Error Messages
Error Messages
XS Series S/M i
August 5, 2008
6.3.7.7 RBC BUBBLE ERROR .............................................................. 36 6.3.12.1 Off Line (DSR Off)!....................................................................47
6.3.7.8 Low Count Error......................................................................... 36 6.3.12.2 Response Time Error!................................................................47
6.3.7.9 HGB ERROR .............................................................................. 37 6.3.12.3 Can't Send TEXT .......................................................................47
6.3.7.10 WBC-CH ERROR....................................................................... 37 6.3.12.4 Response Time Error!................................................................47
6.3.7.11 DIFF-CH ERROR ....................................................................... 37 6.3.12.5 Response Code Error! ...............................................................47
6.3.7.12 RBC-CH ERROR........................................................................ 38 6.3.12.6 NOT STX ....................................................................................47
6.3.7.13 PLT-CH ERROR......................................................................... 38 6.3.12.7 NOT ETX ....................................................................................47
6.3.7.14 Data Error ................................................................................... 38 6.3.12.8 Host Connection Error!! .............................................................47
6.3.7.15 Analysis Error ............................................................................. 38 6.3.12.9 Receive Error..............................................................................47
6.3.12.10 Send Error ................................................................................48
6.3.8 Laser Relating Errors ........................................................ 38 6.3.12.11 Sample ID# does not match Host Computer........................48
6.3.8.1 Laser Tube Aged ....................................................................... 39 6.3.12.12 Rack tube pos does not match Host Computer...................48
6.3.8.2 Close FCM Detect Cover .......................................................... 39
6.3.8.3 Laser Output Error ..................................................................... 39 6.3.13. Host Computer (ASTM Format) Message...................... 49
6.3.13.1 There isn't astm script file.........................................................49
6.3.9 System Errors..................................................................... 40 6.3.13.2 Response Time Error!................................................................49
6.3.9.1 Barcode Reader Com. Error (XS-1000i with Sampler)........... 40 6.3.13.3 Response Record Error! ............................................................49
6.3.9.2 ID Read Error (XS-1000i with Sampler) .................................. 40 6.3.13.4 Impossibile Inviare TEXT...........................................................49
6.3.9.3 System Configuration Error ....................................................... 40 6.3.13.5 Sample ID# does not match Host Computer..........................49
6.3.10 QC Errors .......................................................................... 41 6.3.13.6 Rack tube pos does not match Host Computer.....................49
6.3.10.1 X-barM Limit Error ..................................................................... 41 6.3.13.7 Referent rack tube pos of inquiry has returned from host co
6.3.10.2 L-J LIMIT ERROR...................................................................... 41 mputer. .......................................................................................................49
6.3.10.3 X-bar Limit Error ........................................................................ 41
6.4 Error Codes ...................................................................... 50
6.3.10.4 Control Expired........................................................................... 41
6.3.10.5 Control Entry ERR ................................................................... 42
6.3.11 Operator Maintenance Warnings...................................... 43
6.3.11.1 Replace Piercer .......................................................................... 43
6.3.11.2 Execute Weekly Rinse............................................................... 43
6.3.11.3 Execute Weekly Rinse(Warning)............................................... 43
6.3.11.4 Exchange the Air pump ............................................................ 44
6.3.11.5 Expired Reagent [CELLPACK(EPK)] ........................................ 44
6.3.11.6 Expired Reagent [SULFOLYSER(SLS)].................................... 44
6.3.11.7 Expired Reagent [STROMATOLYSER-4DL(FFD)] ................... 44
6.3.11.8 Expired Reagent [[STROMATOLYSER-4DS(FFS)] .................. 44
6.3.11.9 Expired Reagent [CELLSHEATH(C)(CSC)] (XS-1000iC only) 44
6.3.11.9 Right side cover is open........................................................... 45
6.3.11.10 Right side cover has opened. ................................................ 45
6.3.11.11 Tube Holder has opened. (XS-1000i) .................................... 46
6.3.12 Host Computer (Standard and DPS Format) Massage 47
XS Series S/M ii
SECTION 6 ERROR MESSAGES (6) “Alarm” sounds to notify the error status based on the rule of each
error. To stop the alarm sound, click “Alarm Stop” button on “Help
In this section, error monitoring method and error recovery method on Dialog Box” window.
XS series are discribed.
NOTE:
1. Descripition of analysis mode in this section
In XS series, there are 2 analysis modes “Manual Mode” and
“Sampler Mode” (XS-1000i). “Capillary Mode” is based on
“Manual Mode” and “Closed Mode” is based on “Sampler Mode”,
except where specifically noted.
2. When error occurs in XS series, each categorized error name is
shown on IPU status display area to prompt users to operate
settings or perform maintenance. XS series error is classified
under following groups (these errors are not described in this
section in particular).
-0.03Mpa +0 06Mpa
M57
PCB NO.3061
M51
Rear
Action Message: “Press [Execute] button. Air Pump test will start.”
Cause: 1) Pneumatic system leakage, disconnection, etc.
2) Relief Valve for -0.03MPa vacuum defective
Action to be taken:
1) Check 0.03MPa Relief Valve and Pneumatic System
for leakage, disconnection, etc.
2) After checking, select corresponding error from “Error
List” on “Help Dialog Box” window and click
“Execute”.
Help [Execute]: Click “Execute” button in “Help Dialog Box” window to
run Air Pump test sequence. If the result is inside the
capable range, the status turns to “Ready”.
