0020-410 MicroLab Software Manual
0020-410 MicroLab Software Manual
Operation Manual
Notices
Manual Part Number Warranty Safety Notices
0020-410 The material contained in this CAUTION
Edition 8, November 2019 document is provided “as is,” and is
subject to being changed, without A CAUTION notice denotes a hazard. It
notice, in future editions. Further, to calls attention to an operating
Copyright the maximum extent permitted by
procedure, practice, or the like that, if
© Agilent Technologies, Inc. 2019 applicable law, Agilent disclaims all
warranties, either express or implied, not correctly performed or adhered to,
No part of this manual may be with regard to this manual and any could result in damage to the product
reproduced in any form or by any information contained herein, or loss of important data. Do not
means (including electronic storage including but not limited to the implied
and retrieval or translation into a proceed beyond a CAUTION notice
warranties of merchantability and
foreign language) without prior fitness for a particular purpose. until the indicated conditions are fully
agreement and written consent from Agilent shall not be liable for errors or understood and met.
Agilent Technologies, Inc. as governed for incidental or consequential
by United States and international damages in connection with the
copyright laws. furnishing, use, or performance of this
document or of any information
contained herein. Should Agilent and
Agilent Technologies Australia [M] Pty the user have a separate written
Ltd agreement with warranty terms
679 Springvale Road covering the material in this document
Mulgrave, Victoria 3170, Australia that conflict with these terms, the
warranty terms in the separate
www.agilent.com agreement shall control.
Technology Licenses
Software Revision The hardware and/or software
This guide is valid for the 5.6.2064 described in this document are
revision or higher of Agilent MicroLab, furnished under a license and may be
until superseded. used or copied only in accordance with
the terms of such license.
2 MicroLab Mobile 9
Connecting the PDA to the instrument for Bluetooth Communication 9
Logging on for the first time 11
Software IQ on 4300 Handheld FTIR 12
Managing users 13
Adding 13
Role definitions 14
Common icons 15
Home screen 15
Status 15
Start 15
Method 16
Reference Method 16
Log Off 16
Lock 16
Menu 16
Method screen 17
Methods list 18
Activate 18
Editing 18
3 MicroLab PC 19
Home screen 19
Status 19
Logoff 19
Lock 19
Previous Results 20
Advanced Features 20
Methods 20
Editing 21
Multivariate method transfer from GRAMS PLS/IQ to MicroLab PC 28
Exporting the PLS/IQ *.CAL file to ASCII 28
Generating the method file 28
MicroLab quantitative model transfer from MicroLab Expert to MicroLab
PC 30
Using Pre-Sample Alignment Check 31
Transferring edited methods to MicroLab Mobile 32
Using MicroLab Lite Synchronize 32
Copying via My Computer 33
Moving files in a 21 CFR part 11 enabled environment 34
System Check 34
4 MicroLab Quant 35
Home screen 35
Logoff 36
Lock 36
Start New Model 36
Import Models 40
21 CFR Part 11 Features 41
21 CFR part 11 41
E-Sign 41
5 MicroLab OQ 43
Operational Qualification (OQ) 43
Diagnostic 43
Performance Validation 44
Perform the OQ 44
OQ Report 45
7 Measuring Samples 53
Cleaning the sample interface 53
Collecting a background measurement 54
Collecting the sample measurement 54
8 Reviewing Results 57
MicroLab Mobile 57
Results screen 57
Handling data 57
Previous results 59
MicroLab PC/MicroLab Lite 60
Results screen 61
Importing spectra into GRAMS 64
MicroLab Report Creation 65
Assigning a Report Template to a Method 65
Creating a new Report Template 65
1 Introduction
Supported Operating Systems 8
Supported Instruments 8
The Agilent MicroLab software suite is designed to run Agilent 4100 ExoScan,
4200 FlexScan, 4300 Handheld FTIR, 4500 Series, 5500 Series and Cary 630 FTIR
spectrometers. The software is available in two different packages:
MicroLab FTIR Software, and MicroLab Mobile Software.
MicroLab FTIR software package consists of the following four modules:
MicroLab PC
MicroLab Quant for quantitative method development
MicroLab OQ for operational qualification
MicroLab Lite for data review and method development
NOTE MicroLab Mobile software package is only applicable with 4300 Handheld FTIR and
instruments that used to ship with a handheld PC (PDA): Agilent 4100 ExoScan,
4200 FlexScan and 4500 Series.
Agilent no longer ships PDA with these instruments.
The software incorporates a user logon; each user has defined ‘roles’ that are
assigned by the administrator. These roles enable administrators to limit the
methods and actions allowed for lower level users.
The MicroLab PC and Lite Software can be used on the following operating
systems:
Windows 7 Pro/Enterprise 64-bit SP1
Windows 10 Pro/Enterprise 64-bit
English, Japanese, and Simplified Chinese
MicroLab Mobile requires the Windows Mobile 6.0 or higher.
The 4300 Handheld FTIR requires Windows CE 6 installed on the 4300
Handheld FTIR's integrated CPU.
The 4500 Portable FTIR is supported on the Getac RX-10 tablet with English
Windows 10 Pro 64-bit.
MicroLab 21 CFR Part 11 Pharma Software can only be used with the following
English only operating systems:
Microsoft Windows 7 Pro or Enterprise 64-bit SP1
Microsoft Windows 10 Pro or Enterprise 64-bit operating systems
MicroLab 21 CFR Part 11 Pharma Software cannot be used with:
Microsoft Windows 7 Pro or Enterprise 32-bit SP1
Microsoft Windows 10 Pro or Enterprise 32-bit operating systems
Operating systems other than English
Supported Instruments
This software build is suitable for use with the Agilent Cary 630, 4100 ExoScan,
4200 FlexScan, 4300 Handheld, 4500 Portable and 5500 Compact FTIR
instruments.
2 MicroLab Mobile
Connecting the PDA to the instrument for Bluetooth Communication 9
Logging on for the first time 11
Software IQ on 4300 Handheld FTIR 12
Managing users 13
Common icons 15
Home screen 15
Method screen 17
NOTE MicroLab Mobile software package is only applicable with 4300 Handheld FTIR and
instruments that used to ship with PDA: Agilent 4100 ExoScan, 4200 FlexScan and
4500 Series.
NOTE This section is only applicable to Agilent 4100 ExoScan, 4200 FlexScan and 4500
FTIR Series instruments.
1 To configure the Bluetooth connection between the instrument and the PDA,
click on the Bluetooth icon from the home page. If “flight mode” is enabled,
disable it. Click on the Bluetooth bar to turn the Bluetooth Communication on if
it is not already in the “ON” mode.
