Calibration and Use of Artificial Mains Networks and Absorbing Clamp
Calibration and Use of Artificial Mains Networks and Absorbing Clamp
1. Introduction 5
-1-
Calibration and use of AMNs and clamps
7. References 59
-2-
Calibration and use of AMNs and clamps
List of figures
Figure 1 - The five commercial units investigated 9
Figure 2 – The equivalent circuit for conducted emissions 10
Figure 3 – AMN/LISN frequency response 11
Figure 4 – AMN/LISN circuit 11
Figure 5 – AMN/LISN impedance (theoretical) showing ground strap resonance 13
Figure 6 – Effects of AMN/LISN distance from vertical reference plane, 30cm ground strap 14
Figure 7 – Change of earth strap length, AMN/LISN 1cm from vertical plane, CM coupling 14
Figure 8 – Change of earth strap length, AMN/LISN 1cm from vertical plane, DM/plate coupling 15
Figure 9 – Effect of AMN/LISN distance from horizontal ground plane, short earth strap 15
Figure 10 – View of AMN/LISN bonding to the horizontal ground reference plane 15
Figure 11 - The effect of raising the front feet 16
Figure 12 – Conducted emissions layout for table-top equipment (CISPR 22) 16
Figure 13 - Difference between lengths of mains cable 18
Figure 14 - Difference in routing of 2.5m bundled mains cable near horizontal plane 18
Figure 15 - Difference in routing of 1m mains cable near vertical plane 18
Figure 16 - Effect of AMN/LISN-to-EUT distance 18
Figure 17 – Comparison of use of earth lead inductor 20
Figure 18 - 10 kHz to 1 MHz differential mode low Z, mains O/C and high Z, mains S/C 22
Figure 19 - 1 MHz to 30 MHz differential mode high Z, and plate 22
Figure 20 - The 3 pin to coaxial adaptor 23
Figure 21 - View of adaptor and setup used for the calibration experiments 23
Figure 22 - Measured adaptor impedance error and insertion loss 24
Figure 23 - Results of impedance measurements 25
Figure 24 - LF impedance measurements versus mains termination 25
Figure 25 - Insertion loss measurement layout 26
Figure 26 - Results of insertion loss measurements 27
Figure 27 - Mains to EUT port isolation 28
Figure 28 - Recommendations for AMN/LISN design 29
Figure 29 – Equivalent circuit for the disturbance power test 30
Figure 30 - Clamp impedance versus frequency 31
Figure 31 - Clamp construction 32
Figure 32 - The three commercial clamps investigated 32
Figure 33 – Standing waves 33
Figure 34 – Side view of calibration set-up 34
Figure 35 - Effect of wire diameter 35
Figure 36 – Centralising the wire within the clamp 35
Figure 37 – Wire position measurements 35
Figure 38 – Effect of loading with ferrites and clamps 36
Figure 39 – Measured output voltage reflection coefficient 37
Figure 40 - Effect of cable position 38
Figure 41 - Effect of absorbing load at far end of mains lead 39
Figure 42 - Differences between open lab area and screened room 39
-3-
Calibration and use of AMNs and clamps
-4-
Calibration and use of AMNs and clamps
1. Introduction
This booklet has been written to advise test engineers and others on the best practice techniques for the use and
calibration of transducers for RF conducted emissions and disturbance power measurements. These
measurements are inherent in many European EMC standards that are based on the CISPR emissions methods,
and form an important part of testing to meet the compliance requirements of the European EMC Directive.
There are two particular aspects of these tests which are of great relevance to equipment manufacturers who
must make a declaration of compliance based on the test results:
• accuracy of measurement: allowing too great a tolerance on the measurement result has the effect of
artificially tightening the emissions limits, which translates into a greater cost for the manufacturer to
ensure that his product emits below the limit level minus the measurement uncertainty. Uncertainties of
measurement should be minimised as far as possible.
• repeatability of measurement: a manufacturer should expect that, if his compliance statement is ever
challenged, he can have a product re-tested at a different test facility to the same standard as the
original compliance test, and achieve the same measurement result. Otherwise, allowing for
uncertainties due to lack of repeatability has the same cost impact as discussed above.
These factors are directly affected by a number of parameters that relate to the test method and the test
equipment. Among them are the calibration and method of use of the transducers. Of these transducers, the most
important are the artificial mains network (AMN) or line impedance stabilising network (LISN) for the
conducted emissions test on the mains lead up to 30MHz, and the absorbing clamp for the disturbance power test
on the mains and other leads above 30MHz. This document therefore sets out best practice in the use and
calibration of these items, with the intent of minimising uncertainties which can be attributed to these particular
aspects.
The project of which this document is the result has drawn on a number of sources of information:
• A literature review to investigate the existing state of the art in use and calibration of these items;
• A questionnaire sent to a number of UKAS accredited EMC test laboratories, asking for their
experience in using and calibrating the items;
• A programme of experimental work using a variety of clamps and AMN/LISNs to determine and
compare the factors which may contribute to their uncertainties.
It should be noted that, to limit the work programme to an acceptable degree, only frequency domain
investigations were carried out. Measurements of noise which is predominantly broadband and pulsed in nature
may well be affected by the time domain parameters of the transducers. These would be a fruitful subject for
future investigation.
This programme of work was carried out by the partners in the project, Schaffner Chase EMC and the National
Physical Laboratory, both of whom are accredited by UKAS to perform these types of calibration.
-5-
Calibration and use of AMNs and clamps
These recommendations result from our investigations as described in the main part of this guide, as well as
other sources that are referenced where relevant. They are presented roughly in order of priority. The comments
are intended to amplify the instructions contained in the various CISPR standards; test procedures and design
aspects that are already commonplace or typical are not discussed.
2.1. AMNs/LISNs
2.1.1. Results
The anticipated contributions to measurement uncertainty for calibration of and testing with the standard CISPR
16-1 AMN/LISN have been systematically investigated. The uncertainties referred to below are expanded
uncertainties for k = 2. The investigation has shown that
• calibration of AMN/LISNs by different organisations can be reproduced to within 2% for impedance
and 0.1dB for insertion loss, provided that the correct method is used and certain straightforward
precautions are taken
• the five commercial units investigated are generally within the CISPR 16-1 specification, with a few
excursions outside the ±20% impedance limit at the frequency extremes
• over the frequency range 25kHz to 15MHz it is possible to achieve test results with an expanded
uncertainty of ±2dB, assuming typical system contributions, with all five AMN/LISNs; the
AMN/LISN itself is not a major contributor to this figure
• the uncertainty degrades slightly below 25kHz, principally because of worsening isolation from the
mains supply impedance variations
• the uncertainty degrades substantially above 15MHz to more than ±3dB, due to several factors
associated with the AMN/LISN's design, construction and use; if a vertical rather than horizontal
ground reference plane is used, with most designs of commercial AMN/LISN this figure may be
doubled
• even greater variations are possible above 15MHz if tight control is not exercised over several aspects
of the test setup and layout; the amplitude of the variations depends on the coupling modes and source
impedance of the equipment under test, and cannot easily be factored into the uncertainty quoted above
The investigation has made it possible to recommend improvements to best practice in testing and calibration,
and certain changes to the design of commercial AMN/LISNs.
2.1.2. Use
• positioning on top of a horizontal ground reference plane (rather than against a vertical ground
reference plane) is preferred (3.2.2)
• use the shortest, most direct wide strap from the earth bonding post to the ground reference plane
(3.2.2)
• do not raise the unit on its feet (if provided) (3.2.2)
• never switch in the earth inductor (if provided), either accidentally or deliberately; if possible, modify
the unit so that it cannot be accidentally selected (3.2.5)
• ensure that the mains (and other) cables never drop to the ground reference plane but are spaced from it
by >10cm or as per the standard requirement (3.2.3)
• prefer standard length mains cables wherever possible (3.2.3)
-6-
Calibration and use of AMNs and clamps
2.1.3. Calibration
• a standard design of test adaptor (as recommended in the appropriate section) should be adopted for the
calibration (3.3.1)
• both the impedance and insertion loss measurements should pay careful attention to the earthing
arrangements to ensure repeatability and reproducibility (3.3.2, 3.3.3)
• the correct way to make the insertion loss calibration is to feed the AMN/LISN and measuring system
in parallel through a Tee adaptor, providing effectively a zero source impedance (3.3.3)
• although impedance and insertion loss calibration results need only be reported at spot frequencies, the
measurements should be swept in frequency to detect any resonances (3.3.3)
• impedance measurements in the range 9kHz up to 25kHz should be made with the mains input port
terminated with open circuit, short circuit and 50Ω; above this frequency an open circuit termination is
sufficient (3.3.4)
• non-mandatory tests can usefully be performed by a calibration laboratory to ensure proper and safe
performance of the unit, including isolation between the mains input port and the EUT port, isolation
between lines, and DC or AC 50 Hz resistance (3.3.4)
2.2.1. Results
The expected uncertainty contributions for calibration of and testing with the CISPR 16-1 absorbing clamp have
been investigated. The results are:
• repeatability of calibration by different organisations of absorbing clamp insertion loss is generally
possible to within ±0.5dB, given control over the method used, and an adequate test site and setup
• the CISPR 16-1 specification allows wide variations in construction, evident in the three units
investigated; nevertheless it is possible to achieve test results within ±1dB with the most commonly
used clamps on the same EUT up to 300MHz, and within ±4dB up to 1GHz
• uncertainties in calibration and testing below 100MHz are dominated by reflections from the far end of
the cable or wire under test, and by the proximity of large conducting objects; as long as testing is not
-7-
Calibration and use of AMNs and clamps
performed within a screened room, an expanded uncertainty for testing of ±2.5dB is achievable with
proper control of the test set-up
• uncertainties in calibration and testing above 300MHz are dominated by variability in wire position
within the clamp, departures of the clamp impedance from specified values and a worsening of the
clamp's output reflection coefficient; with attention to these sources of error, an expanded uncertainty
for testing of around ±2.5dB up to 700MHz is possible, degrading to around ±4dB above this
frequency
The investigation has made it possible to clarify the important parameters in the test and calibration setups and
methods, and to recommend some improvements to best practice in testing and calibration.
