_out_air
7 Connector 6 • T_massflow_air
8 Connector 7 • T_heatexchanger_air
9 Connector 8 • dp_collector_in/out_air
10 Connector 9 • dp_collector_in/room_air
53
3.1 Electrical Cabinets
11 Connector 10 • Humidity/
• T_amb_air
12 Connector 11 • safety_valve_pressure
• frequency_inverter
54
3 Electronics
Figure 3.2:
Electrical
Cabinet: ATLAS
Lamp
controller.
55
3.3 Emergency stop
Figure 3.4 shows an example of such a limit switch. Depending on which one of them gets
activated, the corresponding actual movement(s) will stop directly. Movements can start
again when they are released. This is done automatically by the software.
56
3 Electronics
Figure 3.5:
Safety rope pull
switch
2
1
No. Description
Figure 3.6:
Operation
pendant station
The push button must be pressed down during every movement of the test stand. By
releasing it, the motion will stop instantaneous.
57
3.4 Maintenance
Pressing down the emergency stop operator will activate the emergency shut off chain. Then
the following actuators are stopped by interrupting the power supply:
1. Lamp field height drive
2. Lamp field inclination drive
3. Collector support position drive
4. Collector support inclination drive
5. Ventilation collector support
6. Artificial sky
The same happens if you push down the emergency shut off operator at the SS1
3.4. Maintenance
Please check every 12 month if all limit switches, emergency
shut off chains and kill switches work.
58
4. Sensors
In this chapter all the Sensors connected to the data logger are described. They are all
connected to the data logger and displayed in the LabVIEW application.
A little 60 W splashing water proof fan (IP54) creates a constant air volume flow round the
housing of the Pt100-Sensor according to EN 12975-2. For the temperature calculation out of
the measured resistivity values DIN EN 60751 parameters are used. See also the tabular in
the following section.
59
4.3 Pyranometer
Their tolerance class is 1/3 DIN B. In Figure 2.31 they are marked by this sign:
For more accuracy and to eliminate the resistivity of the cables, all temperature sensors
(also these for air temperature measurement) are wired by 4-point probes method except
that one for the cooling water temperature (T_cooling_water). That's why each sensor needs
two channels on the multiplexer card 34921A. The connection on the modules for the 4-wire-
configuration is always 2 wires on channel 1…20 (Bank 1) and two wires on the channel
20…40 (Bank 2).
The sensors delivered are calibrated and the certificates can be found in the annex (see also
section 4.9). The following table gives a survey of all Pt100 Sensors, their denomination in
LabVIEW and their fitting parameters calculated out of the calibration certificates. These
parameters are needed for the calculation of the temperature T out of a measured
resistivity value R . LabVIEW handles this calculation and the corresponding formula is
[ ]
1
− R0 A + ( R0 A) 2 − 4 R0 B ( R0 − R ) 2
T ( R) =
2 R0 B
Also Sensors of the solar air collector test stand are listed. For the calibration parameters
which have to be entered in the LabVIEW VI please refer to the Excel sheet in the annex.
• T_in_collector
• T_out_collector
• T_massflow
• T_ambient
• T_ventilation_left
• T_ventilation_right
• T_in_air
• T_out_air
• T_massflow_air
• T_heat_exchanger_air
• T_cooling_water
4.3. Pyranometer
A pyranometer (see Figure 2.25) is a wavelength-integrating radiometer and measures
irradiance values.
60
4 Sensors
The homogeneity of the irradiance on the collector surface must be monitored before and
after each test according to EN 12975:2006 to enable reproducibility of the test conditions.
The irradiance measurements have to be made on the same level as the transparent
collector cover. By entering the collector thickness in the software-menu "measure" you may
accomplish a correction movement, so that the collector surface gets driven into the
previously scanned plane.
The pyranometer is mounted on the X-Y-scanner head. Thus one can gather irradiance data
points in a grid inside of the test-area. The scan time for every data point of the grid has to
be at least 5 seconds resulting from the response time of the pyranometer. Moreover you've
to add circa 3 seconds for the linear movement of the scanner — dependent of the grid
fineness.
The Pyranometer type CMP 21 is fabricated by KIPP & ZONEN. It provides small voltage
signals to the data logger. Afterwards the software computes the irradiance value by means
of the sensitivity. The CMP21 has an individual calibration certificate containing the
sensitivity value.
