Model Ap/W High Voltage Holiday Detector: Roduct Nstructions
Model Ap/W High Voltage Holiday Detector: Roduct Nstructions
PRODUCT INSTRUCTIONS
MODEL AP/W
HIGH VOLTAGE HOLIDAY DETECTOR
(NEW PRODUCT RELEASE 2015)
ALWAYS CONNECT CABLES PRIOR TO POWERING ON THE INSTRUMENT
The purpose of this instrument manual is to inform the user on the operation of this
particular instrument and is not meant as an instruction guide to holiday detection in
general. It is recommended that only knowledgeable and trained operators use this
equipment. We suggest NACE International (www.nace.org) for this type of training.
* Instrument ships with electrode of choice. Electrode may be Full Circle Spring Electrode (up to 36” pipe diameter),
Half Circle Spring Electrode (up to 8” pipe diameter) or Silicone Rubber Electrode (up to 8” pipe diameter) or
Web: www.tinker-rasor.com E-mail: [email protected]
–1–
CORROSION MITIGATION INSTRUMENTATION
CORROSION MITIGATION INSTRUMENTATION
2828 FM 758, NEW BRAUNFELS, TX 78130 TEL: (830) 253-5621
PRODUCT INSTRUCTIONS
Wire Brush Electrodes (up to 8” width). Larger size electrodes and additional electrodes available at additional
cost.
AND……… ALWAYS READ THE INSTRUCTIONS THROGHLY PRIOR TO USING ANY TEST
EQUIPMENT!
W ARN I NG! This is a HI GH VOLTAGE device capable of producing an electrical shock if not
properly grounded and/ or operated in accordance w ith the instructions and procedures
prescribed in this m anual!
–2–
CORROSION MITIGATION INSTRUMENTATION
CORROSION MITIGATION INSTRUMENTATION
2828 FM 758, NEW BRAUNFELS, TX 78130 TEL: (830) 253-5621
PRODUCT INSTRUCTIONS
electrode. It is therefore recommended that the user does not activate the high voltage within the vicinity
of sensitive electronics or radio equipment.
DO NOT SHORTEN THE GROUND CABLE. NEVER TOUCH THE BARE GROUND WIRE WHEN THE
DETECTOR IS TURNED ON.
WARNING! USE CAUTION WHEN RAINING. If it is raining then there is a safety concern for the operator.
Surface water on the sub straight is usually not conductive enough to transmit voltage, but almost any
contaminant will cause water to become more conductive. If water covers the wand handle while
connected to the APW Power-Pak or the APS Instrument, then the operator could become the return path
for the high voltage. While the output voltage will cause harm to a healthy person, according to OSHA,
your environment must be taken into consideration when operating this type of equipment. For example;
if you are standing on top of a flag pole and receive a jolt of high voltage, it will most likely cause you to
jump, and you will fall off the flag pole. Stay Alert, Stay Aw are and Stay Alive .
If it isn’t raining, but the coating surface is wet, Tinker & Rasor Holiday Detectors are designed to
minimize the effects of damp or wet coatings. In extreme cases, “ghost” holidays might be witnessed,
should a continual path of water reach a holiday. If this continues, dry the sub straight and continue
testing.
I t is im portant to note that all accidents are preventable. Take caution w hen using Tink er &
Rasor holiday detectors or any other high voltage test equipm ent.
3. CHECK-OUT INSTRUCTIONS
• Remove all accessories from the shipping carton.
• Attach the ground cable to instrument. The cable connector is pushed into its
mating connector located on the front of the instrument case, in the lower Left
corner. Insert and twist clockwise to lock.
NOTE: Never connect or disconnect the ground cable or PowerPak when the
instrument is turned on.
• The Model AP/W Holiday Detector is now ready to be set for a specific voltage.
• Remove the dust cover with the supplied screwdriver, to reveal the voltage
adjustment knob.
• Push and HOLD the Power button on the panel of the Instrument case. Then
push either the LOW or HIGH button to set the voltage range.
