33% found this document useful (9 votes)
5K views15 pages

Service Manual Model: Gentlemax Pro

This document is a service manual for the GentleLASE PRO laser system. It provides information on installing, maintaining, and servicing the system. The manual contains sections on theory of operation, service procedures, engineering schematics, assembly drawings, and warnings for safety during maintenance. Final steps when servicing include performing tests to ensure proper functioning and filling out a performance report.

Uploaded by

mahsa
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
33% found this document useful (9 votes)
5K views15 pages

Service Manual Model: Gentlemax Pro

This document is a service manual for the GentleLASE PRO laser system. It provides information on installing, maintaining, and servicing the system. The manual contains sections on theory of operation, service procedures, engineering schematics, assembly drawings, and warnings for safety during maintenance. Final steps when servicing include performing tests to ensure proper functioning and filling out a performance report.

Uploaded by

mahsa
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 15

GentleLASE PRO Service Manual

Service Manual
Model: GentleMAX PRO

Candela Corporation
530 Boston Post Road
Wayland, MA 01778-1886
Telephone (508) 358-7400
Toll Free (800) 733-8550 ext. 734 (Technical Assistance)
Toll Free (800) 733-7400 ext. 740 (Customer Service)

P/N 8501-00-9035, Revision 01


December 2011

8501-00-1965, Revision 01 Candela Corporation Proprietary I


GentleLASE PRO Service Manual

Table of Contents

SECTION ONE THEORY OF OPERATION

8501-00-4020 Spectrum Theory of Operation

SECTION TWO SERVICE PROCEDURES

8503-01-0930 GMAX PRO Installation Procedures


8503-01-0931 GMAX PRO System Software Initialization Procedure
8503-01-0932 GMAX PRO Preventative Maintenance Procedure
8503-01-0933 GMAX PRO Laser Head Replacement Procedure
8503-01-0934 GMAX PRO Flash Lamp Replacement Procedure
8503-01-0935 GMAX PRO Laser Rail Alignment Procedure
8503-01-0936 GMAX PRO Focus Lens Replacement Procedure
8503-01-0937 GMAX PRO Calport Verification Procedure
8503-01-0938 GMAX PRO Performance Test Procedure
8503-01-0939 GMAX PRO HVPS Verification Procedure
8503-01-0940 GMAX PRO Aiming Beam Test Procedure
8503-01-0941 GMAX PRO Inspect and Clean Optics
8503-01-0942 GMAX PRO Internal Circuit Calibration Procedure
8503-01-0943 GMAX PRO Efficiency and Ramp Test Procedure
8503-01-0944 GMAX PRO DCD Pressure Verification Procedure
8503-01-0945 GMAX PRO DI Water Temperature Calibration Procedure
10-400-00151 GMAX PRO Performance and Usage Data Sheet

SECTION THREE ENGINEERING SCHEMATICS

9914-80-9035 System Schematic, GMAX PRO


7111-80-2792 CPU I/O Schematics
7111-80-2793 AC Distribution PVB Schematic
7111-80-2798 HV Daughter Board Schematic
7111-80-2799 PCB ASSY HV Split Cap Spectrum Schematic
7111-80-2804 DHP HV Dump PCB Schematic

SECTION FOUR ENGINEERING DRAWINGS

9914-91-9035 GMAX PRO Family Tree


9914-84-9035 Block Diagram Control System 1 of 9
9914-84-9035 Block Diagram HV Modulator Section 2 of 9
9914-84-9035 Block Diagram Modulator Control 3 of 9
9914-84-9035 Block Diagram DCD Delivery System 4 of 9
9914-84-9035 Block Diagram ACC Delivery System 5 of 9
9914-84-9035 Block Diagram CALPORT 6 of 9
9914-84-9035 Block Diagram DCD Section 7 of 9
9914-84-9035 Block Diagram Laser Rail Section 8 of 9
9914-84-9035 Block Diagram Fluid Section 9 of 9

