100% found this document useful (1 vote)
63 views

NF148 Protocole-General-Validation en

This document provides validation protocols for alternative commercial methods for water microbiology analysis as compared to reference methods. It outlines procedures for preliminary studies, interlaboratory studies, and criteria for validating qualitative and quantitative methods. The preliminary study involves assessing method selectivity, accuracy, sensitivity, specificity, detection limits, and practicality. The interlaboratory study further evaluates the methods across multiple laboratories by measuring accuracy, repeatability, reproducibility, and constructing accuracy profiles to determine if validation criteria are met. The document provides statistical calculations and interpretations for validating both qualitative and quantitative alternative water microbiology analysis methods.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
100% found this document useful (1 vote)
63 views

NF148 Protocole-General-Validation en

This document provides validation protocols for alternative commercial methods for water microbiology analysis as compared to reference methods. It outlines procedures for preliminary studies, interlaboratory studies, and criteria for validating qualitative and quantitative methods. The preliminary study involves assessing method selectivity, accuracy, sensitivity, specificity, detection limits, and practicality. The interlaboratory study further evaluates the methods across multiple laboratories by measuring accuracy, repeatability, reproducibility, and constructing accuracy profiles to determine if validation criteria are met. The document provides statistical calculations and interpretations for validating both qualitative and quantitative alternative water microbiology analysis methods.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 47

Application identification No.

: NF148
Issue date: 30 October 2013

Validation of analysis methods

Application to water microbiology

Validation protocol for an alternative commercial method


as compared with a reference method

Revision 2 – Adopted by AFNOR Certification on 17 May 2013


( after approval by the relevant Technical Board)
NF148 - NF Mark VALIDATION (application to the analysis of water)
General validation protocol – Revision 2 (Issue dated 30 October 2013 – Approval of 17 May 2013)

CONTENTS

Introduction ......................................................................................................................... 4
1 Purpose ......................................................................................................................... 4
2 Normative references ................................................................................................... 4
3 Terms, definitions and symbols .................................................................................. 4
3.1 General terms......................................................................................................................... 5
3.1.1 Reference method ............................................................................................................... 5
3.1.2 Alternative method ............................................................................................................... 5
3.1.3 Validation of an alternative method ..................................................................................... 5
3.1.4 Analyte ................................................................................................................................. 5
3.1.5 Quantitative method............................................................................................................. 5
3.1.6 Qualitative method ............................................................................................................... 5
3.1.7 Organizing laboratory / expert laboratory ............................................................................ 5
3.1.8 Collaborating laboratory ...................................................................................................... 6
3.2 Validation criteria for qualitative methods .......................................................................... 6
3.2.1 Positive deviation (PD) ........................................................................................................ 6
3.2.2 Negative deviation (ND) ...................................................................................................... 6
3.2.3 Relative sensitivity (SE) ....................................................................................................... 6
3.2.4 Relative accuracy (AC) ........................................................................................................ 6
3.2.5 Relative specificity (SP) ....................................................................................................... 7
3.2.6 Inclusivity and exclusivity (Selectivity) ................................................................................. 7
3.2.7 Relative detection level or LOD50 ........................................................................................ 7
3.2.8 Practicability......................................................................................................................... 7
3.3 Validation criteria for quantitative methods ....................................................................... 7
3.3.1 Linearity ............................................................................................................................... 7
3.3.2 Accuracy .............................................................................................................................. 7
3.3.3 Relative accuracy ................................................................................................................ 7
3.3.4 Repeatability ........................................................................................................................ 7
3.3.5 Reproducibility ..................................................................................................................... 8
3.3.6 Limit of detection.................................................................................................................. 8
3.3.7 Quantification limit (or determination limit) .......................................................................... 8
3.3.8 Inclusivity and exclusivity (Selectivity) ................................................................................. 8
3.3.9 Practicability (see 3.2.8) ...................................................................................................... 8
3.3.10 Acceptability limit ............................................................................................................. 8
4 Principle ........................................................................................................................ 8
5 Preliminary study ......................................................................................................... 9

Page 2/47
NF148 - NF Mark VALIDATION (application to the analysis of water)
General validation protocol – Revision 2 (Issue dated 30 October 2013 – Approval of 17 May 2013)

5.1 Qualitative methods .............................................................................................................. 9


5.1.1 Protocol for the assessment of inclusivity and exclusivity (selectivity) ................................ 9
5.1.2 Protocol to measure relative accuracy, sensitivity and specificity ..................................... 10
5.1.3 Protocol to measure the relative detection level ................................................................ 12
5.1.4 Practicability....................................................................................................................... 13
5.2 Quantitative methods .......................................................................................................... 14
5.2.1 Principle and definitions ..................................................................................................... 14
5.2.2 Protocol for the assessment of inclusivity and exclusivity (selectivity) .............................. 14
5.2.3 Protocol to measure linearity and relative accuracy .......................................................... 15
5.2.4 Detection and quantification limits ..................................................................................... 16
5.2.5 Practicability of the alternative method .............................................................................. 16
6 Interlaboratory study...................................................................................................17
6.1 Qualitative methods ............................................................................................................ 17
6.1.1 Design ................................................................................................................................ 17
6.1.2 Calculations and interpretation .......................................................................................... 17
6.1.3 Interpretation ...................................................................................................................... 19
6.2 Quantitative methods .......................................................................................................... 19
6.2.1 Accuracy profile: general principles ................................................................................... 19
6.2.2 Design ................................................................................................................................ 22
6.2.3 Statistical calculations........................................................................................................ 23
6.2.4 Construction of the accuracy profile .................................................................................. 26
6.3 Summary of the interlaboratory test ................................................................................. 27
6.3.1 Interpretation of the results ................................................................................................ 27
6.3.2 Limit of quantification ......................................................................................................... 27
6.4 Bibliography for section 6 .................................................................................................. 28

Annex 1. Classification by categories of water ................................................................................ 29


Annex 2. Stress to apply by type of water ........................................................................................ 30
Annex 3. Organisation of analyses carried out in parallel depending on the type of method .... 31
Annex 4. Qualitative methods. Test applied to the rejection of discordant results ..................... 33
Annex 5. Qualitative methods. Interlaboratory study, criteria of accordance, concordance and
concordance odds ratio ...................................................................................................................... 34
Annex 6. Quantitative methods. Calculation of repeatability and reproducibility standard
deviations ............................................................................................................................................. 39
Annex 7. Example of an accuracy profile calculation ..................................................................... 40
Annex 8. Guidelines for the organisation and conducting of interlaboratory studies................. 43

Page 3/47
NF148 - NF Mark VALIDATION (application to the analysis of water)
General validation protocol – Revision 2 (Issue dated 30 October 2013 – Approval of 17 May 2013)

Introduction
This document sets out a procedure for the characterization and validation of a method of
microbiological analysis (qualitative or quantitative) as compared with a standard reference
method when it exists.

1 Purpose
This document provides guidelines that shall lead to the validation, for a specified scope (e.g.
water intended for human consumption and/or treated water and/or untreated water and/or
surface water and/or deep water and/or bathing water and/or swimming pool water and/or
mineral water and/or bottled water, etc.), of a "rapid" or commercial method of microbiological
analysis designed to find (qualitative method) or count (quantitative method) target
microorganisms as compared with a standard reference method targeting the same
microorganisms.
The procedure is organized in the following way:
1. A preliminary study carried out by an expert laboratory which supplies an initial
characterization of the method in terms of classical validation criteria (repeatability,
practicability ...).
2. An inter-laboratory study designed to estimate various criteria for qualitative methods
(§ 6.1.3) and to construct the accuracy profile (§ 6.2.4), for quantitative methods.
3. Final interpretation of the results.
The methods to be validated may be direct counting methods, for example of colonies, or
indirect counting, for example counting the most probable number MPN.

2 Normative references
The following normative documents contain provisions that, owing to the reference made in
them, constitute valid provisions for this document. In the case of dated references, later
amendments or revisions of these publications do not apply. However, stakeholders in
agreements founded on this document are invited to research the option of applying the most
recent editions of the normative documents indicated below. In the case of undated
references, the latest edition of the normative document referred to applies.
ISO/TR 13843:2000, Water quality – Guidance on validation of microbiological methods.
EN ISO 16140:2003, Microbiology of food – Protocol for the validation of alternative
methods.

3 Terms, definitions and symbols


For the purposes of this document, the following terms and definitions apply.

Page 4/47
NF148 - NF Mark VALIDATION (application to the analysis of water)
General validation protocol – Revision 2 (Issue dated 30 October 2013 – Approval of 17 May 2013)

3.1 General terms


3.1.1 Reference method
A standardized method of analysis specified for the purpose of detecting or counting a
species or a given group of microorganisms.

3.1.2 Alternative method


A commercial method used to detect or count, for a given type of water, the same
microorganisms as those measured by a standardized method, and which has for example,
one or more of the following criteria:
 Speed of analysis and/or response.
 Ease of execution and/or automation.
 Analytical characteristics (precision, detection limit, etc.).
The term "alternative method" designates the product, equipment and procedure as a whole.
It includes all the ingredients, whether material or not, required to implement the alternative
method.

3.1.3 Validation of an alternative method


According to VIM:2006 subclauses 2.44 and 2.45
Provision of tangible evidence that the alternative method meets the requirements specified,
that is results comparable to the results obtained with the reference method, taking into
account any measurement uncertainty and that these requirements specified are suitable for
a pre-determined use.

3.1.4 Analyte
Component measured using the method of analysis (microorganisms).

3.1.5 Quantitative method


Method of analysis whose response is the amount of analyte measured either directly
(counting in a mass or volume), or indirectly (colour absorbance, impedance, etc.) in a
certain amount of sample.

3.1.6 Qualitative method


Method of analysis whose response is qualitative (presence/absence of the analyte) and
measured either directly or indirectly (colour absorbance, impedance, etc.) in a certain
amount of sample.

3.1.7 Organizing laboratory / expert laboratory


Laboratory having the qualified staff and skills required to perform the preliminary study and
the inter-laboratory study. It is selected according to criteria defined by AFNOR Certification.

Page 5/47
NF148 - NF Mark VALIDATION (application to the analysis of water)
General validation protocol – Revision 2 (Issue dated 30 October 2013 – Approval of 17 May 2013)

3.1.8 Collaborating laboratory


Laboratory participating in the inter-laboratory study, according to the instructions defined by
the organizing /expert laboratory.

