ML Alpha 7400 ServiceManual R01
ML Alpha 7400 ServiceManual R01
01
March 2012
7400
SERVICE MANUAL
359006300
SERVICE MANUAL
Introduction
This manual describes the 7400 system functions and the allowed service procedures
for the Service technicians.
The manual is addressed to all the Service centres directly authorised by ESAOTE. All
the operations described in this manual are not critical from a safety point of view.
Do not attempt to service 7400 unless this service manual has been consulted and
understood.
The enter and context menu keys are respectively indicated as ENTER and UNDO keys
in this manual.
ii
SERVICE MANUAL
Section Overview
This Service Manual is composed of the following sections:
• Section 2: Disassembling/Assembling
• Section 7: Troubleshooting
WARNING In this manual WARNING identifies a risk for the patient and/or the operator.
CAUTION The word CAUTION describes the precautions necessary for protecting the
equipment.
iii
SERVICE MANUAL
iv
SERVICE MANUAL
Table of Contents
1 Service Characteristics.......................................................................1-1
Identifying the Boards ...........................................................................1-1
System Information ...............................................................................1-1
2 Equipment ........................................................................................2-1
Disassembling Equipment .....................................................................2-1
Service Procedures ...............................................................................2-1
USB Medium and Re-Writable CD.......................................................2-1
Upgrading Disks....................................................................................2-2
Safety Test Equipment ..........................................................................2-2
4 Precautions ........................................................................................4-1
5 System Packaging..............................................................................5-1
How to Package the System ..................................................................5-1
Section 2 - Disassembling/Assembling
1 How to Remove the LCD Covers, the LCD Display and Batteries.. 1-1
LCD Front Cover...................................................................................1-1
Tools ............................................................................................1-1
Removal Procedure...........................................................................1-1
Assembly Procedure ....................................................................1-2
LCD Display..........................................................................................1-2
Tools ............................................................................................1-2
Removal Procedure...........................................................................1-2
Assembly Procedure ....................................................................1-3
LCD Rear Cover....................................................................................1-3
Tools ............................................................................................1-3
Assembly Procedure ....................................................................1-4
Batteries ............................................................................................1-4
Tools ............................................................................................1-4
v
SERVICE MANUAL
1 Traceability ........................................................................................1-1
4 Power Cords......................................................................................4-1
1 Block Diagram...................................................................................1-1
Probe Adapter (PB) Board (B1) ...........................................................1-3
Front End (FE) Board (B2)....................................................................1-3
Probe Transmission Section .............................................................1-3
Probe Reception Section...................................................................1-3
CW Acquisition Section ...................................................................1-3
Power Supply Section.......................................................................1-3
Battery Bridge Board (B3) ....................................................................1-4
Back End (BE) Board (B4)....................................................................1-4
Keyboard Group (B5-B6)......................................................................1-5
Antenna Boards (B7-B8).......................................................................1-5
vi
SERVICE MANUAL
AC Filter ...........................................................................................1-2
Frame Rate........................................................................................1-2
Reset Touchscreen Calibration at Next Reboot................................1-2
Tools.............................................................................................1-2
Procedure .....................................................................................1-3
System Logo .....................................................................................1-4
2 Licences .............................................................................................2-1
Disabling Active Licence ......................................................................2-1
Demo Licence........................................................................................2-2
5 Printer Installation.............................................................................5-1
Procedure...............................................................................................5-1
Printing Test Images..............................................................................5-2
WScale.bmp......................................................................................5-2
vii
SERVICE MANUAL
Section 7 - Troubleshooting
3 Format Disk......................................................................................3-1
When Using the Format Disk ................................................................3-1
Format Procedure ..................................................................................3-2
Tools .................................................................................................3-2
Procedure ..........................................................................................3-2
Boot Configuration ......................................................................3-2
Format Procedure.........................................................................3-3
Boot Re-configuration..................................................................3-4
Burning Utility Installation ..........................................................3-4
Software Installation ....................................................................3-7
System Reconfiguration...............................................................3-9
Settings Installation....................................................................3-10
viii
SERVICE MANUAL
Chapter
1 - Service
1
Characteristics
CHARACTERISTICS
SERVICE
The aim of this introductory chapter is to illustrate some of the characteristics of
the system of fundamental importance to the technical staff, in particular, where to
find the equipment status and how to keep it up to date.
Refer to the next section for information on how to disassembly the unit and
which board can be removed.