2.2.2. Use
• apply a secondary absorber at the end of the cable (6 or more large ferrite clip-on sleeve absorbers are
acceptable in lieu of a second clamp) (4.3.2)
• the cable under test should be kept central within the clamp (4.3.1)
• other objects including personnel should be kept at least 1m away from the setup when the
measurement is made (4.3.4)
• for measurements close to the limit, the clamp output should be taken immediately through a 6dB pad
before connecting to the cable to the measuring instrument; this pad can be left out for initial scans
(4.3.4)
• the output cable should extend away from the set-up at right angles, above or to the side rather than
straight down, and should carry ferrite absorbers (4.3.4)
• the test area should not incorporate a ground plane on the floor (4.3.3)
2.2.3. Calibration
• apply a secondary absorber at the end of the calibration wire (10 or more large ferrite clip-on sleeve
absorbers are acceptable in lieu of a second clamp) (4.2.3)
• the calibration wire should be kept central within the clamp via centralising guide(s), and tensioned to
keep it taut (4.2.2)
• the calibration should be performed in an open area devoid of large metal structures, with no ground
reference plane on the floor, and with no objects including personnel closer than 1.25m to the wire
when a calibration measurement is made (4.2.4)
• the calibration wire should have at least 2mm2 cross section, with 4mm2 cross section preferred if the
standard is changed to allow this (4.2.1)
• the clamp output should be taken immediately through a 6dB pad; the output cable should extend away
from the set-up at right angles, above or to the side rather than straight down, and should carry ferrite
absorbers; this output cable assembly is to be regarded as part of the calibrated equipment (4.2.5, 4.2.6)
-8-
Calibration and use of AMNs and clamps
For the specific purpose of making conducted emissions voltage measurements on the mains input of apparatus,
CISPR 16-1 [3] defines a transducer known as an Artificial Mains Network (AMN). The US term Line
Impedance Stabilisation Network (LISN) is in more general use and usually means the same thing, although
strictly speaking it can be applied to a stabilising network intended for any type of line. The AMN/LISN has
three main purposes:
• to define the RF impedance seen by the EUT’s mains port. Otherwise, tests done on the same EUT
with different mains supplies would not be repeatable.
• to couple the interference signal from the EUT mains terminals to the measuring instrument in a
defined fashion, and to prevent the mains voltage from being directly applied to the measuring
instrument.
• to reduce extraneous ambient noise that might be present on the incoming mains circuit.
Several variants of AMN/LISN are specified in CISPR 16-1, but one in particular (the “50Ω/50µH + 5Ω”
version) has become established as the norm and is widely available from commercial suppliers. Other
impedance stabilising networks are specified for military, aerospace or automotive tests or are defined in draft
specifications for signal or control ports. These are not discussed in this document.
For the investigations which form the basis of this section, five manufacturers’ AMN/LISNs were used and
compared:
• MN2050C, Chase EMC, UK
• L2-16, PMM, Italy
• LISN 1600, Thurlby Thandar Instruments, UK, supplied by Laplace Instruments
• ESH3-Z5, Rohde & Schwarz, Germany
• 4825/2, EMCO, USA
A view of these five units is shown in Figure 1.
A purpose-designed surrogate EUT was used for all comparisons of these five units. The salient features of this
EUT are that it provides a fixed and stable comb output voltage consisting of harmonics of 150kHz or 10kHz up
to and beyond 30MHz; the dominant coupling mode can be selected as differential or common mode directly
coupled, or common mode via a capacitive plate on one side; the direct coupling source impedance can be
-9-
Calibration and use of AMNs and clamps
selected as high (around 100Ω differential, 1nF common mode) or low (around 1Ω differential, 10nF common
mode); the interference source is self-powered, but the unit can accept applied mains voltage or can draw a load
current of up to 20A without affecting its interference characteristics.
ancillary
equipment
equipment under test
AMN/LISN
ZS(sig) ZS(L) mains cable
ZL(sig)
ZS(N)
VS(sig) VS(L) VS(N) LC
CS(sig) CS
ground bonding strap
Any typical EUT will contain multiple noise sources. These can be represented by those which are coupled
directly to live and neutral mains connections, and those which appear in series with other connections.
Referring to Figure 2, the mains-coupled noise signals are VS(L) and VS(N), applied to the live and neutral lines
respectively through ZS(L) and ZS(N), and referenced to the EUT’s chassis or enclosure. The mains lead connects
these signals to the AMN/LISN where they are measured across the 50Ω/50µH impedance with respect to the
ground connection. If the EUT enclosure is metallic and is grounded via a safety earth lead, this refers the
enclosure back to the ground reference plane at the LISN. If the enclosure is non-metallic and there is no safety
earth, the coupling is dominated by stray capacitance CS between the EUT and the ground plane. This stray
capacitance may also form a resonance with the inductance of the earth lead LC if it is present.
For diagnostic purposes it is often easier to divide the sources into differential (symmetric) mode and common
(asymmetric) mode. In this case the common-mode voltage is given by ½·(VS(L) + VS(N)) appearing between both
live – chassis and neutral – chassis, and the differential mode is given by ½·(VS(L) - VS(N)) appearing between live
– neutral. Neither of these voltages are measured directly by the AMN/LISN, although it can be modified to do
so [10]. As well as resonances occurring between the mains cable and stray capacitances, there is also coupling
between the conductors of the mains cable which may become significant at the higher frequencies.
As well as the noise directly coupled to the power leads, noise voltages at VS(sig) may drive current through any
signal lines connected to the EUT. Although the signal line emissions are not measured directly, such current can
flow through the impedances which are common to the power line measurement circuit (CS and/or the safety
earth) and therefore contribute to emissions measured on the power lines. This is why the connection and layout
of any ancillary equipment (represented by CS(sig), LC(sig) and ZL(sig)) can have an impact on the conducted
emissions measured on the mains port.
- 10 -
Calibration and use of AMNs and clamps
100Ω
±20% tolerance
50Ω
Impedance
w.r.t. 50Ω/50µΗ
reference
150kHz
10Ω
9kHz
50Ω/50µΗ + 5Ω
1Ω
10kHz 100kHz 1MHz 10MHz 30MHz
CISPR 16-1 only actually specifies the impedance curve, but suggests how this may be achieved using the circuit
and component values shown in Figure 4. Because of the component values used the standard has become
known as the “50Ω/50µH + 5Ω” AMN/LISN. Two variants are defined; one of these, the “cut-down” version,
has fewer components and its impedance stabilisation is only good above 150kHz. The components to the mains
side of the 50µH inductor are replaced by a single 1µF capacitor, and the coupling capacitor to the receiver is
0.1µF rather than 0.25µF. This is perfectly adequate for measurements to the generic emissions standards or
many other common product standards which do not define tests below 150kHz.
N N
250µH 50µH
L L
4µF 8µF 0.25µF
Equipment
Mains under test
input 10Ω 5Ω E
50Ω
E ground
reference
plane
9kHz high pass filter
50Ω
advisable but not mandatory HPF
receiver
The main impedance determining factors are the test receiver/spectrum analyser impedance, the 50µH inductor
and the 5Ω resistor. A common addition to the basic circuit is a high-pass filter (HPF in Figure 4) with a cut-off
at 9kHz following the 0.25µF coupling capacitor, to attenuate low frequency noise such as switching and 50Hz
harmonics that would otherwise be passed to the receiver input with a consequent risk of overload. This is now a
specific requirement according to clause 2.4.3.1.3 of CISPR 16-2 [4]. When a line is not being measured, it must
be terminated in 50Ω to maintain the correct impedance for the rest of the system; this is normally achieved by
default through line switching in most commercial units, though in one of the units investigated it was necessary
to manually load the line that was not being measured. If this is not done, a substantial fixed error (dependent on
the EUT's source impedance) is introduced.
The remaining components serve to decouple the mains supply, both to reduce feedthrough of mains-borne
interference and to prevent variations in the RF impedance of the mains connection from affecting the
impedance at the EUT port. Construction of the inductors is critical; they must be able to carry the required
- 11 -
Calibration and use of AMNs and clamps
supply current without saturating, yet should be physically small to keep their self-resonant frequency above
30MHz. Air-cored inductors are preferred, especially for higher currents.
Inspection of the circuits of the five different commercial AMN/LISNs shows that the 8µF capacitor and 5Ω
resistor are uniformly implemented. The construction of the chokes is different in each case, with some
manufacturers opting for ferrite-cored chokes in some positions. The 4µF capacitor and 10Ω resistor are not
always present with these values, and the coupling from each line through to the receiver terminal varies widely
from the 0.25µF capacitor recommended. Some manufacturers provide an in-built switchable limiter; all include
a high pass filter, but their implementations differ.
3.2.1. Safety
Two particular safety precautions are necessary when using an AMN/LISN on the mains supply. The first has to
do with personnel safety and is due to the presence of approximately 12µF capacitance within the unit from the
live to the earth terminal. With an applied voltage of 240V 50Hz this allows a current of about 0.9A to flow in
the supply earth conductor. This level of current can easily be lethal if it flows in the body. Even with the smaller
version which has 1.1µF from live to earth, the earth current is too high for safety.
If the AMN/LISN is not properly bonded to the supply earth and it becomes disconnected for any reason then its
case (including, for example, any RF connectors attached to it) will immediately become live. THE AMN/LISN
CASE MUST BE SOLIDLY BONDED TO THE SUPPLY EARTH AND TO THE GROUND PLANE. Best
practice is to permanently install the unit at the test facility. Portable AMN/LISNs for on-site work demand
especial care in installation.
A secondary result of this earth current is that AMN/LISNs cannot be used on mains circuits that are protected
by earth leakage or residual current contact breakers. To allow this, and for best safety practice, use an isolating
transformer in the mains feed to the LISN. This will not affect the performance of the network itself, but may of
course limit the power that can be supplied to the EUT.
The second precaution concerns the safety of the measuring instrumentation.
The supply mains is a fruitful source of transients, which can easily exceed 1kV on occasion. These transients
are attenuated to some extent by the LISN circuitry but it cannot guarantee to keep them all within safe limits.
More importantly, supply switching operations within the EUT itself are likely to generate large transients due to
interruption of current through the LISN chokes and these are fed directly to the measuring instrument without
attenuation.
For this reason it is essential to incorporate a transient limiter at the output of the AMN/LISN when a spectrum
analyser is used (two of the five commercial units investigated included a switchable limiter in the output signal
path). This adds an extra 10 dB loss to the signal, as well as contributing a small factor to the measurement
uncertainty, but this can normally be tolerated and is a much cheaper option than expensive repair bills. Note that
some limiters also incorporate a low-pass filter to restrict the frequency range transmitted.
A limiter is less necessary, though may still be advisable, when a measuring receiver is used since the receiver’s
front end is narrowband and already protected.
3.2.2. The ground reference plane and its connection to the AMN/LISN
The ground reference plane (GRP) is a crucial part of the conducted emissions test set-up. It provides the
reference for the measurement and offers a defined coupling to the EUT. CISPR standards specify a minimum
size for the GRP of 2m x 2m, or that it should be the wall or floor of a screened room. It should extend at least
0.5m beyond the boundaries of the EUT. This ensures adequate electric field coupling.
The GRP should be a conductive material such as aluminium, steel or copper. The thickness is unimportant in
most applications since its prime function is to provide capacitive coupling to the EUT. Skin depth in aluminium
at 150kHz is 0.2mm and therefore for tests above this frequency, extra thickness does not substantially affect the
impedance of the plane. However if the EUT generates strong LF magnetic fields, such as might be caused by
switchmode power magnetic components, these will induce eddy currents in the GRP which could in turn create
- 12 -
Calibration and use of AMNs and clamps
voltage differentials across it, and hence measurement errors, if the impedance of the plane at these frequencies
is too high. For these applications the use of copper or aluminium sheet with a thickness greater than 1mm is
recommended.
