C1.9 PLUS Ultrasound System: Service Manual
C1.9 PLUS Ultrasound System: Service Manual
9 PLUS
Ultrasound System
Service Manual
P02287-03 01/2003
Copyright 2003 by SonoSite, Inc.
All rights reserved. Printed in the USA.
CAUTION:
` United States federal law restricts this device to sale by or on the order of a
physician.
“SiteCharge,” “SitePack,” “SiteStand,” and “SonoHeart,” are trademarks of SonoSite, Inc.
SonoSite products may be covered by one or more of the following U.S. patents: 4454884, 4462408,
4469106, 4474184, 4475376, 4515017, 4534357, 4542653, 4543960, 4552607, 4561807, 4566035,
4567895, 4581636, 4591355, 4603702, (4607642), 4644795, 4670339, 4773140, 4817618,
4883059, 4887306, 5016641, 5050610, 5095910, 5099847, 5123415, 5158088, 5197477, 5207225,
5215094, 5226420, 5226422, 5233994, 5255682, (5275167), 5287753, 5305756, 5353354,
5365929, 5381795, 5386830, 5390674, 5402793, (5,423,220), 5438994, 5450851, 5456257,
5471989, 5471990, 5474073, 5476097, 5479930, 5482045, 5482047, 5485842, 5492134, 5517994,
5529070, 5546946, 5555887, 5603323, 5606972, 5617863, (5634465), 5634466, 5636631,
5645066, 5648942, 5669385, (5706819), 5715823, 5718229, 5720291, 5722412, 5752517,
5762067, 5782769, 5800356, 5817024, 5833613, 5846200, 5860924, 5893363, 5916168, 5951478,
6036643, 6102863, 6104126, 6113547, 6117085, 6142946, 6203498 B1, D0280762, D0285484,
D0286325, D0300241, D0306343, D0328095, D0369307, D0379231. Other patents pending.
iii
iv C1.9 PLUS Ultrasound System Service Manual
Table of Contents
CHAPTER 1 Introduction 1
CHAPTER 2 Safety 5
v
3.5.5 Image Storage...................................................................18
3.5.6 Temperature, Pressure, and Humidity Limits ..................18
3.5.6.1 System Operating .........................................................18
3.5.6.2 System Shipping/Storage ............................................18
3.5.6.3 Battery Operating .........................................................18
3.5.6.4 Battery Shipping/Storage .............................................18
3.5.6.5 Transducers Operating .................................................18
3.5.6.6 Transducers Shipping/Storage .....................................19
3.5.7 Electrical...........................................................................19
3.6 Battery.....................................................................................19
3.7 Safety Requirements ...............................................................19
3.7.1 Meets Electromechanical Safety Standards .....................19
3.7.2 Meets EMC/EMI Standards .............................................20
3.7.3 Meets Airborne Equipment Standards (without ECG Cable
Attached) ..........................................................................20
3.7.4 Meets ECG Standard ........................................................20
CHAPTER 6 Troubleshooting 33
vii
CHAPTER 8 Performance Tests 65
8.1 Overview.................................................................................65
8.2 Test Equipment: ......................................................................65
8.3 Setting Up Performance Tests ................................................66
8.3.1 Scan Reference Orientation..............................................66
8.4 Testing 2D Performance .........................................................66
8.4.1 2D Image Quality .............................................................66
8.4.2 Axial Measurement Accuracy ..........................................67
8.4.3 Lateral Measurement Accuracy .......................................67
8.4.4 Penetration........................................................................68
8.5 Additional Performance Tests ................................................68
8.5.1 CPD ..................................................................................68
8.5.2 PowerMap DCPD.............................................................69
8.5.3 M-mode Imaging..............................................................69
8.5.4 Tissue Harmonic Imaging ................................................70
8.5.5 Pulsed Wave (PW) Doppler Imaging...............................70
8.5.6 ECG Monitoring...............................................................70
8.5.7 Image Quality Verification Test.......................................71
8.5.8 Image Review...................................................................71
8.5.9 Printer ...............................................................................71
8.5.10 Battery Charging ..............................................................71
8.5.11 Video Output ...................................................................71
8.6 Returning Products to SonoSite ..............................................72
8.6.1 Contacting SonoSite Technical Support ..........................72
8.6.2 Shipping Instructions........................................................72
Index 87
Chapter 1: Introduction 1
System accessories include a SiteStand® mobile docking station and accessories,
SiteCharge™ dual battery charger, a power adapter, a battery, ECG cable, video and
printer cables, SiteLink image manager software, SiteStand® basket, Basic Stand,
CRT Stand, ScanPack quick access carrier, and SitePack™ protective carry pack.