The sensitivity is temperature-dependent and therefore the calibration certificate contains
data for this dependency. The resulting effect isn't very large. Nevertheless the GUI-software
regards this factor. Other correction terms or offset-effects — e.g. cosine-deviation,
nonlinearity or radiation transfer effects — won't be comprised 'cause they're too small or
hard to estimate. There's a detailed user's guide in the annex containing information related
to that measurement errors and offsets. Look here for a short instruction sheet.
Please clean the quartz glass dome regularly with soft cloth
and water or alcohol.
Keep the instrument leveled!
61
4.4 Solar cell
Response tome 5s
62
4 Sensors
4.5. Anemometer
The wind speed in the collector plane has to be measured by an anemometer for valid test
sequences. Like all the other sensors the velocity transmitter D12-65V C is wired to the
AGILENT data logger. The values delivered by the sensor lies in a range of 0 V to 10 V. The
wind speed can be calculated by a linear relation from 0 up to 10 m/s with an accuracy of
±0.1 m/s.
The anemometer is mounted on the X-Y-scanner-head as like as the pyranometer and the
solar cell (see Figure 2.25 and Figure 2.26). The height has to be set manually (please refer
to chapter X-Y-scanner).
For more information have a look to the corresponding datasheet.
63
4.6 Volume flow meter
Figure 4.3:
In the annex you'll find a datasheet, quickstart instructions and a detailed handbook for the
Motorized
Optiflux 4000. Also there are datasheet, quickstart instructions and handbook for the IFC
valve in the
300.
collector
circuit.
Sensors must be recalibrated from time to time!
The calibration intervals have to be defined by the user.
To guarantee as constant as possible volume flow, i.e. to minimize fluctuations a bypass has
been installed in the collector circuit. The cut off valve in the bypass must always be opened
during testing operation. For a volume flow higher than 300 l/h the valve has to be closed.
The volume flow can be regulated by a driving speed change. Moreover there's a motorized
valve as showed in Figure 4.3 to do this.
This valve gets controlled by voltage signals from LabVIEW. A manual adjustment can be
done by means of the black colored hand wheel. The actual condition of the valve can be
read by a little indicator on the scale at the bottom side. There's a. If the indicator is at the
top, the valve is open and contrariwise it's at the bottom by closed valve.
64
4 Sensors
4.7. Manometer
There's an analog manometer in the collector circuit. It shows the actual pressure in the
collector circuit. Operating pressure should be 1.5 bar.
Figure 4.4:
Analog
manometer in
the collector
circuit.
65
4.9 Data logger
T_cooling_water 011
66
4 Sensors
<disengaged> 031
T_amb_air 012
<disengaged> 032
<disengaged> 013
<disengaged> 033
dp_room/volume_air 016
<disengaged> 036
dp_collector_in/out_air 017 x
<disengaged> 037
dp_collector_in/room_air 018 x
<disengaged> 038
Humidity/Temp_air 019 x
<disengaged> 039
dp_collector_in/room_air 020 x
<disengaged> 040
The following table shows some technical specifications of the data logger (switch 34980A):
67
4.9 Data logger
Power 60 W
68
5. LabVIEW – Measurement & control
The measurement and the control according to collector tests are driven by a LabVIEW VI.
The functions are shown in a separate manual.
69
6. Graphical User Interface (GUI)
6.1. Overview
The main test stand and all additional components get controlled by a graphical user
interface (GUI). This GUI can be operated via a touch screen (SS1). This enables controlling
not only of the main lamp field, artificial sky and the collector test platform but also of the
mobile lamp field and the mobile scanner. The control includes operation of all mechanically
moving components plus dimming of the lamps. The additional testing devices have their
own manual. Control of the solar air collector test stand happens via LabVIEW. The mobile
climate chamber has an own control panel and isn't connected to the central touch. Figure
6.1 gives you an overview about the control interconnections between all testing devices.
• And finally a pyranometer scan can be done in the Main Scanner menu (section 6.7)
before starting the intended collector measurement (e.g. characteristic curve).
71
6.2 Menu bar
• Affiliated to this are the control items for Mobile Lamp field (section 6.8) and Mobile
Scanner (section 6.9) control.
• In the Setup panel you may do different adjustments and special operations for
maintenance as well as entry calibration parameters (see section 6.10).
