USB2000 Fiber Optic Spectrometer Operating Instructions: Ocean Optics, Inc
USB2000 Fiber Optic Spectrometer Operating Instructions: Ocean Optics, Inc
05
11062003
USB2000
Fiber Optic Spectrometer
Operating Instructions
Trademarks
Microsoft, Windows, Windows 95, Windows 98, Windows ME, Windows NT, Windows 2000, Windows XP and Excel are either registered
trademarks or trademarks of Microsoft Corporation.
Limit of Liability
Every effort has been made to make this manual as complete and as accurate as possible, but no warranty or fitness is implied. The
information provided is on an “as is” basis. Ocean Optics, Inc. shall have neither liability nor responsibility to any person or entity with respect
to any loss or damages arising from the information contained in this manual.
Table of Contents
1 USB2000 Introduction .................................................................................................................. 1
Product Overview ...................................................................................................................................................1
Interface Options ....................................................................................................................................................3
Shipment Components ...........................................................................................................................................3
2 USB2000 Specifications ............................................................................................................... 5
How the USB2000 Works .......................................................................................................................................5
USB2000 Specifications .........................................................................................................................................7
3 Installing the USB2000 ............................................................................................................... 10
USB2000 Installation ............................................................................................................................................10
External Triggering Options ..................................................................................................................................13
4 USB2000 Compatible Products.................................................................................................. 14
5 Troubleshooting ......................................................................................................................... 18
USB2000 Connected to PC Prior to OOIBase32 Installation ...............................................................................18
Troubleshooting the Serial Port Configuration......................................................................................................21
Older Version of OOIBase32 Installed .................................................................................................................21
6 Sample Experiments................................................................................................................... 22
Preparing for Experiments ....................................................................................................................................22
Absorbance Experiments......................................................................................................................................24
Transmission Experiments ...................................................................................................................................26
Reflection Experiments.........................................................................................................................................28
Relative Irradiance Experiments...........................................................................................................................30
Time Acquisition Experiments ..............................................................................................................................32
Appendix A: Calibrating the Wavelength of the USB2000 .......................................................... 35
About Wavelength Calibration ..............................................................................................................................35
Calibrating the Spectrometer ................................................................................................................................35
Appendix B: USB2000-FLG Spectrometer ...................................................................................... 40
About the USB2000-FLG......................................................................................................................................40
Using the USB2000-FLG ......................................................................................................................................40
Index .....................................................................................................................................................................42
USB2000 Introduction
1 USB2000 Introduction
The following chapter contains introductory information about the USB2000 Spectrometer, shipment information,
and spectrometer connectivity.
Product Overview
The USB2000 Spectrometer connects to a notebook or desktop PC via USB port or serial port. When connected
to the USB port of a PC, the USB2000 draws power from the host PC, eliminating the need for an external power
supply.
System Requirements
You can use the USB2000’s USB connectivity with any PC that meets the following requirements:
• Windows 98/Me/2000/XP operating system (or Windows CE 2.11 or later for palm-sized PCs)
• Ocean Optics OOIBase32 software application installed and configured for use with the USB2000
(OOIPS2000 software if using a palm-sized PC). Consult the “Configuring the USB2000 in OOIBase32”
section of Chapter 3: Installing the USB2000 for specific configuration instructions.
Alternately, the USB2000 has a serial port for connecting to PCs, PLCs, and other devices that support the RS-
232 communication protocol. However, this connection method requires an external power supply to power the
USB2000.
EEPROM Utilization
An EEPROM memory chip in each USB2000 contains wavelength calibration coefficients, linearity coefficients,
and a serial number unique to each individual spectrometer. The OOIBase32 software application reads these
values directly from the spectrometer, enabling the ability to “hot-swap” spectrometers between PCs without
entering the spectrometer coefficients manually on each PC.
About OOIBase32
OOIBase32 is the latest generation of operating software for all Ocean Optics spectrometers and is available free
to all customers. OOIBase32 is a user-customizable, advanced acquisition and display program that provides a
real-time interface to a variety of signal-processing functions. With OOIBase32, you have the ability to perform
spectroscopic measurements (such as absorbance, reflectance, and emission), control all system parameters,
collect and display data in real time, and perform reference monitoring and time acquisition experiments.
