Proline Promass 500 Modbus RS485 Table of contents
Table of contents
Endress+Hauser 3
Table of contents Proline Promass 500 Modbus RS485
Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 313
4 Endress+Hauser
Proline Promass 500 Modbus RS485 About this document
Endress+Hauser 5
About this document Proline Promass 500 Modbus RS485
Parameter 1
Operator
Parameter n
Submenu 1
Submenu n
Wizard 1 / Parameter 1
Maintenance
Task-oriented
Wizard n / Parameter n
Parameter 1
Parameter n
Submenu 1
Submenu n
Diagnostics Parameter 1
Parameter n
Submenu 1
Submenu n
Parameter n
System Function-oriented
Sensor
Expert
Input
Output
Communication
Application
Diagnostics
A0029160-EN
6 Endress+Hauser
Proline Promass 500 Modbus RS485 About this document
Navigation Navigation path to the parameter via the local display (direct access code) or web browser
Navigation path to the parameter via the operating tool
The names of the menus, submenus and parameters are abbreviated to the form in which they appear on
the display and in the operating tool.
Symbol Meaning
Tip
Indicates additional information.
Reference to documentation
A Reference to page
Reference to graphic
Write-protected parameter
Endress+Hauser 7
About this document Proline Promass 500 Modbus RS485
1.5 Documentation
Operating Instructions
Special Documentation
Radio approvals for WLAN interface for A309/A310 display module SD01793D
Petroleum SD02014D
8 Endress+Hauser
Proline Promass 500 Modbus RS485 Overview of the Expert operating menu
Expert
‣ System → 15
‣ Display → 15
‣ Config. backup → 30
‣ Diagn. handling → 33
‣ Administration → 44
‣ Sensor → 50
‣ Measured val. → 50
‣ System units → 65
‣ Process param. → 82
‣ Calculated value → 95
‣ Measurement mode → 90
‣ External comp. → 92
‣ Sensor adjustm. → 98
‣ Calibration → 106
Endress+Hauser 9
Overview of the Expert operating menu Proline Promass 500 Modbus RS485
‣ Input → 110
‣ Output → 115
‣ Curr.output 1 to n → 116
‣ Communication → 164
‣ Application → 181
‣ Totalizer 1 to n → 182
‣ Viscosity → 187
‣ Concentration → 187
‣ Petroleum → 187
10 Endress+Hauser
Proline Promass 500 Modbus RS485 Overview of the Expert operating menu
‣ Diagnostics → 195
‣ Cust.transf.log. → 204
‣ Heartbeat → 237
‣ Simulation → 237
Endress+Hauser 11
Description of device parameters Proline Promass 500 Modbus RS485
Expert
‣ System → 15
‣ Sensor → 50
‣ Input → 110
‣ Output → 115
‣ Communication → 164
‣ Application → 181
‣ Diagnostics → 195
Direct access
Description Use this function to enter the access code to enable direct access to the desired parameter
via the local display. A parameter number is assigned to each parameter for this purpose.
12 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters
00914-2 1
A0029414
Locking status
Selection
Options Description
None The access status displayed in the Access status parameter (→ 14) applies .
Only appears on local display.
Hardware locked The DIP switch for hardware locking is activated on the PCB board. This locks write
(priority 1) access to the parameters (e.g. via local display or operating tool) .
Cust.trans. act.
(priority 2) Only available for Promass F, O, Q and X.
The DIP switch for custody transfer mode is activated on the PCB board. This locks
write access to all the parameters (e.g. via local display or operating tool).
Endress+Hauser 13
Description of device parameters Proline Promass 500 Modbus RS485
Options Description
CT act.-def.par.
(priority 3) Only available for Promass F, O, Q and X.
The DIP switch for custody transfer mode is activated on the PCB board. This locks
write access to the defined parameters (e.g. via local display or operating tool).
Temp. locked Write access to the parameters is temporarily locked on account of internal
(Priority 4) processes running in the device (e.g. data upload/download, reset etc.). Once the
internal processing has been completed, the parameters can be changed once
again.
Access status
Description Displays the access authorization to the parameters via the local display, Web browser or
operating tool.
Display
Detailed information on access authorization is provided in the "User roles and
associated access authorization" and "Operating concept" sections of the Operations
Instructions for the device → 8
Description Use this function to enter the user-specific release code to remove parameter write
protection.
User entry Max. 16-digit character string comprising numbers, letters and special characters
14 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters
‣ System
‣ Display → 15
‣ Config. backup → 30
‣ Diagn. handling → 33
‣ Administration → 44
‣ Display
0% bargraph 1 (0123) → 21
0% bargraph 3 (0124) → 24
Endress+Hauser 15
Description of device parameters Proline Promass 500 Modbus RS485
Header (0097) → 28
Separator (0101) → 29
Backlight (0111) → 29
Display language
Description Use this function to select the configured language on the local display.
Selection • English
• Deutsch
• Français
• Español
• Italiano
• Nederlands
• Portuguesa
• Polski
• русский язык(Ru)
• Svenska
• Türkçe
• 中文 (Chinese)
• 日本語 (Japanese)
• 한국어 (Korean)
• (الْعَرَبيّةAra) *
• Bahasa Indonesia
• ภาษาไทย (Thai) *
• tiếng Việt (Vit)
• čeština (Czech)
Factory setting English (alternatively, the ordered language is preset in the device)
16 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters
Format display
Description Use this function to select how the measured value is shown on the local display.
• Once the measuring device has been enabled for custody transfer mode, depending on
the custody transfer approval selected the display can switch between showing the
relevant information and the custody transfer counter.
• In addition, a padlock symbol appears in the header of the display ().
For detailed information on custody transfer mode, see the Special Documentation for
the device → 8
Endress+Hauser 17
Description of device parameters Proline Promass 500 Modbus RS485
XXXXXXXXX
900.00
kg/h
A0013099
XXXXXXXXX
900.00 kg/h
900.00
l/h
A0013098
XXXXXXXXX
900.00
kg/h
900.00
l/h
A0013100
XXXXXXXXX
900.00
kg/h
900.00 l/h
1.00 kg/l
A0013102
XXXXXXXXX
900.00 kg/h
900.00 l/h
1.0 kg/l
213.94 kg
A0013103
18 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters
Value 1 display
Description Use this function to select one of the measured values shown on the local display.
Endress+Hauser 19
Description of device parameters Proline Promass 500 Modbus RS485
• Exc. current 0 *
• Exc. current 1 *
• HBSI *
• Totalizer 1
• Totalizer 2
• Totalizer 3
• Curr.output 1 *
• Curr.output 2 *
• Curr.output 3 *
• Curr.output 4 *
• Pressure
• Spec. output 1 *
• Index inh.medium
• Spec. output 0 *
• Index sus.bubble *
Once the measuring device has been enabled for custody transfer mode, depending on the
custody transfer approval selected the display can switch to show the relevant information.
For detailed information on custody transfer mode, see the Special Documentation for
the device → 8
Dependency
The unit of the displayed measured value is taken from the System units submenu
(→ 65).
Selection
• Oscil. frequency option
Displays the current oscillation frequency of the measuring tubes. This frequency
depends on the density of the medium.
• Oscil. amplitude option
Displays the relative oscillation amplitude of the measuring tubes in relation to the
preset value. This value is 100 % under optimum conditions.
• Oscil. damping option
Displays the current oscillation damping. Oscillation damping is an indicator of the
sensor's current need for excitation power.
• Signal asymmetry option
Displays the relative difference between the oscillation amplitude at the inlet and outlet
of the sensor. The measured value is the result of production tolerances of the sensor
coils and should remain constant over the life time of a sensor.
20 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters
0% bargraph 1
Description Use this function to enter the 0% bar graph value to be shown on the display for the
measured value 1.
User entry
The unit of the displayed measured value is taken from the System units submenu
(→ 65).
100% bargraph 1
Description Use this function to enter the 100% bar graph value to be shown on the display for the
measured value 1.
User entry
The unit of the displayed measured value is taken from the System units submenu
(→ 65).
Endress+Hauser 21
Description of device parameters Proline Promass 500 Modbus RS485
Decimal places 1
Description Use this function to select the number of decimal places for measured value 1.
Selection •x
• x.x
• x.xx
• x.xxx
• x.xxxx
Value 2 display
Description Use this function to select one of the measured values shown on the local display.
Selection For the picklist, see the Value 2 display parameter (→ 22)
Dependency
The unit of the displayed measured value is taken from the System units submenu
(→ 65).
22 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters
Decimal places 2
Description Use this function to select the number of decimal places for measured value 2.
Selection •x
• x.x
• x.xx
• x.xxx
• x.xxxx
Value 3 display
Description Use this function to select one of the measured values shown on the local display.
Selection For the picklist, see the Value 2 display parameter (→ 22)
Selection
The unit of the displayed measured value is taken from the System units submenu
(→ 65).
Endress+Hauser 23
Description of device parameters Proline Promass 500 Modbus RS485
0% bargraph 3
Description Use this function to enter the 0% bar graph value to be shown on the display for the
measured value 3.
User entry
The unit of the displayed measured value is taken from the System units submenu
(→ 65).
100% bargraph 3
Description Use this function to enter the 100% bar graph value to be shown on the display for the
measured value 3.
Factory setting 0
User entry
The unit of the displayed measured value is taken from the System units submenu
(→ 65).
24 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters
Decimal places 3
Description Use this function to select the number of decimal places for measured value 3.
Selection •x
• x.x
• x.xx
• x.xxx
• x.xxxx
Value 4 display
Description Use this function to select one of the measured values shown on the local display.
Selection For the picklist, see the Value 2 display parameter (→ 22)
Endress+Hauser 25
Description of device parameters Proline Promass 500 Modbus RS485
Selection
The unit of the displayed measured value is taken from the System units submenu
(→ 65).
Once the measuring device has been enabled for custody transfer mode, depending on the
custody transfer approval selected the display can switch to showing the custody transfer
counter.
For detailed information on custody transfer mode, see the Special Documentation for
the device → 8
Decimal places 4
Description Use this function to select the number of decimal places for measured value 4.
Selection •x
• x.x
• x.xx
• x.xxx
• x.xxxx
Display interval
Description Use this function to enter the length of time the measured values are displayed if the
values alternate on the display.
26 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters
User entry 1 to 10 s
Factory setting 5s
Once the measuring device has been enabled for custody transfer mode, depending on the
custody transfer approval selected the display can switch between showing the relevant
information and the custody transfer counter.
For detailed information on custody transfer mode, see the Special Documentation for
the device → 8
Display damping
Description Use this function to enter a time constant for the reaction time of the local display to
fluctuations in the measured value caused by process conditions.
Endress+Hauser 27
Description of device parameters Proline Promass 500 Modbus RS485
Header
Description Use this function to select the contents of the header of the local display.
1 XXXXXXXXX
A0029422
Selection
• Device tag
Is defined in the Device tag parameter (→ 205).
• Free text
Is defined in the Header text parameter (→ 28).
Header text
Prerequisite In the Header parameter (→ 28), the Free text option is selected.
Description Use this function to enter a customer-specific text for the header of the local display.
User entry Max. 12 characters such as letters, numbers or special characters (e.g. @, %, /)
28 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters
1 XXXXXXXXX
A0029422
User entry
The number of characters displayed depends on the characters used.
Separator
Selection • . (point)
• , (comma)
Contrast display
Description Use this function to enter a value to adapt the display contrast to the ambient conditions
(e.g. the lighting or viewing angle).
User entry 20 to 80 %
Backlight
Endress+Hauser 29
Description of device parameters Proline Promass 500 Modbus RS485
Description Use this function to switch the backlight of the local display on and off.
Selection • Disable
• Enable
‣ Config. backup
Operating time
Description Use this function to display the length of time the device has been in operation.
User interface Days (d), hours (h), minutes (m) and seconds (s)
Last backup
Description Displays the time since a backup copy of the data was last saved to the device memory.
User interface Days (d), hours (h), minutes (m) and seconds (s)
30 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters
Config. managem.
Description Use this function to select an action to save the data to the device memory.
Selection • Cancel
• Execute backup
• Restore *
• Compare *
• Clear backup
Options Description
Execute backup A backup copy of the current device configuration is saved from the HistoROM
backup to the memory of the device. The backup copy includes the transmitter data
of the device.
The following message appears on local display: Backup active, please wait!
Restore The last backup copy of the device configuration is restored from the device
memory to the device's HistoROM backup. The backup copy includes the
transmitter data of the device.
The following message appears on local display: Restore active! Do not interrupt
power supply!
Compare The device configuration saved in the device memory is compared with the current
device configuration of the HistoROM backup.
The following message appears on local display: Comparing files
The result can be viewed in Compar. result parameter.
Clear backup The backup copy of the device configuration is deleted from the memory of the
device.
The following message appears on local display: Deleting file
HistoROM
A HistoROM is a "non-volatile" device memory in the form of an EEPROM.
Backup state
Endress+Hauser 31
Description of device parameters Proline Promass 500 Modbus RS485
• Comp. in progr.
• Restoring failed
• Backup failed
Compar. result
Description Displays the last result of the comparison of the data records in the device memory and in
the HistoROM.
Selection
Options Description
Set. identical The current device configuration of the HistoROM is identical to the backup copy in
the device memory.
If the transmitter configuration of another device has been transmitted to the
device via HistoROM in the Config. managem. parameter, the current device
configuration of the HistoROM is only partially identical to the backup copy in the
device memory: The settings for the transmitter are not identical.
Set. not ident. The current device configuration of the HistoROM is not identical to the backup
copy in the device memory.
No backup There is no backup copy of the device configuration of the HistoROM in the device
memory.
Backup corrupt The current device configuration of the HistoROM is corrupt or not compatible with
the backup copy in the device memory.
Check not done The device configuration of the HistoROM has not yet been compared to the
backup copy in the device memory.
Dataset incomp. The backup copy in the device memory is not compatible with the device.
HistoROM
A HistoROM is a "non-volatile" device memory in the form of an EEPROM.
32 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters
‣ Diagn. handling
‣ Diagn. behavior → 33
Alarm delay
Description Use this function to enter the time interval until the device generates a diagnostic
message.
The diagnostic message is reset without a time delay.
User entry 0 to 60 s
Factory setting 0s
Endress+Hauser 33
Description of device parameters Proline Promass 500 Modbus RS485
The following options are available in the Assign behavior of diagnostic no. xxx
parameters:
Options Description
Alarm The device stops measurement. The measured value output via Modbus RS485 and the
totalizers assume the defined alarm condition. A diagnostic message is generated.
The background lighting changes to red.
Warning The device continues to measure. The measured value output via Modbus RS485 and the
totalizers are not affected. A diagnostic message is generated.
Logbook only The device continues to measure. The diagnostic message is displayed only in the Event
logbook submenu (→ 202) (Event list submenu (→ 203)) and is not displayed in
alternation with the operational display.
Off The diagnostic event is ignored, and no diagnostic message is generated or entered.
For a list of all the diagnostic events, see the Operating Instructions for the device
→ 8
‣ Diagn. behavior
34 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters
Navigation Expert → System → Diagn. handling → Diagn. behavior → Diagnostic no. 046 (0709)
Description Use this function to change the diagnostic behavior of the diagnostic message 046 Sensor
limit.
Selection • Off
• Alarm
• Warning
• Logbook only
Navigation Expert → System → Diagn. handling → Diagn. behavior → Diagnostic no. 140 (0708)
Description Use this function to change the diagnostic behavior of the diagnostic message 140 Sensor
sig.asym..
Endress+Hauser 35
Description of device parameters Proline Promass 500 Modbus RS485
Selection • Off
• Alarm
• Warning
• Logbook only
Navigation Expert → System → Diagn. handling → Diagn. behavior → Diagnostic no. 144 (0731)
Description Use this function to change the diagnostic behavior of the diagnostic message
144 MeasErrorTooHigh.
Selection • Off
• Alarm
• Warning
• Logbook only
Navigation Expert → System → Diagn. handling → Diagn. behavior → Diagnostic no. 374 (0710)
Description Use this function to change the diagnostic behavior of the diagnostic message 374 Sensor
electron..
Selection • Off
• Alarm
• Warning
• Logbook only
36 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters
Navigation Expert → System → Diagn. handling → Diagn. behavior → Diagnostic no. 302 (0739)
Description Option for changing the diagnostic behavior of the diagnostic message 302 Verific. active.
Selection • Alarm
• Warning
Navigation Expert → System → Diagn. handling → Diagn. behavior → Diagnostic no. 441 (0657)
Description Option for changing the diagnostic behavior of the diagnostic message 441 Curr.output
1 to n.
Selection • Off
• Alarm
• Warning
• Logbook only
Navigation Expert → System → Diagn. handling → Diagn. behavior → Diagnostic no. 442 (0658)
Description Option for changing the diagnostic behavior of the diagnostic message 442 Freq. output
1 to n.
Selection • Off
• Alarm
• Warning
• Logbook only
Endress+Hauser 37
Description of device parameters Proline Promass 500 Modbus RS485
Navigation Expert → System → Diagn. handling → Diagn. behavior → Diagnostic no. 443 (0659)
Description Option for changing the diagnostic behavior of the diagnostic message 443 Pulse output
1 to n.
Selection • Off
• Alarm
• Warning
• Logbook only
Navigation Expert → System → Diagn. handling → Diagn. behavior → Diagnostic no. 444 (0740)
Description Option for changing the diagnostic behavior of the diagnostic message 444 Current input
1 to n.
Selection • Off
• Alarm
• Warning
• Logbook only
Navigation Expert → System → Diagn. handling → Diagn. behavior → Diagnostic no. 543 (0643)
Description Option for changing the diagnostic behavior of the diagnostic message 543 Double pulse
out.
Selection • Off
• Alarm
• Warning
• Logbook only
38 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters
Navigation Expert → System → Diagn. handling → Diagn. behavior → Diagnostic no. 830 (0800)
Description Use this function to change the diagnostic behavior of the diagnostic message 830 Sensor
temp..
Selection • Off
• Alarm
• Warning
• Logbook only
Navigation Expert → System → Diagn. handling → Diagn. behavior → Diagnostic no. 831 (0641)
Description Use this function to change the diagnostic behavior of the diagnostic message 831 Sensor
temp..
Selection • Off
• Alarm
• Warning
• Logbook only
Navigation Expert → System → Diagn. handling → Diagn. behavior → Diagnostic no. 832 (0681)
Description Option for changing the diagnostic behavior of the diagnostic message 832 Electronic
temp..
Endress+Hauser 39
Description of device parameters Proline Promass 500 Modbus RS485
Selection • Off
• Alarm
• Warning
• Logbook only
Navigation Expert → System → Diagn. handling → Diagn. behavior → Diagnostic no. 833 (0682)
Description Option for changing the diagnostic behavior of the diagnostic message 833 Electronic
temp..
Selection • Off
• Alarm
• Warning
• Logbook only
Navigation Expert → System → Diagn. handling → Diagn. behavior → Diagnostic no. 834 (0700)
Description Option for changing the diagnostic behavior of the diagnostic message 834 Process temp..
Selection • Off
• Alarm
• Warning
• Logbook only
Navigation Expert → System → Diagn. handling → Diagn. behavior → Diagnostic no. 835 (0702)
Description Option for changing the diagnostic behavior of the diagnostic message 835 Process temp..
40 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters
Selection • Off
• Alarm
• Warning
• Logbook only
Navigation Expert → System → Diagn. handling → Diagn. behavior → Diagnostic no. 842 (0638)
Description Option for changing the diagnostic behavior of the diagnostic message S842 Process
limit.
Selection • Off
• Alarm
• Warning
• Logbook only
Navigation Expert → System → Diagn. handling → Diagn. behavior → Diagnostic no. 862 (0679)
Description Use this function to change the diagnostic behavior of the diagnostic message 862 Empty
pipe.
Selection • Off
• Alarm
• Warning
• Logbook only
Endress+Hauser 41
Description of device parameters Proline Promass 500 Modbus RS485
Navigation Expert → System → Diagn. handling → Diagn. behavior → Diagnostic no. 912 (0703)
Description Use this function to change the diagnostic behavior of the diagnostic message
912 Medium inhomog..
Selection • Off
• Alarm
• Warning
• Logbook only
Navigation Expert → System → Diagn. handling → Diagn. behavior → Diagnostic no. 913 (0712)
Description Use this function to change the diagnostic behavior of the diagnostic message
913 Medium unsuitab..
Selection • Off
• Alarm
• Warning
• Logbook only
Navigation Expert → System → Diagn. handling → Diagn. behavior → Diagnostic no. 941 (0635)
Description Use this function to change the diagnostic behavior of the diagnostic message API
temperature.
Selection • Off
• Alarm
• Warning
• Logbook only
42 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters
Navigation Expert → System → Diagn. handling → Diagn. behavior → Diagnostic no. 942 (0636)
Description Use this function to change the diagnostic behavior of the diagnostic message API
density.
Selection • Off
• Alarm
• Warning
• Logbook only
Navigation Expert → System → Diagn. handling → Diagn. behavior → Diagnostic no. 943 (0637)
Description Use this function to change the diagnostic behavior of the diagnostic message API
pressure.
Selection • Off
• Alarm
• Warning
• Logbook only
Endress+Hauser 43
Description of device parameters Proline Promass 500 Modbus RS485
Navigation Expert → System → Diagn. handling → Diagn. behavior → Diagnostic no. 944 (0732)
Description Use this function to change the diagnostic behavior of the diagnostic message
944 MonitoringFailed.
Selection • Off
• Alarm
• Warning
• Logbook only
Navigation Expert → System → Diagn. handling → Diagn. behavior → Diagnostic no. 948 (0744)
Description Use this function to change the diagnostic behavior of the diagnostic message 948 Oscill.
damping.
Selection • Off
• Alarm
• Warning
• Logbook only
‣ Administration
44 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters
Confirm code → 46
Navigation Expert → System → Administration → Def. access code → Def. access code
Description Use this function to enter a user-specific release code to restrict write-access to the
parameters. This protects the device configuration against any inadvertent modifications
via the local display, Web browser, FieldCare or DeviceCare (via CDI-RJ45 service
interface).