• With the Lock function you may protect the touch screen against unwanted intervention
of a third party (section 0).
Each of these menu items will be explained in a separate section in the following. Please
note: in every interface all selectable buttons arrows or input fields are marked in light
Figure 6.1: purple.
Overview of all
controllable
devices.
72
6 Graphical User Interface (GUI)
9
3 4 5
No. Description
1 Menu bar.
2 Status frame.
4 Buttons for setting mobile lamp field and transparent PV collision control
73
6.3 Stand control
Figure 6.3:
Stand control.
Desired
position shown
in yellow,
actual or
current
position shown
as gray
shadow.
In the right upper corner of the status frame you may read precise values for lamp field
height and inclination (first two lines) and collector support position and angle (bottom two
lines).
74
6 Graphical User Interface (GUI)
Figure 6.4:
Stand control
with status
frame in current
view.
75
6.3 Stand control
Figure 6.5:
The file
manager when
saving.
Figure 6.6:
The file
manager when
loading.
76
6 Graphical User Interface (GUI)
Figure 6.7:
Stand menu
with messages
in the error
panel.
77
6.3 Stand control
How the GUI looks has already been showed in Figure 6.2. In this mode only entries for the
inclination of both – lamp field and collector support – as well as for the distance between
lamps and measurement plane of the pyranometer are possible.
The set distance between the lamps and the test platform will be maintained if you change
the angle value. The same applies for distance changing.
Distance and angle value may be increased or reduced by multiple tapping on the little blue
triangles on the left and the right side of each input field.
Alternatively you may tap on the value itself. This will bring up a numeric keypad where an
arbitrary value may be entered.
Please note: In the upper part of the keypad the permitted interval for values is shown. For
each input field and corresponding keypad there are of course different intervals. Apart from
that the program will let you know, if an entered value is forbidden.
Figure 6.8:
Stand panel
(in direct
control mode).
78
6 Graphical User Interface (GUI)
In the collector support box you may enter position, angle and height. Let's see: Position of
the CTP means the horizontal distance to the wall which is the vertical carriage is hanging
on. Angle is the inclination of the collector support and Hgt (height) means the height of the
electrically drive collector support ventilation. Changing this value enables the adjustment
of the air flow to the collector surface.
By means of the direct control also e.g. (implausible required) configurations with different
inclinations of test surface and lamp field are possible. Have fun!
The Start Movement button will not be activated if any errors are detected. If you try to start
the movement while there are still error-messages, the test stand drawing on the screen will
get red colored and nothing will move like shown in Figure 6.9. Then you've got to tap the
Reset Error button and eliminate all error messages before again start the movement.
But if no error messages exist anymore and if all settings for the target position are done,
tap on that button and you'll get requested to press down the green push button on the
pendant operation station (Figure 3.6).
A warning signal sounds and finally the test stand will start to move.
During the movement the screen will turn like in Figure 6.10. The current position is appears
displayed in blue and the target position is shown as light gray contour in the background.
While the system is moving it's not possible to change the target.
Hold down the button during the whole movement. The movement is finished, if the stand on
the screen turns yellow again.
The movement may be stopped at any time either by releasing the bush button or by tapping
on Stop Movement.
79
6.3 Stand control
Figure 6.10:
Stand menu
during
movement.
80
6 Graphical User Interface (GUI)
6.4. Lamp
Tap on Lamp in the menu bar and you'll reach the screen shown in Figure 6.11. The Lamp
interface is used for positioning of the lamps in the main lamp field. It contains 6 panels
whose functions will be explained in the following.
Figure 6.11:
2 1 3
Lamp control
interface.
No. Description
3 Target panel.
6 Actions panel.
81
6.4 Lamp
Figure 6.12:
Lamp Panel
after selecting
a lamp row.
Note the
selection box
surrounding
the selected
lamps and the
highlighted
arrow controls.
82
6 Graphical User Interface (GUI)
Figure 6.13:
Change lamp
positioning by
drag-and-drop.
83
6.4 Lamp
The lamps may only be positioned pair wise. This means that only one row at a time may be
selected.
When the target position has been set, both the target and current positions are indicated
on the center panel. The current position is indicated with a gray outline and the target
position with a yellow square (see Figure 6.14).
Figure 6.14:
Lamp Panel
after changing
the target
position.
Please note the
gray outline
indication a
current lamp
position.