3.2 Validation criteria for qualitative methods


The different situations of agreement or deviation between the two methods are summarized
in Table 1.

Table 1. Types of agreement and deviation of results between the reference and alternative
methods.
Reference method Reference method
Results Total
positive negative
Alternative method PA PD
A+
positive Positive agreement +/+ Positive deviation -/+
NA
Alternative method ND
Negative agreement A–
negative Negative deviation +/-
-/-
Total N+ = PA+ND N– = NA+PD N

3.2.1 Positive deviation (PD)


The alternative method presents a positive deviation if it gives a positive result when the
reference method gives a negative result. A positive deviation becomes a false positive result
when the true result can be proven as being negative. A positive deviation is considered as a
true positive when the true result can be proven as being positive.

3.2.2 Negative deviation (ND)


The alternative method presents a negative deviation if it gives a negative result when the
reference method gives a positive result. A negative deviation becomes a false negative
result when the true result can be proved as being positive.

3.2.3 Relative sensitivity (SE)


Ability of the alternative method to detect the analyte when it is detected by the reference
method.
PA
SE  100 
N

3.2.4 Relative accuracy (AC)


Relative accuracy
PA  NA
AC  100 
N

Page 6/47
NF148 - NF Mark VALIDATION (application to the analysis of water)
General validation protocol – Revision 2 (Issue dated 30 October 2013 – Approval of 17 May 2013)

3.2.5 Relative specificity (SP)


Ability of the alternative method to not detect the analyte when it is not detected by the
reference method.
NA
SP  100 
N

3.2.6 Inclusivity and exclusivity (Selectivity)


Inclusivity is the ability of the alternative method to detect the target analyte from a relevant
range of strains.
Exclusivity is the ability of the alternative method not to detect non-target analytes from a
relevant range of strains.

3.2.7 Relative detection level or LOD50


The relative detection level is the smallest number of culturable microorganisms that can be
detected in the sample, with a probability of 50 % by the alternative and reference methods.
LOD50 is estimated by a statistical model but never directly measured.

3.2.8 Practicability
Usability of a method, defined by a set of criteria given by the user.

3.3 Validation criteria for quantitative methods


3.3.1 Linearity
Ability of the method when used with a given matrix to give results that are in proportion to
the amount of analyte present in the sample, that is an increase in analyte corresponds to a
linear or proportional increase in results.

3.3.2 Accuracy
Closeness of agreement between a test result and the accepted reference value.

3.3.3 Relative accuracy


Degree of correspondence between the response obtained by the reference method and the
response obtained by the alternative method on the same samples.
NOTE The term "relative accuracy" used here is complementary to 3.3.2. For the purpose of this
document, the accepted reference value is chosen as the value obtained by the reference method.
The term "relative" implies therefore that the reference method does not automatically provide the
accepted reference value.

3.3.4 Repeatability
Closeness of agreement between successive measurements of the same quantity, carried
out under the same conditions of measurement.

Page 7/47
NF148 - NF Mark VALIDATION (application to the analysis of water)
General validation protocol – Revision 2 (Issue dated 30 October 2013 – Approval of 17 May 2013)

3.3.5 Reproducibility
Closeness of agreement between successive measurements of the same quantity, carried
out under conditions of reproducibility.

3.3.6 Limit of detection


(VIM. 4.18) Measured value, obtained by a given procedure, for which the probability of
falsely claiming the absence of a component in a material is β, given a probability α of falsely
claiming its presence.
Or
(Draft standard XP T 90-465-1) The lowest degree of bacterial load for which the probability
of a positive result (observation of at least one germ) is greater than or equal to 95 % or
99 %. Note: this definition involves a probability theory law of the Poisson or negative
binomial type.

3.3.7 Quantification limit (or determination limit)


The lowest average concentration of an analyte that may be measured and quantified with
known uncertainty, under the experimental conditions of the method considered.

3.3.8 Inclusivity and exclusivity (Selectivity)


Inclusivity is the ability of the alternative method to detect the target analyte from a wide
range of strains.
Exclusivity is the lack of interference by a relevant range of non-target strains with the
alternative method.
NOTE This criterion does not apply to counting total flora.

3.3.9 Practicability (see 3.2.8)

3.3.10 Acceptability limit


Maximum acceptable deviation between the value assigned by the reference method and a
result obtained by the alternative method. It is recorded as λ. In the case of bacterial
counting, it is expressed in log units.

4 Principle
The validation protocol comprises two phases:
1. A preliminary study carried out by a single, so-called "expert laboratory" to estimate the
performance of the alternative method against the reference method, carried out on
natural or artificial samples or on reference materials.
2. An interlaboratory study organised by the expert laboratory with a minimum set number
of laboratories involving the use of artificially spiked samples analysed in parallel using
the alternative method and the reference method.
Before starting the study, whether the scope of the alternative method is compatible with that
of the reference method should be defined.

Page 8/47
NF148 - NF Mark VALIDATION (application to the analysis of water)
General validation protocol – Revision 2 (Issue dated 30 October 2013 – Approval of 17 May 2013)

In particular, this includes:


 identification of the microorganism or microorganisms targeted.
 the categories of water concerned and the concentration range envisaged.
The different categories of water that can be covered by the study are summarised at Annex
1. Classification by categories of water.

5 Preliminary study

5.1 Qualitative methods


5.1.1 Protocol for the assessment of inclusivity and exclusivity
(selectivity)
To study the selectivity of the alternative method, analyse the strains in the list in Table 2.

Table 2. Minimum number of strains for the selectivity study.


Types of strains
Target strains (species) (for example E. coli or Legionella
20
pneumophila)
Inclusivity
Target strains belonging to a genus, a group or a family (for example
30
coliforms, Legionella spp., or intestinal enterococci).
Non-target strains known to cause interference with target strains
20
(genus-species).
Exclusivity
Non-target strains known to cause interference with target strains
30
belonging to a group or a family.

NOTE The number of strains to test may be increased according to the target (germ) of the
alternative method to be validated (for example: 50 target strains on an inclusive basis when looking
for salmonella).
Choose the target and non-target strains that regularly contaminate the class or classes,
and/or categories of water selected. In addition, strains shall if possible come mainly from
natural samples representative of the water's envisaged scope. The origin of each strain
shall be known, recorded and traceable.
The level of inoculation of each microorganism is defined as follows:
 Target microorganisms (inclusivity): shall be around 10 to 100 times greater than the
minimum relative detection level of the alternative method. The alternative method's
complete protocol – without including the test portion – shall be implemented, but
including pre-enrichment if stipulated. If the principle of the reference method is different
from that of the alternative method, both methods are implemented in parallel.

Page 9/47
NF148 - NF Mark VALIDATION (application to the analysis of water)
General validation protocol – Revision 2 (Issue dated 30 October 2013 – Approval of 17 May 2013)

 Non-target microorganisms (exclusivity): shall be around 1,000 to 100,000 times greater


than the minimum relative detection level of the alternative method. If the enrichment
medium is a selective broth, replace it with a non-selective broth When the alternative
method gives positive or doubtful results with non-target microorganisms, the test shall
be repeated using the complete protocol. The laboratory may carry out the reference
method in parallel if it so desires.
When there are discordant results between the alternative method and the reference
method, the laboratory shall investigate to explain the results.
All these results shall be recorded in the final report and their interpretation shall be the
subject of a comment by the expert laboratory.

5.1.2 Protocol to measure relative accuracy, sensitivity and


specificity
5.1.2.1 Selection and preparation of samples
It is important to find samples of water that are naturally contaminated. In a general manner,
all the categories/types for which validation is requested by the manufacturer must be
studied. As a consequence, if validation is requested for all categories, they must all be
studied. Please refer to Annex 1. Classification by categories of water.
When analysing naturally contaminated samples, the levels and distribution of contamination
of the samples should be representative of the levels usually found in these classes of water.
If it is not possible to acquire a sufficient number of naturally contaminated samples for each
of the categories, artificial contamination is permissible. The spiking method and levels
targeted should as far as possible result in samples behaving similarly to naturally
contaminated ones (for example, choose a low level of contamination for production or tap
water). Thus, microorganisms used for spiking shall in particular have been stressed, a
description of which is included in the final report.
Stress applied to microbial suspensions used for spiking artificial samples shall be relevant to
the type of water concerned and the strain chosen. The types of stress applicable by water
type are defined in Annex 1.
The types and modalities of stress, as well as the intensity of stress applied (a minimum of
0,5 log) shall be included in the expert laboratory's preliminary study report.
The total number of samples to be analysed is 60 per category of water. It is desirable to
obtain approximately 50 % positive results and 50 % negative results. For each category of
water, all the types of water listed in Annex 1. Classification by categories of water shall be
tested, with equitable distribution of the samples between the types representative of the
scope of the method being validated.
The reference and alternative methods shall always be implemented from two test portions
from the same sample. Finally, each test portion is subject to the whole of the method's
procedure. However, the first phases of both methods are often common; in this case, the
two test portions are produced at the end of the final common phase (see Annex 3).
Comment With artificial contamination, directly inoculate the sample then homogenize it before
taking the test portions.

5.1.2.2 Interpretation of the results


Collect the pairs of results obtained with the alternative and reference methods in the form of
a table, like Table 3, using the conventions described in Table 1.

Page 10/47
NF148 - NF Mark VALIDATION (application to the analysis of water)
General validation protocol – Revision 2 (Issue dated 30 October 2013 – Approval of 17 May 2013)

Table 3. Calculation of the relative accuracy, relative sensitivity and the relative specificity
Relative Relative Relative
Category PA NA ND PD N accuracy N+ sensitivity N- specificity
AC SE SP
Water 1
Water 2

Total

For each category of water, calculate the following parameters:


PA  NA
AC  100  Relative accuracy
N
PA
SE  100  Relative sensitivity
N
NA
SP  100  Relative specificity
N
Discordant results can be examined by the PD and ND count, using the McNemar test for
example, as described in Annex F of the EN ISO 16140 standard.
When the values for PD and ND are high and almost equal, no statistical difference between
the two methods can be detected using the McNemar test. In this case, the expert laboratory
shall pay particular attention to the reasons for these high values of PD and ND.

5.1.2.3 Interpretation
Taking into account the number of positive deviations and the number of negative deviations,
the capability of the alternative method to give more or fewer positive results than the
reference method is evaluated.
The study report shall distinguish the results obtained with naturally contaminated and
artificially contaminated samples.
By way of illustration of the values of the validation criteria, Figure 1 presents 10 extreme
cases in order to understand how to interpret the results better. The left-hand scale gives the
counts for 100 samples, and the right-hand scale gives the values of the criteria in %.

