MA 53 Operating Instructions
MA 53 Operating Instructions
MAICO Diagnostic GmbH, Salzufer 13/14, D 10587 Berlin, Tel.: ++4930/70 71 46 50, Fax: ++4930/70 71 46 99
MAICO Diagnostics, 7625 Golden Triangle Drive, Eden Prairie, MN 55344, Toll Free ++1/888.941.4201, Fax ++1/952.903.4200
MAICO
Operating Instructions MA 53
Table of Contents Page
1. Introduction 2
2. Description 3
3. Getting started 5
3.1 Unpacking your instrument 5
3.2 Preparing the MA 53 for use 6
3.3 Getting familiar with your MA 53 7
3.3.1 Using the keyboard of the MA 53 8
3.3.2 The display of the MA 53 9
4. Pure tone audiometry 11
4.1 Air Conduction (AC) Testing 11
4.2 Uncomfortable Hearing Level (UCL) Testing 12
4.3 Bone Conduction (BC) Testing 13
4.4 Masking of the opposite ear 14
4.5 Diagnostic Tests 16
4.6 Dual Frequency Mode 21
4.7 High Frequency Audiometry (optional) 22
6. Speech Audiometry 23
6.1 Connecting the speech source (microphone, tape or CD-Player) 23
6.2 Speech Calibration 24
6.3 Connection and adjustment of the monitor phone 25
6.4 Speech Audiometry 26
6.5 Masking 27
7. Advanced features 28
8. Quick reference guide 31
9. Recommended literature 36
10. Computer interface 37
11. Care and Maintenance of the instrument 38
12. Trouble shooting 39
13. Technical data 40
14. Warranty, Maintenance and After-Sales Service 47
15. Safety Regulations 48
16. List for subjective instrument check 49
17. Certificate of conformity 50
18. Front panel layout and rear panel connections 52
Geba53_e.03.b.wpd 1 8 5 1 4 3 0 / 1 0 2 / 0 3
MAICO
Operating Instructions MA 53
1 Introduction
Thank you very much for purchasing a quality product from the MAICO family.
The MA 53 audiometer is manufactured to meet all quality and safety
requirements, and has been certified with the CE-symbol according to Medical
Directive 93/42/EEC.
This user manual should make it as easy as possible for you to become familiar
with the functions of the MA 53. Please open out the flap of illustrations on the
last page. The description of the position (e.g. ! ) of controls, displays and
connections, found again in the text, will make it easier for you to learn how to
operate the MA 53.
If you have problems or have ideas for further improvements, please get in touch
with us. Simply call.
Your MAICO-team
Geba53_e.03.b.wpd 2 8 5 1 4 3 0 / 1 0 2 / 0 3
MAICO
Operating Instructions MA 53
2 Description
The MA 53 is a real two-channel audiometer for advanced pure tone and
speech audiometric tests. It can be used for ENT diagnostics and hearing aid
fittings in the office, and for mobile audiometry in clinics, homes and aboard.
Tests can be performed using the TDH 39 headphones (AC), the optional high
frequency headphones HDA 200, B 71 bone conduction receiver (BC) or optional
insert phones and loudspeakers (FF). Built-in test signals are pure tone, pulse
tone, warble tone, narrow band and broadband noise. All signals and test
frequencies are individually adjustable on both channels. The MA 53 has
stereophonic microphone inputs for live speech audiometry, an input for an
optional tape/CD player with speech test material and an additional input for a
second tape/CD player (i.e. for external masking signals).
The MA 53 audiometer with the standard TDH 39 delivers 12 air conduction (AC)
test frequencies from 125 Hz to 12 kHz, with levels from -10 dBHL to 120 dBHL.
With the optional high frequency HDA 200 it delivers 17 air conduction (AC-HF)
test frequencies from 125 Hz to 16 kHz, with levels from -20 dBHL to 120 dBHL.
Bone conduction (BC) can be tested with 11 test frequencies from 125 Hz to
8 kHz with levels from -10 dBHL to 70 dB HL (with the standard bone conductor
B 71 the frequency range is limited from 250 Hz to 6 kHz).
The optional insert phones EAR 3A submit levels from -10 dBHL to 120 dBHL with
11 test frequencies from 125 Hz to 8 kHz.
Optional loudspeakers are available for free sound field measurements (FF). The
new, portable loudspeakers MAICO SBC have a dynamic range from -10 dBHL to
80 dBHL at 1m distance, for nine test frequencies from 500 Hz to 8 kHz. The
frequency range for speech is 500 Hz to 8 kHz with levels up to 80 dBSPL.
The loudspeakers Canton Plus XL (not available in the USA), deliver levels up to
90 dBHL at 1m distance over the entire frequency range from 125 Hz to 8 kHz.
The MA 53 has also separate line level outputs for an external amplifier.
The hearing level is controlled independently for each channel by two detented
dials on the left and right of the instrument which can be operated from the side
or the top. The level steps are 5 dB and can be changed to 2 dB or 1 dB. The
signal STIM bar and the frequency up/down keys are just beside the level control
knobs for easy one handed control of level, frequency and signal presentation.
The large backlighted LCD-display shows level, frequency, transducer, signal and
other information for each channel.
Geba53_e.03.b.wpd 3 8 5 1 4 3 0 / 1 0 2 / 0 3
MAICO
Operating Instructions MA 53
As a result of modern microprocessor technology, the MA 53 is easy to use and
is extremely reliable.
The audiometer is designed to be extremely service friendly. Automatic test
programs make trouble shooting and the yearly calibration as easy as possible.
PC-Interface:
A serial RS 232C interface for data transfer to a connected computer is built in.
The optional MAICO-audiometry module for NOAH and PAX enables the
automatic data transfer of all test results of the MA 53, like speech audiogram
and ABLB, DLI, SISI, tone decay etc.
For more information see chapter 10.
Geba53_e.03.b.wpd 4 8 5 1 4 3 0 / 1 0 2 / 0 3
MAICO
Operating Instructions MA 53
3 Getting started
3.1 Unpacking your instrument
Your MA 53 was carefully inspected and packed for shipping. However, it is
a good practice to thoroughly inspect the outside of the shipping box for
signs of damage. If any damage is noted, please notify the carrier
immediately.
Carefully remove the instrument from the shipping box. Remove the plastic
bag from the instrument and inspect the case for any damage.
Notify the carrier immediately if any mechanical damage is noted. This will
assure that a proper claim is made. Save all packing material so the claim
adjuster can inspect it as well. Notify your dealer or MAICO when the
adjuster has completed the inspection.
SAVE ALL THE ORIGINAL PACKING MATERIAL AND THE SHIPPING
CONTAINER SO THE INSTRUMENT CAN BE PROPERLY PACKED IF IT NEEDS
TO BE RETURNED FOR SERVICE OR CALIBRATION.
All accessories are already packaged in the compartment connected with the
MA 53. Please check that all accessories listed below are received in good
condition. If any accessories are missing or damaged, immediately notify your
dealer or MAICO.
Standard accessories:
- 1 Headphone TDH 39
- 1 Bone conduction receiver B 71 with headband
- 1 Patient resonse switch
- 1 power cable
- 1 microphone with stand (only for export outside USA)
- 1 audiogram form (50 sheets)
- 1 monitor headset with microphone (USA only)
- 5 sound room cords (USA only)
Calibration of the device:
The instrument, headphones, bone conduction receiver as well as the optional
insert phone and loudspeaker match one another and have the same serial
number (e.g. 6631520). As they have been calibrated with this particular
instrument, using of other transducers is not allowed. If one of the acoustic
transducers is replaced, the instrument must be recalibrated.
