CPC 100 PTM User Manual ENU
CPC 100 PTM User Manual ENU
User Manual
CPC 100 PTM User Manual
2 OMICRON
Contents
Contents
About this manual 7
Related documents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Conventions and symbols used . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
1 Safety instructions 9
1.1 Operator qualifications and safety standards . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
1.2 Orderly measures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
1.3 Disclaimer. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
1.3.1 Automatic assessment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
1.3.2 Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
2 System requirements 11
3 Introduction 12
3.1 The test set. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
3.2 Installing Primary Test Manager. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
3.3 Firewall configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
3.4 Starting Primary Test Manager and connecting to the CPC 100 . . . . . . . . . . . . . . . . . . . . . . . . . . 17
3.4.1 Starting via the PTM icon . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
3.4.2 Starting via the CPC Start Page. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
3.4.3 Connecting to the CPC 100 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
3.4.4 Updating the CPC 100 embedded software. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
3.5 Primary Test Manager licensing. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
4 Home view 20
4.1 Title bar . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
4.1.1 Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
4.1.2 Send data to technical support. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
4.1.3 About . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
4.2 Manage connection to the test system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
4.3 Data synchronization . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
4.3.1 Server settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
4.3.2 Managing subscriptions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
4.3.3 Database synchronization . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
4.4 Import data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
4.5 Data backup and restoring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
4.6 Status bar . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
4.6.1 Connecting to a test system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
4.6.2 Manual connection to a test system. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
5 Create new guided jobs 41
5.1 Guided test workflow . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
5.2 Job overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
5.2.1 Job data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
5.2.2 Approving jobs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
5.2.3 Assessment summary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
5.2.4 Managing attachments. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
5.3 Location view . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
5.3.1 Location data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
5.3.2 Setting the geo coordinates . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
5.3.3 Managing attachments. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
5.4 Asset view . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
5.4.1 General asset data. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
5.4.2 Managing attachments. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
OMICRON 3
CPC 100 PTM User Manual
4 OMICRON
Contents
OMICRON 5
CPC 100 PTM User Manual
6 OMICRON
About this manual
This manual provides information on how to use Primary Test Manager safely, properly and efficiently.
The CPC 100 PTM User Manual gets you familiar with operating Primary Test Manager and guides you
step by step through the test procedures with the OMICRON test systems.
This manual and the technical documentation shipped with the test systems always have to be available
on the site where Primary Test Manager is used. Read all manuals carefully before using Primary Test
Manager and observe the safety, installation, and operation instructions therein. Reading this manual
alone does not release you from the duty to comply with all national and international safety regulations
relevant to working on the high-voltage equipment.
Note: This User Manual describes Primary Test Manager with all available features. Please note that
the features are only available if you have bought corresponding license keys. For more information, see
3.5 "Primary Test Manager licensing" on page 19.
Related documents
The following documents complete the information covered in this manual.
Title Description
CPC 100 User Manual Contains safety and operation instructions on the CPC 100
test system.
CPC 100 Reference Manual Contains detailed hardware and software information on the
CPC 100 including relevant safety instructions.
CPC 80 User Manual Contains safety and operation instructions on the CPC 80 test
system.
CP CR500 User Manual Contains safety and operation instructions on the CP CR500
test system.
CP CR600 User Manual Contains safety and operation instructions on the CP CR600
test system.
CP TD1 User Manual Contains hardware and software information about the
CP TD1 test system.
CP TD12/15 User Manual Contains hardware and software information about the
CP TD12 and CP TD15 test systems.
CP SB1 User Manual Contains hardware and software information about the
CP SB1 transformer switch box.
HGT1 User Manual Contains safety and operation instructions on the HGT1.
CP CU1 User Manual Contains safety and operation instructions on the CP CU1
test system.
CPOL 2 User Manual Contains safety and operation instructions on the CPOL 2 test
system.
OMICRON 7
CPC 100 PTM User Manual
DANGER
Death or severe injury will occur if the appropriate safety instructions are not observed.
WARNING
Death or severe injury can occur if the appropriate safety instructions are not observed.
CAUTION
Minor or moderate injury may occur if the appropriate safety instructions are not observed.
NOTICE
Equipment damage or loss of data possible
8 OMICRON
Safety instructions
1 Safety instructions
Before operating Primary Test Manager and the test set, read the following safety instructions carefully.
When testing with Primary Test Manager, observe the safety rules in this chapter and all relevant
safety instructions in the technical documentation shipped with the test systems.
Do not turn on or operate the CPC 100 if you do not fully understand the safety information. If you have
any questions regarding the safety instructions, contact OMICRON before proceeding.
Before connecting any test leads to a test object, always observe the five safety rules:
► Disconnect completely
► Secure from reconnection
► Verify that the installation is dead
► Carry out grounding and short-circuiting
► Provide protection against adjacent live parts
Safety accessories
OMICRON offers a range of accessories for added safety during the operation of our test systems. For
further information and specifications, refer to the corresponding Supplementary Sheet or contact
OMICRON Support (see "Support" on page 209).
OMICRON 9
CPC 100 PTM User Manual
1.3 Disclaimer
1.3.1 Automatic assessment
The applied auto-assessment rules (for example, indicated colors, indications) and limits within the
Program are defined according to industry standards and/or are based on the experience of industry
experts. The purpose of the auto-assessment is to indicate to the person that uses the Program – "User"
any deviation of the measured parameters from the expected values based on the recommended limits
provided by OMICRON or defined by the User. Any decision to return the apparatus to service or remove
the apparatus from service cannot be based on the results of the auto-assessment alone. This decision
is the sole responsibility of the owner or end user of the apparatus.
1.3.2 Safety
If Primary Test Manager and the test set are used in a manner not specified by the manufacturer, the
protection provided by the test systems may be impaired.
10 OMICRON
System requirements
2 System requirements
Table 2-1: Primary Test Manager system requirements
OMICRON 11
CPC 100 PTM User Manual
3 Introduction
Primary Test Manager is a management tool for testing primary assets such as power transformers,
circuit breakers, and current transformers with the OMICRON test systems. Primary Test Manager
provides a computer interface to the test set, controls the automated test procedures, and facilitates
testing of primary assets by guiding you through the test workflow.
Primary Test Manager uses the concept of jobs. A job contains all relevant information about the
location, the asset under test, and the tests. With Primary Test Manager, you can process jobs as
separate entities.
With Primary Test Manager, you can create new jobs, manage locations, assets, jobs, and reports, and
execute prepared jobs. For a specified job, you can make measurements on the asset under test by just
pressing the I/0 button on the front panel of the CPC 100 test system. After you have performed a test,
you can generate exhaustive test reports. Primary Test Manager runs on a computer and communicates
with the test set through the Ethernet connection.
12 OMICRON
Introduction
I/0 button
Network connector
For detailed information about the test set, see the technical documentation shipped with the test
systems.