Page 11/47
NF148 - NF Mark VALIDATION (application to the analysis of water)
General validation protocol – Revision 2 (Issue dated 30 October 2013 – Approval of 17 May 2013)

100 0 0 5 0 100
5 15 13 18 %
25 20
80 5 12 80%
12
N
br 60 25 80 80 60%
e
de 60
ré 40 40%
po
ns
es 20 0 20%

0 0%
1 2 3 4 5 6 7 8 9 10

PA NA ND PD Accuracy
Exactitud Sensitivity
Sensibilit Specificity
Spécificit

Figure 1. Example of the calculation of validation criteria for different situations.

5.1.3 Protocol to measure the relative detection level


5.1.3.1 Design
The relative detection level is defined in 3.2.7. To measure it, test in parallel with the
reference method and alternative method according to the following design. For each
category of water, choose a strain that is normally present. However, ensure that the strains
are diversified, so that two categories of water are not associated with the same strain.
Initially determine the level of flora contamination associated with each category of water
chosen.
For each "strain/category" combination, independently prepare 6 spiked samples for a
minimum of 4 levels of contamination, including a negative control; that is, a total number of
at least 24 spiked samples per combination. The choice of levels of contamination follows the
following recommendation:
1. Level 1 corresponds to the absence of contamination.
2. Level 2 corresponds to contamination that gives less than 50 % positive results
(≤ 3/6) with one of the two methods.
3. Level 3 shall give over 50 % positive results.
4. Level 4 shall be such that 100 % of results are positive.

Comment The expert laboratory may produce an inoculum level of less than 1 cell per test
portion. Low contaminations are controlled by precise calibration and experience acquired with the
strains tested.

Depending on the methods, arrange the measurements according to the recommendations


in Annex .

Page 12/47
NF148 - NF Mark VALIDATION (application to the analysis of water)
General validation protocol – Revision 2 (Issue dated 30 October 2013 – Approval of 17 May 2013)

5.1.3.2 Calculation and interpretation


For each level of contamination and each "category/strain" combination, estimate the relative
detection level setting the limit point to 50 %, called LOD50. This calculation is carried out
according to the Spearman-Kärber method which also supplies an associated confidence
interval. LOD50 can also be interpreted as the level of contamination that theoretically
corresponds to a recovery rate of 50 %.
The method's overall relative detection level is expressed by taking the minimum and
maximum of all the confidence intervals of the various LOD50.

5.1.4 Practicability
The following 13 criteria are to be completed by the expert laboratory.

Table 4. List of practicability criteria identified to characterise the alternative method


Communication
Criteria to check concerning the Method of checking the criteria
criteria
1 Procedure for conditioning Packaging or Verification by the expert laboratory
the elements of the method instructions

2 Reagent volume Packaging or Verification by the expert laboratory


instructions

3 Conditions of storage of the Packaging or Verification by the expert laboratory that


elements (expiry date for instructions the conditions exist
unopened products)

4 Modalities of use after the Packaging or Verification by the expert laboratory that
first use (expiry dates for use) instructions the modalities exist

5 Specific equipment or Instructions Verification by the expert laboratory of the


premises required veracity of the written information

6 Reagents ready-to-use or to Packaging or Verification by the expert laboratory of the


be reconstituted (in this case, instructions veracity of the written information
is there a procedure)

7 Period required to train an Report Measured by the expert laboratory (option


operator not initiated into the of using periods implemented by
method collaborating laboratories) and allocated
into one of the following 3 categories: less
than 1 day, between 1 day and a week,
more than a week.

8 Real-time handling and Report Handling time measured in comparison


flexibility of the method with the reference method: less than,
depending on the number of equal to or greater than the reference
samples to be analysed. method handling time.

Page 13/47
NF148 - NF Mark VALIDATION (application to the analysis of water)
General validation protocol – Revision 2 (Issue dated 30 October 2013 – Approval of 17 May 2013)

Communication
Criteria to check concerning the Method of checking the criteria
criteria
9 Time required for obtaining Report and 2 cycles are established describing each
the results certification phase of the method only in terms of
time: 1st cycle: negative samples- 2nd
cycle: positive samples

10 Operator qualification type Report Specified by the expert laboratory as


compared with the level required for the
reference method: level identical to, or
different from, that required for the
reference method (the expert laboratory
may use data from the collaborating
laboratories)

11 Phases shared with the Report Verification by the expert laboratory


reference method
12 Means or traceability of the Instructions Verification by the expert laboratory
analysis results for the user
13 Obligation to maintain Report Period and frequency
specific apparatus for the
user

5.2 Quantitative methods


5.2.1 Principle and definitions
In the case of quantitative methods, the purpose of the preliminary study is to measure the
performance of the alternative method, and if need be, the reference method for the following
parameters:
 selectivity, specificity (inclusivity/exclusivity)
 linearity, relative accuracy
 detection limit, quantification limit
 practicability

5.2.2 Protocol for the assessment of inclusivity and exclusivity


(selectivity)
To study the selectivity of the alternative method, analyse the strains in the list in Table 5.
This protocol does not apply to non-selective methods (for example, counting revivifiable
aerobic germs).

Page 14/47
NF148 - NF Mark VALIDATION (application to the analysis of water)
General validation protocol – Revision 2 (Issue dated 30 October 2013 – Approval of 17 May 2013)

Table 5. Minimum number of strains for the selectivity study


Types of strains
Target strains (species) (for example E. coli or Legionella
20
pneumophila)
Inclusivity
Target strains belonging to a genus, a group or a family (for example
30
coliforms, Legionella spp., or intestinal enterococci).
Non-target strains known to cause interference with target strains
20
(genus-species).
Exclusivity
Non-target strains known to cause interference with target strains
30
belonging to a group or a family.

Choose the target and non-target strains that regularly contaminate the class or classes,
and/or categories of water selected. In addition, strains shall if possible come mainly from
natural samples representative of the water's envisaged scope. The origin of each strain
shall be known, recorded and traceable.
The degree of inoculation of each microorganism is defined as follows:
 Target microorganisms (inclusivity): count a suspension of the target germ by the
alternative method and in a non-selective medium. If the principle of the reference
method is different from that of the alternative method, both methods are implemented in
parallel.
 Non-target microorganisms (exclusivity): use non-target germ concentrations around 104
to 105 CFU / volume analysed.
When there are discordant results between the alternative method and the reference
method, the laboratory shall investigate to explain the results.
All these results shall be recorded in the final report and their interpretation shall be the
subject of a comment by the expert laboratory.

5.2.3 Protocol to measure linearity and relative accuracy


NOTE Validation applies to the whole range of the alternative method. The levels of contamination
chosen shall include the regulatory limits, if such exist, and the maximum range. If the calculation of
the limit of quantification is high, a level once the assumed LOQ and another twice the assumed LOQ
should be set.

5.2.3.1 Linearity
To check the linearity, for each category of water studied as described in Annex 1, prepare J
samples of known concentration, such as reference materials or spiked natural samples. The
number J is called the number of levels and it depends on the scope envisaged by the
alternative method and shall be ≥ 3.
At each level and for each category of water, carry out k complete and independent
replications with each method (k ≥ 2). Note yjk the result obtained with the alternative method
and xjk the result obtained with the reference method. These two measurements are paired.
The linearity can generally be studied graphically by transferring the pairs of values obtained
with the 2 methods (reference and alternative) onto a graph. If need be, the straight line
y jk  a0  a1 x jk that passes between the points of the graph can be calculated with a linear
regression method. Its slope a1 shall be close to 1 and its y-intercept a0 close to 0. However,
it is not advantageous to carry out statistical compliance tests.

Page 15/47
NF148 - NF Mark VALIDATION (application to the analysis of water)
General validation protocol – Revision 2 (Issue dated 30 October 2013 – Approval of 17 May 2013)

The raw values obtained for the samples analysed are presented in the form of two-
dimensional graphs by categories of water and possibly by type of water if necessary. The
counting results per category are presented on two-dimensional graphs as log N and as
CFU.

5.2.3.2 Relative accuracy


For each category of water tested, a minimum of 20 samples are analysed in duplicate with
both methods, that is, 80 analyses in all per category, with homogenous distribution for each
type of water and a minimum of 6 to 7 samples per type.
The stress applied to microbial suspensions used for spiking artificial samples shall be
relevant to the type of water concerned and the strain chosen. The types of stress applicable
by water type are defined in Annex 2.
The types and modalities of stress, as well as the intensity of stress applied (a minimum of
0.5 log) shall be included in the expert laboratory's preliminary study report.
The relative accuracy between the alternative method and the reference method is evaluated
by calculating the differences d jk  y jk  x jk for each pair of values. For each level j, group
the values of djk and calculate a relative accuracy for the level via the median:
Di  mediane(d jk )
The relationship between the relative accuracies Dj and the values of the levels (xj) may be
modelled. A null value (D = 0) is expected for each level. If the values D are too large for all
or part of the scope, it may be concluded that the alternative method differs too much from
the reference method and does not merit further validation. In addition, the values of D can
be used to start to define a value for the limit of acceptability (λ) that will be used to interpret
the accuracy profile (section 6.2.1). However, the repeatability and purpose of the method
must also be borne in mind when fixing the value λ.
The raw values obtained for the samples analysed are presented in the form of two-
dimensional graphs by categories of water and possibly by type of water if necessary. The
counting results per category are presented on two-dimensional graphs as log N and as
CFU.
When there are discordant results between the alternative method and the reference
method, the laboratory shall investigate to explain the results.

5.2.4 Detection and quantification limits


NOTE The calculation shall take the type of matrix and regulatory threshold values into
consideration.

The detection and quantification limits are checked, according to the manufacturer's
recommendations and the reference method, in accordance with 6.2.2.4 of EN ISO 16140.
The detection and quantification limits are determined by analysing a pure culture of the
target strain. As a minimum, 6 replicates for each level are tested in sterile water. As a
minimum, 3 to 4 levels shall be tested (in addition to level 0), of which, at least one level with
less than 50 % positive responses and one level with more than 50 % positive responses.

5.2.5 Practicability of the alternative method


The 13 criteria to be completed by the expert laboratory are listed in Table 4.

