Fundamentals Medical Imaging
Fundamentals Medical Imaging
6 Ultrasound
imaging
submarines, and in
Introduc seismology to
tion locate gas fields.
The basic
Ultrasound
principle of
imaging has been
ultrasound
used in clinical
imaging is simple.
practice for more
A propagating
than half a
wave partially
century. It is
reflects at the
noninvasive,
interface between
relatively
different tissues. If
inexpensive,
these reflections
portable, and has
are measured as a
an excellent
function of time,
temporal
information is
resolution.
obtained on the
Imaging by means
position of the
of acoustic waves
tissue if the
is not restricted to
velocity of the
medical imaging.
wave in the
It is used in
medium is known.
several other
However, besides
applications such
reflection, other
as in the field of
phenomena such
nondestructive
as diffraction,
testing of
refraction,
materials to check
attenuation,
for microscopic
dispersion, and
cracks in, for
scattering appear
example, airplane
when ultrasound
wings or bridges,
propagates through
in sound
matter. All these
navigation ranging
effects are
(SONAR) to
discussed below.
locate fish, in the
Ultrasound
study of the
imaging is used
seabed or to detect
Chapter 6: Ultrasound imaging
Z
1
(6.1) 2
∂
∂
where ρ is the mass 2 p
t
density and c the
acoustic wave 2 =
velocity in the 0,
medium. Table 6.1
(6.
illustrates that c and
consequently also Z 2)
c
typically increase with
ρ. where ∇2 is the
Table6.1 Values of
Laplacian. The
the acoustic wave
velocity c and
velocity c of sound
acoustic in soft tissue is
impedance Z of very close to that
some substances in water and is
Substance c approximately
(m/s)1540 m/s. In air
Air (25◦ C) 346 the sound velocity
Fat 1450is approximately
Water (25◦ C) 1493
300 m/s and in
bone
Soft tissue 1540
approximately
Liver 1550
4000 m/s.
Blood (37◦ C) 1570
Equation (6.2)
Bone 4000
is the basic
Aluminum 6320differential
equation for the
mathematical
13
2
Chapter 6: Ultrasound imaging
description of superposition of a
wave propagation. left and right
A general solution progressive wave
of the linear wave with velocity c.
equation in one The shape of
dimension is p(x, t) the waveform is
= A1f1(x − ct) + irrelevant for the
A2f2(x + ct), (6.3) wave equation. In
other words, any
waveform that can
where f1(x) and
be generated can
f2(x) are arbitrary
propagate through
functions of x that
matter. A simple
are twice
example is the
differentiable. This
following
can easily be
sinusoidal plane
verified by
wave p(x, t) = p0
substituting Eq.
(6.3) into the wave
equation Eq. (6.2). sin
This solution is
illustrated in
Figure 6.2 for f1(x)
= f2(x) = f (x) and A1 , (6.4)
= A2. It represents
the where x is the
direction of the
p(x,t)
acoustic
2f (x) wave
propagation, λ the
wavelength, and T x
the period (c =
f (x + c0t ) f (x – c0t )
λ/T). In practice,
the shape of the
propagating wave
is defined by the
t characteristics of
the transducer.
Figure6.2 Schematic
representation of a Acoustic intensity
solution of the linear and loudness
wave equation. This
solution represents the The acoustic
superposition of a left intensity I (in units
and right propagating
wave. W/m2) of a wave
is the average
energy per time
unit through a unit
area perpendicular
13
3
Chapter 6: Ultrasound imaging
to the propagation = I
direction. Hence, = 12
W/m2. (6.7)
T L 10 log10 I0 ,
1 with I0 10
I= p(
(6.5)
Using Eq. (6.6) L can
T also be written as
0
p
For the plane
progressive wave L = 20 log10 p 0 ,
of Eq. (6.4), the with p0 = 20 µPa.
