0% found this document useful (0 votes)
66 views

Dta 100 C

Copyright
© © All Rights Reserved
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
66 views

Dta 100 C

Copyright
© © All Rights Reserved
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 90

User manual

Oil breakdown voltage tester


DTA 100 C
phd

B A U R G m b H ▪ R a i f f e i s e n s t r . 8 ▪ 6 8 3 2 S u l z , A u s t r i a
T +43 5522 4941-0 ▪ F +43 5522 4941-3 ▪ www.baur.eu ▪ [email protected]
Copyright © 2019
All rights reserved.
Reproduction, circulation in any form whatsoever, publishing on online services or the Internet, as well
as duplication on data carriers, even in part or in an amended format, is allowed only with prior written
permission of BAUR GmbH, 6832 Sulz, Austria.
We reserve the right in the interests of our customers to make amendments as a result of further
technical development. Illustrations, descriptions and scope of supply are therefore not binding.
The names of products and companies are the trademarks or brand names of the relevant companies.
DTA 100 C Table of contents

Table of contents

1 About this manual ....................................................................................... 6


1.1 Using this manual ................................................................................ 6
1.2 Applicability of the instructions ............................................................. 6
1.3 Structure of safety instructions ............................................................ 6
1.4 View Settings ....................................................................................... 7
1.5 Note on the screenshots and graphics used ....................................... 8
2 For your safety ............................................................................................. 9
2.1 Instructions to the user......................................................................... 9
2.2 Intended use ...................................................................................... 10
2.3 Avoid dangers, take safety measures................................................ 10
2.3.1 Dangers when working with electric voltage ........................ 11
2.4 Special personal protective equipment .............................................. 12
3 Product information .................................................................................. 13
3.1 Full illustration .................................................................................... 13
3.2 Operating and display elements ........................................................ 16
3.3 USB interface ..................................................................................... 17
3.4 Power supply ..................................................................................... 17
3.5 Rating plate ........................................................................................ 18
4 Technical data ............................................................................................ 19
5 Basic insulating oil test procedure .......................................................... 20
6 Operating the oil breakdown voltage tester............................................ 21
6.1 Menu navigation ................................................................................. 21
6.2 Entering numbers............................................................................... 21
6.3 Symbols and abbreviations on the display ........................................ 22
6.4 Main menu ......................................................................................... 23
6.5 Information on the oil breakdown voltage tester ................................ 24
6.5.1 Information on the USB interface ......................................... 24
7 Commissioning .......................................................................................... 25
7.1 Checks to perform before commissioning ......................................... 25
7.2 Installing the oil breakdown voltage tester ......................................... 26
7.3 Earthing the oil breakdown voltage tester.......................................... 26
7.4 Replacing the electrodes ................................................................... 27
7.5 Clean the electrodes .......................................................................... 28
7.6 Setting an electrode gap .................................................................... 28

822-129-8 iii / 90
Table of contents DTA 100 C

7.7 Fill and use the test vessel ................................................................ 29


7.7.1 Instructions for sampling ...................................................... 30
7.7.2 Insert the test vessel ............................................................ 31
7.8 Turning on the oil breakdown voltage tester ...................................... 32
8 Device settings .......................................................................................... 33
8.1 Set display brightness ........................................................................ 33
8.2 Select language ................................................................................. 34
8.3 Switch printer on/off ........................................................................... 34
8.4 Overwrite measurement logs ............................................................. 35
8.5 Configure the IEC 60156:2018 settings ............................................. 35
8.5.1 Setting the hold time before the first measurement ............. 36
8.5.2 Adjusting the stirrer .............................................................. 37
8.5.3 Setting the electrode shape ................................................. 38
8.6 Configure the ASTM settings ............................................................. 39
8.6.1 Display individual breakdown values ................................... 39
8.6.2 Set the hold time before the first measurement ................... 40
8.7 Configure settings for user-defined measurements ........................... 41
8.8 Set date .............................................................................................. 42
8.9 Set time .............................................................................................. 43
8.10 Reset settings .................................................................................... 43
9 Standardised measurement ..................................................................... 44
9.1 Overview of standards ....................................................................... 44
9.2 Carry out a standardised measurement ............................................ 45
10 Quick test ................................................................................................... 48
11 User-defined measurement ...................................................................... 51
11.1 Creating a user-defined measurement .............................................. 52
11.1.1 Overview of templates.......................................................... 53
11.1.2 Selecting a template............................................................. 54
11.1.3 Setting the hold time before the first measurement ............. 54
11.1.4 Setting the stirring time during the hold time ....................... 55
11.1.5 Setting the slew rate for the test voltage .............................. 55
11.1.6 Setting the duration of the pauses between
measurements ..................................................................... 56
11.1.7 Setting the stirring time in the pauses .................................. 56
11.1.8 Setting the number of measurements .................................. 57
11.1.9 Setting the maximum output voltage .................................... 57
11.1.10 Setting the withstand voltage ............................................... 58
11.1.11 Selecting which measured values are not to be
evaluated .............................................................................. 59

iv / 90 822-129-8
DTA 100 C Table of contents

11.1.12 Saving user-defined measurements .................................... 60


11.2 Running a user-defined measurement .............................................. 61
11.3 Editing or deleting a user-defined measurement ............................... 63
12 Display of measurement results .............................................................. 64
12.1 Measurement log as printout ............................................................. 65
12.2 Measurement log on the display ........................................................ 66
13 Putting the device out of operation ......................................................... 67
14 BAUR ITS Lite software for measurement data management .............. 68
14.1 Description of function ....................................................................... 68
14.2 Installing ITS Lite ............................................................................... 69
14.3 Starting ITS Lite ................................................................................. 69
15 Checking the measurement accuracy of the oil breakdown voltage
tester (optional) .......................................................................................... 70
16 Maintenance ............................................................................................... 71
16.1 Cleaning the oil breakdown voltage tester ......................................... 72
16.2 Replacing the printer paper roll .......................................................... 74
16.3 Replacing the printer ink ribbon ......................................................... 76
16.4 Replacing the fuse ............................................................................. 77
16.5 Calibration .......................................................................................... 77
16.6 Ordering accessories and spare parts ............................................... 78
17 Faults .......................................................................................................... 79
17.1 Troubleshooting ................................................................................. 79
17.2 Error messages and corrective measures ......................................... 80
18 Transportation and storage ...................................................................... 81
18.1 Packaging .......................................................................................... 81
18.2 Transportation .................................................................................... 81
18.3 Storage .............................................................................................. 82
19 Warranty and After Sales .......................................................................... 82
20 Disposal ...................................................................................................... 83
20.1 Disposing of the device ...................................................................... 83
20.2 Disposing of the insulating oil ............................................................ 83
21 Delivery includes and Options ................................................................. 84
22 Declaration of conformity ......................................................................... 85
23 Index............................................................................................................ 86

822-129-8 v / 90
About this manual DTA 100 C

1 A BOUT THIS M ANUAL

1.1 Using this manual


This user manual contains all necessary information that is needed for the commissioning
and operation of the described product.
 Read this user manual completely before operating the product for the first time.
 Consider this user manual to be a part of the product and store it in an easily accessible
location.
 If this user manual is lost, please contact BAUR GmbH or your nearest BAUR
representative (http://www.baur.eu/baur-worldwide).

1.2 Applicability of the instructions


These operating instructions apply to the DTA 100 C with the firmware version from 1.16.
The details of the firmware version currently installed can be found under:
Main menu > Tools > Info

1.3 Structure of safety instructions


The safety instructions in this user manual are presented as follows:

Danger
symbol SIGNAL WORD

Type of danger and its source


Possible consequences of violation.
 Measure to prevent the danger.

If a dangerous situation could arise at a specific step, the safety instruction is displayed
immediately before this dangerous step and is shown as follows:

SIGNAL WORD
Type of danger and its source. Possible consequences of violation.
1. Measure to prevent the danger.

6 / 90 822-129-8
DTA 100 C About this manual

Danger levels
Signal words in the safety instructions specify the danger levels.

DANGER Will lead to severe injuries or death.

WARNING May lead to severe injuries or death.

CAUTION May lead to light to moderate injuries.

NOTICE May lead to material damage.

Danger symbols

General danger

Risk of electric shock

Warning about combustible materials

1.4 View Settings

Symbol Meaning

 You are requested to perform an action.

1. Perform the actions in this sequence.


2. ...

a. If an operation consists of several operating steps, they are specified


with "a, b, c". Perform the operating steps in this sequence.
b. ...

1 Numbering in the legend


2 ...
 List

Indicates further information on the topic.

Indicates tools required for the subsequent tasks.

Indicates spare parts required for the subsequent tasks.

Indicates required cleaning agents.

822-129-8 7 / 90
About this manual DTA 100 C

1.5 Note on the screenshots and graphics used


The screenshots and graphics used are intended to illustrate the procedure and may differ
slightly from the actual state.

8 / 90 822-129-8
DTA 100 C For your safety

2 F OR YOUR SAFETY

All BAUR devices and systems are manufactured according to the state of the art and are
safe to operate. The individual parts and the finished devices are subject to continuous
testing by our qualified personnel as part of our quality assurance system. Each device and
system is tested before delivery.
However, the operational safety and reliability in practice can be achieved only when all
necessary measures have been taken. The responsible body1 and operator2 of the device or
system are responsible for planning these measures and monitoring their implementation.
Make sure that the responsible body and persons working with the device or system have
carefully read through and understood the user manual for the DTA 100 C, as well as the
user manuals for all associated devices, before starting work.
The responsible body and operator of the device or system are responsible for any injuries
or damage resulting from non-compliance with this user manual.

2.1 Instructions to the user


The product may only be operated by authorised and trained electrical engineers. An
electrical engineer is a person who, owing to his professional education (electrical
engineering), knowledge, experience and familiarity with the applicable standards and
regulations, can assess the tasks assigned to him and detect possible dangers.
In addition, the user must have:
 Knowledge of the technical equipment and operation of the DTA 100 C
 Knowledge of the testing and measurement procedures
 Knowledge of the electrical insulating materials, in particular the insulating oil, and how
to handle them.