User entry Max. 16-digit character string comprising numbers, letters and special characters
User entry
A message is displayed if the access code is not in the input range.
Factory setting
If the factory setting is not changed or 0 is defined as the access code, the parameters are
not write-protected and the device configuration data can be modified. The user is logged
on in the "Maintenance" role.
Endress+Hauser 45
Description of device parameters Proline Promass 500 Modbus RS485
Confirm code
Description Enter the defined release code a second time to confirm the release code.
User entry Max. 16-digit character string comprising numbers, letters and special characters
Operating time
Navigation Expert → System → Administration → Reset acc. code → Operating time (0652)
Description Use this function to display the length of time the device has been in operation.
User interface Days (d), hours (h), minutes (m) and seconds (s)
Navigation Expert → System → Administration → Reset acc. code → Reset acc. code (0024)
Description Use this function to enter a reset code to reset the user-specific release code to the factory
setting.
User entry Character string comprising numbers, letters and special characters
46 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters
User entry
The reset code can only be entered via:
• Web browser
• DeviceCare, FieldCare (via interface CDI RJ45)
• Fieldbus
Device reset
Description Use this function to choose whether to reset the device configuration - either entirely or in
part - to a defined state.
Selection • Cancel
• To delivery set.
• Restart device
• Rest.S-DATBackup *
Options Description
To delivery set. Every parameter for which a customer-specific default setting was ordered is reset
to this customer-specific value. All other parameters are reset to the factory
setting.
Restart device The restart resets every parameter whose data are in the volatile memory (RAM)
to the factory setting (e.g. measured value data). The device configuration remains
unchanged.
Rest.S-DATBackup Restore the data that are saved on the S-DAT. The data record is restored from the
electronics memory to the S-DAT.
Endress+Hauser 47
Description of device parameters Proline Promass 500 Modbus RS485
Activate SW opt.
Description Use this function to enter an activation code to enable an additional, ordered software
option.
User entry
To activate a software option subsequently, please contact your Endress+Hauser sales
organization.
NOTE!
The activation code is linked to the serial number of the measuring device and varies
according to the device and software option.
If an incorrect or invalid code is entered, this results in the loss of software options that
have already been activated.
▸ Before you enter a new activation code, make a note of the current activation code .
▸ Enter the new activation code provided by Endress+Hauser when the new software
option was ordered.
▸ Once the activation code has been entered, check if the new software option is displayed
in the SW option overv. parameter (→ 49).
The new software option is active if it is displayed.
If the new software option is not displayed or all software options have been deleted,
the code entered was either incorrect or invalid.
▸ If the code entered is incorrect or invalid, enter the old activation code .
▸ Have your Endress+Hauser sales organization check the new activation code
remembering to specify the serial number or ask for the code again.
Web browser
Once a software option has been activated, the page must be loaded again in the Web
browser.
48 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters
SW option overv.
Description Displays all the software options that are enabled in the device.
"Concentration" option
Order code for "Application package", option ED "Concentration" and option EE "Special
density"
"Viscosity" option
Only available for Promass I.
"Petroleum" option
Only available for Promass E, F, O, Q and X.
Endress+Hauser 49
Description of device parameters Proline Promass 500 Modbus RS485
‣ Sensor
‣ Measured val. → 50
‣ System units → 65
‣ Process param. → 82
‣ Calculated value → 95
‣ Measurement mode → 90
‣ External comp. → 92
‣ Sensor adjustm. → 98
‣ Calibration → 106
‣ Measured val.
‣ Process variab. → 50
‣ Totalizer → 58
‣ Input values → 59
‣ Output values → 61
‣ Process variab.
Correct.vol.flow (1851) → 52
50 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters
Density (1850) → 52
Ref.density (1852) → 52
Temperature (1853) → 53
TempCompDynVisc (1872) → 54
TempCompKinVisc (1863) → 54
Concentration (1887) → 55
Targ.corr.vol.fl (1893) → 56
Carr.corr.vol.fl (1894) → 56
Mass flow
Navigation Expert → Sensor → Measured val. → Process variab. → Mass flow (1838)
Endress+Hauser 51
Description of device parameters Proline Promass 500 Modbus RS485
Volume flow
Navigation Expert → Sensor → Measured val. → Process variab. → Volume flow (1847)
Dependency
The unit is taken from the Volume flow unit parameter (→ 67)
Correct.vol.flow
Density
Ref.density
52 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters
Temperature
Pressure value
Navigation Expert → Sensor → Measured val. → Process variab. → Pressure value (6129)
Dynam. viscosity
Navigation Expert → Sensor → Measured val. → Process variab. → Dynam. viscosity (1854)
Endress+Hauser 53
Description of device parameters Proline Promass 500 Modbus RS485
Kinematic visc.
Navigation Expert → Sensor → Measured val. → Process variab. → Kinematic visc. (1857)
TempCompDynVisc
Description Displays the temperature compensation that is currently calculated for the viscosity.
TempCompKinVisc
Description Displays the temperature compensation that is currently calculated for the kinetic
viscosity.
54 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters
Concentration
Navigation Expert → Sensor → Measured val. → Process variab. → Target mass flow (1864)
Description Displays the mass flow that is currently measured for the target medium.
Endress+Hauser 55
Description of device parameters Proline Promass 500 Modbus RS485
Navigation Expert → Sensor → Measured val. → Process variab. → Carrier mass fl. (1865)
Description Displays the mass flow that is currently measured for the carrier medium.
Targ.corr.vol.fl
Description Displays the corrected volume flow that is currently measured for the target fluid.
Carr.corr.vol.fl
Description Displays the corrected volume flow currently measured for the carrier fluid.
56 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters
Navigation Expert → Sensor → Measured val. → Process variab. → Target vol. flow (1895)
Description Displays the volume flow currently measured for the target medium.
Navigation Expert → Sensor → Measured val. → Process variab. → Carrier vol. fl. (1896)
Description Use this function to display the volume flow currently measured for the carrier medium.
Endress+Hauser 57
Description of device parameters Proline Promass 500 Modbus RS485
"Totalizer" submenu
‣ Totalizer
Totalizer val. 1 to n
Prerequisite A process variable is selected in the Assign variable parameter (→ 182) of the
Totalizer 1 to n submenu.
User interface
The value of the process variable totalized since measuring began can be positive or
negative. This depends on the settings in the Operation mode parameter (→ 184).
The unit of the selected process variable is specified for the totalizer in the Unit
totalizer parameter (→ 183).
Example
Calculation of the current totalizer reading when the value exceeds the 7-digit display
range of the operating tool:
• Value in the Totalizer val. 1 parameter: 1 968 457 m3
• Value in the Tot. overflow 1 parameter: 1 ⋅ 107 (1 overflow) = 10 000 000 [m3]
• Current totalizer reading: 11 968 457 m3
Tot. overflow 1 to n
Prerequisite A process variable is selected in the Assign variable parameter (→ 182) of the
Totalizer 1 to n submenu.
58 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters
User interface
The unit of the selected process variable is specified for the totalizer in the Unit
totalizer parameter (→ 183).
Example
Calculation of the current totalizer reading when the value exceeds the 7-digit display
range of the operating tool:
• Value in the Totalizer val. 1 parameter: 1 968 457 m3
• Value in the Tot. overflow 1 parameter: 2 ⋅ 107 (2 overflows) = 20 000 000 [m3]
• Current totalizer reading: 21 968 457 m3
‣ Input values
‣ Current input 1 to n → 59
‣ Val.stat.inp. 1 to n → 60
‣ Current input 1 to n
Endress+Hauser 59
Description of device parameters Proline Promass 500 Modbus RS485
Measured val. 1 to n
Measur. curr. 1 to n
Navigation Expert → Sensor → Measured val. → Input values → Current input 1 to n → Measur.
curr. 1 to n (1604–1 to n)
‣ Val.stat.inp. 1 to n
Val.stat.inp. (1353–1 to n) → 60
Val.stat.inp.
60 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters
‣ Output values
‣ Value curr.out 1 to n → 61
‣ PFS output 1 to n → 62
‣ Relay output 1 to n → 63
‣ Value curr.out 1 to n
Output curr. 1 to n
Navigation Expert → Sensor → Measured val. → Output values → Value curr.out 1 to n → Output
curr. 1 to n (0361–1 to n)
Description Displays the current value currently calculated for the current output.
Measur. curr. 1 to n
Description Use this function to display the actual measured value of the output current.
User interface 0 to 30 mA
Endress+Hauser 61
Description of device parameters Proline Promass 500 Modbus RS485
‣ PFS output 1 to n
Output freq. 1 to n
Navigation Expert → Sensor → Measured val. → Output values → PFS output 1 to n → Output
freq. 1 to n (0471–1 to n)
Prerequisite In the Operating mode parameter (→ 132), the Frequency option is selected.
Description Displays the actual value of the output frequency which is currently measured.
Pulse output 1 to n
Navigation Expert → Sensor → Measured val. → Output values → PFS output 1 to n → Pulse
output 1 to n (0456–1 to n)
Prerequisite The Pulse option is selected in the Operating mode parameter (→ 132) parameter.
62 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters
1 1
0 t 0 t
NC NO
A0028726
0 Non-conductive
1 Conductive
NC NC contact (normally closed)
NO NO contact (normally open)
The output behavior can be reversed via the Invert outp.sig. parameter (→ 152) i.e.
the transistor does not conduct for the duration of the pulse.
In addition, the behavior of the output in the event of a device alarm (Failure mode
parameter (→ 136)) can be configured.
Switch status 1 to n
Navigation Expert → Sensor → Measured val. → Output values → PFS output 1 to n → Switch
status 1 to n (0461–1 to n)
Prerequisite The Switch option is selected in the Operating mode parameter (→ 132).
‣ Relay output 1 to n
Endress+Hauser 63
Description of device parameters Proline Promass 500 Modbus RS485
Switch status
Navigation Expert → Sensor → Measured val. → Output values → Relay output 1 to n → Switch
status (0801–1 to n)
Switch cycles
Navigation Expert → Sensor → Measured val. → Output values → Relay output 1 to n → Switch
cycles (0815–1 to n)
Navigation Expert → Sensor → Measured val. → Output values → Relay output 1 to n → Max.
cycles no. (0817–1 to n)
64 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters
Pulse output
Navigation Expert → Sensor → Measured val. → Output values → Double pulse out → Pulse
output (0987)
Description Displays the pulse frequency of the double pulse output which is currently output.
Additional information For a detailed description and example: Pulse output parameter (→ 62)
‣ System units
Description Use this function to select the unit for the mass flow.
Endress+Hauser 65
Description of device parameters Proline Promass 500 Modbus RS485
Selection
For an explanation of the abbreviated units: → 254
Customer-specific units
The unit for the customer-specific mass is specified in the Mass text parameter
(→ 76).
Mass unit
Description Use this function to select the unit for the mass.
66 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters
Customer-specific units
The unit for the customer-specific mass is specified in the Mass text parameter
(→ 76).
Description Use this function to select the unit for the volume flow.
or
Endress+Hauser 67
Description of device parameters Proline Promass 500 Modbus RS485
Selection
For an explanation of the abbreviated units: → 254
Customer-specific units
The unit for the customer-specific volume is specified in the Volume text parameter
(→ 75).
Volume unit
Description Use this function to select the unit for the volume.
68 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters
or
Customer-specific units
The unit for the customer-specific volume is specified in the Volume text parameter
(→ 75).
Cor.volflow unit
Description Use this function to select the unit for the corrected volume flow.
Endress+Hauser 69
Description of device parameters Proline Promass 500 Modbus RS485
Selection
For an explanation of the abbreviated units: → 254
Description Use this function to select the unit for the corrected volume.
Density unit
Description Use this function to select the unit for the density.
70 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters
or
or
Selection
• SD = specific density
The specific density is the ratio of the medium density to the water density at a water
temperature of +4 °C (+39 °F), +15 °C (+59 °F), +20 °C (+68 °F).
• SG = specific gravity
The specific gravity is the ratio of the medium density to the water density at a water
temperature of +4 °C (+39 °F), +15 °C (+59 °F), +20 °C (+68 °F).
For an explanation of the abbreviated units: → 254
Customer-specific units
The unit for the customer-specific density is specified in the Density text parameter
(→ 78).
Endress+Hauser 71
Description of device parameters Proline Promass 500 Modbus RS485
Description Use this function to select the unit for the reference density.
Selection
For an explanation of the abbreviated units: → 254
Temperature unit
Description Use this function to select the unit for the temperature.
72 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters
Selection
For an explanation of the abbreviated units: → 254
Pressure unit
Description Use this function to select the unit for the pipe pressure.
Selection
For an explanation of the abbreviated units: → 254
Date/time format
Description Use this function to select the desired time format for calibration history.
Endress+Hauser 73
Description of device parameters Proline Promass 500 Modbus RS485
‣ User-spec. units
74 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters
Volume text
Navigation Expert → Sensor → System units → User-spec. units → Volume text (0567)
Description Use this function to enter a text for the user-specific unit of volume and volume flow. The
corresponding time units (s, min, h, d) for volume flow are generated automatically.
User entry Max. 10 characters such as letters, numbers or special characters (@, %, /)
Example
If the text GLAS is entered, the choose list of the Volume flow unit parameter (→ 67)
shows the following options:
• GLAS/s
• GLAS/min
• GLAS/h
• GLAS/d
Volume offset
Navigation Expert → Sensor → System units → User-spec. units → Volume offset (0569)
Description Use this function to enter the offset for adapting the user-specific volume unit and volume
flow unit (without time).
Factory setting 0
Endress+Hauser 75
Description of device parameters Proline Promass 500 Modbus RS485
Volume factor
Navigation Expert → Sensor → System units → User-spec. units → Volume factor (0568)
Description Use this function to enter a quantity factor (without time) for the user-specific volume and
volume flow unit.
Mass text
Navigation Expert → Sensor → System units → User-spec. units → Mass text (0560)
Description Use this function to enter a text for the user-specific unit of mass and mass flow. The
corresponding time units (s, min, h, d) for mass flow are generated automatically.
User entry Max. 10 characters such as letters, numbers or special characters (@, %, /)
Example
If the text CENT for "centner" is entered, the following options are displayed in the picklist
for the Mass flow unit parameter (→ 65):
• CENT/s
• CENT/min
• CENT/h
• CENT/d
Mass offset
Navigation Expert → Sensor → System units → User-spec. units → Mass offset (0562)
Description Use this function to enter the zero point shift for the user-specific mass and mass flow
unit.
Factory setting 0
76 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters
Mass factor
Navigation Expert → Sensor → System units → User-spec. units → Mass factor (0561)
Description Use this function to enter a quantity factor (without time) for the user-specific mass and
mass flow unit.
Navigation Expert → Sensor → System units → User-spec. units → Corr. vol. text (0592)
Description Use this function to enter a text for the user-specific unit of the corrected volume and
corrected volume flow. The corresponding time units (s, min, h, d) for mass flow are
generated automatically.
User entry Max. 10 characters such as letters, numbers or special characters (@, %, /)
Example
If the text GLAS is entered, the choose list of the Cor.volflow unit parameter (→ 69)
shows the following options:
• GLAS/s
• GLAS/min
• GLAS/h
• GLAS/d
Endress+Hauser 77
Description of device parameters Proline Promass 500 Modbus RS485
Navigation Expert → Sensor → System units → User-spec. units → Corr vol. offset (0602)
Description Use this function to enter the offset for adapting the user-specific corrected volume unit
and corrected volume flow unit (without time).
Value in user-specific unit = (factor × value in base unit) + offset
Factory setting 0
Cor.vol. factor
Navigation Expert → Sensor → System units → User-spec. units → Cor.vol. factor (0590)
Description Use this function to enter a quantity factor (without time) for the user-specific corrected
volume unit and corrected volume flow unit.
Density text
Navigation Expert → Sensor → System units → User-spec. units → Density text (0570)
Description Use this function to enter a text or the user-specific unit of density.
User entry Max. 10 characters such as letters, numbers or special characters (@, %, /)
Example
Enter text “CE_L” for centners per liter
78 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters
Density offset
Navigation Expert → Sensor → System units → User-spec. units → Density offset (0571)
Description Use this function to enter the zero point shift for the user-specific density unit.
Value in user-specific unit = (factor × value in basic unit) + offset
Factory setting 0
Density factor
Navigation Expert → Sensor → System units → User-spec. units → Density factor (0572)
Description Use this function to enter a quantity factor for the user-specific density unit.
Navigation Expert → Sensor → System units → User-spec. units → Spec. enth. text (0585)
User entry Max. 10 characters such as letters, numbers or special characters (@, %, /)
Example
If the text CAL is entered, the choose list of the Cal. value unit parameter shows the
following options:
• CAL/Nm3
• CAL/m3
• CAL/ft3
• CAL/Sft3
Endress+Hauser 79
Description of device parameters Proline Promass 500 Modbus RS485
Navigation Expert → Sensor → System units → User-spec. units → Spec. enth. off. (0584)
Description Use this function to enter the offset for adapting the user-specific calorific value unit
(without volume).
Factory setting 0
Navigation Expert → Sensor → System units → User-spec. units → Spec. enth. fac. (0583)
Description Use this function to enter a quantity factor (without volume) for the user-specific calorific
value unit.
Energy text
Navigation Expert → Sensor → System units → User-spec. units → Energy text (0600)
Description Use this function to enter a text for the user-specific energy unit.
User entry Max. 10 characters such as letters, numbers or special characters (@, %, /)
Example
If the text W is entered, the choose list of the Energy flow unit parameter shows the
following options:
• W/s
• W/min
• W/h
• W/d
80 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters
Energy offset
Navigation Expert → Sensor → System units → User-spec. units → Energy offset (0599)
Description Use this function to enter the offset for adapting the user-specific energy unit (without
time).
Factory setting 0
Energy factor
Navigation Expert → Sensor → System units → User-spec. units → Energy factor (0586)
Description Use this function to enter a quantity factor for the user-specific energy unit.
Pressure text
Navigation Expert → Sensor → System units → User-spec. units → Pressure text (0581)
Description Use this function to enter a text for the user-specific pressure unit.
User entry Max. 10 characters such as letters, numbers or special characters (@, %, /)
Pressure offset
Navigation Expert → Sensor → System units → User-spec. units → Pressure offset (0580)
Description Use this function to enter the offset for adapting the user-specific pressure unit.
Factory setting 0
Endress+Hauser 81
Description of device parameters Proline Promass 500 Modbus RS485
Pressure factor
Navigation Expert → Sensor → System units → User-spec. units → Pressure factor (0579)
Description Use this function to enter a quantity factor for the user-specific pressure unit.
‣ Process param.
Flow damping
Description Use this function to enter a time constant for flow damping (PT1 element). Reduction of
the variability of the flow measured value (in relation to interference). For this purpose,
the depth of the flow filter is adjusted: when the filter setting increases, the reaction time
of the device also increases.
Factory setting 0s
82 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters
User entry
• Value = 0: no damping
• Value > 0: damping is increased
Damping is switched off if 0 is entered (factory setting).
Result
The damping affects the following variables of the device:
• Outputs → 115
• Low flow cut off → 84
• Totalizers → 182
Density damping
Description Use this function to enter a time constant for the damping (PT1 element) of the density
measured value.
Factory setting 0s
User entry
• Value = 0: no damping
• Value > 0: damping is increased
Damping is switched off if 0 is entered (factory setting).
Temp. damping
Description Use this function to enter a time constant for the damping (PT1 element) of the
temperature measured value.
Endress+Hauser 83
Description of device parameters Proline Promass 500 Modbus RS485
Factory setting 0s
User entry
• Value = 0: no damping
• Value > 0: damping is increased
Damping is switched off if 0 is entered (factory setting).
Flow override
Description Use this function to select whether to interrupt the evaluation of measured values. This is
useful for the cleaning processes of a pipeline, for example.
Selection • Off
• On
On value (1805) → 85
84 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters
Assign variable
Navigation Expert → Sensor → Process param. → Low flow cut off → Assign variable (1837)
Description Use this function to select the process variable for low flow cutoff detection.
Selection • Off
• Mass flow
• Volume flow
• Correct.vol.flow *
On value
Navigation Expert → Sensor → Process param. → Low flow cut off → On value (1805)
Description Use this function to enter a switch-on value for low flow cut off. Low flow cut off is
activated if the value entered is not equal to 0 → 85.
Off value
Navigation Expert → Sensor → Process param. → Low flow cut off → Off value (1804)
Description Use this function to enter a switch-off value for low flow cut off. The off value is entered as
a positive hysteresis from the on value→ 85.
Endress+Hauser 85
Description of device parameters Proline Promass 500 Modbus RS485
Factory setting 50 %
Q 2 2
4
H
3
t
1 A 1 A
A0012887
Q Flow
t Time
H Hysteresis
A Low flow cut off active
1 Low flow cut off is activated
2 Low flow cut off is deactivated
3 On value entered
4 Off value entered
Navigation Expert → Sensor → Process param. → Low flow cut off → Pres. shock sup. (1806)
Description Use this function to enter the time interval for signal suppression (= active pressure shock
suppression).
Factory setting 0s
86 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters
Example
When closing a valve, momentarily strong fluid movements may occur in the pipeline,
which are registered by the measuring system. These totalized flow values lead to a false
totalizer status, particularly during batching processes.
Q D C D
1 2 3 4
6
5
t
A B
A0012888
Q Flow
t Time
A Drip
B Pressure shock
C Pressure shock suppression active as specified by the time entered
D Pressure shock suppression inactive
1 Valve closes
2 Flow falls below the on-value of the low flow cut off: pressure shock suppression is activated
3 The time entered has elapsed: pressure shock suppression is deactivated
4 The actual flow value is now displayed and output
5 On-value for low flow cut off
6 Off-value for low flow cut off
Endress+Hauser 87
Description of device parameters Proline Promass 500 Modbus RS485
Assign variable
Navigation Expert → Sensor → Process param. → Partial pipe det → Assign variable (1860)
Description Use this function to select a process variable to detect empty or partially filled measuring
tubes.
For gas measurement: Deactivate monitoring due to low gas density.
