UV Spectroscopy-Notes
UV Spectroscopy-Notes
UV/Visible Spectroscopy
Swapnil J. Dengale
Introduction:
Spectroscopy per se involves the measurement and interpretation of EMR (Electro Magnetic
Radiations) absorbed or emitted when the molecules or atoms or ions of sample moves from one
energy state to another energy state. On passing EMR in the UV and visible region through compound
with multiple bonds, a portion of radiation is normally absorbed by the compound. The amount of
absorption depends on the wavelength of the radiation and the structure of the compound. The
absorption of radiation is due to the subtraction of energy from the radiation beam when electrons in
orbitals of lower energy are excited into orbitals of higher energy. Since this is an electron excitation
phenomenon, UV spectroscopy is sometimes called as electronic spectroscopy.
Definition of spectroscopy:
The interaction of EM radiation and matter are the subject of the science called “spectroscopy”.
Spectroscopic analytical methods are based on measuring the amount of radiation produced or
absorbed by molecular or atomic species of interest.
By looking at the definition of spectroscopy, it is imperative to understand the EMR spectrum and its
properties.
EMR Spectrum:
• EMR or light, is a form of energy. Whose behavior is described by the properties of both waves
and particles
• The optical properties of EMR such as diffraction are explained best by describing light as a
wave.
• Many of the interactions between EMR and matter such as absorption and emission, however,
are better described by treating light as particle or photon
• EMR consists of oscillating electric and magnetic fields that propagate through space along a
linear path with constant velocity
1
Dept. of Pharmaceutical Quality Assurance, MCOPS Swapnil Dengale
• Moves through medium other than a vacuum with a velocity V, which is less than the speed
of light in vacuum.
In order to use light wave for any meaningful purposes, the qualification (characterization) of waves
have to be understood first.
Fig. 2
Wavelength (λ):
The distance between consecutive crests (or troughs) is called as wavelength. It can be measured in
cm (10-2 m) , mm, µm (10-6m), nm (mµ or 10-9mt), A° (10-10).
2
Dept. of Pharmaceutical Quality Assurance, MCOPS Swapnil Dengale
Frequency (ν):
The number of waves per second or Number of wavelength unit passing through a given point in unit
time. It can be measured in Hz, KHz, MHz, GHz ... or CPS.
Wavenumber (ῡ):
E = h.v eq(1)
Where,
H = Planks constant
v = C/λ ,
where,
E = h. C/λ eq(2)
Eq (1) proves that energy of electromagnetic radiation (E) is directly proportional to the frequency (v).
Eq (2) proves that energy of electromagnetic radiation (E) is inversely proportional to wavelength (λ).
That means as wavelength of light increases, the energy within it decreases and vice-versa.
3
Dept. of Pharmaceutical Quality Assurance, MCOPS Swapnil Dengale
The following table lists the names of different spectral regions, the range of frequencies and
wavelengths in those regions, and the type of transition that can occur when a photon in these spectral
ranges interacts with matter.
11 13 molecular rotations,
microwaves 3x10 -10 1 mm-25 µm
electron spin flips*
11
radio waves <3x10 >1 mm nuclear spin flips*
4
Dept. of Pharmaceutical Quality Assurance, MCOPS Swapnil Dengale
Types of spectroscopy:
Spectroscopy is classified based on the region of EMR spectrum used for the interactions with the
matter.
Spectroscopy
1. Absorption spectroscopy: Energy levels have well defined values (i.e they are quantized)
absorption only occurs when the photon’s energy matches the difference in energy, ΔE,
between two energy levels. A plot of absorbance as a function of the photon’s energy is called
an “absorbance spectrum”.
E
n
h
5
E
o
Absorption
Dept. of Pharmaceutical Quality Assurance, MCOPS Swapnil Dengale
E
n
h
E
o
Emission
1. Atomic spectroscopy: Changes in the energy take place at atomic level. Further subdivided
into Atomic Absorption spectroscopy (AAS) & Atomic Emission Spectroscopy (AES).
2. Molecular spectroscopy: Changes in energy takes place at molecular level. Further subdivided
into Molecular absorption & Molecular emission spectroscopy
Spectroscopy
When electrons absorb radiations (photons) and become excited, they move from bonding orbital
(ground state) to anti-bonding orbital (excited state). Bonding electrons like σ and π move to excited
states like σ* and π* respectively. While non-bonding electrons (n) move to either of the antibonding
orbitals (excite levels) like σ* and π*.