84
6 Graphical User Interface (GUI)
Figure 6.15:
Lamp formation
panel with
errors indicated
in the top in red
and detailed on
the bottom
error list.
85
6.4 Lamp
Figure 6.16:
Lamp interface
after starting a
movement.
86
6 Graphical User Interface (GUI)
6.5. Measure
Do the following steps only after you have set the lamps and
did the scanning process.
The reference plane for calculating the distance between collector support and lamp field is
the scanning plane of the pyranometer. Exactly in this scanning plane the irradiance
distribution gets measured. Maybe this plane is special distinguished 'cause you've got
optimized the lamp constellation to get there a very well irradiance homogeneity or a
particular mean irradiance value.
Unfortunately this measurement plane isn't the same as the collector surface, because the
pyranometer is running above the test sample. So that’s why you've to carry the collector
surface in the analyzed measurement plane. This has to be done always after scanning and
before beginning a collector test program. Otherwise your identified parameters are not
valid.
Hence a measure interface has been designed. Tap on Measure in the menu bar and the
interface shown in Fig 6.17 will open.
All you've got to do now is enter the accurate collector thickness in the provided input field
on the right side. Then tap on Start Movement.
Similar to the Stand interface you'll get requested to press and hold down the green push
button on the pendant operating station. Don't release the button until the movement has
finished completely. The error panel shows you, what's actually going on.
87
6.6 Main Lamp field
Figure 6.18:
Measurement
interface.
Figure 6.19:
The Main Lamp
field interface.
88
6 Graphical User Interface (GUI)
1 4
6
No. Description
1 Temperatures panel.
5 Actions panel.
89
6.6 Main Lamp field
Figure 6.20:
Artificial sky
panel, when
switched on.
90
6 Graphical User Interface (GUI)
Figure 6.21:
Pop-up-panel
for entering
new dimming
values.
Figure 6.22:
A desired
operation
status will be
shine yellow in
the Target
Position panel.
91
6.6 Main Lamp field
You may possibly hear the switching in the Atlas cabinet before igniting. During the lamps
are in operation they will send out a noticeable buzzing. But that is normal.
The operation status of the lamps is shown in the operation status panel as given exemplary
in Fehler! Verweisquelle konnte nicht gefunden werden.. For more details have a look to the
upcoming section.
Tapping on Lamps Off will deactivate the whole lamp field. Please note: After switch off the
lamps you've got to wait 10 minutes before reactivation the lamps. This is necessary to
avoid damage of the lamp bulbs.
92
6 Graphical User Interface (GUI)
Figure 6.23:
Operation
panel with
lamps at
different
intensities.
93
6.6 Main Lamp field
Setting Description
On / … % The lamp is switched on. The current intensity is also shown as percentage of
the maximum output intensity. Each lamp square is green colored. The lower
the intensity, the darker the color.
Error A lamp error has occurred. If this happens please contact PSE.
Cool The lamp is cooling down. It takes up to 10 minutes before the lamp may be
switched on again.
94
6 Graphical User Interface (GUI)
1 3 4 Figure 6.24:
The Main
Scanner
interface.
6
2
No. Description
1 Measurements panel.
2 Ventilation panel.
3 Visualization display.
95
6.7 Main scanner control
Tap on any position in the display. The scanner head will immediately start to move to this
position. The screen turns in a way shown in Figure 6.26.
As you can see, the target position gets shown as circular outline, whereas the actual
position is indicated by a blue filled and moving circle.
The movement starts without delay. Pressing any start movement button isn't necessary. All
control buttons except the Stop Movement button will turn gray implying: locked. Tap on
Stop Movement will break the actual movement immediately!
96
6 Graphical User Interface (GUI)
Figure 6.26:
Main Scanner
interface
during scanner
head is moving.
The visualization display also will show other activities and settings explained in the
following sections.
97
6.7 Main scanner control
Then tap on any other point in the visualization display. If the scanner head movement has
been completed, tap the Set End Position button. Now the area defined by start and end
position will appear as a blue rectangle. Figure 6.27 illustrates that.
Modifying scan area dimensions is multiple possible by choosing any other end position in
the same way as described above. Start position will remain the same. Only if you change
the start position, also additionally the end position has to been set again.
The described steps have to be performed in this order.
Figure 6.27:
A defined scan
area will
appear blue in
the
visualization
display.