Environmental
Temperature
Sensor
6.3.2.15 FFD RH Therm Sens ERR (XS-1000iC only) defective. Turn OFF the power and contact your Sysmex
service representative.”
Description: Temperature control thermistor in FFD reagent heater is Action to be taken:
defective or broken wire. When it is not recovered by power ON/OFF, check the
Purpose: To ensure the temperature control in the FFD reagent reagent heater thermistor for breaking/ disconnection.
heater. Replace the thermistor if it is obviously defective.
Method: 1) In Ready: The signal from the driver board is
monitored every 1 second. System generates an
error when detect defective signal for more than 3.0
seconds continuously.
2) In Analysis Operation:
This is monitored immediately after the analysis starts
XS’s Action: Status: Not Ready Data Mask: Yes
Sampler Operation: Interrupted Alarm Sound: Yes
Action Message: “One of the thermal sensors (FFD Reagent Heater) is
defective. Turn OFF the power and contact your Sysmex
service representative.”
Action to be taken:
When it is not recovered by power ON/OFF, check the
reagent heater thermistor for breaking/ disconnection.
Replace the thermistor if it is obviously defective.
Description: EPK could not be replenished while the external reagent Description: The reagent container is empty and needs to be
container is filled with enough amount of EPK. replaced.
Purpose: To ensure the required quantity of EPK for instrument Purpose: To ensure the required quantity of reagent for analysis.
operation. Method: Air bubble sensor at SLS replenish line is monitored
Method: The float switch installed on the reagent chamber is when SLS reagent is replenished.
checked. XS’s Action: Status: Not Ready Data Mask: No
Only in Analysis sequence: Timing at immediately after Sampler Operation: Interrupted Alarm Sound: Yes
replenishing EPK chamber completed. Action Message:
XS’s Action: Status: Not Ready Data Mask: No "Press [Execute] button. Reagent replace dialog will
Sampler Operation: Interrupted Alarm Sound: Yes appear."
Action Message: Cause: 1) Reagent shortage
“Press [Execute] button. Reagent replace dialog will 2) Hydraulic trouble in SLS replenish path
appear.” Help [Execute]: Run Reagent Replacement Sequence to recover from
Cause: Hydraulic trouble in EPK chamber replenish path the error.
Help [Execute]: The error is recovered after executing Reagent Replace Action to be taken: If this error occurs again even though the reagent is
Sequence. connected properly, check SLS replenish line for
Corrective Action: disconnection and leakage, and also check the function
If this error occurs again even though the reagent is of SLS Diaphragm Pump, SV31-32, and SLS Bubble
replaced/ connected properly, check EPK replenish line Sensor.
for disconnection and leakage, SV15 function, and also
check the Float Switch in EPK Chamber.
Description: The reagent container is empty and needs to be Description: The reagent container is empty and needs to be
replaced. replaced.
Purpose: To ensure the required quantity of reagent for analysis. Purpose: To ensure the required quantity of reagent for analysis.
Method: Air bubble sensor at reagent replenish line is checked Method: Error occurs when cycle count of FFS diaphragm pump
during FFD is aspirated. reaches 1200 times.
XS’s Action: Status: Not Ready Data Mask: No XS’s Action: Status: Not Ready Data Mask: No
Sampler Operation: Interrupted Alarm Sound: Yes Sampler Operation: Interrupted Alarm Sound: Yes
Action Message: Action Message: “Press [Execute] button. Reagent replace dialog will
“Press [Execute] button. Reagent replace dialog will appear.”
appear.” Cause: 1) Reagent shortage
Cause: 1) Reagent shortage 2) Defect in reagent flow path
2) Defect in reagent flow path Help [Execute]: System resets error after performing Reagent Replace
Help [Execute]: Run Reagent Replacement Sequence to recover from Sequence.
the error. Corrective Action: Replace FFS reagent package with new one, then
Action to be taken: If this error occurs again even though the reagent is reset the counter of FFS diaphragm pump cycle counter.
connected properly, check reagent replenish line for
disconnection and leakage, and also check the function
of FFD Diaphragm Pump, SV27-28 and FFD Bubble
Sensor.
Description: The liquid inside the RBC/HGB chamber could not be Description: The liquid inside the WBC chamber could not be
drained. drained.
Purpose: To ensure the drain function of RBC/HGB Chamber. Purpose: To ensure the drain function of WBC Chamber.
Method: The vacuum value just before RBC/HGB chamber drain Method: In analysis operation, the value at the vacuum sensor
is taken as VA, and if the vacuum (-0.03MPa) doesn’t go just before RBC/HGB chamber drain is taken as VA,
down below 90% of VA, when the sample of WBC Chamber drains, if the vacuum
System defines the status as error. Refer to Timing (-0.03MPa) doesn’t go down below 90 % of VA, System
Chart for the details of monitor timing. define the status as error. Refer to Timing Chart for the
XS’s Action: Status: Not Ready Data Mask: Yes details of monitor timing.
Sampler Operation: Interrupted Alarm Sound: Yes XS’s Action: Status: Not Ready Data Mask: Yes
Action Message: “Press [Execute] button. Sampler Operation: Interrupted Alarm Sound: Yes
RBC/HGB Chamber Drain test will start.” Action Message: “Press [Execute] button.
Cause: Bending or clogging in tube of RBC/HGB Chamber Drain WBC Chamber Drain test will start.”