System Information
Press the MENU key and select “SYSTEM INFO” option. The unit will show:
1-1
SERVICE MANUAL
Note
The BIOS revision is displayed on the touchscreen at the unit start
up.
1-2
SERVICE MANUAL
Chapter
2 - Equipment
2
CHARACTERISTICS
SERVICE
In this chapter the required tools to access the system internal parts are listed
together with the tools to activate the Service procedures and to perform a
functional test of the unit.
Disassembling Equipment
The disassembling In order to access the system, the following tools are needed:
procedures are detailed in
Section 2 Tool Dimension
The use of an electric screwdriver allows the access time to be minimized. Low
torques are recommended.
Service Procedures
Service key The service procedures that require an interaction with the unit, as for example the
P/N 8610264000 language setting, can be activated only through the service key. The service key is
provided with an USB connector and can be inserted in any of the USB ports.
The key must be inserted before switching on the unit, to be able to access to the
several service menus. If the procedure is not carried out properly, the system
doesn’t allow the access to the menu.
The symbol shown on the left is used in this manual to indicate all the procedures
that require the use of the key.
2-1
SERVICE MANUAL
Upgrading Disks
It is suggested to always have the last revision of the unit upgrading disk together
with the pertaining upgrading instructions. These tools allow the service technician
to install again the software on the unit, should it be necessary.
2-2
SERVICE MANUAL
Chapter
CONFIGURATIONS AND
Service Key
CALIBRATIONS
When the system is turned on with the service key inserted, MyLab stops its
inizialization phase at the desk top, showing the tool bar on the main display and
the icons of the utilities on the touch screen (see figure below).
Note
If the shutdown modality has been set to “Stand By” (“General
Setup” option of the MENU key), the system forces a complete
shutdown the first time the service key is logged. The same applies
the first time the system is turned on without the service key
inserted and immediately after having used it.
To move the cursor on the touchscreen, completely scroll the trackball towards the
left side of the screen. To move the cursor back on the main display, completely
scroll the trackball towards the right side of the touchscreen.
3-1
SERVICE MANUAL
Place the cursor on the “Workflow” icon and press the ENTER key twice to
complete the initialization phase and activate the standard interface to access to the
system menu.
3-2
SERVICE MANUAL
Chapter
4 - Precautions
4
CHARACTERISTICS
SERVICE
A series of precautions that must be observed each time you perform an operation
on the system are emphasised below:
The chapter entitled “Parts List” details the sub-assemblies that are to be
considered sensitive to electro-static discharges (ESD).
When the board is ESD sensitive, you can find this symbol on the PCB master.
4-1
SERVICE MANUAL
4-2
SERVICE MANUAL
Chapter
5 - System Packaging
5
CHARACTERISTICS
SERVICE
Whenever the 7400 has to be shipped, the system has to be properly packaged
to prevent traumatic damage. The procedure below explains how to do it.
Note
The original package (composed of lateral protections, carton
cover, foam support for battery location, manual box and system
carton box) has to be stored should it be necessary to ship the
system.
• Place the unit with the protection into the system carton box.
5-1
SERVICE MANUAL
Note
All the items composing the 7400 packaging are necessary: the
system does not have to move within the system carton box.
CAUTION The system could be damaged during transportation if the above rules are
not strictly followed.
5-2
SERVICE MANUAL
Chapter
In order to carry out all the removal procedures (with the exception of the battery
replacement), all connected probes, peripherals and cables must be disconnected.
DISASSEMBLING/
ASSEMBLING
CAUTION For battery replacement carefully follow the instruction provided in this
chapter.
Tool Dimension
1/4 h
Removal Procedure
• Remove the two (2) little rubber plugs located on the left and
right ide of the LCD front covers.
• Start in the lower side and work your way around the LCD
cover loosening the tongue and groove joints.
1-1
SERVICE MANUAL
Assembly Procedure
Carry out the inverse of the removal procedure.
Once the front cover has been removed, you get access to LCD Display and the
Rear Cover.
Tool Dimension
1/4 h
Removal Procedure
1-2
SERVICE MANUAL
• Unscrew the three (3) Phillips screws placed on the left and
right sides of the display, shown on the filgure below.
DISASSEMBLING/
ASSEMBLING
• Rise the LCD display.
Assembly Procedure
In reassembling the display, pay attention.to correctly fit the ultraminitaure cables
between the ridges around the LCD display itself. When the cables have been
posiotioned, fix them with one drop of hot melt glue.