Geba53_e.03.b.wpd 5 8 5 1 4 3 0 / 1 0 2 / 0 3
MAICO
Operating Instructions MA 53
3.2 Preparing the MA 53 for use
The MA 53 with its integrated compartment for the
accessories is perfect for portable use. The rugged
housing, light weight and the comfortable handling
make it easy and convenient to transport the
instrument. Carry it, as shown, with the bottom away
from the leg. That helps protect the front cover from
damage, and due to the asymmetrical handle position
it provides more space for your leg.
To get started first move the handle under the
housing. Unlatch the cover by pressing in the two
black locks located on the left and right sides near the Picture 1 Transport of MA 53
front of the instrument. Open the front cover and rest
it on the back cover of the accessory compartment. To
open the compartment, press the two black locks in
the upper middle of the instrument. Open the
compartment cover as seen in picture 2. Take the
headphone, the bone conduction receiver, the patient
response switch and the microphone out of the
compartment and connect the power cable to power.
The MA 53 operates with voltages from 100 to
240 V~ AC, 50/60 Hz. To avoid pinching the cables
when closing the cover, lay the cables in the slots Picture 2 MA 53 with open cover
provided. Close the back cover and latch it with slight
pressure. You can place the microphone or a tape or
CD-player on top of the cover (see picture 3).
The MA 53 should be operated in a quiet room, so
that the audiometric examinations are not influenced
by outside noises. For use in noisier environments
headphones with optional sound insulation muffs are
available.
Electro-medical instruments, which emit strong
electromagnetic fields (e.g. microwaves - radiotherapy
devices), can influence the function of the audiometer.
Therefore the use of these instruments is not allowed Picture 3 MA 53 in operation
in close proximity to the audiometer.
The test room must be at normal temperature, usually 15/C/ 59/F to 35/C / 95/F,
and the instrument should be switched on about 10 minutes before the first
measurement to guarantee precise measuring results. If the device has been
cooled down (e.g. during transport), please wait until it has warmed up to room
temperature.
Geba53_e.03.b.wpd 6 8 5 1 4 3 0 / 1 0 2 / 0 3
MAICO
Operating Instructions MA 53
3.3 Getting familiar with your MA 53
Turn on the instrument with the
power switch " which is located at
the right side of the MA 53. The L R
PAT
STIM
Hz
FM PULSE SWAP TRACK
+
Hz
Geba53_e.03.b.wpd 7 8 5 1 4 3 0 / 1 0 2 / 0 3
MAICO
Operating Instructions MA 53
3.3.1 Using the keyboard of the MA 53
All main functions of the MA 53 are directly accessable by pressing a single
button. Not frequently used procedures like the calibration of the speech source
or the level step selection are hidden as a “second level function”. To select this
function just press the addressed button more than two seconds. A user menu
for the customization of the MA 53 is available for advanced users (see
chapter 8).
Please find below the description of
the main and second level functions of
each button: L R
Hz
FM PULSE SWAP TRACK
+
Hz
frequency mode
! left (blue) channel frequency down:
INSERT NOISE NOISE INSERT
FF MIC MIC FF
BC TAPE STIM TAPE BC
Talk Over AC TONE MODE TONE AC Talk Over
Geba53_e.03.b.wpd 8 8 5 1 4 3 0 / 1 0 2 / 0 3
MAICO
Operating Instructions MA 53
affects both channels.
5 swap function:
exchanges the outputs for both ears (i.e. the right controls now affect the left
ear).
6 tracking function:
enables the automatic tracking (see chapter 4.4 Masking).
2nd level: enters level step selection menu. The actual level step is displayed at the lower
center of the LCD-display #. Choose the level steps with the frequency up/down buttons
1and 0
' right (red) channel signal selector key:
changes the test signal from pure tone to tape/CD or microphone or noise or
switches the channel off.
2nd level: adjustment of talkback microphone and monitor phone (see chapter 6.2)
% right (red) channel receiver selector:
changes the receiver from AC to BC or FF or INSERT (if option assembled). The
lighted LED above shows the current selection.
2nd level: selects test mode HL, UCL,
MCL for selected receiver (see chapter
4.2)
7 right (red) channel talk over L R
microphone:
switches on talk over microphone STIM STIM
Geba53_e.03.b.wpd 9 8 5 1 4 3 0 / 1 0 2 / 0 3
MAICO
Operating Instructions MA 53
threshold test. The left channel is bone conduction with 70 dBHL , tone and
tracking function enabled. The right channel is set to 90 dBHL narrow band noise
with air conduction for masking. The test frequency is 1500 Hz and the test
signal is a warble tone. The lower section of the display shows that the talk over
microphone button 2 or 7 is pressed
and the audiologist talks to the patient 60 dB L 60% SRT
R 60 Air dB
with 70 dBHL. The HL in the upper row Air Tape/CD
Correct: 6
Wrong: 4 Speech-Noise
before 1500 Hz indicates that the
current measurement is stored as
Hearing Level (threshold) test.
Picture 10 shows the display for a Picture 10 MA 53 display for speech test
speech test with speech test material from tape or CD of the left ear and masking
of the opposite ear. The left part of the LCD-display # shows the settings for the
left channel:
level = 60 dBHL, receiver = AC, signal = tape/CD.
The middle of the LCD-display # shows the actual test and the result:
MS stands for multi syllable speech test,
speech discrimination = 60%, correct words = 6, wrong words = 4.
The right part of the LCD-display #shows the setting for the right channel:
level = 60 dBHL, receiver = AC, signal = speech-noise (masking).
The LCD-display # has an energy saver function: the backlight of the display is
automatically switched off after approximately three minutes. Any action with
the MA 53 such as pressing a button or turning the knobs switches the backlight
on again immediately.
Geba53_e.03.b.wpd 10 8 5 1 4 3 0 / 1 0 2 / 0 3
MAICO
Operating Instructions MA 53
4 Pure tone audiometry
4.1 Air Conduction (AC) Testing
In the hearing threshold test, the hearing threshold of the patient is measured in
comparison to the normal hearing threshold for air conduction. The test is started
on the ear with better hearing.
- The patient should sit at a distance of at least 1 m from the device.
- Eliminate any obstructions which will interfere with the placement of the
earphone cushions on the ear (i.e. hair, eyeglasses).
- Ensure that the headphones are put on correctly. Red
side on the right, blue side on the left. Adjust the
headband of the headphones so that the receivers are
at the correct height (the sound output grid 8 exactly
facing the ear canal).
- Start with the “better” ear at 1 kHz. (After switching
on, the frequency is automatically set to 1 kHz.) In the
following example we assume that the left is the
Picture 11 Headphone
“better” ear.
- Set the receiver selector $ to AC and the signal selector & to TONE.
- Set the left level control knob ( to a value just below expected hearing
loss.
- Explain to the patient that he
should press the button of the L R
PAT
STIM
Hz
FM PULSE SWAP TRACK
+
Hz
presenting the test signal as Picture 12 MA 53 controls and display for air
conduction threshold test of the left ear
described before.
- If the patient doesn't hear the test tone, increase the level with the level
control knob (. Proceed with presenting the test signal as described
before.
Geba53_e.03.b.wpd 11 8 5 1 4 3 0 / 1 0 2 / 0 3
MAICO
Operating Instructions MA 53
- When you find a stable threshold value, note the level and frequency. If the
MA 53 is connected to a PC, the value was stored with your last STIM
presentation.
- Test through the frequencies. Starting at 1 kHz, set the higher frequencies
first then the lower frequencies.
- Use the frequency up key / or 0 to select the higher frequencies and use
the frequency down ! or 1 to select the lower frequencies.
- Select the next frequency, increase the level again and proceed with
presenting the test signal as described before.