Line Cause: Bending or clogging in tube of WBC Chamber drain line
Action to be taken: Check RBC/HGB drain path for tube bending or Action to be taken:
clogging, and rinse or replace the tube accordingly. Check WBC Drain Line for tube bending or clogging, and
Help [Execute]: Run WBC Chamber Drain Sequence by clicking rinse or replace the tube accordingly.
“Execute" button, and clear the error. Help [Execute]: Run WBC Chamber Drain Sequence to recover from the
error.
RBC/HBG Chamber WBC Chamber
6.3.4.1 WB Asp Motor Error Description: The sheath syringe motor does not rotate smoothly.
Purpose: To ensure the quantity of the sample measurement.
Description: The WB aspiration motor does not rotate smoothly. Method: System monitors whether drive mechanism returned
Purpose: To ensure the quantity of WB sample aspiration and back to its lower limit and stopped after syringe
rinse reagent in WB line. ascending operation in Analysis sequence.
Method: System monitors whether the Upper Limit Switch is ON In test operation, system monitors the remaining pulse
immediately after syringe ascended in Analysis number additionally when the motor stops.
sequence. XS’s Action: Status: Not Ready Data Mask: Yes
XS’s Action: Status: Not Ready Data Mask: Yes Sampler Operation: Interrupted Alarm Sound: Yes
Sampler Operation: Interrupted Alarm Sound: Yes Action Message: Press [Execute] button.
Action Message: Sheath Motor test will start.
“Press [Execute] button. Cause: Defect in Stepping Motor, Operational trouble by foreign
WB Aspiration Motor test will start.” objects.
Cause: Defect in Stepping Motor, operational trouble by foreign Action to be taken:
objects Check if there is no operational trouble by foreign objects,
Action to be taken: and take an corrective action or replace Sheath Syringe
Check if there is no operational trouble by foreign objects, Unit accordingly.
and take an corrective action or replace WB Aspiration Help [Execute]: The error is recovered after executing the Test
Motor Unit accordingly. Sequence in Sheath Motor.
Help [Execute]: The error is recovered after executing the Test Sheath Syringe
Sequence in WB Aspiration Motor. Lower Sensor
WB Pump
Upper Sensor
Right
Figure 6-16 Sheath Syringe Lower Sensor
Right
Description: Due to heavy loading by up and down movement of Description: The Aspiration Unit Front Back motor does not rotate
Aspiration Unit, motor doesn’t move smoothly. smoothly.
Purpose: Ensure up and down movement of Aspiration Unit. Purpose: To ensure forward - backward operation of Aspiration
Method: System monitors whether Drive Mechanism returned Unit Front-Back Mechanism WB sample aspiration and
back to its upper limit and stopped during Aspiration Unit rinse reagent in WB line.
ascending operation in Analysis Sequence. In test Method: When Aspiration Unit returns to its limit position, the
operation, system monitors the remaining pulse number system monitors whether Aspiration Drive Mechanism
when the motor stops. shielded the photo interrupter and stopped. In test
XS’s Action: Status: Not Ready Data Mask: Yes operation, the system also monitors the remaining pulse
Sampler Operation: Interrupted Alarm Sound: Yes number when the motor stops.
Action Message: “Press [Execute] button. Aspiration Unit Up Down XS’s Action: Status: Not Ready Data Mask: Yes
Motor test will start.” Sampler Operation: Interrupted Alarm Sound: Yes
Cause: Defect in Stepping Motor, Operational trouble by foreign Action Message: "Press [Execute] button.
objects. Aspiration Unit Front/Back Motor test will start."
Action to be taken: Cause: Defect in Stepping Motor, Operational trouble by foreign
Check if there is no operational trouble by foreign objects, objects.
and take an corrective action or replace Aspiration Unit Action to be taken:
accordingly. Check if there is no operational trouble by foreign objects,
Help [Execute]: The error is recovered after executing the Test and take an corrective action or replace defective part
Sequence in Aspiration Unit. accordingly.
Help [Execute]: Run Test Sequence of front and back motors in
Home Position Sensor Aspiration Unit to recover from the error.
of Aspiration Unit Home Position Sensor
Up Down Motor of Aspiration Unit
Front
Right
Figure 6-##
Figure 6-17 Home position sensor of Aspiration Unit
Description: The operation of pinch valve of Rinse Cup (PV-3) is Description: Waste Chamber1 pinch valve for draining (PV-2) is
defective. defective.
Purpose: To ensure the rinse aspiration operation of rinse cup. Purpose: To ensure Waste Chamber1 draining operation.
Method: During start-up sequence, the pinch valve function is Method: During start-up sequence, the pinch valve function is
checked as below; checked as below;
The value of ” –0.03MPa” vacuum when the pinch value The value of ” +0.06MPa” pressure when the pinch value
closed is taken as VA. closed is taken as PA.
The value of ” –0.03MPa” vacuum when the pinch value The value of ” +0.06MPa” pressure when the pinch value
open is taken as VB. open is taken as PB.
When “VA – VB < 0.01MPa”, system define the status as When “PA – PB < 0.02MPa”, system define the status as
the error. the error.
XS’s Action: Status: Not Ready Data Mask: Yes XS’s Action: Status: Not Ready Data Mask: Yes
Sampler Operation: Interrupted Alarm Sound: Yes Sampler Operation: Interrupted Alarm Sound: Yes
Action Message: Action Message: "Press [Execute] button.
