Good Practice Guide For Improving Accuracy of Dissolved Oxygen Measurements
Good Practice Guide For Improving Accuracy of Dissolved Oxygen Measurements
Abstract: Dissolved oxygen concentration is a key parameter for characterizing natural and
wastewaters and for assessing the global state of the environment in general. The decrease of
dissolved oxygen levels in the world’s oceans, which is becoming increasingly obvious, is
expected to have an impact on the whole ecosystem of the Earth, including the carbon cycle,
the climate, etc. Dissolved oxygen measurements by sensors are often deemed easy
measurements by routine laboratories. In reality, the physical and chemical processes
underlying the measurements are complex and these measurements are not at all as robust as
often considered.
Based on the analysis of the results of the intercomparison a set of tools and
recommendations are given to the participants of how to improve the quality of their results.
Table of contents
1. Introduction .......................................................................................................................3
2.8 Using the proficiency test results for checking self-declared measurement uncertainty 8
2.9 Using the proficiency test results for measurement uncertainty estimation .............10
Acknowledgments ................................................................................................................15
References ............................................................................................................................16
Good Practice Guide… 3
1. Introduction
Dissolved oxygen (below DO) content in natural waters is an indispensable quantity whenever data are
collected for investigations of nature from a hydrobiological, ecological or environmental protection
viewpoint [1]. A sufficient concentration of DO is critical for the survival of most aquatic plants and
animals as well as for waste water treatment [2]. DO concentration is a key parameter for characterizing
natural and wastewaters and for assessing the global state of the environment in general [3]. The
decrease of DO levels in the world’s oceans, which is becoming increasingly obvious [3–6], is expected
to have an impact on the whole ecosystem of the Earth, including the carbon cycle [7], the climate [3,5],
etc. The current understanding of the dynamics of the processes and their interrelation is still far from
sufficient. Measurement and monitoring of dissolved oxygen concentration is essential for improving
that understanding.
The majority of dissolved oxygen measurements are made with the use of amperometric [8] and
optical sensors [9]. The performance of these sensors has dramatically improved over the years [10].
Nevertheless, accurate DO measurement with sensors is not easy because it is influenced by numerous
uncertainty sources [8,10,11]. Therefore, the agreement between the sensor-based DO data from
different laboratories has long been an issue and has caused a negative perception of the data using
sensors in the oceanography community. Because of this, the recent issue of the World Ocean Atlas [12]
was compiled taking into account only DO concentrations obtained with chemical titration methods
(first of all the Winkler titration method, WM) and rejecting all sensor-based data. A similar decision
was made in a recent study of DO decline rates in coastal oceans [6]. Yet, oceanographers need large
amounts of DO data, collected continuously around the clock during lengthy time periods (months),
often far away from any human settlement. Only sensor-based automatic measurements can satisfy this
need. It is thus important to make every effort to underpin the quality of sensor-based measurements.
DO concentration is a highly unstable parameter of water. Thus preparation of reference solutions
that are stable for extended periods of time is almost impossible. This complicates the standardization of
the measurements and preparation of certified reference materials (below CRM). This is as true for
Winkler titration as it is for sensor measurement of DO concentration. Also Nordtest TR 537 [13]
pointed out that there is a “long-term” uncertainty component from the variation in the calibration,
which is hard to measure, as no stable reference material or CRM is available for DO measurement. The
method suggested in Nordtest TR 537 for Winkler titration was to calibrate the same thiosulfate solution
several times during a few days and use the variation between the results for the uncertainty estimation.
Nevertheless, the highly important long-term variation component is only estimated by educated guess in
Nordtest TR 537, which cannot be considered fully satisfactory and the bias component of uncertainty is
not addressed at all.
However, in this guide we present a tool for laboratories: A robust method to prepare an in-house
reference material—water saturated with air—for DO measurement, which will help laboratories to
estimate their measurement bias. This will enable the use of the increasingly popular Nordtest
approach [13,14] and the MUkit software [15] for measurement uncertainty evaluation. In the Nordtest
approach, among other things, control sample and routine sample replicate results are utilized for
uncertainty evaluation. Also intercomparison measurement results may be exploited for uncertainty
estimation with the Nordtest approach.