1 Responsible body is the person or group that is responsible for the safe operation of the device and its
maintenance (EN 61010-1, 3.5.12).
2 Operator is the person who uses the device for its intended purpose (according to the definition of user in
compliance with EN 61010-1, 3.5.11).

822-129-8 9 / 90
For your safety DTA 100 C

2.2 Intended use


The powerful BAUR DTA 100 C oil breakdown voltage tester automatically measures the
electrical breakdown strength of insulating liquids fully.
Note: With the DTA 100 C you can test insulating liquids with tan δ values < 4.5 and specific
resistance ρ > 30 MΩm.
If the device is used without observing this condition, safe operation cannot be guaranteed.
The operator or user is liable for any damage to persons and property resulting from
incorrect operation.
Proper use also includes
 Compliance with all instructions in this user manual,
 Compliance with the technical data and connection requirements given on the rating
plate and in the user manual,
 Compliance with the inspection and maintenance tasks.

2.3 Avoid dangers, take safety measures


 When operating the DTA 100 C, observe the following rules and guidelines:
 Accident prevention and environment protection rules applicable for your country
 Safety instructions and regulations of the country where the DTA 100 C is being
used (according to the latest version)
 Any relevant national and international standards and guidelines in the latest
applicable version:
 Local safety and accident prevention regulations
 Employers’ liability insurance association regulations (if any)

Technically secure state of the product


Safety, function and availability depend on the proper condition of the product. Upgrades,
modifications or alterations to the product are essentially prohibited.
 Operate the product only in a technically perfect condition.
 In the event of damage or malfunctions, take the product out of operation immediately,
mark it accordingly and have the faults rectified by appropriately qualified and authorised
personnel.
 Comply with the inspection and maintenance conditions.
 Use only accessories and original spare parts recommended by BAUR. The use of spare
parts, accessories and special fittings that are not tested and approved by BAUR could
adversely affect the safety, function and characteristics of the product.
 Never take apart the product. Inside the product there are no components that could be
serviced or repaired by the user.

No operation with condensation


Condensation can form in devices and systems due to temperature fluctuations and high air
humidity, which in some components can lead from leakage currents and flashovers up to
short-circuit.
Maximum danger arises when relatively high air humidity and temperature fluctuations occur
in a device consecutively, e.g. which is the case when storing the device in an unheated
room or when placed outdoors. When the device is then exposed to a high ambient

10 / 90 822-129-8
DTA 100 C For your safety

temperature, the cold device surfaces cool the air in the immediate vicinity, which leads to
formation of condensation even inside the device.
In this process, two factors are crucial:
 The higher the relative air humidity, the faster the dew point is reached and water is
condensed.
 The higher the temperature difference between the surfaces and the ambient air, the
stronger the tendency for condensation.
 Always prevent condensation in devices. Temper the device and system before and
during the measurements so that no condensation occurs.

No operation in areas with risk of explosion and fire


Measurements in direct contact with water, in environments with explosive gases and in
areas with fire risks are not permitted.

Cleanliness of oil breakdown voltage tester


To achieve reliable measurement results, the oil breakdown voltage tester, the test chamber,
the HV isolators, the test containers and all tools that come into contact with the oil must be
clean.
BAUR GmbH recommends cleaning the HV isolators at least once a year to guarantee
maximum accuracy.
HV isolators may be cleaned only by qualified technical staff authorised by BAUR.
In case of questions, please contact BAUR GmbH or your local BAUR representative
(http://www.baur.eu/baur-worldwide).

Lifting and carrying the oil breakdown voltage tester


 The DTA 100 C oil breakdown voltage tester weighs up to 39 kg. It is recommended to
seek the help of another person to lift or carry the device.

2.3.1 Dangers when working with electric voltage


A dangerous voltage of up to 100 kV is generated during measurements with the
DTA 100 C. Operating personnel need to pay special attention and must be very careful
while working with electric voltage.

DANGER

Dangerous electric voltage


Danger to life or risk of injury due to electric shock
 Connect the oil breakdown voltage tester as described in this user
manual.
 Take particular care to ensure the oil breakdown voltage tester is
earthed properly.
 Before carrying out any cleaning or maintenance, switch off the
device and remove the mains plug to ensure the oil tester is
completely disconnected from the mains voltage.

822-129-8 11 / 90
For your safety DTA 100 C

2.4 Special personal protective equipment


Personal protective equipment based on the risk assessment for the relevant working
conditions is part of the DTA 100 C safety concept.
 Observe the national safety regulations and your company's working and operating
instructions.
Dependent on the conditions of the work place, use the following protective equipment:

Protection against electrostatic charging,


 Safety footwear
crushing, slipping and other accidents:

Protection against electrical hazards (arcing  Certified safety clothing


fault):
 Hard hat with visor
 Insulating protective gloves
 LV HRC fuse handle with sleeve

Protection against noise:  Ear protection

Protection against dangers from road traffic:  High visibility vest according to EN 471
(Protection class 2) or according to the
applicable standards in your country for high
visibility clothing for commercial use.
Important: No high visibility vest during tasks
with risk of arcs!

Hand protection:  Safety gloves

12 / 90 822-129-8
DTA 100 C Product information

3 P RODUCT INFORM ATION

3.1 Full illustration


Front view

No. Element Function

1 Handle Used to open and close the protective cover

2 Protective cover Used to protect against dust and oil

3 Safety switch Used as a protective device


The protective cover must be closed in order to perform a
measurement.

822-129-8 13 / 90
Product information DTA 100 C

No. Element Function

4 Test vessel with electrodes Filled with insulating oil


The test vessels have a standard holding capacity of 0.4 litres and are
made of glass. They comply with the following standards:
 IEC 60156 Fig. I or Fig. II
 ASTM D877
 ASTM D1816
The electrode shape is specified in the standards and is indicated on
the display before a measurement begins.
The figure illustrates the 0.4 litre test vessel in accordance with
IEC 60156 Fig. II with cover.
Further information:
 Test vessel shapes – Chapter Delivery includes and Options (on
page 84)
 Replacing the electrodes – Chapter Replacing the electrodes (on
page 27)

5 Oil collecting tray Used to collect insulating oil

6 Integrated plain paper printer Used to print the measurement logs

7 Fold-out operating unit Used to operate the oil breakdown voltage tester
The operating unit contains the operator control panel and the
membrane keyboard.
Further information: Chapter Operating and display elements (on page
16)

Standard accessories

No. Element Function

1 Setting gauge Used to set the electrode gap

2 Magnetic stirrer Used to stir the oil sample

14 / 90 822-129-8
DTA 100 C Product information

No. Element Function

3 Lifting stick for magnetic Used to remove the magnetic stirrer from the oil sample
stirrer

Rear view

No. Element Function

1 USB connection Used to connect to a PC


Type of interface:USB Type B

2 Protective earthing Is used to connect the protective earthing


connection

3 Mains connection Is used to connect the device to the mains voltage (90 – 264 V,
50/60 Hz)
The power supply connection comprises:
 Power supply socket
 Mains fuse, Type: T 4 AH
 Mains switch

822-129-8 15 / 90
Product information DTA 100 C

3.2 Operating and display elements

No. Element Function

1 Standardised measurement Opens the Standardised measurement menu


button
Further information: Chapter Standardised measurement (on page 44)

2 Measurement logs button Opens the Measurement logs menu


Further information: Chapter Display of measurement results (on page
64)

3 Info button Displays the device information.


Further information: Chapter Information on the oil breakdown voltage
tester (on page 24)

4 Input button Used to confirm the input

5 Delete button Used to delete a character at the cursor position

6 Display Show the menu of the device

7 Control keys Used to navigate through the menu


Further information: Chapter Operating the oil breakdown voltage tester
(on page 21)

8 Printer operating keys  The paper feed button LF/SEL is used to feed in the printer paper
when the paper roll is replaced.
Further information: Chapter Replacing the printer paper roll (on
page 74)
 The SET button is not assigned.

16 / 90 822-129-8
DTA 100 C Product information

3.3 USB interface


The USB interface is used:
 to connect to a PC for communicating with the BAUR ITS Lite software for managing
measurement data (free download: http://www.baur.eu),
 for firmware updates performed by a BAUR representative.
Type of interface:: USB Type B

3.4 Power supply


The voltage supply for DTA 100 C can be provided via an existing mains supply on site.
Permissible mains voltage: 90 – 264 V
Permissible mains frequency: 50/60 Hz

NOTICE

Too high or too low mains voltage


A low mains voltage adversely affects the function of the system, a high mains
voltage can cause damage.
 Ensure that the supply voltage matches the specifications on the rating plate.

822-129-8 17 / 90
Product information DTA 100 C

3.5 Rating plate


The rating plate is located on the back of the oil breakdown voltage tester.

Element Description

Type Device type

Nr. Serial number

U Supply voltage
If several supply voltages are possible, these are given consecutively one after
another.

Time characteristics and nominal current of the device fuse


Time characteristics:
 Very Quick Acting (FF)
 Quick Acting (F)
 Medium (M)
 Slow Blow (T)
 Very Slow Blow (TT)

f Mains frequency

VA Max. recorded apparent output

AC ueff: 0 - 100 kV Effective output AC voltage

General warning sign


Indicates that there is a potential risk of danger when using the product and hence
the user manual must be observed

CE mark
Indicates that the device or system conforms to CE.

