Gp01062den 0519

This document provides a description of device parameters for a Proline Promass 500 Coriolis flowmeter that communicates via Modbus RS485. It includes: - An overview of the device's Expert operating menu structure with descriptions of system, sensor, I/O configuration, and application submenus. - Details on country-specific factory settings for units of measurement. - Explanations of abbreviated units for measurements. - Information on accessing device parameters via Modbus registers.

Uploaded by

throdis
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
65 views324 pages

Gp01062den 0519

This document provides a description of device parameters for a Proline Promass 500 Coriolis flowmeter that communicates via Modbus RS485. It includes: - An overview of the device's Expert operating menu structure with descriptions of system, sensor, I/O configuration, and application submenus. - Details on country-specific factory settings for units of measurement. - Explanations of abbreviated units for measurements. - Information on accessing device parameters via Modbus registers.

Uploaded by

throdis
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 324

GP01062D/06/EN/05.

19 Products Solutions Services


71446678
2019-09-01
Valid as of version
01.05.zz (Device firmware)

Description of Device Parameters


Proline Promass 500
Modbus RS485
Coriolis flowmeter

Menu/Variable
Proline Device
P Access status tooling:
Operation
Setup
Diagnostics
Expert
P Locking status:
P Access status tooling:
P Enter access code:
System
Sensor
Input
Output
Communication
Application
Diagnostics
Proline Promass 500 Modbus RS485 Table of contents

Table of contents

1 About this document . . . . . . . . . . . . . . . . 5 3.7.3 "Concentration" submenu . . . . . . . . . 187


3.7.4 "Custody transfer" submenu . . . . . . . 187
1.1 Document function . . . . . . . . . . . . . . . . . . . . . 5 3.7.5 "Petroleum" submenu . . . . . . . . . . . . 187
1.2 Target group . . . . . . . . . . . . . . . . . . . . . . . . . . 5 3.7.6 "Appl.spec. calc." submenu . . . . . . . . 187
1.3 Using this document . . . . . . . . . . . . . . . . . . . . 5 3.7.7 "Medium index" submenu . . . . . . . . . 193
1.3.1 Information on the document 3.8 "Diagnostics" submenu . . . . . . . . . . . . . . . . . 195
structure . . . . . . . . . . . . . . . . . . . . . . . 5 3.8.1 "Diagnostic list" submenu . . . . . . . . . 198
1.3.2 Structure of a parameter 3.8.2 "Event logbook" submenu . . . . . . . . . 202
description . . . . . . . . . . . . . . . . . . . . . 7 3.8.3 "Custody transfer logbook" submenu 204
1.4 Symbols used . . . . . . . . . . . . . . . . . . . . . . . . . . 7 3.8.4 "Device information" submenu . . . . . 205
1.4.1 Symbols for certain types of 3.8.5 "Main elec.+I/O1" submenu . . . . . . . 208
information . . . . . . . . . . . . . . . . . . . . 7 3.8.6 "Sens. electronic" submenu . . . . . . . . 209
1.4.2 Symbols in graphics . . . . . . . . . . . . . . . 8 3.8.7 "I/O module 1" submenu . . . . . . . . . . 210
1.5 Documentation . . . . . . . . . . . . . . . . . . . . . . . . 8 3.8.8 "I/O module 2" submenu . . . . . . . . . . 211
1.5.1 Standard documentation . . . . . . . . . . . 8 3.8.9 "I/O module 3" submenu . . . . . . . . . . 213
1.5.2 Supplementary device-dependent 3.8.10 "I/O module 4" submenu . . . . . . . . . . 214
documentation . . . . . . . . . . . . . . . . . . 8 3.8.11 "Display module" submenu . . . . . . . . 215
3.8.12 "Data logging" submenu . . . . . . . . . . 216
2 Overview of the Expert operating 3.8.13 "Min/max val." submenu . . . . . . . . . 225
menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 3.8.14 "Heartbeat" submenu . . . . . . . . . . . . 237
3.8.15 "Simulation" submenu . . . . . . . . . . . 237

3 Description of device parameters . . . 12 4 Country-specific factory settings . . 248


3.1 "System" submenu . . . . . . . . . . . . . . . . . . . . . 15
4.1 SI units . . . . . . . . . . . . . . . . . . . . . . . . . . . . 248
3.1.1 "Display" submenu . . . . . . . . . . . . . . . 15
4.1.1 System units . . . . . . . . . . . . . . . . . . 248
3.1.2 "Configuration backup" submenu . . . . 30
4.1.2 Full scale values . . . . . . . . . . . . . . . 248
3.1.3 "Diagn. handling" submenu . . . . . . . . . 33
4.1.3 Output current span . . . . . . . . . . . . 249
3.1.4 "Administration" submenu . . . . . . . . . 44
4.1.4 Pulse value . . . . . . . . . . . . . . . . . . . 249
3.2 "Sensor" submenu . . . . . . . . . . . . . . . . . . . . . . 50
4.1.5 On value low flow cut off . . . . . . . . . 249
3.2.1 "Measured values" submenu . . . . . . . . 50
4.2 US units . . . . . . . . . . . . . . . . . . . . . . . . . . . 250
3.2.2 "System units" submenu . . . . . . . . . . . 65
4.2.1 System units . . . . . . . . . . . . . . . . . . 250
3.2.3 "Process param." submenu . . . . . . . . . 82
4.2.2 Full scale values . . . . . . . . . . . . . . . 251
3.2.4 "Measurement mode" submenu . . . . . 90
4.2.3 Output current span . . . . . . . . . . . . 251
3.2.5 "External comp." submenu . . . . . . . . . 92
4.2.4 Pulse value . . . . . . . . . . . . . . . . . . . 251
3.2.6 "Calculated value" submenu . . . . . . . . 95
4.2.5 On value low flow cut off . . . . . . . . 252
3.2.7 "Sensor adjustment" submenu . . . . . . . 98
3.2.8 "Calibration" submenu . . . . . . . . . . . 106
3.3 "I/O configuration" submenu . . . . . . . . . . . . . 107 5 Explanation of abbreviated units . . 254
3.4 "Input" submenu . . . . . . . . . . . . . . . . . . . . . . 110 5.1 SI units . . . . . . . . . . . . . . . . . . . . . . . . . . . . 254
3.4.1 "Current input 1 to n" submenu . . . . . 110 5.2 US units . . . . . . . . . . . . . . . . . . . . . . . . . . . 254
3.4.2 "Status input 1 to n" submenu . . . . . . 113 5.3 Imperial units . . . . . . . . . . . . . . . . . . . . . . . 256
3.5 "Output" submenu . . . . . . . . . . . . . . . . . . . . . 115
3.5.1 "Current output 1 to n" submenu . . . . 116
3.5.2 "Pulse/frequency/switch output
6 Modbus RS485 Register
1 to n" submenu . . . . . . . . . . . . . . . 130 Information . . . . . . . . . . . . . . . . . . . . . . 257
3.5.3 "Relay output 1 to n" submenu . . . . . 152 6.1 Notes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 257
3.5.4 "Double pulse output" submenu . . . . . 159 6.1.1 Structure of the register
3.6 "Communication" submenu . . . . . . . . . . . . . . 164 information . . . . . . . . . . . . . . . . . . 257
3.6.1 "Modbus config." submenu . . . . . . . . 165 6.1.2 Address model . . . . . . . . . . . . . . . . 257
3.6.2 "Modbus info" submenu . . . . . . . . . . 170 6.2 Overview of the Expert operating menu . . . . 258
3.6.3 "Modbus data map" submenu . . . . . . 171 6.3 Register information . . . . . . . . . . . . . . . . . . 271
3.6.4 "Web server" submenu . . . . . . . . . . . 171 6.3.1 "System" submenu . . . . . . . . . . . . . . 272
3.6.5 "WLAN settings" submenu . . . . . . . . 174 6.3.2 "Sensor" submenu . . . . . . . . . . . . . . 278
3.7 "Application" submenu . . . . . . . . . . . . . . . . . 181 6.3.3 "I/O config." submenu . . . . . . . . . . . . 288
3.7.1 "Totalizer 1 to n" submenu . . . . . . . . 182 6.3.4 "Input" submenu . . . . . . . . . . . . . . . . 289
3.7.2 "Viscosity" submenu . . . . . . . . . . . . . 187

Endress+Hauser 3
Table of contents Proline Promass 500 Modbus RS485

6.3.5 "Output" submenu . . . . . . . . . . . . . . 290


6.3.6 "Communication" submenu . . . . . . . . 300
6.3.7 "Application" submenu . . . . . . . . . . . 302
6.3.8 "Diagnostics" submenu . . . . . . . . . . . 304

Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 313

4 Endress+Hauser
Proline Promass 500 Modbus RS485 About this document

1 About this document

1.1 Document function


The document is part of the Operating Instructions and serves as a reference for
parameters, providing a detailed explanation of each individual parameter of the Expert
operating menu.
It is used to perform tasks that require detailed knowledge of the function of the device:
• Commissioning measurements under difficult conditions
• Optimal adaptation of the measurement to difficult conditions
• Detailed configuration of the communication interface
• Error diagnostics in difficult cases

1.2 Target group


The document is aimed at specialists who work with the device over the entire life cycle
and perform specific configurations.

1.3 Using this document

1.3.1 Information on the document structure


The document lists the submenus and their parameters according to the structure from the
Expert menu (→  9), which is displayed when the "Maintenance" user role is
enabled.

Endress+Hauser 5
About this document Proline Promass 500 Modbus RS485

Operating menu for operators and maintenances


Language !
Operatation Language

Parameter 1

Operator
Parameter n

Submenu 1

Submenu n

Setup Device tag

Wizard 1 / Parameter 1
Maintenance

Task-oriented
Wizard n / Parameter n

Advanced setup Enter access code

Parameter 1

Parameter n

Submenu 1

Submenu n

Diagnostics Parameter 1

Parameter n

Submenu 1

Submenu n

Operating menu for experts


Expert Access status display

Parameter n

System Function-oriented
Sensor
Expert

Input

Output

Communication

Application

Diagnostics

A0029160-EN

1 Sample graphic for the schematic layout of the operating menu

Additional information regarding:


• The arrangement of the parameters according to the menu structure of the
Operation menu, Setup menu, Diagnostics menu with a brief description:
Operating Instructions →  8
• Operating concept of the operating menus: Operating Instructions →  8

6 Endress+Hauser
Proline Promass 500 Modbus RS485 About this document

1.3.2 Structure of a parameter description


The individual parts of a parameter description are described in the following section:

Complete parameter name Write-protected parameter =

Navigation Navigation path to the parameter via the local display (direct access code) or web browser
Navigation path to the parameter via the operating tool
The names of the menus, submenus and parameters are abbreviated to the form in which they appear on
the display and in the operating tool.

Prerequisite The parameter is only available under these specific conditions

Description Description of the parameter function

Selection List of the individual options for the parameter


• Option 1
• Option 2

User entry Input range for the parameter

User interface Display value/data for the parameter

Factory setting Default setting ex works

Additional information Additional explanations (e.g. in examples):


• On individual options
• On display values/data
• On the input range
• On the factory setting
• On the parameter function

1.4 Symbols used

1.4.1 Symbols for certain types of information

Symbol Meaning

Tip
Indicates additional information.

Reference to documentation

A Reference to page

Reference to graphic

Operation via local display

Operation via operating tool

Write-protected parameter

Endress+Hauser 7
About this document Proline Promass 500 Modbus RS485

1.4.2 Symbols in graphics

Symbol Meaning Symbol Meaning

1, 2, 3 ... Item numbers A, B, C, ... Views

A-A, B-B, C-C, ... Sections

1.5 Documentation

1.5.1 Standard documentation

Operating Instructions

Measuring device Documentation

Promass A 500 (8A5B**-…) BA01537D

Promass A 500 (8A5C**-…) BA01884D

Promass E 500 BA01539D

Promass F 500 BA01540D

Promass H 500 BA01541D

Promass I 500 BA01542D

Promass O 500 BA01543D

Promass P 500 BA01544D

Promass Q 500 BA01545D

Promass S 500 BA01546D

Promass X 500 BA01547D

1.5.2 Supplementary device-dependent documentation

Special Documentation

Contents Documentation code

Information on the Pressure Equipment Directive SD01614D

Radio approvals for WLAN interface for A309/A310 display module SD01793D

Web server SD01667D

Heartbeat Technology SD01704D

Concentration measurement SD01710D

Petroleum SD02014D

Viscosity measurement Promass I SD01724D

Custody transfer SD01691D

8 Endress+Hauser
Proline Promass 500 Modbus RS485 Overview of the Expert operating menu

2 Overview of the Expert operating menu


The following table provides an overview of the menu structure of the expert operating
menu and its parameters. The page reference indicates where the associated description of
the submenu or parameter can be found.

 Expert

Direct access (0106) →  12

Locking status (0004) →  13

Access status (0005) →  14

Ent. access code (0003) →  14

‣ System →  15

‣ Display →  15

‣ Config. backup →  30

‣ Diagn. handling →  33

‣ Administration →  44

‣ Sensor →  50

‣ Measured val. →  50

‣ System units →  65

‣ Process param. →  82

‣ Calculated value →  95

‣ Measurement mode →  90

‣ External comp. →  92

‣ Sensor adjustm. →  98

‣ Calibration →  106

‣ I/O config. →  107

I/O 1 to n terminals (3902–1 to n) →  108

I/O 1 to n info (3906–1 to n) →  108

Endress+Hauser 9
Overview of the Expert operating menu Proline Promass 500 Modbus RS485

I/O 1 to n type (3901–1 to n) →  109

Apply I/O config (3907) →  109

I/O alterat.code (2762) →  109

‣ Input →  110

‣ Current input 1 to n →  110

‣ Status input 1 to n →  113

‣ Output →  115

‣ Curr.output 1 to n →  116

‣ PFS output 1 to n →  130

‣ Relay output 1 to n →  152

‣ Double pulse out →  159

‣ Communication →  164

‣ Modbus config. →  165

‣ Modbus info →  170

‣ Modbus data map →  171

‣ Web server →  171

‣ WLAN settings →  174

‣ Application →  181

Reset all tot. (2806) →  181

‣ Totalizer 1 to n →  182

‣ Viscosity →  187

‣ Concentration →  187

‣ Custody transfer →  187

‣ Petroleum →  187

10 Endress+Hauser
Proline Promass 500 Modbus RS485 Overview of the Expert operating menu

‣ Appl.spec. calc. →  187

‣ Medium index →  193

‣ Diagnostics →  195

Actual diagnos. (0691) →  196

Prev.diagnostics (0690) →  196

Time fr. restart (0653) →  197

Operating time (0652) →  197

‣ Diagnostic list →  198

‣ Event logbook →  202

‣ Cust.transf.log. →  204

‣ Device info →  205

‣ Main elec.+I/O1 →  208

‣ Sens. electronic →  209

‣ I/O module 2 →  211

‣ I/O module 3 →  213

‣ I/O module 4 →  214

‣ Display module →  215

‣ Min/max val. →  225

‣ Data logging →  216

‣ Heartbeat →  237

‣ Simulation →  237

Endress+Hauser 11
Description of device parameters Proline Promass 500 Modbus RS485

3 Description of device parameters


In the following section, the parameters are listed according to the menu structure of the
local display. Specific parameters for the operating tools are included at the appropriate
points in the menu structure.

 Expert

Direct access (0106) →  12

Locking status (0004) →  13

Access status (0005) →  14

Ent. access code (0003) →  14

‣ System →  15

‣ Sensor →  50

‣ I/O config. →  107

‣ Input →  110

‣ Output →  115

‣ Communication →  164

‣ Application →  181

‣ Diagnostics →  195

Direct access 

Navigation  Expert → Direct access (0106)

Description Use this function to enter the access code to enable direct access to the desired parameter
via the local display. A parameter number is assigned to each parameter for this purpose.

User entry 0 to 65 535

Additional information User entry


The direct access code consists of a 5-digit number (at maximum) and the channel
number, which identifies the channel of a process variable: e.g. 00914-2. In the navigation
view, this appears on the right-hand side in the header of the selected parameter.

12 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters

00914-2 1

A0029414

1 Direct access code

Note the following when entering the direct access code:


• The leading zeros in the direct access code do not have to be entered.
Example: Enter "914" instead of "00914"
• If no channel number is entered, channel 1 is accessed automatically.
Example: Enter 00914 → Assign variable parameter
• If a different channel is accessed: Enter the direct access code with the corresponding
channel number.
Example: Enter 00914-2 → Assign variable parameter

Locking status

Navigation  Expert → Locking status (0004)

Description Displays the active write protection.

User interface • Hardware locked


• CT act.-def.par.
• CT act.-all par.
• Temp. locked

Additional information Display


If two or more types of write protection are active, the write protection with the highest
priority is shown on the local display. In the operating tool all active types of write
protection are displayed.
Detailed information on access authorization is provided in the "User roles and
associated access authorization" and "Operating concept" sections of the Operations
Instructions for the device →  8

Selection

Options Description

None The access status displayed in the Access status parameter (→  14) applies .
Only appears on local display.

Hardware locked The DIP switch for hardware locking is activated on the PCB board. This locks write
(priority 1) access to the parameters (e.g. via local display or operating tool) .

Cust.trans. act.
(priority 2)  Only available for Promass F, O, Q and X.
The DIP switch for custody transfer mode is activated on the PCB board. This locks
write access to all the parameters (e.g. via local display or operating tool).

 For detailed information on custody transfer mode, see the Special


Documentation for the device →  8

Endress+Hauser 13
Description of device parameters Proline Promass 500 Modbus RS485

Options Description

CT act.-def.par.
(priority 3)  Only available for Promass F, O, Q and X.
The DIP switch for custody transfer mode is activated on the PCB board. This locks
write access to the defined parameters (e.g. via local display or operating tool).

 For detailed information on custody transfer mode, see the Special


Documentation for the device →  8

Temp. locked Write access to the parameters is temporarily locked on account of internal
(Priority 4) processes running in the device (e.g. data upload/download, reset etc.). Once the
internal processing has been completed, the parameters can be changed once
again.

Access status

Navigation  Expert → Access status (0005)

Description Displays the access authorization to the parameters via the local display, Web browser or
operating tool.

User interface • Operator


• Maintenance

Factory setting Maintenance

Additional information Description


Access authorization can be modified via the Ent. access code parameter
(→  14).
If additional write protection is active, this restricts the current access authorization
even further.

Display
Detailed information on access authorization is provided in the "User roles and
associated access authorization" and "Operating concept" sections of the Operations
Instructions for the device →  8

Ent. access code

Navigation  Expert → Ent. access code (0003)

Description Use this function to enter the user-specific release code to remove parameter write
protection.

User entry Max. 16-digit character string comprising numbers, letters and special characters

14 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters

3.1 "System" submenu


Navigation  Expert → System

‣ System

‣ Display →  15

‣ Config. backup →  30

‣ Diagn. handling →  33

‣ Administration →  44

3.1.1 "Display" submenu

Navigation  Expert → System → Display

‣ Display

Display language (0104) →  16

Format display (0098) →  17

Value 1 display (0107) →  19

0% bargraph 1 (0123) →  21

100% bargraph 1 (0125) →  21

Decimal places 1 (0095) →  22

Value 2 display (0108) →  22

Decimal places 2 (0117) →  23

Value 3 display (0110) →  23

0% bargraph 3 (0124) →  24

100% bargraph 3 (0126) →  24

Decimal places 3 (0118) →  25

Value 4 display (0109) →  25

Decimal places 4 (0119) →  26

Endress+Hauser 15
Description of device parameters Proline Promass 500 Modbus RS485

Display interval (0096) →  26

Display damping (0094) →  27

Header (0097) →  28

Header text (0112) →  28

Separator (0101) →  29

Contrast display (0105) →  29

Backlight (0111) →  29

Display language

Navigation  Expert → System → Display → Display language (0104)

Prerequisite A local display is provided.

Description Use this function to select the configured language on the local display.

Selection • English
• Deutsch
• Français
• Español
• Italiano
• Nederlands
• Portuguesa
• Polski
• русский язык(Ru)
• Svenska
• Türkçe
• 中文 (Chinese)
• 日本語 (Japanese)
• 한국어 (Korean)
• ‫(الْعَرَبيّة‬Ara) *
• Bahasa Indonesia
• ภาษาไทย (Thai) *
• tiếng Việt (Vit)
• čeština (Czech)

Factory setting English (alternatively, the ordered language is preset in the device)

* Visibility depends on order options or device settings

16 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters

Format display

Navigation  Expert → System → Display → Format display (0098)

Prerequisite A local display is provided.

Description Use this function to select how the measured value is shown on the local display.

Selection • 1 value, max.


• Bargr. + 1 value
• 2 values
• Val. large+2val.
• 4 values

Factory setting 1 value, max.

Additional information Description


The display format (size, bar graph etc.) and number of measured values displayed
simultaneously (1 to 4) can be configured. This setting only applies to normal operation.
• The Value 1 display parameter (→  19) to Value 4 display parameter
(→  25) are used to specify which measured values are shown on the local
display and in what order.
• If more measured values are specified than the display mode selected permits, then
the values alternate on the device display. The display time until the next change is
configured via the Display interval parameter (→  26).

Custody transfer mode


Only available for Promass F, O, Q and X.

• Once the measuring device has been enabled for custody transfer mode, depending on
the custody transfer approval selected the display can switch between showing the
relevant information and the custody transfer counter.
• In addition, a padlock symbol appears in the header of the display ().
For detailed information on custody transfer mode, see the Special Documentation for
the device →  8

Endress+Hauser 17
Description of device parameters Proline Promass 500 Modbus RS485

Possible measured values shown on the local display:

"1 value, max." option

XXXXXXXXX

900.00
kg/h

A0013099

"Bargr. + 1 value" option

XXXXXXXXX

900.00 kg/h
900.00
l/h
A0013098

"2 values" option

XXXXXXXXX
900.00
kg/h
900.00
l/h
A0013100

"Val. large+2val." option

XXXXXXXXX
900.00
kg/h
900.00 l/h
1.00 kg/l
A0013102

"4 values" option

XXXXXXXXX
900.00 kg/h
900.00 l/h
1.0 kg/l
213.94 kg
A0013103

18 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters

Value 1 display 

Navigation  Expert → System → Display → Value 1 display (0107)

Prerequisite A local display is provided.

Description Use this function to select one of the measured values shown on the local display.

Selection • Mass flow


• Volume flow
• Correct.vol.flow *
• Target mass flow *
• Carrier mass fl. *
• Target vol. flow *
• Carrier vol. fl. *
• Targ.corr.vol.fl *
• Carr.corr.vol.fl *
• Density
• Ref.density *
• Ref.dens.altern. *
• GSV flow *
• GSVa *
• NSV flow *
• NSVa *
• S&W volume flow *
• Water cut *
• Oil density *
• Water density *
• Oil mass flow *
• Water mass flow *
• Oil volume flow *
• Water vol. flow *
• Oil corr.vol.fl. *
• Water corr.v.fl. *
• Density average *
• Temp. average *
• Concentration *
• Dynam. viscosity *
• Kinematic visc. *
• TempCompDynVisc *
• TempCompKinVisc *
• Temperature
• Carr. pipe temp. *
• Electronic temp.
• Osc. freq. 0
• Osc. freq. 1 *
• Osc. ampl. 0 *
• Osc. ampl. 1 *
• Freq. fluct. 0 *
• Freq. fluct. 1 *
• Osc. damping 0 *
• Osc. damping 1 *
• Osc.damp.fluct 0 *
• Osc.damp.fluct 1 *
• Signal asymmetry *

* Visibility depends on order options or device settings

Endress+Hauser 19
Description of device parameters Proline Promass 500 Modbus RS485

• Exc. current 0 *
• Exc. current 1 *
• HBSI *
• Totalizer 1
• Totalizer 2
• Totalizer 3
• Curr.output 1 *
• Curr.output 2 *
• Curr.output 3 *
• Curr.output 4 *
• Pressure
• Spec. output 1 *
• Index inh.medium
• Spec. output 0 *
• Index sus.bubble *

Factory setting Mass flow

Additional information Description


If several measured values are displayed at once, the measured value selected here will be
the first value to be displayed. The value is only displayed during normal operation.
The Format display parameter (→  17) is used to specify how many measured
values are displayed simultaneously and how.

Custody transfer mode


Only available for Promass F, O, Q and X.

Once the measuring device has been enabled for custody transfer mode, depending on the
custody transfer approval selected the display can switch to show the relevant information.
For detailed information on custody transfer mode, see the Special Documentation for
the device →  8

Dependency
The unit of the displayed measured value is taken from the System units submenu
(→  65).

Selection
• Oscil. frequency option
Displays the current oscillation frequency of the measuring tubes. This frequency
depends on the density of the medium.
• Oscil. amplitude option
Displays the relative oscillation amplitude of the measuring tubes in relation to the
preset value. This value is 100 % under optimum conditions.
• Oscil. damping option
Displays the current oscillation damping. Oscillation damping is an indicator of the
sensor's current need for excitation power.
• Signal asymmetry option
Displays the relative difference between the oscillation amplitude at the inlet and outlet
of the sensor. The measured value is the result of production tolerances of the sensor
coils and should remain constant over the life time of a sensor.

* Visibility depends on order options or device settings

20 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters

0% bargraph 1 

Navigation  Expert → System → Display → 0% bargraph 1 (0123)

Prerequisite A local display is provided.

Description Use this function to enter the 0% bar graph value to be shown on the display for the
measured value 1.

User entry Signed floating-point number

Factory setting Country-specific:


• 0 kg/h
• 0 lb/min

Additional information Description


The Format display parameter (→  17) is used to specify that the measured value
is to be displayed as a bar graph.

User entry
The unit of the displayed measured value is taken from the System units submenu
(→  65).

100% bargraph 1 

Navigation  Expert → System → Display → 100% bargraph 1 (0125)

Prerequisite A local display is provided.

Description Use this function to enter the 100% bar graph value to be shown on the display for the
measured value 1.

User entry Signed floating-point number

Factory setting Depends on country and nominal diameter →  248

Additional information Description


The Format display parameter (→  17) is used to specify that the measured value
is to be displayed as a bar graph.

User entry
The unit of the displayed measured value is taken from the System units submenu
(→  65).

Endress+Hauser 21
Description of device parameters Proline Promass 500 Modbus RS485

Decimal places 1 

Navigation  Expert → System → Display → Decimal places 1 (0095)

Prerequisite A measured value is specified in the Value 1 display parameter (→  19).

Description Use this function to select the number of decimal places for measured value 1.

Selection •x
• x.x
• x.xx
• x.xxx
• x.xxxx

Factory setting x.xx

Additional information Description


This setting does not affect the measuring or computational accuracy of the device.
The arrow displayed between the measured value and the unit indicates that the
device computes with more digits than are shown on the local display.

Value 2 display 

Navigation  Expert → System → Display → Value 2 display (0108)

Prerequisite A local display is provided.

Description Use this function to select one of the measured values shown on the local display.

Selection For the picklist, see the Value 2 display parameter (→  22)

Factory setting None

Additional information Description


If several measured values are displayed at once, the measured value selected here will be
the second value to be displayed. The value is only displayed during normal operation.
The Format display parameter (→  17) is used to specify how many measured
values are displayed simultaneously and how.

Dependency
The unit of the displayed measured value is taken from the System units submenu
(→  65).

22 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters

Decimal places 2 

Navigation  Expert → System → Display → Decimal places 2 (0117)

Prerequisite A measured value is specified in the Value 2 display parameter (→  22).

Description Use this function to select the number of decimal places for measured value 2.

Selection •x
• x.x
• x.xx
• x.xxx
• x.xxxx

Factory setting x.xx

Additional information Description


This setting does not affect the measuring or computational accuracy of the device.
The arrow displayed between the measured value and the unit indicates that the
device computes with more digits than are shown on the local display.

Value 3 display 

Navigation  Expert → System → Display → Value 3 display (0110)

Prerequisite A local display is provided.

Description Use this function to select one of the measured values shown on the local display.

Selection For the picklist, see the Value 2 display parameter (→  22)

Factory setting None

Additional information Description


If several measured values are displayed at once, the measured value selected here will be
the third value to be displayed. The value is only displayed during normal operation.
The Format display parameter (→  17) is used to specify how many measured
values are displayed simultaneously and how.

Selection
The unit of the displayed measured value is taken from the System units submenu
(→  65).

Endress+Hauser 23
Description of device parameters Proline Promass 500 Modbus RS485

0% bargraph 3 

Navigation  Expert → System → Display → 0% bargraph 3 (0124)

Prerequisite A selection was made in the Value 3 display parameter (→  23).

Description Use this function to enter the 0% bar graph value to be shown on the display for the
measured value 3.

User entry Signed floating-point number

Factory setting Country-specific:


• 0 kg/h
• 0 lb/min

Additional information Description


The Format display parameter (→  17) is used to specify that the measured value
is to be displayed as a bar graph.

User entry
The unit of the displayed measured value is taken from the System units submenu
(→  65).

100% bargraph 3 

Navigation  Expert → System → Display → 100% bargraph 3 (0126)

Prerequisite A selection was made in the Value 3 display parameter (→  23).

Description Use this function to enter the 100% bar graph value to be shown on the display for the
measured value 3.

User entry Signed floating-point number

Factory setting 0

Additional information Description


The Format display parameter (→  17) is used to specify that the measured value
is to be displayed as a bar graph.

User entry
The unit of the displayed measured value is taken from the System units submenu
(→  65).

24 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters

Decimal places 3 

Navigation  Expert → System → Display → Decimal places 3 (0118)

Prerequisite A measured value is specified in the Value 3 display parameter (→  23).

Description Use this function to select the number of decimal places for measured value 3.

Selection •x
• x.x
• x.xx
• x.xxx
• x.xxxx

Factory setting x.xx

Additional information Description


This setting does not affect the measuring or computational accuracy of the device.
The arrow displayed between the measured value and the unit indicates that the
device computes with more digits than are shown on the local display.

Value 4 display 

Navigation  Expert → System → Display → Value 4 display (0109)

Prerequisite A local display is provided.

Description Use this function to select one of the measured values shown on the local display.

Selection For the picklist, see the Value 2 display parameter (→  22)

Factory setting None

Endress+Hauser 25
Description of device parameters Proline Promass 500 Modbus RS485

Additional information Description


If several measured values are displayed at once, the measured value selected here will be
the fourth value to be displayed. The value is only displayed during normal operation.
The Format display parameter (→  17) is used to specify how many measured
values are displayed simultaneously and how.

Selection
The unit of the displayed measured value is taken from the System units submenu
(→  65).

Custody transfer mode


Only available for Promass F, O, Q and X.

Once the measuring device has been enabled for custody transfer mode, depending on the
custody transfer approval selected the display can switch to showing the custody transfer
counter.
For detailed information on custody transfer mode, see the Special Documentation for
the device →  8

Decimal places 4 

Navigation  Expert → System → Display → Decimal places 4 (0119)

Prerequisite A measured value is specified in the Value 4 display parameter (→  25).

Description Use this function to select the number of decimal places for measured value 4.

Selection •x
• x.x
• x.xx
• x.xxx
• x.xxxx

Factory setting x.xx

Additional information Description


This setting does not affect the measuring or computational accuracy of the device.
The arrow displayed between the measured value and the unit indicates that the
device computes with more digits than are shown on the local display.

Display interval

Navigation  Expert → System → Display → Display interval (0096)

Prerequisite A local display is provided.

Description Use this function to enter the length of time the measured values are displayed if the
values alternate on the display.

26 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters

User entry 1 to 10 s

Factory setting 5s

Additional information Description


This type of alternating display only occurs automatically if the number of measured
values defined exceeds the number of values the selected display format can display
simultaneously.
• The Value 1 display parameter (→  19) to Value 4 display parameter (→  25)
are used to specify which measured values are shown on the local display.
• The display format of the displayed measured values is specified using the Format
display parameter (→  17).

Custody transfer mode


Only available for Promass F, O, Q and X.

Once the measuring device has been enabled for custody transfer mode, depending on the
custody transfer approval selected the display can switch between showing the relevant
information and the custody transfer counter.
For detailed information on custody transfer mode, see the Special Documentation for
the device →  8

Display damping 

Navigation  Expert → System → Display → Display damping (0094)

Prerequisite A local display is provided.

Description Use this function to enter a time constant for the reaction time of the local display to
fluctuations in the measured value caused by process conditions.

User entry 0.0 to 999.9 s

Factory setting 0.0 s

Additional information User entry


Use this function to enter a time constant (PT1 element 1)) for display damping:
• If a low time constant is entered, the display reacts particularly quickly to fluctuating
measured variables.
• On the other hand, the display reacts more slowly if a high time constant is entered.
Damping is switched off if 0 is entered (factory setting).

1) proportional transmission behavior with first order delay

Endress+Hauser 27
Description of device parameters Proline Promass 500 Modbus RS485

Header 

Navigation  Expert → System → Display → Header (0097)

Prerequisite A local display is provided.

Description Use this function to select the contents of the header of the local display.

Selection • Device tag


• Free text

Factory setting Device tag

Additional information Description


The header text only appears during normal operation.

1 XXXXXXXXX

A0029422

1 Position of the header text on the display

Selection
• Device tag
Is defined in the Device tag parameter (→  205).
• Free text
Is defined in the Header text parameter (→  28).

Header text 

Navigation  Expert → System → Display → Header text (0112)

Prerequisite In the Header parameter (→  28), the Free text option is selected.

Description Use this function to enter a customer-specific text for the header of the local display.

User entry Max. 12 characters such as letters, numbers or special characters (e.g. @, %, /)

Factory setting ------------

Additional information Description


The header text only appears during normal operation.

28 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters

1 XXXXXXXXX

A0029422

1 Position of the header text on the display

User entry
The number of characters displayed depends on the characters used.

Separator 

Navigation  Expert → System → Display → Separator (0101)

Prerequisite A local display is provided.

Description Use this function to select the decimal separator.

Selection • . (point)
• , (comma)

Factory setting . (point)

Contrast display

Navigation  Expert → System → Display → Contrast display (0105)

Prerequisite A local display is provided.

Description Use this function to enter a value to adapt the display contrast to the ambient conditions
(e.g. the lighting or viewing angle).

User entry 20 to 80 %

Factory setting Depends on the display

Backlight

Navigation  Expert → System → Display → Backlight (0111)

Prerequisite One of the following conditions is met:


• Order code for "Display; operation", option F "4-line, illum.; touch control"
• Order code for "Display; operation", option G "4-line, illum.; touch control +WLAN"

Endress+Hauser 29
Description of device parameters Proline Promass 500 Modbus RS485

Description Use this function to switch the backlight of the local display on and off.

Selection • Disable
• Enable

Factory setting Enable

3.1.2 "Configuration backup" submenu

Navigation  Expert → System → Config. backup

‣ Config. backup

Operating time (0652) →  30

Last backup (2757) →  30

Config. managem. (2758) →  31

Backup state (2759) →  31

Compar. result (2760) →  32

Operating time

Navigation  Expert → System → Config. backup → Operating time (0652)

Description Use this function to display the length of time the device has been in operation.

User interface Days (d), hours (h), minutes (m) and seconds (s)

Additional information User interface


The maximum number of days is 9999, which is equivalent to 27 years.

Last backup

Navigation  Expert → System → Config. backup → Last backup (2757)

Description Displays the time since a backup copy of the data was last saved to the device memory.

User interface Days (d), hours (h), minutes (m) and seconds (s)

30 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters

Config. managem. 

Navigation  Expert → System → Config. backup → Config. managem. (2758)

Description Use this function to select an action to save the data to the device memory.

Selection • Cancel
• Execute backup
• Restore *
• Compare *
• Clear backup

Factory setting Cancel

Additional information Selection

Options Description

Cancel No action is executed and the user exits the parameter.

Execute backup A backup copy of the current device configuration is saved from the HistoROM
backup to the memory of the device. The backup copy includes the transmitter data
of the device.
The following message appears on local display: Backup active, please wait!

Restore The last backup copy of the device configuration is restored from the device
memory to the device's HistoROM backup. The backup copy includes the
transmitter data of the device.
The following message appears on local display: Restore active! Do not interrupt
power supply!

Compare The device configuration saved in the device memory is compared with the current
device configuration of the HistoROM backup.
The following message appears on local display: Comparing files
The result can be viewed in Compar. result parameter.

Clear backup The backup copy of the device configuration is deleted from the memory of the
device.
The following message appears on local display: Deleting file

HistoROM
A HistoROM is a "non-volatile" device memory in the form of an EEPROM.

Backup state

Navigation  Expert → System → Config. backup → Backup state (2759)

Description Displays the status of the data backup process.

User interface • None


• Backup in progr.
• Restore in progr
• Delete in progr.

* Visibility depends on order options or device settings

Endress+Hauser 31
Description of device parameters Proline Promass 500 Modbus RS485

• Comp. in progr.
• Restoring failed
• Backup failed

Factory setting None

Compar. result

Navigation  Expert → System → Config. backup → Compar. result (2760)

Description Displays the last result of the comparison of the data records in the device memory and in
the HistoROM.

User interface • Set. identical


• Set. not ident.
• No backup
• Backup corrupt
• Check not done
• Dataset incomp.

Factory setting Check not done

Additional information Description


The comparison is started via the Compare option in the Config. managem.
parameter (→  31).

Selection

Options Description

Set. identical The current device configuration of the HistoROM is identical to the backup copy in
the device memory.
If the transmitter configuration of another device has been transmitted to the
device via HistoROM in the Config. managem. parameter, the current device
configuration of the HistoROM is only partially identical to the backup copy in the
device memory: The settings for the transmitter are not identical.

Set. not ident. The current device configuration of the HistoROM is not identical to the backup
copy in the device memory.

No backup There is no backup copy of the device configuration of the HistoROM in the device
memory.

Backup corrupt The current device configuration of the HistoROM is corrupt or not compatible with
the backup copy in the device memory.

Check not done The device configuration of the HistoROM has not yet been compared to the
backup copy in the device memory.

Dataset incomp. The backup copy in the device memory is not compatible with the device.

HistoROM
A HistoROM is a "non-volatile" device memory in the form of an EEPROM.

32 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters

3.1.3 "Diagn. handling" submenu

Navigation  Expert → System → Diagn. handling

‣ Diagn. handling

Alarm delay (0651) →  33

‣ Diagn. behavior →  33

Alarm delay 

Navigation  Expert → System → Diagn. handling → Alarm delay (0651)

Description Use this function to enter the time interval until the device generates a diagnostic
message.
The diagnostic message is reset without a time delay.

User entry 0 to 60 s

Factory setting 0s

Additional information Result


This setting affects the following diagnostic messages:
• 046 Sensor limit
• 140 Sensor sig.asym.
• 144 MeasErrorTooHigh
• 830 Sensor temp.
• 831 Sensor temp.
• 832 Electronic temp.
• 833 Electronic temp.
• 834 Process temp.
• 835 Process temp.
• 843 Process limit
• 862 Partly filled
• 912 Medium inhomog.
• 913 Medium unsuitab.
• 944 MonitoringFailed

"Diagn. behavior" submenu


Each item of diagnostic information is assigned a specific diagnostic behavior at the
factory. The user can change this assignment for specific diagnostic information in the
Diagn. behavior submenu (→  33).

Endress+Hauser 33
Description of device parameters Proline Promass 500 Modbus RS485

The following options are available in the Assign behavior of diagnostic no. xxx
parameters:

Options Description

Alarm The device stops measurement. The measured value output via Modbus RS485 and the
totalizers assume the defined alarm condition. A diagnostic message is generated.
The background lighting changes to red.

Warning The device continues to measure. The measured value output via Modbus RS485 and the
totalizers are not affected. A diagnostic message is generated.

Logbook only The device continues to measure. The diagnostic message is displayed only in the Event
logbook submenu (→  202) (Event list submenu (→  203)) and is not displayed in
alternation with the operational display.

Off The diagnostic event is ignored, and no diagnostic message is generated or entered.

For a list of all the diagnostic events, see the Operating Instructions for the device
→ 8

Navigation  Expert → System → Diagn. handling → Diagn. behavior

‣ Diagn. behavior

Diagnostic no. 046 (0709) →  35

Diagnostic no. 140 (0708) →  35

Diagnostic no. 144 (0731) →  36

Diagnostic no. 374 (0710) →  36

Diagnostic no. 302 (0739) →  37

Diagnostic no. 441 (0657) →  37

Diagnostic no. 442 (0658) →  37

Diagnostic no. 443 (0659) →  38

Diagnostic no. 444 (0740) →  38

Diagnostic no. 543 (0643) →  38

Diagnostic no. 830 (0800) →  39

Diagnostic no. 831 (0641) →  39

Diagnostic no. 832 (0681) →  39

Diagnostic no. 833 (0682) →  40

Diagnostic no. 834 (0700) →  40

34 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters

Diagnostic no. 835 (0702) →  40

Diagnostic no. 842 (0638) →  41

Diagnostic no. 862 (0679) →  41

Diagnostic no. 912 (0703) →  42

Diagnostic no. 913 (0712) →  42

Diagnostic no. 941 (0635) →  42

Diagnostic no. 942 (0636) →  43

Diagnostic no. 943 (0637) →  43

Diagnostic no. 944 (0732) →  44

Diagnostic no. 948 (0744) →  44

Diagnostic no. 046 (Sensor limit) 

Navigation  Expert → System → Diagn. handling → Diagn. behavior → Diagnostic no. 046 (0709)

Description Use this function to change the diagnostic behavior of the diagnostic message 046 Sensor
limit.

Selection • Off
• Alarm
• Warning
• Logbook only

Factory setting Alarm

Additional information Detailed description of the options available for selection: →  34

Diagnostic no. 140 (Sensor sig.asym.) 

Navigation  Expert → System → Diagn. handling → Diagn. behavior → Diagnostic no. 140 (0708)

Description Use this function to change the diagnostic behavior of the diagnostic message 140 Sensor
sig.asym..

Endress+Hauser 35
Description of device parameters Proline Promass 500 Modbus RS485

Selection • Off
• Alarm
• Warning
• Logbook only

Factory setting Alarm

Additional information Detailed description of the options available for selection: →  34

Diagnostic no. 144 (MeasErrorTooHigh) 

Navigation  Expert → System → Diagn. handling → Diagn. behavior → Diagnostic no. 144 (0731)

Description Use this function to change the diagnostic behavior of the diagnostic message
144 MeasErrorTooHigh.

Selection • Off
• Alarm
• Warning
• Logbook only

Factory setting Alarm

Additional information Detailed description of the options available for selection: →  34

Diagnostic no. 374 (Sensor electron.) 

Navigation  Expert → System → Diagn. handling → Diagn. behavior → Diagnostic no. 374 (0710)

Description Use this function to change the diagnostic behavior of the diagnostic message 374 Sensor
electron..

Selection • Off
• Alarm
• Warning
• Logbook only

Factory setting Warning

Additional information Detailed description of the options available for selection: →  34

36 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters

Diagnostic no. 302 (Verific. active) 

Navigation  Expert → System → Diagn. handling → Diagn. behavior → Diagnostic no. 302 (0739)

Description Option for changing the diagnostic behavior of the diagnostic message 302 Verific. active.

Selection • Alarm
• Warning

Factory setting Warning

Additional information Detailed description of the options available for selection: →  34

Diagnostic no. 441 (Curr.output 1 to n) 

Navigation  Expert → System → Diagn. handling → Diagn. behavior → Diagnostic no. 441 (0657)

Description Option for changing the diagnostic behavior of the diagnostic message 441 Curr.output
1 to n.

Selection • Off
• Alarm
• Warning
• Logbook only

Factory setting Warning

Additional information Detailed description of the options available for selection: →  34

Diagnostic no. 442 (Freq. output 1 to n) 

Navigation  Expert → System → Diagn. handling → Diagn. behavior → Diagnostic no. 442 (0658)

Prerequisite The measuring device has a pulse/frequency/switch output.

Description Option for changing the diagnostic behavior of the diagnostic message 442 Freq. output
1 to n.

Selection • Off
• Alarm
• Warning
• Logbook only

Factory setting Warning

Additional information Detailed description of the options available for selection: →  34

Endress+Hauser 37
Description of device parameters Proline Promass 500 Modbus RS485

Diagnostic no. 443 (Pulse output 1 to n) 

Navigation  Expert → System → Diagn. handling → Diagn. behavior → Diagnostic no. 443 (0659)

Prerequisite The measuring device has a pulse/frequency/switch output.

Description Option for changing the diagnostic behavior of the diagnostic message 443 Pulse output
1 to n.

Selection • Off
• Alarm
• Warning
• Logbook only

Factory setting Warning

Additional information Detailed description of the options available for selection: →  34

Diagnostic no. 444 (Current input 1 to n) 

Navigation  Expert → System → Diagn. handling → Diagn. behavior → Diagnostic no. 444 (0740)

Prerequisite The device has one current input.

Description Option for changing the diagnostic behavior of the diagnostic message 444 Current input
1 to n.

Selection • Off
• Alarm
• Warning
• Logbook only

Factory setting Warning

Additional information Detailed description of the options available for selection: →  34

Diagnostic no. 543 (Double pulse out) 

Navigation  Expert → System → Diagn. handling → Diagn. behavior → Diagnostic no. 543 (0643)

Description Option for changing the diagnostic behavior of the diagnostic message 543 Double pulse
out.

Selection • Off
• Alarm
• Warning
• Logbook only

38 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters

Factory setting Warning

Additional information Detailed description of the options available for selection: →  34

Diagnostic no. 830 (Sensor temp.) 

Navigation  Expert → System → Diagn. handling → Diagn. behavior → Diagnostic no. 830 (0800)

Description Use this function to change the diagnostic behavior of the diagnostic message 830 Sensor
temp..

Selection • Off
• Alarm
• Warning
• Logbook only

Factory setting Warning

Additional information Detailed description of the options available for selection: →  34

Diagnostic no. 831 (Sensor temp.) 

Navigation  Expert → System → Diagn. handling → Diagn. behavior → Diagnostic no. 831 (0641)

Description Use this function to change the diagnostic behavior of the diagnostic message 831 Sensor
temp..

Selection • Off
• Alarm
• Warning
• Logbook only

Factory setting Warning

Additional information Detailed description of the options available for selection: →  34

Diagnostic no. 832 (Electronic temp.) 

Navigation  Expert → System → Diagn. handling → Diagn. behavior → Diagnostic no. 832 (0681)

Description Option for changing the diagnostic behavior of the diagnostic message 832 Electronic
temp..

Endress+Hauser 39
Description of device parameters Proline Promass 500 Modbus RS485

Selection • Off
• Alarm
• Warning
• Logbook only

Factory setting Logbook only

Additional information Detailed description of the options available for selection: →  34

Diagnostic no. 833 (Electronic temp.) 

Navigation  Expert → System → Diagn. handling → Diagn. behavior → Diagnostic no. 833 (0682)

Description Option for changing the diagnostic behavior of the diagnostic message 833 Electronic
temp..

Selection • Off
• Alarm
• Warning
• Logbook only

Factory setting Logbook only

Additional information Detailed description of the options available for selection: →  34

Diagnostic no. 834 (Process temp.) 

Navigation  Expert → System → Diagn. handling → Diagn. behavior → Diagnostic no. 834 (0700)

Description Option for changing the diagnostic behavior of the diagnostic message 834 Process temp..

Selection • Off
• Alarm
• Warning
• Logbook only

Factory setting Warning

Additional information Detailed description of the options available for selection: →  34

Diagnostic no. 835 (Process temp.) 

Navigation  Expert → System → Diagn. handling → Diagn. behavior → Diagnostic no. 835 (0702)

Description Option for changing the diagnostic behavior of the diagnostic message 835 Process temp..

40 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters

Selection • Off
• Alarm
• Warning
• Logbook only

Factory setting Warning

Additional information Detailed description of the options available for selection: →  34

Diagnostic no. 842 (Process limit) 

Navigation  Expert → System → Diagn. handling → Diagn. behavior → Diagnostic no. 842 (0638)

Description Option for changing the diagnostic behavior of the diagnostic message S842 Process
limit.

Selection • Off
• Alarm
• Warning
• Logbook only

Factory setting Off

Additional information Detailed description of the options available for selection: →  34

Diagnostic no. 862 (Empty pipe) 

Navigation  Expert → System → Diagn. handling → Diagn. behavior → Diagnostic no. 862 (0679)

Description Use this function to change the diagnostic behavior of the diagnostic message 862 Empty
pipe.

Selection • Off
• Alarm
• Warning
• Logbook only

Factory setting Warning

Additional information Detailed description of the options available for selection: →  34

Endress+Hauser 41
Description of device parameters Proline Promass 500 Modbus RS485

Diagnostic no. 912 (Medium inhomog.) 

Navigation  Expert → System → Diagn. handling → Diagn. behavior → Diagnostic no. 912 (0703)

Description Use this function to change the diagnostic behavior of the diagnostic message
912 Medium inhomog..

Selection • Off
• Alarm
• Warning
• Logbook only

Factory setting Warning

Additional information Detailed description of the options available for selection: →  34

Diagnostic no. 913 (Medium unsuitab.) 

Navigation  Expert → System → Diagn. handling → Diagn. behavior → Diagnostic no. 913 (0712)

Description Use this function to change the diagnostic behavior of the diagnostic message
913 Medium unsuitab..

Selection • Off
• Alarm
• Warning
• Logbook only

Factory setting Warning

Additional information Detailed description of the options available for selection: →  34

Diagnostic no. 941 (API temperature) 

Navigation  Expert → System → Diagn. handling → Diagn. behavior → Diagnostic no. 941 (0635)

Prerequisite For the following order code:


"Application package", option EJ "Petroleum"

Description Use this function to change the diagnostic behavior of the diagnostic message API
temperature.

Selection • Off
• Alarm
• Warning
• Logbook only

42 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters

Factory setting Warning

Additional information Detailed description of the options available for selection: →  34

Diagnostic no. 942 (API density) 

Navigation  Expert → System → Diagn. handling → Diagn. behavior → Diagnostic no. 942 (0636)

Prerequisite For the following order code:


"Application package", option EJ "Petroleum"

Description Use this function to change the diagnostic behavior of the diagnostic message API
density.

Selection • Off
• Alarm
• Warning
• Logbook only

Factory setting Warning

Additional information Detailed description of the options available for selection: →  34

Diagnostic no. 943 (API pressure) 

Navigation  Expert → System → Diagn. handling → Diagn. behavior → Diagnostic no. 943 (0637)

Prerequisite For the following order code:


"Application package", option EJ "Petroleum"

Description Use this function to change the diagnostic behavior of the diagnostic message API
pressure.

Selection • Off
• Alarm
• Warning
• Logbook only

Factory setting Warning

Additional information Detailed description of the options available for selection: →  34

Endress+Hauser 43
Description of device parameters Proline Promass 500 Modbus RS485

Diagnostic no. 944 (MonitoringFailed) 

Navigation  Expert → System → Diagn. handling → Diagn. behavior → Diagnostic no. 944 (0732)

Description Use this function to change the diagnostic behavior of the diagnostic message
944 MonitoringFailed.

Selection • Off
• Alarm
• Warning
• Logbook only

Factory setting Warning

Additional information Detailed description of the options available for selection: →  34

Diagnostic no. 948 (Oscill. damping) 

Navigation  Expert → System → Diagn. handling → Diagn. behavior → Diagnostic no. 948 (0744)

Description Use this function to change the diagnostic behavior of the diagnostic message 948 Oscill.
damping.

Selection • Off
• Alarm
• Warning
• Logbook only

Factory setting Warning

Additional information Detailed description of the options available for selection: →  34

3.1.4 "Administration" submenu

Navigation  Expert → System → Administration

‣ Administration

‣ Def. access code →  45

‣ Reset acc. code →  46

Device reset (0000) →  47

44 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters

Activate SW opt. (0029) →  48

SW option overv. (0015) →  49

"Def. access code" wizard


The Def. access code wizard (→  45) is only available when operating via the local
display or Web browser.
If operating via the operating tool, the Def. access code parameter can be found
directly in the Administration submenu. There is no Confirm code parameter if the
device is operated via the operating tool.

Navigation  Expert → System → Administration → Def. access code

‣ Def. access code

Def. access code →  45

Confirm code →  46

Def. access code 

Navigation  Expert → System → Administration → Def. access code → Def. access code

Description Use this function to enter a user-specific release code to restrict write-access to the
parameters. This protects the device configuration against any inadvertent modifications
via the local display, Web browser, FieldCare or DeviceCare (via CDI-RJ45 service
interface).

User entry Max. 16-digit character string comprising numbers, letters and special characters

Additional information Description


The write protection affects all parameters in the document marked with the  symbol.
On the local display, the  symbol in front of a parameter indicates that the parameter is
write-protected.
The parameters that cannot be write-accessed are grayed out in the Web browser.
Once the access code has been defined, write-protected parameters can only be
modified if the access code is entered in the Ent. access code parameter (→  14).
If you lose the access code, please contact your Endress+Hauser sales organization.

User entry
A message is displayed if the access code is not in the input range.

Factory setting
If the factory setting is not changed or 0 is defined as the access code, the parameters are
not write-protected and the device configuration data can be modified. The user is logged
on in the "Maintenance" role.

Endress+Hauser 45
Description of device parameters Proline Promass 500 Modbus RS485

Confirm code 

Navigation  Expert → System → Administration → Def. access code → Confirm code

Description Enter the defined release code a second time to confirm the release code.

User entry Max. 16-digit character string comprising numbers, letters and special characters

"Reset access code" submenu

Navigation  Expert → System → Administration → Reset acc. code

‣ Reset acc. code

Operating time (0652) →  46

Reset acc. code (0024) →  46

Operating time

Navigation  Expert → System → Administration → Reset acc. code → Operating time (0652)

Description Use this function to display the length of time the device has been in operation.

User interface Days (d), hours (h), minutes (m) and seconds (s)

Additional information User interface


The maximum number of days is 9999, which is equivalent to 27 years.

Reset acc. code

Navigation  Expert → System → Administration → Reset acc. code → Reset acc. code (0024)

Description Use this function to enter a reset code to reset the user-specific release code to the factory
setting.

User entry Character string comprising numbers, letters and special characters

Factory setting 0x00

46 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters

Additional information Description


For a reset code, contact your Endress+Hauser service organization.

User entry
The reset code can only be entered via:
• Web browser
• DeviceCare, FieldCare (via interface CDI RJ45)
• Fieldbus

Additional parameters in the "Administration" submenu

Device reset 

Navigation  Expert → System → Administration → Device reset (0000)

Description Use this function to choose whether to reset the device configuration - either entirely or in
part - to a defined state.

Selection • Cancel
• To delivery set.
• Restart device
• Rest.S-DATBackup *

Factory setting Cancel

Additional information Selection

Options Description

Cancel No action is executed and the user exits the parameter.

To delivery set. Every parameter for which a customer-specific default setting was ordered is reset
to this customer-specific value. All other parameters are reset to the factory
setting.

Restart device The restart resets every parameter whose data are in the volatile memory (RAM)
to the factory setting (e.g. measured value data). The device configuration remains
unchanged.

Rest.S-DATBackup Restore the data that are saved on the S-DAT. The data record is restored from the
electronics memory to the S-DAT.

 This option is displayed only in an alarm condition.

* Visibility depends on order options or device settings

Endress+Hauser 47
Description of device parameters Proline Promass 500 Modbus RS485

Activate SW opt. 

Navigation  Expert → System → Administration → Activate SW opt. (0029)

Description Use this function to enter an activation code to enable an additional, ordered software
option.

User entry Max. 10-digit string consisting of numbers.

Factory setting Depends on the software option ordered

Additional information Description


If a measuring device was ordered with an additional software option, the activation code
is programmed in the device at the factory.

User entry
To activate a software option subsequently, please contact your Endress+Hauser sales
organization.
NOTE!
The activation code is linked to the serial number of the measuring device and varies
according to the device and software option.
If an incorrect or invalid code is entered, this results in the loss of software options that
have already been activated.
▸ Before you enter a new activation code, make a note of the current activation code .
▸ Enter the new activation code provided by Endress+Hauser when the new software
option was ordered.
▸ Once the activation code has been entered, check if the new software option is displayed
in the SW option overv. parameter (→  49).
 The new software option is active if it is displayed.
 If the new software option is not displayed or all software options have been deleted,
the code entered was either incorrect or invalid.
▸ If the code entered is incorrect or invalid, enter the old activation code .
▸ Have your Endress+Hauser sales organization check the new activation code
remembering to specify the serial number or ask for the code again.

Example for a software option


Order code for "Application package", option EA "Extended HistoROM"
The software options currently enabled are displayed in the SW option overv.
parameter (→  49).

Web browser
Once a software option has been activated, the page must be loaded again in the Web
browser.

48 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters

SW option overv.

Navigation  Expert → System → Administration → SW option overv. (0015)

Description Displays all the software options that are enabled in the device.

User interface • Extend. HistoROM *


• HBT Monitoring *
• Custody transfer
• Appl.spec. calc.
• HBT Verification *
• Concentration *
• Petroleum *
• Viscosity *
• Custody transfer

Additional information Description


Displays all the options that are available if ordered by the customer.

"Extend. HistoROM" option


Order code for "Application package", option EA "Extended HistoROM"

"HBT Verification" option and "HBT Monitoring" option


Order code for "Application package", option EB "Heartbeat Verification + Monitoring"

"Concentration" option
Order code for "Application package", option ED "Concentration" and option EE "Special
density"

"Viscosity" option
Only available for Promass I.

Order code for "Application package", option EG "Viscosity"

"Custody transfer" option


The measuring device has an approval for custody transfer measurement.
Detailed information on the national and international approvals for custody transfer
that are currently available can be supplied by your Endress+Hauser sales
organization.

"Petroleum" option
Only available for Promass E, F, O, Q and X.

Order code for "Application package", option EJ "Petroleum"

* Visibility depends on order options or device settings

Endress+Hauser 49
Description of device parameters Proline Promass 500 Modbus RS485

3.2 "Sensor" submenu


Navigation  Expert → Sensor

‣ Sensor

‣ Measured val. →  50

‣ System units →  65

‣ Process param. →  82

‣ Calculated value →  95

‣ Measurement mode →  90

‣ External comp. →  92

‣ Sensor adjustm. →  98

‣ Calibration →  106

3.2.1 "Measured values" submenu

Navigation  Expert → Sensor → Measured val.

‣ Measured val.

‣ Process variab. →  50

‣ Totalizer →  58

‣ Input values →  59

‣ Output values →  61

"Process variab." submenu

Navigation  Expert → Sensor → Measured val. → Process variab.