- Once all frequencies are tested choose the poorer ear and repeat the
hearing threshold test. You can do this with the corresponding controls on
the other side or just by pressing
the SWAP button 5. The SWAP 60 dB R 1500 HL
Hz L 30 AirdB
function enabled means that you Air Tone
Warble
Pulse Tone
measure the right ear using the
controls on the left side. The
ouput to the left and right
earphones are exchanged. Also, Picture 13 MA 53 display for AC-test with
the left and right display text is enabled SWAP function, Pulse and Warble tone
exchanged (see picture 13).
- The correct marks in an audiogram are: O (red) = right and X (blue) = left
Pulse tone: If required, the test can also be performed with a pulsed tone.
Press the PULSE button 4 and the pure tone will be switched 0.25s on and
0.25s off.
Warble tone: If required, the test can also be performed with a warble
tone. Press the FM button 3 and the pure tone will modulate frequency.
The warble tone can also be pulsed as described before.
For hygienic reasons it is important to clean the ear cushions of the
headphone (see chapter 11).
4.2 Uncomfortable Hearing Level (UCL) Testing
Testing of UCL can be measured using pure tone stimuli or speech (live voice
or tape/CD). The purpose is to determine the dB level at which the stimuli
becomes uncomfortable to the patient. This information is valuable for
determining the patient's upper dynamic range limit for proper hearing aid
fittings and for determining symptoms of recruitment.
Warning! Because this test uses high sound pressure levels, it is
extremely important to perform this test using the utmost caution to
avoid damaging the ear. To prevent the possibility of extreme
discomfort by the patient, it is important to start the test with levels
near the patient‘s MCL (Most Comfortable Level).
Geba53_e.03.b.wpd 12 8 5 1 4 3 0 / 1 0 2 / 0 3
MAICO
Operating Instructions MA 53
The UCL level is described as the level between very loud and loud perception of
the test signal.
- Press the receiver selector button UL
$ or % longer than 2 seconds. Air60 dB L 1000 Hz R 30 Air dB
The LCD-display in the upper row Tone Tone
changes from HL to UL (see
picture 14).
- Start as described in chapter 4.1
with a test level of 60 dBHL. Picture 14 MA 53 display for UCL-test left
- Present the tone briefly (max. 1s).
- If the signal was recognized by the patient as not uncomfortable
increase the level and proceed as described before.
- If the signal was uncomfortable for the patient note the values.
- Proceed accordingly with other test frequencies.
Geba53_e.03.b.wpd 15 8 5 1 4 3 0 / 1 0 2 / 0 3
MAICO
Operating Instructions MA 53
4.5 Diagnostic Tests
4.5.1 Suprathreshold tests
The MA 53 offers additional suprathreshold tests which can offer more
information about the kind of the hearing loss.
To access suprathreshold tests HL
press the PULSE button 4 longer Air70 dB L 1000 Hz R 30 Air dB
than two seconds. The Tone Tone
suprathreshold test menu is now ABL 0.5 1.0 sec SISI
displayed on the lower part of DLI 4.0 3.0 2.0 1.5 1.0 0.5 dB
Geba53_e.03.b.wpd 16 8 5 1 4 3 0 / 1 0 2 / 0 3
MAICO
Operating Instructions MA 53
4.5.1.2 The DLI (Lüscher) Test
The Intensity Difference LImen for loudness test (or Just Noticeable
Difference, JND) is another recruitment test. A pure tone is amplitude
modulated at a rate of 2/second. The patient has to determine whether this
signal was steady or undulating (beating).
- To start the test select a modulation amplitude from 0.5 dB to 4.0 dB in the
suprathreshold menu (as decribed before under 4.5.1) and press the STIM
bar * or +.
- The LCD-display # shows DLI, 50 dB L 2000 HL
Hz R 50 dB
along with the chosen modu- Air Tone DLI 1.5 dB
Air
Tone
lation amplitude (see picture 22).
- Now adjust the required test
level with the level control knob
Picture 22 MA 53 display with DLI test
( or ) and present the test
signal by pressing the STIM bar * or +.
- To exit the test briefly press the PULSE button 4.
4.5.1.3 The SISI Test
The SISI (Short Increment Sensitivity Index) is guided by the principle that
patients with cochlear impairment are hypersensitive to small intensity
increments. The continuous test tone is increased by 1 dB for a period of
0.2 seconds every 4.8 seconds. Whenever the patient hears the increment,
he/she presses the patient response switch. The SISI test information and
the test score are shown on the LCD-display #. The test will end
automatically after 20 presented increments. The score is expressed as a
percentage of ratio of the increments heard to the delivered increments (all
increments heard = 100% - no increments heard = 0%). A high score
indicates a cochlear impairment. A low score is related to normal hearing or
conductive or retrocochlear pathology.
4.5.1.3.1 Preparation of the SISI Test
- To start the test select SISI in the suprathreshold menu (as decribed before
under 4.5.1) and press the STIM bar * or +. The display # changes to the
SISI conditioning screen (as shown in picture 23).
- Select the test frequency with HL
the frequency keys //! or Air 65 dB L 2000 Hz R -10 Air
dB
1/0. For the test you should Tone SISI Conditioning Tone
Geba53_e.03.b.wpd 17 8 5 1 4 3 0 / 1 0 2 / 0 3
MAICO
Operating Instructions MA 53
- Adjust the test level with the level control knob ( or ). The level should
be set to a value 20 dB above the individual hearing threshold (measured
under chapter 4.1). It must reach at least 60 dBHL.
- Recognize that in the SISI mode the test signal is presented continuously.
You can interrupt the test by pressing the STIM bar * or + for the time
you hold it down.
- The patient must be instructed: "You will now hear a continuous tone.
Every time it becomes louder, immediately push the switch”.
4.5.1.3.2 Training of the Patient
Only with careful training during the following conditioning phase a valid
test result can be achieved.
- The training starts with a HL
modulation level of 5 dB. That Tone
65
Air
dB L 2000 Hz R -10 dB
Air
SISI Conditioning Tone
means a continuous tone, e.g.
65 dBHL, is briefly increased every Tone now Answer now
4.8 seconds for 0.2 seconds by
5 dB Presented: 4 Detected: 2
5 dB, e.g. to 70 dBHL. During the Picture 24 MA 53 display with
increment time the text “Tone SISI conditioning 5 dB and presented increment
now” is shown on the lower area of the LCD display # (see picture 24).
The text “Answer now” appears for approx. 1.5 seconds. During this
period the patient may respond. If he/she answers too early or too late the
answer will not be accepted by the MA 53.
- The number of detected increments is counted and shown on the lower
right area of the display. Furthermore, the number of presented increments
is shown (see picture 24).
- When you are sure that the patient has understood the object of the test,
press the the PULSE button 4. The modulation level will be changed to
3 dB.
- If the patient has also understood this part of the test, press the PULSE
button 4 again. The modulation level will be reduced to 2 dB.
4.5.1.3.3 SISI Test
- If the patient has understood the test up to here, press the PULSE button
4 again to start the SISI test.
- The LCD-display indicates SISI HL
Test (see picture 25) and the Tone
65
Air
dB L 2000 Hz R -10 dB
Air
SISI Test Tone
modulation level is adjusted to
1 dB. Now 20 increments will 1 dB Presented: 0 Detected: 0
be presented to the patient.
- You can delay the test for the Picture 25 MA 53 display with SISI test
Geba53_e.03.b.wpd 18 8 5 1 4 3 0 / 1 0 2 / 0 3
MAICO
Operating Instructions MA 53
time you press STIM bar * or +. This could be helpful in determining if the
patient malingers.