Good Practice Guide… 4
results of this intercomparison. At the same time carrying out calibration is very easy, so that frequent
calibration is not a problem even at routine laboratory level.
Using an in-house reference material/environment (as is presented in Chapter 2.4) the calibration
(usually 20 or 25 °C) is performed and then measurements in a wider temperature range (example ± 5,
±10 and ± 15 °C, in the range 1 - 40 °C, depending the sensor type) are also performed. All measured
values are compared to the saturation values (according to the Chapter 2.4 equations). If the
measurements are always carried out at low or high temperatures (example below or over 20 °C), also
the calibration should preferably be carried out at near or even at the same temperature.
a) an artificially saline solution is prepared (for example using a KCl), which is saturated with air
and the oxygen content is measured with the sensor and the salinity is calculated or the measured
as conductivity/salinity. The measured DO value is compared with the theoretical value of the
saturation value (in that particular salinity).
b) the real sea water or ocean sample is measured with DO sensor and the results are compared
with the Winkler titration result.
tests and found that the equilibrium saturation concentration can change in the range of
±0.03 mg/L if the diameters of the bubbles change in the range of 1.3–10 mm (at a temperature of
20 °C this means a relative difference of 0.33%). The bubble diameter was estimated from
photographs of the bubbles taken against a ruler positioned in the water.
5. Measure the temperature of the water (Step 2) and the air pressure next to the water bath as
accurately as is possible for your laboratory.
6. Calculate the dissolved oxygen saturation concentration (CO2_saturation, in Table 1 and in the main
text this is equal to Cref) using Equations (6–8) or check the reference value from the ISO
5814:2012 Tables A.4 and A.5 in [19]. ISO 5814 gives the solubility of oxygen in water vs.
temperature and atmospheric pressure with 1 °C temperature interval and for air pressures such
as 967 hPa, 1000 hPa, 1013 hPa and 1033 hPa.
A A A A
C O2_saturation exp A1 2 32 43 54 W (6)
T T T T
where T is the temperature [K] of the water; A1, A2, A3, A4, A5 are constants and the values are
−1.393441 × 102, 1.575701 × 105, −6.642308 × 107, 1.2438 × 1010, −8.621949 × 1011,
respectively [20, 21]. W is the pressure correction factor [19]:
p pH2O
W
pn pH2O_100% (7)
where B1, B2, B3 are constants and the values are 11.8571, −3840.7, −216961 respectively.
An example of Excel spreadsheet for the calculation of the DO concentration in saturation
conditions is given in Appendix Tables A1 and A2.
7. In order to achieve the DO concentration in the reference medium with a k = 2 expanded
uncertainty of 0.2 mg/L or lower, a measuring temperature with k = 2 expanded uncertainty in the
range of 0.1 to 0.2 °C is safe. Sufficient k = 2 uncertainty for air pressure measurement is 500 Pa.
Good Practice Guide… 7
Scheme 1. The experimental setup for the preparation of dissolved oxygen (DO) in-house
reference material.
Second bath
Heating system
Cooling
5 cm
system
Based on the previous proficiency tests, the results of optical and electrochemical sensors generally
deviated from each other so that the results of electrochemical oxygen sensors were lower than the
results of optical oxygen sensors [18, 24]. Two possible reasons can be envisaged for the behavior of the
electrochemical sensors. Firstly, it is possible that the sensors’ parameters have drifted during the time
period from the last calibration to the intercomparison. Such drift almost always leads to lowering of the
values (not increasing), which is also observed here. Secondly, all electrochemical DO sensors have
some flow dependency because they consume oxygen at the membrane surface. Therefore the
electrochemical sensors need water movement and if this is not sufficient then lowered readings are
observed. In slow-moving water, mechanical stirring is necessary for most models.