BAUR GmbH Name and address of the manufacturer


6832 Sulz / Austria

Made in Austria Indicates the country in which the device was manufactured.
Austria: Austria

18 / 90 822-129-8
DTA 100 C Technical data

4 T ECHNICAL DATA

Further information on the possible test vessels: Chapter Delivery includes and Options (on
page 84)

822-129-8 19 / 90
Basic insulating oil test procedure DTA 100 C

5 B ASIC INSULATING OIL TEST PROCEDURE

1. Installing the oil breakdown voltage tester


2. Connecting the oil breakdown voltage tester to the
power supply
3. Earthing the oil breakdown voltage tester (if
operated in a laboratory)

4. Turning on the oil breakdown voltage tester

5. Select the electrodes specified in the standard


6. Insert the electrodes in the test vessel
7. Set the electrode distance

8. Fill the test vessel with the oil sample


9. Insert the test vessel into the oil breakdown voltage
tester

10. Performing the measurement


11. Print and/or save the measurement log

12. Remove the test vessel from the oil breakdown


voltage tester
13. Dispose of the oil sample

14. Switch off the oil breakdown voltage tester

20 / 90 822-129-8
DTA 100 C Operating the oil breakdown voltage tester

6 O PERATING THE OIL BREAKDOWN VOLTAGE TESTER

6.1 Menu navigation


The control menu is displayed on the colour display. The menu is operated with the control
keys under the colour display and the function keys on the membrane keypad. The meaning
of the relevant control key is displayed in plain text or as a symbol in the lowest row of the
display.

6.2 Entering numbers


There are two options for entering or changing numbers:
 Enter a value with the number keys on the membrane keypad.
 Press the arrow keys to change a value:
 To increase a value, press the <up arrow>.
 To decrease a value, press the <down arrow>.

822-129-8 21 / 90
Operating the oil breakdown voltage tester DTA 100 C

6.3 Symbols and abbreviations on the display


Symbols

Symbol Description

Indicates an information message

Indicates a request

Indicates that the menu for setting the time is open

Indicates that the menu for setting the date is open

Indicates that the menu for setting the withstand voltage is open

Indicates whether an option has been enabled


/

/ Used to toggle between menu items

/ Used to move the cursor to the left/right

Indicates an important process for achieving a precise measurement

Indicates that there is high voltage at the electrodes

Indicates which electrode shape is required to perform a measurement


Further information: Chapter Replacing the electrodes (on page 27)

Abbreviations

Stand. dev.: Standard deviation


The standard deviation shows how close the individual recorded values are
to the average value.

Std.dev./Avg.: Ratio of the standard deviation to the average value in % (coefficient of


variation).
In IEC 60156, diagram 3 in section 11 illustrates the ratio of the standard
deviation to the average value. This diagram and the determined values can
be used to check whether the measurement result lies within the permitted
range.

22 / 90 822-129-8
DTA 100 C Operating the oil breakdown voltage tester

6.4 Main menu


All functions and user settings can be accessed from the main menu.

Menu item Description

Standardised Configure settings for measurements in compliance with specific standards and
measurement start measurements.

Quick test Start test for quick evaluation of the insulating oil status

User-defined Configure settings for user-defined measurements and start measurements


measurement

Device settings You can define or change general system settings for the device:
 Set display brightness
 Selecting the language
 Switch printer on or off
 Define how to proceed if the measurement log memory is full and a new
measurement is to be performed (overwrite logs yes/no)
 Configure settings for IEC 60156:2018 and the standards based on it
 Configure ASTM D standard settings
 Configure settings for user-defined measurements
 Setting the date
 Setting the time
 View information about the USB interface

Tools  Perform electrode cleaning


 View device information on the oil breakdown voltage tester: Serial number,
date of last calibration and the manufacturer’s contact details
 Activating options
 Check the measurement accuracy of the oil breakdown voltage tester with the
calibrator KA DPA/DTA C (option).

Measurement logs Display, print or delete measurement logs.

822-129-8 23 / 90
Operating the oil breakdown voltage tester DTA 100 C

6.5 Information on the oil breakdown voltage tester


Under Main menu > Tools > Info, you will find the following information on the oil
breakdown voltage tester:
 Date on which the oil breakdown voltage tester was last calibrated
 Oil breakdown voltage tester serial number
 Oil breakdown voltage tester firmware version
 Hardware version
 Available options
 Correction factors
Main menu > Tools > Info

1. Select the menu item Info by pressing the arrow keys.


2. Select the menu item Next to confirm the selection.
Information on the oil breakdown voltage tester is displayed.

6.5.1 Information on the USB interface


Main menu > Device settings > About USB port

1. Select the position by pressing the arrow keys.


2. Select the menu item Next to confirm the selection.
The version and technical data of the USB port are displayed.

24 / 90 822-129-8
DTA 100 C Commissioning

7 C OMMISSIONING

 Observe the following information:


 The safety instructions in the chapter For your safety (on page 9)
 Local safety and accident prevention regulations
 Safety instructions and regulations according to the state-of-the-art
 Any relevant national and international standards and guidelines in the latest
applicable version

7.1 Checks to perform before commissioning

CAUTION

Safety defects due to use of damaged device.


 Never use devices that are visibly damaged or clearly have a
malfunction.
 Secure devices that are visibly damaged or clearly have a
malfunction against unintentional switching-on.
 Rectify faults immediately.

1. Check the oil breakdown voltage tester for mechanical damage.


2. Check electrical connections and connection cables for damage.
Use only undamaged connection cables.
3. Check the electrodes of the test vessel for pitting. If corrosion was detected, replace the
test vessel.
Further information: Chapter Ordering accessories and spare parts (on page 78)
4. Always keep the oil breakdown voltage tester clean.
The cleanliness of the oil breakdown voltage tester and the test cell will have a strong
influence on the measurement results.
5. Always wipe off spilled or leaked oil with petroleum ether or other similar mild solvent
and a lint-free cloth, in particular:
 on the test vessel isolators
 in the oil collecting tray (test chamber floor)
 on the test vessel
6. Allow the oil breakdown voltage tester to dry well before closing it.

822-129-8 25 / 90
Commissioning DTA 100 C

7.2 Installing the oil breakdown voltage tester


1. Select the place of installation for the oil breakdown voltage tester in such a way that
 a stable base with sufficient air circulation is guaranteed,
 the oil breakdown voltage tester is standing on a flat, horizontal surface,
 the oil breakdown voltage tester is accessible for making connections and for use.
2. If the oil breakdown voltage tester is in a transport case (option), remove the lid of the
case. The oil breakdown voltage tester can be operated in the transport case.

7.3 Earthing the oil breakdown voltage tester

1 Earth connection screw


2 Cable lug
3 Protective earthing connection

Prerequisite
Earth cable (not included in standard delivery): Diameter min. 1.5 mm 2, cable lug for M6

Procedure
1. Connect a protective earthing cable to the station earth.
2. The earth connection screw is located on the back of the oil tester. Unscrew the earth
connection screw (1).
3. Connect the cable lug of the protective earthing cable (2) to the earth connection screw
(1).
4. Re-screw the earth connection screw into the protective earthing connection (3).

26 / 90 822-129-8
DTA 100 C Commissioning

7.4 Replacing the electrodes


The electrode shape is defined by standards and is shown pictographically on the display
before a measurement begins.
Note:
 Before each measurement, check that the electrodes used meet the selected standard.
 If the wrong electrodes are used, replace them.

1. Open the dust cover.


2. Remove the cover of the test vessel.
3. First, turn one and then the other electrode anticlockwise until they are released from the
thread.
4. Take the other electrode pair that must be used. First, turn one and then the other
electrode clockwise on the thread.
5. Close the cover of the test vessel.
Note: Clean the electrodes when using the electrodes for the first time, or if the electrodes
have been replaced, e.g. if another electrode shape is required for the measurement. Further
information: Chapter Clean the electrodes (on page 28)

822-129-8 27 / 90
Commissioning DTA 100 C

7.5 Clean the electrodes


Clean the electrodes:
 Before using new electrodes for the first time
 After having replaced the electrodes, e.g. if another electrode shape is required for the
measurement
Main menu > Tools > Electrode cleaning

1. In the main menu, select the Tools menu item.


2. Select the menu item Next to confirm the selection.
3. In the Tools menu, select the Electrode cleaning menu item and then Next.
4. Fill the test vessel with clean, unused oil of the same type that will be tested next.
5. Select Start menu item to start the electrode cleaning process.
24 cleaning breakdowns are performed. The voltage required to clean the electrodes is
built up automatically and shown in the display.

7.6 Setting an electrode gap


The electrode gap is defined by standards and is shown pictographically on the display
before a measurement begins:

 Set an electrode gap according to the selected standard. Proceed as follows:

28 / 90 822-129-8
DTA 100 C Commissioning

1. Open the dust cover.


2. Set the Vernier scale to zero (1).
3. On the other side of the ring with Vernier scale, there is an adjusting ring without scale
(2). Turn the adjusting ring clockwise until you hear an acoustic signal. The electrodes
are in contact with each other.
4. Slowly turn the adjusting ring anticlockwise until the acoustic signal stops.
The electrode gap is set to zero.
5. Set the Vernier scale to the desired electrode gap.
1 rotation (360°) = 1 mm
6. Remove the cover of the test vessel.
7. Use a setting gauge to check the set electrode gap.
8. Close the cover of the test vessel.

7.7 Fill and use the test vessel

NOTICE

Measurement errors caused by improper sampling and handling of the oil


sample
Improper sampling and handling of the oil sample can have a strong influence on the
measurement results which can lead to incorrect conclusions being drawn regarding
the quality of the insulating oil.
 The sampling must only be performed by experienced, qualified personnel.
 Follow the instructions for sampling below.

822-129-8 29 / 90
Commissioning DTA 100 C

7.7.1 Instructions for sampling


Important: In order to achieve reliable measurement results, all devices, containers and
tools that come into contact with the oil must be clean.