Note: When using a palm-sized PC, you will use the OOIPS2000 software instead of OOIBase32. The
functionality of OOIPS2000 is similar to OOIBase32, but it is specifically for the palm-sized PC.
Interface Options
The USB2000 has both USB and serial port connectors, enabling you to connect the spectrometer to a desktop or
notebook PC via a USB port or to a desktop, notebook, or to a palm-sized PC via a serial port.
Computer Operating System
Part Needed Description of Part
Interface Requirements
Desktop or
Windows 98/Me/ USB-CBL-1 Cable that connects from USB port on USB2000 to USB port on
Notebook PC
2000/XP (included) desktop or notebook PC
via USB Port
Desktop or Any 32-bit Adapter block that enables connection from serial port on
USB-ADP-PC
Notebook PC Windows operating USB2000 to serial port on desktop or notebook PC; comes with
(not included)
via Serial Port system 5 VDC power supply (required when connecting to serial port)
Adapter block that enables connection (with standard 9-pin
Palm-sized PC Windows CE 2.11 USB-ADP-H serial cable) from serial port on USB2000 to serial port on palm-
via Serial Port or higher (not included) sized PC; comes with 5 VDC power supply (required when
connecting to serial port)
Shipment Components
The following information and documentation ships with the USB2000 Spectrometer:
Packing List
The packing list is inside a plastic bag attached to the outside of the shipment box (the invoice arrives separately).
The packing slip lists all items in the order, including customized components in the spectrometer (such as the
grating, detector collection lens, and slit). The packing list also includes the shipping and billing addresses, as well
as any items on back order.
Other Documentation
You can find detailed instructions for the OOIBase32 Spectrometer Operating Software at
http://www.oceanoptics.com/technical/ooibase32bit.pdf.
Note: You can find instructions for the OOIPS2000 Operating Software for the palm-sized PC at the following
location: http://www.oceanoptics.com/products/ooips2000.asp.
To find operating instructions on some of Ocean Optics’ most popular spectroscopic accessories (including light
sources, sampling chambers, sampling optics, and probes), consult the USB2000-compatible products list in
Chapter 4. You can also access operating instructions for Ocean Optics products from the Software and
Technical Resources CD that ships with the product.
For those engineers and programmers needing the serial port commands of the USB2000, consult the USB2000
Technical Datasheet on the Ocean Optics web site. You can find engineering-level documentation at
http://OceanOptics.com/Technical/EngineeringDocs.asp.
2 USB2000 Specifications
This chapter contains information on spectrometer operation, specifications, and system compatibility. It also
includes accessory connector pinout diagrams and pin-specific information.
Note: The optical bench has no moving parts that can wear or break. Ocean Optics secures all components in
place permanently at the time of manufacture.
The USB Component Table on the following page explains the function of each numbered component in the
USB2000 Spectrometer diagram (Figure 2-1).
USB2000 Specifications
The following sections provide specification information for the CCD Detector in the USB2000, as well as the
USB2000 Spectrometer itself.
System Compatibility
The following sections provide information on hardware and software requirements for the USB2000:
10-pin Connector
10 8 6 4 2
9 7 5 3 1
If you have already connected the USB2000 to the PC prior to installing OOIBase32, consult Chapter 5:
Troubleshooting for information on correcting a corrupt USB2000 installation.
USB2000 Installation
USB Mode Serial Port Mode
To connect the USB2000 to a PC via the USB port, To use the serial port capacity of the USB2000
the PC must be running the Windows 98/ME/2000/XP Spectrometer, the PC must be running a 32-bit
operating system. version of the Windows operating system (or Windows
CE 2.11 or higher for palm-sized PCs).
Note: The USB port on a PC can power up to five
USB2000 spectrometer channels. Systems Follow the steps below to connect the USB2000 to the
with more than five channels require a PC via serial port:
powered USB hub. 1. Connect the serial cable adapter block to the
USB2000.