Tool Dimension
1/4 h
Removal Procedure
1-3
SERVICE MANUAL
• Unscrew the four (4) slotted screws (two for each side) that
fix the LCD display to its bracket.
Assembly Procedure
Carry out the inverse of the removal procedure.
Tools
Tool Dimension
1/4 h
Removal Procedure
1-4
SERVICE MANUAL
• Connect the unit to the mains. The unit must be OFF and
powered (the ON/OFF button on the control panel is
lighted).
• Unscrew the n.1 (one) screws that fix the battery cover.
CAUTION The system could be damaged if the above procedure is not attended.
Assembly Procedure
Carry out the inverse of the removal procedure.
Spare Reference
The spare reference includes two batteries.
DISASSEMBLING/
ASSEMBLING
1-5
SERVICE MANUAL
1-6
SERVICE MANUAL
Chapter
1 - Traceability
1
The traceability of a system part is based on the Serial Number (S/N).
The Serial Number of a system part is composed of two (2) or more figures. It is
reported in a label affixed to:
In the following chapter all parts that are traceable are identified.
PART LISTS
1-1
SERVICE MANUAL
1-2
SERVICE MANUAL
Chapter
2 - Parts Lists
2
LCD Display
LCD Covers
No cover is considered a traceable part or to be sent in advance.
PART LISTS
AC/DC Adapter
Batteries
9103316000 Batteries 2 N
2-1
SERVICE MANUAL
The batteries installed in the 7400 model are manufactured by Insipired Energy
(www.inspiredenergy.com), P/N NL2024HD22.
CAUTION Do not use different batteries form the ones specified above.
2-2
SERVICE MANUAL
Chapter
3 - System Replacement
3
When the system has to be replaced in the field, it could be necessary to configure
the new equipment as the replaced one is, that is with the same clinical and system
ssettings and the same archive.
This operation can be done only when it is possible to access to the system
internal folders.
Note
All system licences are correlated to the ID button (Hardware ID):
make sure to have all the licences associated to the Hardware ID of
the new equipment.
PART LISTS
• Service key P/N 8610264000
Procedure
• Using the trackball select the “work” folder and press ENTER
twice.
3-1
SERVICE MANUAL
Tools
Procedure
• Open the contextual menu with the UNDO key and select the
“Copy” option.
• Using the trackball select the “work” folder and press ENTER.
3-2
SERVICE MANUAL
• Open the contextual menu with the UNDO key and select the
“Copy” option.
• Using the trackball select the “work” folder and press ENTER.
PART LISTS
3-3
SERVICE MANUAL
3-4
SERVICE MANUAL
Chapter
4 - Power Cords
4
The 7400 system is equipped with the power cord having the following
characteristics:
Cond.
Device Socket Plug Sect. Length
#
CEE Plug IEC 60320/C13 type; CEE (7) VII type; 3 1 mm2 2,5 m
10A-250V 16A-250V
CEI Plug IEC 60320/C13 type; CEI 23-50 S11 type 3 1 mm2 2,5 m
12A-250V
According to the IEC 60601-1 standard these cables are marked by a competent
body (TUV, UL, IMQ,..).
If these cables can’t be used in your country because of the plug, they can be only
replaced by other power cord cable having the same characteristics including the
PART LISTS
Safety mark.
4-1
SERVICE MANUAL
4-2
SERVICE MANUAL
Chapter
1 - Block Diagram
1
This chapter contains the general block diagram description. The diagram shows
the relationships of each board having a unique function within the overall unit.
BLOCK DIAGRAM
1-1
SERVICE MANUAL
1-2
SERVICE MANUAL
The board mounts n.2 (two) small connector together with the switching logic to
enable the active connector.
The main function of the board is to switch among the four probe connectors
selecting the active one for the acquisition.
The board manages the transmission to the probe controlling the pulses to be
transmitted to the elements. The transmitters are individually controlled in term of
enabling, delaying and pulse shape by a dedicated FPGA.
The board manages also the reception of the signals implementing the receiving
beam former.
The received signals are summed together, amplified, digitally converted and then
sent to dedicated RX FPGAs that implement the digital focalization.
CW Acquisition Section
This dedicated section manages the acquisition of the CW signals from their
BLOCK DIAGRAM
The FE implements the power supply function distributing the supply voltage
within both FE and BE boards so that all the other needed supply voltages can be
generated. It supplies also the AT voltages for probe transmission and the standby
voltages for PC module.
