Big Lab Analyzer Performance - Designed For Small Workload Labs
Big Lab Analyzer Performance - Designed For Small Workload Labs
With key components such as optics, reagents and calibrators taken from larger COBAS IntegraT and
cobas blood gas systems used in reference labs and hospitals, the cobas c 111 analyzer joins a family
of analyzers designed to deliver consistent results from the smallest to the largest system.
Test Menu Technical Specifications
Test Throughput 65 tests/hour for photometry tests only
(Theoretical Max) 180 tests/hour for ISE tests only
Up to 100 tests/hour for mixed requests
Test panels:
Number of Channels 27 slots (or approximately 14 assays) per reagent disk,
• Basic Metabolic Panel (Reagent Slots) up to 8 disks
• Comprehensive Metabolic Panel 3 channels on ISE module
• Electrolyte Panel
Programmable Maximum 25 photometric with software ver 1.0
• Hepatic Panel Parameters 3 ISE tests, 10 profiles
• Lipid Panel
Sample Types Serum, plasma, urine, whole blood (HbA1c)
• Renal Panel
Sample Input Load capacity: 8 positions, continuous loading
(Sample Disk) STAT: Real STAT interrupt is available
Continuous operation: Additional sample setting during
Individual tests: operation
Sample cooling: Not available
• Albumin • CRPhs* Sample Container Primary tubes: 5 – 10 mL; 16 x 100; 16 x 75;
• ALT • D-dimer* Types 13 x 100; 13 x 75 mm
• ALP • Ethanol* Sample cup: 2.5 mL
• Ammonia* • GGT Micro cup: 1.5 mL
Cups on tube: Cup on top of a 16 x 75/100 mm
• Amylase • Glucose tube
• Amylase, • HbA1c
pancreatic • HDL cholesterol Sample Volume 2.0 – 16.5 µL in 0.25 µL steps
• AST • Homocysteine* Sample Dilution 1.5 – 100 times
• Bicarbonate • Lactate* Sample Clot Detection Not available
• Bilirubin direct • LDH
• Bilirubin total • LDL cholesterol Minimum Sample Primary tubes: 500 µL
Volume Sample cup: 40 µL
• BUN • Lipase Micro cup: 40 µL
• Calcium • Magnesium*
System Interfaces RS 232 serial interface, bi-directional
• Chloride • Microalbumin*
• Cholesterol • Phosphorus Sample Data Base 300 test results
• CK • Potassium QC Analysis QC history provides QC results of previous 60 days
• CK-MB • Sodium Graphic analysis of results over time
• Creatinine Jaffe • Total Protein
Rerun Manual rerun with predefined dilution
• Creatinine plus • Triglycerides
(enzymatic) • Uric acid Electrical Power requirements: 100 – 125V and 200 – 240V
Requirements (-15% to +10%)
• CRP
250 VA (320 VA with ISE)
Frequency: 50 Hz or 60 Hz ± 0.5%
*In development
Water/Waste Minimum: Reagent Grade NCCLS Type II
Requirements Recommended: Reagent Grade NCCLS Type I
Water consumption: 2.5 – 3.0 L/day
Water temperature: 15° – 32° C
Biohazards waste: Separate container
Operating Conditions Ambient temperature: 15° – 32° C
Ambient humidity: 30 – 80% (RH, without
condensation)
For more information, contact your Heat output: Maximum 320 W/hr
or 270 Kcal/hr
Roche Diagnostics representative Noise output: <60 dB
or call 1-800-852-8766 (option 7).
Physical Dimensions Width: 28.4 in (720 mm)
Depth: 21.7 in (550 mm)
Height: 18.9 in (480 mm)
Weight 85.1 lbs (38.6 kg)
COBAS, COBAS INTEGRA and LIFE NEEDS ANSWERS are trademarks of Roche.
© 2007 Roche Diagnostics. All rights reserved. 317-37898-0907
Roche Diagnostics
9115 Hague Road
Indianapolis, IN 46256
http://us.labsystems.roche.com