8501-00-1965, Revision 01 Candela Corporation Proprietary II


GentleLASE PRO Service Manual

ASSEMBLY DRAWINGS/BOM

7111-99-2510 ASSY, PCB HP, Bulkhead Drawing


7111-99-2792 CPU I/O PCB Drawing
7111-99-2793 AC Distribution PCB Assembly Drawing
7111-99-2798 HV Daughter Board Assembly Drawing
7111-99-2799 HV Split Capacitor Board Assembly Drawing
7111-99-2800 LCD Display Interface PCB Drawing
7111-99-2807 DHP Dump PCB Drawing
7122-99-9250 AC Section Wire Kit Drawing

7122-00-9216 BOM AC Section Assembly


7122-99-9216 AC Section Assembly

7122-00-9228 BOM Heat Exchanger Assembly


7122-99-9228 Heat Exchanger Assembly

7122-00-9234 BOM Tubing Kit Assembly


7122-99-9234 Tubing Kit Assembly

7122-00-9235 BOB DCD Assembly


7122-99-9235 DCD Assembly

7122-00-9400 BOM Assembly, ACC Electrical Cable


7122-99-9400 Assembly, ACC Electrical Cable

7122-00-9401 BOM DCD Hand Cable Assembly


7122-99-9401 DCD Hand Cable Assembly

7122-00-9403 BOM Assembly, ACC Electrical Cable


7122-99-9403 Assembly, ACC Electrical Cable

7122-00-9404 BOM ACC HP Cable Assembly


7122-99-9404 ACC HP Cable Assembly

7122-00-9416 BOM Slider Attachment 12/15/18


7122-99-9416 Slider Attachment 12/15/18

7122-00-9417 BOM Slider Attachment 6/8/10


7122-99-9417 Slider Attachment 6/8/10

7122-00-9419 BOM Fiber Assembly


7122-99-9419 Fiber Assembly

7122-00-9432 BOM Aiming Beam Assembly


7122-99-9432 Aiming Beam Assembly

7122-00-9435 BOM Fiber Pole Assembly


7122-99-9435 Fiber Pole Assembly

7122-00-9441 BOM GMAX DHP Display Bezel Assembly


8501-00-1965, Revision 01 Candela Corporation Proprietary III
GentleLASE PRO Service Manual
7122-99-9441 GMAX DHP Display Bezel Assembly

7122-00-9457 BOM Head Detector Assembly


7122-99-9457 Head Detector Assembly

7122-00-9460 BOM Display Bezel, Rear


7122-99-9460 Display Bezel, Rear

7122-00-9529 BOM Beam Blocker Assembly


7122-99-9529 Beam Blocker Assembly

7122-00-9537 BOM Fiber Pole Accessory Kit


7122-99-9537 Fiber Pole Accessory Kit

7122-00-9557 BOM GMAX DHP Head Rail Assembly


7122-99-9557 GMAX DHP Head Rail Assembly

7122-00-9560 BOM 1.5MM Slider Attachment Assembly


7122-99-9560 1.5MM Slider Attachment Assembly

7122-00-9561 BOM 3MM Slider Attachment Assembly


7122-99-9561 3MM Slider Attachment Assembly

7122-00-9567 BOM Upper Chassis Assembly


7122-99-9567 Upper Chassis Assembly

7122-00-9571 BOM GMAX DHP Tubing Kit


7122-99-9571 GMAX DHP Tubing Kit

7122-00-9572 BOM GMAX DHP Alex Head Assembly


7122-99-9572 GMAX DHP Alex Head Assembly

7122-00-9574 BOM Fiber Receptacle Assembly


7122-99-9574 Fiber Receptacle Assembly

7122-00-9575 BOM Focus Lens Housing Assembly


7122-99-9575 Focus Lens Housing Assembly

7122-00-9576 BOM GMAX DHP Beam Combiner Assembly


7122-99-9576 GMAX DHP Beam Combiner Assembly

7122-00-9577 BOM GMAX DHP Turning Mirror Assembly


7122-99-9577 GMAX DHP Turning Mirror Assembly

7122-00-9578 BOM GMAX DHP YAG Head Assembly


7122-99-9578 GMAX DHP YAG Head Assembly

7122-00-9579 BOM HV Section Assembly


7122-99-9579 HV Section Assembly

7122-00-9580 BOM HV Wire Harness Assembly


7122-99-9580 HV Wire Harness Assembly
8501-00-1965, Revision 01 Candela Corporation Proprietary IV
GentleLASE PRO Service Manual