The GRP should be electrically unbroken under the EUT and between it and the point of connection to the
AMN/LISN. Where seams occur they should be welded or bonded with fasteners at short intervals. A thin floor-
covering material is acceptable since it will not materially affect the capacitive coupling at distances of 40cm,
but appropriate provision for electrical contact from the plane to the AMN/LISN is necessary.
A short, direct connection from the AMN/LISN to the ground reference plane is vital for repeatability above 10-
15MHz. Experiments show that this aspect is probably the most important in allowing a repeatable test. At
close distances, the case exhibits a capacitance to the GRP according to
where area is the footprint of the LISN in cm2, and distance is the separation of the AMN/LISN from the ground
plane in cm. The ground connection exhibits an inductance according to
where length is the strap length and dia its diameter, both in cm. If these variables are incorporated into the
equivalent circuit shown in Figure 2 then the actual input impedance with respect to the ground plane will show
a resonant peak somewhere in the 10–100MHz region, which in turn will upset the measurement in this region.
(The impedance is plotted in Figure 5 for Cstray = 500pF and Lstrap = 100nH.)
For a typical LISN-to-ground plane capacitance of 500pF the resonance will be adequately above 30MHz, the
top of the measuring frequency range, if the ground strap inductance is less than about 40nH, which requires a
length no more than 5cm. Clause 2.4.4.1 of CISPR 16-2 states that the impedance should preferably be less than
10Ω at 30MHz, and suggests a connection strap with length-to-width ratio of not more than 3:1.
Several experiments were done to attempt to quantify the effect of the ground strap. The issue is complicated by
the difference in construction between commercial AMN/LISNs, and by the alternatives offered by different test
standards, to bond either to a horizontal or a vertical reference plane. Four out of the five commercial units
investigated had ground reference connections provided by a front panel binding post, between 33 and 60mm
from the base of the unit; the fifth had a bare metal strip along its side. The latter has the ability to provide a very
low inductance connection if it is used correctly, but the binding post requires an inductive strap of variable
length depending on the unit. Some units included retractable front feet, which further increased the distance to
the ground reference when extended. If a connection to a vertical (rather than horizontal) reference plane was
required, a ground strap of up to 30cm depending on the unit would be necessary.
The following page shows a number of graphs which illustrate the following aspects:
- 13 -
Calibration and use of AMNs and clamps
• Figure 6 - with a 30cm x 10mm2 cable to the vertical ground reference plane, the differences which
exist when the AMN/LISN is moved away from the vertical plane by 10cm (changing Cstray in Figure
5);
• Figure 7, Figure 8 - the difference between a 30cm and 60cm cable to the vertical plane (changing Lstrap
in Figure 5);
• Figure 9 - with a 7cm x 3cm strap to the horizontal ground plane, the effect of lifting the AMN/LISN
5mm off the horizontal ground plane, as might occur on a covered floor, for example.
These graphs lead to the following conclusions:
• the effect of the ground connection and the shape, size and proximity to the reference plane of the
AMN/LISN are largely negligible below 15MHz, but dramatic above this frequency;
• the greatest effect is found when the EUT coupling is primarily by stray capacitance (“plate” in the
figures);
• the effect is least for differential mode and/or low impedance coupling from the EUT, but not
negligible in either case;
• different AMN/LISNs show substantial variations from each other due to case size and construction
when the ground connection is long, but the variation is almost negligible when the units are solidly
bonded to the horizontal ground plane.
We can make two recommendations resulting from these investigations. One is that, in using present commercial
designs of AMN/LISN, test houses ensure that the unit is solidly and permanently bonded to the horizontal
ground reference plane using the shortest possible right-angled metal plate (Figure 10 shows a photograph of the
optimum construction that was used in the experiments), and that the unit is placed as close as possible to the
plane. The other is that future commercial designs of AMN/LISN should concentrate on providing the facility for
a low inductance ground connection, preferably by using a mounting plate or bracket (see Figure 28), even if this
means a less attractive outward appearance.
20 20
15 15
10 10
Difference dB
Difference dB
5 5
0 0
-5 -5
-10 -10
-15 1 2 3 4 5 -15 1 2 3 4 5
-20 -20
Frequency MHz Frequency MHz
-25 -25
0 5 10 15 20 25 30 0 5 10 15 20 25 30
Figure 6 – Effects of AMN/LISN distance from vertical reference plane, 30cm ground strap
10 10
8 8
6 6
4 4
Difference dB
Difference dB
2 2
0 0
-2 -2
-4 -4
-6 Frequency MHz -6 Frequency MHz
-8 -8
-10 -10
0 5 10 15 20 25 30 0 5 10 15 20 25 30
1 2 3 4 5 1 2 3 4 5
Figure 7 – Change of earth strap length, AMN/LISN 1cm from vertical plane, CM coupling
- 14 -
Calibration and use of AMNs and clamps
10 25
8
6 15
4
Difference dB
Difference dB
2 5
0
-2 -5
-4
-6 Frequency MHz -15 Frequency MHz
-8
-10 -25
0 5 10 15 20 25 30 0 5 10 15 20 25 30
1 2 3 4 5 1 2 3 4 5
Figure 8 – Change of earth strap length, AMN/LISN 1cm from vertical plane, DM/plate coupling
CM HI Z: DIFFERENCE 5 mm - 0 mm CM LOW Z: DIFFERENCE 5 mm - 0 mm
2 2
1 1
Difference dB
Difference dB
0 0
-1 -1
1 2 3 4 5 1 2 3 4 5
Figure 9 – Effect of AMN/LISN distance from horizontal ground plane, short earth strap
A related issue is that several of the commercially available AMN/LISNs, including three of the five that were
investigated, used enclosures that had retractable feet, presumably for aesthetic purposes. When the unit is used
on the horizontal ground plane and these are extended, the front panel of the unit is raised away from the plane
by up to 5cm. The effect of this is quite significant and means that the ground reference plane connection is
longer than it need be. Also, the AMN/LISN's distance from the ground reference plane is uncertain. The graphs
in Figure 11 show the impact this has on the measurement. In the light of this, we must recommend that these
feet are never used in actual testing, and that the construction of commercial units does not include them in
future.
- 15 -
Calibration and use of AMNs and clamps
5 5
Difference dB
Difference dB
0 0
-5 -5
2 3 4 2 3 4
3.2.3. The test layout: lead arrangement and EUT distance from the ground plane
The test layout is most rigorously defined in CISPR 22 Amd 2 [6] (now re-published in the third edition of
CISPR 22 [7]), with several alternative layouts offered for different EUT situations. Other product standards give
layout instructions but are not as prescriptive as CISPR 22; CISPR 16-2 does give fairly prescriptive instructions
(which differ in detail from other standards) but it has no status directly as a product standard and can only be
implemented by reference. The preferred layout for table-top equipment in CISPR 22 Ed. 3 is shown in Figure
12. Note that although it allows AMN/LISNs to be bonded to the vertical plane, our experiments suggest that
with many current commercial designs this is not to be preferred.
Investigations for this project centred on the AMN/LISN to EUT distance, and the length and layout of the mains
lead between the two. [12] points out that the mains cable can present a substantial impedance variation from the
required 50Ω/50µH at its far end; a 2.5m cable has a resonant impedance of 300Ω at around 15MHz, and even
short cables can create impedance discontinuities exceeding the 20% tolerance. The standards dictate a 1m
length cable with excess length bundled non-inductively. Therefore we investigated the difference between a 1m
- 16 -
Calibration and use of AMNs and clamps
cable and a 2.5m bundled cable (Figure 13), with the AMN/LISN closely bonded to the horizontal ground
reference plane. Also, the routing of the mains cable can have a significant effect: if it is run close to the ground
reference plane the stray capacitance increases sufficiently to modify the cable’s transfer characteristic. If the
cable is more than 1m long but not long enough to bundle, the lead can sag and touch the floor. The graphs in
Figure 14 show the difference between the lead touching the ground plane for 40cm of its length, versus being
raised at least 10cm above the plane. Further graphs in Figure 15 show the differences for distances of 10cm and
60cm from the vertical plane.
The standards universally require the AMN/LISN-to-EUT distance to be maintained at 0.8m, but this is not
always easy to ensure, particularly with EUTs of complex shape and with mains leads exiting in awkward places.
We investigated the impact of varying this distance from 0.8m to 0.7m (Figure 16).
From these graphs, the following conclusions can be drawn:
• as with the issue of AMN/LISN connection to and distance from the ground reference plane, the effect
of variations in the mains cable are generally negligible below 15-20MHz, but significant above this
frequency;
• the mains cable length, even when longer leads are non-inductively bundled as per the standard, has a
substantial effect on the result; up to 18dB variation between AMN/LISNs can be found (Figure 13),
although this depends on the mode of coupling within the EUT, with low-impedance common mode
coupling being the worst case;
• routing the mains cable near to the ground reference plane can affect the result by 5-10dB near the
resonant frequencies of the whole assembly; up to 5dB difference in this effect can be found between
different AMN/LISNs, with the method of bonding the unit to the ground reference plane also
affecting the result (in Figure 15, the unit was connected via a 30cm strap);
• the effect of a 10cm variation in the EUT – to – AMN/LISN distance is moderate by comparison to the
other effects, being confined to the region above 25MHz, whether a horizontal or vertical ground
reference plane is used; this aspect should not affect measurement uncertainty seriously if the distance
can be controlled to 1cm as is the case for most set-ups.
A further experiment was carried out to test the effect of resonances on the 1m mains lead by varying its distance
above the horizontal ground plane between 4 and 6cm. Four out of the five AMN/LISNs exhibited a resonance
resulting in differences of up to 4dB between 28 and 29MHz when the mains lead was moved by only 2cm.
Clearly the separation of the mains lead from the ground plane is very important. This aspect is not helped by the
design of most of the AMN/LISNs investigated, with respect to the placement of the 13A EUT socket on the
front panel. These are all mounted with the earth pin at the top. Since the 3 pin plug has the mains lead outlet at
the bottom this means that the mains lead is routed directly towards the horizontal ground plane. It would be
better if the EUT socket on the AMN/LISN is mounted the other way up which would route the mains lead away
from the ground plane and place the earth pin closer to the ground reference plane terminal, which should be as
near to the base of the AMN/LISN as practical. Ideally the EUT socket should be near the top face of the
AMN/LISN. This proposal is incorporated into the recommendations for AMN/LISN design given in 3.4.
A final factor which was investigated was the effect of length and position of the output cable, that is the RF
cable that takes the output of the AMN/LISN to the measuring instrument. On some commercial units this lead
may come into close proximity to the mains lead and there is some possibility of coupling. Experiments showed
that this does occur but that the effects are not substantial. Nevertheless, the disposition of the output cable
should be a factor when recording the test configuration and layout.
