High Performance Liquid Chromatography
High Performance Liquid Chromatography
chapter
High-Performance Liquid
Chromatography
S.S. Nielsen, Food Analysis, Food Science Texts Series, DOI 10.1007/978-1-4419-1478-1_28, 499
°c Springer Science+Business Media, LLC 2010
500 Part V • Chromatography
Schematic representation of a system for high-performance liquid chromatography (not drawn to scale).
28-1 Column(s) and detector may be thermostatted, as indicated by the dashed line, for operation at elevated
figure temperature.
28.2.2 Injector
The role of the injector is to place the sample into the
flowing mobile phase for introduction onto the col-
umn. Virtually all HPLC systems use valve injectors,
which separate sample introduction from the high-
pressure eluent system. With the injection valve in the
LOAD position (Fig. 28-2a), the sample is loaded into
an external, fixed-volume loop using a syringe. Elu-
ent, meanwhile, flows directly from the pump to the
column at high pressure. When the valve is rotated Valve-type injector. The valve allows the sam-
to the INJECT position (Fig. 28-2b), the loop becomes 28-2 ple loop to be (a) isolated from the pump elu-
figure
part of the eluent flow stream and sample is carried ent stream (LOAD position) or (b) positioned
onto the column. Such injectors are generally trouble in it (INJECT position). [from (9), used with
permission.]
free and afford good precision.
Chapter 28 • High-Performance Liquid Chromatography 503
Changing the loop allows different volumes to 28.2.3.1.2 Analytical Columns The most commonly
be injected. Although injection volumes of 10–100 µl used analytical HPLC columns are 10, 15, or 25-cm
are typical, both larger (e.g., 1–10 ml) and smaller long with an internal diameter of 4.6 or 5 mm (9).
(e.g., ≤2 µl) sample volumes can be loaded by utiliz- Short (3 cm) columns, packed with ≤3 µm particles,
ing special hardware. An important advantage of the are gaining popularity for fast separations; for exam-
loop valve design is that it is readily adapted to auto- ple, in method development or process monitoring. In
matic operation. Thus, automated sample injectors, or recent years, the use of columns with smaller inter-
autosamplers, may be used to store and inject large nal diameters (<0.5–2.0 mm), including wall-coated
numbers of samples. Samples are placed in uniform- capillary columns, has increased. The advantages of
size vials, sealed with a septum, and held in a (pos- using smaller diameter columns include a decreased
sibly refrigerated) tray. A computer actuated needle consumption of mobile phase, an increased peak con-
penetrates the septum to withdraw solution from the centration, increased resolution, and the ability to
vial, and a mechanically or pneumatically operated couple HPLC with mass spectrometry (MS) (13).
valve introduces it onto the column. Autosamplers Various names have been used for the reduced-
can reduce the tedium and labor costs associated with volume columns. Dorsey et al. (14) refer to columns
routine HPLC analyses and improve assay precision. with internal diameters of 0.5–2.0 mm as microbore,
However, because samples may remain unattended while packed or open tubular columns having inter-
for 12–24 h prior to automatic injection, sample stabil- nal diameters of <0.5 mm are termed microcolumns
ity is a limiting factor for using this accessory. or capillary columns (a capillary column is a narrow-
bore open tubular column, in which the inner surface
28.2.3 Column is coated with a thin layer of stationary phase). In
the case of the packed columns, the microbore or
28.2.3.1 Column Hardware microcolumns contain very small particle size pack-
An HPLC column is usually constructed of stainless ing material. Because of the extremely high operating
steel tubing with terminators that allow it to be con- pressures of these systems, they are often referred
nected between the injector and detector of the system to as ultra-HPLC (UHPLC). To achieve good perfor-
(Fig. 28-1). Columns also are made from glass, fused mance from microcolumns, it is essential to have an
silica, titanium, and polyether ether ketone (PEEK) HPLC system with very low dead-volume, so that
resin; the PEEK columns are essential for the high peak broadening outside the column does not destroy
pH, high salt concentrations necessary for the power- resolution achieved within the column. Pumps, injec-
ful ion exchange HPLC systems. Many types and sizes tors, and other hardware designed specifically for use
of columns are commercially available, ranging from with these columns are available from commercial
5 cm × 50 cm (or larger) preparative columns down to suppliers.
wall-coated capillary columns.
