The Oscilloscope: Operation and Applications
The Oscilloscope: Operation and Applications
Applications
1. The Oscilloscope
Oscilloscope Operation (X vs Y mode)
An oscilloscope can be used to measure voltage. It does this by measuring the voltage drop
across a resistor and in the process draws a small current. The voltage drop is amplified and
used to deflect an electron beam in either the X (horizontal) or Y (vertical) axis using an
electric field. The electron beam creates a bright dot on the face of the Cathode Ray Tube
(CRT) where it hits the phosphorous. The deflection, due to an applied voltage, can be
measured with the aid of the calibrated lines on the graticule.
First we will consider the circuitry that amplifies and conditions the voltage to be measured (the
“Amp” block in figure 1).
The input signal can be ac or dc coupled. Ac coupling involves adding a series capacitor. This
has the effect of blocking (removing) the dc bias and low frequency components of a signal.
Dc coupling does not have this problem and therefore allows you to measure voltages right
down to 0 Hz. Ac coupling is useful when you are trying to measure a small ac voltage that is
“on-top” of a large dc voltage. A typical example is trying to measure the noise of a dc power
supply.
INV - There is almost always a control which lets you invert one channel. This can be used
along with the ADD function to subtract two voltages. This is necessary because the common
input (black lead of the oscilloscope cable) can only be connected to a 0V node. If channel A
has V1 + V2 and channel B has voltage V1 then the reading of channel A + (-channel B) = (V1
+ V2) + (-V1) = V2
Position - For each axis there is a control which lets you shift the electron beam. With this you
can set the zero voltage point to anywhere that is convenient for you.
Oscilloscope Inputs
The input of the oscilloscope can usually be modelled as a resistance and a parallel capacitance
(see figure 3). The resistance is usually 1MS but it and the capacitance can vary greatly. The
total or effective capacitance includes the oscilloscope circuitry (approx. 30 pF), cables
(approx. 30 pF/m) and stray capacitance. The resistance will draw current from the circuit while
the capacitance will add an RC time constant with its associated time delay, frequency response
and distortion of some waveforms.
The common connection (black lead or shield) at the input of the oscilloscope goes to the metal
case as the symbol by the input connector shows. Because of this, the common input can only
be connected to a 0V point in the circuit. Since the common inputs for both the A and B
channels are connected to the case, they are effectively shorted together.
Due to limitations in the amplifiers, the oscilloscope's frequency response is limited. The
manufacturer simply lists the half-power point for the oscilloscope without any external effects.
Half power is also called the -3dB point. At this point, the voltage has decreased to 70.7% of its
maximum. This means that only one-half of the maximum power would be dissipated in a
resistive load. Keep in mind that an oscilloscope that is rated at 20 MHz is usually only accurate
to 4 MHz for non-sinusoidal waveforms before distortion becomes a problem.
With ac coupling (figure 4), an oscilloscope has another series RC circuit. It acts like a high
pass filter (HPF). If you are viewing low frequency signals when ac coupled, not only will you
not be able to measure any dc offset, but you will also be removing some low frequency
information.
To obtain a stable picture on the CRT screen, the ramp waveform has to be in phase with the
signal that you want to observe. This is done with a triggering circuit. The triggering circuit
allows the oscilloscope to draw repeatedly the same waveform over and over by identifying the
same point on a repetitive waveform.
Not only do you have control over the starting point of the ramp, but the amount of time that the
ramp takes to reach its maximum voltage (the right hand side of the CRT screen) can be
adjusted with the timebase control. In essence, you have a “window”. You can move the
window to any point on a waveform with the triggering circuit and you can change the size of
the window with the timebase.
The time-base control allows you to set the time / division that the beams takes to scan across
When set to AUTO (automatic) triggering, the oscilloscope will always show a trace. However,
when you use a manual triggering mode (DC, AC), many strange things can happen. For
example, if the triggering voltage or level is set to +10V and the waveform never exceeds +5V,
the triggering circuit will never trigger and the screen will stay blank.
You may think that in a condition of no triggering, you would still have a bright dot on the
screen because the electron beam would go to its 'home' or undeflected position. Since the
oscilloscope is designed to work with a moving electron beam, a stationary beam can very
quickly 'burn' a hole in the phosphorous coating of the screen. To prevent this, there is a '
blanking' circuit which turns off the electron beam. Blanking occurs when there is no triggering
or when the electron beam is sweeping from the right edge back to the left side of the screen.
Time measurements are done the same way as voltage measurements. As long as the timebase
is calibrated you multiply the number of divisions by the number of seconds per division to get
the total time difference. Phase measurements are done by comparing the measured time to the
period of the waveform.
When in the chop mode (figure 8), the oscilloscope displays a little bit of channel A, then a
little bit of B, then A, then B ....during a single sweep of the electron beam. If you increase the
timebase to about 1:s/division, you can start to see the individual pieces as it chops between
one channel and the other channel.
In the alternate mode (figure 9), the oscilloscope will sweep the electron beam twice across the
screen. The first time it will draw the signal from channel A and the next time from channel B.
At very low timebase settings, you can see it draw one channel and then the other in successive
passes.
Note: When you use the alternate function, the two waveforms that you see are from different
points in time and the triggering circuit has to trigger twice.
The reason that you can see a non-flickering image on the screen is because the phosphorous
coating on the CRT has persistence. In essence, the phosphorous acts like a low pass filter and
averages several images that are drawn on the screen.
By viewing two signals at a time, you can measure relative time differences. By combining a
voltage and phase measurement (relative to the appropriate reference), you can measure a
phasor value.
