Synapse 3D V4.1US: Product Specifications
Synapse 3D V4.1US: Product Specifications
Synapse 3D V4.1US
Base Tools
Cardiac Tools (option)
Brain Perfusion (option)
Lung and Abdomen Analysis (option)
Liver and Kidney Analysis (option)
Nodule Analysis (option)
Colon Analysis (option)
Tensor Analysis (option)
Product Specifications
September 2014
Contact Information:
United States of America:
FUJIFILM MEDICAL SYSTEMS, U.S.A., Inc.
419 West Avenue, Stamford, CT 06902, U.S.A.
Manufacturer:
FUJIFILM Corporation
26-30 Nishiazabu, 2 Chome
Minato-ku, Tokyo, 106-8620, Japan
Proprietary Information: For Distribution within FUJIFILM Corporation and its Authorized Affiliates, Distributors, Partners and Customers Only. Duplication or
Distribution of this document without agreement from FUJIFILM Corporation is strictly prohibited.
FUJIFILM Corporation
Synapse 3D (US) Product Specifications: Z30N1138B
Revision History
Date Comment
March 25, 2014 First version
July 16, 2014 Second edition
September 29, 2014 Third edition
2/62
Proprietary Information: For Distribution within FUJIFILM Corporation and its Authorized Affiliates, Distributors, Partners and Customers Only. Duplication or
Distribution of this document without agreement from FUJIFILM Corporation is strictly prohibited.
FUJIFILM Corporation
Synapse 3D (US) Product Specifications: Z30N1138B
Table of Contents
3/62
Proprietary Information: For Distribution within FUJIFILM Corporation and its Authorized Affiliates, Distributors, Partners and Customers Only. Duplication or
Distribution of this document without agreement from FUJIFILM Corporation is strictly prohibited.
FUJIFILM Corporation
Synapse 3D (US) Product Specifications: Z30N1138B
(Note) The name Synapse 3D is the system name of related products. The “Synapse 3D Base Tools” product is
the baseline product package of Synapse 3D. This Product Specification is created specific for “Synapse 3D
Base Tools” and other packages, such as Cardiac Tools. However throughout this specification, the name
“Synapse 3D” is used generally where product name needs to be referred except where the full product name
“Synapse 3D Base Tools” or " Synapse 3D Cardiac Tools " is necessary.
This computer program is protected by copyright law and international treaties. Unauthorized reproduction or
distribution of this program, or any portion of it, is strictly prohibited.
Synapse is a registered trademark of FUJIFILM Medical Systems U.S.A., Inc. and FUJIFILM Corporation.
Precautions
1. Unauthorized transfer or duplication of all or any part of the contents of this Product Specification is strictly
prohibited.
2. The contents of this Product Specification are subject to change without prior notice.
3. FUJIFILM Corporation accepts no liability whatsoever with regard to faults or damage arising from
installation, relocation, modifications, maintenance or repairs performed by anyone other than FUJIFILM
Corporation or a contractor approved by FUJIFILM Corporation.
4. FUJIFILM Corporation accepts no liability whatsoever with regard to faulty functioning of or damage to this
product caused by third-party products other than those supplied by FUJIFILM Corporation.
5. FUJIFILM Corporation accepts no liability whatsoever with regard to faults or damage arising from
modifications, maintenance or repairs in which replacement parts other than the genuine parts approved by
4/62
Proprietary Information: For Distribution within FUJIFILM Corporation and its Authorized Affiliates, Distributors, Partners and Customers Only. Duplication or
Distribution of this document without agreement from FUJIFILM Corporation is strictly prohibited.
FUJIFILM Corporation
Synapse 3D (US) Product Specifications: Z30N1138B
5/62
Proprietary Information: For Distribution within FUJIFILM Corporation and its Authorized Affiliates, Distributors, Partners and Customers Only. Duplication or
Distribution of this document without agreement from FUJIFILM Corporation is strictly prohibited.
FUJIFILM Corporation
Synapse 3D (US) Product Specifications: Z30N1138B
• Basic imaging tools for general images, including 2D viewing, volume rendering and 3D volume viewing,
orthogonal / oblique / curved Multi-Planar Reconstructions (MPR), Maximum (MIP), Average (RaySum)
and Minimum (MinIP) Intensity Projection, 4D volume viewing, image fusion, image subtraction, surface
rendering, sector and rectangular shape MPR image viewing, MPR for dental images, creating and
displaying multiple MPR images along an object time-density distribution, basic image processing, CINE,
measurements, annotations, reporting, printing, storing, distribution, and general image management
and administration tools, etc.
• Tools for regional segmentation of anatomical structures within the image data, path definition through
vascular and other tubular structures, and boundary detection.
• Image viewing tools for modality specific images, including CT PET fusion and ADC image viewing for MR
studies.
• Functional cardiac analysis for CT left ventriculography images: which is intended to evaluate the
functional characteristics of heart.
• Functional cardiac analysis for non-contrast MR heart images: which is intended to evaluate the
functional characteristics of heart.
• Coronary artery analysis for CT coronary arteriography images: which is intended for the qualitative and
quantitative analysis of coronary arteries.
6/62
Proprietary Information: For Distribution within FUJIFILM Corporation and its Authorized Affiliates, Distributors, Partners and Customers Only. Duplication or
Distribution of this document without agreement from FUJIFILM Corporation is strictly prohibited.
FUJIFILM Corporation
Synapse 3D (US) Product Specifications: Z30N1138B
• Coronary artery analysis for MR heart images: which is intended for the qualitative and quantitative
analysis of coronary arteries.
• Calcium scoring for non-contrast CT heart images: which is intended for non-invasive identification and
quantification of calcified atherosclerotic plaques in the coronary arteries using tomographic medical
image data and clinically accepted calcium scoring algorithms.
• Cardiac Fusion: which is intended to analyze cardiac anatomy and pathology with a fused image of
functional data (e.g. NM image, Bulls eye) and anatomical data.
• Aortic Valve Analysis for contrast CT heart images: which is intended for visualization of the heart, aorta
regions, and contour of the aorta, measurement of the vicinity of the aortic valve, measurement of the
calcification area in the aorta.
• use non-contrast and contrast enhanced computed tomographic images of the lung, provide custom
workflows and UI, and reporting functions for lung analysis including boundary detection and volume
calculation for pulmonary nodules in the lung based on the location specified by the user, segmentation
of bronchial tubes in the lung, approximation of air supply region by the user specified bronchial tube,
identifying, displaying and processing low absorption regions in the lung.
• use non-contrast CT images and calculate subcutaneous fat and visceral fat areas in 2D and both
volumes in 3D.
• analyze a bronchus path to reach a lung nodule using the volume data collected with CT, and simulate
insertion of bronchoscope into the path.
7/62
Proprietary Information: For Distribution within FUJIFILM Corporation and its Authorized Affiliates, Distributors, Partners and Customers Only. Duplication or
Distribution of this document without agreement from FUJIFILM Corporation is strictly prohibited.
FUJIFILM Corporation
Synapse 3D (US) Product Specifications: Z30N1138B
• use non-contrast and contrast enhanced computed tomography (CT) images, provide custom workflows
and UI, and reporting functions for nodule analysis including boundary detection of nodules based on the
location specified by the user, measurement of nodules, temporal comparison of nodule images, and
fusion of a CT nodule image on a PT nodule image.
• use positron emission tomography (PT) images, provide custom workflows and UI, and reporting
functions for nodule analysis including SUV measurement of nodules with sphere VOI based on the
location specified by the user, temporal comparison of nodule images, and fusion of a PT nodule image
on a CT nodule image.
8/62
Proprietary Information: For Distribution within FUJIFILM Corporation and its Authorized Affiliates, Distributors, Partners and Customers Only. Duplication or
Distribution of this document without agreement from FUJIFILM Corporation is strictly prohibited.
FUJIFILM Corporation
Synapse 3D (US) Product Specifications: Z30N1138B
Addition to the general 2D and 3D image processing and measurement tools available in Synapse 3D Base
Tools, Synapse 3D Tensor Analysis provides custom workflows, UI, and reporting functions for tensor analysis
with neck and head MR images. It includes display of diffusion and FA color map images, white matter
tractography, dynamic review in MR, vessel and body visualization with registration of MR, CT, XA, PET and
NM.
