Technical Tip: Overview of A Parametric Release Cycle For Eo Sterilization
Technical Tip: Overview of A Parametric Release Cycle For Eo Sterilization
OV E RV I E W O F A PA RA M E T R I C R E L E AS E CYC L E
F O R E O S T E R I L I Z AT I O N
F O R M O R E I N F O R M AT I O N
STERIS Applied Sterilization Technologies
Web: www.steris-ast.com // Email: [email protected]
(EMEAA) +44 (0) 8456 88 99 70
(Americas) 877.783.7479
OV E RV I E W O F A PA RA M E T R I C R E L E AS E CYC L E
F O R E O S T E R I L I Z AT I O N
• Temperature and pressure in the chamber throughout the Simulation of Winter Conditions
sterilization cycle
To assure that preconditioning (if used) and/or conditioning is
• If pressure is used as the primary control measure, the effective regardless of climatic conditions or facility location, each
requirement for the secondary measure is only to confirm half cycle validation load is exposed to simulated winter conditions
admission of EO to the chamber by at least one of the prior to half cycle processing. Successful biological results when
following: the mass of EO used; the direct measurement of the challenged with this worst case load conditioning assures that the
concentration of EO in the sterilization chamber; volume of EO process is effective for all conditions regardless of the time of the
used year or the location of the facility.
• Inert gas injection (if used) Control of loading patterns and densities is required when
practicing parametric release. For those customers that present
• Exposure time
mixed loads to the sterilizing facility, a fourth half cycle is required
• Time taken to evacuate the chamber using the minimum loading configuration at minimum density.
This load is processed in a biological challenged half cycle using
• Time and pressure changes during post exposure flushing
minimum cycle conditions. Successful biological results using this
Aeration Phase minimum loading condition when supported by the successful
maximum loading half cycles, serves to validate the process for
• Time, temperature pressure changes (if any) during aeration
any load mix from the minimum loading qualified to the maximum
In addition, the following physical parameters must be included as loading qualified.
part of the parametric release criteria as directed in section 10.5 of
the ANSI/AAMI/ISO 11135 document:
F O R M O R E I N F O R M AT I O N
STERIS Applied Sterilization Technologies
Web: www.steris-ast.com // Email: [email protected]
(EMEAA) +44 (0) 8456 88 99 70
(Americas) 877.783.7479
OV E RV I E W O F A PA RA M E T R I C R E L E AS E CYC L E
F O R E O S T E R I L I Z AT I O N
REFERENCES
2 . ANSI/AAMI/ISO 11135, Medical devices – Validation and routine control of ethylene oxide sterilization, 2014.
F O R M O R E I N F O R M AT I O N
STERIS Applied Sterilization Technologies
Web: www.steris-ast.com // Email: [email protected]
(EMEAA) +44 (0) 8456 88 99 70
(Americas) 877.783.7479