In connection to Fieldoxy2014 proficiency test [18], tests were carried out for checking the flow
dependency of DO measurement results on water flow rate in the surface of the sensor (sensor type
SBE43; Seabird). It was noticed that flow velocities ca. 6-14 cm/s yielded DO results within 1.5% at 8.9
mg/l concentration level. If the flow rate of water was 0 cm/s, then the DO results were decreased
dramatically resulting DO concentration ca 65% lower than compared to flow speed of 14 cm/s.
Before DO measurements, the operator of electrochemical sensor should check that the movement of the
water is sufficient for the purpose.
2.8 Using the proficiency test results for checking self-declared measurement
uncertainty
If proficiency test provider reports z- and zeta-scores to the participants, this information is very useful,
when participant checks, if the self-declared measurement uncertainty is realistic or not.
zeta = ( x X ) / ulab
2
uc2 , where
The bias can be calculated using e.g. proficiency test results. The Nordtest handbook [13] suggests
having at least six different PT results. As the use of PT results for bias estimate is inferior to use of
certified reference material for same purpose, the participants should also consider setting up facility for
production of “in-house” reference material, water saturated with air, for dissolved oxygen
determination. This is described in detail in Sections 2.2 and 3 and references [17, 23].
Titration measurement:
1. Analyze at least 10 samples as duplicates.
2. Denoting the pairs of DO values measured in the natural water samples i with Ri,1 and
Ri,2, calculate their absolute difference Di (Equation 9).
| | (9)
3. Calculate the average difference, Dmean, of all individual Di values. (Equation 10)
∑
(10)
where M denotes the number of sample pairs.
4. Calculate the standard deviation as uRw = Dmean/1.128 (Equation 11), where the factor
1.128 accounts for converting the mean difference to the standard deviation when two
replicates are used for Di determination [13, 14].
Estimation of systematic component (ub):
Since several PT results are used for bias estimate,
1. Calculate Root Mean Square of bias (RMSb)
∑(𝑏 i )2
𝑀𝑆b =
𝑀 (12)
where bi is the bias estimate obtained from individual PTi measurement results, i=1, 2
and 3 (in this PT), M is the number of different PT results used.
2. Take also into account the average of the uncertainties of the assigned values. ( ̅ cref).
3. Calculate ub using Equation 13
1
This approach should be applied with caution, because it only gives the within-day variation (repeatability) for sampling
and measurement, and there will also be a “long-term” uncertainty component from the variation in the calibration. For DO
titration method Nordtest TR537 [13] suggests to calibrate the same thiosulphate solution several times during a few days
time, and use the variation between the results as additional component for reproducibility in calibration.
Good Practice Guide… 11
b = 𝑀𝑆b2 + ̅cref
2
(13)
Apply coverage factor of “2” for calculating the expanded measurement uncertainty (Equation 15)
(15)
12
Table 1. Example of Excel® spreadsheet for the calculation of measurement uncertainty of applying the
results of the proficiency test.
Good Practice Guide… 13
Figure 1. Photo of the experimental setup of creating the in-house reference material.
a
In the front of the photo: thermostat with the bath for the reference medium, the second saturation bottle and
the stirrer motor. The sensors are immersed concentrically through the lid (cover) of the reference medium
bath. The readout devices are on the right side of the thermostat. The temperature sensors are inside the foam
(shown by an arrow).
The saturation conditions were obtained as follows. Air-saturated MilliQ water was used as the
reference medium (equilibrium saturation medium). The pressure, humidity and temperature of the air
used for saturation were controlled and taken into account. The saturation medium was created in a
modified (a second bath and a mechanical stirrer were added) thermostat CC2-K12 (Peter Huber
Kältemaschinenbau GmbH, Germany) in MilliQ water with overall volume 3.9 L (Scheme 1). The obtained
temperature variability was lower than 0.01 C (expressed as standard deviation). The air used for
saturation of the reference medium was taken from the air inlet situated on the roof of the building. The
air used for saturation was first saturated with water to achieve relative humidity of 100% for the air. The
air flow velocity during calibration was around 1 dm3 min−1. The ordinary aquarium spray was used for
bubbling (at a depth of 13 cm). The estimated diameter of the bubbles was between 0.8 and 1.8 mm.