Contraindications
None known.
9/62
Proprietary Information: For Distribution within FUJIFILM Corporation and its Authorized Affiliates, Distributors, Partners and Customers Only. Duplication or
Distribution of this document without agreement from FUJIFILM Corporation is strictly prohibited.
FUJIFILM Corporation
Synapse 3D (US) Product Specifications: Z30N1138B
Before using this workstation, please read “For Safe Operation” carefully so that you can operate it correctly.
Whenever you operate this workstation, be sure to observe all precautions. Failure to do so may subject you to
injuries or cause property damage.
The medical institution where the equipment is installed is responsible for its use and maintenance. This
workstation should not be used by persons other than suitably trained physicians or staffs.
Rx only in the United States (Federal law restricts this device to sale by or on the order of a physician)
Synapse 3D (workstation) must operate on the specified computers and monitors that satisfy the Safety
Standards and Regulations, Electromagnetic Interference Standards and Regulations, as well as Environmental
Regulations valid in each country or region where Synapse 3D is used. When discarding the hardware, the user
must comply with environmental regulations valid in each country or region. The environmental regulations
include, but not limited to, the EC directive 2002/96/EC of the European Parliament and of the Council of 27
January 2003 on waste electrical and electronic equipment (WEEE Directive).
Computers and monitors specified in the Synapse 3D product specification are designed to meet the
requirements of IEC60950 or equivalent standards but not in accordance with IEC60601-1 required for medical
10/62
Proprietary Information: For Distribution within FUJIFILM Corporation and its Authorized Affiliates, Distributors, Partners and Customers Only. Duplication or
Distribution of this document without agreement from FUJIFILM Corporation is strictly prohibited.
FUJIFILM Corporation
Synapse 3D (US) Product Specifications: Z30N1138B
equipment installed in an environment where a patient may accidentally touch the equipment. To use the
workstation in such an environment, specific safety measures related to the computer to be used are necessary.
Consult the computer dealer for measures to be taken.
The computer to be used with Synapse 3D is classified as electrical equipment. To use the equipment correctly
and prevent any danger, please observe the following electrical equipment use precautions:
2. If there is trouble with the computer, do not attempt to fix it randomly. Instead, indicate the problem as
appropriate and entrust repairs to a professional.
3. The supply voltage to the computer is 100 - 240 V depending on the model and region. Do not remove the
cover or remodel the computer.
4. Explosion Precautions
Since this equipment is not explosion-proof, do not use combustible and explosive gases near the equipment.
11/62
Proprietary Information: For Distribution within FUJIFILM Corporation and its Authorized Affiliates, Distributors, Partners and Customers Only. Duplication or
Distribution of this document without agreement from FUJIFILM Corporation is strictly prohibited.
FUJIFILM Corporation
Synapse 3D (US) Product Specifications: Z30N1138B
1.
Synapse 3D is intended for use only as a supplement to traditional methods for interpreting radiological
images. Synapse 3D is only an aid or adjunct to processes or decisions that can be made without the use
of Synapse 3D. No diagnostic or treatment decisions or other decisions that may affect patient care should
be based exclusively on Synapse 3D.
2.
For exported images;
For images exported from Synapse 3D using e-mail, printing or any other mechanism, users must
ensure that recipients of exported images are familiar with this material. The precise appearance of the
image is dependent on the characteristics and settings of the hardware, including monitors, used. The
notes and cautions in this document related to the interpretation of displayed images apply equally to
these exported images.
3.
For studies to be used;
Users must be fully conversant with the limitations of the basic imaging modality and of ensuing image
processing. This includes understanding the limitations of the initial series acquisition, image processing
technology used, and image display methods. Also, be aware that medical imaging is valid only when
appropriate measures have been taken to obtain optimal images with correct orientation and correct
patient identifiers.
4.
For Image processing;
a. The user must keep the limitations of 3D image processing in mind when viewing 3D images. Small
pathologies may go undetected because the interslice distance is too great relative to the size of the
pathology. The creation of a 3D image using interpolation of data points between image slices based
on the original image can create a smooth-looking 3D image. The user must interpret the apparent
lack of pathology in the context of the inherent limitation in the resolution of the original slice images
as scanned.
b. Artifacts may appear in the image depending on the amount of image processing. When using an
image processing method, fully understand its characteristics.
c. The visualization settings of the 3D views may cause some anatomical features to be completely
transparent and therefore invisible. Synapse 3D allows users to hide anatomical features to
emphasize anatomy of interest. Users must take into account any actions and software behaviors of
this type when interpreting displayed images.
12/62
Proprietary Information: For Distribution within FUJIFILM Corporation and its Authorized Affiliates, Distributors, Partners and Customers Only. Duplication or
Distribution of this document without agreement from FUJIFILM Corporation is strictly prohibited.
FUJIFILM Corporation
Synapse 3D (US) Product Specifications: Z30N1138B
5.
When installing, servicing and administrating Synapse 3D;
a. Set and install medical imaging equipment connected to the workstation properly, using the
installation specifications. If not, some errors may occur.
b. The installation, upgrade, system maintenance, system administration, and administration shall
only be performed by trained personnel.
c. Training materials must be created or updated using the correct version of service manual, user
manual, or other formal documents released as a part of the Synapse 3D product.
6.
While operating Synapse 3D;
a. Check the workstation before use to make sure that it operates normally.
b. In extended use of the workstation, take a 10 to 15 minute break after every hour to rest your eyes
and hands.
c. Do not turn off the power to the computer while using the workstation.
d. Perform the shutdown procedure specified in the operating system to end operation.
f. If the computer is forcibly terminated while using the system due to a power outage or other reasons,
data may be lost. It is recommended to use an uninterruptible power supply system. If a power outage
occurs and an uninterruptible power supply system is in use, terminate the workstation as soon as
possible.
g. When leaving an image displayed for a long time during operation, use the screensaver. If it does
not start running, contact your service engineer. If an image has been displayed for a long time, even
after the display changes, the image may be visible as a residual image on the next display. The video
card may become quite hot depending on the screensaver in use, causing a malfunction.
7.
For Hardware and software being used with Synapse 3D;
a. Do not install any hardware onto the workstation other than those designated by FUJIFILM
Corporation.
b. Do not install any software onto the workstation other than those designated by FUJIFILM
Corporation.
c. Do not uninstall originally installed software. Do not install any software other than that originally
installed.
13/62
Proprietary Information: For Distribution within FUJIFILM Corporation and its Authorized Affiliates, Distributors, Partners and Customers Only. Duplication or
Distribution of this document without agreement from FUJIFILM Corporation is strictly prohibited.
FUJIFILM Corporation
Synapse 3D (US) Product Specifications: Z30N1138B
8.
Since the backlight (fluorescent tube) of the LCD monitor has a limited lifetime, the luminance (emitted
light amount) of the screen decreases as running time increases, and it may eventually make viewing
difficult. Check that the emitted light amount of the LCD backlight is appropriate before starting operation.
If not, contact your service engineer.
9.
When using a monitor or a LCD monitor;
a. White or black spots may appear on the LCD monitor. This is part of the normal wear and tear of an
LCD, and it is not a failure of the LCD monitor. These spots caused by LCD characteristics appear in
the fixed place at any time. Recognize their existence to prevent misinterpreting them as possible
medical abnormalities.
b. When the display resolution is too low, it may make viewing difficult. Make sure the settings are
appropriate for use.
c. When the luminance or contrast of the screen is set excessively high or low, it may make viewing
difficult. Make sure the settings are appropriate for use.
d. A color image may not be displayed with the correct gradation (grayscale) when viewed on a
grayscale monitor. Use a monitor that is appropriate for handling the images.
e. When interpreting an image on the monitor, be aware of the purpose and differences in the image
on the film. Image interpretation should be performed at responsibility of the user. For image
interpretation via the monitor, use high-definition and high-quality monitors as much as possible.
f. If the angle of view is not appropriate for image interpretation, it may affect correct image
interpretation. Check the angle of view, and adjust the monitor angle if necessary.
10.
Before interpreting an image, operate the workstation so that images are displayed at the appropriate size.
Always check the image size in the display scale.
11.
