This document provides specifications for the Beneview T5, T6, and T8 patient monitors. The T8 is the largest monitor with a 17" display, weighing up to 14.5kg. The T6 has a 15" display and weighs up to 9.8kg. The smallest, the T5, has a 12.1" display and weighs up to 6.6kg. All monitors can display ECG, respiration, SpO2, NIBP, IBP, temperature and other parameters. Features include arrhythmia analysis, defibrillation protection, and connectivity to external displays. Accuracy and ranges are provided for all measured parameters.

Uploaded by

fbeerc
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
67 views

Beneview T5/T6/T8: Patient Monitor

This document provides specifications for the Beneview T5, T6, and T8 patient monitors. The T8 is the largest monitor with a 17" display, weighing up to 14.5kg. The T6 has a 15" display and weighs up to 9.8kg. The smallest, the T5, has a 12.1" display and weighs up to 6.6kg. All monitors can display ECG, respiration, SpO2, NIBP, IBP, temperature and other parameters. Features include arrhythmia analysis, defibrillation protection, and connectivity to external displays. Accuracy and ranges are provided for all measured parameters.

Uploaded by

fbeerc
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 2

TM

Beneview T5/T6/T8
PATIENT MONITOR
Physical Dimensions
BeneView T8
Monitor size: 400mm x 370mm x 193mm
14.6” x 15.7” x 7.6”
Weight: Less than 14.5kg (31.9lbs) including 17" touchscreen, MPM,
AG module, two lithium batteries, a recorder
BeneView T6
Monitor size: 400mm x 370mm x 193mm
14.6” x 15.7” x 7.6”
Weight: Less than 9.8kg (21.6lbs) including 15" touchscreen, MPM
BeneView T5
Monitor size: 297mm x 336mm x 186mm
11.7” x 13.2” x 7.3”
Weight: Less than 6.6kg (15.8lbs) including 12.1" touchscreen