acoustic intensity (6.8)
is
T I0 is the threshold
at 1000 Hz of human
= 1 hearing. Hence,
(t) dt absolute silence for
I p humans corresponds
Z to 0 dB. Increasing the
intensity by a factor of
ten corresponds to an
·
increase in sound level
of 10 dB and
T
0
approximately twice
T the perceived loudness
in
p 02
= · sin
− dt
= 0
Z 2Z
T humans. Doubling the
T intensity causes an
0 increase of 3 dB.
p2 The wave
. (6.6) frequency also has an
A tenfold increase effect on the perceived
in intensity sounds a loudness of sound. To
little more than twice compensate for this
as loud to the human effect, the phon scale,
ear. To express the shown in Figure 6.3,
sound level L (in is used. The phon is a
decibels (dB)) a unit of the perceived
logarithmic scale was loudness level for
therefore introduced pure, i.e., single-
frequency, tones. By
definition, the number
13
4
Chapter 6: Ultrasound imaging
Interference
Interference between
waves is a well-
known phenomenon.
For an infinite (or a
very large) number of
coherentsources(i.e.,
sourceswiththesamefr
equency
andaconstantphaseshif
t)theresultingcomplexi
nterference pattern is
called diffraction. The
shape of this 3D
pattern is closely
related to the
geometry of the
acoustic source.
Figure 6.4 shows
how two coherent
point sources interfere
in an arbitrary point P.
They can interfere
constructively or
destructively,
depending on the
difference in traveled
13
5
Chapter 6: Ultrasound imaging
Figure6.3 (a) The phon scale compensates for the effect of frequency on the perceived loudness in humans. By definition, x phon is equal
to x dB at a frequency of 1 kHz. 0 phon corresponds to the threshold of audibility for humans, and 120 phon to the pain threshold. An
increase of 10 phon corresponds approximately to twice the perceived loudness. (b) A-weighting. This curve is an approximation of the
isophones in the range 20–40 phons, shown in (a). It expresses the shift of the sound level (in dB) as a function of frequency to obtain an
approximation of
O1
O2
contributions of all
wavelets from all
point sources will
interfere
constructively because
their phase difference
13
6
Chapter 6: Ultrasound imaging
–10 results in an
–20 exponential decay
–30 of the amplitude of
–40 the propagating
10
wave. 96
5 72
0 Typically,
48 Depth (mm) the
24
Lateral distance (mm) –5 attenuation is a
120
function of the
0 wave frequency.
Therefore, it is
Figure6.5 Simulated spatial
distribution of the maximal often modeled as a
pressure generated by a function of the
circular, planar source with
a diameter of 10 mm, form
transmitting a pulse with a
center frequency of 5 MHz. H(f , z) = e−αz ≡ e−α0f nz,
(Courtesy of Professor J.
D’hooge, Department of (6.9)
Cardiology. Reprinted with
permission of Leuven
University Press.) where f is the
frequency and z
the distance
13
7
Chapter 6: Ultrasound imaging
waveform. This S
effect is visible in o
the frequency m
e
domain by the
generation of
t
higherharmonics(i.
y
e., p
integermultiplesoft i
heoriginal c
frequency). a
To illustrate l
this phenomenon,
a propagating v
wave was a
l
measured as a
u
function of time. e
Figure 6.6 s
represents these
recorded time o
signals and their f
spectra. The origin
of the time axis α
(time = 0) is the 0
moment at which
the pulse was f
generated. The o
r
distortion of the
waveform and the
d
introduction of i
harmonic f
frequencies f
increase with e
increasing r
pressure e
amplitude, which n
can be noticed by t
comparing the top
s
row with
u
T
b
a
b s
l t
e a
n
6 c
. e
2 s
Substance α0 (dB/(cm MHz))
13
9
Chapter 6: Ultrasound imaging
14
0
Chapter 6: Ultrasound imaging
Figure6.6 (a) The nonlinear propagation of a wave yields a distortion of the original waveform. (b) This distortion is visible in the
frequency domain as higher harmonics. The amount of distortion depends on both the amplitude (central versus top row) and the
propagation distance (bottom versus central row). (Courtesy of Professor J. D’hooge, Department of Cardiology. Reprinted with
permission of Leuven University
Press.) and
Table 6.3
Nonlinearity
B .