28.2.3.2 HPLC Column Packing Materials
28.2.3.1.1 Precolumns Auxiliary columns that pre-
cede the analytical HPLC column are termed pre- The development of a wide variety of column packing
columns. Short (≤5 cm) expendable columns, called materials has contributed substantially to the success
guard columns, often are used to protect the analytical and widespread use of HPLC.
column from strongly adsorbed sample components.
A guard column (or cartridge) is installed between the
injector and analytical column via short lengths of cap- 28.2.3.2.1 General Requirements A packing mate-
illary tubing (or a cartridge holder). They may be filled rial serves, first of all, to form the chromatographic
with either pellicular media (see Sect. 28.2.3.2.2) of the bed; however, in most modes of chromatography the
same bonded phase as the analytical column, or with column packing material serves as both support and
microparticulate (≤10 µm) packing material identi- the stationary phase. Requirements for HPLC column
cal to that of the analytical column. Microparticulate packing materials are good chemical stability, suffi-
guard columns are usually purchased as prepacked, cient mechanical strength to withstand pressure gener-
disposable inserts for use in a special holder, and ated during use, and the availability of a well-defined
cost much less than replacing an analytical column. particle size, with a narrow particle size distribution
A guard column (or cartridge) should be repacked or (10). Two materials that meet the above criteria are
replaced before its binding capacity is exceeded and porous silica and synthetic organic resins (see Sects.
contaminants pollute the analytical column. 28.2.3.2.2 and 28.2.3.2.3, respectively).
504 Part V • Chromatography
28.2.3.2.2 Silica-Based Column Packings Porous skeleton slowly dissolves in aqueous solutions, and
silica meets the above criteria quite well and can be the rate of this process becomes prohibitive at pH<2
prepared in a wide range of particle and pore sizes, and >8.
with a narrow particle size distribution. Both par- A pellicular packing material (Fig. 28-3b) is made
ticle size and pore diameter are important: Small by depositing a thin layer or coating onto the surface
particles reduce the distance a solute must travel of an inert, usually nonporous, microparticulate core.
between stationary and mobile phases, which facil- Core material may be either inorganic, such as silica, or
itates equilibration and results in good column effi- organic, such as poly(styrene-divinylbenzene) or latex.
ciencies (Chap. 27, Sect. 27.5.2.2.2). However, small Functional groups, such as ion-exchange sites, are then
particles also yield greater flow resistance and higher present at the surface only. The rigid core ensures good
pressure at equivalent flow rates. Spherical particles physical strength, whereas the thin stationary phase
of 3, 5, or 10-µm diameter are utilized in analytical provides for rapid mass transfer and favorable column
columns. One-half or more of the volume of porous efficiency.
silica consists of the pores (10). Use of the smallest
possible pore diameter will maximize surface area and 28.2.3.2.3 Porous Polymeric Column Packings Syn-
sample capacity, which is the amount of sample that thetic organic resins offer the advantages of good
can be separated on a given column. Packing materi- chemical stability and the possibility to vary interac-
als with a pore diameter of 50–100 Å and surface area tive properties through direct chemical modification.
of 200–400 m2 /g are used for low-molecular-weight Two major categories of porous polymeric packing
(<500 Da) solutes. For increasingly larger molecules, materials exist.
such as proteins and polysaccharides, it is necessary to Microporous or gel-type resins (Fig. 28-3c) are
use wider pore materials (pore diameter ≥300 Å), so comprised of crosslinked copolymers in which the
that internal surface is accessible to the solute (10). apparent porosity, evident only when the gel is in its
Bonded phases (Fig. 28-3a) are made by cova- swollen state, is determined by the degree of crosslink-
lently bonding hydrocarbon moieties to −OH groups ing. These gel-type packings undergo swelling and
(silanols) on the surface of silica particles (10, 16). contraction with changes in the chromatographic
Often, the silica is reacted with an organochlorosilane: mobile phase. Microporous polymers of less than ca.
8% crosslinking are not sufficiently rigid for HPLC use.