2 From the Menu, choose Bluetooth Settings.
3 In the Devices tab, clear any devices by selecting the device with the Stylus and
holding it in place until the menu appears. Then click Delete.
4 Once the devices are cleared, click on Add a New Device. Wait for the list of
Bluetooth enabled devices to appear and select the appropriate device. Click
Next at the bottom of the screen.
5 The connection with the handheld devices requires a universal passcode. The
passcode required is 7890. Enter this value and select Next. The system will
then test the connection to the portable FTIR Device, and add the device.
6 Once the device is added, select Advanced and ensure that the Serial Port
configuration box is checked. The box defaults to unchecked, and must be
turned on in order for the device to work properly. Click Save.
The advanced option only appears right after the connection. If it
disappears and returns to the device tab, select the device just connected
by clicking on it with the stylus. This will bring you to the Advanced Portion
above for the serial port communication setting.
7 Once the serial port has been set, select the COM Ports tab at the bottom of the
Bluetooth settings screen.
8 Select New Outgoing Port. Select the device just added to the Bluetooth
communication configuration and click Next.
9 The default COM Port for version 4.0 and later of MicroLab Mobile is COM 1.
Set the COM Port to COM 1 and UNCHECK the Secure Connection box. Once
you have done this, select finish in the bottom section of the screen.
If there is a communication issue with the COM Port selected, the PDA will
alert you via an error message. The COM Port default for communication
can be changed via communication file for the portable/handheld device.
Please contact your Agilent support personnel if additional installation
instructions are needed to change the COM Port settings.
10 This will bring you back to the initial Bluetooth setting Configuration screen.
Click OK in the upright corner and close any programs open until you return to
the main home screen.
11 Choose the Start Menu and MicroLab Mobile should appear as a selection.
For 21 CFR Part 11 environments: Login with a profile that has the MicroLab
Mobile privilege. Enter the appropriate user name for the profile created in SCM.
The Groups and Projects fields will populate automatically. Enter the password
and then click Login.
NOTE This section is only applicable to the Agilent 4300 Handheld FTIR.
Managing users
NOTE Only users with administration privileges can access User Management.
This feature is not available when the 21 CFR Part 11 environment is enabled. All
user management functions are performed in the Spectroscopy Configuration
Manager (SCM) software. For more information see the SCM User’s Guide provided
with the software.
Adding
To add a user to the system:
1 Tap the Advanced Features button on the MicroLab Mobile Home screen.
2 Click User Management.
3 Click Add New User.
4 Enter a personalized UserID in the text box.
5 Enter the user’s complete name.
6 Insert a new password.
7 Identify the role of the new user (Admin, Service, Developer, and Technician).
8 Click Apply Changes.
9 A window will appear notifying you that you will be logged off the system. You
will be forced to login again after the changes have been applied.
NOTE The system will automatically log off the user. The new UserID will appear in the
User Name drop-down menu on the Logon screen.
Role definitions
Service: The Service role provides access to all instrument parameters within a
method including the sampling technology applicable for the method.
Admin: The Admin (administration) role is reserved for the person(s) responsible for
continual maintenance and changes to MicroLab software, including editing
methods, creating new methods, defining public methods, conducting instrument
verification, and controlling user access.
Developer: The Developer role allows access to system performance validation in
addition the ability to edit and run methods. However, this role does not have
access to User Management functions or system diagnostics.
Technician: The Technician role is for standard users of the system. The technician
only has access to run methods defined by Admin as public. In addition, the
technician cannot review data, conduct system validation or modify methods.
E-Sign: Designation used to allow users to e-sign methods and results at various
levels of data acquisition and data review. This level is typically only used in the 21
CFR Part 11 version of the MicroLab Software.
NOTE Only the Admin and Developer roles have access to Advanced Features.
Common icons
Home screen
The Home screen is the first displayed screen in MicroLab following logon. From
the Home screen, you can start a method, measure a reference sample, open the
Method Selection dialog box, open the Previous Results dialog box or open the
Advanced Features dialog box. The sampling module or interface should be
changed at the Home screen All procedures begin and end at the Home screen.
Status
In the upper right corner of the screen, a small circle appears. This circle gives a
color indicating the status of the instrument connected to the MicroLab software.
The circle will be green when the system is operating within the factory-defined
parameters. A red or yellow light indicates a parameter outside of specification.
When a red or yellow light condition cannot be resolved, contact Agilent for
assistance. All parameters used to determine system readiness are listed in
Advanced Features > Diagnostics. Parameters outside limits will be listed in red on
the Diagnostics screen.
Just above the ready light is a battery indicator. The battery indicator shows the
level of charge remaining on the instrument batteries. The charge level of the
batteries running the handheld PC is not indicated on the MicroLab software.
Start
The Start button begins a sample measurement using the currently selected
method. The currently selected method is shown in the upper left corner of the
Home screen.
Method
The Method button opens the Method Selection dialog box. Users without Admin
privileges can only select methods designated as Public by the administrator. See
the Method Selection instructions below.
Reference Method
The Reference Method button enables the collection of a reference sample for a
component method. Some component methods present concentrations as a
percentage increase or decrease. For these calculations, a reference of the in-
specification sample is needed to calculate these starting values. Also, the
Reference Method saves the spectrum of the in specification (new) material for
later visual comparison.
Log Off
The Log Off button exits the current user and returns the MicroLab Mobile software
to the Logon screen. From there, the Exit button can be tapped to close the
software.
Lock
This is available in 21 CFR Part 11 environments only. Use to Lock the software to
control access while unattended. See the MicroLab 21 CFR Part 11 section on page
47 for more information.
Menu
The menu contains:
Previous Results
Advanced Features
Previous Results
Enables viewing of previously collected data. Only Service, Admin and Developer
roles have access to Previous Results.
21 CFR part 11
With a Results file open, click the 21 CFR part 11 button to display an audit trail with
a list of actions performed while creating or editing a model.
See the MicroLab 21 CFR part 11 section on page 47 for more information on this
feature.
Advanced Features
The Advanced Features screen provides access to Diagnostics, System Check,
User Management, and Library Management for Admin, Service and Developer
roles.
System Check includes the following instrument tests: Performance (Signal to
Noise), Stability Test, Laser Frequency Calibration Check, and Spectral Resolution
Check. Refer to the appropriate instrument operation manual for more information
on Diagnostics and System Check.
Sampling Interface
The Advanced Features screen also allows the user to access the Sampling
Interface selection for the 4100 ExoScan FTIR systems. This is important for
methods that are tied directly to the sampling interface as well as to link
performance specifications to the sampling device. The appropriate sampling
interface must be selected for either of those to work appropriately.
