Portaflow C Software Manual
Portaflow C Software Manual
PORTAFLOW C
PC LOADER SOFTWARE
INSTRUCTION MANUAL
INF-TN5A1784-E
Contents
1. OUTLINE ·········································································································································································1
1.1. COPYRIGHT OF THIS SOFTWARE·························································································································1
1.2. GENERAL ·································································································································································1
1.3. SYSTEM REQUIREMENTS······································································································································1
1.3.1. PC main unit ·······················································································································································1
1.3.2. Memory capacity·················································································································································1
1.3.3. Interface······························································································································································1
1.3.4. OS ···································································································································································1
1.4. SCOPE OF DELIVERY ·············································································································································1
3. Communication Specification·········································································································································15
6. Measurement·································································································································································19
7. Logger data····································································································································································21
9. Maintenance ··································································································································································27
9.1. RECEIVED SIGNAL················································································································································28
9.2. Adjustment of graph scale ·······································································································································29
13. APPENDIX···································································································································································34
13.1. If an USB device is recognized as a COM port ·····································································································34
13.1.1. How to check ··················································································································································34
13.1.2. Restoration procedure ····································································································································36
13.1.3. Check after installation····································································································································40
Thank you very much for purchasing our PortaFlow-C Loader Software.Read this instruction manual carefully before using
this personal computer communication software.
-i-
1. OUTLINE
1.1. COPYRIGHT OF THIS SOFTWARE
The copyright of this software belongs to Fuji Electric Systems Co., Ltd.No part of this software may be reproduced or
transmitted in any form.
1.2. GENERAL
Using this software, you can set, read and display relevant graphs of the PortaFlow-C on your PC with ease.
1.3.3. Interface
More than one USB port is necessary.
1.3.4. OS
Microsoft Windows 2000 Professional (SP4 or more)
Microsoft Windows XP Professional (SP2 or more)
Microsoft Windows Vista Susiness: Use it in Basic mode.
It is not avairable for Windows Aero.
Microsoft Windows 7 (Home Premium, Professional)
-1-
2. INSTALLING OF SOFTWARE
2.1. Installing Microsoft Windows Installer 3.1
Note) this software is not required for PC which already installed Windows installer.
(1) Insert the setup disk into the drive, and double-click “WindowsInstaller-KB893803-v2-x86.exe” in the
WindowsInstaller3.1 folder.
(2) Setting wizard will start up. Click the [Next] button.Click the [Cancel] button to cancel the installation.
-2-
(3) There is a query about the License Agreement, check the [I Agree] and click the [Next] button.Click the [Cancel]
button to cancel the installation.The [Back] and the [Print] buttons are not used here.
(4) Screen for updating your system is displayed.Wait until installation is completed.
-3-
(5) After installation, the installation complete screen appears. Check the “Do not restart new”, click the [Finish]
button.That is the end of installation for Microsoft Windows Installer 3.1.
-4-
2.2. Installing Microsoft .NET Framework Version 2.0
(1) Insert the setup disk into the drive, and double-click “dotnetfx.exe” in the FrameWork2.0 folder.
(2) Setting wizard will start up. Click the [Next] button. Click the [Cancel] button to cancel the installation.
-5-
(3) There is a query about the End-User Liecemse Agreement, check the [I accept the terms of the License Agreement]
and click the [Install] button. Click the [Cancel] button to cancel the installation. The [< Back] and the [Print] buttons
are not used here.
(4) Screen is displayed to confirm installation. Click the [Cancel] button to cancel the installation.
-6-
(5) After installation, the Completion of Process Setting screen appears. Click the [Finish] button, and the installation of
Microsoft .NET Framework Version 2.0 is completed.
-7-
(2) Installer will start up. Click the [Install] button. Click the [Cancel] button to cancel the installation. The [Change Install
Location] button is not used.
(3) If the warning on the windows logo test is indicated, click the [Continue Anyway] button to continue the operation.
(4) After installation, the installation complete screen appears. Click the [OK] button, and the installation of USBExpress
is completed.
-8-
2.4. Device Recognition
(1) After the setup in “Section 2.1, 2.2 and 2.3” are normally completed, connect the PortaFlow-C to the PC with USB
cable. Found New Hardware Wizard will start up. Check the [Install from a list or specific location (Advanced) (S)]
button and click the [Next] button. Click the [Cancel] button to cancel the installation.