R1
Macroporous resins are highly crosslinked (e.g.,
R1
\ | \ | ≥50%) and consist of a network of microspheric
— Si — OH + C1 —Si—R3 —Si— O — Si—R3 + HC1 gel beads joined together to form a larger bead
/ | / |
R2 R2 (Fig. 28-3d). Large, permanent pores, ranging from
100 to 4000 Å or more in diameter, and large surface
[1] areas (≥100 m2 /g) are the result of interstitial spaces
between the microbeads (16). Rigid microparticulate
Substituents R1 and R2 may be halides or methyl poly(styrene-divinylbenzene) packing materials of the
groups. The nature of R3 determines whether the macroporous type are popular for HPLC use. They are
resulting bonded phase will exhibit normal-phase, stable from pH 1 to 14 and are available in a variety
reversed-phase, or ion-exchange chromatographic of particle and pore sizes. These resins can be used in
behavior. The main disadvantage of silica and silica- unmodified form for reversed-phase chromatography
based bonded-phase column packings is that the silica or functionalized for use in other HPLC modes.
Some types of packing materials utilized in HPLC. (a) Bonded-phase silica; (b) pellicular packing; (c) microporous
28-3 polymeric resin; (d) macroporous polymeric resin. [Adapted from (16), p. 621, by courtesy of Marcel Dekker, Inc.]
figure
Chapter 28 • High-Performance Liquid Chromatography 505
by the application of voltage across electrodes of the drugs. Decay of a radioactive nucleus leads to exci-
flow cell. This method is highly selective (nonreac- tation of a scintillator, which subsequently loses
tive compounds give no response) and very sensitive. its excess energy by photon emission. Photons are
A major application of electrochemical detection has counted by a photomultiplier tube and the number
been for the routine determination of catecholamines, of counts per second is proportional to radiolabeled
which are phenolic compounds of clinical importance analyte (9).
that are present in blood and tissues at very low lev- A chemiluminescent nitrogen detector (CLND)
els. The development of a triple-pulsed amperometric allows nitrogen-containing compounds, such as amino
detector, which overcame the problem of electrode acids, to be detected without using chemical derivati-
poisoning (accumulation of oxidized product on the zation (Sect. 28.2.4.7). This nitrogen-specific detection
electrode surface), has allowed electrochemical detec- system has been used to quantify caffeine in coffee and
tion to be applied to the analysis of carbohydrates soft drink beverages, and to analyze capsaicin in hot
(see Sect. 28.3.3.2.2). Pulsed electrochemical detection peppers (2).
also has excellent sensitivity for the quantification of
flavor-active alcohols, particularly terpenols (2).
Analytes that are ionized and carry a charge can 28.2.4.6 Coupled Analytical Techniques
be detected by measuring the change in eluent conduc-
To obtain more information about the analyte(s), elu-
tivity between two electrodes. Conductivity detection
ent from an HPLC system can be passed on to a second
has been used mainly to detect inorganic anions and
analytical instrument, such as infrared (IR), nuclear
cations and organic acids upon elution from weak
magnetic resonance (NMR), or MS [see Chaps. 23, 25,
ion-exchange columns. Its principal application has
and 26, respectively, or reference (6)]. The coupling of
been the basis of ion chromatography (Sect. 28.3.3.2.1).
spectrometers with liquid chromatography (LC) was
An excellent overview of electrochemical detection is
initially slow to gain application, due to many prac-
provided by Swedesh (8).
tical problems. For example, in the case of HPLC
with mass spectrometric detection (LC–MS), the liq-
uid mobile phase affected the vacuum in the MS. This
problem was addressed by the development of com-
28.2.4.5 Other HPLC Detectors mercial interfaces that allow solvent to be evaporated
so that only analyte is carried to the spectrometer. Two
Unfortunately, there is no truly universal HPLC detec-
commonly used interface techniques are discussed in
tor with high sensitivity. Thus, there have been many
detail by Harris (6). The use of microbore or capillary
attempts to find new principles that could lead to
HPLC columns with a low flow volume also facilitates
improved instrumentation. One interesting concept is
direct coupling of the two instruments (13). LC–MS
the evaporative light scattering detector. The mobile
systems continue to improve, and the applications
phase is sprayed into a heated air stream, evapo-
are expanding to nearly every class of relatively low
rating volatile solvents and leaving nonvolatile ana-
molecular weight compounds, including bioactives
lytes as aerosols. These droplets or particles can be
and contaminants.