OMICRON 13
CPC 100 PTM User Manual
Windows firewall
The configuration of the Windows firewall is carried out automatically during the installation of Primary
Test Manager. However, in certain cases this may have no immediate effect.
► To prevent the Windows firewall from blocking communication, (temporarily) disable it via the
Windows Control panel.
If you are now able to successfully establish communication, the Windows firewall was the reason for
the blocked communication between your test set and your computer.
► Reconfigure the Windows firewall in order to enable a permanent use of the test set without having
to disable the Windows firewall.
Third-party firewall
► If you are using a firewall other than the Windows firewall, temporarily disable it to see if this firewall
may be the cause for the blocked communication.
Note: Numerous computer security programs or anti-virus packages also contain an integrated firewall
function. Double-check and, if applicable, remove all such programs that may be installed on your
computer.
14 OMICRON
Introduction
Program/service Rule name Protocol Local Port Remote Port Local IP Remote IP
name type
OMICRON 15
CPC 100 PTM User Manual
16 OMICRON
Introduction
2. On the CPC start page, select the CPC 100 device to which you want to connect.
3. Under Tools, click Primary Test Manager.
OMICRON 17
CPC 100 PTM User Manual
If you were not able to connect to your CPC 100 device and the green light is permanently on, wait a few
seconds, and then do one of the following:
► Click More (under the Connect button), then click Refresh.
► Press F5.
If the CPC 100 device to which you want to connect is not displayed in the list of available devices,
proceed as described in 4.6.2 "Manual connection to a test system" on page 40.
Alternatively, you can manage the connection to CPC 100 in the Primary Test Manager status bar (see
4.3 "Data synchronization" on page 35).
18 OMICRON
Introduction
License Description
PTM Standard Manual control mode with tests according to your CPC 100 license.
Additional 30 testing days with guided workflow according to your
CPC 100 license.
PTM Advanced Timely unlimited testing with guided workflow and manual tests according
to your CPC 100 license.
The PTM Advanced license key is on the CPC 100. You can upgrade your CPC 100 on the CPC start
page (see Figure 3-3: "CPC start page" on page 17) or via Device Link.
OMICRON 19
CPC 100 PTM User Manual
4 Home view
After starting Primary Test Manager, the home view opens. In the home view, you can select different
user tasks designed to support you during diagnostic testing and management of test objects and test
data.
Primary Test Manager processes data of different workflow importance. This is indicated by balloons of
different categories as described in the following table.
20 OMICRON
Home view
OMICRON 21
CPC 100 PTM User Manual
The following table describes the available user interface commands in the Home view.
Synchronize1 Click to synchronize your local database with the Primary Test Manager
server database.
(see 4.6 "Status bar" on page 38).
Import files Click to import Primary Test Manager data.
(see 4.4 "Import data" on page 37).
Back up your data Click to backup the Primary Test Manager database.
(see 4.5 "Data backup and restoring" on page 38).
Restore data from a Click to restore your data in the database
backup (see 4.5 "Data backup and restoring" on page 38).
Information
Click a list item to get information about your test system and its application.
Recent guided jobs/Recent manual jobs
Click a list item to open a recently created guided or manual job.
Status bar
In the status bar, you can connect to and disconnect from a test system and view the test set
information (see 4.6 "Status bar" on page 38).
1. Only enabled with the appropriate license.
22 OMICRON
Home view
4.1.1 Settings
In the Settings dialog box, you can make a number of Primary Test Manager settings to match your
regional conventions, manage the job templates, and set the Primary Test Manager server settings for
data synchronization (see 4.6 "Status bar" on page 38). To open the Settings dialog box, click Settings
in the title bar.
NOTICE
Equipment damage or loss of data possible
Changing the settings in the Settings dialog box affects all data in Primary Test Manager.
► Only change settings if you are qualified to do so.
► Review your changes before clicking OK.
Note: After changing a setting, you must restart Primary Test Manager for the setting to take effect.
General
The General tab lists the general settings of Primary Test Manager.
► To set the Primary Test Manager language, select your preferred language from the Language list.
OMICRON 23
CPC 100 PTM User Manual
► To set the logging level, select your preferred level from the Logging PTM and Logging Device lists.
The logging function provides information to help find the cause for an error in cooperation with an
OMICRON support engineer. Logging PTM collects information on PTM while Logging Device
focuses on your device.
Note: Log files do not contain any personal information.
► To set the types of available devices, select the respective check boxes.
Location settings
In this section you can activate the Use location information check box to allow GPS coordinates to
be shared by a smartphone running the PTMate app connected via Bluetooth or a Wi-Fi access point.
Note: Further information regarding how to connect via Wi-Fi access point can be found later in this
chapter.
Primary Test Manager displays status icons in the bottom bar:
24 OMICRON
Home view
Remote control
Certain features of Primary Test Manager can be controlled via the PTMate app. Complete the steps
below to establish the connection between your smartphone and your computer.
1. Activate the Allow remote control via PTMate check box in the Remote control section of the PTM
Settings. Primary Test Manager will establish a Wi-Fi access point.
► If both your smartphone and your computer are already connected to the same Wi-Fi network,
proceed to step 2 below.
► If you are not connected to a Wi-Fi network, press the Start Wi-Fi access point button. PTM will
attempt to create a Wi-Fi access point and refresh the displayed QR code.
Note: If your computer does not support ad hoc Wi-Fi access point creation, you can use an external
Wi-Fi device supporting this functionality or create a hotspot on your smartphone. Be aware that
using a smartphone hotspot can lead to additional costs.
2. Open the PTMate app on your smartphone, go to Settings and scan the QR code displayed in the
Remote control section in PTM.
Primary Test Manager displays status icons in the bottom bar:
OMICRON 25
CPC 100 PTM User Manual
Profiles
In the Profiles tab, you can set your profile, the default rated frequency, the loss index, the units of your
own profiles, and make the test system settings.
With Primary Test Manager, you can use predefined profiles and create your own profiles for naming
conventions.
Note: Primary Test Manager sets the default profile according to the regional settings of your computer.
► To set a profile, select the profile you want to use from the Profiles list.
To create your own profile:
1. Click Create custom profile.
2. In the Create custom profile dialog box, type the profile name, and then click Create.
3. Under Global settings, set the default rated frequency, the loss index, and your preferred units.
26 OMICRON
Home view
Profiles: Transformer
4. Under Transformer, set the transformer terminal name schemes and preferences such as the
names of some tests, the oil measure, and the short-circuit impedance abbreviation.
With Primary Test Manager, you can use predefined transformer naming conventions according to the
established standards and create your own terminal name schemes.
► To set a terminal names scheme, select the scheme you want to use from the
Terminal names scheme list.
To create your own terminal names scheme:
a) Click Create custom scheme.
b) In the Enter scheme name dialog box, type the scheme name.
c) Set the transformer terminal names, scheme options, and preferences.
► To delete your own terminal name scheme, select the scheme from the Terminal names scheme
list, and then click Delete current scheme.
