P R O D U C T DATA

Piezoelectric Charge Accelerometer Types 4371 and 4371-V

Uses
• General purpose vibration testing and analysis
• High-frequency measurements
• Measurements in high-temperature environments

Features
• High sensitivity
• High resonance frequency

Characteristics
Description This piezoelectric accelerometer may be treated as a charge
Type 4371 DeltaShear Unigain* accelerometer. It features a source. Its sensitivity is expressed in terms of charge per unit
10 – 32 UNF-2A side connector and a 10 – 32 UNF-2B threaded acceleration (pC/ms–2, pC/g).
hole for mounting. Type 4371-V† has the same specifications
and long-term stability as Type 4371, but it has a relaxed The DeltaShear design consists of three piezoelectric elements
sensitivity tolerance. and three seismic masses arranged in a triangular configuration
around a triangular centre post. They are held in place by a
Fig. 1 Dimensions of Type 4371
clamping ring that isolates the configuration from the base. The
Spanner, ‡ 13.5 ring also prestresses the piezoelectric elements to give a high
14.0 across flats degree of linearity. This design provides a high sensitivity-to-
Recommended mass ratio, a relatively high resonance frequency and high
clearance for plug, isolation from base strains and temperature transients.
19.6

15.0 (min. 12.0)

The piezoelectric element is a PZ 23 lead zirconate titanate


7.2

element, and the housing material is titanium.


2.0 Fastening thread,
3.7

10–32 UNF-2B
depth 3.2 Calibration
Miniature coax. 5.5 ‡ 13.5
connector, Recess, The sensitivity is calibrated to a convenient value such as 1, 3.16
10–32 UNF-2A
Mounting surface
‡ 6.4, depth 0.5 or 31.6 pC/ms– 2 for Unigain accelerometers. The sensitivity
180089
given in the calibration chart has been measured at 159.2 Hz
All dimensions in millimetres
with 95% confidence level, using the coverage factor k = 2.

Fig. 2 Frequency response curves for Type 4371


Individual Frequency Response Typical High-frequency Response
Sensitivity [%] Phase [Degrees] Sensitivity [dB] Phase [Degrees]
Amplitude 30 Amplitude
10 Phase 10 Phase

20

0 0 10
10
0 0
–10
–10 –10
040040/1

–10 –20

5 10 20 50 100 200 500 1k 2k 5k 10k [Hz] 1 2 5 10 20 50 100 [kHz]

* Unigain: The individual measured sensitivity is within ± 2% of the specified sensitivity


† V-type: The individual measured sensitivity is within ± 15% of the specified sensitivity
Specifications – Charge Accelerometer Types 4371 and 4371-V Ordering Information