1-3
SERVICE MANUAL
The bridge includes the power voltages, the SMB logic connections and the
monitoring thermal signals.
The BE board processes the RF signal through n.2 (two) FPGAs and n.2 (two)
DSPs. The first FPGA processes the RF signal and the Doppler data, the second
processes the CFM data.
The first DSP controls the processes on the two FPGAs, the acquisition sequences
and processes both Doppler data and Physio signal. The second filters and mixes
the image data.
1-4
SERVICE MANUAL
The BE Board also processes the ECG PHISYIO analogic signal up to its AD
conversion.
These n.2 (two) boards realize the antenna for the Wi-Fi connection.
1-5
SERVICE MANUAL
1-6
SERVICE MANUAL
Chapter
1 - Service Menu
1
CONFIGURATIONS AND
CALIBRATIONS
The MENU key provides the access to the service menu. The key can be pressed in
every environment.
When the system is initialized with the service key inserted, the “Service” tab,
Service key is required displayed on the top of the menu, is added to the System and Clinical
Configuration Menu.
Place the cursor on the “Service” tab and press ENTER to confirm. The system
displays the service menu, that makes it possible:
• to set whether to show the frame rate on the real time image
or not.
1-1
SERVICE MANUAL
Language Setting
This option sets the default language that can be selected among the available ones.
Note
With the exception of countries whose language is managed by
7400, all 7400 units delivered by Esaote have English as the default
language.
AC Filter
According to the mains frequency used in the country, the AC filter for the ECG
signal must be selected. This is the filter that removes the mains frequency
interference from the ECG signal.
Frame Rate
When enabled, the frame rate is displayed on the bottom right side of the image.
This procedure sets the alignment between the touchscreen and the touchscreen
display.
The procedure is in two phases: in the first phase the operator has to confirm the
position of targets; in the second phase the operator has to validate the targets
alignment.
Note
During the procedure pay attention to touch the screen only with
the pen. Not following this rule could compromise the procedure.
Tools
• Touchscreen pen.
1-2
SERVICE MANUAL
Procedure
First Phase • Check this option in the service menu.
CONFIGURATIONS AND
• At the start up, the following message is displayed on the
CALIBRATIONS
touchscreen:
• Repeat the same operation for all the displayed green targets.
1-3
SERVICE MANUAL
Second Phase • The system displays a pink target with a white cross inside.
Using the touchscreen pen, press the first target at the cross
center.
• Once the target has been pressed, the system displays a blue
cross within the target.
• Repeat the same operation for all the displayed green targets.
• The greatest is the distance between the white and the blue
crosses, the biggest is the disalignement between the
touchscreen and its display.
System Logo
Either Esaote or the model name can be set as logo. The default logo is Esaote.
1-4
SERVICE MANUAL
Chapter
2 - Licences
2
CONFIGURATIONS AND
CALIBRATIONS
When the system is initialized with the service key inserted, the “LICENCES”
option of the MENU key allows the operator to disable active licences, to activate
or to reset a demo licence.
Service key is required Place the cursor on the “LICENCES” option and press ENTER to confirm. The
system displays the following window:
Note
The licence number has to be input again to reactivate the licence.
2-1
SERVICE MANUAL
Demo Licence
Demo licenses last five hundreds (500) working hours before expiring.
Note
The Reset Demo licence has to be installed within 2 (two) months
from its purchase.
To activate the demo license for the first time, position the trackball on the desired
field and press ENTER to check it. Press SAVE to confirm.
If at least a demo license has been activated, the “LICENCES” option of the
MENU key allows the user to check its expiring date.
Reset Demo When the demo licence is expired, it can be reactivated through the Reset Demo
licence to be inserted in the “RESET DEMO” tab of the Service menu. Once
expired, it can be re-activated five times indipendently from the enabled demo
licence.
The Reset Demo licence is always related to the system Hardware ID, shown on
the left upper corner of the licences menu.
2-2
SERVICE MANUAL
Chapter
3 - Software Release
3
CONFIGURATIONS AND
Installation
CALIBRATIONS
The support used to install the software release on the unit is tipically a DVD
containing all necessary files.
Some software releases could have specific procedures (for instance the Bios Set
Service key is required Up could be run before software installation): refer to the protected service web
site to know if a software release requires specific instructions.