 Follow the manufacturer's instructions and safety instructions for the relevant sampling
container or electrical device.
 When sampling, ensure you follow the instructions given in the relevant standard.
 Only take oil samples during dry weather. Avoid sampling during rain or if it is very
humid. High humidity adversely affects the measurement results.
 Remove any impurities at the sampling port (e.g. on the transformer) and in the sampling
container.
 Use separate sampling containers for each oil type.
 Take the sample during regular operation of the electrical equipment.
 Once the oil samples have been received by the laboratory, they must be left to stand in
the room for approximately 24 hours in order for any bubbling to subside and for the oil
to adjust to the room temperature.
 Protect the sample against moisture and solar radiation.
 Condensation adversely affects the measurement results. To prevent condensation,
ensure that the temperature of the oil to be tested corresponds to the ambient
temperature of 20°C ± 5°C before measuring.
 Heat the sampling container to above the ambient temperature.
Note: Some standards stipulate maintaining specific temperature conditions during the
measurement process. Further information: Chapter Overview of standards (on page 44)

Further information on sampling, handling the oil sample and cleaning can be
found in the current version of each of the following standards:
 IEC 60475
 IEC 60156
 ASTM D923
 ASTM D877
 ASTM D1816

30 / 90 822-129-8
DTA 100 C Commissioning

7.7.2 Insert the test vessel

1. Open the protective cover.


2. Remove the test vessel.
Note: To avoid impurities in the test vessel isolators, place the test vessel on an even,
clean and dry base when you remove the test vessel from the oil breakdown voltage
tester.
3. Remove the cover of the test vessel.
4. Clean the test vessel and the lid with a clean, lint-free cloth.
5. Rinse the test vessel and lid with petroleum ether or acetone.
6. Dry the test vessel and lid.
7. Rinse the test vessel and lid three times with the oil to be tested.
8. Allow the oil to be tested to flow down the inside wall of the test vessel slowly without
bubbles until the vessel is 95 to 98% full.
9. Replace the full test vessel in the oil breakdown voltage tester.
10. Close the lid of the test vessel.

822-129-8 31 / 90
Commissioning DTA 100 C

7.8 Turning on the oil breakdown voltage tester


1. Ensure that the mains voltage matches the specifications on the rating plate.
2. Connect the oil breakdown voltage tester to the mains voltage.
3. There is a mains switch on the back of the device. Use it to switch on the device.
4. Fold out the operating unit.
After the operating unit has been opened, the firmware starts and the oil breakdown
voltage tester performs a self test. On completion of a successful self test, the Start
window of the last measurement opens.

32 / 90 822-129-8
DTA 100 C Device settings

8 D EVICE SETTINGS

The oil breakdown voltage tester settings are adjusted via the main menu item Device
settings.

8.1 Set display brightness


Main menu > Device settings > Display brightness

1. In the main menu, select the menu item Device settings.


2. Select the menu item Next to confirm the selection.
3. In the Device settings menu select the menu item Display brightness and then menu
item Change.
4. Select the time after which the brightness of the display light automatically reduces if the
device is inactive.
5. Click Save to confirm the entry.

822-129-8 33 / 90
Device settings DTA 100 C

8.2 Select language


Main menu > Device settings > Language

1. Choose a language by pressing the arrow keys.


2. Click Save to confirm the entry.

8.3 Switch printer on/off


Main menu > Device settings > Printer

1. Choose one of the following settings by pressing the arrow keys:


 off: The printer is switched off.
 on: The printer is switched on.
2. Click Save to confirm the entry.

34 / 90 822-129-8
DTA 100 C Device settings

8.4 Overwrite measurement logs


The device can store up to ten measurement logs. When the measurement log memory is
full, old measurement logs must be deleted before you can save new ones. Under Clear
meas. logs you can define how to proceed when the measurement log memory is full and a
measurement is to be performed.
Main menu > Device settings > Clear meas. logs

1. Choose one of the following settings by pressing the arrow keys.


 yes: When you start the next measurement log, the oldest measurement log will be
automatically deleted.
 no: Before starting the next measurement log, a saved measurement log must be
deleted manually.
Further information: Chapter Display of measurement results (on page 64)
2. Click Save to confirm the entry.

8.5 Configure the IEC 60156:2018 settings


According to the standard IEC 60156:2018, its Annex A and the standards based on
IEC 60156:2018, it is possible to adjust the hold time prior to the first measurement, various
stirrer settings and electrode shapes. The settings configured under this menu item are
applied automatically for the following standards:
 IEC 60156:2018
 IEC 60156:2018 Annex A
 AS 1767.2.1
 BS EN 60156
 CEI EN 60156
 SEV EN 60156
 UNE EN 60156
 NF EN 60156
 SABS EN 60156
 VDE 0370-5:1996

822-129-8 35 / 90
Device settings DTA 100 C

Main menu > Device settings > IEC 60156:2018 settings

8.5.1 Setting the hold time before the first measurement


Main menu > Device settings > IEC 60156:2018 settings > Hold time

1. In the main menu, select the menu item Device settings.


2. Select the menu item Next to confirm the selection.
3. In the Device settings menu, select the IEC 60156:2018 settings menu item and then
Next.
4. Select Hold time menu item.
5. Select menu item Change.
6. Choose one of the following settings by pressing the arrow keys:
 5 min: The hold time is 5 minutes.
 15...60 min: The hold time can be set to between 15 and 60 minutes.
7. If you have enabled the 15...60 min setting, select the Next menu item.
8. Enter the desired hold time.
For insulating liquids with a standard viscosity greater than 15 mm2/s (40°C), the hold
time prior to the first measurement should be 15-30 minutes.
9. Click Save to confirm the entry.

36 / 90 822-129-8
DTA 100 C Device settings

8.5.2 Adjusting the stirrer


Main menu > Device settings > IEC 60156:2018 settings > Stirrer

1. In the main menu, select the menu item Device settings.


2. Select the menu item Next to confirm the selection.
3. In the Device settings menu, select the IEC 60156:2018 settings menu item and then
Next.
4. Select Stirrer menu item.
5. Choose one of the following settings by pressing the arrow keys:
 without stirring: The stirrer is inactive when a standard based on the IEC 60156 is
selected.
 contin. stirring: The stirrer runs constantly when a standard based on the
IEC 60156 is selected.
6. Click Save to confirm the entry.

822-129-8 37 / 90
Device settings DTA 100 C

8.5.3 Setting the electrode shape


Main menu > Device settings > ... > Electrode shape
1. In the main menu, select the menu item Device settings.
2. Select the menu item Next to confirm the selection.
3. In the Device settings menu, select the IEC 60156:2018 settings menu item and then
Next.
4. Select Electrode shape menu item.
5. Select menu item Change.
A prompt to select an electrode shape appears on the display.
6. Choose one of the following settings by pressing the arrow keys:
 IEC 60156 Fig. I: Ball-shaped electrodes according to IEC 60156 Fig. I
 IEC 60156 Fig. II: Mushroom-shaped electrodes according to IEC 60156 Fig. II

7. Choose an electrode shape by pressing the arrow keys.


8. Click Save to confirm the entry.

38 / 90 822-129-8
DTA 100 C Device settings

8.6 Configure the ASTM settings


With the ASTM D standards, it is possible to display the individual breakdown values and set
the hold time before the first measurement. The settings configured under this menu item are
applied automatically for the following standards:
 ASTM D1816:2012 1 mm
 ASTM D1816:2012 2 mm
 ASTM D877/D877M:2013 PA
 ASTM D877/D877M:2013 PB
Main menu > Device settings > ASTM settings

8.6.1 Display individual breakdown values


Main menu > Device settings > ... > Display indiv. breakdown values

1. In the main menu, select the menu item Device settings.


2. Select the menu item Next to confirm the selection.
3. In the Device settings menu select the menu item ASTM settings and then menu item
Next.
4. Select menu item Display indiv. breakdown values.
5. Select menu item Change.
6. Choose one of the following settings by pressing the arrow keys:
 yes: The individual breakdown values recorded during a series are listed in the
measurement log.
 no: The average value of the recorded breakdown values is calculated and
displayed in the measurement log.
7. Click Save to confirm the entry.

822-129-8 39 / 90
Device settings DTA 100 C

8.6.2 Set the hold time before the first measurement


Main menu > Device settings > ... > Hold time

1. In the main menu, select the menu item Device settings.


2. Select the menu item Next to confirm the selection.
3. In the Device settings menu select the menu item ASTM settings and then menu item
Next.
4. Select menu item Hold time.
5. Select menu item Change.
6. Choose one of the following settings by pressing the arrow keys:
 3 min: The hold time is 3 minutes.
 30...60 min: The hold time can be set to between 30 and 60 minutes.
7. If you have enabled the 30...60 min setting, select the menu item Next.
8. Enter the desired hold time.
When testing ester liquids in accordance with ASTM D1816:2012 the hold time before
the first measurement must be at least 30 mins.
9. Click Save to confirm the entry.

40 / 90 822-129-8
DTA 100 C Device settings

8.7 Configure settings for user-defined measurements


With user-defined measurements, it is possible to set the electrode shapes and the distance
between the electrodes.

Prerequisite
The desired distance between electrodes has been set.
Further information: Chapter Setting an electrode gap (on page 28)

Procedure
Main menu > Device settings > User-defined measurement settings

1. In the main menu, select the menu item Device settings.


2. Select the menu item Next to confirm the selection.
3. In the Device settings menu, select the User-defined measurement settings menu
item and then Next.
A prompt to select an electrode shape appears on the display.

4. Choose an electrode shape by pressing the arrow keys.

822-129-8 41 / 90
Device settings DTA 100 C

5. Click Save to confirm the entry.


A window appears on the display with an input field for the distance between electrodes.

6. Use the number keys on the membrane keypad to enter the desired distance between
electrodes.
A distance between electrodes of between 0.5 and 10 mm can be entered. The entered
distance between electrodes is displayed in the measurement log.
7. Click Save to confirm the entry.

8.8 Set date


Main menu > Device settings > Date

1. Select the position by pressing the arrow keys.


2. Change the date with the membrane keypad.
3. Click Save to confirm the entry.

42 / 90 822-129-8
DTA 100 C Device settings

8.9 Set time


Main menu > Device settings > Time

1. Select the position by pressing the arrow keys.


2. Change the time with the membrane keypad.
3. Click Save to confirm the entry.

8.10 Reset settings


Note: This menu item resets all settings to the factory defaults. All user-defined
measurements and measurement logs will be deleted!
Main menu > Device settings > Reset settings

1. Select the menu item by pressing the arrow keys.


2. Click Next to confirm the entry.
A warning is displayed that all settings and user data will be deleted upon resetting the
settings.
3. Confirm with Yes.