Regarding the direction and the object orientation in image;
a. The image display direction depends on the DICOM information attached to the image by the
medical equipment used to produce the image. When interpreting an image, make sure the direction
of the image is correct.
b. When displaying images in which it is hard to differentiate the left from the right (such as images of
kidneys and head), use the "L" or "R" displayed with the cube to confirm the orientation of the image.
The following image is displayed by reversing the right and left orientation of original image.
14/62
Proprietary Information: For Distribution within FUJIFILM Corporation and its Authorized Affiliates, Distributors, Partners and Customers Only. Duplication or
Distribution of this document without agreement from FUJIFILM Corporation is strictly prohibited.
FUJIFILM Corporation
Synapse 3D (US) Product Specifications: Z30N1138B
12.
For measurements;
a. Inappropriate analysis results and measurements may be output if inappropriate points or lines are
used for the basis of analysis or measurement in applications. Set them appropriately.
b. When performing measurement, specify the coordinates and the units carefully.
c. For measurement, the workstation uses the DICOM information attached to the image by the
medical equipment used to produce the image. Medical practice using the measurement results
should be performed at the user's own responsibility.
d. Use the workstation taking into account errors between an image and the measurement result. An
image used for measurement may have errors in the length of the subject depending on the image
capturing method.
e. Note that if you measure the distance or other things on a saved image of CPR, the measurement
is performed on a curved surface.
f. Accuracy of the measurement functions is described in the measurement section in the User
Manual. Read and understand the accuracy and limitation information of measurement before use.
13.
Operate the security functions and procedures as specified to protect the workstation or patient
information from an unauthorized access.
14.
For installation environment;
15/62
Proprietary Information: For Distribution within FUJIFILM Corporation and its Authorized Affiliates, Distributors, Partners and Customers Only. Duplication or
Distribution of this document without agreement from FUJIFILM Corporation is strictly prohibited.
FUJIFILM Corporation
Synapse 3D (US) Product Specifications: Z30N1138B
a. When disposing of the computer, make sure to erase patient information that may be recorded in
the magnetic disk (HDD) before disposal.
b. Installing the monitor in a place where strong light is emitted may prevent correct viewing because
the display is either directly exposed to the light or reflects the light. Install the equipment in a place
free from such influence.
c. Using the computer in a place where a strong magnetic field is present may cause the displayed
image to be distorted. Install the equipment in a place free from such influence.
d. A plasma monitor constitutively emits a small amount of infrared radiation from the panel surface.
Therefore, any equipment operating near the monitor that uses infrared rays may be affected. When
installing a plasma monitor, check that it does not affect neighboring devices. If the monitor interferes
with devices nearby, change the orientation of their transmitting/receiving section to prevent the
interference.
15.
When an error occurs;
a. If the workstation malfunctions, take action by following the messages displayed on the screen.
b. If an error occurs in the workstation, turn off the power, indicate that the equipment is faulty by posting
a notice or other means, and then contact your service engineer.
16/62
Proprietary Information: For Distribution within FUJIFILM Corporation and its Authorized Affiliates, Distributors, Partners and Customers Only. Duplication or
Distribution of this document without agreement from FUJIFILM Corporation is strictly prohibited.
FUJIFILM Corporation
Synapse 3D (US) Product Specifications: Z30N1138B
1 Product overview
The product “Synapse 3D Base Tools” is a system that receives and retrieves DICOM data from modalities including
CT, MR, PET, XA, etc. and constructs 3D images and performs image analysis.
▀ Optimum operability
Users can configure a wide range of settings to enable optimum operations. They can enjoy stable
operations at an optimal speed without noticing any slowdowns thanks to optimized processing
based on the latest multi-core CPU and native 64-bit OS for data loading and 3D reconstruction.
▀ 3D image processing
The 3D processing is optimized to the sub-voxel level to achieve the best imaging performance while
providing the freedom to set object surface textures, 3D ray directions, and light attribute settings.
Useful operations, such as clipping and interactive processing, can be performed in real time.
Rendering can be performed at multiple resolutions.
▀ Various applications
Various, useful applications are available to use for different body parts and images from many
modalities. Extremely large images and image series (e.g. cardiac images) can be loaded directly
into memory. Operation is simple and easy using Image Intelligence™, Fujifilm’s proprietary image
processing technology.
17/62
Proprietary Information: For Distribution within FUJIFILM Corporation and its Authorized Affiliates, Distributors, Partners and Customers Only. Duplication or
Distribution of this document without agreement from FUJIFILM Corporation is strictly prohibited.
FUJIFILM Corporation
Synapse 3D (US) Product Specifications: Z30N1138B
2 Product configurations
The product “Synapse 3D Base Tools” consists of the following two modules: Synapse 3D Server or Synapse 3D
Standalone for 3D image construction and analysis, and Synapse 3D Client.
(Standalone type is newly-included from ‘Synapse 3D V4.x’.)
There are two types of Synapse 3D client, one is dedicated client software which can launch Synapse 3D directly,
another is Synapse integration client software which can launch Synapse 3D through Synapse integration.
The following configuration is a connection example among Synapse 3D Server or Synapse 3D Standalone, image
modalities, Synapse Server (image data server) and Synapse 3D Client.
The number of Synapse 3D Servers varies according to the maximum number of concurrent connections. In the
case of multiple servers, high performance is realized with a server cluster where one machine works as the master
server and the others work as slave servers. When the master server receives a login request from a Synapse 3D
Client, the master server can select a login server automatically by confirming each server’s login status.
Image Diagnostic
Device (CT, MR etc) Synapse 3D NAS
i Master server
(DICOM server)
Non-DICOM Image
DICOM Transfer
Diagnostic Device
G/W device ii
(CT, MR etc)
v
vi iii iv
Generic PC
Synapse Server
DICOM Printer
Image Server
Synapse クライアント
Synapse クライアント
Synapse Client
18/62
Proprietary Information: For Distribution within FUJIFILM Corporation and its Authorized Affiliates, Distributors, Partners and Customers Only. Duplication or
Distribution of this document without agreement from FUJIFILM Corporation is strictly prohibited.
FUJIFILM Corporation
Synapse 3D (US) Product Specifications: Z30N1138B
i. DICOM Storage
Receive DICOM-compliant images such as CT and MR images
ii. Receive images through interface
For modalities that do not support the DICOM standard, separate DICOM conversion GW device is
required
iii. DICOM Q/R Storage
Display and analyze images by DICOM query and retrieval
iv. DICOM Printer
Enables DICOM printing
v. Local Data Import
Import DICOM data from CD/DVD
vi. Integration with Synapse
Refer DICOM database of Synapse
<Standalone Configuration>
i. DICOM Storage
Receive DICOM-compliant images such as CT and MR images
ii. Receive images through interface
For modalities that do not support the DICOM standard, separate DICOM conversion GW device is
required
iii. DICOM Q/R Storage
Display and analyze images by DICOM query and retrieval
iv. DICOM Printer
Enables DICOM printing
v. Local Data Import
Import DICOM data from CD/DVD
Notes
• With respect to connections to other systems, this covers those devices confirmed in advance by
connection testing etc.
• The standalone model does not support connections from tablet clients.
19/62
Proprietary Information: For Distribution within FUJIFILM Corporation and its Authorized Affiliates, Distributors, Partners and Customers Only. Duplication or
Distribution of this document without agreement from FUJIFILM Corporation is strictly prohibited.
FUJIFILM Corporation
Synapse 3D (US) Product Specifications: Z30N1138B
□ License key
20/62
Proprietary Information: For Distribution within FUJIFILM Corporation and its Authorized Affiliates, Distributors, Partners and Customers Only. Duplication or
Distribution of this document without agreement from FUJIFILM Corporation is strictly prohibited.
FUJIFILM Corporation
Synapse 3D (US) Product Specifications: Z30N1138B
4 Requirements
The following are the off-the-shelf software and hardware requirements used for Synapse 3D Base Tools.
