Beneview T5/T6/T8: Patient Monitor
Beneview T5/T6/T8: Patient Monitor
Beneview T5/T6/T8
PATIENT MONITOR
Physical Dimensions
BeneView T8
Monitor size: 400mm x 370mm x 193mm
14.6” x 15.7” x 7.6”
Weight: Less than 14.5kg (31.9lbs) including 17" touchscreen, MPM,
AG module, two lithium batteries, a recorder
BeneView T6
Monitor size: 400mm x 370mm x 193mm
14.6” x 15.7” x 7.6”
Weight: Less than 9.8kg (21.6lbs) including 15" touchscreen, MPM
BeneView T5
Monitor size: 297mm x 336mm x 186mm
11.7” x 13.2” x 7.3”
Weight: Less than 6.6kg (15.8lbs) including 12.1" touchscreen
Display
Type: BeneView T8: 17” color TFT
BeneView T6: 15” color TFT
BeneView T5: 12.1” color TFT Unspecified (251-300 bpm)
Resolution: BeneView T8/T6: 1280 x 1024 pixels IBP Module: ±1bpm or ±1%, whichever is greater
BeneView T5: 800 x 600 pixels NIBP Module: ±3bpm or ±3%, whichever is greater
Waveforms: BeneView T8/T6: up to 12 Resolution 1 bpm
BeneView T5: up to 8 Refreshing rate 1s
External display: 1 display through DVI_D connector
2 displays through RDD NIBP
Method Automatic Oscillometric
ECG Operation mode Manual, Auto, STAT
Leads: 3-lead: I, II, III Parameters Systolic, Diastolic, Mean
5-lead: I, II, III, aVR, aVL, aVF, V Systolic range Adu: 40 to 270 mmHg
12-lead: I, II, III, aVR, aVL, aVF, V1 to V6 Ped: 40 to 200 mmHg
Gain: x0.125, x0.25, x0.5, x1, x2, x4, Auto Neo: 40 to 135 mmHg
Sweep speed: 6.25 mm/s, 12.5 mm/s, 25 mm/s, 50 mm/s Diastolic range Adu: 10 to 210 mmHg
Bandwidth: Diagnostic Mode: 0.05-150Hz Ped: 10 to 150 mmHg
Monitor Mode: 0.5-40Hz Neo: 10 to 100 mmHg
Surgical Mode: 1-20Hz Mean range Adu: 20 to 230 mmHg
Defib.protection: Withstand 5000V (360J)defibrillation Ped: 20 to 165 mmHg
Recovery time: ≤10 s Neo: 20 to 110 mmHg
CMRR: Diagnostic Mode: ≥90dB Accuracy Max mean error: ±5 mmHg
Monitor Mode: ≥105dB Max standard deviation: 8 mmHg
Surgical Mode: ≥105dB (Notch filter set to off ) Resolution 1 mmHg
ST analysis: -2.0 to 2.0 mV
Arr analysis: Yes Temperature
Range 0 to 50 ℃ (32 to 122 F)
Heart Rate Resolution 0.1 ℃
Range: Adu: 15 to 300 bpm
Accuracy ±0.1 ℃ or ±0.2 F (without probe)
Ped: 15 to 350 bpm
Neo: 15 to 350 bpm Parameters T1, T2 and △ T
Resolution: 1 bpm
Accuracy: ±1 bpm or ±1%, whichever is greater IBP
Channel up to 8 channels
Respiration Range -50 to 300 mmHg
Range: Adu: 0 to 120 rpm Resolution 1 mmHg
Ped/Neo: 0 to 150 rpm Accuracy ±2% or ±1 mmHg, whichever is greater (without sensor)
Resolution: 1 rpm Sensitivity 5 uV/mmHg/V
Accuracy: 7 to 150 rpm: ±2 rpm or ±2%, whichever is greater Impedance range 300 to 3000Ω
0 to 6 rpm: Not specified
C.O.