- 17 -
Calibration and use of AMNs and clamps
15 15
10 10
Difference dB
Difference dB
5 5
0 0
-5 -5
-10 -10
-15 -15
5 5
Difference dB
0 Difference dB 0
-5 -5
-10 -10
Frequency MHz Frequency MHz
-15 -15
0 5 10 15 20 25 30 0 5 10 15 20 25 30
Figure 14 - Difference in routing of 2.5m bundled mains cable near horizontal plane
5 5
Difference dB
Difference dB
0 0
-5 1 2 3 4 5 -5 1 2 3 4 5
Frequency MHz
Frequency MHz
-10 -10
0 5 10 15 20 25 30 0 5 10 15 20 25 30
CM LOW Z: DIFFERENCE 80 cm - 70 cm, VERT GRP CM LOW Z: DIFFERENCE 80 cm - 70 cm, HORIZ GRP
10 10
5 5
Difference dB
Difference dB
0 0
-5 1 2 3 4 5 -5 1 2 3 4 5
- 18 -
Calibration and use of AMNs and clamps
- 19 -
Calibration and use of AMNs and clamps
measured at the phase connections, with respect to the ground reference plane. None of the current CISPR-based
standards require use of this inductor in the test, although some older national emissions test standards did.
CISPR 16-1 is equivocal regarding the value of this inductor. It may be either 1.6mH or it may be 50Ω in parallel
with 50µH. Two of the AMN/LISNs investigated for this project were found to have the one value, two of them
had the other, and one did not have an earth lead inductor at all. Naturally, when this inductor was switched into
circuit and the EUT was emitting primarily in common mode, different AMN/LISNs gave substantially different
results (see Figure 17). The effect of implementing this inductor is that there must be a small impedance in series
with the EUT earth connection even when it is switched out, due to the finite length of wiring to the front panel
switch, which compromises the unit's impedance curve at high frequencies, even if it remains within the
specification limits. Figure 17 shows the differences between the four units with the switchable inductor, with
the EUT emitting in common mode, high impedance. The two units with the 50Ω/50µH combination showed
essentially the same result, a small increase of around 3dB over most of the frequency range. The units with the
1.6mH choke showed much greater increases of up to 40dB, and were also substantially different from each
other; this difference might be explained by the fact that the unit with the greatest increase also had a second
1.6mH choke in the incoming mains earth line, whereas the other unit did not. Self-resonances in the choke
construction are also noticeable.
The inductor is not now a requirement of any international standards. Since its inclusion compromises the HF
impedance from EUT earth connection to ground reference plane (because of the need for a switch), best
engineering practice would dictate that it is left out of modern designs of AMN/LISNs.
40
1 2 4 5
30
Difference dB
20
10
-10
Frequency MHz
-20
0 5 10 15 20 25 30
- 20 -
Calibration and use of AMNs and clamps
• over-temperature of the internal components, principally the chokes, leading to equipment damage;
• saturation of the chokes at current peaks, leading to inadequate impedance stabilisation and inaccurate
measurements;
• excessive voltage drop through the AMN/LISN, leading to an out-of-specification power supply
voltage at the EUT.
Experiments were done to attempt to discover any effect on the measurement results when the maximum rated
AMN/LISN mains current was passed through the unit, compared to no current. Differences for any particular
unit were found to be a maximum of 1.2dB and these could have been due to other random effects in the test.
Three out of the five units investigated used only air-cored chokes which should be immune to saturation effects.
The fact that the other two also showed no serious effect is encouraging and confirms that the chokes are
properly designed for their purpose.
CISPR 16-1 requires (clause 11.8) that the mains voltage drop across the AMN/LISN should not exceed 5% of
the total applied. All those investigated were able to meet this specification. But note that some EUTs may draw
a current which though its RMS value is within the AMN/LISN’s rating, nevertheless has a high crest factor and
takes a peak current several times its RMS value. Such a waveform might still cause a high voltage drop at the
peaks with a consequent unacceptable reduction in the voltage supplied to the EUT and possible erroneous
measurements.
The DC resistance was not measured as part of this project but is a useful indicator of any connection problems
in the AMN/LISN circuit. The DC resistance between the mains input and the EUT port is specified by the
manufacturer for some of the AMN/LISNs tested while others give the voltage drop, at 50 Hz, for a specified
current. It is recommended that the DC resistance is checked when calibration is performed and also between
calibrations by the user as a safeguard against incorrect operation caused by corrosion of connection points
within the unit.
- 21 -
Calibration and use of AMNs and clamps
COMPARISON OF LISNs: DM LOW Z (MAINS O/C) COMPARISON OF LISNs: DM HIGH Z (MAINS S/C)
4 4
1 1
3 3
2 2
2
3 3
1 1
4 4
0 0
5 5
-1 -1
+ Unc -Unc
-2 -2
- Unc +Unc
-3 -3
Frequency MHz Frequency MHz
-4 -4
0.01 0.1 1 0.01 0.1 1
Figure 18 - 10 kHz to 1 MHz differential mode low Z, mains O/C and high Z, mains S/C
0 4 0 4
-1 5 -1 5
-2 -2
+unc +unc
-3 Frequency MHz -3
-unc Frequency MHz
-unc
-4 -4
0 5 10 15 20 25 30 0 5 10 15 20 25 30
Comments on comparisons
Figure 18 clearly shows that the error between units rises steeply at frequencies below 25 kHz and is almost
certainly due to the lower isolation at these frequencies (see 3.3.4) and the higher reactance of the EUT port,
which makes the voltage developed at the EUT port more dependent on small variations in the impedance
between units.
Figure 19 differential mode shows that despite the possibility of significant errors, particularly above 15 MHz,
there is very good agreement between all the units tested with the errors well within the estimated uncertainty. In
general the differences between AMN/LISNs was the lowest with the EUT operating in differential mode.
The second plot in Figure 19 shows the increase in the difference between AMN/LISNs when the EUT was
coupled via its “plate” circuit, that is, the coupling is mediated entirely by stray capacitance and hence is much
more susceptible to small layout variations, particularly those causing changes in parasitic inductance and
capacitance. This represents the likely worst case and demonstrates the significance that must be attached to
variations between EUTs that are largely unknown.
The calibrations required for an AMN/LISN to ensure compliance with CISPR 16-1 and ANSI C63.4 are as
follows:
• impedance at the EUT port
• insertion loss between the EUT port and the receiver output port
Other calibrations and performance tests can usefully be carried out by a calibration laboratory to ensure proper
and safe performance of the unit, these include:
• isolation between the mains input port and the EUT port
- 22 -
Calibration and use of AMNs and clamps
Figure 21 - View of adaptor and setup used for the calibration experiments
- 23 -
Calibration and use of AMNs and clamps
The adaptor is used for both the loss measurements and the impedance measurements and it is therefore
necessary to assess its effect on these. Since the measuring systems are coaxial it is convenient to measure the
adaptor in combination with a mating one so that the input and output are coaxial, a back-to-back pair. This
arrangement is not ideal because some assumptions have to be made concerning the asymmetry of the two
adaptors. For this project a trailing 3 pin socket was used for the female adaptor, with a similar construction to
the male adaptor.
The results for the impedance and loss measurements for the back-to-back pair up to a frequency of 100 MHz
(Figure 22) indicate, perhaps surprisingly, that the adaptor is not a significant contribution to the uncertainty. For
this reason further work to try to resolve the asymmetry problem was not considered worthwhile.
0.15
Insertion Loss dB
2 Freq Limit
Freq Limit
1
0.1
-1 0.05
-2 Frequency MHz
0
-3 0 10 20 30 40 50 60 70 80 90 100 110
0 10 20 30 40 50 60 70 80 90 100 110 Frequency MHz
3.3.2. Impedance
For this project an impedance analyser was used for the frequency range 9 kHz to 5 MHz and a network analyser
between 5 MHz and 30 MHz [14]. There is a frequency overlap between the two instruments so that a
comparison could be made. The measurement uses the coaxial to 3 pin adaptor to obtain the input impedance at
the EUT port. The required impedance is between the live terminal and the RF ground reference and also
between the neutral line and RF ground reference. The receiver output port is terminated in a precision 50Ω load
(VRC < 0.01), this is referred to as the remote load. Where an AMN/LISN is provided with an internal (local)
50Ω termination the input impedance should be measured using this load as well; this is achieved by switching to
the neutral line when measuring the live line and vice versa.
The RF ground reference is not the same as the EUT earth (as discussed in 3.2.2), and it is essential to ensure
that the coax outer is connected to the AMN/LISN's reference terminal using a low impedance connection. One
of the units investigated is fitted with two 4mm sockets, one each side of the EUT port, specifically for
impedance measurement purposes. For this particular unit an adaptor was manufactured from copper sheet that
connects directly to this earth terminal. For the other units it was necessary to use the normal earth binding posts
provided on the front panel, unfortunately these are at difference distances from the EUT port on different
AMN/LISNs so either separate adaptors need to be manufactured for each type or the earth strap has to be
flexible. Experiments showed that the difference between the two earthing arrangements was less than 0.3% in
the measured impedance.
When making the impedance measurements it is recommended that the earthing arrangements are checked to
ensure repeatability and reproducibility. An indication that this has been achieved is that the measurements are
stable when connecting leads are moved and hand contact has no effect. Also, the unit being calibrated should be
located at least 1m away from the network analyser, and should not sit near or on top of it.
- 24 -
Calibration and use of AMNs and clamps
Impedance Ohms
5 5
LOWER LIMIT LOWER LIMIT
30 30
UPPER LIMIT
UPPER LIMIT
20 20
10 10
IMPEDANCE OHMS
IMPEDANCE OHMS
UPPER LIMIT
12 12
8 8
4 4
The results obtained for all the AMN/LISNs between 9 kHz and 30 MHz are shown in Figure 23 for the live line
with the mains input open circuit; the results for the neutral lines were very similar for all units. Also plotted are
the limit lines as given in CISPR 16-1: 1993. Figure 24 shows the frequency range 9 kHz to 50 kHz for two of
the units with the mains input line-to-earth open circuit, short circuit and 50Ω.
Most of the measured values fall within the CISPR limits, with some exceptions, particularly at 9 kHz as can be
seen in Figure 24. All units implement the values for CISPR C2, L2 and R2 differently and this is likely to
explain the variations observed at the bottom end, with respect to changes in mains termination (see also 3.3.4).
Simulating the CISPR circuit at these frequencies suggests that C2 and R2 could best be fixed at 2µF and 5Ω.
Adding a 1.6mH inductor in series with the mains input earth pin, as is done by units #2 and #5 (not shown in
the above figures) eliminates variations of impedance with mains termination, but it is difficult to recommend
this as a universal practice since it means that the incoming mains supply earth must float at RF with respect to
the ground reference plane.
The unusually high value in Figure 23 for unit #5 with a local load is traceable directly to this unit's use of a 56Ω
resistor as the load, rather than the 51Ω or 50Ω values used by the other manufacturers. (Unit #4 does not
provide an internal local load and so is omitted from the relevant graph.)
