M4140 Application Note
M4140 Application Note
Application Note
This application note discusses using the M4100 and the M4140 clamp-on CT probe to
test individual capacitors in a capacitor bank without disconnecting them from the
bank.
For this application, the M4100 serves solely as the measurement instrument. The power
source is the M4110, which provides much higher AC output (up to 25 A). The M4100
monitors the voltage supplied by the M4110 through the red low-voltage (LV) lead. The
M4100 measures the current drawn by the individual capacitor through a clamp-on CT
probe connected to the blue low-voltage lead.
Advantages of using this application of the M4100 to test capacitor banks are:
• Does not require disconnecting each capacitor unit from the bank, reducing test
time.
• Each capacitor unit test takes about one minute to perform.
• Provides more effective troubleshooting of high-voltage AC capacitors than a
low DC voltage capacitance test.
• Test conditions and test data are automatically recorded and saved. This greatly
facilitates data analysis, management, and exchange, and avoids human error.
• A comprehensive bank of test data and conditions (voltage, current, watts loss,
power factor, capacitance, correction factors, temperature, humidity, date and
time) is essential for comparison and trend analysis.
• For M4100 users, this test requires little investment and further enhances the
usefulness of the M4100.
• The M4110 affords extensive control and protection of the external source
voltage.
The capacitor bank test has been used to successfully troubleshoot an intermittent
voltage unbalance and assess the condition of capacitor banks [1, 4]. The test procedure
and interpretation are simple and can be easily implemented in a maintenance program
or used for troubleshooting.
This test cannot detect fluid leaks in early stages that have not affected the capacitor’s
internal condition. Visual inspection is the best technique for this problem.
Requirements
Hardware:
• M4100 instrument
• M4140 Capacitor Bank Test Kit (clamp-on CT probe and lead)
• M4110 Leakage Reactance Interface
Software:
• M4000 software with M4120 External Reference Module, version 2.1 or later
Test Principle
The M4110 applies an external source voltage to the capacitor bank (Figure 5 through
Figure 8). The M4100 measures the test potential and individual cell current magnitude
and angle (angle reference to the applied voltage). From this data, the software extracts
the electrical parameters (voltage, current, watts, % PF, and capacitance) for each
individual cell and displays it on the External Reference screen of the M4000 software.
The changes in % PF and capacitance indicate the change in condition of the capacitor.
The current of each capacitor Ii is measured through the clamp-on current probe
(Figure 1), which is connected to the M4100 via the blue LV lead (Figure 6). The test
voltage is measured via the red LV lead through the M4110 Red terminal (Figure 6). The
M4100 determines the watts loss, power factor, and capacitance of each tested
capacitor using the measured voltage, current, and phase angle between the voltage and
current, in the same manner as in power factor testing [2].
subjected to the same voltage. Then the group can be connected to other groups, to
form multiple parallel or series-parallel groups. They can also be connected in a serial
string in different racks, and then the strings are connected together in parallel (Figure 8
through Figure 10 and Figure 2d).
(a) Parallel Group (b) Multiple Parallel Groups (c) Series Parallel Groups (d) Serial String Groups
Unit Identification
Parallel Group
For a parallel group, we use Xn to identify the group and Xn-m to identify a unit in the
group, where:
X = phase identification (ex. A, B, and C)
n = rack position starting from the line end to neutral for series-parallel groups or
from top to bottom starting from the outside rack for parallel groups
m = unit position in the rack, starting from left to right when facing the capacitor
rack
Serial String
For a serial string, we use Xk-n to identify a string in the group and Xk-n-m to identify a
unit in the string, where:
X = phase identification (ex. A, B, and C)
k = position of the string group
n = string position in the group, starting from the current-transformer side
m = unit position in the string, starting from phase end (top) to neutral end (bottom)
Figure 3 and Figure 4 are examples of capacitor identification for a capacitor bank that
consists of five-series parallel groups.
A3-1
Rack Position Number Unit Position
1 1 2 ... 14
3
2
A3-14
4
5
1 2 3 ... 14 5 ...
5 4
Grounded Neutral
Grounded Neutral Terminal
Terminal
Figure 4. One Phase of the Five Series-Parallel Groups
Test Procedures
WARNING Before attempting to test a capacitor bank, establish a safety procedure that complies
with your policies, standards, and regulations. Isolate, ground, and discharge the
capacitor bank.
Parallel Group
To test a unit in a parallel or series-parallel group: Move the current probe from unit to
unit until the end of the group (Figure 5 and Figure 6). The same process applies to each
group for multiple parallel or series-parallel groups (Figure 7).
Bottom Bar
M4100
M4110
Top Bar
Bottom Bar
M4100
Sensor PC
Figure 6. Test Connection for a Unit in a Parallel Group
Line
M4110
M4100
Neutral
Serial String
To test an entire serial string: Move the current probe from one string to the next until
the end of the group (Figure 8). The same process applies to each string in the group.