OMICRON 27
CPC 100 PTM User Manual
28 OMICRON
Home view
OMICRON 29
CPC 100 PTM User Manual
Templates
In the Templates tab, you can edit, export and import job templates for transformers.
► For information on how to process the templates, see 5.5.6 "Processing templates" on page 79.
The right pane of the template workspace displays the template preview.
To manage the job templates, select Transformer from the Display templates for list, and then do one
of the following:
► To assign a template to a different asset type or a phase group, or to edit template properties (name,
comment), click the respective Edit button.
► Press the Delete button to delete templates from the Asset type or Phases list.
► To export a template, select the template, and click the Export button.
► To import a template, click the Import button, then browse to the template you want to import.
► To set a template as favorite, click the star icon .
Note: All future test lists with the same asset and number of phases will by default be loaded with the
tests defined in this favorite template.
Note: If you own a CBTL (Circuit Breaker Testing Library) licence, additional options are available in
the Circuit breaker section:
• Show template
• Show OMICRON CBTL
• Show custom CBTL
30 OMICRON
Home view
Databases
In the Databases tab you can create, manage and switch between different databases for Primary Test
Manager. Under Properties, you can adjust the server settings for DataSync. For more information, see
4.3.1 "Server settings" on page 35.
OMICRON 31
CPC 100 PTM User Manual
1. In the Omicron Assist dialog box, select your region, and then click Next.
2. Omicron Assist will automatically collect your system information. Once this step is complete, click
Next.
3. Click Add files.
4. Browse to the data you want to send, and then click Next.
5. Click Prepare e-mail.
32 OMICRON
Home view
4.1.3 About
In the About Primary Test Manager dialog box, you can enter a license key to upgrade your Primary
Test Manager and enhance its functionality by installing additional features. To open the
About Primary Test Manager dialog box, click About in the title bar.
To activate a license:
1. Enter the license key in the About Primary Test Manager dialog box, and then click
Add license key.
The About Primary Test Manager dialog box displays the available licenses and a new
Enter license key box.
2. Repeat step 1 for all license keys you want to enter.
Alternatively, you can enter license keys from files. To a enter a license key from a file, click
Add license from file, and then browse to the file containing the license you want to add.
For detailed information about the Primary Test Manager licensing, contact your OMICRON local sales
representative or distributor.
OMICRON 33
CPC 100 PTM User Manual
If you could not connect to your device and the green light is permanently on, wait a few seconds, and
then proceed as follows:
1. Click More next to the Connect button, and then click Refresh.
2. Select the test system from the list, and then click Connect.
If the device to which you want to connect is not displayed in the list of available devices, proceed as
described in 4.6.2 "Manual connection to a test system" on page 40.
Alternatively, you can manage the connection to devices in the Primary Test Manager status bar (see
4.6 "Status bar" on page 38).
34 OMICRON
Home view
OMICRON 35
CPC 100 PTM User Manual
DataSync on premises
► For the service URL and certificate for DataSync on premises, contact your system administrator.
2. In the Databases tab, enter the Service URL and upload the Certificate.
3. To test the connection to the server, click Test next to the Connection status.
2. In the manage view, click Manage subscriptions on the top of the workspace.
3. In the Subscriptions dialog box, select the data on the server you want to synchronize with your local
data.
You can synchronize the data anytime.
► To synchronize the data, click Synchronize in the title bar.
After that, Primary Test Manager shows the synchronization progress.
36 OMICRON
Home view
OMICRON 37
CPC 100 PTM User Manual
38 OMICRON
Home view
2. In the Connect to device dialog box, select the test system from the list, and then click Connect.
Note: Select the Automatic connection if only one device detected check box if only one device is
available. Then Primary Test Manager connects to the available device automatically.
If you could not connect to your CPC 100 device and the green light is permanently on, wait a few
seconds, and then proceed as follows:
1. Click More next to the Connect button.
2. Click Refresh.
3. In the Connect to device dialog box, select the test system from the list, and then click Connect.
After you have connected to the test system, the following dialog box appears.
OMICRON 39
CPC 100 PTM User Manual
After you have connected to a test system, right-click the CPC 100 icon in the status bar, and then do
one of the following:
► To disconnect from a test system, click Disconnect.
► To display information about the connected test system, click Show test set information.
► To update the test set information, click Refresh test set information.
Note: You can open the Connect to device and the Connected to device dialog boxes also by double-
clicking the CPC 100 icon.
40 OMICRON
Create new guided jobs
In the new guided job view, you can configure and execute jobs. A job contains all relevant information
about the location, the asset under test, and the tests. With Primary Test Manager, you can process jobs
as separate entities. During the guided test workflow, the job status displayed in the left pane of the new
guided job view changes. The following table describes the job statuses.
Status Description
New Location has been defined.
Prepared Asset has been defined.
Partially executed At least one measurement has been executed.
Executed All tests of the job have been executed.
Approved Job has been approved.
OMICRON 41
CPC 100 PTM User Manual
Note: You can anytime interrupt the test workflow and return to any view by clicking the corresponding
navigation button.
42 OMICRON
Create new guided jobs
By using the commands on the menu bar, you can process jobs. The following table describes the
available operations.
Command Action
Home/Manage Closes a job displayed in the create new job view and leads you back to
home or manage view respectively.
Save job Saves the job displayed in the create new job view.
Export job Exports the job displayed in the new guided job view into a Microsoft
Excel spreadsheet.
Load existing location1 Load an existing location available in Primary Test Manager.
Load existing asset2 Load an existing asset available in Primary Test Manager.
Copy test3 Adds another test of the same kind and with the same settings to the test
list. Results are not copied.
Delete test Deletes a test
Take screenshot Takes screenshot of the selected area of the Primary Test Manager
workspace. The screenshot appears as attachment in the General area
and can be attached to the test report (see 10 "Generate test reports" on
page 97).
1. Only available if the Location view is open and job has not been saved yet.
2. Only available if the Asset view is open and job has not been saved yet.
3. Only available if a test is open
For more information about operations on the jobs, see 6 "Manage objects" on page 85.
OMICRON 43
CPC 100 PTM User Manual
Data Description
44 OMICRON
Create new guided jobs
Data Description
Approved by Person who approved the test
Approval date Date the job was approved (see 5.2.2 "Approving jobs" later in this
section)
Summary Text field to summarize the job data in own words.
Attachments Attachments to the job (see 5.2.4 "Managing attachments" on page 47)
1. Mandatory data
Status Description
Fail The status was automatically set to Fail by Primary Test Manager.
Manual fail The status was manually set to Fail.
Investigate The status was automatically set to Investigate by Primary Test Manager.
Manual investigate The status was manually set to Investigate.
Pass The status was automatically set to Pass by Primary Test Manager.
Partial pass Some measurements have not been assessed.
OMICRON 45
CPC 100 PTM User Manual
Status Description
Manual pass The status was manually set to Pass.
Manual partial pass Some measurements have not been assessed and at least one
assessment status was changed manually.
