Not Required: Verification of Equipment (Weekly)

1. The document summarizes the results of a weekly calibration of ultrasonic equipment used to inspect ITER vacuum vessel butt welds. 2. Key tests included verifying horizontal and vertical linearity, sensitivity, signal-to-noise ratio, pulse duration, and squint angle for three inspection techniques using a 2.25DM7X4PM-19X12-A17 probe and SA17-DN55L0-IHC wedge. 3. All tests were within specified tolerances, with minor deviations observed for some sensitivity and signal-to-noise ratio measurements, indicating the equipment is performing as expected.

Uploaded by

asdasdasdasd
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as XLSX, PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
59 views

Not Required: Verification of Equipment (Weekly)

1. The document summarizes the results of a weekly calibration of ultrasonic equipment used to inspect ITER vacuum vessel butt welds. 2. Key tests included verifying horizontal and vertical linearity, sensitivity, signal-to-noise ratio, pulse duration, and squint angle for three inspection techniques using a 2.25DM7X4PM-19X12-A17 probe and SA17-DN55L0-IHC wedge. 3. All tests were within specified tolerances, with minor deviations observed for some sensitivity and signal-to-noise ratio measurements, indicating the equipment is performing as expected.

Uploaded by

asdasdasdasd
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as XLSX, PDF, TXT or read online on Scribd
You are on page 1/ 3

Doc. No. (Rev.

) UCDS(W)_QC-002511_22-06-2021_v01_00
ULTRASONIC Form Rev. 5
CALIBRATION DATA
Date 6/28/2021
SHEET
Page 1 of 3

VERIFICATION OF EQUIPMENT (WEEKLY)


ITER VV butt welds

1. EQUIPMENT
1.1 Instrument
Model No. Manufacturer Serial number Calibration Date
Mainframe OMNISCAN-MX2 Olympus OMNI2-101098 1/26/2021
UT module OMNI-M2-PA32128PR Olympus QC-002511 1/26/2021
1.2 Probes and Wedges
TECHNIQUE T1-T3 Model No. Manufacturer Serial number
Probe 2.25DM7X4PM-19X12-A17 Olympus U4107
Wedge SA17-DN55L0-IHC Olympus ABWX2786
TECHNIQUE T5 Model No. Manufacturer Serial number
Probe 2.25DM7X4PM-19X12-A17 NOT REQUIRED
Olympus
Wedge SA17-DN55L0-IHC Olympus

2. HORIZONTAL LINEARITY (EN 12668-3, Subclause 3.2.1), 0ºL, T AUX


Full Screen Width B0 B1 B2 B3 B4 B5
Expected timebase (mm) B0 B0 + 60 B0 + 120 B0 + 180 B0 + 240 B0 + 300

Measured timebase (mm) 60.00 119.90 179.80 239.90 299.90 360.20


Deviation (mm) NA 0.1 0.2 0.1 0.1 -0.2
Results of Calibration OK
Tolerance: The deviation from linearity shall not exceed ± 2% of full screen width.

3. VERTICAL LINEARITY (ISO 18563-3, Subclause 9.4.2), 60ºL, T1 / H40, SCB-03


Expected Screen Height
Gain, dB Measured Signal Amplitude (%) Limits
(%)
2 101 101.8 Not less than 95%
0 80 80.2 (Reference line)
-6 40 40.1 37% to 43%
-12 20 19.6 17% to 23%
-18 10 9.8 8% to 12%
-24 5 4.9 Visible, below 8%
Results OK
Tolerance: The signal amplitude shall be within the limits.