Good Practice Guide… 14
The measured environment was stirred using a four-bladed stirrer with constant speed (160 rpm).
Thus the DO probes of the participants were arranged concentrically in the bath and were immersed
approximately to the same depth for achieving the same velocity of water movement in the location of
each sensor. According to our experience over the years [16], this setup permits achieving the best
possible uniformity of the measurement conditions between the participants, and the differences
between the DO concentrations in the vicinity of different sensors are negligible. Stirrer dimensions and
its location in the reference medium are also given in Scheme 1.
Optical sensors are in general more robust and require less skill in usage than the amperometric
sensors. The latter need more maintenance (replacement of the membrane and electrolyte
solution) and are more sensitive to stirring of the solution and temperature.
The sensors should be regularly calibrated.
The zero current of the sensors should periodically be checked.
Use of “in-house” reference material and routine sample replicate results together with control
charts enable laboratories to estimate measurement uncertainty of the DO measurement with the
Nordtest approach [13] more easily and more reliably than before. Näykki et al. recently
published a paper presenting software support for the measurement of uncertainty evaluation
based on the Nordtest approach [15].
Control charts should be set up for daily quality control
Participation in interlaboratory comparisons or proficiency tests may be exploited to check if the
self-declared uncertainty estimate is realistic [26].
Participation in interlaboratory comparisons or proficiency tests may be exploited in estimation of
measurement uncertainty.
Acknowledgments
References
1. Nagy, S.A.; Dévai, G.Y.; Grigorszky, I.; Schitchen, C.S.; Tóth, A.; Balogh, E.; Andrikovics, S. The
measurement of dissolved oxygen today—Tradition and topicality. Acta Zool. Acad. Sci. Hung.
2008, 54, 13–21.
2. Hobbs, J.P.A.; McDonald, C.A. Increased seawater temperature and decreased dissolved oxygen
triggers fish kill at the Cocos (Keeling) Islands, Indian Ocean. J. Fish Biol. 2010, 77, 1219–1229.
3. Keeling, R.F.; Garcia, H.E. The change in oceanic O2 inventory associated with recent global
warming. Proc. Natl. Acad. Sci. USA 2002, 99, 7848–7853.
4. Shaffer, G.; Olsen, S.M.; Pedersen, J.O.P. Long-term ocean oxygen depletion in response to carbon
dioxide emissions from fossil fuels. Nat. Geosci. 2009, 2, 105–109.
5. Keeling, R.F.; Körtzinger, A.; Gruber, N. Ocean Deoxygenation in a Warming World. Annu. Rev.
Marter. Sci. 2010, 2, 199–229.
6. Gilbert, D.; Rabalais, N.N.; Diaz, R.J.; Zhang, J. Evidence for greater oxygen decline rates in the
coastal ocean than in the open ocean. Biogeosci. Discuss. 2009, 6, 9127–9160.
7. Joos, F.; Plattner, G.-K.; Stocker, T.F.; Körtzinger, A.; Wallace, D.W.R. Trends in marine
dissolved oxygen: Implications for ocean circulation changes and the carbon budget. EOS Trans.
AGU 2003, 84, 197–204.
8. Helm, I.; Jalukse, L.; Leito, I. Measurement uncertainty estimation in amperometric sensors: A
tutorial review. Sensors 2010, 10, 4430–4455.
9. McDonagh, C.; Burke, C.S.; MacCraith, B.D. Optical chemical sensors. Chem. Rev. 2008, 108,
400–422.
10. Johnson, K.S.; Needoba, J.A.; Riser, S.C.; Showers, W.J. Chemical sensor networks for the aquatic
environment. Chemical sensor networks for the aquatic environment. Chem. Rev. 2007, 107, 623–
640.
11. Jalukse, L.; Leito, I. Model-Based measurement uncertainty estimation in amperometric dissolved
oxygen concentration measurement. Meas. Sci. Technol. 2007, 18, 1877–1886.