Not assessed The measurement has not been assessed.
Not rated The status was automatically set to Not Rated by Primary Test Manager.
46 OMICRON
Create new guided jobs
OMICRON 47
CPC 100 PTM User Manual
In the Select Location dialog box, you can search for locations (see 6.1 "Search for objects" on
page 86).
Data Description
48 OMICRON
Create new guided jobs
Data Description
Plant Plant where the asset is located
Address Address of the location
City City where the asset is located
State/Province State or province where the asset is located
Postal code Postal code of the location
Country Country where the asset is located
Geo coordinates Geo coordinates of the location
(see 5.3.2 "Setting the geo coordinates" later in this section)
Location system code Location code used by maintenance planning systems
Contact Person
Name Name of the contact person
Phone no. 1 Phone number of the contact person
Phone no. 2 Alternative phone number of the contact person
E-mail E-mail address of the contact person
Company
Company Company where the asset is located
Department Department of the company
Address Address of the company
City City where the company is located
State/Province State or province where the company is located
Postal code Postal code of the company
Country Country where the company is located
Phone no. Phone number of the contact person
Fax no. Fax number of the contact person
E-mail E-mail address of the contact person
1. Mandatory data
In the location view, you can enter additional addresses of, for example, a client, owner or utility. To enter
additional addresses, click Add address under Additional addresses. For the additional address data,
see Table 5-6: "Location data" on page 48.
OMICRON 49
CPC 100 PTM User Manual
2. In the Edit Coordinates dialog box, enter the latitude and longitude of the location.
Note: If you have a smartphone running the PTMate app connected via Bluetooth to Primary Test
Manager, you have the option to add coordinates through live location sharing by clicking the Geo
coordinates button .
50 OMICRON
Create new guided jobs
The asset view depends on the asset you want to specify with Primary Test Manager. To specify an
asset, do one of the following:
► Enter the asset data. The asset data includes the general asset data common to all assets (see
5.4.1 "General asset data" on page 52 and the asset-specific data described in chapter 11 "Asset
data" on page 99).
Note: If you enter location or asset data for a prepared job that differ from those of the master location
or master asset, a notification bar will be displayed. In this case, choose between the following options:
► To import the previously defined location or asset data to this job, click Import from master
location or Import from master asset in the notification bar.
► To update the previously defined location or asset data with the data you have entered for this job,
click Update master location or Update master asset in the notification bar.
► For more information on operations on the jobs, see 6 "Manage objects" on page 85.
OMICRON 51
CPC 100 PTM User Manual
► To load the asset data available in Primary Test Manager, click Load existing asset, and then select
the asset you want to load in the Select Asset dialog box.
In the Select Asset dialog box, you can search for assets (see 6.1 "Search for objects" on page 86) and
sort them alphabetically or in the chronological order.
Data Description
1. Mandatory data
2. Only available for current transformers, voltage transformers and miscellaneous assets.
52 OMICRON
Create new guided jobs
OMICRON 53
CPC 100 PTM User Manual
3. In the transformer view, enter the general asset data (see Table 5-7: "General asset data" on
page 52).
4. Under Winding configuration, set the transformer’s vector group (see "Setting the vector group"
later in this section).
5. Under Ratings, Impedances and Others, enter the transformer data (see 11.1 "Transformer" on
page 99).
6. Optionally, specify the bushing mounted on the transformer (see "Bushings tab (Transformer)" on
page 56).
7. Optionally, specify the tap changers of the transformer (see "Tap changers tab" on page 57).
8. Optionally, specify the surge arresters mounted on the transformer (see "Surge arresters tab" on
page 58).
54 OMICRON
Create new guided jobs
Note: Primary Test Manager sets the vector group of an autotransformer without tertiary winding
automatically.
OMICRON 55
CPC 100 PTM User Manual
Specifying a bushing
1. From the Asset type list, select the type of the bushing.
2. Enter the bushing data (see 11.2 "Spare bushing" on page 102).
Under Copy bushing data, you can copy data of a bushing to other bushings. To copy the bushing data,
select the respective bushings from the From and To lists, and then click Copy.
56 OMICRON
Create new guided jobs
OMICRON 57
CPC 100 PTM User Manual
58 OMICRON
Create new guided jobs
DGA Trending
DGA Trending is a licensed feature that visualizes a transformer’s historic Oil analysis data in various
charts and offers a comparison of data recorded at different points in time.
► Refer to 12.1.1 "Oil Analysis" on page 117 for more detailed information on the Oil analysis test.
3. In the spare bushing view, enter the general asset data (see Table 5-7: "General asset data" on
page 52).
4. Enter the spare bushing data (see 11.2 "Spare bushing" on page 102).
OMICRON 59
CPC 100 PTM User Manual
3. In the circuit breaker view, enter further general asset data (see Table 5-7: "General asset data" on
page 52).
4. In the Circuit breaker area, enter the circuit breaker data (see 11.3 "Circuit breaker" on page 103).
5. Specify the operating mechanism of the circuit breaker (see "Operating mechanism tab" on page 61).
6. Optionally, specify the bushings mounted on the circuit breaker (see "Bushings tab (Circuit breaker)"
on page 62).
7. Set the assessment limits of the circuit breaker (see "Assessment limits tab" on page 63).
60 OMICRON
Create new guided jobs
To specify an operating mechanism, enter the operating mechanism data (see 11.3.1 "Operating
mechanism" on page 104).
OMICRON 61
CPC 100 PTM User Manual
To specify a bushing:
1. From the Asset type list, select the type of the bushing for all circuit breaker’s terminals.
2. Enter the bushing data (see 11.2 "Spare bushing" on page 102).
Under Copy bushing data, you can copy data of a bushing to other bushings. To copy the bushing data,
select the respective bushings from the From and To lists, and then click Copy.
62 OMICRON
Create new guided jobs
OMICRON 63
CPC 100 PTM User Manual
2. In the current transformer view, enter the general asset data (see Table 5-7: "General asset data"
on page 52).
3. Enter the current transformer data (see 11.4 "Current transformer" on page 107).
64 OMICRON
Create new guided jobs
3. In the voltage transformer view, enter the general asset data (see Table 5-7: "General asset data"
on page 52).
4. Enter the voltage transformer data (see 11.5 "Voltage transformer" on page 113).
OMICRON 65
CPC 100 PTM User Manual
3. In the rotating machine view, enter the general asset data (see Table 5-7: "General asset data" on
page 52).
4. In the Configuration area and the Ratings area, enter the rotating machine data (see 11.6 "Rotating
machine" on page 113).
66 OMICRON
Create new guided jobs
2. In the miscellaneous view, enter the general asset data (see Table 5-7: "General asset data" on
page 52).
OMICRON 67
CPC 100 PTM User Manual
68 OMICRON
Create new guided jobs
For example, if the CPC 100 is connected to your computer and you configured a job with tests for the
CPC 100 and the CIBANO 500 as shown in the following figure, the symbol left next the
CIBANO 500 tests indicate that you must connect the CIBANO 500.