‣ Process variab.

Mass flow (1838) →  51

Volume flow (1847) →  52

Correct.vol.flow (1851) →  52

50 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters

Density (1850) →  52

Ref.density (1852) →  52

Temperature (1853) →  53

Pressure value (6129) →  53

Dynam. viscosity (1854) →  53

Kinematic visc. (1857) →  54

TempCompDynVisc (1872) →  54

TempCompKinVisc (1863) →  54

Concentration (1887) →  55

Target mass flow (1864) →  55

Carrier mass fl. (1865) →  56

Targ.corr.vol.fl (1893) →  56

Carr.corr.vol.fl (1894) →  56

Target vol. flow (1895) →  57

Carrier vol. fl. (1896) →  57

Mass flow

Navigation  Expert → Sensor → Measured val. → Process variab. → Mass flow (1838)

Description Displays the mass flow that is currently measured.

User interface Signed floating-point number

Additional information Dependency


The unit is taken from the Mass flow unit parameter (→  65)

Endress+Hauser 51
Description of device parameters Proline Promass 500 Modbus RS485

Volume flow

Navigation  Expert → Sensor → Measured val. → Process variab. → Volume flow (1847)

Description Displays the volume flow currently calculated.

User interface Signed floating-point number

Additional information Description


The volume flow is calculated from the mass flow currently measured and the density
currently measured.

Dependency
The unit is taken from the Volume flow unit parameter (→  67)

Correct.vol.flow

Navigation  Expert → Sensor → Measured val. → Process variab. → Correct.vol.flow (1851)

Description Displays the corrected volume flow that is currently measured.

User interface Signed floating-point number

Additional information Dependency


The unit is taken from the Cor.volflow unit parameter (→  69)

Density

Navigation  Expert → Sensor → Measured val. → Process variab. → Density (1850)

Description Displays the density that is currently measured.

User interface Signed floating-point number

Additional information Dependency


The unit is taken from the Density unit parameter (→  70)

Ref.density

Navigation  Expert → Sensor → Measured val. → Process variab. → Ref.density (1852)

Description Displays the reference density that is currently calculated.

52 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters

User interface Signed floating-point number

Additional information Dependency


The unit is taken from the Ref. dens. unit parameter (→  72)

Temperature

Navigation  Expert → Sensor → Measured val. → Process variab. → Temperature (1853)

Description Displays the medium temperature that is currently measured.

User interface Signed floating-point number

Additional information Dependency


The unit is taken from the Temperature unit parameter (→  72)

Pressure value

Navigation  Expert → Sensor → Measured val. → Process variab. → Pressure value (6129)

Description Displays the fixed or external pressure value.

User interface Signed floating-point number

Additional information Dependency


The unit is taken from the Pressure unit parameter (→  73)

Dynam. viscosity

Navigation  Expert → Sensor → Measured val. → Process variab. → Dynam. viscosity (1854)

Prerequisite For the following order code:


"Application package", option EG "Viscosity"
The software options currently enabled are displayed in the SW option overv.
parameter (→  49).

Description Displays the dynamic viscosity that is currently calculated.

User interface Signed floating-point number

Additional information Dependency


The unit is taken from the Dyn. visc. unit parameter.

Endress+Hauser 53
Description of device parameters Proline Promass 500 Modbus RS485

Kinematic visc.

Navigation  Expert → Sensor → Measured val. → Process variab. → Kinematic visc. (1857)

Prerequisite For the following order code:


"Application package", option EG "Viscosity"
The software options currently enabled are displayed in the SW option overv.
parameter (→  49).

Description Displays the kinematic viscosity that is currently calculated.

User interface Signed floating-point number

Additional information Dependency


The unit is taken from the Kin. visc. unit parameter (0578).

TempCompDynVisc

Navigation  Expert → Sensor → Measured val. → Process variab. → TempCompDynVisc (1872)

Prerequisite For the following order code:


"Application package", option EG "Viscosity"
The software options currently enabled are displayed in the SW option overv.
parameter (→  49).

Description Displays the temperature compensation that is currently calculated for the viscosity.

User interface Signed floating-point number

Additional information Dependency


The unit is taken from the Dyn. visc. unit parameter.

TempCompKinVisc

Navigation  Expert → Sensor → Measured val. → Process variab. → TempCompKinVisc (1863)

Prerequisite For the following order code:


"Application package", option EG "Viscosity"
The software options currently enabled are displayed in the SW option overv.
parameter (→  49).

Description Displays the temperature compensation that is currently calculated for the kinetic
viscosity.

User interface Signed floating-point number

54 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters

Additional information Dependency


The unit is taken from the Kin. visc. unit parameter (0578).

Concentration

Navigation  Expert → Sensor → Measured val. → Process variab. → Concentration (1887)

Prerequisite For the following order code:


Order code for "Application package", option ED "Concentration"
The software options currently enabled are displayed in the SW option overv.
parameter (→  49).

Description Displays the concentration that is currently calculated.

User interface Signed floating-point number

Additional information Dependency


The unit is taken from the Concentr. unit parameter (0613).

Target mass flow

Navigation  Expert → Sensor → Measured val. → Process variab. → Target mass flow (1864)

Prerequisite With the following conditions:


Order code for "Application package", option ED "Concentration"
The software options currently enabled are displayed in the SW option overv.
parameter (→  49).

Description Displays the mass flow that is currently measured for the target medium.

User interface Signed floating-point number

Additional information Dependency


The unit is taken from the Mass flow unit parameter (→  65)

Endress+Hauser 55
Description of device parameters Proline Promass 500 Modbus RS485

Carrier mass fl.

Navigation  Expert → Sensor → Measured val. → Process variab. → Carrier mass fl. (1865)

Prerequisite With the following conditions:


Order code for "Application package", option ED "Concentration"
The software options currently enabled are displayed in the SW option overv.
parameter (→  49).

Description Displays the mass flow that is currently measured for the carrier medium.

User interface Signed floating-point number

Additional information Dependency


The unit is taken from the Mass flow unit parameter (→  65)

Targ.corr.vol.fl

Navigation  Expert → Sensor → Measured val. → Process variab. → Targ.corr.vol.fl (1893)

Prerequisite With the following conditions:


• Order code for "Application package", option ED "Concentration"
• The Ethanol in water option or %mass / %volume option is selected in the Liquid type
parameter.
The software options currently enabled are displayed in the SW option overv.
parameter (→  49).

Description Displays the corrected volume flow that is currently measured for the target fluid.

User interface Signed floating-point number

Additional information Dependency


The unit is taken from the Volume flow unit parameter (→  67)

Carr.corr.vol.fl

Navigation  Expert → Sensor → Measured val. → Process variab. → Carr.corr.vol.fl (1894)

Prerequisite With the following conditions:


• Order code for "Application package", option ED "Concentration"
• In the Liquid type parameter, the Ethanol in water option or %mass / %volume option
is selected.
The software options currently enabled are displayed in the SW option overv.
parameter (→  49).

Description Displays the corrected volume flow currently measured for the carrier fluid.

56 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters

User interface Signed floating-point number

Additional information Dependency


The unit is taken from the Volume flow unit parameter (→  67)

Target vol. flow

Navigation  Expert → Sensor → Measured val. → Process variab. → Target vol. flow (1895)

Prerequisite With the following conditions:


• Order code for "Application package", option ED "Concentration"
• The Ethanol in water option or %mass / %volume option is selected in the Liquid type
parameter.
• The %vol option is selected in the Concentr. unit parameter.
The software options currently enabled are displayed in the SW option overv.
parameter (→  49).

Description Displays the volume flow currently measured for the target medium.

User interface Signed floating-point number

Additional information Dependency


The unit is taken from the Volume flow unit parameter (→  67)

Carrier vol. fl.

Navigation  Expert → Sensor → Measured val. → Process variab. → Carrier vol. fl. (1896)

Prerequisite With the following conditions:


• Order code for "Application package", option ED "Concentration"
• The Ethanol in water option or %mass / %volume option is selected in the Liquid type
parameter.
• The %vol option is selected in the Concentr. unit parameter.
The software options currently enabled are displayed in the SW option overv.
parameter (→  49).

Description Use this function to display the volume flow currently measured for the carrier medium.

User interface Signed floating-point number

Additional information Dependency


The unit is taken from the Volume flow unit parameter (→  67)

Endress+Hauser 57
Description of device parameters Proline Promass 500 Modbus RS485

"Totalizer" submenu

Navigation  Expert → Sensor → Measured val. → Totalizer

‣ Totalizer

Totalizer val. 1 to n (0911–1 to n) →  58

Tot. overflow 1 to n (0910–1 to n) →  58

Totalizer val. 1 to n 

Navigation  Expert → Sensor → Measured val. → Totalizer → Totalizer val. 1 to n (0911–1 to n)

Prerequisite A process variable is selected in the Assign variable parameter (→  182) of the
Totalizer 1 to n submenu.

Description Displays the current totalizer reading.

User interface Signed floating-point number

Additional information Description


As it is only possible to display a maximum of 7 digits in the operating tool, the current
counter value is the sum of the totalizer value and the overflow value from the Tot.
overflow 1 to n parameter if the display range is exceeded.
In the event of an error, the totalizer adopts the mode defined in the Failure mode
parameter (→  186).

User interface
The value of the process variable totalized since measuring began can be positive or
negative. This depends on the settings in the Operation mode parameter (→  184).
The unit of the selected process variable is specified for the totalizer in the Unit
totalizer parameter (→  183).

Example
Calculation of the current totalizer reading when the value exceeds the 7-digit display
range of the operating tool:
• Value in the Totalizer val. 1 parameter: 1 968 457 m3
• Value in the Tot. overflow 1 parameter: 1 ⋅ 107 (1 overflow) = 10 000 000 [m3]
• Current totalizer reading: 11 968 457 m3

Tot. overflow 1 to n 

Navigation  Expert → Sensor → Measured val. → Totalizer → Tot. overflow 1 to n (0910–1 to n)

Prerequisite A process variable is selected in the Assign variable parameter (→  182) of the
Totalizer 1 to n submenu.

58 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters

Description Displays the current totalizer overflow.

User interface Integer with sign

Additional information Description


If the current totalizer reading exceeds 7 digits, which is the maximum value range that
can be displayed by the operating tool, the value above this range is output as an overflow.
The current totalizer value is therefore the sum of the overflow value and the totalizer
value from the Totalizer val. 1 to n parameter.

User interface
The unit of the selected process variable is specified for the totalizer in the Unit
totalizer parameter (→  183).

Example
Calculation of the current totalizer reading when the value exceeds the 7-digit display
range of the operating tool:
• Value in the Totalizer val. 1 parameter: 1 968 457 m3
• Value in the Tot. overflow 1 parameter: 2 ⋅ 107 (2 overflows) = 20 000 000 [m3]
• Current totalizer reading: 21 968 457 m3

"Input values" submenu

Navigation  Expert → Sensor → Measured val. → Input values

‣ Input values

‣ Current input 1 to n →  59

‣ Val.stat.inp. 1 to n →  60

"Current input 1 to n" submenu

Navigation  Expert → Sensor → Measured val. → Input values → Current


input 1 to n

‣ Current input 1 to n

Measured val. 1 to n (1603–1 to n) →  60

Measur. curr. 1 to n (1604–1 to n) →  60

Endress+Hauser 59
Description of device parameters Proline Promass 500 Modbus RS485

Measured val. 1 to n

Navigation  Expert → Sensor → Measured val. → Input values → Current input 1 to n


→ Measured val. 1 to n (1603–1 to n)

Description Displays the current input value.

User interface Signed floating-point number

Measur. curr. 1 to n

Navigation  Expert → Sensor → Measured val. → Input values → Current input 1 to n → Measur.
curr. 1 to n (1604–1 to n)

Description Displays the current value of the current input.

User interface 0 to 22.5 mA

"Value status input 1 to n" submenu

Navigation  Expert → Sensor → Measured val. → Input values → Val.stat.inp.


1 to n

‣ Val.stat.inp. 1 to n

Val.stat.inp. (1353–1 to n) →  60

Val.stat.inp.

Navigation  Expert → Sensor → Measured val. → Input values → Val.stat.inp. 1 to n


→ Val.stat.inp. (1353–1 to n)

Description Displays the current input signal level.

User interface • High


• Low

60 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters

"Output values" submenu

Navigation  Expert → Sensor → Measured val. → Output values

‣ Output values

‣ Value curr.out 1 to n →  61

‣ PFS output 1 to n →  62

‣ Relay output 1 to n →  63

‣ Double pulse out →  64

"Value current output 1 to n" submenu

Navigation  Expert → Sensor → Measured val. → Output values → Value


curr.out 1 to n

‣ Value curr.out 1 to n

Output curr. 1 to n (0361–1 to n) →  61

Measur. curr. 1 to n (0366–1 to n) →  61

Output curr. 1 to n

Navigation  Expert → Sensor → Measured val. → Output values → Value curr.out 1 to n → Output
curr. 1 to n (0361–1 to n)

Description Displays the current value currently calculated for the current output.

User interface 0 to 22.5 mA

Measur. curr. 1 to n

Navigation  Expert → Sensor → Measured val. → Output values → Value curr.out 1 to n


→ Measur. curr. 1 to n (0366–1 to n)

Description Use this function to display the actual measured value of the output current.

User interface 0 to 30 mA

Endress+Hauser 61
Description of device parameters Proline Promass 500 Modbus RS485

"Pulse/frequency/switch output 1 to n" submenu

Navigation  Expert → Sensor → Measured val. → Output values → PFS output


1 to n

‣ PFS output 1 to n

Output freq. 1 to n (0471–1 to n) →  62

Pulse output 1 to n (0456–1 to n) →  62

Switch status 1 to n (0461–1 to n) →  63

Output freq. 1 to n

Navigation  Expert → Sensor → Measured val. → Output values → PFS output 1 to n → Output
freq. 1 to n (0471–1 to n)

Prerequisite In the Operating mode parameter (→  132), the Frequency option is selected.

Description Displays the actual value of the output frequency which is currently measured.

User interface 0.0 to 12 500.0 Hz

Pulse output 1 to n

Navigation  Expert → Sensor → Measured val. → Output values → PFS output 1 to n → Pulse
output 1 to n (0456–1 to n)

Prerequisite The Pulse option is selected in the Operating mode parameter (→  132) parameter.

Description Displays the pulse frequency currently output.

User interface Positive floating-point number

Additional information Description


• The pulse output is an open collector output.
• This is configured at the factory in such a way that the transistor is conductive for the
duration of the pulse (NO contact) and is safety-oriented.

62 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters

1 1

0 t 0 t
NC NO
A0028726

0 Non-conductive
1 Conductive
NC NC contact (normally closed)
NO NO contact (normally open)

The output behavior can be reversed via the Invert outp.sig. parameter (→  152) i.e.
the transistor does not conduct for the duration of the pulse.
In addition, the behavior of the output in the event of a device alarm (Failure mode
parameter (→  136)) can be configured.

Switch status 1 to n

Navigation  Expert → Sensor → Measured val. → Output values → PFS output 1 to n → Switch
status 1 to n (0461–1 to n)

Prerequisite The Switch option is selected in the Operating mode parameter (→  132).

Description Displays the current switch status of the status output.

User interface • Open


• Closed

Additional information User interface


• Open
The switch output is not conductive.
• Closed
The switch output is conductive.

"Relay output 1 to n" submenu

Navigation  Expert → Sensor → Measured val. → Output values → Relay


output 1 to n

‣ Relay output 1 to n

Switch status (0801–1 to n) →  64

Switch cycles (0815–1 to n) →  64

Max. cycles no. (0817–1 to n) →  64

Endress+Hauser 63
Description of device parameters Proline Promass 500 Modbus RS485

Switch status

Navigation  Expert → Sensor → Measured val. → Output values → Relay output 1 to n → Switch
status (0801–1 to n)

Description Displays the current status of the relay output.

User interface • Open


• Closed

Additional information User interface


• Open
The relay output is not conductive.
• Closed
The relay output is conductive.

Switch cycles

Navigation  Expert → Sensor → Measured val. → Output values → Relay output 1 to n → Switch
cycles (0815–1 to n)

Description Displays all the switch cycles performed.

User interface Positive integer

Max. cycles no.

Navigation  Expert → Sensor → Measured val. → Output values → Relay output 1 to n → Max.
cycles no. (0817–1 to n)

Description Displays the maximum number of guaranteed switch cycles.

User interface Positive integer

"Double pulse output" submenu

Navigation  Expert → Sensor → Measured val. → Output values → Double


pulse out

‣ Double pulse out

Pulse output (0987) →  65

64 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters

Pulse output

Navigation  Expert → Sensor → Measured val. → Output values → Double pulse out → Pulse
output (0987)

Description Displays the pulse frequency of the double pulse output which is currently output.

User interface Positive floating-point number

Additional information For a detailed description and example: Pulse output parameter (→  62)

3.2.2 "System units" submenu

Navigation  Expert → Sensor → System units

‣ System units

Mass flow unit (0554) →  65

Mass unit (0574) →  66

Volume flow unit (0553) →  67

Volume unit (0563) →  68

Cor.volflow unit (0558) →  69

Corr. vol. unit (0575) →  70

Density unit (0555) →  70

Ref. dens. unit (0556) →  72

Temperature unit (0557) →  72

Pressure unit (0564) →  73

Date/time format (2812) →  73

Mass flow unit 

Navigation  Expert → Sensor → System units → Mass flow unit (0554)

Description Use this function to select the unit for the mass flow.

Endress+Hauser 65
Description of device parameters Proline Promass 500 Modbus RS485

Selection SI units US units


• g/s • oz/s
• g/min • oz/min
• g/h • oz/h
• g/d • oz/d
• kg/s • lb/s
• kg/min • lb/min
• kg/h • lb/h
• kg/d • lb/d
• t/s • STon/s
• t/min • STon/min
• t/h • STon/h
• t/d • STon/d

Factory setting Country-specific:


• kg/h (DN > 150 (6"): t/h)
• lb/min

Additional information Result


The selected unit applies for:
• Target mass flow parameter (→  55)
• Carrier mass fl. parameter (→  56)
• Mass flow parameter (→  51)

Selection
For an explanation of the abbreviated units: →  254

Customer-specific units
The unit for the customer-specific mass is specified in the Mass text parameter
(→  76).

Mass unit 

Navigation  Expert → Sensor → System units → Mass unit (0574)

Description Use this function to select the unit for the mass.

Selection SI units US units


•g • oz
• /g • /oz
• kg • lb
• /kg • /lb
•t • STon
• /t • /STon

Factory setting Country-specific:


• kg (DN > 150 (6"): t)
• lb

66 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters

Additional information Selection


For an explanation of the abbreviated units: →  254

Customer-specific units
The unit for the customer-specific mass is specified in the Mass text parameter
(→  76).

Volume flow unit 

Navigation  Expert → Sensor → System units → Volume flow unit (0553)

Description Use this function to select the unit for the volume flow.

Selection SI units US units Imperial units


• cm³/s • af/s • gal/s (imp)
• cm³/min • af/min • gal/min (imp)
• cm³/h • af/h • gal/h (imp)
• cm³/d • af/d • gal/d (imp)
• dm³/s • ft³/s • Mgal/s (imp)
• dm³/min • ft³/min • Mgal/min (imp)
• dm³/h • ft³/h • Mgal/h (imp)
• dm³/d • ft³/d • Mgal/d (imp)
• m³/s • MMft³/s • bbl/s (imp;oil)
• m³/min • MMft³/min • bbl/min (imp;oil)
• m³/h • MMft³/h • bbl/h (imp;oil)
• m³/d • Mft³/d • bbl/d (imp;oil)
• ml/s • fl oz/s (us)
• ml/min • fl oz/min (us)
• ml/h • fl oz/h (us)
• ml/d • fl oz/d (us)
• l/s • gal/s (us)
• l/min • gal/min (us)
• l/h • gal/h (us)
• l/d • gal/d (us)
• hl/s • Mgal/s (us)
• hl/min • Mgal/min (us)
• hl/h • Mgal/h (us)
• hl/d • Mgal/d (us)
• Ml/s • bbl/s (us;oil)
• Ml/min • bbl/min (us;oil)
• Ml/h • bbl/h (us;oil)
• Ml/d • bbl/d (us;oil)
• bbl/s (us;tank)
• bbl/min (us;tank)
• bbl/h (us;tank)
• bbl/d (us;tank)
• kgal/s (us)
• kgal/min (us)
• kgal/h (us)
• kgal/d (us)

or

Endress+Hauser 67
Description of device parameters Proline Promass 500 Modbus RS485

US units Imperial units


• bbl/s (us;liq.) * • bbl/s (imp;beer) *
• bbl/min (us;liq.) * • bbl/min (imp;beer) *
• bbl/h (us;liq.) * • bbl/h (imp;beer) *
• bbl/d (us;liq.) * • bbl/d (imp;beer) *
• bbl/s (us;beer) *
• bbl/min (us;beer) *
• bbl/h (us;beer) *
• bbl/d (us;beer) *

* Visibility depends on order options or device settings

Factory setting Country-specific:


• l/h (DN > 150 (6"): m³/h)
• gal/min (us)

Additional information Result


The selected unit applies for:
Volume flow parameter (→  52)

Selection
For an explanation of the abbreviated units: →  254

Customer-specific units
The unit for the customer-specific volume is specified in the Volume text parameter
(→  75).

Volume unit 

Navigation  Expert → Sensor → System units → Volume unit (0563)

Description Use this function to select the unit for the volume.

Selection SI units US units Imperial units


• cm³ • af • gal (imp)
• /cm³ • /af • /gal (imp)
• dm³ • ft³ • Mgal (imp)
• /dm³ • /ft³ • /Mgal (imp)
• m³ • Mft³ • bbl (imp;oil)
• /m³ • /MMft³ • /bbl (imp;oil)
• ml • fl oz (us)
• /ml • /fl oz (us)
•l • gal (us)
• /l • /gal (us)
• hl • kgal (us)
• /hl • /kgal (us)
• Ml Mega • Mgal (us)
• /Ml • /Mgal (us)
• bbl (us;oil)
• /bbl (us;oil)
• bbl (us;tank)
• /bbl (us;tank)

68 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters

or

US units Imperial units


• bbl (us;liq.) * • bbl (imp;beer) *
• /bbl (us;liq.) * • /bbl (imp;beer) *
• bbl (us;beer) *
• /bbl (us;beer) *

* Visibility depends on order options or device settings

Factory setting Country-specific:


• l (DN > 150 (6"): m³)
• gal (us)

Additional information Selection


For an explanation of the abbreviated units: →  254

Customer-specific units
The unit for the customer-specific volume is specified in the Volume text parameter
(→  75).

Cor.volflow unit 

Navigation  Expert → Sensor → System units → Cor.volflow unit (0558)

Description Use this function to select the unit for the corrected volume flow.

Selection SI units US units Imperial units


• Nl/s • Sft³/s • Sgal/s (imp)
• Nl/min • Sft³/min • Sgal/min (imp)
• Nl/h • Sft³/h • Sgal/h (imp)
• Nl/d • Sft³/d • Sgal/d (imp)
• Nhl/s • Sgal/s (us)
• Nhl/min • Sgal/min (us)
• Nhl/h • Sgal/h (us)
• Nhl/d • Sgal/d (us)
• Nm³/s • Sbbl/s (us;liq.)
• Nm³/min • Sbbl/min (us;liq.)
• Nm³/h • Sbbl/h (us;liq.)
• Nm³/d • Sbbl/d (us;liq.)
• Sl/s • MMSft³/s
• Sl/min • MMSft³/min
• Sl/h • MMSft³/h
• Sl/d • Sbbl/s (us;oil)
• Sm³/s • Sbbl/min (us;oil)
• Sm³/min • Sbbl/h (us;oil)
• Sm³/h • Sbbl/d (us;oil)
• Sm³/d
• MSft³/d

Factory setting Country-specific:


• Nl/h (DN > 150 (6"): Nm³/h)
• Sft³/min

Endress+Hauser 69
Description of device parameters Proline Promass 500 Modbus RS485

Additional information Result


The selected unit applies for:
Correct.vol.flow parameter (→  52)

Selection
For an explanation of the abbreviated units: →  254

Corr. vol. unit 

Navigation  Expert → Sensor → System units → Corr. vol. unit (0575)

Description Use this function to select the unit for the corrected volume.

Selection SI units US units Imperial units


• Nl • Sft³ • Sgal (imp)
• /Nl • /Sft³ • /Sgal (imp)
• Nhl • MMSft³
• /Nhl • /MMSft³
• Nm³ • Sgal (us)
• /Nm³ • /Sgal (us)
• Sl • Sbbl (us;liq.)
• /Sl • /Sbbl (us;liq.)
• Sm³ • Sbbl (us;oil)
• /Sm³ • /Sbbl (us;oil)

Factory setting Country-specific:


• Nl (DN > 150 (6"): Nm³)
• Sft³

Additional information Selection


For an explanation of the abbreviated units: →  254

Density unit 

Navigation  Expert → Sensor → System units → Density unit (0555)

Description Use this function to select the unit for the density.

70 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters

Selection SI units US units Imperial units


• g/cm³ • lb/ft³ • lb/gal (imp)
• g/m³ • lb/gal (us) • lb/bbl (imp;oil)
• g/ml • lb/bbl (us;oil)
• kg/l • lb/bbl (us;tank)
• kg/dm³
• kg/m³
• SD4°C
• SD15°C
• SD20°C
• SG4°C
• SG15°C
• SG20°C

or

US units Other units


SG60°F * °API *

* Visibility depends on order options or device settings

or

US units Imperial units


• lb/bbl (us;liq.) * lb/bbl (imp;beer) *
• lb/bbl (us;beer) *

* Visibility depends on order options or device settings

Factory setting Country-specific:


• kg/l
• lb/ft³

Additional information Result


The selected unit applies for:
• Density setpt 1 parameter
• Density setpt 2 parameter
• Density parameter (→  52)

Selection
• SD = specific density
The specific density is the ratio of the medium density to the water density at a water
temperature of +4 °C (+39 °F), +15 °C (+59 °F), +20 °C (+68 °F).
• SG = specific gravity
The specific gravity is the ratio of the medium density to the water density at a water
temperature of +4 °C (+39 °F), +15 °C (+59 °F), +20 °C (+68 °F).
For an explanation of the abbreviated units: →  254

Customer-specific units
The unit for the customer-specific density is specified in the Density text parameter
(→  78).

Endress+Hauser 71
Description of device parameters Proline Promass 500 Modbus RS485

Ref. dens. unit 

Navigation  Expert → Sensor → System units → Ref. dens. unit (0556)

Description Use this function to select the unit for the reference density.

Selection SI units US units Other units


• kg/Nm³ • lb/Sft³ °APIbase
• kg/Nl • RD60°F
• g/Scm³
• kg/Sm³
• RD15°C
• RD20°C

Factory setting Country-dependent


• kg/Nl
• lb/Sft³

Additional information Result


The selected unit applies for:
• Ext. ref.density parameter (→  96)
• Fix ref.density parameter (→  97)
• Ref.density parameter (→  52)

Selection
For an explanation of the abbreviated units: →  254

Temperature unit 

Navigation  Expert → Sensor → System units → Temperature unit (0557)

Description Use this function to select the unit for the temperature.

Selection SI units US units


• °C • °F
•K • °R

Factory setting Country-specific:


• °C
• °F

Additional information Result


The selected unit applies for:
• Maximum value parameter (→  228)
• Minimum value parameter (→  228)
• Maximum value parameter (→  229)
• Minimum value parameter (→  229)
• Maximum value parameter (→  230)
• Minimum value parameter (→  230)
• External temp. parameter (→  94)

72 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters

• Ref. temperature parameter (6222)


• Temperature parameter (→  53)
• Ref. temperature parameter (→  97)

Selection
For an explanation of the abbreviated units: →  254

Pressure unit 

Navigation  Expert → Sensor → System units → Pressure unit (0564)

Description Use this function to select the unit for the pipe pressure.

Selection SI units US units


• Pa a • psi a
• kPa a • psi g
• MPa a
• bar
• Pa g
• kPa g
• MPa g
• bar g

Factory setting Country-specific:


• bar a
• psi a

Additional information Result


The unit is taken from:
• Pressure value parameter (→  93)
• External press. parameter (→  93)
• Pressure value parameter (→  53)

Selection
For an explanation of the abbreviated units: →  254

Date/time format 

Navigation  Expert → Sensor → System units → Date/time format (2812)

Description Use this function to select the desired time format for calibration history.

Selection • dd.mm.yy hh:mm


• dd.mm.yy am/pm
• mm/dd/yy hh:mm
• mm/dd/yy am/pm

Factory setting dd.mm.yy hh:mm

Endress+Hauser 73
Description of device parameters Proline Promass 500 Modbus RS485

Additional information Selection


For an explanation of the abbreviated units: →  254

"User-spec. units" submenu

Navigation  Expert → Sensor → System units → User-spec. units

‣ User-spec. units

Volume text (0567) →  75

Volume offset (0569) →  75

Volume factor (0568) →  76

Mass text (0560) →  76

Mass offset (0562) →  76

Mass factor (0561) →  77

Corr. vol. text (0592) →  77

Corr vol. offset (0602) →  78

Cor.vol. factor (0590) →  78

Density text (0570) →  78

Density offset (0571) →  79

Density factor (0572) →  79

Spec. enth. text (0585) →  79

Spec. enth. off. (0584) →  80

Spec. enth. fac. (0583) →  80

Energy text (0600) →  80

Energy offset (0599) →  81

Energy factor (0586) →  81

Pressure text (0581) →  81

74 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters

Pressure offset (0580) →  81

Pressure factor (0579) →  82

Volume text 

Navigation  Expert → Sensor → System units → User-spec. units → Volume text (0567)

Description Use this function to enter a text for the user-specific unit of volume and volume flow. The
corresponding time units (s, min, h, d) for volume flow are generated automatically.

User entry Max. 10 characters such as letters, numbers or special characters (@, %, /)

Factory setting User vol.

Additional information Result


The defined unit is shown as an option in the choose list of the following parameters:
• Volume flow unit parameter (→  67)
• Volume unit parameter (→  68)

Example
If the text GLAS is entered, the choose list of the Volume flow unit parameter (→  67)
shows the following options:
• GLAS/s
• GLAS/min
• GLAS/h
• GLAS/d

Volume offset 

Navigation  Expert → Sensor → System units → User-spec. units → Volume offset (0569)

Description Use this function to enter the offset for adapting the user-specific volume unit and volume
flow unit (without time).

User entry Signed floating-point number

Factory setting 0

Additional information Description


Value in user-specific unit = (factor × value in base unit) + offset

Endress+Hauser 75
Description of device parameters Proline Promass 500 Modbus RS485

Volume factor 

Navigation  Expert → Sensor → System units → User-spec. units → Volume factor (0568)

Description Use this function to enter a quantity factor (without time) for the user-specific volume and
volume flow unit.

User entry Signed floating-point number

Factory setting 1.0

Mass text 

Navigation  Expert → Sensor → System units → User-spec. units → Mass text (0560)

Description Use this function to enter a text for the user-specific unit of mass and mass flow. The
corresponding time units (s, min, h, d) for mass flow are generated automatically.

User entry Max. 10 characters such as letters, numbers or special characters (@, %, /)

Factory setting User mass

Additional information Result


The defined unit is shown as an option in the choose list of the following parameters:
• Mass flow unit parameter (→  65)
• Mass unit parameter (→  66)

Example
If the text CENT for "centner" is entered, the following options are displayed in the picklist
for the Mass flow unit parameter (→  65):
• CENT/s
• CENT/min
• CENT/h
• CENT/d

Mass offset 

Navigation  Expert → Sensor → System units → User-spec. units → Mass offset (0562)

Description Use this function to enter the zero point shift for the user-specific mass and mass flow
unit.

User entry Signed floating-point number

Factory setting 0

76 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters

Additional information Description


Value in user-specific unit = (factor × value in basic unit) + offset

Mass factor 

Navigation  Expert → Sensor → System units → User-spec. units → Mass factor (0561)

Description Use this function to enter a quantity factor (without time) for the user-specific mass and
mass flow unit.

User entry Signed floating-point number

Factory setting 1.0

Additional information Example


Mass of 1 Zentner = 50 kg → 0.02 Zentner = 1 kg → entry: 0.02

Corr. vol. text 

Navigation  Expert → Sensor → System units → User-spec. units → Corr. vol. text (0592)

Description Use this function to enter a text for the user-specific unit of the corrected volume and
corrected volume flow. The corresponding time units (s, min, h, d) for mass flow are
generated automatically.

User entry Max. 10 characters such as letters, numbers or special characters (@, %, /)

Factory setting UserCrVol.

Additional information Result


The defined unit is shown as an option in the choose list of the following parameters:
• Cor.volflow unit parameter (→  69)
• Corr. vol. unit parameter (→  70)

Example
If the text GLAS is entered, the choose list of the Cor.volflow unit parameter (→  69)
shows the following options:
• GLAS/s
• GLAS/min
• GLAS/h
• GLAS/d

Endress+Hauser 77
Description of device parameters Proline Promass 500 Modbus RS485

Corr vol. offset 

Navigation  Expert → Sensor → System units → User-spec. units → Corr vol. offset (0602)

Description Use this function to enter the offset for adapting the user-specific corrected volume unit
and corrected volume flow unit (without time).
Value in user-specific unit = (factor × value in base unit) + offset

User entry Signed floating-point number

Factory setting 0

Cor.vol. factor 

Navigation  Expert → Sensor → System units → User-spec. units → Cor.vol. factor (0590)

Description Use this function to enter a quantity factor (without time) for the user-specific corrected
volume unit and corrected volume flow unit.

User entry Signed floating-point number

Factory setting 1.0

Density text 

Navigation  Expert → Sensor → System units → User-spec. units → Density text (0570)

Description Use this function to enter a text or the user-specific unit of density.

User entry Max. 10 characters such as letters, numbers or special characters (@, %, /)

Factory setting User dens.

Additional information Result


The defined unit is shown as an option in the choose list of the Density unit
parameter (→  70).

Example
Enter text “CE_L” for centners per liter

78 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters

Density offset 

Navigation  Expert → Sensor → System units → User-spec. units → Density offset (0571)

Description Use this function to enter the zero point shift for the user-specific density unit.
Value in user-specific unit = (factor × value in basic unit) + offset

User entry Signed floating-point number

Factory setting 0

Density factor 

Navigation  Expert → Sensor → System units → User-spec. units → Density factor (0572)

Description Use this function to enter a quantity factor for the user-specific density unit.

User entry Signed floating-point number

Factory setting 1.0

Spec. enth. text 

Navigation  Expert → Sensor → System units → User-spec. units → Spec. enth. text (0585)

User entry Max. 10 characters such as letters, numbers or special characters (@, %, /)

Factory setting User enth.

Additional information Result

Example
If the text CAL is entered, the choose list of the Cal. value unit parameter shows the
following options:
• CAL/Nm3
• CAL/m3
• CAL/ft3
• CAL/Sft3

Endress+Hauser 79
Description of device parameters Proline Promass 500 Modbus RS485

Spec. enth. off. 

Navigation  Expert → Sensor → System units → User-spec. units → Spec. enth. off. (0584)

Description Use this function to enter the offset for adapting the user-specific calorific value unit
(without volume).

User entry Signed floating-point number

Factory setting 0

Spec. enth. fac. 

Navigation  Expert → Sensor → System units → User-spec. units → Spec. enth. fac. (0583)

Description Use this function to enter a quantity factor (without volume) for the user-specific calorific
value unit.

User entry Signed floating-point number

Factory setting 1.0

Additional information Example


1 W × min = 60 J → 0.166 W × min = 1 J → user entry: 0.0166

Energy text 

Navigation  Expert → Sensor → System units → User-spec. units → Energy text (0600)

Description Use this function to enter a text for the user-specific energy unit.

User entry Max. 10 characters such as letters, numbers or special characters (@, %, /)

Factory setting User en.

Additional information Result


The defined unit is shown as an option in the choose list of the following parameters:
• Energy unit parameter
• Energy flow unit parameter

Example
If the text W is entered, the choose list of the Energy flow unit parameter shows the
following options:
• W/s
• W/min
• W/h
• W/d

80 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters

Energy offset 

Navigation  Expert → Sensor → System units → User-spec. units → Energy offset (0599)

Description Use this function to enter the offset for adapting the user-specific energy unit (without
time).

User entry Signed floating-point number

Factory setting 0

Energy factor 

Navigation  Expert → Sensor → System units → User-spec. units → Energy factor (0586)

Description Use this function to enter a quantity factor for the user-specific energy unit.

User entry Signed floating-point number

Factory setting 1.0

Pressure text 

Navigation  Expert → Sensor → System units → User-spec. units → Pressure text (0581)

Description Use this function to enter a text for the user-specific pressure unit.

User entry Max. 10 characters such as letters, numbers or special characters (@, %, /)

Factory setting User pres.

Additional information Result


The defined unit is shown as an option in the choose list of the Pressure unit
parameter (→  73).

Pressure offset 

Navigation  Expert → Sensor → System units → User-spec. units → Pressure offset (0580)

Description Use this function to enter the offset for adapting the user-specific pressure unit.

User entry Signed floating-point number

Factory setting 0

Endress+Hauser 81
Description of device parameters Proline Promass 500 Modbus RS485

Pressure factor 

Navigation  Expert → Sensor → System units → User-spec. units → Pressure factor (0579)

Description Use this function to enter a quantity factor for the user-specific pressure unit.

User entry Signed floating-point number

Factory setting 1.0

Additional information Example


1 Dyn/cm² = 0.1 Pa → 10 Dyn/cm² = 1 Pa → user entry: 10

3.2.3 "Process param." submenu

Navigation  Expert → Sensor → Process param.

‣ Process param.

Flow damping (1802) →  82

Density damping (1803) →  83

Temp. damping (1822) →  83

Flow override (1839) →  84

‣ Low flow cut off →  84

‣ Partial pipe det →  87

Flow damping 

Navigation  Expert → Sensor → Process param. → Flow damping (1802)

Description Use this function to enter a time constant for flow damping (PT1 element). Reduction of
the variability of the flow measured value (in relation to interference). For this purpose,
the depth of the flow filter is adjusted: when the filter setting increases, the reaction time
of the device also increases.

User entry 0 to 100.0 s

Factory setting 0s

82 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters

Additional information Description


The damping is performed by a PT1 element 2).

User entry
• Value = 0: no damping
• Value > 0: damping is increased
Damping is switched off if 0 is entered (factory setting).

Result
The damping affects the following variables of the device:
• Outputs →  115
• Low flow cut off →  84
• Totalizers →  182

Density damping 

Navigation  Expert → Sensor → Process param. → Density damping (1803)

Description Use this function to enter a time constant for the damping (PT1 element) of the density
measured value.

User entry 0 to 999.9 s

Factory setting 0s

Additional information Description


The damping is performed by a PT1 element 3).

User entry
• Value = 0: no damping
• Value > 0: damping is increased
Damping is switched off if 0 is entered (factory setting).

Temp. damping 

Navigation  Expert → Sensor → Process param. → Temp. damping (1822)

Description Use this function to enter a time constant for the damping (PT1 element) of the
temperature measured value.

User entry 0 to 999.9 s

2) Proportional behavior with first-order lag


3) Proportional behavior with first-order lag

Endress+Hauser 83
Description of device parameters Proline Promass 500 Modbus RS485

Factory setting 0s

Additional information Description


The damping is performed by a PT1 element 4).

User entry
• Value = 0: no damping
• Value > 0: damping is increased
Damping is switched off if 0 is entered (factory setting).

Flow override 

Navigation  Expert → Sensor → Process param. → Flow override (1839)

Description Use this function to select whether to interrupt the evaluation of measured values. This is
useful for the cleaning processes of a pipeline, for example.

Selection • Off
• On

Factory setting Off

Additional information Description


Flow override is active
• The diagnostic message C453 Flow override is output.
• Output values
• Temperature: continues to be output
• Totalizers 1-3: stop being totalized
The Flow override option can also be activated in the Status input submenu: Assign
stat.inp. parameter (→  114).

"Low flow cut off" submenu

Navigation  Expert → Sensor → Process param. → Low flow cut off

‣ Low flow cut off

Assign variable (1837) →  85

On value (1805) →  85

4) Proportional behavior with first-order lag

84 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters

Off value (1804) →  85

Pres. shock sup. (1806) →  86

Assign variable 

Navigation  Expert → Sensor → Process param. → Low flow cut off → Assign variable (1837)

Description Use this function to select the process variable for low flow cutoff detection.

Selection • Off
• Mass flow
• Volume flow
• Correct.vol.flow *

Factory setting Mass flow

On value 

Navigation  Expert → Sensor → Process param. → Low flow cut off → On value (1805)

Prerequisite A process variable is selected in the Assign variable parameter (→  85).

Description Use this function to enter a switch-on value for low flow cut off. Low flow cut off is
activated if the value entered is not equal to 0 →  85.

User entry Positive floating-point number

Factory setting Depends on country and nominal diameter→  249

Additional information Dependency


The unit depends on the process variable selected in the Assign variable parameter
(→  85).

Off value 

Navigation  Expert → Sensor → Process param. → Low flow cut off → Off value (1804)

Prerequisite A process variable is selected in the Assign variable parameter (→  85).

Description Use this function to enter a switch-off value for low flow cut off. The off value is entered as
a positive hysteresis from the on value→  85.

* Visibility depends on order options or device settings

Endress+Hauser 85
Description of device parameters Proline Promass 500 Modbus RS485

User entry 0 to 100.0 %

Factory setting 50 %

Additional information Example

Q 2 2

4
H
3
t
1 A 1 A

A0012887

Q Flow
t Time
H Hysteresis
A Low flow cut off active
1 Low flow cut off is activated
2 Low flow cut off is deactivated
3 On value entered
4 Off value entered

Pres. shock sup. 

Navigation  Expert → Sensor → Process param. → Low flow cut off → Pres. shock sup. (1806)

Prerequisite A process variable is selected in the Assign variable parameter (→  85).

Description Use this function to enter the time interval for signal suppression (= active pressure shock
suppression).

User entry 0 to 100 s

Factory setting 0s

Additional information Description


Pressure shock suppression is enabled
• Prerequisite:
• Flow rate < on-value of low flow cut off
or
• Changing the flow direction
• Output values
• Flow displayed: 0
• Totalizer: the totalizers are pegged at the last correct value

86 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters

Pressure shock suppression is disabled


• Prerequisite: the time interval set in this function has elapsed.
• If the flow also exceeds the switch-off value for low flow cut off, the device starts
processing the current flow value again and displays it.

Example
When closing a valve, momentarily strong fluid movements may occur in the pipeline,
which are registered by the measuring system. These totalized flow values lead to a false
totalizer status, particularly during batching processes.

Q D C D

1 2 3 4

6
5
t

A B

A0012888

Q Flow
t Time
A Drip
B Pressure shock
C Pressure shock suppression active as specified by the time entered
D Pressure shock suppression inactive
1 Valve closes
2 Flow falls below the on-value of the low flow cut off: pressure shock suppression is activated
3 The time entered has elapsed: pressure shock suppression is deactivated
4 The actual flow value is now displayed and output
5 On-value for low flow cut off
6 Off-value for low flow cut off

"Partial pipe det" submenu

Navigation  Expert → Sensor → Process param. → Partial pipe det

‣ Partial pipe det

Assign variable (1860) →  88

Low value (1861) →  88

High value (1858) →  89

Endress+Hauser 87
Description of device parameters Proline Promass 500 Modbus RS485

Response time (1859) →  89

Max. damping (6040) →  89

Assign variable 

Navigation  Expert → Sensor → Process param. → Partial pipe det → Assign variable (1860)

Description Use this function to select a process variable to detect empty or partially filled measuring
tubes.
For gas measurement: Deactivate monitoring due to low gas density.

Selection • Off
• Density
• Ref.density

Factory setting Off

Low value 

Navigation  Expert → Sensor → Process param. → Partial pipe det → Low value (1861)

Prerequisite A process variable is selected in the Assign variable parameter (→  88).

Description Use this function to enter a lower limit value to enable detection of empty or partially filled
measuring tubes. If the measured density falls below this value, monitoring is enabled.

User entry Signed floating-point number

Factory setting 200

Additional information User entry


The lower limit value must be less than the upper limit value defined in the High value
parameter (→  89).
The unit depends on the process variable selected in the Assign variable parameter
(→  88).

Limit value
If the displayed value is outside the limit value, the measuring device displays the
diagnostic message S862 Partly filled.

88 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters

High value 

Navigation  Expert → Sensor → Process param. → Partial pipe det → High value (1858)

Prerequisite A process variable is selected in the Assign variable parameter (→  88).

Description Use this function to enter an upper limit value to enable detection of empty or partially
filled measuring tubes. If the measured density exceeds this value, detection is enabled.

User entry Signed floating-point number

Factory setting 6 000

Additional information User entry


The upper limit value must be greater than the lower limit value defined in the Low value
parameter (→  88).
The unit depends on the process variable selected in the Assign variable parameter
(→  88).

Limit value
If the displayed value is outside the limit value, the measuring device displays the
diagnostic message S862 Partly filled.

Response time 

Navigation  Expert → Sensor → Process param. → Partial pipe det → Response time (1859)

Prerequisite A process variable is selected in the Assign variable parameter (→  88).

Description Use this function to enter the minimum length of time (debouncing time) the signal must
be present for the diagnostic message S862 Partly filled to be triggered if the
measuring pipe is empty or partially full.

User entry 0 to 100 s

Factory setting 1s

Max. damping 

Navigation  Expert → Sensor → Process param. → Partial pipe det → Max. damping (6040)

Description Use this function to enter a damping value to enable detection of empty or partially filled
measuring tubes.

User entry Positive floating-point number

Factory setting 0

Endress+Hauser 89
Description of device parameters Proline Promass 500 Modbus RS485

Additional information Description


If oscillation damping exceeds the specified value, the measuring device presumes that the
pipe is partially filled and the flow signal is set to 0. The measuring device displays the
diagnostic message S862 Partly filled. In the case of non-homogeneous media or air
pockets, the damping of the measuring tubes increases.

User entry
• Damping is disabled if 0 is entered (factory setting).
• Damping is enabled if the value entered is greater than 0.
• The value entered depends on application-specific influence variables, such as the
medium, nominal diameter, sensor etc.

Example
• If the pipe is filled normally the value of the oscillation damping is 500.
• If the pipe is partially filled the value of the oscillation damping is > 5000.
• A practical damping value would then be 2000: enter 2000 as the value.

3.2.4 "Measurement mode" submenu

Navigation  Expert → Sensor → Measurement mode

‣ Measurement mode

Select medium (6062) →  90

Select gas type (6074) →  91

Sound velocity (6147) →  91

Temp. coeff. SV (6181) →  92

Gas Frac Handler (6377) →  92

Select medium 

Navigation  Expert → Sensor → Measurement mode → Select medium (6062)

Description Use this function to select the type of medium.

Selection • Liquid
• Gas

Factory setting Liquid

90 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters

Select gas type 

Navigation  Expert → Sensor → Measurement mode → Select gas type (6074)

Prerequisite The Gas option is selected in the Select medium parameter (→  90).

Description Use this function to select the type of gas for the measuring application.

Selection • Air
• Ammonia NH3
• Argon Ar
• Sulf. hex.fl.SF6
• Oxygen O2
• Ozone O3
• Nitrog. ox. NOx
• Nitrogen N2
• Nitrous ox. N2O
• Methane CH4
• Hydrogen H2
• Helium He
• Hydrog.chlor.HCl
• Hydrog.sulf. H2S
• Ethylene C2H4
• Carbon diox. CO2
• Carbon monox. CO
• Chlorine Cl2
• Butane C4H10
• Propane C3H8
• Propylene C3H6
• Ethane C2H6
• Others

Factory setting Methane CH4

Additional information Description


The gas type needs to be selected so that it is possible to comply with accuracy
specifications in gas applications.

Sound velocity 

Navigation  Expert → Sensor → Measurement mode → Sound velocity (6147)

Prerequisite In the Select gas type parameter (→  91), the Others option is selected.

Description Use this function to enter the sound velocity of the gas at 0 °C (+32 °F).

User entry 1 to 99 999.9999 m/s

Factory setting 415.0 m/s

Endress+Hauser 91
Description of device parameters Proline Promass 500 Modbus RS485

Temp. coeff. SV 

Navigation  Expert → Sensor → Measurement mode → Temp. coeff. SV (6181)

Prerequisite The Others option is selected in the Select gas type parameter (→  91).

Description Use this function to enter a temperature coefficient for the sound velocity of the gas.

User entry Positive floating-point number

Factory setting 0 (m/s)/K

Gas Frac Handler 

Navigation  Expert → Sensor → Measurement mode → Gas Frac Handler (6377)

Description Activates the Gas Fraction Handler function for two phase media.

Selection • Off
• Moderate
• Powerful

Factory setting Off

3.2.5 "External comp." submenu

Navigation  Expert → Sensor → External comp.

‣ External comp.

Pressure compen. (6130) →  93

Pressure value (6059) →  93

External press. (6209) →  93

Temp.corr.source (6184) →  94

External temp. (6080) →  94

Spec. source 0 (6401)

Spec. source 1 (6402)

92 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters

Pressure compen. 

Navigation  Expert → Sensor → External comp. → Pressure compen. (6130)

Description Use this function to select the type of pressure compensation.

Selection • Off
• Fixed value
• External value
• Current input 1 *
• Current input 2 *
• Current input 3 *

Factory setting Off

Additional information Selection


• Fixed value
A fixed pressure value is used for compensation: Pressure value parameter (→  93)
• Current input 1 option, Current input 2 option , Current input 3 option
The pressure value read in via the current input is used for compensation.

Pressure value 

Navigation  Expert → Sensor → External comp. → Pressure value (6059)

Prerequisite The Fixed value option or the Current input 1…n option is selected in the Pressure
compen. parameter (→  93).

Description Use this function to enter a value for the process pressure that is used for pressure
correction.

User entry Positive floating-point number

Factory setting 1.01325 bar

Additional information User entry


The unit is taken from the Pressure unit parameter (→  73)

External press.

Navigation  Expert → Sensor → External comp. → External press. (6209)

Prerequisite The Fixed value option or the Current input 1…n option is selected in the Pressure
compen. parameter (→  93).

* Visibility depends on order options or device settings

Endress+Hauser 93
Description of device parameters Proline Promass 500 Modbus RS485

Description Use this function to enter an external pressure value.

User entry Positive floating-point number

Factory setting 1.01325 bar

Additional information User entry


The unit is taken from the Pressure unit parameter (→  73)

Temp.corr.source 

Navigation  Expert → Sensor → External comp. → Temp.corr.source (6184)

Description Use this function to select the temperature mode.

Selection • Internal value


• External value
• Current input 1 *
• Current input 2 *
• Current input 3 *

Factory setting Internal value

Additional information Description


Use this function to select the type of temperature compensation.

Selection
All the options available for selection are used for measured value compensation.
• Internal value
The temperature value measured internally (temperature sensor of the measuring
sensor) is used for compensation.
• Current input 1 optionVisibility depends on order options or device settings, Current
input 2 optionVisibility depends on order options or device settings ,Current input 3
optionVisibility depends on order options or device settings
The temperature value read in via the current input is used for compensation.

External temp.

Navigation  Expert → Sensor → External comp. → External temp. (6080)

Prerequisite The External value option or the Current input 1…n option is selected in the
Temperature mode parameter (→  94).

Description Use this function to enter the external temperature.

* Visibility depends on order options or device settings

94 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters

User entry –273.15 to 99 999 °C

Factory setting Country-specific:


• 0 °C
• +32 °F

Additional information Description


The unit is taken from the Temperature unit parameter (→  72)

Temperature mode 

Navigation  Expert → Sensor → External comp. → Temperature mode (6341)

Description Select temperature mode for temperature compensation.

Selection • Internal value


• External value

Factory setting Internal value

3.2.6 "Calculated value" submenu

Navigation  Expert → Sensor → Calculated value

‣ Calculated value

‣ Corr. vol.flow. →  95

"Corr. vol.flow." submenu

Navigation  Expert → Sensor → Calculated value → Corr. vol.flow.

‣ Corr. vol.flow.

Corr. vol.flow. →  96

Ext. ref.density →  96

Fix ref.density →  97

Ref. temperature →  97

Endress+Hauser 95
Description of device parameters Proline Promass 500 Modbus RS485

Linear exp coeff →  98

Square exp coeff →  98

Corr. vol.flow. 

Navigation  Expert → Sensor → Calculated value → Corr. vol.flow. → Corr. vol.flow. (1812)

Description Use this function to select the reference density for calculating the corrected volume flow.

Selection • Fix ref.density


• Calc ref density
• Current input 1 *
• Current input 2 *
• Current input 3 *

Factory setting Calc ref density

Additional information Selection


The Ref. dens API 53 option is suitable only for applications involving LPG 5), where the
flow rate is measured on the basis of the corrected volume flow.
Selecting this option means that the reference density is used, taking into account the
values in table 53 E of API MPMS section 11.2. Temperature measurement (measured
internally or read into the device from an external source→  92→  92) and density
measurement take place during operation while the medium is flowing. The mass flow is
divided by the reference density to give the corrected volume flow and is issued as an
output signal.

Ext. ref.density

Navigation  Expert → Sensor → Calculated value → Corr. vol.flow. → Ext. ref.density (6198)

Prerequisite In the Corr. vol.flow. parameter (→  96), the Ext. ref.density option is selected.

Description Displays the reference density which is read in externally, e.g. via the current input.

User interface Floating point number with sign

Additional information Dependency


The unit is taken from the Ref. dens. unit parameter (→  72)

* Visibility depends on order options or device settings


5) liquefied petroleum gas

96 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters

Fix ref.density 

Navigation  Expert → Sensor → Calculated value → Corr. vol.flow. → Fix ref.density (1814)

Prerequisite The Fix ref.density option is selected in the Corr. vol.flow. parameter (→  96)
parameter.

Description Use this function to enter a fixed value for the reference density.

User entry Positive floating-point number

Factory setting 1 kg/Nl

Additional information Dependency


The unit is taken from the Ref. dens. unit parameter (→  72)

Ref. temperature 

Navigation  Expert → Sensor → Calculated value → Corr. vol.flow. → Ref. temperature (1816)

Prerequisite The Calc ref density option is selected in the Corr. vol.flow. parameter (→  96)
parameter.

Description Use this function to enter a reference temperature for calculating the reference density.

User entry –273.15 to 99 999 °C

Factory setting Country-specific:


• +20 °C
• +68 °F

Additional information Dependency


The unit is taken from the Temperature unit parameter (→  72)

Reference density calculation

ρn = ρ ⋅ (1 + α ⋅ ∆t + β ⋅ ∆t2)
A0023403

• ρN: reference density


• ρ: fluid density currently measured
• t: fluid temperature currently measured
• tN: reference temperature at which the reference density is calculated (e.g. 20 °C)
• ∆t: t - tN
• α: linear expansion coefficient of the fluid, unit = [1/K]; K = Kelvin
• β: square expansion coefficient of the fluid, unit = [1/K2]

Endress+Hauser 97
Description of device parameters Proline Promass 500 Modbus RS485

Linear exp coeff 

Navigation  Expert → Sensor → Calculated value → Corr. vol.flow. → Linear exp coeff (1817)

Prerequisite The Calc ref density option is selected in the Corr. vol.flow. parameter (→  96)
parameter.

Description Use this function to enter a linear, fluid-specific expansion coefficient for calculating the
reference density.

User entry Signed floating-point number

Factory setting 0.0 1/K

Square exp coeff 

Navigation  Expert → Sensor → Calculated value → Corr. vol.flow. → Square exp coeff (1818)

Prerequisite The Calc ref density option is selected in the Corr. vol.flow. parameter (→  96)
parameter.

Description For fluid with a non-linear expansion pattern: use this function to enter a quadratic, fluid-
specific expansion coefficient for calculating the reference density.

User entry Signed floating-point number

Factory setting 0.0 1/K²

3.2.7 "Sensor adjustment" submenu

Navigation  Expert → Sensor → Sensor adjustm.

‣ Sensor adjustm.

Install. direct. (1809) →  99

Inst. angle roll (6282) →  99

Inst.angle pitch (6236) →  99

‣ Zero point adj. →  100

‣ Variable adjust →  101

98 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters

Install. direct. 

Navigation  Expert → Sensor → Sensor adjustm. → Install. direct. (1809)

Description Use this function to change the sign of the medium flow direction.

Selection • In arrow direct.


• Against arrow

Factory setting In arrow direct.

Additional information Description


Before changing the sign: ascertain the actual direction of fluid flow with reference to
the direction indicated by the arrow on the sensor nameplate.

Inst. angle roll 

Navigation  Expert → Sensor → Sensor adjustm. → Inst. angle roll (6282)

Prerequisite Available only with Promass Q.

Description Use this function to enter the roll installation angle in degrees.

User entry –180 to 180 °

Factory setting 0°

Additional information FIGURE: ROLL ANGLE


Roll angle
• The roll angle is the angle β from the vertical V to align the central axis Z of the
transmitter.
• The roll angle can be between –180 to +180 °.

Inst.angle pitch 

Navigation  Expert → Sensor → Sensor adjustm. → Inst.angle pitch (6236)

Prerequisite Available only with Promass Q.

Description Use this function to enter the installation angle pitch in degrees.

User entry –180 to 180 °

Factory setting 0°

Additional information FIGURE: PITCH ANGLE

Endress+Hauser 99
Description of device parameters Proline Promass 500 Modbus RS485

Pitch angle
• The pitch angle is the angle α from the horizontal H to align the central axis Z of the
measuring device.
• The pitch angle can be between –90 to +90 °.

"Zero point adj." submenu


• It is generally not necessary to perform zero point adjustment.
• However, this function may be needed in some applications with low flow and strict
accuracy requirements.
• A zero point adjustment cannot increase repeatability.
• The following conditions should be met to perform a zero point adjustment
successfully without the adjustment finishing in an error:
• The real flow must be 0.
• The pressure must be at least 15 psi g.
• The adjustment takes a maximum of 60 s. The more stable the conditions, the
faster the adjustment is completed.
• This function can also be used to check the health of the measuring device.
A healthy measuring device has a maximum zero point deviation of ±100 compared
to the factory setting of the measuring device (calibration report).

Navigation  Expert → Sensor → Sensor adjustm. → Zero point adj.

‣ Zero point adj.