detected because they will scatter a beam of light
(8). HPLC with light scattering detection has been
applied to the analysis of wheat flour lipids. Also,
28.2.4.7 Chemical Reactions
light scattering detectors are quite useful for the char-
acterization of polymers by size-exclusion chromatog- Detection sensitivity or specificity may sometimes be
raphy. Improvements in laser applications brought enhanced by converting the analyte to a chemical
about the development of low-angle laser light scat- derivative with different or additional characteristics.
tering (LALLS) and multi-angle laser light scattering An appropriate reagent can be added to the sample
(MALLS) detectors. With these detectors, there is no prior to injection (i.e., precolumn derivatization) or
need to evaporate the mobile phase, as the laser beam combined with column effluent before it enters the
is directed at the flow cell, and scattered laser light detector (i.e., postcolumn derivatization). Automated
is then monitored by photo detectors set at specific amino acid analyzers utilize postcolumn derivatiza-
angles to the cell. In MALLS there may be as many tion, usually with ninhydrin, for reliable and repro-
as 18 different photo detectors at discrete angles, each ducible analysis of amino acids. Precolumn derivati-
continuously collecting and analyzing the scattered zation of amino acids with o-phthalaldehyde or similar
light; from this data, the computer can determine the reagents permits highly sensitive HPLC determination
molecular weight of the eluting sample. of amino acids using fluorescence detection (Chap. 15,
Radioactive detectors are widely used for phar- Sect. 15.3.1.2). In addition, fractions may be collected
macokinetic and metabolism studies with radiolabeled after passing through the detector and aliquots of each
Chapter 28 • High-Performance Liquid Chromatography 507
fraction analyzed by various means, including chem- developing the gradient via control of the proportion-
ical/colorimetric assays, such as the Lowry protein ing pump systems. When combined with an autosam-
assay (Chap. 9, Sect. 9.2.5) or a total carbohydrate pler, the data station can carry out the entire operation,
assay (Chap. 10, Sect. 10.3.2). The results can then on hundreds of samples, in the absence of an operator,
be plotted and overlaid with the detector plot, yield- and, with networking, deliver the analysis file/report
ing very important information about the compounds to an office computer.
eluting in various peaks.
28-1
table Example Applications of HPLC in the Analysis of Various Food Constituents
strength refers to the way a solvent affects the migra- 28.3.2.2 Applications of Reversed-Phase
tion rate of the sample. Weak solvents increase reten- HPLC
tion (large k′ values) and strong solvents decrease
Reversed-phase has been the HPLC mode most used
retention (small k′ values).
for analysis of plant proteins. Cereal proteins, among
the most difficult of these proteins to isolate and char-
28.3.1.2 Applications of Normal-Phase HPLC acterize, are now routinely analyzed by this method
In the past, normal-phase HPLC was used for the (10). Both water- and fat-soluble vitamins (Chap. 11)
analysis of fat-soluble vitamins, although reverse can be analyzed by reversed-phase HPLC (2–5), and
phase is currently applied more frequently for these the availability of fluorescence detectors has enabled
analyses (see Table 28-1). Normal phase is currently researchers to quantitate very small amounts of the
used for the analyses of biologically active polyphe- different forms of vitamin B6 (vitamers) in foods and
nols from natural plant sources, such as grape and
cocoa. It is also used for the analysis of relatively
polar vitamins, such as vitamins A, D, E, and K
(see Chap. 11), and also natural carotenoid pigments,
which impart both color and health benefits to foods.
Highly hydrophilic species, such as carbohydrates (see
Chap. 10, Sect. 10.3.4.1), also may be resolved by
normal-phase chromatography, using amino bonded-
phase HPLC columns (10).
biological samples. Figure 28-4 shows the separation 28.3.3 Ion Exchange
of several of these vitamers in a rice bran extract
28.3.3.1 Stationary and Mobile Phases
achieved by reversed-phase ion-pair HPLC (17).
