Uni-Directional and Bi-Directional Ranges: PX2300 Series
Uni-Directional and Bi-Directional Ranges: PX2300 Series
PRESSURE Transducer
UNI-DIRECTIONAL AND BI-DIRECTIONAL RANGES
0-1 to 0-100 psid Uni-Directional
±0.5 to ±50 psid Bi-Directional 77 Dimensions:
mm (inch)
PRESSURE TRANSDUCERS
(3.01)
55
PX2300 Series
(2.15)
DIFFERENTIAL
50
(1.94) 22.2 DIA. Conduit
(0.875) Opening
78
(3.05) Bleed
Screws
25.4
B
(2.44) Low Pressure Port
High Pressure Port
Metric thread
U Wet/Wet Corrosive adaptors
Environments Sensor: Capacitive and snubbers
available, PX2300-1DI,
U Ideal for Measuring Maximum Line Pressure: 250 psig
visit us online. shown smaller
Pressure Drop Maximum Overpressure: than actual size.
Across Filters High Side: 1 to 5 psi = 20 x FS,
10 to 25 psi = 10 x FS, 50 psi = 5 x FS,
100 psi = 2.5 x FS
Low Side: 2.5 x FS (shift recoverable) Wetted Parts: Air and fluids compatible
The zero will shift slightly when high with 17-4 and 300 stainless steel, FKM,
OMEGA’s PX2300 Series and silicone O-rings
differential pressure is applied. The shift
high-output, low differential pressure may be as much as ±10% FS with Cavity Volume: 0.27/0.08" pos./neg. port
transducers are compatible with overpressure applied to the low port. Case: NEMA 4 (IP65)
most media, from dry air to corrosive (Other parameters will not shift.) The Pressure Port: 1⁄4 NPT internal
liquids. All wetted parts are shift may be recovered by a positive
stainless steel with elastomer seals. overpressure or if the overpressure is Electrical Connection: Internal
barrier strip with 22.2 mm (0.875")
The electronics are housed in a always in one direction, the user may
conduit opening
NEMA 4 (IP65) enclosure. A apply this pressure to pre-set the sensor.
Subsequent overloads of less pressure Response Time: 50 ms (water)
high working pressure and high
overpressure ratings ensure will not cause any further shift. Weight: 410 g (14.4 oz)
dependability in harsh industrial
environments. These transducers
are ideal for measuring pressure
To Order
drop across filters and other RANGE
(psid) MODEL NO. COMPATIBLE METERS
process devices.
0 to 1 PX2300-1DI DP41-E, DP25B-E, DP-7700
Specifications 0 to 2 PX2300-2DI DP41-E, DP25B-E, DP-7700
Excitation: 24 Vdc nominal 0 to 5 PX2300-5DI DP41-E, DP25B-E, DP-7700
Max: 30 + 0.004 x 0 to 10 PX2300-10DI DP41-E, DP25B-E, DP-7700
(loop resistance Ω) Vdc 0 to 25 PX2300-25DI DP41-E, DP25B-E, DP-7700
Min: 11 + 0.02 x
(loop resistance Ω) Vdc 0 to 50 PX2300-50DI DP41-E, DP25B-E, DP-7700
Loop Resistance: 0 to 1000 Ω 0 to 100 PX2300-100DI DP41-E, DP25B-E, DP-7700
Output: 4 to 20 mA bi-directional RANGE
Accuracy: ±0.25% RSS FS at ±0.5 PX2300-0.5BDI DP41-E, DP25B-E, DP-7700
constant temperature (includes linearity, ±1 PX2300-1BDI DP41-E, DP25B-E, DP-7700
repeatability and hysteresis)
±2.5 PX2300-2.5BDI DP41-E, DP25B-E, DP-7700
Linearity: ±0.20% FS
Hysteresis: 0.10% FS ±5 PX2300-5BDI DP41-E, DP25B-E, DP-7700
Repeatability: ±0.05% FS ±10 PX2300-10BDI DP41-E, DP25B-E, DP-7700
Operating Temperature Range: ±25 PX2300-25BDI DP41-E, DP25B-E, DP-7700
-18 to 80°C (0 to 176°F) ±50 PX2300-50BDI DP41-E, DP25B-E, DP-7700
Compensated Temperature Range: Comes complete with operator’s manual.
-1 to 65°C (30 to 149°F) Ordering Examples: PX2300-5BDI, bi-directional range -5 to 5 psid transducer with
Thermal Zero Effect: <±0.02% FS/°F current output.
Thermal Span Effect: <±0.02% FS/°F PX2300-1DI, 0 to 1 psid range transducer with current output.
B-242