"Press [Execute] button. Pinch Valve test will start."
Pinch Valve test will start." Cause: Draining trouble due to defect in Drive Motor, clogging
Cause: Draining trouble due to defect in Drive Motor, clogging and stuck in Drain Line Tube.
and stuck in Drain Line Tube. Action to be taken:
Action to be taken: Check if there is any obstruction to the motor operation.
Check if there is any obstruction to the motor operation. Also check clogging and stuck in Drain Tube, and rinse
Also check clogging and stuck in Drain Tube, and rinse or replace the parts accordingly.
or replace the parts accordingly. Help [Execute]: Run Test Sequence of pinch valve to recover from the
Help [Execute]: Run Test Sequence of pinch valve to recover from the error.
error.
Pinch Valve for
Pinch Valve for
Waste Chamber 1
Rinse Cup
(PV-2)
Drain Line (PV 3)
Waste Chamber1
Description: Waste Chamber 2 pinch valve for draining (PV-1) is Description: The Tube Holder motor does not rotate smoothly.
defective. Purpose: To ensure Forward – Backward operation of Aspiration
Purpose: To ensure Waste Chamber 2 draining operation. Unit Front-Back Mechanism WB sample aspiration and
Method: During start-up sequence, the pinch valve function is rinse reagent in WB line.
checked as below; Method: After motor operation completed in Test Sequence, the
The value of ” –0.03MPa” vacuum when the pinch value system monitors whether both the photo interrupter is
closed is taken as VA. shielded and stopped.
The value of ” –0.03MPa” vacuum when the pinch value Timing:
open is taken as VB. 1) On start-up of the analyzer
When “VA – VB < 0.01MPa”, system define the status as 2) When Service Sequence executed
the error. 3) When Analysis Operation executed
XS’s Action: Status: Not Ready Data Mask: Yes XS’s Action: Status: Not Ready Data Mask: No
Sampler Operation: Interrupted Alarm Sound: Yes Sampler Operation: Interrupted Alarm Sound: Yes
Action Message: "Press [Execute] button. Action Message: Remove the tube and Press [Execute] button.
Pinch valve test will start." Tube Holder Motor test will start.
Cause: Draining trouble due to defect in Drive Motor, clogging Cause: Draining trouble due to obstruction in drive mechanism
and stuck in Drain Line Tube. or defective drive motor.
Action to be taken: Action to be taken:
Check if there is any obstruction to the motor operation. Check if there is any obstruction to the motor operation.
Also check clogging and stuck in Drain Tube, and rinse Also check clogging and stuck in Drain Tube, and rinse
or replace the parts accordingly. or replace the parts accordingly.
Help [Execute]: Run Test Sequence of pinch valve to recover from the Help [Execute]: Run Test Sequence for Sample Set Unit to recover from
error. the error.
Sample Set Unit
Home Position Sensor
(Consisted of two
Rear
photo interrupters)
Waste Chamber 2
(WC2)
Waste Chamber2
pinch valve (PV-1)
Front
Figure 6-21: Home Position Sensor of Sample Set Unit
Figure 6-20 Waste Chamber 2 pinch valve
Description: Hand (Left/Right) Drive Mechanism does not operate Description: Hand (Up / Down) Drive Mechanism does not operate
smoothly. smoothly.
Purpose: To ensure the operation of Hand (Left/Right) drive Purpose: To ensure the operation of Hand (Up / Down) drive
mechanism. mechanism.
Monitoring condition: Method: 1) When the sensor is shielded ** seconds after Hand
1) Corresponding position sensor (photo interrupter) was started ascending
not light-shaded when Hand mech. should position at 2) When the sensor is not shielded ** seconds after
left initial position. Hands started descending.
2) Hand (Left/Right) Drive did not returned to Home XS’s Action: Status: Not Ready Data Mask: No
position in specified Remained Pulse Numbers. Sampler Operation: Interrupted Alarm Sound: Yes
Action Message: “Execute sample test menu.
XS’s Action: Status: Not Ready Data Mask: No Open the sampler cover and remove the tube. Then,
Sampler Operation: Interrupted Alarm Sound: Yes close the sampler cover and press [execute] button.”
Action Message:“Execute sample test menu. Cause: 1) Operational trouble by foreign objects (Blood
Open the sampler cover and remove the tube. Then, collection tube etc.).
close the sampler cover and press [execute] button.” 2) Defective motor, slack drive belt etc.
Parameter 1 Sensor Position: Action to be taken:
(0:Home Position, 1: Right, 2: Left) 1) Remove foreign objects (Blood collection tube etc.)
Parameter2 Remained Pulse Numbers 2) Find out the areas which cause malfunction, and
Cause: 1) Operational trouble by foreign objects (blood adjust or replace them.
collection tube etc.). Help [Execute]: Sampler initialization.
2) Defective motor, slack drive belt, and etc
Action to be taken:
1) Remove foreign objects (Blood collection tube etc.)
2) Specify the defective part as the cause, and take
corrective action adjustment or replacement.
Help [Execute]: Sampler initialization.
6.3.7.1 Background Error Description: During counting, the variation of sampling data exceeds
the specified range.
Description: Because the background counts are high, the Purpose: To monitor the uniformity of the sample concentration, or
measurement error might become wide. the sudden clogs at the aperture.