In the case of unit #4 with a remote load at 24 MHz the measured value is right on the edge of the CISPR
specification. This unit is the only one of those investigated whose inductors (CISPR L1 and L2) are arranged
coaxially rather than side-by-side; also, in common with unit #2, which shows similar but less extreme
behaviour, its L1 uses a single solenoidal construction with no resistive damping. The other units all implement
some form of damping on L1. Even if the measured value is within the CISPR specification, when the
calibration measurement uncertainty is added to this value the limit could be exceeded. Under these
circumstances the calibration laboratory could not declare with certainty that the AMN/LISN meets the
impedance requirements given in CISPR 16-1, leaving the user in some doubt about the validity of using the unit
for compliance testing. This is a difficult problem to resolve, but as shown earlier in this guide there are
significantly larger contributions to measurement uncertainty, especially at high frequencies. These departures
from the impedance specification should be of little significance for the test results, causing a maximum
deviation of 1.6 dB from the correct measurement for EUTs with high source impedance, reducing for lower
source impedances. The question remains, though, to what extent any departure is acceptable in declaring
calibration results.
- 25 -
Calibration and use of AMNs and clamps
coax-to-3-pin adapter
Output
AMN/
Attenuation EUT
LISN
measuring 10dB attenuators port
system
coax-to-3-pin adapter
Output
AMN/
Attenuation T-adapter EUT
port LISN
measuring
system 50Ω load
∗
According to ANSI C63.4: 1992 [8] and a proposed revision to CISPR 16-1, CISPR/A/201/CDV [9].
- 26 -
Calibration and use of AMNs and clamps
5
1
4 2
Insertion Loss dB
3 4
5
2
0
0.001 0.01 0.1 1 10 100
Frequency MHz
The results of the insertion loss measurements for the live line are shown in Figure 26, the neutral line gave very
similar results. This shows that all but one of the AMN/LISNs have negligible loss between 100 kHz and 5 MHz
(< 0.2 dB) with the exception being approximately 0.5 dB over this range (the circuit of this unit shows a small
series resistance in the signal path, absent from the other units). The insertion loss increases for all units below
100 kHz generally rising to a maximum at 9 kHz, the variation between units at this lowest frequency is 0.5 dB
to 6 dB. This variation can be traced to the use of different values and circuits for C3, the coupling capacitor in
the CISPR circuit. There is also an increase above 25 MHz for most of the units with the maximum being 1.4 dB
at 30 MHz.
3.3.4. Isolation
The output impedance of the AMN/LISN as measured at the EUT port will depend to some extent on the
impedance terminating the mains connection. For most of the impedance measurements the mains input was
open circuit but the effect was investigated by making the measurements with the mains connection terminated
with a short circuit and with 50Ω. It was found that for all units tested the mains impedance has a negligible
effect above 30 kHz but below this frequency there is a significant variation on some models, and so we
recommend that impedance measurements in this range are made with the mains input terminated with open
circuit, short circuit and 50Ω. Note that ANSI C63.4 [8] requires AMN/LISN calibration with a mains RF filter
in place if this will be used in the test environment; the proposed revision to CISPR 16-1 [9] requires the
sensitivity of insertion loss to the mains terminating impedance to be checked across the whole frequency range,
and if this exceeds 0.1 dB, then each line of the mains port must be terminated with 50Ω.
The isolation that the LISN provides between the mains input and the EUT port is not specified in CISPR 16-1
but was investigated for this project. The measurements of isolation were made in a 50Ω system and the results
are given in Figure 27. This shows that the isolation is greater than 25 dB over most of the frequency range
although one model gave results below 20 dB at 20 MHz. All units have reducing isolation below 100 kHz
falling to between 13 dB and 19 dB at 9 kHz. When this measurement is made with a zero impedance source
instead of 50Ω the figure for isolation is in the order of 8 dB at 9 kHz. The actual value applicable to a test
situation will depend on the source impedance of the mains supply. The isolation between the mains input and
receiver port was also measured and generally gave higher values than that to the EUT port.
The lower isolation at frequencies below 100 kHz could have some effect on the test results if the mains supply
carries LF noise and is not sufficiently filtered. See also Figure 18 and associated discussion.
Isolation between lines also is not a specified requirement but was investigated for this project. The isolation
between live and neutral was measured with the input applied to the live line of the EUT port from a zero
impedance source and measuring the difference in insertion loss at the receiver port when the input was switched
to the neutral line. With one exception the isolation between lines for all units was greater than 30 dB over the
full frequency range, the one exception decreased to 25 dB at 30 MHz. The use of the calibration adaptor should
not noticeably degrade line-to-line isolation.
- 27 -
Calibration and use of AMNs and clamps
3
60
4
50 5
Isolation dB
40
30
20
10
Frequency MHz
0
0.001 0.01 0.1 1 10 100
Resulting from the experiments described in this document, we can offer some guidelines regarding the actual
construction of a standard AMN/LISN. These were summarised earlier but are amplified here, with a drawing
(Figure 28) for clarification.
• the enclosure construction should emphasize a low impedance bond to the ground reference plane. The
design should concentrate on providing the facility for low inductance ground connections to which the
EUT earth connection is tightly coupled, preferably by using a mounting plate, even if this means a
less attractive outward appearance. Connections both to a horizontal and a vertical plane should be
envisaged.
• there should be minimum spacing between the reference earth connection(s) and the bottom of the unit;
with no extendable feet. This allows the lowest possible impedance to be realised by the connecting
strap, as above.
• the EUT connector should be mounted near the top of the front panel, upside down (for BS1363 13A
types) to encourage the mains lead to exit away from the ground reference plane and to give a short
connection from the earth pin to the reference earth connection(s). When the mains lead exits towards
the ground plane, it is harder to maintain the necessary separation.
• a reference earth connection should be provided close to the EUT port's live and neutral pins that is
specifically used for calibration purposes.
• no earth lead inductor should be included. It is not a requirement of current measurement standards,
and its inclusion compromises the impedance between the EUT earth and the reference earth
connection(s); the value is not properly defined in CISPR 16-1, and this results in different
implementations between AMN/LISN manufacturers. Providing a switch to control the earth coupling
allows an extra source of accidental error in the test procedure.
• provision of locally switched 50Ω loads is preferable to external loads, as long as the load value is
correct. Requiring an external load to be fitted to the unused receiver port(s) allows an extra source of
accidental error in the test procedure. Internal loads appear to give a better-behaved impedance curve at
high frequencies.
• the CISPR 50µH inductor L1 should be resistively damped if it is a single solenoidal winding, and
should not be coaxial with L2; this appears to minimise resonances in the high frequency impedance
response.
• care should be taken over the selection of component values for CISPR C2, L2 and R2, in order to
ensure that the impedance curve at 9kHz is maintained whatever the terminating impedance of the
mains input port.
- 28 -
Calibration and use of AMNs and clamps
- 29 -
Calibration and use of AMNs and clamps
For the investigations which form the basis of this section, three manufacturers’ absorbing clamps were used:
• Luthi MDS-21, manufactured in Switzerland by Schwarzbeck
• Fischer F-201, manufactured in the USA by Fischer Custom Communications Inc
• Laplace RF-400, manufactured in the UK by Laplace Instruments Ltd
The principal standards which require disturbance power tests are EN 55013 and EN 55014-1, derived from
CISPR 13 [1] and CISPR 14-1 [2] respectively. Other standards require the use of the clamp for other purposes,
notably for cable screening effectiveness tests, but these are not considered here.
VS
ZC ZT
CS
The essential assumption of the disturbance power measurement method is that the primary source of high
frequency emissions is the signal that is coupled onto a connected cable from interference generated within the
EUT. Then, an indication of the disturbing capability of the EUT is obtained by measuring the current that is fed
in common-mode onto this cable, and which will be coupled away from the cable by radiation. The assumption
is valid for many types of apparatus that have only one cable (typically the mains cable) and whose dimensions
are small compared to a wavelength. When the equipment has several connected cables, or when its dimensions
are greater than a quarter wavelength (that is, it is physically large, and/or the frequency of interest is high) the
assumption is harder to justify and a radiated test is preferred. Nevertheless this measurement method has a good
record as a means of assessing the interference potential of many types of equipment, and it is generally simpler
to carry out than the radiated method.
The equivalent circuit for high frequency emissions in the standard absorbing clamp test set-up is shown in
Figure 29. A length (normally >5m) of the cable under test is stretched out at a constant height above ground and
thus forms a transmission line of characteristic impedance Z0, a value which is determined by the geometry of
the set-up. The far end of this cable is terminated in an impedance ZT, which is generally unknown, being fixed
by the high frequency impedance of the mains supply or the ancillary equipment which may be connected.
(Placing a second absorber at this end stabilises ZT, as discussed later.) The impedance presented to the EUT is
ZC, which is the transmission line impedance Z0 modified by ZT and any other mismatches that are present on the
line.
The disturbing source in any EUT can be represented by a voltage source VS in series with an impedance ZS,
referred to the enclosure (this simplification is valid whether or not the equipment has a Class II insulated or
Class I earthed metal enclosure). Of course, both VS and ZS are frequency dependent. ZS is coupled in common
- 30 -
Calibration and use of AMNs and clamps
mode to the cable under test – that is, the resulting current ICM flows equally and in the same sense in all
conductors in the cable. The mechanism of coupling depends on the design detail and may be complicated, but
need not concern this discussion. The EUT is isolated from the ground but coupled to it via stray capacitance CS,
which is a function of the equipment’s geometry and its distance from the floor and other surrounding objects.
Matching considerations dictate that the maximum power transfer from the EUT to the cable will occur when
Re(ZC) = Re(ZS + 1/jωCS)
ZS and CS are fixed by the EUT and the set-up, so power matching can be achieved by varying ZC. This is the
purpose of applying the clamp to the cable and moving it through a half wavelength, as discussed in the next
section.
250 #1 R
Resistance (ohms), reactance (%)
250 #1 X%
#1 X%
#2 R
#2 R
200 #2 X% 200
#2 X%
#3 R
#3 R
150
#3 X% 150
#3 X%
100 100
50 50
0 0
10 Frequency MHz 100 1000 10 Frequency MHz 100 1000
The ferrite rings are halved, and the clamp is so hinged that the cable under test can be laid through the whole
assembly and the other halves brought down to enclose it completely. Good mating between the halves is
essential and the ferrite housings are sprung to ensure this. Typically, the ferrites are mounted in the body of the
clamp which is made of insulating material; there are no galvanic connections to be made either to the cable
under test or to the transducer output cable. A typical construction is shown in Figure 31 and photographs of
those used for the comparative measurements are shown in Figure 32. CISPR 16-1 shows two alternative
- 31 -
Calibration and use of AMNs and clamps
construction possibilities, but in practice all those commercially available use variants of the form shown in Fig
39 of the standard.
lead to be
measured
In use, the clamp is placed round the cable under test with the current probe facing the emitting source. The far
end of the cable is connected to the mains supply or to ancillary equipment, or left open in calibration; a second
absorber may be placed over the lead. The clamp output is connected to the measuring receiver or spectrum
analyser.