(2) The Found New Hardware Wizard appears. Check the [Don’t search. I will choose the driver to install (D)] and click
the [Next] button. Click the [Cancel] button to cancel the installation. The [Back (B)] button is not used here.
-9-
(3) The Found New Hardware Wizard appears. Select the device driver to be installed for this hardware. Select the
[USBExpress Driver] indicated in the [Model], and click the [Next] button. Click the [Cancel] button to cancel the
installation. The [Back] button is not used here.
(4) If the warning on the windows logo test is indicated, click the [Continue Anyway] button to continue the operation.
Screen is displayed to confirm installation. Wait until installation is completed.
-10-
Fig. 2-18 Installation confirmation screen
(5) After installation, the installation complete screen appears. Click the [OK] button, and the device recognition is
completed.
-11-
2.5. Installing of Loader Software
(1) Insert the setup disk into the drive, and double-click “PFC_SETUP_ENG.msi”.
(2) Setting wizard will start up. Click the [Next] button. Click the [Cancel] button to cancel the installation.
-12-
(3) There is a query about selection of installation folder. Click the [Next] button to install the software in that folder. Click
the [Cancel] button to cancel the installation. The [Back (B)], the [Print (P)] and the [Disk range (D)] buttons are not
used here.
(4) Screen is displayed to confirm installation. Click the [Next] button to execute the installation. Click the [Cancel] button
to cancel the installation. The [Back (B)] button is not used here.
-13-
(5) Execute installation. Click the [Cancel] button to cancel the installation.
(6) The Installation Complete screen is displayed. Click the [Close] button to exit the installation screen.
(7) After installation, the start menu and the application (“PortaFlow-C”) that has been installed in the desktop are
created.
-14-
3. Communication Specification
See the following table for communication specifications.
-15-
4. Startup Method
Select “PORTAFLOW-C” from the Start menu and the following Menu screen appears.
Click the Main menu and each function button to execute a desired function
-16-
4.1. Version
On the Menu screen, click “Version” of the main menu item, and the following screen appears.
-17-
5. Setting Screen
Set the check box of each setting item to ON “” so that the item can be input or selected. When it is set to OFF “”, the
item cannot be input or selected. Click the [Set up] and send the setting value of the item (check box set to ON “”) one by
one, and reflect the response value to the setting value. Click the [load] and read the setting value of the item (check box
set to ON “”), and reflect the response value to the setting value. Click the [Save] and write the current setting value into
Flash ROM. If you change your setting, this operation above must be required.
When the unit in the system setting is changed, it reflects the corresponding item. Be sure to reflect the setting value or the
range unit after reading is completed. (The flowmeter returns values in a unit specified.) The values are displayed in a set
unit. (Read or change the unit when the flowmeter is started. You cannot change the unit by key operation at the loader.)
In the case of the unit whose input range is limited, the input value is not checked by the loader and sent as it is. The range
is checked on the flowmeter side. If the range is exceeded, the flowmeter does not update the request and respond the
current value.
CAUTION
For changing Flow meter setting, make sure to switch from flow meter screen to measurement
screen.
If the setting change is operated except measurement screen, communication error will occur
-18-
6. Measurement
Click the “MEASURE” button on the Menu screen, and the following screen appears.
First, select instantaneous trend value from [FLOW RATE], [FLOW RATE%], [VELOCITY], [THERMAL FLOW],
[THERMAL FLOW%], [ANALOG INPUT1], [ANALOG INPUT2], [FEEDING TEMP], [RETURNING TEMP] and
[DIFFEREIAL TEMP].
Up to 2 instantaneous values can be selected.
Click the [SCALE SETTING] button to set [Max.] and [Min.] of Y scale and [Cycle] and [Point] of X scale. Press the
[Setting] button to reflect the setting.Press the [Cancel] button to cancel the setting. When setting is not made, the initial
value is reflected.
-19-
Click the [Save as CSV] button and you are prompted to enter the name of a file to which the data is to be saved.
Specify the destination to save and enter the file name, and a CSV file is created. When the save file is set, the
[START] button can be pressed.If number of data exceeds 65531, create a separate file to save the data.Note that the
number of files to be created (serial number) is up to 500.
