Kinematic Viscosity of Asphalts (Bitumens) : Standard Method of Test For
Kinematic Viscosity of Asphalts (Bitumens) : Standard Method of Test For
AASHTO T 201-03 is identical to ASTM D 2170-01a except for the following provisions:
1. All references to ASTM Standard D 92 contained in ASTM D 2170-01a shall be replaced
with AASHTO Standard T 48. Add reference to R 16, Regulatory Information for Chemicals
Used in AASHTO Tests.
2. Delete the 1 in front of cm2/s in the fourth sentence of Section 3.1.2.
3. Replace Section 8.2 of ASTM D 2170-01a with the following:
8.2 Maintain the bath at the test temperature within ±0.03ºC (±0.05ºF). Apply the necessary
corrections, if any, to all thermometer readings.
4. Add a last sentence to Section A2.3.3.1 of ASTM D 2170-01a as follows:
“When the sample travels through R and fills bulb A approximately half full, arrest its flow by
placing a stopper in tube L.’’
5. Add a last sentence to Section A2.3.3.3 of ASTM D 2170-01a as follows:
“This 15-minute period is part of and not in addition to the equilibration time of A2.3.4.”
6. Revise the second sentence of A2.3.5 to read:
“For the Zeitfuchs cross-arm viscometer apply slight vacuum to tube M (or pressure to tube
N) to cause…’’
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.
ANNEXES
(Mandatory Information)
A1.1 The viscosity of a Newtonian liquid is calculated from A1.3 When the specific gravity of the asphalt is not known,
its kinematic viscosity by multiplying the kinematic viscosity a satisfactory determination of the viscosity at 135°C (275°F)
by the density of the liquid at the test temperature. can be obtained by multiplying the kinematic viscosity at
135°C (275°F) by an assumed density of 0.948 g/cm3. This
A1.2 For paving grade asphalts, density at 135°C (275°F)
density value is equivalent to a specific gravity of 1.015 at
can be conveniently determined from specific gravity at
25/25°C (77/77°F). Tests on a large number of asphalts indicate
25/25°C (77/77°F) by multiplying by a factor of 0.934 g/cm3;
that the error introduced by this assumed density will not
or from the specific gravity at 15.5/15.5°C (60/60°F) by
multiplying by a factor of 0.931 g/cm3. These factors are based exceed 63 %.
on an average coefficient of expansion for asphalt cement of
0.00061/°C (0.00034/°F).
A2.1 Scope A2.3.2 Select a clean, dry viscometer that will give a flow
A2.1.1 The reverse-flow viscometers for transparent and time greater than 60 s.
opaque liquids include the Cannon-Fenske Opaque,7 Zeitfuchs A2.3.3 Charge the viscometer in the manner dictated by the
Cross-Arm,8 BS/IP/RF, and Lantz-Zeitfuchs viscometers. Un- design of the instrument, this operation being in conformity
like the modified Ostwald and suspended-level viscometers, with that employed when the instrument was calibrated. If the
the sample of liquid flows into a timing bulb not previously wet sample may contain lint, solid particles, etc., filter as prescribed
by sample, thus allowing the timing of liquids whose thin films in Test Method D 445.
are opaque. Reverse-flow viscometers are used for the mea- A2.3.3.1 To charge the Cannon-Fenske Opaque viscometer,
surement of kinematic viscosities of opaque and transparent invert the viscometer and apply suction to the tube L, immers-
liquids up to 100 000 mm2/s (cSt). ing tube N in the liquid sample. Draw liquid through tube N,
filling bulb D to fill mark G. Wipe excess sample off tube N and
A2.2 Apparatus invert the viscometer to its normal position. Mount the vis-
A2.2.1 For the reverse-flow viscometers, detailed drawings cometer in the constant-temperature bath, keeping tube L
of the size designations, approximate constants, kinematic vertical.
viscosity range, capillary diameter, and bulb volumes for each A2.3.3.2 Mount the Zeitfuchs Cross-Arm viscometer in the
viscometer are shown in Figs. A2.1-A2.4. The kinematic constant-temperature bath, keeping tube N vertical. Introduce
viscosity range is based on a 60-s minimum flow-time in this sample through tube N taking care not to wet the sides of tube
method; the same viscometers are also described in Specifica- N, into the cross-arm D until the leading edge stands within 0.5
tions D 446 with a different viscosity range based on a 200-s mm of fill mark G on the siphon tube.
minimum flow-time as prescribed in Test Method D 445. The NOTE A2.1—The volume of the test specimen is dependent on the
lengths of the upper tubes K, L, M and N in Figs. A2.1-A2.4 location of the fill mark G. When the flow time of the lower meniscus is
may be increased (or decreased) to conform to the immersion being measured between timing marks E and F, the upper meniscus shall
requirements of 6.4. be in the horizontal cross-arm D, thus making the location of fill mark G
critical.