- The test will end automatically HL
after 20 presented increments. Air 65 dB L 2000 Hz R -10 Air dB
The number of detected Tone SISI-Result Tone
increments is counted and
indicated as percentage of the 20% Presented: 20 Detected: 4
presented increments (see
picture 26). Picture 26 MA 53 display with SISI result
- If you are sure that the result of HL
the test is consistent you can Air 80 dB L 1500 Hz R -10 Air dB
abort the test at any time by Tone SISI-Aborted Tone
pressing the PULSE button 4.
The result is then indicated 50% Presented: 10 Detected: 5
accordingly (picture 27).
- Retrocochlear hearing impair- Picture 27 MA 53 display with aborted SISI test
ment is represented by low
scores (< 25 %), cochlear defects by high scores (> 70 %).
- Press any button to leave the SISI test.
4.5.2 The Stenger Test
The Stenger test is conducted as a malingering test in cases of alleged
hearing loss or unilateral deafness. It is based on the premise that when
two tones of the same frequency are presented simultaneously into both
ears, only the louder tone will be perceived.
- The patient with a feigned hearing loss will not be able to hear the quiet
tone, but will also not admit being able to hear the loud tone.
- Select the desired frequency with the frequency keys //! or 1/0 and
adjust the sound intensities with the level control knobs ( and ).
- Present the tone with the STIM bars * and +.
- The easiest procedure is to present the desired frequency into to the better
ear at a level of 10 dB above the threshold. At the same time present the
signal into the poorer ear (the one with the possible feigned loss) at a level
10 dB below the admitted threshold. If there is a real loss in the poorer ear,
the patient will be unaware of the signal in that ear and will respond only
to the tone in the good ear. If the patient does not respond at all it is likely
that the admitted threshold is incorrect and you have a “positive” Stenger.
4.5.2 The Threshold Decay (Carhart) Test
This examination is conducted as a fatigue test in cases of sensitivity loss.
- Select the desired frequency with the frequency keys //! or 1/0.
Geba53_e.03.b.wpd 19 8 5 1 4 3 0 / 1 0 2 / 0 3
MAICO
Operating Instructions MA 53
- Adjust the sound level to a value of about 5 dB above the hearing threshold
with the level control knob ( or ).
- Instruct the patient to start pressing the patient response switch when he
hears a tone and to release the switch when he doesn't hear the tone
anymore.
- Enter continuous tone presentation by pressing the STIM MODE button .
to present the test signal or press the corresponding STIM bar * or +.
- The PAT response LED ; lights HL
and after two seconds a bar stop Air 35 dB L 500 Hz R -10 Air dB
watch with a scale from 0 to 60 Tone Tone
seconds appears on the lower Decay
area of the LCD-display # (see 0 10 20 30 40 50 60sec
picture 28). Picture 28 MA 53 display with
- When the patient releases the initial tone decay test
switch the bar of the stop watch stops and you can see the time the patient
heard the tone (see example
HL
picture 29). 35
Air
dB L 500 Hz R -10 Air dB
- If the time was less than 60 Tone Tone
seconds increase the level with Decay
the level control knob ( or ) by 0 10 20 30 40 50 60sec
5 dB. Picture 29 MA 53 display with
example tone decay test
- Press any button to erase the
stop watch on the display # and start the test again.
- Continue this procedure until the patient can hear the tone at least for
60 seconds.
- The threshold shift can be entered on the tone audiogram pad by drawing
a wavy line under the selected frequency.
Geba53_e.03.b.wpd 20 8 5 1 4 3 0 / 1 0 2 / 0 3
MAICO
Operating Instructions MA 53
4.6 Dual Frequency Mode
The MA 53 offers the dual frequency mode to the right and the left ear
with different frequencies. This “true two channel” feature can be used to
detect frequency disorders.
- To enable the dual frequency mode press the left frequency up key / and
frequency down key ! simul-
HL
taneously. 50
Air
dB L 1000 2000 Hz R 50 dB
Air
- The LCD-display # now shows Tone Tone
the test frequency of the left
channel on the left (in example
picture 30 1000 Hz) and the test
Picture 30 MA 53 display with
frequency of the right channel on dual frequency mode
the right (in example picture 30
2000 Hz) of the center.
- Use the frequency up key / to select the higher frequencies and the
frequency down key ! to select the lower frequencies for the left channel.
- The frequencies of the right channel could be selected independent from
the left with the frequency up key 0 and the frequency down key 1.
- To exit the dual frequency mode press the left frequency up key / and
frequency down key ! simultaneously again.
- The LCD-display # now shows HL
the test frequency for both Air50 dB L 1500 Hz R 50 Air
dB
channels in the center (in Tone Tone
example picture 31 1500 Hz).
Geba53_e.03.b.wpd 21 8 5 1 4 3 0 / 1 0 2 / 0 3
MAICO
Operating Instructions MA 53
4.7 High Frequency Audiometry (optional)
The optional High Frequency Audiometry offers tone audiometry with the
additional test frequencies 9000 Hz, 10000 Hz, 11200 Hz, 12500 Hz,
14000 Hz and 16000 Hz. Tests using this advanced frequencies give
additional information of the individual hearing performance. They are
especially usefull for the early detection of hearing damages of younger
persons.
- The headphone Sennheiser HDA
200 is designed for high Air 30 dB L 16000 HzHL R -20 Air dB
frequency audiometry. The HDA Tone Tone
200 works over the full
frequency range from 125 Hz to
16000 Hz.
Picture 32 MA 53 display for
- It is possible to use the HDA 200 high frequency audiometry
as standard AC receivers for all
AC tests or to use two AC-receivers with the MA 53. In this case the
normal AC receiver (TDH 39) goes up to 8000 Hz. The HDA 200 is selected
with the receiver selector key $ or % as “insert phone”.
- The high frequency bone conduction receiver KLH 96 is optionally available
for BC measurements from 125 Hz to 16000 Hz.
- The tests with high frequencies are done as described before in chapter 4.
- The minimum hearing level starts for frequencies from 9000 Hz to
16000 Hz at -20 dBHL.
Geba53_e.03.b.wpd 22 8 5 1 4 3 0 / 1 0 2 / 0 3
MAICO
Operating Instructions MA 53
6 Speech Audiometry
6.1 Connecting the speech source
6.1.1 Connecting a CD or tape-player
To conduct speech tests using speech test
material you can connect a CD- or tape-
player to the MA 53. Use a standard
phono plug cable to connect the right and R FF L BC MASK BC R INSERT L R AC L
left outputs of the device with the SIGNAL TAPE / CD MONITOR TB MIC
PAT
picture 50).
- On every CD or tape with speech 2 3 7
10 18
19 23
24
test material you have a reference
signal, such as a reference tone Picture 49 MA 53 controls and displays
or speech simulating noise. for speech calibration
- Perform the reference signal with Tape or CD.
- Use the left knob ( for the left SRT
channel and the right knob ) for Air30 dB L Correct:0% 0
R 30 Air dB
the right channel and adjust the Tape/CD Wrong: 0 Tape/CD
levels until both VU-meters @ Calibration
and # show all yellow and one 95% 95%
green light. Picture 50 MA 53 display with
- If one or more red lights are on, speech calibration of CD-player or tape
reduce the level using the corresponding knobs ( or ).
- Store the calibration by pressing
the STIM bar * or +. SRT
5.2.1 Calibration of the microphone
30 dB L Air
0% R Correct:
30 dB
0 Air
Microphone Wrong: 0 Microphone
for live speech test
- To calibrate the SPEECH MIC
input select MIC with the left Picture 51 MA 53 display with
SIGNAL selector key &. speech calibration of the microphone
Geba53_e.03.b.wpd 24 8 5 1 4 3 0 / 1 0 2 / 0 3
MAICO
Operating Instructions MA 53
- Press the left SIGNAL key & longer than 2 seconds.
- In the lower area of the display # the text Calibration appears. You also
see the actual settings in % of the maximum calibration for both channels.