Example) Setting file name _ YYYYMMDDHHMMSS _ XXX (3-digit serial number. 1 START)
Click the [START] button, and the trend is displayed (X-axis is a collection time. When the specified number of items is
stored, the oldest data is deleted and the time display is changed so that the latest value is displayed.), the data is read in
the specified cycle, and the following are updated: [FLOW RATE], [FLOW RATE%], [VELOCITY], [THERMAL FLOW],
[THERMAL FLOW%], [ANALOG INPUT1], [ANALOG INPUT2], [FEEDING TEMP], [RETURNING TEMP], [DIFFERELAL
TEMP], [+TOTAL(ACTUAL)], [–TOTAL(ACTUAL)], [+TOTAL(THERMAL)], [–TOTAL(THERMAL)], [RAS INFO.].The line of
graph in the display 1 is blue and that in the display 2 is red. For stop collecting, click [Stop].
-20-
7. Logger data
Click the ”Logger data” button on the Menu screen, and the following screen appears.
CAUTION
The time of reading logger data to PC will take about 25 min. in case of 1.6MB.
The reading will take time for a while; regularly reading SD memory card directly via a PC is
recommended.
-21-
8. Site Setting
Click the “SITE SETUP” button on the Menu screen, and the following screen appears.
-22-
8.1. Site Memory
Select Tab Page 1 on the SITE SETUP screen and the following screen appears.
Table 8-1
Item Contents
Name list for site memory number 1 to 32.The number of lists to be
Site memory setting displayed is fixed to 32 and multiple selections are not allowed.
Check the site number 1 in case of initial display.
Site name Displays the name of site memory selected in the list.
Pressed when the site memory list is being acquired. Acquire the name
Reading of Site list list for site memory number 1 to 32 to be displayed in the site memory
setting.
Registers the site memory as a new entry. On the site memory setting list,
REGISTRATION select the site number that has not been registered in the site memory,
and press it after entering the site name.
Deletes the site memory. On the site memory setting list, select the site
DELETE
number registered in the site memory, and press it.
Press the [Setting] button, and set all the tab pages of the site setting only on this screen. The setting is performed only
for checked setting items in the [PROCESS SETTING] and [UNIT OF OUTPUT/OUTPUT CONTROL] pages.
Press the [READ] button, and set all the tab pages of the site setting only on this screen. The setting is performed only
for checked setting items in the [PROCESS SETTING] and [UNITOF OUTPUT/OUTPUT CONTROL] pages.
-23-
8.2. Establish Setting
Select Tab Page 2 on the SITE SETUP screen and the following screen appears.
-24-
Table 8-3 Sensor specifications
Item Contents
SENSOR MOUNT Select from Z method and V method.
TRANS. VOLTAGE Select from 20, 40, 80 and 160.
Select from
FSD12/FLD12, FSSD/FSD22/FLD22, FSSH, FSSC, FSSF,
FLD11/FSG31/FLW11, FSG32/FLW12, FSD32/FLD32/FLW32,
SENSOR TYPE
FSD41/FLD41/FSG41/FLW41, FSD51/FLD51/FSG51/FLW51,
FSSE/FSG50/FLW50, FSSA/FSSG, FLS_12/FLS_22, FSW12,
FSW21, FSW40 and FSW50.
The site name which belongs to reading/writing information is displayed in SITE NAME.
-25-
8.3. Output unit/Output control
Select Tab Page 3 on SITE SETUP screen and the following screen appears.
The site name which belongs to reading/writing information is displayed in the SITE NAME.
-26-
9. Maintenance
Click the “MAINTENANCE” button on the Menu screen, and the following screen appears.
-27-
9.1. RECEIVED SIGNAL
Select Tab Page 2 on the MAINTE screen and the following screen appears.
Received
Trigger level
signal
Select one from forward direction received wave, reverse direction received wave, forward direction filter, reverse
direction filter and calculation waveform.
Click the [START] button to read in the specified cycle and click the [Stop] button to cancel the operation. Display is
updated with each read.
Save the received waveform in the specified file by [SAVE As CSV] button. Trigger level is also displayed.
Display the RAS information in the STATUS DISPLAY.