Figure 5-24: Job configured with test for the CPC 100 and the CIBANO 500 test systems
OMICRON 69
CPC 100 PTM User Manual
If you, for example, try to perform a test for the CIBANO 500 with CPC 100 connected, a warning
message appears.
Figure 5-25: Warning message when trying to perform a test for the CIBANO 500
The optional manual tests are asset-independent. You can perform the tests for any asset described in
this User Manual, but Primary Test Manager does not guide you through the tests and provides no test
settings data. These tests offer a large amount of flexibility to define test procedures and specify test
settings according to your specific needs. For more information about the manual tests, see 7 "Create
new manual jobs" on page 91.
The Selected tests area displays the tests you want to perform. By default, Primary Test Manager
displays the tests recommended by OMICRON. To add a test into the Selected tests area, click the
symbol next to the test name in the Available tests area. To add all tests of a category into the
Selected tests area, click the symbol. The Selected tests area displays the test to be performed in
the recommended order. You can change the order of the tests by dragging them or by using the and
symbols. To remove a test from the Selected tests area, click the symbol next to the test name.
The General test settings area displays the reason of the job, the global test conditions, and some asset
specific data.
70 OMICRON
Create new guided jobs
Data Description
Reason for the job Reason why the job has been created
Global test conditions
Weather Weather conditions on site
Unit location Location of the unit under test
Humidity Relative ambient humidity on site
Ambient temperature Ambient temperature on site
Number of operations
Counter reading found Counter reading before starting the tests
Counter reading left Counter reading after the tests has finished
OMICRON 71
CPC 100 PTM User Manual
Data Description
Contact bounce filter
Main contact Threshold value of the time interval between two consequent bounces of
the main contact. For time intervals equal or below the threshold, the
contact is considered as closed.
Setting the value to 0.0 ms deactivates the contact bounce filter.
Auxiliary contact Threshold value of the time interval between two consequent bounces of
the auxiliary contact. For time intervals equal or below the threshold, the
contact is considered as closed.
Setting the value to 0.0 ms deactivates the contact bounce filter.
Update all tests Click Update all tests to set the contact bounce filter for all tests.
You can also import tests in JPG, PDF and any Microsoft Office file format.
To import test data:
1. In the Selected tests area, click Add from file.
2. In the Open dialog box, browse to the file you want the import.
72 OMICRON
Create new guided jobs
3. In the left pane of the test view, click the imported test.
4. In the workspace of the test view, you can change the test title and the test type.
5. To open the test, click the Open button under Test data.
Note: To open a test, you must have installed the associated application software on your computer.
6. You can attach files as described earlier in this chapter and add comments to the test.
For information about importing and exporting jobs, see 6.6 "Export and import jobs" on page 90.
OMICRON 73
CPC 100 PTM User Manual
3. In the Settings and conditions pane, enter the test settings (see the chapter on the asset tests later
in this manual).
74 OMICRON
Create new guided jobs
4. In the Assessment pane, enter the automatic assessment parameters and limits, if applicable.
OMICRON 75
CPC 100 PTM User Manual
5. Connect the test setup to the asset under test according to the wiring diagram displayed in the
General pane. For information about connecting the test setup to the asset under test, see the
technical documentation shipped with the CPC 100, CP TD, and CP SB1 test systems.
DANGER
Death or severe injury caused by high voltage or current
► Before connecting the test set to the test object, observe the five safety rules listed in 1 "Safety
instructions" on page 9.
► Observe the relevant laws and internal safety standards.
76 OMICRON
Create new guided jobs
DANGER
Death or severe injury caused by high voltage or current
The flashing lightning symbol in the Primary Test Manager test view indicates that an output of the
CPC 100 is active.
► Do not touch any of the CPC 100’s outputs or cables while the lightning symbol is displayed.
► If in doubt, press the Emergency Stop button.
OMICRON 77
CPC 100 PTM User Manual
7. Press the I/0 button on the front panel of the CPC 100.
After the measurement has been finished, Primary Test Manager displays the numeric measurement
data and the automatic assessment, if available, in the Measurements pane. To view graphical diagrams
of the measurement results, click the Plot tab.
8. Repeat steps 6 and 7 for all test measurements.
Note: Some tests support starting all measurements at once by clicking the Start all button.
Note: After the test has been performed some asset data relevant for the test configuration cannot be
edited anymore.
78 OMICRON
Create new guided jobs
OMICRON 79
CPC 100 PTM User Manual
To open a template:
1. In the Selected tests area of the test view, click Open template.
2. In the Choose Template dialog box, select the asset type, the number of phases and the template
you want to open.
Note: If you added a Microsoft Excel report template to the job template, its location is displayed
under Customized Excel report template.
3. Press Replace existing testplan to replace the tests already listed in the test list.
Press Append to existing testplan to add the tests listed in the template to the end of the existing
testplan.
Note: If you click Append to existing testplan, the Microsoft Excel report template will not be added
the selected job template to the currently opened job.
80 OMICRON
Create new guided jobs
In the chart, measurements performed with 10 kV at rated frequency are displayed as circles . All other
data are displayed as triangles .
Note: If several tests are performed on one day, the most recent test of that day is connected to the
curve in the Trending chart. The others are displayed in the same chart but are not connected.
Status Description
Fail The status was automatically set to Fail by Primary Test Manager.
Manual fail The status was manually set to Fail.
Investigate The status was automatically set to Investigate by Primary Test
Manager.
Manual investigate The status was manually set to Investigate.
Pass The status was automatically set to Pass by Primary Test Manager.
Manual pass The status was manually set to Pass.
Not assessed The measurement has not been assessed.
Not rated The status was automatically set to Not Rated by Primary Test
Manager.
OMICRON 81
CPC 100 PTM User Manual
1. In the home view (see 4 "Home view" on page 20), click the New guided job button .
2. In the Create new job view, configure a job as described earlier in this chapter.
3. In the test view (see 5.5 "Test view" on page 68), select the test you want to perform.
4. In the Measurements pane, click the + Select comparison button, if available.
5. In the Select a test window, select the test you want to compare with the current test.
82 OMICRON
Create new guided jobs
6. The Measurements pane displays the measurement results of the selected test.
OMICRON 83
CPC 100 PTM User Manual
9. Primary Test Manager displays the measurement results of both tests in real time.
84 OMICRON
Manage objects
6 Manage objects
In the manage view, you can manage locations, assets, jobs, and reports available in Primary Test
Manager. After you have opened a job, Primary Test Manager leads you through the guided test
workflow (see 5 "Create new guided jobs" on page 41).
► To open the manage view, click the Manage button in the home view.
Note: In this chapter, the locations, assets, jobs, and reports are collectively called objects.
Note: The mounted assets are displayed in italics. To hide them, expand the Search area under Asset,
and then select the Hide mounted assets check box.