WARNING If the software upgrade has been downloaded from the dedicated Esaote
ftp site, be sure that the PC used for the download operations is protected
from viruses.
To copy the software upgrade always and only use removable devices with
a safe content (devices used only on systems protected by malware).
If the equipment is equipped with the batteries, be sure that the system is
connected to the main before starting the upgrading procedure. The
system must remain connected to the main during the upgrade, including
the first reboot to complete the firmware upgrade.
Note
Before proceeding with the procedure save both the clinical and
system configurations and back up the archive.
3-1
SERVICE MANUAL
Installation Procedure
Tool
Procedure
3-2
SERVICE MANUAL
CONFIGURATIONS AND
•
CALIBRATIONS
Press the “Ok, continue” button to proceed with the
installation.
• Enter the Service menu and verify that the monitor model is
correctly set; if necessary, set it.
• Place the cursor on the “Start” option of the Tool bar and
press ENTER.
3-3
SERVICE MANUAL
FPGA Rescue
The system has installed a tool that can restore the Algol device, named
“FpgaRescue”.
Note
Before running the procedure check which configuration index has
the Back End board.
inst\update\Board\FpgaRescue
3-4
SERVICE MANUAL
CONFIGURATIONS AND
• Place the cursor on the “Start” option of the Tool bar and
press ENTER.
CALIBRATIONS
• Select the “Control Panel” option.
• Place the cursor on the “Workflow” icon and press the ENTER
to launch the software.
If FPGA rescue procedure doesn’t solve the problem or if the Algol device was
listed and the installation continously fails, the HD has to be replaced as first
service attempt.
3-5
SERVICE MANUAL
3-6
SERVICE MANUAL
Chapter
4 - Thermal Printer
4
CONFIGURATIONS AND
Installation
CALIBRATIONS
The system software automatically installs the driver for the SONY UP-D897
thermal printer.
Note
All the systems delivered by ESAOTE are configured to correctly
manage the Sony UP-D897 printer.
Should it be necessary to re-configure this printer in the field, the procedure below
describes the steps to be done.
Note
The procedure must be done without any service key inserted: the
settings apply only to the user active at the time of the
configuration.
Procedure
• Connect the printer to one of the unit USB connectors.
• Place the cursor on the printer icon and press the UNDO key.
4-1
SERVICE MANUAL
• “Paper” to 1920x1280
• “Orientation” to Portrait
4-2
SERVICE MANUAL
Chapter
5 - Printer Installation
5
CONFIGURATIONS AND
CALIBRATIONS
The system supports both laser and inkjet USB printers: please refer to the web site
for the list of the printers that can be connected to the system.
Before starting the installation procedure be sure to have the printer installation
Service key is required driver CD with you. The procedure runs like a printer installation in Windows.
Note
The operative system installed in the system is Windows®
Embedded.
Procedure
• Connect the printer to one of the unit USB connectors.
• Turn both the printer and the system (with the service key
inserted) on.
• Select “No, not this time” option and press Next to proceed.
Note
Please note that only the printer driver has to be installed! Any
other printer program which might be listed or proposed during
this installation phase has to be disactivated.
5-1
SERVICE MANUAL
• Switch the unit off, remove the service key and then switch
the unit on.
The printer models tested by Esaote, whose installation differs from the procedure
described above, have specific instructions available in the protected service web
site.
Esaote does not test all printer models available on the market. Other printers may
be therefore compatible with our systems, and require different installation
procedures to ensure their complete range of performances.
Procedure • Insert the Service Key into the unit and switch it on.
• Place the cursor on the “Start” option of the Tool bar and
press ENTER.
The driver contains several bmp images to be used as reference for the printing
quality setting.
WScale.bmp
The WScale.bmp image has a pattern with 256 different grey levels.
5-2
SERVICE MANUAL
CONFIGURATIONS AND
CALIBRATIONS
These levels are uniformly distributed in the four columns. This uniform
distribution has to be kept in the test printing. If not, printer contrast and
brightness adjustments are required.
5-3
SERVICE MANUAL
5-4
SERVICE MANUAL
Chapter
1 - Maintenance
1
Procedures
The maintenance procedures are:
• Trackball cleaning
• Battery replacement
Trackball Cleaning
MAINTENANCE
PROCEDURES
The trackball has to be cleaned at least once a year.