- Position yourself the customary distance from the microphone and speak
test words.
- Use the left knob ( to adjust the level until both VU-meters @ and # show
all yellow and one green light.
- If one or more red lights are on, reduce the level using the knob (.
- Store the calibration by pressing the STIM bar * or +.
6.3 Connection and adjustment of the monitor phone
6.3.1 Connection of the monitor phone
Connect the monitor phone plug E to the
MONITOR connector A at the rear panel
of the MA 53 (see picture 52). To avoid
pinching the cables when closing the
cover, lay the cables in the slots provided.
The monitor phone allows you and the R FF L BC MASK BC R INSERT L R AC L
patient to hear the speech test material SIGNAL TAPE / CD TAPE / CD MONITOR TB MIC
R SPEECH L
Geba53_e.03.b.wpd 25 8 5 1 4 3 0 / 1 0 2 / 0 3
MAICO
Operating Instructions MA 53
The speech-recognition threshold SRT is the hearing threshold for speech.
It is the lowest level at which the patient correctly recognizes the stimuli
50% of the time. Usually, recognition is indicated by repetition of the
speech-stimulus item. The speech test can be done with recorded speech
test material from CD or tape or with the microphone and live voice using
standardized word lists. Different methods for speech tests are standardized
worldwide such as the ASHA guidelines for determining the threshold level
for speech (Asha 21, page 353-356).
- Explain to the patient that he should repeat each word he hears.
- The patient should sit at a distance of at least 1 m from the device.
- Eliminate any obstructions which will interfere with the
placement of the earphone cushions on the ear (i.e.
hair, eyeglasses).
- Ensure the headphones are put on correctly. Red side
on the right, blue side on the left. Adjust the
headband of the headphones so that the receivers are
at the correct height (the sound output grid 8 exactly Picture 54 Headphone
facing the ear canal).
- Set the receiver selectors $ and SRT
% to AC and the signal selectors Air30 dB L Correct:0% 0
R 30 Air dB
Geba53_e.03.b.wpd 26 8 5 1 4 3 0 / 1 0 2 / 0 3
MAICO
Operating Instructions MA 53
discrimination score. In the
SRT
example shown in picture 56, 40 dB L 50% R -10 AirdB
Air Correct: 5
one half of the words were heard Tape/CD Wrong: 5 Tape/CD
correctly.
- If you are working online with a
PC press the STIM bar * or + to
Picture 56 MA 53 display with
transfer the test result to the PC typical result for speech audiometry
(see also chapter 10).
- To clear the score for the next test, press the FM key 3.
It is also possible to perform the speech test with loudspeakers (e.g. for the
verification of hearing aid fitting) or bone conduction.
- Select the required transducer with the RECEIVER selection button $ or %
and proceed as described before.
6.5 Masking
Speech audiometry has similar masking rules to those desribed in chapter
4.4 for pure tone audiometry.
When TAPE/CD or MIC is selected on one channel, and NOISE masking on
the opposite channel, the MA 53 automatically performs speech masking
noise.
Geba53_e.03.b.wpd 27 8 5 1 4 3 0 / 1 0 2 / 0 3
MAICO
Operating Instructions MA 53
7 Advanced features
7.1 Complex mixing
The complex mixing feature of the MA 53 allows the operator to route two
different test signals to the same ear. The test signals and the test levels are
independently adjustable but will be presented to one ear (ipsilateral). This
feature could be used to test with speech in noise, competing message
tests, tone on noise or several other combinations.
- To enter the complex mixing
press the swap key 5 longer Air30 dB L 1000 HzHL R 30 Air dB
than 2 seconds. Tone NB-Noise
Geba53_e.03.b.wpd 28 8 5 1 4 3 0 / 1 0 2 / 0 3
MAICO
Operating Instructions MA 53
7.2 Advanced Noise Selection
The MA 53 offers for advanced masking or testing the selection of narrow
band noise, speech masking noise and white noise as test signal.
- Enable this feature from the user setup menu (see chapter 8.17)
- The additional signals are now available as test signals. Select the required
signal by pressing the signal selector key & for the left channel and key '
for the right channel.
- When the noise-LED lights you can choose between NB-Noise, Speech-
Noise and White Noise by pressing the signal selector key.
7.3 Talk-back microphone
The optional talk-back microphone allows the patient to speak to the
operator. This feature is used if the
patient sits in a sound excluding
booth. The communication from the
patient to the operator will be heard
with the monitor phone. R FF L BC MASK BC R INSERT L R AC L
Picture 63
MA 53 display with
adjustment of the assistant phone
Geba53_e.03.b.wpd 30 8 5 1 4 3 0 / 1 0 2 / 0 3
MAICO
Operating Instructions MA 53
8 Quick reference guide
8.1 Startup settings
Air conduction on both channels
Pure tone on both channels
30 dBHL on both channels
Presenter mode
8.2 Receiver selection
Press either the right % or left $ RECEIVER key to select the receiver (AC,
BC, FF, INSERT). The display # shows the selected receiver below the levels.
Not available receivers will be skipped. It is not possible to select bone on
both channels simultaneously.
After a new selection the level will be reset to 30 dBHL.
8.3 Signal selection
Press either the right ' or left & SIGNAL key to select the signal. The kind
of NOISE signal depends on the signal of the opposite channel. Speech
masking noise is used if on the opposite channel the speech signal is
selected (see 8.4 below). The display # shows the selected signal below the
receiver text.
The 5. Position is signal off. The light at the signal key & or ' and the
STIM LED , or - is off and the level in the display # is erased.
8.4 How to select tone or speech audiometry mode
There are two modes of audiometry.
Tone audiometry : shows level and frequency on the display #
Speech audiometry : shows level, correct and incorrect words and
speech discrimination in percentage on the display
This table shows how to select one of these modes:
Geba53_e.03.b.wpd 31 8 5 1 4 3 0 / 1 0 2 / 0 3
MAICO
Operating Instructions MA 53
8.5 Frequency selection
Use one of the two sets of frequency keys //! or 1/0 to select the
frequency. The maximum and minimum frequency depends on the receiver
you have selected.
8.6 Warble tone (FM)
Press the FM key 3 to activate the warble tone. Warble is shown in the
display #.
Press the FM key 3 again to switch back to pure tone.
8.7 Pulse tone
Press the PULSE key 4 to activate the pulse tone. Pulse is shown in the
display # and the light above the PULSE key 4 is on.
Press the PULSE key # again to deactivate pulsing.
8.8 Suprathreshold tests
Press the PULSE key # longer than 2 seconds. In the lower area of the
display # the parameters of the ABLB, DLI and SISI test are shown.
Using the PULSE key 4, select the test and parameter.
Press one of the STIM bars * or + to select your choice. The lower area of
the display # is cleared and the selected test is shown on the display #.
The light above the PULSE key 4 is on.
To change to another parameter press the PULSE key 4 again and hold
longer than 2 seconds.
To end the selected test, press the PULSE key 4 briefly.
8.8.1 SISI Test
Once SISI is selected, the test begins immediately by conditioning in
5 dB increments. Press the PULSE key again to enter the next levels
of SISI conditioning (3 dB, then 2 dB) and to start the SISI test (1dB).
During the test the display # shows the increment level, the number
of presented increments and the number of correct answers given by
the patient. In addition, “Tone Now“ is displayed when an
increment is presented and “Answer Now“ is displayed during the
answer acceptance time.
During SISI-conditioning it is possible to clear the counters by
pressing the FM key 3.
To cancel the conditioning and the SISI test and to return to standard
operation press the SWAP key 5.
After 20 increments the SISI test is finished. The text SISI Result with
the result in percentage is shown.