parameter (6.13)
B/A of some Eq. (6.10) can be
biologic rewritten as
tissues
Medium B/A
A
(6.12)
14
1
Chapter 6: Ultrasound imaging
Wave
propagatio
n in
inhomogen
eous media
Tissues are
inhomogeneous
media. When
inhomogeneities
are present,
additional
phenomena occur,
which can be
explained using
Huygens’ Figure6.7 Schematic
principle, as shown representation of
in Figure 6.7. This Huygens’ principle. The
concentric lines in this
states that any figure represent the
point on a wavefronts, i.e.,
wavefrontcanbeco surfaces where the
waves have the same
nsideredasthesourc phase. Any point on a
eofsecondary wavefront can be
waves, and the considered as the
source of secondary
surface tangent to waves and the surface
these secondary tangent to these
waves determines secondary waves
determines the future
the future position position of the
of the wavefront. wavefront.
A wavefront is a
surface where the
waves have the
same phase.
Reflection and
refraction
When a wave
propagating in a
medium with
density ρ1 and
sound velocity c1
meets another
14
2
Chapter 6: Ultrasound imaging
sin θi
sin θ
, (6.15)
c
ui
Aincident
t 2Z2 cos
θi wavefront
T≡ =c +
1
1 2 3 4 (6.17)
c2
refracted
wavefront Ai Z2 cos
θi Z1 cos θt
u
andt
It Ar Z2 cos
−
Figure6.9 Schematic θi Z1 cos θt
representation of refraction R≡ = +
of an incident wave at a
planar interface of two , (6.18)
different media. The
relationships between the Ai Z2 cos
angles θi is given in Eq. θi Z1 cos θt
(6.15).
3 15
2
10
Amplitude (mV)
Amplitude (mV)
1
5
0
0
–1
–5
–2
–10
–3
–4 –15
–5 –20
20 30 40 50 60 70 80 90 100110 20 30 40 50 60 70 80 90 100110
Time (ms) Time (ms)
(a) (b )
14
6
Chapter 6: Ultrasound imaging
t
P P
t
t
t
t
(a) (b)
observer, the =−
fD = fR − fT +
frequencies of the
observed and fT.
transmitted waves
are different. This
(6.20
is the Doppler
effect. A well- ) c
known
va
exampleisthatofaw
histlingtrainpassin
Proof of Eq. (6.20)
ganobserver. The
The position Ps(t)
observed pitch of
of the start point of
the whistle is
the transmitted
higher when the
pulse at time t can
train approaches
be written as
than when it
moves in the other
direction. Ps(t) = ct,
Consider the (6.21)
l
schematic
representation of a
v
transducer, a
u
transmitted pulse,
Transducer
and a point Pe Ps P va
scatterer in Figure
6.12. Assume that
the scatterer d 0 + va . t
moves away from
Figure6.12 Geometry used
the static to derive an expression for
transducer with an the frequency shift due to
axial velocity the motion of the scatterer
(i.e., the Doppler
component va = |v| frequency).
· cos θ. If fT is the
frequency of the
where c is the sound
pulse transmitted
velocity within the
by the transducer,
medium and t the time
the moving
since the pulse was
scatterer reflects
transmitted. The
this pulse at a
position P(t) of the
different frequency
scatterer can be
fR. The frequency
written as
shift fD = fR − fT is
the Doppler
P(t) = d0 + vat,
frequency and can
(6.22)
be written as
2va
14
8
Chapter 6: Ultrasound imaging
c(tie − T) = d0 + vatie
.c d
−
cT
va
tie = 0
+ c − va
Without loss of c
generality, assume that
the pulse length equals = tis + −
one wavelength λ. If T
T
is the period of the
transmitted pulse, the .
position Pe(t) of the
(
end point can be
written as 6
.
