Procedure For Determining Permeability of Rock Core - Flow Pump Permeability Test
Procedure For Determining Permeability of Rock Core - Flow Pump Permeability Test
1.Scope
2
USBR 6310-09
6.Apparatus
3
USBR 6258-09
4
USBR 6310-09
5
USBR 6258-09
6
USBR 6310-09
7
USBR 6258-09
8
USBR 6310-09
(5 lbf/in2) are to be used for the test, hole in the center of the triaxial test
then completely submerge the end chamber bottom plate.
plate, connecting lines, and porous
insert in de-aired water and boil them 12.2.3 To ensure that pressure
from 8 to 16 hours to ensure that all does not build up in the system when
entrapped air has been removed. the end plate connecting lines are
attached to the fittings in the triaxial
12.1.2 If new porous inserts test chamber bottom plate, open
with a bubbling pressure of 35 kPa (5 valves to allow a small amount of
lbf/in2) or less are to be used for the water to flow from the appropriate
test, then completely submerge the drain tube and flushing tank through
end plates, connecting lines, and the fittings in the triaxial test chamber
porous inserts in de-aired water and bottom plate while the appropriate
boil them for a minimum of 2 hours to end plate connecting lines are being
ensure that all entrapped air has been attached.
removed.
12.2.4 Attach the bottom end
12.1.3 If end plates with plate connecting lines to the
previously saturated porous inserts appropriate fittings in the triaxial test
with a bubbling pressure greater than chamber bottom plate. Close valves.
35 kPa (5 lbf/in2) are to be reused for
the test, then remove end plates, 12.2.5 Attach the top end plate
connecting lines, and porous inserts connecting lines to the appropriate
from the triaxial assembly, fittings in the triaxial test chamber
completely submerge them in bottom plate. Close valves.
de-aired water, and boil them for a
minimum of 4 hours to remove any 12.2.6 Place the triaxial test
entrapped air. chamber into the groove in the triaxial
test chamber bottom plate.
12.1.4 If end plates with
previously saturated porous inserts 12.2.7 Check that all system
with a bubbling pressure of 35 kPa (5 valves are closed.
lbf/in2) or less are to be reused for the
test, then these end plates, connecting 12.2.8 Fill the triaxial test
lines, and porous inserts can be left in chamber about one-half full with
the triaxial assembly for subsequent de-aired water.
tests and de-aired using vacuum.
12.2.9 Open triaxial test
12.2 De-airing the Triaxial Test chamber cap relief valve.
Chamber:
12.2.1 Attach the saturated end 12.2.10 Set the triaxial test
plates with porous inserts to the chamber cap in place and force the
triaxial test chamber bottom plate. cap down until it is properly seated on
the triaxial test chamber.
12.2.2 Place the bottom end
plate with centering dowel into the
9
USBR 6258-09
10
USBR 6310-09
11
USBR 6258-09
12
USBR 6310-09
13
USBR 6258-09
increase over a period of two hours check and the compliance test until no
would clearly indicate a leakage in air is entrapped in the system.
the system. If the pressure rate
decreases with time as registered by 13.20 Select Syringe Speed:
the back pressure transducer, the Open all valves on the triaxial shear
specimen is still consolidating and the triaxial test chamber except the two
permeability measurement should not outlet valves on the injection and
be attempted before a steady-state back pressure lines. Set the pump
condition is reached. driving mechanism speed at 9 (0.13
mm/min [0.005 in/min]), and switch
Note 4.-A pressure increase inside on the flow pump. Wait at least 10
the specimen does not necessarily seconds and then close the ball valve.
correspond to a voltage increase. A Simultaneously monitor the voltage,
pressure decrease within the specimen being careful not to exceed a 50-mV
suggests that saturation is still taking voltage differential. If a satisfactory
place. steady voltage change occurs (not
exceeding 50 mV), then speed
13.19 Perform Compliance Test number 9 is acceptable for the
This step checks for remaining experiment. However, if the voltage
undissolved air in the system. Close change is too small (lower than 10
the valve connecting the flow pump mV) or too large (greater than 50
and triaxial shear triaxial test mv), then adjust the speed of the
chamber. Set the syringe speed to 9 pump driving mechanism until a
(0.13 mm/min [0.005 in/min]) and satisfactory steady voltage occurs
start the pump. Wait at least 10 (between 15 and 50 mV).
seconds, then close the ball valve
while simultaneously monitoring the Note 5.-If unfamiliar with the
time (t = 0 is the instant the valve is material behavior, a general rule of
closed). Read and record the time it thumb is to always begin with a very
takes for the voltmeter reading to low speed setting and progress with
reach 50 mV to the nearest 1 minute. higher speeds until a measurable
Open the valve and repeat the pressure is obtained.
compliance test with a different back
pressure. If the system is entirely 13.21 Perform Permeability Test:
saturated, the time necessary to attain 13.21.1 After selecting an
a 50-mV reading for the two different appropriate speed setting, close the
back pressures should be identical. A ball valve and start the flow pump to
±2-second deviation is usually inject water into the specimen.
acceptable given the inaccuracy Determine the steady-state pressure
introduced by opening and closing the differential across the rock specimen.
ball valve in addition to using a stop Monitor the voltage readings.
watch. If the two compliance times Materials of very high permeability
do not fall within this range, the (i.e., 1 x 10-2 cm/s or higher) will
system still contains a substantial almost instantaneously jump to a
amount of air. Repeat the saturation "steady-state" condition upon closing
the ball valve. On the other hand,
14
USBR 6310-09
13.24 After completion of the test, Note 5.-To obtain k in ft/yr, divide
disassemble the test setup and clean the value of k in equation 1 by 9.67 x
the test area and the equipment. 10-7
15
USBR 6258-09
16
USBR 6310-09
15.1.3 Moisture.-General
indication of the moisture condition
of the test specimen at the time of
testing, such as saturated, as received,
laboratory air dry, or oven dry. Some
cases may require reporting the actual
moisture content as determined by
procedure USBR 5300.
15.1.7Picture.-Color photograph of
the specimen after the test.
17
USBR 6310
INDEX CONT.
65W
SAMPLE # CONT SPECIMEN# CONT
42 4
HEIGHT cm DIA cm MASS gm LAT PRESS BACK FLOW
6.80 5.1 253.7 140 PRESS RATE #
125 12
SUB TITLE CONT CONT CONT CONT
TUCSON TER MINAT STOR AGE CAP
PRESS 1i PRESS 2i PRESS 1f PRESS 2f B LVDT i LVDT f
122.833 130.870 126.278 134.196 0.988 .753 .632
19
USBR 6310
0.0 -32.599
.5 -1.941
1.0 0.000
1.5 0.477
2.0 0.623
2.5 0.731
3.0 0.522
3.5 0.052
4.0 0.265
4.5 -0.094
5.0 -0.351
5.5 -0.288
6.0 -0.114
6.5 -0.049
7.0 -0.264
7.5 -0.172
8.0 -0.487
8.5 -0.564
9.0 0.528
10.0 -0.133
21