Zero point adj. (6196) →  100

Zero pt adj.stat (6253) →  101

Progress (2808) →  101

Zero point adj. 

Navigation  Expert → Sensor → Sensor adjustm. → Zero point adj. → Zero point adj. (6196)
 Expert → Sensor → Sensor adjustm. → Zero point adj. → Zero point adj. (6196)

Description Use this function to select the start of the zero point adjustment.
Observe conditions →  100.

Selection • Cancel
• Start

Factory setting Cancel

100 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters

Additional information Description


• Cancel
If zero point adjustment has failed, select this option to cancel zero point adjustment.
• Busy
Is displayed during zero point adjustment.
• Zero adjust fail
Is displayed if zero point adjustment has failed.
• Start
Select this option to start zero point adjustment.

Zero pt adj.stat

Navigation  Expert → Sensor → Sensor adjustm. → Zero point adj. → Zero pt adj.stat (6253)
 Expert → Sensor → Sensor adjustm. → Zero point adj. → Zero pt adj.stat (6253)

Description Shows the status of the zero point adjustment.

User interface • Busy


• Zero adjust fail
• Ok

Factory setting Ok

Progress

Navigation  Expert → Sensor → Sensor adjustm. → Zero point adj. → Progress (2808)

Description The progress of the process is indicated.

User interface 0 to 100 %

"Variable adjust" submenu

Navigation  Expert → Sensor → Sensor adjustm. → Variable adjust

‣ Variable adjust

Mass flow offset (1831) →  102

Mass flow factor (1832) →  102

Vol. flow offset (1841) →  103

Vol. flow factor (1846) →  103

Endress+Hauser 101
Description of device parameters Proline Promass 500 Modbus RS485

Density offset (1848) →  103

Density factor (1849) →  104

Corr. vol offset (1866) →  104

Corr. vol factor (1867) →  104

Ref.dens. offset (1868) →  105

Ref.dens. factor (1869) →  105

Temp. offset (1870) →  105

Temp. factor (1871) →  106

Mass flow offset 

Navigation  Expert → Sensor → Sensor adjustm. → Variable adjust → Mass flow offset (1831)

Description Use this function to enter the zero point shift for the mass flow trim. The mass flow unit
on which the shift is based is kg/s.

User entry Signed floating-point number

Factory setting 0 kg/s

Additional information Description


Corrected value = (factor × value) + offset

Mass flow factor 

Navigation  Expert → Sensor → Sensor adjustm. → Variable adjust → Mass flow factor (1832)

Description Use this function to enter a quantity factor (without time) for the mass flow. This
multiplication factor is applied over the mass flow range.

User entry Positive floating-point number

Factory setting 1

Additional information Description


Corrected value = (factor × value) + offset

102 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters

Vol. flow offset 

Navigation  Expert → Sensor → Sensor adjustm. → Variable adjust → Vol. flow offset (1841)

Description Use this function to enter the zero point shift for the volume flow trim. The volume flow
unit on which the shift is based is m³/s.

User entry Signed floating-point number

Factory setting 0 m³/s

Additional information Description


Corrected value = (factor × value) + offset

Vol. flow factor 

Navigation  Expert → Sensor → Sensor adjustm. → Variable adjust → Vol. flow factor (1846)

Description Use this function to enter a quantity factor (without time) for the volume flow. This
multiplication factor is applied over the volume flow range.

User entry Positive floating-point number

Factory setting 1

Additional information Description


Corrected value = (factor × value) + offset

Density offset 

Navigation  Expert → Sensor → Sensor adjustm. → Variable adjust → Density offset (1848)

Description Use this function to enter the zero point shift for the density trim. The density unit on
which the shift is based is kg/m³.

User entry Signed floating-point number

Factory setting 0 kg/m³

Additional information Description


Corrected value = (factor × value) + offset

Endress+Hauser 103
Description of device parameters Proline Promass 500 Modbus RS485

Density factor 

Navigation  Expert → Sensor → Sensor adjustm. → Variable adjust → Density factor (1849)

Description Use this function to enter a quantity factor for the density. This multiplication factor is
applied over the density range.

User entry Positive floating-point number

Factory setting 1

Additional information Description


Corrected value = (factor × value) + offset

Corr. vol offset 

Navigation  Expert → Sensor → Sensor adjustm. → Variable adjust → Corr. vol offset (1866)

Description Use this function to enter the zero point shift for the corrected volume flow trim. The
corrected volume flow unit on which the shift is based is 1 Nm³/s.

User entry Signed floating-point number

Factory setting 0 Nm³/s

Additional information Description


Corrected value = (factor × value) + offset

Corr. vol factor 

Navigation  Expert → Sensor → Sensor adjustm. → Variable adjust → Corr. vol factor (1867)

Description Use this function to enter a quantity factor (without time) for the corrected volume flow.
This multiplication factor is applied over the corrected volume flow range.

User entry Positive floating-point number

Factory setting 1

Additional information Description


Corrected value = (factor × value) + offset

104 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters

Ref.dens. offset 

Navigation  Expert → Sensor → Sensor adjustm. → Variable adjust → Ref.dens. offset (1868)

Description Use this parameter to enter the zero point shift for the reference density trim. The
reference density unit on which the shift is based is 1 kg/Nm³.

User entry Signed floating-point number

Factory setting 0 kg/Nm³

Additional information Description


Corrected value = (factor × value) + offset

Ref.dens. factor 

Navigation  Expert → Sensor → Sensor adjustm. → Variable adjust → Ref.dens. factor (1869)

Description Use this function to enter a quantity factor (without time) for the reference density. This
multiplication factor is applied over the reference density range.

User entry Positive floating-point number

Factory setting 1

Additional information Description


Corrected value = (factor × value) + offset

Temp. offset 

Navigation  Expert → Sensor → Sensor adjustm. → Variable adjust → Temp. offset (1870)

Description Use this function to enter the zero point shift for the temperature trim. The temperature
unit on which the shift is based is K.

User entry Signed floating-point number

Factory setting 0K

Additional information Description


Corrected value = (factor × value) + offset

Endress+Hauser 105
Description of device parameters Proline Promass 500 Modbus RS485

Temp. factor 

Navigation  Expert → Sensor → Sensor adjustm. → Variable adjust → Temp. factor (1871)

Description Use this function to enter a quantity factor for the temperature. In each case, this factor
refers to the temperature in K.

User entry Positive floating-point number

Factory setting 1

Additional information Description


Corrected value = (factor × value) + offset

3.2.8 "Calibration" submenu

Navigation  Expert → Sensor → Calibration

‣ Calibration

Cal. factor (6025) →  106

Zero point (6195) →  107

Nominal diameter (2807) →  107

C0 to 5 (6022) →  107

Cal. factor

Navigation  Expert → Sensor → Calibration → Cal. factor (6025)

Description Displays the current calibration factor for the sensor.

User interface Signed floating-point number

Factory setting Depends on nominal diameter and calibration.

106 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters

Zero point 

Navigation  Expert → Sensor → Calibration → Zero point (6195)

Description Use this function to enter the zero point correction value for the sensor.

User entry Signed floating-point number

Factory setting Depends on nominal diameter and calibration.

Nominal diameter

Navigation  Expert → Sensor → Calibration → Nominal diameter (2807)

Description Displays the nominal diameter of the sensor.

User interface DNxx / x"

Factory setting Depends on the size of the sensor

Additional information Description


The value is also specified on the sensor nameplate.

C0 to 5

Navigation  Expert → Sensor → Calibration → C0 to 5 (6022)

Description Displays the current density coefficients C0 to 5 of the sensor.

User interface Signed floating-point number

Factory setting 0

3.3 "I/O configuration" submenu


Navigation  Expert → I/O config.

‣ I/O config.

I/O 1 to n terminals (3902–1 to n) →  108

I/O 1 to n info (3906–1 to n) →  108

Endress+Hauser 107
Description of device parameters Proline Promass 500 Modbus RS485

I/O 1 to n type (3901–1 to n) →  109

Apply I/O config (3907) →  109

I/O alterat.code (2762) →  109

I/O 1 to n terminals

Navigation  Expert → I/O config. → I/O 1 to n terminals (3902–1 to n)

Description Displays the terminal numbers used by the I/O module.

User interface • Not used


• 26-27 (I/O 1)
• 24-25 (I/O 2)
• 22-23 (I/O 3)
• 20-21 (I/O 4) *

I/O 1 to n info

Navigation  Expert → I/O config. → I/O 1 to n info (3906–1 to n)

Description Displays information about the plugged in I/O module.

User interface • Not plugged


• Invalid
• Not configurable
• Configurable
• MODBUS

Additional information "Not plugged" option


The I/O module is not plugged in.

"Invalid" option
The I/O module is not plugged correctly.

"Not configurable" option


The I/O module is not configurable.

"Configurable" option
The I/O module is configurable.

"MODBUS" option
The I/O module is configured for Modbus.

* Visibility depends on order options or device settings

108 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters

I/O 1 to n type 

Navigation  Expert → I/O config. → I/O 1 to n type (3901–1 to n)

Prerequisite For the following order code:


• "Output; input 2", option D "Configurable I/O initial setting off"
• "Output; input 3", option D "Configurable I/O initial setting off"
• "Output; input 4", option D "Configurable I/O initial setting off"

Description Use this function to select the I/O module type for the configuration of the I/O module.

Selection • Off
• Curr.output *
• Current input *
• Status input *
• PFS output *
• Double pulse out *
• Relay output *

Factory setting Off

Apply I/O config 

Navigation  Expert → I/O config. → Apply I/O config (3907)

Description Use this function to activate the newly configured I/O module type.

Selection • No
• Yes

Factory setting No

I/O alterat.code 

Navigation  Expert → I/O config. → I/O alterat.code (2762)

Description Use this function to enter the ordered activation code to activate the I/O configuration
change.

User entry Positive integer

Factory setting 0

* Visibility depends on order options or device settings

Endress+Hauser 109
Description of device parameters Proline Promass 500 Modbus RS485

Additional information Description


The I/O configuration is changed in the I/O type parameter (→  109).

3.4 "Input" submenu


Navigation  Expert → Input

‣ Input

‣ Current input 1 to n →  110

‣ Status input 1 to n →  113

3.4.1 "Current input 1 to n" submenu

Navigation  Expert → Input → Current input 1 to n

‣ Current input 1 to n

Terminal no. (1611–1 to n) →  110

Signal mode (1610–1 to n) →  111

Current span (1605–1 to n) →  111

0/4 mA value (1606–1 to n) →  112

20 mA value (1607–1 to n) →  112

Failure mode (1601–1 to n) →  112

Failure value (1602–1 to n) →  113

Terminal no.

Navigation  Expert → Input → Current input 1 to n → Terminal no. (1611–1 to n)

Description Displays the terminal numbers used by the current input module.

110 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters

User interface • Not used


• 24-25 (I/O 2)
• 22-23 (I/O 3)
• 20-21 (I/O 4) *

Additional information "Not used" option


The current input module does not use any terminal numbers.

Signal mode 

Navigation  Expert → Input → Current input 1 to n → Signal mode (1610–1 to n)

Prerequisite The measuring device is not approved for use in the hazardous area with type of
protection Ex-i.

Description Use this function to select the signal mode for the current input.

Selection • Passive
• Active *

Factory setting Active

Current span 

Navigation  Expert → Input → Current input 1 to n → Current span (1605–1 to n)

Description Use this function to select the current range for the process value output and the upper
and lower level for signal on alarm.

Selection • 4...20 mA
• 4...20 mA NAMUR
• 4...20 mA US
• 0...20 mA

Factory setting Country-specific:


• 4...20 mA NAMUR
• 4...20 mA US

Additional information Examples


Sample values for the current range: Current span parameter (→  118)

* Visibility depends on order options or device settings

Endress+Hauser 111
Description of device parameters Proline Promass 500 Modbus RS485

0/4 mA value 

Navigation  Expert → Input → Current input 1 to n → 0/4 mA value (1606–1 to n)

Description Use this function to enter a value for the 4 mA current.

User entry Signed floating-point number

Factory setting 0

Additional information Current input behavior


The current input behaves differently depending on the settings configured in the
following parameters:
• Current span (→  111)
• Failure mode (→  112)

Configuration examples
Pay attention to the configuration examples for 4 mA value parameter (→  120).

20 mA value 

Navigation  Expert → Input → Current input 1 to n → 20 mA value (1607–1 to n)

Description Use this function to enter a value for the 20 mA current.

User entry Signed floating-point number

Factory setting Depends on country and nominal diameter

Additional information Configuration examples


Pay attention to the configuration examples for 4 mA value parameter (→  120).

Failure mode 

Navigation  Expert → Input → Current input 1 to n → Failure mode (1601–1 to n)

Description Use this function to select the input behavior when measuring a current outside the
configured Current span parameter (→  111).

Selection • Alarm
• Last valid value
• Defined value

Factory setting Alarm

112 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters

Additional information Options


• Alarm
An error message is set.
• Last valid value
The last valid measured value is used.
• Defined value
A user-defined measured value is used (Failure value parameter (→  113)).

Failure value 

Navigation  Expert → Input → Current input 1 to n → Failure value (1602–1 to n)

Prerequisite In the Failure mode parameter (→  112), the Defined value option is selected.

Description Use this function to enter the value that the device uses if it does not receive an input
signal from the external device, or if the input signal is invalid.

User entry Signed floating-point number

Factory setting 0

3.4.2 "Status input 1 to n" submenu

Navigation  Expert → Input → Status input 1 to n

‣ Status input 1 to n

Terminal no. (1358–1 to n) →  113

Assign stat.inp. (1352–1 to n) →  114

Val.stat.inp. (1353–1 to n) →  115

Active level (1351–1 to n) →  115

Response time (1354–1 to n) →  115

Terminal no.

Navigation  Expert → Input → Status input 1 to n → Terminal no. (1358–1 to n)

Description Displays the terminal numbers used by the status input module.

Endress+Hauser 113
Description of device parameters Proline Promass 500 Modbus RS485

User interface • Not used


• 24-25 (I/O 2)
• 22-23 (I/O 3)
• 20-21 (I/O 4) *

Additional information "Not used" option


The status input module does not use any terminal numbers.

Assign stat.inp. 

Navigation  Expert → Input → Status input 1 to n → Assign stat.inp. (1352–1 to n)

Description Use this function to select the function for the status input.

Selection • Off
• Reset totaliz. 1
• Reset totaliz. 2
• Reset totaliz. 3
• Reset all tot.
• Flow override
• Zero point adj.

Factory setting Off

Additional information Custody transfer mode


Only available for Promass F, O, Q and X.

NOTE!
Before enabling the measuring device for custody transfer mode, make sure that the Off
option is selected in the Assign stat.inp..
For detailed information on custody transfer mode, see the Special Documentation for
the device →  8

Additional information Selection


• Off
The status input is switched off.
• Reset totaliz. 1…3
The individual totalizers are reset.
• Reset all tot.
All totalizers are reset.
• Flow override
The Flow override (→  84) is activated.
Note on the Flow override (→  84):
• The Flow override (→  84) is enabled as long as the level is at the status input
(continuous signal).
• All other assignments react to a change in level (pulse) at the status input.

* Visibility depends on order options or device settings

114 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters

Val.stat.inp.

Navigation  Expert → Input → Status input 1 to n → Val.stat.inp. (1353–1 to n)

Description Displays the current input signal level.

User interface • High


• Low

Active level 

Navigation  Expert → Input → Status input 1 to n → Active level (1351–1 to n)

Description Use this function to determine the input signal level at which the assigned function is
activated.

Selection • High
• Low

Factory setting High

Response time 

Navigation  Expert → Input → Status input 1 to n → Response time (1354–1 to n)

Description Use this function to enter the minimum time period for which the input signal level must
be present before the selected function is activated.

User entry 5 to 200 ms

Factory setting 50 ms

3.5 "Output" submenu


Navigation  Expert → Output

‣ Output

‣ Curr.output 1 to n →  116

‣ PFS output 1 to n →  130

Endress+Hauser 115
Description of device parameters Proline Promass 500 Modbus RS485

‣ Relay output 1 to n →  152

‣ Double pulse out →  159

3.5.1 "Current output 1 to n" submenu

Navigation  Expert → Output → Curr.output 1 to n

‣ Curr.output 1 to n

Terminal no. (0379–1 to n) →  116

Signal mode (0377–1 to n) →  117

Assign curr. 1 to n (0359–1 to n) →  117

Current span (0353–1 to n) →  118

Fixed current (0365–1 to n) →  119

0/4 mA value (0367–1 to n) →  120

20 mA value (0372–1 to n) →  121

Measuring mode (0351–1 to n) →  122

Damping out. 1 to n (0363–1 to n) →  127

Failure mode (0364–1 to n) →  128

Failure current (0352–1 to n) →  129

Output curr. 1 to n (0361–1 to n) →  130

Measur. curr. 1 to n (0366–1 to n) →  130

Terminal no.

Navigation  Expert → Output → Curr.output 1 to n → Terminal no. (0379–1 to n)

Description Displays the terminal numbers used by the current output module.

116 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters

User interface • Not used


• 24-25 (I/O 2)
• 22-23 (I/O 3)
• 20-21 (I/O 4) *

Additional information "Not used" option


The current output module does not use any terminal numbers.

Signal mode 

Navigation  Expert → Output → Curr.output 1 to n → Signal mode (0377–1 to n)

Description Use this function to select the signal mode for the current output.

Selection • Active *
• Passive *

Factory setting Active

Assign curr. 1 to n 

Navigation  Expert → Output → Curr.output 1 to n → Assign curr. 1 to n (0359–1 to n)

Description Use this function to select a process variable for the current output.
Detailed description of the options Oscil. frequency, Oscil. amplitude, Oscil.
damping and Signal asymmetry: Value 1 display parameter (→  19)

Selection • Off *
• Mass flow
• Volume flow
• Correct.vol.flow *
• Target mass flow *
• Carrier mass fl. *
• Target vol. flow *
• Carrier vol. fl. *
• Targ.corr.vol.fl *
• Carr.corr.vol.fl *
• Density
• Ref.density *
• Ref.dens.altern. *
• GSV flow *
• GSVa *
• NSV flow *
• NSVa *
• S&W volume flow *
• Water cut *
• Oil density *
• Water density *

* Visibility depends on order options or device settings

Endress+Hauser 117
Description of device parameters Proline Promass 500 Modbus RS485

• Oil mass flow *


• Water mass flow *
• Oil volume flow *
• Water vol. flow *
• Oil corr.vol.fl. *
• Water corr.v.fl. *
• Concentration *
• Dynam. viscosity *
• Kinematic visc. *
• TempCompDynVisc *
• TempCompKinVisc *
• Temperature
• Carr. pipe temp. *
• Electronic temp.
• Osc. freq. 0
• Osc. freq. 1 *
• Osc. ampl. 0 *
• Osc. ampl. 1 *
• Freq. fluct. 0 *
• Freq. fluct. 1 *
• Osc. damping 0 *
• Osc. damping 1 *
• Osc.damp.fluct 0 *
• Osc.damp.fluct 1 *
• Signal asymmetry *
• Exc. current 0 *
• Exc. current 1 *
• HBSI *
• Pressure
• Spec. output 0 *
• Spec. output 1 *
• Index inh.medium
• Index sus.bubble *

Factory setting Mass flow

Current span 

Navigation  Expert → Output → Curr.output 1 to n → Current span (0353–1 to n)

Description Use this function to select the current range for the process value output and the upper
and lower level for signal on alarm.

Selection • 4...20 mA NAMUR


• 4...20 mA US
• 4...20 mA
• 0...20 mA
• Fixed current

Factory setting Country-specific:


• 4...20 mA NAMUR (3.8...20.5 mA)
• 4...20 mA US (3.9...20.8 mA)

* Visibility depends on order options or device settings

118 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters

Additional information Description


• In the event of a device alarm, the current output adopts the value specified in the
Failure mode parameter (→  128).
• If the measured value is outside the measuring range, the diagnostic message
S441 Curr.output 1 to n is displayed.
• The measuring range is specified via the 0/4 mA value parameter (→  120) and
20 mA value parameter (→  121).

"Fixed current" option


The current value is set via the Fixed current parameter (→  119).

Example
Shows the relationship between the current span for the output of the process variable and
the lower and upper alarm levels:

2 1 3

I [mA]
A0034351

1 Current span for process value


2 Lower level for signal on alarm
3 Upper level for signal on alarm

Selection

Selection 1 2 3

4...20 mA NAMUR 3.8 to 20.5 mA < 3.6 mA > 21.95 mA

4...20 mA US 3.9 to 20.8 mA US < 3.6 mA > 21.95 mA

4...20 mA 4 to 20.5 mA < 3.6 mA > 21.95 mA

0...20 mA 0 to 20.5 mA < 0 mA > 21.95 mA

If the flow exceeds or falls below the upper or lower signal on alarm level, the
diagnostic message S441 Curr.output 1 to n is displayed.

Fixed current 

Navigation  Expert → Output → Curr.output 1 to n → Fixed current (0365–1 to n)

Prerequisite The Fixed current option is selected in the Current span parameter (→  118).

Description Use this function to enter a constant current value for the current output.

User entry 0 to 22.5 mA

Factory setting 22.5 mA

Endress+Hauser 119
Description of device parameters Proline Promass 500 Modbus RS485

0/4 mA value 

Navigation  Expert → Output → Curr.output 1 to n → 0/4 mA value (0367–1 to n)

Prerequisite In the Current span parameter (→  118), one of the following options is selected:
• 4...20 mA NAMUR
• 4...20 mA US
• 4...20 mA
• 0...20 mA

Description Use this function to enter a value for the 0/4 mA current.

User entry Signed floating-point number

Factory setting Country-specific:


• 0 kg/h
• 0 lb/min

Additional information Description


Positive and negative values are permitted depending on the process variable assigned in
the Assign curr. parameter (→  117). In addition, the value can be greater than or
smaller than the value assigned for the 20 mA current in the 20 mA value parameter
(→  121).

Dependency
The unit depends on the process variable selected in the Assign curr. parameter
(→  117).

Current output behavior


The current output behaves differently depending on the settings configured in the
following parameters:
• Current span (→  118)
• Failure mode (→  128)

Configuration examples
Some examples of parameter settings and their effect on the current output are given in
the following section.
Configuration example A
Measuring mode with Forward flow option
• 0/4 mA value parameter (→  120) = not equal to zero flow (e.g. –250 m3/h)
• 20 mA value parameter (→  121) = not equal to zero flow (e.g. +750 m3/h)
• Calculated current value = 8 mA at zero flow

I [mA]
20
8
4
Q
-250 0 +750
1 1
A0013757

Q Flow
I Current
1 Measuring range is exceeded or undershot

120 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters

The operational range of the measuring device is defined by the values entered for the 0/4
mA value parameter (→  120) and 20 mA value parameter (→  121). If the
effective flow exceeds or falls below this operational range, the diagnostic message
S441 Curr.output 1 to n is displayed.
Configuration example B
Measuring mode with Forward/Reverse option

I [mA]

20

4
0 Q
0
3 1 2
A0013758

I Current
Q Flow
1 Value assigned to the 0/4 mA current
2 Forward flow
3 Reverse flow

The current output signal is independent of the direction of flow (absolute amount of the
measured variable). The values for the 0/4 mA value parameter (→  120) and 20 mA
value parameter (→  121) must have the same sign. The value for the 20 mA value
parameter (→  121) (e.g. reverse flow) corresponds to the mirrored value for the 20
mA value parameter (→  121) (e.g. forward flow).
Configuration example C
Measuring mode with Rev. flow comp. option
If flow is characterized by severe fluctuations (e.g. when using reciprocating pumps), flow
components outside the measuring range are buffered, balanced and output after a
maximum delay of 60 s →  122.

20 mA value 

Navigation  Expert → Output → Curr.output 1 to n → 20 mA value (0372–1 to n)

Prerequisite One of the following options is selected in the Current span parameter (→  118):
• 4...20 mA NAMUR
• 4...20 mA US
• 4...20 mA
• 0...20 mA

Description Use this function to enter a value for the 20 mA current.

User entry Signed floating-point number

Factory setting Depends on country and nominal diameter→  248

Additional information Description


Positive and negative values are permitted depending on the process variable assigned in
the Assign curr. parameter (→  117). In addition, the value can be greater than or

Endress+Hauser 121
Description of device parameters Proline Promass 500 Modbus RS485

smaller than the value assigned for the 0/4 mA current in the 0/4 mA value parameter
(→  120).

Dependency
The unit depends on the process variable selected in the Assign curr. parameter
(→  117).

Example
• Value assigned to 0/4 mA = –250 m3/h
• Value assigned to 20 mA = +750 m3/h
• Calculated current value = 8 mA (at zero flow)
If the Forward/Reverse option is selected in the Measuring mode parameter
(→  122), different signs cannot be entered for the values of the 0/4 mA value
parameter (→  120) and 20 mA value parameter (→  121). The diagnostic message
S441 Curr.output 1 to n is displayed.

Configuration examples
Observe the configuration examples for the 0/4 mA value parameter (→  120).

Measuring mode 

Navigation  Expert → Output → Curr.output 1 to n → Measuring mode (0351–1 to n)

Prerequisite In the Assign curr. parameter (→  117), one of the following options is selected:
• Mass flow
• Volume flow
• Correct.vol.flow
• Target mass flow *
• Carrier mass fl. *
• Density
• Ref.density
• Concentration *
• Dynam. viscosity *
• Kinematic visc. *
• TempCompDynVisc *
• TempCompKinVisc *
• Temperature
• Carr. pipe temp. *
• Electronic temp.
• Osc. freq. 0
• Osc. freq. 1 *
• Osc. ampl. 0 *
• Osc. ampl. 1 *
• Freq. fluct. 0
• Freq. fluct. 1 *
• Osc. damping 0
• Osc. damping 1 *
• Osc.damp.fluct 0
• Osc.damp.fluct 1 *
• Signal asymmetry

* Visibility depends on order options or device settings

122 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters

• Exc. current 0
• Exc. current 1 *
• HBSI *
Detailed description of the options Oscil. frequency, Oscil. amplitude, Oscil.
damping and Signal asymmetry: Value 1 display parameter (→  19)
In the Current span parameter (→  118), one of the following options is selected:
• 4...20 mA NAMUR
• 4...20 mA US
• 4...20 mA
• 0...20 mA

Description Use this function to select the measuring mode for the current output.

Selection • Forward flow


• Forward/Reverse *
• Rev. flow comp.

Factory setting Forward flow

Additional information Description


The process variable that is assigned to the current output via the Assign curr.
parameter (→  117) is displayed below the parameter.

"Forward flow" option


The current output signal is proportional to the process variable assigned. The measuring
range is defined by the values that are assigned to the 0/4 mA and 20 mA current value.
The flow components outside the scaled measuring range are taken into account for signal
output as follows:
• Both values are defined such that they are not equal to zero flow e.g.:
• 0/4 mA current value = –5 m3/h
• 20 mA current value = 10 m3/h
• If the effective flow exceeds or falls below this measuring range, the diagnostic message
S441 Curr.output 1 to n is displayed.

"Forward/Reverse" option

I [mA]

20

4
0 Q
0
3 1 2
A0013758

I Current
Q Flow
1 Value assigned to the 0/4 mA current
2 Forward flow
3 Reverse flow

* Visibility depends on order options or device settings

Endress+Hauser 123
Description of device parameters Proline Promass 500 Modbus RS485

• The current output signal is independent of the direction of flow (absolute amount of the
measured variable). The values for the 0/4 mA value parameter (→  120) and 20 mA
value parameter (→  121) must have the same sign.
• The value for the 20 mA value parameter (→  121) (e.g. reverse flow) corresponds to
the mirrored value for the 20 mA value parameter (→  121) (e.g. forward flow).

"Rev. flow comp." option


The Rev. flow comp. option is primarily used to compensate for abrupt reverse flow which
can occur in connection with positive displacement pumps as a result of wear or high
viscosity. The reverse flows are recorded in a buffer and balanced against forward flow the
next time flow is in the forward direction.
If buffering cannot be processed within approx. 60 s, the diagnostic message
S441 Curr.output 1 to n is displayed.
Flow values can aggregate in the buffer in the event of prolonged and unwanted fluid
reverse flow. However, these flows are not taken into consideration by the current output
configuration, i.e. the reverse flow is not compensated.
If this option is set, the measuring device does not attenuate the flow signal. The flow
signal is not attenuated.

Examples of how the current output behaves


Example 1
Defined measuring range: lower range value and upper range value with the same sign

I [mA]

20

4
0 Q
1 2
A0028084

2 Measuring range
I Current
Q Flow
1 Lower range value (value assigned to 0/4 mA current)
2 Upper range value (value assigned to 20 mA current)

With the following flow response:

A0028091

3 Flow response
Q Flow
t Time

With Forward flow option


The current output signal is proportional to the process variable assigned. The flow
components outside the scaled measuring range are not taken into account for signal
output:.

124 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters

I [mA]

t
A0028092

I Current
t Time

With Forward/Reverse option


The current output signal is independent of the direction of flow.

I [mA]

A0028093

I Current
t Time

With Rev. flow comp. option


Flow components outside the span are buffered, balanced and output after a maximum
delay of 60 s.

I [mA]

A
t
S
S=A
A0028094

I Current
t Time
S Flow components saved
A Balancing of saved flow components

Example 2
Defined measuring range: lower range value and upper range value with different signs

I [mA]

20

4
0 Q
0
1 2
A0028095

4 Measuring range
I Current
Q Flow
1 Lower range value (value assigned to 0/4 mA current)
2 Upper range value (value assigned to 20 mA current)

Endress+Hauser 125
Description of device parameters Proline Promass 500 Modbus RS485

With flow a (—) outside, b (- -) inside the measuring range

Q
1
+
0 t
-
2
a=
b=
A0028098

Q Flow
t Time
1 Lower range value (value assigned to 0/4 mA current)
2 Upper range value (value assigned to 20 mA current)

With Forward flow option


• a (—): The flow components outside the scaled measuring range cannot be taken into
account for signal output.
The diagnostic message S441 Curr.output 1 to n is displayed.
• b (- -): The current output signal is proportional to the process variable assigned.

I [mA]

t
a=
b=
A0028100

I Current
t Time

With Forward/Reverse option


This option is not possible in this case as the values for the 0/4 mA value parameter
(→  120) and 20 mA value parameter (→  121) have different signs.
With Rev. flow comp. option
Flow components outside the span are buffered, balanced and output after a maximum
delay of 60 s.

I [mA]

t
S
S=A
A0028101

I Current
t Time
S Flow components saved
A Balancing of saved flow components

126 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters

Damping out. 1 to n 

Navigation  Expert → Output → Curr.output 1 to n → Damping out. 1 to n (0363–1 to n)

Prerequisite A process variable is selected in the Assign curr. parameter (→  117) and one of the
following options is selected in the Current span parameter (→  118):
• 4...20 mA NAMUR
• 4...20 mA US
• 4...20 mA
• 0...20 mA

Description Use this function to enter a time constant for the reaction time of the current output signal
to fluctuations in the measured value caused by process conditions.

User entry 0.0 to 999.9 s

Factory setting 1.0 s

Additional information User entry


Use this function to enter a time constant (PT1 element 6)) for current output damping:
• If a low time constant is entered, the current output reacts particularly quickly to
fluctuating measured variables.
• On the other hand, the current output reacts more slowly if a high time constant is
entered.
Damping is switched off if 0 is entered (factory setting).

Response time

Navigation  Expert → Output → Curr.output 2 → Response time (0378)

Prerequisite One of the following options is selected in the Assign curr. parameter (→  117):
• Mass flow
• Volume flow
• Correct.vol.flow
• Target mass flow *
• Carrier mass fl. *
• Density
• Ref.density
• Concentration *
• Dynam. viscosity *
• Kinematic visc. *
• TempCompDynVisc *
• TempCompKinVisc *
• Temperature
• Carr. pipe temp. *
• Electronic temp.
• Osc. freq. 0
• Osc. freq. 1 *

6) proportional transmission behavior with first order delay


* Visibility depends on order options or device settings

Endress+Hauser 127
Description of device parameters Proline Promass 500 Modbus RS485

• Osc. ampl. 0 *
• Osc. ampl. 1 *
• Freq. fluct. 0
• Freq. fluct. 1 *
• Osc. damping 0
• Osc. damping 1 *
• Osc.damp.fluct 0
• Osc.damp.fluct 1 *
• Signal asymmetry
• Exc. current 0
• Exc. current 1 *
• HBSI *
Detailed description of the options Oscil. frequency, Oscil. amplitude, Oscil.
damping and Signal asymmetry: Value 1 display parameter (→  19)
One of the following options is selected in the Current span parameter (→  118):
• 4...20 mA NAMUR
• 4...20 mA US
• 4...20 mA
• 0...20 mA

Description Displays the response time. This specifies how quickly the current output reaches the
measured value change of 63 % of 100 % of the measured value change.

User interface Positive floating-point number

Additional information Description


The response time is made up of the time specified for the following dampings:
• Current output damping →  127
and
• Depending on the measured variable assigned to the output.
• Flow damping
or
• Density damping
or
• Temperature damping

Failure mode 

Navigation  Expert → Output → Curr.output 1 to n → Failure mode (0364–1 to n)

Prerequisite A process variable is selected in the Assign curr. parameter (→  117) and one of the
following options is selected in the Current span parameter (→  118):
• 4...20 mA NAMUR
• 4...20 mA US
• 4...20 mA
• 0...20 mA

Description Use this function to select the value of the current output in the event of a device alarm.

* Visibility depends on order options or device settings

128 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters

Selection • Min.
• Max.
• Last valid value
• Actual value
• Defined value

Factory setting Max.

Additional information Description


This setting does not affect the failsafe mode of other outputs and totalizers. This is
specified in separate parameters.

"Min." option
The current output adopts the value of the lower level for signal on alarm.
The signal on alarm level is defined via the Current span parameter (→  118).

"Max." option
The current output adopts the value of the upper level for signal on alarm.
The signal on alarm level is defined via the Current span parameter (→  118).

"Last valid value" option


The current output adopts the last measured value that was valid before the device alarm
occurred.

"Actual value" option


The current output adopts the measured value on the basis of the current flow
measurement; the device alarm is ignored.

"Defined value" option


The current output adopts a defined measured value.
The measured value is defined via the Failure current parameter (→  129).

Failure current 

Navigation  Expert → Output → Curr.output 1 to n → Failure current (0352–1 to n)

Prerequisite The Defined value option is selected in the Failure mode parameter (→  128).

Description Use this function to enter a fixed value that the current output adopts in the event of a
device alarm.

User entry 0 to 22.5 mA

Factory setting 22.5 mA

Endress+Hauser 129
Description of device parameters Proline Promass 500 Modbus RS485

Output curr. 1 to n

Navigation  Expert → Output → Curr.output 1 to n → Output curr. 1 to n (0361–1 to n)

Description Displays the current value currently calculated for the current output.

User interface 3.59 to 22.5 mA

Measur. curr. 1 to n

Navigation  Expert → Output → Curr.output 1 to n → Measur. curr. 1 to n (0366–1 to n)

Description Use this function to display the actual measured value of the output current.

User interface 0 to 30 mA

3.5.2 "Pulse/frequency/switch output 1 to n" submenu

Navigation  Expert → Output → PFS output 1 to n

‣ PFS output 1 to n

Terminal no. (0492–1 to n) →  131

Signal mode (0490–1 to n) →  132

Operating mode (0469–1 to n) →  132

Assign pulse 1 to n (0460–1 to n) →  134

Pulse scaling (0455–1 to n) →  134

Pulse width (0452–1 to n) →  135

Measuring mode (0457–1 to n) →  136

Failure mode (0480–1 to n) →  136

Pulse output 1 to n (0456–1 to n) →  137

Assign freq. (0478–1 to n) →  138

Min. freq. value (0453–1 to n) →  139

130 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters

Max. freq. value (0454–1 to n) →  139

Val. at min.freq (0476–1 to n) →  140

Val. at max.freq (0475–1 to n) →  140

Measuring mode (0479–1 to n) →  140

Damping out. 1 to n (0477–1 to n) →  142

Response time (0491–1 to n) →  143

Failure mode (0451–1 to n) →  144

Failure freq. (0474–1 to n) →  144

Output freq. 1 to n (0471–1 to n) →  145

Switch out funct (0481–1 to n) →  145

Assign diag. beh (0482–1 to n) →  146

Assign limit (0483–1 to n) →  146

Switch-on value (0466–1 to n) →  149

Switch-off value (0464–1 to n) →  149

Assign dir.check (0484–1 to n) →  150

Assign status (0485–1 to n) →  150

Switch-on delay (0467–1 to n) →  150

Switch-off delay (0465–1 to n) →  151

Failure mode (0486–1 to n) →  151

Switch status 1 to n (0461–1 to n) →  151

Invert outp.sig. (0470–1 to n) →  152

Terminal no.

Navigation  Expert → Output → PFS output 1 to n → Terminal no. (0492–1 to n)

Description Displays the terminal numbers used by the pulse/frequency/switch output module.

Endress+Hauser 131
Description of device parameters Proline Promass 500 Modbus RS485

User interface • Not used


• 24-25 (I/O 2)
• 22-23 (I/O 3)
• 20-21 (I/O 4) *

Additional information "Not used" option


The pulse/frequency/switch output module does not use any terminal numbers.

Signal mode 

Navigation  Expert → Output → PFS output 1 to n → Signal mode (0490–1 to n)

Description Use this function to select the signal mode for the pulse/frequency/switch output.

Selection • Passive
• Active
• Passive NAMUR

Factory setting Passive

Operating mode 

Navigation  Expert → Output → PFS output 1 to n → Operating mode (0469–1 to n)

Description Use this function to select the operating mode of the output as a pulse, frequency or switch
output.

Selection • Pulse
• Frequency
• Switch

Factory setting Pulse

Additional information "Pulse" option


Quantity-dependent pulse with configurable pulse width
• Whenever a specific mass, volume, corrected volume, target mass or carrier mass is
reached (pulse value), a pulse is output, the duration of which was set previously (pulse
width).
• The pulses are never shorter than the set duration.
Example
• Flow rate approx. 100 g/s
• Pulse value 0.1 g
• Pulse width 0.05 ms
• Pulse rate 1 000 Impuls/s

* Visibility depends on order options or device settings

132 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters

U[V] B<P
B
24

0 t
P
A0026883

5 Quantity-proportional pulse (pulse value) with pulse width to be configured


B Pulse width entered
P Pauses between the individual pulses

"Frequency" option
Flow-proportional frequency output with 1:1 on/off ratio
An output frequency is output that is proportional to the value of a process variable, such
as mass flow, volume flow, corrected volume flow, target mass flow, carrier mass flow,
density, reference density, concentration, dynamic viscosity, kinematic viscosity,
temperature-compensated dynamic viscosity, temperature-compensated kinematic
viscosity, temperature, carrier tube temperature, electronic temperature, vibration
frequency, frequency fluctuation, oscillation amplitude, oscillation damping, oscillation
damping fluctuation, signal asymmetry or excitation current.
Example
• Flow rate approx. 100 g/s
• Max. frequency 10 kHz
• Flow rate at max. frequency 1 000 g/s
• Output frequency approx. 1 000 Hz

U[V]

24

0 t
A0026886

6 Flow-proportional frequency output

"Switch" option
Contact for displaying a condition (e.g. alarm or warning if a limit value is reached)
Example
Alarm response without alarm

U[V]

24

0 t
A0026884

7 No alarm, high level

Example
Alarm response in case of alarm

Endress+Hauser 133
Description of device parameters Proline Promass 500 Modbus RS485

U[V]
24

0 t
A0026885

8 Alarm, low level

Assign pulse 1 to n 

Navigation  Expert → Output → PFS output 1 to n → Assign pulse 1 to n (0460–1 to n)

Prerequisite The Pulse option is selected in the Operating mode parameter (→  132) parameter.

Description Use this function to select the process variable for the pulse output.

Selection • Off
• Mass flow
• Volume flow
• Correct.vol.flow *
• Target mass flow *
• Carrier mass fl. *
• Target vol. flow *
• Carrier vol. fl. *
• Targ.corr.vol.fl *
• Carr.corr.vol.fl *
• GSV flow *
• GSVa *
• NSV flow *
• NSVa *
• S&W volume flow *
• Oil mass flow *
• Water mass flow *
• Oil volume flow *
• Water vol. flow *
• Oil corr.vol.fl. *
• Water corr.v.fl. *

Factory setting Off

Pulse scaling 

Navigation  Expert → Output → PFS output 1 to n → Pulse scaling (0455–1 to n)

Prerequisite The Pulse option is selected in the Operating mode parameter (→  132) and a process
variable is selected in the Assign pulse parameter (→  134).

Description Use this function to enter the value for the measured value that a pulse is equivalent to.

* Visibility depends on order options or device settings

134 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters

User entry Positive floating point number

Factory setting Depends on country and nominal diameter→  249

Additional information User entry


Weighting of the pulse output with a quantity.
The lower the pulse value, the
• better the resolution.
• the higher the frequency of the pulse response.

Pulse width 

Navigation  Expert → Output → PFS output 1 to n → Pulse width (0452–1 to n)

Prerequisite The Pulse option is selected in the Operating mode parameter (→  132) and a process
variable is selected in the Assign pulse parameter (→  134).

Description Use this function to enter the duration of the output pulse.

User entry 0.05 to 2 000 ms

Factory setting 100 ms

Additional information Description


• Define how long a pulse is (duration).
• The maximum pulse rate is defined by fmax = 1 / (2 × pulse width).
• The interval between two pulses lasts at least as long as the set pulse width.
• The maximum flow is defined by Qmax = fmax × pulse value.
• If the flow exceeds these limit values, the measuring device displays the diagnostic
message S443 Pulse output 1 to n.

U[V] B<P U[V] B=P


B B

t t
P P
A0026882

B Pulse width entered


P Pauses between the individual pulses

Example
• Pulse value: 0.1 g
• Pulse width: 0.1 ms
• fmax: 1 / (2 × 0.1 ms) = 5 kHz
• Qmax: 5 kHz × 0.1 g = 0.5 kg/s

Endress+Hauser 135
Description of device parameters Proline Promass 500 Modbus RS485

Measuring mode 

Navigation  Expert → Output → PFS output 1 to n → Measuring mode (0457–1 to n)

Prerequisite In the Operating mode parameter (→  132), the Pulse option is selected, and one of
the following options is selected in the Assign pulse parameter (→  134):
• Mass flow
• Volume flow
• Correct.vol.flow
• Target mass flow *
• Carrier mass fl. *

Description Use this function to select the measuring mode for the pulse output.

Selection • Forward flow


• Forward/Reverse
• Reverse flow
• Rev. flow comp.

Factory setting Forward flow

Additional information Selection


• Forward flow
Positive flow is output, negative flow is not output.
• Forward/Reverse
Positive and negative flow are output (absolute value), but a distinction is not made
between positive and negative flow.
• Reverse flow
Negative flow is output, positive flow is not output.
• Rev. flow comp.
The flow components outside the span are buffered, balanced and output after a
maximum delay of 60 s.
For a detailed description of the options available, see the Measuring mode
parameter (→  122)

Examples
For a detailed description of the configuration examples, see the Measuring mode
parameter (→  122)

Failure mode 

Navigation  Expert → Output → PFS output 1 to n → Failure mode (0480–1 to n)

Prerequisite The Pulse option is selected in the Operating mode parameter (→  132) and a process
variable is selected in the Assign pulse parameter (→  134).

Description Use this function to select the failure mode of the pulse output in the event of a device
alarm.

* Visibility depends on order options or device settings

136 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters

Selection • Actual value


• No pulses

Factory setting No pulses

Additional information Description


The dictates of safety render it advisable to ensure that the pulse output shows a
predefined behavior in the event of a device alarm.

Selection
• Actual value
In the event of a device alarm, the pulse output continues on the basis of the current
flow measurement. The fault is ignored.
• No pulses
In the event of a device alarm, the pulse output is "switched off".
NOTICE! A device alarm is a measuring device error that must be taken seriously. It can
affect the measurement quality such that the quality can no longer be guaranteed. The
Actual value option is only recommended if it can be guaranteed that all possible alarm
conditions will not affect the measurement quality.

Pulse output 1 to n

Navigation  Expert → Output → PFS output 1 to n → Pulse output 1 to n (0456–1 to n)

Prerequisite The Pulse option is selected in the Operating mode parameter (→  132) parameter.

Description Displays the pulse frequency currently output.

User interface Positive floating-point number

Additional information Description


• The pulse output is an open collector output.
• This is configured at the factory in such a way that the transistor is conductive for the
duration of the pulse (NO contact) and is safety-oriented.

1 1

0 t 0 t
NC NO
A0028726

0 Non-conductive
1 Conductive
NC NC contact (normally closed)
NO NO contact (normally open)

The output behavior can be reversed via the Invert outp.sig. parameter (→  152) i.e.
the transistor does not conduct for the duration of the pulse.
In addition, the behavior of the output in the event of a device alarm (Failure mode
parameter (→  136)) can be configured.

Endress+Hauser 137
Description of device parameters Proline Promass 500 Modbus RS485

Assign freq. 

Navigation  Expert → Output → PFS output 1 to n → Assign freq. (0478–1 to n)

Prerequisite The Frequency option is selected in the Operating mode parameter (→  132).

Description Use this function to select the process variable for the frequency output.
Detailed description of the options Oscil. frequency, Oscil. amplitude, Oscil.
damping and Signal asymmetry: Value 1 display parameter (→  19)

Selection • Off
• Mass flow
• Volume flow
• Correct.vol.flow
• Target mass flow *
• Carrier mass fl. *
• Target vol. flow *
• Carrier vol. fl. *
• Targ.corr.vol.fl *
• Carr.corr.vol.fl *
• Density
• Ref.density
• Ref.dens.altern. *
• GSV flow *
• GSVa *
• NSV flow *
• NSVa *
• S&W volume flow *
• Water cut *
• Oil density *
• Water density *
• Oil mass flow *
• Water mass flow *
• Oil volume flow *
• Water vol. flow *
• Oil corr.vol.fl. *
• Water corr.v.fl. *
• Concentration *
• Dynam. viscosity *
• Kinematic visc. *
• TempCompDynVisc *
• TempCompKinVisc *
• Temperature
• Carr. pipe temp. *
• Electronic temp.
• Osc. freq. 0
• Osc. freq. 1 *
• Osc. ampl. 0 *
• Osc. ampl. 1 *
• Freq. fluct. 0 *
• Freq. fluct. 1 *
• Osc. damping 0 *
• Osc. damping 1 *
• Osc.damp.fluct 0 *

* Visibility depends on order options or device settings

138 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters

• Osc.damp.fluct 1 *
• Signal asymmetry *
• Exc. current 0 *
• Exc. current 1 *
• HBSI *
• Pressure
• Spec. output 0 *
• Spec. output 1 *
• Index inh.medium
• Index sus.bubble *

Factory setting Off

Min. freq. value 

Navigation  Expert → Output → PFS output 1 to n → Min. freq. value (0453–1 to n)

Prerequisite The Frequency option is selected in the Operating mode parameter (→  132) and a
process variable is selected in the Assign freq. parameter (→  138).

Description Use this function to enter the minimum frequency.

User entry 0.0 to 10 000.0 Hz

Factory setting 0.0 Hz

Max. freq. value 

Navigation  Expert → Output → PFS output 1 to n → Max. freq. value (0454–1 to n)

Prerequisite The Frequency option is selected in the Operating mode parameter (→  132) and a
process variable is selected in the Assign freq. parameter (→  138).

Description Use this function to enter the end value frequency.

User entry 0.0 to 10 000.0 Hz

Factory setting 10 000.0 Hz

* Visibility depends on order options or device settings

Endress+Hauser 139
Description of device parameters Proline Promass 500 Modbus RS485

Val. at min.freq 

Navigation  Expert → Output → PFS output 1 to n → Val. at min.freq (0476–1 to n)

Prerequisite The Frequency option is selected in the Operating mode parameter (→  132) and a
process variable is selected in the Assign freq. parameter (→  138).

Description Use this function to enter the measured value for the start value frequency.

User entry Signed floating-point number

Factory setting Depends on country and nominal diameter

Additional information Dependency


The entry depends on the process variable selected in the Assign freq. parameter
(→  138).

Val. at max.freq 

Navigation  Expert → Output → PFS output 1 to n → Val. at max.freq (0475–1 to n)

Prerequisite The Frequency option is selected in the Operating mode parameter (→  132) and a
process variable is selected in the Assign freq. parameter (→  138).

Description Use this function to enter the measured value for the end value frequency.

User entry Signed floating-point number

Factory setting Depends on country and nominal diameter

Additional information Description


Use this function to enter the maximum measured value at the maximum frequency. The
selected process variable is output as a proportional frequency.

Dependency
The entry depends on the process variable selected in the Assign freq. parameter
(→  138).

Measuring mode 

Navigation  Expert → Output → PFS output 1 to n → Measuring mode (0479–1 to n)

Prerequisite In the Operating mode parameter (→  132), the Frequency option is selected, and one
of the following options is selected in the Assign freq. parameter (→  138):
• Mass flow
• Volume flow
• Correct.vol.flow

140 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters

• Target mass flow *


• Carrier mass fl. *
• Density
• Ref.density
• Concentration *
• Dynam. viscosity *
• Kinematic visc. *
• TempCompDynVisc *
• TempCompKinVisc *
• Temperature
• Carr. pipe temp. *
• Electronic temp.
• Osc. freq. 0
• Osc. freq. 1 *
• Freq. fluct. 0
• Freq. fluct. 1 *
• Osc. ampl. 0 *
• Osc. ampl. 1 *
• Osc. damping 0
• Osc. damping 1 *
• Osc.damp.fluct 0
• Osc.damp.fluct 1 *
• Signal asymmetry
• Exc. current 0
• Exc. current 1 *
Detailed description of the options Oscil. frequency, Oscil. amplitude, Oscil.
damping and Signal asymmetry: Value 1 display parameter (→  19)

Description Use this function to select the measuring mode for the frequency output.

Selection • Forward flow


• Forward/Reverse
• Rev. flow comp.

Factory setting Forward flow

Additional information Selection


For a detailed description of the options available, see the Measuring mode
parameter (→  122)

Examples
For a detailed description of the configuration examples, see the Measuring mode
parameter (→  122)

* Visibility depends on order options or device settings

Endress+Hauser 141
Description of device parameters Proline Promass 500 Modbus RS485

Damping out. 1 to n 

Navigation  Expert → Output → PFS output 1 to n → Damping out. 1 to n (0477–1 to n)

Prerequisite In the Operating mode parameter (→  132), the Frequency option is selected, and one
of the following options is selected in the Assign freq. parameter (→  138):
• Mass flow
• Volume flow
• Correct.vol.flow
• Target mass flow *
• Carrier mass fl. *
• Density
• Ref.density
• Concentration *
• Dynam. viscosity *
• Kinematic visc. *
• TempCompDynVisc *
• TempCompKinVisc *
• Temperature
• Carr. pipe temp. *
• Electronic temp.
• Osc. freq. 0
• Osc. freq. 1 *
• Freq. fluct. 0
• Freq. fluct. 1 *
• Osc. ampl. 0 *
• Osc. ampl. 1 *
• Osc. damping 0
• Osc. damping 1 *
• Osc.damp.fluct 0
• Osc.damp.fluct 1 *
• Signal asymmetry
• Exc. current 0
• Exc. current 1 *
Detailed description of the options Oscil. frequency, Oscil. amplitude, Oscil.
damping and Signal asymmetry: Value 1 display parameter (→  19)

Description Use this function to enter a time constant for the reaction time of the output signal to
fluctuations in the measured value.

User entry 0 to 999.9 s

Factory setting 0.0 s

Additional information User entry


Use this function to enter a time constant (PT1 element 7)) for frequency output damping:
• If a low time constant is entered, the current output reacts particularly quickly to
fluctuating measured variables.
• On the other hand, the current output reacts more slowly if a high time constant is
entered.
Damping is switched off if 0 is entered (factory setting).

* Visibility depends on order options or device settings


7) proportional transmission behavior with first order delay

142 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters

The frequency output is subject to separate damping that is independent of all preceding
time constants.

Response time

Navigation  Expert → Output → PFS output 1 to n → Response time (0491–1 to n)

Prerequisite In the Operating mode parameter (→  132), the Frequency option is selected, and one
of the following options is selected in the Assign freq. parameter (→  138):
• Mass flow
• Volume flow
• Correct.vol.flow
• Target mass flow *
• Carrier mass fl. *
• Density
• Ref.density
• Concentration *
• Dynam. viscosity *
• Kinematic visc. *
• TempCompDynVisc *
• TempCompKinVisc *
• Temperature
• Carr. pipe temp. *
• Electronic temp.
• Osc. freq. 0
• Osc. freq. 1 *
• Freq. fluct. 0
• Freq. fluct. 1 *
• Osc. ampl. 0 *
• Osc. ampl. 1 *
• Osc. damping 0
• Osc. damping 1 *
• Osc.damp.fluct 0
• Osc.damp.fluct 1 *
• Signal asymmetry
• Exc. current 0
• Exc. current 1 *
Detailed description of the options Oscil. frequency, Oscil. amplitude, Oscil.
damping and Signal asymmetry: Value 1 display parameter (→  19)

Description Displays the response time. This specifies how quickly the pulse/frequency/switch output
reaches the measured value change of 63 % of 100 % of the measured value change.

User interface Positive floating-point number

* Visibility depends on order options or device settings

Endress+Hauser 143
Description of device parameters Proline Promass 500 Modbus RS485

Additional information Description


The response time is made up of the time specified for the following dampings:
• Damping of pulse/frequency/switch output →  127
and
• Depending on the measured variable assigned to the output.
• Flow damping
or
• Density damping
or
• Temperature damping

Failure mode 

Navigation  Expert → Output → PFS output 1 to n → Failure mode (0451–1 to n)

Prerequisite The Frequency option is selected in the Operating mode parameter (→  132) and a
process variable is selected in the Assign freq. parameter (→  138).

Description Use this function to select the failure mode of the frequency output in the event of a device
alarm.

Selection • Actual value


• Defined value
• 0 Hz

Factory setting 0 Hz

Additional information Selection


• Actual value
In the event of a device alarm, the frequency output continues on the basis of the current
flow measurement. The device alarm is ignored.
• Defined value
In the event of a device alarm, the frequency output continues on the basis of a
predefined value. The Failure freq. (→  144) replaces the current measured value,
making it possible to bypass the device alarm. The actual measurement is switched off
for the duration of the device alarm.
• 0 Hz
In the event of a device alarm, the frequency output is "switched off".
NOTICE! A device alarm is a measuring device error that must be taken seriously. It can
affect the measurement quality such that the quality can no longer be guaranteed. The
Actual value option is only recommended if it can be guaranteed that all possible alarm
conditions will not affect the measurement quality.

Failure freq. 

Navigation  Expert → Output → PFS output 1 to n → Failure freq. (0474–1 to n)

Prerequisite The Frequency option is selected in the Operating mode parameter (→  132) and a
process variable is selected in the Assign freq. parameter (→  138).

144 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters

Description Use this function to enter the value for the frequency output in the event of a device alarm
in order to bypass the alarm.

User entry 0.0 to 12 500.0 Hz

Factory setting 0.0 Hz

Output freq. 1 to n

Navigation  Expert → Output → PFS output 1 to n → Output freq. 1 to n (0471–1 to n)

Prerequisite In the Operating mode parameter (→  132), the Frequency option is selected.

Description Displays the actual value of the output frequency which is currently measured.

User interface 0.0 to 12 500.0 Hz

Switch out funct 

Navigation  Expert → Output → PFS output 1 to n → Switch out funct (0481–1 to n)

Prerequisite The Switch option is selected in the Operating mode parameter (→  132).

Description Use this function to select a function for the switch output.

Selection • Off
• On
• Diag. behavior
• Limit
• Fl. direct.check
• Status

Factory setting Off

Additional information Selection


• Off
The switch output is permanently switched off (open, non-conductive).
• On
The switch output is permanently switched on (closed, conductive).
• Diag. behavior
Indicates if the diagnostic event is present or not. Is used to output diagnostic
information and to react to it appropriately at the system level.

Endress+Hauser 145
Description of device parameters Proline Promass 500 Modbus RS485

• Limit
Indicates if a specified limit value has been reached for the process variable. Is used to
output diagnostic information relating to the process and to react to it appropriately at
the system level.
• Fl. direct.check
Indicates the flow direction (forward or reverse flow).
• Status
Indicates the device status depending on whether empty pipe detection or low flow cut
off is selected.

Assign diag. beh 

Navigation  Expert → Output → PFS output 1 to n → Assign diag. beh (0482–1 to n)

Prerequisite • In the Operating mode parameter (→  132), the Switch option is selected.
• In the Switch out funct parameter (→  145), the Diag. behavior option is selected.

Description Use this function to select the diagnostic event category that is displayed for the switch
output.

Selection • Alarm
• Alarm or warning
• Warning

Factory setting Alarm

Additional information Description


If no diagnostic event is pending, the switch output is closed and conductive.

Selection
• Alarm
The switch output signals only diagnostic events in the alarm category.
• Alarm or warning
The switch output signals diagnostic events in the alarm and warning category.
• Warning
The switch output signals only diagnostic events in the warning category.

Assign limit 

Navigation  Expert → Output → PFS output 1 to n → Assign limit (0483–1 to n)

Prerequisite • The Switch option is selected in the Operating mode parameter (→  132).
• The Limit option is selected in the Switch out funct parameter (→  145).

Description Use this function to select a process variable for the limit function.