Purpose: To ensure that the background counts within the range. Method: During counting, get the count values every 0.5 seconds.
Method: Verify that the value for each parameter in the last And then, error is checked when maximum, minimum,
background counts is less than following value. and total values of sampling value satisfy following
WBC-C: 0.1 [X103/µL] formula.
WBC-D: 1.0 [X10302/µL] Maximum Value - Minimum Value - 150
X100 > 2.0
RBC: 2.0 [X104/µL] Total of the Sampling Value
HGB: 0.2 [g/dL]
(Exclude the first and the last sampling values.)
PLT: 1.0 [X104/µL]
XS’s Action: Status: Ready Data Mask: Yes
XS’s Action: Status: Ready Data Mask: No
Sampler Operation: Interrupted
Sampler Operation: Continued Alarm Sound: Yes
Alarm Sound: Sampler Mode only
Action Message:
Cause: 1) Dirt and/or clogging at RBC/PLT transducer
“Execute Auto Rinse menu. Press the [Execute]
2) Incomplete mixing operation in RBC/Hgb chamber
button.”
3) Malfunction of Sheath Syringe Unit
Help [Execute]: Performs Auto Rinse Sequence.
(Syringe leakage, Defective Drive Motor.
4) Incomplete charging operation of RBC/PLT sample
NOTE::
into sample measurement line.
Error message will be displayed until background check in Auto
5) External inductive noise
Rinse passes.
Action to be taken:
1) Perform Clog Removal or Brush cleaning of RBC
transducer.
2) Confirm that mixing operation at RBC/Hgb Chamber is
performed correctly.
3) Check if there is no leakage of Sheath Syringe and
the surrounding hydraulic system.
4) Check if charging operation of RBC sample into
sample measurement line works properly.
5) Control the noise.
Description: During counting, the difference of sampling data exceeds Description: During counting, the difference of sampling data exceeds
the specified range. the specified range.
Purpose: To monitor the uniformity of the sample concentration Purpose: To monitor the uniformity of the sample concentration
and the sudden clogs at the aperture. and the sudden clogs at the jet nozzle and flow cell.
Method: While counting, get the count values every 0.5 seconds. Method: While counting, get the count values every 1 second.
And then, error is checked when maximum, minimum And then, error is checked when maximum, minimum
and total values of sampling value satisfy following and total values of sampling data satisfy following
formula. formula.
Maximum Value - Minimum Value - 100
Maximum Value - Minimum Value - 100 X100 > 3.0
X100 > 2.0 Total of the Sampling Value
Total of the Sampling Value (Exclude the first and the last sampling values.)
(Exclude the first and the last sampling values.) XS’s Action: Status: Ready Data Mask: Yes
XS’s Action: Status: Ready Data Mask: Yes Sampler Operation: Interrupted
Sampler Operation: Interrupted Alarm Sound: Sampler Mode only
Alarm Sound: Only in Sampler analysis XS’s Action: Status: Not Ready Data Mask: Yes
Cause: 1) Dirt or Clog at RBC/PLT Transducer Sampler Operation: Interrupted Alarm Sound: Yes
2) Incomplete mixing in RBC/Hgb Chamber Cause: 1) Dirt at liquid path in FCM Flow Cell
3) Leakage of Sheath syringe or surrounding hydraulic 2) Incomplete mixing operation in WBC/DIFF Chamber
line. 3) Leakage of Sheath syringe or surrounding hydraulic
4) Defective Sheath Motor. line.
5) Incomplete charge operation of RBC/PLT sample. 4) Defective Sheath Motor.
5) External inductive noise 5) Incomplete charge operation of WBC sample.
Corrective Action: 6) External inductive noise
1) Perform Clog Removal or RBC transducer cleaning Action to be taken:
with brush. 1) Perform Flow Cell cleaning.
2) Check if mixing in RBC/Hgb Chamber is performed 2) Confirm if mixing operation at WBC Chamber is
properly. performed correctly.
3) Leakage of Sheath syringe or surrounding hydraulic 3) Check if there is no leakage of Sheath Syringe and
line. the surrounding hydraulic system.
4) Defective Sheath Motor. 4) Check if charging operation of WBC/DIFF diluted
5) Check if charging operation of RBC sample into sample into sample measurement line works properly.
sample measurement line works properly. 5) Control the noise.
6) Control the noise.
Description: During counting, the differences of sampling data exceed Description: Indicates that the RBC transducer aperture has clogging.
the specified range. Purpose: To ensure sheath flow and RBC/PLT analysis.
Purpose: To monitor the uniformity of the sample concentration Method: System monitors the A/D converted value of the clogging
and the sudden clogs at the jet nozzle and flow cell. signal from the both electrodes at the transducer, and
Method: While counting, get the count values every 1 second. checks the value is within the following range.
And then, error is checked when maximum, minimum Clogging Rate: Error occurs at C>105.
and total values of sampling data satisfy following C = 3.832 X 10 X D X 3.9/256-2.110 X TTD2 X 10-2
formula.
+1.992XTTD –3.660X10
Maximum Value - Minimum Value - 100 C: Clogging Rate (integral value, round to decimal point)
X100 > 3.0
Total of the Sampling Value D: A/D converted value of the clogging voltage System
(Exclude the first and the last sampling values.) monitors at the timing described in the timing chart.