Point
-28-
9.2. Adjustment of graph scale
For adjusting scale, right-click on “Measurement”, “Waves” or “Velocity” screen.
Example) For adjusting the scale of X axis, select X axis, input “Max.” and “Min.” on the scale tab. Click [OK] button.
-29-
10. Flow velocity profile
When 6th digit of the type of main unit is 1(Velocity profile measurement opton), this function should be available.
Click the“Velocity Profile” button on the Menu screen and the following screen appears.
-30-
10.1. Flow velocity profile
Select Tab Page 2 on the Velocity Profile screen and the following screen appears.
First, press the “GRAPH DISPLAY SETTING” button, and select either from VELOCITY / STANDARD DEVIATION/ POWER
or Moment / Moving Average in the following screen. In case of Moving Average, set the number of times and select one
from SENSOR-U/ SENSOUR-D/ SENSOUR-U/D on the graph display.
-31-
[KIND]
[VELOCITY]
[Collection]
Moment :Displays data by each read.
Moving Average :Displays data in moving average with the number of times set by channel in each read data.
[MODE]
SENSOR-U :Displays only upstream sensor that connected cable with the upstream side.
SENSOR-D :Displays only downstream sensor that connected cable with the upstream side.
SENSOR-U/D :Displays both upstream/downstream sensor above.
[Moment]
VEL. / F.RATE :Displays flow velocity or flow rate with each read.
[RAS]
RAS...Displaying RAS with each read.
[DEMO feature]
Reads the file logged by flow rate or [SAVE As CSV], displays the flow velocity profile.
[Check Box]
Displays repeatedly by setting check box to ON (“☑”).
[OFFSET Calculation]
Displays flow velocity profile detected by the sensor on the upper stream side and that on the lower stream side overlapped
at the center of the piping.
Click the [START] button to read in the specified cycle and click the [Stop] button to cancel the operation.
Click the [SAVE As CSV] button to save the data in the CSV file.
For saving format, refer to “16.5.3 Flow Velocity Profile” in the manual.
-32-
11. QUIT
Click the Quit on the main menu screen or [].
-33-
13. APPENDIX
13.1. If an USB device is recognized as a COM port
If there is an error of reorganization of a PC device driver, an USB device is not installed properly. In this case, the
communication between Portaflow-C and a loader software can not be performed.
This section provides the information about the restoration procedure how to make a PC recognize an USB device.
Fig. 13-1
-34-
(2) Select the [Ports (COM & LPT)] on the displayed screen.
Fig. 13-2
(3) Next, select the [Universal Serial Bus controllers].
Fig. 13-3
If the items of the Fig. 13-2 [CP2101 USB to UART Bridge Controller] and the Fig. 13-3 [CP2101 USB Composite Device]
are displayed when checked the error according to the procedure above, the restoration procedure provided in the next
section is applied.
-35-
13.1.2. Restoration procedure
Before proceeding to the following procedure, make sure to connect the PC used and the Portaflow-C with an USB cable.
(1) Right click on the Fig. 13-3, and then click the [Properties].
Fig. 13-4
Fig. 13-5
-36-
(3) The [Hardware Update Wizard] is launched.
Select the【Install from a list or specific location(Advanced)】, and click the【Next】button to transit to the [Hardware
Update Wizard].
Fig. 13-6
(4) Select the【Don’t search. I will choose the driver to install.】, and then click the【Next】button.
Fig. 13-7
-37-
(5) Click the【Have Disk…】button to install a device driver.
Fig. 13-8
Fig. 13-9
Fig. 13-10
-38-
(8) Now, installation starts. If the warning like the Fig. 13-12 concerning Windows Logo test appears during installation,
click【Continue Anyway】button.
Fig. 13-11
Fig. 13-12
(9) When installation has completed, the completion screen is displayed. Click【Finish】button. Now, restoration
procedure for USB communication error has completed.
Fig. 13-13
-39-
13.1.3. Check after installation
Check【Universal Serial Bus controller】in the Device Manager.
Check if the device type has changed from Fig. 13-3[CP2101 USB Composite Device]in the procedure “13.1.1. How to
check” to [USBXpress Device] that is highlighted in the Fig. 13-14 below.
Fig. 13-14
-40-