VERIFIED BY: APPROVED BY:


Name: V. Gagliardi Name: D. Pavoni
Date: 6/28/2021 Date: 6/28/2021
Signature: Signature:
Doc. No. UCDS(W)_QC-002511_22-06-2021_v01_00
ULTRASONIC Rev. 5
CALIBRATION DATA
Date 6/28/2021
SHEET
Page 2 of 3

VERIFICATION OF PROBE 2.25DM7X4PM-19X12-A17 (WEEKLY)


ITER VV butt welds

4. TECHNIQUE T1, 48ºL, H40 (BLOCK SCB-03)


4.1 Sensitivity and Signal-to-Noise (ISO 18563-3, Subclause 9.5.3)
Description Base measurement Measured values Deviation
Sensivity (dB), 20% 24.5 21.6 -2.9
Noise Gain (dB), 20% [15:70] mm 73.5 74.8 1.3
Signal-to-Noise (dB) 49.0 53.2 4.2
Result OK
4.2 Pulse Duration (EN 12668-3, Subclause 3.4.4)
Base measurement Measured values Deviation
Pulse Duration (mm), 10% / 100% 3.1 3.7 0.6
Result OK
4.3 Squint Angle (ISO 18563-3, Subclause 9.9)
Perpendicular Swivelling Deviation
Maximum Amplitude (dB), 80%FSH 33.5 33.5 0.0
Result Conform

5. TECHNIQUE T2, 60ºL, H20 (BLOCK SCB-03)


5.1 Sensitivity and Signal-to-Noise (ISO 18563-3, Subclause 9.5.3)
Description Base measurement Measured values Deviation
Sensivity (dB), 20% 28.0 24.8 -3.2
Noise Gain (dB), 20% [3:35] mm 69.5 73.8 4.3
Signal-to-Noise (dB) 41.5 44.3 2.8
Result OK
5.2 Pulse Duration (EN 12668-3, Subclause 3.4.4)
Base measurement Measured values Deviation
Pulse Duration (mm), 10% / 100% 2.5 2.9 0.4
Result OK
5.3 Squint Angle (ISO 18563-3, Subclause 9.9)
Perpendicular Swivelling Deviation
Maximum Amplitude (dB), 80%FSH 36.8 36.6 -0.2
Result Conform

6. TECHNIQUE T3, 80ºL, N11 (BLOCK SCB-03)


6.1 Sensitivity and Signal-to-Noise (ISO 18563-3, Subclause 9.5.3)
Description Base measurement Measured values Deviation
Sensivity (dB), 20% 33.0 31.1 -1.9
Noise Gain (dB), 20% [1:6] mm 75.5 72 -3.5
Signal-to-Noise (dB) 42.5 40.9 -1.6
Result OK
6.2 Pulse Duration (EN 12668-3, Subclause 3.4.4)
Base measurement Measured values Deviation
Pulse Duration (mm), 10% / 100% 0.6 0.8 0.2
Result OK
6.3 Squint Angle (ISO 18563-3, Subclause 9.9)
Perpendicular Swivelling Deviation
Maximum Amplitude (dB), 80%FSH 43.5 43.5 0.0
Result Conform

Tolerance: The sensivity and signal-to-noise ratio shall be within ± 6 dB of the base measurement.
Tolerance: The pulse duration shall not be greater than 1.5 times the base measurement.
Tolerance: The deviation for the squint angle shall be within ± 1 dB
Doc. No. UCDS(W)_QC-002511_22-06-2021_v01_00
ULTRASONIC Rev. 5
CALIBRATION DATA
Date 6/28/2021
SHEET
Page 3 of 3

VERIFICATION OF PROBE 2.25DM7X4PM-19X12-A17 (WEEKLY)


ITER VV butt welds

7. TECHNIQUE T5, 50ºL, H40 (BLOCK SCB-03)


7.1 Sensitivity and Signal-to-Noise (ISO 18563-3, Subclause 9.5.3)
Description Base measurement Measured values Deviation
Sensivity (dB), 20% 28.0
Noise Gain (dB), 20% [15:70] mm 73.0
Signal-to-Noise (dB) NOT REQUIRED45.0
Result OK
7.2 Pulse Duration (EN 12668-3, Subclause 3.4.4)
Base measurement Measured values Deviation
Pulse Duration (mm), 10% / 100% 4.1
Result OK
7.3 Squint Angle (ISO 18563-3, Subclause 9.9)
Perpendicular Swivelling Deviation
Maximum Amplitude (dB), 80%FSH
Result Conform

Tolerance: The sensivity and signal-to-noise ratio shall be within ± 6 dB of the base measurement.
Tolerance: The pulse duration shall not be greater than 1.5 times the base measurement.
Tolerance: The deviation for the squint angle shall be within ± 1 dB