A Single-Shot Method For Capillary Pressure Curve Measurement Using Centrifuge
A Single-Shot Method For Capillary Pressure Curve Measurement Using Centrifuge
the rotational speeds permit a quantification of the capillary length of the core was negligible compared to the radius of
pressure as a function of fluid saturation [8-10]. A full rotation.
capillary pressure curve determined with traditional
In a centrifuge capillary pressure experiment, a fluid
centrifugal methods requires approximately 10 different
saturated core plug, confined in a special core-holder, is
centrifuge speeds, thus the measurement for a capillary
rotated at different rotational speeds. The relevant distances
pressure curve takes several days to more than a week
are denoted as r1, r, and r2, as illustrated in Figure 1, where r1,
depending on equilibrium time. In addition, the experiment
r2, and r are the distances from the rotational axis to the inlet
requires a very expensive ultracentrifuge with precise speed face, the outlet face, and any point along the core length,
control over a wide range of speeds. This technique has been
The average fluid saturation of the core after from '0' to '-kz', corresponding to a gradient change from 0 to
centrifugation can be expressed as minus maximum gradient (-Gmax), after a delay of 5 times
1 r2 T 1, the other of half k-space data is collecting from '0' to
S= ∫
r2 − r1 r1
S (r )dr (8) '+kz ', corresponding to a gradient change from 0 to a
maximum gradient (Gmax). Fourier transformation of the k-
space data yields a real space image. In the centric scan
Equation (8) may be rewritten by changing the integration SPRITE method, the image signal intensity (S) is given by:
variable Pc(r2)=0 and Pc(r1)=PcL, with additional tp
mathematical manipulation, which yields the Hassler- S (r ) = M 0 (r ) exp(− ) sin α (12)
spin echo based MRI methods, due to multi-exponential T2 by increasing the image matrix size and/or decreasing the field
decay in rocks [11]. of view of the image, which results in more data points on the
capillary pressure curve.
100 Figures 4 and 5 show the 1D water content
FID signal (a.u.)
dried sample was determined. The samples were kept under 0.2
vacuum conditions for more than 24 hours, and then saturated
with distilled and degassed water under vacuum conditions 0.0
0 2 4 6 8 10
until no bubbles came from the core samples. For the Berea Z (cm)
sandstone sample, the porosity was 0.19, and the permeability
was 0.20 Darcy. For the oilfield sandstone core #126, the Figure 4. One dimensional water content distribution along the
porosity was 0.28, and the permeability was 0.37 Darcy. Each length (Z) of the fully water saturated Berea sandstone core,
sample was 5.2 cm in length and 2.5 cm in diameter. determined with 1D centric scan SPRITE MRI.
0.8
TJ6R tabletop centrifuge with arm length of 18.6 cm, at 1500
RPM for 140 minutes. The temperature in the centrifuge was 0.6
controlled to be 3°C to avoid water evaporation during the
centrifuge experiment. The duration of centrifugation is 0.4
9. The core sample was put into a D2O filled sample determined in practice. The current methodology shows they
tube for centrifugation at 3000 rpm in centrifuge for may now be determined with some ease.
48 hours for imbibition capillary pressure curve
measurement. 100
10. Steps 6 to 7 were repeated.
11. The core sample was flipped and spun in the 80
Sw (% )
centrifuge, then flipped and spun again several times 60
to reach a relative uniform residual oil saturation (Sor)
distribution along the length of the core sample to 40
Sw (% )
experimental procedures in the previous section. 60
The centrifuge experiment was carried out with a
Z513K tabletop centrifuge (Hermle Labortechnik, Germany) 40
with arm length of 23.5 cm, at 3000 RPM for 24 or 48 hours. 20
The MRI experiments were carried out in a 0.2 Tesla
0
permanent magnet with an Apollo console (Tecmag Inc.,
0 2 4 6 8 10
Houston, TX). A 3 cm inner diameter solenoid probe was
Z (cm)
employed. The core sample was wrapped with Teflon tape to
decrease the evaporation of liquid within the samples during Figure 10. Water saturation distribution along the length (Z) of
core #126 after centrifugation at 3000 rpm in water for 48 hours
MRI measurements. The advantages of using a low-field MRI for imbibition capillary pressure measurement.
instrument are that (1) the instrument is cheap, and (2) the
effective spin-spin relaxation time (T2*) is much longer than
100
the phase encoding time (tp). This ensures the SPRITE MRI
image is a simple fluid distribution image. 80
Sw (% )