Follow the steps below to connect the USB2000 to a
PC via the USB port: 2. Connect one end of the 9-pin serial cable to
the adapter block on the USB2000, and then
1. Install OOIBase32 on the destination PC.
connect the other end to a serial port on the
2. Locate the USB cable (USB-CBL-1) provided PC.
with the USB2000.
3. Note the number of the serial port (COM Port)
3. Insert the square end of the cable into the to which you connected the USB2000 (some
side of the USB2000. PCs may not have numbered ports; palm-
sized PCs typically have only one serial port).
4. Insert the rectangular end of the cable into the
USB port of the PC. 4. Plug the 5 VDC external power supply into an
outlet and connect it to the USB2000.
If you installed OOIBase32 prior to connecting the
USB2000, the Add New Hardware Wizard appears
and installs the USB2000 drivers. If the drivers do not
successfully install (or if you connected the USB2000
to the PC before installing OOIBase32), consult
Chapter 5: Troubleshooting.
The following sections contain instructions on initially configuring the USB2000 the first time you start OOIBase32:
The spectrometer should now be able to acquire data and respond to light. Exit and restart OOIBase32 to save
configuration data to disk.
Note: For information on using the USB2000 with the OOIPS2000 Operating Software for the palm-sized PC,
visit http://www.oceanoptics.com/technical/palmspec.pdf.
Software
The USB2000 is compatible with all Ocean Optics software.
• OOIBase32 Spectrometer Operating Software Manual
• OOIChem Chemistry Teaching Software Manual
• OOIColor Color Measurement Software Manual
• OOIIrrad Absolute Irradiance Software Manual
• OOIPS2000 Palm-SPEC Operating Software Manual
• OOISensors Sensor Software Manual
Device Drivers
The USB2000 is compatible with all Ocean Optics device drivers.
• OOIWinIP Windows Interface Package
• OOILVD LabVIEW Software Device Driver Package
• OOIHIP Palm-sized Interface Package
• OOIHSD High-speed Driver Library
A/D Converters
The USB2000 has an A/D converter integrated with the spectrometer. Thus, it does not require an external A/D
Converter.
Light Sources
• D-2000 Deuterium Source
• DH-2000 Deuterium Tungsten Halogen Source
• HG-1 Mercury Argon Calibration Source
• LS-1 Tungsten Halogen Source
• LS-1-CAL Calibrated Source
• LS-1-CAL-INT Calibrated Source
• LS-450 Blue LED Pulsed Source
• Mini-D2T Deuterium Tungsten Source
• PX-2 Pulsed Xenon Lamp
Sampling Accessories
• 74-90-UV Right Angle Reflector
• 74-ACH Adjustable Collimating Lens Holder
• 74-OPM Optical Post Mount
• Collimating Lenses
• CUV-ALL-UV 4-way Cuvette Holder
• CUV-CCE Electrophoresis Sample Cell
• Cuvettes
• CUV-FL-DA Direct Attach Cuvette Holder
• CUV-UV, CUV-VIS Cuvette Holders
• CUV-UV-10, CUV-VIS-10 Cuvette Holders
• FHS-UV In-line Filter Holder
• FIA Flow Cells
• FOIS-1 Integrating Sphere
• FVA-UV Variable Attenuator
• ISP-REF Integrating Sphere
• LED-PS Power Supply
• LPC Long Pass Flow Cells
• NANO-1 Surface Tension Microcell
• PIP-10 SpectroPipetter Microcell
• RPH-1 Reflection Probe Holder
• STAN-ABS-UV and STAN-ABS-VIS Absorbance Standards
• WS-1 Diffuse Reflectance Standard
5 Troubleshooting
The following sections contain information on troubleshooting issues you may encounter when using the
USB2000 Spectrometer.
Note: For issues encountered when using a palm-sized PC, consult the OOIPS2000 manual.
If you connect your Ocean Optics USB or PCI device to the computer prior to installing your Ocean Optics
software application, you may encounter installation issues that you must correct before your Ocean Optics
device will operate properly.
