M/s.Sabui Medical Stores (MFG) 3, Sardar Ghat Lane, Bhadreshwar, Hooghly
M/s.Sabui Medical Stores (MFG) 3, Sardar Ghat Lane, Bhadreshwar, Hooghly
PURPOSE
To the determine the actual Airborne Particle count Cleanliness Level in a clean room "At Rest" Occupancy State
CUSTOMER DETAILS
Client M/s.Sabui Medical Stores (MFG) Document No GMPLCS/SML/PCT/21-22/B001
3,Sardar Ghat Lane,Bhadreshwar,Hooghly. Test Reference EUGMP-ISO:2015
telangana-503102 Date of Test 13/09/2021
Equipment Name Laminar Air Flow Next Due Date 12/09/2022
Equipment Id RLS/WH/RLAF/003 Area of test Microbiology
INSTRUMENT USED FOR TESTING
Instrument Name Airborne Particle counter Make TSI
Model 9350-03 Serial No 93501922002
Flow Rate 50.0 lpm Sampling Time 01 Minute
Calibrated On 14/07/2021 Calibration Due On 14/07/2022
OBTAINED TEST RESULT ACCEPTANCE CRITERIA
ISO Accepatance Accepatance
Obtained No.of Particles ≥ 0.5 μm³ Obtained No.of Particles ≥ 5.0 μm³ Result
Classification Location Criteria for the Criteria for the
Number Particle size ≥ Particle size ≥
S1 S1 0.5 µm /ft³ 5.0 µm / ft³
1 0 0 100 0 Pass
2 0 0 100 0 Pass
3 0 0 100 0 Pass
4 0 0 100 0 Pass
5 0 0 100 0 Pass
ISO Class "5"
6 0 0 100 0 Pass
7 0 0 100 0 Pass
8 0 0 100 0 Pass
9 0 0 100 0 Pass
10 0 0 100 0 Pass
Mean of the average 0 0 100 0 Pass
INFERENCE The above Specified Confirms to ISO Class "A" for Particle Count Test as per -ISO:2015
Mr.Balaji ENGINEERING QA
M/s.GMP laboratories And Clean Room Solutions M/s.Sabui Medical Stores (MFG)
Page 1 of 6
PARTICLE COUNT TEST CERTIFICATE
PURPOSE
To the determine the actual Airborne Particle count Cleanliness Level in a clean room "At Rest" Occupancy State
CUSTOMER DETAILS
Client M/s.Sabui Medical Stores (MFG) Document No GMPLCS/SML/PCT/21-22/B002
3,Sardar Ghat Lane,Bhadreshwar,Hooghly. Test Reference EUGMP-ISO:2015
telangana-503102 Date of Test 13/09/2021
Room Name Filling Sealing Next Due Date 12/09/2022
Room ID Filling sealing Room Area of test Filling Area
INSTRUMENT USED FOR TESTING
Instrument Name Airborne Particle counter Make TSI
Model 9350-03 Serial No 93501922002
Flow Rate 50.0 lpm Sampling Time 01 Minute
Calibrated On 14/07/2021 Calibration Due On 14/07/2022
OBTAINED TEST RESULT ACCEPTANCE CRITERIA
ISO Accepatance Accepatance
Obtained No.of Particles ≥ 0.5 μm³ Obtained No.of Particles ≥ 5.0 μm³ Result
Classification Location Criteria for the Criteria for the
Number Particle size ≥ Particle size ≥
S1 S1 0.5 µm /ft³ 5.0 µm / ft³
INFERENCE The above Specified Confirms to ISO Class "D" for Particle Count Test as per -ISO:2015
Mr.Balaji ENGINEERING QA
M/s.GMP laboratories And Clean Room Solutions M/s.Sabui Medical Stores (MFG)
Page 2 of 6
PARTICLE COUNT TEST CERTIFICATE
PURPOSE
To the determine the actual Airborne Particle count Cleanliness Level in a clean room "At Rest" Occupancy State
CUSTOMER DETAILS
Client M/s.Sabui Medical Stores (MFG) Document No GMPLCS/SML/PCT/21-22/B003
3,Sardar Ghat Lane,Bhadreshwar,Hooghly. Test Reference EUGMP-ISO:2015
telangana-503102 Date of Test 13/09/2021
Room Name Bulk Filling Next Due Date 12/09/2022
Room ID Bulk Filling Room Area of test Filling Area
INSTRUMENT USED FOR TESTING
Instrument Name Airborne Particle counter Make TSI
Model 9350-03 Serial No 93501922002
Flow Rate 50.0 lpm Sampling Time 01 Minute
Calibrated On 14/07/2021 Calibration Due On 14/07/2022
OBTAINED TEST RESULT ACCEPTANCE CRITERIA
ISO Accepatance Accepatance
Obtained No.of Particles ≥ 0.5 μm³ Obtained No.of Particles ≥ 5.0 μm³ Result
Classification Location Criteria for the Criteria for the
Number Particle size ≥ Particle size ≥
S1 S1 0.5 µm /ft³ 5.0 µm / ft³