12. Garcia, H.E.; Locarnini, R.A.; Boyer, T.P.; Antonov, J.I. World Ocean Atlas 2005, Volume 3:
Dissolved Oxygen, Apparent Oxygen Utilization, and Oxygen Saturation; Levitus, S., Ed.; NOAA
Atlas NESDIS 63; US Government Printing Office: Washington, DC, USA, 2006.
13. Magnusson, B.; Näykki, T.; Hovind, H.; Krysell, M. Handbook for Calculation of Measurement
Uncertainty in Environmental Laboratories, 3.1 ed.; Nordtest TR 537; Nordic Innovation: Oslo,
Norway, 2012.
14. International Standard Organization (ISO). Water Quality—Estimation of Measurement
Uncertainty based on Validation and Quality Control Data; ISO 11352:2012; ISO: Geneva,
Switzerland, 2012.
15. Näykki, T.; Virtanen, A.; Leito, I. Software support for the Nordtest method of measurement
uncertainty evaluation. Accred. Qual. Assur. 2012, 6, 603–612.
16. Jalukse, L.; Vabson, V.; Leito, I. In situ interlaboratory comparisons for dissolved oxygen
concentration and pH. Accred. Qual. Assur. 2006, 10, 562–564.
Good Practice Guide… 17
17. Jalukse, L.; Helm, I.; Leito, I. Dissolved Oxygen Concentration Interlaboratory Comparison
Measurement; ESTDO-2012 Final Report; Available online: http://www.ut.ee/katsekoda/ILC/
O2_2012/ESTDO-2012_final_report.pdf (accessed on 8 April 2013).
18. Leivuori, M.; Näykki, T.; Leito I.; Helm I.; Jalukse L.; Kaukonen L.; Hänninen P.; Ilmakunnas M.,
Field measurement intercomparison, Field measurements of dissolved oxygen concentration,
Reports of Finnish environment institute 24/2014, Finnish environment institute, Helsinki, Finland,
2014
19. International Standard Organization (ISO). Water Quality—Determination of Dissolved
Oxygen—Electrochemical Probe Method; ISO 5814:2012; ISO: Geneva, Switzerland, 2012.
20. Benson, B.B.; Krause, D., Jr. The concentration and isotopic fractionation of gases dissolved in
freshwater in equilibrium with the atmosphere. 1. Oxygen. Limnol. Oceanogr. 1980, 25, 662–671.
21. Mortimer, C.H. The oxygen content of air-saturated freshwaters over ranges of temperature and
atmospheric pressure of limnological interest. Mitt. Int. Ver. Theor. Angew. Limnol. 1981, 22, 1–23.
22. Hovind, H.; Magnusson, B.; Krysell, M.; Lund, U.; Mäkinen, I. Internal Quality Control, 4th ed.;
Nordtest Technical Report 569; Nordic Innovation: Oslo, Norway, 2011. ‘
23. Näykki T,; Jalukse L,; Helm I,; Leito I,; Dissolved oxygen concentration interlaboratory
comparison: What can we learn?, Water, 2013, 5:420–442.
24. Leivuori, M.; Björklöf, K.; Näykki, T.; Väisänen, R.; Laboratorioiden välinen pätevyyskoe 5/2013,
proficiency test for field measurements of oxygen, temperature, pH and electrical conductivity (In
Finnish) Suomen ympäristökeskuksen raportteja 30/2013, Finnish environment institute, Helsinki,
Finland, 2013
25. Helm, I.; Jalukse, L.; Leito, I. A highly accurate method for determination of dissolved oxygen:
Gravimetric Winkler method. Anal. Chim. Acta 2012, 741, 21–31.
26. Analytical Method Committee. Is My Uncertainty Estimate Realistic? AMC Technical Brief No. 15;
Available online: http://www.rsc.org/images/realistic-estimate-technical-brief-15_tcm18-214874.pdf
(accessed on 8 April 2013).