Manual Estufa de Cultivo Celular - Model 310 Series Forma Direct Heat Co2 Incubator
Manual Estufa de Cultivo Celular - Model 310 Series Forma Direct Heat Co2 Incubator
Important Read this instruction manual. Failure to read, understand and follow the instructions in this manual
may result in damage to the unit, injury to operating personnel, and poor equipment performance. ▲
Warning If this unit is not used in the manner specified in this operating manual, the protection provided by the
equipment design may be impaired. ▲
Caution All internal adjustments and maintenance must be performed by qualified service personnel. ▲
Material in this manual is for information purposes only. The contents and the product it describes are subject to
change without notice. Thermo Fisher Scientific makes no representations or warranties with respect to this
manual. In no event shall Thermo be held liable for any damages, direct or incidental, arising out of or related to
the use of this manual.
Important operating and/or maintenance instructions. Read the accompanying text carefully.
Potential electrical hazards. Only qualified persons should perform procedures associated with this
symbol.
Equipment being maintained or serviced must be turned off and locked off to prevent possible injury.
Hot surface(s) present which may cause burns to unprotected skin, or to materials which may be
damaged by elevated temperatures.
Marking of electrical and electronic equipment, which applies to electrical and electronic equipment
falling under the Directive 2002/96/EC (WEEE) and the equipment that has been put on the market
after 13 August 2005.
This product is required to comply with the European Union’s Waste Electrical & Electronic
Equipment (WEEE) Directive 2002/96/EC. It is marked with the WEEE symbol. Thermo Fisher
Scientific has contracted with one or more recycling/disposal companies in each EU Member State
European Country, and this product should be disposed of or recycled through them. Further
information on Thermo’s compliance with this directive, the recyclers in your country and
information on Thermo products will be available at www.thermo.com.
✔ Always use the proper protective equipment (clothing, gloves, goggles, etc.)
If you do, please contact us 8:00 a.m. to 6:00 p.m. (Eastern Time) at:
1-740-373-4763 Direct
1-800-438-4851 Toll Free, U.S. and Canada
1-877-213-8051 FAX
http://www.thermoscientific.com Internet Worldwide Web Home Page
[email protected] Service E-Mail Address
Our Sales Support staff can provide information on pricing and give you quotations. We can
take your order and provide delivery information on major equipment items or make
arrangements to have your local sales representative contact you. Our products are listed on the
Internet and we can be contacted through our Internet home page.
Our Service Support staff can supply technical information about proper setup, operation or
troubleshooting of your equipment. We can fill your needs for spare or replacement parts or
provide you with on-site service. We can also provide you with a quotation on our Extended
Warranty for your Thermo Scientific products.
Whatever Thermo Scientific products you need or use, we will be happy to discuss your
applications. If you are experiencing technical problems, working together, we will help you
locate the problem and, chances are, correct it yourself...over the telephone without a service
call.
When more extensive service is necessary, we will assist you with direct factory trained
technicians or a qualified service organization for on-the-spot repair. If your service need is
covered by the warranty, we will arrange for the unit to be repaired at our expense and to your
satisfaction.
Regardless of your needs, our professional telephone technicians are available to assist you
Monday through Friday from 8:00 a.m. to 6:00 p.m. Eastern Time. Please contact us by
telephone or fax. If you wish to write, our mailing address is:
Warranty Notes
Information You Should Know Before Requesting Warranty Service
• Locate the model and serial numbers. A serial tag is located on the unit itself.
• For equipment service or maintenance, or with technical or special application inquiries, contact Technical
Services at 1-800-438-4851 or 1-740-373-4763 (USA and Canada). Outside the USA, contact your local
distributor.
• Chamber Gas Sample Port - Used for sampling chamber CO2 content
using a FYRITE or similar instrument.
Keypad Operation The Model 310 Series direct heat incubator has four basic modes which
allow incubator setup: Run, Setpoints, Calibration and System
Configuration.
Run is the default mode which the incubator will normally be in during
operation.
Set is used to enter system setpoints for incubator operation.
Calibration is used to calibrate various system parameters.
Configuration allows for custom setup of various options.
The chart below shows the selections under each of the modes.
RUN SETPOINT CALIBRATION CONFIGURATION
Default Mode Temperature Temp Offset Audible
Overtemp CO2 Cal1 Access Code
CO2 IR Cal2 Temp Lo Alarm
RH Cal Temp Relay
CO2 Lo Alarm
CO2 Hi Alarm
CO2 Relay
CO2 Z & S #'s *
RH Lo Alarm
RH Relay
Display Temp
Display Rh
Tank Select
Gas Guard
RS485 Address
1
T/C units only Base Unit Displays
2
IR units only Option Displays
*T/C units only
Scroll for Parameters Arrows: Steps the operator through the parameters
of SET, CAL and CONFIG Modes. The right arrow goes to the next
parameter, the left arrow returns to the previous parameter.
Up Arrow: Increases or toggles the parameter value that has been selected
in the SET, CAL, and CONFIG Modes.
Enter: Must press Enter key to save to memory all changed values.
Down Arrow: Decreases or toggles the parameter values that have been
selected in the SET, CAL and CONFIG Modes.
Silence Key: Press to silence the audible alarm. See Section 4 for alarm
ringback times.
Displays Message Center: Displays the system status (Mode) at all times. Displays
SYSTEM OK during normal operation, or alarm messages if the system
detects an alarm condition. See Section 4, Alarms.
Upper and Lower Displays: These 7 segment displays vary depending upon
the options present and the configuration chosen. The upper display can
display temp or RH, or toggle between them. The bottom display shows
CO2 continuously.
Installing the Warning Single and stacked units must be installed against a wall or
Incubator similar structure. Maintain a three-inch clearance behind the incubator for
electrical and gas hook-ups. ▲
1. Locate the unit on a firm level surface capable of supporting the unit’s
weight of 205 lbs.
2. Locate the unit away from doors and windows and heating and air
conditioning ducts.
3. Lift the unit only by the sides of the cabinet base . Do not attempt to
lift it by the front and back . This places stress on the outer door
hinges.