Display
Type: BeneView T8: 17” color TFT
BeneView T6: 15” color TFT
BeneView T5: 12.1” color TFT Unspecified (251-300 bpm)
Resolution: BeneView T8/T6: 1280 x 1024 pixels IBP Module: ±1bpm or ±1%, whichever is greater
BeneView T5: 800 x 600 pixels NIBP Module: ±3bpm or ±3%, whichever is greater
Waveforms: BeneView T8/T6: up to 12 Resolution 1 bpm
BeneView T5: up to 8 Refreshing rate 1s
External display: 1 display through DVI_D connector
2 displays through RDD NIBP
Method Automatic Oscillometric
ECG Operation mode Manual, Auto, STAT
Leads: 3-lead: I, II, III Parameters Systolic, Diastolic, Mean
5-lead: I, II, III, aVR, aVL, aVF, V Systolic range Adu: 40 to 270 mmHg
12-lead: I, II, III, aVR, aVL, aVF, V1 to V6 Ped: 40 to 200 mmHg
Gain: x0.125, x0.25, x0.5, x1, x2, x4, Auto Neo: 40 to 135 mmHg
Sweep speed: 6.25 mm/s, 12.5 mm/s, 25 mm/s, 50 mm/s Diastolic range Adu: 10 to 210 mmHg
Bandwidth: Diagnostic Mode: 0.05-150Hz Ped: 10 to 150 mmHg
Monitor Mode: 0.5-40Hz Neo: 10 to 100 mmHg
Surgical Mode: 1-20Hz Mean range Adu: 20 to 230 mmHg
Defib.protection: Withstand 5000V (360J)defibrillation Ped: 20 to 165 mmHg
Recovery time: ≤10 s Neo: 20 to 110 mmHg
CMRR: Diagnostic Mode: ≥90dB Accuracy Max mean error: ±5 mmHg
Monitor Mode: ≥105dB Max standard deviation: 8 mmHg
Surgical Mode: ≥105dB (Notch filter set to off ) Resolution 1 mmHg
ST analysis: -2.0 to 2.0 mV
Arr analysis: Yes Temperature
Range 0 to 50 ℃ (32 to 122 F)
Heart Rate Resolution 0.1 ℃
Range: Adu: 15 to 300 bpm
Accuracy ±0.1 ℃ or ±0.2 F (without probe)
Ped: 15 to 350 bpm
Neo: 15 to 350 bpm Parameters T1, T2 and △ T
Resolution: 1 bpm
Accuracy: ±1 bpm or ±1%, whichever is greater IBP
Channel up to 8 channels
Respiration Range -50 to 300 mmHg
Range: Adu: 0 to 120 rpm Resolution 1 mmHg
Ped/Neo: 0 to 150 rpm Accuracy ±2% or ±1 mmHg, whichever is greater (without sensor)
Resolution: 1 rpm Sensitivity 5 uV/mmHg/V
Accuracy: 7 to 150 rpm: ±2 rpm or ±2%, whichever is greater Impedance range 300 to 3000Ω
0 to 6 rpm: Not specified
C.O.
Lead: I or II (default: lead II)
Method Thermodilution
Sweep speed: 6.25 mm/s, 12.5 mm/s or 25 mm/s
Range C.O.: 0 to 30 L/min
SpO 2 TB: 23 to 43 ℃
Range: 0 to 100% TI: 0 to 27 ℃
Resolution: 1% Accuracy C.O.: ±5% or ±0.1 L /min, whichever is greater
Mindray accuracy: ±2% (70-100%, Adu/Ped, non-motion) TB, TI: ±0.1 ℃ (without sensor)
±3% (70-100%, Neo, non-motion)
±3% (70-100%, motion) Resolution: C.O.: 0.1 L/min
Unspecified (0-69%) TB, TI: 0.1 ℃
Masimo accuracy ±2% (70-100%, Adu/Ped, non-motion)
±3% (70-100%, Neo, non-motion) Sidestream CO 2
±3% (70-100%, motion) CO 2 Rang e 0 to 99 mmHg
Unspecified (0-69%) Accurac y 0 to 40 mmHg : ±2 mmHg
Nellcor accuracy Actual accuracy depends on probe. 41 to 76 mmHg : ±5% of the reading
Refer to the operator’s manual. 77 to 99 mmHg : ±10% of the reading
Refreshing rate 1s Sample flowrat e 70, 100 ml/min
Accurac y ±15% or ±15 ml/min, whichever is greater.
Pulse rate Warm-up tim e <1 min, enter the ISO accuracy mode
Range Mindray SpO 2: 20 to 254 bpm After 1 min, enters the full accuracy mode
Masimo SpO 2: 25 to 240 bpm awRR rang e 0 to 120 rpm
Nellcor SpO 2: 20 to 300 bpm awRR precisio n 0 to 70 rpm : ±2 rpm
IBP Module: 25 to 350 bpm 71 to 120 rpm : ±5 rpm
NIBP Module: 40 to 240 bpm Response tim e When using neonatal watertrap and 2.5 m
Accuracy Mindray SpO 2: ±3 bpm (non-motion) neonatal sampling line
±5 bpm (motion) <3.5 s @ 100 ml/min
Masimo SpO 2: ±3 bpm (non-motion) <4 s @ 70 ml/min
±5 bpm (motion)
Nellcor SpO 2: ±3 bpm (20-250 bpm)