Tool
Tool Dimension
Paintbrush -
Procedure
1-1
SERVICE MANUAL
Battery Replacement
Batteries duration is higher than 300 charging cycles and their replacement is
recommended at least every three years.
Note
Both batteries must be always replaced together.
CAUTION Do not use different batteries from the ones specified above.
The battery is Lithium Ion (Li-Ion) battery, FCC approved; the cells, composing
the battery, are UL marked.
Note
The batteries must be treated as special waste according to the
applicable local regulations.
The batteries REF 9103316000, provided by Esaote, are charged: the last
recharging operation is indicated in the recharging card. The batteries must be fully
recharged every six (6) months when they are kept on stock.
CAUTION The system could be damaged if the battery disassembling procedure is not
attended.
1-2
SERVICE MANUAL
Charging Procedure
The batteries can be charged either by installing them on the 7400 or by using the
charger manufactured by Inspired Energy:
Refer to the manufacturer “Instructions for use” manual to properly manage and
charge the batteries with the charger.
GS Refer to the “Getting Started” manual to charge the batteries using the 7400.
MAINTENANCE
PROCEDURES
CAUTION Do not expose the battery to high temperature. Keep the battery away from
fire and other heat sources.
Do not charge the batteries near a heat surface such as fire or heater.
Storage
1-3
SERVICE MANUAL
1-4
SERVICE MANUAL
Chapter
2 - Safety Tests
2
Every unit complies with EN60601-1 (IEC 60601-1) standard. The system is Class
I Type B and BF (ultrasound probe) and CF (ECG) applied parts.
MAINTENANCE
PROCEDURES
WARNING Whenever the measured values exceed the reference ones (see the following
table) don’t use the unit and send it to ESAOTE.
This chapter defines safety parameters and their range according to EN60601-1
standard and describes the safety tests to be carried out on the equipment.
The impedance between the Protective Earth (PE) terminal of the mains input
connector and any accessible metal part.
The current that flows from the mains terminals (P=Phase, N=Neutral) to the
Protective Earth (PE) through the insulation.
2-1
SERVICE MANUAL
The current that flows between the enclosure and the Protective Earth (PE)
terminal.
The current that flows through the applied parts (ECG and US probe) towards the
Protective Earth (PE).
The current that flows between two different applied parts (for instance between
two ECG electrodes).
The table below provides the user with a list of the parameters to be checked, the
maximum values and references to the IEC 60601-1 standard (1988), II Edition
and connected amendments.
Max Value
Parameter Max Value N.C.a EN 60601-1
S.F.C.b
2-2
SERVICE MANUAL
Note
The test must be carried out by skilled personnel using equipment
compliant with the reference standard indicated. ESAOTE
recommends the use of the automatic BIO-TEK 601-PRO or
Fluke Esa620 equipment manufactured by Fluke Biomedical
(http://global.flukebiomedical.com/busen/home/default.htm) or
equivalent equipment.
Note
MAINTENANCE
PROCEDURES
Before proceeding with the safety test, be sure of the calibration of
the testing equipment .
Procedure
2-3
SERVICE MANUAL
If the measured value is higher than the value in the table, do not use the equipment
and send to ESAOTE.
If any measured value is higher than the value in the table, do not use the
equipment and replace the power cable. Repeat the procedure with the new cable
and if the values continue to be higher, send the equipment to ESAOTE.
Place the metal foil in close contact with the unit and
activate the procedure for measuring the enclosure
leakage current on the automatic equipment; varying the
foil position, check that the value indicated complies
2-4
SERVICE MANUAL
MAINTENANCE
• Repeat the above measurement procedures inverting the
PROCEDURES
polarity of the power conductors, by means of the automatic
equipment.
If any of the measured values are higher than the value in the table, do not use the
equipment and send it to ESAOTE.
2-5
SERVICE MANUAL
If any of the measured values are higher than the value in the table, do not use the
equipment and replace the ECG patient cable. Repeat the procedure with the new
cable and if the values continue to be higher, send the equipment to ESAOTE.
If any of the measured values are higher than the value in the table, do not use the
equipment and replace the ECG patient cable. Repeat the procedure with the new
cable and if the values continue to be higher, send the equipment to ESAOTE.
WARNING The defective patient cable or power cable must not be used.
2-6
SERVICE MANUAL
wrapping the probe enclosure with the metal foil and connecting the test lead to it.
The test must be performed in any Real Time mode application. The probe must
be connected to the unit.