The manage view displays the objects in a hierarchical structure as follows:
• If you select a location, the manage view displays the assets, jobs, and reports associated with the
selected location.
• If you select an asset, the manage view displays the jobs and reports associated with the selected
asset.
• If you select a job, the manage view displays the reports associated with the selected job.
You can sort the objects:
► To sort the objects alphabetically, click Alphabetically on the ribbon.
► To sort the objects in the chronological order, click Recently used on the ribbon.
► To sort the objects according to the job status, click Status on the ribbon.
OMICRON 85
CPC 100 PTM User Manual
Data Description
Name Name of the location
Address Address of the location
City City where the asset is located
State/Province State or province where the asset is located
Postal code Postal code of the location
Country Country where the asset is located
Data Description
Asset Asset under test
Asset type Type of the asset
Serial no. Serial number of the asset
86 OMICRON
Manage objects
Data Description
Manufacturer Manufacturer of the asset
Manufacturer type Type of the asset according to the manufacturer
Asset system code Code of the asset used by the maintenance planning systems
Apparatus ID Identifier of the asset
The following table describes the job search data.
Data Description
Name/WO Name of the job or work order
Tested by Person who performed the test
Executed between Time period between the job was executed
Status Status of the job
Data Description
Name Name of the report
Client Customer for which the report is designated
Created between Time period between the report was created
OMICRON 87
CPC 100 PTM User Manual
► To export multiple jobs, select the check boxes next to the jobs you want to export, and then click the
Export button .
88 OMICRON
Manage objects
3. In the Relocate Asset dialog box, select the location you want to move the asset to.
4. If the asset you want to relocate is mountable, select an asset where the moved asset is to be
mounted.
Note: You can filter the locations and assets by searching for keywords (see 6.1 "Search for objects" on
page 86).
OMICRON 89
CPC 100 PTM User Manual
90 OMICRON
Create new manual jobs
The workspace of the new manual job view depends on the selected button in the left pane (see
Figure 7-2: "Left-pane buttons" on page 92):
• Initially, the workspace is divided into the Available tests area and the Selected tests area.
• If you click the Asset button, Primary Test Manager displays the general asset data (see
5.4.1 "General asset data" on page 52).
• If you click the Tests button, Primary Test Manager displays the Available tests area and the
Selected tests area. In this workspace, you can add tests to a job (see 7.1 "Add tests to a job" on
page 92).
If you have added a test to a job and click the Report button, Primary Test Manager displays the report
view. In the report view, you can configure and generate test reports.
OMICRON 91
CPC 100 PTM User Manual
If you click the Tests button, the workspace displays the General area and the Tests area again.
Note: You can change the default test names. To rename a test, click the corresponding button in the
left pane, and then click the test name.
92 OMICRON
Create new manual jobs
For the description of the test settings and measurement data, see 19 "Optional manual tests" on
page 248).
Command Action
Home Closes the current job and leads you to the Primary Test Manager home
view. Before closing the job, you can save the changes for all tests.
Save job Saves the current job. When saving for the first time, you must specify the
filename and directory of the job.
Save as Saves the current job in a newly specified directory.
Export job Exports the job displayed in the new manual job view into a Microsoft
Excel spreadsheet.
Copy test1 Adds another test of the same kind and with the same settings to the test
list. Results are not copied.
Note: Clicking Home in the title bar and Close in the menu bar have the same functionality.
OMICRON 93
CPC 100 PTM User Manual
8 Open jobs
With Primary Test Manager, you can open existing guided and manual jobs. To open a job:
1. Click the Open job button in the home view (see Figure 4-1: "Primary Test Manager home view"
on page 20).
2. In the Open dialog box, browse to the file you want to open.
94 OMICRON
Open jobs
The open job view displays the tests in the left pane. To view the test results, click the corresponding
test button. You can add new tests to the job and process tests as described in 5 "Create new guided
jobs" on page 41 and 7 "Create new manual jobs" on page 91.
OMICRON 95
CPC 100 PTM User Manual
9 Control tests
In the Measurements area of Primary Test Manager, you can control the test execution. The following
table describes the available commands.
Command Action
Start Starts the selected measurement of the currently open test.
Start all Starts all measurements of the currently open test.
Stop Stops the running measurement.
Stop all Stops all running measurements.
Clear all Deletes all measurement results of the currently open test.
Clear result Deletes the results of the selected measurement.
Delete measurement Deletes the selected measurement row.
Add measurement Adds a new measurement row to the currently open test.
96 OMICRON
Generate test reports
The report view is divided into the New report area, the Settings area and the Existing reports area.
In the New report area, you can set the report data. The following table describes the report data.
Table 10-1: Report data
Data Description
Title Title of the report. Appears as the report header.
Report language Language the report is created in
OMICRON 97
CPC 100 PTM User Manual
98 OMICRON
Asset data
11 Asset data
This section describes the asset specific data.
11.1 Transformer
The following tables describe the transformer data.
Table 11-1: Winding configuration
Data Description
Phases Number of the transformer’s phases
Vector group Vector group of the transformer
(see "Setting the vector group" on page 55)
Unsupported vector group Vector group not supported by Primary Test Manager as text for
(for documentation) documentation
Data Description
Rated frequency Rated frequency of the transformer
Voltage ratings
Winding Transformer’s winding
Voltage L-L L-L voltage of the transformer’s winding
Voltage L-N L-N voltage of the transformer’s winding
Insul. level L-L (BIL) L-L basic impulse level rating of the transformer’s winding
Power ratings
Rated power Power rating of the transformer
Cooling class Cooling class of the transformer
Temp. rise wind. Temperature rise of the transformer’s winding
Current ratings at rated power
Short-circuit rating
Max. short-circuit current Maximum short-circuit current of the transformer in kA during a given time
in seconds
1. Set by the regional conventions (see 4.1.1 "Settings" on page 23).
OMICRON 99
CPC 100 PTM User Manual
Data Description
Ref. temp. Reference temperature
Data Description
Category Application category of the transformer
Status Usage status of the transformer
Tank type Type of the transformer’s tank
Insulation medium Insulation medium of the transformer
Weight Weight of the transformer’s insulation
Insulation
Volume Volume of the transformer’s insulation
Total weight Total weight of the transformer
Winding Transformer’s winding
Conductor material Conductor material of the transformer’s winding
11.1.1 Bushing
For the data of the transformer’s bushings, see 11.2 "Spare bushing" on page 102.
100 OMICRON
Asset data
Data Description
OLTC/DETC Select the OLTC check box to set the OLTC data.
Select the DETC check box to set the DETC data.
Tap changer configuration
Winding Transformer’s winding to which the tap changer is connected
Tap scheme Notation scheme for tap identification
No. of taps Number of the tap changer’s taps
Voltage table
Tap Number of the tap
Voltage Voltage on the tap
1. Only available for the DETC
Data Description
Ratings
Units in stack Number of the surge arrester’s units
Numerical positions Select the Numerical positions check box to set numerical positions of the
surge arrester.