1. σ σ* Transition:
A transition of an electron from a bonding σ orbital to the higher energy antibonding (excited) σ*
orbital is designated σ σ* Transition. Sigma bonds are, in general, very strong, this therefore, is a
high energy process and these transitions require very short wavelength i.e. high energy UV light
approximately around 150 nm. Alkanes eg. Methane, Ethane, Cyclopropane show this type of
transition.
2. n σ* Transition:
This transition involves saturated compounds with one hetero atom (S, O, N) with lone pair of
electrons. The energy required for the transition is less when compared to σ σ* transition. The peaks
due to absorption bands of such transition occur in near UV region i.e. between 170 nm – 190 nm.
Eg. Water absorbs at 167 nm. Methanol at 174 nm & methyl chloride at 169 nm.
These kind of transitions are particularly sensitive to hydrogen bonding and thus exert solvent effect
on UV spectrum.
7
Dept. of Pharmaceutical Quality Assurance, MCOPS Swapnil Dengale
3. ππ* transition:
This transition is available in compounds with unsaturated centres eg. Simple alkenes, carbonyl
compounds etc. This transition requires lesser energy than n σ* transition. In simple unconjugated
alkene, an absorption band at 190 nm is due to such transition. The band due to ππ* transition in a
compound with conjugated π system is usually intense (ɛmax > 10000) and is frequently referred to
as K-band (German Konjugierte).
4. n π* transitions:
Unsaturated molecules having atoms like sulphur, oxygen and nitrogen exhibit n π* transition.
These atoms contain lone pair of electrons which are promoted from ground state to antibonding
excited state π*. Aldehydes, Ketones and nitro group compounds show this type of transition. This
transition involves least amount of energy than all other transitions and therefore, this transition gives
rise to an absorption band at longer wavelength.
Eg. In saturated aliphatic ketones, n π* transition around 280 nm is the lowest energy transition.
n π* transition is forbidden by symmetry consideration, thus the intensity of the band due to this
transition is low, although the wavelength is longer.
This kind of transitions arise due to the presence of radical group containing lone pair of electrons
within a structure, that’s why the absorption band due to such transition is called as R-band. (R stands
for Radikal).
Laws of photometry:
Beer’s Law:
“The intensity of a beam of parallel monochromatic radiation decreases exponentially with the
number of absorbing molecules. More simply it is stated that the absorbance is proportional to the
concentration”.
OR
Lambert’s law:
when a beam of monochromatic light is passed through a substance, a absorption of light is directly
proportional to the path length of the sample of substance.
OR
OR
More simply it is stated that the absorbance is proportional to the thickness (pathlength) of the
solution
Before learning the derivation of Beer-Lambert’s law, it is imperative to understand the relationship
between Absorbance and Transmittance.
9
Dept. of Pharmaceutical Quality Assurance, MCOPS Swapnil Dengale
P0 P
(power in) (power out)
The diagram shows a beam of monochromatic radiation of radiant power P0 directed at a sample
solution. Absorption takes place and the beam of radiation leaving the sample has radiant power P.
the amount of radiation absorbed may be measured in a number of ways.
% Transmittance = P/P0*100
Example: If no light is absorbed then P=Po and A=0. If P0 is 100 and P is 10 then Transmittance is 10%
whilst Absorbance is 1.
So,
10
Dept. of Pharmaceutical Quality Assurance, MCOPS Swapnil Dengale
So,
We have Beer-Lambert’s law which describes the dependence of T on both the path length and the
concentration
It is more convenient to omit the negative sign on the right hand side of the equation and to define
the new term absorbance
A = a.b.C or/ A = ε. b. C
Other common names for absorptivity are absorption index and extinction coefficient.
If the absorbance is divided by the product of concentration of the solution (in g/L) and solution path
length (in cm) the dividend is known as absorptivity,
a = A/bc
E1%1cm It is the absorption [log10 (Po/P)] of a 1% w/v solution in a cell with 1cm path length.
11
Dept. of Pharmaceutical Quality Assurance, MCOPS Swapnil Dengale
• Molar absorption index or Molar absorption coefficient, ε. It is the product of absrptivity (a)
and molecular weight (M),
ε = a.M
• Molar absorptivity can be obtained by dividing absorbance with the product of concentration
(expressed in g moles per litre) and path length (cm). it has the unit of 1000 cm2 mol-1 but the
units are, by convention, never expressed
If the specifies conditions are not met then there are deviations from BL law, meaning the absorbance
is not proportional to the concentration of an analyte.