Method screen
Methods define instrument collection parameters and data analysis within the
MicroLab software. Methods can be defined to only collect data, collect and
compare collected data to a predefined library or collect data and perform a
quantitative analysis on that data. Predefined methods include data collection,
qualitative library searches and quantitative methods for oil and fuel analysis on the
Agilent 4500t/5500t and Cary 630 FTIR spectrometers. For questions and
additional information on specifications for predefined methods, contact Agilent.
MicroLab Mobile only allows for selection of predefined methods. Editing of
methods must be conducted on a separate computer operating MicroLab PC or
MicroLab Lite software. To be viewable within the Method dialog box, methods
developed with PC and Lite versions of the software should be transferred to the
following directory on the handheld or embedded controller: @\Program Files\
MicroLab Mobile\Methods.
In a 21 CFR part 11 environment, method transfer occurs during synchronization.
Methods list
Displays a list of methods created for the MicroLab PC application software. To
move up or down the method list, simply touch the stylus to the list and drag the list
across the screen.
Activate
The Activate button selects the highlighted method and closes the screen. The
software returns to the Home screen. Check to make sure the correct method is
active. If the incorrect method appears, return to the Methods screen and choose
an alternative method. To do this, tap the Select Methods button, highlight the
desired method and tap Activate.
Editing
Method editing is not available on the MicroLab Mobile software. In order to edit a
method, you must use the MicroLab Lite software installed on a separate
computer. Once the method is established, it can be transferred to the handheld or
embedded PC for use with MicroLab Mobile.
NOTE For instruments with PDA, MicroLab Lite software should be installed on a
computer containing Windows Mobile Device Center in order to easily transfer
methods and data. Windows Mobile Device Center can be installed from the
MicroLab software installation disk.
3 MicroLab PC
Home screen 19
Methods 20
Multivariate method transfer from GRAMS PLS/IQ to MicroLab PC 28
MicroLab quantitative model transfer from MicroLab Expert to MicroLab PC 30
Using Pre-Sample Alignment Check 31
Transferring edited methods to MicroLab Mobile 32
System Check 34
Home screen
Once logged on, you will see the MicroLab Home screen. This is the main screen for
the MicroLab PC software. From here, you can access method editing and
reviewing data.
Status
In the upper left of the screen, the Status button should be green and display
READY. If the field is yellow or red, click the Diagnostics button under the Advanced
Features option to identify the problem. When a red or yellow light condition cannot
be resolved, contact Agilent for assistance.
Logoff
The Logoff screen enables different users to log on and off the system. The Exit
button closes the MicroLab PC software.
Lock
This is available in 21 CFR part 11 environments only. Use to Lock the software to
control access while unattended. See the MicroLab 21 CFR part 11 section on page
47 for more information.
Previous Results
Once you have performed your analysis, you can access your previous results via
the Home screen. After clicking Previous Results, a screen will appear listing the
samples that have been run. The samples will be named based on the chosen
naming conventions.
Advanced Features
From the Advanced Features screen, you may perform Diagnostics, System Check,
Library Management, User Management, and the Change Password option. This
feature enables a user to perform administrator tasks.
System Check includes the following instrument tests: Performance (Signal to
Noise), Stability Test, Laser Frequency Calibration Check, and Spectral Resolution
Check. Refer to the appropriate instrument operation manual for more information
on Diagnostics and System Check.
Methods
Methods list: Displays a list of methods created for the MicroLab PC application
software.
Delete: Deletes the currently selected method; only accessible by users with Admin
rights. This is not available in 21 CFR part 11 environments.
Import: This is available in 21 CFR part 11 environments only. Import CFR and non-
CFR methods. CFR methods will retain the old audit trail although all signing will be
removed. Once imported, these methods can be used in a CFR environment.
Imported non-CFR methods are read-only.
New: Creates a new method; only accessible by users with Admin rights.
Print: This system feature enables the user to print a copy of the selected methods.
Edit: Opens the currently selected method for editing; only accessible by users with
Admin rights.
Activate: Selects the currently highlighted method as the method to be used. In
order to collect data using a particular method, it must first be activated.
NOTE Only methods selected in the Public column will be viewable or selectable by non-
Admin level users.
Editing
Info page
Method Name: This field specifies the method name; the method name will be
displayed in the method list.
Developer: This field specifies the user who developed the method.
Comment: This field is made available to the user to input any special information
concerning the method.
Type page
Method Type: Three types of methods are available in the MicroLab PC software:
Components: Quantitative prediction methods. An infrared spectrum is
collected then a predefined quantitative prediction is made from that data.
Several components can be defined, each with their own quantitative
calibration.
Data Collect Only: Only collect and display the infrared spectrum. No further
analysis is performed on the collected data.
Qualitative Search: Collect an infrared spectrum and search it against a
predefined spectral library. The results of this method will be the top matches
to the spectral library.
Check for Clean Crystal Prior to Collecting Background: This option enables or
disables the system checks of the cleanliness of the ATR crystal or sample cell
prior to collecting a background. This option is recommended but not required.
Background: Two options exist for collection of the background:
Collect new Background on every Sample: This requires the collection of a new
background spectrum for every sample. This is the recommended option, as it
provides the best correction for water vapor, carbon dioxide and other
atmospheric variations.
Background Valid time limit: If this option is selected, background is collected at
time intervals, and samples can be collected in sequence without measuring a
background spectrum in between samples. The time limit can be set in
minutes, hours and days.
Y-Axis Units: Specifies the Y-axis scale: Absorbance, Transmittance or Reflectance.
Clamp Max. Y-value: This option will display values of absorbance over the specified
value as the clamp max y-value. This is a useful option when bands that are not
used analytically may be off-scale. If this feature is selected, the off-scale bands will
appear ‘flat-topped’ and not adversely affect the displayed scale. This option is only
available when Absorbance is selected as Y-Axis Units.
Minimum Y Threshold: Set the minimum threshold on the Y-axis. This option is only
available when Absorbance is selected as Y-Axis Units.
Show Pre-Sample Alignment Check: This option enables and starts the pre sample
alignment check. Refer to page 31 for more information on Pre-Sample Alignment
Check.
Show Sampling Continuity Check: This option enables or disables the sample
continuity check. This check is typically enabled for handheld devices to make sure
the sample is kept in the focus of the sampling technology during measurement.
Instrument page
Spectral Range (cm-1): This specifies the upper and lower limits of the spectral
range in wavenumbers. Select Full to enable full detector range for data acquisition.
The range might vary according to the instrument series.