7122-00-9588 BOM GMAX DHP Intermediate Lens Assembly


7122-99-9588 GMAX DHP Intermediate Lens Assembly

7122-00-9598 BOM Fluid Section Assembly


7122-99-9598 Fluid Section Assembly

7122-00-9601 BOM Frame Assembly


7122-99-9601 Frame Assembly

7122-00-9604 BOM Skins Kit GMAX PRO Assembly


7122-99-9604 Skins Kit GMAX PRO Assembly

7122-00-9608 BOM CALPORT Assembly


7122-99-9608 CALPORT Assembly

8501-00-1965, Revision 01 Candela Corporation Proprietary V


GentleLASE PRO Service Manual

INTRODUCTION

At the conclusion of each service call final closing-up steps should be performed
including the completion of the GMAX PRO Performance and Usage Data Sheet.

This includes exiting Maintenance Mode and replacing the system panels. The
GMAX PRO laser system must be turned off when replacing all panels/dust covers.

Whenever any service is performed on a GMAX PRO laser system, the Performance
and Usage Data Sheet should be filled out completely after the system is closed up.
The laser must be pulsed and tested in User Mode to make sure that nothing has
been disturbed during the installation of all dust covers and panels. You must verify
that the system is ready to be used by the customer and that all previous
parameters set by the customer are the same as when you arrived, including
the footswitch to handpiece setting, if customer was not using the footswitch.

______________________________________________________________________
WARNING
FOLLOW THE INSTALLATION STEPS IN SEQUENCE TO ELIMINATE ANY
POSSIBLE DAMAGE TO THE SYSTEM.

Warnings, Cautions, and Precautions

Observe the following warnings and cautions when using the Candela Corporation
GMAX PRO Laser system.

Warnings:

WARNING!
The electrical and laser radiation hazards present during servicing of the GMAX
PRO Laser can be extremely dangerous if proper safety precautions are not taken.
Consequently, the GMAX PRO Laser is to be serviced only by those qualified
technicians who have received appropriate training on the GMAX PRO Laser from
Candela, and who are familiar with the safety considerations discussed in this
section.

WARNING!
The GMAX PRO Laser system has been designed for the safest possible operation
and maintenance. However, any laser system can cause injury if it is not properly
installed, operated, moved or serviced, and the GMAX PRO Laser is no exception.
The potential hazards associated with the GMAX PRO Laser are: ocular (vision)
damage resulting from exposure to direct or reflected laser radiation, electrical shock
from contact with high voltage electrical components inside the system, and physical
injury incurred while moving the system.

WARNING!
8501-00-1965, Revision 01 Candela Corporation Proprietary VI
GentleLASE PRO Service Manual
Invisible and visible laser radiation. Avoid eye or skin exposure to direct or scattered
radiation. Class 4 laser product (Per EN60825-1: 2002-07)

WARNING!
The Preset Treatment Parameters and Clinical Treatment Guidelines do not take the
place of the procedures and instructions found in the Operator’s Manual. Failure to
use this laser in accordance with such procedures and instructions could result in
serious injury to the operator, the patient, and others, as well as damage to the laser
system.

Follow OSHA and ANSI standards for laser safety. Protective eyewear must be worn
by all persons in the treatment room during laser operation.

Check the delivery system for any damage (i.e. dropped).

Discontinue use of your laser delivery system if you suspect a problem.

WARNING!
Always put the laser system into Standby or turn it off before attempting to check, clean,
and/or replace the Delivery System, Handpiece, Slider Attachments, or Distance Gauge.