Stacking the Incubators Caution When stacking incubators, the direct heat incubator must be the
top unit. Never stack a water-jacketed incubator on top of a Model 310
Series unit. ▲
Warning If the units have been in operation, shut them both off and
disconnect from the power source before beginning any service work. ▲
Figure 1-3.
Stacking Brackets
5. Align the sides of the top unit with the bottom unit and slide the top
unit forward until the notches in the base of the top unit align with
the bolts in the top of the bottom unit. The flatwashers should slide
over the base frame. Tighten bolts.
6. Remove the four nylon plugs on the lower back of the upper incubator.
7. Insert the stacking brackets into the slots on the back of the control
panel of the bottom unit as shown in Figure 1-6.
Stacking (continued) 8. Align the slots in the brackets with the mounting holes on the back of
the top incubator. Secure the brackets with the screws provided in the
stacking kit.
9. Make sure the hinge pins are in the hinge brackets. Slide the door onto
the hinge pins.
10. Secure the door heater cable to the connector below the door and
replace the kick plate on the top unit, taking care not to crimp cable.
11. The stacked incubators are ready for service.
Preliminary Cleaning and Remove vinyl from shelf channels, duct sheets, and air duct, if present.
Disinfecting Using a suitable laboratory disinfectant, disinfect all interior surfaces
including shelves and shelf supports, door gaskets, blower wheel and CO2
sensor. Refer to Section 5.
Caution Before using any cleaning or decontamination method except
those recommended by the manufacturer, users should check with the
manufacturer that the proposed method will not damage the equipment.
Accidental spills of hazardous materials on or inside this unit are the
responsibility of the user. ▲
Installing the Shelves 1. Install the large sheet metal shelf supports with the tabs facing into the
center of the chamber with their slots up. There are no right side or left
side supports. Simply turn one of them to fit the opposite side. Tilt the
shelf supports as they are placed in the chamber so the tops fit into the
top air duct, then guide them into the vertical position. Figure 1-7
shows the support as it would be oriented for right side of chamber.
Installing the Shelves 2. Note there is no difference in the left side and right side shelf channels.
(continued) 3. Install the shelf channels by placing the channel’s rear slot over the
appropriate rear tab on the shelf support. Pull the shelf channel
forward and engage the channel’s front slot into the shelf support’s
appropriate forward tab. Refer to Figure 1-8.
Installing Access Port Locate the opening in the top left corner on the inside of the chamber.
Filter Remove the tape covering the opening on the outside of the unit. Locate
the stopper with filter in the hardware bag. Install the assembly in the
opening inside the chamber. See Figure 1-9.
Installing (optional) HEPA Caution Be careful when handling the filter. The media can be damaged if
Filter it is mishandled.
To avoid damage to the incubator, do not operate the unit without the
HEPA filter in place. ▲
Installing the Air Sample 1. Remove the filter from the shipping bag.
Filter
2. Install the air sample filter assembly to the black hose barb behind the
top duct.
3. Insert the other end of the filter assembly into the larger hole in the air
intake to the blower. Refer to Figure 1-10.
4. Verify that the temperature sensor is installed into the smaller hole in
the intake, as in Figure 1-10.
Leveling the Unit Check for level by placing a bubble style level on one of the shelves. Turn
the hex nut on the leveler counterclockwise to lengthen the leg or
clockwise to shorten it. Level the unit front-to-back and left-to-right.
See the serial tag on the side of the unit for electrical specifications or refer
Connecting the Unit to to the electrical schematics at the end of this manual.
Electrical Power
Caution Connect the incubator to a grounded, dedicated circuit.
The power cord connector is the mains disconnect device for the
incubator. Position the incubator so the unit can be easily disconnected. ▲
Plug the provided power cord into the power inlet connector on the back
of the cabinet (See Figure 1-11), then into the grounded, dedicated,
electrical circuit.
Filling the Humidity Pan Fill the humidity pan with sterile pure water to within 1/2” of the top.
Place the pan directly in the center of the incubator floor to ensure best
humidity and temperature response.
Check the level and change the water frequently to avoid contamination.
Do not allow the water level to fluctuate significantly. “Dry-outs” will have
an adverse effect on the humidity level and CO2 calibration of the T/C
units.
Connecting the CO2 Gas Warning High concentrations of CO2 gas can cause asphyxiation! OSHA
Supply Standards specify that employee exposure to carbon dioxide in any eight-
hour shift of a 40-hour work week shall not exceed the eight-hour time
weighted average of 5000 PPM (0.5% CO2). The short term exposure
limit for 15 minutes or less is 30,000 ppm (3% CO2). Carbon dioxide
monitors are recommended for confined areas where concentrations of
carbon dioxide gas can accumulate. ▲
Connecting the CO2 gas The CO2 gas supply being connected should be industrial grade 99.5%
Supply (continued) pure and should not contain siphon tubes. Install a two-stage pressure
regulator at the cylinder outlet. The high pressure gauge at the tank should
have 0-2000 psig range. The low pressure gauge, at the incubator inlet,
should have a 0-30 psig range. Input pressure to the incubator must be
maintained at 15 psig (103.4 kPa).
The incubator has serrated fittings on the back of the cabinet to connect
the gas supply. Refer to Figure 1-11. The fitting is labeled CO2 Inlet #1
Tank. Make sure that the connections are secured with clamps. Check all
fittings for leaks.
For units having the CO2 Gas Guard option, refer to Section 6.
Incubator Start-Up With the incubator properly installed and connected to power, the
humidity pan filled, and the unit connected to gas supplies, system
setpoints can be entered. The following setpoints can be entered in Set
mode: Temperature, Overtemperature and CO2. To enter Set mode, press
the Mode key until the Set indicator lights. Press the right and/or left
arrow keys until the proper parameter appears in the message center. See
Chart 1-1 for more detail.
Setting the Operating Model 310 Series incubators have an operating temperature range of 10°C
Temperature to 50°C, depending on ambient temperature. The incubator is shipped
from the factory with a temperature setpoint of 10°C. At this setting, all
heaters are turned off. To change the operating temperature setpoint:
2. Press the right arrow until “Temp XX.X” is displayed in the message
center.