146 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters

Selection • Mass flow


• Volume flow
• Correct.vol.flow *
• Target mass flow *
• Carrier mass fl. *
• Target vol. flow *
• Carrier vol. fl. *
• Targ.corr.vol.fl *
• Carr.corr.vol.fl *
• Density
• Ref.density *
• Ref.dens.altern. *
• GSV flow *
• GSVa *
• NSV flow *
• NSVa *
• S&W volume flow *
• Water cut *
• Oil density *
• Water density *
• Oil mass flow *
• Water mass flow *
• Oil volume flow *
• Water vol. flow *
• Oil corr.vol.fl. *
• Water corr.v.fl. *
• Dynam. viscosity *
• Concentration *
• Kinematic visc. *
• TempCompDynVisc *
• TempCompKinVisc *
• Temperature
• Totalizer 1
• Totalizer 2
• Totalizer 3
• Oscil. damping
• Pressure
• Spec. output 0 *
• Spec. output 1 *
• Index inh.medium
• Index sus.bubble *

Factory setting Mass flow

Additional information Description


Behavior of status output when Switch-on value > Switch-off value:
• Process variable > Switch-on value: transistor is conductive
• Process variable < Switch-off value: transistor is non-conductive

* Visibility depends on order options or device settings

Endress+Hauser 147
Description of device parameters Proline Promass 500 Modbus RS485

A
1

2
t

3
4 t
A0026891

1 Switch-on value
2 Switch-off value
3 Conductive
4 Non-conductive
A Process variable
B Status output

Behavior of status output when Switch-on value < Switch-off value:


• Process variable < Switch-on value: transistor is conductive
• Process variable > Switch-off value: transistor is non-conductive

A
1

2
t

3
4 t
A0026892

1 Switch-off value
2 Switch-on value
3 Conductive
4 Non-conductive
A Process variable
B Status output

Behavior of status output when Switch-on value = Switch-off value:


• Process variable > Switch-on value: transistor is conductive
• Process variable < Switch-off value: transistor is non-conductive

2
3 t
A0026893

1 Switch-on value = Switch-off value


2 Conductive
3 Non-conductive
A Process variable
B Status output

148 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters

Switch-on value 

Navigation  Expert → Output → PFS output 1 to n → Switch-on value (0466–1 to n)

Prerequisite • The Switch option is selected in the Operating mode parameter (→  132).
• The Limit option is selected in the Switch out funct parameter (→  145).

Description Use this function to enter the measured value for the switch-on point.

User entry Signed floating-point number

Factory setting Country-specific:


• 0 kg/h
• 0 lb/min

Additional information Description


Use this function to enter the limit value for the switch-on value (process variable >
switch-on value = closed, conductive).
When using a hysteresis: Switch-on value > Switch-off value.

Dependency
The unit depends on the process variable selected in the Assign limit parameter
(→  146).

Switch-off value 

Navigation  Expert → Output → PFS output 1 to n → Switch-off value (0464–1 to n)

Prerequisite • The Switch option is selected in the Operating mode parameter (→  132).
• The Limit option is selected in the Switch out funct parameter (→  145).

Description Use this function to enter the measured value for the switch-off point.

User entry Signed floating-point number

Factory setting Country-specific:


• 0 kg/h
• 0 lb/min

Additional information Description


Use this function to enter the limit value for the switch-off value (process variable <
switch-off value = open, non-conductive).
When using a hysteresis: Switch-on value > Switch-off value.

Dependency
The unit depends on the process variable selected in the Assign limit parameter
(→  146).

Endress+Hauser 149
Description of device parameters Proline Promass 500 Modbus RS485

Assign dir.check 

Navigation  Expert → Output → PFS output 1 to n → Assign dir.check (0484–1 to n)

Prerequisite • The Switch option is selected in the Operating mode parameter (→  132).
• The Fl. direct.check option is selected in the Switch out funct parameter (→  145).

Description Use this function to select a process variable for monitoring the flow direction.

Selection • Off
• Volume flow
• Mass flow
• Correct.vol.flow *

Factory setting Mass flow

Assign status 

Navigation  Expert → Output → PFS output 1 to n → Assign status (0485–1 to n)

Prerequisite • The Switch option is selected in the Operating mode parameter (→  132).
• The Status option is selected in the Switch out funct parameter (→  145).

Description Use this function to select a device status for the switch output.

Selection • Partial pipe det


• Low flow cut off

Factory setting Partial pipe det

Additional information Options


If empty pipe detection or low flow cut off are enabled, the output is conductive.
Otherwise, the switch output is non-conductive.

Switch-on delay 

Navigation  Expert → Output → PFS output 1 to n → Switch-on delay (0467–1 to n)

Prerequisite • The Switch option is selected in the Operating mode parameter (→  132).
• The Limit option is selected in the Switch out funct parameter (→  145).

Description Use this function to enter a delay time for switching on the switch output.

User entry 0.0 to 100.0 s

Factory setting 0.0 s

* Visibility depends on order options or device settings

150 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters

Switch-off delay 

Navigation  Expert → Output → PFS output 1 to n → Switch-off delay (0465–1 to n)

Prerequisite • The Switch option is selected in the Operating mode parameter (→  132).
• The Limit option is selected in the Switch out funct parameter (→  145).

Description Use this function to enter a delay time for switching off the switch output.

User entry 0.0 to 100.0 s

Factory setting 0.0 s

Failure mode 

Navigation  Expert → Output → PFS output 1 to n → Failure mode (0486–1 to n)

Description Use this function to select a failsafe mode for the switch output in the event of a device
alarm.

Selection • Actual status


• Open
• Closed

Factory setting Open

Additional information Options


• Actual status
In the event of a device alarm, faults are ignored and the current behavior of the input
value is output by the switch output. The Actual status option behaves in the same way
as the current input value.
• Open
In the event of a device alarm, the switch output's transistor is set to non-conductive.
• Closed
In the event of a device alarm, the switch output's transistor is set to conductive.

Switch status 1 to n

Navigation  Expert → Output → PFS output 1 to n → Switch status 1 to n (0461–1 to n)

Prerequisite The Switch option is selected in the Operating mode parameter (→  132).

Description Displays the current switch status of the status output.

User interface • Open


• Closed

Endress+Hauser 151
Description of device parameters Proline Promass 500 Modbus RS485

Additional information User interface


• Open
The switch output is not conductive.
• Closed
The switch output is conductive.

Invert outp.sig. 

Navigation  Expert → Output → PFS output 1 to n → Invert outp.sig. (0470–1 to n)

Description Use this function to select whether to invert the output signal.

Selection • No
• Yes

Factory setting No

Additional information Selection


No option (passive - negative)

0
A0026693

Yes option (passive - positive)

0
A0026692

3.5.3 "Relay output 1 to n" submenu

Navigation  Expert → Output → Relay output 1 to n

‣ Relay output 1 to n

Terminal no. →  153

Relay outp.func. →  153

Assign dir.check →  154

152 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters

Assign limit →  154

Assign diag. beh →  155

Assign status →  156

Switch-off value →  156

Switch-off delay →  157

Switch-on value →  157

Switch-on delay →  158

Failure mode →  158

Switch status →  158

Powerless relay →  159

Terminal no.

Navigation  Expert → Output → Relay output 1 to n → Terminal no. (0812–1 to n)

Description Displays the terminal numbers used by the relay output module.

User interface • Not used


• 24-25 (I/O 2)
• 22-23 (I/O 3)
• 20-21 (I/O 4)

Additional information "Not used" option


The relay output module does not use any terminal numbers.

Relay outp.func. 

Navigation  Expert → Output → Relay output 1 to n → Relay outp.func. (0804–1 to n)

Description Use this function to select an output function for the relay output.

Selection • Closed
• Open
• Diag. behavior
• Limit
• Fl. direct.check
• Digital Output

Endress+Hauser 153
Description of device parameters Proline Promass 500 Modbus RS485

Factory setting Closed

Additional information Selection


• Closed
The relay output is permanently switched on (closed, conductive).
• Open
The relay output is permanently switched off (open, non-conductive).
• Diag. behavior
Indicates if the diagnostic event is present or not. Is used to output diagnostic
information and to react to it appropriately at the system level.
• Limit
Indicates if a specified limit value has been reached for the process variable. Is used to
output diagnostic information relating to the process and to react to it appropriately at
the system level.
• Fl. direct.check
Indicates the flow direction (forward or reverse flow).
• Digital Output
Indicates the device status depending on whether empty pipe detection or low flow cut
off is selected.

Assign dir.check 

Navigation  Expert → Output → Relay output 1 to n → Assign dir.check (0808–1 to n)

Prerequisite In the Relay outp.func. parameter (→  153), the Fl. direct.check option is selected.

Description Use this function to select a process variable for monitoring the flow direction.

Selection • Off
• Volume flow
• Mass flow
• Correct.vol.flow *

Factory setting Mass flow

Assign limit 

Navigation  Expert → Output → Relay output 1 to n → Assign limit (0807–1 to n)

Prerequisite The Limit option is selected in the Relay outp.func. parameter (→  153).

Description Use this function to select a process variable for the limit value function.

Selection • Mass flow


• Volume flow
• Correct.vol.flow *
• Target mass flow *
• Carrier mass fl. *

* Visibility depends on order options or device settings

154 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters

• Target vol. flow *


• Carrier vol. fl. *
• Targ.corr.vol.fl *
• Carr.corr.vol.fl *
• Density
• Ref.density *
• Ref.dens.altern. *
• GSV flow *
• GSVa *
• NSV flow *
• NSVa *
• S&W volume flow *
• Water cut *
• Oil density *
• Water density *
• Oil mass flow *
• Water mass flow *
• Oil volume flow *
• Water vol. flow *
• Oil corr.vol.fl. *
• Water corr.v.fl. *
• Dynam. viscosity *
• Concentration *
• Kinematic visc. *
• TempCompDynVisc *
• TempCompKinVisc *
• Temperature
• Totalizer 1
• Totalizer 2
• Totalizer 3
• Oscil. damping
• Pressure
• Spec. output 0 *
• Spec. output 1 *
• Index inh.medium
• Index sus.bubble *

Factory setting Mass flow

Assign diag. beh 

Navigation  Expert → Output → Relay output 1 to n → Assign diag. beh (0806–1 to n)

Prerequisite In the Relay outp.func. parameter (→  153), the Diag. behavior option is selected.

Description Use this function to select the category of the diagnostic events that are displayed for the
relay output.

Selection • Alarm
• Alarm or warning
• Warning

* Visibility depends on order options or device settings

Endress+Hauser 155
Description of device parameters Proline Promass 500 Modbus RS485

Factory setting Alarm

Additional information Description


If no diagnostic event is pending, the relay output is closed and conductive.

Selection
• Alarm
The relay output signals only diagnostic events in the alarm category.
• Alarm or warning
The relay output signals diagnostic events in the alarm and warning category.
• Warning
The relay output signals only diagnostic events in the warning category.

Assign status 

Navigation  Expert → Output → Relay output 1 to n → Assign status (0805–1 to n)

Prerequisite In the Relay outp.func. parameter (→  153), the Digital Output option is selected.

Description Use this function to select the device status for the relay output.

Selection • Partial pipe det


• Low flow cut off

Factory setting Partial pipe det

Switch-off value 

Navigation  Expert → Output → Relay output 1 to n → Switch-off value (0809–1 to n)

Prerequisite In the Relay outp.func. parameter (→  153), the Limit option is selected.

Description Use this function to enter the measured value for the switch-off point.

User entry Signed floating-point number

Factory setting Country-specific:


• 0 kg/h
• 0 lb/min

156 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters

Additional information Description


Use this function to enter the limit value for the switch-off value (process variable <
switch-off value = open, non-conductive).
When using a hysteresis: Switch-on value > Switch-off value.

Dependency
The unit is dependent on the process variable selected in the Assign limit parameter
(→  154).

Switch-off delay 

Navigation  Expert → Output → Relay output 1 to n → Switch-off delay (0813–1 to n)

Prerequisite In the Relay outp.func. parameter (→  153), the Limit option is selected.

Description Use this function to enter a delay time for switching off the switch output.

User entry 0.0 to 100.0 s

Factory setting 0.0 s

Switch-on value 

Navigation  Expert → Output → Relay output 1 to n → Switch-on value (0810–1 to n)

Prerequisite The Limit option is selected in the Relay outp.func. parameter (→  153).

Description Use this function to enter the measured value for the switch-on point.

User entry Signed floating-point number

Factory setting Country-specific:


• 0 kg/h
• 0 lb/min

Additional information Description


Use this function to enter the limit value for the switch-on value (process variable >
switch-on value = closed, conductive).
When using a hysteresis: Switch-on value > Switch-off value.

Dependency
The unit is dependent on the process variable selected in the Assign limit parameter
(→  154).

Endress+Hauser 157
Description of device parameters Proline Promass 500 Modbus RS485

Switch-on delay 

Navigation  Expert → Output → Relay output 1 to n → Switch-on delay (0814–1 to n)

Prerequisite In the Relay outp.func. parameter (→  153), the Limit option is selected.

Description Use this function to enter a delay time for switching on the switch output.

User entry 0.0 to 100.0 s

Factory setting 0.0 s

Failure mode 

Navigation  Expert → Output → Relay output 1 to n → Failure mode (0811–1 to n)

Description Use this function to select the failure mode of the relay output in the event of a device
alarm.

Selection • Actual status


• Open
• Closed

Factory setting Open

Additional information Selection


• Actual status
In the event of a device alarm, faults are ignored and the current behavior of the input
value is output by the relay output. The Actual status option behaves in the same way as
the current input value.
• Open
In the event of a device alarm, the relay output's transistor is set to non-conductive.
• Closed
In the event of a device alarm, the relay output's transistor is set to conductive.

Switch status

Navigation  Expert → Output → Relay output 1 to n → Switch status (0801–1 to n)

Description Displays the current status of the relay output.

User interface • Open


• Closed

158 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters

Additional information User interface


• Open
The relay output is not conductive.
• Closed
The relay output is conductive.

Powerless relay 

Navigation  Expert → Output → Relay output 1 to n → Powerless relay (0816–1 to n)

Description Use this function to select the quiescent state for the relay output.

Selection • Open
• Closed

Factory setting Open

Additional information Selection


• Open
The relay output is not conductive.
• Closed
The relay output is conductive.

3.5.4 "Double pulse output" submenu

Navigation  Expert → Output → Double pulse out

‣ Double pulse out

Master term. no. (0981) →  160

Slave term. no. (0990) →  160

Signal mode (0991) →  160

Assign pulse 1 (0982–1) →  161

Value per pulse (0983) →  161

Pulse width (0986) →  162

Phase shift (0992) →  162

Measuring mode (0984) →  162

Endress+Hauser 159
Description of device parameters Proline Promass 500 Modbus RS485

Failure mode (0985) →  163

Pulse output (0987) →  164

Invert outp.sig. (0993) →  164

Master term. no.

Navigation  Expert → Output → Double pulse out → Master term. no. (0981)

Description Displays the master terminal number for the double pulse output.

User interface • Not used


• 24-25 (I/O 2)
• 22-23 (I/O 3)

Additional information "Not used" option


The double pulse output does not use any terminal numbers.

Slave term. no.

Navigation  Expert → Output → Double pulse out → Slave term. no. (0990)

Description Displays the slave terminal number for the double pulse output.

User interface • Not used


• 24-25 (I/O 2)
• 22-23 (I/O 3)

Additional information "Not used" option


The double pulse output does not use any terminal numbers.

Signal mode 

Navigation  Expert → Output → Double pulse out → Signal mode (0991)

Description Use this function to select the signal mode for the double pulse output.

Selection • Passive
• Active
• Passive NAMUR

Factory setting Passive

160 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters

Assign pulse 1 

Navigation  Expert → Output → Double pulse out → Assign pulse 1 (0982–1)

Description Use this function to select a process variable for the double pulse output.

Selection • Off
• Mass flow
• Volume flow
• Correct.vol.flow *
• Target mass flow *
• Carrier mass fl. *
• Target vol. flow *
• Carrier vol. fl. *
• Targ.corr.vol.fl *
• Carr.corr.vol.fl *
• GSV flow *
• GSVa *
• NSV flow *
• NSVa *
• S&W volume flow *
• Oil mass flow *
• Water mass flow *
• Oil volume flow *
• Water vol. flow *
• Oil corr.vol.fl. *
• Water corr.v.fl. *

Factory setting Off

Value per pulse 

Navigation  Expert → Output → Double pulse out → Value per pulse (0983)

Description Use this function to enter the value for the measured value that a pulse is equivalent to.

User entry Signed floating-point number

Factory setting Depends on country and nominal diameter →  249

Additional information User entry


Weighting of the pulse output with a quantity.
The lower the pulse value, the
• better the resolution.
• the higher the frequency of the pulse response.

* Visibility depends on order options or device settings

Endress+Hauser 161
Description of device parameters Proline Promass 500 Modbus RS485

Pulse width 

Navigation  Expert → Output → Double pulse out → Pulse width (0986)

Description Use this function to enter the duration of the output pulse.

User entry 0.5 to 2 000 ms

Factory setting 0.5 ms

Additional information For a detailed description and example: Pulse width parameter (→  135)

Phase shift 

Navigation  Expert → Output → Double pulse out → Phase shift (0992)

Description Use this function to select the degree of phase shift.

Selection • 90°
• 180°

Factory setting 90°

Additional information Selection


• 90°
Phase shift by a quarter period.
• 180°
Phase shift by a half period, which is equivalent to a phase reversal.

Measuring mode 

Navigation  Expert → Output → Double pulse out → Measuring mode (0984)

Description Use this function to select the measuring mode for the double pulse output.

Selection • Forward flow


• Forward/Reverse
• Reverse flow
• Rev. flow comp.

Factory setting Forward flow

162 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters

Additional information Selection


• Forward flow
Positive flow is output, negative flow is not output.
• Forward/Reverse
Positive and negative flow are output (absolute value), but a distinction is not made
between positive and negative flow.
• Reverse flow
Negative flow is output, positive flow is not output.
• Rev. flow comp.
The flow components outside the span are buffered, balanced and output after a
maximum delay of 60 s.
For a detailed description of the options available, see the Measuring mode
parameter (→  122)

Examples
For a detailed description of the configuration examples, see the Measuring mode
parameter (→  122)

Failure mode 

Navigation  Expert → Output → Double pulse out → Failure mode (0985)

Description Use this function to select the failure mode of the double pulse output in the event of a
device alarm.

Selection • Actual value


• No pulses

Factory setting No pulses

Additional information Description


The dictates of safety render it advisable to ensure that the double pulse output shows a
predefined behavior in the event of a device alarm.

Selection
• Actual value
In the event of a device alarm, the double pulse output continues on the basis of the
current flow measurement. The fault is ignored.
• No pulses
In the event of a device alarm, the double pulse output is "switched off".
NOTICE! A device alarm is a measuring device error that must be taken seriously. It can
affect the measurement quality such that the quality can no longer be guaranteed. The
Actual value option is only recommended if it can be guaranteed that all possible alarm
conditions will not affect the measurement quality.

Endress+Hauser 163
Description of device parameters Proline Promass 500 Modbus RS485

Pulse output

Navigation  Expert → Output → Double pulse out → Pulse output (0987)

Description Displays the pulse frequency of the double pulse output which is currently output.

User interface Positive floating-point number

Additional information For a detailed description and example: Pulse output parameter (→  62)

Invert outp.sig. 

Navigation  Expert → Output → Double pulse out → Invert outp.sig. (0993)

Description Use this function to select whether to invert the output signal.

Selection • No
• Yes

Factory setting No

Additional information Selection


No option (passive - negative)

0
A0026693

Yes option (passive - positive)

0
A0026692

3.6 "Communication" submenu


Navigation  Expert → Communication

‣ Communication

‣ Modbus config. →  165

164 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters

‣ Modbus info →  170

‣ Modbus data map →  171

‣ Web server →  171

‣ WLAN settings →  174

3.6.1 "Modbus config." submenu

Navigation  Expert → Communication → Modbus config.

‣ Modbus config.

Bus address (7112) →  165

Baudrate (7111) →  166

Data trans. mode (7115) →  166

Parity (7122) →  166

Byte order (7113) →  167

Telegram delay (7146) →  168

Failure mode (7116) →  168

Bus termination (7155) →  169

Fieldb.writ.acc. (7156) →  169

Bus address 

Navigation  Expert → Communication → Modbus config. → Bus address (7112)

Description For entering the device address.

User entry 1 to 247

Factory setting 247

Endress+Hauser 165
Description of device parameters Proline Promass 500 Modbus RS485

Baudrate 

Navigation  Expert → Communication → Modbus config. → Baudrate (7111)

Description Use this function to select a transmission rate.

Selection • 1200 BAUD


• 2400 BAUD
• 4800 BAUD
• 9600 BAUD
• 19200 BAUD
• 38400 BAUD
• 57600 BAUD
• 115200 BAUD

Factory setting 19200 BAUD

Data trans. mode 

Navigation  Expert → Communication → Modbus config. → Data trans. mode (7115)

Description Use this function to select the data transmission mode.

Selection • ASCII
• RTU

Factory setting RTU

Additional information Options


• ASCII
Transmission of data in the form of readable ASCII characters. Error protection via LRC.
• RTU
Transmission of data in binary form. Error protection via CRC16.

Parity 

Navigation  Expert → Communication → Modbus config. → Parity (7122)

Description Use this function to select the parity bit.

Selection • Odd
• Even
• None/1 stop bit
• None/2 stop bits

Factory setting Even

166 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters

Additional information Options


Picklist ASCII option:
• 0 = Even option
• 1 = Odd option
Picklist RTU option:
• 0 = Even option
• 1 = Odd option
• 2 = None/1 stop bit option
• 3 = None/2 stop bits option

Byte order 

Navigation  Expert → Communication → Modbus config. → Byte order (7113)

Description Use this function to select the sequence in which the bytes are transmitted. The
transmission sequence must be coordinated with the Modbus master.

Selection • 0-1-2-3
• 3-2-1-0
• 1-0-3-2
• 2-3-0-1

Factory setting 1-0-3-2

Additional information Description


The byte sequence is not standardized by the Modbus protocol. However, if the host
system and the measuring device do not use the same byte sequence, correct data
exchange is not possible.
Changing the byte sequence in the host system often requires a extensive knowledge and
significant programming efforts. Endress+Hauser introduced the Byte order parameter
(→  167) for this reason.
This makes it possible to use the standard settings of the host system and change the byte
sequence on the measuring device by trial and error. If correct data exchange cannot be
achieved by changing the byte sequence, the settings for the byte sequence of the host
system must be adapted accordingly.

Byte transmission sequence


Byte addressing, i.e. the transmission sequence of the bytes, is not specified in the Modbus
specification. For this reason, it is important to coordinate or match the addressing
method between the master and slave during commissioning. This can be configured in
the measuring device using the Byte order parameter (→  167).
The bytes are transmitted depending on the selection in the Byte order parameter
(→  167):

FLOAT

Sequence

Options 1. 2. 3. 4.

1–0–3–2* Byte 1 Byte 0 Byte 3 Byte 2


(MMMMMMMM) (MMMMMMMM) (SEEEEEEE) (EMMMMMMM)

0–1–2–3 Byte 0 Byte 1 Byte 2 Byte 3


(MMMMMMMM) (MMMMMMMM) (EMMMMMMM) (SEEEEEEE)

Endress+Hauser 167
Description of device parameters Proline Promass 500 Modbus RS485

2–3–0–1 Byte 2 Byte 3 Byte 0 Byte 1


(EMMMMMMM) (SEEEEEEE) (MMMMMMM) (MMMMMMM)

3–2–1–0 Byte 3 Byte 2 Byte 1 Byte 0


(SEEEEEEE) (EMMMMMMM) (MMMMMMMM) (MMMMMMMM)

* = factory setting, S = sign, E = exponent, M = mantissa

INTEGER

Sequence

Options 1. 2.

1–0–3–2* Byte 1 Byte 0


3–2–1–0 (MSB) (LSB)

0–1–2–3 Byte 0 Byte 1


2–3–0–1 (LSB) (MSB)

* = factory setting, MSB = most significant byte, LSB = least significant byte

STRING
Presentation taking the example of a device parameter with a data length of 18 bytes.

Sequence

Options 1. 2. ... 17. 18.

1–0–3–2* Byte 17 Byte 16 ... Byte 1 Byte 0


3–2–1–0 (MSB) (LSB)

0–1–2–3 Byte 16 Byte 17 ... Byte 0 Byte 1


2–3–0–1 (MSB) (LSB)

* = factory setting, MSB = most significant byte, LSB = least significant byte

Telegram delay 

Navigation  Expert → Communication → Modbus config. → Telegram delay (7146)

Description Use this function to enter a delay time after which the measuring device replies to the
request telegram of the Modbus master. This allows communication to be adapted to slow
Modbus RS485 masters.

User entry 0 to 100 ms

Factory setting 6 ms

Failure mode 

Navigation  Expert → Communication → Modbus config. → Failure mode (7116)

Description Use this function to select the measured value output in the event of a diagnostic message
via Modbus communication.

Selection • NaN value


• Last valid value

168 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters

Factory setting NaN value

Additional information Options


• NaN value
The device outputs the NaN value 8).
• Last valid value
The device outputs the last valid measured value before the fault occurred.
This effect of this parameter depends on the option selected in the Assign diag. beh
parameter.

Bus termination

Navigation  Expert → Communication → Modbus config. → Bus termination (7155)

Description Displays whether the terminating resistor is enabled or disabled.

User interface • Off


• On

Factory setting Off

Additional information Selection


• Off
The terminating resistor is disabled.
• On
The terminating resistor is enabled.
For detailed information about enabling the terminating resistor, see the Operating
Instructions for the device, "Enabling the terminating resistor" section→  8

Fieldb.writ.acc.

Navigation  Expert → Communication → Modbus config. → Fieldb.writ.acc. (7156)

Description Use this function to restrict access to the measuring device via fieldbus (Modbus protocol).

Selection • Read + write


• Read only

Factory setting Read + write

8) Not a Number

Endress+Hauser 169
Description of device parameters Proline Promass 500 Modbus RS485

Additional information Description


If read and/or write protection is enabled, the parameter can only be controlled and reset
via local operation. Access is no longer possible via operating tools.
This does not affect cyclic measured value transmission to the higher-order system,
which is always guaranteed.

Selection
• Read + write
The parameters are read and write parameters.
• Read only
• The parameters are read only parameters.

3.6.2 "Modbus info" submenu

Navigation  Expert → Communication → Modbus info

‣ Modbus info

Device ID (7153) →  170

Device revision (7154) →  170

Device ID

Navigation  Expert → Communication → Modbus info → Device ID (7153)

Description Displays the device ID for identifying the measuring device.

User interface 4-digit hexadecimal number

Device revision

Navigation  Expert → Communication → Modbus info → Device revision (7154)

Description Displays the device revision.

User interface 4-digit hexadecimal number

170 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters

3.6.3 "Modbus data map" submenu

Navigation  Expert → Communication → Modbus data map

‣ Modbus data map

Scan list reg.0 to 15 (7114) →  171

Scan list reg.0 to 15 

Navigation  Expert → Communication → Modbus data map → Scan list reg.0 to 15 (7114)

Description Use this function to enter the scan list register. By entering the register address (1-based),
up to 16 device parameters can be grouped by assigning them to the scan list registers 0 to
15. The data of the device parameters assigned here are read out via the register addresses
5051 to 5081.

User entry 1 to 65 535

Factory setting 1

3.6.4 "Web server" submenu

Navigation  Expert → Communication → Web server

‣ Web server

Webserv.language (7221) →  172

MAC Address (7214) →  172

DHCP client (7212) →  172

IP address (7209) →  173

Subnet mask (7211) →  173

Default gateway (7210) →  173

Webserver funct. (7222) →  174

Login page (7273) →  174

Endress+Hauser 171
Description of device parameters Proline Promass 500 Modbus RS485

Webserv.language

Navigation  Expert → Communication → Web server → Webserv.language (7221)

Description Use this function to select the Web server language setting.

Selection • English
• Deutsch
• Français
• Español
• Italiano
• Nederlands
• Portuguesa
• Polski
• русский язык(Ru)
• Svenska
• Türkçe
• 中文 (Chinese)
• 日本語 (Japanese)
• 한국어 (Korean)
• Bahasa Indonesia
• tiếng Việt (Vit)
• čeština (Czech)

Factory setting English

MAC Address

Navigation  Expert → Communication → Web server → MAC Address (7214)

Description Displays the MAC 9) address of the measuring device.

User interface Unique 12-digit character string comprising letters and numbers

Factory setting Each measuring device is given an individual address.

Additional information Example


For the display format
00:07:05:10:01:5F

DHCP client 

Navigation  Expert → Communication → Web server → DHCP client (7212)

Description Use this function to activate and deactivate the DHCP client functionality.

9) Media Access Control

172 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters

Selection • Off
• On

Factory setting Off

Additional information Result


If the DHCP client functionality of the Web server is activated, the IP address (→  173),
Subnet mask (→  173) and Default gateway (→  173) are set automatically.
• Identification is via the MAC address of the measuring device.
• The IP address (→  173) in the IP address parameter (→  173) is ignored as
long as the DHCP client parameter (→  172) is active. This is also the case, in
particular, if the DHCP server cannot be reached. The IP address (→  173) in the
parameter of the same name is only used if the DHCP client parameter (→  172)
is inactive.

IP address 

Navigation  Expert → Communication → Web server → IP address (7209)

Description Display or enter the IP address of the Web server integrated in the measuring device.

User entry 4 octet: 0 to 255 (in the particular octet)

Factory setting 192.168.1.212

Subnet mask 

Navigation  Expert → Communication → Web server → Subnet mask (7211)

Description Display or enter the subnet mask.

User entry 4 octet: 0 to 255 (in the particular octet)

Factory setting 255.255.255.0

Default gateway 

Navigation  Expert → Communication → Web server → Default gateway (7210)

Description Display or enter the Default gateway (→  173).

User entry 4 octet: 0 to 255 (in the particular octet)

Factory setting 0.0.0.0

Endress+Hauser 173
Description of device parameters Proline Promass 500 Modbus RS485

Webserver funct. 

Navigation  Expert → Communication → Web server → Webserver funct. (7222)

Description Use this function to switch the Web server on and off.

Selection • Off
• HTML Off
• On

Factory setting On

Additional information Description


Once disabled, the Webserver funct. can only be re-enabled via or the operating tool
FieldCare.

Selection

Option Description

Off • The web server is completely disabled.


• Port 80 is locked.

On • The complete functionality of the web server is available.


• JavaScript is used.
• The password is transferred in an encrypted state.
• Any change to the password is also transferred in an encrypted state.

Login page 

Navigation  Expert → Communication → Web server → Login page (7273)

Description Use this function to select the format of the login page.

Selection • Without header


• With header

Factory setting With header

3.6.5 "WLAN settings" submenu

Navigation  Expert → Communication → WLAN settings

‣ WLAN settings

WLAN (2702) →  175

WLAN mode (2717) →  176

174 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters

SSID name (2714) →  176

Network security (2705) →  176

Sec. identific. (2718) →  177

User name (2715) →  177

WLAN password (2716) →  177

WLAN IP address (2711) →  177

WLAN MAC address (2703) →  178

WLAN subnet mask (2709) →  178

WLAN MAC address (2703) →  178

WLAN passphrase (2706) →  178

WLAN MAC address (2703) →  178

Assign SSID name (2708) →  179

SSID name (2707) →  179

WLAN channel (2704) →  179

Select antenna (2713) →  180

Connection state (2722) →  180

Rec.sig.strength (2721) →  180

WLAN IP address (2711) →  177

Gateway IP addr. (2719) →  180

IP address DNS (2720) →  181

WLAN 

Navigation  Expert → Communication → WLAN settings → WLAN (2702)

Description Use this function to enable and disable the WLAN connection.

Endress+Hauser 175
Description of device parameters Proline Promass 500 Modbus RS485

Selection • Disable
• Enable

Factory setting Enable

WLAN mode 

Navigation  Expert → Communication → WLAN settings → WLAN mode (2717)

Description Use this function to select the WLAN mode.

Selection • Access point


• WLAN Client

Factory setting Access point

SSID name 

Navigation  Expert → Communication → WLAN settings → SSID name (2714)

Prerequisite The client is activated.

Description Use this function to enter the user-defined SSID name (max. 32 characters) of the WLAN
network.

User entry –

Factory setting –

Network security 

Navigation  Expert → Communication → WLAN settings → Network security (2705)

Description Use this function to select the type of security for the WLAN interface.

Selection • Unsecured
• WPA2-PSK
• EAP-PEAP MSCHAP2 *
• EAP-PEAP NoAuth. *
• EAP-TLS *

Factory setting WPA2-PSK

* Visibility depends on order options or device settings

176 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters

Additional information Selection


• Unsecured
Access the WLAN connection without identification.
• WPA2-PSK
Access the WLAN connection with a network key.

Sec. identific.

Navigation  Expert → Communication → WLAN settings → Sec. identific. (2718)

Description Use this function to select the security settings (download via the menu: Data
Management > Security > Download WLAN).

User interface • Trust. iss.cert.


• Device certific.
• Dev. private key

User name 

Navigation  Expert → Communication → WLAN settings → User name (2715)

Description Use this function to enter the username of the WLAN network.

User entry –

Factory setting –

WLAN password 

Navigation  Expert → Communication → WLAN settings → WLAN password (2716)

Description Use this function to enter the WLAN password for the WLAN network.

User entry –

Factory setting –

WLAN IP address 

Navigation  Expert → Communication → WLAN settings → WLAN IP address (2711)

Description Use this function to enter the IP address of the measuring device's WLAN connection.

Endress+Hauser 177
Description of device parameters Proline Promass 500 Modbus RS485

User entry 4 octet: 0 to 255 (in the particular octet)

Factory setting 192.168.1.212

WLAN MAC address

Navigation  Expert → Communication → WLAN settings → WLAN MAC address (2703)

Description Displays the MAC 10) address of the measuring device.

User interface Unique 12-digit character string comprising letters and numbers

Factory setting Each measuring device is given an individual address.

Additional information Example


For the display format
00:07:05:10:01:5F

WLAN subnet mask 

Navigation  Expert → Communication → WLAN settings → WLAN subnet mask (2709)

Description Use this function to enter the subnet mask.

User entry 4 octet: 0 to 255 (in the particular octet)

Factory setting 255.255.255.0

WLAN passphrase 

Navigation  Expert → Communication → WLAN settings → WLAN passphrase (2706)

Prerequisite The WPA2-PSK option is selected in the Security type parameter (→  176).

Description Use this function to enter the network key.

User entry 8 to 32-digit character string comprising numbers, letters and special characters (without
spaces)

Factory setting Serial number of the measuring device (e.g. L100A802000)

10) Media Access Control

178 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters

Assign SSID name 

Navigation  Expert → Communication → WLAN settings → Assign SSID name (2708)

Description Use this function to select which name is used for the SSID 11).

Selection • Device tag


• User-defined

Factory setting User-defined

Additional information Selection


• Device tag
The device tag name is used as the SSID.
• User-defined
A user-defined name is used as the SSID.

SSID name 

Navigation  Expert → Communication → WLAN settings → SSID name (2707)

Prerequisite • The User-defined option is selected in the Assign SSID name parameter (→  179).
• The Access point option is selected in the WLAN mode parameter (→  176).

Description Use this function to enter a user-defined SSID name.

User entry Max. 32-digit character string comprising numbers, letters and special characters

Factory setting EH_device designation_last 7 digits of the serial number (e.g.


EH_Promass_500_A802000)

WLAN channel 

Navigation  Expert → Communication → WLAN settings → WLAN channel (2704)

Description Use this function to enter the WLAN channel.

User entry 1 to 11

Factory setting 6

Additional information Description


• It is only necessary to enter a WLAN channel if multiple WLAN devices are in use.
• If just one measuring device is in use, it is recommended to keep the factory setting.

11) Service Set Identifier

Endress+Hauser 179
Description of device parameters Proline Promass 500 Modbus RS485

Select antenna 

Navigation  Expert → Communication → WLAN settings → Select antenna (2713)

Description Use this function to select whether the external or internal antenna is used for reception.

Selection • External antenna


• Internal antenna

Factory setting Internal antenna

Connection state

Navigation  Expert → Communication → WLAN settings → Connection state (2722)

Description The connection status is displayed.

User interface • Connected


• Not connected

Factory setting Not connected

Rec.sig.strength

Navigation  Expert → Communication → WLAN settings → Rec.sig.strength (2721)

Description Displays the signal strength received.

User interface • Low


• Medium
• High

Factory setting High

Gateway IP addr.

Navigation  Expert → Communication → WLAN settings → Gateway IP addr. (2719)

Description Use this function to enter the IP address of the gateway.

Factory setting 192.168.1.212

180 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters

IP address DNS

Navigation  Expert → Communication → WLAN settings → IP address DNS (2720)


 Expert → Communication → WLAN settings → IP address DNS (2720)

Description Use this function to enter the IP address of the domain name server.

Factory setting 192.168.1.212

3.7 "Application" submenu


Navigation  Expert → Application

‣ Application

Reset all tot. (2806) →  181

‣ Totalizer 1 to n →  182

‣ Viscosity →  187

‣ Concentration →  187

‣ Custody transfer →  187

‣ Petroleum →  187

‣ Appl.spec. calc. →  187

‣ Medium index →  193

Reset all tot.

Navigation  Expert → Application → Reset all tot. (2806)

Description Use this function to reset all totalizers to the value 0 and restart the totaling process. This
deletes all the flow values previously totalized.

Selection • Cancel
• Reset + totalize

Factory setting Cancel

Endress+Hauser 181
Description of device parameters Proline Promass 500 Modbus RS485

Additional information Selection

Options Description

Cancel No action is executed and the user exits the parameter.

Reset + totalize Resets all totalizers to 0 and restarts the totaling process. This deletes all the flow
values previously totalized.

3.7.1 "Totalizer 1 to n" submenu

Navigation  Expert → Application → Totalizer 1 to n

‣ Totalizer 1 to n

Assign variable (0914–1 to n) →  182

Unit totalizer 1 to n (0915–1 to n) →  183

Operation mode (0908–1 to n) →  184

Control Tot. 1 to n (0912–1 to n) →  185

Preset value 1 to n (0913–1 to n) →  185

Failure mode (0901–1 to n) →  186

Assign variable 

Navigation  Expert → Application → Totalizer 1 to n → Assign variable (0914–1 to n)

Description Use this function to select a process variable for the Totalizer 1 to n.

Selection • Off
• Mass flow
• Volume flow
• Correct.vol.flow *
• Target mass flow *
• Carrier mass fl. *
• Target vol. flow *
• Carrier vol. fl. *
• Targ.corr.vol.fl *
• Carr.corr.vol.fl *
• GSV flow *
• GSVa *
• NSV flow *

* Visibility depends on order options or device settings

182 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters

• NSVa *
• S&W volume flow *
• Oil mass flow *
• Water mass flow *
• Oil volume flow *
• Water vol. flow *
• Oil corr.vol.fl. *
• Water corr.v.fl. *

Factory setting Volume flow

Additional information Description


If the option selected is changed, the device resets the totalizer to 0.

Selection
If the Off option is selected, only Assign variable parameter (→  182) is still displayed
in the Totalizer 1 to n submenu. All other parameters in the submenu are hidden.

Unit totalizer 1 to n 

Navigation  Expert → Application → Totalizer 1 to n → Unit totalizer 1 to n (0915–1 to n)

Prerequisite A process variable is selected in the Assign variable parameter (→  182) of the
Totalizer 1 to n submenu.

Description Use this function to select the process variable unit for the Totalizer 1 to n (→  182).

Selection SI units US units


• g* • oz *
• kg * • lb *
• t* • STon *

* Visibility depends on order options or device settings

or

SI units US units Imperial units


• cm³ * • af * • gal (imp) *
• dm³ * • ft³ * • Mgal (imp) *
• m³ * • Mft³ * • bbl (imp;oil) *
• ml * • fl oz (us) *
• l* • gal (us) *
• hl * • kgal (us) *
• Ml Mega * • Mgal (us) *
• bbl (us;oil) *
• bbl (us;tank) *

* Visibility depends on order options or device settings

or

* Visibility depends on order options or device settings

Endress+Hauser 183
Description of device parameters Proline Promass 500 Modbus RS485

US units Imperial units


• bbl (us;liq.) * bbl (imp;beer) *
• bbl (us;beer) *

* Visibility depends on order options or device settings

or

SI units US units Imperial units


• Nl * • Sft³ * Sgal (imp) *
• Nhl * • MMSft³ *
• Nm³ * • Sgal (us) *
• Sl * • Sbbl (us;liq.) *
• Sm³ * • Sbbl (us;oil) *

* Visibility depends on order options or device settings

or

Other units
None *

* Visibility depends on order options or device settings

Factory setting Country-specific:


• kg
• lb

Additional information Description


The unit is selected separately for each totalizer. It is independent of the selection
made in the System units submenu (→  65).

Selection
The selection is dependent on the process variable selected in the Assign variable
parameter (→  182).

Operation mode 

Navigation  Expert → Application → Totalizer 1 to n → Operation mode (0908–1 to n)

Prerequisite A process variable is selected in the Assign variable parameter (→  182) of the
Totalizer 1 to n submenu.

Description Use this function to select how the totalizer summates the flow.

Selection • Net flow total


• Forward total
• Reverse total

Factory setting Net flow total

184 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters

Additional information Selection


• Net flow total
Flow values in the forward and reverse flow direction are totalized and balanced against
one another. Net flow is registered in the flow direction.
• Forward total
Only the flow in the forward flow direction is totalized.
• Reverse total
Only the flow in the reverse flow direction is totalized (= reverse flow quantity).

Control Tot. 1 to n

Navigation  Expert → Application → Totalizer 1 to n → Control Tot. 1 to n (0912–1 to n)

Prerequisite A process variable is selected in the Assign variable parameter (→  182) of the
Totalizer 1 to n submenu.

Description Use this function to select the control of totalizer value 1-3.

Selection • Totalize
• Reset + hold *
• Preset + hold *
• Reset + totalize
• Preset+totalize *
• Hold *

Factory setting Totalize

Additional information Selection

Options Description

Totalize The totalizer is started or continues running.

Reset + hold The totaling process is stopped and the totalizer is reset to 0.

Preset + hold The totaling process is stopped and the totalizer is set to its defined start value
from the Preset value parameter.

Reset + totalize The totalizer is reset to 0 and the totaling process is restarted.

Preset+totalize The totalizer is set to the defined start value from the Preset value parameter and
the totaling process is restarted.

Hold Totalizing is stopped.

Preset value 1 to n

Navigation  Expert → Application → Totalizer 1 to n → Preset value 1 to n (0913–1 to n)

Prerequisite A process variable is selected in the Assign variable parameter (→  182) of the
Totalizer 1 to n submenu.

* Visibility depends on order options or device settings

Endress+Hauser 185
Description of device parameters Proline Promass 500 Modbus RS485

Description Use this function to enter a start value for the Totalizer 1 to n.

User entry Signed floating-point number

Factory setting Country-specific:


• 0 kg
• 0 lb

Additional information User entry


The unit of the selected process variable is specified for the totalizer in the Unit
totalizer parameter (→  183).

Example
This configuration is suitable for applications such as iterative filling processes with a fixed
batch quantity.

Failure mode 

Navigation  Expert → Application → Totalizer 1 to n → Failure mode (0901–1 to n)

Prerequisite A process variable is selected in the Assign variable parameter (→  182) of the
Totalizer 1 to n submenu.

Description Use this function to select how a totalizer behaves in the event of a device alarm.

Selection • Stop
• Actual value
• Last valid value

Factory setting Stop

Additional information Description


This setting does not affect the failsafe mode of other totalizers and the outputs. This
is specified in separate parameters.

Selection
• Stop
The totalizer is stopped in the event of a device alarm.
• Actual value
The totalizer continues to count based on the actual measured value; the device alarm is
ignored.
• Last valid value
The totalizer continues to count based on the last valid measured value before the device
alarm occurred.

186 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters

3.7.2 "Viscosity" submenu


Only available for Promass I.

For detailed information on the parameter descriptions for the Viscosity application
package,refer to the Special Documentation for the device →  8

Navigation  Expert → Application → Viscosity

‣ Viscosity

3.7.3 "Concentration" submenu


For detailed information on the parameter descriptions for the Concentration
application package,refer to the Special Documentation for the device →  8

Navigation  Expert → Application → Concentration

‣ Concentration

3.7.4 "Custody transfer" submenu


Only available for Promass F, O, Q and X.

For detailed information on the parameter descriptions for custody transfer


measurement, see the Special Documentation for the device →  8

Navigation  Expert → Application → Custody transfer

‣ Custody transfer

3.7.5 "Petroleum" submenu


For detailed information on the parameter descriptions for the Petroleum application
package,refer to the Special Documentation for the device →  8

Navigation  Expert → Application → Petroleum

‣ Petroleum

3.7.6 "Appl.spec. calc." submenu

Navigation  Expert → Application → Appl.spec. calc.

‣ Appl.spec. calc.

‣ Appl.spec.param. →  188

‣ Process variab. →  191

Endress+Hauser 187
Description of device parameters Proline Promass 500 Modbus RS485

"Appl.spec.param." submenu

Navigation  Expert → Application → Appl.spec. calc. → Appl.spec.param.

‣ Appl.spec.param.

Parameter 0 (6358) →  188

Parameter 1 (6359) →  188

Parameter 2 (6360) →  189

Parameter 3 (6361) →  189

Parameter 4 (6345) →  189

Parameter 5 (6346) →  189

Parameter 6 (6347) →  190

Parameter 7 (6348) →  190

Parameter 8 (6349) →  190

Parameter 9 (6350) →  190

Parameter 0 

Navigation  Expert → Application → Appl.spec. calc. → Appl.spec.param. → Parameter 0 (6358)

Description Enter application specific value 0 for application specific calculation.

User entry Signed floating-point number

Factory setting 0

Parameter 1 

Navigation  Expert → Application → Appl.spec. calc. → Appl.spec.param. → Parameter 1 (6359)

Description Enter application specific value 1 for application specific calculation.

User entry Signed floating-point number

Factory setting 0

188 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters

Parameter 2 

Navigation  Expert → Application → Appl.spec. calc. → Appl.spec.param. → Parameter 2 (6360)

Description nter application specific value 2 for application specific calculation.

User entry Signed floating-point number

Factory setting 0

Parameter 3 

Navigation  Expert → Application → Appl.spec. calc. → Appl.spec.param. → Parameter 3 (6361)

Description Enter application specific value 3 for application specific calculation.

User entry Signed floating-point number

Factory setting 0

Parameter 4 

Navigation  Expert → Application → Appl.spec. calc. → Appl.spec.param. → Parameter 4 (6345)

Description Enter application specific value 4 for application specific calculation.

User entry Signed floating-point number

Factory setting 0

Parameter 5 

Navigation  Expert → Application → Appl.spec. calc. → Appl.spec.param. → Parameter 5 (6346)

Description Enter application specific value 5 for application specific calculation.

User entry Signed floating-point number

Factory setting 0

Endress+Hauser 189
Description of device parameters Proline Promass 500 Modbus RS485

Parameter 6 

Navigation  Expert → Application → Appl.spec. calc. → Appl.spec.param. → Parameter 6 (6347)

Description Enter application specific value 6 for application specific calculation.

User entry Signed floating-point number

Factory setting 0

Parameter 7 

Navigation  Expert → Application → Appl.spec. calc. → Appl.spec.param. → Parameter 7 (6348)

Description Enter application specific value 7 for application specific calculation.

User entry Signed floating-point number

Factory setting 0

Parameter 8 

Navigation  Expert → Application → Appl.spec. calc. → Appl.spec.param. → Parameter 8 (6349)

Description Enter application specific value 8 for application specific calculation.

User entry Signed floating-point number

Factory setting 0

Parameter 9 

Navigation  Expert → Application → Appl.spec. calc. → Appl.spec.param. → Parameter 9 (6350)

Description Enter application specific value 9 for application specific calculation.

User entry Signed floating-point number

Factory setting 0

190 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters

"Process variab." submenu

Navigation  Expert → Application → Appl.spec. calc. → Process variab.

‣ Process variab.

Spec. input 0 (6366) →  191

Spec. input 1 (6367) →  192

Spec. output 0 (6364) →  192

Spec. output 1 (6365) →  193

Spec. input 0

Navigation  Expert → Application → Appl.spec. calc. → Process variab. → Spec. input 0 (6366)

Description Shows the application specific input value 0 used for the application specific calculation.

User interface Signed floating-point number

Factory setting 0

FSTypeAppSpec 0

Navigation  Expert → Application → Appl.spec. calc. → Process variab. → FSTypeAppSpec 0


(2098)

Description Use this function to select the failsafe mode for the application-specific input value 0.

Selection • Fail-safe value


• Fallback value
• Off

Factory setting Off

FSValueAppSpec 0 

Navigation  Expert → Application → Appl.spec. calc. → Process variab. → FSValueAppSpec 0


(2099)

Description Use this function to enter the failsafe value for the application-specific input value 0.

Endress+Hauser 191
Description of device parameters Proline Promass 500 Modbus RS485

User entry Signed floating-point number

Factory setting 0

Spec. input 1

Navigation  Expert → Application → Appl.spec. calc. → Process variab. → Spec. input 1 (6367)

Description Shows the application specific input value 1 used for the application specific calculation.

User interface Signed floating-point number

Factory setting 0

FSTypeAppSpec 1

Navigation  Expert → Application → Appl.spec. calc. → Process variab. → FSTypeAppSpec 1


(2100)

Description Use this function to select the failsafe mode for the application-specific input value 1.

Selection • Fail-safe value


• Fallback value
• Off

Factory setting Off

FSValueAppSpec 1 

Navigation  Expert → Application → Appl.spec. calc. → Process variab. → FSValueAppSpec 1


(65535)

Description Use this function to enter the failsafe value for the application-specific input value 1.

User entry Signed floating-point number

Factory setting 0

Spec. output 0

Navigation  Expert → Application → Appl.spec. calc. → Process variab. → Spec. output 0 (6364)

Description Shows the calculated application specific output value 0.

192 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters

User interface Signed floating-point number

Factory setting 0

Spec. output 1

Navigation  Expert → Application → Appl.spec. calc. → Process variab. → Spec. output 1 (6365)

Description Shows the calculated specific output value 1.

User interface Signed floating-point number

Factory setting 0

3.7.7 "Medium index" submenu

Navigation  Expert → Application → Medium index

‣ Medium index

Index inh.medium (6368) →  194

Cut off inh. gas (6375) →  193

Cut off liquid (6374) →  194

Index sus.bubble (6376) →  194

Cut off bubbles (6370) →  194

Cut off inh. gas 

Navigation  Expert → Application → Medium index → Cut off inh. gas (6375)

Description Enter cut off value for wet gas applications. Below this value the Index inhomogeneous
medium is set to 0.

User entry Positive floating-point number

Factory setting 0.25

Endress+Hauser 193
Description of device parameters Proline Promass 500 Modbus RS485

Cut off liquid 

Navigation  Expert → Application → Medium index → Cut off liquid (6374)

Description Enter cut off value for liquid applications. Below this value the Index inhomogeneous
medium is set to 0.

User entry Positive floating-point number

Factory setting 0.05

Index inh.medium

Navigation  Expert → Application → Medium index → Index inh.medium (6368)

Description Shows the degree of inhomogeneity of the medium.

User interface Signed floating-point number

Factory setting 2

Cut off bubbles 

Navigation  Expert → Application → Medium index → Cut off bubbles (6370)

Description Enter cut off value for suspended bubbles. Below this value the Index suspended bubbles is
set to 0.

User entry Positive floating-point number

Factory setting 0.05

Index sus.bubble

Navigation  Expert → Application → Medium index → Index sus.bubble (6376)

Description Shows the relative amount of suspended bubbles in the medium.

User interface Signed floating-point number

Factory setting 0

194 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters

3.8 "Diagnostics" submenu


Navigation  Expert → Diagnostics

‣ Diagnostics

Actual diagnos. (0691) →  196

Prev.diagnostics (0690) →  196

Time fr. restart (0653) →  197

Operating time (0652) →  197

‣ Diagnostic list →  198

‣ Event logbook →  202

‣ Cust.transf.log. →  204

‣ Device info →  205

‣ Mainboard module →  208

‣ Sens. electronic →  209

‣ I/O module 1 →  210

‣ I/O module 2 →  211

‣ I/O module 3 →  213

‣ I/O module 4 →  214

‣ Display module →  215

‣ Min/max val. →  225

‣ Data logging →  216

‣ Heartbeat →  237

‣ Simulation →  237

Endress+Hauser 195
Description of device parameters Proline Promass 500 Modbus RS485

Actual diagnos.

Navigation  Expert → Diagnostics → Actual diagnos. (0691)

Prerequisite A diagnostic event has occurred.

Description Displays the current diagnostic message. If two or more messages occur simultaneously,
the message with the highest priority is shown on the display.

User interface Symbol for diagnostic behavior, diagnostic code and short message.

Additional information Display


Additional pending diagnostic messages can be viewed in the Diagnostic list
submenu (→  198).
Via the local display: the time stamp and corrective measures referring to the cause of
the diagnostic message can be accessed via the  key.

Example
For the display format:
F271 Main electronics

Timestamp

Navigation  Expert → Diagnostics → Timestamp

Description Displays the operating time when the current diagnostic message occurred.

User interface Days (d), hours (h), minutes (m) and seconds (s)

Additional information Display


The diagnostic message can be viewed via the Actual diagnos. parameter
(→  196).

Example
For the display format:
24d12h13m00s

Prev.diagnostics

Navigation  Expert → Diagnostics → Prev.diagnostics (0690)

Prerequisite Two diagnostic events have already occurred.

Description Displays the diagnostic message that occurred before the current message.

User interface Symbol for diagnostic behavior, diagnostic code and short message.

196 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters

Additional information Display


Via the local display: the time stamp and corrective measures referring to the cause of
the diagnostic message can be accessed via the  key.

Example
For the display format:
F271 Main electronics

Timestamp

Navigation  Expert → Diagnostics → Timestamp

Description Displays the operating time when the last diagnostic message before the current message
occurred.

User interface Days (d), hours (h), minutes (m) and seconds (s)

Additional information Display


The diagnostic message can be viewed via the Prev.diagnostics parameter
(→  196).

Example
For the display format:
24d12h13m00s

Time fr. restart

Navigation  Expert → Diagnostics → Time fr. restart (0653)

Description Use this function to display the time the device has been in operation since the last device
restart.

User interface Days (d), hours (h), minutes (m) and seconds (s)

Operating time

Navigation  Expert → Diagnostics → Operating time (0652)

Description Use this function to display the length of time the device has been in operation.

User interface Days (d), hours (h), minutes (m) and seconds (s)

Endress+Hauser 197
Description of device parameters Proline Promass 500 Modbus RS485

Additional information User interface


The maximum number of days is 9999, which is equivalent to 27 years.

3.8.1 "Diagnostic list" submenu

Navigation  Expert → Diagnostics → Diagnostic list

‣ Diagnostic list

Diagnostics 1 (0692) →  198

Diagnostics 2 (0693) →  199

Diagnostics 3 (0694) →  200

Diagnostics 4 (0695) →  200

Diagnostics 5 (0696) →  201

Diagnostics 1

Navigation  Expert → Diagnostics → Diagnostic list → Diagnostics 1 (0692)

Description Displays the current diagnostics message with the highest priority.

User interface Symbol for diagnostic behavior, diagnostic code and short message.

Additional information Display


Via the local display: the time stamp and corrective measures referring to the cause of
the diagnostic message can be accessed via the  key.

Examples
For the display format:
• F271 Main electronics
• F276 I/O module

Timestamp

Navigation  Expert → Diagnostics → Diagnostic list → Timestamp

Description Displays the operating time when the diagnostic message with the highest priority
occurred.

198 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters

User interface Days (d), hours (h), minutes (m) and seconds (s)

Additional information Display


The diagnostic message can be viewed via the Diagnostics 1 parameter (→  198).

Example
For the display format:
24d12h13m00s

Diagnostics 2

Navigation  Expert → Diagnostics → Diagnostic list → Diagnostics 2 (0693)

Description Displays the current diagnostics message with the second-highest priority.

User interface Symbol for diagnostic behavior, diagnostic code and short message.

Additional information Display


Via the local display: the time stamp and corrective measures referring to the cause of
the diagnostic message can be accessed via the  key.

Examples
For the display format:
• F271 Main electronics
• F276 I/O module

Timestamp

Navigation  Expert → Diagnostics → Diagnostic list → Timestamp

Description Displays the operating time when the diagnostic message with the second-highest priority
occurred.

User interface Days (d), hours (h), minutes (m) and seconds (s)

Additional information Display


The diagnostic message can be viewed via the Diagnostics 2 parameter (→  199).

Example
For the display format:
24d12h13m00s

Endress+Hauser 199
Description of device parameters Proline Promass 500 Modbus RS485

Diagnostics 3

Navigation  Expert → Diagnostics → Diagnostic list → Diagnostics 3 (0694)

Description Displays the current diagnostics message with the third-highest priority.

User interface Symbol for diagnostic behavior, diagnostic code and short message.

Additional information Display


Via the local display: the time stamp and corrective measures referring to the cause of
the diagnostic message can be accessed via the  key.

Examples
For the display format:
• F271 Main electronics
• F276 I/O module

Timestamp

Navigation  Expert → Diagnostics → Diagnostic list → Timestamp

Description Displays the operating time when the diagnostic message with the third-highest priority
occurred.

User interface Days (d), hours (h), minutes (m) and seconds (s)

Additional information Display


The diagnostic message can be viewed via the Diagnostics 3 parameter (→  200).

Example
For the display format:
24d12h13m00s

Diagnostics 4

Navigation  Expert → Diagnostics → Diagnostic list → Diagnostics 4 (0695)

Description Displays the current diagnostics message with the fourth-highest priority.

User interface Symbol for diagnostic behavior, diagnostic code and short message.

200 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters

Additional information Display


Via the local display: the time stamp and corrective measures referring to the cause of
the diagnostic message can be accessed via the  key.

Examples
For the display format:
• F271 Main electronics
• F276 I/O module

Timestamp

Navigation  Expert → Diagnostics → Diagnostic list → Timestamp

Description Displays the operating time when the diagnostic message with the fourth-highest priority
occurred.

User interface Days (d), hours (h), minutes (m) and seconds (s)

Additional information Display


The diagnostic message can be viewed via the Diagnostics 4 parameter (→  200).

Example
For the display format:
24d12h13m00s

Diagnostics 5

Navigation  Expert → Diagnostics → Diagnostic list → Diagnostics 5 (0696)

Description Displays the current diagnostics message with the fifth-highest priority.

User interface Symbol for diagnostic behavior, diagnostic code and short message.

Additional information Display


Via the local display: the time stamp and corrective measures referring to the cause of
the diagnostic message can be accessed via the  key.

Examples
For the display format:
• F271 Main electronics
• F276 I/O module

Endress+Hauser 201
Description of device parameters Proline Promass 500 Modbus RS485

Timestamp

Navigation  Expert → Diagnostics → Diagnostic list → Timestamp

Description Displays the operating time when the diagnostic message with the fifth-highest priority
occurred.

User interface Days (d), hours (h), minutes (m) and seconds (s)

Additional information Display


The diagnostic message can be viewed via the Diagnostics 5 parameter (→  201).