System peripherals include medical grade (conforming to the requirements of
EN60601-1) and non-medical (commercial) grade products. System medical grade
peripherals include an external color monitor, video printers, and VCRs. System
non-medical grade peripherals include a digital video recorder, a battery charger, a
lithium-ion battery, printer, and a handheld monitor. Use of peripherals is covered in
the manufacturers’ instructions, which accompany each peripheral. System setup
instructions for the use of peripherals are covered in Chapter 3 in the SonoSite
Ultasound System User Guide.
1.2 Audience
The intended audience of this manual is properly trained field and in-house service
personnel.
Chapter 1: Introduction 3
This page intentionally left blank.
Please read this information before using the SonoSite ultrasound system. It applies
to the ultrasound system, transducers, accessories, and peripherals.
A WARNING describes precautions necessary to prevent injury or loss of life.
A CAUTION describes precautions necessary to protect the products.
Chapter 2: Safety 5
` To avoid the risk of electrical shock or injury, do not open the system enclosures.
All internal adjustments and replacements, except battery replacement, must be
made by a qualified technician.
` To avoid the risk of injury, do not operate the system in the presence of flammable
gasses or anesthetics. Explosion can result.
` To avoid the risk of electrical shock, use only properly grounded equipment.
Shock hazards exist if the AC power adapter is not properly grounded. Grounding
reliability can only be achieved when equipment is connected to a receptacle
marked “Hospital Only” or “Hospital Grade” or the equivalent. The grounding
wire must not be removed or defeated.
` To avoid the risk of electrical shock, before using the transducer, inspect the
transducer face, housing, and cable. Do not use the transducer, if the transducer or
cable is damaged.
` To avoid the risk of electrical shock, always disconnect the AC power adapter
from the system before cleaning the system.
` To avoid the risk of electrical shock, do not use any transducer that has been
immersed beyond the specified cleaning or disinfection level. Refer to the
SonoSite Ultrasound System User Guide for cleaning and disinfection levels.
` To avoid the risk of electrical shock and fire hazard, inspect the AC power adapter
cord and plug on a regular basis. Ensure they are not damaged.
` To avoid the risk of electrical shock, use only accessories and peripherals
recommended by SonoSite. Connection of accessories and peripherals not
recommended by SonoSite could result in electrical shock. Contact SonoSite or
your local representative for a list of accessories and peripherals available from or
recommend by SonoSite.
` To avoid the risk of electrical shock, use commercial grade peripherals
recommended by SonoSite on battery power only. Do not connect these products
to AC mains power when using the system to scan or diagnose a patient/subject.
Contact SonoSite or your local representative for a list of the commercial grade
peripherals available from or recommended by SonoSite.
` To prevent injury, only use market cleared ECG electrodes and cables.
` To avoid the risk of electrical shock to the patient/subject, ensure proper assembly
of the ECG electrodes and cables.
` To avoid the risk of electrical shock to the patient/subject, do not touch the system
battery contacts while simultaneously touching a patient/subject.
` To prevent injury to the operator/bystander, the transducer must be removed from
patient contact before the application of a high-voltage defibrillation pulse.
CAUTIONS:
` Although your system has been manufactured in compliance with existing EMC/
EMI requirements (EN60601-1-2), use of the system in the presence of
an electromagnetic field can cause degradation of the ultrasound image. If this
occurs often, SonoSite suggests a review of the system environment. Identify and
remove the possible sources of the emissions or move your system.
Chapter 2: Safety 7
2.2 Battery Safety
To avoid the risk of injury, follow the warnings and cautions to make sure that the
battery does not burst, ignite, or generate heat or fumes.
WARNINGS:
` The battery has a safety device. Do not disassemble or alter the battery.
` Charge the batteries only when the ambient temperature is between 0° and 40°C
(32° and 104°F).
