Astm D-4212
Astm D-4212
Copyright © ASTM, 100 Barr Harbor Drive, West Conshohocken, PA 19428-2959, United States.
1
D 4212
NOTE 1—Dimensions are approximate only and may vary with the manufacturer and from batch to batch
FIG. 1 Zahn Cup Nominal Dimensions
relationship whatsoever with a No. 2 Shell cup. 6.2.1 Nominal Shell cup orifice diameters are listed in Table
X2.1. Cup Nos. 1 through 21⁄2 are recommended for use with
6.1.1 Nominal Zahn cup orifice diameters are listed in Table
X2.1. Cup No. 1 with the smallest orifice is used for determin- reduced rotogravure inks; No. 2 is for use with flexographic
ing the viscosity of thin-bodied materials. Cup No. 2 is for use inks; Nos. 3 through 4 are used for industrial enamels,
with clears, lacquers, enamels, and press-side adjustment of lacquers, flexographic, and gravure inks; Nos. 5 and 6 are used
flexographic inks; cups Nos. 3 and 4 are for use with more for heavy materials.
viscous paints and inks (No. 3 for manufacturing of flexo- 6.3 Calibration Thermometer—ASTM Saybolt Viscosity
graphic inks); and cup No. 5 is used for silk screen inks. Thermometer 17F having a range of 66 to 80°F and subdivi-
6.2 Shell Viscosity Cup4—No. 1 through No. 6 Shell vis- sions of 0.2°F, or 17C having a range of 19 to 27°C and
cosity cups made of stainless steel with a capacity of 23 mL subdivisions of 0.1°C, both conforming to the requirements of
and a 1-in. (25-mm) long capillary in the bottom and conform- Specification E 1. Thermometers having subdivisions other
ing to the dimensions shown in Fig. 2. than these may be used depending on the sensitivity of the
material to be tested, the demands of the application, and the
agreement between the purchaser and seller.
4
Shell cups may be obtained from the Norcross Corp., 255 Newtonville Ave.,
Newton, MA 02158. This committee is not aware of any other source for flow cups 6.4 Timer—Any timing device may be used provided that
having properties similar enough to the Shell cup to be included in this test method. the readings can be taken with a discrimination of 0.1 s or
If you have knowledge of a cup that should be considered, please provide details to better.
ASTM Headquarters. Your comments will receive careful consideration at a meeting
of the responsible technical committee,1 which you may attend.
2
D 4212
3
D 4212
NOTE 4—Dip cups are not recommended for use with thixotropic (time between-laboratories coefficient of variation was 11.5 %.
dependent) materials but if used for them (such as gravure or flexographic Based on these coefficients the following criteria should be
inks), more vigorous agitation will be necessary to break up the structure used for judging the acceptability of results at the 95 %
before the measurement is made.
confidence level:
10.3 Lift the cup vertically out of the material in a quick, 13.1.1.1 Repeatability—Two results, each the mean of two
steady motion. As the top edge of the cup breaks the surface, measurements, obtained by the same operator should be
start the timer. During the time of flow, hold the cup vertically considered suspect if they differ by more than 11 % of their
no more than 15.2 cm (6 in.) above the level of the liquid. Stop mean value.
the timer at the first definite break in the stream at the base of 13.1.1.2 Reproducibility—Two results, each the mean of
the cup. The efflux time in seconds constitutes the viscosity. It two measurements, obtained by operators in different labora-
is common to make only a single measurement, but for greater tories should be considered suspect if they differ by more than
precision and accuracy the mean of two or more measurements 33 % of their mean value.
should be taken.
NOTE 6—The values used to determine the precision were obtained
NOTE 5—The cup should not be held by the loop handle during the under ideal conditions (a single set of cups), reproducibility in practice can
measurement process. Most manufacturers equip the cup with a ring be just as good, by employing strict controls and good techniques.
through the loop handle. Holding the cup by this ring will help to ensure
that the cup hangs vertically. 13.1.1.3 Bias—Bias does not apply to this test method as no
acceptable standards exist.