Example
For the display format:
24d12h13m00s

3.8.2 "Event logbook" submenu

Navigation  Expert → Diagnostics → Event logbook

‣ Event logbook

Filter options (0705) →  202

‣ Event list →  203

Filter options 

Navigation  Expert → Diagnostics → Event logbook → Filter options (0705)

Description Use this function to select the category whose event messages are displayed in the event
list of the local display.

Selection • All
• Failure (F)
• Funct. check (C)
• Out of spec. (S)
• Mainten. req.(M)
• Information (I)

Factory setting All

202 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters

Additional information Description


The status signals are categorized in accordance with VDI/VDE 2650 and NAMUR
Recommendation NE 107:
• F = Failure
• C = Function Check
• S = Out of Specification
• M = Maintenance Required

Filter options 

Navigation  Expert → Diagnostics → Event logbook → Filter options

Description Use this function to select the category whose event messages are displayed in the event
list of the operating tool.

Selection • All
• Failure (F)
• Funct. check (C)
• Out of spec. (S)
• Mainten. req.(M)
• Information (I)

Factory setting All

Additional information Description


The status signals are categorized in accordance with VDI/VDE 2650 and NAMUR
Recommendation NE 107:
• F = Failure
• C = Function Check
• S = Out of Specification
• M = Maintenance Required

"Event list" submenu


The Event list submenu is only displayed if operating via the local display.
If operating via the FieldCare operating tool, the event list can be read out with a
separate FieldCare module.
If operating via the Web browser, the event messages can be found directly in the
Event logbook submenu.

Navigation  Expert → Diagnostics → Event logbook → Event list

▸ Event list

Event list →  204

Endress+Hauser 203
Description of device parameters Proline Promass 500 Modbus RS485

Event list

Navigation  Expert → Diagnostics → Event logbook → Event list

Description Displays the history of event messages of the category selected in the Filter options
parameter (→  202).

User interface • For a "Category I" event message


Information event, short message, symbol for event recording and operating time when
error occurred
• For a "Category F, C, S, M" event message (status signal)
Diagnostics code, short message, symbol for event recording and operating time when
error occurred

Additional information Description


A maximum of 20 event messages are displayed in chronological order.
If the Extended HistoROM application package (order option) is enabled in the device, the
event list can contain up to 100 entries .
The following symbols indicate whether an event has occurred or has ended:
• : Occurrence of the event
• : End of the event

Examples
For the display format:
• I1091 Configuration modified
 24d12h13m00s
• F271 Main electronics
 01d04h12min30s

HistoROM
A HistoROM is a "non-volatile" device memory in the form of an EEPROM.

3.8.3 "Custody transfer logbook" submenu


Only available for Promass F, O, Q and X.

For detailed information on the parameter descriptions for custody transfer


measurement, see the Special Documentation for the device →  8

Navigation  Expert → Diagnostics → Cust.transf.log.

‣ Cust.transf.log.

204 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters

3.8.4 "Device information" submenu

Navigation  Expert → Diagnostics → Device info

‣ Device info

Device tag →  205

Serial number →  206

Firmware version →  206

Device name →  206

Order code →  206

Ext. order cd. 1 →  207

Ext. order cd. 2 →  207

Ext. order cd. 3 →  207

ENP version →  208

Device tag

Navigation  Expert → Diagnostics → Device info → Device tag (0011)

Description Displays a unique name for the measuring point so it can be identified quickly within the
plant. The name is displayed in the header.

User interface Max. 32 characters, such as letters, numbers or special characters (e.g. @, %, /).

Factory setting Promass

Additional information User interface

1 XXXXXXXXX

A0029422

1 Position of the header text on the display

The number of characters displayed depends on the characters used.

Endress+Hauser 205
Description of device parameters Proline Promass 500 Modbus RS485

Serial number

Navigation  Expert → Diagnostics → Device info → Serial number (0009)

Description Displays the serial number of the measuring device.


The number can be found on the nameplate of the sensor and transmitter.

User interface Max. 11-digit character string comprising letters and numbers.

Additional information Description


Uses of the serial number
• To identify the measuring device quickly, e.g. when contacting Endress+Hauser.
• To obtain specific information on the measuring device using the Device Viewer:
www.endress.com/deviceviewer

Firmware version

Navigation  Expert → Diagnostics → Device info → Firmware version (0010)

Description Displays the device firmware version installed.

User interface Character string in the format xx.yy.zz

Additional information Display


The Firmware version is also located:
• On the title page of the Operating instructions
• On the transmitter nameplate

Device name

Navigation  Expert → Diagnostics → Device info → Device name (0020)

Description Displays the name of the transmitter. It can also be found on the nameplate of the
transmitter.

User interface Promass 300/500

Order code 

Navigation  Expert → Diagnostics → Device info → Order code (0008)

Description Displays the device order code.

User interface Character string composed of letters, numbers and certain punctuation marks (e.g. /).

206 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters

Additional information Description


The order code can be found on the nameplate of the sensor and transmitter in the
"Order code" field.
The order code is generated from the extended order code through a process of reversible
transformation. The extended order code indicates the attributes for all the device features
in the product structure. The device features are not directly readable from the order code.
Uses of the order code
• To order an identical spare device.
• To identify the device quickly and easily, e.g. when contacting Endress+Hauser.

Ext. order cd. 1 

Navigation  Expert → Diagnostics → Device info → Ext. order cd. 1 (0023)

Description Displays the first part of the extended order code.


On account of length restrictions, the extended order code is split into a maximum of 3
parameters.

User interface Character string

Additional information Description


The extended order code indicates the version of all the features of the product structure
for the measuring device and thus uniquely identifies the measuring device.
The extended order code can also be found on the nameplate of the sensor and
transmitter in the "Ext. ord. cd." field.

Ext. order cd. 2 

Navigation  Expert → Diagnostics → Device info → Ext. order cd. 2 (0021)

Description Displays the second part of the extended order code.

User interface Character string

Additional information For additional information, see Ext. order cd. 1 parameter (→  207)

Ext. order cd. 3 

Navigation  Expert → Diagnostics → Device info → Ext. order cd. 3 (0022)

Description Displays the third part of the extended order code.

User interface Character string

Endress+Hauser 207
Description of device parameters Proline Promass 500 Modbus RS485

Additional information For additional information, see Ext. order cd. 1 parameter (→  207)

ENP version

Navigation  Expert → Diagnostics → Device info → ENP version (0012)

Description Displays the version of the electronic nameplate.

User interface Character string

Factory setting 2.02.00

Additional information Description


This electronic nameplate stores a data record for device identification that includes more
data than the nameplates attached to the outside of the device.

3.8.5 "Main elec.+I/O1" submenu

Navigation  Expert → Diagnostics → Main elec.+I/O1

‣ Main elec.+I/O1

Software rev. →  208

Build no. softw. →  209

Bootloader rev. →  209

Software rev.

Navigation  Expert → Diagnostics → Main elec.+I/O1 → Software rev. (0072)

Description Use this function to display the software revision of the module.

User interface Positive integer

208 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters

Build no. softw.

Navigation  Expert → Diagnostics → Main elec.+I/O1 → Build no. softw. (0079)

Description Use this function to display the software build number of the module.

User interface Positive integer

Bootloader rev.

Navigation  Expert → Diagnostics → Main elec.+I/O1 → Bootloader rev. (0073)

Description Use this function to display the bootloader revision of the software.

User interface Positive integer

3.8.6 "Sens. electronic" submenu

Navigation  Expert → Diagnostics → Sens. electronic

‣ Sens. electronic

Software rev. (0072) →  209

Build no. softw. (0079) →  210

Bootloader rev. (0073) →  210

Software rev.

Navigation  Expert → Diagnostics → Sens. electronic → Software rev. (0072)

Description Use this function to display the software revision of the module.

User interface Positive integer

Endress+Hauser 209
Description of device parameters Proline Promass 500 Modbus RS485

Build no. softw.

Navigation  Expert → Diagnostics → Sens. electronic → Build no. softw. (0079)

Description Use this function to display the software build number of the module.

User interface Positive integer

Bootloader rev.

Navigation  Expert → Diagnostics → Sens. electronic → Bootloader rev. (0073)

Description Use this function to display the bootloader revision of the software.

User interface Positive integer

3.8.7 "I/O module 1" submenu

Navigation  Expert → Diagnostics → I/O module 1

‣ I/O module 1

I/O 1 terminals (3902–1) →  210

Software rev. (0072) →  211

Build no. softw. (0079) →  211

Bootloader rev. (0073) →  211

I/O 1 terminals

Navigation  Expert → Diagnostics → I/O module 1 → I/O 1 terminals (3902–1)

Description Displays the terminal numbers used by the I/O module.

210 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters

User interface • Not used


• 26-27 (I/O 1)
• 24-25 (I/O 2)
• 22-23 (I/O 3)
• 20-21 (I/O 4) *

Software rev.

Navigation  Expert → Diagnostics → I/O module → Software rev. (0072)

Description Use this function to display the software revision of the module.

User interface Positive integer

Build no. softw.

Navigation  Expert → Diagnostics → I/O module → Build no. softw. (0079)

Description Use this function to display the software build number of the module.

User interface Positive integer

Bootloader rev.

Navigation  Expert → Diagnostics → I/O module → Bootloader rev. (0073)

Description Use this function to display the bootloader revision of the software.

User interface Positive integer

3.8.8 "I/O module 2" submenu

Navigation  Expert → Diagnostics → I/O module 2

‣ I/O module 2

I/O 2 terminals (3902–2) →  212

Software rev. (0072) →  212

* Visibility depends on order options or device settings

Endress+Hauser 211
Description of device parameters Proline Promass 500 Modbus RS485

Build no. softw. (0079) →  212

Bootloader rev. (0073) →  212

I/O 2 terminals

Navigation  Expert → Diagnostics → I/O module 2 → I/O 2 terminals (3902–2)

Description Displays the terminal numbers used by the I/O module.

User interface • Not used


• 26-27 (I/O 1)
• 24-25 (I/O 2)
• 22-23 (I/O 3)
• 20-21 (I/O 4) *

Software rev.

Navigation  Expert → Diagnostics → I/O module → Software rev. (0072)

Description Use this function to display the software revision of the module.

User interface Positive integer

Build no. softw.

Navigation  Expert → Diagnostics → I/O module → Build no. softw. (0079)

Description Use this function to display the software build number of the module.

User interface Positive integer

Bootloader rev.

Navigation  Expert → Diagnostics → I/O module → Bootloader rev. (0073)

Description Use this function to display the bootloader revision of the software.

* Visibility depends on order options or device settings

212 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters

User interface Positive integer

3.8.9 "I/O module 3" submenu

Navigation  Expert → Diagnostics → I/O module 3

‣ I/O module 3

I/O 3 terminals (3902–3) →  213

Software rev. (0072) →  213

Build no. softw. (0079) →  214

Bootloader rev. (0073) →  214

I/O 3 terminals

Navigation  Expert → Diagnostics → I/O module 3 → I/O 3 terminals (3902–3)

Description Displays the terminal numbers used by the I/O module.

User interface • Not used


• 26-27 (I/O 1)
• 24-25 (I/O 2)
• 22-23 (I/O 3)
• 20-21 (I/O 4) *

Software rev.

Navigation  Expert → Diagnostics → I/O module → Software rev. (0072)

Description Use this function to display the software revision of the module.

User interface Positive integer

* Visibility depends on order options or device settings

Endress+Hauser 213
Description of device parameters Proline Promass 500 Modbus RS485

Build no. softw.

Navigation  Expert → Diagnostics → I/O module → Build no. softw. (0079)

Description Use this function to display the software build number of the module.

User interface Positive integer

Bootloader rev.

Navigation  Expert → Diagnostics → I/O module → Bootloader rev. (0073)

Description Use this function to display the bootloader revision of the software.

User interface Positive integer

3.8.10 "I/O module 4" submenu

Navigation  Expert → Diagnostics → I/O module 4

‣ I/O module 4

I/O 4 terminals (3902–4) →  214

Software rev. (0072) →  215

Build no. softw. (0079) →  215

Bootloader rev. (0073) →  215

I/O 4 terminals

Navigation  Expert → Diagnostics → I/O module 4 → I/O 4 terminals (3902–4)

Description Displays the terminal numbers used by the I/O module.

214 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters

User interface • Not used


• 26-27 (I/O 1)
• 24-25 (I/O 2)
• 22-23 (I/O 3)
• 20-21 (I/O 4) *

Software rev.

Navigation  Expert → Diagnostics → I/O module → Software rev. (0072)

Description Use this function to display the software revision of the module.

User interface Positive integer

Build no. softw.

Navigation  Expert → Diagnostics → I/O module → Build no. softw. (0079)

Description Use this function to display the software build number of the module.

User interface Positive integer

Bootloader rev.

Navigation  Expert → Diagnostics → I/O module → Bootloader rev. (0073)

Description Use this function to display the bootloader revision of the software.

User interface Positive integer

3.8.11 "Display module" submenu

Navigation  Expert → Diagnostics → Display module

‣ Display module

Software rev. (0072) →  216

* Visibility depends on order options or device settings

Endress+Hauser 215
Description of device parameters Proline Promass 500 Modbus RS485

Build no. softw. (0079) →  216

Bootloader rev. (0073) →  216

Software rev.

Navigation  Expert → Diagnostics → Display module → Software rev. (0072)

Description Use this function to display the software revision of the module.

User interface Positive integer

Build no. softw.

Navigation  Expert → Diagnostics → Display module → Build no. softw. (0079)

Description Use this function to display the software build number of the module.

User interface Positive integer

Bootloader rev.

Navigation  Expert → Diagnostics → Display module → Bootloader rev. (0073)

Description Use this function to display the bootloader revision of the software.

User interface Positive integer

3.8.12 "Data logging" submenu

Navigation  Expert → Diagnostics → Data logging

‣ Data logging

Assign chan. 1 (0851) →  217

Assign chan. 2 (0852) →  219

Assign chan. 3 (0853) →  219

216 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters

Assign chan. 4 (0854) →  219

Logging interval (0856) →  220

Clear logging (0855) →  220

Data logging (0860) →  221

Logging delay (0859) →  221

Data log.control (0857) →  222

Data log. status (0858) →  222

Logging duration (0861) →  223

‣ Displ.channel 1 →  223

‣ Displ.channel 2 →  224

‣ Displ.channel 3 →  225

‣ Displ.channel 4 →  225

Assign chan. 1 

Navigation  Expert → Diagnostics → Data logging → Assign chan. 1 (0851)

Prerequisite The Extended HistoROM application package is available.


The software options currently enabled are displayed in the SW option overv.
parameter (→  49).

Description Use this function to select a process variable for the data logging channel.

Selection • Off
• Mass flow
• Volume flow
• Correct.vol.flow *
• Target mass flow *
• Carrier mass fl. *
• Target vol. flow *
• Carrier vol. fl. *
• Targ.corr.vol.fl *
• Carr.corr.vol.fl *
• Density
• Ref.density *
• Ref.dens.altern. *

* Visibility depends on order options or device settings

Endress+Hauser 217
Description of device parameters Proline Promass 500 Modbus RS485

• GSV flow *
• GSVa *
• NSV flow *
• NSVa *
• S&W volume flow *
• Water cut *
• Oil density *
• Water density *
• Oil mass flow *
• Water mass flow *
• Oil volume flow *
• Water vol. flow *
• Oil corr.vol.fl. *
• Water corr.v.fl. *
• Concentration *
• Dynam. viscosity *
• Kinematic visc. *
• TempCompDynVisc *
• TempCompKinVisc *
• Temperature
• Carr. pipe temp. *
• Electronic temp.
• Osc. freq. 0
• Osc. freq. 1 *
• Freq. fluct. 0 *
• Freq. fluct. 1 *
• Oscil. amplitude *
• Osc. ampl. 1 *
• Freq. fluct. 0 *
• Osc. damping 0 *
• Osc. damping 1 *
• Osc.damp.fluct 0 *
• Osc.damp.fluct 1 *
• Signal asymmetry *
• Exc. current 0 *
• Exc. current 1 *
• HBSI *
• Curr.output 1 *
• Curr.output 2 *
• Curr.output 3 *
• Pressure
• Spec. output 1 *
• Index inh.medium
• Spec. output 0 *
• Index sus.bubble *

Factory setting Off

Additional information Description


A total of 1000 measured values can be logged. This means:
• 1000 data points if 1 logging channel is used
• 500 data points if 2 logging channels are used
• 333 data points if 3 logging channels are used
• 250 data points if 4 logging channels are used

* Visibility depends on order options or device settings

218 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters

Once the maximum number of data points is reached, the oldest data points in the data log
are cyclically overwritten in such a way that the last 1000, 500, 333 or 250 measured
values are always in the log (ring memory principle).
The log contents are cleared if the option selected is changed.

Assign chan. 2 

Navigation  Expert → Diagnostics → Data logging → Assign chan. 2 (0852)

Prerequisite The Extended HistoROM application package is available.


The software options currently enabled are displayed in the SW option overv.
parameter (→  49).

Description Options for the assignment of a process variable to the data logging channel.

Selection Picklist, see Assign channel 1 parameter (→  217)

Factory setting Off

Assign chan. 3 

Navigation  Expert → Diagnostics → Data logging → Assign chan. 3 (0853)

Prerequisite The Extended HistoROM application package is available.


The software options currently enabled are displayed in the SW option overv.
parameter (→  49).

Description Options for the assignment of a process variable to the data logging channel.

Selection Picklist, see Assign channel 1 parameter (→  217)

Factory setting Off

Assign chan. 4 

Navigation  Expert → Diagnostics → Data logging → Assign chan. 4 (0854)

Prerequisite The Extended HistoROM application package is available.


The software options currently enabled are displayed in the SW option overv.
parameter (→  49).

Description Options for the assignment of a process variable to the data logging channel.

Selection Picklist, see Assign channel 1 parameter (→  217)

Endress+Hauser 219
Description of device parameters Proline Promass 500 Modbus RS485

Factory setting Off

Logging interval 

Navigation  Expert → Diagnostics → Data logging → Logging interval (0856)

Prerequisite The Extended HistoROM application package is available.


The software options currently enabled are displayed in the SW option overv.
parameter (→  49).

Description Use this function to enter the logging interval Tlog for data logging.

User entry 0.1 to 3 600.0 s

Factory setting 1.0 s

Additional information Description


This defines the interval between the individual data points in the data log, and thus the
maximum loggable process time Tlog:
• If 1 logging channel is used: Tlog = 1000 × tlog
• If 2 logging channels are used: Tlog = 500 × tlog
• If 3 logging channels are used: Tlog = 333 × tlog
• If 4 logging channels are used: Tlog = 250 × tlog
Once this time elapses, the oldest data points in the data log are cyclically overwritten such
that a time of Tlog always remains in the memory (ring memory principle).
The log contents are cleared if the length of the logging interval is changed.

Example
If 1 logging channel is used:
• Tlog = 1000 × 1 s = 1 000 s ≈15 min
• Tlog = 1000 × 10 s = 10 000 s ≈ 3 h
• Tlog = 1000 × 80 s = 80 000 s ≈ 1 d
• Tlog = 1000 × 3 600 s = 3 600 000 s ≈ 41 d

Clear logging 

Navigation  Expert → Diagnostics → Data logging → Clear logging (0855)

Prerequisite The Extended HistoROM application package is available.


The software options currently enabled are displayed in the SW option overv.
parameter (→  49).

Description Use this function to clear the entire logging data.

Selection • Cancel
• Clear data

220 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters

Factory setting Cancel

Additional information Selection


• Cancel
The data is not cleared. All the data is retained.
• Clear data
The logging data is cleared. The logging process starts from the beginning.

Data logging 

Navigation  Expert → Diagnostics → Data logging → Data logging (0860)

Description Use this function to select the data logging method.

Selection • Overwriting
• Not overwriting

Factory setting Overwriting

Additional information Selection


• Overwriting
The device memory applies the FIFO principle.
• Not overwriting
Data logging is canceled if the measured value memory is full (single shot).

Logging delay 

Navigation  Expert → Diagnostics → Data logging → Logging delay (0859)

Prerequisite In the Data logging parameter (→  221), the Not overwriting option is selected.

Description Use this function to enter the time delay for measured value logging.

User entry 0 to 999 h

Factory setting 0h

Additional information Description


Once measured value logging has been started with the Data log.control parameter
(→  222), the device does not save any data for the duration of the time delay entered.

Endress+Hauser 221
Description of device parameters Proline Promass 500 Modbus RS485

Data log.control 

Navigation  Expert → Diagnostics → Data logging → Data log.control (0857)

Prerequisite In the Data logging parameter (→  221), the Not overwriting option is selected.

Description Use this function to start and stop measured value logging.

Selection • None
• Delete + start
• Stop

Factory setting None

Additional information Selection


• None
Initial measured value logging status.
• Delete + start
All the measured values recorded for all the channels are deleted and measured value
logging starts again.
• Stop
Measured value logging is stopped.

Data log. status

Navigation  Expert → Diagnostics → Data logging → Data log. status (0858)

Prerequisite In the Data logging parameter (→  221), the Not overwriting option is selected.

Description Displays the measured value logging status.

User interface • Done


• Delay active
• Active
• Stopped

Factory setting Done

Additional information Selection


• Done
Measured value logging has been performed and completed successfully.
• Delay active
Measured value logging has been started but the logging interval has not yet elapsed.
• Active
The logging interval has elapsed and measured value logging is active.
• Stopped
Measured value logging is stopped.

222 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters

Logging duration

Navigation  Expert → Diagnostics → Data logging → Logging duration (0861)

Prerequisite In the Data logging parameter (→  221), the Not overwriting option is selected.

Description Displays the total logging duration.

User interface Positive floating-point number

Factory setting 0s

"Displ.channel 1" submenu

Navigation  Expert → Diagnostics → Data logging → Displ.channel 1

▸ Displ.channel 1

Display channel 1 →  223

Display channel 1

Navigation  Expert → Diagnostics → Data logging → Displ.channel 1

Prerequisite The Extended HistoROM application package is available.


The software options currently enabled are displayed in the SW option overv.
parameter (→  49).
In the Assign chan. 1 parameter (→  217), one of the following options is selected:
• Mass flow
• Volume flow
• Correct.vol.flow
• Target mass flow *
• Carrier mass fl. *
• Density
• Ref.density
• Concentration *
• Dynam. viscosity *
• Kinematic visc. *
• TempCompDynVisc *
• TempCompKinVisc *
• Carr. pipe temp. *
• Electronic temp.
• Curr.output 1
• Osc. freq. 0
• Osc. freq. 1 *

* Visibility depends on order options or device settings

Endress+Hauser 223
Description of device parameters Proline Promass 500 Modbus RS485

• Freq. fluct. 0
• Freq. fluct. 1 *
• Oscil. amplitude *
• Osc. ampl. 1 *
• Osc. damping 0
• Osc. damping 1 *
• Osc.damp.fluct 0
• Osc.damp.fluct 1 *
• Signal asymmetry
• Exc. current 0
• Exc. current 1 *

Description Displays the measured value trend for the logging channel in the form of a chart.

Additional information Description

/ ../XXXXXXXX
175.77

40.69 kg/h
-100s 0

A0016357

9 Chart of a measured value trend

• x-axis: depending on the number of channels selected displays 250 to 1000 measured
values of a process variable.
• y-axis: displays the approximate measured value span and constantly adapts this to the
ongoing measurement.

"Displ.channel 2" submenu

Navigation  Expert → Diagnostics → Data logging → Displ.channel 2

▸ Displ.channel 2

Display channel 2 →  224

Display channel 2

Navigation  Expert → Diagnostics → Data logging → Displ.channel 2

Prerequisite A process variable is defined in the Assign chan. 2 parameter.

Description See the Display channel 1 parameter →  223

* Visibility depends on order options or device settings

224 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters

"Displ.channel 3" submenu

Navigation  Expert → Diagnostics → Data logging → Displ.channel 3

▸ Displ.channel 3

Display channel 3 →  225

Display channel 3

Navigation  Expert → Diagnostics → Data logging → Displ.channel 3

Prerequisite A process variable is defined in the Assign chan. 3 parameter.

Description See the Display channel 1 parameter →  223

"Displ.channel 4" submenu

Navigation  Expert → Diagnostics → Data logging → Displ.channel 4

▸ Displ.channel 4

Display channel 4 →  225

Display channel 4

Navigation  Expert → Diagnostics → Data logging → Displ.channel 4

Prerequisite A process variable is defined in the Assign chan. 4 parameter.

Description See the Display channel 1 parameter →  223

3.8.13 "Min/max val." submenu

Navigation  Expert → Diagnostics → Min/max val.

‣ Min/max val.

Reset min/max (6151) →  226

Endress+Hauser 225
Description of device parameters Proline Promass 500 Modbus RS485

‣ Main elect.temp. →  227

‣ Sensor elec.temp →  228

‣ Medium temp. →  229

‣ Carr. pipe temp. →  229

‣ Oscil. frequency →  231

‣ Tors.oscil.freq. →  231

‣ Oscil. amplitude →  232

‣ Tor. osc. amp. →  233

‣ Oscil. damping →  234

‣ Tors.oscil.damp. →  234

‣ Signal asymmetry →  235

‣ Tors.sig.asymm. →  236

Reset min/max 

Navigation  Expert → Diagnostics → Min/max val. → Reset min/max (6151)

Description Use this function to select measured variables whose minimum, maximum and average
measured values are to be reset.

Selection • Cancel
• Oscil. amplitude *
• Osc. ampl. 1 *
• Oscil. damping
• Tors.oscil.damp. *
• Oscil. frequency
• Tors.oscil.freq. *
• Signal asymmetry
• Tors.sig.asymm. *

Factory setting Cancel

* Visibility depends on order options or device settings

226 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters

"Main electronic temperature" submenu

Navigation  Expert → Diagnostics → Min/max val. → Main elect.temp.

‣ Main elect.temp.

Minimum value (0688) →  227

Maximum value (0665) →  227

Minimum value

Navigation  Expert → Diagnostics → Min/max val. → Main elect.temp. → Minimum value (0688)

Description Displays the lowest previously measured temperature value of the electronics module in
the transmitter.

User interface Signed floating-point number

Additional information Dependency


The unit is taken from the Temperature unit parameter (→  72)

Maximum value

Navigation  Expert → Diagnostics → Min/max val. → Main elect.temp. → Maximum value


(0665)

Description Displays the highest previously measured temperature value of the electronics module in
the transmitter.

User interface Signed floating-point number

Additional information Dependency


The unit is taken from the Temperature unit parameter (→  72)

Endress+Hauser 227
Description of device parameters Proline Promass 500 Modbus RS485

"Sensor electronic temperature (ISEM)" submenu

Navigation  Expert → Diagnostics → Min/max val. → Sensor elec.temp

‣ Sensor elec.temp

Minimum value (6052) →  228

Maximum value (6051) →  228

Maximum value

Navigation  Expert → Diagnostics → Min/max val. → Sensor elec.temp → Maximum value


(6051)

Description Displays the highest previously measured temperature value of the electronics module in
the sensor connection housing.

User interface Signed floating-point number

Additional information Dependency


The unit is taken from the Temperature unit parameter (→  72)

Minimum value

Navigation  Expert → Diagnostics → Min/max val. → Sensor elec.temp → Minimum value (6052)

Description Displays the lowest previously measured temperature value of the electronics module in
the sensor connection housing.

User interface Signed floating-point number

Additional information Dependency


The unit is taken from the Temperature unit parameter (→  72)

228 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters

"Medium temp." submenu

Navigation  Expert → Diagnostics → Min/max val. → Medium temp.

‣ Medium temp.

Minimum value (6109) →  229

Maximum value (6108) →  229

Minimum value

Navigation  Expert → Diagnostics → Min/max val. → Medium temp. → Minimum value (6109)

Description Displays the lowest previously measured medium temperature value.

User interface Signed floating-point number

Additional information Dependency


The unit is taken from the Temperature unit parameter (→  72)

Maximum value

Navigation  Expert → Diagnostics → Min/max val. → Medium temp. → Maximum value (6108)

Description Displays the highest previously measured medium temperature value.

User interface Signed floating-point number

Additional information Dependency


The unit is taken from the Temperature unit parameter (→  72)

"Carr. pipe temp." submenu

Navigation  Expert → Diagnostics → Min/max val. → Carr. pipe temp.

‣ Carr. pipe temp.

Minimum value (6030) →  230

Maximum value (6029) →  230

Endress+Hauser 229
Description of device parameters Proline Promass 500 Modbus RS485

Minimum value

Navigation  Expert → Diagnostics → Min/max val. → Carr. pipe temp. → Minimum value (6030)

Prerequisite Only available for:


• Promass A
• Promass F
• Promass H
• Promass I
• Promass O
• Promass P
• PromassQ
• Promass S
• Promass X
For the following order code
"Application package", option EB "Heartbeat Verification + Monitoring"

Description Displays the lowest previously measured temperature value of the carrier pipe.

User interface Signed floating-point number

Additional information Dependency


The unit is taken from the Temperature unit parameter (→  72)

Maximum value

Navigation  Expert → Diagnostics → Min/max val. → Carr. pipe temp. → Maximum value (6029)

Prerequisite Only available for:


• Promass A
• Promass F
• Promass H
• Promass I
• Promass O
• Promass P
• PromassQ
• Promass S
• Promass X
For the following order code
"Application package", option EB "Heartbeat Verification + Monitoring"

Description Displays the highest previously measured temperature value of the carrier pipe.

User interface Signed floating-point number

Additional information Dependency


The unit is taken from the Temperature unit parameter (→  72)

230 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters

"Oscil. frequency" submenu

Navigation  Expert → Diagnostics → Min/max val. → Oscil. frequency

‣ Oscil. frequency

Minimum value (6071) →  231

Maximum value (6070) →  231

Minimum value

Navigation  Expert → Diagnostics → Min/max val. → Oscil. frequency → Minimum value (6071)

Description Displays the lowest previously measured oscillation frequency.

User interface Signed floating-point number

Maximum value

Navigation  Expert → Diagnostics → Min/max val. → Oscil. frequency → Maximum value (6070)

Description Displays the highest previously measured oscillation frequency.

User interface Signed floating-point number

"Tors.oscil.freq." submenu

Navigation  Expert → Diagnostics → Min/max val. → Tors.oscil.freq.

‣ Tors.oscil.freq.

Minimum value (6069) →  232

Maximum value (6068) →  232

Endress+Hauser 231
Description of device parameters Proline Promass 500 Modbus RS485

Minimum value

Navigation  Expert → Diagnostics → Min/max val. → Tors.oscil.freq. → Minimum value (6069)

Prerequisite Only available for Promass I.

For the following order code:


"Application package", option EB "Heartbeat Verification + Monitoring"

Description Displays the lowest previously measured torsion oscillation frequency.

User interface Signed floating-point number

Maximum value

Navigation  Expert → Diagnostics → Min/max val. → Tors.oscil.freq. → Maximum value (6068)

Prerequisite Only available for Promass I.

For the following order code:


"Application package", option EB "Heartbeat Verification + Monitoring"

Description Displays the highest previously measured torsion oscillation frequency.

User interface Signed floating-point number

"Oscil. amplitude" submenu

Navigation  Expert → Diagnostics → Min/max val. → Oscil. amplitude

‣ Oscil. amplitude

Minimum value (6010) →  232

Maximum value (6009) →  233

Minimum value

Navigation  Expert → Diagnostics → Min/max val. → Oscil. amplitude → Minimum value (6010)

Description Displays the lowest previously measured oscillation amplitude.

User interface Signed floating-point number

232 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters

Maximum value

Navigation  Expert → Diagnostics → Min/max val. → Oscil. amplitude → Maximum value (6009)

Description Displays the highest previously measured oscillation amplitude.

User interface Signed floating-point number

"Tor. osc. amp." submenu

Navigation  Expert → Diagnostics → Min/max val. → Tor. osc. amp.

‣ Tor. osc. amp.

Minimum value (6008) →  233

Maximum value (6007) →  233

Minimum value

Navigation  Expert → Diagnostics → Min/max val. → Tor. osc. amp. → Minimum value (6008)

Prerequisite Only available for Promass I.

For the following order code:


"Application package", option EB "Heartbeat Verification + Monitoring"

Description Displays the lowest previously measured torsion oscillation amplitude.

User interface Signed floating-point number

Maximum value

Navigation  Expert → Diagnostics → Min/max val. → Tor. osc. amp. → Maximum value (6007)

Prerequisite Only available for Promass I.

For the following order code:


"Application package", option EB "Heartbeat Verification + Monitoring"

Description Displays the highest previously measured torsion oscillation amplitude.

Endress+Hauser 233
Description of device parameters Proline Promass 500 Modbus RS485

User interface Signed floating-point number

"Oscil. damping" submenu

Navigation  Expert → Diagnostics → Min/max val. → Oscil. damping

‣ Oscil. damping

Minimum value (6122) →  234

Maximum value (6121) →  234

Minimum value

Navigation  Expert → Diagnostics → Min/max val. → Oscil. damping → Minimum value (6122)

Description Displays the lowest previously measured oscillation damping.

User interface Signed floating-point number

Maximum value

Navigation  Expert → Diagnostics → Min/max val. → Oscil. damping → Maximum value (6121)

Description Displays the highest previously measured oscillation damping.

User interface Signed floating-point number

"Tors.oscil.damp." submenu

Navigation  Expert → Diagnostics → Min/max val. → Tors.oscil.damp.

‣ Tors.oscil.damp.

Minimum value (6120) →  235

Maximum value (6119) →  235

234 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters

Minimum value

Navigation  Expert → Diagnostics → Min/max val. → Tors.oscil.damp. → Minimum value (6120)

Prerequisite Only available for Promass I.

For the following order code:


"Application package", option EB "Heartbeat Verification + Monitoring"

Description Displays the lowest previously measured torsion oscillation damping.

User interface Signed floating-point number

Maximum value

Navigation  Expert → Diagnostics → Min/max val. → Tors.oscil.damp. → Maximum value (6119)

Prerequisite Only available for Promass I.

For the following order code:


"Application package", option EB "Heartbeat Verification + Monitoring"

Description Displays the highest previously measured torsion oscillation damping.

User interface Signed floating-point number

"Signal asymmetry" submenu

Navigation  Expert → Diagnostics → Min/max val. → Signal asymmetry

‣ Signal asymmetry

Minimum value (6015) →  235

Maximum value (6014) →  236

Minimum value

Navigation  Expert → Diagnostics → Min/max val. → Signal asymmetry → Minimum value


(6015)

Description Displays the lowest previously measured signal asymmetry.

Endress+Hauser 235
Description of device parameters Proline Promass 500 Modbus RS485

User interface Signed floating-point number

Maximum value

Navigation  Expert → Diagnostics → Min/max val. → Signal asymmetry → Maximum value


(6014)

Description Displays the highest previously measured signal asymmetry.

User interface Signed floating-point number

"Torsion signal asymmetry" submenu

Navigation  Expert → Diagnostics → Min/max val. → Tors.sig.asymm.

‣ Tors.sig.asymm.

Minimum value (6284) →  236

Maximum value (6283) →  236

Minimum value

Navigation  Expert → Diagnostics → Min/max val. → Tors.sig.asymm. → Minimum value (6284)

Prerequisite Only available for Promass I.

For the following order code:


"Application package", option EB "Heartbeat Verification + Monitoring"

Description Displays the lowest previously measured torsion signal asymmetry.

User interface Signed floating-point number

Maximum value

Navigation  Expert → Diagnostics → Min/max val. → Tors.sig.asymm. → Maximum value (6283)

Prerequisite Only available for Promass I.

236 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters

For the following order code:


"Application package", option EB "Heartbeat Verification + Monitoring"

Description Displays the highest previously measured torsion signal asymmetry.

User interface Signed floating-point number

3.8.14 "Heartbeat" submenu


For detailed information on the parameter descriptions for the Heartbeat
Verification+Monitoringrefer to the Special Documentation for the device →  8

Navigation  Expert → Diagnostics → Heartbeat

‣ Heartbeat

3.8.15 "Simulation" submenu

Navigation  Expert → Diagnostics → Simulation

‣ Simulation

Assign proc.var. (1810) →  238

Proc. var. value (1811) →  239

Status inp.sim (1355) →  239

Signal level (1356) →  240

Curr.inp 1 to n sim. (1608–1 to n) →  240

Value curr.inp 1 to n (1609–1 to n) →  241

Curr.out. 1 to n sim. (0354–1 to n) →  241

Value curr.out 1 to n (0355–1 to n) →  241

FreqOutputSim 1 to n (0472–1 to n) →  242

Freq value 1 to n (0473–1 to n) →  242

Puls.outp.sim. 1 to n (0458–1 to n) →  243

Pulse value 1 to n (0459–1 to n) →  243

Switch sim. 1 to n (0462–1 to n) →  243

Endress+Hauser 237
Description of device parameters Proline Promass 500 Modbus RS485

Switch status 1 to n (0463–1 to n) →  244

Relay out. 1 to n sim (0802–1 to n) →  244

Switch status 1 to n (0803–1 to n) →  245

Puls.outp.sim. (0988) →  245

Pulse value (0989) →  246

Dev. alarm sim. (0654) →  246

Event category (0738) →  247

Diag. event sim. (0737) →  247

Assign proc.var. 

Navigation  Expert → Diagnostics → Simulation → Assign proc.var. (1810)

Description Use this function to select a process variable for the simulation process that is activated.
The display alternates between the measured value and a diagnostic message of the
"Function check" category (C) while simulation is in progress.

Selection • Off
• Mass flow
• Volume flow
• Correct.vol.flow *
• Target vol. flow *
• Carrier vol. fl. *
• Targ.corr.vol.fl *
• Carr.corr.vol.fl *
• Density
• Ref.density *
• Ref.dens.altern. *
• GSV flow *
• GSVa *
• NSV flow *
• NSVa *
• S&W volume flow *
• Water cut *
• Oil density *
• Water density *
• Oil mass flow *
• Water mass flow *
• Oil volume flow *
• Water vol. flow *
• Oil corr.vol.fl. *
• Water corr.v.fl. *

* Visibility depends on order options or device settings

238 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters

• Temperature
• Dynam. viscosity *
• Kinematic visc. *
• TempCompDynVisc *
• TempCompKinVisc *
• Concentration *
• Target mass flow *
• Carrier mass fl. *

Factory setting Off

Additional information Description


The simulation value of the process variable selected is defined in the Proc. var. value
parameter (→  239).

Proc. var. value 

Navigation  Expert → Diagnostics → Simulation → Proc. var. value (1811)

Prerequisite A process variable is selected in the Assign proc.var. parameter (→  238).

Description Use this function to enter a simulation value for the selected process variable. Subsequent
measured value processing and the signal output use this simulation value. In this way,
users can verify whether the measuring device has been configured correctly.

User entry Depends on the process variable selected

Factory setting 0

Additional information User entry


The unit of the displayed measured value is taken from the System units submenu
(→  65).

Status inp.sim 1 to n 

Navigation  Expert → Diagnostics → Simulation → Status inp.sim 1 to n (1355–1 to n)

Description Use this function to switch simulation of the status input on and off. The display alternates
between the measured value and a diagnostic message of the "Function check" category (C)
while simulation is in progress.

Selection • Off
• On

Factory setting Off

* Visibility depends on order options or device settings

Endress+Hauser 239
Description of device parameters Proline Promass 500 Modbus RS485

Additional information Description


The desired simulation value is defined in the Signal level parameter (→  240).

Selection
• Off
Simulation for the status input is switched off. The device is in normal measuring mode
or another process variable is being simulated.
• On
Simulation for the status input is active.

Signal level 1 to n 

Navigation  Expert → Diagnostics → Simulation → Signal level 1 to n (1356–1 to n)

Prerequisite In the Status inp.sim parameter (→  239), the On option is selected.

Description Use this function to select the signal level for the simulation of the status input. In this
way, users can verify the correct configuration of the status input and the correct function
of upstream feed-in units.

Selection • High
• Low

Curr.inp 1 to n sim. 

Navigation  Expert → Diagnostics → Simulation → Curr.inp 1 to n sim. (1608–1 to n)

Description Option for switching simulation of the current input on and off. The display alternates
between the measured value and a diagnostic message of the "Function check" category (C)
while simulation is in progress.
The desired simulation value is defined in the Value curr.inp 1 to n parameter.

Selection • Off
• On

Factory setting Off

Additional information Selection


• Off
Current simulation is switched off. The device is in normal measuring mode or another
process variable is being simulated.
• On
Current simulation is active.

240 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters

Value curr.inp 1 to n 

Navigation  Expert → Diagnostics → Simulation → Value curr.inp 1 to n (1609–1 to n)

Prerequisite In the Curr.inp 1 to n sim. parameter, the On option is selected.

Description Use this function to enter the current value for the simulation. In this way, users can verify
the correct configuration of the current input and the correct function of upstream feed-in
units.

User entry 0 to 22.5 mA

Curr.out. 1 to n sim. 

Navigation  Expert → Diagnostics → Simulation → Curr.out. 1 to n sim. (0354–1 to n)

Description Use this function to switch simulation of the current output on and off. The display
alternates between the measured value and a diagnostic message of the "Function check"
category (C) while simulation is in progress.

Selection • Off
• On

Factory setting Off

Additional information Description


The desired simulation value is defined in the Value curr.out 1 to n parameter.

Selection
• Off
Current simulation is switched off. The device is in normal measuring mode or another
process variable is being simulated.
• On
Current simulation is active.

Value curr.out 1 to n 

Navigation  Expert → Diagnostics → Simulation → Value curr.out 1 to n (0355–1 to n)

Prerequisite In the Curr.out. 1 to n sim. parameter, the On option is selected.

Description Use this function to enter a current value for the simulation. In this way, users can verify
the correct adjustment of the current output and the correct function of downstream
switching units.

User entry 3.59 to 22.5 mA

Endress+Hauser 241
Description of device parameters Proline Promass 500 Modbus RS485

Additional information Dependency


The input range is dependent on the option selected in the Current span parameter
(→  118).

FreqOutputSim 1 to n 

Navigation  Expert → Diagnostics → Simulation → FreqOutputSim 1 to n (0472–1 to n)

Prerequisite In the Operating mode parameter (→  132), the Frequency option is selected.

Description Use this function to switch simulation of the frequency output on and off. The display
alternates between the measured value and a diagnostic message of the "Function check"
category (C) while simulation is in progress.

Selection • Off
• On

Factory setting Off

Additional information Description


The desired simulation value is defined in the Freq value 1 to n parameter.

Selection
• Off
Frequency simulation is switched off. The device is in normal measuring mode or
another process variable is being simulated.
• On
Frequency simulation is active.

Freq value 1 to n 

Navigation  Expert → Diagnostics → Simulation → Freq value 1 to n (0473–1 to n)

Prerequisite In the FreqOutputSim 1 to n parameter, the On option is selected.

Description Use this function to enter a frequency value for the simulation. In this way, users can verify
the correct adjustment of the frequency output and the correct function of downstream
switching units.

User entry 0.0 to 12 500.0 Hz

242 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters

Puls.outp.sim. 1 to n 

Navigation  Expert → Diagnostics → Simulation → Puls.outp.sim. 1 to n (0458–1 to n)

Prerequisite In the Operating mode parameter (→  132), the Pulse option is selected.

Description Use this function to switch simulation of the pulse output on and off. The display
alternates between the measured value and a diagnostic message of the "Function check"
category (C) while simulation is in progress.

Selection • Off
• Fixed value
• Down-count. val.

Factory setting Off

Additional information Description


The desired simulation value is defined in the Pulse value 1 to n parameter.

Selection
• Off
Pulse simulation is switched off. The device is in normal measuring mode or another
process variable is being simulated.
• Fixed value
Pulses are continuously output with the pulse width specified in the Pulse width
parameter (→  135).
• Down-count. val.
The pulses specified in the Pulse value parameter (→  243) are output.

Pulse value 1 to n 

Navigation  Expert → Diagnostics → Simulation → Pulse value 1 to n (0459–1 to n)

Prerequisite In the Puls.outp.sim. 1 to n parameter, the Down-count. val. option is selected.

Description Use this function to enter a pulse value for the simulation. In this way, users can verify the
correct adjustment of the pulse output and the correct function of downstream switching
units.

User entry 0 to 65 535

Switch sim. 1 to n 

Navigation  Expert → Diagnostics → Simulation → Switch sim. 1 to n (0462–1 to n)

Prerequisite In the Operating mode parameter (→  132), the Switch option is selected.

Endress+Hauser 243
Description of device parameters Proline Promass 500 Modbus RS485

Description Use this function to switch simulation of the switch output on and off. The display
alternates between the measured value and a diagnostic message of the "Function check"
category (C) while simulation is in progress.

Selection • Off
• On

Factory setting Off

Additional information Description


The desired simulation value is defined in the Switch status 1 to n parameter.

Selection
• Off
Switch simulation is switched off. The device is in normal measuring mode or another
process variable is being simulated.
• On
Switch simulation is active.

Switch status 1 to n 

Navigation  Expert → Diagnostics → Simulation → Switch status 1 to n (0463–1 to n)

Description Use this function to select a switch value for the simulation. In this way, users can verify
the correct adjustment of the switch output and the correct function of downstream
switching units.

Selection • Open
• Closed

Additional information Selection


• Open
Switch simulation is switched off. The device is in normal measuring mode or another
process variable is being simulated.
• Closed
Switch simulation is active.

Relay out. 1 to n sim 

Navigation  Expert → Diagnostics → Simulation → Relay out. 1 to n sim (0802–1 to n)

Description Use this function to switch simulation of the relay output on and off. The display
alternates between the measured value and a diagnostic message of the "Function check"
category (C) while simulation is in progress.

Selection • Off
• On

Factory setting Off

244 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters

Additional information Description


The desired simulation value is defined in the Switch status 1 to n parameter.

Selection
• Off
Relay simulation is switched off. The device is in normal measuring mode or another
process variable is being simulated.
• On
Relay simulation is active.

Switch status 1 to n 

Navigation  Expert → Diagnostics → Simulation → Switch status 1 to n (0803–1 to n)

Prerequisite The On option is selected in the Switch sim. 1 to n parameter parameter.

Description Use this function to select a relay value for the simulation. In this way, users can verify the
correct adjustment of the relay output and the correct function of downstream switching
units.

Selection • Open
• Closed

Additional information Selection


• Open
Relay simulation is switched off. The device is in normal measuring mode or another
process variable is being simulated.
• Closed
Relay simulation is active.

Puls.outp.sim. 

Navigation  Expert → Diagnostics → Simulation → Puls.outp.sim. (0988)

Description Use this function to switch simulation of the double pulse output on and off. The display
alternates between the measured value and a diagnostic message of the "Function check"
category (C) while simulation is in progress.

Selection • Off
• Fixed value
• Down-count. val.

Factory setting Off

Endress+Hauser 245
Description of device parameters Proline Promass 500 Modbus RS485

Additional information Description


The desired simulation value is defined in the Pulse value parameter (→  246).

Selection
• Off
Simulation of the double pulse output is switched off. The device is in normal measuring
mode or another process variable is being simulated.
• Fixed value
Pulses are continuously output with the pulse width specified in the Pulse width
parameter (→  162).
• Down-count. val.
The pulses specified in the Pulse value parameter (→  246) are output.

Pulse value 

Navigation  Expert → Diagnostics → Simulation → Pulse value (0989)

Prerequisite In the Puls.outp.sim. parameter (→  245), the Down-count. val. option is selected.

Description Use this function to enter a pulse value for simulation of the double pulse output. In this
way, users can verify the correct adjustment of the double pulse output and the correct
function of downstream switching units.

User entry 0 to 65 535

Dev. alarm sim. 

Navigation  Expert → Diagnostics → Simulation → Dev. alarm sim. (0654)

Description Use this function to switch the device alarm on and off.

Selection • Off
• On

Factory setting Off

Additional information Description


The display alternates between the measured value and a diagnostic message of the
"Function check" category (C) while simulation is in progress.

246 Endress+Hauser
Proline Promass 500 Modbus RS485 Description of device parameters

Event category 

Navigation  Expert → Diagnostics → Simulation → Event category (0738)

Description Use this function to select the category of the diagnostic events that are displayed for the
simulation in the Diag. event sim. parameter (→  247).

Selection • Sensor
• Electronics
• Configuration
• Process

Factory setting Process

Diag. event sim. 

Navigation  Expert → Diagnostics → Simulation → Diag. event sim. (0737)

Description Use this function to select a diagnostic event for the simulation process that is activated.

Selection • Off
• Diagnostic event picklist (depends on the category selected)

Factory setting Off

Additional information Description


For the simulation, you can choose from the diagnostic events of the category selected
in the Event category parameter (→  247).

Endress+Hauser 247
Country-specific factory settings Proline Promass 500 Modbus RS485

4 Country-specific factory settings

4.1 SI units
Not valid for USA and Canada.

4.1.1 System units

Mass kg

Mass flow kg/h

Volume l

Volume flow l/h

Corrected volume Nl

Corrected volume flow Nl/h

Density kg/l

Reference density kg/Nl

Temperature °C

Pressure bar a

4.1.2 Full scale values


The factory settings applie to the following parameters:
• 20 mA value (full scale value of the current output)
• 100% bar graph value 1
For detailed information about the full scale value for measuring devices for custody
transfer, see the Special Documentation for the device →  8

Nominal diameter
[mm] [kg/h]

1 4

2 20

4 90

8 400

15 1 300

15 FB 3 600

25 3 600

25 FB 9 000

40 9 000

40 FB 14 000

50 14 000

50 FB 36 000

80 36 000

100 60 000

150 130 t/h

250 360 t/h

350 650 t/h

248 Endress+Hauser
Proline Promass 500 Modbus RS485 Country-specific factory settings

4.1.3 Output current span

Current output 1 to n 4 to 20 mA NAMUR

4.1.4 Pulse value


For detailed information about the pulse value for measuring devices for custody
transfer, see the Special Documentation for the device →  8

Nominal diameter
[mm] [kg/p]

1 0.001

2 0.01

4 0.01

8 0.1

15 0.1

15 FB 1

25 1

25 FB 1

40 1

40 FB 10

50 10

50 FB 10

80 10

100 10

150 100

250 100

350 100

4.1.5 On value low flow cut off


The switch-on point depends on the type of medium and the nominal diameter.

Nominal diameter On-value for liquid


[mm] [kg/h]

1 0.08

2 0.4

4 1.8

8 8

15 26

15 FB 72

25 72

25 FB 180

40 180

40 FB 300

50 300

50 FB 720

Endress+Hauser 249
Country-specific factory settings Proline Promass 500 Modbus RS485

Nominal diameter On-value for liquid


[mm] [kg/h]

80 720

100 1 200

150 2.6 t/h

250 7.2 t/h

350 13 t/h

Nominal diameter Switch-on value for gas


[mm] [kg/h]

1 0.02

2 0.1

4 0.45

8 2

15 6.5

15 FB 18

25 18

25 FB 45

40 45

40 FB 75

50 75

50 FB 180

80 180

100 300

150 650

250 1.8 t/h

350 3.25 t/h

4.2 US units
Only valid for USA and Canada.

4.2.1 System units

Mass lb

Mass flow lb/min

Volume gal (us)

Volume flow gal/min (us)

Corrected volume Sft³

Corrected volume flow Sft³/min

Density lb/ft³

Reference density lb/Sft³

Temperature °F

Pressure psi a

250 Endress+Hauser
Proline Promass 500 Modbus RS485 Country-specific factory settings

4.2.2 Full scale values


The factory settings applie to the following parameters:
• 20 mA value (full scale value of the current output)
• 100% bar graph value 1
For detailed information about the full scale value for measuring devices for custody
transfer, see the Special Documentation for the device →  8

Nominal diameter
[in] [lb/min]

¹/₂₄ 0.15

¹/₁₂ 0.75

¹/₈ 3.3

³/₈ 15

½ 50

½ FB 130

1 130

1 FB 330

1½ 330

1½ FB 550

2 550

2 FB 1 300

3 1 300

4 2 200

6 4 800

10 13 000

14 23 500

4.2.3 Output current span

Current output 1 to n 4 to 20 mA US

4.2.4 Pulse value


For detailed information about the pulse value for measuring devices for custody
transfer, see the Special Documentation for the device →  8

Nominal diameter
[in] [lb/p]

¹/₂₄ 0.002

¹/₁₂ 0.02

¹/₈ 0.02

³/₈ 0.2

½ 0.2

½ FB 2

1 2

1 FB 2

1½ 2

Endress+Hauser 251
Country-specific factory settings Proline Promass 500 Modbus RS485

Nominal diameter
[in] [lb/p]

1½ FB 20

2 20

2 FB 20

3 20

4 20

6 200

10 200

14 200

4.2.5 On value low flow cut off


The switch-on point depends on the type of medium and the nominal diameter.

Nominal diameter On-value for liquid


[in] [lb/min]

¹/₂₄ 0.003

¹/₁₂ 0.015

¹/₈ 0.066

³/₈ 0.3

½ 1

½ FB 2.6

1 2.6

1 FB 6.6

1½ 6.6

1½ FB 11

2 11

2 FB 26

3 26

4 44

6 95

10 260

14 470

Nominal diameter Switch-on value for gas


[in] [lb/min]

¹/₂₄ 0.001

¹/₁₂ 0.004

¹/₈ 0.016

³/₈ 0.075

½ 0.25

½ FB 0.65

1 0.65

1 FB 1.65

252 Endress+Hauser
Proline Promass 500 Modbus RS485 Country-specific factory settings

Nominal diameter Switch-on value for gas


[in] [lb/min]

1½ 1.65

1½ FB 2.75

2 2.75

2 FB 6.5

3 6.5

4 11

6 23.75

10 65

14 117.5

Endress+Hauser 253
Explanation of abbreviated units Proline Promass 500 Modbus RS485

5 Explanation of abbreviated units

5.1 SI units
Process variable Units Explanation

Density g/cm³, g/m³ Gram/volume unit

kg/dm³, kg/l, kg/m³ Kilogram/volume unit

SD4°C, SD15°C, SD20°C Specific density: The specific density is the ratio of the
density of the fluid to the density of water at a water
temperature of 4 °C (39 °F), 15 °C (59 °F), 20 °C (68 °F).

SG4°C, SG15°C, SG20°C Specific gravity: The specific gravity is the ratio of the
density of the fluid to the density of water at a water
temperature of 4 °C (39 °F), 15 °C (59 °F), 20 °C (68 °F).

Pressure Pa a, kPa a, MPa a Pascal, kilopascal, megapascal (absolute)

bar Bar

Pa g, kPa g, MPa g Pascal, kilopascal, megapascal (relative/gauge)

bar g Bar (relative/gauge)

Mass g, kg, t Gram, kilogram, metric ton

Mass flow g/s, g/min, g/h, g/d Gram/time unit

kg/s, kg/min, kg/h, kg/d Kilogram/time unit

t/s, t/min, t/h, t/d Metric ton/time unit

Ref.density kg/Nm³, kg/Nl, g/Scm³, kg/Sm³ Kilogram, gram/standard volume unit

Corrected Nl, Nm³, Sm³ Normal liter, normal cubic meter, standard cubic meter
volume

Correct.vol.flow Nl/s, Nl/min, Nl/h, Nl/d Normal liter/time unit

Nm³/s, Nm³/min, Nm³/h, Nm³/d Normal cubic meter/time unit

Sm³/s, Sm³/min, Sm³/h, Sm³/d Standard cubic meter/time unit

Temperature °C , K Celsius, Kelvin

Volume cm³, dm³, m³ Cubic centimeter, cubic decimeter, cubic meter

ml, l, hl, Ml Mega Milliliter, liter, hectoliter, megaliter

Volume flow cm³/s, cm³/min, cm³/h, cm³/d Cubic centimeter/time unit

dm³/s, dm³/min, dm³/h, dm³/d Cubic decimeter/time unit

m³/s, m³/min, m³/h, m³/d Cubic meter/time unit

ml/s, ml/min, ml/h, ml/d Milliliter/time unit

l/s, l/min, l/h, l/d Liter/time unit

hl/s, hl/min, hl/h, hl/d Hectoliter/time unit

Ml/s, Ml/min, Ml/h, Ml/d Megaliter/time unit

Time s, m, h, d, y Second, minute, hour, day, year

5.2 US units
Process variable Units Explanation

Density lb/ft³, lb/gal (us) Pound/cubic foot, pound/gallon

lb/bbl (us;liq.), lb/bbl (us;beer), Pound/volume unit


lb/bbl (us;oil), lb/bbl (us;tank)

254 Endress+Hauser
Proline Promass 500 Modbus RS485 Explanation of abbreviated units

Process variable Units Explanation

Pressure psi a Pounds per square inch (absolute)

psi g Pounds per square inch (gauge)

Mass oz, lb, STon Ounce, pound, standard ton

Mass flow oz/s, oz/min, oz/h, oz/d Ounce/time unit

lb/s, lb/min, lb/h, lb/d Pound/time unit

STon/s, STon/min, STon/h, STon/d Standard ton/time unit

Ref.density lb/Sft³ Weight unit/standard volume unit

Corrected Sft³, Sgal (us), Sbbl (us;liq.) Standard cubic foot, standard gallon, standard barrel
volume

Correct.vol.flow Sft³/s, Sft³/min, Sft³/h, Sft³/d Standard cubic foot/time unit

Sgal/s (us), Sgal/min (us), Sgal/h Standard gallon/time unit


(us), Sgal/d (us)

Sbbl/s (us;liq.), Sbbl/min (us;liq.), Barrel/time unit (normal liquids)


Sbbl/h (us;liq.), Sbbl/d (us;liq.)