Follow the applicable steps in this document to remove the incorrectly installed device, device driver, and
installation files.
Note: If these procedures do not correct your device driver problem, you will need to obtain the “Correcting
Device Driver Issues” document from the Ocean Optics website. Visit the following web address:
http://www.oceanoptics.com/technical/engineering/correctingdevicedriverissues.pdf
Windows 98/ME:
Windows 2000/XP:
2. Locate the Other Devices option and expand the Other Devices selection by clicking on the "+" sign to
the immediate left.
Note: Improperly installed USB devices can also appear under the Universal Serial Bus Controller
option. Be sure to check this location if you cannot locate the unknown device.
3. Locate the unknown device (marked with a large question mark). Right-click on the Unknown Device
listing and select the Uninstall or Remove option.
4. Click the OK button to continue. A warning box appears confirming the removal of the Unknown Device.
Click the OK button to confirm the device removal.
6. Locate the section in this chapter that is appropriate to your operating system and perform the steps in
the “Remove Improperly Installed Files” section.
Windows 98:
2. Navigate to the Windows | INF directory. If the INF directory is not visible, you will need to disable the
“Hide System Files and Folders” option on in Windows Folder Options.
Note: If the INF directory is not visible, you will need to disable the “Hide System Files and Folders” and
“Hide File Extensions for Known File Types” options in Windows Folder Options.
You can access Windows Folder Options from Windows Explorer, under the View | Options
menu selection.
6. Reinstall your Ocean Optics application and reboot the system when prompted.
The system will now be able to locate and install the correct drivers for the USB device.
Windows 2000/XP:
2. Navigate to the Windows | INF directory. If the INF directory is not visible, you will need to disable the
“Hide System Files and Folders” option on in Windows Folder Options.
Note: If the INF directory is not visible, you will need to disable the “Hide System Files and Folders” and
“Hide File Extensions for Known File Types” options in Windows Folder Options.
You can access Windows Folder Options from Windows Explorer, under the Tools | Folder
Options menu selection.
6. Reinstall your Ocean Optics application and reboot the system when prompted.
The system will now be able to locate and install the correct drivers for the USB device.
3. Double-click on the Ports (COM & LPT) option to display COM port numbers. Ensure that no warning icon
appears next to the USB2000’s COM port.
4. Verify that the COM port to which the USB2000 is interfaced is active. If the ports on the PC are not
labeled and you do not know the COM port number, use trial-and-error to find the correct COM port.
Open OOIBase32 and view the displayed graph. If the correct COM port is selected, you will see a
dynamic trace responding to light near the bottom of the graph. If the correct COM port is not selected,
you will see a straight line at zero counts.
5. Disable virus protection to ensure timely and complete data transfer (optional – some computers require
this step).
6 Sample Experiments
The following sections contain information on conducting sample experiments using the USB2000 and
OOIBase32. For information on experiments with OOIPS2000, consult the OOIPS2000 Operating Instructions.
Locate the Wavelength Calibration Data sheet that came with the USB2000. Select Spectrometer |
Configure from the menu and choose the Wavelength Calibration page. For each spectrometer
channel in the system, enable the channel and make sure the First Coefficient, Second Coefficient, Third
Coefficient and Intercept correspond to those of the system.
4. Adjust the acquisition parameters using the Acquisition Parameters dialog bar or select Spectrum |
Configure Data Acquisition from the menu.
If you have followed the previous steps and started OOIBase32, the spectrometer is already acquiring
data. Even with no light in the spectrometer, there should be a dynamic trace displayed in the bottom of
the graph. If you allow light into the spectrometer, the graph trace should rise with increasing light
intensity. This means the software and hardware are correctly installed.
Once you install the hardware, configure the software, and establish your sampling system, you are ready to take
measurements.
There are four basic optical measurements from which to choose:
• Absorbance
• Transmission
• Reflection
• Relative irradiance
The type of measurement you will take determines the configuration of the sampling optics for your system.
Furthermore, your choice of reference and data analysis determines how the OOIBase32 presents the results.
