WORKSHEET Phy lab-II
WORKSHEET Phy lab-II
AIM OF THE
EXPERIMENT–
Calculate the refrative Index of the Prism for various wavelengths of the
Mercury Spectrum and then plot a Dispersion and Calibration Curves
using a Prism Spectrometer.
APPARATUS-
Sprectometer
Prism
Spirit level
Theory of experiment-
sinA
2
• First the telescope has to be focussed distant objects i.e infinity and
this has to be maintained until the experiment is over, so as not to
refocus again. Then, the cross-wires should be focussed by moving the
eye-piece of the telescope.
• Adjust the Collimater such that the image seen in the telescope is
sharp of the slit without the prism.
• Measuring the Angle of PrismA:Place the prism on the Prism Table and
lock the prism table in the position so the the incident beam falls on one
of the edges of the prism. Now, move the telescope and locate the
images of the slit and note down the angles. The difference beteen both
the angles is 2A. Hence, half of the diffece will give us A.
• Now, choose an angle of incidence other than the preious chosen one
and with eye locate aprroximately the angle at which the spectrum starts
to move in the opposite direction as the prism table is rotated, and lock
the prism table. Now, using the telescope, fix the telescope on one of the
sprectrum lines, and then use the fine adjustment for the movement of
prism tabe to move the table so that we get the precise location of the
angle where the line starts to move in the opposite direction, and note
the angle for this.
• Without disturbing anything, remove the prism and get the measure of
the angle of the direct image of the slit in the telescope. The difference
between these two angles is the Angle of Minimum Deviation δ m for this
spectral line λ. Repeat the same for all the spectral lines that are given by
the mercury lamp.
• From above data we can calculate the refractive index n of the prism
for various wavelengths. For the Calibration Curve, plot a graph of δ m
versus λ. For the Dispersion Curve, plot a graph of n versus λ.
OBSERVATIONS-
Measurement for RED light
• R3-the image of the slit without the prism on the prism table
Angle of minimum deviation Dm = R1 ∼ R3
−
• The behaviour of the Dispersion curve can be seen that the fall is
not rapid over these range of wavlengths, hence, it is not a very
heavily sloping line which implies that the dispersion of various
spectral line donot vary a lot from each other i.e which is
manifested by the closeness of the refractive index for the range of
wavelengths.It can be observed that the curve is roughly parabolic
in nature.The dispersion curve is as follows:
Now, if µ be the refractive index of the medium, then by Cauchy’s for-mula,
µ = a + λb2
Now, we take two arbitray readings, say for RED and YELLOW light. Then, we put
the values of µ and λ and get two simultaneous equation. We solve for a and b.
The values are :
SOURCES OF ERROR-
1. The telescope and collimator should be individually set for parallel rays.
4. The prism should be properly placed on the prism table for the measurement of angle of the
prism as well as for the angle of minimum deviation.
•It must be ensured that the light rays coming out of collimator are parallel.
Hence, the collimator must be focussed properly before the experiment.