Literal positions Select the Literal positions check box to set alphabetical positions of the surge
arrester.
Position Position of the surge arrester
Serial no. Serial number of the surge arrester
Voltage L-L Values needed to calculate the maximum test voltages
Voltage L-N
MCOV rating Maximum continuous operating voltage between the terminals of the surge
arrester
Unit catalog no. Identifier of the surge arrester unit
OMICRON 101
CPC 100 PTM User Manual
Data Description
102 OMICRON
Asset data
Data Description
Other
Insulation type Insulation type of the spare bushing
Outer insulation type Outer insulation type of the spare bushing
1. Only available for spare bushings mounted on another assets
2. Set by the regional conventions (see 4.1.1 "Settings" on page 23).
Data Description
Circuit breaker
Number of phases Number of the circuit breaker’s phases
Number of interrupters Number of the circuit breaker’s interrupters per phase
per phase
Pole operation Pole operation of the circuit breaker
Pre-insertion resistors Select the Pre-insertion resistors (PIR) check box if the circuit breaker
(PIR) contains pre-insertion resistors.
PIR value Pre-insertion resistor value
Grading capacitors Select the Grading capacitors check box if the circuit breaker contains
grading capacitors.
Capacitor value Grading capacitor value
Interrupting medium Interrupting medium of the circuit breaker
Tank type Type of the circuit breaker’s tank
Ratings
Rated frequency Rated frequency of the circuit breaker
Rated voltage L-L Rated voltage of the circuit breaker
Rated current Rated current of the circuit breaker
Rated short-circuit Rated short-circuit breaking current of the circuit breaker
breaking current
Short-circuit nominal Nominal duration of the short-circuiting
duration
Rated insulation level Basic impulse level rating of the circuit breaker
(BIL)
OMICRON 103
CPC 100 PTM User Manual
Data Description
Rated interrupting time Rated interrupting time of the circuit breaker
Interrupting duty cycle Interrupting duty cycle of the circuit breaker
Rated power at closing Rated power of the circuit breaker at closing
Rated power at opening Rated power of the circuit breaker at opening
Rated power at motor Rated power of the circuit breaker at motor charge
charge
Contact system
Total travel Total distance traveled by the contact during operation (excluding possible
overshoot)
Damping time Time in which the damping units are engaged to decelerate the circuit
breaker’s moving contacts
Nozzle length Length of the circuit breaker’s nozzle
Others
Total weight with oil/gas Total weight of the circuit breaker with oil or gas
Weight of oil/gas Weight of the circuit breaker’s oil or gas
Volume of oil/gas Volume of the circuit breaker’s oil or gas
Rated gas pressure Rated gas pressure of the circuit breaker at given temperature
Data Description
Number of trip coils Number of trip coils to operate the circuit breaker
Number of close coils Number of close coils to operate the circuit breaker
Component Operating mechanism’s component
Rated voltage Rated voltage of the operating mechanism’s component
Rated current Rated current of the operating mechanism’s component
DC Select the DC check box to set the component’s DC operation
AC Select the AC check box to set the component’s AC operation
Frequency AC operation frequency
Rated operating pressure Rated operating pressure of the operating mechanism at given
temperature
104 OMICRON
Asset data
11.3.2 Bushing
For the data of the circuit breaker’s bushings, see Table 11-7: "Spare bushing data" on page 102.
Setting Description
Contact resistance
R min Minimum limit of the contact resistance
R max Maximum limit of the contact resistance
Operating times
t min Minimum limit of the operating time
t max Maximum limit of the operating time
Contact travel1
d min Minimum assessment limit
d max Maximum assessment limit
Add velocity zone Click Add velocity zone to define a velocity zone for assessment of the
contact travel velocity.
Coil currents
I min Minimum limit of the coil current
I max Maximum limit of the coil current
Pickup voltage
V min Minimum limit of the pickup voltage
V max Maximum limit of the pickup voltage
1. Only available if testing with the CB TN3 modules
The following table describes the circuit breaker’s relative assessment limits.
Table 11-11: Relative assessment limits
Setting Description
Contact resistance
R ref Reference contact resistance
R dev Allowed deviation from the reference contact resistance
Operating times
t ref Reference operating time
OMICRON 105
CPC 100 PTM User Manual
Setting Description
–t dev Allowed negative deviation from the reference operating time
+t dev Allowed positive deviation from the reference operating time
Contact travel1
d ref Reference velocity
d dev Allowed deviation from the reference velocity
Add velocity zone Click Add velocity zone to define a velocity zone for assessment of the
contact travel velocity.
Coil currents
I ref Reference coil current
–I dev Allowed negative deviation from the reference coil current
+I dev Allowed positive deviation from the reference coil current
Pickup voltage
V ref Reference pickup voltage
V dev Allowed deviation from the reference pickup voltage
1. Only available if testing with the CB TN3 modules
106 OMICRON
Asset data
Data Description
Standard Standard used for testing the current transformer
Rated frequency Rated frequency of the current transformer
Rated Ipn : Isn1 Rated primary- to-secondary current ratio of the current transformer
Rated Ipr : Isr 2
Primary windings Number of primary windings on the current transformer (for primary
reconnection)
IEEE C57.13
System voltage System voltage of the current transformer
Rated insulation level Basic impulse level rating of the current transformer
(BIL)
Rating factor (RF) Continuous thermal current rating factor of the current transformer
IEC 60044/IEC 61869
Um (r.m.s.) Highest voltage for equipment
U withstand (r.m.s.) Rated power-frequency withstand voltage
U lightning (peak) Rated lightning impulse withstand voltage
Icth Rated continuous thermal current
Idyn (peak) Rated dynamic current
Ith (r.m.s.) Rated short time thermal current
Duration Duration of rated short time thermal current
1. According to IEC 60444
2. According to IEC 61869 and IEEE C57.13
OMICRON 107
CPC 100 PTM User Manual
Data Description
Cores Number of the current transformer’s cores
Name Inter-tap identification
Kx Dimensioning factor. x
108 OMICRON
Asset data
Table 11-14: Specific parameters and settings displayed for IEC 60044 protection CTs
Available for
IEC 60044 Available for
Param. Description protection CTs, IEC 60044, class
class
xP xPR PX TPS TPX TPY TPZ
K Dimensioning factor. x
The following parameters are only displayed if the standard IEC 60044 is selected with the type
"Metering CT".