3. Press the up/down arrow key until the desired temperature setpoint is
displayed.
5. Press the Mode key until the Run indicator lights for Run mode or
press the right/left arrow keys to go to next/previous parameter.
Setting the Overtemp Caution The independent overtemp system is designed as a safety to
Setpoint protect the incubator only. It is not intended to protect or limit the
maximum temperature of the cell cultures or customer’s equipment inside
the incubator if an overtemp condition occurs. ▲
2. Press the right arrow until Otemp XX.X is displayed in the message
center
3. Press the up or down arrow key until the desired Overtemp setpoint is
displayed
5. Press the Mode key until the Run indicator lights or press the right or
left arrow to go to the next or previous parameter.
Setting the CO2 Setpoint All T/C CO2 cells are calibrated at the factory at 37°C, high humidity, and
10% CO2. Therefore, if a temperature setpoint of 37°C has been entered,
the humidity pan has been filled and the CO2 control is to run between 0-
10% with a T/C CO2 sensor, the CO2 setpoint may be entered
immediately. Otherwise, it is important to allow the unit 12 hours to
stabilize at the temperature setpoint before entering the CO2 setpoint.
All models of the incubator have a CO2 setpoint range of 0.0% to 20.0%.
The incubator is shipped from the factory with a CO2 setpoint of 0.0%.
At this setting, all CO2 control and alarms are turned off. To change the
CO2 setpoint:
2. Press the right arrow until “CO2 XX.X” is displayed in the message
center.
3. Press the up/down arrows until the desired CO2 setpoint is displayed.
5. Press the Mode key until the Run indicator lights to go to Run mode
or press the right/left arrow keys to go to next/previous parameter.
Calibrating the Before calibration, allow the cabinet temperature to stabilize. Place the
calibrated instrument in the center of the chamber. The instrument
Temperature should be in the airflow, not against the shelf.
Calibrating Temp 1. Press the Mode key until Cal indicator lights.
(continued) 2. Press the right arrow until “TEMPCAL XX.X” appears in the message
center.
Calibrating the T/C Models 310, 311, 350 and 351 have a thermal conductivity (T/C) CO2
sensor. Thermal conductivity of the incubator atmosphere is not only
CO2 System effected by the quantity of CO2 present, but also by the air temperature
and water vapor present in the incubator atmosphere. In monitoring the
effects of CO2, air temperature and absolute humidity must be held
constant so any change in thermal conductivity is caused by a change in
CO2 concentration.
Start -Up - The CO2 sensor has been calibrated at the factory for 37°C
and elevated humidity. Allow the temperature, humidity, and CO2 levels in
the chamber to stabilize at least 12 hours before checking the CO2
concentration with an independent instrument.
Presently Operating - Make sure the chamber doors are closed. Allow at
least 2 hours after the temperature and CO2 displays reach their setpoints
for chamber atmosphere stabilization.
Calibrating T/C CO2 1. Make sure the stabilization periods outlined above are followed.
System (continued) 2. Sample the chamber atmosphere through the sample port with an
independent instrument. Sample the atmosphere at least 3 times to
ensure accuracy of the instrument.
4. Press the right arrow until “CO2 CAL XX.X” is displayed in the
message center.
7. Press the Mode key to return to Run or the right or left arrows to go to
the next/ previous parameter.
Calibrating the Models 320, 321, 360 and 361 have an infrared (IR) CO2 sensor. Infrared
Infrared CO2 System CO2 sensors are not effected by chamber atmosphere temperature or
humidity. However, the light detector in the sensor is effected by wide
temperature changes. Therefore, changing temperature setpoints could
necessitate a recalibration of the CO2. Chamber temperature should be
allowed to stabilize before checking CO2 concentrations with an
independent instrument, especially on start-up.
Start-Up- Allow the temperature and the CO2 of the cabinet to stabilize at
least 12 hours before proceeding.
Calibrating the IR CO2 1. Measure the CO2 concentration in the chamber through the gas sample
port with a FYRITE or other independent instrument. Several readings
System (continued) should be taken to ensure accuracy.
3. Press the right arrow until “IR CAL XX.X” appears in the message
center.
6. After Enter is pressed, the unit will go into a calibration cycle that will
last 5 to 6 minutes. The control panel is locked during this cycle.
Calibrating Relative All 310 Series incubators can be equipped with an optional direct readout
relative humidity (RH) sensor. This is a readout only of the chamber
Humidity relative humidity level. It does not provide any control of the relative
humidity in the cabinet.
Start-Up - Allow 12 hours for the relative humidity and temperature in the
chamber to stabilize before proceeding.
3. Press the right arrow key until “RH CAL XX” appears in the message
center.
2. With a full humidity pan and stable temperature, the relative humidity
in the chamber will be 95%.
3. Using Step 3-5 of the relative humidity sensor adjustment, adjust the
display to 95%.
Turn the Audible The audible alarm can be turned on or off. The factory setting is ON.
Alarm ON/OFF 1. Press the Mode key until the Config indicator lights.
2. Press the right arrow until Audible XXX is displayed in the message
center.
Set an Access Code A 3-digit Access Code can be entered to avoid unauthorized personnel
from changing the setpoints, calibration, or configuration. A setting of
000 will bypass the access code. The factory setting is 000.
2. Press the right arrow until Acc Code XXX is displayed in the message
center.
Set a Low Temp The low temp alarm limit (tracking alarm) is the deviation from the
temperature setpoint which will cause a low temp alarm. The low temp
Alarm Limit alarm is variable from 0.5° below setpoint to 5° below setpoint. The
factory setting is 1° below setpoint. A minus sign (-) in the display
indicates that the alarm setting is below the setpoint.
2. Press the right arrow until Temp Lo -X.X is displayed in the message
center.
Enable the Low Temp The low temperature alarm can be programmed to trip the remote alarm
Alarm to Trip Contacts contacts. A setting of ON allows it, a setting of OFF will not allow the
temperature alarm to trip the contacts. The factory setting is ON.