Note: For each measurement, you must first take a reference and dark spectrum. After you take a reference
and a dark spectrum, you can take as many measurement scans as needed. However, if you change any
sampling variable (integration time, averaging, smoothing, angle, temperature, fiber size, etc.), you must
store a new dark and reference spectrum.
Application Tips
If the signal you collect is saturating the spectrometer (intensity greater than 4000 counts), you can decrease the
light level on scale in scope mode by:
• Decreasing the integration time
• Attenuating the light going into the spectrometer
• Using a smaller diameter fiber
• Using a neutral density filter with the correct optical density
If the signal you collect has too little light, you can increase the light level on scale in scope mode by:
• Increasing the integration time
• Using a larger diameter fiber
• Removing any optical filters
Absorbance Experiments
Absorbance spectra are a measure of how much light a sample absorbs. For most samples, absorbance is
linearly related to the concentration of the substance. OOIBase32 calculates absorbance (Aλ) using the following
equation…
Sλ - Dλ
Aλ = - log10 ( R -D )
λ λ
…where Sλ is the sample intensity at wavelength λ, Dλ is the dark intensity at wavelength λ, Rλ is the reference
intensity at wavelength λ.
Figure 6-1: Typical absorbance setup. The light source (far right) sends light via an input fiber into a cuvette in a cuvette
holder (bottom center). The light interacts with the sample. The output fiber carries light from the sample to the
spectrometer (top center), which is connected to the PC (far left).
Absorbance is also proportional to the concentration of the substance interacting with the light (this is known as
Beer’s Law). Common absorption applications include the quantification of chemical concentrations in aqueous or
gaseous samples.
Follow the steps below to take an absorbance measurement using OOIBase32:
1. Place OOIBase32 in scope mode by clicking the scope mode icon on the toolbar or selecting Spectrum |
Scope Mode from the menu bar.
2. Ensure that the entire signal is on scale. The intensity of the reference signal should peak at about 3500
counts. If necessary, adjust the integration time until the intensity is approximately 3500 counts.
(Continued)
3. Place a sample of the solvent into a cuvette and take a reference spectrum. You must take a reference
spectrum before measuring absorbance.
Note: Do not put the sample itself in the path when taking a reference spectrum, only the solvent.
Click the Store Reference spectrum icon on the toolbar or select Spectrum | Store Reference from the
menu bar to store the reference. This command merely stores a reference spectrum in memory. You
must select File | Save | Reference from the menu bar to permanently save the spectrum to disk.
4. Block the light path to the spectrometer. Then, take a dark spectrum by clicking the Store Dark
Spectrum icon on the toolbar or by selecting Spectrum | Store Dark from the menu bar. This command
merely stores a dark spectrum in memory. You must select File | Save | Dark from the menu to
permanently save the spectrum to disk.
Note: If possible, do not turn off the light source when taking a dark spectrum. If you must turn off your
light source to store a dark spectrum, allow enough time for the lamp to warm up again before
continuing your experiment. After the lamp warms up again, store a new reference (Step 3).
You must take a dark spectrum before measuring absorbance.
5. Put the sample in place and ensure that the light path is clear. Then, take an absorbance measurement
by clicking on the Absorbance Mode icon on the toolbar or selecting Spectrum | Absorbance Mode
from the menu. To permanently save the spectrum to disk, click the Save icon on the toolbar or select
File | Save | Processed from the menu bar.
Note: If you change any sampling variable (integration time, averaging, smoothing, angle, temperature,
fiber size, etc.), you must store a new dark and reference spectrum.
Transmission Experiments
Transmission is the percentage of energy passing through a sample relative to the amount that passes through
the reference. Transmission Mode also displays the portion of light reflected from a sample, since transmission
and reflection measurements use the same mathematical calculations. The transmission is expressed as a
percentage (%Tλ) relative to a standard substance (such as air). OOIBase32 calculates %Tλ (or %Rλ) with the
following equation…
Sλ - Dλ
%Tλ = x 100%
Rλ - Dλ
…where Sλ is the sample intensity at wavelength λ, Dλ is the dark intensity at wavelength λ, Rλ is the reference
intensity at wavelength λ.
