S T K SQ1: Set Up and User Guide
S T K SQ1: Set Up and User Guide
Set Up And
User Guide
Distributed by
Warning
Warranty
4
Additional Equipment
Requirements
To enable you to complete the steam tests, additional
equipment is required.
5
Steam Test Points
In order to test the steam quality specific test points on
the steam line are required.
Fig. 1 Illustrates the location of the three test points on
the steam supply pipe and they are fitted between the
steam distribution system and the sterilizer. It is expected
that the pressure at this point would be 2-5BarA.
Non-condensable
gas Take Off
Min 400 mm
Steam
Temperature
Probe entry Point
Fig. 1 To
Sterilizer
Note All entry points are 1/4 NPT female sockets with full
size holes drilled through the pipes.
6
Assembling the Steam Test
Kit
NON-CONDENSABLE GAS KIT
1. Burette Bulb
2. Burette Cock
3. Burette
4. Burette Guide
5. Condense Collection Cylinder
6. 250 ml Measuring Cylinder
7. Steam Valve
8. Cooling Water Valve
1
9. Steam Condensing Unit
10. Cooling Water Inlet 2
11. Cooling Water Outlet
12. Steam inlet
3
5
6
Fig 2
7 8
7
SUPERHEAT KIT
3
Fig 4
1. Pitot Tube
2. Tubing
3. Tube Clamps
4. Rubber Bung and Tube Assembly
5. 1 Litre Flask
Fig 5
5
8
The Coolant Transfer Unit
To condense the steam in the non-condensable gas testing unit, a
supply of cold water is needed. Due to the difficulty that is often found
in finding an accessible supply of water in a sterilizer plant room, a
coolant transfer unit has been included in this kit.
The unit is a 110 volt, 60 Hz, alternating current pump that needs to be
totally submerged in water.
9
Non-condensable Gas Test
S E T T I N G U P
P E R F O R M I N G T H E T E S T
1. Before starting the test ensure the cooling water valve is fully open
and the steam valve is closed.
2. Turn on the water supply by supplying power to the pump. After
ensuring that the cooling water is leaving the outlet pipe, the main
steam valve may be slowly opened. Live steam and boiling water
could discharge from the condense collecting chamber if
insufficient cooling water is available and present a hazard. In any
event, precautions should be taken. Do not look into the collecting
chamber and wear eye protection.
10
3. Slowly open the steam valve and by reducing or increasing
the flow through the steam and water valves, obtain a flow of
condense that will give a temperature of between 70o C and
90o C, as indicated on the dial temperature gauge.
4. Fill the condense collection chamber with water either by
filling with water from an external source, or allowing
condense to accumulate.
5. Open the burette cock and draw up condense into the burette
with the rubber bulb provided, to get a water level near the
top. Remember to isolate the burette from the rubber bulb by
shutting the burette cock before testing commences.
6. Fill the condense collection chamber with more water, until it
overflows.
7. Ensure that the sterilizer chamber is empty except the normal
furniture etc. Select a porous load/equipment cycle and start a
run.
8. When the steam supply to chamber first opens, ensure the
measuring cylinder is empty, by emptying it.
9. Make a note of the water level in the burette.
10. Any non-condensable gases present in the steam being
sampled will rise to the top of the burette. The overflow
formed by the condensate and the water displaced by the
gases, will collect in the measuring cylinder.
11. When the at least 100ml of condense has been collected in
the measuring cylinder note the volume of gas collected in the
burette (Vb) and the volume of water collected in the
measuring cylinder (Vc).
12. Calculate the fraction of non-condensable gases as a
percentage as follows:
Percentage of non-condensable gas = 100 x (Vb/Vc)
A C C E P T A N C E C R I T E R I A
The test should be done three times in total to check consistency. If the
results of the tests differ significantly, then the cause should be
investigated before proceeding further.
