Uma Co., Ltd. Pepsinogen (PG) Calibrator: 2-19-6 Yokosuka Matsudo, Chiba, Japan

This document provides information about a pepsinogen calibrator kit produced by UMA CO., LTD. for quantitative determination of pepsinogen levels in blood. The kit contains 6 levels of standardized pepsinogen solutions for use as calibration controls. The calibrator is ready to use and stable for 18 months when stored between 2-8 degrees C and protected from light. Proper handling and safety precautions are described to avoid infection when working with samples containing human blood.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as DOCX, PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
49 views1 page

Uma Co., Ltd. Pepsinogen (PG) Calibrator: 2-19-6 Yokosuka Matsudo, Chiba, Japan

This document provides information about a pepsinogen calibrator kit produced by UMA CO., LTD. for quantitative determination of pepsinogen levels in blood. The kit contains 6 levels of standardized pepsinogen solutions for use as calibration controls. The calibrator is ready to use and stable for 18 months when stored between 2-8 degrees C and protected from light. Proper handling and safety precautions are described to avoid infection when working with samples containing human blood.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as DOCX, PDF, TXT or read online on Scribd
You are on page 1/ 1

UMA CO., LTD.

2-19-6 Yokosuka Pepsinogen (PG) Calibrator


Matsudo, Chiba, Japan

 2 ~ 8 °C Packages
 18 months/block from light 0.5ml x 6 Vial (Level 1 to Level 6)
 DO NOT freeze

1. PURPOSE OF USE solution (active chloride concentration of over 1000 ppm).


Calibration for quantitative determination of Measure - Absolutely do not freeze this product.
Pepsinogen (PG) in blood by Immunoassay. 5. ASSIGNED VALUE
This product can be used for both PG I/PG II. Please refer to the package insert for each lot number.
For in Vitro Diagnostics use only.
6. STORAGE AND SHELF LIFE
2. STRUCTURE AND COMPOSITION OF THE KIT
Storage: Store in cool and dark place (2 ~ 8 C).
Standard solution of PG and stabilizers to maintain PG for
Shelf Life: 18 months. Shelf life is printed on the label of kit
the accurate controls.
box.
Reconstituted Calibrator 0.5 mL × 6 levels set (Level 1,
7. CONTACT INFORMATION AND MANUFACTURER
Level 2, Level 3, Level 4, Level 5, Level 6).
UMA CO., LTD
3. DIRECTION FOR USE
2-19-6 Yokosuka, Matsudo city, Chiba Prefecture, 270-
Preparation method of PG Calibrator.
0031, Japan
PG Calibrator is ready for use.
Tel: 047-710-4871 Fax: 047-710-4872
Storage
Can be used for one week under storage in refrigerator (2
~ 8 C).
Bring back the refrigerated standard to room temperature
before use.

4. PRECAUTION FOR USE AND HANDLING


- This product is made of human blood as a base. Take
same precautions against infection as patient samples
such as to wear gloves.
Do not pipette by mouth.
- If this product is spilled from the container, disinfect with
80% of alcohol spray and wipe it out wearing gloves for
protection of infection.
- All samples and calibrators, as well as their containers
must be treated according to the handling manual for
discarding infectious waste materials   or treat them
according to the following methods.
 Sterilize with an autoclave, subjecting to high pressure
saturated steam at 121 C for 20 minutes, however do not
process wastes containing sodium hypochlorite with an
autoclave.
 Immerse at least one hour in sodium hypochlorite

1/1 Revision 12/2016