Web: www.tinker-rasor.com E-mail: [email protected]
–3–
CORROSION MITIGATION INSTRUMENTATION
CORROSION MITIGATION INSTRUMENTATION
2828 FM 758, NEW BRAUNFELS, TX 78130 TEL: (830) 253-5621
PRODUCT INSTRUCTIONS
• Continue holding the Power button on the panel of the AP/W Instrument case
and use the screwdriver to adjust the voltage setting, using the digital display for
reference.
NOTE: When the Power button on the panel of the Instrument case is held,
there is no output from the PowerPak. However, all attachments from the nose
of the PowerPak and beyond should always be considered a potential hazard.
• Once the voltage has been set, as shown by the LCD on the panel of the
Instrument Case, the Power button on the panel of the Instrument case may be
released.
• Attach electrode to the high voltage wand handle assembly and apply to the
structure to be inspected. The electrode should always make intimate contact
with the surface under inspection.
• Turn the instrument ON by holding the safety switch handle of the PowerPak
firmly against the instrument handle. (Instrument handle has neoprene rubber
grip), then pressing the “ON” button on the PowerPak.
NOTE: The instrument will turn OFF when the safety handle is released.
• The instrument will “remember” the last HIGH or LOW and voltage setting
selected, after being turned OFF and then ON again.
4. INTEGRATED VOLTMETER
The Model AP/W Holiday Detector includes an integrated voltmeter displayed on the
front panel of the main instrument.
The LCD of the voltmeter measures and displays the output voltage of the holiday
detector. This display gives the user the ability to tune the AP/W to a specific voltage
within the 800v – 8,000v Low range and the 3,500v – 35,000v High range. Voltages
increment in 100v steps.
Because the Model AP/W includes this integrated peak reading voltmeter, it is not
suggested that an external peak reading voltmeter be used with this instrument. Most
–4–
CORROSION MITIGATION INSTRUMENTATION
2828 FM 758, NEW BRAUNFELS, TX 78130 TEL: (830) 253-5621
PRODUCT INSTRUCTIONS
external peak reading voltmeters currently available are not as accurate as the
integrated voltmeter of the AP/W, and will not show as accurate results.
It is recommended that the Model AP/W follow an annual calibration cycle to ensure
the instrument is in good working order and that the LCD of the integration voltmeter
is accurate.
6. OPERATING METHODS
A good ground return system for both the pipe and the detector will always provide the
best and most reliable inspection. The pipe to be inspected must be grounded from
the bare pipe to earth at some point along the pipe. If individual joints of the pipe are
to be inspected that are not electrically connected, each joint must be grounded.
The speed of the electrode's travel along the pipe should never be excessive, as faulty
inspection may result.
The most accurate and recommended inspection voltage should be obtained from the
coating manufacturer spec or by contacting the coating manufacture or rep. Correct
voltage output for a given thickness of coating has long been a matter of controversy.
However, formulas have been suggested which may be used as a guide for correct
peak voltages on various coating thickness. The calculation is as follows:
V = K √𝐓𝐓
–5–
CORROSION MITIGATION INSTRUMENTATION
2828 FM 758, NEW BRAUNFELS, TX 78130 TEL: (830) 253-5621
PRODUCT INSTRUCTIONS
NOTE: Formula taken from NACE International Standard RP0274-04. This standard and others available for free
download to members at: www.nace.org
A common practice used in setting inspection voltages in the field is to adjust the
output voltage by visual observation. It is the general consensus that a spark
discharge across a gap distance of at least twice the thickness of the coating will give
adequate inspection voltage and compensate for any irregularity in coating thickness
and grounding conditions. If this practice is desired for determining inspection voltage,
it should be done while the electrode is in the normal operating position and under
actual grounding conditions.
The formula and suggestions for setting voltage are supplied as a guide, and are not
necessarily recommended by the manufacturer of this detector.
NOTE: Do not use the AP/W while battery is being charged. Significant
damage can occur to battery and charger.
The Green Power LED on the main panel of the AP/W will blink when the battery is
low. This indicates the battery needs charging.