1. True deviations
2. Chemical deviations
3. Instrumental deviations
True deviations:
At high concentrations, solute molecules can cause different charge distribution on their neighboring
species in the solution. Since UV-visible absorption is an electronic phenomenon, high concentrations
would possibly result in a shift in the absorption wavelength of the analyte. At times, even electrolyte
concentrations (such as those present in buffers) play an important role in altering the charge
distributions and affecting UV-visible absorbance. Some large ions or molecules show deviations even
at very low concentrations. For e.g. methylene blue absorptivity at 436 nm fails to observe Beer
Lambert law even at concentrations as low as 10μM.
12
Dept. of Pharmaceutical Quality Assurance, MCOPS Swapnil Dengale
High analyte concentrations can also possibly alter the refractive index (η) of the solution which in
turn could affect the absorbance obtained. If the addition of solute causes a significant change in the
refractive index of the solution.
Chemical deviations:
Chemical deviations occur due to chemical phenomenon involving the analyte molecules due to
association, dissociation and interaction with the solvent to produce a product with different
absorption characteristics. For example, phenol red undergoes a resonance transformation when
moving from the acidic form (yellow) to the basic form (red). Due to this resonance, the electron
distribution of the bonds of molecule changes with the pH of the solvent in which it is dissolved. Since
UV-visible spectroscopy is an electron-related phenomenon, the absorption spectrum of the sample
changes with the change in pH of the solvent.
Instrumental deviations:
Due to polychromatic radiation (Bandpass effect):
13
Dept. of Pharmaceutical Quality Assurance, MCOPS Swapnil Dengale
Equation to calculate absorbance of a sample with polychromatic light source.
When the molar absorptivities are the same at both wavelengths (i.e. ε’ = ε”) , the relationship
between absorbance and concentration follows Beer-Lambert law to obtain a straight line.
However, as the difference between ε’ and ε” increases, the deviations from linearity also
increases.
14
Dept. of Pharmaceutical Quality Assurance, MCOPS Swapnil Dengale
Stray radiation or scattered radiation is defined as radiation from the instrument that is outside
the nominal wavelength band selected. Usually the wavelength of the stray radiation is very
different from the wavelength band selected. It is known that radiation exiting from a
monochromator is often contaminated with minute quantities of scattered or stray radiation.
Usually, this radiation is due to reflection and scattering by the surfaces of lenses, mirrors,
gratings, filters and windows. If the analyte absorbs at the wavelength of the stray radiation, a
deviation from Beer-Lambert law is observed similar to the deviation due to polychromatic
radiation.
If the cells holding the analyte and the blank solutions are having different path-lengths, or unequal
optical characteristics, it is obvious that there would be a deviation observed in Beer-Lambert law.
This is due mainly to the change in the path-length. In today’s instrument this problem is generally not
observed, however if it is present, appropriate linear regression to quantify this deviation must be
made.
Chromophore:
It is a covalently unsaturated group which is responsible for absorption of UV or visible radiation and
impart colour to the compound. Chromophores can be identified by unsaturated linkages within a
compound such as -C=C-, -N=N-. In unsaturated linkages, the π electrons are loosely bound. These
loosely bound electrons require less energy for electronic transition and the absorption bond occur in
near UV region.
Eg; ethylenic, acetylenie, carbonyl, acids, esters, nitro and nitrile group. C=C, C=O, NO2 etc
I. Independent chromophore: single chromophore is sufficient to import color to the compound e.g.
Azo group
II. Dependent chromophore: Where more than one chromophore is required to produce color. e.g.
acetone having one ketone group is colorless where as diacetyl having two ketone group is yellow.
15
Dept. of Pharmaceutical Quality Assurance, MCOPS Swapnil Dengale
Auxochrome:
An auxochrome represents a saturated group containing unshared electrons which when attached to
a chromophore changes both the intensity as well as the wavelength of absorption maximum.
The n-π* transition for carbonyl compounds experiences bathochromic shift when the polarity of
the solvent is decreased.