A2.3 Operating Instructions A2.3.3.3 Mount the Lantz-Zeitfuchs viscometer in the
A2.3.1 A standard operating procedure is contained in constant-temperature bath keeping tube N vertical. Introduce
Section 7. Additional operating instructions for the reverse- sufficient sample through tube N to completely fill bulb D,
flow viscometers are outlined in A2.3.2-A2.3.8 with emphasis overflowing slightly into overflow bulb K. If the sample is
on procedures that are specific to a particular instrument or this poured at a temperature above test temperature, wait 15 min for
group of instruments. the sample in the viscometer to attain bath temperature and add
more sample to overflow slightly into bulb K.
A2.3.3.4 Mount the BS/IP/RF viscometer in the constant-
7
Cannon, M. R., and Fenske, M. R., “Viscosity Measurement—Opaque Liq- temperature bath keeping tube L vertical. Pour sample through
uids,” Industrial and Engineering Chemistry, Analytical Edition, Vol 13, 1941, p.
2910.
tube N to a point just above filling mark G; allow the sample
8
Zeitfuchs, E. H., “Kinematic Viscometer for Opaque and Very Viscous to flow freely through capillary R, taking care that the liquid
Liquids,” Oil and Gas Journal, Vol 44, No. 36, 1946, p. 98. column remains unbroken, until the lower meniscus is about 5
FIG. A2.1 Cannon-Fenske Opaque Viscometer for Opaque and Transparent Liquids
mm below the filling mark H and then arrest its flow by closing A2.3.5 For the Cannon-Fenske Opaque and BS/IP/RF vis-
the timing tube with a cork or rubber stopper in tube L. Add cometers, remove the stopper in tubes N and L, respectively,
more liquid if necessary to bring the upper meniscus slightly and allow the sample to flow by gravity. For the Zeitfuchs
above mark G. After allowing the sample to attain bath Cross-Arm viscometer, apply slight vacuum to tube N (or
temperature (see A2.3.4) and any air bubbles to rise to the pressure to tube L) to cause the meniscus to move over the
surface, gently loosen the stopper allowing the sample to flow siphon tube and about 30 mm below the level of tube D in
to the lower filling mark H and again arrest flow. Remove the capillary R; gravity flow is thus initiated. For the Lantz-
excess sample above filling mark G by inserting the special Zeitfuchs viscometer, apply slight vacuum to tube M (or
pipet until its cork rests on top of tube N; apply gentle suction pressure to tube N with tube K closed) until the lower meniscus
until air is drawn through. The upper meniscus shall coincide is opposite the lower timing mark E; allow the sample to flow
with mark G. by gravity.
A2.3.4 Allow the viscometer to remain in the constant-
temperature bath a sufficient time to ensure that the sample A2.3.6 Measure to the nearest 0.1 s the time required for the
reaches temperature equilibrium (10 min minimum, 30 min leading edge of the meniscus to pass from timing mark E to
maximum). timing mark F as prescribed in Section 8 of this method.
5
TS-2b T 201-6 AASHTO
D 2170
FIG. A2.2 Zeitfuchs Cross-Arm Viscometer for Transparent and Opaque Liquids
FIG. A2.3 Lantz-Zeitfuchs Type Reverse Flow Viscometer for Opaque Liquids
FIG. A2.4 BS/IP/RF U-Tube Reverse Flow Viscometer for Opaque Liquids
A3.1 Scope cosity oil standards9 having the approximate kinematic viscosi-
ties shown in Table A3.1. Certified kinematic viscosity values
A3.1.1 This annex describes the materials and procedures
established are supplied with each sample.
used for calibrating or checking the calibration of viscometers
used in this method.
9
The viscosity oil standards are available in 1-pt containers. Purchase orders
A3.2 Reference Materials should be addressed to the Cannon Instrument Co., P. O. Box 16, State College, Pa.