INFERENCE The above Specified Confirms to ISO Class "D" for Particle Count Test as per -ISO:2015
Mr.Balaji ENGINEERING QA
M/s.GMP laboratories And Clean Room Solutions M/s.Sabui Medical Stores (MFG)
Page 3 of 6
PARTICLE COUNT TEST CERTIFICATE
PURPOSE
To the determine the actual Airborne Particle count Cleanliness Level in a clean room "At Rest" Occupancy State
CUSTOMER DETAILS
Client M/s.Sabui Medical Stores (MFG) Document No GMPLCS/SML/PCT/21-22/B004
3,Sardar Ghat Lane,Bhadreshwar,Hooghly. Test Reference EUGMP-ISO:2015
telangana-503102 Date of Test 13/09/2021
Room Name LAF Room Next Due Date 12/09/2022
Room ID LAF Room Area of test Microbiology
INSTRUMENT USED FOR TESTING
Instrument Name Airborne Particle counter Make TSI
Model 9350-03 Serial No 93501922002
Flow Rate 50.0 lpm Sampling Time 01 Minute
Calibrated On 14/07/2021 Calibration Due On 14/07/2022
OBTAINED TEST RESULT ACCEPTANCE CRITERIA
ISO Accepatance Accepatance
Obtained No.of Particles ≥ 0.5 μm³ Obtained No.of Particles ≥ 5.0 μm³ Result
Classification Location Criteria for the Criteria for the
Number Particle size ≥ Particle size ≥
S1 S1 0.5 µm /ft³ 5.0 µm / ft³
INFERENCE The above Specified Confirms to ISO Class "D" for Particle Count Test as per -ISO:2015
Mr.Balaji ENGINEERING QA
M/s.GMP laboratories And Clean Room Solutions M/s.Sabui Medical Stores (MFG)
Page 4 of 6
PARTICLE COUNT TEST CERTIFICATE
PURPOSE
To the determine the actual Airborne Particle count Cleanliness Level in a clean room "At Rest" Occupancy State
CUSTOMER DETAILS
Client M/s.Sabui Medical Stores (MFG) Document No GMPLCS/SML/PCT/21-22/B005
3,Sardar Ghat Lane,Bhadreshwar,Hooghly. Test Reference EUGMP-ISO:2015
telangana-503102 Date of Test 13/09/2021
Room Name Air Lock Next Due Date 12/09/2022
Room ID Air Lock Room Area of test Microbiology
INSTRUMENT USED FOR TESTING
Instrument Name Airborne Particle counter Make TSI
Model 9350-03 Serial No 93501922002
Flow Rate 50.0 lpm Sampling Time 01 Minute
Calibrated On 14/07/2021 Calibration Due On 14/07/2022
OBTAINED TEST RESULT ACCEPTANCE CRITERIA
ISO Accepatance Accepatance
Obtained No.of Particles ≥ 0.5 μm³ Obtained No.of Particles ≥ 5.0 μm³ Result
Classification Location Criteria for the Criteria for the
Number Particle size ≥ Particle size ≥
S1 S1 0.5 µm /ft³ 5.0 µm / ft³
INFERENCE The above Specified Confirms to ISO Class "D" for Particle Count Test as per -ISO:2015
Mr.Balaji ENGINEERING QA
M/s.GMP laboratories And Clean Room Solutions M/s.Sabui Medical Stores (MFG)
Page 5 of 6
PARTICLE COUNT TEST CERTIFICATE
PURPOSE
To the determine the actual Airborne Particle count Cleanliness Level in a clean room "At Rest" Occupancy State
CUSTOMER DETAILS
Client M/s.Sabui Medical Stores (MFG) Document No GMPLCS/SML/PCT/21-22/B006
3,Sardar Ghat Lane,Bhadreshwar,Hooghly. Test Reference EUGMP-ISO:2015
telangana-503102 Date of Test 13/09/2021
Room Name Incubator Next Due Date 12/09/2022
Room ID Incubator Room Area of test Microbiology
INSTRUMENT USED FOR TESTING
Instrument Name Airborne Particle counter Make TSI
Model 9350-03 Serial No 93501922002
Flow Rate 50.0 lpm Sampling Time 01 Minute
Calibrated On 14/07/2021 Calibration Due On 14/07/2022
OBTAINED TEST RESULT ACCEPTANCE CRITERIA
ISO Accepatance Accepatance
Obtained No.of Particles ≥ 0.5 μm³ Obtained No.of Particles ≥ 5.0 μm³ Result
Classification Location Criteria for the Criteria for the
Number Particle size ≥ Particle size ≥
S1 S1 0.5 µm /ft³ 5.0 µm / ft³
INFERENCE The above Specified Confirms to ISO Class "D" for Particle Count Test as per -ISO:2015
Mr.Balaji ENGINEERING QA
M/s.GMP laboratories And Clean Room Solutions M/s.Sabui Medical Stores (MFG)
Page 6 of 6