5. Press the Mode key to return to Run or the right/left arrow key to go
to next/previous parameter.
Set a Low CO2 Alarm The low CO2 alarm limit (tracking alarm) is the deviation from the CO2
setpoint which will cause a low CO2 alarm. The setpoint is variable from
Limit 0.5% CO2 below setpoint to 5.0% CO2 below setpoint. The factory
setting is 1.0% CO2 below setpoint. A minus sign (-) in the display
indicates that the alarm setting is below the setpoint.
2. Press the right arrow until CO2 Lo -X.X is displayed in the message
center.
Setting a High CO2 The high CO2 alarm limit (tracking alarm) is the deviation from the CO2
setpoint which will cause a high CO2 alarm. The setpoint is variable from
Alarm Limit 0.5% CO2 above setpoint to 5.0% CO2 above setpoint. The factory
setting is 1.0% CO2 above setpoint.
2. Press the right arrow until CO2 Hi X.X is displayed in the message
center.
Enable CO2 Alarms to High and Low CO2 alarms can be programmed to trip the remote alarm
contacts. A setting of ON allows it, a setting of OFF will not allow the
Trip Contacts CO2 alarms to trip the contacts. The factory setting is ON.
2. Press the right arrow until CO2 Rly XXX is displayed in the message
center.
Set New Zero Number If a new T/C CO2 sensor is being installed, the two numbers on the
factory installed sticker on the T/C cell must be entered to calibrate the
for T/C CO2 Sensors CO2 in the unit.
Note For the technician’s convenience, a label containing the two numbers
on the T/C cell is affixed inside the electronics drawer. ▲
2. Press the right arrow until T/CZ# XXXX is displayed in the message
center.
3. Press up/down arrow to change the zero number to match the sticker.
Set New Span Number If a new T/C CO2 sensor is being installed, the two numbers on the
factory installed sticker on the T/C cell must be entered to calibrate the
for T/C CO2 Sensors CO2 in the unit.
Note For the technician’s convenience, a label containing the two numbers
on the T/C cell is affixed inside the electronics drawer. ▲
2. Press the right arrow until T/CS# XXXX is displayed in the message
center.
3. Press up/down arrow to change the span number to match the sticker.
Set a Low RH Alarm On units that have the RH option installed, a low RH alarm limit may be
Limit entered. The low RH alarm limit is the %RH in the cabinet which will
cause a low RH alarm. The setpoint is variable from setpoint 0 to 90%
RH The factory setting is 0% RH which disables the alarm.
Enable RH Alarms to The low RH alarm can be programmed to trip the remote alarm contacts.
A setting of ON will cause this, a setting of OFF will not allow the RH
Trip Contacts alarm to trip the contacts. The factory setting is ON.
2. Press the right arrow until RH Rly XXX is displayed in the message
center.
Enable Temp/RH to be On units that are equipped with the RH option, the upper seven segment
display on the control panel can be configured to display Temp
Displayed continuously, RH continuously, or toggle between Temp and RH. If the
units does not have RH, the upper display will always display temperature.
If temperature is set to ON and the RH is set OFF, temperature will be
displayed continuously. If temperature is set to OFF and RH is set to ON,
RH will be displayed continuously. If both are turned ON, the display
will toggle between the two. The factory setting will default to toggle
mode if the RH option is present.
2. Press the right arrow until Disp Tmp XXX or Disp RH XXX is
displayed in the message center.
Select Primary Tank On units equipped with the Gas Guard option, a primary tank can be
selected. The primary tank will be either Tank 1 or 2. The factory setting
w/ Gas Guard Option is Tank1.
2. Press the right arrow until Tnk Sel X is displayed in the message center.
Disable the Gas On units equipped with the Gas Guard option, the Gas Guard system may
Guard System be turned ON, or OFF if not in use. The factory setting is ON.
2. Press the right arrow until Gas Grd XX is displayed in the message
center.
Communications On units that have the RS485 option, direct communication with the
Model 1535 alarm system can be established. Each piece of equipment
Address for RS485 connected to the 1535 must have a unique address. An address of 0-24 can
be entered for the incubator. A setting of 0 is an invalid address that the
1535 will ignore. The factory setting for the RS485 address is 0.
When multiple alarm conditions occur, active messages are displayed in the
message center one at a time, updating at 5 second intervals. Pressing
Silence during multiple alarms causes all active alarms to be silenced and to
ring back in 15 minutes.
The temperature alarms are disabled when the Temp set point is 10°C. The
CO2 alarms are disabled when the CO2 set point is 0.0%.
Temp Control Failure In addition to other safety features designed into Model 310 Series
incubators, a thermostat is also provided to monitor the cabinet’s
Alarm TMP CNTR ERR temperature. In the unlikely event of a temperature control failure, the
thermostat will turn off all heaters at a cabinet temperature of 65°C,
±10%. This is intended to be a safety feature to protect the incubator, and
is not intended to protect the cell cultures or the equipment inside the
chamber should a temperature control failure occur.
Sensor Fault Alarms The microprocessor in Model 310 Series incubators continually scans all
available sensors to ensure that they are operating properly. Should an error
be detected, the incubator sounds an alarm and displays the appropriate
message. If such an alarm occurs, contact your local distributor or the
Technical Services department at 740-373-4763 or 1-800-438-4851 (USA
and Canada) or fax 740-373-4189.
CO2 SNSR ERR If the cables or connectors between the main microprocessor board and the
CO2 sensor or between the CO2 board and the sensor head on the I/R
CO2 units become loose or disconnected, the CO2 SENS ERR alarm
occurs.
I/R Units
On I/R incubators, the red light on the I/R module (see Figure 5-8 for
module circuit board location) will be lit continuously if communication is
lost between the CO2 board and the sensor head. The CO2 display will also
be locked at 00.0 without injection. Turning the incubator Off and On
will not clear the alarm. Only proper connection of all components will
correct the alarm.
