Design KVP Meter
Design KVP Meter
Abstract The kVp setting is one of the major factors affecting the image quality in X-ray imaging and should be an-
nually measured and calibrated if necessary. In this work, a kVp-meter is designed around the ATmega16 (Atmel) micro-
controller, based on the physical principle that the linear attenuation coefficient of materials, namely copper has a smooth
dependence on the energy level of the X-ray photons. Based on the logarithm of the ratio of the radiation intensities
through 0.5mm and 1mm thick copper filters, a look-up table is generated in the range 60-120kVp. Logarithmic operation
increased the precision at higher kVp values. Since sampling is performed over the exposure period in a continuous manner,
the measurement is not affected by the X-ray waveform. A prototype unit was built and the performance was tested in
terms of accuracy, precision and reliability.
Keywords kVp Meter, Linear Attenuation Coefficient, X-ray Imaging
cassettes. 0
Electronic kVp-meters provide a measurement based on 20 40 60 80 100 120
the change in X-ray transmission through varying thick-
Photon Energy (keV)
nesses of filtration. kVp-meters in general do not measure
the true peak voltage; but rather an effective kV value by Figure 1. The relation between the linear attenuation coefficient of cop-
per and the incident photon energy.
integrating the detector's outputs over exposure time and
then taking their ratio. The instrument then corrects this However this behaviour is valid only for monochromatic
value, usually according to a switch set by the user to indi- X-ray beam. In practice, the energy spectrum generated at a
cate what type of X-ray machine (1Φ or 3Φ or DC) is being certain kVp exhibits a continuous behaviour called Brehmss
measured, to give an effective kVp by knowing the amount trahlung continuum up to the largest possible keV level
of ripple in the waveform. which is in fact the kVp, with a few peaks called characte-
The method employed in this article relays on the same ristic X-rays. Filtration eliminates the relatively lower
principle, namely the dependence of the linear attenuation energy photons and the effective peak level is shifted to the
coefficient µ of copper on the energy level of the X-ray pho right, where the overall intensity, namely the number of
photons decreases. Therefore it is important to have a
* Corresponding author: µ-kVp relationship rather than µ-photon energy.
[email protected] (Y. Ülgen)
Published online at http://journal.sapub.org/ajbe
Copyright © 2011 Scientific & Academic Publishing. All Rights Reserved 2. Method and Measurement
42 Y. Ülgen et al.: Design of a Microcontroller Based and X-Ray Waveform Independent kVp-Meter
2.2
2.0
1.8
1.6
1.4
1.2
1.0
60 70 80 90 100 110 120
Tube voltage (kVp)
surement, the microcontroller finds the corresponding kVp To test the dependency of the measurements on the ex-
value between the reference values. posure level, several measurements were conducted at con-
C. Logarithm Operation stant kVp, but varying mAs (5 to 60mAs). There were no
The ratio I1/I2 is already informative on the kVp level deviations in the readings. It is concluded that the device
since logarithm and division are reversible operations. Most measurements are independent of mAs settings on the X-ray
of the previous designs used this ratio and a look-up table to unit over a wide range of mAs.
find the corresponding kVp. The graph of this ratio vs. kVp
Table 1. Accuracy of the KVP-Meter
exhibits a more curved behaviour which may result in con-
gested values in the look-up table for the higher kVp levels. kVp-meter
set kVp error %
reading
Logarithmic operation on the other hand, makes the graph
60 60.0 0.0
more linear, thus increases the precision especially at higher
61.5 61.5 0.0
kVp levels (Figure 5).
63 62.9 0.2
64.5 64.5 0.0
66 65.8 0.3
70 69.9 0.1
73 72.8 0.3
75 74.8 0.3
77 76.8 0.3
79 78.7 0.4
81 80.7 0.4
85 84.6 0.5
90 88.4 1.8
96 94.2 1.9
Figure 4. Placement of the photodetectors inside a lead case to allow
only beams from the X-ray source to reach the photodiodes. 102 99.8 2.2
109 106.5 2.3
4.0
1.4 113 109.9 2.7
3.5 117 113.6 2.9
ratio 1.1
121 117.4 3.0
ln ratio
3.0
ratio