Figure 6-2: Typical transmission setup. The light source (far right) sends light via the input leg of a transmission probe
into a container (bottom center). The light interacts with the sample. The output leg of the transmission probe carries the
information to the spectrometer (top center), which transmits the information to the PC (far left).
Common transmission applications include measuring light through solutions, optical filters, optical coatings, and
other optical elements (such as lenses and fibers).
Perform the following steps to take a transmission measurement using OOIBase32:
1. Place OOIBase32 in scope mode by clicking the Scope Mode icon on the toolbar or by selecting
Spectrum | Scope Mode from the menu bar.
2. Ensure that the entire signal is on scale. The intensity of the reference signal should peak at about 3500
counts. If necessary, adjust the integration time until the intensity is approximately 3500 counts.
(Continued)
6. Place a sample of the solvent into a cuvette and take a reference spectrum. You must take a reference
spectrum before measuring transmission.
Note: Do not put the sample itself in the path when taking a reference spectrum, only the solvent.
Click the Store Reference spectrum icon on the toolbar or select Spectrum | Store Reference from the
menu bar to store the reference. This command merely stores a reference spectrum in memory. You
must select File | Save | Reference from the menu bar to permanently save the spectrum to disk.
3. Block the light path to the spectrometer. Then, take a dark spectrum by clicking the Store Dark
Spectrum icon on the toolbar or by selecting Spectrum | Store Dark from the menu bar. This command
merely stores a dark spectrum in memory. You must select File | Save | Dark from the menu to
permanently save the spectrum to disk.
Note: If possible, do not turn off the light source when taking a dark spectrum. If you must turn off your
light source to store a dark spectrum, allow enough time for the lamp to warm up again before
continuing your experiment.
You must take a dark spectrum before measuring transmission.
4. Put the sample in place and verify that the light path is clear. Then, take a transmission measurement by
clicking the Transmission Mode icon on the toolbar or selecting Spectrum | Transmission Mode from
the menu bar. To save the spectrum to disk, click the Save icon on the toolbar or select File | Save |
Processed from the menu bar.
Note: If you change any sampling variable (integration time, averaging, smoothing, angle, temperature,
fiber size, etc.), you must store a new dark and reference spectrum.
Reflection Experiments
Reflection is the return of radiation by a surface, without a change in wavelength. Reflection can be:
• Specular (the angle of incidence is equal to the angle of reflection)
• Diffuse (the angle of incidence is not equal to the angle of reflection)
Every surface returns both specular and diffuse reflections. Some surfaces may return mostly specular reflection,
while others may return mostly diffuse reflection. Specular reflection increases proportionately with the amount of
gloss on a surface.
Reflection is expressed as a percentage (%Rλ) relative to the reflection from a standard reference substance…
Sλ - Dλ
%Rλ = x 100%
Rλ - Dλ
…where Sλ is the sample intensity at wavelength λ, Dλ is the dark intensity at wavelength λ, Rλ is the reference
intensity at wavelength λ.
Figure 6-3: Typical reflection setup. A light source (far right) sends light via the input leg of a reflection probe onto a
sample (bottom center). A reflection probe holder holds the probe in either a 90 or 45-degree angle from the surface. The
output leg of the reflection probe carries light from the sample to the spectrometer (top center), which is connected to the
PC (far left).
Common reflection applications include measuring the properties of mirrors and coatings. Other applications
include measuring the visual properties of the color in paints, plastics, and food products.
Sλ - Dλ
Iλ = Bλ ( R -D )
λ λ
…where Bλ is the relative energy of the reference (calculated from the color temperature) at wavelength λ, Sλ is
the sample intensity at wavelength λ, Dλ is the dark intensity at wavelength λ, Rλ is the reference intensity at
wavelength λ.