Lead: I or II (default: lead II)
Method Thermodilution
Sweep speed: 6.25 mm/s, 12.5 mm/s or 25 mm/s
Range C.O.: 0 to 30 L/min
SpO 2 TB: 23 to 43 ℃
Range: 0 to 100% TI: 0 to 27 ℃
Resolution: 1% Accuracy C.O.: ±5% or ±0.1 L /min, whichever is greater
Mindray accuracy: ±2% (70-100%, Adu/Ped, non-motion) TB, TI: ±0.1 ℃ (without sensor)
±3% (70-100%, Neo, non-motion)
±3% (70-100%, motion) Resolution: C.O.: 0.1 L/min
Unspecified (0-69%) TB, TI: 0.1 ℃
Masimo accuracy ±2% (70-100%, Adu/Ped, non-motion)
±3% (70-100%, Neo, non-motion) Sidestream CO 2
±3% (70-100%, motion) CO 2 Rang e 0 to 99 mmHg
Unspecified (0-69%) Accurac y 0 to 40 mmHg : ±2 mmHg
Nellcor accuracy Actual accuracy depends on probe. 41 to 76 mmHg : ±5% of the reading
Refer to the operator’s manual. 77 to 99 mmHg : ±10% of the reading
Refreshing rate 1s Sample flowrat e 70, 100 ml/min
Accurac y ±15% or ±15 ml/min, whichever is greater.
Pulse rate Warm-up tim e <1 min, enter the ISO accuracy mode
Range Mindray SpO 2: 20 to 254 bpm After 1 min, enters the full accuracy mode
Masimo SpO 2: 25 to 240 bpm awRR rang e 0 to 120 rpm
Nellcor SpO 2: 20 to 300 bpm awRR precisio n 0 to 70 rpm : ±2 rpm
IBP Module: 25 to 350 bpm 71 to 120 rpm : ±5 rpm
NIBP Module: 40 to 240 bpm Response tim e When using neonatal watertrap and 2.5 m
Accuracy Mindray SpO 2: ±3 bpm (non-motion) neonatal sampling line
±5 bpm (motion) <3.5 s @ 100 ml/min
Masimo SpO 2: ±3 bpm (non-motion) <4 s @ 70 ml/min
±5 bpm (motion)
Nellcor SpO 2: ±3 bpm (20-250 bpm)
When using adult watertrap and 2.5 m adult Flow accuracy Adu/Ped: 1.2 L/min or ±10% of the reading, whichever is greater
sampling line Inf: 0.5 L/min or ±10% of the reading, whichever is greater
<5.5 s @ 100 ml/min Resolution 0.1 L/min
<7 s @ 70 ml/min Paw range -20 to 120 cmH 2O
Apnea tim e 10 s, 15 s, 20 s, 25 s, 30 s, 35 s, 40 s Paw accuracy ±3%
Resolution 0.1 cmH2 O
Microstream CO 2 MVe/MVi range: Adu/Ped: 2 to 60 L/min
CO 2 Rang e 0 to 99 mmHg Inf: 0.5 to 15 L/min
Accurac y 0 to 38 mmHg : ±2 mmHg MVe/MVi accuracy ±10%×reading
39 to 99 mmHg : ±5% of the reading+0.08% of TVe/TVi range: Adu/Ped: 100 to 1500 ml
(the reading-38) Inf: 20 to 500 ml
+15
Sample flow rate 50 -7.5 ml/min TVe/TVi accuracy Adu/Ped: ±10% or 15 ml, whichever is greater
Initialization tim e 30 s (typical) Inf: ±10% or 6 ml, whichever is greater
awRR rang e 0 to 150 rpm Resolution 1 ml
awRR precisio n 0 to 70 rpm : ±1 rpm RR range 4 to 120 rpm
71 to 120 rpm : ±2 rpm RR accuracy ±1 rpm (4 to 99 rpm)
121 to 150 rpm : ±3 rpm ±2 rpm (100 to 120 rpm)
Response tim e 2.9 s (typical) I:E 4:1 to 1:8
Apnea tim e 10 s, 15 s, 20 s, 25 s, 30 s, 35 s, 40 s FEV1.0% 0 to 100%
Others: Pmean, TV, MV, PEEP, PEF, PIF, PIP, Pplat
Mainstream CO 2 Compl, Raw, RSBI, NIP, WOB
CO 2Rang e 0 to 150 mmHg
Accurac y 0 to 40 mmHg : ±2 mmHg CCO/Sv O 2
41 to 70 mmHg : ±5% of the reading Mode Interfaces with Edwards Vigilance II monitor
71 to 100 mmHg : ±8% of the reading Display CCO, EDV, SVR, SV, SvO2 , SQI, DO2 , etc.