` Do not short-circuit the battery by directly connecting the positive and negative
terminals with metal objects.
` Do not heat the battery or discard it in a fire.
` Do not expose the battery to temperatures over 60°C (140°F). Keep it away from
fire and other heat sources.
` Do not charge the battery near a heat source, such as a fire or heater.
` Do not leave the battery in direct sunlight.
` Recharge the battery only with the SiteCharge™ dual battery charger or the
system.
` Do not pierce the battery with a sharp object, hit it, or step on it.
` Do not use a damaged battery.
` Do not solder a battery.
` When connecting the battery to the SiteCharge™ dual battery charger or to the
system, never reverse the polarity of the battery terminals.
` The polarity of the battery terminals are fixed and cannot be switched or reversed.
Do not force the battery into the system or the SiteCharge™ dual battery charger.
` Do not connect the battery to an electrical power outlet.
` Do not continue recharging the battery if it does not recharge after two successive
six hour charging cycles.
CAUTIONS:
` To avoid the battery bursting, igniting, or fumes from the battery and causing
equipment damage, observe the following precautions:
` Do not immerse the battery in water or allow it to get wet.
` Do not put the battery into a microwave oven or pressurized container.
` If the battery leaks or emits an odor, remove it from all possible flammable
sources.
` If the battery emits an odor or heat, is deformed or discolored, or in any way
appears abnormal during use, recharging or storage, immediately remove it and
stop using it. If you have any questions about the battery, consult SonoSite or your
local representative.
` Store the battery between -20°C (-4°F) and 60°C (140°F).
` Use only SonoSite batteries.
Chapter 2: Safety 9
This page intentionally left blank.
Processing Display
subsystem subsystem Video bus
User
interface
subsystem
Transducer
Power
Display power subsystem Battery
To all pack
subsystems
Logic power
Pulser voltage
Power
adapter
External power
3.2 Components
The SonoSite system components are identified in Chapter 1, section 1.2.
10
2
rear view 11
3
12
4 13
6 14
7 15
8 16
17
1 2 3 4
NUMBER FEATURE
1 DC line voltage connector
2 I/O connector
3 Print control connector for a recommended printer
4 Video out connector for a recommended VCR, printer, or external video monitor
Power Cord
Table 3.4 Power Cord Specifications
FEMALE
CONFIGU- APPRO
LENGTH RATING MALE PLUG CONNEC- MARKING
RATION -VALS
TOR
100-120 V ac 9 ft. 10 in. 250 V ac MA 5-15P CEE-22, UL, Manufac-
ACC/ 60 Hz Hospital grade, molded on CSA turer,
grounding type Agency
molded on Approvals
230 V ac /50 Hz / 3 m 250 V ac CEE-7/VII CEE-22, EU Manufac-
PAL grounding type molded on turer,
230 V ac / 60 Hz with 4.8 mm pins Agency
molded on Approvals
3.5.2 Monitor
Height: 3.1 in. (7.9 cm)
Width: 4.3 in. (10.9 cm)
Diagonal: 5 in. (12.7 cm)
Brightness control
Contrast control
3.5.7 Electrical
• System optional: 100-120/220-240 V ac, 50/60 Hz input, 16.0 V dc output
power adapter
• SiteCharge dual battery charger input voltage: 16.0 V dc, 2.8 A
• SiteCharge dual battery charger output voltage: 12.6 V dc, 3.0 A (2x)
• AC power adapter input: 100-120/220-240 V ac, 50/60 Hz, 1.0-0.50 A
• AC power adapter output: + 16.0 V dc, 2.8 A
3.6 Battery
• 6-cell, 11.1 V dc, 3.0 amp-hours, rechargeable, lithium-ion battery pack
• Run time: 1.5 to 4 hours, depending upon operating conditions
WARNING:
` CRITICAL TEST FUNCTION - Skipping the functional tests in this procedure
could adversely affect safety or effectiveness of the system.
Operation of the SonoSite ultrasound system can be found in the user guide provided
with each system.
To turn on power:
1. Locate the Power switch on the back of the left side of the system handle. See
Figure 3.2, SonoSite Ultrasound System Controls, on page 14.