11. Care of Cups
NOTE 7—Since the precision values were obtained under ideal condi-
11.1 Following each determination, clean the cup with a tions (a single set of cups), reproducibility in practice probably is poorer
suitable solvent and a soft brush. Use no metal tools in contact than that given (perhaps as bad as 50 %).
with the instrument as nicks or wear of the drilled orifice affect
13.1.2 Shell Cups—Precision was determined on the basis
the accuracy of the cup.
of an interlaboratory test in which four laboratories tested
12. Report seven paints covering a broad range of viscosities. The
12.1 Report the efflux time to the nearest 0.2 s for Zahn or within-laboratory coefficient of variation was 3.2 % and the
Shell cup No. ___, manufactured by _____, (in the case of between-laboratories coefficient of variation was 6.3 %. Based
Zahn cups) the temperature of the fluid (where measured), and on these coefficients the following criteria should be used for
whether the result is from a single measurement or the mean of judging the acceptability of results at the 95 % confidence
two of more measurements. level:
13.1.2.1 Repeatability—Two results, each the mean of two
13. Precision and Bias measurements, obtained by the same operator should be
13.1 The most satisfactory results when using dip cups are considered suspect if they differ by more than 9 % of their
obtained when viscosity is being controlled at a single location mean value.
only. However, when comparisons between locations are made, 13.1.2.2 Reproducibility—Two results, each the mean of
cups from the same manufacturer must be used or other action two measurements, obtained by operators in different labora-
taken to ensure compatibility of results. The following criteria tories should be considered suspect if they differ by more than
can be used for judging the acceptability of results at the 95 % 18 % of their mean value.
confidence level: 13.1.2.3 Bias—Bias does not apply to this test method as no
13.1.1 Zahn Cups—Precision was determined on the basis acceptable standards exist.
of an interlaboratory test in which six laboratories used new
Zahn cups (all from the same set from the same manufacturer) 14. Keywords
to test eight paints covering a broad range of viscosities. The 14.1 dip cup(s); flow cup(s); Shell cup(s); viscosity; Zahn
within-laboratory coefficient of variation was 3.7 % and the cup(s)
4
D 4212
APPENDIXES
(Nonmandatory Information)
X1.1 Zahn cups were designed (and manufactured by X1.2 Zahn cup patents have expired and Zahn-type cups
General Electric) as simple flow property devices for use in dip are available from several sources. Each manufacturer makes
tanks, flow coat reservoirs, etc. They were not designed to be cups that are somewhat different from those made by others.
viscometers or to be used as such. Zahn cup use usually Considerable batch-to-batch difference has been noted in the
involves thinning or maintaining a coating, adhesive, or ink at past from some manufacturers. Cup variations are not a
a certain consistency (so many Zahn-seconds) regardless of problem for consistent control of a bath or tank as long as
temperature. This is the beauty of dip cups. The operator appropriate comparisons are made with old cups, if required, to
knows that his or her tank, bath, or coater runs well over a maintain continuity of data. However, cup differences can
certain range of Zahn-seconds whether the plant temperature is cause great difficulty if cups are used to set producer-user
50°F or 100°F. The operator does whatever is necessary to keep specifications. Comparisons under such conditions require
the fluid within the range. In such an application the viscosity considerable attention to detail and practiced expertise.
at 25.0°C (77.0°F) is not important.
X2.1 The viscosity versus efflux time formulas published TABLE X2.1 Viscosity Standards Recommended for Checking
here are in wide use, however, not all Zahn-type cups are Cups
designed to comply with these formulas. Techniques for Cup Number Approximate Oil Viscosity
Nominal Orifice Diameter,
at 77°F (25°C), cST
ensuring continued calibration of cups that do not claim (mm)
(mm2/s)
compliance with these formulas include; comparison of current
Zahn
flow times for a standard fluid against the original flow time, 1 2.0 20
and comparison against a verified standard cup. The same 2 2.7 120
3 3.8 480
calibration verification methods of comparing a cup against 4 4.3 480
one in known condition applied to Zahn-type cups can be used 5 5.3 900, 1600
with Shell cups.