Selection • Off
• Density
• Ref.density
Low value
Navigation Expert → Sensor → Process param. → Partial pipe det → Low value (1861)
Description Use this function to enter a lower limit value to enable detection of empty or partially filled
measuring tubes. If the measured density falls below this value, monitoring is enabled.
Limit value
If the displayed value is outside the limit value, the measuring device displays the
diagnostic message S862 Partly filled.
88 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters
High value
Navigation Expert → Sensor → Process param. → Partial pipe det → High value (1858)
Description Use this function to enter an upper limit value to enable detection of empty or partially
filled measuring tubes. If the measured density exceeds this value, detection is enabled.
Limit value
If the displayed value is outside the limit value, the measuring device displays the
diagnostic message S862 Partly filled.
Response time
Navigation Expert → Sensor → Process param. → Partial pipe det → Response time (1859)
Description Use this function to enter the minimum length of time (debouncing time) the signal must
be present for the diagnostic message S862 Partly filled to be triggered if the
measuring pipe is empty or partially full.
Factory setting 1s
Max. damping
Navigation Expert → Sensor → Process param. → Partial pipe det → Max. damping (6040)
Description Use this function to enter a damping value to enable detection of empty or partially filled
measuring tubes.
Factory setting 0
Endress+Hauser 89
Description of device parameters Proline Promass 500 Modbus RS485
User entry
• Damping is disabled if 0 is entered (factory setting).
• Damping is enabled if the value entered is greater than 0.
• The value entered depends on application-specific influence variables, such as the
medium, nominal diameter, sensor etc.
Example
• If the pipe is filled normally the value of the oscillation damping is 500.
• If the pipe is partially filled the value of the oscillation damping is > 5000.
• A practical damping value would then be 2000: enter 2000 as the value.
‣ Measurement mode
Select medium
Selection • Liquid
• Gas
90 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters
Prerequisite The Gas option is selected in the Select medium parameter (→ 90).
Description Use this function to select the type of gas for the measuring application.
Selection • Air
• Ammonia NH3
• Argon Ar
• Sulf. hex.fl.SF6
• Oxygen O2
• Ozone O3
• Nitrog. ox. NOx
• Nitrogen N2
• Nitrous ox. N2O
• Methane CH4
• Hydrogen H2
• Helium He
• Hydrog.chlor.HCl
• Hydrog.sulf. H2S
• Ethylene C2H4
• Carbon diox. CO2
• Carbon monox. CO
• Chlorine Cl2
• Butane C4H10
• Propane C3H8
• Propylene C3H6
• Ethane C2H6
• Others
Sound velocity
Prerequisite In the Select gas type parameter (→ 91), the Others option is selected.
Description Use this function to enter the sound velocity of the gas at 0 °C (+32 °F).
Endress+Hauser 91
Description of device parameters Proline Promass 500 Modbus RS485
Temp. coeff. SV
Prerequisite The Others option is selected in the Select gas type parameter (→ 91).
Description Use this function to enter a temperature coefficient for the sound velocity of the gas.
Description Activates the Gas Fraction Handler function for two phase media.
Selection • Off
• Moderate
• Powerful
‣ External comp.
Temp.corr.source (6184) → 94
92 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters
Pressure compen.
Selection • Off
• Fixed value
• External value
• Current input 1 *
• Current input 2 *
• Current input 3 *
Pressure value
Prerequisite The Fixed value option or the Current input 1…n option is selected in the Pressure
compen. parameter (→ 93).
Description Use this function to enter a value for the process pressure that is used for pressure
correction.
External press.
Prerequisite The Fixed value option or the Current input 1…n option is selected in the Pressure
compen. parameter (→ 93).
Endress+Hauser 93
Description of device parameters Proline Promass 500 Modbus RS485
Temp.corr.source
Selection
All the options available for selection are used for measured value compensation.
• Internal value
The temperature value measured internally (temperature sensor of the measuring
sensor) is used for compensation.
• Current input 1 optionVisibility depends on order options or device settings, Current
input 2 optionVisibility depends on order options or device settings ,Current input 3
optionVisibility depends on order options or device settings
The temperature value read in via the current input is used for compensation.
External temp.
Prerequisite The External value option or the Current input 1…n option is selected in the
Temperature mode parameter (→ 94).
94 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters
Temperature mode
‣ Calculated value
‣ Corr. vol.flow. → 95
‣ Corr. vol.flow.
Corr. vol.flow. → 96
Ext. ref.density → 96
Fix ref.density → 97
Ref. temperature → 97
Endress+Hauser 95
Description of device parameters Proline Promass 500 Modbus RS485
Corr. vol.flow.
Navigation Expert → Sensor → Calculated value → Corr. vol.flow. → Corr. vol.flow. (1812)
Description Use this function to select the reference density for calculating the corrected volume flow.
Ext. ref.density
Navigation Expert → Sensor → Calculated value → Corr. vol.flow. → Ext. ref.density (6198)
Prerequisite In the Corr. vol.flow. parameter (→ 96), the Ext. ref.density option is selected.
Description Displays the reference density which is read in externally, e.g. via the current input.
96 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters
Fix ref.density
Navigation Expert → Sensor → Calculated value → Corr. vol.flow. → Fix ref.density (1814)
Prerequisite The Fix ref.density option is selected in the Corr. vol.flow. parameter (→ 96)
parameter.
Description Use this function to enter a fixed value for the reference density.
Ref. temperature
Navigation Expert → Sensor → Calculated value → Corr. vol.flow. → Ref. temperature (1816)
Prerequisite The Calc ref density option is selected in the Corr. vol.flow. parameter (→ 96)
parameter.
Description Use this function to enter a reference temperature for calculating the reference density.
ρn = ρ ⋅ (1 + α ⋅ ∆t + β ⋅ ∆t2)
A0023403
Endress+Hauser 97
Description of device parameters Proline Promass 500 Modbus RS485
Navigation Expert → Sensor → Calculated value → Corr. vol.flow. → Linear exp coeff (1817)
Prerequisite The Calc ref density option is selected in the Corr. vol.flow. parameter (→ 96)
parameter.
Description Use this function to enter a linear, fluid-specific expansion coefficient for calculating the
reference density.
Navigation Expert → Sensor → Calculated value → Corr. vol.flow. → Square exp coeff (1818)
Prerequisite The Calc ref density option is selected in the Corr. vol.flow. parameter (→ 96)
parameter.
Description For fluid with a non-linear expansion pattern: use this function to enter a quadratic, fluid-
specific expansion coefficient for calculating the reference density.
‣ Sensor adjustm.
98 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters
Install. direct.
Description Use this function to change the sign of the medium flow direction.
Description Use this function to enter the roll installation angle in degrees.
Factory setting 0°
Inst.angle pitch
Description Use this function to enter the installation angle pitch in degrees.
Factory setting 0°
Endress+Hauser 99
Description of device parameters Proline Promass 500 Modbus RS485
Pitch angle
• The pitch angle is the angle α from the horizontal H to align the central axis Z of the
measuring device.
• The pitch angle can be between –90 to +90 °.
Navigation Expert → Sensor → Sensor adjustm. → Zero point adj. → Zero point adj. (6196)
Expert → Sensor → Sensor adjustm. → Zero point adj. → Zero point adj. (6196)
Description Use this function to select the start of the zero point adjustment.
Observe conditions → 100.
Selection • Cancel
• Start
100 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters
Zero pt adj.stat
Navigation Expert → Sensor → Sensor adjustm. → Zero point adj. → Zero pt adj.stat (6253)
Expert → Sensor → Sensor adjustm. → Zero point adj. → Zero pt adj.stat (6253)
Factory setting Ok
Progress
Navigation Expert → Sensor → Sensor adjustm. → Zero point adj. → Progress (2808)
‣ Variable adjust
Endress+Hauser 101
Description of device parameters Proline Promass 500 Modbus RS485
Navigation Expert → Sensor → Sensor adjustm. → Variable adjust → Mass flow offset (1831)
Description Use this function to enter the zero point shift for the mass flow trim. The mass flow unit
on which the shift is based is kg/s.
Navigation Expert → Sensor → Sensor adjustm. → Variable adjust → Mass flow factor (1832)
Description Use this function to enter a quantity factor (without time) for the mass flow. This
multiplication factor is applied over the mass flow range.
Factory setting 1
102 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters
Navigation Expert → Sensor → Sensor adjustm. → Variable adjust → Vol. flow offset (1841)
Description Use this function to enter the zero point shift for the volume flow trim. The volume flow
unit on which the shift is based is m³/s.
Navigation Expert → Sensor → Sensor adjustm. → Variable adjust → Vol. flow factor (1846)
Description Use this function to enter a quantity factor (without time) for the volume flow. This
multiplication factor is applied over the volume flow range.
Factory setting 1
Density offset
Navigation Expert → Sensor → Sensor adjustm. → Variable adjust → Density offset (1848)
Description Use this function to enter the zero point shift for the density trim. The density unit on
which the shift is based is kg/m³.
Endress+Hauser 103
Description of device parameters Proline Promass 500 Modbus RS485
Density factor
Navigation Expert → Sensor → Sensor adjustm. → Variable adjust → Density factor (1849)
Description Use this function to enter a quantity factor for the density. This multiplication factor is
applied over the density range.
Factory setting 1
Navigation Expert → Sensor → Sensor adjustm. → Variable adjust → Corr. vol offset (1866)
Description Use this function to enter the zero point shift for the corrected volume flow trim. The
corrected volume flow unit on which the shift is based is 1 Nm³/s.
Navigation Expert → Sensor → Sensor adjustm. → Variable adjust → Corr. vol factor (1867)
Description Use this function to enter a quantity factor (without time) for the corrected volume flow.
This multiplication factor is applied over the corrected volume flow range.
Factory setting 1
104 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters
Ref.dens. offset
Navigation Expert → Sensor → Sensor adjustm. → Variable adjust → Ref.dens. offset (1868)
Description Use this parameter to enter the zero point shift for the reference density trim. The
reference density unit on which the shift is based is 1 kg/Nm³.
Ref.dens. factor
Navigation Expert → Sensor → Sensor adjustm. → Variable adjust → Ref.dens. factor (1869)
Description Use this function to enter a quantity factor (without time) for the reference density. This
multiplication factor is applied over the reference density range.
Factory setting 1
Temp. offset
Navigation Expert → Sensor → Sensor adjustm. → Variable adjust → Temp. offset (1870)
Description Use this function to enter the zero point shift for the temperature trim. The temperature
unit on which the shift is based is K.
Factory setting 0K
Endress+Hauser 105
Description of device parameters Proline Promass 500 Modbus RS485
Temp. factor
Navigation Expert → Sensor → Sensor adjustm. → Variable adjust → Temp. factor (1871)
Description Use this function to enter a quantity factor for the temperature. In each case, this factor
refers to the temperature in K.
Factory setting 1
‣ Calibration
C0 to 5 (6022) → 107
Cal. factor
106 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters
Zero point
Description Use this function to enter the zero point correction value for the sensor.
Nominal diameter
C0 to 5
Factory setting 0
‣ I/O config.
Endress+Hauser 107
Description of device parameters Proline Promass 500 Modbus RS485
I/O 1 to n terminals
I/O 1 to n info
"Invalid" option
The I/O module is not plugged correctly.
"Configurable" option
The I/O module is configurable.
"MODBUS" option
The I/O module is configured for Modbus.
108 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters
I/O 1 to n type
Description Use this function to select the I/O module type for the configuration of the I/O module.
Selection • Off
• Curr.output *
• Current input *
• Status input *
• PFS output *
• Double pulse out *
• Relay output *
Description Use this function to activate the newly configured I/O module type.
Selection • No
• Yes
Factory setting No
I/O alterat.code
Description Use this function to enter the ordered activation code to activate the I/O configuration
change.
Factory setting 0
Endress+Hauser 109
Description of device parameters Proline Promass 500 Modbus RS485
‣ Input
‣ Current input 1 to n
Terminal no.
Description Displays the terminal numbers used by the current input module.
110 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters
Signal mode
Prerequisite The measuring device is not approved for use in the hazardous area with type of
protection Ex-i.
Description Use this function to select the signal mode for the current input.
Selection • Passive
• Active *
Current span
Description Use this function to select the current range for the process value output and the upper
and lower level for signal on alarm.
Selection • 4...20 mA
• 4...20 mA NAMUR
• 4...20 mA US
• 0...20 mA
Endress+Hauser 111
Description of device parameters Proline Promass 500 Modbus RS485
0/4 mA value
Factory setting 0
Configuration examples
Pay attention to the configuration examples for 4 mA value parameter (→ 120).
20 mA value
Failure mode
Description Use this function to select the input behavior when measuring a current outside the
configured Current span parameter (→ 111).
Selection • Alarm
• Last valid value
• Defined value
112 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters
Failure value
Prerequisite In the Failure mode parameter (→ 112), the Defined value option is selected.
Description Use this function to enter the value that the device uses if it does not receive an input
signal from the external device, or if the input signal is invalid.
Factory setting 0
‣ Status input 1 to n
Terminal no.
Description Displays the terminal numbers used by the status input module.
Endress+Hauser 113
Description of device parameters Proline Promass 500 Modbus RS485
Assign stat.inp.
Description Use this function to select the function for the status input.
Selection • Off
• Reset totaliz. 1
• Reset totaliz. 2
• Reset totaliz. 3
• Reset all tot.
• Flow override
• Zero point adj.
NOTE!
Before enabling the measuring device for custody transfer mode, make sure that the Off
option is selected in the Assign stat.inp..
For detailed information on custody transfer mode, see the Special Documentation for
the device → 8
114 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters
Val.stat.inp.
Active level
Description Use this function to determine the input signal level at which the assigned function is
activated.
Selection • High
• Low
Response time
Description Use this function to enter the minimum time period for which the input signal level must
be present before the selected function is activated.
Factory setting 50 ms
‣ Output
‣ Curr.output 1 to n → 116
Endress+Hauser 115
Description of device parameters Proline Promass 500 Modbus RS485
‣ Curr.output 1 to n
Terminal no.
Description Displays the terminal numbers used by the current output module.
116 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters
Signal mode
Description Use this function to select the signal mode for the current output.
Selection • Active *
• Passive *
Assign curr. 1 to n
Description Use this function to select a process variable for the current output.
Detailed description of the options Oscil. frequency, Oscil. amplitude, Oscil.
damping and Signal asymmetry: Value 1 display parameter (→ 19)
Selection • Off *
• Mass flow
• Volume flow
• Correct.vol.flow *
• Target mass flow *
• Carrier mass fl. *
• Target vol. flow *
• Carrier vol. fl. *
• Targ.corr.vol.fl *
• Carr.corr.vol.fl *
• Density
• Ref.density *
• Ref.dens.altern. *
• GSV flow *
• GSVa *
• NSV flow *
• NSVa *
• S&W volume flow *
• Water cut *
• Oil density *
• Water density *
Endress+Hauser 117
Description of device parameters Proline Promass 500 Modbus RS485
Current span
Description Use this function to select the current range for the process value output and the upper
and lower level for signal on alarm.
118 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters
Example
Shows the relationship between the current span for the output of the process variable and
the lower and upper alarm levels:
2 1 3
I [mA]
A0034351
Selection
Selection 1 2 3
If the flow exceeds or falls below the upper or lower signal on alarm level, the
diagnostic message S441 Curr.output 1 to n is displayed.
Fixed current
Prerequisite The Fixed current option is selected in the Current span parameter (→ 118).
Description Use this function to enter a constant current value for the current output.
Endress+Hauser 119
Description of device parameters Proline Promass 500 Modbus RS485
0/4 mA value
Prerequisite In the Current span parameter (→ 118), one of the following options is selected:
• 4...20 mA NAMUR
• 4...20 mA US
• 4...20 mA
• 0...20 mA
Description Use this function to enter a value for the 0/4 mA current.
Dependency
The unit depends on the process variable selected in the Assign curr. parameter
(→ 117).
Configuration examples
Some examples of parameter settings and their effect on the current output are given in
the following section.
Configuration example A
Measuring mode with Forward flow option
• 0/4 mA value parameter (→ 120) = not equal to zero flow (e.g. –250 m3/h)
• 20 mA value parameter (→ 121) = not equal to zero flow (e.g. +750 m3/h)
• Calculated current value = 8 mA at zero flow
I [mA]
20
8
4
Q
-250 0 +750
1 1
A0013757
Q Flow
I Current
1 Measuring range is exceeded or undershot
120 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters
The operational range of the measuring device is defined by the values entered for the 0/4
mA value parameter (→ 120) and 20 mA value parameter (→ 121). If the
effective flow exceeds or falls below this operational range, the diagnostic message
S441 Curr.output 1 to n is displayed.
Configuration example B
Measuring mode with Forward/Reverse option
I [mA]
20
4
0 Q
0
3 1 2
A0013758
I Current
Q Flow
1 Value assigned to the 0/4 mA current
2 Forward flow
3 Reverse flow
The current output signal is independent of the direction of flow (absolute amount of the
measured variable). The values for the 0/4 mA value parameter (→ 120) and 20 mA
value parameter (→ 121) must have the same sign. The value for the 20 mA value
parameter (→ 121) (e.g. reverse flow) corresponds to the mirrored value for the 20
mA value parameter (→ 121) (e.g. forward flow).
Configuration example C
Measuring mode with Rev. flow comp. option
If flow is characterized by severe fluctuations (e.g. when using reciprocating pumps), flow
components outside the measuring range are buffered, balanced and output after a
maximum delay of 60 s → 122.
20 mA value
Prerequisite One of the following options is selected in the Current span parameter (→ 118):
• 4...20 mA NAMUR
• 4...20 mA US
• 4...20 mA
• 0...20 mA
Endress+Hauser 121
Description of device parameters Proline Promass 500 Modbus RS485
smaller than the value assigned for the 0/4 mA current in the 0/4 mA value parameter
(→ 120).
Dependency
The unit depends on the process variable selected in the Assign curr. parameter
(→ 117).
Example
• Value assigned to 0/4 mA = –250 m3/h
• Value assigned to 20 mA = +750 m3/h
• Calculated current value = 8 mA (at zero flow)
If the Forward/Reverse option is selected in the Measuring mode parameter
(→ 122), different signs cannot be entered for the values of the 0/4 mA value
parameter (→ 120) and 20 mA value parameter (→ 121). The diagnostic message
S441 Curr.output 1 to n is displayed.
Configuration examples
Observe the configuration examples for the 0/4 mA value parameter (→ 120).
Measuring mode
Prerequisite In the Assign curr. parameter (→ 117), one of the following options is selected:
• Mass flow
• Volume flow
• Correct.vol.flow
• Target mass flow *
• Carrier mass fl. *
• Density
• Ref.density
• Concentration *
• Dynam. viscosity *
• Kinematic visc. *
• TempCompDynVisc *
• TempCompKinVisc *
• Temperature
• Carr. pipe temp. *
• Electronic temp.
• Osc. freq. 0
• Osc. freq. 1 *
• Osc. ampl. 0 *
• Osc. ampl. 1 *
• Freq. fluct. 0
• Freq. fluct. 1 *
• Osc. damping 0
• Osc. damping 1 *
• Osc.damp.fluct 0
• Osc.damp.fluct 1 *
• Signal asymmetry
122 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters
• Exc. current 0
• Exc. current 1 *
• HBSI *
Detailed description of the options Oscil. frequency, Oscil. amplitude, Oscil.
damping and Signal asymmetry: Value 1 display parameter (→ 19)
In the Current span parameter (→ 118), one of the following options is selected:
• 4...20 mA NAMUR
• 4...20 mA US
• 4...20 mA
• 0...20 mA
Description Use this function to select the measuring mode for the current output.
"Forward/Reverse" option
I [mA]
20
4
0 Q
0
3 1 2
A0013758
I Current
Q Flow
1 Value assigned to the 0/4 mA current
2 Forward flow
3 Reverse flow
Endress+Hauser 123
Description of device parameters Proline Promass 500 Modbus RS485
• The current output signal is independent of the direction of flow (absolute amount of the
measured variable). The values for the 0/4 mA value parameter (→ 120) and 20 mA
value parameter (→ 121) must have the same sign.
• The value for the 20 mA value parameter (→ 121) (e.g. reverse flow) corresponds to
the mirrored value for the 20 mA value parameter (→ 121) (e.g. forward flow).
I [mA]
20
4
0 Q
1 2
A0028084
2 Measuring range
I Current
Q Flow
1 Lower range value (value assigned to 0/4 mA current)
2 Upper range value (value assigned to 20 mA current)
A0028091
3 Flow response
Q Flow
t Time
124 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters
I [mA]
t
A0028092
I Current
t Time
I [mA]
A0028093
I Current
t Time
I [mA]
A
t
S
S=A
A0028094
I Current
t Time
S Flow components saved
A Balancing of saved flow components
Example 2
Defined measuring range: lower range value and upper range value with different signs
I [mA]
20
4
0 Q
0
1 2
A0028095
4 Measuring range
I Current
Q Flow
1 Lower range value (value assigned to 0/4 mA current)
2 Upper range value (value assigned to 20 mA current)
Endress+Hauser 125
Description of device parameters Proline Promass 500 Modbus RS485
Q
1
+
0 t
-
2
a=
b=
A0028098
Q Flow
t Time
1 Lower range value (value assigned to 0/4 mA current)
2 Upper range value (value assigned to 20 mA current)
I [mA]
t
a=
b=
A0028100
I Current
t Time
I [mA]
t
S
S=A
A0028101
I Current
t Time
S Flow components saved
A Balancing of saved flow components
126 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters
Damping out. 1 to n
Prerequisite A process variable is selected in the Assign curr. parameter (→ 117) and one of the
following options is selected in the Current span parameter (→ 118):
• 4...20 mA NAMUR
• 4...20 mA US
• 4...20 mA
• 0...20 mA
Description Use this function to enter a time constant for the reaction time of the current output signal
to fluctuations in the measured value caused by process conditions.
