Principles and Safety Measures of Electrosurgery in Laparos
Principles and Safety Measures of Electrosurgery in Laparos
ABSTRACT INTRODUCTION
Background: Electrosurgical units are the most common A basic understanding of electricity is needed to safely
type of electrical equipment in the operating room. A apply electrosurgical technology for patient care.1 Electro-
basic understanding of electricity is needed to safely apply surgery is one of the most commonly used energy systems
electrosurgical technology for patient care. in laparoscopic surgery.2 The surgical team should have a
good understanding of the principles of electrosurgery
Methods: We reviewed the literature concerning the es-
and tissue effects to avoid complications. The risk of
sential biophysics, the incidence of electrosurgical inju-
complications is linked to the surgeon’s fundamental
ries, and the possible mechanisms for injury. Various
knowledge of instruments, surgical technique, biophysics,
safety guidelines pertaining to avoidance of injuries were
relevant anatomy, and safe technical equipment. The risk
also reviewed.
of complications is linked to fundamental surgical knowl-
Results: Electrothermal injury may result from direct ap- edge of instruments, surgical technique, biophysics, and
plication, insulation failure, direct coupling, capacitive relevant anatomy. Appropriately applied, electrosurgery is
coupling, and so forth. safe and effective. Electrothermal injury may result from
direct application, insulation failure, direct coupling, and
Conclusion: A thorough knowledge of the fundamentals capacitive coupling.3
of electrosurgery by the entire team in the operating room
is essential for patient safety and for recognizing potential
History
complications. Newer hemostatic technologies can be
used to decrease the incidence of complications. The conception of electrosurgery began in the early 19th
century when the French physicist Becquerel first used
Key Words: Electrosurgery, Electrosurgical safety, Lapa- electrocautery. Rather than using boiled oil to achieve
roscopic electrosurgery. hemostasis, he passed direct current through a wire
thereby heating it and effectively cauterizing tissue upon
contact. In 1881, D. Arsonoval pioneered the use of alter-
nating current.
It was not until the late 1920s that collaboration between
the physicist, William T. Bovie and the neurosurgeon
Harvey Cushing resulted in the predecessor of today’s
electrosurgical unit. This model was used until 1968 when
a smaller model was developed by Valleylab, which has
since produced today’s platform of electrosurgical units.3
electrode in order to disperse the current returning to the away from the tissue to create a spark gap or steam
electrosurgical unit and minimize heat production at this envelope through which the current arcs to the tissue.
return electrode site.6 – 8 This spark gap results from heating up the atmosphere
between the electrode and the tissue.13 The coagulation
4. Tissue Conductivity current is effective with the power settings in the range of
30W and 50W.6
Various tissue types have a different electrical resistance,
which affects the rate of heating. Adipose tissue and bone Fulguration (Spray) is a noncontact coagulation that also
have high resistance and are poor conductors of electric- utilizes spark gap to mediate tissue effects, which results
ity, whereas muscle and skin are good conductors of in heating and necrosis as well as greater thermal spread.
electricity and have low resistance.7,10 Desiccation (Deep) is another form of coagulation in
which direct contact is made with the tissue, resulting in
5. Current Waveforms electrical energy being converted into heat within the
The final determinant of how tissue responds to elec- tissue. The end result is deeper necrosis and greater ther-
trosurgery is the current type. Electrosurgical units pro- mal spread.9
duce 3 different waveforms: cut, blend, and coagulation Table 1 is giving an overview of the tissue effects of the
(Figure 2).9 two traditional and two innovative energy modalities.