Page 16/47
NF148 - NF Mark VALIDATION (application to the analysis of water)
General validation protocol – Revision 2 (Issue dated 30 October 2013 – Approval of 17 May 2013)

6 Interlaboratory study
The interlaboratory study shall be used to evaluate the performance of the method in several
laboratories under real conditions that represent its routine application. It is organized under
the control of the expert laboratory and generally follows the preliminary study. Its
interpretation is "method" oriented and not "laboratory" oriented.

6.1 Qualitative methods


6.1.1 Design
The interlaboratory study applicable to qualitative methods must involve at least 8
collaborating laboratories having results without outliers.
The collaborating laboratory must demonstrate its competence in the use of the alternative
method and of the reference method prior to participating in the study.
- one relevant water matrix is used to prepare the test samples.
- each sample shall be individually inoculated.
- at least three different levels of contamination shall be used: a negative control (L0),
one level slightly above the relative detection level of the alternative method (L1), and
one level about 10 times greater than the detection level (L2) (for example 0, 3, and 30
cells/25 mL).
- the stability of the microorganism in the water matrix used shall be demonstrated.
- blind replicates at each level of contamination are analysed with the reference method
and with the alternative method in each of the collaborating laboratories, that is X=24
samples analysed per laboratory.
- if the alternative method and the reference method do not have a common phase, the
expert laboratory prepares and sends the 24 samples to each collaborating laboratory
in duplicate (2X).
- the samples shall be analysed in each laboratory on the date stipulated.
- the temperature of the control sample shall not exceed 10 °C during transport, until the
samples are received by the collaborating laboratory. A temperature recorder must be
used to record changes in temperature during shipment (for example to create heat
profiles).
The expert laboratory using all recorded data shall determine which results are suitable and
which are outliers for use in calculating the precision data.

6.1.2 Calculations and interpretation


For each level, put the positive results obtained with each method as illustrated in

Page 17/47
NF148 - NF Mark VALIDATION (application to the analysis of water)
General validation protocol – Revision 2 (Issue dated 30 October 2013 – Approval of 17 May 2013)

Table 6 and Table 7.

Page 18/47
NF148 - NF Mark VALIDATION (application to the analysis of water)
General validation protocol – Revision 2 (Issue dated 30 October 2013 – Approval of 17 May 2013)

Table 6. Positive results obtained by the reference method


Laboratories Contamination level
L0 L1 L2
Laboratory 1 /8 /8 /8
Laboratory 2 /8 /8 /8
Laboratory 3 /8 /8 /8
Etc. /8 /8 /8
Total FP TP1r TP2r
FP: false positive by the reference method
TP1r: true positive at level 1 by the reference method
TP2r: true positive at level 2 by the reference method

Table 7. Positive results obtained by the alternative method


Laboratories Contamination level
L0 L1 L2
Laboratory 1 /8 /8 /8
Laboratory 2 /8 /8 /8
Laboratory 3 /8 /8 /8
Etc. /8 /8 /8
Total FP TP1r TP2r
FP: false positive by the alternative method
TP1r: true positive at level 1 by the alternative method
TP2r: true positive at level 2 by the alternative method

For level L0 and for each method, calculate the percentage specificity SP as in the following
equation:
 FP 
SP  100  1  
 N 
Where N- is the total number of all tests L0.
FP is the number of false positives.
For each positive contamination level (L1 and L2), calculate the percentage sensitivity SE
using the following equation:
 TP 
SE  100   
 N
Where N+ is the total number of L1 or L2 tests respectively.
TP is the number of true positives.

For each level of contamination and the totality of the results, compare the alternative
method and the reference method in order to calculate the relative accuracy and to examine
the discordant results.
Each pair of results from a sample measured by the alternative and the reference methods
shall be reported as in Table 8.

Page 19/47
NF148 - NF Mark VALIDATION (application to the analysis of water)
General validation protocol – Revision 2 (Issue dated 30 October 2013 – Approval of 17 May 2013)

Table 8. Paired results of the alternative method and the reference method in the context of
the collaborative study.
Alternative method Reference method
Total
+ -
+ PA PD
- ND NA
Total N+ N- N

Calculate the relative accuracy AC expressed in percentage, as in the following equation:


 PA  NA 
AC  100   
 N 
Calculate the confidence intervals for each proportion (see 5.1.2.2.2 of EN ISO 16140).
Examine the discordant results (see 5.1.2.2.3 of EN ISO 16140).

6.1.3 Interpretation
Compare AC, SE and SP with their corresponding values, obtained in the context of the
preliminary study. These criteria do not really address the variability of the method within a
laboratory and between laboratories.
The degree of agreement, concordance and odds ratio criteria are used to help address this
notion of variability (repeatability and reproducibility) (cf. Annex 5 of this document).

6.2 Quantitative methods


6.2.1 Accuracy profile: general principles
The accuracy profile allows a statistical approach to validation to be used. It is based on the
direct application of the principles described in the standards in the ISO 5725 series, parts 1,
2 and 3. These standards describe a statistical method for estimating the accuracy (trueness
and precision) of a method or results. In this case, the model applied to explain a
measurement z of a measurand Z from a laboratory is of the type:
z  m Be
where m is the general mean of the homogenous sample sent to the laboratories, B the
laboratory's bias component under conditions of repeatability and e the random error
occurring in each measurement under conditions of repeatability. The bias component
comes from the laboratory, in the case of an interlaboratory study, but it can also come from
another source of uncertainty, like the day, the operator, the instrument, etc. in the case of an
intralaboratory study. The best known application of the ISO 5725 standard is the calculation
of precision criteria in an interlaboratory study that results in standard deviations for
repeatability and reproducibility. Apart from methods for the calculation of accuracy criteria,
these standards also specify the organization of collection in detail and the precautions to be
observed.
Based on the reproducibility standard deviation calculated according to the recommendations
of the ISO 5725 standard, a so-called tolerance (or prediction) interval is constructed which
contains a proportion β of future results. It is important to note that the tolerance interval
shall not be confused with the confidence interval. A confidence interval describes the
dispersion of results around a parameter, such as the mean, based on past experiments.
Thus the confidence interval for the mean is expressed as follows, with kIC its coverage factor

Page 20/47
NF148 - NF Mark VALIDATION (application to the analysis of water)
General validation protocol – Revision 2 (Issue dated 30 October 2013 – Approval of 17 May 2013)

and s x the standard deviation from the mean:

x  kIC  sx
A tolerance interval describes the dispersion of future results. It takes account of past
experiments and uncertainty concerning the mean. It is expressed with kIT another coverage
factor calculated in a completely different manner from kIC and sR, the reproducibility standard
deviation:
x  kIT  sR
This calculation is repeated for each level of concentration j to obtain a series of tolerance
intervals which, together, cover the whole of the method's scope and form what is called the
accuracy profile.
Figure 2 gives an example of an accuracy profile constructed with 3 levels of
concentration, between 2.2 and 4.2 log units, which define the validation range or scope of
the method to validate.
The horizontal axis represents the theoretical concentration of the levels and the vertical axis
the difference between the theoretical concentration and the concentration found expressed
as a log, that is, the absolute bias. The tolerance interval limits define a range where a
proportion β of future results is located (in this case 80 %).
Finally, the accuracy profile may be compared to an interval of acceptability, defined
according to the method's purpose. The limits of the intervals of acceptability, recorded as
±λ, are also expressed by a log difference. The interpretation strategy is described in detail in
Annex 7. Example of an accuracy profile calculation. It is defined as informative since it calls
on a limit of acceptability λ, which depends on the method's context of use and the β
proportion chosen.
In the range marked by the broken vertical line the method is capable of producing a
proportion β of results between the limits of acceptability. The method is then said to be valid
in this range indicated by the double arrow. The scope represents the range initially chosen
in which to conduct validation.

Page 21/47
NF148 - NF Mark VALIDATION (application to the analysis of water)
General validation protocol – Revision 2 (Issue dated 30 October 2013 – Approval of 17 May 2013)

0.3
Limit of quantification

0.2

0.1
Bias (log unit)
0.0
2.0 2.5 3.0 3.5 4.0 4.5
-0.1

-0.2

-0.3 Validation range of the method

-0.4
Target value in log10 (cfu)

Figure 2. Accuracy profile calculated for 3 levels of concentration. Each grey circle
represents the average concentration found for the level and quantifies the
trueness. The dotted lines mark the interval of acceptability and the continuous
lines the tolerance interval calculated from the intermediate precision standard
deviations for each level. The vertical lines mark the range of validity, in which the
method is capable of producing a known proportion (80 %) of acceptable results.

The operations leading to the construction of the accuracy profile may be summarised as
follows:
1. Organise the collection of data according to the design described in Section 6.2.2.
2. Transform the counts into log10.
IMPORTANT NOTE Present the accuracy profiles in log form and as a number of
microorganisms, except in specific cases validated by the Technical Board.
3. For each level (low, medium and high) calculate the reproducibility standard deviation
with Z replicates of the alternative method, according to ISO 5725.
4. For each level, calculate the median of X replicates of the reference method; these
medians are the target values that shall be achieved by the alternative method, taking
the acceptability limits into account.
5. For each level, calculate the mean of Z replicates of the alternative method.
6. For each level, calculate the tolerance interval limits according to the publication's
formulae (Hubert, 2006). Remember that the tolerance interval defines a range where
microbiologists can guarantee they can produce a known percentage (80 % in the
example) of future measurements.
7. Summarise all the calculations in a table that combines data from Table 10 and from
Table 11 and a figure designed as in Figure 2.

Page 22/47
NF148 - NF Mark VALIDATION (application to the analysis of water)
General validation protocol – Revision 2 (Issue dated 30 October 2013 – Approval of 17 May 2013)

6.2.2 Design
6.2.2.1 Organisation of the interlaboratory study
The interlaboratory study is used to evaluate the alternative method in relation to the
reference method. It shall be organized by the expert laboratory which selects a sufficient
number of laboratories (preferably at least 10) to obtain a minimum of 8 laboratories with no
measurement rejected as an outlier. It is based on sending samples which will be analysed in
duplicate and in parallel by the alternative method and the reference method. The purpose of
these tests is to calculate the performance criteria for the alternative method which will be
used to construct the accuracy profile and make a decision concerning its validity. Guidelines
and requirements for organising, dispatching and conducting interlaboratory studies are
given in Annex 8.
The expert laboratory is responsible for the preparation of the test protocol and the
measurement data sheet for recording all measurements and critical experimental conditions
used by each laboratory. The analyst in each participating laboratory shall demonstrate
his/her competence in the use of the alternative method and of the reference method prior to
participating in the study.
The protocol is the following:
 Choose a relevant matrix generally defined after the preliminary study (see 6.2.2.3).
 Choose the levels of concentration for artificially or naturally contaminated samples
and which cover at least the lower, middle and upper levels of the expected scope
of the alternative method. A negative control may also be included but the
measurements carried out are not included in the calculations.
 The expert laboratory shall demonstrate the homogenous nature and stability of the
three concentration levels tested.
 Use the same samples with each method.
 The organising laboratory prepares 4 aliquots from each level (that is, two aliquots
each measured by both methods) for each laboratory. These aliquots are coded
blind, such that two are measured by the reference method and two measured by
the alternative method. In some special cases, subdivide the aliquots, just before
measurement with the two methods (see Annex 3. Organisation of analyses carried
out in parallel depending on the type of method).
 Each participating laboratory shall analyse the aliquots on a stipulated date using
common batches of media, reagents and/or kits.
 The expert laboratory shall define the rules for rounding results.
 The expert laboratory pools the laboratories' results and sorts them by level j and by
laboratory i as shown in Table 9.