2. Press and hold the Power switch until the system beeps or until you see the
image display.
3. Release the Power switch.
1 2 3 4
NUMBER FEATURE
1 DC line voltage connector
2 I/O connector
3 Print control connector for a recommended printer
4 Video out connector for a recommended VCR, printer, or external video
monitor
1 2 3 4
NUMBER FEATURE
1 DC line voltage connector
2 I/O connector
3 Print control connector for a recommended printer
4 Video out connector for a recommended VCR, printer, or external video monitor
Chapter 6: Troubleshooting 33
6.4 Failures
6.4.1 Display
Attach an external monitor to the external video connector to verify display failures.
For example, if the system display is blank and the external monitor works properly,
the system display requires servicing.
6.4.3 Trackball
Intermittent function or loss of control indicates trackball failures. Clean the
trackball by removing the retainer ring and then removing the ball.
Assert code
6.4.6 Battery
If the system does not operate or does not run for the expected duration for a given
charge, battery failure is likely.
Chapter 6: Troubleshooting 35
If System
Display Is
Blank, Attach
An External
Monitor
Ext. Replace
Monitor No Main PCBA Corrected?
OK?
Yes
No
Yes
No
Replace
Display
Return
System to
SonoSite
Replace Perform
Corrected? No Control Panel Corrected? Image Quality
Tests
Yes Yes
Perform Perform
Image Quality Image Quality
Tests Tests
No
No No No
Replace
Correct
No External
Monitor?
Replace Connect Monitor
Corrected?
Main PCBA Cable
Yes
Yes
Replace
Stop Cable
No
Replace
No Main PCBA
Return Return
System to Monitor to No Corrected? Yes Stop
SonoSite SonoSite
Chapter 6: Troubleshooting 37
LED Burned Out?
Key Not
Working?
Replace
Control Panel
Replace
Corrected? No Main PCBA
Yes
Return
Corrected? No System to
SonoSite
Stop
Yes
Perform
Image Quality
Tests
Keybd. Clean
Yes Corrected? Yes Stop
OK? Trackball
No
Replace
Corrected? Yes Stop
Trackball
Replace
No Control No
Panel
No
Replace
Corrected? Yes Stop
Main PCB
No
Return
System to
SonoSite
Chapter 6: Troubleshooting 39
No Indication Of
Listen For Two Beeps If System Works On
Power When You
And Look For A Battery, Verify
Press The ON
Video Display. Operation On AC
Button?
Power
Try Battery
Power Only
First Remove
Battery And
Make Sure Then Connect
Battery Is AC Power
Fully Inserted
No Problem
Try AC Working? Yes
Working? Yes Found
Power
No
No
Replace
Charge or AC
Battery
No Replace Adapter
Charged?
Battery
Yes
No
No
Replace
Replace Main PCB
Main PCB
Perform Perform
Image Quality Image Quality
Tests And Tests And
Corrected? Yes Corrected? Yes
Verify AC Verify AC
Power Power
Operation Operation
No
Return
No System To
SonoSite
No
First Remove
Replace
Is Battery Battery And
Yes Working? Yes Battery and
Fully Inserted? Then Connect
Try Again
AC Power
No No
Insert
Battery and Go To System
Try Power Diagram
Again
No Problem
Working? Yes
Found
No
Charge or
Replace
Battery
Problem
Working? Yes
Solved
No
Go To System
Diagram
Chapter 6: Troubleshooting 41
This page intentionally left blank.
Screws (2)
Battery
compartment
3. Insert an anti-static mat between the Control Panel Subassembly and the Display
Subassembly to prevent damaging the display LCD (Figure 7.2).
4. Lay the Control Panel Subassembly over onto the Display Subassembly to
expose the wire harness and flex circuits.
5. Carefully disconnect the 100-pin flex circuit from the bottom module.
6. Disconnect the Display Subassembly Wire Harness and unlatch two flex circuits
connectors from the Control Panel Subassembly (Figure 7.2).
7. Set the Control Panel Subassembly aside.
Anti-static mat
Figure 7.2 Disconnecting the Display Wire Harness and Flex Circuits
Display hinges
Display Wire
Harness
40-pin flex
circuits (2)
Display
100-pin flex
circuit
Display Wire
Harness
Control Panel
Trackball
Anti-static mat
Figure 7.5 Placing the Control Panel Subassembly onto the Display
6. Carefully replace the Control Panel Subassembly by inserting the tabs on the top
of the Control Panel Subassembly into the slots.