Reversed-phase ion-pair HPLC can be used to Packing materials for ion-exchange HPLC are usu-
resolve carbohydrates on C18 bonded-phase columns ally functionalized organic resins, such as sulfonated
(10), and the constituents of soft drinks (caffeine, or aminated poly(styrene-divinylbenzene) (Chap. 27,
aspartame, etc.) can be rapidly separated. Reversed- Sect. 27.4.3). Macroporous resins are most effective
phase HPLC using a variety of detection methods, for HPLC columns due to their rigidity and per-
including RI, UV, and light scattering, has been manent pore structure. Pellicular packings also are
applied to the analysis of lipids (2–5,10). Antioxidants, TM
utilized, particularly in the CarboPac (Dionex)
such as butylated hydroxyanisole (BHA) and buty- series, in which the nonporous, latex resin beads are
lated hydroxytoluene (BHT), can be extracted from dry coated with functionalized microbeads. The mobile
foods and analyzed with simultaneous UV and flu- phase in ion-exchange HPLC is usually an aqueous
orescence detection (3). Phenolic flavor compounds buffer, and solute retention is controlled by changing
(such as vanillin) and pigments (such as chlorophylls, mobile phase ionic strength and/or pH. Gradient elu-
carotenoids, and anthocyanins) are also easily ana- tion (gradually increasing ionic strength) is frequently
lyzed (2–5,10). A typical chromatogram of carotenoids employed.
present in a carrot extract is shown in Fig. 28-5.
Reversed-phase ion-pair chromatography also is used
for the separation of synthetic food colors (e.g., FD&C 28.3.3.2 Applications of Ion-Exchange HPLC
Red No. 40 and FD&C Blue No. 1) (5). Ion-exchange HPLC has many applications, ranging
from the detection of simple inorganic ions, to analysis
of carbohydrates and amino acids, to the preparative
purification of proteins oligosaccharides.
infant formula; and trace metals, phosphates, and 28.3.4 Size Exclusion
sulfites in foods. Figure 28-6 illustrates the simulta-
neous determination of organic acids and inorganic Size-exclusion chromatography (SEC) fractionates
anions in coffee by ion chromatography. solutes solely on the basis of size, with larger
molecules eluting first. Due to the limited separation
volume available in this chromatographic mode, as
explained in Chap. 27 (Sect. 27.4.4), the peak capac-
28.3.3.2.2 Ion Exchange Chromatography of Car- ity of a size-exclusion column is relatively small. Thus,
bohydrates and Proteins Both cation- and anion- the “high-performance” aspect of HPLC is not really
exchange stationary phases have been applied to applicable in the case of size exclusion. The main
HPLC of carbohydrates. The advantage of separating advantage gained from use of small particle pack-
carbohydrates by anion exchange is that retention and ing materials is speed. Relatively small amounts of
selectivity may be altered by changes in eluent com- sample can be analyzed or separated and collected in
position. Carbohydrate analysis has benefited greatly ≤60 min, compared to ≤24 h separations using clas-
by the development of a system that involves anion- sic low pressure systems (10). A second advantage is
exchange HPLC at high pH (≥12), and detection that the sample concentration is higher and the rela-
by a pulsed amperometric detector (PAD). Pellicular tive volume is lower, so there is much less eluent to
column packings (see Sect. 28.2.3.2.2), consisting of remove.
nonporous latex beads coated with a thin film of strong
anion exchanger, provide the necessary fast exchange,
high efficiency, and resistance to strong alkali. These 28.3.4.1 Column Packings and Mobile
systems may be used in a variety of applications, Phases
from routine quality control to basic research. One Size-exclusion packing materials or columns are
common application is the determination of oligosac- selected so that matrix pore size matches the molecular
charide distributions in corn syrups and other starch weight range of the species to be resolved. Prepacked
hydrolysates (Fig. 28-7). columns of microparticulate media are available in a
Amino acids have been resolved on polymeric wide variety of pore sizes. Hydrophilic packings, for
ion exchangers for more than 40 years (see Chap. 15, use with water-soluble samples and aqueous mobile
Sect. 15.3.1.2). Ion exchange is one of the most effec- phases, may be surface-modified silica or methacrylate
tive modes for HPLC of proteins and, recently, has resins. Poly(styrene-divinylbenzene) resins are use-
been recognized as valuable for the fractionation of ful for nonaqueous size-exclusion chromatography of
peptides. synthetic polymers.