Dwell time
Each measurement instrument has a specific response time. The CMP 21 pyranometer has a
95% response time of 5 s. So that's why you should grant a minimum 5 s dwell time to the
instrument for getting reliable measurement results. Longer dwell times will probably rise
the precision but unfortunately also massive prolong the total measurement time of the
whole scanning procedure. Statistically speaking it’s anyway more practical to make several
measurements to minimize measurement errors.
98
6 Graphical User Interface (GUI)
In practice the scanner head will be driven to a desired point. Then it waits for the declared
dwell time and after expiration of that time, the actual measurement value is transmitted to
the data logger.
You can change the dwell time by tapping on the Time (s) input text field. A keypad will pop
up as shown in Figure 6.28.
Figure 6.28:
A pop up key
pad for
entering scan
area settings.
Alternatively the value can get de- or increased by tapping to the arrows on the left and the
right of the input text field.
Anemometer 4s
Pyranometer 5s
Grid fineness
This input text field enables adjustment of the grid fineness of the data point scan. The grid
is symbolized by small yellow dots in the central visualization display. During a scan, the
scanner head will move incrementally in X- and Y-direction on this grid.
Also the direct scanner head control by arrow accesses to that grid.
Please note: the scan area may only be an integer multiple of the grid. For instance, if the
grid is 100 mm and the user enters a collector width of 1050 mm, the system will round the
scan area to an integral multiple of 100 mm (in this case 1100 mm).
99
6.7 Main scanner control
Park Scanner
Tap on Park Scanner to move the scanner head to the park position at the bottom right
corner.
Goto Start
Tapping on that button will move the scanner head to the user defined start position. If no
start position has been set so far, the start position is the ultimate down left grid point.
Fast Scan
You can only accomplish a scan, if a scan area has been declared.
Sometimes the user only wants to get a rough overview of the irradiance distribution in the
measurement plane. Therefore a fast scan facility is given. In this case the scanner will not
stop at each scan point. How to watch the values and switch between different view modes
will be described in the following section 6.7.8.
You better only observe the measurement values of the solar cell during a fast scan, 'cause
its response time is considerable shorter than that one of the pyranometer.
However the results are less accurate. Pyranometer values will still be recorded but they will
be inaccurate and hence not very meaningful.
Start the fast scan by tapping on Fast Scan.
Pyranometer Scan
You can only accomplish a scan, if a scan area has been declared.
The pyranometer scan is the most accurate integrating irradiance distribution measurement
method, given in this test stand. The accuracy partially depends on the dwell time, as
explained above.
A pyranometer scan will scan each grid point and wait for the dwell time indicated in the
time setting before recording a measurement value. How to view the results will be
explained in the upcoming section.
100
6 Graphical User Interface (GUI)
Start a pyranometer Scan by tapping on Pyranometer Scan. The software will record raw data
(voltage signals) from the pyranometer as well as thermistor signals allowing a calculation of
the housing temperature of the pyranometer and a temperature correction of its sensitivity.
The Measurement panel shows temperature-corrected raw irradiance data! The temperature
corrected data can also be displayed.
While the scanner is moving, all buttons shine light gray except the Stop Movement button.
Tap on this button interrupts every movement of the scanner head instantaneous.
Scan
If a scan is in progress or has been completed, the scan area in the visualization display will
contain colored dots as shown in Figure 6.29. The color expresses the deviation Δ from the
mean value. Red means Δ ≥ ±10%, yellow symbolizes Δ ≥ ±5% and green is Δ ≤ ± 5%.
Tapping on Scan always brings you back to this view mode.
The representation of measurement values by an small circle or dot is physically correct.
Because strictly speaking you can (particularly with the pyranometer) only measure the
(area-related) irradiance referring to the finite, circular and black the sensor element
surface.
According to DIN EN 60904-9 it is a well sufficient assumption, that an irradiance
measurement at a single point represents excellent the mean irradiance for the whole
surface of the related mesh of the measurement grid.
But this indeed a lot depends on the fineness of the grid. For example the standard
mentioned above contains a precise specification the required fineness.
Hence the representation in the zoom view is a tessellation of squares (see below).
You may switch between different views of anemometer, pyranometer and solar cell
101
6.7 Main scanner control
Figure 6.29:
Main scanner
panel during a
scan.