Example: the conjugation of lone pair of electrons on the nitrogen atom of aniline with the π bond
system of the benzene ring is removed on protonation. Aniline absorbs at 280nm, but in acid solution
the main peak is almost identical with that of benzene, being at 203nm due to blue shift. In case of
C6H5N+ H3 ions formed in acidic solution, the electron pair is no longer present and hence conjugation
is removed.
16
Dept. of Pharmaceutical Quality Assurance, MCOPS Swapnil Dengale
Hyperchromic shift:
This effect leads to increased absorption intensity i.e., ε max increases. The introduction of
auxochrome usually increases the intensity of absorption.
Example: pyridine absorbs at 257 nm εmax = 2750 is shifted to 262nm εmax = 3560 for 2- methyl
pyridine. The introduction of auxochrome usually increases the intensity of absorption.
Hypochromic shift:
This effect leads to decreased absorption intensity i.e., ε max decreases. The introduction of a group
which distorts the geometry of the molecule causes hypochromic effect.
Example: biphenyl absorbs at 250nm, εmax = 19000 where 2-methyl biphenyl absorbs at 237nm, εmax
= 10250 It is due to distortion caused by methyl group in 2-methyl biphenyl
17
Dept. of Pharmaceutical Quality Assurance, MCOPS Swapnil Dengale
Instrumentation
H2 discharge lamp
Tungsten lamp
18
Dept. of Pharmaceutical Quality Assurance, MCOPS Swapnil Dengale
Radiation (Light)sources:
UV region:
a) D2 discharge lamp
b) H2 discharge lamp
Visible region:
Needs to cover region from 200-800 nm with sufficient power and must be constant over the
time taken to make the measurement.
The source should provide continuous radiation
The source must generate beam with sufficient power for ready detection and measurement
It should be stable. Either a fully charged high capacity battery or an A.C supply with a
transformer or voltage regulator is necessary to provide constant intensity.
Deuterium lamp:
This lamp consists of a silica or quartz envelope containing D2 filled under low pressure. The
mechanism by which continuous radiation is produced by this source involves the formation of an
excited molecule D2* by absorption of electrical energy. This species then dissociates to give 2
deuterium atoms plus an Ultraviolet photon.
19
Dept. of Pharmaceutical Quality Assurance, MCOPS Swapnil Dengale
When voltage is applied across electrodes, an arc is formed between heated, oxide coated filament
and a metal electrode, which gives away high intensity beam. The intensity of D2 discharge lamp is
about 3-5 times the intensity of H2 discharge lamp. The D2 lamp provides continuous spectrum in the
region of 160 to 375 nm. At longer wavelengths (> 360 nm), the lamp generates emission lines. These
lines are nuisance but they are useful for wavelength calibration.
H2 discharge lamp:
It is widely used UV radiation source, consists of 2 electrodes enclosed in a glass tube with a quartz
window. The glass tube is filled with H2 gas under relatively high pressure. When voltage is applied
across the electrodes, the hydrogen molecules are excited to higher energy level. As like D2 discharge
lamp, while returning to the ground state, the electrons emit radiation in the near UV region (180-350
nm). H2 discharge lamps are stable and robust but their intensity is significantly less than D2 lamp.
Xenon lamp is an electric discharge lamp which utilizes ionized xenon gas to produce an extremely
intense white light light for short duration. It consists of a sealed tube made up of glass or fuse quartz
which is filled with xenon gas at a pressure of 10-30 atmospheres. Additionally it contains a pair of
tungesten metal electrodes which can carry electric current to the gas. The electrodes are connected
to a capacitor which is charged at a high voltage.
20
Dept. of Pharmaceutical Quality Assurance, MCOPS Swapnil Dengale
The xenon lamp generates significant amounts of UV radiation and shows spectral lines in the UV
region. Its intensity is higher than the hydrogen discharge lamp. The disadvantage of xenon lamp is
that the UV radiation released by it causes generation of ozone by ionization of the oxygen molecules.
Tungsten lamp:
It is simmilar to an electric bulb. Tungsten is the most widely used and most suitable material for lamp
filaments. It is used when polychromatic light is required. It consists of tungsten filament enclosed in
a vacuum bulb, which is heated electrically at a higher temperature to produce white light. It requires
a potential of 3-220 volts. It emits continuous radiations over wide wavelenght region (350 – 2500
nm). The energy obtained from a tungsten lamp can be used in wavelength regions above 375 nm. It
is available in different shapes and sizes suitable for fixing into different instruments.