Chapter 28 • High-Performance Liquid Chromatography 511
The mobile phase in this mode is chosen for has been shown to be a rapid, one-step method for
sample solubility, column compatibility, and minimal assessing soybean cultivars on the basis of protein con-
solute–stationary phase interaction. Otherwise, it has tent (proteins in the extracts of nondefatted flours from
little effect on the separation. Aqueous buffers are used five soybean cultivars were separated into six common
for biopolymers, such as proteins and polysaccharides, peaks, and cultivars could be identified by the per-
both to preserve biological activity and to prevent cent total area of the fifth peak). A size-exclusion liquid
adsorptive interactions. Tetrahydrofuran or dimethyl- chromatographic method also has been applied to the
formamide is generally used for size-exclusion chro- determination of polymerized triacylglycerols in oils
matography of other polymer samples, to ensure and fats (2).
sample solubility.
28.3.5 Affinity
28.3.4.2 Applications of High
Performance SEC Affinity chromatography is based on the principle
that the molecules to be purified can form a selec-
Hydrophilic polymeric size-exclusion packings are tive but reversible interaction with another molecular
used for the rapid determination of average molecu- species that has been immobilized on a chromato-
lar weight and molecular weight range of polysac- graphic support. Although almost any material can
charides, including amylose, amylopectin, and other be immobilized on a suitably activated support, the
soluble gums such as xanthan, pullulan, guar, and major ligands are proteins, including lectins (Chap. 15,
water-soluble cellulose derivatives. Molecular weight Sect. 15.2.3.2.2), nucleic acids, and dyes (Chap. 27,
distribution can be determined directly from high per- Sect. 27.4.5). Affinity chromatography is used to purify
formance size-exclusion chromatography, if LALLS many glycoproteins. Affinity chromatography using
or MALLS is used for detection (8, 10). The appli- immobilized folate-binding protein is an effective tool
cation of aqueous size-exclusion chromatography to in purifying sample extracts for HPLC analysis of
two commercially important polysaccharides, xanthan folates in foods [e.g., reference (19)].
and carboxymethyl cellulose, is discussed in detail in
reference (8).
SEC analysis has been useful to better understand
numerous food components and systems. SEC analy- 28.4 SUMMARY
sis of tomato cell-wall pectin from hot-and cold-break
tomato preparations (Fig. 28-8) showed that the cell- HPLC is a chromatographic technique of great ver-
wall pectin was not differentially degraded by the satility and analytical power. A basic HPLC system
different processing procedures. Size-exclusion HPLC consists of a pump, injector, column, detector, and data
system. The pump delivers mobile phase through the
system. An injector allows sample to be placed into
the flowing mobile phase for introduction onto the
column. The HPLC column consists of stainless steel
or polymer hardware filled with a separation pack-
ing material. Various auxiliary columns, particularly
guard columns, may be used prior to the analytical col-
umn. Detectors used in HPLC include UV-Vis absorp-
tion, fluorescence, RI, electrochemical, and light scat-
tering, as well as coupled analytical systems, such as a
mass spectrometer. Detection sensitivity or specificity
sometimes can be enhanced by chemical derivatiza-
tion of the analyte. Computer-controlled data station
systems offer data collection and processing capabili-
ties, and can run the instrument when an automated
system is needed. A broad variety of column packing
Analysis of tomato cell-wall pectin from materials have contributed greatly to the widespread
28-8 hot- and cold-break tomato preparations by use of HPLC. These column packing materials may
figure
size-exclusion chromatography. The solid lines be categorized as silica-based (porous silica, bonded
are from a refractive index detector response.
The lines with markers result from a postchro-
phases, pellicular packings) or polymeric (microp-
matography analysis. Aliquots of collected orous, macroporous, or pellicular/nonporous). The
fractions were analyzed by a colorimetric success of silica-based bonded phases has expanded
chemical assay that is specific for pectic sugars. the applications of normal-phase and reversed-phase
512 Part V • Chromatography