11
The Superheat Test
S E T T I N G U P
12
Dryness Test
S E T T I N G U P
P E R F O R M I N G T H E T E S T
Important
Before performing the test ensure the pitot hole is not restricted as
this will lead to false results. This is best done prior to fitment for
safety reasons. A long heat up time for the flask reaching ~80oC
is not normal. It should take no more than 5 minutes.
1. Weigh the whole assembly including pipe and clips and note the
weight (M1).
2. Remove the stopper and tube assembly and pour 650 +/- 50 ml of
cold water (below 27o C) into the flask. Replace the stopper and
tube assembly, weigh the flask and record the mass (M2).
3. Support the flask close to the pitot tube taking care to avoid the
issuing steam. Ensure that the rubber tube and flask are protected
from excess heat and draughts. Do not connect it to the pitot tube
yet.
4. Introduce the second temperature sensor through the shorter of
the two pipes into the water in the flask. Agitate the flask and note
the temperature of the water in the flask (T0).
5. Ensure the sterilizer is empty except for the usual chamber
furniture. Select and start a porous load/equipment cycle.
6. When the steam supply valve to the chamber first opens, connect
the rubber tube to the pitot tube. This will require the tester to be in
close proximity to the steam issuing from the pitot tube and
extreme care is required to avoid scalding and/or burning. Gloves,
overalls and eye protection must be worn.
13
7. 0 Observe/record the steam temperature for the duration of the test
0 and on completion of the test calculate the average temperature
0 (TS).
8. 0 When the temperature in the flask is approximately 80o C,
disconnect the rubber tube from the stainless steel tube taking the
same precautions as when fitting. Agitate the water in the flask to
make sure it is thoroughly mixed. Note the temperature of the
water (T1).
9. 0 Weigh the flask and stopper assembly including pipe and clips and
note the mass (M3).
10. Calculate the dryness value by using the following formula.
0
Where:
D = Dryness value
MW = The initial mass of water in the flask is M2 - M1
MC = The mass of condense collected is M3 - M2
T0 = Initial temperature of the water in the flask (o C)
T1 = Final temperature of the water and condense in the flask
TS = Average temperature of the steam delivered to the flask
L = Latent heat of dry saturated steam at temperature TS (KJ Kg-1) - see
Appendix 1 for latent heat tables.
Note. The formula has been modified for stainless steel flasks
and dip tubes instead of glass, as detailed in HTM 2010.
A C C E P T A N C E C R I T E R I A
14
Health and Safety
Care should always be taken when working on or near steam pipes due
to the very high temperatures involved. Thermal gloves, overalls that
cover arms and eye protection must be used.
While every effort has been made to provide protection from the ex-
treme temperatures, care should be taken when testing with this equip-
ment.
Do not plug or restrict the “Cooling Out” flow as this may pressurize the
heat exchanger, which it has not been designed for.
Do not look into the condense collection cylinder while the steam valve
is open, as hot condense and/or steam could be ejected.
The insulation surrounding the steam pipe is made from glass fibre and
can cause irritation to the skin in some instances. It is therefore
recommended that gloves or a good quality barrier cream be applied as
a precaution before handling the material. The insulation is designed to
prevent burns from transient contact with the pipe and is not designed
to protect against prolonged contact.
The submersible pump is electrically powered and care must be taken
to prevent hazards arising from the use of electricity in a potentially wet
environment.
Care and maintenance
No special care is required for the steam test kit. However, the 4 mm
copper tubing, will with bending, become work hardened, which will
lead to difficulty in returning it to the tight radius required for storage. In
this instance, it may be easily replaced.
Should difficulty be found in obtaining a replacement or any other
components, please contact your local Distributor in the first instance or
Keith Shuttleworth & Associates Ltd for replacements.
Equipment specifications: -
Non-condensable gas test kit 0-20% N/C per 100ml condensed
steam
Dryness test Full range
Superheat test Full range
15
Appendix 1 Steam Tables