You can stop the SISI test with the PULSE key 4. The text SISI
aborted and the actual result in percentage is shown.
Geba53_e.03.b.wpd 32 8 5 1 4 3 0 / 1 0 2 / 0 3
MAICO
Operating Instructions MA 53
Press the PULSE key 4 to quit the SISI test.
8.8.2 Decay test (Carhart)
If the patient presses the patient response switch longer than 3 seconds, in
the lower area of the display # a decay bar appears and shows the counted
time until the patient releases the patient response switch.
To clear the display # and reset the decay function press one of the STIM
bars * or +.
8.9 Swap function
Press the SWAP key 5 to exchange the left and right channel. The big
letters R and L will be swapped in the display # and the light above the
SWAP key 5 is on.
Press the SWAP key 5 again to deactivate swapping.
8.10 Tracking function
Press the TRACK key 6 to activate tracking. Both the left and right sides of
the display # will say TRACKED, meaning that both channels are tracking
each other. If NOISE is selected on one channel, the text TRACKED is
erased, because a noise channel is not allowed to track a signal channel.
Press the TRACK key 6 again to deactivate tracking.
8.11 STIM Mode selection
Press STIM MODE . key to switch between continuous presentation and
presentation by pressing the STIM bar * or +. When in continuous
presentation mode the STIM bars * and + function as interrupters. The
light above the STIM MODE key . is on if continuous presentation is
selected.
Speech and noise are always in continuous presentation, except when both
channels are selected as noise.
Press the STIM MODE key . again to deactivate continuous presentation.
8.12 Interlock function
Press the STIM MODE key . longer than 2 seconds to activate the interlock
function. The display # shows the text Interlock.
Interlock works only if both channels have the same kind of signal (speech,
tone or noise). In modes where no interlock is available the display # shows
No Interlock.
Press the STIM MODE key . longer than 2 seconds to deactivate the
interlock function.
8.13 Talk-over microphone
Press one of the TALK OVER keys 2 and 7. In the lower area of the
display # the volume level of the talk-over microphone is shown.
Adjust the talk forward volume with the left ( or right ) knob.
Release the TALK OVER key 2 or 7 if you have finished talking to the
patient.
Geba53_e.03.b.wpd 33 8 5 1 4 3 0 / 1 0 2 / 0 3
MAICO
Operating Instructions MA 53
8.14 Talk-back microphone and Monitor volume control
Press the right SIGNAL key ' longer than 2 seconds and the volume control
bars for TB and monitor are shown.
Adjust the monitor volume with the right knob ).
Adjust the talk-back microphone volume with the left knob (.
To store the actual settings press one of the STIM bars * or +.
8.15 Speech audiometry
To select speech audiometry switch both channels to TAPE/CD or MIC by
using the SIGNAL selector keys $ and '. In speech mode the display #
shows the percentage of correct repeated words. Below this percentage the
number of correct and wrong repeated words is shown.
To count the correct words press one of the frequency up keys / or 0.
To count the wrong words press one of the frequency down keys ! or 1.
To clear the counter press the FM key 3.
8.16 Speech calibration
To calibrate the speech inputs TAPE/CD or MIC select the signal you'd like
to calibrate with the left SIGNAL selector key &.
Press the left SIGNAL key & longer than 2 seconds. In the lower area of
the display # the text Calibration appears. Perform the reference signal
with Tape, CD or microphone. Use the left knob ( for the left channel and
the right knob ) for the right channel and adjust the levels until both VU-
meters @ and # show all yellow and one green light. If one or more red
lights are on, reduce the level using the corresponding knobs ( or ).
Store the calibration by pressing the STIM bar * or +.
8.17 User menu
To enter the User Menu you have to press the FM key &.
To choose an item from the menu use the FREQUENCY keys //! or 1/0.
To select the chosen item press one of the STIM bars * or +.
Change the actual setting with the FREQUENCY keys //! or 1/0.
To confirm the changed setting press one of the STIM bars * or +.
To return without change press the FM key &.
To return to the standard operation press the FM key & at the user menu.
Geba53_e.03.b.wpd 34 8 5 1 4 3 0 / 1 0 2 / 0 3
MAICO
Operating Instructions MA 53
These menu items are available :
FF Set 1/2* Select the calibration for the first or the second loudspeaker set
Bin Monaural Select the signal source channel for CD/TAPE (Bin/Left/Right)
User Setup Setup of special functions
User Preference Customization of the instrument
* = only accessable if option is purchased
Geba53_e.03.b.wpd 35 8 5 1 4 3 0 / 1 0 2 / 0 3
MAICO
Operating Instructions MA 53
9 Recommended literature
Audiometric Interpretation: A Manual of Basic Audiometry
Lloyd, Lyle L., and Harriet Kaplan
Baltimore: University Park Press, 1980
Auditory Diagnosis
Silam, Shlomo and Carol A. Silvermann
San Diego / London: Singular Publishing Group, 1997
Geba53_e.03.b.wpd 36 8 5 1 4 3 0 / 1 0 2 / 0 3
MAICO
Operating Instructions MA 53
10 Computer interface
The MA 53 has a built-in serial RS 232 PC-interface for data transfer to a
connected PC. The serial cable to connect the MA 53 with the PC should be a
normal, non twisted 9-pin extension cable. It has a male and a female connector
with all wires directly connected 1:1. It is also sold as monitor extension cable.
Caution: The computer you are using must meet electric safety
requirements, such as IEC 601-1 or UL. This is to avoid electric shock of
either the patient or you.
All test results are transferred online to the PC and stored in the database.
- In pure tone audiometry initially
all measured data is transferred Air30 dB L 1000 HL Hz L 30 Air
dB
as hearing threshold data. You Tone Tone
can see the letters HL meaning
threshold measurement in the
upper middle of the LCD- Picture 64 MA 53 display with
display # (see picture 64). pure tone threshold setting
- To test the uncomfortable level (UCL) press the receiver selector button $
or % longer than 2 seconds and the LCD-display # in the upper row
changes from HL to UL. All measured data is now transferred as UCL data.
- To test the most comfortable level (MCL) press the receiver selector
button $ or % longer than 2 seconds and the LCD-display # in the upper
row changes from UL to ML. All measured data is now transferred as MCL
data.
- In speech audiometry initially all SRT
measured data is transferred as Air30 dB L Correct:0% 0
R 30 AirdB
Geba53_e.03.b.wpd 37 8 5 1 4 3 0 / 1 0 2 / 0 3
MAICO
Operating Instructions MA 53
11 Care and maintenance of the instrument
Disconnect the power plug before cleaning!
- To clean the instrument, headphones,
bone conduction receiver, loud-
speaker and other accessories use a
soft cloth dampened with a little
warm soapy water or washing-up
liquid; no alcohol or spirits should
be used.
- The ear cushions Z of the headphone
can be detached for cleaning. To
remove the ear cushion Z pull it
gently from the earphone. To
assemble the cleaned or changed ear
cushion press it on the earphone and
make sure that it sits properly. The
sound outlet hole must be exactly in
the middle of the earphone.
Picture 66 Changing of the
ear cushions
- During cleaning, please ensure that no liquid runs into the switches, level
control, headphone capsules or loudspeaker openings.
Geba53_e.03.b.wpd 38 8 5 1 4 3 0 / 1 0 2 / 0 3
MAICO
Operating Instructions MA 53
12 Trouble shooting
If you should find that your instrument is no longer working properly during a
test run, please check the following points:
1 AT RS 232
If your instrument still does not work properly after this short check, please
consult your dealer or service center.
Geba53_e.03.b.wpd 39 8 5 1 4 3 0 / 1 0 2 / 0 3
MAICO
Operating Instructions MA 53
13 Technical Data
The MA 53 audiometer is an active, diagnostic medical
product according to the class IIa of the EU medical directive
0124 93/42/EEC.