Zoom
Sometimes the normal view mode showing only colored dots will be unsatisfying. You want
to get further information such as numerical values? Or enlarge details of the scan area –
especially if the grid fineness is high?
Then you may obtain a zoomed view of the scan area by tapping on the Zoom button. The
screen flips to that one shown exemplarily in Figure 6.30. On the zoomed display you see
the raw measurement data as numbers directly superimposed over the scan area containing
colored squares. The color scheme has the same meaning as with the dots in normal view.
Additionally the average (mean) of all measured values so far is shown in the left upper
corner as well as current minimum and maximum.
In the zoom view mode there's also a button bar at the bottom of the screen. Its functions
are principally self-explanatory…
So Zoom Out and Zoom In will enlarge or shrink the view. The three buttons Pyranometer,
Solar Cell and Anemometer are once more destined for switching between the results of
these instruments. The None, Value and Difference button triplet enables viewing the raw
values itself, their difference to the mean or none values at all.
Tapping the Back button brings you back to the un-zoomed main scanner interface.
102
6 Graphical User Interface (GUI)
Move the regarded excerpt of the scan area either by sliding it around with your finger
directly on the screen or by using the arrow keys in the right upper corner of the screen.
Figure 6.30:
Zoomed view of
the scan area.
Clear
Tapping this button will delete all not saved measurement data on the screen. As a
precaution a pop-up-dialogue will demand for your confirmation.
103
6.7 Main scanner control
Tap on Filename … and a keyboard will appear for entering a favored file name (only small
letters allowed).
It's also possible to view previous scan results. Simply tap on Load and a dialog box will
come up asking you if you want to load a scan. Tap on Yes and the file manager will open.
Select a desired file by tapping on it. It will be emphasized in dark blue like shown in Figure
6.6. Tap on load.
Please note: Save the actual scan before loading another one. Otherwise it will be deleted.
Loading will not affect the scan area start and end position settings.
Lamp
With the Lamp-button you may fade in or out the as light shadows. Look at this: The lamp
positions are horizontally mirrored by contrast with the view in the Lamp positioning and
dimming interfaces (compare the numbers). But that is intended that way, 'cause now you're
looking from behind through the lamp field onto the scan surface. So everything is as it
should be.
Even the operation status coloration and realistic lamp positions have been taken into
account.
The fade in function is disabled, if lamp field and collector support aren't in parallel position
(not in angle / distance mode).
Figure 6.31:
Lamps can be
superimposed
to the scan
area.
104
6 Graphical User Interface (GUI)
1 6 Figure 6.32:
4
Controls of the
mobile lamp
field.
No. Description
105
6.8 Mobile lamp field control
Attention!
Make sure that no persons or objects are located in the
operation radius of the mobile lamp field!
Danger of death or serious injuries respectively destruction or
damages at man and machine!
In the mechanical configuration visualization below the start movement button you can
track, what's actually going on. The current height is indicated in millimeters in the upper
section of the panel.
During the movement there's also a short description of the current processes in the error
panel g in Figure 6.32.
After tapping the Start Movement button this button turns into Stop Movement. It is
obviously, that tapping this button will stop the actual movement. Even easier stopping will
be achieved by release the dead man push button. However in this case an error message
will appear and it's necessary to reset this error.
Height adjustment during lamp operation is allowed.
106
6 Graphical User Interface (GUI)
Don't worry about the windows and the glass panes of the big climate chamber. They will not
be affected, if you're positioning the lamps directly in front of them.
Figure 6.33:
Mobile lamp
field interface,
when lamps are
in operation.
107
6.9 Mobile scanner control
Tap 2 on Mobile Scanner in the menu bar and you'll see the screen shown in Figure 6.34.
Actually it's very similar to that one of the main scanner. So please have a look to
section 6.7 to get an explanation of the main functions and controls.
Let's see the main differences: First there's only one measurement instrument. On delivery it
is a CMP 11 from Kipp & Zonen. Moreover the maximum scan area is smaller. The mobile
scanner is neither equipped with a ventilation unit. Finally mobile lamp field and mobile
scanner aren't structurally jointed among each other. So an automatic detection of their
relative positions isn't possible. Furthermore lamp positioning can only be done by hand.
That's why the lamps can't be faded in to the scan area.
Figure 6.34:
Mobile scanner
interface.
2
The term "tap" will be maintained, but of course when you're working with the laptop
you've got to "click", 'cause it's not equipped with a touch sensitive screen.