7. Set the Control Panel into place and replace the two screws inside the battery
compartment (Figure 7.6). Torque screws to 7.1 in./lb. (0.8 newton meter).
8. Place the battery in the battery compartment.
9. Turn on the system.
10. Perform the Display tests in Chapter 8, Performance Tests, on page 65 to
verify that the Display Subassembly is functioning properly.
Screws (2)
Battery
compartment
3. Insert an anti-static mat between the Control Panel Subassembly and the Display
Subassembly to prevent damaging the Display LCD.
4. Carefully lay the Control Panel Subassembly over onto the display to expose the
display wire harness and the two 40-pin flex circuits.
5. Carefully disconnect the 100-pin flex circuit from the bottom module.
6. Disconnect the display wire harness and the two 40-pin flex circuits from the
Control Panel Subassembly (Figure 7.7).
7. Remove the Control Panel Subassembly.
Anti-static mat
Display Wire
Harness
Screws (2)
Battery
compartment
Foam backing
Screws (4)
13. Holding the Trackball in place, carefully turn the Control Panel Subassembly
over and allow the Trackball to drop off of the Control Panel Subassembly, to
prevent losing the Trackball parts (Figure 7.12).
14. Remove the Trackball and note the location of its components.
Trackball
Trackball Wire
Harness
Screws (4)
Foam backing
7. Turn the Control Panel Subassembly over and secure the trackball wire harness
to the Control Panel Subassembly.
8. Place the Control Panel Subassembly onto the display LCD.
9. Install the display wire harness and the two 40-pin flex circuits (Figure 7.16).
10. Replace the 100-pin flex circuit on the bottom module.
100-pin flex
circuit
30-pin flex
circuits (2)
Figure 7.16 Installing the Display Wire Harness and Flex Circuits
11. Carefully replace the Control Panel Subassembly by inserting the tabs on the top
of the Control Panel Subassembly into the slots.
Screws (2)
Battery
compartment
Screws (8)
Figure 7.18 Removing the Bottom Housing from the Top Housing
4. Carefully separate the bottom housing from the top housing; do not damage the
100-pin flex circuit that connects the two.
5. Disconnect the 100-pin flex circuit from the Main PCBA Subassembly.
6. Remove the bottom housing that holds the Main PCBA Subassembly.
Screws (4)
40-pin flex
Figure 7.19 Removing the Main PCBA Subassembly from the Bottom Housing
8. Carefully disconnect the 40-pin flex circuit that connects the Main PCBA
Subassembly to the Dock Interface PCBA (located under the Main PCBA).
9. Remove the Main PCBA Subassembly.
40 Pin flex
2. Secure the four screws that connect the Main PCBA Subassembly to the bottom
housing (Figure 7.21). Torque the screws to 3.6 in./lb. (0.4 newton meter).
Screws (4)
100-pin flex
circuit
Top housing
Figure 7.22 Setting the Top Housing Onto the Bottom Housing
Screws (8)
7.5 Transducers
The transducers have no user-serviceable parts; return transducers
to SonoSite for exchange.
8.1 Overview
WARNING:
` Critical Test Function - A failure of the system functions tested in this section
could affect safety or effectiveness of the system adversely. While performing the
steps in this section, verify that the images on the system display and on the
external monitor are acceptable.
• Verify that all controls operate smoothly over their full range and that the system
responds properly.
• To obtain 2D images, SonoSite recommends using the RMI 413A Soft Tissue
Phantom or, the RMI 403 GS Multipurpose Phantom. Any equivalent Phantom
is acceptable.
• To obtain Power Doppler images, SonoSite recommends using the RMI 425
Doppler Phantom or the RMI 1425A Doppler Phantom. Any equivalent
Phantom is acceptable.
• When making penetration measurements on a phantom, apply the phantom
reference value and tolerance to the measurement.