A pure cutting (vaporization) waveform is continuous,
In monopolar electrodes, radiofrequency current flows from
unmodulated, and undamped. A coagulation waveform is
the generator through the active electrode, into the target
interrupted, modulated, and damped current.11,12 A blend
tissue, through the patient, the dispersive electrode and then
waveform is a modification of the cutting waveform and is
returns to the generator.13 The most common site of injury is
used when hemostasis is needed while cutting.5,6 This
at the patient return electrode. The return electrode must be
waveform type consists of a combination of both cutting
of low resistance with a large enough surface area to dis-
and coagulation waveforms.4 Higher blend settings trans-
perse the electrical current without generating heat. If the
late into more time between bursts of current and greater
patient’s return electrode is not large enough or is not com-
coagulation, as seen in the following examples: Blend 1
pletely in contact with the patient’s skin, then the current
(80% cut, 20% coagulation); Blend 2 (60% cut, 40% coag-
exiting the body can have enough density to produce unin-
ulation); and Blend 3 (50% cut, 50% coagulation).9
tended burns. Excessive hair, adipose tissue, bony promi-
A cutting current power setting must be between 50W and nences, and the presence of fluid and scar tissue compro-
80W to be effective. Ideally, the electrode is held slightly mise the quality of contact. To avoid this type of injury,
Direct Coupling instrument, which is outside the view of the monitor but
distal to the protective cannula, carries the highest risk for
Direct coupling occurs when the electrosurgical unit is
creating an injury that even the most attentive surgeon is
accidentally activated while the active electrode is in close
unable to detect. Disposable instruments have a lower
proximity to another metal instrument. Current from the
incidence of insulation failure compared with reusable
active electrode flows through the adjacent instrument
instruments. The distal third of laparoscopic instruments is
through the pathway of least resistance, and potentially
the most common site of insulation failure (Figure 7).20
damages adjacent structures or organs not within the vi-
sual field that are in direct contact with the secondary
Capacitive Coupling
instrument.6 It can be prevented with visualization of the
electrode in contact with the target tissue and avoiding Capacitive coupling is electrical current that is established
contact with any other conductive instruments prior to in tissue or in metal instruments running parallel to but not
activating the electrode (Figure 6).4 directly in contact with the active electrode. This occurs
when electric current is transferred from one conductor
Insulation Failure (the active electrode) through intact insulation and into
adjacent conductive materials (eg, bowel) without direct
This is now thought to be a main cause of laparoscopic
contact.21
electrosurgical injuries. It is defined as a break or defect in
the insulation that coats the instrument. Insulation failure In monopolar mode, an alternate current flowing through
is caused by excessive use of reusable instruments, par- an active monopolar electrode and back to the electrosur-
ticularly with repetitive passage through trocars and fre- gical generator through the patient and the return pad
quent mechanized sterilization.19 By lowering the concen- induces an unintended current in any conductors in close
tration of the current used, coagulation with cutting proximity. The degree of current induced will depend on
current and use of an active electrode monitoring system, the proximity of the conductors, the voltage, and the
the risk of accidental burns can be reduced.6 insulation. Any conductor in the operating room is at risk
of inheriting a stray current by becoming capacitively
Eighteen percent of insulation defects are located in the
coupled to the current coming from the active electrode.
section of the instrument most likely to create a cata-
strophic electrosurgical injury. Originally described as
‘‘Zone 2 ” by Voyles and Tucker, the location along the
NEWER TECHNOLOGIES
Figure 8. Capacitive Coupling. A capacitively coupled current Tissue response generators are the next step in the evo-
typically returns through a metal trocar sheath to the grounding lution of electrosurgical generators. By using a computer-
pad. If a plastic trocar sheath is used, the current will accumulate controlled tissue feedback system that senses tissue im-
at the junction of the plastic and metal and seek an alternate pedance or resistance, a consistent electrosurgical clinical
path. effect is obtained through all tissue types.6,27
Vessel Sealing Technology Its higher cost is a significant handicap for its use in
surgery in developing countries.
The most recent advancement in electrosurgery has been
the introduction of vessel sealing technology. Core to this The EnSeal instruments adjust dose energy simultaneously
technology is the use of bipolar electrosurgery that relies to various tissue types in a tissue bundle each with its own
on tissue response generators. This advanced electrical impedance characteristics. Less heat is required to accom-
current is combined with optimal mechanical pressure plish fusion, as the tissue volume is minimized through
delivery by the instruments to fuse vessel walls and create compression; energy is focused on the captured segment;
a seal. Specifically, high current and low voltage are de- and the vessel walls are fused through compression, pro-
livered to the targeted tissue and denature the collagen tein denaturation, and then renaturation.