IR AUTOZ ERR On incubators equipped with I/R CO2 control, calibration is done
automatically using an auto zero system. Auto zero occurs once every 24
hours. If, during the auto zero cycle, a CO2 correction of more than 0.45%
is detected, the IR AUTOZ ERR alarm occurs.
(continued) • Auto Zero pump, orifice, filter or tubing will not let air to the sensor.
Possibilities are:
Defective or electrically disconnected air pump
Kinked auto zero vinyl tubing
Disconnected tubing between the air pump and the sensor
Plugged filter or orifice on the auto zero assembly
Defective auto zero circuit
• High CO2 sensor calibration drift. This will require replacement of the
sensor.
• Calibration of the CO2 system to less than 3% actual CO2. If this has
occurred, contact the Technical Services Department.
Your equipment has been thoroughly tested and calibrated before shipment. Regular preventive maintenance is important to keep your unit
functioning properly. The operator should perform routine cleaning and maintenance on a regular basis. For maximum performance and
efficiency, it is recommended the unit be checked and calibrated periodically by a qualified service technician.
The following is a condensed list of preventive maintenance requirements. See the specified section of the operating manual for further
details.
Thermo Electron Corporation has qualified service technicians, using NIST traceable instruments, available in many areas. For more
information on Preventive Maintenance or Extended Warranties, please contact us at the number listed below.
Cleaning and calibration adjustment intervals are dependent upon use, environmental conditions and accuracy required.
Thermo Scientific
Thermo Scientific
Preventive Maintenance for Direct Heat Incubators
Check water level in the humidity pan, ½” from top. See “Filling the 9
Humidity Pan”.
Change HEPA and gas filters, if applicable (or as needed). See “Installing 9
Air Sample Filter”, “Installing HEPA Filter”.
Warning If the unit has been in service, turn it off and disconnect the
power cord connector before proceeding with any maintenance. ▲
Disinfecting the Caution Before using any cleaning or decontamination method except
Incubator Interior those recommended by the manufacturer, users must check with the
manufacturer that the proposed method will not damage the equipment. ▲
2. Set the HEPA filter aside. Wash the shelves, ducts, wing nuts and
stopper with disinfectant. Optional: The shelves, ducts and wing nuts
may be autoclaved.
3. Remove the blower scroll held on by four screws. Remove the blower
wheel and the access port filter. If a new blower wheel and scroll are
going to be used, discard the old ones. If the old ones are being
reused, wash and rinse all parts with disinfectant.
5. Wash the inner door gasket with disinfectant. The inner door gasket
may be removed and/or replaced.
6. Wash the cabinet interior with disinfectant, starting at the top and
working down. Refer to the disinfectant directions for length of time
needed before rinsing. Wash the inner door both inside and out. The
cabinet and door must be rinsed with sterile water until the
disinfectant has been removed. After the cabinet has been rinsed, spray
with 70% alcohol.
7. If the CO2 sensor gasket was removed, spray it with 70% alcohol and
install it. Install the sample filter.
8. Install the top duct, left and right ducts, inner door gasket, and access
port stopper spraying each with 70% alcohol.
Cleaning the Cabinet Clean the incubator exterior with a damp sponge or soft, well-wrung cloth
Exterior and mild detergent dissolved in water. Dry with a soft cloth.
Cleaning the Glass The chamber glass door and the optional independent inner glass doors
Doors may be cleaned using the same disinfectant used on the incubator interior.
It is imperative that they be rinsed with sterile distilled water to remove the
disinfectant residue. The doors should then be dried with a soft cloth.
Cleaning the Humidity Clean the humidity pan with soap and water and a general use laboratory
disinfectant. Rinse with sterile water and spray with 70% alcohol. The
Pan humidity pan may be autoclaved.
Reversing the Door For side-by-side operation or changing lab layouts, the inner and outer
doors are field reversible. The procedure is written from the prospective of
Swing changing the door swing from a left-hand to a right-hand swing. All screw
holes are pre-drilled for reversing the door. The tools required are a Phillips
screwdriver, a 1/8” Allen wrench and a flatblade screwdriver.
Reversing the Door 1. Remove the kick plate at the bottom of the cabinet by removing the
two Phillips screws indicated by the black arrows in the illustration.
Swing (continued) Disconnect the cable inside the cabinet.
Note The hinge pins are not connected to the brackets and may fall out
when the door is lifted off. ▲
4. Remove the door hinges, rotate them 180° and secure them on the left
side of the door. Install the nylon screws into the newly exposed holes.
5. Remove the two plastic clips securing the door heater cable. These clips
will be rotated 180° when installed on the left lower edge of the door
frame.
6. Rotate the kick plate and attach the door heater cable with the clips as
shown in Figure 5-3. The insert in the illustration shows the proper
positioning of the clips.
8. Refer to Figure 5-5 for the next steps. Remove the nylon screws along
the right side of the cabinet. Remove the outer door hinges. Rotate
them 180° and install them on the right side of the cabinet. Install the
nylon screws in the holes vacated by the hinges.
9. With a 1/8” Allen wrench, remove the door open switch target plate
on the top of the inner door.
11. Hold the door securely and remove the upper hinge bracket on the
cabinet. This hinge bracket will be rotated 180° and be placed in the
lower right corner of the cabinet where indicated.
12. Lift the glass door from the lower hinge and set it aside. Then remove
the lower hinge from the cabinet. This hinge bracket will be rotated
180° and be placed in the upper right corner of the cabinet at where
indicated.
13. Remove the screws and washers from the new striker plate location.
These screws and washers can be threaded into the holes vacated by the
striker plate where indicated.
14. Rotate the striker plate 180° from its original position and attach it to
the cabinet where indicated.
15. Using a flatblade screwdriver, remove the threaded nylon plugs from
the new hinge bracket locations. Install the lower hinge bracket where
indicated.
16. Rotate the door 180° and place it in the lower hinge bracket. Install
the upper hinge bracket where indicated.
17. Screw the threaded plugs into the holes vacated by the door hinge
brackets where indicated in Figure 5-5.
18. Install the door open switch target plate. Refer to Figure 5-6 and 5-7.
The plate must be positioned exactly 12.5 inches from the outside
edge of the cabinet to the leading edge of the plate. Do not
overtighten.