The battery charger has an LED as well. RED color LED indicates the battery is being
charged. GREEN color LED indicates that battery is fully charged. GREEN color LED
also means that the battery is on trickle charge, and can be left in this condition. The
charger can be used with 110v / 240v AC. It is an auto sensing charger that can be
used throughout the world.
–6–
CORROSION MITIGATION INSTRUMENTATION
2828 FM 758, NEW BRAUNFELS, TX 78130 TEL: (830) 253-5621
PRODUCT INSTRUCTIONS
Older Model AP/W instruments used a different type of battery and charger. See below
for details on Serial Number less than 1333.
Beginning with AP/W serial number 588, the battery charger included with the
instrument has a slide switch that will allow the charger to operate from 115V AC or
230V AC. The factory setting of this slide switch is set for 115V AC, which is correct
for the US. To adjust for 230V AC, slide the switch so that 230V AC Shows.
The new battery charger also now comes with a standard computer –style AC cord
that plugs into the battery charger and the wall outlet. When the charger is being used
outside of the US, a different plug configuration may be required. This type of computer
–style cord should make it easy to find an appropriate AC cord throughout the world.
Earlier Model AP/W units were supplied with either a 110V or 240V charger with a
hardwired cable. Check the packing list or invoice of your order to verify which battery
charger was included with your Model AP/W.
The battery charger has a special plug that is designed to fit into the connector on the
side of the battery, so that polarity of the connection is not an issue.
The battery charger has an LED on its side. This LED will glow steady while charging
the battery. The LED will begin BLINKING when the battery is fully charged.
On earlier Model AP/W instruments (up to s/n 575), to charge the battery, first connect
the PowerPak to the instrument. The PowerPak connector must be connected to the
main instrument to charge the battery. On newer Model AP/W instruments the
PowerPak does not need to be connected to charge the battery.
Connect the charger to the instrument body, using the special connector, fitted into
the charge port on the front and lower Right of the AP/W case. Attach the battery
charger to an AC power source, ensuring the slide switch is in the appropriate AC
supply position: 115v AC or 230V AC (or, on earlier versions, 120v or 240v, depending
on which charger you have. Each charger has an “Inspected By” label which shows
this information.) Leave on charge for 8 to 10 hours between uses. The battery
charger is equipped with a trickle charge feature, so the battery will not be harmed by
leaving the instrument on charge for too long.
–7–
CORROSION MITIGATION INSTRUMENTATION
2828 FM 758, NEW BRAUNFELS, TX 78130 TEL: (830) 253-5621
PRODUCT INSTRUCTIONS
9. INSTRUMENT SERVICING INSTRUCTIONS
A. Cleaning: Keep the instrument clean and dry. Clean the instrument case with
a soft, damp cloth, then wipe dry. Do not use solvents such as lacquer thinner,
methyl ethyl keytone, etc.
KEEP ALL ELECTRICAL CONTACTS CLEAN
B. Voltage Output Checks
a. In case of LOW output voltage:
i. Check the LED indicator for the HIGH or LOW Range selector,
to ensure the instrument is in the correct Range.
ii. Check the display for voltage output. Voltage displayed is in
kilovolts. (22,000volts = 22.0 on display)
iii. Check green ON LED. If flashing, this indicated low battery
voltage.
–8–
CORROSION MITIGATION INSTRUMENTATION
2828 FM 758, NEW BRAUNFELS, TX 78130 TEL: (830) 253-5621
PRODUCT INSTRUCTIONS
SPECIFICATIONS
–9–
CORROSION MITIGATION INSTRUMENTATION
2828 FM 758, NEW BRAUNFELS, TX 78130 TEL: (830) 253-5621
PRODUCT INSTRUCTIONS
OPERATIONAL CHECK
Prior to use, the Model AP/W Holiday Detector should be given an operational check to
ensure it is working properly. The Operational Check ensures that the audible and visual
alerts are working properly and the instrument has voltage output. The Operational Check is
not a calibration check and should not replace the recommendation of annual service.
The instrument should be completely assembled and ready for use in the application in
which it is required.
Operational Check
NOTE: Spark may or may not be seen and heard. When using lower voltage
outputs and on bright sunny days, the spark may not be seen or heard.
– 10 –