Response time
Prerequisite One of the following options is selected in the Assign curr. parameter (→ 117):
• Mass flow
• Volume flow
• Correct.vol.flow
• Target mass flow *
• Carrier mass fl. *
• Density
• Ref.density
• Concentration *
• Dynam. viscosity *
• Kinematic visc. *
• TempCompDynVisc *
• TempCompKinVisc *
• Temperature
• Carr. pipe temp. *
• Electronic temp.
• Osc. freq. 0
• Osc. freq. 1 *
Endress+Hauser 127
Description of device parameters Proline Promass 500 Modbus RS485
• Osc. ampl. 0 *
• Osc. ampl. 1 *
• Freq. fluct. 0
• Freq. fluct. 1 *
• Osc. damping 0
• Osc. damping 1 *
• Osc.damp.fluct 0
• Osc.damp.fluct 1 *
• Signal asymmetry
• Exc. current 0
• Exc. current 1 *
• HBSI *
Detailed description of the options Oscil. frequency, Oscil. amplitude, Oscil.
damping and Signal asymmetry: Value 1 display parameter (→ 19)
One of the following options is selected in the Current span parameter (→ 118):
• 4...20 mA NAMUR
• 4...20 mA US
• 4...20 mA
• 0...20 mA
Description Displays the response time. This specifies how quickly the current output reaches the
measured value change of 63 % of 100 % of the measured value change.
Failure mode
Prerequisite A process variable is selected in the Assign curr. parameter (→ 117) and one of the
following options is selected in the Current span parameter (→ 118):
• 4...20 mA NAMUR
• 4...20 mA US
• 4...20 mA
• 0...20 mA
Description Use this function to select the value of the current output in the event of a device alarm.
128 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters
Selection • Min.
• Max.
• Last valid value
• Actual value
• Defined value
"Min." option
The current output adopts the value of the lower level for signal on alarm.
The signal on alarm level is defined via the Current span parameter (→ 118).
"Max." option
The current output adopts the value of the upper level for signal on alarm.
The signal on alarm level is defined via the Current span parameter (→ 118).
Failure current
Prerequisite The Defined value option is selected in the Failure mode parameter (→ 128).
Description Use this function to enter a fixed value that the current output adopts in the event of a
device alarm.
Endress+Hauser 129
Description of device parameters Proline Promass 500 Modbus RS485
Output curr. 1 to n
Description Displays the current value currently calculated for the current output.
Measur. curr. 1 to n
Description Use this function to display the actual measured value of the output current.
User interface 0 to 30 mA
‣ PFS output 1 to n
130 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters
Terminal no.
Description Displays the terminal numbers used by the pulse/frequency/switch output module.
Endress+Hauser 131
Description of device parameters Proline Promass 500 Modbus RS485
Signal mode
Description Use this function to select the signal mode for the pulse/frequency/switch output.
Selection • Passive
• Active
• Passive NAMUR
Operating mode
Description Use this function to select the operating mode of the output as a pulse, frequency or switch
output.
Selection • Pulse
• Frequency
• Switch
132 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters
U[V] B<P
B
24
0 t
P
A0026883
"Frequency" option
Flow-proportional frequency output with 1:1 on/off ratio
An output frequency is output that is proportional to the value of a process variable, such
as mass flow, volume flow, corrected volume flow, target mass flow, carrier mass flow,
density, reference density, concentration, dynamic viscosity, kinematic viscosity,
temperature-compensated dynamic viscosity, temperature-compensated kinematic
viscosity, temperature, carrier tube temperature, electronic temperature, vibration
frequency, frequency fluctuation, oscillation amplitude, oscillation damping, oscillation
damping fluctuation, signal asymmetry or excitation current.
Example
• Flow rate approx. 100 g/s
• Max. frequency 10 kHz
• Flow rate at max. frequency 1 000 g/s
• Output frequency approx. 1 000 Hz
U[V]
24
0 t
A0026886
"Switch" option
Contact for displaying a condition (e.g. alarm or warning if a limit value is reached)
Example
Alarm response without alarm
U[V]
24
0 t
A0026884
Example
Alarm response in case of alarm
Endress+Hauser 133
Description of device parameters Proline Promass 500 Modbus RS485
U[V]
24
0 t
A0026885
Assign pulse 1 to n
Prerequisite The Pulse option is selected in the Operating mode parameter (→ 132) parameter.
Description Use this function to select the process variable for the pulse output.
Selection • Off
• Mass flow
• Volume flow
• Correct.vol.flow *
• Target mass flow *
• Carrier mass fl. *
• Target vol. flow *
• Carrier vol. fl. *
• Targ.corr.vol.fl *
• Carr.corr.vol.fl *
• GSV flow *
• GSVa *
• NSV flow *
• NSVa *
• S&W volume flow *
• Oil mass flow *
• Water mass flow *
• Oil volume flow *
• Water vol. flow *
• Oil corr.vol.fl. *
• Water corr.v.fl. *
Pulse scaling
Prerequisite The Pulse option is selected in the Operating mode parameter (→ 132) and a process
variable is selected in the Assign pulse parameter (→ 134).
Description Use this function to enter the value for the measured value that a pulse is equivalent to.
134 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters
Pulse width
Prerequisite The Pulse option is selected in the Operating mode parameter (→ 132) and a process
variable is selected in the Assign pulse parameter (→ 134).
Description Use this function to enter the duration of the output pulse.
t t
P P
A0026882
Example
• Pulse value: 0.1 g
• Pulse width: 0.1 ms
• fmax: 1 / (2 × 0.1 ms) = 5 kHz
• Qmax: 5 kHz × 0.1 g = 0.5 kg/s
Endress+Hauser 135
Description of device parameters Proline Promass 500 Modbus RS485
Measuring mode
Prerequisite In the Operating mode parameter (→ 132), the Pulse option is selected, and one of
the following options is selected in the Assign pulse parameter (→ 134):
• Mass flow
• Volume flow
• Correct.vol.flow
• Target mass flow *
• Carrier mass fl. *
Description Use this function to select the measuring mode for the pulse output.
Examples
For a detailed description of the configuration examples, see the Measuring mode
parameter (→ 122)
Failure mode
Prerequisite The Pulse option is selected in the Operating mode parameter (→ 132) and a process
variable is selected in the Assign pulse parameter (→ 134).
Description Use this function to select the failure mode of the pulse output in the event of a device
alarm.
136 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters
Selection
• Actual value
In the event of a device alarm, the pulse output continues on the basis of the current
flow measurement. The fault is ignored.
• No pulses
In the event of a device alarm, the pulse output is "switched off".
NOTICE! A device alarm is a measuring device error that must be taken seriously. It can
affect the measurement quality such that the quality can no longer be guaranteed. The
Actual value option is only recommended if it can be guaranteed that all possible alarm
conditions will not affect the measurement quality.
Pulse output 1 to n
Prerequisite The Pulse option is selected in the Operating mode parameter (→ 132) parameter.
1 1
0 t 0 t
NC NO
A0028726
0 Non-conductive
1 Conductive
NC NC contact (normally closed)
NO NO contact (normally open)
The output behavior can be reversed via the Invert outp.sig. parameter (→ 152) i.e.
the transistor does not conduct for the duration of the pulse.
In addition, the behavior of the output in the event of a device alarm (Failure mode
parameter (→ 136)) can be configured.
Endress+Hauser 137
Description of device parameters Proline Promass 500 Modbus RS485
Assign freq.
Prerequisite The Frequency option is selected in the Operating mode parameter (→ 132).
Description Use this function to select the process variable for the frequency output.
Detailed description of the options Oscil. frequency, Oscil. amplitude, Oscil.
damping and Signal asymmetry: Value 1 display parameter (→ 19)
Selection • Off
• Mass flow
• Volume flow
• Correct.vol.flow
• Target mass flow *
• Carrier mass fl. *
• Target vol. flow *
• Carrier vol. fl. *
• Targ.corr.vol.fl *
• Carr.corr.vol.fl *
• Density
• Ref.density
• Ref.dens.altern. *
• GSV flow *
• GSVa *
• NSV flow *
• NSVa *
• S&W volume flow *
• Water cut *
• Oil density *
• Water density *
• Oil mass flow *
• Water mass flow *
• Oil volume flow *
• Water vol. flow *
• Oil corr.vol.fl. *
• Water corr.v.fl. *
• Concentration *
• Dynam. viscosity *
• Kinematic visc. *
• TempCompDynVisc *
• TempCompKinVisc *
• Temperature
• Carr. pipe temp. *
• Electronic temp.
• Osc. freq. 0
• Osc. freq. 1 *
• Osc. ampl. 0 *
• Osc. ampl. 1 *
• Freq. fluct. 0 *
• Freq. fluct. 1 *
• Osc. damping 0 *
• Osc. damping 1 *
• Osc.damp.fluct 0 *
138 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters
• Osc.damp.fluct 1 *
• Signal asymmetry *
• Exc. current 0 *
• Exc. current 1 *
• HBSI *
• Pressure
• Spec. output 0 *
• Spec. output 1 *
• Index inh.medium
• Index sus.bubble *
Prerequisite The Frequency option is selected in the Operating mode parameter (→ 132) and a
process variable is selected in the Assign freq. parameter (→ 138).
Prerequisite The Frequency option is selected in the Operating mode parameter (→ 132) and a
process variable is selected in the Assign freq. parameter (→ 138).
Endress+Hauser 139
Description of device parameters Proline Promass 500 Modbus RS485
Val. at min.freq
Prerequisite The Frequency option is selected in the Operating mode parameter (→ 132) and a
process variable is selected in the Assign freq. parameter (→ 138).
Description Use this function to enter the measured value for the start value frequency.
Val. at max.freq
Prerequisite The Frequency option is selected in the Operating mode parameter (→ 132) and a
process variable is selected in the Assign freq. parameter (→ 138).
Description Use this function to enter the measured value for the end value frequency.
Dependency
The entry depends on the process variable selected in the Assign freq. parameter
(→ 138).
Measuring mode
Prerequisite In the Operating mode parameter (→ 132), the Frequency option is selected, and one
of the following options is selected in the Assign freq. parameter (→ 138):
• Mass flow
• Volume flow
• Correct.vol.flow
140 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters
Description Use this function to select the measuring mode for the frequency output.
Examples
For a detailed description of the configuration examples, see the Measuring mode
parameter (→ 122)
Endress+Hauser 141
Description of device parameters Proline Promass 500 Modbus RS485
Damping out. 1 to n
Prerequisite In the Operating mode parameter (→ 132), the Frequency option is selected, and one
of the following options is selected in the Assign freq. parameter (→ 138):
• Mass flow
• Volume flow
• Correct.vol.flow
• Target mass flow *
• Carrier mass fl. *
• Density
• Ref.density
• Concentration *
• Dynam. viscosity *
• Kinematic visc. *
• TempCompDynVisc *
• TempCompKinVisc *
• Temperature
• Carr. pipe temp. *
• Electronic temp.
• Osc. freq. 0
• Osc. freq. 1 *
• Freq. fluct. 0
• Freq. fluct. 1 *
• Osc. ampl. 0 *
• Osc. ampl. 1 *
• Osc. damping 0
• Osc. damping 1 *
• Osc.damp.fluct 0
• Osc.damp.fluct 1 *
• Signal asymmetry
• Exc. current 0
• Exc. current 1 *
Detailed description of the options Oscil. frequency, Oscil. amplitude, Oscil.
damping and Signal asymmetry: Value 1 display parameter (→ 19)
Description Use this function to enter a time constant for the reaction time of the output signal to
fluctuations in the measured value.
142 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters
The frequency output is subject to separate damping that is independent of all preceding
time constants.
Response time
Prerequisite In the Operating mode parameter (→ 132), the Frequency option is selected, and one
of the following options is selected in the Assign freq. parameter (→ 138):
• Mass flow
• Volume flow
• Correct.vol.flow
• Target mass flow *
• Carrier mass fl. *
• Density
• Ref.density
• Concentration *
• Dynam. viscosity *
• Kinematic visc. *
• TempCompDynVisc *
• TempCompKinVisc *
• Temperature
• Carr. pipe temp. *
• Electronic temp.
• Osc. freq. 0
• Osc. freq. 1 *
• Freq. fluct. 0
• Freq. fluct. 1 *
• Osc. ampl. 0 *
• Osc. ampl. 1 *
• Osc. damping 0
• Osc. damping 1 *
• Osc.damp.fluct 0
• Osc.damp.fluct 1 *
• Signal asymmetry
• Exc. current 0
• Exc. current 1 *
Detailed description of the options Oscil. frequency, Oscil. amplitude, Oscil.
damping and Signal asymmetry: Value 1 display parameter (→ 19)
Description Displays the response time. This specifies how quickly the pulse/frequency/switch output
reaches the measured value change of 63 % of 100 % of the measured value change.
Endress+Hauser 143
Description of device parameters Proline Promass 500 Modbus RS485
Failure mode
Prerequisite The Frequency option is selected in the Operating mode parameter (→ 132) and a
process variable is selected in the Assign freq. parameter (→ 138).
Description Use this function to select the failure mode of the frequency output in the event of a device
alarm.
Factory setting 0 Hz
Failure freq.
Prerequisite The Frequency option is selected in the Operating mode parameter (→ 132) and a
process variable is selected in the Assign freq. parameter (→ 138).
144 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters
Description Use this function to enter the value for the frequency output in the event of a device alarm
in order to bypass the alarm.
Output freq. 1 to n
Prerequisite In the Operating mode parameter (→ 132), the Frequency option is selected.
Description Displays the actual value of the output frequency which is currently measured.
Prerequisite The Switch option is selected in the Operating mode parameter (→ 132).
Description Use this function to select a function for the switch output.
Selection • Off
• On
• Diag. behavior
• Limit
• Fl. direct.check
• Status
Endress+Hauser 145
Description of device parameters Proline Promass 500 Modbus RS485
• Limit
Indicates if a specified limit value has been reached for the process variable. Is used to
output diagnostic information relating to the process and to react to it appropriately at
the system level.
• Fl. direct.check
Indicates the flow direction (forward or reverse flow).
• Status
Indicates the device status depending on whether empty pipe detection or low flow cut
off is selected.
Prerequisite • In the Operating mode parameter (→ 132), the Switch option is selected.
• In the Switch out funct parameter (→ 145), the Diag. behavior option is selected.
Description Use this function to select the diagnostic event category that is displayed for the switch
output.
Selection • Alarm
• Alarm or warning
• Warning
Selection
• Alarm
The switch output signals only diagnostic events in the alarm category.
• Alarm or warning
The switch output signals diagnostic events in the alarm and warning category.
• Warning
The switch output signals only diagnostic events in the warning category.
Assign limit
Prerequisite • The Switch option is selected in the Operating mode parameter (→ 132).
• The Limit option is selected in the Switch out funct parameter (→ 145).
Description Use this function to select a process variable for the limit function.
146 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters
Endress+Hauser 147
Description of device parameters Proline Promass 500 Modbus RS485
A
1
2
t
3
4 t
A0026891
1 Switch-on value
2 Switch-off value
3 Conductive
4 Non-conductive
A Process variable
B Status output
A
1
2
t
3
4 t
A0026892
1 Switch-off value
2 Switch-on value
3 Conductive
4 Non-conductive
A Process variable
B Status output
2
3 t
A0026893
148 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters
Switch-on value
Prerequisite • The Switch option is selected in the Operating mode parameter (→ 132).
• The Limit option is selected in the Switch out funct parameter (→ 145).
Description Use this function to enter the measured value for the switch-on point.
Dependency
The unit depends on the process variable selected in the Assign limit parameter
(→ 146).
Switch-off value
Prerequisite • The Switch option is selected in the Operating mode parameter (→ 132).
• The Limit option is selected in the Switch out funct parameter (→ 145).
Description Use this function to enter the measured value for the switch-off point.
Dependency
The unit depends on the process variable selected in the Assign limit parameter
(→ 146).
Endress+Hauser 149
Description of device parameters Proline Promass 500 Modbus RS485
Assign dir.check
Prerequisite • The Switch option is selected in the Operating mode parameter (→ 132).
• The Fl. direct.check option is selected in the Switch out funct parameter (→ 145).
Description Use this function to select a process variable for monitoring the flow direction.
Selection • Off
• Volume flow
• Mass flow
• Correct.vol.flow *
Assign status
Prerequisite • The Switch option is selected in the Operating mode parameter (→ 132).
• The Status option is selected in the Switch out funct parameter (→ 145).
Description Use this function to select a device status for the switch output.
Switch-on delay
Prerequisite • The Switch option is selected in the Operating mode parameter (→ 132).
• The Limit option is selected in the Switch out funct parameter (→ 145).
Description Use this function to enter a delay time for switching on the switch output.
150 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters
Switch-off delay
Prerequisite • The Switch option is selected in the Operating mode parameter (→ 132).
• The Limit option is selected in the Switch out funct parameter (→ 145).
Description Use this function to enter a delay time for switching off the switch output.
Failure mode
Description Use this function to select a failsafe mode for the switch output in the event of a device
alarm.
Switch status 1 to n
Prerequisite The Switch option is selected in the Operating mode parameter (→ 132).
Endress+Hauser 151
Description of device parameters Proline Promass 500 Modbus RS485
Invert outp.sig.
Description Use this function to select whether to invert the output signal.
Selection • No
• Yes
Factory setting No
0
A0026693
0
A0026692
‣ Relay output 1 to n
152 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters
Terminal no.
Description Displays the terminal numbers used by the relay output module.
Relay outp.func.
Description Use this function to select an output function for the relay output.
Selection • Closed
• Open
• Diag. behavior
• Limit
• Fl. direct.check
• Digital Output
Endress+Hauser 153
Description of device parameters Proline Promass 500 Modbus RS485
Assign dir.check
Prerequisite In the Relay outp.func. parameter (→ 153), the Fl. direct.check option is selected.
Description Use this function to select a process variable for monitoring the flow direction.
Selection • Off
• Volume flow
• Mass flow
• Correct.vol.flow *
Assign limit
Prerequisite The Limit option is selected in the Relay outp.func. parameter (→ 153).
Description Use this function to select a process variable for the limit value function.
154 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters
Prerequisite In the Relay outp.func. parameter (→ 153), the Diag. behavior option is selected.
Description Use this function to select the category of the diagnostic events that are displayed for the
relay output.
Selection • Alarm
• Alarm or warning
• Warning
Endress+Hauser 155
Description of device parameters Proline Promass 500 Modbus RS485
Selection
• Alarm
The relay output signals only diagnostic events in the alarm category.
• Alarm or warning
The relay output signals diagnostic events in the alarm and warning category.
• Warning
The relay output signals only diagnostic events in the warning category.
Assign status
Prerequisite In the Relay outp.func. parameter (→ 153), the Digital Output option is selected.
Description Use this function to select the device status for the relay output.
Switch-off value
Prerequisite In the Relay outp.func. parameter (→ 153), the Limit option is selected.
Description Use this function to enter the measured value for the switch-off point.
156 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters
Dependency
The unit is dependent on the process variable selected in the Assign limit parameter
(→ 154).
Switch-off delay
Prerequisite In the Relay outp.func. parameter (→ 153), the Limit option is selected.
Description Use this function to enter a delay time for switching off the switch output.
Switch-on value
Prerequisite The Limit option is selected in the Relay outp.func. parameter (→ 153).
Description Use this function to enter the measured value for the switch-on point.
Dependency
The unit is dependent on the process variable selected in the Assign limit parameter
(→ 154).
Endress+Hauser 157
Description of device parameters Proline Promass 500 Modbus RS485
Switch-on delay
Prerequisite In the Relay outp.func. parameter (→ 153), the Limit option is selected.
Description Use this function to enter a delay time for switching on the switch output.
Failure mode
Description Use this function to select the failure mode of the relay output in the event of a device
alarm.
Switch status
158 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters
Powerless relay
Description Use this function to select the quiescent state for the relay output.
Selection • Open
• Closed
Endress+Hauser 159
Description of device parameters Proline Promass 500 Modbus RS485
Navigation Expert → Output → Double pulse out → Master term. no. (0981)
Description Displays the master terminal number for the double pulse output.
Navigation Expert → Output → Double pulse out → Slave term. no. (0990)
Description Displays the slave terminal number for the double pulse output.
Signal mode
Description Use this function to select the signal mode for the double pulse output.
Selection • Passive
• Active
• Passive NAMUR
160 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters
Assign pulse 1
Description Use this function to select a process variable for the double pulse output.
Selection • Off
• Mass flow
• Volume flow
• Correct.vol.flow *
• Target mass flow *
• Carrier mass fl. *
• Target vol. flow *
• Carrier vol. fl. *
• Targ.corr.vol.fl *
• Carr.corr.vol.fl *
• GSV flow *
• GSVa *
• NSV flow *
• NSVa *
• S&W volume flow *
• Oil mass flow *
• Water mass flow *
• Oil volume flow *
• Water vol. flow *
• Oil corr.vol.fl. *
• Water corr.v.fl. *
Navigation Expert → Output → Double pulse out → Value per pulse (0983)
Description Use this function to enter the value for the measured value that a pulse is equivalent to.
Endress+Hauser 161
Description of device parameters Proline Promass 500 Modbus RS485
Pulse width
Description Use this function to enter the duration of the output pulse.
Additional information For a detailed description and example: Pulse width parameter (→ 135)
Phase shift
Selection • 90°
• 180°
Measuring mode
Description Use this function to select the measuring mode for the double pulse output.
162 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters
Examples
For a detailed description of the configuration examples, see the Measuring mode
parameter (→ 122)
Failure mode
Description Use this function to select the failure mode of the double pulse output in the event of a
device alarm.
Selection
• Actual value
In the event of a device alarm, the double pulse output continues on the basis of the
current flow measurement. The fault is ignored.
• No pulses
In the event of a device alarm, the double pulse output is "switched off".
NOTICE! A device alarm is a measuring device error that must be taken seriously. It can
affect the measurement quality such that the quality can no longer be guaranteed. The
Actual value option is only recommended if it can be guaranteed that all possible alarm
conditions will not affect the measurement quality.
Endress+Hauser 163
Description of device parameters Proline Promass 500 Modbus RS485
Pulse output
Description Displays the pulse frequency of the double pulse output which is currently output.