108
6 Graphical User Interface (GUI)
6.10. Setup
In the setup interface – available by tapping on Setup in the menu bar – diverse settings for
calibration or maintenance can be done. Figure 6.35 shows that screen. In the following
subsections each button in the array on the left will be treated. Tap on a button and the
corresponding panel will appear in the middle of the screen.
Figure 6.35:
Setup screen.
6.10.1. Language
Here you can change the language used for all messages and buttons in the whole GUI.
Choose from one language: English as shown in Figure 6.35. Other languages are possible.
6.10.2. Maintenance
Release brake
In this panel you can release the brakes of both scanners by tapping on the corresponding
button.
Attention!
The collector test platform has to be driven in 0°-position, i.e.
horizontal. The mobile Scanner has to be placed in a horizontal position
to be the floor. Otherwise the cross axis respective scanner head may
crash down uncontrollable and generate serious damage at person and
109
6.10 Setup
machinery.
Stop Horn
When starting any test stand movement, which requires pressing down the dead mans push
button on the operation pendant station, a warning signal will sound emitted by a horn. This
contributes to your security!
Nevertheless you're allowed to turn off this signal by tapping on the Stop Horn button.
This panel shown in Figure 6.36 enables changing the sensitivity of the different
measurement instruments on the scanner head.
This is possibly necessary after a recalibration or exchange. To change tap on the desired
text field and enter a new value by means of the pop-up-keypad.
Figure 6.36:
Calibration
panel.
You can lock the screen against operation by third parties as described in section 0. To
unlock, a PIN-input is required. You may change the current PIN to an arbitrary personal one.
Therefore tap on the Change PIN button. First you'll be requested to confirm the actual PIN.
After that the input of a new PIN is possible.
Memorize this PIN well!
110
6 Graphical User Interface (GUI)
6.10.10. Restart
111
6.11 Lock
6.11. Lock
With the Lock function you may protect the touch screen against unwanted intervention of a
third party. If you tap on Lock in the menu bar a dialogue panel will pop up as illustrates.
Only click on YES if you know the correct PIN for unlocking the screen!
Figure 6.37:
Pop-up-panel
when locking
the touch
screen.
112
6 Graphical User Interface (GUI)
113
6.13 Web interface and working with it
114
6 Graphical User Interface (GUI)
115
6.14 File format for Stand Configuration
116
6 Graphical User Interface (GUI)
117
6.15 File format for scan files
FAQ
6.15.1. Why does not the scanner move anymore and a error is shown in the GUI?
Reason 1: Scanner was running to a limit switch. Act as described in section 2.5.7.
Reason 2: Software bug or software breakdown. Save all files and restart the GUI.
6.15.5. Why the irradiance values between pyranometer and solar cell are differing?
Because pyranometer and solar cell have different sensitivities. Moreover they are based on
different functional principles and physical effects. An pyranometer is a wavelength
integrating radiation detector containing a thermopile. However in a solar cell charge carriers
get lifted directly into the conduction band due to the photoelectric effect inducing a voltage
/ current.
Moreover a pyranometer is spectral sensitive over a wider range than a solar cell. The
measurement accuracy of the pyranometer is higher.
118
6 Graphical User Interface (GUI)
6.15.6. Some parts of the X-Y-Scanner motors are getting very hot. Is that normal?
Yes! These parts are the brakes of the step motors. Don’t worry about that. That's normal.
6.15.7. How do I get my scan files (*.scn) from the touch screen to other computers?
Via the web interface (see chapter 6.13) or via LabVIEW (please refer to the LabVIEW
Manual).
6.15.13. May I switch on the electrical cabinets of the different test stand components
in an arbitrary order?
Switch on the SS1 cabinet at the latest 5 minutes after the Atlas cabinet. Otherwise there
will be no connection from the centralized control with the lamps.
119
6.15 File format for scan files
6.15.14. May I move the main or mobile lamp field during lamp operation?
@Main lamp field: Yes.
All movements such as lamp positioning, height adjustment or
inclination is allowed during lamp operation.
@Mobile lamp field: Yes and No.
Electrically height adjustment is allowed during lamp operation.
Manual positioning of the lamps (including loosing and arresting
several screws) is NOT allowed during lamp operation!
Loosing the brakes and driving the lamp field around is NOT allowed
during lamp operation!
120