19. Connect the door heater cable and install the kick plate with the door
heater cable on the right side.
20. Installing the outer door onto the hinge pins completes the door
reversal procedure.
HEPA Filter (Factory Replace the HEPA filter routinely on an annual basis, or more often if it
changes color. The filter should also be changed when the chamber is
Installed Option) cleaned.
Replacing the Power There are only two replaceable fuses in the incubator. To access the fuses:
Fuses 1. Turn off the incubator’s power switch and unplug the power cord.
6. Slide the control panel back in, replace the screws and return the unit
to service. If the fuse(s) blow after restoring power to the incubator,
contact Technical Services Department (see Page iv near the front of
this manual).
Major Components Expanded descriptions for some of the components indicated in Figure 5-8
follow.
Power Switch - Located beneath the control panel. The outer door must
be open to access the switch.
RS 485 and Remote Alarm Contacts - Connectors and auxiliary wire
through port
CO2 manifold and supply tank connectors - Part of the Gas Guard sys-
tem, a factory installed option.
Humidity sensor (optional)
Analog output board - 4-20 mA, 0-1 Volt, 0-5 Volt
Electronics Section 1. Turn off the incubator’s power switch and unplug the power cord.
2. Remove the two screws from the front of the cabinet top and the two
screws in the uppermost corners of the back panel.
3. Slide the top about an inch to the rear and lift it off.
Note There are small locking tabs in the rear which will be bent if the top
is lifted off without sliding it first to the rear. ▲
Replacing the Sample 1. Connect one end of the sample port filter to the ceiling port.
Air Filter 2. Connect the other end of the filter to the back of the blower fan inlet.
Connect Remote A set of relay contacts is provided to monitor alarms through an RJ11
telephone style connector on the back of the cabinet. Refer to Figure 6-3
Alarm Contacts for the location of the alarm connector. The 12-foot telephone cord (P/N
190388) and RJ11-to-screw terminal conversion box (190392) are
available through our service department.
Connect RS485 All incubator models can be purchased with the RS485 communications
Interface option (P/N 190523). This option allows the incubator to be directly
connected to a Model 1535 alarm system without the use of a
communications module. A junction box is provided with each RS485
option. See Figure 6-2 for wiring details. Figure 6-3 shows the RS485
connector location on the back of the incubator cabinet.
Connecting the Warning The electronics section contains hazardous voltages. Only
Analog Output Boards qualified personnel should access this area. ▲
The analog output board option (P/N 190512, 190543, 190544) allows
the incubator to output analog signals representing the air temperature of
the cabinet, CO2 content, and relative humidity, depending upon which
systems are in the incubator. There are three different analog output board
options available : 0-1V, 0-5V, or 4-20mA signals. Negative display
readings will output 0V. The outputs do not have isolated grounds. Refer
to Table 6-1 for output specifications of the three boards.
For the 0-1V and 0-5V boards, the recording device must supply a load
>/= 1000 ohm. For the 4-20mA board, the recording device must supply a
load of </= 100 ohm.
1. Turn off the incubator and unplug it from the wall outlet.
3. Locate the Analog Output board, Figure 6-4 and Figure 5-8.
* Alternate access to the electronics is to open the outer door and remove the two
screws in the upper corners of the cabinet just under the display top. Lifting up
the top section, the electronics drawer can be pulled out to the limit of the
internal wiring.
7. Replace the screws removed earlier and return the unit to service.
CO2 Gas Guard Warning CO2 can be lethal in high concentrations. Refer to warnings in
Section 1 of this manual. ▲
The Model 310 Series incubators can be equipped with a built-in Gas
Guard system (P/N 190640) that will operate with a CO2 gas supply. The
Gas Guard uses two pressure switches to continuously monitor the
pressures of two independent CO2 supplies and automatically switches
from one supply to the other when the supply pressure drops below 10
psig (0.690 bar). The Gas Guard’s design does not facilitate use by
multiple incubators.
The CO2 gas supplies must be equipped with two-stage pressure regulators
with gauges. The high pressure gauge at the tank should have a 0-2000
psig range and the low pressure gauge should have a 0-30 psig range. The
gas supply to the incubator must be maintained at 15 psig (1.034 bar), ±5
psig. Gas pressures below 15 psig will cause nuisance alarms to occur on
incubators equipped with the built-in Gas Guard.
Connecting the CO2 Gas The CO2 inlets for the Gas Guard are located on the rear of the cabinet.
Supplies Using 1/4” ID tubing, connect one of the CO2 supply tanks to the fitting
labeled CO2 Inlet #1 Tank. Connect the second CO2 supply tank to the
fitting labeled CO2 Inlet #2 Tank. Install 3/8” hose clamps to secure the
1/4” ID tubing to the fittings on the rear of the drawer. (Refer to Figure 6-
6.)
De-activating the Gas The built-in Gas Guard is turned ON when shipped from the factory. In
Guard addition, the Tank Sel for the Gas Guard is specified as Tank 1 when
shipped. Refer to Section 3, Configuration, to de-activate the Gas Guard
or change the Tank Sel from #1 to #2. If the Gas Guard system is not
used, the incubator will function normally by supplying CO2 from the
supply tank connected to Inlet #1.
Operation of the CO2 Gas With the Gas Guard in operation, the incubator will use the gas supplied
by the tank connected to Inlet #1 until the pressure drops below 10 psig
Guard
(0.690 bars). At this time, the Gas Guard automatically switches to gas
supplied through CO2 Inlet #2.
Caution Both the audible and visual alarms described above do not ring
back once the Silence key is pressed. ▲
If the Gas Guard system does not detect an adequate gas supply at CO2
Inlet #1 or Inlet #2, a visual and audible alarm will again occur on the
control panel. The visual alarm in the Message Center will read Tank 1&2
Low. The audible alarm will continue to ring until the Silence key is
pressed. The audible alarm will ring back every 15 minutes after the alarm
is silenced if the Gas Guard continues to detect that both gas supply
pressures are below 10 psig (0.690 bars).