Additional information For a detailed description and example: Pulse output parameter (→ 62)
Invert outp.sig.
Description Use this function to select whether to invert the output signal.
Selection • No
• Yes
Factory setting No
0
A0026693
0
A0026692
‣ Communication
164 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters
‣ Modbus config.
Bus address
Endress+Hauser 165
Description of device parameters Proline Promass 500 Modbus RS485
Baudrate
Selection • ASCII
• RTU
Parity
Selection • Odd
• Even
• None/1 stop bit
• None/2 stop bits
166 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters
Byte order
Description Use this function to select the sequence in which the bytes are transmitted. The
transmission sequence must be coordinated with the Modbus master.
Selection • 0-1-2-3
• 3-2-1-0
• 1-0-3-2
• 2-3-0-1
FLOAT
Sequence
Options 1. 2. 3. 4.
Endress+Hauser 167
Description of device parameters Proline Promass 500 Modbus RS485
INTEGER
Sequence
Options 1. 2.
* = factory setting, MSB = most significant byte, LSB = least significant byte
STRING
Presentation taking the example of a device parameter with a data length of 18 bytes.
Sequence
* = factory setting, MSB = most significant byte, LSB = least significant byte
Telegram delay
Description Use this function to enter a delay time after which the measuring device replies to the
request telegram of the Modbus master. This allows communication to be adapted to slow
Modbus RS485 masters.
Factory setting 6 ms
Failure mode
Description Use this function to select the measured value output in the event of a diagnostic message
via Modbus communication.
168 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters
Bus termination
Fieldb.writ.acc.
Description Use this function to restrict access to the measuring device via fieldbus (Modbus protocol).
8) Not a Number
Endress+Hauser 169
Description of device parameters Proline Promass 500 Modbus RS485
Selection
• Read + write
The parameters are read and write parameters.
• Read only
• The parameters are read only parameters.
‣ Modbus info
Device ID
Device revision
170 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters
Navigation Expert → Communication → Modbus data map → Scan list reg.0 to 15 (7114)
Description Use this function to enter the scan list register. By entering the register address (1-based),
up to 16 device parameters can be grouped by assigning them to the scan list registers 0 to
15. The data of the device parameters assigned here are read out via the register addresses
5051 to 5081.
Factory setting 1
‣ Web server
Endress+Hauser 171
Description of device parameters Proline Promass 500 Modbus RS485
Webserv.language
Description Use this function to select the Web server language setting.
Selection • English
• Deutsch
• Français
• Español
• Italiano
• Nederlands
• Portuguesa
• Polski
• русский язык(Ru)
• Svenska
• Türkçe
• 中文 (Chinese)
• 日本語 (Japanese)
• 한국어 (Korean)
• Bahasa Indonesia
• tiếng Việt (Vit)
• čeština (Czech)
MAC Address
User interface Unique 12-digit character string comprising letters and numbers
DHCP client
Description Use this function to activate and deactivate the DHCP client functionality.
172 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters
Selection • Off
• On
IP address
Description Display or enter the IP address of the Web server integrated in the measuring device.
Subnet mask
Default gateway
Endress+Hauser 173
Description of device parameters Proline Promass 500 Modbus RS485
Webserver funct.
Description Use this function to switch the Web server on and off.
Selection • Off
• HTML Off
• On
Factory setting On
Selection
Option Description
Login page
Description Use this function to select the format of the login page.
‣ WLAN settings
174 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters
WLAN
Description Use this function to enable and disable the WLAN connection.
Endress+Hauser 175
Description of device parameters Proline Promass 500 Modbus RS485
Selection • Disable
• Enable
WLAN mode
SSID name
Description Use this function to enter the user-defined SSID name (max. 32 characters) of the WLAN
network.
User entry –
Factory setting –
Network security
Description Use this function to select the type of security for the WLAN interface.
Selection • Unsecured
• WPA2-PSK
• EAP-PEAP MSCHAP2 *
• EAP-PEAP NoAuth. *
• EAP-TLS *
176 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters
Sec. identific.
Description Use this function to select the security settings (download via the menu: Data
Management > Security > Download WLAN).
User name
Description Use this function to enter the username of the WLAN network.
User entry –
Factory setting –
WLAN password
Description Use this function to enter the WLAN password for the WLAN network.
User entry –
Factory setting –
WLAN IP address
Description Use this function to enter the IP address of the measuring device's WLAN connection.
Endress+Hauser 177
Description of device parameters Proline Promass 500 Modbus RS485
User interface Unique 12-digit character string comprising letters and numbers
WLAN passphrase
Prerequisite The WPA2-PSK option is selected in the Security type parameter (→ 176).
User entry 8 to 32-digit character string comprising numbers, letters and special characters (without
spaces)
178 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters
Description Use this function to select which name is used for the SSID 11).
SSID name
Prerequisite • The User-defined option is selected in the Assign SSID name parameter (→ 179).
• The Access point option is selected in the WLAN mode parameter (→ 176).
User entry Max. 32-digit character string comprising numbers, letters and special characters
WLAN channel
User entry 1 to 11
Factory setting 6
Endress+Hauser 179
Description of device parameters Proline Promass 500 Modbus RS485
Select antenna
Description Use this function to select whether the external or internal antenna is used for reception.
Connection state
Rec.sig.strength
Gateway IP addr.
180 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters
IP address DNS
Description Use this function to enter the IP address of the domain name server.
‣ Application
‣ Totalizer 1 to n → 182
‣ Viscosity → 187
‣ Concentration → 187
‣ Petroleum → 187
Description Use this function to reset all totalizers to the value 0 and restart the totaling process. This
deletes all the flow values previously totalized.
Selection • Cancel
• Reset + totalize
Endress+Hauser 181
Description of device parameters Proline Promass 500 Modbus RS485
Options Description
Reset + totalize Resets all totalizers to 0 and restarts the totaling process. This deletes all the flow
values previously totalized.
‣ Totalizer 1 to n
Assign variable
Description Use this function to select a process variable for the Totalizer 1 to n.
Selection • Off
• Mass flow
• Volume flow
• Correct.vol.flow *
• Target mass flow *
• Carrier mass fl. *
• Target vol. flow *
• Carrier vol. fl. *
• Targ.corr.vol.fl *
• Carr.corr.vol.fl *
• GSV flow *
• GSVa *
• NSV flow *
182 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters
• NSVa *
• S&W volume flow *
• Oil mass flow *
• Water mass flow *
• Oil volume flow *
• Water vol. flow *
• Oil corr.vol.fl. *
• Water corr.v.fl. *
Selection
If the Off option is selected, only Assign variable parameter (→ 182) is still displayed
in the Totalizer 1 to n submenu. All other parameters in the submenu are hidden.
Unit totalizer 1 to n
Prerequisite A process variable is selected in the Assign variable parameter (→ 182) of the
Totalizer 1 to n submenu.
Description Use this function to select the process variable unit for the Totalizer 1 to n (→ 182).
or
or
Endress+Hauser 183
Description of device parameters Proline Promass 500 Modbus RS485
or
or
Other units
None *
Selection
The selection is dependent on the process variable selected in the Assign variable
parameter (→ 182).
Operation mode
Prerequisite A process variable is selected in the Assign variable parameter (→ 182) of the
Totalizer 1 to n submenu.
Description Use this function to select how the totalizer summates the flow.
184 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters
Control Tot. 1 to n
Prerequisite A process variable is selected in the Assign variable parameter (→ 182) of the
Totalizer 1 to n submenu.
Description Use this function to select the control of totalizer value 1-3.
Selection • Totalize
• Reset + hold *
• Preset + hold *
• Reset + totalize
• Preset+totalize *
• Hold *
Options Description
Reset + hold The totaling process is stopped and the totalizer is reset to 0.
Preset + hold The totaling process is stopped and the totalizer is set to its defined start value
from the Preset value parameter.
Reset + totalize The totalizer is reset to 0 and the totaling process is restarted.
Preset+totalize The totalizer is set to the defined start value from the Preset value parameter and
the totaling process is restarted.
Preset value 1 to n
Prerequisite A process variable is selected in the Assign variable parameter (→ 182) of the
Totalizer 1 to n submenu.
Endress+Hauser 185
Description of device parameters Proline Promass 500 Modbus RS485
Description Use this function to enter a start value for the Totalizer 1 to n.
Example
This configuration is suitable for applications such as iterative filling processes with a fixed
batch quantity.
Failure mode
Prerequisite A process variable is selected in the Assign variable parameter (→ 182) of the
Totalizer 1 to n submenu.
Description Use this function to select how a totalizer behaves in the event of a device alarm.
Selection • Stop
• Actual value
• Last valid value
Selection
• Stop
The totalizer is stopped in the event of a device alarm.
• Actual value
The totalizer continues to count based on the actual measured value; the device alarm is
ignored.
• Last valid value
The totalizer continues to count based on the last valid measured value before the device
alarm occurred.
186 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters
For detailed information on the parameter descriptions for the Viscosity application
package,refer to the Special Documentation for the device → 8
‣ Viscosity
‣ Concentration
‣ Custody transfer
‣ Petroleum
‣ Appl.spec. calc.
‣ Appl.spec.param. → 188
Endress+Hauser 187
Description of device parameters Proline Promass 500 Modbus RS485
"Appl.spec.param." submenu
‣ Appl.spec.param.
Parameter 0
Factory setting 0
Parameter 1
Factory setting 0
188 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters
Parameter 2
Factory setting 0
Parameter 3
Factory setting 0
Parameter 4
Factory setting 0
Parameter 5
Factory setting 0
Endress+Hauser 189
Description of device parameters Proline Promass 500 Modbus RS485
Parameter 6
Factory setting 0
Parameter 7
Factory setting 0
Parameter 8
Factory setting 0
Parameter 9
Factory setting 0
190 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters
‣ Process variab.
Spec. input 0
Navigation Expert → Application → Appl.spec. calc. → Process variab. → Spec. input 0 (6366)
Description Shows the application specific input value 0 used for the application specific calculation.
Factory setting 0
FSTypeAppSpec 0
Description Use this function to select the failsafe mode for the application-specific input value 0.
FSValueAppSpec 0
Description Use this function to enter the failsafe value for the application-specific input value 0.
Endress+Hauser 191
Description of device parameters Proline Promass 500 Modbus RS485
Factory setting 0
Spec. input 1
Navigation Expert → Application → Appl.spec. calc. → Process variab. → Spec. input 1 (6367)
Description Shows the application specific input value 1 used for the application specific calculation.
Factory setting 0
FSTypeAppSpec 1
Description Use this function to select the failsafe mode for the application-specific input value 1.
FSValueAppSpec 1
Description Use this function to enter the failsafe value for the application-specific input value 1.
Factory setting 0
Spec. output 0
Navigation Expert → Application → Appl.spec. calc. → Process variab. → Spec. output 0 (6364)
192 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters
Factory setting 0
Spec. output 1
Navigation Expert → Application → Appl.spec. calc. → Process variab. → Spec. output 1 (6365)
Factory setting 0
‣ Medium index
Navigation Expert → Application → Medium index → Cut off inh. gas (6375)
Description Enter cut off value for wet gas applications. Below this value the Index inhomogeneous
medium is set to 0.
Endress+Hauser 193
Description of device parameters Proline Promass 500 Modbus RS485
Description Enter cut off value for liquid applications. Below this value the Index inhomogeneous
medium is set to 0.
Index inh.medium
Factory setting 2
Description Enter cut off value for suspended bubbles. Below this value the Index suspended bubbles is
set to 0.
Index sus.bubble
Factory setting 0
194 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters
‣ Diagnostics
‣ Cust.transf.log. → 204
‣ Heartbeat → 237
‣ Simulation → 237
Endress+Hauser 195
Description of device parameters Proline Promass 500 Modbus RS485
Actual diagnos.
Description Displays the current diagnostic message. If two or more messages occur simultaneously,
the message with the highest priority is shown on the display.
User interface Symbol for diagnostic behavior, diagnostic code and short message.
Example
For the display format:
F271 Main electronics
Timestamp
Description Displays the operating time when the current diagnostic message occurred.
User interface Days (d), hours (h), minutes (m) and seconds (s)
Example
For the display format:
24d12h13m00s
Prev.diagnostics
Description Displays the diagnostic message that occurred before the current message.
User interface Symbol for diagnostic behavior, diagnostic code and short message.
196 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters
Example
For the display format:
F271 Main electronics
Timestamp
Description Displays the operating time when the last diagnostic message before the current message
occurred.
User interface Days (d), hours (h), minutes (m) and seconds (s)
Example
For the display format:
24d12h13m00s
Description Use this function to display the time the device has been in operation since the last device
restart.
User interface Days (d), hours (h), minutes (m) and seconds (s)
Operating time
Description Use this function to display the length of time the device has been in operation.
User interface Days (d), hours (h), minutes (m) and seconds (s)
Endress+Hauser 197
Description of device parameters Proline Promass 500 Modbus RS485
‣ Diagnostic list
Diagnostics 1
Description Displays the current diagnostics message with the highest priority.
User interface Symbol for diagnostic behavior, diagnostic code and short message.
Examples
For the display format:
• F271 Main electronics
• F276 I/O module
Timestamp
Description Displays the operating time when the diagnostic message with the highest priority
occurred.
198 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters
User interface Days (d), hours (h), minutes (m) and seconds (s)
Example
For the display format:
24d12h13m00s
Diagnostics 2
Description Displays the current diagnostics message with the second-highest priority.
User interface Symbol for diagnostic behavior, diagnostic code and short message.
Examples
For the display format:
• F271 Main electronics
• F276 I/O module
Timestamp
Description Displays the operating time when the diagnostic message with the second-highest priority
occurred.
User interface Days (d), hours (h), minutes (m) and seconds (s)
Example
For the display format:
24d12h13m00s
Endress+Hauser 199
Description of device parameters Proline Promass 500 Modbus RS485
Diagnostics 3
Description Displays the current diagnostics message with the third-highest priority.
User interface Symbol for diagnostic behavior, diagnostic code and short message.
Examples
For the display format:
• F271 Main electronics
• F276 I/O module
Timestamp
Description Displays the operating time when the diagnostic message with the third-highest priority
occurred.
User interface Days (d), hours (h), minutes (m) and seconds (s)
Example
For the display format:
24d12h13m00s
Diagnostics 4
Description Displays the current diagnostics message with the fourth-highest priority.
User interface Symbol for diagnostic behavior, diagnostic code and short message.
200 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters
Examples
For the display format:
• F271 Main electronics
• F276 I/O module
Timestamp
Description Displays the operating time when the diagnostic message with the fourth-highest priority
occurred.
User interface Days (d), hours (h), minutes (m) and seconds (s)
Example
For the display format:
24d12h13m00s
Diagnostics 5
Description Displays the current diagnostics message with the fifth-highest priority.
User interface Symbol for diagnostic behavior, diagnostic code and short message.
Examples
For the display format:
• F271 Main electronics
• F276 I/O module
Endress+Hauser 201
Description of device parameters Proline Promass 500 Modbus RS485
Timestamp
Description Displays the operating time when the diagnostic message with the fifth-highest priority
occurred.
User interface Days (d), hours (h), minutes (m) and seconds (s)
Example
For the display format:
24d12h13m00s
‣ Event logbook
Filter options
Description Use this function to select the category whose event messages are displayed in the event
list of the local display.
Selection • All
• Failure (F)
• Funct. check (C)
• Out of spec. (S)
• Mainten. req.(M)
• Information (I)
202 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters
Filter options
Description Use this function to select the category whose event messages are displayed in the event
list of the operating tool.
Selection • All
• Failure (F)
• Funct. check (C)
• Out of spec. (S)
• Mainten. req.(M)
• Information (I)
▸ Event list
Endress+Hauser 203
Description of device parameters Proline Promass 500 Modbus RS485
Event list
Description Displays the history of event messages of the category selected in the Filter options
parameter (→ 202).
Examples
For the display format:
• I1091 Configuration modified
24d12h13m00s
• F271 Main electronics
01d04h12min30s
HistoROM
A HistoROM is a "non-volatile" device memory in the form of an EEPROM.
‣ Cust.transf.log.
204 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters
‣ Device info
Device tag
Description Displays a unique name for the measuring point so it can be identified quickly within the
plant. The name is displayed in the header.
User interface Max. 32 characters, such as letters, numbers or special characters (e.g. @, %, /).
1 XXXXXXXXX
A0029422
Endress+Hauser 205
Description of device parameters Proline Promass 500 Modbus RS485
Serial number
User interface Max. 11-digit character string comprising letters and numbers.
Firmware version
Device name
Description Displays the name of the transmitter. It can also be found on the nameplate of the
transmitter.
Order code
User interface Character string composed of letters, numbers and certain punctuation marks (e.g. /).
206 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters
Additional information For additional information, see Ext. order cd. 1 parameter (→ 207)
Endress+Hauser 207
Description of device parameters Proline Promass 500 Modbus RS485
Additional information For additional information, see Ext. order cd. 1 parameter (→ 207)
ENP version
‣ Main elec.+I/O1
Software rev.
Description Use this function to display the software revision of the module.
208 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters
Description Use this function to display the software build number of the module.
Bootloader rev.
Description Use this function to display the bootloader revision of the software.
‣ Sens. electronic
Software rev.
Description Use this function to display the software revision of the module.
Endress+Hauser 209
Description of device parameters Proline Promass 500 Modbus RS485
Description Use this function to display the software build number of the module.
Bootloader rev.
Description Use this function to display the bootloader revision of the software.
‣ I/O module 1
I/O 1 terminals
210 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters
Software rev.
Description Use this function to display the software revision of the module.
Description Use this function to display the software build number of the module.
Bootloader rev.
Description Use this function to display the bootloader revision of the software.
‣ I/O module 2
Endress+Hauser 211
Description of device parameters Proline Promass 500 Modbus RS485
I/O 2 terminals
Software rev.
Description Use this function to display the software revision of the module.
Description Use this function to display the software build number of the module.
Bootloader rev.
Description Use this function to display the bootloader revision of the software.
212 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters
‣ I/O module 3
I/O 3 terminals
Software rev.
Description Use this function to display the software revision of the module.
Endress+Hauser 213
Description of device parameters Proline Promass 500 Modbus RS485
Description Use this function to display the software build number of the module.
Bootloader rev.
Description Use this function to display the bootloader revision of the software.
‣ I/O module 4
I/O 4 terminals
214 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters
Software rev.
Description Use this function to display the software revision of the module.
Description Use this function to display the software build number of the module.
Bootloader rev.
Description Use this function to display the bootloader revision of the software.
‣ Display module
Endress+Hauser 215
Description of device parameters Proline Promass 500 Modbus RS485
Software rev.
Description Use this function to display the software revision of the module.
Description Use this function to display the software build number of the module.
Bootloader rev.
Description Use this function to display the bootloader revision of the software.
‣ Data logging
216 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters
‣ Displ.channel 1 → 223
‣ Displ.channel 2 → 224
‣ Displ.channel 3 → 225
‣ Displ.channel 4 → 225
Assign chan. 1
Description Use this function to select a process variable for the data logging channel.
Selection • Off
• Mass flow
• Volume flow
• Correct.vol.flow *
• Target mass flow *
• Carrier mass fl. *
• Target vol. flow *
• Carrier vol. fl. *
• Targ.corr.vol.fl *
• Carr.corr.vol.fl *
• Density
• Ref.density *
• Ref.dens.altern. *
Endress+Hauser 217
Description of device parameters Proline Promass 500 Modbus RS485
• GSV flow *
• GSVa *
• NSV flow *
• NSVa *
• S&W volume flow *
• Water cut *
• Oil density *
• Water density *
• Oil mass flow *
• Water mass flow *
• Oil volume flow *
• Water vol. flow *
• Oil corr.vol.fl. *
• Water corr.v.fl. *
• Concentration *
• Dynam. viscosity *
• Kinematic visc. *
• TempCompDynVisc *
• TempCompKinVisc *
• Temperature
• Carr. pipe temp. *
• Electronic temp.
• Osc. freq. 0
• Osc. freq. 1 *
• Freq. fluct. 0 *
• Freq. fluct. 1 *
• Oscil. amplitude *
• Osc. ampl. 1 *
• Freq. fluct. 0 *
• Osc. damping 0 *
• Osc. damping 1 *
• Osc.damp.fluct 0 *
• Osc.damp.fluct 1 *
• Signal asymmetry *
• Exc. current 0 *
• Exc. current 1 *
• HBSI *
• Curr.output 1 *
• Curr.output 2 *
• Curr.output 3 *
• Pressure
• Spec. output 1 *
• Index inh.medium
• Spec. output 0 *
• Index sus.bubble *
218 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters
Once the maximum number of data points is reached, the oldest data points in the data log
are cyclically overwritten in such a way that the last 1000, 500, 333 or 250 measured
values are always in the log (ring memory principle).
The log contents are cleared if the option selected is changed.
Assign chan. 2
Description Options for the assignment of a process variable to the data logging channel.
Assign chan. 3
Description Options for the assignment of a process variable to the data logging channel.
Assign chan. 4
Description Options for the assignment of a process variable to the data logging channel.
Endress+Hauser 219
Description of device parameters Proline Promass 500 Modbus RS485
Logging interval
Description Use this function to enter the logging interval Tlog for data logging.
Example
If 1 logging channel is used:
• Tlog = 1000 × 1 s = 1 000 s ≈15 min
• Tlog = 1000 × 10 s = 10 000 s ≈ 3 h
• Tlog = 1000 × 80 s = 80 000 s ≈ 1 d
• Tlog = 1000 × 3 600 s = 3 600 000 s ≈ 41 d
Clear logging
Selection • Cancel
• Clear data
220 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters
Data logging
Selection • Overwriting
• Not overwriting
Logging delay
Prerequisite In the Data logging parameter (→ 221), the Not overwriting option is selected.
Description Use this function to enter the time delay for measured value logging.
Factory setting 0h
Endress+Hauser 221
Description of device parameters Proline Promass 500 Modbus RS485
Data log.control
Prerequisite In the Data logging parameter (→ 221), the Not overwriting option is selected.