P
e 2
( 5
t
) )
c
14
9
Chapter 6: Ultrasound imaging
a
ts =
2tis
The scatterer reflects
(6.27)
the pulse back to the
receiver. The position te = 2tie − T.
where the points Ps
and Pe meet the Consequently,
scatterer are
thedurationTR = te−ts
respectively Ps(tis) and
Pe(tie). These are also ofthereceived pulse
the distances these can easily be obtained
points have to travel
back to the transducer. from Eq. (6.25):
The corresponding c
travel times trs and tre TR = 2
are
Ps(tis)
−
t
r
s
T
=
−
=
t T
i
s
c c
Pe(tie
) v
tre = = tie − T, a
(6.26) c
which have to be
(
added to tis and tie
6.2
respectively to 8)
calculate the travel T.
time back and forth of
Ps and Pe, that is, ts = tis Substituting TR by 1/fR
and T by 1/fT in this
15
0
Chapter 6: Ultrasound imaging
ultrasonic
Crystal
transducer, which
electrical force, it consists of four
can be shown that important elements
the amplitude of (i.e., a backing layer,
electrodes, a
the vibration is piezoelectric crystal,
maximal when the and a matching
thickness of the layer).
crystal is exactly
half the
Electrodes
wavelength of the
induced wave. This
phenomenon is
called resonance, medium. However,
and the because the
correspond- acoustic
ing frequency is called impedance of
the fundamental solids is very
resonance frequency. different from that
Obviously, as of fluids (which is
much of the a good model for
acoustic energy as biologic tissue),
possible should part of the energy
emerge from the is reflected into
crystal through the the crystal again.
surface on the This problem is
15
2
Chapter 6: Ultrasound imaging
solved by using a
so-called matching
Gray
layer (Figure scale
6.13),√
which has an imaging
acoustic
impedance equal
Data
to ZcZt, with Zc and acquisitio
Zt being the
acoustic n
impedance of the Insteadofapplyinga
crystal and the continuouselectric
tissue, alsignaltothe
respectively. It can crystal, pulses are
be shown that if used to obtain
this layer has a spatial
thickness equal to information.
an odd number of Data acquisition is
quarter done in three
wavelengths, different ways.
complete
A-mode
transmission of
energy from the Immediately after
the transmission of
crystal to the
the pulse, the
tissue can be
transducer is used
obtained.
as a receiver. The
An ultrasonic
reflected (both
transducer can
specular and
only generate and scattered) waves
receive a limited are recorded as a
band of function of time.
frequencies. This An example has
band is called the already been
bandwidth of the shown in Figure
transducer. 6.10. Note that
time and depth are
equivalent in
echography
because the sound
velocity is
approximately
constant
throughout the
tissue. In other
words, c
multiplied by the
travel time of the
15
3
Chapter 6: Ultrasound imaging
Amplitude (volt)
principle, is called
A-mode
(amplitude)
imaging. The
detected signal is 0.02
15
4
Chapter 6: Ultrasound imaging
Transducer
Tra
nsd
uce
r
Object
Object cer
du
ns
Tra
Transducer
(a) (b)
Figure6.16 (a) B-mode image of a fetus. The dark region is the uterus, which is filled with fluid. (Courtesy of Professor M. H. Smet,
Department of Radiology) (b) B-mode image of a normal heart in a four-chamber view showing the two ventricles (LV left ventricle; RV
right ventricle), the two atria (LA left atrium; RA right atrium) and the origin of the aorta (outflow tract). Besides the anatomy of the
whole heart, the morphology of the valves (e.g., mitral valve) can be visualized. (Courtesy of the Department of Cardiology.)
reconstruc Envelope
detection
tion Because the very
Reconstructing fast fluctuations of
ultrasound images the RF signal (as
based on the acquired illustrated in
RF data as shown in Figure 6.19(a)) are
Figure 6.14, involves not relevant for
the following steps: gray scale
filtering, envelope imaging, the high-
detection, attenuation frequency
correction, log- information is
compression, and scan removed by
conversion. envelope
15
6
Chapter 6: Ultrasound imaging
RV LV
RA
LA
(a) (b)
40
5
30 10
Lateral displacement (mm)
Amplitude (mV)
20 15
10 20
25
0
30
–10
35
–20 40
–30 45
–40 50
0 2 4 6 8 10 2 4 6 8 10 12 14 16 18 20
Time (ms) Time ( s)
(a) (b)
200
Processed gray value
(a) (b)
and depth are linearly ultrasound scanners
related in echography, therefore enable the
attenuation correction user to modify the
is often called time gain manually at
gain compensation. different depths.