Shell
X2.2 Monitoring Cup Characteristics—A useful checking 1 1.8 9
2 2.4 9, 20
technique is to measure the efflux time for a new cup with a 21⁄2 2.7 35
given standard fluid, then check the cup periodically with the 3 3.1 35, 120
same oil at the same temperature to determine if the efflux time 31⁄2 3.5 120
4 3.8 120
has changed. If the time changes more than 20 % (or more than 5 4.6 120, 480
your process tolerance permits), the cup should be replaced. If 6 5.8 480
the change is small, it may be appropriate to use the cup as is,
or by applying a correction factor to subsequent efflux times,
depending on the degree of precision required. The correction
have a tendency to contaminate cups and containers and to change the
factor is equal to original standard–fluid efflux time divided by drainage characteristics of cups (especially cups with capillaries).
the current one. Recommended viscosities for standard fluids
for such tests are given in Table X2.1. X2.4 Comparing Cups—It sometimes is necessary to
compare one Zahn-type cup with another in order to settle a
X2.3 The viscosity of many standard fluids is very sensitive dispute, determine whether a new cup will give similar times to
to temperature variations, therefore, the temperature of the an old one, etc. The most usual technique is to dip both cups
fluid must be controlled closely during calibration testing. It is into the same container of standard fluid at the same time and
recommended that the fluid (and the cup) be held at the test once the temperature of both cups has stabilized, pull them out
temperature for at least 15 min prior to testing. together, timing both of them. The efflux times are compared
NOTE X2.1—Silicone fluid viscosity standards are not recommended. and a correction factor can be calculated. Sometimes paint or
Although they exhibit very little viscosity change with temperature, they another material is substituted for the oil.
5
D 4212
X3.1 Select the appropriate standard oil for the cup to be TABLE X3.1 ConstantsA for Use with Viscosity Formulas
checked (see Table X2.1). Cup K C
X3.2 Bring the cup and the standard fluid to a constant Zahn
1 1.1 29
temperature as close as possible to 77.0°F (25.0°C). Some 2 3.5 14
fluids sold as viscosity standards can vary in viscosity by 2 to 3 11.7 7.5
18 % per degree centigrade, higher viscosities typically having 4 14.8 5
5 23 0
the higher rate of change. Determine the time of efflux to the Shell
nearest 0.2 s using the procedure detailed in Section 10. Record 1 0.226 13
2 0.576 5
the temperature of the efflux stream. If it is not 77°F, correct the
21⁄2 0.925 3
viscosity of the standard fluid to the actual temperature. 3 1.51 2
Temperature versus viscosity data is available from most 31⁄2 2.17 1.5
4 3.45 1
suppliers of viscosity standards. 5 6.5 1
X3.3 Convert the time of flow in seconds to kinematic 6 16.2 0.5
A
viscosity as follows: Cup constants from Patton, T. C., Paint Flow and Pigment Dispersion, second
edition, John Wiley & Sons, New York, 1979, p. 82.
V 5 K ~t 2 c! (X3.1)
The American Society for Testing and Materials takes no position respecting the validity of any patent rights asserted in connection
with any item mentioned in this standard. Users of this standard are expressly advised that determination of the validity of any such
patent rights, and the risk of infringement of such rights, are entirely their own responsibility.
This standard is subject to revision at any time by the responsible technical committee and must be reviewed every five years and
if not revised, either reapproved or withdrawn. Your comments are invited either for revision of this standard or for additional standards
and should be addressed to ASTM Headquarters. Your comments will receive careful consideration at a meeting of the responsible
technical committee, which you may attend. If you feel that your comments have not received a fair hearing you should make your
views known to the ASTM Committee on Standards, 100 Barr Harbor Drive, West Conshohocken, PA 19428.
This standard is copyrighted by ASTM, 100 Barr Harbor Drive, West Conshohocken, PA 19428-2959, United States. Individual
reprints (single or multiple copies) of this standard may be obtained by contacting ASTM at the above address or at 610-832-9585
(phone), 610-832-9555 (fax), or [email protected] (e-mail); or through the ASTM website (http://www.astm.org).