XS’s Action: Status: Ready Data Mask: Yes XS’s Action: Status: Ready Data Mask: Yes
Sampler Operation: Interrupted Sampler Operation: Interrupted
Alarm Sound: Sampler Mode only Alarm Sound: Sampler Mode only
Cause: 1) Dirt at liquid path in the Flow Cell Action Message:
2) Incomplete mixing operation in WBC/DIFF Chamber “Execute Clog Removal menu.
3) Leakage of Sheath syringe or surrounding hydraulic Press the [Execute] button.”
line. Cause: Dir or clog at RBC/PLT transducer
4) Defective Sheath Motor. Corrective Action:
5) Incomplete charge operation of WBC sample. Perform Clog Removal or brush cleaning of RBC/PLT
6) External inductive noise Transducer
Action to be taken: Help [Execute]: Performs Clog Removal sequence and then resets
1) Perform Flow Cell cleaning. Error state to Ready.
2) Check if mixing in WBC Chamber is performed
properly.
3) Check if there is no leakage of Sheath Syringe and
the surrounding hydraulic system.
4) Check if the charging operating into WBC/DIFF
measurement line is performed correctly.
5) Take corrective measures to reduce inductive noise.
Description: Indicates that the air bubbles are on the RBC aperture. Description: Although the blood volume sensor detects blood volume
Purpose: To ensure RBC/PLT analysis. enough and aspiration error is not detected with the
Method System counts the output pulse of bubble monitoring blood aspiration sensor, the counting value is too low.
circuit during counting. Verify the total output Purpose: To detect the abnormalities in the dilution system.
pulse of bubble monitoring circuit should be less than Method: Error is checked if every data below is less than the
128. specified value even though the blood volume sensor
XS’s Action: Status: Ready Data Mask: Yes monitors enough sample volume and blood aspiration
Sampler Operation: Interrupted sensor does not detect the aspiration error during
Alarm Sound: Sampler Mode only sampler analysis.
Action Message: “Execute Air Bubble removal menu. WBC 1.0 [X103/µL]
Press the [Execute] button.” RBC: 0.3 [X106/µL]
Cause: 1) Air bubbles stuck due to the contamination in HGB: 1.0 [g/dL]
RBC/PLT Transducer and the surrounding hydraulic PLT: 20 [X103/µL]
system path. XS’s Action: Status: Ready Data Mask: No
2) Air bubbles due to looseness or deterioration of tubes Sampler Operation:
of hydraulic line around RBC Transducer. depends on “Sampler stop condition setting”
Action to be taken: Alarm Sound:
1) Clear the clog or rinse RBC Transducer Pellet with depends on “Sampler stop condition setting”
brush. Action Message:
2) Find out the place where air bubbles got in, and “Blood sample is not aspirated correctly.”
replace the tube accordingly. Cause: 1) RBC or sample diluted in WBC Reaction Chamber
Help [Execute]: System performs Clog Removal sequence and then was not charged properly to Main Unit due to the
resets Error state to Ready. trouble of hydraulic system such as Diaphragm Pump
or Solenoid Valve.
2) Leakage of Sheath syringe or surrounding hydraulic
line.
Action to be taken:
1) Find out the defect areas, and rinse or replace the
parts accordingly.
2) Find out the cause of looseness of Sheath Syringe
and the surrounding tubes, and take the corrective
measure.
Description: A/D converted HGB blank value or sample value Description: WBC channel in FCM detects abnormalities. It means
exceeds the specified range. bad hemolysis, etc.
Purpose: To ensure that the HGB value is analyzed without any Purpose: To ensure the WBC data.
problem. Method: If the WBC value is in the following range.
Method: If the A/D converted HGB blank value or HGB sample WBC count > 999.99 x 103µL
value satisfies either of the following conditions. XS’s Action: Status: Ready Data Mask: Yes
Blank converted value<50 Sampler Operation: Interrupted Alarm Sound: Yes
Sample converted value>12000 Action Message: “Check the sample.”
Sample converted value - Blank converted value< -50 Cause: Lyse process in WBC Chamber could not be performed
Sample converted value - Blank converted value> 3600 correctly due to Incomplete operation of relating
XS’s Action: Status: Ready Data Mask: Yes Diaphragm Pumps or Solenoid Valves etc.
Sampler Operation: Interrupted Alarm Sound: Yes Inductive noise or A/D conversion trouble due to board
Action Message: defect.
“If the same error message is displayed again, contact
your Sysmex service representative.
Cause: Air bubbles got into HGB Flow Cell during Blank or 6.3.7.11 DIFF-CH ERROR
Sample Analysis. (caused by flow path clogging or
malfunction of related Diaphragm Pump or Solenoid Description: WBC channel in FCM detects abnormalities.
Valve) Purpose: To ensure the WBC data.
Action to be taken: Method: If the WBC value is in the following range.
Check Diluent flow to RBC Chamber or pull-in operation WBC-D > 999.99×103μL
to HGB Flow Cell, and then find out the defective area to
XS’s Action: Status: Ready Data Mask: Yes
take the corrective action.
Sampler Operation: Interrupted Alarm Sound: Yes
Action Message:
“Check the volume of STROMATOLYSER-4DS(FFS)”
Cause: Lyse process in WBC reaction chamber could not be
Hgb Flow Cell performed correctly.
(incomplete operation of related diaphragm pump,
solenoid valves, etc)
Inductive noise or A/D conversion trouble due to board
defect.
Front
Description: RBC channel detects abnormalities. Description: Data that Sampler Analysis should be interrupted occurs.