Typically, a simple
model is used – for Log-compression
example, an Figure 6.19(b) mainly
exponential decay – shows the specular
but in practice several reflections. However,
tissueswithdifferentatt the scatter reflections
enuationpropertiesarei are almost invisible.
nvolved. Most Thereasonisthelargedi
15
8
Chapter 6: Ultrasound imaging
fferenceinamplitudebe
tween the specular
Acquisition
and the scatter and
reflections, yielding a
large dynamic range. reconstructio
In order to overcome
this problem, a
n time
suitable gray level To have an idea of the
transformation can be acquisition time of a
applied (see Chapter typical ultrasound
1, p. 4). Typically, a image, a simple
logarithmic function is calculation can be
used (Figure 6.20(a)). made. Typically, each
The log-compressed line in the image
version of the image corresponds to a depth
in Figure 6.19(b) is of 20 cm. Because the
shown in Figure velocity of sound is
6.20(b). The scatter or approximately 1540
speckle can now m/s and the travel
easily be perceived. distance to and from
Note that different the transducer is 40
tissues generate cm, the acquisition of
different speckle each line takes 267 µs.
patterns. A typical image with
120 image lines then
Scan conversion requires an acquisition
If the image is time of about 32 ms.
acquired by tilting the The reconstruction of
transducer instead of the images can be
translating it, samples done in real time.
on a polar grid are Consequently, a
obtained. Converting temporal resolution of
the polar into a 30 Hz (i.e., 30 images
rectangular grid needs per second) can be
interpolation. This obtained. This
process is called scan temporal
conversion or sector resolutioncanbeincrea
reconstruction. sedatthecostofspatialre
solution by decreasing
the number of scan
lines. However,
current clinical
scanners are able to
acquire multiple scan
lines simultaneously
with little influence on
the spatial resolution
15
9
Chapter 6: Ultrasound imaging
p(t) = ymax
Line number
(6.31)
Amplitude (volt)
The received signal
then is
0.02
0 25
0.02 20
20
. (6.32) 15
40 Line number
range 60 10
The interval t is the 80 Depth (mm)
gate tR 5
time between the 100
transmission and the 120 0
If the scattering
object moves away
from the transducer
with a constant axial
velocity v a, the
distance d increases
between subsequent
pulses with vaTPRF,
where TPRF is the pulse
repetition period, TPRF
=
16
3
Chapter 6: Ultrasound imaging
sinusoidal function
US pulse moving scatterer
with frequency
2va
fD = −
samples at tR
fT, (6.35)
c
which is exactly
the Doppler
frequency defined
in Eq. (6.30).
tR Hence, the velocity
reflected pulsesof the scattering
object at a specific
Figure6.23 Schematic
representation of the PW depth, defined by
Doppler principle. Pulses the range gate, can
are transmitted at a fixed be calculated from
pulse repetition frequency
(PRF). They are reflected at the sampled signal
a scatterer in motion. sj(tR). As in CW
Because of this motion, the
reflected pulses are Doppler, this
dephased. Measuring each signal can be made
reflected pulse at the range
gate tR yields a sampled
audible or can be
sinusoidal signal with represented as a
frequency fD. spectrogram.
Often, the
spectrogram is
1/PRF (see Figure displayed together
6.23). Consequently, with a B-mode
Eq. (6.33) can be gray scale image in
written as which the position
sj of the range gate is
indicated. Such a
combined scan is
called a duplex.