Server(*1) Recommended requirements Minimum requirements(*6)
OS Microsoft® Windows Server® 2012 R2 Microsoft® Windows Server® 2008
Standard Edition (x64) Standard Edition SP2 (x64) (*7)
CPU (*5) Intel® Xeon® processor E5-2630 v2 Quad-core Intel® Xeon® processor
2.6GHz or greater, 2CPUs or more 2.0GHz or greater, 2 CPUs or more
Main memory 64GB
HDD SAS 10000rpm SATA 7200rpm
21/62
Proprietary Information: For Distribution within FUJIFILM Corporation and its Authorized Affiliates, Distributors, Partners and Customers Only. Duplication or
Distribution of this document without agreement from FUJIFILM Corporation is strictly prohibited.
FUJIFILM Corporation
Synapse 3D (US) Product Specifications: Z30N1138B
22/62
Proprietary Information: For Distribution within FUJIFILM Corporation and its Authorized Affiliates, Distributors, Partners and Customers Only. Duplication or
Distribution of this document without agreement from FUJIFILM Corporation is strictly prohibited.
FUJIFILM Corporation
Synapse 3D (US) Product Specifications: Z30N1138B
Image storage capacity can be expanded by using extra network attached storage (NAS)].
*3 This requirement is only for the master server. The slave server doesn’t need to satisfy this requirement.
*4 Configuration using a 1Gbps network is recommended. If the network load is high, performance may drop
significantly.
*5 Only Intel CPUs are supported.
*6 Other software or problems specific to hardware might affect this software behavior.
*7 This requirement is only for upgrade from V2 to V4.
23/62
Proprietary Information: For Distribution within FUJIFILM Corporation and its Authorized Affiliates, Distributors, Partners and Customers Only. Duplication or
Distribution of this document without agreement from FUJIFILM Corporation is strictly prohibited.
FUJIFILM Corporation
Synapse 3D (US) Product Specifications: Z30N1138B
24/62
Proprietary Information: For Distribution within FUJIFILM Corporation and its Authorized Affiliates, Distributors, Partners and Customers Only. Duplication or
Distribution of this document without agreement from FUJIFILM Corporation is strictly prohibited.
FUJIFILM Corporation
Synapse 3D (US) Product Specifications: Z30N1138B
Image application
● 2D Viewer: Simple viewer for 2D reading
● 3D Viewer: Displays 2D cross sections, 3D images
● 4D Viewer: Displays 4D cine play of multi-phase images, e.g., cardiac images, edits phase time
information
● 3D Comparison: Compares multiple series images in 2D cross section and 3D display
● Slicer: Application for observation of arbitrary cross section
● Dynamic Data: Displays profile curve and color map of dynamic images
● Fusion: Viewer superimposes two 3D intra/intermodality images
● 3D Compositor: Viewer displays multiple 3D images in the same coordinate space
● Vessel Extraction: Uses images pre-contrast and post-contrast agent, extracts and displays cerebral
blood vessels
●Nuclear Medicine Viewer: Overlay display of CT images with NM or PET images
● Combination: Combines multiple CT or MR images taken in two or more times.
● Sector MPR: Using single/multiple phase CT, MR images, observes sector or rectangular MPR images
● Dental MPR: Using CT image, creates and displays panorama image along dental arch and cross
section of Alveolar bone
● ADC Viewer: Using MR diffusion image, calculates and displays ADC map
● 2D Fusion: Viewer superimposes two 2D intra/intermodality images
●General CPR: Extraction and CPR display of vessel paths and the aorta; stenosis rate and color
mapping analysis; virtual stent graft
● MPR Reformat: Observation of arbitrary cross sections along the reference line using either line,
sector, or circle
● Tx Map: Creates a T2* Map using T2 images
25/62
Proprietary Information: For Distribution within FUJIFILM Corporation and its Authorized Affiliates, Distributors, Partners and Customers Only. Duplication or
Distribution of this document without agreement from FUJIFILM Corporation is strictly prohibited.
FUJIFILM Corporation
Synapse 3D (US) Product Specifications: Z30N1138B
cardiac function
● Calcium Scoring: Using cardiac CT image, displays color mapping of calcifications of the coronary
arteries , and calculates Agatston score
● Cardiac Fusion: Displays fusion image between cardiac structure image and cardiac function image
● Aortic Valve Analysis: Using single or multiple phase cardiac CT images, measure an aortic valve
● MR Flow Analysis: Calculates the blood flow volume and the flow velocity per heart rate velocity of an
arranged ROI
● 4-Chamber Analysis: Evaluates cardiac function in the four chambers using a single cardiac CT image
or multiple phase images
● Cardiac Ablation Analysis: Pre-operation simulation for ablation and post-operation observation
5.4 Function specifications summary for "Synapse 3D Lung and Abdomen Analysis (option)"
The following is the function specifications summary for "Synapse 3D Lung and Abdomen Analysis (option)".
Image application
● 2D Fat Analysis (CT): Calculates percentage of visceral fat and subcutaneous fat
● 3D Fat Analysis (CT): Using abdominal 3D CT image, calculates volume and percentage of visceral fat
and subcutaneous fat
● Lung Analysis (CT): Using CT image, calculates volume of lung nodules and displays low attenuation
areas
●Lung Analysis Scope (CT): Extracts the bronchus from CT chest images and find the shortest path to
the tumor
●Lung Analysis Resection (CT): Using contrast-enhanced CT images, extract the lung, bronchus, PA,
and PV and display the results in 3D:
5.5 Function specifications summary for "Synapse 3D Liver and Kidney Analysis (option)"
The following is the function specifications summary for "Synapse 3D Liver and Kidney Analysis (option)".
Image application
● Liver Analysis (CT): Using contrast CT image, extracts liver and blood vessel regions and displays
results in 3D
● Liver Analysis (MR): Viewer and analysis dedicated to captured images using an EOB contrast agent
26/62
Proprietary Information: For Distribution within FUJIFILM Corporation and its Authorized Affiliates, Distributors, Partners and Customers Only. Duplication or
Distribution of this document without agreement from FUJIFILM Corporation is strictly prohibited.
FUJIFILM Corporation
Synapse 3D (US) Product Specifications: Z30N1138B
●Kidney Analysis (CT): Using contrast-enhanced CT images of the kidney, extract the kidney and each
vessel area and display the results in 3D
27/62
Proprietary Information: For Distribution within FUJIFILM Corporation and its Authorized Affiliates, Distributors, Partners and Customers Only. Duplication or
Distribution of this document without agreement from FUJIFILM Corporation is strictly prohibited.
FUJIFILM Corporation
Synapse 3D (US) Product Specifications: Z30N1138B
6 System restrictions
The following are restrictions for “Synapse 3D Base Tools.”
Item Description
Number of modality The maximum number of modality connections is 30,
connections And the maximum number of concurrent connections is 10.
Client connections In case of multiple servers’ configuration, the number of slave servers can be
expanded to three units.
The number of servers and the maximum number of concurrent connections
are as follows.
Number of Maximum number of
servers concurrent connections
1 unit 8
2 units 16
3 units 24
4 units 32
28/62
Proprietary Information: For Distribution within FUJIFILM Corporation and its Authorized Affiliates, Distributors, Partners and Customers Only. Duplication or
Distribution of this document without agreement from FUJIFILM Corporation is strictly prohibited.
FUJIFILM Corporation
Synapse 3D (US) Product Specifications: Z30N1138B
the terminal.
*Confirms next chapter ‘7 Upgrade’ about upgrade of this software from ‘Synapse 3D V2.x’ and ‘Synapse 3D V3.x’.
It is possible to connect to this product by installing dedicated client software on a pre-existing generic PC in the
facility.
29/62
Proprietary Information: For Distribution within FUJIFILM Corporation and its Authorized Affiliates, Distributors, Partners and Customers Only. Duplication or
Distribution of this document without agreement from FUJIFILM Corporation is strictly prohibited.
FUJIFILM Corporation
Synapse 3D (US) Product Specifications: Z30N1138B
The recommended specifications for a general-purpose PC are listed below. Even if the PC meets these
requirements, it might not work due to hardware-specific issues, compatibility with other software, etc.
If these requirements are met by existing PCs in the medical facilities, they can be used after the following
checks are performed:
・ Confirmation of the above specifications
・ Actual operational verification
Note that a separate license must be purchased in order to install the software on a PC in the facilities.
30/62
Proprietary Information: For Distribution within FUJIFILM Corporation and its Authorized Affiliates, Distributors, Partners and Customers Only. Duplication or
Distribution of this document without agreement from FUJIFILM Corporation is strictly prohibited.