Temperature °F, °R Fahrenheit, Rankine

Volume af Acre foot

ft³ Cubic foot

fl oz (us), gal (us), kgal (us), Mgal Fluid ounce, gallon, kilogallon, million gallon
(us)

bbl (us;liq.), bbl (us;beer), bbl Barrel (normal liquids), barrel (beer), barrel
(us;oil), bbl (us;tank) (petrochemicals), barrel (filling tanks)

Volume flow af/s, af/min, af/h, af/d Acre foot/time unit

ft³/s, ft³/min, ft³/h, ft³/d Cubic foot/time unit

fl oz/s (us), fl oz/min (us), fl oz/h Fluid ounce/time unit


(us), fl oz/d (us)

gal/s (us), gal/min (us), gal/h (us), Gallon/time unit


gal/d (us)

kgal/s (us), kgal/min (us), kgal/h Kilogallon/time unit


(us), kgal/d (us)

Mgal/s (us), Mgal/min (us), Million gallon/time unit


Mgal/h (us), Mgal/d (us)

bbl/s (us;liq.), bbl/min (us;liq.), Barrel/time unit (normal liquids)


bbl/h (us;liq.), bbl/d (us;liq.) Normal liquids: 31.5 gal/bbl

bbl/s (us;beer), bbl/min (us;beer), Barrel /time unit (beer)


bbl/h (us;beer), bbl/d (us;beer) Beer: 31.0 gal/bbl

bbl/s (us;oil), bbl/min (us;oil), Barrel/time unit (petrochemicals)


bbl/h (us;oil), bbl/d (us;oil) Petrochemicals: 42.0 gal/bbl

bbl/s (us;tank), bbl/min (us;tank), Barrel/time unit (filling tank)


bbl/h (us;tank), bbl/d (us;tank) Filling tanks: 55.0 gal/bbl

Time s, m, h, d, y Second, minute, hour, day, year

am, pm Ante meridiem ( before midday), post meridiem (after


midday)

Endress+Hauser 255
Explanation of abbreviated units Proline Promass 500 Modbus RS485

5.3 Imperial units


Process variable Units Explanation

Density lb/gal (imp), lb/bbl (imp;beer), Pound/volume unit


lb/bbl (imp;oil)

Corrected Sgal (imp) Standard gallon


volume

Correct.vol.flow Sgal/s (imp), Sgal/min (imp), Standard gallon/time unit


Sgal/h (imp), Sgal/d (imp)

Volume gal (imp), Mgal (imp) Gallon, mega gallon

bbl (imp;beer), bbl (imp;oil) Barrel (beer), barrel (petrochemicals)

Volume flow gal/s (imp), gal/min (imp), gal/h Gallon/time unit


(imp), gal/d (imp)

Mgal/s (imp), Mgal/min (imp), Mega gallon/time unit


Mgal/h (imp), Mgal/d (imp)

bbl/s (imp;beer), bbl/min Barrel /time unit (beer)


(imp;beer), bbl/h (imp;beer), bbl/d Beer: 36.0 gal/bbl
(imp;beer)

bbl/s (imp;oil), bbl/min (imp;oil), Barrel/time unit (petrochemicals)


bbl/h (imp;oil), bbl/d (imp;oil) Petrochemicals: 34.97 gal/bbl

Time s, m, h, d, y Second, minute, hour, day, year

am, pm Ante meridiem ( before midday), post meridiem (after


midday)

256 Endress+Hauser
Proline Promass 500 Modbus RS485 Modbus RS485 Register Information

6 Modbus RS485 Register Information

6.1 Notes

6.1.1 Structure of the register information


The individual parts of a parameter description are described in the following section:

Navigation: navigation path to the parameter

Parameter Register Data type Access type Selection/input →

Name of parameter Indicated in decimal • Float length = Possible type of access Options Page number information
numerical format 4 byte to parameter: List of the individual and cross-reference to the
• Integer length = • Read access via options for the standard parameter
2 byte function codes 03, parameter description
• String length, 04 or 23 • Option 1
depending on • Write access via • Option 2
parameter function codes 06, • Option 3 (+)
16 or 23
 • Factory
setting
highlighted in
bold
• (+) = Factory
setting
depends on
country, order
options or
device
settings
User entry
Input range for the
parameter

NOTICE
If non-volatile device parameters are modified via the MODBUS RS485 function codes
06, 16 or 23, the change is saved in the EEPROM of the measuring device.
The number of writes to the EEPROM is technically restricted to a maximum of 1 million.
‣ Make sure to comply with this limit since, if it is exceeded, data loss and measuring
device failure will result.
‣ Avoid constantly writing non-volatile device parameters via the MODBUS RS485.

6.1.2 Address model


The Modbus RS485 register addresses of the measuring device are implemented in
accordance with the "Modbus Applications Protocol Specification V1.1".
In addition, systems are used that work with the register address model "Modicon Modbus
Protocol Reference Guide (PI-MBUS-300 Rev. J)".
Depending on the function code used, a number is added at the start of the register
address with this specification:
• "3" → "Read" access
• "4" → "Write" access

Endress+Hauser 257
Modbus RS485 Register Information Proline Promass 500 Modbus RS485

Function code Access type Register in accordance with "Modbus Register in accordance with
Applications Protocol Specification" "Modicon Modbus Protocol
Reference Guide"

03 Read XXXX 3XXXX


04 Example: mass flow = 2007 Example: mass flow = 32007
23

06 Write XXXX 4XXXX


16 Example: reset totalizer = 6401 Example: reset totalizer = 46401
23

6.2 Overview of the Expert operating menu


The following table provides an overview of the menu structure of the expert operating
menu and its parameters. The page reference indicates where the associated description of
the submenu or parameter can be found.

 Expert

Direct access (0106) →  271

Locking status (0004) →  271

Access status (0005) →  271

Ent. access code (0003) →  271

‣ System →  272

‣ Display →  272

Display language (0104) →  272

Format display (0098) →  272

Value 1 display (0107) →  273

0% bargraph 1 (0123) →  273

100% bargraph 1 (0125) →  273

Decimal places 1 (0095) →  274

Value 2 display (0108) →  274

Decimal places 2 (0117) →  274

Value 3 display (0110) →  274

0% bargraph 3 (0124) →  274

100% bargraph 3 (0126) →  274

258 Endress+Hauser
Proline Promass 500 Modbus RS485 Modbus RS485 Register Information

Decimal places 3 (0118) →  274

Value 4 display (0109) →  274

Decimal places 4 (0119) →  274

Display interval (0096) →  274

Display damping (0094) →  274

Header (0097) →  274

Header text (0112) →  274

Separator (0101) →  274

Contrast display (0105) →  274

Backlight (0111) →  274

‣ Config. backup →  274

Operating time (0652) →  274

Last backup (2757) →  274

Config. managem. (2758) →  275

Backup state (2759) →  275

Compar. result (2760) →  275

‣ Diagn. handling →  275

Alarm delay (0651) →  275

‣ Diagn. behavior →  275

‣ Administration →  277

‣ Def. access code →  277

‣ Reset acc. code →  277

Device reset (0000) →  277

Activate SW opt. (0029) →  277

SW option overv. (0015) →  277

Endress+Hauser 259
Modbus RS485 Register Information Proline Promass 500 Modbus RS485

‣ Sensor →  278

‣ Measured val. →  278

‣ Process variab. →  278

‣ Totalizer →  278

‣ Input values →  278

‣ Output values →  279

‣ System units →  280

Mass flow unit (0554) →  280

Mass unit (0574) →  280

Volume flow unit (0553) →  281

Volume unit (0563) →  282

Cor.volflow unit (0558) →  283

Corr. vol. unit (0575) →  283

Density unit (0555) →  284

Ref. dens. unit (0556) →  284

Temperature unit (0557) →  284

Pressure unit (0564) →  284

Date/time format (2812) →  284

‣ User-spec. units →  284

‣ Process param. →  285

Flow damping (1802) →  285

Density damping (1803) →  285

Temp. damping (1822) →  285

Flow override (1839) →  285

260 Endress+Hauser
Proline Promass 500 Modbus RS485 Modbus RS485 Register Information

‣ Low flow cut off →  285

‣ Partial pipe det →  286

‣ Measurement mode →  286

Select medium (6062) →  286

Select gas type (6074) →  286

Sound velocity (6147) →  286

Temp. coeff. SV (6181) →  286

‣ External comp. →  286

Pressure compen. (6130) →  286

Pressure value (6059) →  286

External press. (6209) →  286

Temperature mode (6184) →  287

External temp. (6080) →  287

‣ Calculated value →  287

‣ Corr. vol.flow. →  287

‣ Sensor adjustm. →  287

Install. direct. (1809) →  287

Inst. angle roll (6282) →  287

Inst.angle pitch (6236) →  287

‣ Zero point adj. →  287

‣ Variable adjust →  288

‣ Calibration →  288

Cal. factor (6025) →  288

Zero point (6195) →  288

Endress+Hauser 261
Modbus RS485 Register Information Proline Promass 500 Modbus RS485

Nominal diameter (2807) →  288

C0 to 5 (6022) →  288

‣ I/O config. →  288

I/O 1 to n terminals (3902–1 to n) →  288

I/O 1 to n info (3906–1 to n) →  288

I/O 1 to n type (3901–1 to n) →  288

Apply I/O config (3907) →  289

Conversion code (2762) →  289

‣ Input →  289

‣ Current input 1 to n →  289

Terminal no. (1611–1 to n) →  289

Signal mode (1610–1 to n) →  289

Current span (1605–1 to n) →  289

0/4 mA value (1606–1 to n) →  289

20 mA value (1607–1 to n) →  289

Failure mode (1601–1 to n) →  289

Failure value (1602–1 to n) →  289

‣ Status input 1 to n →  289

Terminal no. (1358–1 to n) →  289

Assign stat.inp. (1352–1 to n) →  289

Val.stat.inp. (1353–1 to n) →  290

Active level (1351–1 to n) →  290

Response time (1354–1 to n) →  290

262 Endress+Hauser
Proline Promass 500 Modbus RS485 Modbus RS485 Register Information

‣ Output →  290

‣ Curr.output 1 to n →  290

Terminal no. (0379–1 to n) →  290

Signal mode (0377–1 to n) →  290

Assign curr. 1 to n (0359–1 to n) →  291

Current span (0353–1 to n) →  291

Fixed current (0365–1 to n) →  291

0/4 mA value (0367–1 to n) →  291

20 mA value (0372–1 to n) →  292

Measuring mode (0351–1 to n) →  292

Damping out. 1 to n (0363–1 to n) →  292

Response time (0378–1 to n) →  292

Failure mode (0364–1 to n) →  292

Failure current (0352–1 to n) →  292

Output curr. 1 to n (0361–1 to n) →  292

Measur. curr. 1 to n (0366–1 to n) →  292

‣ PFS output 1 to n →  292

Terminal no. (0492–1 to n) →  292

Signal mode (0490–1 to n) →  292

Operating mode (0469–1 to n) →  292

Assign pulse 1 to n (0460–1 to n) →  293

Value per pulse (0455–1 to n) →  293

Pulse width (0452–1 to n) →  293

Measuring mode (0457–1 to n) →  293

Failure mode (0480–1 to n) →  293

Endress+Hauser 263
Modbus RS485 Register Information Proline Promass 500 Modbus RS485

Pulse output 1 to n (0456–1 to n) →  293

Assign freq. (0478–1 to n) →  294

Min. freq. value (0453–1 to n) →  294

Max. freq. value (0454–1 to n) →  294

Val. at min.freq (0476–1 to n) →  294

Val. at max.freq (0475–1 to n) →  295

Measuring mode (0479–1 to n) →  295

Damping out. 1 to n (0477–1 to n) →  295

Response time (0491–1 to n) →  295

Failure mode (0451–1 to n) →  295

Failure freq. (0474–1 to n) →  295

Output freq. 1 to n (0471–1 to n) →  295

Switch out funct (0481–1 to n) →  295

Assign diag. beh (0482–1 to n) →  295

Assign limit (0483–1 to n) →  296

Switch-on value (0466–1 to n) →  296

Switch-off value (0464–1 to n) →  296

Assign dir.check (0484–1 to n) →  296

Assign status (0485–1 to n) →  296

Switch-on delay (0467–1 to n) →  296

Switch-off delay (0465–1 to n) →  296

Failure mode (0486–1 to n) →  296

Switch status 1 to n (0461–1 to n) →  297

Invert outp.sig. (0470–1 to n) →  297

264 Endress+Hauser
Proline Promass 500 Modbus RS485 Modbus RS485 Register Information

‣ Relay output 1 to n →  297

Terminal no. (0812–1 to n) →  297

Relay outp.func. (0804–1 to n) →  297

Assign dir.check (0808–1 to n) →  297

Assign limit (0807–1 to n) →  298

Assign diag. beh (0806–1 to n) →  298

Assign status (0805–1 to n) →  298

Switch-off value (0809–1 to n) →  298

Switch-off delay (0813–1 to n) →  298

Switch-on value (0810–1 to n) →  298

Switch-on delay (0814–1 to n) →  298

Failure mode (0811–1 to n) →  298

Switch status (0801–1 to n) →  299

Act. relay state (0816–1 to n) →  299

‣ Double pulse out →  299

Master term. no. (0981) →  299

Slave term. no. (0990) →  299

Signal mode (0991) →  299

Assign pulse 1 (0982–1) →  299

Value per pulse (0983) →  299

Pulse width (0986) →  299

Phase shift (0992) →  299

Measuring mode (0984) →  299

Failure mode (0985) →  299

Endress+Hauser 265
Modbus RS485 Register Information Proline Promass 500 Modbus RS485

Pulse output (0987) →  300

Invert outp.sig. (0993) →  300

‣ Communication →  300

‣ Modbus config. →  300

Bus address (7112) →  300

Baudrate (7111) →  300

Data trans. mode (7115) →  300

Parity (7122) →  300

Byte order (7113) →  300

Telegram delay (7146) →  300

Failure mode (7116) →  300

Bus termination (7155) →  300

Fieldb.writ.acc. (7156) →  300

‣ Modbus info →  300

Device ID (7153) →  300

Device revision (7154) →  300

‣ Modbus data map →  301

Scan list reg.0 to 15 (7114) →  301

‣ Web server →  301

Webserv.language (7221) →  301

MAC Address (7214) →  301

IP address (7209) →  301

Subnet mask (7211) →  301

Default gateway (7210) →  301

266 Endress+Hauser
Proline Promass 500 Modbus RS485 Modbus RS485 Register Information

Webserver funct. (7222) →  301

Login page (7273) →  301

‣ WLAN settings →  302

WLAN (2702) →  302

WLAN IP address (2711) →  302

WLAN subnet mask (2709) →  302

WLAN MAC address (2703) →  302

Security type (2705) →  302

WLAN passphrase (2706) →  302

Assign SSID name (2708) →  302

SSID name (2707) →  302

Select antenna (2713) →  302

WLAN channel (2704) →  302

Apply changes (2712) →  302

‣ Application →  302

Reset all tot. (2806) →  302

‣ Totalizer 1 to n →  303

Assign variable (0914–1 to n) →  303

Unit totalizer 1 to n (0915–1 to n) →  303

Operation mode (0908–1 to n) →  303

Control Tot. 1 to n (0912–1 to n) →  304

Preset value 1 to n (0913–1 to n) →  304

Failure mode (0901–1 to n) →  304

‣ Diagnostics →  304

Actual diagnos. (0691) →  304

Endress+Hauser 267
Modbus RS485 Register Information Proline Promass 500 Modbus RS485

Prev.diagnostics (0690) →  304

Time fr. restart (0653) →  304

Operating time (0652) →  304

‣ Diagnostic list →  304

Diagnostics 1 (0692) →  304

Diagnostics 2 (0693) →  304

Diagnostics 3 (0694) →  304

Diagnostics 4 (0695) →  304

Diagnostics 5 (0696) →  304

‣ Event logbook →  305

Filter options (0705) →  305

‣ Device info →  305

Device tag (0011) →  305

Serial number (0009) →  305

Firmware version (0010) →  305

Device name (0020) →  305

Order code (0008) →  305

Ext. order cd. 1 (0023) →  305

Ext. order cd. 2 (0021) →  305

Ext. order cd. 3 (0022) →  305

ENP version (0012) →  305

‣ Mainboard module →  305

Software rev. (0072) →  305

‣ Sens. electronic →  305

Software rev. (0072) →  305

268 Endress+Hauser
Proline Promass 500 Modbus RS485 Modbus RS485 Register Information

‣ I/O module 1 →  306

I/O 1 terminals (3902–1) →  306

Software rev. (0072) →  306

‣ I/O module 2 →  306

I/O 2 terminals (3902–2) →  306

Software rev. (0072) →  306

‣ I/O module 3 →  306

I/O 3 terminals (3902–3) →  306

Software rev. (0072) →  306

‣ I/O module 4 →  306

I/O 4 terminals (3902–4) →  306

Software rev. (0072) →  306

‣ Display module →  307

Software rev. (0072) →  307

‣ Min/max val. →  307

Reset min/max (6151) →  307

‣ Main elect.temp. →  307

‣ Sensor elec.temp →  307

‣ Medium temp. →  307

‣ Carr. pipe temp. →  307

‣ Oscil. frequency →  308

‣ Tors.oscil.freq. →  308

‣ Oscil. amplitude →  308

‣ Tor. osc. amp. →  308

‣ Oscil. damping →  308

Endress+Hauser 269
Modbus RS485 Register Information Proline Promass 500 Modbus RS485

‣ Tors.oscil.damp. →  308

‣ Signal asymmetry →  309

‣ Tors.sig.asymm. →  309

‣ Data logging →  310

Assign chan. 1 (0851) →  310

Assign chan. 2 (0852) →  310

Assign chan. 3 (0853) →  310

Assign chan. 4 (0854) →  311

Logging interval (0856) →  311

Clear logging (0855) →  311

Data logging (0860) →  311

Logging delay (0859) →  311

Data log.control (0857) →  311

Data log. status (0858) →  311

Logging duration (0861) →  311

‣ Simulation →  311

Assign proc.var. (1810) →  311

Proc. var. value (1811) →  311

Status inp.sim (1355) →  312

Signal level (1356) →  312

Curr.inp 1 to n sim. (1608–1 to n) →  312

Value curr.inp 1 to n (1609–1 to n) →  312

Curr.out. 1 to n sim. (0354–1 to n) →  312

Value curr.out 1 to n (0355–1 to n) →  312

FreqOutputSim 1 to n (0472–1 to n) →  312

270 Endress+Hauser
Proline Promass 500 Modbus RS485 Modbus RS485 Register Information

Freq value 1 to n (0473–1 to n) →  312

Puls.outp.sim. 1 to n (0458–1 to n) →  312

Pulse value 1 to n (0459–1 to n) →  312

Switch sim. 1 to n (0462–1 to n) →  312

Switch status 1 to n (0463–1 to n) →  312

Relay out. 1 to n sim (0802–1 to n) →  312

Switch status 1 to n (0803–1 to n) →  312

Puls.outp.sim. (0988) →  312

Pulse value (0989) →  312

Dev. alarm sim. (0654) →  312

Event category (0738) →  312

Diag. event sim. (0737) →  312

6.3 Register information


Navigation: Expert

Parameter Register Data type Access Selection / User entry / User interface → 

Direct access (0106) 3878 Integer Read / Write 0 to 65 535 12

Locking status (0004) 4918 Integer Read 256 = Hardware locked 13


512 = Temp. locked
2048 = CT act.-def.par.
32768 = CT act.-all par.

Access status (0005) 2178 Integer Read 0 = Operator 14


1 = Maintenance

Ent. access code (0003) 2177 Integer Read / Write Max. 16-digit character string 14
comprising numbers, letters and special
characters

Endress+Hauser 271
Modbus RS485 Register Information Proline Promass 500 Modbus RS485

6.3.1 "System" submenu

"Display" submenu

Navigation: Expert → System → Display

Parameter Register Data type Access Selection / User entry / User interface → 

Display language (0104) 3673 Integer Read / Write 0 = English 16


1 = Deutsch
2 = Français
3 = Español
4 = Italiano
5 = Nederlands
8 = Svenska
10 = Bahasa Indonesia
11 = 日本語 (Japanese)
12 = Portuguesa
13 = Polski
14 = русский язык(Ru)
15 = čeština (Czech)
16 = 中文 (Chinese)
17 = ภาษาไทย (Thai) *
18 = Türkçe
19 = tiếng Việt (Vit)
20 = 한국어 (Korean)
21 = ‫(الْعَرَبيّة‬Ara) *

Format display (0098) 3625 Integer Read / Write 0 = 1 value, max. 17


1 = Bargr. + 1 value
2 = 2 values
3 = Val. large+2val.
4 = 4 values

272 Endress+Hauser
Proline Promass 500 Modbus RS485 Modbus RS485 Register Information

Navigation: Expert → System → Display

Parameter Register Data type Access Selection / User entry / User interface → 

Value 1 display (0107) 3963 Integer Read / Write 1 = Mass flow 19


2 = Volume flow
3 = Correct.vol.flow *
4 = Density
5 = Ref.density *
6 = Osc. ampl. 1 *
7 = Temperature
8 = Osc. ampl. 0 *
9 = Signal asymmetry *
16 = Totalizer 1
17 = Totalizer 2
18 = Totalizer 3
23 = Carr. pipe temp. *
31 = Osc.damp.fluct 1 *
32 = Exc. current 0 *
33 = Exc. current 1 *
39 = Electronic temp.
45 = Kinematic visc. *
46 = Dynam. viscosity *
48 = Osc. freq. 0
50 = Osc. freq. 1 *
52 = Freq. fluct. 1 *
58 = Spec. output 0 *
59 = Spec. output 1 *
60 = Index inh.medium
61 = Index sus.bubble *
63 = Osc. damping 0 *
64 = Osc. damping 1 *
66 = Pressure
67 = Osc.damp.fluct 0 *
68 = Freq. fluct. 0 *
73 = Concentration *
74 = Target mass flow *
75 = Carrier mass fl. *
76 = TempCompDynVisc *
77 = TempCompKinVisc *
78 = Carrier vol. fl. *
79 = Carr.corr.vol.fl *
81 = HBSI *
83 = Oil density *
84 = Water density *
85 = Density average *
86 = GSV flow *
87 = GSVa *
88 = Oil mass flow *
89 = Water mass flow *
90 = NSV flow *
91 = NSVa *
92 = S&W volume flow *
93 = Ref.dens.altern. *
94 = Oil corr.vol.fl. *
95 = Water corr.v.fl. *
96 = Targ.corr.vol.fl *
97 = Target vol. flow *
98 = Temp. average *
99 = Oil volume flow *
101 = Water cut *
102 = Water vol. flow *
121 = Curr.output 1 *
122 = Curr.output 2 *
123 = Curr.output 3 *
124 = Curr.output 4 *

0% bargraph 1 (0123) 4136 to 4137 Float Read / Write Signed floating-point number 21

100% bargraph 1 (0125) 4142 to 4143 Float Read / Write Signed floating-point number 21

Endress+Hauser 273
Modbus RS485 Register Information Proline Promass 500 Modbus RS485

Navigation: Expert → System → Display

Parameter Register Data type Access Selection / User entry / User interface → 

Decimal places 1 (0095) 3365 Integer Read / Write 0=x 22


1 = x.x
2 = x.xx
3 = x.xxx
4 = x.xxxx

Value 2 display (0108) 3964 Integer Read / Write For the picklist, see the Value 2 display 22
parameter (→  22)

Decimal places 2 (0117) 4049 Integer Read / Write 0=x 23


1 = x.x
2 = x.xx
3 = x.xxx
4 = x.xxxx

Value 3 display (0110) 3966 Integer Read / Write For the picklist, see the Value 2 display 23
parameter (→  22)

0% bargraph 3 (0124) 4138 to 4139 Float Read / Write Signed floating-point number 24

100% bargraph 3 (0126) 4140 to 4141 Float Read / Write Signed floating-point number 24

Decimal places 3 (0118) 4050 Integer Read / Write 0=x 25


1 = x.x
2 = x.xx
3 = x.xxx
4 = x.xxxx

Value 4 display (0109) 3965 Integer Read / Write For the picklist, see the Value 2 display 25
parameter (→  22)

Decimal places 4 (0119) 4051 Integer Read / Write 0=x 26


1 = x.x
2 = x.xx
3 = x.xxx
4 = x.xxxx

Display interval (0096) 3604 to 3605 Float Read / Write 1 to 10 s 26

Display damping (0094) 3554 to 3555 Float Read / Write 0.0 to 999.9 s 27

Header (0097) 3624 Integer Read / Write 0 = Device tag 28


1 = Free text

Header text (0112) 3968 to 3973 String Read / Write Max. 12 characters such as letters, 28
numbers or special characters (e.g. @,
%, /)

Separator (0101) 3671 Integer Read / Write • . (point) 29


• , (comma)

Contrast display (0105) 3674 to 3675 Float Read / Write 20 to 80 % 29

Backlight (0111) 3967 Integer Read / Write 0 = Disable 29


1 = Enable

* Visibility depends on order options or device settings

"Config. backup" submenu

Navigation: Expert → System → Config. backup

Parameter Register Data type Access Selection / User entry / User interface → 

Operating time (0652) 2631 Integer Read Days (d), hours (h), minutes (m) and 30
seconds (s)

Last backup (2757) 6430 Integer Read Days (d), hours (h), minutes (m) and 30
seconds (s)

274 Endress+Hauser
Proline Promass 500 Modbus RS485 Modbus RS485 Register Information

Navigation: Expert → System → Config. backup

Parameter Register Data type Access Selection / User entry / User interface → 

Config. managem. (2758) 5500 Integer Read / Write 0 = Cancel 31


1 = Execute backup
2 = Restore *
4 = Clear backup
5 = Compare *

Backup state (2759) 5502 Integer Read 1 = Backup in progr. 31


2 = Restore in progr
4 = Delete in progr.
5 = Comp. in progr.
6 = Restoring failed
7 = Backup failed
251 = None

Compar. result (2760) 5514 Integer Read 0 = Set. identical 32


1 = Set. not ident.
2 = No backup
3 = Check not done
4 = Backup corrupt
5 = Dataset incomp.

* Visibility depends on order options or device settings

"Diagn. handling" submenu

Navigation: Expert → System → Diagn. handling

Parameter Register Data type Access Selection / User entry / User interface → 

Alarm delay (0651) 6808 to 6809 Float Read / Write 0 to 60 s 33

"Diagn. behavior" submenu

Navigation: Expert → System → Diagn. handling → Diagn. behavior

Parameter Register Data type Access Selection / User entry / User interface → 

Diagnostic no. 046 (0709) 2756 Integer Read / Write 0 = Off 35


1 = Logbook only
2 = Warning
3 = Alarm

Diagnostic no. 140 (0708) 2757 Integer Read / Write 0 = Off 35


1 = Logbook only
2 = Warning
3 = Alarm

Diagnostic no. 144 (0731) 2081 Integer Read / Write 0 = Off 36


1 = Logbook only
2 = Warning
3 = Alarm

Diagnostic no. 374 (0710) 2755 Integer Read / Write 0 = Off 36


1 = Logbook only
2 = Warning
3 = Alarm

Diagnostic no. 441 (0657) 4742 Integer Read / Write 0 = Off 37


1 = Logbook only
2 = Warning
3 = Alarm

Diagnostic no. 442 (0658) 4919 Integer Read / Write 0 = Off 37


1 = Logbook only
2 = Warning
3 = Alarm

Endress+Hauser 275
Modbus RS485 Register Information Proline Promass 500 Modbus RS485

Navigation: Expert → System → Diagn. handling → Diagn. behavior

Parameter Register Data type Access Selection / User entry / User interface → 

Diagnostic no. 443 (0659) 5000 Integer Read / Write 0 = Off 38


1 = Logbook only
2 = Warning
3 = Alarm

Diagnostic no. 444 (0740) 5120 Integer Read / Write 0 = Off 38


1 = Logbook only
2 = Warning
3 = Alarm

Diagnostic no. 543 (0643) 2362 Integer Read / Write 0 = Off 38


1 = Logbook only
2 = Warning
3 = Alarm

Diagnostic no. 830 (0800) 6805 Integer Read / Write 0 = Off 39


1 = Logbook only
2 = Warning
3 = Alarm

Diagnostic no. 831 (0641) 6806 Integer Read / Write 0 = Off 39


1 = Logbook only
2 = Warning
3 = Alarm

Diagnostic no. 832 (0681) 2759 Integer Read / Write 0 = Off 39


1 = Logbook only
2 = Warning
3 = Alarm

Diagnostic no. 833 (0682) 2762 Integer Read / Write 0 = Off 40


1 = Logbook only
2 = Warning
3 = Alarm

Diagnostic no. 834 (0700) 2761 Integer Read / Write 0 = Off 40


1 = Logbook only
2 = Warning
3 = Alarm

Diagnostic no. 835 (0702) 2760 Integer Read / Write 0 = Off 40


1 = Logbook only
2 = Warning
3 = Alarm

Diagnostic no. 862 (0679) 6441 Integer Read / Write 0 = Off 41


1 = Logbook only
2 = Warning
3 = Alarm

Diagnostic no. 912 (0703) 2758 Integer Read / Write 0 = Off 42


1 = Logbook only
2 = Warning
3 = Alarm

Diagnostic no. 913 (0712) 2754 Integer Read / Write 0 = Off 42


1 = Logbook only
2 = Warning
3 = Alarm

Diagnostic no. 944 (0732) 2082 Integer Read / Write 0 = Off 44


1 = Logbook only
2 = Warning
3 = Alarm

Diagnostic no. 948 (0744) 5179 Integer Read / Write 0 = Off 44


1 = Logbook only
2 = Warning
3 = Alarm

276 Endress+Hauser
Proline Promass 500 Modbus RS485 Modbus RS485 Register Information

"Administration" submenu

Navigation: Expert → System → Administration

Parameter Register Data type Access Selection / User entry / User interface → 

Device reset (0000) 6817 Integer Read / Write 0 = Cancel 47


1 = Restart device
2 = To delivery set.
5 = Del. p.fail data
21 = Delete T-DAT
22 = Res.fault.param.
23 = Del. deliv.sett.
24 = Delete HistoROM
25 = Rest.S-DATBackup *

Activate SW opt. (0029) 2795 Integer Read / Write Max. 10-digit string consisting of 48
numbers.

SW option overv. (0015) 2902 Integer Read 1 = Extend. HistoROM * 49


4 = Concentration *
64 = Viscosity *
128 = Custody transfer
128 = Custody transfer
256 = Petroleum *
1024 = Appl.spec. calc.
16384 = HBT Monitoring *
32768 = HBT Verification *

* Visibility depends on order options or device settings

"Def. access code" wizard

Navigation: Expert → System → Administration → Def. access code

Parameter Register Data type Access Selection / User entry / User interface → 

Def. access code 8677 to 8684 String Read / Write Max. 16-digit character string 45
comprising numbers, letters and special
characters

Confirm code 8685 to 8692 String Read / Write Max. 16-digit character string 46
comprising numbers, letters and special
characters

"Reset acc. code" submenu

Navigation: Expert → System → Administration → Reset acc. code

Parameter Register Data type Access Selection / User entry / User interface → 

Operating time (0652) 2631 Integer Read Days (d), hours (h), minutes (m) and 46
seconds (s)

Reset acc. code (0024) 8880 to 8895 String Read / Write Character string comprising numbers, 46
letters and special characters

Endress+Hauser 277
Modbus RS485 Register Information Proline Promass 500 Modbus RS485

6.3.2 "Sensor" submenu

"Measured val." submenu


"Process variab." submenu

Navigation: Expert → Sensor → Measured val. → Process variab.

Parameter Register Data type Access Selection / User entry / User interface → 

Mass flow (1838) 2007 to 2008 Float Read Signed floating-point number 51

Volume flow (1847) 2009 to 2010 Float Read Signed floating-point number 52

Correct.vol.flow (1851) 2011 to 2012 Float Read Signed floating-point number 52

Density (1850) 2013 to 2014 Float Read Signed floating-point number 52

Ref.density (1852) 2015 to 2016 Float Read Signed floating-point number 52

Temperature (1853) 2017 to 2018 Float Read Signed floating-point number 53

Pressure value (6129) 2089 to 2090 Float Read Signed floating-point number 53

Dynam. viscosity (1854) 2019 to 2020 Float Read Signed floating-point number 53

Kinematic visc. (1857) 2083 to 2084 Float Read Signed floating-point number 54

TempCompDynVisc (1872) 2093 to 2094 Float Read Signed floating-point number 54

TempCompKinVisc (1863) 2095 to 2096 Float Read Signed floating-point number 54

Concentration (1887) 2598 to 2599 Float Read Signed floating-point number 55

Target mass flow (1864) 2797 to 2798 Float Read Signed floating-point number 55

Carrier mass fl. (1865) 2799 to 2800 Float Read Signed floating-point number 56

"Totalizer" submenu

Navigation: Expert → Sensor → Measured val. → Totalizer

Parameter Register Data type Access Selection / User entry / User interface → 

Totalizer val. 1 to n (0911–1 to n) 1: 2610 to 2611 Float Read Signed floating-point number 58
2: 2810 to 2811
3: 3010 to 3011

Tot. overflow 1 to n (0910–1 to n) 1: 2612 to 2613 Float Read Integer with sign 58
2: 2812 to 2813
3: 3012 to 3013

"Input values" submenu


"Current input 1 to n" submenu

Navigation: Expert → Sensor → Measured val. → Input values → Current input 1 to n

Parameter Register Data type Access Selection / User entry / User interface → 

Measured val. 1 to n (1603–1 to n) 1: 6151 to 6152 Float Read Signed floating-point number 60
2: 6153 to 6154
3: 6155 to 6156

Measur. curr. 1 to n (1604–1 to n) 1: 6131 to 6132 Float Read 0 to 22.5 mA 60


2: 6133 to 6134
3: 6135 to 6136

278 Endress+Hauser
Proline Promass 500 Modbus RS485 Modbus RS485 Register Information

"Val.stat.inp. 1 to n" submenu

Navigation: Expert → Sensor → Measured val. → Input values → Val.stat.inp. 1 to n

Parameter Register Data type Access Selection / User entry / User interface → 

Val.stat.inp. (1353–1 to n) 1: 2746 Integer Read 0 = Low 60


2: 4699 1 = High
3: 4700

"Output values" submenu


"Value curr.out 1 to n" submenu

Navigation: Expert → Sensor → Measured val. → Output values → Value curr.out 1 to n

Parameter Register Data type Access Selection / User entry / User interface → 

Output curr. 1 to n (0361–1 to n) 1: 5931 to 5932 Float Read 0 to 22.5 mA 61


2: 5933 to 5934
3: 5935 to 5936

Measur. curr. 1 to n (0366–1 to n) 1: 5779 to 5780 Float Read 0 to 30 mA 61


2: 5781 to 5782
3: 5783 to 5784

"PFS output 1 to n" submenu

Navigation: Expert → Sensor → Measured val. → Output values → PFS output 1 to n

Parameter Register Data type Access Selection / User entry / User interface → 

Output freq. 1 to n (0471–1 to n) 1: 3462 to 3463 Float Read 0.0 to 12 500.0 Hz 62


2: 3464 to 3465
3: 9910 to 9911

Value per pulse (0455–1 to n) Float Read / Write

Switch status 1 to n (0461–1 to n) 1: 2485 Integer Read 1 = Open 63


2: 2486 6 = Closed
3: 9917

"Relay output 1 to n" submenu

Navigation: Expert → Sensor → Measured val. → Output values → Relay output 1 to n

Parameter Register Data type Access Selection / User entry / User interface → 

Switch status (0801–1 to n) 1: 3518 Integer Read 1 = Open 64


2: 3519 6 = Closed
3: 9875

Switch cycles (0815–1 to n) 1: 7625 Integer Read Positive integer 64


2: 7627
3: 7629

Max. cycles no. (0817–1 to n) 1: 21919 Integer Read Positive integer 64


2: 21921
3: 21923

"Double pulse out" submenu

Navigation: Expert → Sensor → Measured val. → Output values → Double pulse out

Parameter Register Data type Access Selection / User entry / User interface → 

Pulse output (0987) 7041 to 7042 Float Read Positive floating-point number 65

Endress+Hauser 279
Modbus RS485 Register Information Proline Promass 500 Modbus RS485

"System units" submenu

Navigation: Expert → Sensor → System units

Parameter Register Data type Access Selection / User entry / User interface → 

Mass flow unit (0554) 2101 Integer Read / Write 0 = g/s 65


1 = g/min
2 = g/h
3 = g/d
4 = kg/s
5 = kg/min
6 = kg/h (+)
7 = kg/d
8 = t/s
9 = t/min
10 = t/h
11 = t/d
12 = oz/s
13 = oz/min
14 = oz/h
15 = oz/d
16 = lb/s
17 = lb/min
18 = lb/h
19 = lb/d
20 = STon/s
21 = STon/min
22 = STon/h
23 = STon/d

Mass unit (0574) 2102 Integer Read / Write 50 = g 66


51 = kg (+)
52 = t
53 = oz
54 = lb
55 = STon
60 = /g
61 = /kg
62 = /t
63 = /lb
64 = /STon
125 = /oz

280 Endress+Hauser
Proline Promass 500 Modbus RS485 Modbus RS485 Register Information

Navigation: Expert → Sensor → System units

Parameter Register Data type Access Selection / User entry / User interface → 

Volume flow unit (0553) 2103 Integer Read / Write 0 = cm³/s 67


1 = cm³/min
2 = cm³/h
3 = cm³/d
4 = dm³/s
5 = dm³/min
6 = dm³/h
7 = dm³/d
8 = m³/s
9 = m³/min
10 = m³/h
11 = m³/d
12 = ml/s
13 = ml/min
14 = ml/h
15 = ml/d
16 = l/s
17 = l/min
18 = l/h (+)
19 = l/d
20 = hl/s
21 = hl/min
22 = hl/h
23 = hl/d
24 = Ml/s
25 = Ml/min
26 = Ml/h
27 = Ml/d
32 = af/s
33 = af/min
34 = af/h
35 = af/d
36 = ft³/s
37 = ft³/min
38 = ft³/h
39 = ft³/d
40 = fl oz/s (us)
41 = fl oz/min (us)
42 = fl oz/h (us)
43 = fl oz/d (us)
44 = gal/s (us)
45 = gal/min (us)
46 = gal/h (us)
47 = gal/d (us)
48 = Mgal/s (us)
49 = Mgal/min (us)
50 = Mgal/h (us)
51 = Mgal/d (us)
52 = bbl/s (us;liq.) *
53 = bbl/min (us;liq.) *
54 = bbl/h (us;liq.) *
55 = bbl/d (us;liq.) *
56 = bbl/s (us;beer) *
57 = bbl/min (us;beer) *
58 = bbl/h (us;beer) *
59 = bbl/d (us;beer) *
60 = bbl/s (us;oil)
61 = bbl/min (us;oil)
62 = bbl/h (us;oil)
63 = bbl/d (us;oil)
64 = bbl/s (us;tank)
65 = bbl/min (us;tank)
66 = bbl/h (us;tank)
67 = bbl/d (us;tank)
68 = gal/s (imp)
69 = gal/min (imp)
70 = gal/h (imp)

Endress+Hauser 281
Modbus RS485 Register Information Proline Promass 500 Modbus RS485

Navigation: Expert → Sensor → System units

Parameter Register Data type Access Selection / User entry / User interface → 


71 = gal/d (imp)
72 = Mgal/s (imp)
73 = Mgal/min (imp)
74 = Mgal/h (imp)
75 = Mgal/d (imp)
76 = bbl/s (imp;beer) *
77 = bbl/min (imp;beer) *
78 = bbl/h (imp;beer) *
79 = bbl/d (imp;beer) *
80 = bbl/s (imp;oil)
81 = bbl/min (imp;oil)
82 = bbl/h (imp;oil)
83 = bbl/d (imp;oil)
88 = kgal/s (us)
89 = kgal/min (us)
90 = kgal/h (us)
91 = kgal/d (us)
92 = MMft³/s
93 = MMft³/min
94 = MMft³/h
96 = Mft³/d

Volume unit (0563) 2104 Integer Read / Write 0 = cm³ 68


1 = dm³
2 = m³
3 = ml
4 = l (+)
5 = hl
6 = Ml Mega
8 = af
9 = ft³
10 = fl oz (us)
11 = gal (us)
12 = Mgal (us)
13 = bbl (us;liq.) *
14 = bbl (us;beer) *
15 = bbl (us;oil)
16 = bbl (us;tank)
17 = gal (imp)
18 = Mgal (imp)
19 = bbl (imp;beer) *
20 = bbl (imp;oil)
22 = kgal (us)
23 = Mft³
110 = /fl oz (us)
111 = /gal (us)
112 = /Mgal (us)
113 = /bbl (us;liq.) *
114 = /bbl (us;beer) *
115 = /bbl (us;oil)
116 = /bbl (us;tank)
117 = /gal (imp)
118 = /Mgal (imp)
119 = /bbl (imp;beer) *
120 = /bbl (imp;oil)
122 = /kgal (us)
123 = /MMft³
173 = /cm³
174 = /dm³
175 = /m³
176 = /ml
177 = /hl
178 = /l
179 = /Ml
180 = /ft³
181 = /af

282 Endress+Hauser
Proline Promass 500 Modbus RS485 Modbus RS485 Register Information

Navigation: Expert → Sensor → System units

Parameter Register Data type Access Selection / User entry / User interface → 

Cor.volflow unit (0558) 2105 Integer Read / Write 0 = Nl/s 69


1 = Nl/min
2 = Nl/h (+)
3 = Nl/d
4 = Nm³/s
5 = Nm³/min
6 = Nm³/h
7 = Nm³/d
8 = Sm³/s
9 = Sm³/min
10 = Sm³/h
11 = Sm³/d
12 = Sft³/s
13 = Sft³/min
14 = Sft³/h
15 = Sft³/d
16 = Sgal/s (us)
17 = Sgal/min (us)
18 = Sgal/h (us)
19 = Sgal/d (us)
20 = Sbbl/s (us;liq.)
21 = Sbbl/min (us;liq.)
22 = Sbbl/h (us;liq.)
23 = Sbbl/d (us;liq.)
24 = Sgal/s (imp)
25 = Sgal/min (imp)
26 = Sgal/h (imp)
27 = Sgal/d (imp)
28 = MMSft³/s
29 = MMSft³/min
30 = MMSft³/h
31 = MSft³/d
32 = Sbbl/s (us;oil)
33 = Sbbl/min (us;oil)
34 = Sbbl/h (us;oil)
35 = Sbbl/d (us;oil)
36 = Nhl/s
37 = Nhl/min
38 = Nhl/h
39 = Nhl/d
40 = Sl/s
41 = Sl/min
42 = Sl/h
43 = Sl/d

Corr. vol. unit (0575) 2106 Integer Read / Write 100 = Nl (+) 70


101 = Nm³
102 = Sm³
103 = Sft³
104 = Sl
105 = Sgal (us)
106 = Sbbl (us;liq.)
107 = Sgal (imp)
108 = Sbbl (us;oil)
109 = MMSft³
110 = Nhl
164 = /Sl
166 = /Nm³
167 = /Nl
168 = /Sft³
169 = /Sm³
170 = /Sbbl (us;liq.)
171 = /Sgal (us)
172 = /Sgal (imp)
178 = /Sbbl (us;oil)
179 = /MMSft³
180 = /Nhl

Endress+Hauser 283
Modbus RS485 Register Information Proline Promass 500 Modbus RS485

Navigation: Expert → Sensor → System units

Parameter Register Data type Access Selection / User entry / User interface → 

Density unit (0555) 2107 Integer Read / Write 0 = g/cm³ 70


2 = kg/dm³
3 = kg/l (+)
4 = kg/m³
5 = SD4°C
6 = SD15°C
7 = SD20°C
8 = SG4°C
9 = SG15°C
10 = SG20°C
11 = lb/ft³
12 = lb/gal (us)
13 = lb/bbl (us;liq.) *
14 = lb/bbl (us;beer) *
15 = lb/bbl (us;oil)
16 = lb/bbl (us;tank)
17 = lb/gal (imp)
18 = lb/bbl (imp;beer) *
19 = lb/bbl (imp;oil)
21 = g/m³
22 = g/ml
23 = °API *
24 = SG60°F *

Ref. dens. unit (0556) 2108 Integer Read / Write 0 = g/Scm³ 72


1 = kg/Nl (+)
2 = kg/Nm³
3 = kg/Sm³
4 = lb/Sft³
5 = °APIbase
6 = RD15°C
7 = RD20°C
8 = RD60°F

Temperature unit (0557) 2109 Integer Read / Write 0 = °C (+) 72


1=K
2 = °F
3 = °R

Pressure unit (0564) 2130 Integer Read / Write 0 = bar 73


1 = psi a
2 = bar g
3 = psi g
4 = Pa a
5 = kPa a
6 = MPa a
7 = Pa g
8 = kPa g
9 = MPa g

Date/time format (2812) 2150 Integer Read / Write 0 = dd.mm.yy hh:mm 73


1 = mm/dd/yy am/pm
2 = dd.mm.yy am/pm
3 = mm/dd/yy hh:mm

* Visibility depends on order options or device settings

"User-spec. units" submenu

Navigation: Expert → Sensor → System units → User-spec. units

Parameter Register Data type Access Selection / User entry / User interface → 

Mass text (0560) String Read / Write Max. 10 characters such as letters, 76


numbers or special characters (@, %, /)

Mass factor (0561) Float Read / Write Signed floating-point number 77

284 Endress+Hauser
Proline Promass 500 Modbus RS485 Modbus RS485 Register Information

Navigation: Expert → Sensor → System units → User-spec. units

Parameter Register Data type Access Selection / User entry / User interface → 

Mass offset (0562) Float Read / Write Signed floating-point number 76

Volume text (0567) String Read / Write Max. 10 characters such as letters, 75


numbers or special characters (@, %, /)

Volume factor (0568) Float Read / Write Signed floating-point number 76

Corr. vol. text (0592) String Read / Write Max. 10 characters such as letters, 77
numbers or special characters (@, %, /)

Cor.vol. factor (0590) Float Read / Write Signed floating-point number 78

Corr vol. offset (0602) Float Read / Write Signed floating-point number 78

Volume offset (0569) Float Read / Write Signed floating-point number 75

Density text (0570) String Read / Write Max. 10 characters such as letters, 78


numbers or special characters (@, %, /)

Density factor (0572) Float Read / Write Signed floating-point number 79

Density offset (0571) Float Read / Write Signed floating-point number 79

Pressure text (0581) String Read / Write Max. 10 characters such as letters, 81


numbers or special characters (@, %, /)

Pressure factor (0579) Float Read / Write Signed floating-point number 82

Pressure offset (0580) Float Read / Write Signed floating-point number 81

"Process param." submenu

Navigation: Expert → Sensor → Process param.

Parameter Register Data type Access Selection / User entry / User interface → 

Flow damping (1802) 5510 to 5511 Float Read / Write 0 to 100.0 s 82

Density damping (1803) 5508 to 5509 Float Read / Write 0 to 999.9 s 83

Temp. damping (1822) 5127 to 5128 Float Read / Write 0 to 999.9 s 83

Flow override (1839) 5503 Integer Read / Write 0 = Off 84


1 = On

"Low flow cut off" submenu

Navigation: Expert → Sensor → Process param. → Low flow cut off

Parameter Register Data type Access Selection / User entry / User interface → 

Assign variable (1837) 5101 Integer Read / Write 0 = Off 85


1 = Mass flow
2 = Volume flow
3 = Correct.vol.flow *

On value (1805) 5138 to 5139 Float Read / Write Positive floating-point number 85

Off value (1804) 5104 to 5105 Float Read / Write 0 to 100.0 % 85

Pres. shock sup. (1806) 5140 to 5141 Float Read / Write 0 to 100 s 86

* Visibility depends on order options or device settings

Endress+Hauser 285
Modbus RS485 Register Information Proline Promass 500 Modbus RS485

"Partial pipe det" submenu

Navigation: Expert → Sensor → Process param. → Partial pipe det

Parameter Register Data type Access Selection / User entry / User interface → 

Assign variable (1860) 5106 Integer Read / Write 0 = Off 88


4 = Density
5 = Ref.density

Low value (1861) 5110 to 5111 Float Read / Write Signed floating-point number 88

High value (1858) 5112 to 5113 Float Read / Write Signed floating-point number 89

Response time (1859) 5108 to 5109 Float Read / Write 0 to 100 s 89

Max. damping (6040) 2414 to 2415 Float Read / Write Positive floating-point number 89

"Measurement mode" submenu

Navigation: Expert → Sensor → Measurement mode

Parameter Register Data type Access Selection / User entry / User interface → 

Select medium (6062) 2442 Integer Read / Write 0 = Liquid 90


1 = Gas

Select gas type (6074) 5229 Integer Read / Write 0 = Air 91


1 = Nitrogen N2
2 = Argon Ar
3 = Helium He
4 = Carbon diox. CO2
5 = Oxygen O2
6 = Methane CH4
7 = Ammonia NH3
9 = Hydrogen H2
10 = Ethane C2H6
11 = Propane C3H8
12 = Butane C4H10
13 = Chlorine Cl2
14 = Hydrog.chlor.HCl
15 = Carbon monox. CO
16 = Nitrous ox. N2O
17 = Nitrog. ox. NOx
18 = Hydrog.sulf. H2S
19 = Sulf. hex.fl.SF6
20 = Propylene C3H6
21 = Ozone O3
22 = Others
23 = Ethylene C2H4

Sound velocity (6147) 7413 to 7414 Float Read / Write 1 to 99 999.9999 m/s 91

Temp. coeff. SV (6181) 7411 to 7412 Float Read / Write Positive floating-point number 92

"External comp." submenu

Navigation: Expert → Sensor → External comp.

Parameter Register Data type Access Selection / User entry / User interface → 

Pressure compen. (6130) 5184 Integer Read / Write 0 = Off 93


1 = Fixed value
2 = External value
11 = Current input 1 *
12 = Current input 2 *
13 = Current input 3 *

Pressure value (6059) 5185 to 5186 Float Read / Write Positive floating-point number 93

External press. (6209) 2440 to 2441 Float Read / Write Positive floating-point number 93

286 Endress+Hauser
Proline Promass 500 Modbus RS485 Modbus RS485 Register Information

Navigation: Expert → Sensor → External comp.

Parameter Register Data type Access Selection / User entry / User interface → 

Temperature mode (6184) 5515 Integer Read / Write 0 = Internal value 94


1 = External value
11 = Current input 1 *
12 = Current input 2 *
13 = Current input 3 *

External temp. (6080) 2507 to 2508 Float Read / Write –273.15 to 99 999 °C 94

* Visibility depends on order options or device settings

"Calculated value" submenu


"Corr. vol.flow." submenu

Navigation: Expert → Sensor → Calculated value → Corr. vol.flow.

Parameter Register Data type Access Selection / User entry / User interface → 

Corr. vol.flow. (1812) 5129 Integer Read / Write 0 = Calc ref density 96


1 = Fix ref.density
11 = Current input 1 *
12 = Current input 2 *
13 = Current input 3 *

Ext. ref.density (6198) 2509 to 2510 Float Read / Write Floating point number with sign 96

Fix ref.density (1814) 5130 to 5131 Float Read / Write Positive floating-point number 97

Ref. temperature (1816) 5136 to 5137 Float Read / Write –273.15 to 99 999 °C 97

Linear exp coeff (1817) 5132 to 5133 Float Read / Write Signed floating-point number 98

Square exp coeff (1818) 5134 to 5135 Float Read / Write Signed floating-point number 98

* Visibility depends on order options or device settings

"Sensor adjustm." submenu

Navigation: Expert → Sensor → Sensor adjustm.

Parameter Register Data type Access Selection / User entry / User interface → 

Install. direct. (1809) 5501 Integer Read / Write 0 = In arrow direct. 99


1 = Against arrow

Inst. angle roll (6282) 2660 to 2661 Float Read / Write –180 to 180 ° 99

Inst.angle pitch (6236) 6529 to 6530 Float Read / Write –180 to 180 ° 99

"Zero point adj." wizard

Navigation: Expert → Sensor → Sensor adjustm. → Zero point adj.

Parameter Register Data type Access Selection / User entry / User interface → 

Zero point adj. (6196) 5121 Integer Read / Write 0 = Cancel 100


1 = Start

Progress (2808) 6797 Integer Read 0 to 100 % 101

Endress+Hauser 287
Modbus RS485 Register Information Proline Promass 500 Modbus RS485

"Variable adjust" submenu

Navigation: Expert → Sensor → Sensor adjustm. → Variable adjust

Parameter Register Data type Access Selection / User entry / User interface → 

Mass flow offset (1831) 5521 to 5522 Float Read / Write Signed floating-point number 102

Mass flow factor (1832) 5519 to 5520 Float Read / Write Positive floating-point number 102

Vol. flow offset (1841) 5525 to 5526 Float Read / Write Signed floating-point number 103

Vol. flow factor (1846) 5523 to 5524 Float Read / Write Positive floating-point number 103

Density offset (1848) 5529 to 5530 Float Read / Write Signed floating-point number 103

Density factor (1849) 5527 to 5528 Float Read / Write Positive floating-point number 104

Corr. vol offset (1866) 2044 to 2045 Float Read / Write Signed floating-point number 104

Corr. vol factor (1867) 2076 to 2077 Float Read / Write Positive floating-point number 104

Ref.dens. offset (1868) 2046 to 2047 Float Read / Write Signed floating-point number 105

Ref.dens. factor (1869) 2042 to 2043 Float Read / Write Positive floating-point number 105

Temp. offset (1870) 5533 to 5534 Float Read / Write Signed floating-point number 105

Temp. factor (1871) 5531 to 5532 Float Read / Write Positive floating-point number 106

"Calibration" submenu

Navigation: Expert → Sensor → Calibration

Parameter Register Data type Access Selection / User entry / User interface → 

Cal. factor (6025) 7513 to 7514 Float Read Signed floating-point number 106

Zero point (6195) 7527 to 7528 Float Read / Write Signed floating-point number 107

Nominal diameter (2807) 2048 to 2057 String Read DNxx / x" 107

C0 to 5 (6022) 0: 7501 to 7502 Float Read Signed floating-point number 107


1: 7503 to 7504
2: 7505 to 7506
3: 7507 to 7508
4: 7509 to 7510
5: 7511 to 7512

6.3.3 "I/O config." submenu

Navigation: Expert → I/O config.

Parameter Register Data type Access Selection / User entry / User interface → 

I/O 1 to n terminals (3902–1 to n) 1: 6541 Integer Read 0 = Not used 108


2: 6542 1 = 26-27 (I/O 1)
3: 6543 2 = 24-25 (I/O 2)
4: 6544 3 = 22-23 (I/O 3)
4 = 20-21 (I/O 4) *

I/O 1 to n info (3906–1 to n) 1: 8659 Integer Read 1 = MODBUS 108


2: 8660 2 = Configurable
3: 8661 3 = Not configurable
4: 8662 254 = Not plugged
255 = Invalid

I/O 1 to n type (3901–1 to n) 1: 6417 Integer Read / Write 0 = Off 109


2: 6418 1 = Curr.output *
3: 6419 2 = Current input *
4: 6420 3 = PFS output *
4 = Double pulse out *
5 = Status input *
6 = Relay output *

288 Endress+Hauser
Proline Promass 500 Modbus RS485 Modbus RS485 Register Information

Navigation: Expert → I/O config.