Procedure
MAINTENANCE
PROCEDURES
(SFC).
If any of the measured values are higher than the value in the table, do not use the
probe and send it to ESAOTE.
WARNING Any break in the probe case or in the probe cable can cause an electrical
hazard. Do not use the probe and send it back to ESAOTE for repair.
2-7
SERVICE MANUAL
2-8
SERVICE MANUAL
Chapter
• System information
• Log files
• Fault description
System Information
TROUBLESHOOTINGS
The first thing to be done is to identify the unit, this means to know the unit serial
number, the model and the installed software release: this information is listed in
the “SYSTEM INFO” option of the MENU key.
Log Files
From the user interface If the system allows the user to access to the system menu, the same option allows
the user to export the log files on an USB medium.
1-1
SERVICE MANUAL
• Press OK to finish.
Note
The log files are saved already zipped.
From the operative system If the system is not working and it is not possible to access to the system menu,
follow the procedure below to save the log files:
• Place the cursor on the “Start” option of the Tool bar and
press ENTER.
• Using the trackball select the “work” folder and press ENTER.
Fault Description
The description of the fault has to be the more detailed as possible, since this
simplifies the problem identification and solution. The following information
should always be reported:
1-2
SERVICE MANUAL
Together with this information, the description should state all the steps that cause
the problem (for instance during 2D-CFM scanning with SP2430 probe in
Cardiac, the 2D sector was reduced to 45°, the CFM sector was set to concurrent.
After having saved a clip the scanning sector doesn’t follow the movements of the
trackball). Images and/or Clips showing the problems are of course other useful
items.
TROUBLESHOOTINGS
below to copy the archive:
• Press ARCHIVE.
• Press OK to confirm.
From the operative system If the system is not working and it is not possible to access to the archive, follow
the procedure below to copy it:
1-3
SERVICE MANUAL
• Place the cursor on the “Start” option of the Tool bar and
press ENTER.
• Using the trackball select the “work” folder and press ENTER.
• Using the trackball select the “work” folder and press ENTER
twice.
1-4
SERVICE MANUAL
Chapter
2 - Log Files
2
The log file has to be saved soon after the problem occurs or, at the latest, at the
next start up. The file has a FIFO structure with a limited memory availability:
When the log file memory is full, the oldest events registered in the file are
eliminated to free space for the recent ones.
For this reason the log file has to be saved as soon as possible.
TROUBLESHOOTINGS
• LogInstallMS folder that traces the updates done on the
system.
• Dump folder.
LogParser Executable
This file allows the technician to look at the log files for a first error analysis. To
launch the executable, place the cursor on it and press the ENTER key twice. This
utility can be run both directly in the system or in any PC having the following
operating systems:
• Windows XP Home
• Windows XP Professional
• Windows Vista
• Windows 7
2-1
SERVICE MANUAL
The TextLog folder, containing the log files, is displayed in the upper left area
(indicated by the number “1” in the above figure). Place the cursor on the “+”
symbol displayed beside the name and press ENTER to display its contents.
TextLog Folder This folder contains the list of all entries done on the system, that is the list of all
times the unit has been switched on. Each entry is so listed:
The first date and time indicate when the unit has been turned on; the second date
and time when the unit has been turned off.
When an entry is selected, the areas “2” and “3” show the logged information
related to this entry.
In the left bottom area (“2”) the following system information is listed:
• model name
• installed build
• hardware characteristics
2-2
SERVICE MANUAL
In the main area (“3”) the list of the operations logged for the selected entry is
displayed.
The three buttons displayed at the bottom of the area “3” allow to change the
zoom factor (buttons “-” and “+”) and to refresh the log of the operations.
Criteria for displaying the Displaying criteria can be selected from the upper left and right lists. The upper
logs left list contains all the system software components; the upper right list contains
logic log groups with which the list of logged operations can be organized for an
easiest analysis.
Place the cursor on the desired element of one of the lists to display the related logs.
User Session
Whenever the log has to be analysed, the simplest way is to follow the procedure
TROUBLESHOOTINGS
below:
The logged operations are displayed in chronologic order, showing the operations
done by the user.
2-3
SERVICE MANUAL
The main phases of the exam are clearly indicated by data delimiters, like:
When an error occurs the delimiters allow to identify in which phase happens.
Logged Errors The errors are reported in the upper right list, as shown in the figure below.