To determine the maximum power that the disturbing source can deliver to a lead of indeterminate length, the
clamp is moved along the cable under test to find a maximum at any given frequency [15]. The output level
varies periodically with distance, the distance between peaks being around half the wavelength of the measured
frequency. The horizontal cable is acting against the ground as a transmission line, on which current standing
waves are induced, and the clamp is measuring these standing waves (Figure 33). For a standing wave to form,
there must be a mismatch along the transmission line. Such a mismatch will be caused by either of two sources:
• the termination (or lack of it) at the end of the line, or
• the absorbing clamp itself.
- 32 -
Calibration and use of AMNs and clamps
Zline
reflected wave Zline
Variation in |Zline| and ∠Zline as clamp is moved, allows maximum power transfer impedance to be found
Zline
Zline combined standing wave
If the absorbing section of the clamp is fully effective, the end of line termination should be invisible to the line
between the clamp and the disturbing source, and should not therefore contribute to the standing wave measured
by the current transformer. Nevertheless, the absorber itself will be mis-matched with respect to the transmission
line impedance and will therefore cause a reflection, which in turn creates a standing wave. The current
transformer is fixed in relation to the absorber and therefore remains at the same point on the standing wave, yet
still senses a periodic variation in output; this is because as the clamp is moved, the amplitude of the signal fed to
the cable varies, because the impedance presented to the disturbing source is varied. Therefore, moving the
clamp will present an envelope of load impedances to the source which will allow, within limits, the maximum
power transfer to be achieved.
At frequencies where the absorbing section is less effective (typically, below 50-100MHz) the far-end
termination mis-match becomes visible to the measurement. The probe output still varies as the clamp is moved,
and the maximum position can still be found; thus this condition is indistinguishable from correct operation at
the higher frequencies. But the pattern on the cable is now partly due to the absorber mis-match as described
above, and partly to the stationary standing wave caused by the far-end reflection. The maxima are thus not
entirely due to the optimisation of power transfer. This is why typical insertion loss curves show patterns of
peaks and nulls at the lower frequencies as well as the higher. It is also why placing a secondary absorber at the
far end of the cable under test (or the calibration wire) will give different and more repeatable results at the lower
frequencies.
The calibration method is specified in CISPR 16-1 Annex H and Fig. 40. This section reviews sources of error
that may be found in this method. Figure 34 shows the basic set-up. As well as the experiments carried out
especially for this report, reference has also been made to a CISPR/A document [16] proposing improvements in
the absorbing clamp calibration method. At the time of writing, CISPR are organising a round-robin comparison
programme for clamp calibration, and the comments made here are aligned as far as possible with the methods
proposed in that programme.
- 33 -
Calibration and use of AMNs and clamps
reference
plane
4m
secondary
ferrite absorber
output cable
> 0.8m assembly
signal generator measuring
under
calibration ferrite receiver
ground
Alternative setup with cable above clamp (recommendations of new draft method in bold)
The calibration procedure is to compare the clamp's insertion loss with the reference loss through the 10dB and
6dB attenuators, i.e. points A and B (in red, above) are joined and the system loss is calibrated at spot
frequencies in the range 30 - 1000MHz. The clamp set-up is then reconnected and at each frequency, the
minimum insertion loss de is determined while the clamp is moved through about 3m (from next to the reference
plane). During calibration personnel should be further than 1.25m from the cable. The calibration factor is then
derived from the following formula:
K [dB(pW/µV)] = de[dB] - 17 [dB(pW/µV)]
The factor 17dB converts between microvolts and picowatts in a 50Ω system. Strictly speaking, the calibration
factor applies to the clamp only when it is connected via a 6dB attenuator and the output cable used in this set-
up. If no attenuator, or a different cable is used, the calibration factor is likely to diverge.
- 34 -
Calibration and use of AMNs and clamps
22
2.5 mm^2
1.0 mm^2
INSERTION LOSS dB
20
18
16
14
10 100 1000
FREQUENCY MHz
clam p bo dy ferrites
existing
guide
central
to ± 1m m
external
ce ntralising
guide ne eded
at either en d
The cable under test should be maintained central within the clamp. However, when calibrating the clamp with a
much thinner wire there can be significant variation in the position of the wire relative to the centre of the ferrites
in the clamp. Offsetting the wire up to 10mm (where the construction of the clamp allowed this) showed that the
difference could be up to 5dB with one model of clamp, the effect being greatest at the higher frequency end.
During calibration, the effect can be minimised by using additional guides within the clamp at either end that
will hold the wire central to within ±1mm (see Figure 36). The wire should be maintained taut along its length.
CLAMP #3 - EFFECT OF WIRE POSITION IN CLAMP CLAMP #3 - DIFFERENTIAL EFFECT OF WIRE POSITION
8
23
7
21
WIRE CENTRAL - 5 mm OFFSET
6
INSERTION LOSS dB
DIFFERENCE dB
19
WIRE CENTRAL - 10 mm OFFSET
5
17
4
15
3
WIRE CENTRAL
13 2
WIRE OFFSET 5 mm
11 WIRE OFFSET 10 mm 1
9 0
10 100 1000 10 100 1000
- 35 -
Calibration and use of AMNs and clamps
WITH 1 CLAMP 4
22 Clamp #3
WITH 2 CLAMPS 3.5
INSERTION LOSS dB
Clamp #1
FERRITES 3
20
2.5
2
18
1.5
1
16
0.5
0
14 0 2 4 6 8 10 12
10 100 1000 NUMBER OF FERRITES
FREQUENCY MHz
- 36 -
Calibration and use of AMNs and clamps
dropping to the floor directly as it exits from the clamp. The length of this cable, and the impedance of the ferrite
sleeves, will both affect the transmission line impedance of the wire. This effect will become more marked the
more closely coupled the output cable is to the wire, although it is lessened by occurring on the downstream end
of the clamp. Minimum coupling will occur when the cable is immediately brought away from the wire at right
angles to it, and this geometry should always be maintained in calibration (see Figure 34).
Since the output cable and its ferrites can have an effect on the result, the clamp and its cable assembly should
always be treated as a unit and calibrated and used together.
0.8
VOLTAGE REFLECTION
0.7
COEFFICIENT
0.6
0.5
0.4 2
0.3 1
0.2
3
0.1
0
10 100 1000
FREQUENCY MHz
Many of the issues discussed under this heading are similar to those which appear under calibration of the
clamp. For the investigations described here, a comb generator was used as a surrogate EUT which provided a
measurable output at 10 MHz intervals between 30 MHz and 1 GHz. The centre conductor of the comb
generator's coaxial output was connected directly to two wires of a three core mains lead with the screen
connected to the third wire; this would give part differential, part common mode excitation as might be expected
in a real EUT. The output of the absorbing clamp was connected, via a 6 dB attenuator to a spectrum analyser
which was used to measure the level of each 10 MHz harmonic over the 30 MHz to 1 GHz range.
The mains lead was 6 m long with the far end open circuit. 10 clip on ferrites were placed 4.5 m from the EUT,
and the output cable had 3 clip on ferrites close to the clamp output and a further 6 spaced along its length (5m).
The output cable was brought away from the clamp perpendicular to the mains lead and was suspended at least
0.4 m above the floor. The clamps were moved along the wire by means of a pulley arrangement to avoid
personnel contact.
- 37 -
Calibration and use of AMNs and clamps
4 1 2 3
2
Difference dB
-1
-2
Frequency MHz
-3
10 100 1000
- 38 -
Calibration and use of AMNs and clamps
1 1
2 2
2 2
Ferrites - No Ferrites dB
1 1 3
Ferrites - Clamp dB
3
0 0
-1 -1
-2 -2
1
Difference dB
1
0
0
-1
-1
-2
-2
-3
-3
1 2 3 -4 1 2 3
-4 Frequency MHz
Frequency MHz
-5
-5
10 100 1000
10 100 1000
The results for this particular laboratory and screened room are probably typical but smaller screened rooms may
well show greater differences. In general when measurements have to be made in a screened room, because of
ambient signals, then the uncertainty needs to be increased by at least 5 dB at frequencies up to 100 MHz and
1 dB above this frequency. If a result obtained in a screened room is within ± 5 dB of the limit at frequencies up
to 100 MHz then this measurement should be repeated outside the screened room, if ambient signals levels are
acceptable.
- 39 -
Calibration and use of AMNs and clamps
HEIGHT ABOVE FLOOR (IN LAB) HEIGHT ABOVE FLOOR (IN SCREENED ROOM)
5
5
4 1 2 3
4 1 2 3
3
3
2
2
1
1
0
0
-1
-1
-2
-2
-3
-3
-4 Frequency MHz
-4 Frequency MHz
-5
-5
10 100 1000
10 100 1000
The output cable appears as a conducting structure coupled with the transmission line impedance of the cable
under test. The routing of this cable away from the clamp might be expected to affect the results. Measurements
were made in the lab and in the screened room with the output cable suspended at least 0.4 m above the floor and
with it lying on the floor. In each case the cable was perpendicular to the mains lead as soon as it cleared the 6dB
attenuator.
The results show that in the laboratory the difference between suspending the cable and having it on the floor is
insignificant. In the screened room the maximum difference is approximately 0.8 dB. For measurements in a
laboratory it appears that if it is more convenient to allow the output cable to rest on the floor then this is
acceptable. In a screened room, it is preferable that arrangements should be made to suspend the cable above the
floor by at least 0.4 m.
As quoted above, EN 55014-1 requires metal objects to be more than 0.4m from the EUT but says nothing
further about the set-up. EN 55013 extends this distance to 0.8m and includes personnel. In order to check these
effects, measurements were made with all three clamps, first using the pulley and then pushing the clamps by
hand; and then with clamp #2 only to determine the effect of metal objects near the mains lead. A 1 m x 2 m
metal sheet was placed parallel to the wire with the long dimension vertical and approximately 1 m from the
EUT. The plate was progressively moved closer to the mains lead in steps of 0.2 m from 1.2 m to 0.2 m. Figure
44 shows the result of this experiment. The resonance introduced by this plate at around 65MHz is clearly
apparent. The maximum difference rises to 2.7 dB at 0.2 m distance. The second plot shows the variation of
difference with respect to distance from the mains lead and indicates that maximum difference at 1.2 m is less
than 0.4 dB.
- 40 -
Calibration and use of AMNs and clamps
0-.2
2
0-.4
Maximum Difference dB
1 2
0-.6
Difference dB
0
0-.8
-1 0-1 1
-2 0-1.2
-3 0
10 100 1000 0 0.5 1 1.5
The size of the metal object has a distinct bearing on the effect and therefore the safe distance from the wire. The
distances specified in the standards should be regarded as the limit for relatively small objects, probably no
larger than 0.4 m3. Larger objects should be further away and very large objects such as the wall of a screened
room at least 3 m away for the effects to be less than 0.5 dB.