822-129-8 43 / 90
Standardised measurement DTA 100 C

9 S TANDARDISED MEASUREM ENT

9.1 Overview of standards


The factory calibrated settings for the standardised measurements are configured as follows:

1)
Can be selected under Main menu > Device settings > IEC 60156:2018 settings
For insulating liquids with a standard viscosity greater than 15 mm2/s (40°C), the hold time
prior to the first measurement should be 15-30 minutes.

44 / 90 822-129-8
DTA 100 C Standardised measurement

2)
Can be selected under Main menu > Device settings > ASTM settings
When testing ester liquids in accordance with ASTM D1816:2012 the hold time before the
first measurement must be at least 30 mins.
3)
Can be deactivated under Main menu > Device settings > IEC 60156:2018 settings
4)
Measurements per filling / number of measurements
The electrode shapes and their dimensions:

: IEC 60156 Fig. I, ball Ø 12.5 to 13.0

: IEC 60156 Fig. II, Ø 36.0 / radius 25.0

: ASTM D877, disc Ø 25.4

9.2 Carry out a standardised measurement


Main menu > Standardised measurement

1. In the main menu, select the menu item Standardised measurement.


2. Select the menu item Next to confirm the selection.

822-129-8 45 / 90
Standardised measurement DTA 100 C

3. Select a standard, then menu item Next.


The electrode shape and gap are defined by the relevant standard.
A prompt appears on the display to set the electrode gap according to the selected
standard. The required electrode shape is displayed pictographically.

4. Check that the electrodes used meet the selected standard.


If the wrong electrodes are used, replace them. Further information: Chapter Replacing
the electrodes (on page 27)
5. Set the indicated electrode gap.
Further information: Chapter Setting an electrode gap (on page 28)
6. Select the menu item Start.
A prompt to enter a sample number appears on the display. The oil breakdown voltage
tester will suggest a sample number composed of the date and time in DDMMYYhhmm
format. This sample number can be extended as required.

7. If you want to amend the suggested sample number, do so or enter any sample number
with the number keys on the membrane keyboard.

46 / 90 822-129-8
DTA 100 C Standardised measurement

8. Click Save to confirm the entry.


The measurement begins. A window displays the details of the measurement process.
The symbol indicates that there is high voltage at the electrodes.

When the oil breakdown voltage tester has completed the measurement, the
measurement log is shown on the display. If the printer is switched on, the measurement
log will automatically be printed first.
9. Select the menu item Details in the lower menu bar for a detailed view of the
measurement results.
10. Select the menu item Back to perform further measurements or to stop the
measurement process.
The oil breakdown voltage tester switches to the beginning of the measurement. A
prompt appears on the display to set the electrode gap.
11. Repeat the process to perform additional measurements.
Select the menu item Back to stop the measurement process.
The oil breakdown voltage tester switches to the main menu.

Cancelling the measurement manually


 Select menu item Stop.

822-129-8 47 / 90
Quick test DTA 100 C

10 Q UICK TEST

A quick test is used to make a quick evaluation of the oil status. In this process, an
orientation value will be determined for the breakdown strength of the insulating oil.
Quick test parameters:
 Standard: none;
 Stirrer: inactive;
 Output voltage: max. possible output voltage.
It is not possible to adjust the quick test parameters. To adjust individual parameters, create
a user-defined measurement. Further information: Chapter User-defined measurement (on
page 51)
Main menu > Quick test

1. In the main menu, select the menu item Quick test.


2. Select the menu item Next to confirm the selection.
A prompt appears on the display to set the electrode gap. You can choose any electrode
shape you want during a quick test.
Recommendation: Select an electrode shape based on the selected standard.

48 / 90 822-129-8
DTA 100 C Quick test

3. Set the electrode gap according to the selected standard.


Further information:
 Chapter Setting an electrode gap (on page 28)
 Chapter Overview of standards (on page 44)
4. Set the slew rate for the test voltage. Please use the arrow keys to make entries.
5. Select the menu item Start.
A prompt to enter a sample number appears on the display. The oil breakdown voltage
tester will suggest a sample number composed of the date and time in DDMMYYhhmm
format. This sample number can be extended as required.

6. If you want to amend the suggested sample number, do so or enter any sample number
with the number keys on the membrane keyboard.
7. Click Save to confirm the entry.
The measurement begins. A window displays the details of the measurement process.
The symbol indicates that there is high voltage at the electrodes.

When the oil breakdown voltage tester has completed the measurement, the
measurement log is shown on the display. If the printer is switched on, the measurement
log will automatically be printed first.
8. Select the menu item Details in the lower menu bar for a detailed view of the
measurement results.

822-129-8 49 / 90
Quick test DTA 100 C

9. Select the menu item Back to perform further measurements or to stop the
measurement process.
The oil breakdown voltage tester switches to the beginning of the measurement. A
prompt appears on the display to set the electrode gap.
10. Repeat the process to perform additional measurements.
Select the menu item Back to stop the measurement process.
The oil breakdown voltage tester switches to the main menu.
11. Select the menu item Back to perform further measurements or to stop the
measurement process.

Cancelling the measurement manually


 Select menu item Stop.

50 / 90 822-129-8
DTA 100 C User-defined measurement

11 U SER - DEFINED MEASUREMENT

Main menu > User-defined measurement

1. Select the menu item by pressing the arrow keys.


2. Select the menu item Next to confirm the selection.
The User-defined measurement menu opens.

3. You have the following options:


 Add a new measurement
Further information: Chapter Creating a user-defined measurement (on page 52)
 Select an existing measurement
Further information: Chapter Editing or deleting a user-defined measurement (on
page 63) or Running a user-defined measurement (on page 61)

822-129-8 51 / 90
User-defined measurement DTA 100 C

11.1 Creating a user-defined measurement


Main menu > User-defined measurement

1. Select the menu item by pressing the arrow keys.


2. Select the menu item Next to confirm the selection.
A prompt appears on the display to enter a name for the measurement.

3. Enter a name for the measurement. If you do not enter a name, the device will issue a
name based on the date and time in a DDMMYYhhmm format.
4. Click Save to confirm the entry.
The menu for defining measurement parameters will open.

52 / 90 822-129-8
DTA 100 C User-defined measurement

11.1.1 Overview of templates


If your measurement is similar to a standardised measurement, you can load the settings
from a standard as a template and customise them in the other menu items.
The following table provides an overview of the settings according to the template.

x: Adjustment possible
–: Adjustment not possible
Note: When a template is selected, only the settings relevant for that specific template will
be displayed. If you make an adjustment to one setting which then makes another setting
impossible, the setting which can no longer be selected will automatically be greyed out.
The following chapter explains in more detail how to adjust each parameter.

Continuous measurement template


A continuous measurement consists of blocks of 6 measurements each. The blocks will be
repeated until the menu item Stop is selected.

Withstand voltage template


In the case of a measurement with a withstand voltage, the voltage is raised gradually until
the desired voltage is achieved. The level of the withstand voltage, the number of steps and
the time the withstand voltage remains at a particular level, are all freely selectable.
The oil sample is stirred constantly during the measurement taken with a withstand voltage.
The parameter Stirring in hold time is not active in this template and is not displayed.
Note: The measurement with withstand voltage can also be used to check whether the oil
breakdown voltage tester is capable of generating the required high voltage.
Further information: Chapter Setting the withstand voltage (on page 58)

822-129-8 53 / 90
User-defined measurement DTA 100 C

11.1.2 Selecting a template


Main menu > User-defined measurement > ... > Template

1. In the main menu, select the menu item User-defined measurement.


2. Select the menu item Next to confirm the selection.
3. In the User-defined measurement menu select the menu item Template and then
menu item Change.
4. Select a template by pressing the arrow keys.
5. Confirm the selection with Load.
The selected template is registered in the measurement log.
Note: If you define additional settings for the measurement which deviate from the selected
standard, a message appears stating that: You cannot change this setting in the selected
template! If you apply the different setting despite this, the selected template will be
disabled.

11.1.3 Setting the hold time before the first measurement


Main menu > User-defined measurement > ... > Hold time

1. Select menu item Hold time.


2. Select menu item Change.
3. Enter the hold time before the first measurement using the number keys on the
membrane keyboard.
4. Click Save to confirm the entry.

54 / 90 822-129-8
DTA 100 C User-defined measurement

11.1.4 Setting the stirring time during the hold time


Recommendation: As the homogeneity of the oil sample has a strong influence on the
quality of the measurement results, stir the oil sample during the measurement process.
Main menu > User-defined measurement > ... > Stirring in hold time

1. Select menu item Stirring in hold time.


2. Select menu item Change.
3. Enter the stirring time during the hold time before the first measurement using the
number keys on the membrane keyboard.
4. Click Save to confirm the entry.

11.1.5 Setting the slew rate for the test voltage


Main menu > User-defined measurement > ... > Volt. rise

1. Select the menu item by pressing the arrow keys.


2. Select menu item Change.
3. Enter the slew rate for the test voltage with the number keys on the membrane keypad.
4. Click Save to confirm the entry.

822-129-8 55 / 90
User-defined measurement DTA 100 C

11.1.6 Setting the duration of the pauses between measurements


Main menu > User-defined measurement > ... > Pause

1. Select the menu item by pressing the arrow keys.


2. Select menu item Change.
3. Enter the duration of the pauses between measurements with the number keys on the
membrane keypad.
4. Click Save to confirm the entry.

11.1.7 Setting the stirring time in the pauses


Recommendation: As the homogeneity of the oil sample has a strong influence on the
quality of the measurement results, stir the oil sample during the measurement process.
Main menu > User-defined measurement > ... > Stirrer

1. Select menu item Stirrer.


2. Select menu item Change.
3. Choose one of the following settings by pressing the arrow keys:
 off: The stirrer is inactive.
 conti.: The stirrer runs continuously.
 0...x min: Define how long the oil sample is stirred during the pauses
Note: The max. stirring time equates to the duration of the pause
4. Click Save to confirm the entry.