Description Use this function to start and stop measured value logging.
Selection • None
• Delete + start
• Stop
Prerequisite In the Data logging parameter (→ 221), the Not overwriting option is selected.
222 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters
Logging duration
Prerequisite In the Data logging parameter (→ 221), the Not overwriting option is selected.
Factory setting 0s
▸ Displ.channel 1
Display channel 1
Endress+Hauser 223
Description of device parameters Proline Promass 500 Modbus RS485
• Freq. fluct. 0
• Freq. fluct. 1 *
• Oscil. amplitude *
• Osc. ampl. 1 *
• Osc. damping 0
• Osc. damping 1 *
• Osc.damp.fluct 0
• Osc.damp.fluct 1 *
• Signal asymmetry
• Exc. current 0
• Exc. current 1 *
Description Displays the measured value trend for the logging channel in the form of a chart.
/ ../XXXXXXXX
175.77
40.69 kg/h
-100s 0
A0016357
• x-axis: depending on the number of channels selected displays 250 to 1000 measured
values of a process variable.
• y-axis: displays the approximate measured value span and constantly adapts this to the
ongoing measurement.
▸ Displ.channel 2
Display channel 2
224 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters
▸ Displ.channel 3
Display channel 3
▸ Displ.channel 4
Display channel 4
‣ Min/max val.
Endress+Hauser 225
Description of device parameters Proline Promass 500 Modbus RS485
‣ Tors.oscil.freq. → 231
‣ Tors.oscil.damp. → 234
‣ Tors.sig.asymm. → 236
Reset min/max
Description Use this function to select measured variables whose minimum, maximum and average
measured values are to be reset.
Selection • Cancel
• Oscil. amplitude *
• Osc. ampl. 1 *
• Oscil. damping
• Tors.oscil.damp. *
• Oscil. frequency
• Tors.oscil.freq. *
• Signal asymmetry
• Tors.sig.asymm. *
226 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters
‣ Main elect.temp.
Minimum value
Navigation Expert → Diagnostics → Min/max val. → Main elect.temp. → Minimum value (0688)
Description Displays the lowest previously measured temperature value of the electronics module in
the transmitter.
Maximum value
Description Displays the highest previously measured temperature value of the electronics module in
the transmitter.
Endress+Hauser 227
Description of device parameters Proline Promass 500 Modbus RS485
‣ Sensor elec.temp
Maximum value
Description Displays the highest previously measured temperature value of the electronics module in
the sensor connection housing.
Minimum value
Navigation Expert → Diagnostics → Min/max val. → Sensor elec.temp → Minimum value (6052)
Description Displays the lowest previously measured temperature value of the electronics module in
the sensor connection housing.
228 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters
‣ Medium temp.
Minimum value
Navigation Expert → Diagnostics → Min/max val. → Medium temp. → Minimum value (6109)
Maximum value
Navigation Expert → Diagnostics → Min/max val. → Medium temp. → Maximum value (6108)
Endress+Hauser 229
Description of device parameters Proline Promass 500 Modbus RS485
Minimum value
Navigation Expert → Diagnostics → Min/max val. → Carr. pipe temp. → Minimum value (6030)
Description Displays the lowest previously measured temperature value of the carrier pipe.
Maximum value
Navigation Expert → Diagnostics → Min/max val. → Carr. pipe temp. → Maximum value (6029)
Description Displays the highest previously measured temperature value of the carrier pipe.
230 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters
‣ Oscil. frequency
Minimum value
Navigation Expert → Diagnostics → Min/max val. → Oscil. frequency → Minimum value (6071)
Maximum value
Navigation Expert → Diagnostics → Min/max val. → Oscil. frequency → Maximum value (6070)
"Tors.oscil.freq." submenu
‣ Tors.oscil.freq.
Endress+Hauser 231
Description of device parameters Proline Promass 500 Modbus RS485
Minimum value
Maximum value
‣ Oscil. amplitude
Minimum value
Navigation Expert → Diagnostics → Min/max val. → Oscil. amplitude → Minimum value (6010)
232 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters
Maximum value
Navigation Expert → Diagnostics → Min/max val. → Oscil. amplitude → Maximum value (6009)
Minimum value
Navigation Expert → Diagnostics → Min/max val. → Tor. osc. amp. → Minimum value (6008)
Maximum value
Navigation Expert → Diagnostics → Min/max val. → Tor. osc. amp. → Maximum value (6007)
Endress+Hauser 233
Description of device parameters Proline Promass 500 Modbus RS485
‣ Oscil. damping
Minimum value
Navigation Expert → Diagnostics → Min/max val. → Oscil. damping → Minimum value (6122)
Maximum value
Navigation Expert → Diagnostics → Min/max val. → Oscil. damping → Maximum value (6121)
"Tors.oscil.damp." submenu
‣ Tors.oscil.damp.
234 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters
Minimum value
Maximum value
‣ Signal asymmetry
Minimum value
Endress+Hauser 235
Description of device parameters Proline Promass 500 Modbus RS485
Maximum value
‣ Tors.sig.asymm.
Minimum value
Maximum value
236 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters
‣ Heartbeat
‣ Simulation
Endress+Hauser 237
Description of device parameters Proline Promass 500 Modbus RS485
Assign proc.var.
Description Use this function to select a process variable for the simulation process that is activated.
The display alternates between the measured value and a diagnostic message of the
"Function check" category (C) while simulation is in progress.
Selection • Off
• Mass flow
• Volume flow
• Correct.vol.flow *
• Target vol. flow *
• Carrier vol. fl. *
• Targ.corr.vol.fl *
• Carr.corr.vol.fl *
• Density
• Ref.density *
• Ref.dens.altern. *
• GSV flow *
• GSVa *
• NSV flow *
• NSVa *
• S&W volume flow *
• Water cut *
• Oil density *
• Water density *
• Oil mass flow *
• Water mass flow *
• Oil volume flow *
• Water vol. flow *
• Oil corr.vol.fl. *
• Water corr.v.fl. *
238 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters
• Temperature
• Dynam. viscosity *
• Kinematic visc. *
• TempCompDynVisc *
• TempCompKinVisc *
• Concentration *
• Target mass flow *
• Carrier mass fl. *
Description Use this function to enter a simulation value for the selected process variable. Subsequent
measured value processing and the signal output use this simulation value. In this way,
users can verify whether the measuring device has been configured correctly.
Factory setting 0
Status inp.sim 1 to n
Description Use this function to switch simulation of the status input on and off. The display alternates
between the measured value and a diagnostic message of the "Function check" category (C)
while simulation is in progress.
Selection • Off
• On
Endress+Hauser 239
Description of device parameters Proline Promass 500 Modbus RS485
Selection
• Off
Simulation for the status input is switched off. The device is in normal measuring mode
or another process variable is being simulated.
• On
Simulation for the status input is active.
Signal level 1 to n
Description Use this function to select the signal level for the simulation of the status input. In this
way, users can verify the correct configuration of the status input and the correct function
of upstream feed-in units.
Selection • High
• Low
Curr.inp 1 to n sim.
Description Option for switching simulation of the current input on and off. The display alternates
between the measured value and a diagnostic message of the "Function check" category (C)
while simulation is in progress.
The desired simulation value is defined in the Value curr.inp 1 to n parameter.
Selection • Off
• On
240 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters
Value curr.inp 1 to n
Description Use this function to enter the current value for the simulation. In this way, users can verify
the correct configuration of the current input and the correct function of upstream feed-in
units.
Curr.out. 1 to n sim.
Description Use this function to switch simulation of the current output on and off. The display
alternates between the measured value and a diagnostic message of the "Function check"
category (C) while simulation is in progress.
Selection • Off
• On
Selection
• Off
Current simulation is switched off. The device is in normal measuring mode or another
process variable is being simulated.
• On
Current simulation is active.
Value curr.out 1 to n
Description Use this function to enter a current value for the simulation. In this way, users can verify
the correct adjustment of the current output and the correct function of downstream
switching units.
Endress+Hauser 241
Description of device parameters Proline Promass 500 Modbus RS485
FreqOutputSim 1 to n
Prerequisite In the Operating mode parameter (→ 132), the Frequency option is selected.
Description Use this function to switch simulation of the frequency output on and off. The display
alternates between the measured value and a diagnostic message of the "Function check"
category (C) while simulation is in progress.
Selection • Off
• On
Selection
• Off
Frequency simulation is switched off. The device is in normal measuring mode or
another process variable is being simulated.
• On
Frequency simulation is active.
Freq value 1 to n
Description Use this function to enter a frequency value for the simulation. In this way, users can verify
the correct adjustment of the frequency output and the correct function of downstream
switching units.
242 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters
Puls.outp.sim. 1 to n
Prerequisite In the Operating mode parameter (→ 132), the Pulse option is selected.
Description Use this function to switch simulation of the pulse output on and off. The display
alternates between the measured value and a diagnostic message of the "Function check"
category (C) while simulation is in progress.
Selection • Off
• Fixed value
• Down-count. val.
Selection
• Off
Pulse simulation is switched off. The device is in normal measuring mode or another
process variable is being simulated.
• Fixed value
Pulses are continuously output with the pulse width specified in the Pulse width
parameter (→ 135).
• Down-count. val.
The pulses specified in the Pulse value parameter (→ 243) are output.
Pulse value 1 to n
Description Use this function to enter a pulse value for the simulation. In this way, users can verify the
correct adjustment of the pulse output and the correct function of downstream switching
units.
Switch sim. 1 to n
Prerequisite In the Operating mode parameter (→ 132), the Switch option is selected.
Endress+Hauser 243
Description of device parameters Proline Promass 500 Modbus RS485
Description Use this function to switch simulation of the switch output on and off. The display
alternates between the measured value and a diagnostic message of the "Function check"
category (C) while simulation is in progress.
Selection • Off
• On
Selection
• Off
Switch simulation is switched off. The device is in normal measuring mode or another
process variable is being simulated.
• On
Switch simulation is active.
Switch status 1 to n
Description Use this function to select a switch value for the simulation. In this way, users can verify
the correct adjustment of the switch output and the correct function of downstream
switching units.
Selection • Open
• Closed
Description Use this function to switch simulation of the relay output on and off. The display
alternates between the measured value and a diagnostic message of the "Function check"
category (C) while simulation is in progress.
Selection • Off
• On
244 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters
Selection
• Off
Relay simulation is switched off. The device is in normal measuring mode or another
process variable is being simulated.
• On
Relay simulation is active.
Switch status 1 to n
Description Use this function to select a relay value for the simulation. In this way, users can verify the
correct adjustment of the relay output and the correct function of downstream switching
units.
Selection • Open
• Closed
Puls.outp.sim.
Description Use this function to switch simulation of the double pulse output on and off. The display
alternates between the measured value and a diagnostic message of the "Function check"
category (C) while simulation is in progress.
Selection • Off
• Fixed value
• Down-count. val.
Endress+Hauser 245
Description of device parameters Proline Promass 500 Modbus RS485
Selection
• Off
Simulation of the double pulse output is switched off. The device is in normal measuring
mode or another process variable is being simulated.
• Fixed value
Pulses are continuously output with the pulse width specified in the Pulse width
parameter (→ 162).
• Down-count. val.
The pulses specified in the Pulse value parameter (→ 246) are output.
Pulse value
Prerequisite In the Puls.outp.sim. parameter (→ 245), the Down-count. val. option is selected.
Description Use this function to enter a pulse value for simulation of the double pulse output. In this
way, users can verify the correct adjustment of the double pulse output and the correct
function of downstream switching units.
Description Use this function to switch the device alarm on and off.
Selection • Off
• On
246 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters
Event category
Description Use this function to select the category of the diagnostic events that are displayed for the
simulation in the Diag. event sim. parameter (→ 247).
Selection • Sensor
• Electronics
• Configuration
• Process
Description Use this function to select a diagnostic event for the simulation process that is activated.
Selection • Off
• Diagnostic event picklist (depends on the category selected)
Endress+Hauser 247
Country-specific factory settings Proline Promass 500 Modbus RS485
4.1 SI units
Not valid for USA and Canada.
Mass kg
Volume l
Corrected volume Nl
Density kg/l
Temperature °C
Pressure bar a
Nominal diameter
[mm] [kg/h]
1 4
2 20
4 90
8 400
15 1 300
15 FB 3 600
25 3 600
25 FB 9 000
40 9 000
40 FB 14 000
50 14 000
50 FB 36 000
80 36 000
100 60 000
248 Endress+Hauser
Proline Promass 500 Modbus RS485 Country-specific factory settings
Nominal diameter
[mm] [kg/p]
1 0.001
2 0.01
4 0.01
8 0.1
15 0.1
15 FB 1
25 1
25 FB 1
40 1
40 FB 10
50 10
50 FB 10
80 10
100 10
150 100
250 100
350 100
1 0.08
2 0.4
4 1.8
8 8
15 26
15 FB 72
25 72
25 FB 180
40 180
40 FB 300
50 300
50 FB 720
Endress+Hauser 249
Country-specific factory settings Proline Promass 500 Modbus RS485
80 720
100 1 200
350 13 t/h
1 0.02
2 0.1
4 0.45
8 2
15 6.5
15 FB 18
25 18
25 FB 45
40 45
40 FB 75
50 75
50 FB 180
80 180
100 300
150 650
4.2 US units
Only valid for USA and Canada.
Mass lb
Density lb/ft³
Temperature °F
Pressure psi a
250 Endress+Hauser
Proline Promass 500 Modbus RS485 Country-specific factory settings
Nominal diameter
[in] [lb/min]
¹/₂₄ 0.15
¹/₁₂ 0.75
¹/₈ 3.3
³/₈ 15
½ 50
½ FB 130
1 130
1 FB 330
1½ 330
1½ FB 550
2 550
2 FB 1 300
3 1 300
4 2 200
6 4 800
10 13 000
14 23 500
Current output 1 to n 4 to 20 mA US
Nominal diameter
[in] [lb/p]
¹/₂₄ 0.002
¹/₁₂ 0.02
¹/₈ 0.02
³/₈ 0.2
½ 0.2
½ FB 2
1 2
1 FB 2
1½ 2
Endress+Hauser 251
Country-specific factory settings Proline Promass 500 Modbus RS485
Nominal diameter
[in] [lb/p]
1½ FB 20
2 20
2 FB 20
3 20
4 20
6 200
10 200
14 200
¹/₂₄ 0.003
¹/₁₂ 0.015
¹/₈ 0.066
³/₈ 0.3
½ 1
½ FB 2.6
1 2.6
1 FB 6.6
1½ 6.6
1½ FB 11
2 11
2 FB 26
3 26
4 44
6 95
10 260
14 470
¹/₂₄ 0.001
¹/₁₂ 0.004
¹/₈ 0.016
³/₈ 0.075
½ 0.25
½ FB 0.65
1 0.65
1 FB 1.65
252 Endress+Hauser
Proline Promass 500 Modbus RS485 Country-specific factory settings
1½ 1.65
1½ FB 2.75
2 2.75
2 FB 6.5
3 6.5
4 11
6 23.75
10 65
14 117.5
Endress+Hauser 253
Explanation of abbreviated units Proline Promass 500 Modbus RS485
5.1 SI units
Process variable Units Explanation
SD4°C, SD15°C, SD20°C Specific density: The specific density is the ratio of the
density of the fluid to the density of water at a water
temperature of 4 °C (39 °F), 15 °C (59 °F), 20 °C (68 °F).
SG4°C, SG15°C, SG20°C Specific gravity: The specific gravity is the ratio of the
density of the fluid to the density of water at a water
temperature of 4 °C (39 °F), 15 °C (59 °F), 20 °C (68 °F).
bar Bar
Corrected Nl, Nm³, Sm³ Normal liter, normal cubic meter, standard cubic meter
volume
5.2 US units
Process variable Units Explanation
254 Endress+Hauser
Proline Promass 500 Modbus RS485 Explanation of abbreviated units
Corrected Sft³, Sgal (us), Sbbl (us;liq.) Standard cubic foot, standard gallon, standard barrel
volume
fl oz (us), gal (us), kgal (us), Mgal Fluid ounce, gallon, kilogallon, million gallon
(us)
bbl (us;liq.), bbl (us;beer), bbl Barrel (normal liquids), barrel (beer), barrel
(us;oil), bbl (us;tank) (petrochemicals), barrel (filling tanks)
Endress+Hauser 255
Explanation of abbreviated units Proline Promass 500 Modbus RS485
256 Endress+Hauser
Proline Promass 500 Modbus RS485 Modbus RS485 Register Information
6.1 Notes
Name of parameter Indicated in decimal • Float length = Possible type of access Options Page number information
numerical format 4 byte to parameter: List of the individual and cross-reference to the
• Integer length = • Read access via options for the standard parameter
2 byte function codes 03, parameter description
• String length, 04 or 23 • Option 1
depending on • Write access via • Option 2
parameter function codes 06, • Option 3 (+)
16 or 23
• Factory
setting
highlighted in
bold
• (+) = Factory
setting
depends on
country, order
options or
device
settings
User entry
Input range for the
parameter
NOTICE
If non-volatile device parameters are modified via the MODBUS RS485 function codes
06, 16 or 23, the change is saved in the EEPROM of the measuring device.
The number of writes to the EEPROM is technically restricted to a maximum of 1 million.
‣ Make sure to comply with this limit since, if it is exceeded, data loss and measuring
device failure will result.
‣ Avoid constantly writing non-volatile device parameters via the MODBUS RS485.
Endress+Hauser 257
Modbus RS485 Register Information Proline Promass 500 Modbus RS485
Function code Access type Register in accordance with "Modbus Register in accordance with
Applications Protocol Specification" "Modicon Modbus Protocol
Reference Guide"
Expert
‣ System → 272
‣ Display → 272
258 Endress+Hauser
Proline Promass 500 Modbus RS485 Modbus RS485 Register Information
‣ Administration → 277
Endress+Hauser 259
Modbus RS485 Register Information Proline Promass 500 Modbus RS485
‣ Sensor → 278
‣ Totalizer → 278
260 Endress+Hauser
Proline Promass 500 Modbus RS485 Modbus RS485 Register Information
‣ Calibration → 288
Endress+Hauser 261
Modbus RS485 Register Information Proline Promass 500 Modbus RS485
C0 to 5 (6022) → 288
‣ Input → 289
262 Endress+Hauser
Proline Promass 500 Modbus RS485 Modbus RS485 Register Information
‣ Output → 290
‣ Curr.output 1 to n → 290
Endress+Hauser 263
Modbus RS485 Register Information Proline Promass 500 Modbus RS485
264 Endress+Hauser
Proline Promass 500 Modbus RS485 Modbus RS485 Register Information
Endress+Hauser 265
Modbus RS485 Register Information Proline Promass 500 Modbus RS485
‣ Communication → 300
266 Endress+Hauser
Proline Promass 500 Modbus RS485 Modbus RS485 Register Information
‣ Application → 302
‣ Totalizer 1 to n → 303
‣ Diagnostics → 304
Endress+Hauser 267
Modbus RS485 Register Information Proline Promass 500 Modbus RS485
268 Endress+Hauser
Proline Promass 500 Modbus RS485 Modbus RS485 Register Information
‣ Tors.oscil.freq. → 308
Endress+Hauser 269
Modbus RS485 Register Information Proline Promass 500 Modbus RS485
‣ Tors.oscil.damp. → 308
‣ Tors.sig.asymm. → 309
‣ Simulation → 311
270 Endress+Hauser
Proline Promass 500 Modbus RS485 Modbus RS485 Register Information
Ent. access code (0003) 2177 Integer Read / Write Max. 16-digit character string 14
comprising numbers, letters and special
characters
Endress+Hauser 271
Modbus RS485 Register Information Proline Promass 500 Modbus RS485
"Display" submenu
272 Endress+Hauser
Proline Promass 500 Modbus RS485 Modbus RS485 Register Information
100% bargraph 1 (0125) 4142 to 4143 Float Read / Write Signed floating-point number 21
Endress+Hauser 273
Modbus RS485 Register Information Proline Promass 500 Modbus RS485
Value 2 display (0108) 3964 Integer Read / Write For the picklist, see the Value 2 display 22
parameter (→ 22)
Value 3 display (0110) 3966 Integer Read / Write For the picklist, see the Value 2 display 23
parameter (→ 22)
100% bargraph 3 (0126) 4140 to 4141 Float Read / Write Signed floating-point number 24
Value 4 display (0109) 3965 Integer Read / Write For the picklist, see the Value 2 display 25
parameter (→ 22)
Header text (0112) 3968 to 3973 String Read / Write Max. 12 characters such as letters, 28
numbers or special characters (e.g. @,
%, /)
Operating time (0652) 2631 Integer Read Days (d), hours (h), minutes (m) and 30
seconds (s)
Last backup (2757) 6430 Integer Read Days (d), hours (h), minutes (m) and 30
seconds (s)
274 Endress+Hauser
Proline Promass 500 Modbus RS485 Modbus RS485 Register Information
Endress+Hauser 275
Modbus RS485 Register Information Proline Promass 500 Modbus RS485
276 Endress+Hauser
Proline Promass 500 Modbus RS485 Modbus RS485 Register Information
"Administration" submenu
Activate SW opt. (0029) 2795 Integer Read / Write Max. 10-digit string consisting of 48
numbers.