Type No. 4371 4371-V Type 4371


includes the following accessories:
General • Carrying box
g 11 • Calibration chart
Weight • AO-0038: Low-noise coaxial cable, 10 – 32 UNF connectors,
oz 0.39
length 1.2 m
pC /ms– 2 1 ± 2% 1 ± 15% • 10 – 32 UNF threaded steel stud, length 12.7 mm
Charge Sensitivity (at 159.2 Hz)
pC /g 9.8 ± 2% 9.8 ± 15% Type 4371-V
includes the following accessories:
Frequency Range (± 10% limit) Hz 0.1 to 12600
• Carrying box
Mounted Resonance Frequency kHz 42 • Calibration chart
–2 • 10 – 32 UNF threaded steel stud, length 12.7 mm
Max. Transverse Sensitivity (at 30 Hz, 100 ms ) % <4
Transverse Resonance Frequency kHz 15
Optional Accessories
Max. Operational Continuous Sinusoidal kms– 2 60 *
AO-0038-x-yyy Low-noise coaxial cable, 10 – 32 UNF
Acceleration (peak) g 6000 connectors, 250 °C (482 °F)
Electrical AO-0122-x-yyy* Super low-noise cable, 10 – 32 UNF
Residual Noise Level (measured with NEXUS mms– 2 2.4 connectors, 250 °C (482 °F)
Type 2692-001 in the specified frequency range) mg 0.24 AO-0231-x-yyy* Super low-noise cable, 10 – 32 UNF to
TNC, 180 °C (356 °F)
Capacitance (excluding cable) pF 1100
Min. Leakage Resistance (at 20 °C) GΩ 20 AO-1382-x-yyy* Flexible double-screened coaxial cable,
10 – 32 UNF connectors, 250 °C (482 °F)
Environmental
DB-0544 Probe with round tip, 10 – 32 UNF
°C – 74 to + 250
Operating Temperature Range JJ-0207 Plug adaptor, 10 – 32 UNF to TNC
°F – 101 to + 482 (female)
Temperature Coefficient of Sensitivity %/ °C 0.05* JP-0162 Plug adaptor, 10 – 32 UNF to TNC (male)
– 2/ °C
Temperature Transient Sensitivity ms 0.4 QA-0013 Hexagonal key for 10 – 32 UNF studs
(3 Hz Low. Lim. Freq. (– 3 dB, 6 dB/octave)) g/ °F 0.02 QA-0029 Tap for 10 – 32 UNF thread
Base Strain Sensitivity ms– 2/με 0.02 UA-0078 Accelerometer accessory set
(at 250 με in the base plane) g/με 0.002
UA-0553 Mechanical filter (set of five)

© Brüel & Kjær. All rights reserved.


ms– 2/ T 4
Magnetic Sensitivity (50 Hz, 0.038 T) UA-0642 Mounting magnet and two insulating
g/ kG 0.04 discs
kms– 2 200 UA-0866 Cementing stud, 10 – 32 UNF, dia.
Max. Non-destructive Shock (± peak)
g 20000 14 mm (set of 25)
Mechanical YG-0150 Steel stud, double-ended with flange,
10 – 32 UNF, length 5.3 mm
Housing Material Titanium ASTM Grade 2
YJ-0216 Beeswax for mounting
Piezoelectric Sensing Element PZ 23
YP-0080 Probe with sharp tip, 10 – 32 UNF
Construction DeltaShear
YP-0150 Insulated stud, fully threaded,

2018-08
Sealing Welded
10 – 32 UNF, length 13 mm
Electrical Connector 10 – 32 UNF-2A
YQ-2960 Set screw, 10 – 32 UNF 1/ 2 (12.8 mm)
10 – 32 UNF-2B × 3.2 mm
Mounting YQ-2962 Set screw, 10 – 32 UNF 5 /16 (7.7 mm)
threaded hole
Type 4294 Calibration Exciter BP 2036 – 14
Max. 3.5 (31)
Mounting Torque Nm (lbfin) Calibration Services
Min. 0.5 (4.4)
ACC-M-CAI Accredited initial calibration
* In the temperature range –25 to +125 °C (– 13 to + 257 °F)
ËBP-2036----Î

ACC-M-CAF Accredited calibration


All values are typical at 25 °C (77 °F) unless measurement uncertainty is specified ACC-M-CFF Factory standard calibration
COMPLIANCE WITH STANDARDS ACC-M-CTF Traceable calibration
* x = D (decimetres) or M (metres)
yyy = length in decimetres or metres
Please specify cable length when ordering

Brüel & Kjær and all other trademarks, service marks, trade names, logos and product names are the property of Brüel & Kjær or a third-party company.

Brüel & Kjær Sound & Vibration Measurement A/S


DK-2850 Nærum · Denmark · Telephone: + 45 77 41 20 00 · Fax: + 45 45 80 14 05
www.bksv.com · [email protected]
Local representatives and service organizations worldwide
Although reasonable care has been taken to ensure the information in this document is accurate, nothing
herein can be construed to imply representation or warranty as to its accuracy, currency or completeness, nor
is it intended to form the basis of any contract. Content is subject to change without notice – contact