Page 23/47
NF148 - NF Mark VALIDATION (application to the analysis of water)
General validation protocol – Revision 2 (Issue dated 30 October 2013 – Approval of 17 May 2013)

Table 9. Presentation of the results of the interlaboratory study


Reference method X Alternative method Z
Laboratory i Level j
Replicate 1 Replicate 2 Replicate 1 Replicate 2
Lab 1 1 x111 x112 z111 z112
Lab 2 1
Labo i 1
Labo I 1

Labo i J
Labo I J

6.2.2.2 Choice of the numbers of levels, participating laboratories and


replicates
The minimum conditions consist of choosing a number of levels J = 3 and a number of
replicates K = 2. It is essential that J ≥ 3 because the linearity between the reference values
and the values found must be checked. Three levels are needed for this verification. For the
value of the number of laboratories I, reference can be made to part 1 of ISO 5725. To
compensate for low values of J and K, it was decided to take I  8 laboratories. Preferably,
these laboratories shall be accredited to EN ISO 17025 but above all have good experience
of microbiological methods and finally accept the participation and quality assurance
(traceability) rules described by the organising laboratory.
Finally, the number of measurements collected shall therefore be at least equal to
8223 = 96.

6.2.2.3 Type of samples


The interlaboratory study only covers a single matrix. The choice of category to which it
belongs shall be defined following the preliminary study. Concentration levels may be
prepared by spiking in order to achieve the values required.

6.2.3 Statistical calculations


NOTE A laboratory's results shall only be excluded as an exception and this shall be
systematically documented (enquiry carried out by the organising laboratory with the collaborating
laboratory).

Two Excel files are provided in a controlled manner by AFNOR Certification and shall be
used to construct the accuracy profile, one for the interpretation of the CFU data and the
other for the Log data.

6.2.3.1 Verification of linearity


Before any computer processing, it is always a good idea to make a prior illustration in graph
form and carry out a visual examination of the data in order to detect any obvious errors,
such as incorrectly recorded data. These graphs may be placed in the appendix of the
assessment file. The presence of abnormal data leads to the suspicion of incorrect
development of the method and may invalidate the validation procedure or a particular
laboratory.

Page 24/47
NF148 - NF Mark VALIDATION (application to the analysis of water)
General validation protocol – Revision 2 (Issue dated 30 October 2013 – Approval of 17 May 2013)

In microbiology, the measurements X and Z are not always distributed according to normal
distribution, and in particular a high level of asymmetry of distribution can be observed. Often
in order to get a more symmetric distribution, counts are better transformed into logarithms.
Other transformations may also be used for the small numbers found in water microbiology.

6.2.3.2 Calculation of target reference values per concentration level


Use the data X from Table 9 to calculate the target reference value per level. The simplest
solution consists of using the median:
x j  mediane( xi.k ) (1)

These medians are used to calculate the trueness of the method and the accuracy of the
measurements. It can be seen that it may not be an advantage to carry out as many
replicates with the reference method, in as much as they are not always used as such.

6.2.3.3 Calculation of precision standard deviations per concentration level


Use the data Z from Table 9 to calculate the repeatability sr, inter-series sB and reproducibility
sR standard deviations. This calculation is carried out independently for each concentration
level j (with 1 ≤ j ≤ 3) according to the principle of ISO 5725-2 described in Annex 6. Collect
the data in a table in the form of Table 10.

Table 10. Precision and trueness criteria per level

Level 1 … J
Number of laboratories I
Number of replicates K
Average theoretical target concentration xj
Average concentration found zj
Repeatability standard deviation s( j )r
Inter-series standard deviation s( j ) B
Intermediate precision standard deviation s( j ) R  s( j ) r2  s( j ) 2B
Absolute mean bias z j  xj

Obtaining reliable estimates of the mean or the standard deviation can be problematic in the
presence of outliers. ISO 5725 proposes tests for discarding outlier means or variances
(Grubbs and Cochran tests respectively) which can be used to exclude values with a strong
influence and so obtain a better estimate. This nevertheless reduces the number of values
that can be used for statistical analysis and is not recommended for this type of study.

6.2.3.4 Calculation of the tolerance interval


Reminder. The tolerance interval defines an interval in which there is an expectation of
finding on average a proportion β of future results obtained by applying the method routinely,
that is under the conditions of precision applied (in this case reproducibility). For this reason,
this is also called the prediction interval.

Page 25/47
NF148 - NF Mark VALIDATION (application to the analysis of water)
General validation protocol – Revision 2 (Issue dated 30 October 2013 – Approval of 17 May 2013)

The method proposed by Mee [Mee, 1984] is the one chosen for the protocol. It has also
been the subject of several publications [Hubert, 2003, Hubert 2006a, Hubert, 2006b]. The
calculation is carried out using data Z from Table 10 for each level of concentration j (with 3 ≤
j ≤ J). In order to simplify the formulas, the index j has been omitted from the following
formulae.
The tolerance interval is symmetrical around the average concentration found z and is
expressed by the formula:
z  ktol  sR (2)

where ktol is a coverage factor equal to:

1
ktol  t 1  1 (3)
,
2
KIB 2
with:

R 1
B (4)
IR  1
sB2
R . (5)
sr2
In addition, t 1  is the quantile of the Student t distribution for ν degrees of freedom,
;
2
calculated according to an approximation method proposed by Satterthwaite [Mee, 1984].

 R  1
2

 2
(6)
 1 1
 R   1
 I
 I
K 1 KI
For each level j, calculate the coverage factor ktol and the lower and upper limits of the
tolerance interval. The value chosen for β shall be at least 80 %. All the calculations are
collected in a table in the form of Table 11.

Table 11. Tolerance interval limits per level


Level 1 … J
Average theoretical concentration x
Lower tolerance limit z  ktol  sFI
Upper tolerance limit z  ktol  sFI
Differential lower tolerance limit  z  ktol  sFI   x
Differential upper tolerance limit  z  ktol  sFI   x
NOTE The number ν is rarely an integer and Student tables which accept non-integer degrees of
freedom should be used. However, the quantile value can be approximated via linear interpolation
between the 2 integer numbers of degrees of freedom that surround v.

Page 26/47
NF148 - NF Mark VALIDATION (application to the analysis of water)
General validation protocol – Revision 2 (Issue dated 30 October 2013 – Approval of 17 May 2013)

sB2
The value of ν depends on the relationship R  between the inter-series and repeatability
sr2
variances.
The relationship expresses the relative importance of the series effect. For example, when
the day effect is the series effect, if the method is very stable from one day to the next and
can supply very close results for the same sample, the correction introduced by R will be low.
By contrast, it will increase with this relationship. In fact, the lower the number of degrees of
freedom, the higher the Student t distribution quantile and the more the tolerance interval is
extended.
The following table illustrates this mechanism for a design V with K = 3 and l' = 3 and sFI
constant and equal to 1. When the variance ratio R increases from 0 to 9, the coverage
factor ktol and the scope of the tolerance interval increase from 1.97 to 3.07.

Table 12. Influence of the variance ratio on the scope of the tolerance interval (p = 0.90)
t
K I’ sr sB sFI R ν ;
1 
2
ktol
3 3 1.00 0.00 1.0 0.0 7.714 1.869 1.970
3 3 0.71 0.71 1.0 1.0 4.154 2.109 2.332
3 3 0.58 0.82 1.0 2.0 3.219 2.290 2.569
3 3 0.50 0.87 1.0 3.0 2.842 2.408 2.722
3 3 0.45 0.89 1.0 4.0 2.642 2.489 2.826
3 3 0.41 0.91 1.0 5.0 2.518 2.549 2.902
3 3 0.38 0.93 1.0 6.0 2.434 2.594 2.959
3 3 0.35 0.94 1.0 7.0 2.374 2.629 3.004
3 3 0.33 0.94 1.0 8.0 2.328 2.658 3.041
3 3 0.32 0.95 1.0 9.0 2.292 2.681 3.070

6.2.4 Construction of the accuracy profile


The accuracy profile may be constructed in different ways, depending on the type of data
being processed. The most traditional method, when relative concentrations are involved, is
the one shown in Figure 2 where the results are expressed by the coverage rate or relative
bias. But if the data is a matter of counts expressed in logarithms, it is better to use the
absolute bias in as much as a difference between 2 quantities expressed by their logarithms
is equivalent to their ratio.
To construct the accuracy profile select the following 5 lines from Table 10 and Table 11:
1. Average theoretical concentration (horizontal axis)
2. Average concentration found
3. Relative lower tolerance limit
4. Relative higher tolerance limit
5. Average coverage rate
Transfer the data to a graph using the average theoretical concentration to draw the x-axis.
Construct at least two profiles with β=80 % and β=90 %, for interpretation purposes.

Page 27/47
NF148 - NF Mark VALIDATION (application to the analysis of water)
General validation protocol – Revision 2 (Issue dated 30 October 2013 – Approval of 17 May 2013)

6.3 Summary of the interlaboratory test


6.3.1 Interpretation of the results
Interpret the graph using the following rules:
 Suggest a value for the acceptability criterion λ. Justify this choice according to the
problem being processed and prior experience. The final choice is the Technical
Board's responsibility.
 Depending on the value of λ, the alternative method is said to be valid in the whole
range where the tolerance interval is between the acceptability limits.
 Finally, define the quantification limit as the point where the tolerance interval
intersects one of the two acceptability limits. This is the limit beyond which the
microbiologist can no longer guarantee that a percentage β of the results obtained
by the alternative method are acceptable.