56 / 90 822-129-8
DTA 100 C User-defined measurement

11.1.8 Setting the number of measurements


Main menu > User-defined measurement > ... > No. measurements

1. Select the menu item by pressing the arrow keys.


2. Select menu item Change.
3. Enter the number of measurements with the number keys on the membrane keypad.
4. Click Save to confirm the entry.

11.1.9 Setting the maximum output voltage


Main menu > User-defined measurement > ... > Max. output volt.

1. Select the menu item by pressing the arrow keys.


2. Select menu item Change.
3. Enter the maximum output voltage with the number keys on the membrane keypad.
4. Click Save to confirm the entry.

822-129-8 57 / 90
User-defined measurement DTA 100 C

11.1.10 Setting the withstand voltage


Prerequisite
The template Withstand voltage has been selected.

Procedure
Main menu > User-defined measurement > ... > Template > Withstand voltage

1. Select menu item Withstand voltage.


2. Select menu item Change.
3. Select the number of steps in which the withstand voltage is to be achieved.
For example:

4. Click Save to confirm the entry.


5. Select the withstand voltage value for the first step.

58 / 90 822-129-8
DTA 100 C User-defined measurement

6. Select the menu item Next to confirm the selection.


7. Select the duration for which the withstand voltage is maintained during the first step.

8. Select the menu item Next to confirm the selection.


9. Select the withstand voltage value and the duration of the withstand voltage for all
subsequent steps. To do this, repeat steps 5 to 8 for each further step.

11.1.11 Selecting which measured values are not to be evaluated


Define which measured values are considered in the measurement result.
Main menu > User-defined measurement > ... > Ignore measurement values

1. Select the menu item by pressing the arrow keys.


2. Select menu item Change.
3. Choose one of the following settings by pressing the arrow keys:
 none: The measured values of all measurements are included in the measurement
result.
 first: The first measurement is not included in the measurement result.
 last: The last measurement is not included in the measurement result.
 first & last: The first and last measurements are not included in the measurement
result.
 highest result: The maximum measured value is not included in the measurement
result.
4. Click Save to confirm the entry.

822-129-8 59 / 90
User-defined measurement DTA 100 C

11.1.12 Saving user-defined measurements


After you have defined the parameters for a user-defined measurement, you can save this
measurement.
Main menu > User-defined measurement > selected measurement

1. Select Back.
You will be asked if you want to save the measurement with modified parameters.

2. Select Yes.
The measurement will be saved.

60 / 90 822-129-8
DTA 100 C User-defined measurement

11.2 Running a user-defined measurement


Main menu > User-defined measurement

1. In the main menu, select the menu item User-defined measurement.


2. Select the menu item Next to confirm the selection.
3. In the User-defined measurement menu, select a saved measurement, followed by the
menu item Next.

4. Select the menu item run, followed by the menu item Next.
A prompt appears on the display to set an electrode gap. The required electrode shape
is displayed pictographically.

5. Check whether the correct electrodes have been selected. If the wrong electrodes are
used, replace them.
Further information: Chapter Replacing the electrodes (on page 27)

822-129-8 61 / 90
User-defined measurement DTA 100 C

6. Set the indicated electrode gap.


Further information: Chapter Setting an electrode gap (on page 28)
7. Select menu item Start.
A prompt to enter a sample number appears on the display. The oil breakdown voltage
tester will suggest a sample number composed of the date and time in DDMMYYhhmm
format. This sample number can be extended as required.

8. If you want to amend the suggested sample number, do so or enter any sample number
with the number keys on the membrane keyboard.
9. Click Save to confirm the entry.
The measurement begins. A window displays the details of the measurement process.
The symbol indicates that there is high voltage at the electrodes.
When the oil breakdown voltage tester has completed the measurement, the
measurement log is shown on the display. If the printer is switched on, the measurement
log will automatically be printed first.
10. Select the menu item Details in the lower menu bar for a detailed view of the
measurement results.
11. Select the menu item Back to perform further measurements or to stop the
measurement process.
The oil breakdown voltage tester switches to the beginning of the measurement. A
prompt appears on the display to set the electrode gap.
12. Repeat the process to perform additional measurements.
Select the menu item Back to stop the measurement process.
The oil breakdown voltage tester switches to the main menu.

Cancelling the measurement manually


 Select menu item Stop.

62 / 90 822-129-8
DTA 100 C User-defined measurement

11.3 Editing or deleting a user-defined measurement


Main menu > User-defined measurement

1. Select a saved measurement by pressing the arrow keys.


2. Select the menu item Next to confirm the selection.
3. Choose one of the following settings by pressing the arrow keys:
 edit: This will take you to the menu for editing the parameters of the selected
measurement where you can edit the settings
Further information: Chapter Creating a user-defined measurement (on page 52)
 delete: The device deletes the selected measurement.
4. Select the menu item Next to confirm the selection.

822-129-8 63 / 90
Display of measurement results DTA 100 C

12 D ISPLAY OF MEASUREMENT RESULTS

Main menu > Measurement logs

1. Choose a measurement log by pressing the arrow keys.


2. Select the menu item Next to confirm the selection.
3. Choose one of the following settings by pressing the arrow keys:
 show: The selected measurement log is displayed.
 print: The selected measurement log is printed.
 delete: The selected measurement log is deleted.
 delete all: All measurement protocols that are saved on the device are deleted.
4. Select the menu item Next to confirm the selection.

64 / 90 822-129-8
DTA 100 C Display of measurement results

12.1 Measurement log as printout


Main menu > Measurement logs > print

Measurement log
BAUR DTA 100 C
Version 1.16
06.03.2019 10:12
----------------------------------------------------
Serial number:
0801900001
Sample number:
1002091355

Standardised measurement:
IEC 60156:2018

Electrode shape:
IEC 60156 Fig. II
Distance: 2.5 mm

Test frequency: 60 Hz
----------------------------------------------------
Temperature 20 °C

Measurement 1 90.2 kV
Measurement 2 73.2 kV
Measurement 3 69.3 kV
Measurement 4 82.9 kV
Measurement 5 69.1 kV
Measurement 6 71.7 kV

Avg. value 76.1 kV


Stand. dev.* 8.6 kV
Std.dev./Avg.* 11.3 %

Test completed!

----------------------------------------------------
Test performed by:

..............................................................

* Abbreviations:

Stand. dev.: Standard deviation


The standard deviation shows how close the individual recorded values are
to the average value.

Std.dev./Avg.: Ratio of the standard deviation to the average value in % (coefficient of


variation).
In IEC 60156, diagram 3 in section 11 illustrates the ratio of the standard
deviation to the average value. This diagram and the determined values can
be used to check whether the measurement result lies within the permitted
range.

822-129-8 65 / 90
Display of measurement results DTA 100 C

12.2 Measurement log on the display


Main menu > Measurement logs > show

 Use the arrow keys to scroll through the text.

66 / 90 822-129-8
DTA 100 C Putting the device out of operation

13 P UTTING THE DEVICE OU T OF OPERATION

1. Close the operating unit.


The firmware stops.
2. There is a mains switch on the back. Turn off the oil breakdown voltage tester.
3. To disconnect the device completely from the mains voltage, pull out the mains plug.
4. Open the protective cover.
5. Remove the test vessel.
6. Empty the test vessel and dispose of the oil sample in an environmentally friendly
manner and in accordance with the applicable national regulations.
7. To prevent contamination of the test vessel, store the test vessel as follows:
 Fill the test vessel with new, filtered oil.
or
 Clean the test vessel with petroleum ether and store the closed test vessel in a dust-
free environment.
Further information: IEC 60156, ASTM D877, chapter Cleaning the oil breakdown
voltage tester (on page 72)
8. Close the protective cover.

822-129-8 67 / 90
BAUR ITS Lite software for measurement data management DTA 100 C

14 BAUR ITS L ITE SOFTWARE FOR MEASUREMENT DATA


MANAGEMENT

The BAUR ITS Lite software is used to read measurement results and to design and archive
measurement logs of the BAUR DTA 75 C, DTA 100 C and DTL C insulating oil testers.

14.1 Description of function


Reading of measurement logs from BAUR insulating oil testers and archiving with ITS Lite
takes place automatically.
To read the measurement logs, start ITS Lite and define the save and print settings. The
following settings are available in ITS Lite:
 Language selection for measurement logs. Both the software interface and user manual
are displayed in the selected language.
 Save format selection for measurement logs: PDF or as a text file (*.TXT).
 Integrate logo or header in PDF measurement logs.
Now you only need to connect one or more insulating oil tester to the PC on which ITS Lite
will run. If measurement logs already exist in the device memory of a connected insulating oil
tester, ITS Lite will read the logs as soon as a connection is established between the
insulating oil tester and the PC. These measurement logs will be saved in the predefined
storage location (path) according to the settings, but will not be printed.
When a connected insulating oil tester completes a new measurement, the measurement
data is automatically transmitted to ITS Lite and is processed as follows:
 A measurement log is generated automatically from the received measurement data and
is saved in PDF format and/or as a text file (*.TXT) in the predefined storage location.
 If the Print option is enabled, the measurement log is printed on the selected printer.
The measurement log contains the following details:
 Information on the insulating oil tester (device type, serial number, firmware version),
 Measurement time (date and time),
 Sample number and name of measurement log,
 Measurement parameters and measured values,
 Measurement diagram with the graphical display of the test voltage (DTA 75 C and
DTA 100 C) or with the graphical display of the measured tan δ values (DTL C), as
required.

68 / 90 822-129-8
DTA 100 C BAUR ITS Lite software for measurement data management

14.2 Installing ITS Lite


Prerequisites
 The system requirements are met.
 The ITS Lite installation package has been downloaded (free download at
http://www.baur.eu).
 You have Administrator rights for the PC.