21
Dept. of Pharmaceutical Quality Assurance, MCOPS Swapnil Dengale
Wavelength Selectors:
1) Dispersive type
2) Non-dispersive type
1) Prism monochromators:
a. Refractive type
b. Reflective type
2) Grating monochromators:
a. Diffraction grating
b. Transmission grating
1) Absorption filter
2) Interference filter
Monochromator:
22
Dept. of Pharmaceutical Quality Assurance, MCOPS Swapnil Dengale
Prism monochromators:
The prisms are usually made-up of glass, quartz or fused silica. They disperse the polychromatic light
falling onto it into its component rainbow colours according to their wavelength. Resolution depends
upon the size and refractive index of the prism. Prisms are usually used in inexpensive instruments.
Disadvantage: Significant portion of the radiation is absorbed reducing the intensity of radiation. Very
high bandpass (30-150 nm).
Light from the radiation sources passes through the entrance slit and falls on collimator. The radiations
are collimated and dispersed into component wavelengths by the prism. Entrance slit selects any one
of the seven colours i.e. VIBGYOR. The required wavelength is selected either by rotating the prism or
by keeping the prism stationary and moving the exit slit. Refraction in the prism occurs as per snell’s
law.
Its working is similar refractive type but a reflective surface is present on one side of the prism. The
dispersed radiation gets reflected & can be collected on the same side of the source of light.
23
Dept. of Pharmaceutical Quality Assurance, MCOPS Swapnil Dengale
Grating monochromator:
Gratings are made up of glass, quartz and alkyl halides like KBr. They are highly efficient than prisms
in converting a polychromatic light into monochromatic light. They consist of densely arranged parallel
lines (grooves or rulings). A grating for UV visible region typically contain 300 to 2000 grooves/mm.
They give resolution of 0.1 nm.
It works on the principle of diffraction of light. It is most commonly used grating, where the grooved
surface is made up of glass with reflective coating (Al, Pt). The ray incident upon the grating gets
reinforced with the reflected ray and hence the resulted radiation has wavelength which is governed
by the equation
24
Dept. of Pharmaceutical Quality Assurance, MCOPS Swapnil Dengale
nλ = d (sinβ + sinγ)
where, d= distance between 2 grooves, β= Incident angle, γ= diffraction angle and n = 1, 2, 3 positive
integers.
Constructive and deconstructive interference occurs because light travels different distances when
reflected from each grating.
This is the reflection (diffraction) type of grating monochromator. Entrance slit that provides a narrow
optical image of the radiation source. A collimator (concave mirrors) that renders the radiation
emanating from the entrance slit parallel. A reflection (diffraction) grating disperses the light. A second
collimator re-forms the image of the entrance slit and focus it onto the exit slit. The exit slit to isolate
the desired spectral band by blocking all of the dispersed radiation except that in the desired range.
Transmission grating:
25
Dept. of Pharmaceutical Quality Assurance, MCOPS Swapnil Dengale
Similar to diffraction grating, but refraction takes place instead of reflection. When the radiation is
transmitted through grating, it reinforces with the partially refracted radiation. Wavelength produced
is given by the equation
θ = Angle of diffraction.
Filters:
Filters operate by absorbing all but a restricted band of radiation from a radiation source. 2 types of
filters are used in spectroscopy
1) Absorption filters
2) Interference filters
Absorption filters:
These filters, which are generally less expensive and more rugged than interference filters, are limited
in use to visible region. This type of filter usually consists of a colored glass plate that removes part of
26
Dept. of Pharmaceutical Quality Assurance, MCOPS Swapnil Dengale
the incident radiation by absorption. Absorption filters have effective bandwidth that range from 30
to 250 nm.
1) A circle (Filter wheel) is drawn which is divided into six equal parts (as shown in above fig.).
2) Each division is given a colour name i.e. Violet, Blue, Yellow, Orange and red either in clockwise
or anticlockwise direction.
3) The desired colour of the filter will be complementary to the colour of the solution
Advantages:
Disadvantage:
27
Dept. of Pharmaceutical Quality Assurance, MCOPS Swapnil Dengale
Interference filters are used with UV and Visible radiation. They consist of a middle layer of a dielectric
material such as CaF2, MgF2, SiO. The dielectric film is sandwiched between two semi-reflective silver
films. The thickness of dielectric film can be of 1st order (1λ/2) or 2nd order (2λ/2). Interference filters
have a band pass of 10-15nm and peak transmittance of 40-60%.