Note that a PW
Doppler system as
TPRF described above is
(6.34) not able to detect
the direction of
where sj(tR) is the motion of the
sample of the jth scattering object.
pulse. Hence, the To obtain
values sj(tR) in Eq. directional
(6.34) are samples of information, not
a slowly time-varying one but two
samples, a quarter
16
4
Chapter 6: Ultrasound imaging
? tR ! tR1 tR2
tR tR1 tR2
? ?
tR tR1 tR2
(a) (b)
Figure6.24 If a single sample is acquired at the range gate, no directional information is obtained (a). However, if a second sample is
acquired slightly after the first one, the direction of motion is uniquely determined since a unique couple of samples within the cycle is
obtained (b).
(a) (b)
Figure6.26 (a) Using color Doppler techniques, blood flow within the ventricles can be visualized. This image shows the flow in a normal
left ventricle at the beginning of diastole. Red colors represent flow toward the transducer, coming from the left atrium through the
mitral valve and into the left ventricle. Blue colors show the blood within the left ventricle flowing away from the transducer toward the
aorta. (b) Doppler techniques can be used to acquire the slower, regional velocities of the heart muscle itself. Local velocities in the
direction of the transducer are represented in red, and velocities away from the transducer are in blue. (Courtesy of the Department of
Cardiology.)
Acquisition
and
reconstructio
n time
Continuous wave and
pulsed wave Doppler
require a long
transmission of either
CW or PW ultrasound
in tissue.
Reconstruction then
consists of a simple
Fourier transform,
which can be done in
real time. In practice,
changes in velocities
over time are
investigated, and the
received signal is
subdivided into small
(overlapping)
segments whose
spectral amplitudes
constitute the columns
of a spectrogram.
Figure 6.21 shows an
example of a
16
8
Chapter 6: Ultrasound imaging
spectrogram of the
heart. One such
Image
complete spectrogram quality
is typically acquired
in 3 to 4 seconds, Spatial
which corresponds to
3 to 4 heart cycles.
resolution
In color flow The spatial resolution
imaging, a few pulses in ultrasound imaging
are sent along each distinguishes the
image line. This axial, lateral, and
means that the time elevation resolution,
needed to acquire an i.e.,
theresolutioninthedire
image is the time
ctionofwavepropagati
required to obtain a
on, the
Bmode gray scale
y
image times the
number of pulses sent
along each line. If, for
example, three pulses
per line are used, an x
Elevation
image is obtained in
z
approximately 100
ms. Velocity
calculation can be
done in real time, Figure6.27 Schematic
representation of the axial,
resulting in a frame lateral and elevation
rate of 10 Hz. To directions.
increase this
framerate,
thefieldofview(FOV)i
resolution
susuallydecreased.
perpendicular to
The size of the FOV
the axial direction
can be selected
within the image
independent of the
plane, and the
FOV of the gray scale
resolution
image. This way, the
perpendicular to
velocity calculations
the image plane
are limited to the
(see Figure 6.27).
regions that contain
relevant velocity
information. Axial resolution
Resolution can be
expressed by the
PSF, which
determines the
minimum distance
16
9
Chapter 6: Ultrasound imaging
17
1
Chapter 6: Ultrasound imaging
0.04 0.04
0.02
Amplitude (V)
0.02
Amplitude (V)
0 0
–0.02 –0.02
–0.04 –0.04
51.8
7 50.8
50.8 3.5 3
0 0
–3.5 Depth (mm) –3
Depth (mm)
6
49.8 –7 Lateral distance (mm)
49.8 –6 Lateral distance (mm)
(a) (b)
5 5
10 10
15 15
20 20
25 25
30 30
35 35
40 40
45 45
50 50
Figure6.30 Image obtained with (a) a planar and (b) a
concave transducer. Because of beam focusing, the lateral
resolution in (b) is clearly superior to that in (a). (Courtesy
of Professor J. D’hooge, Department of Cardiology.
Reprinted
with permission of Leuven University
Press.)