Procedure
1. Double click on the ITS Lite Setup.exe file in the folder where you saved the ITS Lite
installation package.
2. In the Welcome to the BAUR ITS Lite Setup Wizard window, click on Next.
3. In the End-User License Agreement window, read the license agreement carefully.
4. Select the I accept the terms in the License Agreement option if you agree with the
terms of the license agreement.
This is the prerequisite for continuing the installation of ITS Lite.
5. Click on Next.
6. In the Configure Setup window, perform the following steps:
 To automatically launch ITS Lite when Windows starts, activate the Run ITS Lite
when Windows is started checkbox.
 To install the driver required to communicate with the insulating oil tester, activate
the Install FTDI driver checkbox (recommended).
 Click on Next.
7. In the Ready to install BAUR ITS Lite window, click Install to start the installation.
ITS Lite will be installed.
8. In the Completed the BAUR ITS Lite Setup Wizard window, click on Finish to end the
installation process.

14.3 Starting ITS Lite


 Select ITS Lite from the Windows program list.
ITS Lite is started. The ITS Lite icon is displayed in the Taskbar.

 Follow the user manual for the BAUR ITS Lite software for
measurement data management (integrated in the software). You
can call up the user manual in the software or download it for free
(http://www.baur.eu).

822-129-8 69 / 90
Checking the measurement accuracy of the oil breakdown voltage tester (optional) DTA 100 C

15 C HECKING THE MEASUREM ENT ACCURACY OF THE OIL


BREAKDOWN VOLTAGE TESTER ( OPTIONAL )

You can check the voltage values with a 0.7 % measuring accuracy using a KA DPA/DTA C
(option) calibrator.
Further information: User manual for the KA DPA/DTA C calibrator.
Main menu > Tools > Measurement with KA DPA/DTA C

70 / 90 822-129-8
DTA 100 C Maintenance

16 M AINTENANCE

DANGER

Dangerous electric voltage


Danger to life or risk of injury due to electric shock
 Work on electrical devices may only be performed by qualified
technical staff.

NOTICE

Damage to device due to improper handling


The user is liable for damages caused due to improper maintenance or care.
 Never take apart the device. This can lead to device damages. Inside the device
there are no components that could be serviced or repaired by the user.
 Maintenance tasks must be carried out only by personnel trained and authorised
by BAUR

NOTICE

Material damage caused by unauthorised spare parts


 Use only accessories and original spare parts recommended by BAUR. The use
of spare parts, accessories and special fittings that have not been tested and
approved by BAUR could adversely affect the safety, function and features of the
product.
The use of any unauthorised spare parts will invalidate the warranty.

822-129-8 71 / 90
Maintenance DTA 100 C

16.1 Cleaning the oil breakdown voltage tester

DANGER

Electrical voltage on device


Danger to life or risk of injury from electrical voltage.
 Switch off the device before cleaning.
 To disconnect the device completely from the mains voltage, pull
out the mains plug.

WARNING

Fire hazard posed by petroleum ether or other cleaning solvent


We recommend using cleaning solvent to clean the test chamber of the
oil breakdown voltage tester and the test vessel. The cleaning solvent
is highly flammable and in some circumstances may cause a fire.
 Do not smoke when working with cleaning solvent.
 Avoid naked flames.

NOTICE

Damage to the device may be caused by using the wrong cleaning agents
 Do not use any abrasive, corrosive cleaning agents or strong solvents.
 Ensure material compatibility.
 Do not clean the product with acetone or thinner.
 Never clean electrical devices with water.

Prerequisites

Safety gloves to prevent coming into contact with insulating oil

 Mild detergent for cleaning the surfaces of the device


 Petroleum ether or other similar mild solvent for cleaning the test chamber
 Lint-free cleaning cloth

72 / 90 822-129-8
DTA 100 C Maintenance

Cleaning the test chamber

1 Test vessel isolators


2 HV isolators
3 Oil collecting tray
4 Test vessel

1. Always keep the oil breakdown voltage tester clean.


The cleanliness of the oil breakdown voltage tester and the test cell will have a strong
influence on the measurement results.
2. Always wipe off spilled or leaked oil with petroleum ether or other similar mild solvent
and a lint-free cloth, in particular:
 on the test vessel isolators
 in the oil collecting tray (test chamber floor)
 on the test vessel
3. Allow the oil breakdown voltage tester to dry well before closing it.
Note: To avoid impurities in the test vessel isolators, place the test vessel on an even, clean
and dry base when you remove the test vessel from the oil breakdown voltage tester.

Cleaning HV isolators
BAUR GmbH recommends cleaning the HV isolators at least once a year to guarantee
maximum accuracy.
HV isolators may be cleaned only by qualified technical staff authorised by BAUR.
In case of questions, please contact BAUR GmbH or your local BAUR representative
(http://www.baur.eu/baur-worldwide).

Cleaning the display


 Clean the displays with a dry or slightly damp lint-free cloth.

822-129-8 73 / 90
Maintenance DTA 100 C

Cleaning the device surfaces and connection cable


1. Clean the device surfaces and connection cable with mild detergent and a lint-free cloth.
2. NOTICE! Damage to device due to leaking fluids. Do not allow liquids to leak into
devices.

16.2 Replacing the printer paper roll


Prerequisites

Paper scissors for cutting the paper roll.

Paper roll for printer, plain paper, Ø 30 cm, width 57 mm


Order number: 565-514

Procedure

1 Cover
2 Latches
3 Paper roll
4 Paper roll holder
5 Groove

1. Remove the cover (1) from the printer.


2. Press both latches (2) inwards and remove the printer.
3. Remove the paper roll holder (4).
4. Remove the empty paper roll.

74 / 90 822-129-8
DTA 100 C Maintenance

5. Use scissors to cut the end of the new paper roll in the shape of a trapeze.

6. Put the paper roll holder with the new paper roll into the printer.

1 Paper roll
2 Paper feed
3 Paper feed button LF/SEL

7. Guide the end of the new paper roll into the paper feed (2).
Make sure that the paper is folded in such a way that the printer can pull it in easily.
8. Press the paper feed button LF/SEL (3) until the end of the paper protrudes by a few
centimetres on the top side of the printer.
9. Tighten the paper roll slightly.
10. Place the printer in the groove so that it snaps into place.
11. Place the cover on the printer.

822-129-8 75 / 90
Maintenance DTA 100 C

16.3 Replacing the printer ink ribbon


Prerequisites

Printer ink ribbon, blue


Order number: 565-513

Procedure

1 Cover
2 Printer ink ribbon

1. Remove the cover (1) from the printer.


2. Replace the printer ink ribbon (2).
3. Place the cover (1) on the printer.

76 / 90 822-129-8
DTA 100 C Maintenance

16.4 Replacing the fuse


Prerequisites

Mains fuse: 250 V / 4 A time lag, rated breaking capacity 1500 A (H)

Procedure

1 Latches
2 Fuse cartridge
3 Fuses
4 Groove

1. Press both latches (1) inwards and remove the fuse cartridge (2).
2. Replace the fuses (3).
3. Place the fuse cartridge (2) in the groove (4) so that it snaps in place.

16.5 Calibration
The product has been inspected, adjusted and calibrated for its functions before delivery.
The customer will receive a log which will also document the feedback of calibration
quantities. If required, a test and calibration log can be delivered with the proof of return of
the calibration to national and international standards.
BAUR GmbH recommends calibrating the device once a year to guarantee maximum
accuracy.
In case of questions, please contact BAUR GmbH or your local BAUR representative
(http://www.baur.eu/baur-worldwide).

822-129-8 77 / 90
Maintenance DTA 100 C

16.6 Ordering accessories and spare parts


 Only use accessories and original spare parts recommended by BAUR.
 Order accessories and spare parts from your nearest BAUR representative
(http://www.baur.eu/baur-worldwide).

Order number Description

670-087 Transport case

670-085 Dust cover

565-513 Printer ink ribbon, blue

565-514 Paper roll for printer, plain paper, Ø 30 mm, width 57 mm

650-209 Face pin wrench for disassembling the test vessel

650-129 Magnetic stirrer

650-229 Lifting stick for magnetic stirrer

432-270 Setting gauge 1 mm in accordance with ASTM D1816

430-109 Setting gauge 2 mm in accordance with ASTM D1816

430-110 Setting gauge 2.5 mm in accordance with IEC 60156

430-105 Setting gauge 2.54 mm in accordance with ASTM D877

431-298 Setting gauge 4 mm in accordance with BS EN 60156

430-111 Setting gauge 5 mm in accordance with SEV EN 60156

415-538 Pair of electrodes in accordance with IEC 60156 Fig. II

415-539 Pair of electrodes in accordance with IEC 60156 Fig. I

415-540 Pair of electrodes in accordance with ASTM D877

415-523 Test vessel 0.4 litre made of glass in accordance with IEC 60156 Fig. I

415-522 Test vessel 0.4 litre made of glass in accordance with IEC 60156 Fig. II

415-525 Test vessel 0.4 litre made of glass in accordance with ASTM D1816

415-524 Test vessel 0.4 litre made of glass in accordance with ASTM D877

415-510 BAUR Report Manager – External USB interface for measurement data
management

78 / 90 822-129-8
DTA 100 C Faults

17 F AULTS

DANGER

Dangerous electric voltage


Danger to life or risk of injury due to electric shock
 Work on electrical devices may only be performed by qualified
technical staff.

NOTICE

Damage to device due to improper handling


The user is liable for damages caused due to repairs.
 Never take apart the device. This can lead to device damages. Inside the device
there are no components that could be serviced or repaired by the user.
 Repairs must be carried out only by personnel trained and authorised by BAUR

17.1 Troubleshooting
When a fault occurs, proceed as follows:
1. Check the supply voltage, the connection cable and earth cable.
2. Pay attention to the message on the display.
Further information: Chapter Error messages and corrective measures (on page 80)
3. Restart the oil breakdown voltage tester.
4. If the error occurs again after the device has been restarted, contact your nearest BAUR
representative (http://www.baur.eu/baur-worldwide).
It may be possible for the BAUR GmbH After Sales Service Team to determine the
cause of the fault remotely. To do so, please specify the following data:
 Oil breakdown voltage tester serial number
 Firmware version
 Message on display
 Procedure that caused the error.