Background Scan: This specifies the number of scans to be collected in the
background measurement. Additional scans will produce higher signal-to-noise;
however, the measurement time increases proportional to the number of scans.
Sample Scans: This specifies the number of scans to be collected in the sample
measurement. As with the background measurement, additional scans increase the
signal-to-noise, but also increase the amount of time. The sample scans should be
equal or less than the number of background scans. Typically, the background and
sample are collected with the same number of scans.
Resolution (cm-1): This specifies the spectral resolution. For condensed phase
samples, most measurements can be made with 8 cm-1 resolution.
Zero Fill Factor: This specifies the zero-filling factor used for Fourier transformation
of the interferogram into an energy spectrum.
Apodization: This specifies the apodization function used for Fourier transformation
of the interferogram into an energy spectrum: None (Boxcar), Triangular or Happ-
Genzel.
Phase Correct: The standard Mertz method is applied.
Sampling Technology: Select the sampling technology: Transmission Cell,
Reflectance, ATR, or Gas Cell.
Set Method Gain: This option is used to save a gain value in the method for routine
analysis. This is typically used for specific instrument setups in combination with
specific analyses.
Calculation Type: This field specifies the type of calculation to be used. Calculation
types include univariate (Peak Height, Peak Height with Single Baseline, Peak Height
with Dual Baseline, Peak Area, Peak Area with Single Baseline, Peak Area with Dual
Baseline), noise calculation (RMS Noise), peak position (Peak Position (Center of
Gravity)) and multivariate (Quant Model). The Quant Model calculation type is also
used with MicroLab Quant univariate or multivariate calibrations.
NOTE The linear fit for a univariate calibration using peak height peak area or peak ratio
must be conducted in third party software (such as Microsoft Excel). The MicroLab
Quant software can be used to create complete univariate calibrations without the
use of third-party software.
NOTE Multivariate calibrations are conducted in the Thermo Galactic PLS/IQ Plus
software and exported to a text format readable by the MicroLab PC software, or
Eigenvector Solo Plus Model Exporter in XML format, or in MicroLab Expert (with
Quant). See instructions below for details on exporting calibrations.
NOTE The use of Agilent MicroLab Expert Quant Models (*.a2q) files requires a separate
registration code, provided as part of a workflow product bundle, such as the
“Measurement Bundle” (G8188AA). Otherwise, an error dialog box will appear.
Scaling: This section defines a linear calibration to be applied to the data. This can
be used with any of the peak definitions (Peak Height, Peak Area, and so on), or to
modify the value generated by a multivariate (Quant) method. Often, this is used to
change the value to an appropriate scale.
Scale (x value): The Scale is the slope of a linear calibration equation. In the
equation for a linear fit, y=mx+b, scale is ‘m’.
Offset: The Offset is the Y-axis offset for a linear calibration. In the equation for
a linear fit, y=mx+b, offset is ‘b’.
Decimal Digits to Report: This defines the number of decimal places that will be
reported in the answer.
Thresholds: This section defines limits that the MicroLab PC software will use to
color-code the displayed results and reports. Values that are between the Marginal
High and Critical High, or the Marginal Low and Critical Low thresholds, will be
displayed in yellow. Values above the Marginal High or below the Marginal Low
thresholds will be displayed in red. All other values will be displayed in green.
Calc Value As: Values can be calculated as either the actual value or a percentage
of either the Critical High or Critical Low values.
Test State: Enables specification of the state of the diagnostic. This is used with
‘Value’ below.
Value: The value of the diagnostic that is used with the test state in order to
determine if the component will be displayed.
Example 1: For a simple univariate method, two components may be specified
for a single analyte; one component would have a valid calibration over a low
range and the other would have a valid calibration over a high range. The
developer may prefer that only the appropriate range be reported. In this
example, the component will be called ‘area’ and the high range will be valid
from 1000 to 5000 ppm. For this example, Area will be selected for the
Component/diagnostic. First, the Test State of >= will be selected, along with a
value of 1000. After this is Added, the “And” qualifier will be selected. Then the
Test State of < will be selected with a value of 5000. In this example, the area
component will now only be reported if the value is between 1000 and 5000.
Example 2: For a multivariate (PLS) method, one may have several components
specified. Each component may have a different multivariate calibration. You
may only want the component that has a calibration set that is most closely
related to the unknown sample. In this case, the developer would select the
component/diagnostic to be MDistance (Mahalanobis Distance), the test state
to be <= and the value to be 3. This would only report values that have a
Mahalanobis Distance less than three, which indicates that the sample is
statistically within the group of the calibration set.
Reports page
Available Report Templates: Report templates are defined for specific types of
methods.
Component methods can use the Oil Report or Component Report templates.
Qualitative search methods can use the Spectrum with Parameter, Stacked
Spectra with Search Results, Stacked Spectrum with Subtraction and
Parameters Summary or the Stacked Spectrum with Zoomed Overlay
templates.
Multiple templates can be selected.
Generation of New Report Templates is covered under Section 8.
Results Folder: This specifies the location within one of the following folders where
the results will be saved.
For all systems, beginning with Version 5.3, all user definable and interactive
call files are kept in C:\Users\Public\Documents\Agilent\MicroLab
If this is left blank, the files are saved directly into the Results folder.
For 21 CFR part 11, when left blank the default results folder is
VAIMDB_Public(000)|Public Results.
A different folder can be entered for saving the files within the Results directory. If
the method is transferred to a system running MicroLab Mobile, the folder will
automatically update to the appropriate folder on the mobile device. There is also
an option for an auto-report feature. Check this box to automatically generate any
and all reports associated with the method in question when a sample is collected
using it.
Results Naming: MicroLab has an auto save feature that creates unique
identification numbers that are exclusive to each analysis. All files are available to
all authorized users and can be reviewed under the Previous Results button. An
administrator has an option of choosing a format that best fits their needs.
Sample ID + increment (Sample ID): The results name will correspond with
Sample ID and time of analysis.
Date Time: The results name will correspond with a date and time stamp of the
analysis.
Sample ID + Date Time: The results name will contain the sample ID, date, and
time of the analysis.
Results Auto-Export: MicroLab has an auto-export function that allows the user to
automatically export the results file at the time of data acquisition. MicroLab can
generate SPC and ASP file types for importing into a variety of third-party software
programs. SPC is a universal spectra data file required for data use in GRAMS or in
Eigenvector. ASP file formats are simply an ASCII output of the Y data.
NOTE The Sample ID is entered by the user during analysis. If the user fails to enter a
Sample ID, the system will automatically create a date and time stamp. The user
must always update the Sample ID when running multiple samples. MicroLab PC
will store the previously entered sample ID.