When a beam of light falls on an interference filter, a part of it is reflected back while the remaining
gets transmitted by the metallic layer (silver film). The reflected rays undergo interference
simultaneously, thus giving monochromatic radiation of wavelength, which is expressed by the
following equation-
λ = 2εt /n
where,
Advantages:
Disadvantage:
1) Because of the band pass, the resolution is not high as that achieved by prism or grating
monochromators.
sample cells (cuvettes) are used to hold sample solutions. Their shape and material of construction
varies depending upon the instrument and nature of the sample being analysed. They may be
rectangular or cylindrical in shape. The thickness (path length) of the cell is normally 1cm, however
29
Dept. of Pharmaceutical Quality Assurance, MCOPS Swapnil Dengale
cells with long path-lengths upto 10cm or short path-length upto 1-2 mm are also available. Before
taking the measurement, the sample cell must be thoroughly cleaned to avoid any contamination. The
level of the sample solution must be upto the mark etched on its surface or in absence of any such
indication, at least 3/4th volume of the total cuvette. Materials used to construct sample cells are
quartz, Alkali metal halide and glass.
1) The material used in the construction of the sample cell should not chemically react with the
solvent.
2) The material should be optically UV/Visible transparent i.e. it must not absorb the UV
radiation by itself.
3) It should have uniform thickness.
To obtain spectroscopic information, the radiant power transmitted or emitted must be detected in
some manner and converted into a measurable quantity. A detector is a device that indicates the
existence of some physical phenomenon. Familiar example of detectors includes the mercury level in
a thermometer (indicating temperature). The human eye is also a detector, it converts visible radiation
into an electric signal that is passed to the brain via a chain of neurones in the optic nerve and produce
vision.
Invariably in modern instruments, the information of interest is encoded and processed as an electric
signal. The term transducer is used to indicate the type of detector that converts quantities, such as
light intensity, pH, mass and temperature, into electrical signals that can be subsequently amplified,
manipulated, and finally converted into numbers proportional to the magnitude of the original
quantity. There are 2 general types of detectors, one type responds to photons (photon detectors),
the other to heat (thermal detectors). All detectors used in UV spectroscopy are photon detectors.
The ideal detector for electromagnetic radiation responds rapidly to low levels of radiant
energy over a broad wavelength range.
It produces an electrical signal that is easily amplified and has a low electrical noise level.
The electrical produced by the detector be directly proportional to the radiant power P of the
beam as shown in below equation
30
Dept. of Pharmaceutical Quality Assurance, MCOPS Swapnil Dengale
G = KP + K’
Where, G is the electrical response of the detector in units of current, voltage or charge. The
proportionality constant K measures the sensitivity of the detector. Many detectors exhibit a small
constant response K’, known as dark current, ever when no radiation strikes their surface.
• It consists of semiconductor, such as selenium, which is deposited on a strong metal base such
as iron.
• A very thin layer of silver or gold sputtered over the surface of the semiconductor to act as a
second collector electrode
Working:
- light of sufficiently high energy passes through the thin transparent silver layer and hits
selenium causing electrons to be released which move across barrier toward silver layer and
31
Dept. of Pharmaceutical Quality Assurance, MCOPS Swapnil Dengale
Advantage:
cheap, rugged.
Disadvantage:
The response of phototube is based on the photoelectric effect. As shown in above figure,
phototube consists of a semi cylindrical photocathode and a wire anode sealed inside an evacuated
32
Dept. of Pharmaceutical Quality Assurance, MCOPS Swapnil Dengale
transparent glass or quartz envelope. The concave surface of the cathode supports a layer of photo-
emissive material, such as an alkali metal or a metal oxide, that emits electrons when irradiated with
light of the appropriate energy.
Working:
When a voltage is applied across the electrodes, the emitted photoelectrons are attracted to the
positively charged wire anode. A photocurrent that produces after the completing circuit is then
easily amplified and measured. The number of photoelectrons ejected from the photocathode per
unit time is directly proportional to the radiant power of the beam striking the surface.
