Error 4021, 4024 Siremobil Compac
Error 4021, 4024 Siremobil Compac
0
From the old Memoskop
1. SAVE IMAGES (optional)
Turn the system on.
Save important patients to CD. They can be restored to the new system. (If CD feature is
bought from Siemens).
2. SAVE CONFIG (required)
Press CTRL + T to enter technical setup. Password = techuser
Insert another blank CD.
Press “S” to save system settings. Label the CD’s.
Save the CD’s and Dongle for the new system installation.
____________________________________________________________________________
______
Regards,
Correo de DOTMED
Meaning:
Gen. tried X-Ray on without XRAY_query signal
Possible trouble:
Generator attempts radiation but the signal XRAY_query was not
present.
Software error D1.
Possible help:
- Faults in the system.
- Sirephos defective
- D1 defective.
Error 4028
Meaning:
Memoskop unexpected acq end signal
Possible trouble:
Host registered unexpected acq end signal of the Memoskop.
Possible help:
- Check faults on the Memoskop acquisition cable (D1.X10.27).
- Check Memoskop.
- Check connection cables to the Memoskop.