2-4
SERVICE MANUAL
The error log group displays all the errors occurred in the system.
TROUBLESHOOTINGS
2-5
SERVICE MANUAL
2-6
SERVICE MANUAL
Chapter
3 - Format Disk
3
This chapter gives the indications on how to use the format disk.
WARNING The format disk will erase all patient data saved in the internal data base:
before running the format procedure, if possible, save the archive folder.
TROUBLESHOOTINGS
Note
The field service technician must inform the system owner that the
format procedure erases all patient data saved in the internal data
base. If no archive back up is available and if the internal Archive
folder cannot be copied, the patient data will be definitively lost
after the format procedure.
Before executing the format procedure, run first the upgrading procedure,
described in this manual (refer to Section 5 for further details) to try to unlock the
system.
WARNING If the equipment is equipped with the batteries, be sure that the system is
connected to the main before starting the format procedure. The system
must remain connected to the main during the format and during all the
next steps, including the first reboot to complete the firmware upgrade.
3-1
SERVICE MANUAL
Format Procedure
The format procedure has six (6) main steps:
• Boot Configuration
• Format Procedure
• Boot Re-configuration
• Software Installation
• System Configuration
Tools
Procedure
Boot Configuration
• Disconnect any digital USB peripherals connected to the
unit.
3-2
SERVICE MANUAL
TROUBLESHOOTINGS
• Repeat the password and confirm it again.
• Usin the arrow keys, select the WRITE CMOS AND EXIT
otpion and press Enter ↵ to confirm.
Format Procedure
• Insert the disk and turn the system off.
3-3
SERVICE MANUAL
Boot Re-configuration
• Insert the BIOS disk and switch the unit off just by pressing
the ON/OFF key.
• Turn the system off by keeping pressed the ON/OFF key for
at least 5 sec.
• Place the cursor on the “Start” option of the Tool bar and
press ENTER.
sw_tools\Nero_7_5\CDS\Nero
• Using the cursor select the setup file and run it by pressing
the ENTER key twice.
3-4
SERVICE MANUAL
TROUBLESHOOTINGS
3-5
SERVICE MANUAL
• NeroBackupIt
• Nero CoverDesigner
• InCD
To exclude a feature, place the cursor on it, press the UNDO key and
select the option “This feature will not be available”.
• In the last window deselect all options and press the Next
button to proceed with the installation.
3-6
SERVICE MANUAL
Software Installation
• Place the cursor on the “Start” option of the Tool bar and
press ENTER.
• Using the cursor select the SETUP file and run it by pressing
the ENTER key twice.
TROUBLESHOOTINGS
• The system starts the installation phase and stops at the
following windows
3-7
SERVICE MANUAL
• Repeat this step for all the USB connectors on the system (on
the control panel, on the left side and the rear panel).
• Switch the unit off and on again, leaving the service key
inserted.
3-8
SERVICE MANUAL
System Reconfiguration
• Switch the unit on with the service key inserted.
TROUBLESHOOTINGS
archive had been saved, follow the corresponding procedure:
• Open the contextual menu with the UNDO key and select
the “Copy” option.
3-9
SERVICE MANUAL
Settings Installation
• Switch the unit on with the service key inserted.
Note
If the settings were neither saved nor exported, they have to be
manually set one by one.
• Open the contextual menu with the UNDO key and select
the “Copy” option.
If the system is equipped with a thermal printer, refer to the next section for the
configuration of the printing settings.
3 - 10
SERVICE MANUAL
TROUBLESHOOTINGS
3 - 11
SERVICE MANUAL
3 - 12
SERVICE MANUAL
Chapter
1 - Check Lists
1
This chapter proposes two different check lists. The first one (Technical Check
List) suggests a set of inspection steps to be performed each time service personnel
modifies the HW/SW composition (for example the rack replacement or
functional upgrade); the other one (Maintenance Check List) suggests a check list
Service key is required intended for a general equipment inspection like yearly maintenance.
1-1
SERVICE MANUAL
8 For all probes, check their external status (cable, “Transducers and
housing, connector.). Consumables” manual
General Check
2 Check that the operator manuals are present and Operator manual
their revisions are consistent with the installed revisions
software release.
Functional check
1-2
SERVICE MANUAL
5 Verify that the CFM mode controls work properly Advanced Operations
manual, Section
“Image Optimization
and Annotation”
CHECK LISTS
1-3
SERVICE MANUAL
1-4