Parameter Register Data type Access Selection / User entry / User interface → 

Apply I/O config (3907) 8665 Integer Read / Write 0 = Yes 109


1 = No

Conversion code (2762) 6427 Integer Read / Write Positive integer 109

* Visibility depends on order options or device settings

6.3.4 "Input" submenu

"Current input 1 to n" submenu

Navigation: Expert → Input → Current input 1 to n

Parameter Register Data type Access Selection / User entry / User interface → 

Terminal no. (1611–1 to n) 1: 6548 Integer Read 0 = Not used 110


2: 6549 2 = 24-25 (I/O 2)
3: 6550 3 = 22-23 (I/O 3)
4 = 20-21 (I/O 4) *

Signal mode (1610–1 to n) 1: 6424 Integer Read / Write 0 = Passive 111


2: 6425 2 = Active *
3: 6426

Current span (1605–1 to n) 1: 6147 Integer Read / Write 0 = 4...20 mA 111


2: 6148 1 = 4...20 mA US
3: 6149 2 = 4...20 mA NAMUR (+)
3 = 0...20 mA

0/4 mA value (1606–1 to n) 1: 6111 to 6112 Float Read / Write Signed floating-point number 112
2: 6113 to 6114
3: 6115 to 6116

20 mA value (1607–1 to n) 1: 6119 to 6120 Float Read / Write Signed floating-point number 112
2: 6121 to 6122
3: 6123 to 6124

Failure mode (1601–1 to n) 1: 6159 Integer Read / Write 1 = Last valid value 112
2: 6160 2 = Alarm
3: 6161 6 = Defined value

Failure value (1602–1 to n) 1: 6163 to 6164 Float Read / Write Signed floating-point number 113
2: 6165 to 6166
3: 6167 to 6168

* Visibility depends on order options or device settings

"Status input 1 to n" submenu

Navigation: Expert → Input → Status input 1 to n

Parameter Register Data type Access Selection / User entry / User interface → 

Terminal no. (1358–1 to n) 1: 6554 Integer Read 0 = Not used 113


2: 6555 2 = 24-25 (I/O 2)
3: 6556 3 = 22-23 (I/O 3)
4 = 20-21 (I/O 4) *

Assign stat.inp. (1352–1 to n) 1: 2506 Integer Read / Write 0 = Off 114


2: 4687 0 = Zero point adj.
3: 4688 1 = Flow override
2 = Reset all tot.
3 = Reset totaliz. 1
4 = Reset totaliz. 2
5 = Reset totaliz. 3

Endress+Hauser 289
Modbus RS485 Register Information Proline Promass 500 Modbus RS485

Navigation: Expert → Input → Status input 1 to n

Parameter Register Data type Access Selection / User entry / User interface → 

Val.stat.inp. (1353–1 to n) 1: 2746 Integer Read 0 = Low 115


2: 4699 1 = High
3: 4700

Active level (1351–1 to n) 1: 2530 Integer Read / Write 0 = Low 115


2: 4690 1 = High
3: 4691

Response time (1354–1 to n) 1: 3404 to 3405 Float Read / Write 5 to 200 ms 115


2: 5753 to 5754
3: 5755 to 5756

* Visibility depends on order options or device settings

6.3.5 "Output" submenu

"Curr.output 1 to n" submenu

Navigation: Expert → Output → Curr.output 1 to n

Parameter Register Data type Access Selection / User entry / User interface → 

Terminal no. (0379–1 to n) 1: 6545 Integer Read 0 = Not used 116


2: 6546 2 = 24-25 (I/O 2)
3: 6547 3 = 22-23 (I/O 3)
4 = 20-21 (I/O 4) *

Signal mode (0377–1 to n) 1: 6421 Integer Read / Write 0 = Passive * 117


2: 6422 2 = Active *
3: 6423

290 Endress+Hauser
Proline Promass 500 Modbus RS485 Modbus RS485 Register Information

Navigation: Expert → Output → Curr.output 1 to n

Parameter Register Data type Access Selection / User entry / User interface → 


*
Assign curr. 1 to n (0359–1 to n) 1: 5927 Integer Read / Write 0 = Off 117
2: 5928 1 = Mass flow
3: 5929 2 = Volume flow
3 = Correct.vol.flow *
4 = Density
5 = Ref.density *
6 = Osc. ampl. 1 *
7 = Temperature
8 = Osc. ampl. 0 *
9 = Signal asymmetry *
23 = Carr. pipe temp. *
31 = Osc.damp.fluct 1 *
32 = Exc. current 0 *
33 = Exc. current 1 *
39 = Electronic temp.
45 = Kinematic visc. *
46 = Dynam. viscosity *
48 = Osc. freq. 0
50 = Osc. freq. 1 *
52 = Freq. fluct. 1 *
58 = Spec. output 0 *
59 = Spec. output 1 *
60 = Index inh.medium
61 = Index sus.bubble *
63 = Osc. damping 0 *
64 = Osc. damping 1 *
66 = Pressure
67 = Osc.damp.fluct 0 *
68 = Freq. fluct. 0 *
73 = Concentration *
74 = Target mass flow *
75 = Carrier mass fl. *
76 = TempCompDynVisc *
77 = TempCompKinVisc *
78 = Carrier vol. fl. *
79 = Carr.corr.vol.fl *
81 = HBSI *
83 = Oil density *
84 = Water density *
86 = GSV flow *
87 = GSVa *
88 = Oil mass flow *
89 = Water mass flow *
90 = NSV flow *
91 = NSVa *
92 = S&W volume flow *
93 = Ref.dens.altern. *
94 = Oil corr.vol.fl. *
95 = Water corr.v.fl. *
96 = Targ.corr.vol.fl *
97 = Target vol. flow *
99 = Oil volume flow *
101 = Water cut *
102 = Water vol. flow *

Current span (0353–1 to n) 1: 5923 Integer Read / Write 0 = 4...20 mA 118


2: 5924 1 = 4...20 mA US
3: 5925 2 = 4...20 mA NAMUR
3 = 0...20 mA
4 = Fixed current

Fixed current (0365–1 to n) 1: 5987 to 5988 Float Read / Write 0 to 22.5 mA 119


2: 5989 to 5990
3: 5991 to 5992

0/4 mA value (0367–1 to n) 1: 6195 to 6196 Float Read / Write Signed floating-point number 120
2: 6197 to 6198
3: 6199 to 6200

Endress+Hauser 291
Modbus RS485 Register Information Proline Promass 500 Modbus RS485

Navigation: Expert → Output → Curr.output 1 to n

Parameter Register Data type Access Selection / User entry / User interface → 

20 mA value (0372–1 to n) 1: 5915 to 5916 Float Read / Write Signed floating-point number 121
2: 5917 to 5918
3: 5919 to 5920

Measuring mode (0351–1 to n) 1: 5899 Integer Read / Write 0 = Forward flow 122


2: 5900 2 = Rev. flow comp.
3: 5901 13 = Forward/Reverse *

Damping out. 1 to n (0363–1 to n) 1: 5903 to 5904 Float Read / Write 0.0 to 999.9 s 127
2: 5905 to 5906
3: 5907 to 5908

Response time (0378–1 to n) Float Read Positive floating-point number 127

Failure mode (0364–1 to n) 1: 5911 Integer Read / Write 0 = Min. 128


2: 5912 1 = Max.
3: 5913 4 = Actual value
5 = Last valid value
6 = Defined value

Failure current (0352–1 to n) 1: 5979 to 5980 Float Read / Write 0 to 22.5 mA 129


2: 5981 to 5982
3: 5983 to 5984

Output curr. 1 to n (0361–1 to n) 1: 5931 to 5932 Float Read 3.59 to 22.5 mA 130
2: 5933 to 5934
3: 5935 to 5936

Measur. curr. 1 to n (0366–1 to n) 1: 5779 to 5780 Float Read 0 to 30 mA 130


2: 5781 to 5782
3: 5783 to 5784

* Visibility depends on order options or device settings

"PFS output 1 to n" submenu

Navigation: Expert → Output → PFS output 1 to n

Parameter Register Data type Access Selection / User entry / User interface → 

Terminal no. (0492–1 to n) 1: 6551 Integer Read 0 = Not used 131


2: 6552 2 = 24-25 (I/O 2)
3: 6553 3 = 22-23 (I/O 3)
4 = 20-21 (I/O 4) *

Signal mode (0490–1 to n) 1: 6235 Integer Read / Write 0 = Passive 132


2: 6236 2 = Active
3: 6237 3 = Passive NAMUR

Operating mode (0469–1 to n) 1: 4479 Integer Read / Write 0 = Pulse 132


2: 4480 1 = Switch
3: 9907 12 = Frequency

292 Endress+Hauser
Proline Promass 500 Modbus RS485 Modbus RS485 Register Information

Navigation: Expert → Output → PFS output 1 to n

Parameter Register Data type Access Selection / User entry / User interface → 

Assign pulse 1 to n (0460–1 to n) 1: 2461 Integer Read / Write 0 = Off 134


2: 2462 1 = Mass flow
3: 4685 2 = Volume flow
3 = Correct.vol.flow *
74 = Target mass flow *
75 = Carrier mass fl. *
78 = Carrier vol. fl. *
79 = Carr.corr.vol.fl *
86 = GSV flow *
87 = GSVa *
88 = Oil mass flow *
89 = Water mass flow *
90 = NSV flow *
91 = NSVa *
92 = S&W volume flow *
94 = Oil corr.vol.fl. *
95 = Water corr.v.fl. *
96 = Targ.corr.vol.fl *
97 = Target vol. flow *
99 = Oil volume flow *
102 = Water vol. flow *

Value per pulse (0455–1 to n) 1: 3034 to 3035 Float Read / Write Positive floating point number 134
2: 3036 to 3037
3: 4714 to 4715

Pulse width (0452–1 to n) 1: 2836 to 2837 Float Read / Write 0.05 to 2 000 ms 135
2: 2838 to 2839
3: 4702 to 4703

Measuring mode (0457–1 to n) 1: 2394 Integer Read / Write 0 = Forward flow 136


2: 2395 1 = Reverse flow
3: 4683 2 = Rev. flow comp.
13 = Forward/Reverse

Failure mode (0480–1 to n) 1: 2948 Integer Read / Write 0 = Actual value 136


2: 2949 1 = No pulses
3: 4708

Pulse output 1 to n (0456–1 to n) 1: 3082 to 3083 Float Read Positive floating-point number 137
2: 3084 to 3085
3: 4718 to 4719

Endress+Hauser 293
Modbus RS485 Register Information Proline Promass 500 Modbus RS485

Navigation: Expert → Output → PFS output 1 to n

Parameter Register Data type Access Selection / User entry / User interface → 

Assign freq. (0478–1 to n) 1: 2614 Integer Read / Write 0 = Off 138


2: 2615 1 = Mass flow
3: 9915 2 = Volume flow
3 = Correct.vol.flow
4 = Density
5 = Ref.density
6 = Osc. ampl. 1 *
7 = Temperature
8 = Osc. ampl. 0 *
9 = Signal asymmetry *
23 = Carr. pipe temp. *
31 = Osc.damp.fluct 1 *
32 = Exc. current 0 *
33 = Exc. current 1 *
39 = Electronic temp.
45 = Kinematic visc. *
46 = Dynam. viscosity *
48 = Osc. freq. 0
50 = Osc. freq. 1 *
52 = Freq. fluct. 1 *
58 = Spec. output 0 *
59 = Spec. output 1 *
60 = Index inh.medium
61 = Index sus.bubble *
63 = Osc. damping 0 *
64 = Osc. damping 1 *
66 = Pressure
67 = Osc.damp.fluct 0 *
68 = Freq. fluct. 0 *
73 = Concentration *
74 = Target mass flow *
75 = Carrier mass fl. *
76 = TempCompDynVisc *
77 = TempCompKinVisc *
78 = Carrier vol. fl. *
79 = Carr.corr.vol.fl *
81 = HBSI *
83 = Oil density *
84 = Water density *
86 = GSV flow *
87 = GSVa *
88 = Oil mass flow *
89 = Water mass flow *
90 = NSV flow *
91 = NSVa *
92 = S&W volume flow *
93 = Ref.dens.altern. *
94 = Oil corr.vol.fl. *
95 = Water corr.v.fl. *
96 = Targ.corr.vol.fl *
97 = Target vol. flow *
99 = Oil volume flow *
101 = Water cut *
102 = Water vol. flow *

Min. freq. value (0453–1 to n) 1: 3526 to 3527 Float Read / Write 0.0 to 10 000.0 Hz 139
2: 3528 to 3529
3: 5767 to 5768

Max. freq. value (0454–1 to n) 1: 2996 to 2997 Float Read / Write 0.0 to 10 000.0 Hz 139
2: 2998 to 2999
3: 4710 to 4711

Val. at min.freq (0476–1 to n) 1: 5887 to 5888 Float Read / Write Signed floating-point number 140
2: 5889 to 5890
3: 5891 to 5892

294 Endress+Hauser
Proline Promass 500 Modbus RS485 Modbus RS485 Register Information

Navigation: Expert → Output → PFS output 1 to n

Parameter Register Data type Access Selection / User entry / User interface → 

Val. at max.freq (0475–1 to n) 1: 3514 to 3515 Float Read / Write Signed floating-point number 140
2: 3516 to 3517
3: 5759 to 5760

Measuring mode (0479–1 to n) 1: 2922 Integer Read / Write 0 = Forward flow 140


2: 2923 2 = Rev. flow comp.
3: 4706 13 = Forward/Reverse

Damping out. 1 to n (0477–1 to n) 1: 3522 to 3523 Float Read / Write 0 to 999.9 s 142


2: 3524 to 3525
3: 5763 to 5764

Response time (0491–1 to n) 1: 5875 to 5876 Float Read Positive floating-point number 143
2: 5877 to 5878
3: 5879 to 5880

Failure mode (0451–1 to n) 1: 2367 Integer Read / Write 0 = Actual value 144


2: 2368 1 = 0 Hz
3: 4681 2 = Defined value

Failure freq. (0474–1 to n) 1: 3510 to 3511 Float Read / Write 0.0 to 12 500.0 Hz 144
2: 3512 to 3513
3: 9908 to 9909

Output freq. 1 to n (0471–1 to n) 1: 3462 to 3463 Float Read 0.0 to 12 500.0 Hz 145
2: 3464 to 3465
3: 9910 to 9911

Switch out funct (0481–1 to n) 1: 3022 Integer Read / Write 0 = Off 145


2: 3023 1 = On
3: 9914 2 = Diag. behavior
3 = Fl. direct.check
4 = Limit
5 = Status

Assign diag. beh (0482–1 to n) 1: 3096 Integer Read / Write 0 = Alarm 146


2: 3097 1 = Warning
3: 9913 2 = Alarm or warning

Endress+Hauser 295
Modbus RS485 Register Information Proline Promass 500 Modbus RS485

Navigation: Expert → Output → PFS output 1 to n

Parameter Register Data type Access Selection / User entry / User interface → 

Assign limit (0483–1 to n) 1: 3184 Integer Read / Write 1 = Mass flow 146


2: 3185 2 = Volume flow
3: 4722 3 = Correct.vol.flow *
4 = Density
5 = Ref.density *
7 = Temperature
16 = Totalizer 1
17 = Totalizer 2
18 = Totalizer 3
45 = Kinematic visc. *
46 = Dynam. viscosity *
58 = Spec. output 0 *
59 = Spec. output 1 *
60 = Index inh.medium
61 = Index sus.bubble *
66 = Pressure
73 = Concentration *
74 = Target mass flow *
75 = Carrier mass fl. *
76 = TempCompDynVisc *
77 = TempCompKinVisc *
78 = Carrier vol. fl. *
79 = Carr.corr.vol.fl *
83 = Oil density *
84 = Water density *
86 = GSV flow *
87 = GSVa *
88 = Oil mass flow *
89 = Water mass flow *
90 = NSV flow *
91 = NSVa *
92 = S&W volume flow *
93 = Ref.dens.altern. *
94 = Oil corr.vol.fl. *
95 = Water corr.v.fl. *
96 = Targ.corr.vol.fl *
97 = Target vol. flow *
99 = Oil volume flow *
100 = Oscil. damping
101 = Water cut *
102 = Water vol. flow *

Switch-on value (0466–1 to n) 1: 3242 to 3243 Float Read / Write Signed floating-point number 149
2: 3244 to 3245
3: 4728 to 4729

Switch-off value (0464–1 to n) 1: 3234 to 3235 Float Read / Write Signed floating-point number 149
2: 3236 to 3237
3: 4724 to 4725

Assign dir.check (0484–1 to n) 1: 3363 Integer Read / Write 0 = Off 150


2: 3364 1 = Mass flow
3: 4732 2 = Volume flow
3 = Correct.vol.flow *

Assign status (0485–1 to n) 1: 3374 Integer Read / Write 0 = Low flow cut off 150
2: 3375 1 = Partial pipe det
3: 4734

Switch-on delay (0467–1 to n) 1: 6247 to 6248 Float Read / Write 0.0 to 100.0 s 150
2: 6249 to 6250
3: 6251 to 6252

Switch-off delay (0465–1 to n) 1: 6239 to 6240 Float Read / Write 0.0 to 100.0 s 151
2: 6241 to 6242
3: 6243 to 6244

Failure mode (0486–1 to n) 1: 3384 Integer Read / Write 0 = Actual status 151


2: 3385 1 = Open
3: 9912 6 = Closed

296 Endress+Hauser
Proline Promass 500 Modbus RS485 Modbus RS485 Register Information

Navigation: Expert → Output → PFS output 1 to n

Parameter Register Data type Access Selection / User entry / User interface → 

Switch status 1 to n (0461–1 to n) 1: 2485 Integer Read 1 = Open 151


2: 2486 6 = Closed
3: 9917

Invert outp.sig. (0470–1 to n) 1: 2583 Integer Read / Write 0 = Yes 152


2: 2584 1 = No
3: 9916

* Visibility depends on order options or device settings

"Relay output 1 to n" submenu

Navigation: Expert → Output → Relay output 1 to n

Parameter Register Data type Access Selection / User entry / User interface → 

Terminal no. (0812–1 to n) 1: 8278 Integer Read 0 = Not used 153


2: 8279 2 = 24-25 (I/O 2)
3: 8280 3 = 22-23 (I/O 3)
4 = 20-21 (I/O 4)

Relay outp.func. (0804–1 to n) 1: 2488 Integer Read / Write 1 = Open 153


2: 2489 2 = Diag. behavior
3: 9876 3 = Fl. direct.check
4 = Limit
5 = Digital Output
6 = Closed

Assign dir.check (0808–1 to n) 1: 8251 Integer Read / Write 0 = Off 154


2: 8252 1 = Mass flow
3: 8253 2 = Volume flow
3 = Correct.vol.flow *

Endress+Hauser 297
Modbus RS485 Register Information Proline Promass 500 Modbus RS485

Navigation: Expert → Output → Relay output 1 to n

Parameter Register Data type Access Selection / User entry / User interface → 

Assign limit (0807–1 to n) 1: 8248 Integer Read / Write 1 = Mass flow 154


2: 8249 2 = Volume flow
3: 8250 3 = Correct.vol.flow *
4 = Density
5 = Ref.density *
7 = Temperature
16 = Totalizer 1
17 = Totalizer 2
18 = Totalizer 3
45 = Kinematic visc. *
46 = Dynam. viscosity *
58 = Spec. output 0 *
59 = Spec. output 1 *
60 = Index inh.medium
61 = Index sus.bubble *
66 = Pressure
73 = Concentration *
74 = Target mass flow *
75 = Carrier mass fl. *
76 = TempCompDynVisc *
77 = TempCompKinVisc *
78 = Carrier vol. fl. *
79 = Carr.corr.vol.fl *
83 = Oil density *
84 = Water density *
86 = GSV flow *
87 = GSVa *
88 = Oil mass flow *
89 = Water mass flow *
90 = NSV flow *
91 = NSVa *
92 = S&W volume flow *
93 = Ref.dens.altern. *
94 = Oil corr.vol.fl. *
95 = Water corr.v.fl. *
96 = Targ.corr.vol.fl *
97 = Target vol. flow *
99 = Oil volume flow *
100 = Oscil. damping
101 = Water cut *
102 = Water vol. flow *

Assign diag. beh (0806–1 to n) 1: 8245 Integer Read / Write 0 = Alarm 155


2: 8246 1 = Warning
3: 8247 2 = Alarm or warning

Assign status (0805–1 to n) 1: 8272 Integer Read / Write 0 = Low flow cut off 156
2: 8273 1 = Partial pipe det
3: 8274

Switch-off value (0809–1 to n) 1: 8260 to 8261 Float Read / Write Signed floating-point number 156
2: 8262 to 8263
3: 8264 to 8265

Switch-off delay (0813–1 to n) 1: 8254 to 8255 Float Read / Write 0.0 to 100.0 s 157
2: 8256 to 8257
3: 8258 to 8259

Switch-on value (0810–1 to n) 1: 8233 to 8234 Float Read / Write Signed floating-point number 157
2: 8235 to 8236
3: 8237 to 8238

Switch-on delay (0814–1 to n) 1: 8266 to 8267 Float Read / Write 0.0 to 100.0 s 158
2: 8268 to 8269
3: 8270 to 8271

Failure mode (0811–1 to n) 1: 8242 Integer Read / Write 0 = Actual status 158


2: 8243 1 = Open
3: 8244 6 = Closed

298 Endress+Hauser
Proline Promass 500 Modbus RS485 Modbus RS485 Register Information

Navigation: Expert → Output → Relay output 1 to n

Parameter Register Data type Access Selection / User entry / User interface → 

Switch status (0801–1 to n) 1: 3518 Integer Read 1 = Open 158


2: 3519 6 = Closed
3: 9875

Act. relay state (0816–1 to n) 1: 7009 Integer Read / Write 1 = Open 159


2: 7010 6 = Closed
3: 7011

* Visibility depends on order options or device settings

"Double pulse out" submenu

Navigation: Expert → Output → Double pulse out

Parameter Register Data type Access Selection / User entry / User interface → 

Master term. no. (0981) 5838 Integer Read 0 = Not used 160
2 = 24-25 (I/O 2)
3 = 22-23 (I/O 3)

Slave term. no. (0990) 5845 Integer Read 0 = Not used 160
2 = 24-25 (I/O 2)
3 = 22-23 (I/O 3)

Signal mode (0991) 5949 Integer Read / Write 0 = Passive 160


2 = Active
3 = Passive NAMUR

Assign pulse 1 (0982–1) 5993 Integer Read / Write 0 = Off 161


1 = Mass flow
2 = Volume flow
3 = Correct.vol.flow *
74 = Target mass flow *
75 = Carrier mass fl. *
78 = Carrier vol. fl. *
79 = Carr.corr.vol.fl *
86 = GSV flow *
87 = GSVa *
88 = Oil mass flow *
89 = Water mass flow *
90 = NSV flow *
91 = NSVa *
92 = S&W volume flow *
94 = Oil corr.vol.fl. *
95 = Water corr.v.fl. *
96 = Targ.corr.vol.fl *
97 = Target vol. flow *
99 = Oil volume flow *
102 = Water vol. flow *

Value per pulse (0983) 7495 to 7496 Float Read / Write Signed floating-point number 161

Pulse width (0986) 6998 to 6999 Float Read / Write 0.5 to 2 000 ms 162

Phase shift (0992) 6089 Integer Read / Write 0 = 90° 162


1 = 180°

Measuring mode (0984) 6001 Integer Read / Write 0 = Forward flow 162


1 = Reverse flow
2 = Rev. flow comp.
13 = Forward/Reverse

Failure mode (0985) 6009 Integer Read / Write 0 = Actual value 163


1 = No pulses

Endress+Hauser 299
Modbus RS485 Register Information Proline Promass 500 Modbus RS485

Navigation: Expert → Output → Double pulse out

Parameter Register Data type Access Selection / User entry / User interface → 

Pulse output (0987) 7041 to 7042 Float Read Positive floating-point number 164

Invert outp.sig. (0993) 6101 Integer Read / Write 0 = Yes 164


1 = No

* Visibility depends on order options or device settings

6.3.6 "Communication" submenu

"Modbus config." submenu

Navigation: Expert → Communication → Modbus config.

Parameter Register Data type Access Selection / User entry / User interface → 

Bus address (7112) 4910 Integer Read / Write 1 to 247 165

Baudrate (7111) 4912 Integer Read / Write 0 = 1200 BAUD 166


1 = 2400 BAUD
2 = 4800 BAUD
3 = 9600 BAUD
4 = 19200 BAUD
5 = 38400 BAUD
6 = 57600 BAUD
7 = 115200 BAUD

Data trans. mode (7115) 4913 Integer Read / Write 0 = RTU 166


1 = ASCII

Parity (7122) 4914 Integer Read / Write 0 = Even 166


1 = Odd
2 = None/2 stop bits
3 = None/1 stop bit

Byte order (7113) 4915 Integer Read / Write 0 = 0-1-2-3 167


1 = 3-2-1-0
2 = 2-3-0-1
3 = 1-0-3-2

Telegram delay (7146) 4916 to 4917 Float Read / Write 0 to 100 ms 168

Failure mode (7116) 4920 Integer Read / Write 1 = Last valid value 168
255 = NaN value

Bus termination (7155) 5774 Integer Read 0 = Off 169


1 = On

Fieldb.writ.acc. (7156) 6807 Integer Read / Write 0 = Read + write 169


1 = Read only

"Modbus info" submenu

Navigation: Expert → Communication → Modbus info

Parameter Register Data type Access Selection / User entry / User interface → 

Device ID (7153) 2547 Integer Read 4-digit hexadecimal number 170

Device revision (7154) 4481 Integer Read 4-digit hexadecimal number 170

300 Endress+Hauser
Proline Promass 500 Modbus RS485 Modbus RS485 Register Information

"Modbus data map" submenu

Navigation: Expert → Communication → Modbus data map

Parameter Register Data type Access Selection / User entry / User interface → 

Scan list reg.0 to 15 (7114) 0: 5001 Integer Read / Write 1 to 65 535 171


1: 5002
2: 5003
3: 5004
4: 5005
5: 5006
6: 5007
7: 5008
8: 5009
9: 5010
10: 5011
11: 5012
12: 5013
13: 5014
14: 5015
15: 5016

"Web server" submenu

Navigation: Expert → Communication → Web server

Parameter Register Data type Access Selection / User entry / User interface → 

Webserv.language (7221) 4219 Integer Read / Write 0 = English 172


1 = Deutsch
2 = Français
3 = Español
4 = Italiano
5 = Nederlands
8 = Svenska
10 = Bahasa Indonesia
11 = 日本語 (Japanese)
12 = Portuguesa
13 = Polski
14 = русский язык(Ru)
15 = čeština (Czech)
16 = 中文 (Chinese)
18 = Türkçe
19 = tiếng Việt (Vit)
20 = 한국어 (Korean)

MAC Address (7214) 4210 to 4217.5 String Read Unique 12-digit character string 172
comprising letters and numbers

IP address (7209) 4155 to 4161.5 String Read 4 octet: 0 to 255 (in the particular octet) 173

Subnet mask (7211) 4163 to 4169.5 String Read 4 octet: 0 to 255 (in the particular octet) 173

Default gateway (7210) 4171 to 4177.5 String Read 4 octet: 0 to 255 (in the particular octet) 173

Webserver funct. (7222) 4220 Integer Read / Write 0 = Off 174


1 = On
2 = HTML Off

Login page (7273) 5802 Integer Read / Write 0 = Without header 174


1 = With header

Endress+Hauser 301
Modbus RS485 Register Information Proline Promass 500 Modbus RS485

"WLAN settings" submenu

Navigation: Expert → Communication → WLAN settings

Parameter Register Data type Access Selection / User entry / User interface → 

WLAN (2702) 6178 Integer Read / Write 0 = Disable 175


1 = Enable

WLAN IP address (2711) 8643 to 8649.5 String Read / Write 4 octet: 0 to 255 (in the particular octet) 177

WLAN subnet mask (2709) 8651 to 8657.5 String Read / Write 4 octet: 0 to 255 (in the particular octet) 178

WLAN MAC address (2703) 8602 to 8609.5 String Read Unique 12-digit character string 178
comprising letters and numbers

Security type (2705) 6206 Integer Read / Write 0 = Unsecured 176


1 = WPA2-PSK
2 = EAP-PEAP MSCHAP2 *
3 = EAP-TLS *
4 = EAP-PEAP NoAuth. *

WLAN passphrase (2706) 8611 to 8626 String Read / Write 8 to 32-digit character string comprising 178
numbers, letters and special characters
(without spaces)

Assign SSID name (2708) 6218 Integer Read / Write 0 = Device tag 179
1 = User-defined

SSID name (2707) 8627 to 8642 String Read / Write Max. 32-digit character string 179
comprising numbers, letters and special
characters

Select antenna (2713) 6102 Integer Read / Write 0 = External antenna 180


1 = Internal antenna

WLAN channel (2704) 6182 Integer Read / Write 1 to 11 179

Apply changes (2712) Integer Read / Write

* Visibility depends on order options or device settings

6.3.7 "Application" submenu

Navigation: Expert → Application

Parameter Register Data type Access Selection / User entry / User interface → 

Reset all tot. (2806) 2609 Integer Read / Write 0 = Cancel 181


1 = Reset + totalize

302 Endress+Hauser
Proline Promass 500 Modbus RS485 Modbus RS485 Register Information

"Totalizer 1 to n" submenu

Navigation: Expert → Application → Totalizer 1 to n

Parameter Register Data type Access Selection / User entry / User interface → 

Assign variable (0914–1 to n) 1: 2601 Integer Read / Write 0 = Off 182


2: 2801 1 = Mass flow
3: 3001 2 = Volume flow
3 = Correct.vol.flow *
74 = Target mass flow *
75 = Carrier mass fl. *
78 = Carrier vol. fl. *
79 = Carr.corr.vol.fl *
86 = GSV flow *
87 = GSVa *
88 = Oil mass flow *
89 = Water mass flow *
90 = NSV flow *
91 = NSVa *
92 = S&W volume flow *
94 = Oil corr.vol.fl. *
95 = Water corr.v.fl. *
96 = Targ.corr.vol.fl *
97 = Target vol. flow *
99 = Oil volume flow *
102 = Water vol. flow *

Unit totalizer 1 to n (0915–1 to n) 1: 4604 Integer Read / Write 0 = cm³ * 183


2: 4605 1 = dm³ *
3: 4606 2 = m³ *
3 = ml *
4 = l*
5 = hl *
6 = Ml Mega *
8 = af *
9 = ft³ *
10 = fl oz (us) *
11 = gal (us) *
12 = Mgal (us) *
13 = bbl (us;liq.) *
14 = bbl (us;beer) *
15 = bbl (us;oil) *
16 = bbl (us;tank) *
17 = gal (imp) *
18 = Mgal (imp) *
19 = bbl (imp;beer) *
20 = bbl (imp;oil) *
22 = kgal (us) *
23 = Mft³ *
50 = g *
51 = kg *
52 = t *
53 = oz *
54 = lb *
55 = STon *
100 = Nl *
101 = Nm³ *
102 = Sm³ *
103 = Sft³ *
104 = Sl *
105 = Sgal (us) *
106 = Sbbl (us;liq.) *
107 = Sgal (imp) *
108 = Sbbl (us;oil) *
109 = MMSft³ *
110 = Nhl *
251 = None *

Operation mode (0908–1 to n) 1: 2605 Integer Read / Write 0 = Net flow total 184
2: 2805 1 = Forward total
3: 3005 2 = Reverse total

Endress+Hauser 303
Modbus RS485 Register Information Proline Promass 500 Modbus RS485

Navigation: Expert → Application → Totalizer 1 to n

Parameter Register Data type Access Selection / User entry / User interface → 

Control Tot. 1 to n (0912–1 to n) 1: 2608 Integer Read / Write 0 = Totalize 185


2: 2808 1 = Reset + totalize
3: 3008 2 = Preset + hold *
3 = Reset + hold *
4 = Preset+totalize *
5 = Hold *

Preset value 1 to n (0913–1 to n) 1: 2590 to 2591 Float Read / Write Signed floating-point number 185
2: 2592 to 2593
3: 2594 to 2595

Failure mode (0901–1 to n) 1: 2606 Integer Read / Write 0 = Stop 186


2: 2806 1 = Actual value
3: 3006 2 = Last valid value

* Visibility depends on order options or device settings

6.3.8 "Diagnostics" submenu

Navigation: Expert → Diagnostics

Parameter Register Data type Access Selection / User entry / User interface → 

Actual diagnos. (0691) 2732 Integer Read Symbol for diagnostic behavior, 196
diagnostic code and short message.

Prev.diagnostics (0690) 2734 Integer Read Symbol for diagnostic behavior, 196
diagnostic code and short message.

Time fr. restart (0653) 2624 Integer Read Days (d), hours (h), minutes (m) and 197
seconds (s)

Operating time (0652) 2631 Integer Read Days (d), hours (h), minutes (m) and 197
seconds (s)

"Diagnostic list" submenu

Navigation: Expert → Diagnostics → Diagnostic list

Parameter Register Data type Access Selection / User entry / User interface → 

Diagnostics 1 (0692) 2736 Integer Read Symbol for diagnostic behavior, 198
diagnostic code and short message.

Diagnostics 2 (0693) 2738 Integer Read Symbol for diagnostic behavior, 199
diagnostic code and short message.

Diagnostics 3 (0694) 2740 Integer Read Symbol for diagnostic behavior, 200
diagnostic code and short message.

Diagnostics 4 (0695) 2742 Integer Read Symbol for diagnostic behavior, 200
diagnostic code and short message.

Diagnostics 5 (0696) 2744 Integer Read Symbol for diagnostic behavior, 201
diagnostic code and short message.

304 Endress+Hauser
Proline Promass 500 Modbus RS485 Modbus RS485 Register Information

"Event logbook" submenu

Navigation: Expert → Diagnostics → Event logbook

Parameter Register Data type Access Selection / User entry / User interface → 

Filter options (0705) 4596 Integer Read / Write 0 = Failure (F) 202


4 = Mainten. req.(M)
8 = Funct. check (C)
12 = Out of spec. (S)
16 = Information (I)
255 = All

"Device info" submenu

Navigation: Expert → Diagnostics → Device info

Parameter Register Data type Access Selection / User entry / User interface → 

Device tag (0011) 2026 to 2041 String Read Max. 32 characters, such as letters, 205
numbers or special characters (e.g. @,
%, /).

Serial number (0009) 7003 to 7007.5 String Read Max. 11-digit character string 206
comprising letters and numbers.

Firmware version (0010) 7277 to 7280 String Read Character string in the format xx.yy.zz 206

Device name (0020) 7238 to 7245 String Read Promass 300/500 206

Order code (0008) 2058 to 2067 String Read Character string composed of letters, 206
numbers and certain punctuation marks
(e.g. /).

Ext. order cd. 1 (0023) 2212 to 2221 String Read Character string 207

Ext. order cd. 2 (0021) 2222 to 2231 String Read Character string 207

Ext. order cd. 3 (0022) 2232 to 2241 String Read Character string 207

ENP version (0012) 4003 to 4010 String Read Character string 208

"Mainboard module" submenu

Navigation: Expert → Diagnostics → Mainboard module

Parameter Register Data type Access Selection / User entry / User interface → 

Software rev. (0072) Integer Read

"Sens. electronic" submenu

Navigation: Expert → Diagnostics → Sens. electronic

Parameter Register Data type Access Selection / User entry / User interface → 

Software rev. (0072) 7039 Integer Read Positive integer 209

Endress+Hauser 305
Modbus RS485 Register Information Proline Promass 500 Modbus RS485

"I/O module 1" submenu

Navigation: Expert → Diagnostics → I/O module 1

Parameter Register Data type Access Selection / User entry / User interface → 

I/O 1 terminals (3902–1) 6541 Integer Read 0 = Not used 210


1 = 26-27 (I/O 1)
2 = 24-25 (I/O 2)
3 = 22-23 (I/O 3)
4 = 20-21 (I/O 4) *

Software rev. (0072) 7039 Integer Read Positive integer 211

* Visibility depends on order options or device settings

"I/O module 2" submenu

Navigation: Expert → Diagnostics → I/O module 2

Parameter Register Data type Access Selection / User entry / User interface → 

I/O 2 terminals (3902–2) 6542 Integer Read 0 = Not used 210


1 = 26-27 (I/O 1)
2 = 24-25 (I/O 2)
3 = 22-23 (I/O 3)
4 = 20-21 (I/O 4) *

Software rev. (0072) 7039 Integer Read Positive integer 211

* Visibility depends on order options or device settings

"I/O module 3" submenu

Navigation: Expert → Diagnostics → I/O module 3

Parameter Register Data type Access Selection / User entry / User interface → 

I/O 3 terminals (3902–3) 6543 Integer Read 0 = Not used 210


1 = 26-27 (I/O 1)
2 = 24-25 (I/O 2)
3 = 22-23 (I/O 3)
4 = 20-21 (I/O 4) *

Software rev. (0072) 7039 Integer Read Positive integer 211

* Visibility depends on order options or device settings

"I/O module 4" submenu

Navigation: Expert → Diagnostics → I/O module 4

Parameter Register Data type Access Selection / User entry / User interface → 

I/O 4 terminals (3902–4) 6544 Integer Read 0 = Not used 210


1 = 26-27 (I/O 1)
2 = 24-25 (I/O 2)
3 = 22-23 (I/O 3)
4 = 20-21 (I/O 4) *

Software rev. (0072) 7039 Integer Read Positive integer 211

* Visibility depends on order options or device settings

306 Endress+Hauser
Proline Promass 500 Modbus RS485 Modbus RS485 Register Information

"Display module" submenu

Navigation: Expert → Diagnostics → Display module

Parameter Register Data type Access Selection / User entry / User interface → 

Software rev. (0072) 7039 Integer Read Positive integer 216

"Min/max val." submenu

Navigation: Expert → Diagnostics → Min/max val.

Parameter Register Data type Access Selection / User entry / User interface → 

Reset min/max (6151) 2525 Integer Read / Write 0 = Cancel 226


6 = Osc. ampl. 1 *
8 = Oscil. amplitude *
9 = Tors.oscil.freq. *
10 = Oscil. damping
11 = Tors.oscil.damp. *
12 = Oscil. frequency
13 = Signal asymmetry
14 = Tors.sig.asymm. *

* Visibility depends on order options or device settings

"Main elect.temp." submenu

Navigation: Expert → Diagnostics → Min/max val. → Main elect.temp.

Parameter Register Data type Access Selection / User entry / User interface → 

Minimum value (0688) 4651 to 4652 Float Read Signed floating-point number 227

Maximum value (0665) 4649 to 4650 Float Read Signed floating-point number 227

"Sensor elec.temp" submenu

Navigation: Expert → Diagnostics → Min/max val. → Sensor elec.temp

Parameter Register Data type Access Selection / User entry / User interface → 

Minimum value (6052) 2421 to 2422 Float Read Signed floating-point number 228

Maximum value (6051) 2419 to 2420 Float Read Signed floating-point number 228

"Medium temp." submenu

Navigation: Expert → Diagnostics → Min/max val. → Medium temp.

Parameter Register Data type Access Selection / User entry / User interface → 

Minimum value (6109) 7529 to 7530 Float Read Signed floating-point number 229

Maximum value (6108) 7531 to 7532 Float Read Signed floating-point number 229

"Carr. pipe temp." submenu

Navigation: Expert → Diagnostics → Min/max val. → Carr. pipe temp.

Parameter Register Data type Access Selection / User entry / User interface → 

Minimum value (6030) 7533 to 7534 Float Read Signed floating-point number 230

Maximum value (6029) 7535 to 7536 Float Read Signed floating-point number 230

Endress+Hauser 307
Modbus RS485 Register Information Proline Promass 500 Modbus RS485

"Oscil. frequency" submenu

Navigation: Expert → Diagnostics → Min/max val. → Oscil. frequency

Parameter Register Data type Access Selection / User entry / User interface → 

Minimum value (6071) 2459 to 2460 Float Read Signed floating-point number 231

Maximum value (6070) 2468 to 2469 Float Read Signed floating-point number 231

"Tors.oscil.freq." submenu

Navigation: Expert → Diagnostics → Min/max val. → Tors.oscil.freq.

Parameter Register Data type Access Selection / User entry / User interface → 

Minimum value (6069) 2519 to 2520 Float Read Signed floating-point number 232

Maximum value (6068) 2517 to 2518 Float Read Signed floating-point number 232

"Oscil. amplitude" submenu

Navigation: Expert → Diagnostics → Min/max val. → Oscil. amplitude

Parameter Register Data type Access Selection / User entry / User interface → 

Minimum value (6010) 2472 to 2473 Float Read Signed floating-point number 232

Maximum value (6009) 2470 to 2471 Float Read Signed floating-point number 233

"Tor. osc. amp." submenu

Navigation: Expert → Diagnostics → Min/max val. → Tor. osc. amp.

Parameter Register Data type Access Selection / User entry / User interface → 

Minimum value (6008) 2515 to 2516 Float Read Signed floating-point number 233

Maximum value (6007) 2480 to 2481 Float Read Signed floating-point number 233

"Oscil. damping" submenu

Navigation: Expert → Diagnostics → Min/max val. → Oscil. damping

Parameter Register Data type Access Selection / User entry / User interface → 

Minimum value (6122) 2478 to 2479 Float Read Signed floating-point number 234

Maximum value (6121) 2423 to 2424 Float Read Signed floating-point number 234

"Tors.oscil.damp." submenu

Navigation: Expert → Diagnostics → Min/max val. → Tors.oscil.damp.

Parameter Register Data type Access Selection / User entry / User interface → 

Minimum value (6120) 2523 to 2524 Float Read Signed floating-point number 235

Maximum value (6119) 2521 to 2522 Float Read Signed floating-point number 235

308 Endress+Hauser
Proline Promass 500 Modbus RS485 Modbus RS485 Register Information

"Signal asymmetry" submenu

Navigation: Expert → Diagnostics → Min/max val. → Signal asymmetry

Parameter Register Data type Access Selection / User entry / User interface → 

Minimum value (6015) 2474 to 2475 Float Read Signed floating-point number 235

Maximum value (6014) 2476 to 2477 Float Read Signed floating-point number 236

"Tors.sig.asymm." submenu

Navigation: Expert → Diagnostics → Min/max val. → Tors.sig.asymm.

Parameter Register Data type Access Selection / User entry / User interface → 

Minimum value (6284) 7000 to 7001 Float Read Signed floating-point number 236

Maximum value (6283) 6924 to 6925 Float Read Signed floating-point number 236

Endress+Hauser 309
Modbus RS485 Register Information Proline Promass 500 Modbus RS485

"Data logging" submenu

Navigation: Expert → Diagnostics → Data logging

Parameter Register Data type Access Selection / User entry / User interface → 

Assign chan. 1 (0851) 2445 Integer Read / Write 0 = Off 217


1 = Mass flow
2 = Volume flow
3 = Correct.vol.flow *
4 = Density
5 = Ref.density *
6 = Osc. ampl. 1 *
7 = Temperature
8 = Oscil. amplitude *
9 = Signal asymmetry *
23 = Carr. pipe temp. *
31 = Osc.damp.fluct 1 *
32 = Exc. current 0 *
33 = Exc. current 1 *
39 = Electronic temp.
45 = Kinematic visc. *
46 = Dynam. viscosity *
48 = Osc. freq. 0
50 = Osc. freq. 1 *
52 = Freq. fluct. 1 *
58 = Spec. output 0 *
59 = Spec. output 1 *
60 = Index inh.medium
61 = Index sus.bubble *
63 = Osc. damping 0 *
64 = Osc. damping 1 *
66 = Pressure
67 = Osc.damp.fluct 0 *
68 = Freq. fluct. 0 *
68 = Freq. fluct. 0 *
73 = Concentration *
74 = Target mass flow *
75 = Carrier mass fl. *
76 = TempCompDynVisc *
77 = TempCompKinVisc *
78 = Carrier vol. fl. *
79 = Carr.corr.vol.fl *
81 = HBSI *
83 = Oil density *
84 = Water density *
86 = GSV flow *
87 = GSVa *
88 = Oil mass flow *
89 = Water mass flow *
90 = NSV flow *
91 = NSVa *
92 = S&W volume flow *
93 = Ref.dens.altern. *
94 = Oil corr.vol.fl. *
95 = Water corr.v.fl. *
96 = Targ.corr.vol.fl *
97 = Target vol. flow *
99 = Oil volume flow *
101 = Water cut *
102 = Water vol. flow *
121 = Curr.output 1 *
122 = Curr.output 2 *
123 = Curr.output 3 *

Assign chan. 2 (0852) 2446 Integer Read / Write Picklist, see Assign channel 1 parameter 219
(→  217)

Assign chan. 3 (0853) 2548 Integer Read / Write Picklist, see Assign channel 1 parameter 219
(→  217)

310 Endress+Hauser
Proline Promass 500 Modbus RS485 Modbus RS485 Register Information

Navigation: Expert → Diagnostics → Data logging

Parameter Register Data type Access Selection / User entry / User interface → 

Assign chan. 4 (0854) 4286 Integer Read / Write Picklist, see Assign channel 1 parameter 219
(→  217)

Logging interval (0856) 4288 to 4289 Float Read / Write 0.1 to 3 600.0 s 220

Clear logging (0855) 4287 Integer Read / Write 0 = Cancel 220


2 = Clear data

Data logging (0860) 5950 Integer Read / Write 0 = Overwriting 221


1 = Not overwriting

Logging delay (0859) 5938 Integer Read / Write 0 to 999 h 221

Data log.control (0857) 5930 Integer Read / Write 0 = None 222


1 = Stop
2 = Delete + start

Data log. status (0858) 5937 Integer Read 0 = Done 222


1 = Stopped
2 = Active
3 = Delay active

Logging duration (0861) 2827 to 2828 Float Read Positive floating-point number 223

* Visibility depends on order options or device settings

"Simulation" submenu

Navigation: Expert → Diagnostics → Simulation

Parameter Register Data type Access Selection / User entry / User interface → 

Assign proc.var. (1810) 6813 Integer Read / Write 0 = Off 238


1 = Mass flow
2 = Volume flow
3 = Correct.vol.flow *
4 = Density
5 = Ref.density *
7 = Temperature
45 = Kinematic visc. *
46 = Dynam. viscosity *
73 = Concentration *
74 = Target mass flow *
75 = Carrier mass fl. *
76 = TempCompDynVisc *
77 = TempCompKinVisc *
78 = Carrier vol. fl. *
79 = Carr.corr.vol.fl *
83 = Oil density *
84 = Water density *
86 = GSV flow *
87 = GSVa *
88 = Oil mass flow *
89 = Water mass flow *
90 = NSV flow *
91 = NSVa *
92 = S&W volume flow *
93 = Ref.dens.altern. *
94 = Oil corr.vol.fl. *
95 = Water corr.v.fl. *
96 = Targ.corr.vol.fl *
97 = Target vol. flow *
99 = Oil volume flow *
101 = Water cut *
102 = Water vol. flow *

Proc. var. value (1811) 6814 to 6815 Float Read / Write Depends on the process variable selected 239

Endress+Hauser 311
Modbus RS485 Register Information Proline Promass 500 Modbus RS485

Navigation: Expert → Diagnostics → Simulation

Parameter Register Data type Access Selection / User entry / User interface → 

Status inp.sim (1355) 2620 Integer Read / Write 0 = Off 239


1 = On

Signal level (1356) 2638 Integer Read / Write 0 = Low 240


1 = High

Curr.inp 1 to n sim. (1608–1 to n) 1: 6127 Integer Read / Write 0 = Off 240


2: 6128 1 = On
3: 6129

Value curr.inp 1 to n (1609–1 to n) 1: 6139 to 6140 Float Read / Write 0 to 22.5 mA 241


2: 6141 to 6142
3: 6143 to 6144

Curr.out. 1 to n sim. (0354–1 to n) 1: 5939 Integer Read / Write 0 = Off 241


2: 5940 1 = On
3: 5941

Value curr.out 1 to n (0355–1 to n) 1: 5995 to 5996 Float Read / Write 3.59 to 22.5 mA 241
2: 5997 to 5998
3: 5999 to 6000

FreqOutputSim 1 to n (0472–1 to n) 1: 6203 Integer Read / Write 0 = Off 242


2: 6204 1 = On
3: 6205

Freq value 1 to n (0473–1 to n) 1: 6207 to 6208 Float Read / Write 0.0 to 12 500.0 Hz 242
2: 6209 to 6210
3: 6211 to 6212

Puls.outp.sim. 1 to n (0458–1 to n) 1: 6215 Integer Read / Write 0 = Off 243


2: 6216 1 = Down-count. val.
3: 6217 2 = Fixed value

Pulse value 1 to n (0459–1 to n) 1: 6219 Integer Read / Write 0 to 65 535 243


2: 6220
3: 6221

Switch sim. 1 to n (0462–1 to n) 1: 6223 Integer Read / Write 0 = Off 243


2: 6224 1 = On
3: 6225

Switch status 1 to n (0463–1 to n) 1: 6227 Integer Read / Write 1 = Open 244


2: 6228 6 = Closed
3: 6229

Relay out. 1 to n sim (0802–1 to n) 1: 7523 Integer Read / Write 0 = Off 244


2: 7524 1 = On
3: 7525

Switch status 1 to n (0803–1 to n) 1: 8239 Integer Read / Write 1 = Open 245


2: 8240 6 = Closed
3: 8241

Puls.outp.sim. (0988) 5957 Integer Read / Write 0 = Off 245


1 = Down-count. val.
2 = Fixed value

Pulse value (0989) 5973 Integer Read / Write 0 to 65 535 246

Dev. alarm sim. (0654) 6812 Integer Read / Write 0 = Off 246


1 = On

Event category (0738) 4261 Integer Read / Write 0 = Sensor 247


1 = Electronics
2 = Configuration
3 = Process

Diag. event sim. (0737) 4259 Integer Read / Write • Off 247


• Diagnostic event picklist (depends on
the category selected)

* Visibility depends on order options or device settings

312 Endress+Hauser
Proline Promass 500 Modbus RS485 Index

Index
0…9 Assign frequency output (Parameter) . . . . . . . . . . . . 138
0/4 mA value (Parameter) . . . . . . . . . . . . . . . . 112, 120 Assign limit (Parameter) . . . . . . . . . . . . . . . . . 146, 154
0% bargraph value 1 (Parameter) . . . . . . . . . . . . . . . 21 Assign process variable (Parameter) . . . . . . . 85, 88, 182
0% bargraph value 3 (Parameter) . . . . . . . . . . . . . . . 24 Assign pulse output 1 (Parameter) . . . . . . . . . . . . . . 161
2.4 GHz WLAN channel (Parameter) . . . . . . . . . . . . 179 Assign pulse output 1 to n (Parameter) . . . . . . . . . . 134
20 mA value (Parameter) . . . . . . . . . . . . . . . . 112, 121 Assign simulation process variable (Parameter) . . . . 238
100% bargraph value 1 (Parameter) . . . . . . . . . . . . . 21 Assign SSID name (Parameter) . . . . . . . . . . . . . . . . 179
100% bargraph value 3 (Parameter) . . . . . . . . . . . . . 24 Assign status (Parameter) . . . . . . . . . . . . . . . . 150, 156
Assign status input (Parameter) . . . . . . . . . . . . . . . . 114
A
Access status (Parameter) . . . . . . . . . . . . . . . . . . . . . 14 B
Activate SW option (Parameter) . . . . . . . . . . . . . . . . 48 Backlight (Parameter) . . . . . . . . . . . . . . . . . . . . . . . . 29
Active level (Parameter) . . . . . . . . . . . . . . . . . . . . . 115 Backup state (Parameter) . . . . . . . . . . . . . . . . . . . . . 31
Actual diagnostics (Parameter) . . . . . . . . . . . . . . . . 196 Baudrate (Parameter) . . . . . . . . . . . . . . . . . . . . . . . 166
Administration (Submenu) . . . . . . . . . . . . . . . . . . . . 44 Bootloader revision (Parameter) 209, 210, 211, 212,
Alarm delay (Parameter) . . . . . . . . . . . . . . . . . . . . . . 33 214, 215, 216
Application (Submenu) . . . . . . . . . . . . . . . . . . . . . . 181 Build no. software (Parameter) 209, 210, 211, 212,
Application specific calculations (Submenu) . . . . . . . 187 214, 215, 216
Application specific input 0 (Parameter) . . . . . . . . . 191 Bus address (Parameter) . . . . . . . . . . . . . . . . . . . . . 165
Application specific input 1 (Parameter) . . . . . . . . . 192 Bus termination (Parameter) . . . . . . . . . . . . . . . . . . 169
Application specific output 0 (Parameter) . . . . . . . . . 192 Byte order (Parameter) . . . . . . . . . . . . . . . . . . . . . . 167
Application specific output 1 (Parameter) . . . . . . . . . 193
Application-specific parameters (Submenu) . . . . . . . 188 C
Apply I/O configuration (Parameter) . . . . . . . . . . . . 109 C0 to 5 (Parameter) . . . . . . . . . . . . . . . . . . . . . . . . . 107
Assign behavior of diagnostic no. 046 (Parameter) . . . 35 Calculated values (Submenu) . . . . . . . . . . . . . . . . . . . 95
Assign behavior of diagnostic no. 140 (Parameter) . . . 35 Calibration (Submenu) . . . . . . . . . . . . . . . . . . . . . . 106
Assign behavior of diagnostic no. 144 (Parameter) . . . 36 Calibration factor (Parameter) . . . . . . . . . . . . . . . . . 106
Assign behavior of diagnostic no. 302 (Parameter) . . . 37 Carrier corrected volume flow (Parameter) . . . . . . . . . 56
Assign behavior of diagnostic no. 374 (Parameter) . . . 36 Carrier mass flow (Parameter) . . . . . . . . . . . . . . . . . . 56
Assign behavior of diagnostic no. 441 (Parameter) . . . 37 Carrier pipe temperature (Submenu) . . . . . . . . . . . . 229
Assign behavior of diagnostic no. 442 (Parameter) . . . 37 Carrier volume flow (Parameter) . . . . . . . . . . . . . . . . 57
Assign behavior of diagnostic no. 443 (Parameter) . . . 38 Clear logging data (Parameter) . . . . . . . . . . . . . . . . 220
Assign behavior of diagnostic no. 444 (Parameter) . . . 38 Communication (Submenu) . . . . . . . . . . . . . . . . . . . 164
Assign behavior of diagnostic no. 543 (Parameter) . . . 38 Comparison result (Parameter) . . . . . . . . . . . . . . . . . 32
Assign behavior of diagnostic no. 830 (Parameter) . . . 39 Concentration (Parameter) . . . . . . . . . . . . . . . . . . . . 55
Assign behavior of diagnostic no. 831 (Parameter) . . . 39 Concentration (Submenu) . . . . . . . . . . . . . . . . . . . . 187
Assign behavior of diagnostic no. 832 (Parameter) . . . 39 Configuration backup (Submenu) . . . . . . . . . . . . . . . . 30
Assign behavior of diagnostic no. 833 (Parameter) . . . 40 Configuration management (Parameter) . . . . . . . . . . 31
Assign behavior of diagnostic no. 834 (Parameter) . . . 40 Confirm access code (Parameter) . . . . . . . . . . . . . . . . 46
Assign behavior of diagnostic no. 835 (Parameter) . . . 40 Connection state (Parameter) . . . . . . . . . . . . . . . . . 180
Assign behavior of diagnostic no. 842 (Parameter) . . . 41 Contrast display (Parameter) . . . . . . . . . . . . . . . . . . . 29
Assign behavior of diagnostic no. 862 (Parameter) . . . 41 Control Totalizer 1 to n (Parameter) . . . . . . . . . . . . . 185
Assign behavior of diagnostic no. 912 (Parameter) . . . 42 Corrected volume flow (Parameter) . . . . . . . . . . . . . . 52
Assign behavior of diagnostic no. 913 (Parameter) . . . 42 Corrected volume flow calculation (Parameter) . . . . . . 96
Assign behavior of diagnostic no. 941 (Parameter) . . . 42 Corrected volume flow calculation (Submenu) . . . . . . 95
Assign behavior of diagnostic no. 942 (Parameter) . . . 43 Corrected volume flow factor (Parameter) . . . . . . . . 104
Assign behavior of diagnostic no. 943 (Parameter) . . . 43 Corrected volume flow offset (Parameter) . . . . . . . . 104
Assign behavior of diagnostic no. 944 (Parameter) . . . 44 Corrected volume flow unit (Parameter) . . . . . . . . . . . 69
Assign behavior of diagnostic no. 948 (Parameter) . . . 44 Corrected volume unit (Parameter) . . . . . . . . . . . . . . 70
Assign channel 1 (Parameter) . . . . . . . . . . . . . . . . . 217 Current input 1 to n (Submenu) . . . . . . . . . . . . . 59, 110
Assign channel 2 (Parameter) . . . . . . . . . . . . . . . . . 219 Current input 1 to n simulation (Parameter) . . . . . . . 240
Assign channel 3 (Parameter) . . . . . . . . . . . . . . . . . 219 Current output 1 to n (Submenu) . . . . . . . . . . . . . . . 116
Assign channel 4 (Parameter) . . . . . . . . . . . . . . . . . 219 Current output 1 to n simulation (Parameter) . . . . . . 241
Assign current output 1 to n (Parameter) . . . . . . . . . 117 Current span (Parameter) . . . . . . . . . . . . . . . . 111, 118
Assign diagnostic behavior (Parameter) . . . . . . 146, 155 Custody transfer (Submenu) . . . . . . . . . . . . . . . . . . 187
Assign flow direction check (Parameter) . . . . . . 150, 154 Custody transfer logbook (Submenu) . . . . . . . . . . . . 204

Endress+Hauser 313
Index Proline Promass 500 Modbus RS485

Cut off inhomogeneous liquid (Parameter) . . . . . . . . 194 Actual diagnostics (0691) . . . . . . . . . . . . . . . . . . 196
Cut off inhomogeneous wet gas (Parameter) . . . . . . 193 Alarm delay (0651) . . . . . . . . . . . . . . . . . . . . . . . 33
Cut off suspended bubbles (Parameter) . . . . . . . . . . 194 Application specific input 0 (6366) . . . . . . . . . . . 191
Application specific input 1 (6367) . . . . . . . . . . . 192
D Application specific output 0 (6364) . . . . . . . . . . 192
Damping output 1 to n (Parameter) . . . . . . . . . 127, 142 Application specific output 1 (6365) . . . . . . . . . . 193
Data logging (Parameter) . . . . . . . . . . . . . . . . . . . . 221 Apply I/O configuration (3907) . . . . . . . . . . . . . 109
Data logging (Submenu) . . . . . . . . . . . . . . . . . . . . . 216 Assign behavior of diagnostic no. 046 (0709) . . . . 35
Data logging control (Parameter) . . . . . . . . . . . . . . . 222 Assign behavior of diagnostic no. 140 (0708) . . . . 35
Data logging status (Parameter) . . . . . . . . . . . . . . . 222 Assign behavior of diagnostic no. 144 (0731) . . . . 36
Data transfer mode (Parameter) . . . . . . . . . . . . . . . 166 Assign behavior of diagnostic no. 302 (0739) . . . . 37
Date/time format (Parameter) . . . . . . . . . . . . . . . . . . 73 Assign behavior of diagnostic no. 374 (0710) . . . . 36
Decimal places 1 (Parameter) . . . . . . . . . . . . . . . . . . 22 Assign behavior of diagnostic no. 441 (0657) . . . . 37
Decimal places 2 (Parameter) . . . . . . . . . . . . . . . . . . 23 Assign behavior of diagnostic no. 442 (0658) . . . . 37
Decimal places 3 (Parameter) . . . . . . . . . . . . . . . . . . 25 Assign behavior of diagnostic no. 443 (0659) . . . . 38
Decimal places 4 (Parameter) . . . . . . . . . . . . . . . . . . 26 Assign behavior of diagnostic no. 444 (0740) . . . . 38
Default gateway (Parameter) . . . . . . . . . . . . . . . . . . 173 Assign behavior of diagnostic no. 543 (0643) . . . . 38
Define access code (Parameter) . . . . . . . . . . . . . . . . . 45 Assign behavior of diagnostic no. 830 (0800) . . . . 39
Define access code (Wizard) . . . . . . . . . . . . . . . . . . . 45 Assign behavior of diagnostic no. 831 (0641) . . . . 39
Density (Parameter) . . . . . . . . . . . . . . . . . . . . . . . . . 52 Assign behavior of diagnostic no. 832 (0681) . . . . 39
Density damping (Parameter) . . . . . . . . . . . . . . . . . . 83 Assign behavior of diagnostic no. 833 (0682) . . . . 40
Density factor (Parameter) . . . . . . . . . . . . . . . . . . . 104 Assign behavior of diagnostic no. 834 (0700) . . . . 40
Density offset (Parameter) . . . . . . . . . . . . . . . . . . . . 103 Assign behavior of diagnostic no. 835 (0702) . . . . 40
Density unit (Parameter) . . . . . . . . . . . . . . . . . . . . . . 70 Assign behavior of diagnostic no. 842 (0638) . . . . 41
Device alarm simulation (Parameter) . . . . . . . . . . . . 246 Assign behavior of diagnostic no. 862 (0679) . . . . 41
Device ID (Parameter) . . . . . . . . . . . . . . . . . . . . . . . 170 Assign behavior of diagnostic no. 912 (0703) . . . . 42
Device information (Submenu) . . . . . . . . . . . . . . . . 205 Assign behavior of diagnostic no. 913 (0712) . . . . 42
Device name (Parameter) . . . . . . . . . . . . . . . . . . . . 206 Assign behavior of diagnostic no. 941 (0635) . . . . 42
Device reset (Parameter) . . . . . . . . . . . . . . . . . . . . . . 47 Assign behavior of diagnostic no. 942 (0636) . . . . 43
Device revision (Parameter) . . . . . . . . . . . . . . . . . . . 170 Assign behavior of diagnostic no. 943 (0637) . . . . 43
Device tag (Parameter) . . . . . . . . . . . . . . . . . . . . . . 205 Assign behavior of diagnostic no. 944 (0732) . . . . 44
DHCP client (Parameter) . . . . . . . . . . . . . . . . . . . . . 172 Assign behavior of diagnostic no. 948 (0744) . . . . 44
Diagnostic behavior (Submenu) . . . . . . . . . . . . . . . . . 33 Assign channel 1 (0851) . . . . . . . . . . . . . . . . . . 217
Diagnostic event category (Parameter) . . . . . . . . . . . 247 Assign channel 2 (0852) . . . . . . . . . . . . . . . . . . 219
Diagnostic event simulation (Parameter) . . . . . . . . . 247 Assign channel 3 (0853) . . . . . . . . . . . . . . . . . . 219
Diagnostic handling (Submenu) . . . . . . . . . . . . . . . . . 33 Assign channel 4 (0854) . . . . . . . . . . . . . . . . . . 219
Diagnostic list (Submenu) . . . . . . . . . . . . . . . . . . . . 198 Assign current output 1 to n (0359–1 to n) . . . . . 117
Diagnostics (Submenu) . . . . . . . . . . . . . . . . . . . . . . 195 Assign diagnostic behavior
Diagnostics 1 (Parameter) . . . . . . . . . . . . . . . . . . . . 198 Pulse/frequency/switch output 1 to n (0482–
Diagnostics 2 (Parameter) . . . . . . . . . . . . . . . . . . . . 199 1 to n) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 146
Diagnostics 3 (Parameter) . . . . . . . . . . . . . . . . . . . . 200 Relay output 1 to n (0806–1 to n) . . . . . . . . . 155
Diagnostics 4 (Parameter) . . . . . . . . . . . . . . . . . . . . 200 Assign flow direction check
Diagnostics 5 (Parameter) . . . . . . . . . . . . . . . . . . . . 201 Pulse/frequency/switch output 1 to n (0484–
Direct access 1 to n) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 150
0/4 mA value Relay output 1 to n (0808–1 to n) . . . . . . . . . 154
Current input 1 to n (1606–1 to n) . . . . . . . . 112 Assign frequency output
Current output 1 to n (0367–1 to n) . . . . . . . 120 Pulse/frequency/switch output 1 to n (0478–
0% bargraph value 1 (0123) . . . . . . . . . . . . . . . . . 21 1 to n) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 138
0% bargraph value 3 (0124) . . . . . . . . . . . . . . . . . 24 Assign limit
2.4 GHz WLAN channel (2704) . . . . . . . . . . . . . 179 Pulse/frequency/switch output 1 to n (0483–
20 mA value 1 to n) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 146
Current input 1 to n (1607–1 to n) . . . . . . . . 112 Relay output 1 to n (0807–1 to n) . . . . . . . . . 154
Current output 1 to n (0372–1 to n) . . . . . . . 121 Assign process variable
100% bargraph value 1 (0125) . . . . . . . . . . . . . . . 21 Totalizer 1 to n (0914–1 to n) . . . . . . . . . . . . 182
100% bargraph value 3 (0126) . . . . . . . . . . . . . . . 24 Assign process variable (1837) . . . . . . . . . . . . . . . 85
Access status (0005) . . . . . . . . . . . . . . . . . . . . . . 14 Assign process variable (1860) . . . . . . . . . . . . . . . 88
Activate SW option (0029) . . . . . . . . . . . . . . . . . . 48 Assign pulse output 1 (0982–1) . . . . . . . . . . . . . 161
Active level Assign pulse output 1 to n (0460–1 to n) . . . . . . 134
Status input 1 to n (1351–1 to n) . . . . . . . . . . 115 Assign simulation process variable (1810) . . . . . 238