Comparing movement of the clamp by hand and by a rope and pulley arrangement, the difference for clamps #1
and #2 were less than 0.5 dB while for clamp #3 it was 1.2 dB at 30 MHz. When hand pushing the clamps
attempts were made to keep the body as far away from the front of the clamp and the mains lead as possible. EN
55013: 1990 says that "No metallic object, including a possible other unit of the equipment under test, or any
person, shall be closer to the lead or unit than 0.8 m". This is difficult to achieve unless there is an arrangement
that moves the clamp without personnel touching it.
We can recommend that using a pulley arrangement is preferable but if hand pushing is used the following
precautions are taken: where a signal is within 3 dB of the limit the clamp should be positioned on the peak by
hand but the final measurement is made when the person is well away from the clamp and lead. The presence of
a person will probably not alter the position of the maximum but it may affect the level. If hand contact is
necessary, it should be made at the end of the clamp away from the current probe.
4.3.5. Differences between clamps, and use of the manufacturer’s calibration figures
Test houses will normally expect to use the manufacturers' calibration figures in default of those provided by an
accredited calibration laboratory. The figures provided in the documentation accompanying those clamps that
were investigated for this project, varied substantially in their quality. Clamp #3 supplied no more than "typical"
values and a plot. Clamp #1 was fully calibrated by the supplier with a certificate and quoted accuracy of
±1.5dB, while #2 came with a correction factor curve that was not explicitly traceable to the serial number of the
unit. None of the three units' documentation quoted a calibration method.
Figure 45 shows the manufacturer's figures versus the actual insertion loss for each of the units, measured in an
open environment following the best practice calibration techniques presented in this guide. It can be seen that
there is a quite severe difference in some instances, approaching 6dB in the worst case, though all units can show
a divergence of 2dB at some frequencies.
In Figure 46 measurements made with the three clamps are compared in optimum conditions, that is:
• the EUT was measured in a laboratory environment
• the EUT and its mains cable were 0.8 m above a non conducting floor
• 10 clip on ferrites of the type described in 4.2.3 were placed on the mains lead 5.5m from the EUT
• a 6 dB attenuator was connected to the output of the clamp
• the output cable was supported at least 0.6 m above the floor and perpendicular to the mains cable
The plots show the differences between each clamp with respect to each other. The uncertainty limits shown are
based on the calculations given in 5.3.2 with the following changes:
• the uncertainty due to the receiver is not included
- 41 -
Calibration and use of AMNs and clamps
• the mismatch uncertainty is calculated using a receiver input reflection coefficient of 0.05 (at the input
to the 6 dB attenuator)
• the remaining errors are substantially independent for each clamp and because the plots show the
difference between each clamp, rather than the difference from the mean, the uncertainty has been
multiplied by 2.
1 actual
2 actual
INSERTION LOSS dB
20 3 actual
1 manuf.
2 manuf.
3 manuf.
15
10
10 100 1000
FREQUENCY MHz
6 3 -1
4
3-2
Difference dB
0 + Unc
-2
- Unc
-4
-6
-8 Frequency MHz
-10
0 100 200 300 400 500 600 700 800 900 1000 1100
Comments
Absorbing clamps #1 and #2 agree well over most of the frequency range but start to differ above 700 MHz,
possibly due to the position of the mains lead in the clamps or to the clamp impedance, however, the differences
remain within the estimated uncertainty.
Absorbing clamp #3 gave results that were significantly different from the other two. The difference shows a
ripple that is characteristic of the correction factor and output reflection coefficient for this clamp. No specific
explanation was found for this difference and, as can be seen, the estimated uncertainty does not include a
contribution to allow for it. It can be speculated that since this is the only clamp with a metal case there is a
substantially different coupling with the wire, in the calibration setup, and with the mains lead in the test setup,
that is the cause of the variation for this unit.
- 42 -
Calibration and use of AMNs and clamps
4.3.6. Directivity
Directivity is the difference between the forward insertion loss - that is, with the clamp used in its usual direction
- and the reverse insertion loss, with the current transformer facing away from the source. Since the current
transformer is largely bi-directional, the directivity is determined by the attenuation offered by the ferrite
absorber material, which of course is dependent on frequency.
Checks were made in the calibration set-up of the directivity of each of the clamps investigated. Figure 47 shows
the results. For all clamps the directivity falls to near zero at the lowest frequency; clamp #3 has a greater
directivity at 30MHz (3dB) but is lower than the other two at the higher frequencies. This is likely to explain the
effect noted in 4.2.3, that loading the far end of the line has a slightly less effect on this unit than on the other
two.
1
25
2
20
DIRECTIVITY dB
3
15
10
-5
10 100 1000
FREQUENCY MHz
4.3.7. Comparison with OATS results, and use for pre-compliance check
There has been considerable interest in using the absorbing clamp as a partial or complete replacement for
radiated emissions tests, though to date no proposal to do this has succeeded in CISPR. Direct correlation
between clamp measurements and radiated emissions on an open area test site (OATS) is not possible; reference
[17] shows a graph of the extreme values of such a relationship versus frequency for small EUTs, which exhibits
a 15 dB range. Nevertheless, as also pointed out in [17], the clamp can be used to provide a peak-detected pre-
scan of significant emission frequencies on particular cables, which are then selectively and fully measured on an
OATS. This method can offer a reduced total measurement time compared to a full scan on an OATS.
A further advantage of the clamp is its suitability for diagnostic work on EUTs before they are subjected to a full
compliance test. Emissions present on different cables can be compared to discover the worst EUT port on which
remedial effort can be concentrated. For this kind of work, absolute accuracy is not required and the clamp
position need not be varied, but other precautions outlined in this document are advisable for repeatability,
particularly the use of a second absorber on the far end of the cable under test, and control of the test layout.
- 43 -
Calibration and use of AMNs and clamps
5.1. Introduction
This section gives some guidance on the estimation of uncertainty involved in both calibrating and using
AMN/LISNs and Absorbing Clamps. The uncertainty budgets are intended only as guidance. Actual
contributions to uncertainty are very dependent on the particular equipment used and the layout of systems,
particularly in the case of testing where the EUT can be a significant contributor. Test houses are expected to
calculate and present their own uncertainty budgets, omitting any consideration of the EUT contribution,
although in those cases where the nature of the EUT is known and consistent (such as in manufacturers' own test
facilities) this could usefully be included.
The method of calculating the total uncertainty is consistent with widely used practice as outlined in the ISO
document “Guide to the Expression of Uncertainty in Measurement” [19] and the United Kingdom Accreditation
Service (UKAS) document M3003, “The Expression of Uncertainty and Confidence in Measurement” [18].
The Expanded Uncertainty, U, for a confidence level of approximately 95% is obtained from:
N
U =2 ∑c u ( x )
i =1
2 2
i i
Where
u(xi ) are standard uncertainties for the individual contributions, expressed in dB
ci are the sensitivity coefficients associated with each standard uncertainty.
In all the measurements made during this investigation the effect of contributions to uncertainty have been
assessed for their influence on the overall result of the measurement and therefore all sensitivity coefficients
have been taken as unity.
The standard uncertainties u(xi) are obtained from:
• assessed limit values with assumed rectangular probability distribution where the semi range limit is
divided by the square root of three,
• uncertainties already expressed as expanded uncertainties where the divisor is two,
• standard deviations where the divisor is one, or
• mismatch uncertainty which has a U-shaped distribution and the semi range limit is divided by the
square root of two.
The calculation given above for expanded uncertainty does not take account of any correlation between
individual contributions. In practice there will inevitably be some correlation but given the nature of the
measurements and the reliability of the estimates made for the magnitude of the contributions a more rigorous
treatment of the calculation was not thought to be warranted.
The calculation used for mismatch uncertainty, UM, is as follows:
(
U M = 20 log 1 ± ΓG . ΓL )
where ΓG and ΓL are the Voltage Reflection Coefficients of the source and load respectively.
The expanded uncertainties shown in the following tables are given to two places of decimal, however, in
practice it is rarely necessary to quote more than one place of decimal, particularly for testing uncertainties.
Since they are intended for guidance, the tables include all sources of uncertainty that might be expected in a
typical test or calibration. Some of the factors are related to the experimental investigations reported in this
document and these are highlighted. Other factors are determined by the lab's own equipment or set-up and these
should be derived for each individual case.
- 44 -
Calibration and use of AMNs and clamps
Note that different uncertainty values may be reported for various sub-ranges of the total frequency span of the
measurement or calibration. The sub-ranges chosen for the tables reflect the experience of this investigation.
5.2. AMN/LISN
For calculation of mismatch uncertainty the following values of VRC were used:
- 45 -
Calibration and use of AMNs and clamps
Impedance
Output 50 Ω load
This uncertainty only applies to the measurement of the remote impedance and is due to the error in the load
used with respect to 50 Ω, a VRC between 0.005 at 9 kHz and 0.01 at 30 MHz has been assumed.
Repeatability of measurement
The values given are the standard uncertainty for 5 repeat measurements which includes disconnection and
reconnection of the measuring system.
Insertion Loss
- 46 -
Calibration and use of AMNs and clamps
Repeatability of measurement
The values given are the standard uncertainty for 5 repeat measurements which includes disconnection and
reconnection of the measuring system.
5.2.3. Tests
The uncertainties given in the following tables have been derived from an analysis of the test data obtained using
the surrogate EUT constructed for this project. However, due to the diverse nature of real EUTs and the
complexity of the RF coupling mechanisms, these budgets can only be considered as general guidance for the
possible values of contributing uncertainties.
Two uncertainty examples are given, one for the use of an AMN/LISN earthed to a vertical ground plane and
one with it earthed to a horizontal ground plane.
Horizontal ground plane, 3 cm long low impedance strap, 1 m mains lead at least 10 cm above ground plane
Uncertainty ± dB
Source of Uncertainty Divisor over frequency range:
9 to 25 25 kHz to 15 to 30
kHz 15 MHz MHz
Insertion Loss of LISN 2 0.07 0.14 0.23
Receiver Level Accuracy √3 1.5 1.5 1.5
Coupling between LISN and Ground Plane √3 0.0 0.10 0.5
Coupling between Mains Lead and GP √3 0.0 0.10 2.0
Coupling between EUT and LISN √3 0.0 0.10 0.5
Length of Earth Strap √3 0.0 0.0 0.5
Effect of mains impedance √3 1.0 0.0 0.0
Mismatch LISN to Receiver √2 0.1 0.52 0.12
Repeatability of measurement 1 0.30 0.30 0.30
Combined Standard Uncertainty 1.09 0.97 1.56
Expanded Uncertainty 2.17 1.94 3.13
For calculation of mismatch uncertainty the following values of VRC were used:
- 47 -
Calibration and use of AMNs and clamps
Vertical ground plane, 30 cm long low impedance strap, 1 m mains lead at least 10 cm from ground plane
Uncertainty ± dB
Source of Uncertainty Divisor over frequency range:
9 to 25 25 kHz to 15 to 30
kHz 15 MHz MHz
Insertion Loss of LISN 2 0.07 0.14 0.23
Receiver Level Accuracy √3 1.5 1.5 1.5
Coupling between LISN and Ground Plane √3 0.0 0.10 2.0
Coupling between Mains Lead and GP √3 0.0 0.10 1.0
Coupling between EUT and LISN √3 0.0 0.10 0.5
Length of Earth Strap √3 0.0 0.0 5.0
Effect of mains impedance √3 1.0 0.0 0.0
Mismatch LISN to Receiver √2 0.1 0.52 0.12
Repeatability of measurement 1 0.30 0.30 0.30
Combined Standard Uncertainty 1.09 0.97 3.30
Expanded Uncertainty 2.17 1.94 6.59
For calculation of mismatch uncertainty the following values of VRC were used:
- 48 -
Calibration and use of AMNs and clamps
In the vertical ground plane configuration the mains lead is less likely to be routed toward the ground plane and
therefore the variation in coupling will probably be less.