Standards: IEC 645-1/ EN 60 645-1: Type 2
IEC 645-2 : Type A
ANSI S3.6-1996 : Type 1 HF A
Test-Frequencies: 125 Hz, 250 Hz, 500 Hz, 750 Hz, 1 kHz, 1.5 kHz, 2 kHz,
(individual selectable 3 kHz, 4 kHz, 6 kHz, 8 kHz, 12 kHz, 9 kHz*, 10 kHz*,
for each channel) 11.2 kHz*, 12.5 kHz*, 14 kHz*, 16 kHz*
* = with optional HDA 200
Level step: 5 dB, 2 dB or 1 dB level steps (user selectable)
Sound Pressure Level: AC with earphone Telephonics TDH 39:
- 10 dBHL ... 120 dBHL ( 125 Hz -10 dBHL .. . 80 dBHL)
( 250 Hz -10 dBHL ... 100 dBHL)
( 500 Hz, 4 kHz, 6 kHz -10 dBHL ... 115 dBHL)
( 8 kHz -10 dBHL ... 105 dBHL)
( 12 kHz -10 dBHL ... 90 dBHL)
BC with bone conduction receiver Radio Ear B 71:
- 10 dBHL ... 80 dBHL ( 250 Hz, 6 kHz -10 dBHL ... 50 dBHL)
( 500 Hz, 750 Hz, -10 dBHL ... 60 dBHL)
( 1 kHz -10 dBHL ... 75 dBHL)
( speech -10 dBHL ... 75 dBHL)
( 125 Hz, 8 kHz no output )
AC with optional earphone Sennheiser HDA 200:
- 10 dBHL ... 120 dBHL
( 125 Hz, 8 kHz, 9 kHz -10 dBHL ... 100 dBHL)
( 250 Hz, 4 kHz, 6 kHZ, 9 kHz -10 dBHL ... 110 dBHL)
( 2 kHz, 3 kHz -10 dBHL ... 115 dBHL)
( 10 kHz, 11.2 kHz, 12,5 kHz -10 dBHL ... 90 dBHL)
( 14 kHz -10 dBHL ... 80 dBHL)
( 16 kHz -10 dBHL ... 60 dBHL)
Insert earphone with optional EAR 3A:
- 10 dBHL ... 110 dBHL ( 125 Hz -10 dBHL ... 75 dBHL)
( 250 Hz, 6 kHz -10 dBHL ... 90 dBHL)
( 750 Hz, 1 kHz -10 dBHL ...115 dBHL)
( 8 kHz -10 dBHL .. . 85 dBHL)
FF with optional loudspeaker Canton Plus XS: (not available in USA)
Geba53_e.03.b.wpd 40 8 5 1 4 3 0 / 1 0 2 / 0 3
MAICO
Operating Instructions MA 53
FF with optional portable loudspeaker MAICO SBC:
- 10 dBHL... 90 dBHL (measurement in 1m distance)
( 125 Hz, 250 Hz no output )
( 750 Hz, 8 kHz -10 dBHL ... 80 dBHL)
( 500 Hz, 1 kHz, 1.5 kHz, speech -10 dBHL ... 85 dBHL)
( 3 kHz, 4 kHz -10 dBHL ... 100 dBHL)
Test Signal: Pure tone, pulse tone, warble tone, narrowband noise,
broadband noise (speech masking noise)
Modulation: Pulse tone: 0.5 s on/off
Warble tone: triangular, freq. deviation 5%,
repetition rate 5 Hz
Tests: ABLB (Fowler): 0.5 s and 1 s
DLI (Lüscher): 4; 3; 2; 1.5; 1 and 0.5 dB
SISI test: 5 dB, 3 dB, 2 dB conditioning;
1 dB test; Modulation time 4.8/0.2 s
Carhart decay test (built-in stop watch)
Stenger test
Functions: Tone presenter / interrupter
Interlock (tone presentation of both channels
simultaneously)
Channel swap
Tracking (fixed level difference between both channels)
Mixing (ipsilateral presentation of both channels)
Talk forward, Talk back, Monitor phone, Assistant phone
Warm-up time: less than 10 min after power on
Environment + 15 ... + 35 /C / + 59 ... + 95 /F (operation)
conditions: + 5 ... + 50 /C / + 41 ... + 122 /F (storage)
Maximum humidity 90 % (storage and operation)
Dimensions: W x D x H: 36 x 46 x 15 cm / 14.2” x 18.1” x 5.9”
Weight: 4.6 kg / 10.1 lbs (with accessories 5.8 kg / 12.8 lbs)
Power Supply: 100 - 240 V~ 50/60 Hz ±10 %
Power Consumption: approx. 25 VA
Device Fuses: 2x 1A slow blow
Geba53_e.03.b.wpd 41 8 5 1 4 3 0 / 1 0 2 / 0 3
MAICO
Operating Instructions MA 53
Connection plugs: Connection Specification
D power socket left/right=power, (100 ... 240 V~ 50/60 Hz)
middle=prot.GND
C power fuses 1 A slow blow
E serial PC-interface 1=GND, 2=RX, 3=TX, RS 232 C
5=GND, 6=DTR
F patient response switch sleeve=GND, tip=in RI= 500 S
G signal output right sleeve=GND, tip=out ZA= 47 kS, UA= 8 Veff
H signal output left sleeve=GND, tip=out ZA= 47 kS, UA= 8 Veff
I FF (loudspeaker) right channel sleeve=GND, tip=out ZA= 4 S, UA= 8 Veff
J tape/CD input right channel sleeve=GND, tip=in ZI= 47 kS, UI= 0.04 - 5 Veff
K FF (loudspeaker) left channel sleeve=GND, tip=out ZA= 4 S, UA= 8 Veff
< tape/CD input left channel sleeve=GND, tip=in ZI= 47 kS, UI= 0.04 - 5 Veff
L masking phone for BC (option) sleeve=GND, tip=out ZA=10 S, UA=1 Veff
= tape/CD input right channel sleeve=GND, tip=in ZI= 47 kS, UI= 0.04 - 5 Veff
< tape/CD input left channel sleeve=GND, tip=in ZI= 47 kS, UI= 0.04 - 5 Veff
M BC (bone conduction receiver) sleeve=GND, tip=out ZA= 4 S, UA= 8 Veff
A monitor phone sleeve=GND, tip=out ZA= 250 S, UA= 8 Veff
N insert phone right channel sleeve=GND, tip=out ZA=10 S, UA=1 Veff
B talk-back microphone sleeve=GND, tip=in ZI= 1 kS, UI= 0.38 - 500 mVeff
assistant phone sleeve=GND, ring=out ZA= 250 S, UA= 8 Veff
O insert phone left channel sleeve=GND, tip=out ZA=10 S, UA=1 Veff
> right live voice microphone sleeve=GND, tip=in ZI= 1 kS, UI= 0.38 - 500 mVeff
P AC (headphone) right channel sleeve=GND, tip=out ZA=10 S, UA=1 Veff
? left live voice microphone sleeve=GND, tip=in ZI= 1 kS, UI= 0.38 - 500 mVeff
Q AC (headphone) left channel sleeve=GND, tip=out ZA=10 S, UA=1 Veff
R FF L BC MASK BC R INSERT L R AC L
100 - 240 V / 50 - 60 Hz
PATIENT
1 AT RS 232 INTERFACE RESPONSE SIGNAL TAPE / CD TAPE / CD MONITOR TB MIC R SPEECH L
R L R 2 L R 1 L MIC
A C E G I K M O Q S U
B D 68 FRear ofHMA 53Jwith connection
Picture L N plugs P R T V
Geba53_e.03.b.wpd 42 8 5 1 4 3 0 / 1 0 2 / 0 3
MAICO
Operating Instructions MA 53
Calibration values: AC-Receiver Telephonics TDH 39 Force: 4 ... 5 N
Geba53_e.03.b.wpd 43 8 5 1 4 3 0 / 1 0 2 / 0 3
MAICO
Operating Instructions MA 53
Calibration values: AC-High Frequency Receiver Sennheiser HDA 200
Force: 9 ... 10N
Frequency [Hz] Reference equivalent threshold Sound damping
sound pressure level
ISO 389 - 5 ANSI S3.6-1
with Coupler IEC 318 with Coupler IEC 318
and Type 1 Adapter and Type 1 Adapter
[dB] (re 20 :Pa) [dB] (re 20 :Pa) [dB]
125 29,5 14.5
250 18.0 16.0
500 9.5 22.5
750 6.5 -
1000 6.5 28.5
1500 5.5 -
2000 3.0 32.0
3000 3.0 -
4000 8.5 45.5
6000 9.5 -
8000 17.0 16.0 44.0
9000 18.0 17.0 -
10000 22.0 21.5 -
11200 22.5 21.0 -
12500 27.5 27.5 -
14000 36.0 37.5 -
16000 58.0 58.0 -
Geba53_e.03.b.wpd 44 8 5 1 4 3 0 / 1 0 2 / 0 3
MAICO
Operating Instructions MA 53
Calibration values: FF : Free sound field (0 degree incidence)
Frequency [Hz] Reference equivalent threshold
sound pressure level
ISO 389 - 7 ANSI S3.6-1996
[dB] (re 20 :Pa) [dB] (re 20 :Pa)
125 22 22
250 11 11
500 4 4