WARNING!
Always recalibrate the laser after fixing, cleaning, or replacing the Delivery System,
Handpiece, or Slider Attachments and/or Windows. Failure to initiate a calibration after
cleaning/replacing the Handpiece, Slider Attachments, Windows, and/or Delivery System
may result in the delivery of excessive laser energy.

WARNING!
Ensure that the spot size on the Slider Attachments and Distance Gauges matches the spot
size displayed in the spot size field on the display panel. Failure to do so can result in the
delivery of improper energy to the patient.

WARNING!
Do not operate the laser if the aiming beam is on but not present when it is turned on! This
may be an indication of a broken fiber optic. If the aiming beam is on but is not present,
replace the Delivery System. If this does not correct the problem, call Technical Support.

Cautions:
Caution!
Use of controls or adjustments or performance of procedures other than those
specified herein may result in hazardous radiation exposure.

Caution!
Federal (USA) law restricts this device to sale by or on the order of a physician.

Caution!
8501-00-1965, Revision 01 Candela Corporation Proprietary VII
GentleLASE PRO Service Manual
Do not enter the Ready state without a Delivery System installed and without having
the proper protective eyewear on.

Caution!
Before the laser system is turned on, a Slider Attachment must be installed in the
Hand piece.

Caution!
Slider Attachment lenses and the tips of the fiber cable may be damaged from
exposure to dust particles or any other foreign particles that may deposit on their
surfaces. Particles on these surfaces will burn and leave a deposit when exposed to
laser energy. This may lead to lower fiber or Slider Attachment transmission and/or
failure of the assembly. In order to reduce the probability of mechanical damage,
please observe the following guidelines.

Caution!
To reduce the risk of personal injury and damage to the Delivery System cables, use
the Fiber Pole to support the delivery system at all times. When not in use, insert the
Handpiece in the holder on the side panel or in the Calibration Port. This removes
excess slack from the Delivery System cables and reduces the possibility of damage
to property and/or personal injury from stepping on or tripping on the cable or
running the wheels over it.

Caution!
When using the Fiber Pole to support the Delivery System, make sure there are no
sharp bends in the Delivery System cables. The laser system can be damaged if the
cable is subjected to excessive bending. To prevent damage, never pulse the laser
system if the Delivery System cable bend radius is less than six inches.

Optical Hazards:

WARNING!
The laser operates at 755 nm and at 1064 nm, which lies in the invisible, near-
infrared region of the electromagnetic spectrum. The laser beam emitted by the
GMAX PRO is capable of causing loss of vision. Energy emitted by this Laser
System that enters the eye will be focused directly on the retina. Direct contact of the
laser beam on the retina can cause temporary clouded vision, retinal lesions, long-
term scotoma (vision absence in an isolated area), long term photophobia
(sensitivity to light) and/or loss of vision.

WARNING!

8501-00-1965, Revision 01 Candela Corporation Proprietary VIII


GentleLASE PRO Service Manual
Use only safety eyewear that is known to have an optical density (O.D.) of > 5.8 for
745 - 765 nm and an optical density (O.D.) of > 6.3 for 1061 – 1065 nm. Safety
eyewear that is designed for use with other laser systems may not provide adequate
protection.

WARNING!
Nominal Ocular Hazard Distance (NOHD). The laser aperture of the GMAX PRO
Laser System is at the distal end of the Handpiece. The beam enlarges as the
distance from the Handpiece increases. The Nominal Ocular Hazard Distance
(NOHD) is the distance at which the beam is so big it is no longer dangerous to the
unprotected eye. The distance along with the full angle beam divergence for each
spot size is shown in the Table below.

WARNING!
To avoid vision hazards, everyone (including service personnel) within the NOHD
where the laser is operating must wear appropriate eye protection available from
Candela. Such eyewear provides adequate protection against reflected or scattered
laser radiation, or inadvertent brief exposure to the laser beam. Laser safety
eyewear should be stored away from direct sunlight at temperatures of 65°F - 75°F
or 18° - 24°C.