FUJIFILM Corporation
Synapse 3D (US) Product Specifications: Z30N1138B
31/62
Proprietary Information: For Distribution within FUJIFILM Corporation and its Authorized Affiliates, Distributors, Partners and Customers Only. Duplication or
Distribution of this document without agreement from FUJIFILM Corporation is strictly prohibited.
FUJIFILM Corporation
Synapse 3D (US) Product Specifications: Z30N1138B
7 Upgrade
It is possible to upgrade to ‘Synapse 3D V4.x’ from ‘Synapse 3D V2.X’ or ‘Synapse 3D V3.X’.
An upgrade option needs to be purchased certainly for the upgrade.
It is possible to upgrade not only the software itself but also the software environments.
DICOM images could be migrated by a script attached with the upgrade kit.
A part of user setting (e.g. Color Template) could be taken over to ‘Synapse 3D V4.x’ by a script attached with the
upgrade kit.
Snapshot created in previous version ‘Synapse 3D V2.x’ can be analyzed on ‘Synapse 3D V4.x’ by using
dedicated application based on ‘Synapse 3D V2.x’, if original DICOM image are migrated. In this case, not the
latest communication method but the previous communication method using Terminal Service is applied.
(Snapshots created in ‘Synapse 3D V3.x’ can be loaded directly in ‘Synapse 3D V4.x’.)
A comparison of main configuration between ‘Synapse 3D V3.x’ and ‘Synapse 3D V4.x’ is shown as follows.
FUJIFILM Corporation
Synapse 3D (US) Product Specifications: Z30N1138B
33/62
Proprietary Information: For Distribution within FUJIFILM Corporation and its Authorized Affiliates, Distributors, Partners and Customers Only. Duplication or
Distribution of this document without agreement from FUJIFILM Corporation is strictly prohibited.
FUJIFILM Corporation
Synapse 3D (US) Product Specifications: Z30N1138B
General
Multi-user environment
● Different settings can be specified per user. The users can share the database
DICOM communication as a service
● DICOM communications can run as a service, and can work without users logging in
Online help
● Displays a help file corresponding to the current operations
Server/Client
Setting up login method
● In the case of multiple servers, the login server can be selected automatically according to the
server’s operating status
Restricting functions available to the client
● Ability to restrict functions. Major functions that can be restricted are:
setting of available printers, adjusting screen resolution, storage of images/reports/snapshots,
image import/delete/transfer/export
Monitoring client connections status
●Server’s login status can be checked through a Web monitor
● Ability to send a message to a connected client and to force a connection to close
Access from tablet devices
● Snapshots can be observed using a tablet-specific UI.
● When communication is disconnected, it can be re-connected and the operating state restored.
DICOM server
Data management with SQL Server
● PostgreSQL Server 2005 manages database
● Database backup and restoration can be performed using SQL Server Management Studio
Setting of local AE
● AE title and port can be configured
Maximum number of concurrent connections
● The number can be configured
Maximum P-DATA-TF length
● Can be specified for receive and send operations separately
34/62
Proprietary Information: For Distribution within FUJIFILM Corporation and its Authorized Affiliates, Distributors, Partners and Customers Only. Duplication or
Distribution of this document without agreement from FUJIFILM Corporation is strictly prohibited.
FUJIFILM Corporation
Synapse 3D (US) Product Specifications: Z30N1138B
Data management
Local data search
● Search by patient ID, patient name, study date, modality, body part examined, and comment
Add comments to local data
● Adds comments to Study, Series
Preset search items for local data
● Search conditions can be set up by user
Display of DICOM header information of local data
35/62
Proprietary Information: For Distribution within FUJIFILM Corporation and its Authorized Affiliates, Distributors, Partners and Customers Only. Duplication or
Distribution of this document without agreement from FUJIFILM Corporation is strictly prohibited.
FUJIFILM Corporation
Synapse 3D (US) Product Specifications: Z30N1138B
Print
Print to DICOM printers
● Settings specified at time of print: page range, number of copies, print direction, film
Prior settings by service engineers: color, print priority, print media, film output destination,
magnification type, structure information, smooth type, edge density, blank image density, edging,
resolution (width, height, length/pixel)
Print to Windows printers
● Settings specified at time of print: page range and number of copies
Print preview
● Ability to print in additional arbitrary layouts by using a settings file
● Manages print queue of DICOM printers (display of data transfer status, cancellation of print
requests)
● Image adjustment from preview by WL/magnification
36/62
Proprietary Information: For Distribution within FUJIFILM Corporation and its Authorized Affiliates, Distributors, Partners and Customers Only. Duplication or
Distribution of this document without agreement from FUJIFILM Corporation is strictly prohibited.
FUJIFILM Corporation
Synapse 3D (US) Product Specifications: Z30N1138B
CPR display
● Straightened CPR
● Stretched CPR
● Projected CPR display
● Multi-path CPR display
● Automatic/manual extraction of path
● Automatic/manual extraction of path diameter
● Displays of graphs of average diameter, projected length of radius, area, and average intensity
● Measures size of stenosis of blood vessels, etc. by specifying at least one normal diameter
● Assigns up to four colors to blood vessels according to intensity values to overlay plaque portions
● SMIP display: MIP rendering display calculated from areas near surface regions
● Ability to change CPR display mode among MIP, MinIP, and Ray summation modes
● Orthogonal cross section view of image with smoothing filter or edge enhancement filter applied
● Connection of path
3D display
● Volume rendering (VR)
● Shaded surface rendering (SSD)
● Maximum intensity projection (MIP)
● Minimum intensity projection (MinIP)
● Ray summation (Raysum)
VR color template
● Changes parameters (ambient light, diffuse light, specular light, shininess, decay level, opacity,
color mixture, light source orientation, self shadow)
● Sets up the shape of opacity function (freehand, increase toward right, decrease toward right,
concave type, convex type)
● Copy and paste of color templates
● Displays 2D cross section in colors specified by the color templates
● Registers color templates as a group
37/62
Proprietary Information: For Distribution within FUJIFILM Corporation and its Authorized Affiliates, Distributors, Partners and Customers Only. Duplication or
Distribution of this document without agreement from FUJIFILM Corporation is strictly prohibited.
FUJIFILM Corporation
Synapse 3D (US) Product Specifications: Z30N1138B
● Displays live color templates (Thumbnail images of the current 3D image after applying registered
templates)
● Export and import color templates
● Share color templates among users
Changing window width and level
● Applies to MIP, MinIP, Ray summation of 2D and 3D views
● Gamma conversion
● 2D cross section view with smoothing filter or edge enhancement filter applied
● Pseudo-color settings such as normal display, reverse display, and individual pseudo-colors can
be specified
● Preset registration and recall of window width, window level, and pseudo-color
● Shared presets among users
Movie creation
● Creates movies by paging 2D cross sections in specified range
● Creates 3D movies by registering key frames based on changing parameters, e.g., orientation,
magnification rate, clipping, color template. Key frames can be registered one by one or
automatically registered when a parameter changes. Selectable presets can be created
● Registers key frames as a preset
● Offline VR: can generate files that enable simple VR operations on any device using a Web
browser
● Converts created movies to private DICOM data and stores them. Can transfers them to other
servers
3D filtering
● Removes noise with a low-pass filter and smoothly displays images in 3D
Measurement
● Measures lines, projected line, polygonal lines, angles, intensity values, rectangles, ellipses,
polygons, freehand, boxes, spheres
● Calculates volume of mask region, and average/max/min/standard deviation of intensity values in
3D images
● Calculates volume of mask region, and average/max/min/standard deviation of intensity values in
2D images
● Displays profile curve or histogram of each measurement
● Semi-automatic measurement of half-width
● Stores measurement results and graph in text file (.csv, .txt) or in capture image
● Ability to create ROI along the contour of an image by detecting contours during freehand
measurement
Annotation
● Annotation by arrows, points, line, cylinder, text, and balloons
● Ability to set up font, color, background color, and frame for text. Ability to register key words
● Ability to set up font, color, and background color for balloons. Ability to register key words
38/62
Proprietary Information: For Distribution within FUJIFILM Corporation and its Authorized Affiliates, Distributors, Partners and Customers Only. Duplication or
Distribution of this document without agreement from FUJIFILM Corporation is strictly prohibited.