314 Endress+Hauser
Proline Promass 500 Modbus RS485 Index

Assign SSID name (2708) . . . . . . . . . . . . . . . . . . 179 Density factor (1849) . . . . . . . . . . . . . . . . . . . . . 104


Assign status Density offset (1848) . . . . . . . . . . . . . . . . . . . . . 103
Pulse/frequency/switch output 1 to n (0485– Density unit (0555) . . . . . . . . . . . . . . . . . . . . . . . 70
1 to n) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 150 Device alarm simulation (0654) . . . . . . . . . . . . . 246
Relay output 1 to n (0805–1 to n) . . . . . . . . . 156 Device ID (7153) . . . . . . . . . . . . . . . . . . . . . . . . 170
Assign status input Device name (0020) . . . . . . . . . . . . . . . . . . . . . . 206
Status input 1 to n (1352–1 to n) . . . . . . . . . . 114 Device reset (0000) . . . . . . . . . . . . . . . . . . . . . . . 47
Backlight (0111) . . . . . . . . . . . . . . . . . . . . . . . . . 29 Device revision (7154) . . . . . . . . . . . . . . . . . . . . 170
Backup state (2759) . . . . . . . . . . . . . . . . . . . . . . . 31 Device tag (0011) . . . . . . . . . . . . . . . . . . . . . . . . 205
Baudrate (7111) . . . . . . . . . . . . . . . . . . . . . . . . 166 DHCP client (7212) . . . . . . . . . . . . . . . . . . . . . . 172
Bootloader revision Diagnostic event category (0738) . . . . . . . . . . . . 247
I/O module (0073) . . . . . . . . 211, 212, 214, 215 Diagnostic event simulation (0737) . . . . . . . . . . 247
Bootloader revision (0073) . . . . . . . . . 209, 210, 216 Diagnostics 1 (0692) . . . . . . . . . . . . . . . . . . . . . 198
Build no. software Diagnostics 2 (0693) . . . . . . . . . . . . . . . . . . . . . 199
I/O module (0079) . . . . . . . . 211, 212, 214, 215 Diagnostics 3 (0694) . . . . . . . . . . . . . . . . . . . . . 200
Build no. software (0079) . . . . . . . . . . 209, 210, 216 Diagnostics 4 (0695) . . . . . . . . . . . . . . . . . . . . . 200
Bus address (7112) . . . . . . . . . . . . . . . . . . . . . . 165 Diagnostics 5 (0696) . . . . . . . . . . . . . . . . . . . . . 201
Bus termination (7155) . . . . . . . . . . . . . . . . . . . 169 Direct access (0106) . . . . . . . . . . . . . . . . . . . . . . . 12
Byte order (7113) . . . . . . . . . . . . . . . . . . . . . . . 167 Display damping (0094) . . . . . . . . . . . . . . . . . . . . 27
C0 to 5 (6022) . . . . . . . . . . . . . . . . . . . . . . . . . . 107 Display interval (0096) . . . . . . . . . . . . . . . . . . . . . 26
Calibration factor (6025) . . . . . . . . . . . . . . . . . . 106 Display language (0104) . . . . . . . . . . . . . . . . . . . 16
Carrier corrected volume flow (1894) . . . . . . . . . . 56 Dynamic viscosity (1854) . . . . . . . . . . . . . . . . . . . 53
Carrier mass flow (1865) . . . . . . . . . . . . . . . . . . . 56 ENP version (0012) . . . . . . . . . . . . . . . . . . . . . . 208
Carrier volume flow (1896) . . . . . . . . . . . . . . . . . 57 Enter access code (0003) . . . . . . . . . . . . . . . . . . . 14
Clear logging data (0855) . . . . . . . . . . . . . . . . . . 220 Entire logging duration (0861) . . . . . . . . . . . . . . 223
Comparison result (2760) . . . . . . . . . . . . . . . . . . . 32 Extended order code 1 (0023) . . . . . . . . . . . . . . 207
Concentration (1887) . . . . . . . . . . . . . . . . . . . . . . 55 Extended order code 2 (0021) . . . . . . . . . . . . . . 207
Configuration management (2758) . . . . . . . . . . . 31 Extended order code 3 (0022) . . . . . . . . . . . . . . 207
Connection state (2722) . . . . . . . . . . . . . . . . . . . 180 External pressure (6209) . . . . . . . . . . . . . . . . . . . 93
Contrast display (0105) . . . . . . . . . . . . . . . . . . . . 29 External reference density (6198) . . . . . . . . . . . . . 96
Control Totalizer 1 to n (0912–1 to n) . . . . . . . . 185 External temperature (6080) . . . . . . . . . . . . . . . . 94
Corrected volume flow (1851) . . . . . . . . . . . . . . . 52 Fail-safe type application specific 0 (2098) . . . . . 191
Corrected volume flow calculation (1812) . . . . . . . 96 Fail-safe type application specific 1 (2100) . . . . . 192
Corrected volume flow factor (1867) . . . . . . . . . . 104 Fail-safe value application specific 0 (2099) . . . . 191
Corrected volume flow offset (1866) . . . . . . . . . . 104 Fail-safe value application specific 1 (65535) . . . 192
Corrected volume flow unit (0558) . . . . . . . . . . . . 69 Failure current
Corrected volume unit (0575) . . . . . . . . . . . . . . . . 70 Current output 1 to n (0352–1 to n) . . . . . . . 129
Current input 1 to n simulation (1608–1 to n) . . . 240 Failure frequency
Current output 1 to n simulation (0354–1 to n) . . 241 Pulse/frequency/switch output 1 to n (0474–
Current span 1 to n) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 144
Current input 1 to n (1605–1 to n) . . . . . . . . 111 Failure mode
Current output 1 to n (0353–1 to n) . . . . . . . 118 Current input 1 to n (1601–1 to n) . . . . . . . . 112
Cut off inhomogeneous liquid (6374) . . . . . . . . . 194 Current output 1 to n (0364–1 to n) . . . . . . . 128
Cut off inhomogeneous wet gas (6375) . . . . . . . . 193 Pulse/frequency/switch output 1 to n (0451–
Cut off suspended bubbles (6370) . . . . . . . . . . . . 194 1 to n) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 144
Damping output 1 to n (0363–1 to n) . . . . . . . . . 127 Pulse/frequency/switch output 1 to n (0480–
Damping output 1 to n (0477–1 to n) . . . . . . . . . 142 1 to n) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 136
Data logging (0860) . . . . . . . . . . . . . . . . . . . . . . 221 Pulse/frequency/switch output 1 to n (0486–
Data logging control (0857) . . . . . . . . . . . . . . . . 222 1 to n) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 151
Data logging status (0858) . . . . . . . . . . . . . . . . . 222 Relay output 1 to n (0811–1 to n) . . . . . . . . . 158
Data transfer mode (7115) . . . . . . . . . . . . . . . . . 166 Totalizer 1 to n (0901–1 to n) . . . . . . . . . . . . 186
Date/time format (2812) . . . . . . . . . . . . . . . . . . . 73 Failure mode (0985) . . . . . . . . . . . . . . . . . . . . . 163
Decimal places 1 (0095) . . . . . . . . . . . . . . . . . . . . 22 Failure mode (7116) . . . . . . . . . . . . . . . . . . . . . 168
Decimal places 2 (0117) . . . . . . . . . . . . . . . . . . . . 23 Failure value
Decimal places 3 (0118) . . . . . . . . . . . . . . . . . . . . 25 Current input 1 to n (1602–1 to n) . . . . . . . . 113
Decimal places 4 (0119) . . . . . . . . . . . . . . . . . . . . 26 Fieldbus writing access (7156) . . . . . . . . . . . . . . 169
Default gateway (7210) . . . . . . . . . . . . . . . . . . . 173 Filter options . . . . . . . . . . . . . . . . . . . . . . . . . . . 203
Density (1850) . . . . . . . . . . . . . . . . . . . . . . . . . . . 52 Filter options (0705) . . . . . . . . . . . . . . . . . . . . . 202
Density damping (1803) . . . . . . . . . . . . . . . . . . . . 83 Firmware version (0010) . . . . . . . . . . . . . . . . . . 206

Endress+Hauser 315
Index Proline Promass 500 Modbus RS485

Fixed current Maximum value (0665) . . . . . . . . . . . . . . . . . . . 227


Current output 1 to n (0365–1 to n) . . . . . . . 119 Maximum value (6009) . . . . . . . . . . . . . . . . . . . 233
Fixed reference density (1814) . . . . . . . . . . . . . . . 97 Maximum value (6014) . . . . . . . . . . . . . . . . . . . 236
Flow damping (1802) . . . . . . . . . . . . . . . . . . . . . . 82 Maximum value (6029) . . . . . . . . . . . . . . . . . . . 230
Flow override (1839) . . . . . . . . . . . . . . . . . . . . . . 84 Maximum value (6051) . . . . . . . . . . . . . . . . . . . 228
Format display (0098) . . . . . . . . . . . . . . . . . . . . . 17 Maximum value (6068) . . . . . . . . . . . . . . . . . . . 232
Frequency output simulation 1 to n (0472–1 to n) Maximum value (6070) . . . . . . . . . . . . . . . . . . . 231
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 242 Maximum value (6108) . . . . . . . . . . . . . . . . . . . 229
Frequency value 1 to n (0473–1 to n) . . . . . . . . . 242 Maximum value (6119) . . . . . . . . . . . . . . . . . . . 235
Gas Fraction Handler (6377) . . . . . . . . . . . . . . . . 92 Maximum value (6121) . . . . . . . . . . . . . . . . . . . 234
Gateway IP address (2719) . . . . . . . . . . . . . . . . . 180 Maximum value (6283) . . . . . . . . . . . . . . . . . . . 236
Header (0097) . . . . . . . . . . . . . . . . . . . . . . . . . . . 28 Measured current 1 to n (0366–1 to n) . . . . . 61, 130
Header text (0112) . . . . . . . . . . . . . . . . . . . . . . . 28 Measured current 1 to n (1604–1 to n) . . . . . . . . . 60
High value partial filled pipe detection (1858) . . . . 89 Measured values 1 to n (1603–1 to n) . . . . . . . . . 60
I/O alteration code (2762) . . . . . . . . . . . . . . . . . 109 Measuring mode
I/O module 1 terminal numbers (3902–1) . . . . . 210 Current output 1 to n (0351–1 to n) . . . . . . . 122
I/O module 1 to n information (3906–1 to n) . . . 108 Pulse/frequency/switch output 1 to n (0457–
I/O module 1 to n terminal numbers (3902– 1 to n) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 136
1 to n) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 108 Pulse/frequency/switch output 1 to n (0479–
I/O module 1 to n type (3901–1 to n) . . . . . . . . . 109 1 to n) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 140
I/O module 2 terminal numbers (3902–2) 212, Measuring mode (0984) . . . . . . . . . . . . . . . . . . . 162
213, 214 Measuring value at maximum frequency
I/O module 3 terminal numbers (3902–3) 212, Pulse/frequency/switch output 1 to n (0475–
213, 214 1 to n) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 140
I/O module 4 terminal numbers (3902–4) 212, Measuring value at minimum frequency
213, 214 Pulse/frequency/switch output 1 to n (0476–
Index inhomogeneous medium (6368) . . . . . . . . 194 1 to n) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 140
Index suspended bubbles (6376) . . . . . . . . . . . . . 194 Minimum frequency value
Input signal level 1 to n (1356–1 to n) . . . . . . . . 240 Pulse/frequency/switch output 1 to n (0453–
Installation angle pitch (6236) . . . . . . . . . . . . . . . 99 1 to n) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 139
Installation angle roll (6282) . . . . . . . . . . . . . . . . 99 Minimum value (0688) . . . . . . . . . . . . . . . . . . . 227
Installation direction (1809) . . . . . . . . . . . . . . . . . 99 Minimum value (6008) . . . . . . . . . . . . . . . . . . . 233
Invert output signal Minimum value (6010) . . . . . . . . . . . . . . . . . . . 232
Pulse/frequency/switch output 1 to n (0470– Minimum value (6015) . . . . . . . . . . . . . . . . . . . 235
1 to n) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 152 Minimum value (6030) . . . . . . . . . . . . . . . . . . . 230
Invert output signal (0993) . . . . . . . . . . . . . . . . . 164 Minimum value (6052) . . . . . . . . . . . . . . . . . . . 228
IP address (7209) . . . . . . . . . . . . . . . . . . . . . . . . 173 Minimum value (6069) . . . . . . . . . . . . . . . . . . . 232
IP address domain name server (2720) . . . . . . . . 181 Minimum value (6071) . . . . . . . . . . . . . . . . . . . 231
Kinematic viscosity (1857) . . . . . . . . . . . . . . . . . . 54 Minimum value (6109) . . . . . . . . . . . . . . . . . . . 229
Last backup (2757) . . . . . . . . . . . . . . . . . . . . . . . 30 Minimum value (6120) . . . . . . . . . . . . . . . . . . . 235
Linear expansion coefficient (1817) . . . . . . . . . . . 98 Minimum value (6122) . . . . . . . . . . . . . . . . . . . 234
Locking status (0004) . . . . . . . . . . . . . . . . . . . . . 13 Minimum value (6284) . . . . . . . . . . . . . . . . . . . 236
Logging delay (0859) . . . . . . . . . . . . . . . . . . . . . 221 Network security (2705) . . . . . . . . . . . . . . . . . . 176
Logging interval (0856) . . . . . . . . . . . . . . . . . . . 220 Nominal diameter (2807) . . . . . . . . . . . . . . . . . . 107
Login page (7273) . . . . . . . . . . . . . . . . . . . . . . . 174 Off value low flow cutoff (1804) . . . . . . . . . . . . . . 85
Low value partial filled pipe detection (1861) . . . . 88 On value low flow cutoff (1805) . . . . . . . . . . . . . . 85
MAC address (7214) . . . . . . . . . . . . . . . . . . . . . 172 Operating mode
Mass flow (1838) . . . . . . . . . . . . . . . . . . . . . . . . . 51 Pulse/frequency/switch output 1 to n (0469–
Mass flow factor (1832) . . . . . . . . . . . . . . . . . . . 102 1 to n) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 132
Mass flow offset (1831) . . . . . . . . . . . . . . . . . . . 102 Operating time (0652) . . . . . . . . . . . . . . 30, 46, 197
Mass flow unit (0554) . . . . . . . . . . . . . . . . . . . . . 65 Operating time from restart (0653) . . . . . . . . . . 197
Mass unit (0574) . . . . . . . . . . . . . . . . . . . . . . . . . 66 Order code (0008) . . . . . . . . . . . . . . . . . . . . . . . 206
Master terminal number (0981) . . . . . . . . . . . . . 160 Output current 1 to n (0361–1 to n) . . . . . . . 61, 130
Max. switch cycles number Output frequency 1 to n (0471–1 to n) . . . . . 62, 145
Relay output 1 to n (0817–1 to n) . . . . . . . . . . 64 Parameter 0 (6358) . . . . . . . . . . . . . . . . . . . . . . 188
Maximum damping partial filled pipe det. (6040) . 89 Parameter 1 (6359) . . . . . . . . . . . . . . . . . . . . . . 188
Maximum frequency value Parameter 2 (6360) . . . . . . . . . . . . . . . . . . . . . . 189
Pulse/frequency/switch output 1 to n (0454– Parameter 3 (6361) . . . . . . . . . . . . . . . . . . . . . . 189
1 to n) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 139 Parameter 4 (6345) . . . . . . . . . . . . . . . . . . . . . . 189

316 Endress+Hauser
Proline Promass 500 Modbus RS485 Index

Parameter 5 (6346) . . . . . . . . . . . . . . . . . . . . . . 189 Signal mode


Parameter 6 (6347) . . . . . . . . . . . . . . . . . . . . . . 190 Current input 1 to n (1610–1 to n) . . . . . . . . 111
Parameter 7 (6348) . . . . . . . . . . . . . . . . . . . . . . 190 Current output 1 to n (0377–1 to n) . . . . . . . 117
Parameter 8 (6349) . . . . . . . . . . . . . . . . . . . . . . 190 Pulse/frequency/switch output 1 to n (0490–
Parameter 9 (6350) . . . . . . . . . . . . . . . . . . . . . . 190 1 to n) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 132
Parity (7122) . . . . . . . . . . . . . . . . . . . . . . . . . . . 166 Signal mode (0991) . . . . . . . . . . . . . . . . . . . . . . 160
Phase shift (0992) . . . . . . . . . . . . . . . . . . . . . . . 162 Slave terminal number (0990) . . . . . . . . . . . . . . 160
Powerless relay status Software option overview (0015) . . . . . . . . . . . . . 49
Relay output 1 to n (0816–1 to n) . . . . . . . . . 159 Software revision
Preset value 1 to n (0913–1 to n) . . . . . . . . . . . . 185 I/O module (0072) . . . . . . . . 211, 212, 213, 215
Pressure compensation (6130) . . . . . . . . . . . . . . . 93 Software revision (0072) . . . . . . . . . . 208, 209, 216
Pressure shock suppression (1806) . . . . . . . . . . . . 86 Square expansion coefficient (1818) . . . . . . . . . . . 98
Pressure unit (0564) . . . . . . . . . . . . . . . . . . . . . . 73 SSID name (2707) . . . . . . . . . . . . . . . . . . . . . . . 179
Pressure value (6059) . . . . . . . . . . . . . . . . . . . . . 93 SSID name (2714) . . . . . . . . . . . . . . . . . . . . . . . 176
Pressure value (6129) . . . . . . . . . . . . . . . . . . . . . 53 Status input simulation 1 to n (1355–1 to n) . . . . 239
Previous diagnostics (0690) . . . . . . . . . . . . . . . . 196 Subnet mask (7211) . . . . . . . . . . . . . . . . . . . . . . 173
Process variable value (1811) . . . . . . . . . . . . . . . 239 Switch cycles
Progress (2808) . . . . . . . . . . . . . . . . . . . . . . . . . 101 Relay output 1 to n (0815–1 to n) . . . . . . . . . . 64
Pulse output (0987) . . . . . . . . . . . . . . . . . . . 65, 164 Switch output function
Pulse output 1 to n (0456–1 to n) . . . . . . . . . 62, 137 Pulse/frequency/switch output 1 to n (0481–
Pulse output simulation (0988) . . . . . . . . . . . . . 245 1 to n) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 145
Pulse output simulation 1 to n (0458–1 to n) . . . 243 Switch output simulation 1 to n (0462–1 to n) . . 243
Pulse scaling Switch status
Pulse/frequency/switch output 1 to n (0455– Relay output 1 to n (0801–1 to n) . . . . . . 64, 158
1 to n) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 134 Switch status 1 to n (0461–1 to n) . . . . . . . . 63, 151
Pulse value (0989) . . . . . . . . . . . . . . . . . . . . . . . 246 Switch status 1 to n (0463–1 to n) . . . . . . . . . . . 244
Pulse value 1 to n (0459–1 to n) . . . . . . . . . . . . . 243 Switch status 1 to n (0803–1 to n) . . . . . . . . . . . 245
Pulse width Switch-off delay
Pulse/frequency/switch output 1 to n (0452– Pulse/frequency/switch output 1 to n (0465–
1 to n) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 135 1 to n) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 151
Pulse width (0986) . . . . . . . . . . . . . . . . . . . . . . 162 Relay output 1 to n (0813–1 to n) . . . . . . . . . 157
Received signal strength (2721) . . . . . . . . . . . . . 180 Switch-off value
Reference density (1852) . . . . . . . . . . . . . . . . . . . 52 Pulse/frequency/switch output 1 to n (0464–
Reference density factor (1869) . . . . . . . . . . . . . 105 1 to n) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 149
Reference density offset (1868) . . . . . . . . . . . . . 105 Relay output 1 to n (0809–1 to n) . . . . . . . . . 156
Reference density unit (0556) . . . . . . . . . . . . . . . 72 Switch-on delay
Reference sound velocity (6147) . . . . . . . . . . . . . . 91 Pulse/frequency/switch output 1 to n (0467–
Reference temperature (1816) . . . . . . . . . . . . . . . 97 1 to n) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 150
Relay output 1 to n simulation (0802–1 to n) . . . 244 Relay output 1 to n (0814–1 to n) . . . . . . . . . 158
Relay output function Switch-on value
Relay output 1 to n (0804–1 to n) . . . . . . . . . 153 Pulse/frequency/switch output 1 to n (0466–
Reset access code (0024) . . . . . . . . . . . . . . . . . . . 46 1 to n) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 149
Reset all totalizers (2806) . . . . . . . . . . . . . . . . . 181 Relay output 1 to n (0810–1 to n) . . . . . . . . . 157
Reset min/max values (6151) . . . . . . . . . . . . . . . 226 Target corrected volume flow (1893) . . . . . . . . . . 56
Response time Target mass flow (1864) . . . . . . . . . . . . . . . . . . . 55
Current output 2 (0378) . . . . . . . . . . . . . . . . 127 Target volume flow (1895) . . . . . . . . . . . . . . . . . . 57
Pulse/frequency/switch output 1 to n (0491– Telegram delay (7146) . . . . . . . . . . . . . . . . . . . . 168
1 to n) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 143 Temp. compensated dynamic viscosity (1872) . . . . 54
Response time part. filled pipe detect. (1859) . . . . 89 Temp. compensated kinematic viscosity (1863) . . . 54
Response time status input Temperature (1853) . . . . . . . . . . . . . . . . . . . . . . 53
Status input 1 to n (1354–1 to n) . . . . . . . . . . 115 Temperature coefficient sound velocity (6181) . . . 92
Scan list register 0 to 15 (7114) . . . . . . . . . . . . . 171 Temperature correction source (6184) . . . . . . . . . 94
Security identification (2718) . . . . . . . . . . . . . . . 177 Temperature damping (1822) . . . . . . . . . . . . . . . 83
Select antenna (2713) . . . . . . . . . . . . . . . . . . . . 180 Temperature factor (1871) . . . . . . . . . . . . . . . . . 106
Select gas type (6074) . . . . . . . . . . . . . . . . . . . . . 91 Temperature mode (6341) . . . . . . . . . . . . . . . . . . 95
Select medium (6062) . . . . . . . . . . . . . . . . . . . . . 90 Temperature offset (1870) . . . . . . . . . . . . . . . . . 105
Separator (0101) . . . . . . . . . . . . . . . . . . . . . . . . . 29 Temperature unit (0557) . . . . . . . . . . . . . . . . . . . 72
Serial number (0009) . . . . . . . . . . . . . . . . . . . . . 206 Terminal number
Current input 1 to n (1611–1 to n) . . . . . . . . 110

Endress+Hauser 317
Index Proline Promass 500 Modbus RS485

Current output 1 to n (0379–1 to n) . . . . . . . 116 Zero point adjustment status (6253) . . . . . . . . . . 101
Pulse/frequency/switch output 1 to n (0492– Direct access (Parameter) . . . . . . . . . . . . . . . . . . . . . 12
1 to n) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 131 Display (Submenu) . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Relay output 1 to n (0812–1 to n) . . . . . . . . . 153 Display channel 1 (Submenu) . . . . . . . . . . . . . . . . . 223
Status input 1 to n (1358–1 to n) . . . . . . . . . . 113 Display channel 2 (Submenu) . . . . . . . . . . . . . . . . . 224
Timestamp . . . . . 196, 197, 198, 199, 200, 201, 202 Display channel 3 (Submenu) . . . . . . . . . . . . . . . . . 225
Totalizer operation mode Display channel 4 (Submenu) . . . . . . . . . . . . . . . . . 225
Totalizer 1 to n (0908–1 to n) . . . . . . . . . . . . 184 Display damping (Parameter) . . . . . . . . . . . . . . . . . . 27
Totalizer overflow 1 to n (0910–1 to n) . . . . . . . . 58 Display interval (Parameter) . . . . . . . . . . . . . . . . . . . 26
Totalizer value 1 to n (0911–1 to n) . . . . . . . . . . . 58 Display language (Parameter) . . . . . . . . . . . . . . . . . . 16
Unit totalizer 1 to n (0915–1 to n) . . . . . . . . . . . 183 Display module (Submenu) . . . . . . . . . . . . . . . . . . . 215
User corrected volume factor (0590) . . . . . . . . . . . 78 Document
User corrected volume offset (0602) . . . . . . . . . . . 78 Explanation of the structure of a parameter
User corrected volume text (0592) . . . . . . . . . . . . 77 description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
User density factor (0572) . . . . . . . . . . . . . . . . . . 79 Function . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
User density offset (0571) . . . . . . . . . . . . . . . . . . 79 Structure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
User density text (0570) . . . . . . . . . . . . . . . . . . . . 78 Symbols used . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
User energy factor (0586) . . . . . . . . . . . . . . . . . . 81 Target group . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
User energy offset (0599) . . . . . . . . . . . . . . . . . . . 81 Using the document . . . . . . . . . . . . . . . . . . . . . . . . 5
User energy text (0600) . . . . . . . . . . . . . . . . . . . . 80 Document function . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
User mass factor (0561) . . . . . . . . . . . . . . . . . . . . 77 Double pulse output (Submenu) . . . . . . . . . . . . . 64, 159
User mass offset (0562) . . . . . . . . . . . . . . . . . . . . 76 Dynamic viscosity (Parameter) . . . . . . . . . . . . . . . . . . 53
User mass text (0560) . . . . . . . . . . . . . . . . . . . . . 76
User name (2715) . . . . . . . . . . . . . . . . . . . . . . . 177 E
User pressure factor (0579) . . . . . . . . . . . . . . . . . 82 ENP version (Parameter) . . . . . . . . . . . . . . . . . . . . . 208
User pressure offset (0580) . . . . . . . . . . . . . . . . . 81 Enter access code (Parameter) . . . . . . . . . . . . . . . . . . 14
User pressure text (0581) . . . . . . . . . . . . . . . . . . . 81 Entire logging duration (Parameter) . . . . . . . . . . . . 223
User specific-enthalpy factor (0583) . . . . . . . . . . . 80 Event list (Submenu) . . . . . . . . . . . . . . . . . . . . . . . . 203
User specific-enthalpy offset (0584) . . . . . . . . . . . 80 Event logbook (Submenu) . . . . . . . . . . . . . . . . . . . . 202
User specific-enthalpy text (0585) . . . . . . . . . . . . 79 Extended order code 1 (Parameter) . . . . . . . . . . . . . 207
User volume factor (0568) . . . . . . . . . . . . . . . . . . 76 Extended order code 2 (Parameter) . . . . . . . . . . . . . 207
User volume offset (0569) . . . . . . . . . . . . . . . . . . 75 Extended order code 3 (Parameter) . . . . . . . . . . . . . 207
User volume text (0567) . . . . . . . . . . . . . . . . . . . . 75 External compensation (Submenu) . . . . . . . . . . . . . . . 92
Value 1 display (0107) . . . . . . . . . . . . . . . . . . . . . 19 External pressure (Parameter) . . . . . . . . . . . . . . . . . . 93
Value 2 display (0108) . . . . . . . . . . . . . . . . . . . . . 22 External reference density (Parameter) . . . . . . . . . . . 96
Value 3 display (0110) . . . . . . . . . . . . . . . . . . . . . 23 External temperature (Parameter) . . . . . . . . . . . . . . . 94
Value 4 display (0109) . . . . . . . . . . . . . . . . . . . . . 25
Value current input 1 to n (1609–1 to n) . . . . . . 241 F
Value current output 1 to n (0355–1 to n) . . . . . 241 Factory settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . 248
Value per pulse (0983) . . . . . . . . . . . . . . . . . . . . 161 SI units . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 248
Value status input US units . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 250
Status input 1 to n (1353–1 to n) . . . . . . . . . . 115 Fail-safe type application specific 0 (Parameter) . . . . 191
Value status input 1 to n (1353–1 to n) . . . . . . 60 Fail-safe type application specific 1 (Parameter) . . . . 192
Volume flow (1847) . . . . . . . . . . . . . . . . . . . . . . . 52 Fail-safe value application specific 0 (Parameter) . . . 191
Volume flow factor (1846) . . . . . . . . . . . . . . . . . 103 Fail-safe value application specific 1 (Parameter) . . . 192
Volume flow offset (1841) . . . . . . . . . . . . . . . . . 103 Failure current (Parameter) . . . . . . . . . . . . . . . . . . . 129
Volume flow unit (0553) . . . . . . . . . . . . . . . . . . . 67 Failure frequency (Parameter) . . . . . . . . . . . . . . . . . 144
Volume unit (0563) . . . . . . . . . . . . . . . . . . . . . . . 68 Failure mode (Parameter) 112, 128, 136, 144, 151,
Web server functionality (7222) . . . . . . . . . . . . . 174 158, 163, 168, 186
Web server language (7221) . . . . . . . . . . . . . . . 172 Failure value (Parameter) . . . . . . . . . . . . . . . . . . . . 113
WLAN (2702) . . . . . . . . . . . . . . . . . . . . . . . . . . 175 Fieldbus writing access (Parameter) . . . . . . . . . . . . . 169
WLAN IP address (2711) . . . . . . . . . . . . . . . . . . 177 Filter options (Parameter) . . . . . . . . . . . . . . . . 202, 203
WLAN MAC address (2703) . . . . . . . . . . . . . . . . 178 Firmware version (Parameter) . . . . . . . . . . . . . . . . . 206
WLAN mode (2717) . . . . . . . . . . . . . . . . . . . . . 176 Fixed current (Parameter) . . . . . . . . . . . . . . . . . . . . 119
WLAN passphrase (2706) . . . . . . . . . . . . . . . . . 178 Fixed reference density (Parameter) . . . . . . . . . . . . . . 97
WLAN password (2716) . . . . . . . . . . . . . . . . . . . 177 Flow damping (Parameter) . . . . . . . . . . . . . . . . . . . . 82
WLAN subnet mask (2709) . . . . . . . . . . . . . . . . 178 Flow override (Parameter) . . . . . . . . . . . . . . . . . . . . . 84
Zero point (6195) . . . . . . . . . . . . . . . . . . . . . . . . 107 Format display (Parameter) . . . . . . . . . . . . . . . . . . . . 17
Zero point adjustment control (6196) . . . . . . . . . 100 Frequency output simulation 1 to n (Parameter) . . . . 242

318 Endress+Hauser
Proline Promass 500 Modbus RS485 Index

Frequency value 1 to n (Parameter) . . . . . . . . . . . . . 242 Mass flow (Parameter) . . . . . . . . . . . . . . . . . . . . . . . 51


Function Mass flow factor (Parameter) . . . . . . . . . . . . . . . . . 102
see Parameter Mass flow offset (Parameter) . . . . . . . . . . . . . . . . . . 102
Mass flow unit (Parameter) . . . . . . . . . . . . . . . . . . . . 65
G Mass unit (Parameter) . . . . . . . . . . . . . . . . . . . . . . . . 66
Gas Fraction Handler (Parameter) . . . . . . . . . . . . . . . 92 Master terminal number (Parameter) . . . . . . . . . . . 160
Gateway IP address (Parameter) . . . . . . . . . . . . . . . 180 Max. switch cycles number (Parameter) . . . . . . . . . . . 64
Maximum damping partial filled pipe det.
H (Parameter) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 89
Header (Parameter) . . . . . . . . . . . . . . . . . . . . . . . . . . 28 Maximum frequency value (Parameter) . . . . . . . . . . 139
Header text (Parameter) . . . . . . . . . . . . . . . . . . . . . . 28 Maximum value (Parameter) 227, 228, 229, 230,
Heartbeat (Submenu) . . . . . . . . . . . . . . . . . . . . . . . 237 231, 232, 233, 234, 235, 236
High value partial filled pipe detection (Parameter) . . 89 Measured current 1 to n (Parameter) . . . . . . 60, 61, 130
Measured values (Submenu) . . . . . . . . . . . . . . . . . . . 50
I Measured values 1 to n (Parameter) . . . . . . . . . . . . . . 60
I/O alteration code (Parameter) . . . . . . . . . . . . . . . . 109 Measurement mode (Submenu) . . . . . . . . . . . . . . . . . 90
I/O configuration (Submenu) . . . . . . . . . . . . . . . . . . 107 Measuring mode (Parameter) . . . . . . 122, 136, 140, 162
I/O module 1 (Submenu) . . . . . . . . . . . . . . . . . . . . . 210 Measuring value at maximum frequency (Parameter)
I/O module 1 terminal numbers (Parameter) . . . . . . 210 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 140
I/O module 1 to n information (Parameter) . . . . . . . 108 Measuring value at minimum frequency (Parameter)
I/O module 1 to n terminal numbers (Parameter) . . . 108 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 140
I/O module 1 to n type (Parameter) . . . . . . . . . . . . . 109 Medium index (Submenu) . . . . . . . . . . . . . . . . . . . . 193
I/O module 2 (Submenu) . . . . . . . . . . . . . . . . . . . . . 211 Medium temperature (Submenu) . . . . . . . . . . . . . . . 229
I/O module 2 terminal numbers (Parameter) 212, Min/max values (Submenu) . . . . . . . . . . . . . . . . . . . 225
213, 214 Minimum frequency value (Parameter) . . . . . . . . . . 139
I/O module 3 (Submenu) . . . . . . . . . . . . . . . . . . . . . 213 Minimum value (Parameter) 227, 228, 229, 230,
I/O module 3 terminal numbers (Parameter) 212, 231, 232, 233, 234, 235, 236
213, 214 Modbus configuration (Submenu) . . . . . . . . . . . . . . 165
I/O module 4 (Submenu) . . . . . . . . . . . . . . . . . . . . . 214 Modbus data map (Submenu) . . . . . . . . . . . . . . . . . 171
I/O module 4 terminal numbers (Parameter) 212, Modbus information (Submenu) . . . . . . . . . . . . . . . 170
213, 214
Index inhomogeneous medium (Parameter) . . . . . . . 194 N
Index suspended bubbles (Parameter) . . . . . . . . . . . 194 Network security (Parameter) . . . . . . . . . . . . . . . . . 176
Input (Submenu) . . . . . . . . . . . . . . . . . . . . . . . . . . . 110 Nominal diameter (Parameter) . . . . . . . . . . . . . . . . 107
Input signal level 1 to n (Parameter) . . . . . . . . . . . . 240
Input values (Submenu) . . . . . . . . . . . . . . . . . . . . . . . 59 O
Installation angle pitch (Parameter) . . . . . . . . . . . . . . 99 Off value low flow cutoff (Parameter) . . . . . . . . . . . . . 85
Installation angle roll (Parameter) . . . . . . . . . . . . . . . 99 On value low flow cutoff (Parameter) . . . . . . . . . . . . . 85
Installation direction (Parameter) . . . . . . . . . . . . . . . 99 Operating mode (Parameter) . . . . . . . . . . . . . . . . . . 132
Invert output signal (Parameter) . . . . . . . . . . . 152, 164 Operating time (Parameter) . . . . . . . . . . . . . 30, 46, 197
IP address (Parameter) . . . . . . . . . . . . . . . . . . . . . . 173 Operating time from restart (Parameter) . . . . . . . . . 197
IP address domain name server (Parameter) . . . . . . . 181 Order code (Parameter) . . . . . . . . . . . . . . . . . . . . . . 206
Oscillation amplitude (Submenu) . . . . . . . . . . . . . . . 232
K Oscillation damping (Submenu) . . . . . . . . . . . . . . . . 234
Kinematic viscosity (Parameter) . . . . . . . . . . . . . . . . . 54 Oscillation frequency (Submenu) . . . . . . . . . . . . . . . 231
Output (Submenu) . . . . . . . . . . . . . . . . . . . . . . . . . . 115
L Output current 1 to n (Parameter) . . . . . . . . . . . 61, 130
Last backup (Parameter) . . . . . . . . . . . . . . . . . . . . . . 30 Output frequency 1 to n (Parameter) . . . . . . . . . 62, 145
Linear expansion coefficient (Parameter) . . . . . . . . . . 98 Output values (Submenu) . . . . . . . . . . . . . . . . . . . . . 61
Locking status (Parameter) . . . . . . . . . . . . . . . . . . . . 13
Logging delay (Parameter) . . . . . . . . . . . . . . . . . . . 221 P
Logging interval (Parameter) . . . . . . . . . . . . . . . . . . 220 Parameter
Login page (Parameter) . . . . . . . . . . . . . . . . . . . . . . 174 Structure of a parameter description . . . . . . . . . . . . 7
Low flow cut off (Submenu) . . . . . . . . . . . . . . . . . . . . 84 Parameter 0 (Parameter) . . . . . . . . . . . . . . . . . . . . . 188
Low value partial filled pipe detection (Parameter) . . . 88 Parameter 1 (Parameter) . . . . . . . . . . . . . . . . . . . . . 188
Parameter 2 (Parameter) . . . . . . . . . . . . . . . . . . . . . 189
M Parameter 3 (Parameter) . . . . . . . . . . . . . . . . . . . . . 189
MAC address (Parameter) . . . . . . . . . . . . . . . . . . . . 172 Parameter 4 (Parameter) . . . . . . . . . . . . . . . . . . . . . 189
Main electronic module + I/O module 1 (Submenu) . 208 Parameter 5 (Parameter) . . . . . . . . . . . . . . . . . . . . . 189
Main electronic temperature (Submenu) . . . . . . . . . 227

Endress+Hauser 319
Index Proline Promass 500 Modbus RS485

Parameter 6 (Parameter) . . . . . . . . . . . . . . . . . . . . . 190 Sensor electronic temperature (ISEM) (Submenu) . . . 228


Parameter 7 (Parameter) . . . . . . . . . . . . . . . . . . . . . 190 Separator (Parameter) . . . . . . . . . . . . . . . . . . . . . . . . 29
Parameter 8 (Parameter) . . . . . . . . . . . . . . . . . . . . . 190 Serial number (Parameter) . . . . . . . . . . . . . . . . . . . 206
Parameter 9 (Parameter) . . . . . . . . . . . . . . . . . . . . . 190 Signal asymmetry (Submenu) . . . . . . . . . . . . . . . . . 235
Parity (Parameter) . . . . . . . . . . . . . . . . . . . . . . . . . 166 Signal mode (Parameter) . . . . . . . . . 111, 117, 132, 160
Partially filled pipe detection (Submenu) . . . . . . . . . . 87 Simulation (Submenu) . . . . . . . . . . . . . . . . . . . . . . . 237
Petroleum (Submenu) . . . . . . . . . . . . . . . . . . . . . . . 187 Slave terminal number (Parameter) . . . . . . . . . . . . . 160
Phase shift (Parameter) . . . . . . . . . . . . . . . . . . . . . . 162 Software option overview (Parameter) . . . . . . . . . . . . 49
Powerless relay status (Parameter) . . . . . . . . . . . . . 159 Software revision (Parameter) 208, 209, 211, 212,
Preset value 1 to n (Parameter) . . . . . . . . . . . . . . . . 185 213, 215, 216
Pressure compensation (Parameter) . . . . . . . . . . . . . . 93 Square expansion coefficient (Parameter) . . . . . . . . . 98
Pressure shock suppression (Parameter) . . . . . . . . . . 86 SSID name (Parameter) . . . . . . . . . . . . . . . . . . 176, 179
Pressure unit (Parameter) . . . . . . . . . . . . . . . . . . . . . 73 Status input 1 to n (Submenu) . . . . . . . . . . . . . . . . . 113
Pressure value (Parameter) . . . . . . . . . . . . . . . . . 53, 93 Status input simulation 1 to n (Parameter) . . . . . . . . 239
Previous diagnostics (Parameter) . . . . . . . . . . . . . . . 196 Submenu
Process parameters (Submenu) . . . . . . . . . . . . . . . . . 82 Administration . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
Process variable adjustment (Submenu) . . . . . . . . . . 101 Application . . . . . . . . . . . . . . . . . . . . . . . . . . . . 181
Process variable value (Parameter) . . . . . . . . . . . . . . 239 Application specific calculations . . . . . . . . . . . . . 187
Process variables (Submenu) . . . . . . . . . . . . . . . 50, 191 Application-specific parameters . . . . . . . . . . . . . 188
Progress (Parameter) . . . . . . . . . . . . . . . . . . . . . . . 101 Calculated values . . . . . . . . . . . . . . . . . . . . . . . . . 95
Pulse output (Parameter) . . . . . . . . . . . . . . . . . . 65, 164 Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 106
Pulse output 1 to n (Parameter) . . . . . . . . . . . . . 62, 137 Carrier pipe temperature . . . . . . . . . . . . . . . . . . 229
Pulse output simulation (Parameter) . . . . . . . . . . . . 245 Communication . . . . . . . . . . . . . . . . . . . . . . . . . 164
Pulse output simulation 1 to n (Parameter) . . . . . . . 243 Concentration . . . . . . . . . . . . . . . . . . . . . . . . . . 187
Pulse scaling (Parameter) . . . . . . . . . . . . . . . . . . . . 134 Configuration backup . . . . . . . . . . . . . . . . . . . . . . 30
Pulse value (Parameter) . . . . . . . . . . . . . . . . . . . . . . 246 Corrected volume flow calculation . . . . . . . . . . . . . 95
Pulse value 1 to n (Parameter) . . . . . . . . . . . . . . . . . 243 Current input 1 to n . . . . . . . . . . . . . . . . . . . 59, 110
Pulse width (Parameter) . . . . . . . . . . . . . . . . . 135, 162 Current output 1 to n . . . . . . . . . . . . . . . . . . . . . 116
Pulse/frequency/switch output 1 to n (Submenu) 62, Custody transfer . . . . . . . . . . . . . . . . . . . . . . . . . 187
130 Custody transfer logbook . . . . . . . . . . . . . . . . . . 204
Data logging . . . . . . . . . . . . . . . . . . . . . . . . . . . 216
R Device information . . . . . . . . . . . . . . . . . . . . . . . 205
Received signal strength (Parameter) . . . . . . . . . . . . 180 Diagnostic behavior . . . . . . . . . . . . . . . . . . . . . . . 33
Reference density (Parameter) . . . . . . . . . . . . . . . . . . 52 Diagnostic handling . . . . . . . . . . . . . . . . . . . . . . . 33
Reference density factor (Parameter) . . . . . . . . . . . . 105 Diagnostic list . . . . . . . . . . . . . . . . . . . . . . . . . . 198
Reference density offset (Parameter) . . . . . . . . . . . . 105 Diagnostics . . . . . . . . . . . . . . . . . . . . . . . . . . . . 195
Reference density unit (Parameter) . . . . . . . . . . . . . . 72 Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Reference sound velocity (Parameter) . . . . . . . . . . . . 91 Display channel 1 . . . . . . . . . . . . . . . . . . . . . . . . 223
Reference temperature (Parameter) . . . . . . . . . . . . . . 97 Display channel 2 . . . . . . . . . . . . . . . . . . . . . . . . 224
Relay output 1 to n (Submenu) . . . . . . . . . . . . . . 63, 152 Display channel 3 . . . . . . . . . . . . . . . . . . . . . . . . 225
Relay output 1 to n simulation (Parameter) . . . . . . . 244 Display channel 4 . . . . . . . . . . . . . . . . . . . . . . . . 225
Relay output function (Parameter) . . . . . . . . . . . . . . 153 Display module . . . . . . . . . . . . . . . . . . . . . . . . . . 215
Reset access code (Parameter) . . . . . . . . . . . . . . . . . . 46 Double pulse output . . . . . . . . . . . . . . . . . . . 64, 159
Reset access code (Submenu) . . . . . . . . . . . . . . . . . . . 46 Event list . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 203
Reset all totalizers (Parameter) . . . . . . . . . . . . . . . . 181 Event logbook . . . . . . . . . . . . . . . . . . . . . . . . . . 202
Reset min/max values (Parameter) . . . . . . . . . . . . . 226 External compensation . . . . . . . . . . . . . . . . . . . . . 92
Response time (Parameter) . . . . . . . . . . . . . . . 127, 143 Heartbeat . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 237
Response time part. filled pipe detect. (Parameter) . . . 89 I/O configuration . . . . . . . . . . . . . . . . . . . . . . . . 107
Response time status input (Parameter) . . . . . . . . . . 115 I/O module 1 . . . . . . . . . . . . . . . . . . . . . . . . . . . 210
I/O module 2 . . . . . . . . . . . . . . . . . . . . . . . . . . . 211
S I/O module 3 . . . . . . . . . . . . . . . . . . . . . . . . . . . 213
Scan list register 0 to 15 (Parameter) . . . . . . . . . . . . 171 I/O module 4 . . . . . . . . . . . . . . . . . . . . . . . . . . . 214
Security identification (Parameter) . . . . . . . . . . . . . . 177 Input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 110
Select antenna (Parameter) . . . . . . . . . . . . . . . . . . . 180 Input values . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
Select gas type (Parameter) . . . . . . . . . . . . . . . . . . . . 91 Low flow cut off . . . . . . . . . . . . . . . . . . . . . . . . . . 84
Select medium (Parameter) . . . . . . . . . . . . . . . . . . . . 90 Main electronic module + I/O module 1 . . . . . . . . 208
Sensor (Submenu) . . . . . . . . . . . . . . . . . . . . . . . . . . . 50 Main electronic temperature . . . . . . . . . . . . . . . . 227
Sensor adjustment (Submenu) . . . . . . . . . . . . . . . . . . 98 Measured values . . . . . . . . . . . . . . . . . . . . . . . . . 50
Sensor electronic module (ISEM) (Submenu) . . . . . . 209 Measurement mode . . . . . . . . . . . . . . . . . . . . . . . 90

320 Endress+Hauser
Proline Promass 500 Modbus RS485 Index

Medium index . . . . . . . . . . . . . . . . . . . . . . . . . . 193 Telegram delay (Parameter) . . . . . . . . . . . . . . . . . . 168


Medium temperature . . . . . . . . . . . . . . . . . . . . . 229 Temp. compensated dynamic viscosity (Parameter) . . 54
Min/max values . . . . . . . . . . . . . . . . . . . . . . . . . 225 Temp. compensated kinematic viscosity (Parameter) . 54
Modbus configuration . . . . . . . . . . . . . . . . . . . . 165 Temperature (Parameter) . . . . . . . . . . . . . . . . . . . . . 53
Modbus data map . . . . . . . . . . . . . . . . . . . . . . . 171 Temperature coefficient sound velocity (Parameter) . . 92
Modbus information . . . . . . . . . . . . . . . . . . . . . . 170 Temperature correction source (Parameter) . . . . . . . . 94
Oscillation amplitude . . . . . . . . . . . . . . . . . . . . . 232 Temperature damping (Parameter) . . . . . . . . . . . . . . 83
Oscillation damping . . . . . . . . . . . . . . . . . . . . . . 234 Temperature factor (Parameter) . . . . . . . . . . . . . . . 106
Oscillation frequency . . . . . . . . . . . . . . . . . . . . . 231 Temperature mode (Parameter) . . . . . . . . . . . . . . . . . 95
Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 115 Temperature offset (Parameter) . . . . . . . . . . . . . . . 105
Output values . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61 Temperature unit (Parameter) . . . . . . . . . . . . . . . . . . 72
Partially filled pipe detection . . . . . . . . . . . . . . . . . 87 Terminal number (Parameter) . 110, 113, 116, 131, 153
Petroleum . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 187 Timestamp (Parameter) 196, 197, 198, 199, 200,
Process parameters . . . . . . . . . . . . . . . . . . . . . . . 82 201, 202
Process variable adjustment . . . . . . . . . . . . . . . . 101 Torsion oscillation amplitude (Submenu) . . . . . . . . . 233
Process variables . . . . . . . . . . . . . . . . . . . . . 50, 191 Torsion oscillation damping (Submenu) . . . . . . . . . . 234
Pulse/frequency/switch output 1 to n . . . . . . 62, 130 Torsion oscillation frequency (Submenu) . . . . . . . . . 231
Relay output 1 to n . . . . . . . . . . . . . . . . . . . . 63, 152 Torsion signal asymmetry (Submenu) . . . . . . . . . . . 236
Reset access code . . . . . . . . . . . . . . . . . . . . . . . . . 46 Totalizer (Submenu) . . . . . . . . . . . . . . . . . . . . . . . . . 58
Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50 Totalizer 1 to n (Submenu) . . . . . . . . . . . . . . . . . . . 182
Sensor adjustment . . . . . . . . . . . . . . . . . . . . . . . . 98 Totalizer operation mode (Parameter) . . . . . . . . . . . 184
Sensor electronic module (ISEM) . . . . . . . . . . . . . 209 Totalizer overflow 1 to n (Parameter) . . . . . . . . . . . . . 58
Sensor electronic temperature (ISEM) . . . . . . . . . 228 Totalizer value 1 to n (Parameter) . . . . . . . . . . . . . . . 58
Signal asymmetry . . . . . . . . . . . . . . . . . . . . . . . . 235
Simulation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 237 U
Status input 1 to n . . . . . . . . . . . . . . . . . . . . . . . 113 Unit totalizer 1 to n (Parameter) . . . . . . . . . . . . . . . 183
System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15 User corrected volume factor (Parameter) . . . . . . . . . 78
System units . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65 User corrected volume offset (Parameter) . . . . . . . . . . 78
Torsion oscillation amplitude . . . . . . . . . . . . . . . 233 User corrected volume text (Parameter) . . . . . . . . . . . 77
Torsion oscillation damping . . . . . . . . . . . . . . . . 234 User density factor (Parameter) . . . . . . . . . . . . . . . . . 79
Torsion oscillation frequency . . . . . . . . . . . . . . . 231 User density offset (Parameter) . . . . . . . . . . . . . . . . . 79
Torsion signal asymmetry . . . . . . . . . . . . . . . . . . 236 User density text (Parameter) . . . . . . . . . . . . . . . . . . 78
Totalizer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58 User energy factor (Parameter) . . . . . . . . . . . . . . . . . 81
Totalizer 1 to n . . . . . . . . . . . . . . . . . . . . . . . . . . 182 User energy offset (Parameter) . . . . . . . . . . . . . . . . . 81
User-specific units . . . . . . . . . . . . . . . . . . . . . . . . 74 User energy text (Parameter) . . . . . . . . . . . . . . . . . . . 80
Value current output 1 to n . . . . . . . . . . . . . . . . . . 61 User mass factor (Parameter) . . . . . . . . . . . . . . . . . . 77
Value status input 1 to n . . . . . . . . . . . . . . . . . . . . 60 User mass offset (Parameter) . . . . . . . . . . . . . . . . . . . 76
Viscosity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 187 User mass text (Parameter) . . . . . . . . . . . . . . . . . . . . 76
Web server . . . . . . . . . . . . . . . . . . . . . . . . . . . . 171 User name (Parameter) . . . . . . . . . . . . . . . . . . . . . . 177
WLAN settings . . . . . . . . . . . . . . . . . . . . . . . . . . 174 User pressure factor (Parameter) . . . . . . . . . . . . . . . . 82
Zero point adjustment . . . . . . . . . . . . . . . . . . . . 100 User pressure offset (Parameter) . . . . . . . . . . . . . . . . 81
Subnet mask (Parameter) . . . . . . . . . . . . . . . . . . . . 173 User pressure text (Parameter) . . . . . . . . . . . . . . . . . 81
Switch cycles (Parameter) . . . . . . . . . . . . . . . . . . . . . 64 User specific-enthalpy factor (Parameter) . . . . . . . . . 80
Switch output function (Parameter) . . . . . . . . . . . . . 145 User specific-enthalpy offset (Parameter) . . . . . . . . . . 80
Switch output simulation 1 to n (Parameter) . . . . . . 243 User specific-enthalpy text (Parameter) . . . . . . . . . . . 79
Switch status (Parameter) . . . . . . . . . . . . . . . . . 64, 158 User volume factor (Parameter) . . . . . . . . . . . . . . . . . 76
Switch status 1 to n (Parameter) . . . . 63, 151, 244, 245 User volume offset (Parameter) . . . . . . . . . . . . . . . . . 75
Switch-off delay (Parameter) . . . . . . . . . . . . . . 151, 157 User volume text (Parameter) . . . . . . . . . . . . . . . . . . 75
Switch-off value (Parameter) . . . . . . . . . . . . . . 149, 156 User-specific units (Submenu) . . . . . . . . . . . . . . . . . . 74
Switch-on delay (Parameter) . . . . . . . . . . . . . . 150, 158
Switch-on value (Parameter) . . . . . . . . . . . . . . 149, 157 V
System (Submenu) . . . . . . . . . . . . . . . . . . . . . . . . . . 15 Value 1 display (Parameter) . . . . . . . . . . . . . . . . . . . 19
System units (Submenu) . . . . . . . . . . . . . . . . . . . . . . 65 Value 2 display (Parameter) . . . . . . . . . . . . . . . . . . . 22
Value 3 display (Parameter) . . . . . . . . . . . . . . . . . . . 23
T Value 4 display (Parameter) . . . . . . . . . . . . . . . . . . . 25
Target corrected volume flow (Parameter) . . . . . . . . . 56 Value current input 1 to n (Parameter) . . . . . . . . . . . 241
Target group . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 Value current output 1 to n (Parameter) . . . . . . . . . . 241
Target mass flow (Parameter) . . . . . . . . . . . . . . . . . . 55 Value current output 1 to n (Submenu) . . . . . . . . . . . 61
Target volume flow (Parameter) . . . . . . . . . . . . . . . . 57 Value per pulse (Parameter) . . . . . . . . . . . . . . . . . . 161

Endress+Hauser 321
Index Proline Promass 500 Modbus RS485

Value status input (Parameter) . . . . . . . . . . . . . 60, 115


Value status input 1 to n (Submenu) . . . . . . . . . . . . . 60
Viscosity (Submenu) . . . . . . . . . . . . . . . . . . . . . . . . 187
Volume flow (Parameter) . . . . . . . . . . . . . . . . . . . . . 52
Volume flow factor (Parameter) . . . . . . . . . . . . . . . . 103
Volume flow offset (Parameter) . . . . . . . . . . . . . . . . 103
Volume flow unit (Parameter) . . . . . . . . . . . . . . . . . . 67
Volume unit (Parameter) . . . . . . . . . . . . . . . . . . . . . . 68

W
Web server (Submenu) . . . . . . . . . . . . . . . . . . . . . . 171
Web server functionality (Parameter) . . . . . . . . . . . 174
Web server language (Parameter) . . . . . . . . . . . . . . 172
Wizard
Define access code . . . . . . . . . . . . . . . . . . . . . . . . 45
WLAN (Parameter) . . . . . . . . . . . . . . . . . . . . . . . . . 175
WLAN IP address (Parameter) . . . . . . . . . . . . . . . . . 177
WLAN MAC address (Parameter) . . . . . . . . . . . . . . 178
WLAN mode (Parameter) . . . . . . . . . . . . . . . . . . . . 176
WLAN passphrase (Parameter) . . . . . . . . . . . . . . . . 178
WLAN password (Parameter) . . . . . . . . . . . . . . . . . 177
WLAN settings (Submenu) . . . . . . . . . . . . . . . . . . . 174
WLAN subnet mask (Parameter) . . . . . . . . . . . . . . . 178

Z
Zero point (Parameter) . . . . . . . . . . . . . . . . . . . . . . 107
Zero point adjustment (Submenu) . . . . . . . . . . . . . . 100
Zero point adjustment control (Parameter) . . . . . . . . 100
Zero point adjustment status (Parameter) . . . . . . . . 101

322 Endress+Hauser
www.addresses.endress.com

You might also like