Advantages:
Limitations:
33
Dept. of Pharmaceutical Quality Assurance, MCOPS Swapnil Dengale
PMT is similar in construction to the phototubes but is significantly more sensitive. Its photocathode
is similar to that of the phototube, with electrons being emitted on exposure to radiation. In place of
a single wire anode, however, the PMT has a series of electrodes called dynodes, as shown in the
figure.
Working:
The electrons emitted from the cathode are accelerated towards the first dynode, which is maintained
90 to 100V positive with respect to cathode. Each accelerated photoelectron that strikes the dynode
surface produces several electrons, called secondary electrons, that are then accelerated to dynode
2, which is held 90 to 100 V more positive than dynode 1. Again electron amplification results. By the
time this process has been repeated at each of the dynode, 105 to 107 electrons have been produced
for each incident photon. This cascade of electrons is finally collected at the anode to provide an
average current that is further amplified electronically and measured.
Advantages:
Even 200 times weaker signals can be processed. Hence useful for fluorescence
measurements.
Limitations:
PMT should be shielded from stray light in order to have accurate results.
Photodiode: a type of diode that allows current to flow only when light strikes the diode. P type:
positive mobile charge carrier. n Type: negative mobile charge carrier. Photodiodes are constructed
from a layer of p-type silicon on a n-type silicon substrate, creating a p-n junction diode, which is
overlaid with a protective a SiO2layer.
34
Dept. of Pharmaceutical Quality Assurance, MCOPS Swapnil Dengale
Working:
A reverse bias is applied, drawing electrons and holes away from the junction. There is a depletion
region at each junction, in which there are few electrons and holes. The junction acts as a capacitor,
with charge stored on either side of the depletion region. At the start of each measurement the diode
is fully charged. When radiation hits the semiconductor, free electrons and holes are created and
migrate to regions of opposite charge resulting in increased conductivity which is easily measured and
is directly proportional to the radiant power.
A silicon-diode detector is more sensitive than a simple vacuum phototube but less sensitive that PMT.
A linear photodiode array comprises many small silicon photodiodes formed on a single silicon
chip. There can be between 64 to 4096 sensor elements on a chip, the most common being
1024 photodiodes.
Advanced technology led to the development and implementation of photodiode detectors,
which when placed closely spaced linear arrays, after rapid and accurate spectrum analysis.
The primary advantage of linear array detectors is that they permit the simultaneous analysis
of an entire spectrum over a period of a few sec.
This advantage when performing kinetic studies involving rapidly changing events
35
Dept. of Pharmaceutical Quality Assurance, MCOPS Swapnil Dengale
In a single beam UV-visible spectrophotometer, light from the radiation source after passing through
a monochromator enters the sample cell containing the sample solution. A part of incident light (I0) is
absorbed by the sample and remaining gets transmitted (It). the transmitted light strikes the detector
and produces electrical signals. the absorbance readings of both the standard and unknown solutions
are recorded after adjusting the instrument to 100% transmittance with a blank solution each time
whenever the wavelength is changed.
Advantages:
Simple in construction.
Easy to operate.
Economical.
Disadvantage:
Any fluctuations in the intensity of the radiation sources affects the absorbance readings.
It requires adjustment of transmittance to 100% whenever wavelength is changed. Hence, a
continuous spectrum is not obtained.
36
Dept. of Pharmaceutical Quality Assurance, MCOPS Swapnil Dengale
Double beam spectrophotometer (DBS) allows direct measurement of the ratio of intensities of the
sample and reference beams. The design of a double beam spectrophotometer is similar to single
beam spectrophotometer except that it contains a beam splitter (selector mirror) or chopper. In DBS,
the radiations from source are allowed to pass through the entrance slit into the monochromator.
Monochromator selects the required wavelength of light which is then passed through the exit slit
and received by rapidly rotating beam splitter or chopper (in the fig. selector mirror). The chopper
splits the monochromatic beam of light into two beams of equal intensities. One beam is passes
through the sample cell and the other through the reference cell. After passing through reference and
sample cells, the transmitted beams reach the detectors and produce a pulsating current which is
proportional to the intensities of incident (I0) and transmitted light (It). The detectors are connected
to an amplifier and readout device which gives the final results in terms of absorbance (-log It/I0) or
transmittance (It/I0) by combining the output by using algorithm.
Advantages:
Disadvantages:
Construction is complicated.
Instrument is expensive compared to Single beam spectrophotometer.