Recommend page
The recommendations will be displayed on the Results screen following sample
analysis. The messages are a unique feature; they give the user the ability to
customize results to their specific needs and parameters.
NOTE For 21 CFR part 11 environment, the model needs to be imported to the database
using MicroLab Quant before including in a method. Refer to Import Models on
page 40 for detailed information.
NOTE The use of Agilent MicroLab Expert Quant Models (*.a2q) files requires a separate
registration code, provided as part of a workflow product bundle, such as the
“Measurement Bundle” (G8188AA). Otherwise, an error dialog box will appear.
9 Choose the model file and click Open. The model is now included in the
method.
10 Enter the other options available in the Add Component dialog box. Click OK.
11 Click Save As. Enter the appropriate method name and save to
C:\Users\Public\Documents\Agilent\MicroLab\Methods.
In 21 CFR part 11, you will be prompted to save the method in a database.
12 The method is now available for use with MicroLab PC. Ensure that the method
is activated in order to use the method.
14 For use on the handheld PC, the method should be transferred to the MicroLab
Mobile/Methods folder on the mobile device.
15 In MicroLab Mobile, the Pre-Sample Alignment Check will be displayed as a
red/yellow/green bar graph on the Sample Alignment page. The sample should
be rotated or aligned in order to reach the maximum value on the bar graph.
NOTE For 21 CFR part 11 environments, non-CFR methods are visible in the mobile
environment when copying via ‘My Computer’, but cannot be activated.
NOTE This procedure will not work on the 4300 Handheld FTIR. Use the Synchronize
feature as specified on page 32.
System Check
4 MicroLab Quant
Home screen 35
Logoff 36
Lock 36
Start New Model 36
Import Models 40
21 CFR Part 11 Features 41
Home screen
When MicroLab PC v 5.1 and higher is installed, the MicroLab Quant Module will
also install. The module is designed to allow the user to create methods and
models that are to be used in the MicroLab Component methods module of
MicroLab PC.
MicroLab Quant uses the same security settings and user administration tools that
are applied in MicroLab PC. To log in, the user enters their username and password
as if they are entering MicroLab PC. For 21 CFR part 11 environments, the group
and project must also be entered.
The following options are available:
Start New Model
Previous Project/Model
Evaluate Model
Import Models (only available with 21 CFR part 11)
Data for the new project/model should be collected prior to using MicroLab Quant.
This should be done in MicroLab PC. All data should have been collected using the
same data collect parameters.
Logoff
The Logoff screen enables different users to log on and off the system. The Exit
button closes the MicroLab PC software.
Lock
This is available in 21 CFR part 11 environments only. Use to Lock the software to
control access while unattended. See the MicroLab 21 CFR part 11 section on page
47 for more information.
1 Right-click in the model plot window and select the desired plot.
2 Click the Select Model Plot button on the bottom of the analysis screen.
h Model Evaluation: By clicking on the Model Evaluation button, the software
presents the user with two functions. These functions are typically used to
determine outliers and robustness of the calibration set.
1 Cross Validation: The software runs through the calibration eliminating
each standard and then determining what value would be returned if
running that standard as an unknown.
2 Independent Set: Allows the user to [Add Files...] of check standards
that have been run independently of the calibration and predict the
values on these samples. The concentrations of each sample are
entered into the table and then the [Predict] button is clicked.
a Total Standard Error, R-Squared for Actual V Predicted are shown
at the bottom of the screen. The Actual V Predicted graph is also
displayed on the screen.
b The user can Export the values to a CSV file format of in the
location of their choice.
c The user now has the option of saving [Save] the model and
returning to the previous data analysis screen [Back].
i Finalize Model: If the user is satisfied with the calibration model, the
[Finalize Model] button is chosen. This then allows the user to save the
Project (*.mpq) which includes the standard and model. The user can also
choose at this time to save the data to a MicroLab method. This will
include the component that the user just defined in the associated model
with all of the calibration parameters.
1 Model Files: If multiple components need to be defined, then this will be
done in individual models/projects. Each component is defined in an
individual model or *.mqm file at the same time the project file is
saved.
2 Save Method: Saves the files to an *.a2m file extension with the
component of interest. The method will have the same data collect
parameters as the calibration files which were used to create the
model. If multiple components are necessary in a methods file, then
they must be added to the method manually through MicroLab Lite or
PC. See Section on Multivariate method transfer under MicroLab PC.
The models are added in the same way as the GRAMS models,
however, they are defined as the *mqm files as noted above.
Import Models
NOTE The use of Agilent MicroLab Expert Quant Models (*.a2q) files requires a separate
registration code, provided as part of a workflow product bundle, such as the
“Measurement Bundle” (G8188AA). Otherwise, an error dialog box will appear.
See the MicroLab 21 CFR part 11 section on page 47 for more information on these
features.
21 CFR part 11
Displays a list of actions performed while creating or editing a model.
E-Sign
Click E-sign to apply an electronic signature and add comments to the model.
5 MicroLab OQ
Operational Qualification (OQ) 43
Perform the OQ 44
OQ Report 45
Diagnostic
The Diagnostic screen displays the following information about the instrument:
Instrument Status: Energy, Battery, Source, and Laser
Temperatures of: Detector, CPU, IR board, Power and Block
Version Information
NOTE Diagnostic can also be viewed from both MicroLab PC and MicroLab Mobile
through the Advanced Features screen.
Performance Validation
The following tests are available:
Signal to Noise (Performance) Test
Stability Test
Laser Frequency Calibration Check (Frequency accuracy and repeatability)
Spectral Resolution Test
NOTE These are the same tests used for System Check in both MicroLab PC and
MicroLab Mobile through the Advanced Features screen.
Perform the OQ
OQ Report
SCM
The use of user identification codes and passwords enables control over who can
log on to the system and who can perform particular functions within the Agilent
application software. It also provides the mechanism to allow electronic signing of
electronic records. The use of databases coupled with SCM, prevents all
unauthorized users from changing or deleting files. The SCM event logs augment
the audit trails resident in the application software. The SCM administrator must
set up the required users. It is important that a number of simple requirements are
followed when this is done to ensure that compliance with the 21 CFR Part 11 rule
is maintained.
For additional information about configuring SCM see the SCM Help. To access the
SCM Help, with the SCM application open, press F1 on your keyboard or navigate to
C:>Program Files (x86) > Varian > 21 CFR 11 > Configuration manager > Server >
Configuration Manager Database Help.
SDA
The Spectroscopy Database Administrator (SDA) has been designed for system
administrators to set up and maintain the databases that are used by the Agilent
application software to store data.