6.3.2 Limit of quantification


The quantification limit is calculated by simple linear interpolation between the points where
the tolerance interval intersects the acceptability limit. Figure 3 illustrates this calculation.

0.3

0.2

0.1
Bias (log unit) Limit of quantification
0.0
2.0 2.5 3.0 3.5 4.0 4.5
-0.1

-0.2

-0.3

-0.4
Target value in log10(cfu)

Figure 3. Calculation of the quantification limit via interpolation


The straight line formed by the lower limit of the tolerance interval (segment shown in bold)
will be considered. Assuming that its equation is formed using the values of the target levels
on the x-axis and the value of the lower limit of the tolerance interval on the y-axis, we have,
y = c0 + c1x

Page 28/47
NF148 - NF Mark VALIDATION (application to the analysis of water)
General validation protocol – Revision 2 (Issue dated 30 October 2013 – Approval of 17 May 2013)

Since the two pairs of values {x1, y1} and {x2, y2} are known, the slope and the y-intercept can
easily be calculated by rules of three and the value of the x-axis for the value –λ can be
predicted. When the intersection occurs with the upper limit, –λ is replaced by +λ:
y2  y1
c1 
x2  x1
c0  y1  c1 x1 (7)
  c0
LOQ 
c1
In the example shown in Figure 3, the values found are as follows:
Points 1 2
Value of the x-axis (target concentration) 2.267 3.230
Value of the y-axis (IT lower limit) -0.211 -0.150
Slope 0.0633
Origin -0.3546
Limit of acceptability (lambda) 0.2
LOQ 2.44
The LOQ, depending on the choices made for acceptability λ and the proportion β, is 2.4 log
units.
If there is no intersection, as shown in Erreur ! Référence non valide pour un signet..

Alternative method average


0.4 Lower tolerance limit
Upper tolerance limit
0.3 Lower limit of acceptability

A Upper limit of acceptability


c 0.2
c
u 0.1
r
a 0.0
c 1.5 2.0 2.5 3.0 3.5 4.0 4.5
y -0.1

-0.2

-0.3

-0.4
Log(cfu)

Figure 4, LOQ corresponds to the lowest value of the validation range.

6.4 Bibliography for section 6


Hubert P., Nguyen-Huu J.J., Boulanger B., Chapuzet E., Chiap P., Cohen N., Compagnon

Page 29/47
NF148 - NF Mark VALIDATION (application to the analysis of water)
General validation protocol – Revision 2 (Issue dated 30 October 2013 – Approval of 17 May 2013)

P.A., Dewe W., Feinberg M., Lallier M., Laurentie M., Mercier N., Muzard G., Nivet C.,
Valat L. (2003) Validation of quantitative analytical procedure, Harmonization of
approaches. Part I. STP Pharma Pratiques 13(3) 101-138.
Hubert P., Nguyen-Huu J.J., Boulanger B., Chapuzet E., Chiap P., Cohen N., Compagnon
P.A., Dewe W., Feinberg M., Lallier M., Laurentie M., Mercier N., Muzard G., Nivet C.,
Valat L. (2006a) Quantative analytical procedures: Harmonization of the approaches.
Part II – Statistics STP Pharma Pratiques 16(1) 28-58.
Hubert P., Nguyen-Huu J.J., Boulanger B., Chapuzet E., Chiap P., Cohen N., Compagnon
P.A., Dewe W., Feinberg M., Lallier M., Laurentie M., Mercier N., Muzard G., Nivet C.,
Valat L. (2006b) Quantitative analytical procedures validation: harmonization of the
approaches. Part III. Examples of application. STP Pharma Pratiques 16(2) 87-121.
Mee R.W. (1984) β-Expectation and β-Content Tolerance Limits for Balanced One-Way
ANOVA Random Model. Technometrics 26:3, 251-254

Page 30/47
NF148 - NF Mark VALIDATION (application to the analysis of water)
General validation protocol – Revision 2 (Issue dated 30 October 2013 – Approval of 17 May 2013)

Annex 1. Classification by categories of water

Water for human consumption:


Water from wells, springs and boreholes
Bottled water: spring water, mineral and other water
Mains and drinking fountain water

Bathing water (untreated):


Sea water
Fresh water

Swimming pool water

Industrial water:
Cooling tower water
Sanitary hot water
Waste water, raw and processed waste water

Others (Specific requests from manufacturers)


For example: thermal waters (spa water, ...)

Page 31/47
NF148 - NF Mark VALIDATION (application to the analysis of water)
General validation protocol – Revision 2 (Issue dated 30 October 2013 – Approval of 17 May 2013)

Annex 2. Stress to apply by type of water

Category Type Stress

Water for human Water from wells, springs none


consumption and boreholes

Bottled water: spring water, thermal (hot and/or positive


mineral and other water cold)

Mains and drinking fountain chlorination, thermal (hot


water and/or positive cold)

Bathing water Sea water thermal (positive cold),


(untreated) osmotic

Fresh water thermal (positive cold)

Swimming pool water chlorination, thermal (hot)

Industrial water Cooling tower water biocide

Sanitary hot water chlorination, thermal (hot)

Waste water, raw and chemical (acid and/or basic


processed waste water pH, etc.)

Others (Specific requests for example:


from manufacturers)
Thermal waters (spa water, chlorination, thermal (hot)
...)

Page 32/47
NF148 - NF Mark VALIDATION (application to the analysis of water)
General validation protocol – Revision 2 (Issue dated 30 October 2013 – Approval of 17 May 2013)

Annex 3. Organisation of analyses carried out in parallel


depending on the type of method

Methods with a common phase

Pure culture

Produce 6 independent
contaminations

25 mL matrix 25 mL matrix 25 mL matrix 25 mL matrix 25 mL matrix 25 mL matrix


+225 mL diluent +225 mL diluent +225 mL diluent +225 mL diluent +225 mL diluent +225 mL diluent

Alternative Reference Alternative Reference Alternative Reference


method method method method method method

Alternative Reference Alternative Reference Alternative Reference


method method method method method method

Page 33/47
NF148 - NF Mark VALIDATION (application to the analysis of water)
General validation protocol – Revision 2 (Issue dated 30 October 2013 – Approval of 17 May 2013)

Methods without common phases

Pure culture

Alternative method Reference method


Produce 6 independent contaminations Produce 6 independent contaminations

25 mL matrix 25 mL matrix 25 mL matrix 25 mL matrix 25 mL matrix 25 mL matrix


+225 mL diluent +225 mL diluent +225 mL diluent +225 mL diluent +225 mL diluent +225 mL diluent

25 mL matrix 25 mL matrix 25 mL matrix 25 mL matrix 25 mL matrix 25 mL matrix


+225 mL diluent +225 mL diluent +225 mL diluent +225 mL diluent +225 mL diluent +225 mL diluent

Méthode alternative Reference method

Page 34/47
NF148 - NF Mark VALIDATION (application to the analysis of water)
General validation protocol – Revision 2 (Issue dated 30 October 2013 – Approval of 17 May 2013)

Annex 4. Qualitative methods. Test applied to the rejection


of discordant results
Procedure
Count the total number of discordant results RD:
RD = PD + ND
For example, if PD = 2 and ND = 10, then RD = 12)
IF RD < 6 (less than 6 discordant results), no test is available.
IF 6  RD  22 (between 6 and 22 discordant results), determine m as the smallest of the
two values of PD and ND (for example m = PD = 2, because PD < ND) and use the
binomial law in accordance with Table 13.
If m  M (M defined in Table 13) for a given Y , the two methods are different at < 0.05
(two-sided).

Table 13. Values of M for Y discordant results (6  Y  22)

Discordant results RD = PD + ND 6-8 9-11 12-14 15-16 17-19 20-22


M = Max(m) with  < 0.05 0 1 2 3 4 5
For example, with Y = 12 discordant results and m = 2, M = 2 and m  M: therefore the two
methods are different with  < 0.05.
if Y > 22 (more than 22 discordant results), use the McNemar test with chi-square
distribution for 1 degree of freedom:
2 = d2/Y, with d = PD - ND and Y = PD + ND
The two methods are different for  < 0.05 (two-sided) if 2 > 3.841
This 2 test corresponds to the minimal d for each Y of the following Table 14 with
 < 0.05 (that is, for a given Y, d shall be greater than or equal to the value given in Table 14
for concluding that the two methods are different).

Table 14. Values of d for Y discordant results (Y > 22)

Discordant results Y= PD + ND 22-26 27-31 32-37 38-44 45-51 52-58


d = PD - ND  10 11 12 13 14 15

Page 35/47
NF148 - NF Mark VALIDATION (application to the analysis of water)
General validation protocol – Revision 2 (Issue dated 30 October 2013 – Approval of 17 May 2013)

Annex 5. Qualitative methods. Interlaboratory study,


criteria of accordance, concordance and concordance
odds ratio
Taken from Annex L of EN ISO 16140

General
The criteria of accuracy, sensitivity and specificity (see 5.2.2) do not really address the
variability of the method within and between laboratories (precision of the method).
This annex provides further criteria (accordance, concordance and odds ratio) which can
help to approach this variability.
Repeatability and reproducibility criteria measure the likely difference between two samples
sent to either the same or to different laboratories. Since the difference for data that are not
quantitative cannot be used, statistics for qualitative methods are instead based on the
probability (expressed as a percentage) that two identical samples both produce the same
result.
These criteria have been developed by the European SMT4 CT 96 2098 project funded by
the European Commission/ DG XII to validate the six main standardised methods used in
food microbiology (co-ordinator: Dr C. Lahellec, AFSSA, France).
The calculations are illustrated by means of the following data (see Table 15).

Table 15. Example of numerical values


Lab 1 2 3 4 5 Number of
positives
(out of 5)
1 + + + + + 5
2 + + + + + 5
3 + + + + + 5
4 + + + + + 5
5 - - + + + 3
6 + + + + + 5
7 - - + + + 3
8 + + + + + 5
9 + + + + + 5
10 + + + + + 5

The data shown in this table are just for one level of one food type. In practice, an
interlaboratory test would be larger, but a smaller dataset makes it easier to explain the
calculations.