To test a unit in a serial string: Move the Voltage Sense lead from unit to unit until the
end of the string (Figure 9 and Figure 10).
Voltage Lead
Current Lead
Clamp-on Is
M4100 Probe
M4110
Source Lead VT
Common Neutral Bar
Figure 8. Test Principle for an Entire Serial String
Voltage Lead
Vi
Current Lead
Is
Clamp-on
M4100 Probe
M4110 To Other
Groups
Source Lead VT
Common Neutral Bar
Figure 9. Test Principle for a Unit in a Serial String
M4110
Voltage Lead Vi
M4100
Current Lead
Is
Sensor PC
VT
Common Neutral Bar
Figure 10. Test Connection for a Unit in a Serial String
Test
M4110 Connections M4100 Connections Description
Blue LV Lead –
Test Voltage Sense Lead Red Lead Black Lead Mode Red LV Lead Current Probe
1 Black to Black Lead Cap. Group Cap. Group UST-B M4110 Red Cap. Unit Fig.1B
White to Red Lead Top Bottom Terminal Bushing and 4
Repeat Test 1 on each unit in the group by moving only the current probe. Repeat the process for each
parallel group in the capacitor bank. The current probe polarity must agree with the current circulation or
source polarity. If the M4110 main breaker trips when the safety switch is depressed, remove the ground on
the capacitor bank neutral.
2 Across the capacitor Cap. String Cap. String UST-B M4110 Red Cap. Unit Fig. 2C
unit Bottom Top Terminal Bushing
Repeat Test 2 on each unit in the string by moving only the Voltage Sense lead. Repeat the process for each
parallel string in the capacitor bank. The current probe and VS lead polarity must agree with the current
circulation or source polarity. The black VS leads are across the capacitor unit under test, with the black lead
closer to the black source lead and the white lead closer to the red source lead.
Running a Test
1. Place the cursor on the desired starting test line.
2. Press F2, click the Single Test (lightning bolt) icon, or select Test > Run The Tests.
The External Reference Test dialog box appears (Figure 12).
4. The Test-In-Progress dialog box appears (Figure 14), with the message SET TEST
VOLTAGE for LINE (x), where (x) is the selected line number.
5. If this is the first test in the capacitor group, use the M4110 voltage knob to
adjust the test voltage to the desired level (240 V), displayed in the voltage
ramping bar.
If this is a subsequent test in the group, go to step 6.
6. When you are ready to take the measurement, press F3.
The message DO NOT CHANGE VOLTAGE – COLLECTING DATA appears.
7. When the message TEST DONE, LINE (x) appears, press one of the function keys
indicated at the bottom of the screen to continue. These options are also
available from the toolbar and menus.
• F1. Opens the Help window.
• F2/F11. F2 restarts a single test; F11 restarts a multiple test.
• F5. Accepts the results of the test. When performing multiple tests, only
those completed before a stopped test are accepted.
• F6. Discards the results.
• F7. Prints the results.
• F8. Saves the results.
NOTE Capacitance in the Test Results column of Figure 12 or Figure 14 becomes Inductance
when the current probe or Voltage Sense lead polarity does not agree with the source
polarity. Reverse the current probe or VS lead to correct the polarity. Abnormally low
capacitance can also be caused by current-probe jaws that are not fully closed.
Test Voltage
The test voltage should not exceed 600 V, due to the rated voltage of the leads and
current probe. The total test current is limited to the rated current of the test source.
In general, the total test current (I) is a function of the test voltage (V), frequency (f), and
total capacitance (C) of the capacitor group under test. Equations 1 and 2 permit
calculating the test current and capacitance of the group under test.
I = 2πfCV (1)
where:
I = total test current
f = test frequency
C = total capacitance
V = test voltage
and
N
C = ∑ Ci (2)
i=1
where:
N = the unit number in the capacitor group
Ci = the capacitance of the unit capacitor
For a 60 Hz system, the preferred test voltage is 240 V. If the total capacitance of the
capacitor group under test is too large, causing the protective devices to operate, reduce
the test voltage to 120 V. For a 50 Hz system, you can use Equation 1 and the above
general guideline. Table 2 shows the summary.
60 Hz System 50 Hz System
120 ≥ 221
If the test current still exceeds the rated current of the source due to the size
of the capacitance of the group capacitor, the test voltage can be reduced.
References
[1] Long Pong, “Field Testing Capacitor Banks with M4000 Test Instrument,” Minutes of
the 73nd Annual International Conference of Doble Clients, 2006, in Arresters,
Capacitors, Cables and Accessories Committee paper ACCA 7.
[2] Doble Engineering Company, M4000 Insulation Analyzer User Guide, PN 500-0110.
[3] Doble Engineering Company, M4110 Leakage Reactance Interface User Guide, PN
500-0396.
[4] Long Pong, “Update – Field Testing Capacitor Bank with M4000 Test Instrument,”
Minutes of the 74th Annual International Conference of Doble Clients, 2007, in
Arresters, Capacitors, Cables and Accessories Committee paper ACCA 3.