When using adult watertrap and 2.5 m adult Flow accuracy Adu/Ped: 1.2 L/min or ±10% of the reading, whichever is greater
sampling line Inf: 0.5 L/min or ±10% of the reading, whichever is greater
<5.5 s @ 100 ml/min Resolution 0.1 L/min
<7 s @ 70 ml/min Paw range -20 to 120 cmH 2O
Apnea tim e 10 s, 15 s, 20 s, 25 s, 30 s, 35 s, 40 s Paw accuracy ±3%
Resolution 0.1 cmH2 O
Microstream CO 2 MVe/MVi range: Adu/Ped: 2 to 60 L/min
CO 2 Rang e 0 to 99 mmHg Inf: 0.5 to 15 L/min
Accurac y 0 to 38 mmHg : ±2 mmHg MVe/MVi accuracy ±10%×reading
39 to 99 mmHg : ±5% of the reading+0.08% of TVe/TVi range: Adu/Ped: 100 to 1500 ml
(the reading-38) Inf: 20 to 500 ml
+15
Sample flow rate 50 -7.5 ml/min TVe/TVi accuracy Adu/Ped: ±10% or 15 ml, whichever is greater
Initialization tim e 30 s (typical) Inf: ±10% or 6 ml, whichever is greater
awRR rang e 0 to 150 rpm Resolution 1 ml
awRR precisio n 0 to 70 rpm : ±1 rpm RR range 4 to 120 rpm
71 to 120 rpm : ±2 rpm RR accuracy ±1 rpm (4 to 99 rpm)
121 to 150 rpm : ±3 rpm ±2 rpm (100 to 120 rpm)
Response tim e 2.9 s (typical) I:E 4:1 to 1:8
Apnea tim e 10 s, 15 s, 20 s, 25 s, 30 s, 35 s, 40 s FEV1.0% 0 to 100%
Others: Pmean, TV, MV, PEEP, PEF, PIF, PIP, Pplat
Mainstream CO 2 Compl, Raw, RSBI, NIP, WOB
CO 2Rang e 0 to 150 mmHg
Accurac y 0 to 40 mmHg : ±2 mmHg CCO/Sv O 2
41 to 70 mmHg : ±5% of the reading Mode Interfaces with Edwards Vigilance II monitor
71 to 100 mmHg : ±8% of the reading Display CCO, EDV, SVR, SV, SvO2 , SQI, DO2 , etc.
101 to 150 mmHg: ±10% of the reading *: Specifications depend on Vigilance II monitor.
awRR rang e 0 to 150 rpm
Accurac y 1 rpm Data Storage
Response tim e <60 ms Trend data: 120 hrs @ 1 min, 4 hrs @ 5 sec
Alarm events: 100 events and associated waveforms
Multi-gas/O 2 Arr. events: 100 Arr. events and associated waveforms
Method Infrared absorption NIBP: 1000 measurements
Gas : CO 2, O 2, N 2O, Des, Iso, Enf, Hal, Sev Waveforms: Max. 48 hrs full disclosure waveforms
Warm-up tim e ISO accuracy mode : 45 s (specific storage time depends on the type and number
Full accuracy mode : 10 min of waveforms stored)
Sample flow rat e Adu/Ped : 120, 150, 200 ml/min
Neo : 70, 90, 120 ml/min Battery
Accurac y ±10 ml/min or ±10%, whichever is greater Type: Chargeable Lithium-Ion
Rang e Co 2: 0 to 30% Number: 2
O 2/N 2O : 0 to 100% Voltage: 11.1 VDC
Hal/Iso/Enf : 0 to 30% Capacity: 4500 mAh
Des : 0 to 30% Run time: T6/T8: 2 hrs
Sev : 0 to 30% T5: 5.5 hrs
awRR rang e 2 to 100 rpm Recharge time: 6 hrs maximum
awRR accurac y 2 to 60 rpm: ±1 rpm
>60 rpm: unspecified Interfacing
Apnea time 10 s, 15 s, 20 s, 25 s, 30 s, 35 s, 40 s Connectors: 1 AC power connector
1 RJ45 network connector (T5)
ICG 2 RJ45 network connectors (T8/T6)
Technique Thoracic electrical bioimpedance (TEB) 1 USB SMR connector
Range SV: 5 to 250 ml/beat USB 1.1 connectors
HR: 44 to 185 bpm 1 CF revision 2.0 connector
C.O.: 1.4 to 15 L/min 1 standard DVI_D video interface connector
Accuracy SV: Not specified 1 BNC connector
HR : ±2 bpm 1 Micro-D connector (output ECG, IBP and Defib.
C.O. : Not specified Synch. Signals)
Recorder
BIS Type: Thermal array
Technique: Bispectral index Speed: 25 mm/s, 50 mm/s
BIS range 0 to 100 Trace: 3
SQI range 0% to 100%
Impedance range 0 to 999 kΩ Power Requirements
EEG bandwidth 0.25 to 100 Hz AC Voltage: 100 to 240 VAC, 50/60Hz
Current: T8/T6: 2.8 to 1.6 A
RM T5: 2.5 to 1.4 A
Flow range Adu/Ped: ± (2 to 120) L/min
Inf: ± (0.5 to 30) L/min

Masimo is the trademark owned by Masimo Corperation.Nellcor is the trademark owned by Nellcor Puritan Bennett LLC