8501-00-1965, Revision 01 Candela Corporation Proprietary IX


GentleLASE PRO Service Manual

WARNING!
During laser procedures, the patient's eyes must be protected. The opaque patient
goggles provided by Candela are appropriate for most patients. It is recommended
to place gauze sponges under the opaque patient goggles to ensure that the patient’
s eyes remain closed. In addition, the opaque goggles do not fit well if used on
infants or small children. Gauze sponges can be used when moistened with water
and taped over the eyelids or a moistened facecloth can be held over the eyes are
recommended. If the patient is asleep, the eyes should be taped closed and covered
with moistened gauze sponges.

WARNING!
Even when wearing protective eyewear, looking directly into the path of the laser
beam may cause permanent eye damage .

Caution!
The laser beam emitted by the GMAX PRO Laser System should never be directed
at any part of the body other than the intended site of treatment or testing.

Caution!
Removal of any of the exterior panels could allow access to hazardous levels of
laser radiation. For this reason, these panels are designed not to be easily
removable; they must not be removed except by authorized Candela trained service
personnel.

Electrical and Mechanical Hazards:

WARNING!
The GMAX PRO Laser converts and amplifies the AC line voltage to produce
extremely high voltages inside the laser system which are very dangerous, even
lethal. It is possible for high voltage components to retain a charge after the power
supply has been turned off, and even after the laser has been disconnected from the
line voltage or wall receptacle. Therefore, no part of the laser interior should be
opened or exposed, except by trained and authorized technicians.

WARNING!
The GMAX PRO laser delivery systems utilize fiber optics that can be damaged if
installed or subjected to excessive bending. To avoid damage to the optical fiber,
limit bends to a radius of 6 inches (15 cm) or greater. Failure to follow recommended
procedures may lead to damage to the fiber or delivery system and/or harm to the
patient or operator. When damaged, the fiber or delivery system becomes a
potential fire hazard.

8501-00-1965, Revision 01 Candela Corporation Proprietary X


GentleLASE PRO Service Manual
WARNING!
Although the GMAX PRO Laser system is well balanced, it weighs more than 250
pounds (almost 150 kg) and may cause injury if proper care is not used when
moving it. The system should always be moved carefully and slowly.

Caution!
To prevent the laser from moving, both front wheels must be locked. To lock the
wheels, step down on the tabs on the front of the wheels. To unlock, step down on
the top (smaller) tabs.

Chemical Hazards
There are no known chemical hazards associated with the GMAX PRO Laser
System.

Cryogen:

This laser system uses a Hydrofluorocarbon (HFC) or cryogen in the Dynamic


Cooling Device (DCD).

Inhalation: If high concentrations are inhaled, immediately move to fresh air. Keep
person calm. If not breathing, give artificial respiration. If breathing is difficult, give
oxygen. Call a physician.

Skin Contact: If large amounts of cryogen contact the skin due to a leak or rupture in
the cryogen system, flush skin immediately with water and call a Physician to check for
frostbite. Treat for frostbite if necessary by gently warming affected area.

Eye Contact: In case of eye contact, immediately flush eyes with plenty of water for at
least 15 minutes. Call a physician.

Ingestion: Ingestion is not considered a potential route of exposure.

WARNING! IMPORTANT NOTE TO PHYSICIANS


Because of possible disturbance of cardiac rhythm, catecholamine drugs such
as epinephrine should only be used with special caution in situations, when
performing emergency life support. See MSDS sheet, Candela P/N 8501-00-1701.

8501-00-1965, Revision 01 Candela Corporation Proprietary XI


GentleLASE PRO Service Manual

8501-00-1965, Revision 01 Candela Corporation Proprietary XII


GentleLASE PRO Service Manual

8501-00-1965, Revision 01 Candela Corporation Proprietary XIII


GentleLASE PRO Service Manual

8501-00-1965, Revision 01 Candela Corporation Proprietary XIV


GentleLASE PRO Service Manual

8501-00-1965, Revision 01 Candela Corporation Proprietary XV

You might also like