With a two channel oscilloscope, you have the ability to trigger on each waveform and
electronically switch (chop or alt) between them as well. A block diagram of a oscilloscope has
now become as shown in figure 10.
Some oscilloscopes offer a way to alternately trigger as depicted in figure 11. When combined
with alternate displaying, you can stably display two waveforms of any frequency by alternately
showing each channel and triggering on the channel that is being drawn. This way, the
oscilloscope is acting like a two beam scope with both waveforms triggering at the same
voltage and slope. However, there is no way to know what the relationship is between one
waveform and the other when using alternate triggering.
If you have two waveforms that do not have the same frequency, it is still possible to show
them as two stable waveforms on a normal oscilloscope. In figure 11, you will notice that if the
triggering occurs at the 'X', both waveforms are in phase (ie. at the same phase each time the
timebase triggers). The condition for a stable display is not that two waveforms have to be of
exactly the same frequency, but that when they are triggered, they have to be in phase.
Or nAf1 = mAf2 where n, m are integers. That is not necessary, but it is sufficient. There are
many other ways to achieve a stable trace when you consider that the trigger circuit will wait for
the next triggering point. There is also a control on some oscilloscopes, called 'hold off', which
allows you to add a delay between the end of the trace being drawn and the time when the
triggering circuit starts to look for the next triggering point. That can be used to stabilize the
display under some circumstances.
Remember that all of this applies only for repetitive waveforms that are properly triggered. If
the triggering is not stable, or the waveform is not repetitive, you will see a constantly moving
image or several images offset and superimposed.
A slightly more complicated block diagram of an oscilloscope, with the typical functions
found in the laboratory, is illustrated in figure 12.
Accuracy
There are many factors affecting the accuracy of oscilloscope measurements.
There are errors due to the input channel voltage divider, timebase control, the use of
magnifiers, the accuracy to which the CRT deflection can be read, beam thickness, temperature
etc. The voltage divider error will be the same for all readings that are done on the same
timebase and voltage range, but may be different each time the range is changed. Measurements
over only two divisions can incur two to three times the error of those made over the centre
eight divisions.
If the phase angle is used in a trigonometric function, this error can be multiplied by the slope
of the function. Consider that the tangent of a 1% phase error entered at 85 degrees is much
worse (20%) than the same 1% error on a sine function (0.2%) at the same angle. To get a feel
for this look at the Taylor expansion of the trigonometric functions.
It is wise to consult the user manual for a particular instrument’s accuracy specifications.
The power on/off, trace rotation, intensity, focus and scale illumination
Trace rotation has to be checked with just a straight line across the screen of the oscilloscope.
CH-1 and CH-2 are the labels for the two vertical inputs.
Each channel has a position control, a range selection switch, a pull “x5” knob, a coupling
selector switch (AC/GND/DC), and an input connector.
Each channel Range switch (VOLTS/DIV) has a smaller knob in the middle of the Range
Selector Switch. And there is an arrow showing that the Range Selector Switch is in the
“CAL” position when rotated fully clockwise.
In the centre of the two channel sections is the channel selection switch. You can choose to
have CH-1, CH-2, both (DUAL), or ADD.
In addition, CH-2 has a “Polarity” switch. You push the polarity switch in to “INVERT” the
polarity of the signal being displayed. The “NORM” or out position is the normal position of
the polarity switch.
Horizontal Inputs
The “EXT” (HORIZONTAL IN) can be supplied a voltage directly via the connector at the
bottom right of the panel, or the Horizontal can be driven by an internal timebase circuit which
generates the voltage.
Timebase
In the timebase mode, the horizontal signal is from an internal source which changes linearly
with respect to time. Hence, the beam is deflected to give us a calibration of time for the
horizontal scale.
The position control, the pull “x5” magnifier switch, the time range selection switch, and the
range “CAL” knob all affect the X-axis of the display.
Trigger Sources
The TRIGGER SOURCE may be selected from one of three sources, CH-1, CH-2, or EXT.
Look at the bottom right of the control panel.
Calibration Source
The Calibration Source is an internal source, available on the oscilloscope.
Look at the bottom right side of the front panel. The output is labelled 0.3 V.
This is a 1000 Hz, square-wave ( the 50 % duty cycle is not accurate ).
Differential Measurements
An important application of the oscilloscope is differential measurements. Such measurements
are necessary because both vertical channels have one terminal connected to the chassis
common (ie single ended). To measure a floating (off ground) voltage you have to use the
“invert and add” feature of the oscilloscope. For example, in figure 14, to measure V1:
Channel A measures (V1 + V2) relative to ground while channel B measures V2 relative to
ground. By pushing the invert button you negate the voltage displayed on channel B. Then you
can add the channels together with the “ADD” display mode. The waveform now displayed is
(V1 + V2) + (- V2) = V1.
The frequency is then simply read with a frequency counter or the oscilloscope. Not by reading
the dial of the signal generator.
Risetime
The risetime indicates how quickly a circuit responds. The risetime is the time it takes a
waveform to go from 10% of the voltage range to 90% of the voltage range. This is in response
to a square wave and the output voltage must settle to a steady-state voltage (0% and 100%).
Most oscilloscopes have dotted lines on the graticule marking the 10% and 90% points to aid in
this measurement. Usually these dotted lines assume that 0% is the lowest line of the graticule
and 100% is the highest line. The measurement, as shown in figure 15, also includes the
risetime of the oscilloscope and the squarewave source.