MEASUREMENTS TOLERANCE
Axial Distance +/- 2%
Lateral Distance +/- 2%
IMAGING
C60 ICT ICT HF C15 L38 C11
PERFORMANCE
2D Penetration 11.5 cm 6.0 cm 5.0 cm 19.0 cm 6.0 cm 5.0 cm
8.5.9 Printer
To test printer operation:
1. Print two images in rapid succession and verify proper operation.
2. Verify that the print control receptacles on the system and on the stand function
correctly.
5
1 8 9 7 6
FIND PART
DESCRIPTION
NUMBER NUMBER
Note: One of the following labels must be ordered for the display.
9 P02362 LCD 5”
FIND PART
DESCRIPTION
NUMBER NUMBER
FIND PART
DESCRIPTION
NUMBER NUMBER
3
4
6
7
FIND PART
DESCRIPTION
NUMBER NUMBER
FIND PART
DESCRIPTION
NUMBER NUMBER
1
2 3
8 9
10
11
INDEX PART
DESCRIPTION
NUMBER NUMBER
3
7
4 8
9
6
5 12 11 10
INDEX PART
DESCRIPTION
NUMBER NUMBER
2 P00361 Foot
INDEX PART
DESCRIPTION
NUMBER NUMBER
The Service Event Report provides information about product failures to the
manufacturer and to authorized service facilities, which provide approved warranty
services for SonoSite products. For all repairs completed, complete the form and
return a copy of it to the following address:
SonoSite, Inc.
Technical Support
21919 30th Drive SE
Bothell, Washington 98021-3904
facsimile: +425-951-6700
e-mail: [email protected]
website: www.sonosite.com
Select Products, then Technical Support.
Device Description
Name: Serial Number:
Part Number: Lot Number: Revision:
Software Version: Other Identifiers:
Event Description
Diagnosis
Service Performed
Performed By: Date:
Actions:
Parts Removed
Part Name Part Number Serial Number Lot Number Rev Replaced By
Parts Installed
Part Name Part Number Serial Number Lot Number Rev Replaced By
C G
C11/7-4 transducer 1 grace period 3
C15/4-2 transducer 1
C60/5-2 transducer 1
cable 2
H
cleaning and disinfecting 31–32 humidity limits 18
precautions 31
Color Power Doppler
adjusting gain 68, 69
I
contaminated materials ICT/7-4 transducer 1
receipt of 31 image
control panel assembly 48 quality verification test 71
removing 49 review 71
replacing 51 storage 18
required parts 48
required tools 48
87
imaging phantoms
modes specified 18 PFS-A-3-1-G2 UHDC Flow System
introduction 1–26 (Doppler) 65
RMI 1425A Doppler 65
RMI 403 GS Multipurpose 65
K RMI 413A Soft Tissue 65
RMI 425 Doppler 65
keyboard
physical dimensions 17
diagnosis figure 38
power
battery charge indicators 15
L battery pack 15
connections 16
L38/10-5 MHz 38-mm 18 external 16
L38/10-5 transducer 1 line cord specifications 17, 29
labeling subsystem 13
symbols 9 supply 17
license key 3 turning off 23
installing 26 turning on 22
obtaining 26 power adapter 2
license update screen PowerMap DCPD
displaying 28 imaging 1
printers
test 71
M processing subsystem 12
main PCBA 57–62 product failures
failures 34 reporting 85
removing 58
products
replacing 59
returning 72
required parts 57
required tools 58
monitor 17 R
external diagnosis 37
specifications 17 replacement parts
list 73
moving image 68, 69 ordering 73
replacement procedures 43–63
P return material authorization number (RMA) 72
parts list 73–82
additional spare parts 82 S
control panel assembly 75
display assembly 74 safety 5–9
main PCBA 77 battery 8
trackball assembly 76 biological 9
performance tests 65–72 electrical 5
overview 65 equipment protection 7
peripheral output 71 labeling symbols 9
printer 71 service event report 85
required test equipment 65 form 86
video output 71 setup and operation 5, 21
peripherals 17, 63 shipping
medical grade 2 instructions 72
non-medical grade 2
shipping specifications 18
printer 63
VCR 63 SiteCharge dual battery charger 2, 62
SiteLink Image Management software 2
SitePack carrying case 2
SiteStand mobile docking station 2
89
90 C1.9 PLUS Ultrasound System Service Manual