Page 36/47
NF148 - NF Mark VALIDATION (application to the analysis of water)
General validation protocol – Revision 2 (Issue dated 30 October 2013 – Approval of 17 May 2013)

Accordance
Definition
Accordance is the percentage chance of finding the same result (i.e. both negative or both
positive) from two identical samples analysed in the same laboratory, under repeatability
conditions (i.e. one operator using the same apparatus and the same reagents within the
shortest feasible time interval).
Accordance is therefore the equivalent of repeatability for quantitative methods.
Calculations
To derive the accordance from the results of an interlaboratory study, the probability that two
identical samples give the same result is calculated for each participating laboratory in turn,
and this probability is then averaged over all the laboratories.
For laboratories (such as laboratory 1) where all the replicates were found to be positive, the
best estimate of the probability of getting the same result is clearly 1.00 or 100 %.
For the other laboratories (in the example, laboratories 5 and 7) the probability that one
replicate will be positive is 3/5=0.60. Square this probability to obtain the probability of a pair
of replicates being positive (0.62 = 0.36). Do the same for the probability of a pair of
replicates both being negative (0.42 = 0.16). Then add these two figures together to get the
overall probability that two replicates will give the same result (0.36 + 0.6 = 0.52). Do this for
all the laboratories (see Table 16).

Table 16. Calculation of accordance


Probability of
Number Probability of Probability of
Probability of Probability of a pair of
Lab of a pair of a pair of
positives a negative identical
positives Positives negatives
results
1 5 1.00 1.00 0.00 0.00 1.00
2 5 1.00 1.00 0.00 0.00 1.00
3 5 1.00 1.00 0.00 0.00 1.00
4 5 1.00 1.00 0.00 0.00 1.00
5 3 0.60 0.36 0.40 0.16 0.52
6 5 1.00 1.00 0.00 0.00 1.00
7 3 0.60 0.36 0.40 0.16 0.52
8 5 1.00 1.00 0.00 0.00 1.00
9 5 1.00 1.00 0.00 0.00 1.00
10 5 1.00 1.00 0.00 0.00 1.00
Average: 0.904 =90.4%
The accordance is the average of the probabilities that two replicates give the same result for
each laboratory: 90.4 % in this case.
NOTE Since repeatability measures difference whereas accordance measures similarity, high
values of accordance indicate a reliable method, in contrast to quantitative methods where low values
of repeatability are desirable.

Page 37/47
NF148 - NF Mark VALIDATION (application to the analysis of water)
General validation protocol – Revision 2 (Issue dated 30 October 2013 – Approval of 17 May 2013)

Concordance
Definition
Concordance is the percentage chance of finding the same result for two identical samples
analysed in two different laboratories.
Concordance is therefore the equivalent of reproducibility for quantitative methods.
Calculations
To calculate concordance from the results of an interlaboratory study, take in turn each
replicate in each participating laboratory, and pair it with identical results from all the other
laboratories.
Concordance is the percentage of all pairings giving the same results for all the possible
pairings of data.
FOR EXAMPLE (see Tables L.1 and L.3 of EN ISO 16140).
Take each replicate in each laboratory in turn, starting with the first replicate from laboratory
1 which is positive. This can be paired with any of the 45 replicates from the other
laboratories, and all but 4 of these pairings (those with two replicates from laboratories 5 and
7) match (i.e. a pair with two positives), 41 pairs thus give the same result.
The same applies to the other four replicates from laboratory 1, so there are a total of 225
(5 x 45) inter-laboratory pairings of replicates involving laboratory 1, of which 205 (5 x 41)
give the same result.
The same applies to all the other laboratories with all replicates found positive.
For laboratory 5, with 3 positive replicates out of 5, the 2 negative replicates each match with
just 2 other negative replicates from laboratory 7, whilst the 3 positive replicates each match
with 43 positive replicates. Thus the total number of pairs with the same result is 133 (2 x 2 +
3 x 43).

Table 17. Calculation of concordance


Number of Interlaboratory pairings with the Total number of interlaboratory
Lab
positives same result pairings
1 5 205 225
2 5 205 225
3 5 205 225
4 5 205 225
5 3 133 225
6 5 205 225
7 3 133 225
8 5 205 225
9 5 205 225
10 5 205 225
Total 1906 2250
Concordance is the percentage of all pairings of replicates giving the same result; in this
example this is 84.7 % (1906/2250 x 100).
NOTE Since reproducibility measures difference whereas concordance measures similarity, high
values of concordance indicate a reliable method, in contrast to quantitative methods where low
values of reproducibility are desirable.

Page 38/47
NF148 - NF Mark VALIDATION (application to the analysis of water)
General validation protocol – Revision 2 (Issue dated 30 October 2013 – Approval of 17 May 2013)

Concordance odds ratio


General remarks and definition
If the concordance is smaller than the accordance, it indicates that two identical samples are
more likely to give the same result if they are analysed by the same laboratory than if they
are analysed by different laboratories, suggesting that there can be a predominant variability
in performance between laboratories. This is the same situation as when reproducibility is
greater than repeatability for a quantitative method.
Unfortunately, the magnitude of the concordance and accordance is highly dependent on the
level of accuracy, making it difficult to assess the degree of interlaboratory variation easily.
It is therefore helpful to calculate the concordance odds ratio (COR) defined as follows:
deg ré d ' accord  100  concordance 
COR  (accordance/concordance)
concordance  100  deg ré d ' accord 

Significance tests
A value for the odds ratio of 1,00 would be expected if accordance and concordance were
equal, and the higher the odds ratio, the more interlaboratory variation is predominant.
Nevertheless, values above 1,00 can occur by chance variation, and so a statistical
significance test should be used to confirm whether the variation between laboratories is
actually predominant. The “exact test” is the one that is most recommended for this
purpose1. The philosophy behind such tests is that the probabilities of occurrence are
calculated for all combinations of replicate results that could have been obtained with the
overall numbers of positives and negatives.
FOR EXAMPLE (see Table L.1 of EN ISO 16140).
With a total of 46 positives and 4 negatives, the possible arrangements are given by columns
in Table L.4 of EN ISO 16140.

Table 18. Possible arrangements of positive replicates between laboratories (columns) to


give a total of 46 positives.
4 3 3 2 1
4 4 3 4 5
4 4 5 5 5
4 5 5 5 5
5 5 5 5 5
5 5 5 5 5
5 5 5 5 5
5 5 5 5 5
5 5 5 5 5
5 5 5 5 5

1) This test can be performed using statistical software such as SAS .

Page 39/47
NF148 - NF Mark VALIDATION (application to the analysis of water)
General validation protocol – Revision 2 (Issue dated 30 October 2013 – Approval of 17 May 2013)

The actual arrangement is shown in bold (third column) in Table M.4. The test adds up the
probabilities P of all the possible arrangements that show at least as much evidence for
interlaboratory variation as the real arrangement – here they correspond to the permutations
in the two columns on the right. If this sum of probabilities is less than the conventional value
of 0,05 or 5 %, it is unlikely that this degree of interlaboratory variation could have occurred
by chance and thus it is concluded that there is a significant variation in performance
between laboratories.
In the example of Table M.4, P = 0,039 indicates that the variation between laboratories is
significant at a level of 5 %.
Where “exact test” software is not available, an ordinary 2 analysis for the contingency
tables can be used as an alternative. The results of this test are less reliable than the “exact
test” with the number of replicates usually used in collaborative studies, but simulations
suggest that the results provide a reasonable guide to the significance of interlaboratory
differences.
With either test, it shall be remembered that the ability to detect interlaboratory differences is
dependent on the number of laboratories and the number of replicates analysed at each
laboratory. A non-significant test result should not be taken to mean that performance does
not vary between laboratories, but rather that such differences have not been proved; this is
particularly true where the P-value is only just above 0,05. The ideal solution would be to
quote the odds ratio with the standard deviation or confidence limits, but the distribution of
the odds ratio is highly skewed making it very difficult to estimate reliable limits.

Page 40/47
NF148 - NF Mark VALIDATION (application to the analysis of water)
General validation protocol – Revision 2 (Issue dated 30 October 2013 – Approval of 17 May 2013)

Annex 6. Quantitative methods. Calculation of repeatability


and reproducibility standard deviations
This annex is based on clause 7 of ISO 5725-2. The calculations are conducted according to
the principle described in ISO 5725. They are applied at each level j to data z. As a result,
the number of replicates per level is considered to be constant and equal to K.
Calculate for each level j:
 the repeatability variance, recorded as s(j)2r .
 the inter-series variance, recorded as s(j)2B .
 the intermediate precision variance s( j )2FI  s( j )2r  s( j )2B .

In order to simplify scoring, the index j for the level is deleted from the formulae that follow.
The starting point consists of writing the following equality which breaks down the sum of the
deviations from the general average into two sums of squared deviations. nk represents…
K I K K I

 ( z
k 1 i 1
ik z )2   ( zk z ) 2   ( zik zk ) 2
k 1 k 1 i 1
(8)

This equation is traditionally written in an abbreviated form, involving three sums of squared
deviations (SSD). General equation for variance analysis
SSDt = SSDB + SSDr
Where each of the sums is defined as follows to facilitate interpretation:
SSDt Sum of total deviations from the level's general mean
SSDB Sum of the inter-series deviations
SSDr Sum of the intra-series deviations

As explained in ISO 5725, the sums of squares are developed to make calculation easier;
SSDB is always calculated by difference. This method poses problems if the result is
negative: the value of SCEB is then forced to 0.
SSDB = SSDt – SSDr
if SSDB > 0
if SSDB ≤ 0 then SSDB = 0
2
NOTE An alternative to this method of calculation is to use an algorithm to estimate s B using the
restricted maximum likelihood method (REML) available in most statistical software.