and elastin in the vessel wall while the mechanical pres-
sure from the instrument allows the denatured protein to Ultrasonic Technology
form a coagulum.28 Vessels up to 7mm in diameter and
large tissue bundles can now be surgically ligated. Addi- The Harmonic scalpel is an ultrasonic surgical instrument
tionally, thermal spread appears to be reduced compared for cutting and coagulating tissue, operating at a fre-
to traditional bipolar electrosurgical systems. Unlike tra- quency of 55.5 kHz/second or 55,500 cycles per second
ditional electrosurgical instruments, these devices re- (Figure 9 and Table 1).34
quire a tension-free application to tissue bundles to There is no electrosurgical current generated. The combi-
successfully obtain the desired tissue effect. Valleylab, nation of mechanical energy and the heat that is generated
Gyrus ACMI, and SurgRx, Inc. are 3 companies that causes protein denaturation and formation of a coagulum
have developed devices for both open and laparo- that seals small blood vessels. Typically, this energy mo-
scopic applications.6,8,28 –31 dality is effective for blood vessels between 2mm and
3mm, although a newer device has demonstrated the
The LigaSure system produces supraphysiological seals ability to coagulate blood vessels up to 5mm in diameter
with significantly higher bursting pressures than the with less heat, charring, and thermal injury to surrounding
plasma kinetics sealer (PK, Gyrus Medical, Maple Grove, tissues.35–37
MN) in vessels ranging from 4mm to 7mm. The plasma
kinetics (PK) seal becomes progressively weaker while The 2 cutting mechanisms of the Harmonic scalpel are
the LigaSure seal increases in strength as the vessel size different from that observed with electrosurgery or laser
increases.32 surgery. The first mechanism is cavitational cutting and
fragmentation. As the blade tip vibrates, it produces large
The LigaSure vessel sealing system is a safe alternative for transient pressure changes, which causes cellular water to
securing pedicles in vaginal hysterectomy compared with vaporize at low temperature, rupturing cells, leading to
conventional suture ligation.33 The LigaSure system re- very precise cutting and dissection. The second mecha-
duces the operating time (by reducing pedicle-securing nism for cutting by Harmonic scalpel is the actual power
time) and blood loss without increasing the postoperative cutting offered by a relatively large blade vibrating 55,500
complication rates of vaginal hysterectomy. This benefi- times per second. The blade edge cuts tissue by stretching
cial effect was more pronounced in difficult procedures. it beyond its elastic limit and on a more microscopic level,
2. Wu MP, Ou CS, Chen SL, Yen EYT, Rowbotham R. Compli- 19. Vancaillie TG. Active electrode monitoring. How to prevent
cations and recommended practices for electrosurgery in lapa- unintentional thermal injury associated with monopolar electro-
roscopy. Am J Surg. 2000;179:67–73. surgery at laparoscopy. Surg Endosc. 1998;12:1009 –1012.
3. Nezhat CH, Nezhat F, Brill I, Nezhat C. Normal variations of 20. Montero PN, Robinson TN, Weaver JS, Stiegmann GV. Insu-
abdominal and pelvic anatomy evaluated at laparoscopy. Obstet lation failure in laparoscopic instruments. Surg Endosc. 2010;24:
Gynecol. 1999;94:238 –242. 462– 465.
4. Jones CM, Pierre KB, Nicoud B, Stain SC, Melvin WV. Elec- 21. Odell RC. Electrosurgery: biophysics, safety and efficacy. In:
trosurgery. Curr Surg. 2006;63:458 – 463. Mann WJ T, Stovall G, eds. Mann/Stovall Gynecologic Surgery.
New York: Churchill Livingstone; 1996;55– 67.
5. Van Way CW, Hinrichs CS. Technology focus: electrosurgery
201:basic electrical principles. Curr Surg. 2000;57:261–264. 22. Vilos G, Latendresse K, Gan BS. Electrophysical properties
of electrosurgery and capacitive induced current. Am J Surg.
6. Valleylab. Principles of electrosurgery. 1999;1–23. 2001;182:222–225.
7. Advincula AP, Wang K. The evolutionary state of electrosur- 23. Soderstrom RM. Bowel injury litigation after laparoscopy.
gery: where are we now? Curr Opin Obstet Gynecol. 2008;20(4): J Am Assoc Gynecol Laparosc. 1993;1:74 –77.
353–358.