Humidity Readout The 310 Series incubators can be equipped with a humidity sensor (P/N
190643) to monitor the relative humidity inside the chamber. The sensor
is mounted to the top air duct and provides a signal that is displayed in
1% increments on the control panel. The humidity readout can be
displayed continuously or toggles with the temperature readout. In
addition, a low alarm limit can be set on the humidity readout, which will
detect when the humidity pan runs dry. Refer to Section 3,
Configuration.
Humidity Readout The sensor is capable of measuring relative humidity from 10% to 100%
Accuracy with an accuracy of ±5% above 90%. See Section 2, Calibration, for
details on calibrating the humidity readout.
Uninterruptible Power Warning The Uninterruptible Power Supply is not rated for life-support
Supply Connections applications. ▲
The remote alarm contacts will not trip during a power failure while
connected to the UPS until battery power is exhausted. There is however,
an audible and visual alarm when the unit switches to battery operation.
The UPS batteries are fully charged when the unit has been connected to
utility power for a minimum of four hours. The UPS power switch does
not have to be on for the batteries to charge. When battery replacement
becomes necessary, refer to the UPS user’s manual.
Caution Use only factory recommended UPS units which produce a pure
sine wave output. Other UPS units could damage the incubator and void
the warranty. Typical facility back-up systems will not effect the efficiency
of the incubator. ▲
Connecting UPS 1. Plug the incubator line cord into one of the receptacles on the back of
P/N 270078, 115V, 50/60Hz the UPS. Then plug the UPS line cord into a suitable wall outlet.
Connecting UPS Use the line cord supplied with the incubator to connect the UPS to a
P/N 270082, 230V, 50/60Hz suitable wall outlet, and the UPS coupler to connect the incubator to the
UPS.
Temperature
Control . . . .±0.1°C Microprocessor PID Control
Setpoint . . . . . . . . . . .Digital - Touch pad, 0.1°C
Range . . . . . . . . . . . . .+5°C above ambient to 50°C
Uniformity . . . . . . . . . . . . . . . . . .±0.2°C @ +37°C
Tracking alarm . . User programmable (low) indicator
Overtemp . .Tracking, user programmable, action, and indicator
Display . . . . . . . . .Digital, LED, 0.1°C increments
Temperature Safety
Type . . Extreme temperature safety, action, and indicator
Sensor . . Thermostat, independent of temperature control system
Indicator . . Message center, audible and visual alarms
Relative Humidity
Control . . . . .Humidity pan - natural vaporization
Humidity with pan . . . . . . . . . . .95% RH at 37°C
Display . . . . . . . . . . . . .Optional in 1% increments
Alarm . . . . . . .Low RH with optional RH monitor
CO2
Control . . . . .±0.1% microprocessor PID control
Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . .T/C or IR
Readability . . . . . . . . . . . . . . . . . . . . . . . . . . .0.1%
Range . . . . . . . . . . . . . . . . . . . . . . . . . . . .0 to 20%
Inlet pressure . . . . . . . . . . . . . . . . .15 psig, ±5 psig
Display . . . . . . . . . .Digital LED, 0.1% increments
Shelves
Dimensions . . . . . . . .18.5” x 18.5” (47cm x 47cm
Construction . . Stainless steel (belt sanded, both sides)
Surface area . . . . . .2.4 sq. ft. (0.22 sq. m) per shelf
Max. per chamber . . . . . . . . .38.4 sq ft (3.6 sq. m)
Loading . . . .35 lbs (16kg) slide in and out, 50 lbs (23kg) stationary
Standard . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4
Maximum . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .16
Construction
Interior volume . . . . . . . . . . . . . . . . . . . .6.5 cu. ft.
Interior . . . . . .Type 304 stainless steel shiny finish
Exterior . . . . . . . . . . . . . . . .18 gauge cold roll steel
Outer door gasket . . Four-sided molded, magnetic vinyl
Inner door gasket . . . . . . .Feather-edge, removable
Insulation . . . . . . . . . . . . . . . . . . . . . . . . .Fiberglass
Fittings
Access port . . . .1-1/4 inch removable silicone plug
CO2 inlet . . . . . . . . . . . . . . . . . . . . . . .1/4” barbed
Sample port . . . . . . . . . . . . .Front mounted barbed
Electrical
115 Volt models -
90-125VAC, 50/60 Hz, 1PH, 2.4 FLA
230 Volt models -
180-250, 50/60 Hz, 1 PH, 2.0 FLA
Power switch/circuit breaker . . . . .2 Pole, 6.0 Amp
Accessory outlet . . Voltage equal to the cabinet input,
75 Watts maximum, 0.5ma leakage current
Alarm contacts . . Deviation of temperature, CO2, power, NO and NC
Dimensions
Exterior . . . . . . . . .26.0” W x 38.5” H x 25.0” F-B
Interior . . . . . . . . .21.4” W x 26.8” H x 20.0” F-B
Weight
Net Weight Stainless Steel . . . . . . . . . . . . .205 lbs
Net Weight Copper . . . . . . . . . . . . . . . . . .230 lbs.
Options/Accessories
HEPA filter . . Factory installed: Stock no. 190858
Battery back-up . . Free Standing: Stock no. 270078, 120VAC, 60Hz
Stock no. 270082, 220VAC, 50Hz
Humidity . . Factory installed: Stock no. 190463 - provides humidity
readout and low alarm
Inner door kit . . Customer installed: Stock no. 190650 - independent
eight separate glass doors
Built-in Gas Guard . . Factory installed: Stock no. 190640 - CO2 only,
automatically switches supply tanks
Recorders . . Customer installed: Standard recorders used on water-
jacket incubators
Decontamination kits
HEPA filter, blower assembly, inner door gasket, etc. -
Stock no. 190868 (with HEPA filter),
Stock no. 190869 (without HEPA filter)
Copper Options
Interior ductwork .Solid copper: Stock no. 190725
Perforated shelfCustomer Installed: Stock no. 224166 - solid copper
Humidity PanCustomer Installed: Stock no. 237015 - solid copper
Copper interior kitFactory installed: Replace stainless steel duct work,
shelves, and
humidity pan with solid copper Stock no. 190870
Certifications
Refer to the Declarations of Conformity at the back of this manual
Safety Specifications
Indoor Use Only
Altitude - 2,000 meters
Temperature - 5°C to 40°C
Humidity - 80% RH at or below 31°C, decreasing linearly to 50% RH
at 40°C
Mains Supply Fluctuations - Mains supply voltage fluctuations not to
exceed ±10% of the nominal voltage
Installation Category II 1 . . . . . . . . . . . . . . . . . . . . .