Figure 6-4: Typical Relative Irradiance Setup. A light source with a known color temperature (such as the LS-1or
LS-1-LL - lower right) is used to take a reference spectrum. The light to measure (lower left) is accumulated through a
CC-3 Cosine Corrector (or FOIS integrating sphere) into an input fiber, which carries the light information to the
spectrometer. The spectrometer then transmits the information to the PC, which compares the measured spectra against
the reference spectrum, thus removing wavelength-dependent instrument response from the measurement.
Common applications include characterizing the light output of LEDs, incandescent lamps, and other radiant
energy sources such as sunlight. Relative irradiance measurements also include fluorescence measurements,
which measure the energy given off by materials that have been excited by light at shorter wavelengths.
Perform the following steps to take a relative irradiance measurement using OOIBase32:
1. Place OOIBase32 is in scope mode by clicking the Scope Mode icon on the toolbar, or by selecting
Spectrum | Scope Mode from the menu bar.
2. Ensure that the entire signal is on scale. The intensity of the reference signal should peak at about 3500
counts.
Note: If you change any sampling variable (integration time, averaging, smoothing, angle, temperature,
fiber size, etc.), you must store a new dark and reference spectrum.
To configure a time acquisition process for the third, fourth, fifth, and sixth single wavelengths, select the
Channel C, Channel D, Channel E, and Channel F pages, respectively, and set the necessary
parameters.
(Continued)
3. Configure a time acquisition process for a combination of two time channels (if desired) by selecting
Combination 1.
Writing data to the disk is a slow process (relative to the speed of some spectral acquisitions) and
causes a decrease in system performance. However, writing data to disk more frequently gives a
larger margin of safety.
4. Enable Show Values in Status Bar to see the time acquisition values in the status bar. These values
replace the cursor values.
5. Name the Stream Filename for the time acquisition process. Clicking on the ellipsis to the right of this
box opens a file save dialog box, allowing you to navigate to a designated folder.
(Continued)
6. Enable Save Every Acquisition to store data for every spectral acquisition during a time acquisition
process (optional).
Note: OOIBase32 has options to either store data for each acquisition, or to collect data only after a
specified delay. Several factors affect the minimum time acquisition frequency, including
integration time, number of spectrometer channels, samples averaged, and computer speed. If
you instruct OOIBase32 to store data every 100 milliseconds, the delay between data
acquisitions will be 100 milliseconds or more, depending on your experimental configuration.
OOIBase32 spends a large amount of time calculating, rendering, and displaying the spectra in a
spectral window. You can suspend the graph display, which greatly improves the performance of
OOIBase32.
7. Enter an Initial Delay to set the delay preceding the time acquisition process. Keep in mind that the delay
countdown does not begin until you start the time acquisition process. Be sure to select Hours, Minutes,
Seconds, or Milliseconds immediately to the right of the initial delay entry.
8. Enter a value to set the Frequency of the data collected in a time acquisition process. OOIBase32
stamps data from a time acquisition with a time accurate to one millisecond. Be sure to select Hours,
Minutes, Seconds, or Milliseconds immediately to the right of the frequency entry. You can enable the
Save Every Acquisition box to store the acquisitions that occur at this frequency. See Step 6 for more
information.
9. Enter a value to set the Duration for the entire time acquisition process. Be sure to select Hours, Minutes,
Seconds, or Milliseconds to the right of the duration entry. Click the OK button to close the Time
Acquisition Configuration dialog box. Then, enable Continue Until Manually Stopped, which instructs
OOIBase32 to store data until you manually stop the acquisition process (optional).
λp = I + C1 p + C2 p2 + C3 p3
...where λ is the wavelength of pixel p, I is the wavelength of pixel 0, C1 is the first coefficient (nm/pixel), C2 is the
2 3
second coefficient (nm/pixel ), and C3 is the third coefficient (nm/pixel ). You will be calculating the value for I and
the three Cs.
• A USB2000 spectrometer
• An optical fiber (for spectrometers without a built-in slit, a 50-µm fiber works best)
• A spreadsheet program (Excel or Quattro Pro, for example) or a calculator that performs third-order linear
regressions
Note: If you are using Microsoft Excel, choose Tools | Add-Ins and check AnalysisToolPak and
AnalysisTookPak-VBA.