16801. Additionally the Cannon Instrument Co. has available viscosity standards
A3.2.1 Viscosity Oil Standards conforming to ASTM vis- N600, N2000, and N8000, which have viscosities at 60°C (140°F) of 180 cSt, 500
cSt, and 1700 cSt, respectively; standards N600 and N2000 have viscosities at
135°C (275°F) of 12 and 25 cSt respectively. Shipment will be made as specified or
by best means.
(Nonmandatory Information)
X1.1 To achieve an accuracy of 60.02°C (60.04°F) for least 1 min apart should agree within one tenth of a division.
calibrated kinematic viscosity thermometers, it is required that X1.2.4 Record the ice point reading and compare it with the
a check at the ice point be made and the corrections altered for previous reading. If the reading is found to be higher or lower
the change seen in the ice point. It is recommended that the than the reading corresponding to a previous calibration,
interval of checking be every six months; for a new thermom- readings at all other temperatures will be correspondingly
eter, check monthly for the first six months. increased or decreased.
X1.2 A detailed procedure for the measurement of the ice X1.2.5 The ice point procedure given in X1.1-X1.2.4 is
point and recalibration of thermometers is described in 6.5 of used for the recalibration of kinematic viscosity thermometers,
Test Method E 77. The suggestions in the following sections of and a complete new calibration of the thermometer is not
this appendix are given specifically for the mercury-in-glass necessary in order to meet the accuracy ascribed to this design
kinematic viscosity thermometers and may not apply to other thermometer.
thermometers.
X1.2.1 The ice point reading of kinematic viscosity ther- X1.3 It is recommended that these kinematic viscosity
mometers shall be taken within 60 min after being at the test thermometers be stored vertically when not in use so as to
temperature for not less than 3 min. The ice point reading shall avoid the separation of the mercury column.
be expressed to the nearest 0.01°C or 0.02°F.
X1.2.2 Select clear pieces of ice, preferably made from X1.4 It is recommended that these kinematic viscosity
distilled or pure water. Discard any cloudy or unsound por- thermometers be read to the nearest 1⁄5 of a division using
tions. Rinse the ice with distilled water and shave or crush into appropriate magnification. Since these thermometers are typi-
small pieces, avoiding direct contact with the hands or any cally in a kinematic viscosity bath (which has vision through
chemically unclean objects. Fill the Dewar vessel with the the front), the thermometer is read by lowering the thermom-
crushed ice and add sufficient distilled and preferably pre- eter such that the top of the mercury column is 5 to 15 mm
cooled water to form a slush, but not enough to float the ice. As below the surface of the bath liquid. Be careful to ensure that
the ice melts, drain off some of the water and add more crushed the expansion chamber at the top of the thermometer is above
ice. Insert the thermometer packing the ice gently about the the lid of the constant temperature bath. If the expansion
stem, to a depth approximately one scale division below the chamber is at elevated or lowered temperatures from ambient
0°C (32°F) graduation. It may be necessary to repack the ice temperatures, a significant error can occur. This error can be as
around the thermometer because of melting. much as one or two thermometer divisions. A reading glass
X1.2.3 After at least 3 min have elapsed, tap the stem such as used for reading books may be useful to ensure reading
gently, and observe the reading. Successive readings taken at the scale to 1⁄5 of a division.
ASTM International takes no position respecting the validity of any patent rights asserted in connection with any item mentioned
in this standard. Users of this standard are expressly advised that determination of the validity of any such patent rights, and the risk
of infringement of such rights, are entirely their own responsibility.
This standard is subject to revision at any time by the responsible technical committee and must be reviewed every five years and
if not revised, either reapproved or withdrawn. Your comments are invited either for revision of this standard or for additional standards
and should be addressed to ASTM International Headquarters. Your comments will receive careful consideration at a meeting of the
responsible technical committee, which you may attend. If you feel that your comments have not received a fair hearing you should
make your views known to the ASTM Committee on Standards, at the address shown below.
This standard is copyrighted by ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959,
United States. Individual reprints (single or multiple copies) of this standard may be obtained by contacting ASTM at the above
address or at 610-832-9585 (phone), 610-832-9555 (fax), or [email protected] (e-mail); or through the ASTM website
(www.astm.org).