750 2 2
1000 2 2
1500 0.5 0.5
2000 -1.5 -1.5
3000 -6 -6
4000 - 6.5 - 6.5
6000 2.5 2.5
8000 11.5 11.5
Geba53_e.03.b.wpd 45 8 5 1 4 3 0 / 1 0 2 / 0 3
MAICO
Operating Instructions MA 53
Standard accessories: 1 air conduction receiver Telephonics TDH 39
1 bone conduction receiver Radio Ear B 71
1 microphone (for live speech)
1 patient response switch
1 power cord
1 pad of audiogram paper (50 sheets)
Geba53_e.03.b.wpd 46 8 5 1 4 3 0 / 1 0 2 / 0 3
MAICO
Operating Instructions MA 53
14 Warranty, Maintenance and After-Sales Service
The MA 53 audiometer is guaranteed for 1 year. This warranty is extended to the
original purchaser of the instrument by MAICO through the Distributor from
whom it was purchased and covers defects in material and workmanship for a
period of one year from date of delivery of the instrument to the original
purchaser.
The audiometer may be repaired only by your dealer or by a service center
recommended by your dealer. We urgently advise you against attempting to
rectify any faults yourself or commissioning non-experts to do so.
In the event of repair during the guarantee period, please enclose evidence of
purchase with the instrument.
In order to ensure that your instrument works properly the audiometer should be
checked and calibrated at least once a year. This check has to be carried out by
your dealer.
When returning the instrument for repairs it is essential to also send the
headphone, bone conduction receiver and other accessories.
Send the device to your dealer or to a service center authorized by your dealer.
In order to prevent damage in transit, please use the original packing if possible
when returning the instrument.
Geba53_e.03.b.wpd 47 8 5 1 4 3 0 / 1 0 2 / 0 3
MAICO
Operating Instructions MA 53
15 Safety Regulations
15.1 Electric Safety: The MA 53 audiometer is constructed to comply
with protection class II of the international standard
IEC 601-1 (EN 60601-1).
Protection from an electric shock is ensured even
without the system earth connection.
The instruments are not intended for operation in
areas with an explosion hazard.
Geba53_e.03.b.wpd 48 8 5 1 4 3 0 / 1 0 2 / 0 3
MAICO
Operating Instructions MA 53
16. List for the subjective instrument checks
- Clean the cushions of the headphones! Model...........................................................................
- If necessary untangle all cords
- Are the cushions of the headphones all right? If not replace Manufacturer................................................................
them!
- Are the connectors and cables damage-free?
- Can all operator elements be handled smoothly ? Serial No.......................................................................
- Are the operator elements over-tightened or damaged?
- Is the patient response system o.k. ? Test person...................................................................
Purity of test signals
Try all test frequencies at the HL settings of the table and if necessary enter :
“B” for humming, “G” for noises, “V” for signal distortion , “S” for switching noises
30 dB HL
AC
50 dB HL
70 dB HL
30 dB HL
BC
50 dB HL
* in the event that disturbed noises “B”, “G”, “V” or “S” have been checked then SERVICE!
* If conduction test signals are also audible in the non-test ear, then SERVICE!
Level
Target
dBHL
Actual
dBHL
If the difference between desired and actual value at a particular frequency or between right and left headphones is more than 10 dB, then SERVICE!
Tested:.............................................................................................
Date:...............................................................................................
Geba53_e.03.b.wpd 49 8 5 1 4 3 0 / 1 0 2 / 0 3
MAICO
Operating Instructions MA 53
Certificate of Conformity
We,
MAICO Diagnostic GmbH
Salzufer 13/14,
D 10587 Berlin
Germany
Diagnostic Audiometer MA 53
U. Ledworuski
Head of quality department
QW-021/2 04/00
Geba53_e.03.b.wpd 50 8 5 1 4 3 0 / 1 0 2 / 0 3
MAICO
Operating Instructions MA 53
Geba53_e.03.b.wpd 51 8 5 1 4 3 0 / 1 0 2 / 0 3
Front panel with controls and displays
(second level function - go to second level function
by pressing the selected key longer than 2 seconds)
! left (blue) STIM Led Led off = signal off
Led on = signal on
" left (blue) channel level control
# left (blue) STIM bar L R
$ left (blue) channel frequency up $ + % press simultaniously for dual frequency mode
% left (blue) channel frequency down $ + % press simultaniously for dual frequency mode
& left (blue) channel TALK OVER Microphone
' left (blue) channel VU-meter
( left (blue) channel receiver selector key ( select testmode HL, UCL, MCL for selected receiver STIM STIM
) FM - modulation (warble tone) ) press key during switch on to enter the user menu PAT
* left (blue) channel signal selector key * calibration of CD/tape or live voice microphone
+ PULSE tone + pulse tone menu: ABLB, DLI and SISI Test +
FM PULSE SWAP TRACK
+
,
Hz Hz
PATient response display
- STIM MODE - interlock function
. SWAP function . mixing function INSERT NOISE NOISE INSERT
/ TRACKING function / level step selection FF
BC
MIC
TAPE STIM
MIC
TAPE
FF
BC
0
Talk Over AC TONE MODE TONE AC Talk Over
right (red) channel signal selector key
1 right (red) channel receiver selector key 1 adjustment of talkback microphone and monitor phone
2 LCD-Display
3 right (red) channel VU-meter
4 right (red) channel TALK OVER Microphone
5 right (red) channel frequency down MAICO MA 53
6 right (red) channel frequency up
7 right (red) channel STIM bar
8 right (red) channel level control
9 right (red) STIM Led Led off = signal off Led on = signal on 1 3 5 7 9 11 13 15 17 19 21 23 25
: power switch 2 4 6 8 10 12 14 16 18 20 22 24 26
; power socket (100 ... 240 V~ 50/60 Hz) < power fuses
= serial PC-interface > patient response switch R FF L BC MASK BC R INSERT L R AC L
? electrical signal output right channel @ electrical signal output left channel 100 - 240 V / 50 - 60 Hz
A C E G I K M O Q S U
B D F H J L N P R T V