2 4 6 8 10 12 14 16 18 20 2 4 6 8 10 12 14 16 18 20
Time (s) Time (s)
(a) (b)
17
3
Chapter 6: Ultrasound imaging
Doppler
image
artifacts
Aliasing
A common artifact
of pulsed Doppler
methods (PW and
CF) is aliasing
(see Figure 6.32).
Aliasing is due to
undersampling
(see Appendix A,
p. 230). The
principle is shown
schematically in
Figure 6.33.
The following
constraint between the
range d0 and the
velocity va can be
deduced:
c
2
|va|.
(6
.3
7)
d
According to the
Nyquist criterion, the
sampling frequency
PRF must be larger
than twice the
maximum
17 Figure6.33 Schematic representation of aliasing encountered in Doppler imaging. (a) Same as Figure 6.23. The reflected pulse is
sampled fast enough to avoid aliasing. (b) The scatterer moves faster than in (a) causing a larger phase difference of the subsequent
6 reflected pulses.
Measuring this signal at the range gate tR results in exactly the same sampled sinusoidal function and Doppler frequency fD as in (a).
Sampling is clearly too slow and the calculated scatterer velocity is too low.
Chapter 6: Ultrasound imaging
spatial frequency c
present in the signal, PRF.
that is, (6.40)
4 va From Eqs. (6.38) and
PRF > 2|f (6.40), it follows that
| |fT.
(6.38) 4|va| c
c fT
<
Because the reflected PRF
ultrasonic pulse must <
be received before the
next one can be (6.4
transmitted, a pulse 1) c
must be able to travel 2d0
to and from the range
gate within the period or
TPRF. Hence, the
c2
following relation
|va| <
between TPRF and the
. (6.42)
depth d0 of the range
gate holds: 8d0fT
2
d0
Equipment
An ultrasound scanner
< is small and mobile.
This is an important
advantage as
T
compared with the
P
other imaging
R
modalities. It consists
F
of a transducer
.
connected to a signal
processing box, which
(
displays the
6
reconstructed images
.
on a monitor in real
3 time (Figure6.34).
9 Laptopandpocket-
) sizeultrasoundscanner
c s are also being
developed. The wide
Because TPRF = 1/PRF applicability of these
by definition, it portable devices will
follows that certainly play an
important role in the
17
7
Chapter 6: Ultrasound imaging
further development
of ultrasound imaging.
A variety of
transducers for 2D and
3D imaging exist for
different applications.
The most important
characteristics are
discussed below.
One-
dimensiona
l array
transducers
A disadvantage of the
acquisition method
shown in Figure 6.17
is that it involves a
mechanical
displacement or
rotation of the
transducer. This yields
several practical
difficulties. For
Figure6.34 Example of a
example, the commercial
displacement between echocardiographic scanner.
two steps must be (Courtesy of the
Department of Cardiology.)
kept constant, and the
contact with the
medium must not be
lost. To reduce this mechanical
motion, electronic
scanning with an
array
transducercanbeus
ed.
Anarraytransduceri
sacollectionofman
ysmallidenticalcry
stalsthatcanbeexcit
ed independently.
Most common in
medical imaging
are the linear-
17
8
Chapter 6: Ultrasound imaging
17
9
Chapter 6: Ultrasound imaging
Radiology.)
(a) (b)
Figure6.36 By using an
array of crystals, a sound
beam can be directed (a) or
focused (b) without
mechanical movement of
the transducer. A time
delay is imposed on the
electrical signals that excite
the crystals, causing a
phase shift of the emitted
pulse. This yields the
specific tilted or focusing
wavefront as shown in this
diagram.