FUJIFILM Corporation
Synapse 3D (US) Product Specifications: Z30N1138B
3D clipping
● Specifies display range for X/Y/Z directions and specifies inside or outside clipping
● Plane clip, sight-line clip, sphere clip, local observation clip
● Centering
Magnifies using visible 3D areas as rotation center
● Ability to paste 2D cross section image on the plane of the border region of clipping
● Mask extraction of clipped areas
Mask editing
● Overlay mask region on 2D cross section view
● Undo/Redo up to 8 times
Extraction/removal of body parts: bone removal, bone extraction, bed removal, heart extraction,
colon extraction, lung field extraction, bronchus extraction, liver extraction, region search, vessel
tracking (all except blood vessel tracking are available for CT images only), spinal extraction,
body surface extraction, tube extraction, presets (registers extraction mode, target mask, and
color template to apply as one job)
● Smart cut (mask extraction/removal by surrounding a region) using polygon, curve, or freehand
● Smart select (extraction/removal by selecting an object)
● Smart tracking (extraction/removal by region growing)
● Dilation, Erosion
● Extracts a region within a specified threshold range as a mask. Applied to the whole image or an
ROI
● Extract a region with a specified opacity range as a mask. Applied to the whole image or an ROI
● Deletes continuous regions that have smaller volume than a specified mask value. Applied to the
whole image or an ROI
● Surface digger (removes surface mask, which is specified by ROI or point, along view direction)
● Adds/Deletes mask region from 2D cross section by using circle cursor
● Adds/Deletes a sphere mask on a clicked position
● Adds/Deletes mask corresponding to the regions created by interpolating multiple ROIs set on the
cross sections
● Function to fill in a mask on a 2D slice
Multi mask
● Registers and displays simultaneously up to 12 layers
● Status confirmation (current mask status) in real time
● Ability to reverse mask, perform ADD, SUB, AND, XOR operations between two layers
● Up to 8 layers can be used as additional mask storage (registration only)
Reformat
● Ability to reformat by line, sector, circle
● Plays cine of reformatted slices
● Saves reformatted slices as sequential images or a movie
SUV evaluation
39/62
Proprietary Information: For Distribution within FUJIFILM Corporation and its Authorized Affiliates, Distributors, Partners and Customers Only. Duplication or
Distribution of this document without agreement from FUJIFILM Corporation is strictly prohibited.
FUJIFILM Corporation
Synapse 3D (US) Product Specifications: Z30N1138B
● Displays SUV (max, min, standard deviation) of sphere VOI of PET images
Supported manufacturers: Siemens, Philips, Shimadzu, GE
Report
● Ability to create reports
Ability to create and save reports as private DICOM data
Snapshot save
● Saves operation status
When the status is saved as private DICOM data, it can be transferred to another server for
restoration
Surface display
● Display polygon model and output polygon model to external files
2D Viewer
Simple viewer for 2D reading
● Displays cine play
● Allows switching among cross section display types (e.g. axial, sagittal, coronal, etc.)
● Real time stack reconstruction (Joins multiple cross-sectional slices together and displays the
result in Ray summation, MIP, or MinIP mode)
● Links coordinates between series
● Applies the same display position, window width and level, and magnification rate to specified
images while displaying them
● Captures each slice of entire selected series
● Can browse a series list of related study data
● Stores and restores image layout
3D Viewer
Display of 2D cross sections, 3D images and virtual Endoscopy views
● Displays 2D cross sections
● 3D view, edit
● Binocular stereo
Simultaneous display of 3D views with small difference in angle
● Inserts image plane
Inserts 2D cross section planes into 3D view and synchronizes with 2D cross section view
● Effective use of 3D display area.
Can be used as an additional MPR slice without using 3D display
Comparative observation using analysis results from past studies
● Can compare saved snapshots of past studies side by side with current study viewer.
● WW/WL, current position, pan, rotation, and zoom rate can be switched between SYNC/ASYNC.
● Series registration of CT images using position recognition.
Virtual endoscope
40/62
Proprietary Information: For Distribution within FUJIFILM Corporation and its Authorized Affiliates, Distributors, Partners and Customers Only. Duplication or
Distribution of this document without agreement from FUJIFILM Corporation is strictly prohibited.
FUJIFILM Corporation
Synapse 3D (US) Product Specifications: Z30N1138B
4D Viewer
Performs 4D play/edit of multi-phase images, e.g., cardiac images
● Cine play with synchronization between 2D cross sections and 3D images
● Ability to specify display time for each phase in milliseconds or % of the total play time
● Creates movies of all images or only 3D images
3D Comparison
Compares multiple series images in 2D and 3D display
● 2D cross section cine play with synchronization among multiple series
● 3D image synchronization among multiple series
● VOI sphere measurement, view comparison, results capture, ability to save measurement values
and graphs to text file(.csv, .txt)
● 3D viewing, editing, measurement
● Real time stack reconstruction (Joins multiple cross-sectional slices together and displays the
result in Ray summation, MIP, or MinIP mode)
● Registration between series using bed position or spine recognition
● Ability to set up synchronization options
● When using multiple monitors, can synchronize study to be analyzed with selected study on study
list.
Dynamic Data
Display of individual parameter images or time-intensity curves of slices of multi-phase data
● Loads single or multiple slice cine images
● Displays time-intensity curves
● Display histogram
● Displays parameter images (difference, time to peak, max-min, area underneath curve)
41/62
Proprietary Information: For Distribution within FUJIFILM Corporation and its Authorized Affiliates, Distributors, Partners and Customers Only. Duplication or
Distribution of this document without agreement from FUJIFILM Corporation is strictly prohibited.
FUJIFILM Corporation
Synapse 3D (US) Product Specifications: Z30N1138B
● Point measurement and ROI measurement as a circle, rectangle, control point, or freehand.
Output of measurement results in .csv format
● Displays and edits time
Fusion
Viewer superimposes two 3D intra/intermodality images
● Comparison reading
● Reference reading
● MPR reading
● Overlay or blending is configurable
● Automatic rigid registration
● Manual rigid registration by translation and rotation
● Registration by specifying reference points
● Composites two images using rigid body registration and supports the following post-processing
reconstructions for 3D viewing: rigid body registration (applied only to overlaid image), subtraction
value, absolute subtraction value, addition value, average value, maximum value, minimum value
In case of different modalities, WL conversion can be specified for reconstruction
● Composites two images using rigid body registration and supports the following post-processing
reconstructions for saving as a DICOM image: rigid body registration (applied only to overlaid
image), subtraction value, absolute subtraction value, addition value, average value, maximum
value, minimum value
In case of different modalities, WL conversion can be specified for reconstruction
Comparative observation using analysis results from past studies
● Can compare saved snapshots of past studies side by side with current study viewer.
● WW/WL, current position, pan, rotation, and zoom rate can be switched between SYNC/ASYNC.
● Series registration of CT images using position recognition.
42/62
Proprietary Information: For Distribution within FUJIFILM Corporation and its Authorized Affiliates, Distributors, Partners and Customers Only. Duplication or
Distribution of this document without agreement from FUJIFILM Corporation is strictly prohibited.
FUJIFILM Corporation
Synapse 3D (US) Product Specifications: Z30N1138B
3D Compositor
3D viewer which displays up to five 3D images in the same coordinate frame
● Automatically or manually adjusts the position of overlay images in reference to a base image
● Registration by specifying reference points
● Perform series registration of CT images using non-rigid registration and display deformed
images
● Displays multiple overlain 3D images in the VR view
● Displays volume mapping
● Synchronizes all 3D views
● Edits masks for each 3D view
● Inserts 2D MPR cross section planes into each 3D image and synchronizes them with the 2D
slice view display
● Stereoscopic view
Slicer
Observes and prints arbitrary cross sections based on set-up slice line
● Creates arbitrary cross sections by a single slice, multiple slices along a line or a curve
● Sets up slice interval, slice thickness and FOV
● Inserts inset image into each images
● Sorting multiple slices according to position
● Sets up lines for intervertebral and centrum automatically by spinal detection
● Outputs Windows/DICOM printer
● Transfers obtained cross sections to capture box or saves in DICOM format
● Displays CPR along slices
● Enter and display related images
Combination
Combines data taken widely as multiple series and creates compound image
● Signal transformation (density adjustment) for pixels
● Manual rigid registration by translation and rotation
● Specifies range of combination
● Displays orthogonal cross section and MIP of compound image
● Processes by 3D Viewer using compound image
● Saves compound image as other series in DICOM dicom format
● Selects direction on saving in DICOM format from axial direction, sagittal direction, coronal
direction or same direction on reading
Settings
● Selects processing method for overlapped region from maximum value or average value
● Specifies size of shortest side of saved image
43/62
Proprietary Information: For Distribution within FUJIFILM Corporation and its Authorized Affiliates, Distributors, Partners and Customers Only. Duplication or
Distribution of this document without agreement from FUJIFILM Corporation is strictly prohibited.