Purpose: To ensure the RBC data. Purpose: To interrupt Sampler Analysis when data that Sampler
Method: If the RBC value is in the following range. Analysis should be interrupted occurs.
RBC count > 999.9×103μL Method: System compares analysis data with Sampler Stop Limit.
If the analysis data exceeds Sampler Stop Limit, the
XS’s Action: Status: Ready Data Mask: Yes
error is generated.
Sampler Operation: Interrupted Alarm Sound: Yes
XS’s Action: Status: Ready Data Mask: No
Action Message: “Check the sample.”
Sampler Operation: Interrupted Alarm Sound: Yes
Cause: Inductive noise or A/D conversion trouble due to board
Action Message:
defect.
Corrective Action
1) Check Stored Data.
2) Correct Error.
6.3.7.13 PLT-CH ERROR
3) Re-analyze the sample.
Description: PLT channel detects abnormalities.
Purpose: To ensure the PLT data.
6.3.7.15 Analysis Error
Method: In the following case, the error is detected.
Number of PLT > 999.9×103μL Description: Error that Sampler Analysis could be interrupted occurs.
XS’s Action: Status: Ready Data Mask: Yes Purpose: To interrupt Sampler Analysis when error that might
Sampler Operation: Interrupted Alarm Sound: Yes affect analysis data while in Sampler Analysis an to
Action Message: “Check the sample.” prompt the situation to the operator.
Cause: Inductive noise or A/D conversion trouble due to board Method: During Sampler Analysis, if any error related to
defect. temperature, pressure, motors and analysis which cause
sampler stop or cancellation affects the data, this error is
detected.
XS’s Action: Status: Ready Data Mask: No
Sampler Operation: Continued Alarm Sound: No
Action Message: Corrective Action
1) Check Stored Data.
2) Correct Error.
3) Re-analyze the sample.
Description: Electric current to drive laser tube is increasing, and the Description: FCM detector cover is open.
malfunction of laser tube is expected. Purpose: To ensure the data by cutting off the light from outside or
Purpose: To predict the laser tube life time and let the operator to to prevent from leakage of the laser light.
contact Sysmex Service Representative to replace the Method: Monitor if the position sensor (photo interrupter) for FCM
laser. detector cover. (ON: when cover is closed.) System
Method: Monitor in accordance with the timing described in monitors cover sensor every 1 second constantly.
Timing Chart. System generates an error when detect error information
Laser Drive Current Value> Y1= LLA×(5/Eta+Ith+LLB) for more than 3.0 seconds continuously.
XS’s Action: Status: Not Ready Data Mask: Yes
LLA=1.28,LLB=18 Sampler Operation: Interrupted Alarm Sound: Yes
(Eta and Ith are values input by the maker when lasers
Action Message: Close FCM Detector Cover.
were replaced as manufactured.) If this error persists after cover is closed, there is
Monitor in accordance with the timing described in
probably something defective in the system.
Timing Chart. Contact your Sysmex Service Representative.
XS’s Action: Status: Ready Data Mask: No
Sampler Operation: Continued Alarm Sound: No
6.3.8.3 Laser Output Error
Action Message:
Laser requires investigation.
Contact your Sysmex Service Representative. Description: The laser output exceeds the specified range.
(Analysis is still possible.) Purpose: To ensure the laser output is stable.
Method: System generates an error when the current value is as
follows.
The laser drive current value > 127 mA
The laser drive current value < 10 mA
XS's Action: Status: Not Ready Data Mask: Yes
Sampler Operation: Interrupted Alarm Sound: Yes
Action Message: Contact your Sysmex Service Representative.
6.3.9.1 Barcode Reader Com. Error (XS-1000i with Sampler) Description: Actual system configuration does not match with the file
as stored in IPU.
Description: Communication between main unit and Barcode reader Purpose: To prevent the possible malfunction with incorrect
on sampler failed. system configuration
Purpose: To ensure that the communication between main unit Method: System check the status of sample set unit immediately
and Barcode reader on sampler performs correctly. after startup.
Method: If the communication between main unit and Barcode Operate Sample Set Unit to check the sensor for
reader on sampler is not properly performed, the error is ON/OFF status. When Sample Set Unit operates
generated. properly, it is confirmed as XS-1000i and check the
XS’s Action: Status: Not Ready Data Mask: No status and configuration.
Sampler Operation: Interrupted Alarm Sound: Yes XS’s Action: Status: Not Ready Data Mask: No
Action Message: Sampler Operation: Interrupted Alarm Sound: Yes
"Restart the system. Action Message: "Restart the system.
Contact your Sysmex service representative, if the error Contact your Sysmex service representative, if the error
occurs again." occurs again."
6.3.11.1 Replace Piercer Description: Warns the user that the time for Weekly Rinse has
already passed.
Description: Indicates that a piercer needle should be replaced. Purpose: To warn the user that the time for Weekly Rinse has
Purpose: To prompt the operator for replacing the piercer needle. already passed and restrict analysis operations until
Method: The warning is displayed if the piercer operating cycles Weekly Rinse is completed.
exceed specified operating cycles, when the power of Method(Cause): If the number of operations is beyond 1400 after last
the main unit is turned ON or during the measurement Weekly Rinse sequence performed, it is informed to the
on the auto mode analysis. user.