You can use SDA together with the Agilent Spectroscopy Configuration Manager
(SCM), the application software, and your own Standard Operating Practices (SOPs)
to form a 21 CFR Part 11 capable environment for controlling your Agilent
instruments.
SDA allows the data collected by Agilent instruments to be stored locally (on the
same PC as the Agilent application software), or remotely in a Client/Server
arrangement.
Configuration must be performed by the system administrator, or a person with
administration rights to run SDA.
For additional information about configuring SDA see the SDA Help. To access the
SDA Help, with the SDA application open, press F1 on your keyboard or click Start >
All Programs > Agilent > Database Utilities > SDA Help.
With the 21 CFR part 11 software installed, additional buttons will be available.
Lock
This dialog allows the user to lock the MicroLab software when using the 21 CFR
Part 11 software package. It is accessed from the Home screen.
When locked, the Unlock MicroLab PC dialog box appears and no actions can be
taken until a validated user logs on and unlocks the software. Users other than the
one that ‘Locked’ the MicroLab PC/Lite/Quant/OQ originally can only unlock the
software when they are on the Home page. Only the user who is originally logged
on to the software can unlock the software from all pages. This ensures privileges
are applied correctly, audit trail information is correct in MicroLab and Configuration
Manager and results file are saved correctly.
User Identification — Type your user identification.
Group — Select your group from the drop-down list.
Project — Select your project from the drop-down list.
Password — Enter your password.
NOTE When MicroLab Mobile 21 CFR part 11 is used, passwords must be alpha numeric
only. Do not use special characters such as $, @, and so on.
OK — Click OK to log on and exit the dialog box. The software is now unlocked.
Cancel — Click Cancel to exit the dialog box without logging on. The software stays
locked.
E-sign
Click E-sign to apply an electronic signature. Any User can “Sign” the method or
results. To perform Approval – 1 and Approval – 2 the User must be given
appropriate permission in Spectroscopy Configuration Manager. You do not need
to login to the MicroLab software to approve, however when the Signature dialog
appears the User will need to enter their credentials to perform the Approval – 1 or
Approval – 2. The original User who was logged in before the Approval is performed
will remain logged in. If the account of the user logged in to the MicroLab software
is deactivated or locked, when cancelling out of the e-Signature dialog the Lock
dialog will appear. The original user account must be unlocked by the Configuration
Manager Administrator before the MicroLab software can be unlocked.
User Identification — This field displays the name of the current user. The
operator/approver must enter their user identification in this field prior to applying
an electronic signature.
Group — The group will automatically be populated once the user identification field
is entered. If the User is linked to more than one group, choose the appropriate
group.
Project — The project will automatically be populated once the user identification
field is entered. If the User is linked to more than one project, choose the
appropriate project.
Password — The operator/approver must enter their password in this field if they
are to apply an electronic signature.
Comment — In this field, type information about the action that is being signed off,
or the meaning of the signature. It is compulsory to enter a comment — if it is left
blank, when the 'Sign', 'Approval-1' or 'Approval-2' button is clicked, a message is
displayed stating that this is a required field.
Approval-1 — Click to apply an electronic signature at the Approval-1 level. A user
who ran the method cannot also approve the data, a message stating this is
displayed.
Approval-2 — Click to apply an electronic signature at the Approval-2 level. A user
who ran the application and created the data cannot also approve the data, a
message stating this is displayed.
The method or results will become locked once Approval – 2 has been performed.
This method or results can be opened and viewed, but no further modifications can
be performed. A new method can be created from this locked method.
MicroLab method and result files cannot be locked unless Approval-2 is performed.
A company SOP may be required to ensure no further modifications are made to
the file once the document has been signed by the Operator and Approver-1. Any
modification between performing the e-signature are detailed in the Audit Trail
including the e-signature events.
The bottom part of the dialog box shows the log of the method including any
changes that have been made.
21 CFR part 11
Click to access the Audit Trails, Collect Parameters and Analysis Parameters tabs.
Audit trail
The method and results audit trails are displayed, showing all recorded actions
including:
Method and results save location
Method used for re-analysis
Method parameter changes
Reason for change
Method run started
Method run stopped
Unexpected instrument failure, method run stopped
Method or results authorized
Residual Applied
For each entry above, a combination of the following information is also logged:
SDAdministrator server/database/file name
Local Date and Time
GMT Date and Time
Workstation name
User: Group/Project/User ID
Username
User Description
Instrument Serial Number
Validated comment
Software version number
Firmware version number
7 Measuring Samples
Cleaning the sample interface 53
Collecting a background measurement 54
Collecting the sample measurement 54
The MicroLab PC and MicroLab Mobile software are designed to walk the user
through sample measurement. The software display images and instructions of the
actions that are required by the user for sample measurement. Following the
instructions on the screen ensures consistent measurement results. For any
sample measurement, the following steps are taken:
Clean the sample interface.
Collect a background.
Place the sample to be measured.
Collect the sample measurement.
The first step in making a sample measurement is to ensure that the sample
mounting interface is clean of any residue from previous samples or from general
use and storage. For specific instructions for cleaning, refer to the appropriate
instrument operation manual supplied with the system.
CAUTION Do not break the spectrometer seal and attempt to clean interior surfaces.
Breaking the seal will void the warranty.
NOTE As mentioned above, before collecting a background, ensure that the sample
interface is clean. To clean the sample interface, a suitable solvent such as
methanol, acetone or isopropyl alcohol should be used. See the instrument
operation manual for more detailed information.
NOTE Only click HOME to go back to the main home screen if you wish to start your
analysis over.
To measure a sample:
1 To begin sample measurement, click the Start button on the Home screen.
2 As the software instructs, clean the sample interface for an ATR or
Transmission measurement.
For a reflectance measurement on the Agilent 4100 ExoScan FTIR, the
reflection standard should be placed on the sample interface.
NOTE The sample surface should be cleaned with a soft tissue and an appropriate solvent
such as methanol, ethanol or acetone.
3 Click the Next button. Clicking the Home button will stop the measurement and
return you to the Home screen.
The Clean Crystal Check will be conducted if it has been selected in the
method. This will check for absorbance over a stored reference. If any
absorbances are found, the software will instruct you to clean the sample
interface and start again.
NOTE If the clean crystal check continually fails, the stored background scan may be
corrupted. If this occurs, the clean background can be reset by clicking Advanced
Features > Diagnostics then Reset Clean Check in the Diagnostics dialog box. This
will clear the stored background.
4 Following the crystal check (if specified), the software will collect the
background sample. Background samples are collected either at every scan or
after a specified time interval, as specified in the method.