24. Levy BS, Soderstrom RM, Dail DH. Bowel injuries during
8. Massarweh N, Cosgriff N, Slakey D. Electrosurgery: history, laparoscopy: gross anatomy and histology. J Reprod Med. 1985;
principles, and current and future uses. Am Coll Surg. 2006;202: 30:168 –172.
520 –530.
25. Reich H. Laparoscopic bowel injury. Surg Laparosc Endosc.
9. Wang K, Advincula AP. Current thoughts in electrosurgery. 1992;2:74 –78.
Int J Gynaecol Obstet. 2007;97:245–250.
26. Saltzstein EC, Schwartz SF, Levinson CJ. Perforation of the
10. Neufield GR. Principles and hazards of electrosurgery in- small intestine secondary to laparoscopic tubal cauterization.
cluding laparoscopy. Surg Gynecol Obstet. 1978;147:705–710. Ann Surg. 1973;178:34 –37.
11. Miller CE. Basic equipment and room setup. In: Shirk GJ, 27. Mayooran Z, Pearce S, Tsaltas J, et al. Ignorance of electro-
Editor, The Video Encyclopedia of Endoscopic Surgery for the surgery among obstetricians and gynaecologists. BJOG. 2004;
Gynecologist. St Louis: Medical Video Productions; 1994; 111:1413–1418.
11–19.
28. Harold KL, Pollinger H, Matthews BD, Kercher KW, Sing RF,
12. Phipps JH. Understanding electrosurgery: safety and effi- Heniford BT. Comparison of ultrasonic energy, bipolar thermal
ciency. In: Lower A, Sutton Cand. Grudzinskas G, eds. Introduc- energy, and vascular clips for the hemostasis of small-, medium-,
tion to Gynecological Endoscopy. Oxford, UK: Iris Medical Me- and large-sized arteries. Surg Endosc. 2003;17:1228 –1230.
dia; 1996;39 –56.
29. Carbonell AM, Joels CS, Kercher KW, Matthews BD, Sing RF,
13. Tucker RD, Schimitt OH, Sievrt CE, et al. Demodulated low Heniford BT. A comparison of laparoscopic bipolar vessel seal-
frequency currents from electrosurgical procedures. Surg Gyne- ing devices in the hemostasis of small-, medium-, and large-sized
col Obstet. 1984;159:39 – 43. arteries. J Laparoendosc Adv Surg Tech A. 2003;13:377–380.
14. Martin DC, SoderstromRM, Hulka JF, et al. Electrosurgery 30. Pietrow PK, Weizer AZ, L’Esperance JO, Auge BK, Silverstein
safety. Am Assoc Gynecol Laparosc Tech Bull. 1995;1–7. A, Cummings T et al. Plasma kinetic bipolar vessel sealing: burst
pressures and thermal spread in an animal model. J Endourol.
15. Ryder RM, Hulka JF. Bladder and bowel injury after electro- 2005;19:107–110.
desiccation with Kleppinger bipolar forceps: a clinicopathologic
study. J Reprod Med. 1993;38:595–598. 31. Richter S, Kollmar O, Schilling MK, Pistorius GA, Menger
MD. Efficacy and quality of vessel sealing: comparison of a
16. Tucker RD, Volyes CR. Laparoscopic electrosurgical compli- reusable with a disposable device and effects of clamp surface
cations and their prevention. AORN. 1995;62:49 –78. geometry and structure. Surg Endosc. 2006;20:890 – 894.
17. Nduka CC, Super PA, Monson JR, Darzi AW. Cause and 32. Lamberton GR, Hsi RS, Jin DH, Lindler TU, Jellison FC,
prevention of electrosurgical injuries in laparoscopy. J Am Coll Baldwin DD. Prospective comparison of four laparoscopic ves-
Surg. 1994;179:161–170. sel ligation devices. J Endourol. 2008;22(10):2307–2312.
18. Hulka JF, Levy BS, Parker WH, Philips JM. Laparoscopic- 33. Hefni MA, Bhaumik J, El-Toukhy T, et al. Safety and efficacy
assisted vaginal hysterectomy: American Association of Gyneco- of using the LigaSure vessel sealing system for securing the
logic Laparoscopists 1995 membership survey. J Am Assoc Gy- pedicles in vaginal hysterectomy: randomised controlled trial.
necol Laparosc. 1997;4:167–171. BJOG. 2005;112(3):329 –333.