Predicted
True Wavelength (nm) Pixel # Pixel # 2 Pixel # 3 Difference
Wavelength
253.65 175 30625 5359375 253.56 0.09
296.73 296 87616 25934336 296.72 0.01
302.15 312 97344 30371328 302.40 -0.25
313.16 342 116964 40001688 313.02 0.13
334.15 402 161604 64964808 334.19 -0.05
365.02 490 240100 117649000 365.05 -0.04
404.66 604 364816 220348864 404.67 -0.01
407.78 613 375769 230346397 407.78 0.00
435.84 694 481636 334255384 435.65 0.19
546.07 1022 1044484 1067462648 546.13 -0.06
576.96 1116 1245456 1389928896 577.05 -0.09
579.07 1122 1258884 1412467848 579.01 0.06
696.54 1491 2223081 3314613771 696.70 -0.15
706.72 1523 2319529 3532642667 706.62 0.10
727.29 1590 2528100 4019679000 727.24 0.06
738.40 1627 2647129 4306878883 738.53 -0.13
751.47 1669 2785561 4649101309 751.27 0.19
(Continued)
5. Use the spreadsheet or calculator to calculate the wavelength calibration coefficients. In the spreadsheet
program, find the functions to perform linear regressions.
• If using Quattro Pro, look under Tools | Advanced Math
• If using Excel, look under Analysis ToolPak
6. Select the true wavelength as the dependent variable (Y). Select the pixel number, pixel number squared,
and the pixel number cubed as the independent variables (X). After executing the regression, you will
obtain an output similar to the one shown below.
Regression Statistics
Multiple R 0.999999831
R Square 0.999999663
Adjusted R Square 0.999999607
Standard Error 0.125540214
Observations 22
Intercept
Third coefficient
Note: This value is entered in microseconds (µs), even though the label shows milliseconds (msec).
7. Enter Variable Delay Mode by placing the USB2000-FLG in External Hardware Trigger mode. In
OOIBase32, select Spectrum | Configure Data Acquisition | External Trigger, and then select the
Hardware Trigger option. You do not need to apply a trigger signal.
(Continued)
8. Enable Single Strobe output by checking the Strobe Enable check box (See Figure B-1 on the following
page).
The strobe signal is a rising edge trigger signal that is TTL High for the entire delay period. See Figure B-2
for a timing diagram.
The USB2000-FLG is now configured for operation in Variable Delay Mode.
A M
Accessories, 14, 15 memory chip, 1
A/D Converters, 16 Mirror, 6
Direct-attach, 15
Integrated Sampling Systems, 17 O
Optical Fiber Assemblies, 18
Probes, 18 OOIBase32, 12
Sampling, 17 Optical Fiber Assemblies, 18
Adobe Acrobat Reader, 4 Overview, 1
AgPlus Mirrors, 41
P
C Packing List, 3
Calibrating, 36 Passwords, 4
CCD Detector, 6 Pin Definitions, 10
Compatibility, 9 Pinout, 9
Compatible Products, 15 Power, 7
Configure Hardware, 12
R
D range, 7
Data transfer rate, 7 resolution, 7
Default Spectrometer Configuration File, 12
Detector Collection Lens, 6 S
Device Drivers, 16
Slit, 6
Dimensions, 7
SMA Connector, 6
Dynamic range, 7
Software, 16
Software and Resources Library CD, 4
E Specifications, 5
EEPROM, 1 Detector, 7
Experiments, 23 Spectrometer, 7
Absorbance, 25 Spectroscopic Accessories, 14
Reflection, 29
Relative Irradiance, 31 T
Time Acquisition, 33
Time Acquisition, 33
Transmission, 27
Timing Diagram, 42
Troubleshooting, 19
F
Filter, 6 U
Focal length, 7
Upgrades, 4
Focusing Mirror, 6
USB2000-FLG, 41
USB-ADP-H, 3
G USB-ADP-PC, 3
Grating, 6 USB-CBL-1, 3
I V