Def. access code 8677 to 8684 String Read / Write Max. 16-digit character string 45
comprising numbers, letters and special
characters
Confirm code 8685 to 8692 String Read / Write Max. 16-digit character string 46
comprising numbers, letters and special
characters
Operating time (0652) 2631 Integer Read Days (d), hours (h), minutes (m) and 46
seconds (s)
Reset acc. code (0024) 8880 to 8895 String Read / Write Character string comprising numbers, 46
letters and special characters
Endress+Hauser 277
Modbus RS485 Register Information Proline Promass 500 Modbus RS485
Mass flow (1838) 2007 to 2008 Float Read Signed floating-point number 51
Volume flow (1847) 2009 to 2010 Float Read Signed floating-point number 52
Pressure value (6129) 2089 to 2090 Float Read Signed floating-point number 53
Dynam. viscosity (1854) 2019 to 2020 Float Read Signed floating-point number 53
Kinematic visc. (1857) 2083 to 2084 Float Read Signed floating-point number 54
Target mass flow (1864) 2797 to 2798 Float Read Signed floating-point number 55
Carrier mass fl. (1865) 2799 to 2800 Float Read Signed floating-point number 56
"Totalizer" submenu
Totalizer val. 1 to n (0911–1 to n) 1: 2610 to 2611 Float Read Signed floating-point number 58
2: 2810 to 2811
3: 3010 to 3011
Tot. overflow 1 to n (0910–1 to n) 1: 2612 to 2613 Float Read Integer with sign 58
2: 2812 to 2813
3: 3012 to 3013
Measured val. 1 to n (1603–1 to n) 1: 6151 to 6152 Float Read Signed floating-point number 60
2: 6153 to 6154
3: 6155 to 6156
278 Endress+Hauser
Proline Promass 500 Modbus RS485 Modbus RS485 Register Information
Navigation: Expert → Sensor → Measured val. → Output values → Double pulse out
Pulse output (0987) 7041 to 7042 Float Read Positive floating-point number 65
Endress+Hauser 279
Modbus RS485 Register Information Proline Promass 500 Modbus RS485
280 Endress+Hauser
Proline Promass 500 Modbus RS485 Modbus RS485 Register Information
Endress+Hauser 281
Modbus RS485 Register Information Proline Promass 500 Modbus RS485
282 Endress+Hauser
Proline Promass 500 Modbus RS485 Modbus RS485 Register Information
Endress+Hauser 283
Modbus RS485 Register Information Proline Promass 500 Modbus RS485
284 Endress+Hauser
Proline Promass 500 Modbus RS485 Modbus RS485 Register Information
Corr. vol. text (0592) String Read / Write Max. 10 characters such as letters, 77
numbers or special characters (@, %, /)
Endress+Hauser 285
Modbus RS485 Register Information Proline Promass 500 Modbus RS485
Low value (1861) 5110 to 5111 Float Read / Write Signed floating-point number 88
High value (1858) 5112 to 5113 Float Read / Write Signed floating-point number 89
Max. damping (6040) 2414 to 2415 Float Read / Write Positive floating-point number 89
Temp. coeff. SV (6181) 7411 to 7412 Float Read / Write Positive floating-point number 92
Pressure value (6059) 5185 to 5186 Float Read / Write Positive floating-point number 93
External press. (6209) 2440 to 2441 Float Read / Write Positive floating-point number 93
286 Endress+Hauser
Proline Promass 500 Modbus RS485 Modbus RS485 Register Information
Ext. ref.density (6198) 2509 to 2510 Float Read / Write Floating point number with sign 96
Fix ref.density (1814) 5130 to 5131 Float Read / Write Positive floating-point number 97
Linear exp coeff (1817) 5132 to 5133 Float Read / Write Signed floating-point number 98
Square exp coeff (1818) 5134 to 5135 Float Read / Write Signed floating-point number 98
Inst. angle roll (6282) 2660 to 2661 Float Read / Write –180 to 180 ° 99
Endress+Hauser 287
Modbus RS485 Register Information Proline Promass 500 Modbus RS485
Mass flow offset (1831) 5521 to 5522 Float Read / Write Signed floating-point number 102
Mass flow factor (1832) 5519 to 5520 Float Read / Write Positive floating-point number 102
Vol. flow offset (1841) 5525 to 5526 Float Read / Write Signed floating-point number 103
Vol. flow factor (1846) 5523 to 5524 Float Read / Write Positive floating-point number 103
Density offset (1848) 5529 to 5530 Float Read / Write Signed floating-point number 103
Density factor (1849) 5527 to 5528 Float Read / Write Positive floating-point number 104
Corr. vol offset (1866) 2044 to 2045 Float Read / Write Signed floating-point number 104
Corr. vol factor (1867) 2076 to 2077 Float Read / Write Positive floating-point number 104
Ref.dens. offset (1868) 2046 to 2047 Float Read / Write Signed floating-point number 105
Ref.dens. factor (1869) 2042 to 2043 Float Read / Write Positive floating-point number 105
Temp. offset (1870) 5533 to 5534 Float Read / Write Signed floating-point number 105
Temp. factor (1871) 5531 to 5532 Float Read / Write Positive floating-point number 106
"Calibration" submenu
Cal. factor (6025) 7513 to 7514 Float Read Signed floating-point number 106
Zero point (6195) 7527 to 7528 Float Read / Write Signed floating-point number 107
Nominal diameter (2807) 2048 to 2057 String Read DNxx / x" 107
288 Endress+Hauser
Proline Promass 500 Modbus RS485 Modbus RS485 Register Information
0/4 mA value (1606–1 to n) 1: 6111 to 6112 Float Read / Write Signed floating-point number 112
2: 6113 to 6114
3: 6115 to 6116
20 mA value (1607–1 to n) 1: 6119 to 6120 Float Read / Write Signed floating-point number 112
2: 6121 to 6122
3: 6123 to 6124
Failure mode (1601–1 to n) 1: 6159 Integer Read / Write 1 = Last valid value 112
2: 6160 2 = Alarm
3: 6161 6 = Defined value
Failure value (1602–1 to n) 1: 6163 to 6164 Float Read / Write Signed floating-point number 113
2: 6165 to 6166
3: 6167 to 6168
Endress+Hauser 289
Modbus RS485 Register Information Proline Promass 500 Modbus RS485
290 Endress+Hauser
Proline Promass 500 Modbus RS485 Modbus RS485 Register Information
0/4 mA value (0367–1 to n) 1: 6195 to 6196 Float Read / Write Signed floating-point number 120
2: 6197 to 6198
3: 6199 to 6200
Endress+Hauser 291
Modbus RS485 Register Information Proline Promass 500 Modbus RS485
20 mA value (0372–1 to n) 1: 5915 to 5916 Float Read / Write Signed floating-point number 121
2: 5917 to 5918
3: 5919 to 5920
Damping out. 1 to n (0363–1 to n) 1: 5903 to 5904 Float Read / Write 0.0 to 999.9 s 127
2: 5905 to 5906
3: 5907 to 5908
Output curr. 1 to n (0361–1 to n) 1: 5931 to 5932 Float Read 3.59 to 22.5 mA 130
2: 5933 to 5934
3: 5935 to 5936
292 Endress+Hauser
Proline Promass 500 Modbus RS485 Modbus RS485 Register Information
Value per pulse (0455–1 to n) 1: 3034 to 3035 Float Read / Write Positive floating point number 134
2: 3036 to 3037
3: 4714 to 4715
Pulse width (0452–1 to n) 1: 2836 to 2837 Float Read / Write 0.05 to 2 000 ms 135
2: 2838 to 2839
3: 4702 to 4703
Pulse output 1 to n (0456–1 to n) 1: 3082 to 3083 Float Read Positive floating-point number 137
2: 3084 to 3085
3: 4718 to 4719
Endress+Hauser 293
Modbus RS485 Register Information Proline Promass 500 Modbus RS485
Min. freq. value (0453–1 to n) 1: 3526 to 3527 Float Read / Write 0.0 to 10 000.0 Hz 139
2: 3528 to 3529
3: 5767 to 5768
Max. freq. value (0454–1 to n) 1: 2996 to 2997 Float Read / Write 0.0 to 10 000.0 Hz 139
2: 2998 to 2999
3: 4710 to 4711
Val. at min.freq (0476–1 to n) 1: 5887 to 5888 Float Read / Write Signed floating-point number 140
2: 5889 to 5890
3: 5891 to 5892
294 Endress+Hauser
Proline Promass 500 Modbus RS485 Modbus RS485 Register Information
Val. at max.freq (0475–1 to n) 1: 3514 to 3515 Float Read / Write Signed floating-point number 140
2: 3516 to 3517
3: 5759 to 5760
Response time (0491–1 to n) 1: 5875 to 5876 Float Read Positive floating-point number 143
2: 5877 to 5878
3: 5879 to 5880
Failure freq. (0474–1 to n) 1: 3510 to 3511 Float Read / Write 0.0 to 12 500.0 Hz 144
2: 3512 to 3513
3: 9908 to 9909
Output freq. 1 to n (0471–1 to n) 1: 3462 to 3463 Float Read 0.0 to 12 500.0 Hz 145
2: 3464 to 3465
3: 9910 to 9911
Endress+Hauser 295
Modbus RS485 Register Information Proline Promass 500 Modbus RS485
Switch-on value (0466–1 to n) 1: 3242 to 3243 Float Read / Write Signed floating-point number 149
2: 3244 to 3245
3: 4728 to 4729
Switch-off value (0464–1 to n) 1: 3234 to 3235 Float Read / Write Signed floating-point number 149
2: 3236 to 3237
3: 4724 to 4725
Assign status (0485–1 to n) 1: 3374 Integer Read / Write 0 = Low flow cut off 150
2: 3375 1 = Partial pipe det
3: 4734
Switch-on delay (0467–1 to n) 1: 6247 to 6248 Float Read / Write 0.0 to 100.0 s 150
2: 6249 to 6250
3: 6251 to 6252
Switch-off delay (0465–1 to n) 1: 6239 to 6240 Float Read / Write 0.0 to 100.0 s 151
2: 6241 to 6242
3: 6243 to 6244
296 Endress+Hauser
Proline Promass 500 Modbus RS485 Modbus RS485 Register Information
Endress+Hauser 297
Modbus RS485 Register Information Proline Promass 500 Modbus RS485
Assign status (0805–1 to n) 1: 8272 Integer Read / Write 0 = Low flow cut off 156
2: 8273 1 = Partial pipe det
3: 8274
Switch-off value (0809–1 to n) 1: 8260 to 8261 Float Read / Write Signed floating-point number 156
2: 8262 to 8263
3: 8264 to 8265
Switch-off delay (0813–1 to n) 1: 8254 to 8255 Float Read / Write 0.0 to 100.0 s 157
2: 8256 to 8257
3: 8258 to 8259
Switch-on value (0810–1 to n) 1: 8233 to 8234 Float Read / Write Signed floating-point number 157
2: 8235 to 8236
3: 8237 to 8238
Switch-on delay (0814–1 to n) 1: 8266 to 8267 Float Read / Write 0.0 to 100.0 s 158
2: 8268 to 8269
3: 8270 to 8271
298 Endress+Hauser
Proline Promass 500 Modbus RS485 Modbus RS485 Register Information
Master term. no. (0981) 5838 Integer Read 0 = Not used 160
2 = 24-25 (I/O 2)
3 = 22-23 (I/O 3)
Slave term. no. (0990) 5845 Integer Read 0 = Not used 160
2 = 24-25 (I/O 2)
3 = 22-23 (I/O 3)
Value per pulse (0983) 7495 to 7496 Float Read / Write Signed floating-point number 161
Pulse width (0986) 6998 to 6999 Float Read / Write 0.5 to 2 000 ms 162
Endress+Hauser 299
Modbus RS485 Register Information Proline Promass 500 Modbus RS485
Pulse output (0987) 7041 to 7042 Float Read Positive floating-point number 164
Failure mode (7116) 4920 Integer Read / Write 1 = Last valid value 168
255 = NaN value
Device revision (7154) 4481 Integer Read 4-digit hexadecimal number 170
300 Endress+Hauser
Proline Promass 500 Modbus RS485 Modbus RS485 Register Information
MAC Address (7214) 4210 to 4217.5 String Read Unique 12-digit character string 172
comprising letters and numbers
IP address (7209) 4155 to 4161.5 String Read 4 octet: 0 to 255 (in the particular octet) 173
Subnet mask (7211) 4163 to 4169.5 String Read 4 octet: 0 to 255 (in the particular octet) 173
Default gateway (7210) 4171 to 4177.5 String Read 4 octet: 0 to 255 (in the particular octet) 173
Endress+Hauser 301
Modbus RS485 Register Information Proline Promass 500 Modbus RS485
WLAN IP address (2711) 8643 to 8649.5 String Read / Write 4 octet: 0 to 255 (in the particular octet) 177
WLAN subnet mask (2709) 8651 to 8657.5 String Read / Write 4 octet: 0 to 255 (in the particular octet) 178
WLAN MAC address (2703) 8602 to 8609.5 String Read Unique 12-digit character string 178
comprising letters and numbers
WLAN passphrase (2706) 8611 to 8626 String Read / Write 8 to 32-digit character string comprising 178
numbers, letters and special characters
(without spaces)
Assign SSID name (2708) 6218 Integer Read / Write 0 = Device tag 179
1 = User-defined
SSID name (2707) 8627 to 8642 String Read / Write Max. 32-digit character string 179
comprising numbers, letters and special
characters
302 Endress+Hauser
Proline Promass 500 Modbus RS485 Modbus RS485 Register Information
Operation mode (0908–1 to n) 1: 2605 Integer Read / Write 0 = Net flow total 184
2: 2805 1 = Forward total
3: 3005 2 = Reverse total
Endress+Hauser 303
Modbus RS485 Register Information Proline Promass 500 Modbus RS485
Preset value 1 to n (0913–1 to n) 1: 2590 to 2591 Float Read / Write Signed floating-point number 185
2: 2592 to 2593
3: 2594 to 2595
Actual diagnos. (0691) 2732 Integer Read Symbol for diagnostic behavior, 196
diagnostic code and short message.
Prev.diagnostics (0690) 2734 Integer Read Symbol for diagnostic behavior, 196
diagnostic code and short message.
Time fr. restart (0653) 2624 Integer Read Days (d), hours (h), minutes (m) and 197
seconds (s)
Operating time (0652) 2631 Integer Read Days (d), hours (h), minutes (m) and 197
seconds (s)
Diagnostics 1 (0692) 2736 Integer Read Symbol for diagnostic behavior, 198
diagnostic code and short message.
Diagnostics 2 (0693) 2738 Integer Read Symbol for diagnostic behavior, 199
diagnostic code and short message.
Diagnostics 3 (0694) 2740 Integer Read Symbol for diagnostic behavior, 200
diagnostic code and short message.
Diagnostics 4 (0695) 2742 Integer Read Symbol for diagnostic behavior, 200
diagnostic code and short message.
Diagnostics 5 (0696) 2744 Integer Read Symbol for diagnostic behavior, 201
diagnostic code and short message.
304 Endress+Hauser
Proline Promass 500 Modbus RS485 Modbus RS485 Register Information
Device tag (0011) 2026 to 2041 String Read Max. 32 characters, such as letters, 205
numbers or special characters (e.g. @,
%, /).
Serial number (0009) 7003 to 7007.5 String Read Max. 11-digit character string 206
comprising letters and numbers.
Firmware version (0010) 7277 to 7280 String Read Character string in the format xx.yy.zz 206
Device name (0020) 7238 to 7245 String Read Promass 300/500 206
Order code (0008) 2058 to 2067 String Read Character string composed of letters, 206
numbers and certain punctuation marks
(e.g. /).
Ext. order cd. 1 (0023) 2212 to 2221 String Read Character string 207
Ext. order cd. 2 (0021) 2222 to 2231 String Read Character string 207
Ext. order cd. 3 (0022) 2232 to 2241 String Read Character string 207
ENP version (0012) 4003 to 4010 String Read Character string 208
Endress+Hauser 305
Modbus RS485 Register Information Proline Promass 500 Modbus RS485
306 Endress+Hauser
Proline Promass 500 Modbus RS485 Modbus RS485 Register Information
Minimum value (0688) 4651 to 4652 Float Read Signed floating-point number 227
Maximum value (0665) 4649 to 4650 Float Read Signed floating-point number 227
Minimum value (6052) 2421 to 2422 Float Read Signed floating-point number 228
Maximum value (6051) 2419 to 2420 Float Read Signed floating-point number 228
Minimum value (6109) 7529 to 7530 Float Read Signed floating-point number 229
Maximum value (6108) 7531 to 7532 Float Read Signed floating-point number 229
Minimum value (6030) 7533 to 7534 Float Read Signed floating-point number 230
Maximum value (6029) 7535 to 7536 Float Read Signed floating-point number 230
Endress+Hauser 307
Modbus RS485 Register Information Proline Promass 500 Modbus RS485
Minimum value (6071) 2459 to 2460 Float Read Signed floating-point number 231
Maximum value (6070) 2468 to 2469 Float Read Signed floating-point number 231
"Tors.oscil.freq." submenu
Minimum value (6069) 2519 to 2520 Float Read Signed floating-point number 232
Maximum value (6068) 2517 to 2518 Float Read Signed floating-point number 232
Minimum value (6010) 2472 to 2473 Float Read Signed floating-point number 232
Maximum value (6009) 2470 to 2471 Float Read Signed floating-point number 233
Minimum value (6008) 2515 to 2516 Float Read Signed floating-point number 233
Maximum value (6007) 2480 to 2481 Float Read Signed floating-point number 233
Minimum value (6122) 2478 to 2479 Float Read Signed floating-point number 234
Maximum value (6121) 2423 to 2424 Float Read Signed floating-point number 234
"Tors.oscil.damp." submenu
Minimum value (6120) 2523 to 2524 Float Read Signed floating-point number 235
Maximum value (6119) 2521 to 2522 Float Read Signed floating-point number 235
308 Endress+Hauser
Proline Promass 500 Modbus RS485 Modbus RS485 Register Information
Minimum value (6015) 2474 to 2475 Float Read Signed floating-point number 235
Maximum value (6014) 2476 to 2477 Float Read Signed floating-point number 236
"Tors.sig.asymm." submenu
Minimum value (6284) 7000 to 7001 Float Read Signed floating-point number 236
Maximum value (6283) 6924 to 6925 Float Read Signed floating-point number 236
Endress+Hauser 309
Modbus RS485 Register Information Proline Promass 500 Modbus RS485
Assign chan. 2 (0852) 2446 Integer Read / Write Picklist, see Assign channel 1 parameter 219
(→ 217)
Assign chan. 3 (0853) 2548 Integer Read / Write Picklist, see Assign channel 1 parameter 219
(→ 217)
310 Endress+Hauser
Proline Promass 500 Modbus RS485 Modbus RS485 Register Information
Assign chan. 4 (0854) 4286 Integer Read / Write Picklist, see Assign channel 1 parameter 219
(→ 217)
Logging interval (0856) 4288 to 4289 Float Read / Write 0.1 to 3 600.0 s 220
Logging duration (0861) 2827 to 2828 Float Read Positive floating-point number 223
"Simulation" submenu
Proc. var. value (1811) 6814 to 6815 Float Read / Write Depends on the process variable selected 239
Endress+Hauser 311
Modbus RS485 Register Information Proline Promass 500 Modbus RS485
Value curr.out 1 to n (0355–1 to n) 1: 5995 to 5996 Float Read / Write 3.59 to 22.5 mA 241
2: 5997 to 5998
3: 5999 to 6000
Freq value 1 to n (0473–1 to n) 1: 6207 to 6208 Float Read / Write 0.0 to 12 500.0 Hz 242
2: 6209 to 6210
3: 6211 to 6212
312 Endress+Hauser
Proline Promass 500 Modbus RS485 Index
Index
0…9 Assign frequency output (Parameter) . . . . . . . . . . . . 138
0/4 mA value (Parameter) . . . . . . . . . . . . . . . . 112, 120 Assign limit (Parameter) . . . . . . . . . . . . . . . . . 146, 154
0% bargraph value 1 (Parameter) . . . . . . . . . . . . . . . 21 Assign process variable (Parameter) . . . . . . . 85, 88, 182
0% bargraph value 3 (Parameter) . . . . . . . . . . . . . . . 24 Assign pulse output 1 (Parameter) . . . . . . . . . . . . . . 161
2.4 GHz WLAN channel (Parameter) . . . . . . . . . . . . 179 Assign pulse output 1 to n (Parameter) . . . . . . . . . . 134
20 mA value (Parameter) . . . . . . . . . . . . . . . . 112, 121 Assign simulation process variable (Parameter) . . . . 238
100% bargraph value 1 (Parameter) . . . . . . . . . . . . . 21 Assign SSID name (Parameter) . . . . . . . . . . . . . . . . 179
100% bargraph value 3 (Parameter) . . . . . . . . . . . . . 24 Assign status (Parameter) . . . . . . . . . . . . . . . . 150, 156
Assign status input (Parameter) . . . . . . . . . . . . . . . . 114
A
Access status (Parameter) . . . . . . . . . . . . . . . . . . . . . 14 B
Activate SW option (Parameter) . . . . . . . . . . . . . . . . 48 Backlight (Parameter) . . . . . . . . . . . . . . . . . . . . . . . . 29
Active level (Parameter) . . . . . . . . . . . . . . . . . . . . . 115 Backup state (Parameter) . . . . . . . . . . . . . . . . . . . . . 31
Actual diagnostics (Parameter) . . . . . . . . . . . . . . . . 196 Baudrate (Parameter) . . . . . . . . . . . . . . . . . . . . . . . 166
Administration (Submenu) . . . . . . . . . . . . . . . . . . . . 44 Bootloader revision (Parameter) 209, 210, 211, 212,
Alarm delay (Parameter) . . . . . . . . . . . . . . . . . . . . . . 33 214, 215, 216
Application (Submenu) . . . . . . . . . . . . . . . . . . . . . . 181 Build no. software (Parameter) 209, 210, 211, 212,
Application specific calculations (Submenu) . . . . . . . 187 214, 215, 216
Application specific input 0 (Parameter) . . . . . . . . . 191 Bus address (Parameter) . . . . . . . . . . . . . . . . . . . . . 165
Application specific input 1 (Parameter) . . . . . . . . . 192 Bus termination (Parameter) . . . . . . . . . . . . . . . . . . 169
Application specific output 0 (Parameter) . . . . . . . . . 192 Byte order (Parameter) . . . . . . . . . . . . . . . . . . . . . . 167
Application specific output 1 (Parameter) . . . . . . . . . 193
Application-specific parameters (Submenu) . . . . . . . 188 C
Apply I/O configuration (Parameter) . . . . . . . . . . . . 109 C0 to 5 (Parameter) . . . . . . . . . . . . . . . . . . . . . . . . . 107