The software will instruct you to apply the sample to the sample interface. The
picture displayed will match the configuration of the system. Once the sample
is in place, click Next.
5 The software will display a graph showing the intensity (absorbance) of the
sample measurement. This screen is used to ensure that the sample has been
correctly placed at the sampling interface.
For oil applications using Agilent 4500t FTIR, the value should be
maximized.
For Agilent 4100 ExoScan FTIR applications using reflectance, the value
should be between 0.1 and 2 AU; ATR measurements should be between
0.05 and 1 AU.
6 Enter a file name in the Sample ID field; this will be the file name under which
the sample data is saved. Enter further information about the sample in the
Comment field.
7 Click the Next button to begin sample measurement. The software will co-add
sample scans as specified in the method.
8 During sample measurement, the sample needs to remain at the focus of the
sampling technology.
For the Agilent 4500t FTIR, the sample cell should not be adjusted during
the measurement.
For the Agilent 4100 ExoScan FTIR, the instrument should not be moved
from the sample surface during the measurement.
The software monitors the sample signal during measurement. If the signal
falls below a preset threshold, the collection will stop. Once the sample focus is
regained, the software will immediately start collecting data.
NOTE The progress bar shows the advancement of the data collection. When the progress
bar is complete, the prompt will change to Transferring Data. Once this occurs, the
measurement is complete, and the instrument can be removed from the sample.
9 When the sample measurement is complete, the software will display the
method results in the Results screen. The sample spectrum and the calculated
method results will automatically be saved. The data can be returned to at any
time from the Previous Results screen.
8 Reviewing Results
MicroLab Mobile 57
MicroLab PC/MicroLab Lite 60
MicroLab Report Creation 65
MicroLab Mobile
Results screen
For component methods, the Results screen shows the calculated component
values. Component values are ordered relative to their critical limits; those values
outside of the critical range are listed first followed by values outside of the
marginal range. The components within the acceptable range are shown last in
green.
Green results show that the sample is within adjustable limits.
Yellow results show that the sample is between marginal and critical limits.
Red results show that the sample is above or below critical limits.
Home: This button returns the software to the Home screen.
Next: This button returns the software to the Sample Collection page for
measurement of additional samples.
Handling data
The Menu button enables you to choose Data Handling. This button allows the user
to input a Sample ID and comments if not previously entered. The system’s auto-
save feature will save the Comment field and generate a new file based on the
sample ID if one is entered.
NOTE The auto save feature will always save the current file with a sample ID and a time
date stamp. This prevents the user from saving over a previous sample result.
NOTE The Re-analyze feature should be used only with data collected at the same
spectral resolution as the method chosen. Differences in spectral resolution can
affect quantitative results.
Export: Enables you to export collected data to other analytical software packages.
The exported file will be saved with the same file name as the sample, except the
extension will be *.csv, *.asp or *.spc. It will be saved into the Results directory.
Peak Picking: Enables the user to define peaks to be marked on the spectrum being
reviewed. This is done by right clicking on the spectrum and choosing Peak Pick
from the menu. To mark peaks on the spectrum, draw a threshold line by clicking,
dragging, and clicking again. This can be repeated multiple times to define several
peaks on the spectrum. The peaks will print on any report generated and will keep
with the data file. Once done, right click again and de-select Peak Pick from the
menu.
Report templates incorporate a peak picking table. For the selected peaks, the peak
values appear in the table in the order they were selected.
Previous results
MicroLab Mobile’s Previous Results enables reviewing of data after collection has
occurred. Previous results are available under the Menu icon on the Home screen. A
list of previously collected samples is displayed on the Previous Results screen. In
order to move through the list, click the list and drag to scroll.
Home: Returns the software to the Home screen.
View: Opens the data file for viewing. The data will be presented in an identical way
to the Results screen shown in the Reviewing Results section on page 57.
Report: From the Menu button, the Report button will generate a PDF report as
specified in the Reports page section on page 26.
MicroLab PC/MicroLab Lite enable further data analysis than is available within
MicroLab Mobile. Due to the screen size, spectra are not currently displayed in the
MicroLab Mobile software. MicroLab PC/MicroLab Lite enable viewing of the
spectra and review of the spectral regions that were used for the component
calculations. Once the data has been transferred to the computer (see Transferring
edited methods to MicroLab Mobile on page 32; transferring results is analogous to
transferring methods), it can be reviewed using the Previous Results feature of
MicroLab PC or MicroLab Lite. From the MicroLab PC Home screen, click the
Previous Results button; a screen will appear listing the all sample file names.
Home: Returns the software to the Home screen.
Import: Enables you to make additions to the Results screen.
Report: Generates a report for the sample if report template(s) is assigned in the
Reports page of the method. Refer to page 65 to assign a report template to a
method.
View: Returns the software to the Sample Results screen. The Sample Results are
always reviewable with the MicroLab software.
Overlay: The Overlay button appears when more than one spectral file is selected. It
allows the user to view up to 5 sample spectra in overlaid or stacked format for
comparison. The files must have the same abscissa type in order to be viewed
together.
Export: Enables you to export collected data to other analytical software packages.
The exported file will be saved with the same file name as the sample, except the
extension will be *.asp or *.spc. It will be saved into the Results directory.
Diag Info: Displays diagnostic information for the spectrometer connected to the
MicroLab software.
NOTE The software from the previous results section will operate as the results section
operates. The screens are identical. These features enable operators to review data.
Results screen
Results box
The results are color-coded for clarity and out-of-spec results are highlighted red
and are moved to the top of the list.
Green results show that sample is within adjustable limits.
Yellow results or close to acceptable limits.
Red results show that sample are outside acceptable limits.
The Results table has two headers: Parameters and Current Level.
NOTE To change between categories, right-click and scroll down to either option. The
screen will reflect the view you indicate.
Details
Click Details to view the spectrum of the sample. The MicroLab software will
display the sample spectra against the reference spectra.
NOTE You can zoom in on any portion of the spectra. Using the mouse or touch pad, click
on the graph and highlight the desired portion. The screen will automatically change
to the zoomed in portion.
To return to the full spectrum, right-click on one graph and choose Undo Zoom.
Params
Click Params to view the following tabs:
Collect Parameters: Shows a summary of method parameters.
Analysis Parameters: Shows a summary of analysis on collected data.
Data handling
Click Data Handling to access the following features.
Report: Generates a report for the sample if report template(s) is assigned in the
Reports page of the method. The report outlines the preset parameters and shows
the results in relevant units compared to the reference oil or ASTM (American
Society for Testing and Materials) method where appropriate. Refer to page 65 to
assign a report template to a method.
Add To a Library: Enables you to add the sample spectra to an appropriate library.