101 to 150 mmHg: ±10% of the reading *: Specifications depend on Vigilance II monitor.
awRR rang e 0 to 150 rpm
Accurac y 1 rpm Data Storage
Response tim e <60 ms Trend data: 120 hrs @ 1 min, 4 hrs @ 5 sec
Alarm events: 100 events and associated waveforms
Multi-gas/O 2 Arr. events: 100 Arr. events and associated waveforms
Method Infrared absorption NIBP: 1000 measurements
Gas : CO 2, O 2, N 2O, Des, Iso, Enf, Hal, Sev Waveforms: Max. 48 hrs full disclosure waveforms
Warm-up tim e ISO accuracy mode : 45 s (specific storage time depends on the type and number
Full accuracy mode : 10 min of waveforms stored)
Sample flow rat e Adu/Ped : 120, 150, 200 ml/min
Neo : 70, 90, 120 ml/min Battery
Accurac y ±10 ml/min or ±10%, whichever is greater Type: Chargeable Lithium-Ion
Rang e Co 2: 0 to 30% Number: 2
O 2/N 2O : 0 to 100% Voltage: 11.1 VDC
Hal/Iso/Enf : 0 to 30% Capacity: 4500 mAh
Des : 0 to 30% Run time: T6/T8: 2 hrs
Sev : 0 to 30% T5: 5.5 hrs
awRR rang e 2 to 100 rpm Recharge time: 6 hrs maximum
awRR accurac y 2 to 60 rpm: ±1 rpm
>60 rpm: unspecified Interfacing
Apnea time 10 s, 15 s, 20 s, 25 s, 30 s, 35 s, 40 s Connectors: 1 AC power connector
1 RJ45 network connector (T5)
ICG 2 RJ45 network connectors (T8/T6)
Technique Thoracic electrical bioimpedance (TEB) 1 USB SMR connector
Range SV: 5 to 250 ml/beat USB 1.1 connectors
HR: 44 to 185 bpm 1 CF revision 2.0 connector
C.O.: 1.4 to 15 L/min 1 standard DVI_D video interface connector
Accuracy SV: Not specified 1 BNC connector
HR : ±2 bpm 1 Micro-D connector (output ECG, IBP and Defib.
C.O. : Not specified Synch. Signals)
Recorder
BIS Type: Thermal array
Technique: Bispectral index Speed: 25 mm/s, 50 mm/s
BIS range 0 to 100 Trace: 3
SQI range 0% to 100%
Impedance range 0 to 999 kΩ Power Requirements
EEG bandwidth 0.25 to 100 Hz AC Voltage: 100 to 240 VAC, 50/60Hz
Current: T8/T6: 2.8 to 1.6 A
RM T5: 2.5 to 1.4 A
Flow range Adu/Ped: ± (2 to 120) L/min
Inf: ± (0.5 to 30) L/min
Masimo is the trademark owned by Masimo Corperation.Nellcor is the trademark owned by Nellcor Puritan Bennett LLC
9
P/N:ENG-BeneView/T6/T8-210285-20090818