Assign behavior of diagnostic no. 046 (Parameter) . . . 35 Calculated values (Submenu) . . . . . . . . . . . . . . . . . . . 95
Assign behavior of diagnostic no. 140 (Parameter) . . . 35 Calibration (Submenu) . . . . . . . . . . . . . . . . . . . . . . 106
Assign behavior of diagnostic no. 144 (Parameter) . . . 36 Calibration factor (Parameter) . . . . . . . . . . . . . . . . . 106
Assign behavior of diagnostic no. 302 (Parameter) . . . 37 Carrier corrected volume flow (Parameter) . . . . . . . . . 56
Assign behavior of diagnostic no. 374 (Parameter) . . . 36 Carrier mass flow (Parameter) . . . . . . . . . . . . . . . . . . 56
Assign behavior of diagnostic no. 441 (Parameter) . . . 37 Carrier pipe temperature (Submenu) . . . . . . . . . . . . 229
Assign behavior of diagnostic no. 442 (Parameter) . . . 37 Carrier volume flow (Parameter) . . . . . . . . . . . . . . . . 57
Assign behavior of diagnostic no. 443 (Parameter) . . . 38 Clear logging data (Parameter) . . . . . . . . . . . . . . . . 220
Assign behavior of diagnostic no. 444 (Parameter) . . . 38 Communication (Submenu) . . . . . . . . . . . . . . . . . . . 164
Assign behavior of diagnostic no. 543 (Parameter) . . . 38 Comparison result (Parameter) . . . . . . . . . . . . . . . . . 32
Assign behavior of diagnostic no. 830 (Parameter) . . . 39 Concentration (Parameter) . . . . . . . . . . . . . . . . . . . . 55
Assign behavior of diagnostic no. 831 (Parameter) . . . 39 Concentration (Submenu) . . . . . . . . . . . . . . . . . . . . 187
Assign behavior of diagnostic no. 832 (Parameter) . . . 39 Configuration backup (Submenu) . . . . . . . . . . . . . . . . 30
Assign behavior of diagnostic no. 833 (Parameter) . . . 40 Configuration management (Parameter) . . . . . . . . . . 31
Assign behavior of diagnostic no. 834 (Parameter) . . . 40 Confirm access code (Parameter) . . . . . . . . . . . . . . . . 46
Assign behavior of diagnostic no. 835 (Parameter) . . . 40 Connection state (Parameter) . . . . . . . . . . . . . . . . . 180
Assign behavior of diagnostic no. 842 (Parameter) . . . 41 Contrast display (Parameter) . . . . . . . . . . . . . . . . . . . 29
Assign behavior of diagnostic no. 862 (Parameter) . . . 41 Control Totalizer 1 to n (Parameter) . . . . . . . . . . . . . 185
Assign behavior of diagnostic no. 912 (Parameter) . . . 42 Corrected volume flow (Parameter) . . . . . . . . . . . . . . 52
Assign behavior of diagnostic no. 913 (Parameter) . . . 42 Corrected volume flow calculation (Parameter) . . . . . . 96
Assign behavior of diagnostic no. 941 (Parameter) . . . 42 Corrected volume flow calculation (Submenu) . . . . . . 95
Assign behavior of diagnostic no. 942 (Parameter) . . . 43 Corrected volume flow factor (Parameter) . . . . . . . . 104
Assign behavior of diagnostic no. 943 (Parameter) . . . 43 Corrected volume flow offset (Parameter) . . . . . . . . 104
Assign behavior of diagnostic no. 944 (Parameter) . . . 44 Corrected volume flow unit (Parameter) . . . . . . . . . . . 69
Assign behavior of diagnostic no. 948 (Parameter) . . . 44 Corrected volume unit (Parameter) . . . . . . . . . . . . . . 70
Assign channel 1 (Parameter) . . . . . . . . . . . . . . . . . 217 Current input 1 to n (Submenu) . . . . . . . . . . . . . 59, 110
Assign channel 2 (Parameter) . . . . . . . . . . . . . . . . . 219 Current input 1 to n simulation (Parameter) . . . . . . . 240
Assign channel 3 (Parameter) . . . . . . . . . . . . . . . . . 219 Current output 1 to n (Submenu) . . . . . . . . . . . . . . . 116
Assign channel 4 (Parameter) . . . . . . . . . . . . . . . . . 219 Current output 1 to n simulation (Parameter) . . . . . . 241
Assign current output 1 to n (Parameter) . . . . . . . . . 117 Current span (Parameter) . . . . . . . . . . . . . . . . 111, 118
Assign diagnostic behavior (Parameter) . . . . . . 146, 155 Custody transfer (Submenu) . . . . . . . . . . . . . . . . . . 187
Assign flow direction check (Parameter) . . . . . . 150, 154 Custody transfer logbook (Submenu) . . . . . . . . . . . . 204
Endress+Hauser 313
Index Proline Promass 500 Modbus RS485
Cut off inhomogeneous liquid (Parameter) . . . . . . . . 194 Actual diagnostics (0691) . . . . . . . . . . . . . . . . . . 196
Cut off inhomogeneous wet gas (Parameter) . . . . . . 193 Alarm delay (0651) . . . . . . . . . . . . . . . . . . . . . . . 33
Cut off suspended bubbles (Parameter) . . . . . . . . . . 194 Application specific input 0 (6366) . . . . . . . . . . . 191
Application specific input 1 (6367) . . . . . . . . . . . 192
D Application specific output 0 (6364) . . . . . . . . . . 192
Damping output 1 to n (Parameter) . . . . . . . . . 127, 142 Application specific output 1 (6365) . . . . . . . . . . 193
Data logging (Parameter) . . . . . . . . . . . . . . . . . . . . 221 Apply I/O configuration (3907) . . . . . . . . . . . . . 109
Data logging (Submenu) . . . . . . . . . . . . . . . . . . . . . 216 Assign behavior of diagnostic no. 046 (0709) . . . . 35
Data logging control (Parameter) . . . . . . . . . . . . . . . 222 Assign behavior of diagnostic no. 140 (0708) . . . . 35
Data logging status (Parameter) . . . . . . . . . . . . . . . 222 Assign behavior of diagnostic no. 144 (0731) . . . . 36
Data transfer mode (Parameter) . . . . . . . . . . . . . . . 166 Assign behavior of diagnostic no. 302 (0739) . . . . 37
Date/time format (Parameter) . . . . . . . . . . . . . . . . . . 73 Assign behavior of diagnostic no. 374 (0710) . . . . 36
Decimal places 1 (Parameter) . . . . . . . . . . . . . . . . . . 22 Assign behavior of diagnostic no. 441 (0657) . . . . 37
Decimal places 2 (Parameter) . . . . . . . . . . . . . . . . . . 23 Assign behavior of diagnostic no. 442 (0658) . . . . 37
Decimal places 3 (Parameter) . . . . . . . . . . . . . . . . . . 25 Assign behavior of diagnostic no. 443 (0659) . . . . 38
Decimal places 4 (Parameter) . . . . . . . . . . . . . . . . . . 26 Assign behavior of diagnostic no. 444 (0740) . . . . 38
Default gateway (Parameter) . . . . . . . . . . . . . . . . . . 173 Assign behavior of diagnostic no. 543 (0643) . . . . 38
Define access code (Parameter) . . . . . . . . . . . . . . . . . 45 Assign behavior of diagnostic no. 830 (0800) . . . . 39
Define access code (Wizard) . . . . . . . . . . . . . . . . . . . 45 Assign behavior of diagnostic no. 831 (0641) . . . . 39
Density (Parameter) . . . . . . . . . . . . . . . . . . . . . . . . . 52 Assign behavior of diagnostic no. 832 (0681) . . . . 39
Density damping (Parameter) . . . . . . . . . . . . . . . . . . 83 Assign behavior of diagnostic no. 833 (0682) . . . . 40
Density factor (Parameter) . . . . . . . . . . . . . . . . . . . 104 Assign behavior of diagnostic no. 834 (0700) . . . . 40
Density offset (Parameter) . . . . . . . . . . . . . . . . . . . . 103 Assign behavior of diagnostic no. 835 (0702) . . . . 40
Density unit (Parameter) . . . . . . . . . . . . . . . . . . . . . . 70 Assign behavior of diagnostic no. 842 (0638) . . . . 41
Device alarm simulation (Parameter) . . . . . . . . . . . . 246 Assign behavior of diagnostic no. 862 (0679) . . . . 41
Device ID (Parameter) . . . . . . . . . . . . . . . . . . . . . . . 170 Assign behavior of diagnostic no. 912 (0703) . . . . 42
Device information (Submenu) . . . . . . . . . . . . . . . . 205 Assign behavior of diagnostic no. 913 (0712) . . . . 42
Device name (Parameter) . . . . . . . . . . . . . . . . . . . . 206 Assign behavior of diagnostic no. 941 (0635) . . . . 42
Device reset (Parameter) . . . . . . . . . . . . . . . . . . . . . . 47 Assign behavior of diagnostic no. 942 (0636) . . . . 43
Device revision (Parameter) . . . . . . . . . . . . . . . . . . . 170 Assign behavior of diagnostic no. 943 (0637) . . . . 43
Device tag (Parameter) . . . . . . . . . . . . . . . . . . . . . . 205 Assign behavior of diagnostic no. 944 (0732) . . . . 44
DHCP client (Parameter) . . . . . . . . . . . . . . . . . . . . . 172 Assign behavior of diagnostic no. 948 (0744) . . . . 44
Diagnostic behavior (Submenu) . . . . . . . . . . . . . . . . . 33 Assign channel 1 (0851) . . . . . . . . . . . . . . . . . . 217
Diagnostic event category (Parameter) . . . . . . . . . . . 247 Assign channel 2 (0852) . . . . . . . . . . . . . . . . . . 219
Diagnostic event simulation (Parameter) . . . . . . . . . 247 Assign channel 3 (0853) . . . . . . . . . . . . . . . . . . 219
Diagnostic handling (Submenu) . . . . . . . . . . . . . . . . . 33 Assign channel 4 (0854) . . . . . . . . . . . . . . . . . . 219
Diagnostic list (Submenu) . . . . . . . . . . . . . . . . . . . . 198 Assign current output 1 to n (0359–1 to n) . . . . . 117
Diagnostics (Submenu) . . . . . . . . . . . . . . . . . . . . . . 195 Assign diagnostic behavior
Diagnostics 1 (Parameter) . . . . . . . . . . . . . . . . . . . . 198 Pulse/frequency/switch output 1 to n (0482–
Diagnostics 2 (Parameter) . . . . . . . . . . . . . . . . . . . . 199 1 to n) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 146
Diagnostics 3 (Parameter) . . . . . . . . . . . . . . . . . . . . 200 Relay output 1 to n (0806–1 to n) . . . . . . . . . 155
Diagnostics 4 (Parameter) . . . . . . . . . . . . . . . . . . . . 200 Assign flow direction check
Diagnostics 5 (Parameter) . . . . . . . . . . . . . . . . . . . . 201 Pulse/frequency/switch output 1 to n (0484–
Direct access 1 to n) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 150
0/4 mA value Relay output 1 to n (0808–1 to n) . . . . . . . . . 154
Current input 1 to n (1606–1 to n) . . . . . . . . 112 Assign frequency output
Current output 1 to n (0367–1 to n) . . . . . . . 120 Pulse/frequency/switch output 1 to n (0478–
0% bargraph value 1 (0123) . . . . . . . . . . . . . . . . . 21 1 to n) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 138
0% bargraph value 3 (0124) . . . . . . . . . . . . . . . . . 24 Assign limit
2.4 GHz WLAN channel (2704) . . . . . . . . . . . . . 179 Pulse/frequency/switch output 1 to n (0483–
20 mA value 1 to n) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 146
Current input 1 to n (1607–1 to n) . . . . . . . . 112 Relay output 1 to n (0807–1 to n) . . . . . . . . . 154
Current output 1 to n (0372–1 to n) . . . . . . . 121 Assign process variable
100% bargraph value 1 (0125) . . . . . . . . . . . . . . . 21 Totalizer 1 to n (0914–1 to n) . . . . . . . . . . . . 182
100% bargraph value 3 (0126) . . . . . . . . . . . . . . . 24 Assign process variable (1837) . . . . . . . . . . . . . . . 85
Access status (0005) . . . . . . . . . . . . . . . . . . . . . . 14 Assign process variable (1860) . . . . . . . . . . . . . . . 88
Activate SW option (0029) . . . . . . . . . . . . . . . . . . 48 Assign pulse output 1 (0982–1) . . . . . . . . . . . . . 161
Active level Assign pulse output 1 to n (0460–1 to n) . . . . . . 134
Status input 1 to n (1351–1 to n) . . . . . . . . . . 115 Assign simulation process variable (1810) . . . . . 238
314 Endress+Hauser
Proline Promass 500 Modbus RS485 Index
Endress+Hauser 315
Index Proline Promass 500 Modbus RS485
316 Endress+Hauser
Proline Promass 500 Modbus RS485 Index
Endress+Hauser 317
Index Proline Promass 500 Modbus RS485
Current output 1 to n (0379–1 to n) . . . . . . . 116 Zero point adjustment status (6253) . . . . . . . . . . 101
Pulse/frequency/switch output 1 to n (0492– Direct access (Parameter) . . . . . . . . . . . . . . . . . . . . . 12
1 to n) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 131 Display (Submenu) . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Relay output 1 to n (0812–1 to n) . . . . . . . . . 153 Display channel 1 (Submenu) . . . . . . . . . . . . . . . . . 223
Status input 1 to n (1358–1 to n) . . . . . . . . . . 113 Display channel 2 (Submenu) . . . . . . . . . . . . . . . . . 224
Timestamp . . . . . 196, 197, 198, 199, 200, 201, 202 Display channel 3 (Submenu) . . . . . . . . . . . . . . . . . 225
Totalizer operation mode Display channel 4 (Submenu) . . . . . . . . . . . . . . . . . 225
Totalizer 1 to n (0908–1 to n) . . . . . . . . . . . . 184 Display damping (Parameter) . . . . . . . . . . . . . . . . . . 27
Totalizer overflow 1 to n (0910–1 to n) . . . . . . . . 58 Display interval (Parameter) . . . . . . . . . . . . . . . . . . . 26
Totalizer value 1 to n (0911–1 to n) . . . . . . . . . . . 58 Display language (Parameter) . . . . . . . . . . . . . . . . . . 16
Unit totalizer 1 to n (0915–1 to n) . . . . . . . . . . . 183 Display module (Submenu) . . . . . . . . . . . . . . . . . . . 215
User corrected volume factor (0590) . . . . . . . . . . . 78 Document
User corrected volume offset (0602) . . . . . . . . . . . 78 Explanation of the structure of a parameter
User corrected volume text (0592) . . . . . . . . . . . . 77 description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
User density factor (0572) . . . . . . . . . . . . . . . . . . 79 Function . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
User density offset (0571) . . . . . . . . . . . . . . . . . . 79 Structure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
User density text (0570) . . . . . . . . . . . . . . . . . . . . 78 Symbols used . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
User energy factor (0586) . . . . . . . . . . . . . . . . . . 81 Target group . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
User energy offset (0599) . . . . . . . . . . . . . . . . . . . 81 Using the document . . . . . . . . . . . . . . . . . . . . . . . . 5
User energy text (0600) . . . . . . . . . . . . . . . . . . . . 80 Document function . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
User mass factor (0561) . . . . . . . . . . . . . . . . . . . . 77 Double pulse output (Submenu) . . . . . . . . . . . . . 64, 159
User mass offset (0562) . . . . . . . . . . . . . . . . . . . . 76 Dynamic viscosity (Parameter) . . . . . . . . . . . . . . . . . . 53
User mass text (0560) . . . . . . . . . . . . . . . . . . . . . 76
User name (2715) . . . . . . . . . . . . . . . . . . . . . . . 177 E
User pressure factor (0579) . . . . . . . . . . . . . . . . . 82 ENP version (Parameter) . . . . . . . . . . . . . . . . . . . . . 208
User pressure offset (0580) . . . . . . . . . . . . . . . . . 81 Enter access code (Parameter) . . . . . . . . . . . . . . . . . . 14
User pressure text (0581) . . . . . . . . . . . . . . . . . . . 81 Entire logging duration (Parameter) . . . . . . . . . . . . 223
User specific-enthalpy factor (0583) . . . . . . . . . . . 80 Event list (Submenu) . . . . . . . . . . . . . . . . . . . . . . . . 203
User specific-enthalpy offset (0584) . . . . . . . . . . . 80 Event logbook (Submenu) . . . . . . . . . . . . . . . . . . . . 202
User specific-enthalpy text (0585) . . . . . . . . . . . . 79 Extended order code 1 (Parameter) . . . . . . . . . . . . . 207
User volume factor (0568) . . . . . . . . . . . . . . . . . . 76 Extended order code 2 (Parameter) . . . . . . . . . . . . . 207
User volume offset (0569) . . . . . . . . . . . . . . . . . . 75 Extended order code 3 (Parameter) . . . . . . . . . . . . . 207
User volume text (0567) . . . . . . . . . . . . . . . . . . . . 75 External compensation (Submenu) . . . . . . . . . . . . . . . 92
Value 1 display (0107) . . . . . . . . . . . . . . . . . . . . . 19 External pressure (Parameter) . . . . . . . . . . . . . . . . . . 93
Value 2 display (0108) . . . . . . . . . . . . . . . . . . . . . 22 External reference density (Parameter) . . . . . . . . . . . 96
Value 3 display (0110) . . . . . . . . . . . . . . . . . . . . . 23 External temperature (Parameter) . . . . . . . . . . . . . . . 94
Value 4 display (0109) . . . . . . . . . . . . . . . . . . . . . 25
Value current input 1 to n (1609–1 to n) . . . . . . 241 F
Value current output 1 to n (0355–1 to n) . . . . . 241 Factory settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . 248
Value per pulse (0983) . . . . . . . . . . . . . . . . . . . . 161 SI units . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 248
Value status input US units . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 250
Status input 1 to n (1353–1 to n) . . . . . . . . . . 115 Fail-safe type application specific 0 (Parameter) . . . . 191
Value status input 1 to n (1353–1 to n) . . . . . . 60 Fail-safe type application specific 1 (Parameter) . . . . 192
Volume flow (1847) . . . . . . . . . . . . . . . . . . . . . . . 52 Fail-safe value application specific 0 (Parameter) . . . 191
Volume flow factor (1846) . . . . . . . . . . . . . . . . . 103 Fail-safe value application specific 1 (Parameter) . . . 192
Volume flow offset (1841) . . . . . . . . . . . . . . . . . 103 Failure current (Parameter) . . . . . . . . . . . . . . . . . . . 129
Volume flow unit (0553) . . . . . . . . . . . . . . . . . . . 67 Failure frequency (Parameter) . . . . . . . . . . . . . . . . . 144
Volume unit (0563) . . . . . . . . . . . . . . . . . . . . . . . 68 Failure mode (Parameter) 112, 128, 136, 144, 151,
Web server functionality (7222) . . . . . . . . . . . . . 174 158, 163, 168, 186
Web server language (7221) . . . . . . . . . . . . . . . 172 Failure value (Parameter) . . . . . . . . . . . . . . . . . . . . 113
WLAN (2702) . . . . . . . . . . . . . . . . . . . . . . . . . . 175 Fieldbus writing access (Parameter) . . . . . . . . . . . . . 169
WLAN IP address (2711) . . . . . . . . . . . . . . . . . . 177 Filter options (Parameter) . . . . . . . . . . . . . . . . 202, 203
WLAN MAC address (2703) . . . . . . . . . . . . . . . . 178 Firmware version (Parameter) . . . . . . . . . . . . . . . . . 206
WLAN mode (2717) . . . . . . . . . . . . . . . . . . . . . 176 Fixed current (Parameter) . . . . . . . . . . . . . . . . . . . . 119
WLAN passphrase (2706) . . . . . . . . . . . . . . . . . 178 Fixed reference density (Parameter) . . . . . . . . . . . . . . 97
WLAN password (2716) . . . . . . . . . . . . . . . . . . . 177 Flow damping (Parameter) . . . . . . . . . . . . . . . . . . . . 82
WLAN subnet mask (2709) . . . . . . . . . . . . . . . . 178 Flow override (Parameter) . . . . . . . . . . . . . . . . . . . . . 84
Zero point (6195) . . . . . . . . . . . . . . . . . . . . . . . . 107 Format display (Parameter) . . . . . . . . . . . . . . . . . . . . 17
Zero point adjustment control (6196) . . . . . . . . . 100 Frequency output simulation 1 to n (Parameter) . . . . 242
318 Endress+Hauser
Proline Promass 500 Modbus RS485 Index
Endress+Hauser 319
Index Proline Promass 500 Modbus RS485
320 Endress+Hauser
Proline Promass 500 Modbus RS485 Index
Endress+Hauser 321
Index Proline Promass 500 Modbus RS485
W
Web server (Submenu) . . . . . . . . . . . . . . . . . . . . . . 171
Web server functionality (Parameter) . . . . . . . . . . . 174
Web server language (Parameter) . . . . . . . . . . . . . . 172
Wizard
Define access code . . . . . . . . . . . . . . . . . . . . . . . . 45
WLAN (Parameter) . . . . . . . . . . . . . . . . . . . . . . . . . 175
WLAN IP address (Parameter) . . . . . . . . . . . . . . . . . 177
WLAN MAC address (Parameter) . . . . . . . . . . . . . . 178
WLAN mode (Parameter) . . . . . . . . . . . . . . . . . . . . 176
WLAN passphrase (Parameter) . . . . . . . . . . . . . . . . 178
WLAN password (Parameter) . . . . . . . . . . . . . . . . . 177
WLAN settings (Submenu) . . . . . . . . . . . . . . . . . . . 174
WLAN subnet mask (Parameter) . . . . . . . . . . . . . . . 178
Z
Zero point (Parameter) . . . . . . . . . . . . . . . . . . . . . . 107
Zero point adjustment (Submenu) . . . . . . . . . . . . . . 100
Zero point adjustment control (Parameter) . . . . . . . . 100
Zero point adjustment status (Parameter) . . . . . . . . 101
322 Endress+Hauser