Applications of UV spectroscopy:
1. Qualitative applications:
Conjugation of a double bond with another double bond or C=O group causes the absorption band to
shift to longer wavelength with greater intensity.
Example:
CH2OH
CH2OH
The presence of benzene within structure can be confirmed by the absorbance at wavelengths 243,
250 and 245 nm.
38
Dept. of Pharmaceutical Quality Assurance, MCOPS Swapnil Dengale
2-pyridone and 2-hydroxy pyridine are tautomeric forms of each other (like phenol red). 2-hydroxy
pyridine shows absorption at 325 nm, while 2-pyridone has exocyclic double bond which
bathochromically shift the absorption maxima to 330 nm.
Trans isomer absorbs at longer wavelength with greater intensity than the cis isomer.
Example: Cis-stilbene has λmax at 280 nm, while trans-stilbene absorbs at 300 nm.
Quantitative applications:
39
Dept. of Pharmaceutical Quality Assurance, MCOPS Swapnil Dengale
The analyte can be quantified by measuring the absorbance and converting it into concentration and
this can be done by 2 ways. First, if absorptivity value of the analyte under consideration is given eg.
715 for paracetamol then the absorbance can directly be converted into concentration using BL law.
this method is called as Single point assay, where only single absorbance is needed for quantification.
Second, if reference standard of the analyte is provided then calibration curve can be prepared by
measuring the absorbance of appropriately prepared calibration standards. The absorbance of
unknown solution can be extrapolated to calculate the concentration with the help of calibration
curve. This method is called as multipoint assay. Eg. Colorimetric estimation of sulpha drug.
Multicomponent analysis:
The spectrophotometric assay of drugs rarely involves the measurement of absorbance of samples
containing only one absorbing component. The pharmaceutical analyst frequently encounters the
situation where the concentration of more than one component is required to be analysed in the
sample known to contain other absorbing substances which potentially interfere in the assay. In such
a scenario, multicomponent analysis techniques come into picture. The basic assumption behind the
all multicomponents analysis methods is that at all wavelengths the absorbance of a solution is the
sum of the absorbances of the individual component.
If a sample contains two absorbing drugs (X and Y) then each of them must absorb at the λmax
of other (i.e λ1 and λ2).
Two equations can be constructed based on the fact that at λ1 and λ2 the absorbance of the mixture
is the sum of the individual absorbances of X and Y.
At λ1,
At λ2,
Cy = A2 – ax2.cx/ay2 (3)
In above equation there are two unknown variables i.e. cy and cx. So let us find out the term cx.
By substituting the absorptivity and absorbances values, one can calculate cx.
• It depends on the property that, for a substance which obeys Beer’s law at all wavelengths,
the ratio of absorbance's at any two wavelengths is a constant value independent of
concentration or path length.
eg: two different dilutions of the same substance give the same absorbance ratio A1 /A2, in the USP
this ratio is referred to as a Q value.
The BP also uses a ratio of absorbance's at specified wavelengths in certain confirmatory test of
identity.
eg: cynocobalmine exhibits three λmax , at 278nm, 361 nm and 550nm. The ratio A361 /A550 is required
to be 3.30 ± 0.15and the A361 / A278 to be 1.79 ± 0.09
In the quantitative assay of two components in admixture by the absorbance ratio method,
absorbances are measured at two wavelengths. One being the λmax of one of the components λ2 and
the other being the wave length of equal absorptivity of the two components λ1 i.e an iso-absorptive
point. Two equations are constructed as described above for the method of simultaneous equations.
And further solved to find out the individual concentrations.
42
Dept. of Pharmaceutical Quality Assurance, MCOPS Swapnil Dengale
References:
1. Beckett AH, Stenlake JB. Practical Pharmaceutical Chemistry: Part II Fourth Edition [Internet].
Bloomsbury Academic; 1988. (Practical Pharmaceutical Chemistry). Available from:
https://books.google.co.in/books?id=Up3L2dAI7k8C
2. Kalsi PS. Spectroscopy of Organic Compounds [Internet]. New Age International Pvt; 2016.
Available from: https://books.google.co.in/books?id=-gTdjwEACAAJ
4. Skoog DA, West DM, Holler FJ, Crouch SR. Fundamentals of Analytical Chemistry [Internet].
Cengage Learning; 2013. Available from:
https://books.google.co.in/books?id=0bBWPgAACAAJ
43