822-129-8 79 / 90
Faults DTA 100 C

17.2 Error messages and corrective measures

Error message Possible cause Corrective measure

Output voltage lies Impaired measurement  Repeat the measurement.


outside tolerance conditions
range! Measurement  Conduct a reference measurement, e. g. with a
cancelled. specific withstand voltage.

No gap between the  Check whether there is a gap between the electrodes.
electrodes
 If the electrodes are touching one another, adjust the
setting to a suitable gap.
Note: If no signal sounds when the electrodes short
circuit, please contact your BAUR representative
(http://www.baur.eu/baur-worldwide).

Test vessel positioned  Check whether the test vessel is seated correctly on
incorrectly on the the isolators and contacts and correct the position if
isolators necessary.

Faulty test vessel  Repeat the measurement with another test vessel and
a slightly higher withstand voltage.

Conductive oil sample  Dispose of the contaminated oil sample in an


due to contamination by environmentally friendly manner and in accordance
water with the applicable national regulations.
 Clean the test vessel with petroleum ether or other
similar mild solvent.
Further information: Chapter Cleaning the oil
breakdown voltage tester (on page 72)
 Repeat the measurement with a new oil sample.

Output voltage lies –  If you want to continue with the measurement despite
outside tolerance the deviation in the output voltage, select Yes.
range! Continue The measurement will be continued. The
anyway? corresponding measured values will be identified with
(during the spot test, an asterisk in the measurement log.
the measurement is
cancelled in this case) Impaired measurement  Repeat the measurement.
conditions
 Conduct a reference measurement, e. g. with a
specific withstand voltage.

The oil breakdown  Clean the oil breakdown voltage tester.


voltage tester is not Further information: Chapter Cleaning the oil
clean breakdown voltage tester (on page 72)

The test vessel is not  Clean the test vessel with petroleum ether or other
clean similar mild solvent.
Further information: Chapter Cleaning the oil
breakdown voltage tester (on page 72)

HV isolators are not  HV isolators may be cleaned only by qualified


clean technical staff authorised by BAUR.
In case of questions, please contact BAUR GmbH or
your local BAUR representative
(http://www.baur.eu/baur-worldwide).

Relative permittivity of  Check the relative permittivity of the oil sample.


the oil sample is too high The relative permittivity εr must be less than 30.

80 / 90 822-129-8
DTA 100 C Transportation and storage

18 T RANSPORTATION AND STORAGE

NOTICE

Damage to the device caused by improper transportation and incorrect storage


 Always transport and store the device as intended.
 Comply with the ambient conditions specified in the technical data for this
device.

18.1 Packaging
1. Keep the original packaging because it provides the best protection for your oil
breakdown voltage tester during transportation.
2. If you would like to dispose of the packaging, ensure you comply with the applicable
national regulations when doing so.

18.2 Transportation
If you are sending the oil breakdown voltage tester to BAUR GmbH, a BAUR representative
or to the Technical Service for repairs or any other reason, ensure the following:
 The DTA 100 C oil breakdown voltage tester weighs up to 39 kg. It is recommended to
seek the help of another person to lift or carry the device.
 To avoid damage during transportation or as a result of faulty packaging, it is best to use
the original packaging.
If none of the original packaging is available, select packaging that is strong enough to
protect against mechanical damage and the ingress of liquids.
 The oil breakdown voltage tester must be transported and shipped on a pallet.

Note: The transport case does not constitute shipping packaging. Do not ship the oil
breakdown voltage tester in the transport case without a pallet.

822-129-8 81 / 90
Warranty and After Sales DTA 100 C

 Dismantle the glass test vessel before transporting the oil breakdown voltage tester.
The glass test vessel is very delicate. To avoid damage during transportation, pack the
glass test vessel as securely as possible to ensure it does not break.
 NOTICE! Damage to device due to improper transport. Transport the oil breakdown
voltage tester in an upright position only.
 Protect the oil breakdown voltage tester against strong vibrations.
 Protect the oil breakdown voltage tester against moisture.

18.3 Storage
 Store the oil breakdown voltage tester in an upright position only.
Storage temperature: -20 °C to +60 °C
 Always store the oil breakdown voltage tester with the cover closed.
The rubber seals integrated in the cover protect the HV isolators and contacts from dirt
and dust.
 Protect the oil breakdown voltage tester against moisture.
 Protect the oil breakdown voltage tester against unauthorised access.

19 W ARRANTY AND A FTER S ALES

Warranty
For warranty claims, please contact BAUR GmbH or your local BAUR representative
(http://www.baur.eu/baur-worldwide). Warranty is cancelled in case of misuse. Wear parts
are excluded from the warranty.

After Sales
For questions contact BAUR GmbH or your BAUR representative (http://www.baur.eu/baur-
worldwide).

BAUR GmbH
Raiffeisenstraße 8
6832 Sulz / Austria
[email protected]
www.baur.eu

82 / 90 822-129-8
DTA 100 C Disposal

20 D ISPOSAL

20.1 Disposing of the device


BAUR devices do not belong in the domestic waste.
 Dispose of the device in an environmentally friendly manner and in accordance with the
applicable national regulations.

20.2 Disposing of the insulating oil


 Dispose of the insulating oil in an environmentally friendly manner and in accordance
with the applicable national regulations.

822-129-8 83 / 90
Delivery includes and Options DTA 100 C

21 D ELIVERY INCLUDES AND O PTIONS

84 / 90 822-129-8
DTA 100 C Declaration of conformity

22 D ECLAR ATION OF CONFORMITY

We

BAUR GmbH
Raiffeisenstraße 8
6832 Sulz / Austria
[email protected]
www.baur.eu
declare, under our sole responsibility, that the BAUR product
BAUR oil breakdown voltage tester DTA 100 C
to which this declaration refers, conforms to the following standards or standard documents:
 Low Voltage Directive 2014/35/EC
EN 61010-1:2010
 EMC Directive 2014/30/EU
EN 55011:2009 + A1:2010
EN 61000-4-2:2009
EN 61000-4-4:2012
EN 61000-4-5:2014
EN 61000-4-11:2004
 Environmental testing
EN 60068-2-ff

Signed: Torsten Berth, Technical Director


Dr. Eberhard Paulus, Director QM/QS

Sulz, 30/11/2015

822-129-8 85 / 90
Index DTA 100 C

23 I NDEX

Declaration of conformity • 85
A
Delivery includes and Options • 84
About this manual • 6
Description of function • 68
Adjusting the stirrer • 37
Device settings • 33
Applicability of the instructions • 6
Display individual breakdown values
Avoid dangers, take safety measures
• 39
• 10
Display of measurement results • 64
B
Disposal • 83
Basic insulating oil test procedure •
Disposing of the device • 83
20
Disposing of the insulating oil • 83
BAUR ITS Lite software for
measurement data management • 68
E
C Earthing the oil breakdown voltage
tester • 26
Calibration • 77
Editing or deleting a user-defined
Carry out a standardised
measurement • 63
measurement • 45
Entering numbers • 21
Checking the measurement accuracy
of the oil breakdown voltage tester Error messages and corrective
(optional) • 70 measures • 80
Checks to perform before
F
commissioning • 25
Faults • 79
Clean the electrodes • 28
Fill and use the test vessel • 29
Cleaning the oil breakdown voltage
tester • 72 For your safety • 9
Commissioning • 25 Full illustration • 13
Configure settings for user-defined
I
measurements • 41
Configure the ASTM settings • 39 Information on the oil breakdown
voltage tester • 24
Configure the IEC 60156
Information on the USB interface • 24
2018 settings • 35
Insert the test vessel • 31
Continuous measurement template •
Installing ITS Lite • 69
53
Creating a user-defined Installing the oil breakdown voltage
measurement • 52 tester • 26
Instructions for sampling • 30
D
Instructions to the user • 9
Dangers when working with electric
Intended use • 10
voltage • 11

86 / 90 822-129-8
DTA 100 C Index

M Select language • 34
Main menu • 23 Selecting a template • 54
Maintenance • 71 Selecting which measured values are
not to be evaluated • 59
Measurement log as printout • 65
Set date • 42
Measurement log on the display • 66
Set display brightness • 33
Menu navigation • 21
Set the hold time before the first
N measurement • 40
Note on the screenshots and Set time • 43
graphics used • 8
Setting an electrode gap • 28
O Setting the duration of the pauses
between measurements • 56
Operating and display elements • 16
Setting the electrode shape • 38
Operating the oil breakdown voltage
tester • 21 Setting the hold time before the first
measurement • 36, 54
Ordering accessories and spare
parts • 78 Setting the maximum output voltage •
57
Overview of standards • 44
Setting the number of measurements
Overview of templates • 53 • 57
Overwrite measurement logs • 35 Setting the slew rate for the test
voltage • 55
P
Setting the stirring time during the
Packaging • 81 hold time • 55
Power supply • 17 Setting the stirring time in the pauses
Product information • 13 • 56
Putting the device out of operation • Setting the withstand voltage • 58
67 Special personal protective
equipment • 12
Q
Standardised measurement • 44
Quick test • 48
Starting ITS Lite • 69
R Storage • 82
Rating plate • 18 Structure of safety instructions • 6
Replacing the electrodes • 27 Switch printer on/off • 34
Replacing the fuse • 77 Symbols and abbreviations on the
Replacing the printer ink ribbon • 76 display • 22
Replacing the printer paper roll • 74 T
Reset settings • 43 Technical data • 19
Running a user-defined Transportation • 81
measurement • 61
Transportation and storage • 81
S Troubleshooting • 79
Saving user-defined measurements • Turning on the oil breakdown voltage
60 tester • 32

822-129-8 87 / 90
Index DTA 100 C

U
USB interface • 17
User-defined measurement • 51
Using this manual • 6

V
View Settings • 7

W
Warranty and After Sales • 82
Withstand voltage template • 53

88 / 90 822-129-8
822-129-8

BAUR GmbH
Raiffeisenstraße 8
6832 Sulz / Austria
[email protected]
www.baur.eu

822-129-8-phd-07.03.2019

You might also like