C-scans
18
2
(a) (b)
Chapter 6: Ultrasound imaging
the number of 2D
Transducers slices that can be
for 3D acquired per second is
limited by the frame
imaging rate, which in turn is
The easiest way to limited by the sound
construct a 3D image velocity. This way a
is to rotate or wobble 3D acquisition cannot
a phased-array be accomplished in
transducer and acquire real time. Hence, a
images sequentially compromise needs to
from different scan be made between the
planes as illustrated in spatial resolution and
Figure 6.37. Three- the temporal
dimensional imaging resolutions. Moreover,
offers the because a 2D array of
opportunitytoobtainsli crystals contains about
cesofanyorientationthr 64 × 64 = 4096
ough the scanned crystals, A/D
volume. Images of conversion and
planes parallel to the cabling are technically
surface of the difficult.
transducer, as
illustrated in Figure
6.37, are known as C-
scans. Obviously, the
patient should not
move during the
scanning process. An
exception is the cyclic
rhythm of the cardiac
contraction, which can
trigger the imaging
process.
Again, the
mechanical motion Figure6.38 The shape of
the transducer is adapted
can be replaced by its to the application: (1)
electronic abdominal transducers –
general purpose; (2)
representation using a intraoperative
2D phasedarray transducers; (3) small
transducer. The focal parts transducers
(muscles, tendons, skin,
point can be thyroid, breast, scrotum);
positioned at any point (4) intrarectal transducer
(rectal wall, prostate); (5)
within a cone instead intravaginal transducer
of a plane and a 3D (uterus, ovaries,
image can be pregnancy); (6) infants
(abdominal, brain).
obtained. However, (Courtesy of Professor R.
18
3
Chapter 6: Ultrasound imaging
Oyen, Department of
Radiology.) Clinical
use
Echography is a
Nevertheless,
safe, transportable,
commercial 3D
and relatively
scanners are already
cheap imaging
available. technique that
does not require
any special
Special infrastructure. In
addition,
purpose sequences of
images are
transducers obtained in real
Several time. For these
transducers exist reasons,
for a variety of ultrasoundimaging
applications, such isusuallythemetho
as intrarectal, dofchoiceifit is
intravaginal, clinically relevant.
transesophageal It is useful if the
and intravascular ultrasonic waves
transducers (see are able to reach
Figure 6.38). the tissues under
Transesophageal examination and if
probes (Figure the specular or
6.39) are used to scatter reflections,
visualize the heart or both, are high
through the enough to be
esophagus. perceived in the
Intravascular image.
probes are inserted Consequently,
into an artery or
vein to make
intravascular
ultrasound (IVUS)
images. The
crystal is mounted
on top of a catheter
(diameter 1 mm).
Complete phased
arrays can be built
on this small tip.
Figure6.39
Transesophageal
18
4
Chapter 6: Ultrasound imaging
18
5
Chapter 6: Ultrasound imaging
(a) (b)
(a)
Left AVERAGE
TL = 22.79mm T+12.3dB
18
6
Chapter 6: Ultrasound imaging
100%
t(µs)
0 10 20 30
(b)
18
9
Chapter 6: Ultrasound imaging
right
diaphragm kidney
liver
lung effusion
acoustic
shadow of the ribs
(a) (b)
19
0
Chapter 6: Ultrasound imaging
(a) (b)
aerobily shadow
(c)
biopsy probe
prostate
transrectal
transducer
(a) (b)
density resulting
from a negative
pressure
(rarefaction
regions),
microscopic gas
bubbles can be
Figure6.47 A transverse
ultrasound image of the formed. When the
abdomen of this fetus pressure increases,
shows a fluid-filled and these bubbles
dilated collecting system of
the left kidney (arrows) collapse. This can
resulting from a renal cause tissue
outflow obstruction.
(Courtesy of Dr. I. Witters, damage. A
Department of Obstetrics mechanical index
and Gynecology, and based on the peak
Professor M. H. Smet,
Department of Radiology.) negative pressure
is shown by the
19
2
Chapter 6: Ultrasound imaging
19
3
Chapter 6: Ultrasound imaging
3.1 cm
(a) (b)
skin
haematoma
deltoid muscle
muscle
fat
supraspinatus m8m
0.
muscle
muscle
humerus
(a) (b)
19
4
Chapter 6: Ultrasound imaging
(a) (b)
(c)
19
5
Chapter 6: Ultrasound imaging
(a) (b)
19
6