Click ‘Browse’ to access the library. A list will appear that shows the libraries
available to you.
Add To Library will complete the process. Adding new spectra to the user’s library
enables the software to determine if the subsequent samples are similar.
Export: Enables you to export collected data to other analytical software packages.
Re-Analyze: Enables you to perform the calculations of a different method on the
currently viewed data. This is useful when it is not possible to recollect the sample
data.
NOTE The Re-analyze feature should be used only with data collected at the same
spectral resolution as the method chosen. Differences in spectral resolution can
affect quantitative results.
E-Sign
This is available in 21 CFR part 11 environments only. Click E-sign to apply an
electronic signature. See page 49 for more information.
21 CFR part 11
Displays a list of actions performed while running or editing a method or results.
See page 51 for more information.
NOTE This option does not allow you to change unit information. If you are running
samples from different machines, you must return to the Home screen.
NOTE If you would like to stop the analysis, click the HOME button.
NOTE This feature is only supported on 32-bit installations of Microsoft Word 2010 or
greater.
The use of this function requires knowledge of Microsoft Word that may not be
covered in this manual.
To create a new report template:
1 Open Microsoft Word and create a new document.
2 Once this document is created, navigate to the View menu/tab.
3 A new feature has been added to the View Menu – MicroLab Reporting. This
includes two buttons: Report Designer and Perform Merge.
4 To begin creating the document. Select the Report Designer function. The
schema associated with MicroLab will appear as a menu on the right of the
screen.
5 Decide the type of report that is needed and select it under Report Type. Types
of reports include the following:
a Component Report: This will only show up if the Method has been
designated as a component method in the method editor.
b Qualitative Search: This report template will only show up if the method has
been designated as a qualitative search method in the method editor.
c Data Collect Only: This report template will only show up if the method has
been designated as a Data Collect method in the method editor.
d Method Report: System level report for printing Method parameters.
e Log Report: System level report for printing the system log file.
6 Give the Report Template a name in the MicroLab Reporting Parameters block.
7 The Report Fields consist of parameter fields that are associated with the
MicroLab Software. Fields can be located in a variety of categories for the user,
from Results to Method Parameters. The user should take time to familiarize
themselves with the different categories. It is important to note that each type
of method has its own Method category and Result category. For example, a
Qualitative Search method has two categories that are related to it – Hit List
and Method Library. Information specific to this type of method and its
resultant files will be found here. Whereas more general information would be
found simply in Results or Method categories.
8 To add fields to the report, simply click and drag the field desired to the report
template page.
a The field added will only print out the information requested. Since this is in
Word, the user may add text at any point in time to identify fields or label
material.
b It is not recommended to use tables other than the predefined tables as
this will cause errors in the Report.
9 Predefined Tables for Component Reports and Search Hit Lists have been
configured for the user to simply click and drag into the report template.
10 At any time, to preview what the report would look like, select the Preview Rep
button located at the bottom of the MicroLab Reporting menu.
a This can also be accomplished through the Perform Merge function.
b It is always best to save the report and run true data against the method
where the reports are assigned to determine if the report is showing all
values necessary.
11 Once the Report has been edited appropriately, the file should be saved to the
report template folder for MicroLab. This folder is typically located in
C:\Users\Public\Documents\Agilent\MicroLab\rptTemplates.
a Save the file as docx file format.
12 There are several report templates that are supplied when the software is
installed. They may be edited. However, it is highly recommended that the user
save edited report templates as a different file name using the “Save As”
function of word, so that the original templates are left intact.
Firmware update
Occasionally Agilent will release updates to the firmware (software contained in the
FTIR spectrometer). These updates will generally correct bugs that have been
noticed by customers, and may expand upon the features that the system can
provide. Firmware updates will be provided as Microsoft Windows executables
files. These files may be provided directly to the customer, or may be made
available in conjunction with an update to the Agilent MicroLab PC application.
For Agilent 4300 Handheld FTIR firmware updates, please refer to the appropriate
instrument operation manual supplied with the system, or contact Agilent for more
information.
Troubleshooting
Initialization failure
If you see this error message:
Instrument failed to initialize
Instrument returned error code: -2.
The application will exit; then power-cycle the instrument and restart the
application.
Upon loading the software, the spectrometer is not connected to the computer. If
using the handheld PC with MicroLab Mobile, verify that the instrument displays a
green light on the power button, and verify that the ‘Bluetooth’ communications are
turned ‘On’ on the handheld PC. The instrument must be connected for either the
MicroLab Mobile or MicroLab PC software. The instrument is not required for
MicroLab Lite.
Agilent Library Files *.a2l All three files are generated and stored together when a library is created in
*.a2x MicroLab PC/Lite. When exporting a library, all three files must be together.
*.lib • *.a2l file contains the actual library contents;
• *.a2x file is the associated blank results holder;
• *.lib file contains the library title for identification.
Agilent Method File *.a2m Method file is used to store one of the following method types:
• Data collect only, or
• Qualitative (library) search, or
• Component type methods.
Agilent Result File *.a2r This is a spectral file format, and generates one of the three possible result file sub-
types according to the method type selected:
• Spectral results
• Library results with normalized spectra
• Component results and spectra
Agilent Security Setting *.a2s This file stores user login settings and password. It is an internal file and should not
File be moved or deleted.
Spectral File *.asp MicroLab PC/Lite can view and import this spectral file type and/or auto export this
file type.
Thermo Galactic / *.spc MicroLab PC/Lite can view and import this spectral file type and/or auto export this
GRAMS file type.
GRAMS PLS1 Text *.txt The text format of GRAMS model’s instructions is implemented into a component
Model method. This file is generated by GRAMS IQ software after appropriate model
building.
Eigen Vector Models *.xml MicroLab PC/Lite uses XML file format to generate a quantitative result within a
component method. Only applicable to PLS1 multivariate model type.
Agilent MicroLab *.mqm Generated by MicroLab Quant, and can be used by MicroLab Quant/PC/Lite in
Quant Model component method to generate a numerical result.
MicroLab Quant *.mqp It is used to save MicroLab Quant workflow and is only applicable to MicroLab
Project Quant.
Agilent MicroLab *.a2q This file is used for MicroLab PC, and is only generated with MicroLab Expert
Expert Quant Model software package v1.1.
File
Introduction
MicroLab Mobile
MicroLab PC
MicroLab Quant
MicroLab OQ
MicroLab with 21 CFR part 11
Measuring Samples
Reviewing Results
Updates and Troubleshooting
www.agilent.com
© Agilent Technologies, Inc. 2019
Edition 8, 11/2019
*0020-410*
0020-410