Pollution Degree 2 2
Class of Equipment I
1
Installation Category (overvoltage category) defines the level of transient
overvoltage which the instrument is designed to withstand safely. It
depends on the nature of the electricity supply and its overvoltage pro-
tection means. For example, in CAT II which is the category used for
instruments in installations supplied from a supply comparable to public
mains such as hospital and research laboratories and most industrial lab-
oratories, the expected transient overvoltage is 2500V for a 230V supply
and 1500V for a 120V supply.
2
Pollution Degree describes the amount of conductive pollution present
in the operating environment. Pollution Degree 2 assumes that normally
only non-conductive pollution such as dust occurs with the exception of
occasional conductivity caused by condensation.
BILL OF MATERIALS
12 ITEM PART NO. PART DESCRIPTION QTY
NO.
1 23052 #10-32 S.S. WING NUT 1
2 22002 #4-40 X 1/2 SS PHP SCREW 4
3 640047 #8-32 x .750" LONG STN STL SELF-CLINCHING STUD 5
4 22142 #8-32 x 3/4 FHP SCREW 4
5 23002 #8-32 ZP LKWASH HEX NUT 4
6 730068 .310 ID TEFLON FLAT WASHER 1
6 7 100113 3.500 X 1.500 BLOWER WHEEL POLY. 1
8 190790-31-1310
310 INCUBATOR BLOWER SUPPORT 1
15 9 1900252 BLOWER MOTOR SHIELD & GUARD 1
10 190846 BLOWER SCROLL -RESTRICTED, 310/3110 1
11 190831 BLOWER SCROLL- NO HEPA, 310 1
12 156126 MOTOR 1
14 13 190832 MOTOR SPACER 1
14 730034 NYLON WASHER 1.500"OD X .500"ID 1
15 730069 SILICONE WASHER 1
5
4
8 13
11
10
During the first year, component parts proven to be non-conforming in materials or workmanship will be repaired or replaced at Thermo's expense,
including labor. In addition, after the initial warranty period of one year the CO2 sensor (IR) will be warranted for a second year, parts only. The CO2
sensor (T/C) will be warranted for four additional years for a total of five years, parts only. Installation and calibration is not covered by this warran-
ty agreement. The Technical Services Department must be contacted for warranty determination and direction prior to performance of any repairs.
Expendable items, glass, filters and gaskets are excluded from this warranty.
Replacement or repair of components parts or equipment under this warranty shall not extend the warranty to either the equipment or to the com-
ponent part beyond the original warranty period. The Technical Services Department must give prior approval for return of any components or
equipment. At Thermo's option, all non-conforming parts must be returned to Thermo postage paid and replacement parts are shipped FOB destina-
tion.
THIS WARRANTY IS EXCLUSIVE AND IN LIEU OF ALL OTHER WARRANTIES, WHETHER WRITTEN, ORAL OR IMPLIED. NO
WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE SHALL APPLY. Thermo shall not be liable for
any indirect or consequential damages including, without limitation, damages relating to lost profits or loss of products.
Your local Thermo Sales Office is ready to help with comprehensive site preparation information before your equipment arrives. Printed instruction
manuals carefully detail equipment installation, operation and preventative maintenance.
If equipment service is required, please call your Technical Service Department at l-800-438-4851 (USA and Canada) or 1-740-373-4763. We're
ready to answer your questions on equipment warranty, operation, maintenance, service and special applications. Outside the USA, contact your local
distributor for warranty information.
ISO
9001
10-1
Warranty Information
Section 10
10-2
Section 10
Warranty Information
During the first year component parts proven to be non-conforming in materials or workmanship will be repaired or replaced at Thermo's expense,
excepting labor. In addition, after the initial warranty period of one year the CO2 sensor (IR) will be warranted for a second year, parts only. The
CO2 sensor (T/C) will be warranted for four additional years for a total of five years, parts only. Installation and calibration is not covered by this
warranty agreement. The Technical Services Department must be contacted for warranty determination and direction prior to performance of any
repairs. Expendable items, glass, filters and gaskets are excluded from this warranty.
Replacement or repair of components parts or equipment under this warranty shall not extend the warranty to either the equipment or to the com-
ponent part beyond the original warranty period. The Technical Services Department must give prior approval for return of any components or
equipment. At Thermo's option, all non-conforming parts must be returned to Thermo postage paid and replacement parts are shipped FOB destina-
tion.
THIS WARRANTY IS EXCLUSIVE AND IN LIEU OF ALL OTHER WARRANTIES, WHETHER WRITTEN, ORAL OR IMPLIED. NO
WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE SHALL APPLY. Thermo shall not be liable for
any indirect or consequential damages including, without limitation, damages relating to lost profits or loss of products.
Your local Thermo Sales Office is ready to help with comprehensive site preparation information before your equipment arrives. Printed instruction
manuals carefully detail equipment installation, operation and preventive maintenance.
If equipment service is required, please call your Technical Services Department at 1-800-438-4851 (USA and Canada) or 1-740-373-4763. We're
ready to answer your questions on equipment warranty, operation, maintenance, service and special application. Outside the USA, contact your local
distributor for warranty information.
ISO
9001
Rev. 4 4/09 REGISTERED
Thermo Scientific
Thermo Scientific Model 310 Series DH Incubator 11-1
11-2 Model 310 Series DH Incubator Thermo Scientific
Thermo Fisher Scientific
Controlled Environment Equipment
401 Millcreek Road
Marietta, Ohio 45750
United States
www.thermo.com