Repeatability of measurement
The figures given for repeatability assume that conditions, lead positions, and other variables remain
substantially the same and is mainly due to receiver stability. Often the repeatability will be dominated by the
stability of the EUT.
- 49 -
Calibration and use of AMNs and clamps
5.3.1. Calibration
Source of Uncertainty Divisor Uncertainty ± dB over frequency range:
30 to 100 100 to 300 300 to 700 700 to 1000
MHz MHz MHz MHz
Measuring System non-linearity √3 0.3 0.3 0.3 0.3
Measuring System resolution √3 0.2 0.2 0.2 0.2
Measuring System noise √3 0.1 0.1 0.1 0.2
Effect of surroundings √3 0.7 0.3 0.1 0.1
Effect of wire cross section √3 0.5 0.5 0.5 0.5
Effect of length of wire √3 0.3 0.3 0.3 0.3
Effect of fixed absorbers √3 0.3 0.2 0.1 0.1
Effect of size of reference plane √3 0.2 0.2 0.1 0.1
Effect of position of wire in clamp √3 0.1 0.2 0.3 0.6
Effect of coupling to output cable √3 0.2 0.2 0.2 0.2
Mismatch at wire input √2 0.23 0.21 0.19 0.18
Mismatch at clamp output √2 0.15 0.15 0.15 0.15
Repeatability of measurement 1 0.1 0.1 0.1 0.1
Combined Standard Uncertainty 0.66 0.54 0.50 0.59
Expanded Uncertainty (k = 2) 1.31 1.07 1.01 1.19
For calculation of the mismatch uncertainty the following values of Voltage Reflection Coefficient (VRC) were
used:
- 50 -
Calibration and use of AMNs and clamps
to allow for the noise offset this is not recommended, steps should be taken to minimise the effect of signal
noise.
Effect of Surroundings
The uncertainties given in the example are for a laboratory environment where the presence of large metal
objects, such as steel used in the construction of the building RSJs, cannot be avoided. However, all other
measures are taken to ensure that metal objects and the operator are at least 1 m from the clamp and wire. If
measurement are made in an area free of large metal objects to a distance of at least 2 m then this uncertainty can
be reduced to ±0.3 below 300 MHz and ±0.1 above this frequency.
Repeatability of measurement
The value given is a typical standard uncertainty obtained when the whole measurement process is repeated 5
times (standard deviation divided by the square root of 5) .
- 51 -
Calibration and use of AMNs and clamps
For calculation of the mismatch uncertainty the following values of Voltage Reflection Coefficient (VRC) were
used, assuming no attenuator between clamps and receiver:
- 52 -
Calibration and use of AMNs and clamps
Effect of surroundings
The uncertainty for normal laboratory conditions apply to a laboratory area that has some metal framework in its
construction and the possibility of screened rooms being within 2m of the test setup. If measurements are made
without these factor being present then the uncertainty below 300 MHz could be reduced, probably to half the
values given. The uncertainties applicable to a screened room would need to be increased for unlined rooms but
it would be difficult to quantify this, the recommendations given in 4.3.3 should be followed so that uncertainties
for measurements near to the test limits can be reduced.
- 53 -
Calibration and use of AMNs and clamps
Random
The values given for random contributions are typical standard deviations that would be applicable to making
only one measurement. If necessary this contribution could be reduced to an insignificant value if repeat
measurements are made. This would normally be done when an emission is measured that is close to the
specification limit.
- 54 -
Calibration and use of AMNs and clamps
Calibrations of two absorbing clamps and one AMN/LISN were made by both the participating
organisations, Schaffner-Chase EMC and the National Physical Laboratory. The calibrations were
made using the best practices that had been established during the investigation stage of the project.
The results were then compared to establish any variances and to see if the uncertainty estimates were
reasonable and valid.
6.1. AMN/LISNs
Unit #2 was calibrated for impedance and insertion loss under the following conditions:
• The adaptors used were manufactured by each laboratory: NPL used a three pin adaptor with
an earthing strap to the chassis earth socket (Figure 48 (a)); Schaffner-Chase used a single pin
adaptor with a plate to the chassis earth socket (Figure 48 (b))
• The impedance was measured with the mains input open circuit
• The insertion loss was measured according to 3.3.3 from effectively a zero impedance source
• Similar impedance analysers (HP4192A/94A) and vector network analysers (HP8753A) were
used in each laboratory
The results of the measurements are shown in Figure 49 and Figure 50.
(a) (b)
4
Impedance Ohms
40 2
30 0
-2
20
-4
10
-6
Frequency MHz Frequency MHz
0 -8
0.001 0.01 0.1 1 10 100 0.001 0.01 0.1 1 10 100
- 55 -
Calibration and use of AMNs and clamps
LISN COMPARISON: INSERTION LOSS (LIVE) LISN COMPARISON: INSERTION LOSS (LIVE)
1.2 0.50
0.20
Insertion Loss dB
0.8
Difference dB
0.10
0.6 0.00
-0.10
0.4
-0.20
-0.30
0.2
-0.40 Frequency MHz
Frequency MHz
0.0 -0.50
0.001 0.01 0.1 1 10 100 0.001 0.01 0.1 1 10 100
6.1.1. Observations
The results show good agreement between the laboratories, the results for the other lines not shown
here were very similar. The differences in impedance are generally less than 2% which could be
accounted for in the uncertainty of the measuring instruments alone. Since the measurements were
made with different designs of adaptors it is clear that these are not a significant contribution to
uncertainty. The assumptions made about the symmetry of a back-to-back pair referred to in 3.3.1 are
supported. The difference in insertion loss measurements are all less than 0.1 dB which is insignificant
in the context of other uncertainties involved in the use of LISNs.
Absorbing clamps #1 and #2 were calibrated for insertion loss under the following conditions.
• The measurements were made using the procedure as recommended by CISPR 16-1:1993
•
The wire cross section was 1.5 mm2
• The wire was held in the centre of the clamp using an additional guide
• Ten clip on ferrites (Fair-Rite 2643-164151) were used on the wire at 4 m from the reference
plane
• The output cable had ferrites at both ends and was perpendicular to the wire
• The measurements were made in an open area above a concrete ground
• The reference plane size was 2m x 2m
• A 6dB attenuator was used at the clamp output
• Spectrum analysers in max hold mode were used at both laboratories
The results of the measurement are shown in Figure 51. The first graph shows the measured insertion
loss for each absorbing clamp and each laboratory; the second shows the difference between each
laboratory with the estimated uncertainty for each laboratory summed.
- 56 -
Calibration and use of AMNs and clamps
20 1
-Unc
18 0
16 -1
14 -2
12 -3
0 100 200 300 400 500 600 700 800 900 1000 0 100 200 300 400 500 600 700 800 900 1000
6.2.1. Observations
Again the results show good agreement between the two laboratories for both models of clamp. The
differences are generally less than 0.5 dB with the occasional spike up to 1 dB different but all within
the combined estimated uncertainty for both sets of measurements.
The results for both the AMN/LISN and the clamps demonstrate that the calibration techniques are
reproducible between laboratories within the estimated uncertainty. It would, though, be inappropriate
to conclude from this result that the uncertainties could be reduced since both laboratories used
substantially the same procedure and equipment configuration so that systematic differences were
minimised.
- 57 -
Calibration and use of AMNs and clamps
- 58 -
Calibration and use of AMNs and clamps
7. References
[1] CISPR 13: , Limits and methods of measurement of radio disturbance characteristics of broadcast
receivers and associated equipment, IEC, Geneva
[2] CISPR 14-1: 1993, Electromagnetic compatibility - Requirements for household appliances, electric tools
and similar apparatus - Part 1: Emission - product family standard, IEC, Geneva
[3] CISPR 16-1: 1993, Specification for radio disturbance and immunity measuring apparatus and methods -
Part 1: Radio disturbance and immunity measuring apparatus, IEC, Geneva
[4] CISPR 16-2: 1996, Specification for radio disturbance and immunity measuring apparatus and methods -
Part 2: Methods of measurement of disturbances and immunity, IEC, Geneva
[5] CISPR 22: 1993 (Second Edition), Limits and methods of measurement of radio disturbance
characteristics of information technology equipment, IEC, Geneva
[6] Amendment 2: 1996 to CISPR 22: 1993, IEC, Geneva
[7] CISPR 22: 1997 (Third Edition), Information technology equipment - radio disturbance characteristics -
limits and methods of measurement, IEC, Geneva
[8] ANSI C63.4: 1992, Methods of Measurement of Radio-Noise Emissions from Low-Voltage Electrical and
Electronic Equipment in the Range of 9 kHz to 40 GHz, ANSI/IEEE
[9] Draft second amendment to CISPR Publication 16-1, 1996, CISPR/A/201/CDV, May 1997
[10] Paul, C R , Hardin, K B: Diagnosis and reduction of conducted noise emissions, IEEE International
Symposium on EMC, August 1988, Washington, pp 19-23
[11] Gisin, F: Designing LISNs for open area test sites, IEEE International Symposium on EMC, August 1992,
Anaheim, pp 491-493
[12] Tsaliovich, A, Moongilan, D: Remote conducted emission testing using matched LISN power cable
assembly, IEEE International Symposium on EMC, August 1990, pp 431-433
[13] Illingworth, G: Comparisons between the radio interference voltages measured on the 150 ohm V-
network and the 50 ohm, 50 µH Artificial Mains Network, ERA Technology, Report No 81-147R,
January 1982
[14] Efficient evaluation of LISNs and Voltage Probes for EMI tests, Hewlett Packard Application Note 339-
7, May 1987
[15] Kwan H.K.: A theory of operation of the CISPR absorbing clamp, IEE 6th International Conference on
EMC, York, September 1988, pp141-143
[16] Amendment to CISPR Publication 16 Part 1 Clause 13: Absorbing clamp for use in the frequency range
30MHz to 1000MHz and Annex H: Calibration of the absorbing clamp, CISPR/A/210/NP, November
1997
[17] Stecher, M: Faster RFI field-strength measurements through prescanning with MDS clamp, News from
Rohde & Schwarz no. 147 (1995/1), pp28-31
[18] The Expression of Uncertainty and Confidence in Measurement, UKAS document M3003, Edition 1,
December 1997
[19] Guide to the Expression of Uncertainty in Measurement, International Organisation for Standardization,
Geneva 1993
- 59 -