Calculate the repeatability variance (intra-series) for the level:


I K
1 1
  z  zk  
2
sr2  SCEr (9)
( I  1) K ( I  1) K
ik
i 1 k 1

Calculate the inter-series variance in the following manner:

 SCEB 
( K  1)   sr2 
sB2   K 1  (10)
I
Finally the intermediate precision variance for the level is equal to: sR2  sr2  sB2 (11)

Page 41/47
NF148 - NF Mark VALIDATION (application to the analysis of water)
General validation protocol – Revision 2 (Issue dated 30 October 2013 – Approval of 17 May 2013)

Annex 7. Example of an accuracy profile calculation

Lab Series Level Alternative (cfu) Reference (cfu) Alter log. Ref log.
A 1 1 60 120 1.778 2.079
A 1 1 40 75 1.602 1.875
B 2 1 70 95 1.845 1.978
B 2 1 90 85 1.954 1.929
C 3 1 80 90 1.903 1.954
C 3 1 50 80 1.699 1.903
D 4 1 90 90 1.954 1.954
D 4 1 170 120 2.230 2.079
E 5 1 40 90 1.602 1.954
E 5 1 70 130 1.845 2.114
F 6 1 50 130 1.699 2.114
F 6 1 120 65 2.079 1.813
G 7 1 90 95 1.954 1.978
G 7 1 130 110 2.114 2.041
H 8 1 90 100 1.954 2.000
H 8 1 110 110 2.041 2.041
I 9 1 140 70 2.146 1.845
I 9 1 90 80 1.954 1.903
J 10 1 90 120 1.954 2.079
J 10 1 110 110 2.041 2.041
L 11 1 90 140 1.954 2.146
L 11 1 120 110 2.079 2.041
M 12 1 110 95 2.041 1.978
M 12 1 130 130 2.114 2.114
N 13 1 90 95 1.954 1.978
N 13 1 150 80 2.176 1.903
O 14 1 110 65 2.041 1.813
O 14 1 50 95 1.699 1.978
A 1 2 810 1300 2.908 3.114
A 1 2 520 1200 2.716 3.079
B 2 2 940 1000 2.973 3.000
B 2 2 950 1100 2.978 3.041
C 3 2 1000 1000 3.000 3.000
C 3 2 1100 1000 3.041 3.000
D 4 2 1000 1000 3.000 3.000
D 4 2 900 940 2.954 2.973
E 5 2 940 930 2.973 2.968
E 5 2 920 1100 2.964 3.041
F 6 2 810 1000 2.908 3.000
F 6 2 1000 1100 3.000 3.041
G 7 2 910 910 2.959 2.959
G 7 2 1000 900 3.000 2.954
H 8 2 1200 1100 3.079 3.041
H 8 2 1100 1000 3.041 3.000
I 9 2 1100 1000 3.041 3.000
I 9 2 980 1100 2.991 3.041
J 10 2 1200 950 3.079 2.978

Page 42/47
NF148 - NF Mark VALIDATION (application to the analysis of water)
General validation protocol – Revision 2 (Issue dated 30 October 2013 – Approval of 17 May 2013)

J 10 2 1100 1100 3.041 3.041


L 11 2 830 1000 2.919 3.000
L 11 2 1300 1000 3.114 3.000
M 12 2 1100 1000 3.041 3.000
M 12 2 830 1200 2.919 3.079
N 13 2 660 940 2.820 2.973
N 13 2 810 980 2.908 2.991
O 14 2 640 820 2.806 2.914
O 14 2 940 1200 2.973 3.079
A 1 3 9300 9700 3.968 3.987
A 1 3 9900 9700 3.996 3.987
B 2 3 7800 9400 3.892 3.973
B 2 3 9700 9000 3.987 3.954
C 3 3 11000 10000 4.041 4.000
C 3 3 11000 12000 4.041 4.079
D 4 3 9500 11000 3.978 4.041
D 4 3 10000 11000 4.000 4.041
E 5 3 11000 9600 4.041 3.982
E 5 3 9400 10000 3.973 4.000
F 6 3 8900 11000 3.949 4.041
F 6 3 9500 10000 3.978 4.000
G 7 3 12000 13000 4.079 4.114
G 7 3 13000 12000 4.114 4.079
H 8 3 9400 12000 3.973 4.079
H 8 3 9500 9400 3.978 3.973
I 9 3 11000 12000 4.041 4.079
I 9 3 12000 12000 4.079 4.079
J 10 3 12000 12000 4.079 4.079
J 10 3 11000 9500 4.041 3.978
N 13 3 9500 12000 3.978 4.079
N 13 3 10000 13000 4.000 4.114
O 14 3 9000 7600 3.954 3.881
O 14 3 9500 8500 3.978 3.929

Table 19. Results of the interlaboratory study


Criteria Level 1 Level 2 Level 3
Target value 1.95 2.97 4.00
Alternative method average 0.018 0.027 0.016
Lower tolerance limit -0.196 -0.117 -0.109
Upper tolerance limit 0.232 0.171 0.140
Lower limit of acceptability -0.3 -0.3 -0.3
Upper limit of acceptability 0.3 0.3 0.3

Construct the accuracy profile like the one illustrated in Erreur ! Référence non valide pour
un signet..

Page 43/47
NF148 - NF Mark VALIDATION (application to the analysis of water)
General validation protocol – Revision 2 (Issue dated 30 October 2013 – Approval of 17 May 2013)

Alternative method average


0.4 Lower tolerance limit
Upper tolerance limit
0.3 Lower limit of acceptability

A Upper limit of acceptability


c 0.2
c
u 0.1
r
a 0.0
c 1.5 2.0 2.5 3.0 3.5 4.0 4.5
y -0.1

-0.2

-0.3

-0.4
Log(cfu)

Figure 4. Accuracy profile The limits of acceptability have been chosen at ±0.3 log

To show the importance of the acceptability criterion,


Alternative method average
0.3 Lower tolerance limit
Upper tolerance limit
Lower limit of acceptability
0.2
Upper limit of acceptability

0.1
Accuracy

0.0
1.5 2.0 2.5 3.0 3.5 4.0 4.5
-0.1

-0.2

-0.3
Log(cfu)

Figure 5 illustrates the situation where the acceptability limits were chosen at ±0.2 log. The
LOQ is then 2.51 log units.

Page 44/47
NF148 - NF Mark VALIDATION (application to the analysis of water)
General validation protocol – Revision 2 (Issue dated 30 October 2013 – Approval of 17 May 2013)

Alternative method average


0.3 Lower tolerance limit
Upper tolerance limit
Lower limit of acceptability
0.2
Upper limit of acceptability

0.1
Accuracy

0.0
1.5 2.0 2.5 3.0 3.5 4.0 4.5
-0.1

-0.2

-0.3
Log(cfu)
Figure 5. Accuracy profile II. The limits of acceptability have been chosen at ±0.2 log

Page 45/47
NF148 - NF Mark VALIDATION (application to the analysis of water)
General validation protocol – Revision 2 (Issue dated 30 October 2013 – Approval of 17 May 2013)

Annex 8. Guidelines for the organisation and conducting of


interlaboratory studies
This annex is based on Annex H (normative) of EN ISO 16140.

Preparation of water samples


Negative (control) samples
The target germ shall not be present in the water matrix. Choose a suitable number of
samples in accordance with 6.1 or 6.2 in this document and analyse them to confirm this.
Negative controls shall also be used during tests where stipulated.
Positive samples containing the target germ.
In the case of qualitative and quantitative studies, samples shall be contaminated artificially
as shown in options 2 and 3 of Annex C of EN ISO 16140. Where reference materials are
used, it is the responsibility of the organising laboratory to send both the reference materials
and details of the inoculation procedure to each collaborating laboratory.
The sample and microorganisms shall be capable of being homogenised and should remain
stable both in transit and for the duration of the analyses. Homogeneity and stability studies
shall be conducted prior to despatching the test material from the organising laboratory.
The water sample should contain representative background microflora or interfering
components that shall also remain stable in transit and for the duration of the analyses.
An alternative method shall be tested using a minimum of one water type, chosen from the
categories defined. The quality and suitability of the water samples shall be determined prior
to the study.
Based on the comparative study, the characteristics of the chosen microbial strain being
used to validate a method shall be representative of the genus/species being sought, for
example growth rate, antigenic characteristics and sensitivity to inimical agents etc.

Transport of samples
The organising laboratory defines the type of water to be used and checks that the transit
packaging is suitable for the intended journey.
Individual samples shall be double sealed to ensure that leakage from any one sample does
not affect the integrity of all the other samples.
Each collaborating laboratory is responsible for supplying the organising laboratory with
relevant details that ensure that the distribution of samples for the study meets international
and national postal regulations.
Concerning refrigeration, the following conditions apply: the temperature of samples during
transport shall be lower than or equal to 10 °C, and, on arrival at the laboratory, between
0 °C and 10 °C.
The packaging conditions and best method of transport shall be determined by the
organising laboratory. An appropriate means of monitoring the samples during transit is
desirable. An additional pack, identical to the test samples, shall be provided in order to
measure their temperature upon receipt.

Organisation of the interlaboratory study


Operating protocols: operating procedures shall be distributed to collaborators for comment
familiarisation prior to the commencement of the test. A final protocol shall be issued in
advance of the evaluation.

Page 46/47
NF148 - NF Mark VALIDATION (application to the analysis of water)
General validation protocol – Revision 2 (Issue dated 30 October 2013 – Approval of 17 May 2013)

Confirmation of sample quality: each collaborating laboratory should be instructed to


undertake enumeration of the total bacterial flora in a specific additional sample prepared
by the expert laboratory. The organising laboratory shall provide each collaborator with a
standard enumeration method.
Alternative method and reference method: the collaborator responsible for carrying out the
analysis in each participating laboratory shall be trained in advance (if necessary) and shall
be fully conversant with the alternative method and the reference method prior to the
commencement of the test.
Reagents and operating conditions: certain additional factors such as the quality and
composition of the culture media and reagents, control of incubation temperatures, etc. can
have a profound effect on the outcome of the test. This variability shall be minimised (for
example, by dispatching media/reagents to all collaborators) or be taken into consideration
when interpreting data.
The degree of tolerance permitted for the different aspects of the germ (for example, times,
temperatures, masses, total plate count and day of analysis), shall be stated in the
protocol. A clear warning shall be issued if no tolerances are permitted.
Guidelines: the organising laboratory shall be available to provide advice or guidance to
collaborators during the whole period of the study.
Data collection: the organising laboratory shall prepare a questionnaire that can be used to
collect information concerning the critical points of the procedure. All collaborators should
be asked to record details such as media pH, times of incubation, incubator start and finish
temperatures, quality control data, state of samples on arrival, time of arrival, temperature
on arrival, storage condition/times, etc. The questionnaire shall be the subject of agreement
by both the organising laboratory and the manufacturer, in order to guarantee its suitability
for the critical phases of the test or test procedure of the alternative method.
Confirmation of the quality of test samples: representative aliquots of the samples should
be analysed by the organising laboratory on the day of the distribution in order to confirm
the presence and homogeneity of the target germ. Samples, previously held under ideal
conditions, should also be analysed by the organising laboratory on the day of the start of
analysis by the collaborating laboratories. These analyses shall be used only to confirm the
inoculation procedure and stability of the germ. The results should not be used in the
statistical evaluation of the method under test.
The fact of using branded media and different batches introduces additional variability.

Page 47/47

You might also like