Piercer Operation Times > 30000 XS’s Action: Status: Not Ready Data Mask: No
XS’s Action: Status: Ready Data Mask: No Sampler Operation: Interrupted Alarm Sound: Yes
Sampler Operation: Continued Alarm Sound: No Action Message:
Action Message: “Turn the main unit power switch off. Replace the Execute the Weekly Rinse.
piercer. Then, reset the piercer counter in the counter Press [Execute] button.
screen of the maintenance menu.” Weekly Rinse dialog will appear.
Help [Execute]: Perform Weekly Rinse sequence and reset the counter.
NOTE::
When Weekly Rinse is performed, cycle conter for this warning
message is reset automatically.
Description: Right side cover of analyzer is open. Description: Right side cover of analyzer has opened during analysis
Purpose: Check if the right side cover is open/ close in Ready operation.
status. If it is open, restrict analysis operations to ensure Purpose: Check if the right side cover is open/ close during
the safety of users. analysis operation. If it opens, analysis operations are
Method(Cause):Every 100ms in Ready Status stopped urgently to ensure the safety of users.
if it is open for 300 ms consecutively, analysis operations Method(Cause): Monitor the open/ close sensor of right side cover
become unavailable. (photo interrupter) every 100 ms during analysis
XS’s Action: Status: Not Ready Data Mask: No operation.
Sampler Operation: Continued Alarm Sound: No If it opens for 300 ms consecutively, the error is
Action Message: “Right side cover is open. Please close it.” detected and analysis sequences are stopped urgently.
Cause: 1) Right side cover is open. XS’s Action: Analysis Operation: Emergency Stop, Next analysis is
2) Defect in the open/ close sensor of right side cover not available
(photo interrupter) Data Mask: No Sampler Operation: Interrupted
Action to be taken: 1) Close right side cover. Alarm Sound: Yes
2) Check the status of open/ close sensor (photo Action Message: Right side cover has opened.
interrupter). Please close it. Then, turn the main unit power switch off,
then turn on.
NOTE: Cause: 1) Right side cover has opened.
When “Right side cover has opened.” error (opened during 2) Defect in the open/ close sensor of right side cover
operation) is active, “Right side cover is open.” error (opened in (photo interrupter)
Ready status) is not monitored. Action to be taken:
Check the status of open/ close sensor (photo
interrupter) and power OFF/ON the Main Unit.
6.3.12.1 Off Line (DSR Off)! Condition The message is displayed when receiving 4th NAK from the
host computer after sending text data on serial
Condition The message is displayed when control signals of Serial Port communication Class B.
(DSR) are not active on serial communication. Action The transmission will be aborted.
Action The transmission will be aborted.
Condition The message is displayed when a frame error, a parity error Condition The message is displayed when the end of text data received
or an over-run error is detected at sending text data to the from host computer is not “ETX(0x03)”.
host computer on serial communication. Action The transmission will be aborted.
Action The transmission will be aborted.
Message Sample ID# does not match Host Computer. This sample
will be measured using sample tube ID# with default
demographic information
Condition The message is displayed when receiving sample No. is
different from the inquired one at order inquiry. (At inquiring
on manual analysis or when the inquiry key is sample No.
on sample analysis.)
Action The transmission will be aborted.
6.3.13.1 There isn't astm script file Message Sample ID# does not match Host Computer. This sample will be
measured using sample tube ID# with default demographic
Condition The message is displayed when ASTM script file does not information.
exist in the install folder. Condition The message is displayed when receiving sample No. is
Action The transmission will be aborted. different from the inquired one at order inquiry. (At inquiring on
manual analysis or when the inquiry key is sample No. on
sample analysis.).
6.3.13.2 Response Time Error! Action The transmission will be aborted.
Message Referent rack tube pos of inquiry has returned from host
6.3.13.4 Impossibile Inviare TEXT computer. Rack tube pos at inquery is used.
Contdition The message is displayed when rack No./ tube position
Condition The message is displayed when the text data is not sent to the received from the host computer is different from inquired No.
host computer because of a communication error and etc. at order inquiry. (When the inquiry key is rack No. / tube
Action The transmission will be aborted. position on sampler analysis.)
Action The transmission will be aborted.
Fllowing table shows the Error codes for XS Series and SNCS parameters.
Category Category Parameter1 Parameter2
Error Description Error Code
-1 -2 XXXXXX ZZZZZ
Pneumatic Errors
Pressure (MPa)
113 Pressure
0.06MPa Error 113030.XXXXX.ZZZZZ Sequence #
Value
11
Vacuum (MPa)
114
-0.03MPa Error 114010.XXXXX.ZZZZZ Vacuum Value Sequence #
Others
119
Pressure Lower Error 119010
Temperature Error
Env Temp High 122510.XXXXX.0 Temperature 0
122
Env Temp Low 122515.XXXXX.0 Temperature 0
FCM RU TEMP HIGH 123220.XXXXX.0 Temperature 0
123
FCM RU TEMP LOW 123225.XXXXX.0 Temperature 0
12 RH Temp High 125020.XXXXX.0 Temperature 0
125
RH Temp Low 125025.XXXXX.0 Temperature 0
ENV THERM SENS ERR 127320.XXXXX.0 Temperature 0
RH Therm Sens ERR 127420.XXXXX.0 Temperature 0
127
FCM SHEATH SENS ERR 127600.XXXXX.0 Temperature 0
Env Therm Sens ERR 127900.XXXXX.0 Temperature 0