FUJIFILM Corporation
Synapse 3D (US) Product Specifications: Z30N1138B
Vessel Extraction
Extracts blood vessels using both pre-contrast and post-contrast CT, and displays the extraction result
● Runs bed removal on loading the image
● Adjustable mixing ratio of pre-contrast images and post-contrast images for displaying 2D cross
section view and 3D view
● Automatic or manual rigid registration
● Can perform registration by specifying reference points
● Can perform non-rigid registration
● After vessels are extracted, they can be viewed as vessel 3D, vessel MIP, vessel+bone,
vessel+calcification, or stereoscopic display
● After extraction, blood vessels are displayable in 3D view, MIP view, or stereoscopic view. Both
blood vessels and bone are displayed in the 3D view
● Saves vessel image to DICOM data
● 3D Viewer is available to open a view of the vessel image
● Can extract cerebral arteries and veins from vessel image
● Can insert image plane into 3D display
Dental MPR
Displays and prints panorama image along dental arch and cross section of Alveolar bone using CT
image
● Sets up position and direction of horizontal cross section
● Displays panorama image
● Displays cross section of Alveolar bone
● Prints using Windows/DICOM printer
● Saves all images (orthogonal cross section, panorama image and cross section of Alveolar bone)
in DICOM format or transfers to capture box
● Can display virtual implant and nerve and set margin of nerve
● VR display of LR view and AP view
ADC Viewer
Calculates and displays ADC Map from MR Diffusion image
● Reads diffusion weighted image
● Reads background image
● Perform registration automatically or manually
● Displays ADC Map and eADC Map
Supported manufacturers: GE, Philips
● Displays maximum and minimum value of ADC
● Measurement of circular, rectangle and freehand ROIs
● Display computed DWI
44/62
Proprietary Information: For Distribution within FUJIFILM Corporation and its Authorized Affiliates, Distributors, Partners and Customers Only. Duplication or
Distribution of this document without agreement from FUJIFILM Corporation is strictly prohibited.
FUJIFILM Corporation
Synapse 3D (US) Product Specifications: Z30N1138B
2D Fusion
Viewer superimposes two 2D intra/intermodality images
● Fuse 2D or 3D images
● Save created fusion images as a DICOM file
Sector MPR
Observation of sector or rectangular MPR images using CT and MR images with single or multiple
phases
● For images with multiple phases, plays them in cine mode in 2D slice, 3D, or creates sector or
rectangular MPR. Duration of each phase is editable
● Displays a sector shaped cross section MPR image based on specified position and orientation
● Reversible display of gray-scale images
● Applies specified image processing and displays pseudo-shadow behind the region that has
signal value outside the threshold
● Positional link of sector or rectangular MPR image to 2D cross section
● Displaying an overlay of the cross section image of sector or rectangular MPR onto the 3D view
● The target display can be selected from skin surface, heart, and liver. While the overlay image
display can be selected from skin surface, heart, liver, and bone
● Enables changes in the range of the slab thickness in the sector or rectangular MPR view
● Provides a choice of view plane types: sector or rectangle
● Measurement of angle and distance between the puncture position and the target point in
centesis mode
General CPR
Extract vessel path and observe and analyze vessels by CPR.
♦ Display straightened CPR.
♦ Display stretched CPR.
♦ Display projected CPR.
♦ Display multi-path CPR.
♦ Automatic path search, and manual settings.
♦ Semi-automatic path search, and manual settings.
♦ Display graph of mean diameter, projection diameter, area, and mean intensity on the path.
♦ Specify one or more normal diameters and measure stenosis rate of vessels, etc.
♦ Color mapping analysis by assigning colors to signal values of up to 6 levels on the vessel.
♦ SMIP display: MIP rendering display of only vicinity along multiple paths.
♦ Can change CPR display to MIP, MinIP, or RaySum.
♦ Orthogonal view of applying smoothing filter or edge enhancement filter.
♦ Connect path route.
♦ Stenosis measurement.
45/62
Proprietary Information: For Distribution within FUJIFILM Corporation and its Authorized Affiliates, Distributors, Partners and Customers Only. Duplication or
Distribution of this document without agreement from FUJIFILM Corporation is strictly prohibited.
FUJIFILM Corporation
Synapse 3D (US) Product Specifications: Z30N1138B
MPR Reformat
Observe and save an arbitrary oblique plane based on scan line that was set
♦ Create arbitrary cross section using axis, line, sector, or circle.
♦ Set slice spacing, thickness, and FOV.
♦ Sort multiple slices according to position.
♦ Transfer obtained slice images to a capture box, or save as DICOM images.
Tx Map
Calculate and display T2* map from MR images
♦ Load diffusion-weighted images.
♦ Display T2* map.
♦ Display max and min T2* values.
♦ ROI measurement using a point, ellipse, rectangle, polygonal line, or free-form shape.
♦ Display time-intensity curve.
♦ 3D display of map image
46/62
Proprietary Information: For Distribution within FUJIFILM Corporation and its Authorized Affiliates, Distributors, Partners and Customers Only. Duplication or
Distribution of this document without agreement from FUJIFILM Corporation is strictly prohibited.
FUJIFILM Corporation
Synapse 3D (US) Product Specifications: Z30N1138B
47/62
Proprietary Information: For Distribution within FUJIFILM Corporation and its Authorized Affiliates, Distributors, Partners and Customers Only. Duplication or
Distribution of this document without agreement from FUJIFILM Corporation is strictly prohibited.
FUJIFILM Corporation
Synapse 3D (US) Product Specifications: Z30N1138B
48/62
Proprietary Information: For Distribution within FUJIFILM Corporation and its Authorized Affiliates, Distributors, Partners and Customers Only. Duplication or
Distribution of this document without agreement from FUJIFILM Corporation is strictly prohibited.
FUJIFILM Corporation
Synapse 3D (US) Product Specifications: Z30N1138B
49/62
Proprietary Information: For Distribution within FUJIFILM Corporation and its Authorized Affiliates, Distributors, Partners and Customers Only. Duplication or
Distribution of this document without agreement from FUJIFILM Corporation is strictly prohibited.
FUJIFILM Corporation
Synapse 3D (US) Product Specifications: Z30N1138B
rendering
● Displays myocardial region
● Can specify how to handle papillary muscle volume (in ventricle or in myocardium)
● Outputs analytical values, bull’s eye plots, ventricular volume curve, observations, and images to
a report
● Calls Cardiac Fusion to overlay the Cardiac Function analysis results onto the heart surface
together with the Coronary Artery analysis results
Calcium Scoring
Using non-contrast CT images, this plugin creates a color-classified view of plaque area in the coronary
arteries and calculates the Agatston score
● Automatic extraction of heart
● Calculates the number of plaques for each vessel (LM/LAD/LCX/RCA), volume (area), Agatston
score, integration score per voxel, mean CT value, and maximum CT value
● Specifies plaque region
● Supports creation of new vessel regions for scoring
● Saves the scoring result to a .csv file or a .txt file, or captures the result
● Outputs scoring results, observations, and images to a report
Cardiac Fusion
Display a bull's eye and a mapping of the nuclear medicine image to the 3D heart surface, which are
obtained by the cardiac function analysis system application
● Overlays bull’s eye images onto the base image, making out of range pixels transparent
● Can perform Stress/Rest calculation and results mapping using nuclear medical images
● Displays coronary artery regions together with nuclear medical images
● Maps images of cardiac function or nuclear medicine onto the left ventricle image
● Selectable view of cardiac 3D images: the entire heart, the origins of arteries, or the left ventricle
● Displays an overlay of the images of cardiac function or nuclear medicine onto the CPR image
● Displays axial/sagittal/coronal/long-axis/short-axis views. When structure image is different from
function image, fusion is available to display a superimposed image
● Registration
● Creates reports
Displays bull's eye
● Transforms bull’s eye plots for display according to the length of the coronary arteries
● Inverse display of bull's eye
● Changes the range of colors by specifying the upper limit and the lower limit
● Displays Topographic map
50/62
Proprietary Information: For Distribution within FUJIFILM Corporation and its Authorized Affiliates, Distributors, Partners and Customers Only. Duplication or
Distribution of this document without agreement from FUJIFILM Corporation is strictly prohibited.
FUJIFILM Corporation
Synapse 3D (US) Product Specifications: Z30N1138B
4 chamber analysis
Perform cardiac function analysis on the four chambers (left ventricle, right ventricle, left atrium, right
atrium) in CT Cine images of the heart
♦ Automatic extraction of heart with manual correction.
♦ Automatic extraction of 4 chambers of heart with manual correction.
♦ Automatic setting of long axis with manual correction.
♦ 4D Cine play.
♦ 4D movie creation.
♦ Chamber list view (2-chamber, 3-chamber, 4-chamber, aortic valve).
♦ Specify apex and base for each phase or all phases, and specify center position of each slice
(short-axis image).
♦ Auto-detect contour of left ventricle and myocardium.
51/62
Proprietary Information: For Distribution within FUJIFILM Corporation and its Authorized Affiliates, Distributors, Partners and Customers Only. Duplication or
Distribution of this document without agreement from FUJIFILM Corporation is strictly prohibited.
FUJIFILM Corporation
Synapse 3D (US) Product Specifications: Z30N1138B
♦ Display results: end-diastolic volume (EDV), end-systolic volume (ESV), end-diastolic volume
index (EDVI), end-systolic volume index (ESVI), stroke volume (SV), cardiac output (CO),
cardiac index (CI), peak filling rate (PFR), time to peak filling (TPF), ejection fraction (EF), body
surface area, myocardial mass.
♦ Bull's eye display: ventricular radius, region EF, wall thickness, wall thickening rate, wall motion,
etc.
♦ Display EF curve of 4 chamber region.
♦ Display surface of each 4 chamber region.
♦ Surface display that maps each bull's eye color to surface of ventricle.
♦ Display myocardial regions.
♦ Choice of whether to include volume of ventricular papillary muscles in either ventricle or
myocardium.
♦ Output analysis values, bull's eye, EF curve, observation, and images to a report.
♦ Non-rigid registration between phases.
♦ Bull's eye display of parameters using information from non-rigid registration between phases.
♦ Can start cardiac fusion in conjunction with results of coronary analysis, and map results of 4
chamber analysis onto surface of heart.
52/62
Proprietary Information: For Distribution within FUJIFILM Corporation and its Authorized Affiliates, Distributors, Partners and Customers Only. Duplication or
Distribution of this document without agreement from FUJIFILM Corporation is strictly prohibited.
FUJIFILM Corporation
Synapse 3D (US) Product Specifications: Z30N1138B
4D Perfusion
Perfusion analysis of 3D CTs of the head
♦ Load and analyze three-dimensional images.
♦ Automatically detect positions of arteries and veins, or set them manually.
♦ Can perform Cine playing of slice images or 3D images.
♦ Calculate CBV/MTT/CBF/TTP images.
♦ Overlay display of a parameter image and original image.
♦ Set vessel removal ratio.
♦ Set ROI (can also set symmetrical ROI).
♦ Display time-intensity curve.
53/62
Proprietary Information: For Distribution within FUJIFILM Corporation and its Authorized Affiliates, Distributors, Partners and Customers Only. Duplication or
Distribution of this document without agreement from FUJIFILM Corporation is strictly prohibited.
FUJIFILM Corporation
Synapse 3D (US) Product Specifications: Z30N1138B
8.4 Detailed functional specifications for "Synapse 3D Lung and Abdomen Analysis (option)"
The following are the detailed functional specifications for “Synapse 3D Lung and Abdomen Analysis (option).”
54/62
Proprietary Information: For Distribution within FUJIFILM Corporation and its Authorized Affiliates, Distributors, Partners and Customers Only. Duplication or
Distribution of this document without agreement from FUJIFILM Corporation is strictly prohibited.
FUJIFILM Corporation
Synapse 3D (US) Product Specifications: Z30N1138B
● Can capture the histogram and the volume list of each region or save them to a file
Cluster analysis
● Classifies low attenuation areas by clustering
● Clustering analysis based on 2D connection components
● Clustering analysis based on 3D connection components
● Overlays colors according to cluster size onto lung region on 2D cross sections
● Displays spheres according to cluster size onto 3D image
● Displays log-log graph made by horizontal axis corresponding to the volume and vertical axis
corresponding to the number.
● Displays regression line for log-log graph
Compare to analysis results of past studies
● Comparison of tumor over time
● Comparison of airway analysis over time
● Comparison of low attenuation area over time
● Comparison of cluster analysis over time
55/62
Proprietary Information: For Distribution within FUJIFILM Corporation and its Authorized Affiliates, Distributors, Partners and Customers Only. Duplication or
Distribution of this document without agreement from FUJIFILM Corporation is strictly prohibited.
FUJIFILM Corporation
Synapse 3D (US) Product Specifications: Z30N1138B
BMI, volume of visceral fat, volume of subcutaneous fat, volume percentage of body fat, volume
percentage of subcutaneous fat, diagnostic result, comments, and images
● Calculation of the specified region volume of the psoas major
56/62
Proprietary Information: For Distribution within FUJIFILM Corporation and its Authorized Affiliates, Distributors, Partners and Customers Only. Duplication or
Distribution of this document without agreement from FUJIFILM Corporation is strictly prohibited.
FUJIFILM Corporation
Synapse 3D (US) Product Specifications: Z30N1138B
8.5 Detailed functional specifications for "Synapse 3D Liver and Kidney Analysis (option)"
The following are the detailed functional specifications for “Synapse 3D Liver and Kidney Analysis (option).”
57/62
Proprietary Information: For Distribution within FUJIFILM Corporation and its Authorized Affiliates, Distributors, Partners and Customers Only. Duplication or
Distribution of this document without agreement from FUJIFILM Corporation is strictly prohibited.
FUJIFILM Corporation
Synapse 3D (US) Product Specifications: Z30N1138B
58/62
Proprietary Information: For Distribution within FUJIFILM Corporation and its Authorized Affiliates, Distributors, Partners and Customers Only. Duplication or
Distribution of this document without agreement from FUJIFILM Corporation is strictly prohibited.
FUJIFILM Corporation
Synapse 3D (US) Product Specifications: Z30N1138B
59/62
Proprietary Information: For Distribution within FUJIFILM Corporation and its Authorized Affiliates, Distributors, Partners and Customers Only. Duplication or
Distribution of this document without agreement from FUJIFILM Corporation is strictly prohibited.
FUJIFILM Corporation
Synapse 3D (US) Product Specifications: Z30N1138B
60/62
Proprietary Information: For Distribution within FUJIFILM Corporation and its Authorized Affiliates, Distributors, Partners and Customers Only. Duplication or
Distribution of this document without agreement from FUJIFILM Corporation is strictly prohibited.
FUJIFILM Corporation
Synapse 3D (US) Product Specifications: Z30N1138B
61/62
Proprietary Information: For Distribution within FUJIFILM Corporation and its Authorized Affiliates, Distributors, Partners and Customers Only. Duplication or
Distribution of this document without agreement from FUJIFILM Corporation is strictly prohibited.
FUJIFILM Corporation
Synapse 3D (US) Product Specifications: Z30N1138B
62/62
Proprietary Information: For Distribution within FUJIFILM Corporation and its Authorized Affiliates, Distributors, Partners and Customers Only. Duplication or
Distribution of this document without agreement from FUJIFILM Corporation is strictly prohibited.
FUJIFILM Corporation