Table 11-15: Specific parameters and settings displayed for IEC 60044 metering CTs
Available for
IEC 60044
metering CTs,
Param. Description class
0.1, 0.2, 0.2s, 0.5,
0.5s, 1, 3, 5
OMICRON 109
CPC 100 PTM User Manual
Kx Dimensioning factor. x x
110 OMICRON
Asset data
Table 11-16: Specific parameters and settings displayed for IEC 61869 protection CTs
Available for IEC 61869
protection CTs, class
Param. Description transient
protection
protection
xP xPR PX PXR TPX TPY TPZ
OMICRON 111
CPC 100 PTM User Manual
112 OMICRON
Asset data
Data Description
Standard Standard used for testing the voltage transformer
Rated frequency Rated frequency of the voltage transformer
Upr Rated primary voltage
Data Description
Windings Number of windings on the voltage transformer
Usr Rated secondary voltage
Rated burden Rated burden of the voltage transformer
cos φ Phase angle of the secondary burden
Data Description
Star point1 Star point connection of the rotating machine during test
Data Description
Rated frequency Rated frequency of the rotating machine
Rated voltage L-L L-L voltage of the rotating machine
Rated current Rated current of the rotating machine
Rated speed (rpm) Rated speed of the rotating machine in revolutions per minute
Rated power Rated power of the rotating machine
OMICRON 113
CPC 100 PTM User Manual
Data Description
Rated power factor Rated power factor of the rotating machine
Rated thermal class Rated thermal class of the rotating machine
Rated excitation current Rated excitation current of the rotor
(rotor)
Rated excitation voltage Rated excitation voltage of the rotor
(rotor)
Data Description
Maximum grid current Maximum current to be diverted to ground via the grounding system
Maximum fault duration Maximum amount of time until the fault is considered to be cleared
Unlimited fault duration Activate the check box for an unlimited fault duration. Maximum fault
duration will be deactivated.
To document your measurement locations, you can add a Grounding system environment (for ground
impedance measurements) and Ground mat plan (for step and touch voltage measurements).
► To add an existing map as an image, press Select image .
► To create a snapshot from Bing Maps, press Show map .
Zoom into the desired area and press Create snapshot.
114 OMICRON
Transformer tests
12 Transformer tests
In this chapter, the settings and measurement data of the transformer tests are described. For more
information about testing with the CPC 100, see the CPC 100 User Manual shipped with your test
system.
Note: Some test names depend on the Profile selected in the Settings (see "Profiles" on page 26). For
your convenience, you can use your preferred naming to, for example, match regional conventions:
• IEEE standard: Power factor (PF) for the loss indicator; H/X/Y for the windings
• IEC standard: Dissipation factor (DF) for the loss indicators; Prim/Sec/Tert for the windings,
• Custom profiles: Power factor (PF), Dissipation factor (DF) or Tangent delta (Tanδ) for the loss
indicator; various naming options for the windings
The dissipation factor and the tangent delta are identical characteristics of the primary asset under test.
1. Test name depends on the Profile selected in the Settings (see "Profiles" on page 26).
OMICRON 115
CPC 100 PTM User Manual
Primary Test Manager supports additionally the TanDelta – PF & CAP manual test.
► For more information, see 19.2 "Manual Tan Delta" on page 251.
116 OMICRON
Transformer tests
Setting Description
Asset
Asset Asset under test – set in the asset data
(see chapter 11 "Asset data" on page 99)
Tank type Type of transformer tank
Insulation medium Insulation medium of the transformer – set in the asset data
(see chapter 11 "Asset data" on page 99)
Note: The DGA is only valid for the insulation medium Mineral oil.
Oil type Type of transformer oil
Test conditions
Sample date Date of sample collection
Oil sample temperature Oil temperature at the time of sampling
Measurement
Analyzed by Information on how the sample was analyzed
• Oil lab: The sample was analyzed by a laboratory.
After selecting Oil lab, you can enter the Name and Address of the
laboratory.
• Mobile DGA: The sample was analyzed using a mobile DGA device.
After selecting Mobile DGA, you can enter the device
Manufacturer/Type and its Serial number.
• Online DGA: The sample was analyzed using a permanently
installed monitoring device.
After selecting Online DGA, you can enter the device
Manufacturer/Type and its Serial number.
Use C3 hydrocarbons ► Activate the Use C3 hydrocarbons check box to add C3H6 and
C3H8 to the list of Gas in oil values, and to activate ratio
assessment according to the MSS scheme.
OMICRON 117
CPC 100 PTM User Manual
Data Description
TDCG Total dissolved combustible gas
TDG Total dissolved gas
TCGe Estimation of the percentage of total combustible gas in the gas space.
It will only correspond to the actually measured value if there is a
balance between the gas blanket and the oil.
Lab. result Assessment result of the laboratory according to the IEEE or IEC
standard.
Assessment Manual Gas-in-oil analysis assessment:
• Manual pass
• Manual fail
• Manual investigate
• Not assessed
Assessment Summary
The results are assessed using the following interpretation methods:
• Duval’s triangles (see Table 12-3 below)
• IEC basic gas ratios
• Roger’s ratios
• Doernenburg’s ratios
• Key gases according to IEEE C57.104 and IEC 60599 (see Table 12-3 below)
• MSS scheme
118 OMICRON
Transformer tests
Condition 4
Condition 3
Condition 1
Indicator
Assumed fault –
see Assessment Summary
table
Duval’s triangle 1
Indicator
Indicator
OMICRON 119
CPC 100 PTM User Manual
Assessment Details
• The Table contains condition ranges and states for individual gases.
• The Ratio Table lists all used gas ratios, depending on the selected standard, and provides an
Interpretation of the recorded values.
Table 12-4: Oil Analysis – Assessment Details
Data Description
Table
Standard Standard used for the condition assessment
Overall assessment Condition fulfilled by the measured value of an individual gas
TDCG units/day Increase in TDCG per day since the last measurement
Recommendation Recommended interval for future measurements
Ratio Table
Sample Date Date of the sampling
Duval Triangle
Duval’s triangles visualize faults in a triangular coordinate system (see Table 12-3).
• Triangle 1: gases formed by faults of low to high energy
• Triangle 4: gases formed more specifically by faults of low energy or temperature
• Triangle 5: gases formed more specifically by faults of high temperature
Pattern
The key gas results are visually compared to four reference patterns. If a reference graph matches the
measured value, it is highlighted.
Data Description
Water content
H2O meas. Measured water content in oil
120 OMICRON
Transformer tests
Data Description
Field strength Field strength
Assessment DC conductivity assessment
Power factor1
Standard Standard underlying the power factor analysis
Meas. value @ 25 ºC Power factor measured at 25 ºC
Meas. value @ 100 ºC Power factor measured at 100 ºC
Assessment Power factor assessment
Dielectric breakdown voltage
Standard Standard underlying the dielectric breakdown voltage analysis
Meas. value Measured dielectric breakdown voltage
Test temperature Oil temperature during dielectric breakdown voltage test
Assessment Dielectric breakdown voltage assessment
Chemical
Interfacial tension Interfacial tension of the oil
Neutralization value Neutralization value of the oil
Particle count Particle count of the oil
Color Color of the oil
Assessment Chemical assessment
1. Term depends on the Profile selected in the Settings (see "Profiles" on page 26).
The following table describes the test status that can be set in the Oil Analysis test:
Table 12-6: Oil Analysis – Test status
Status Description
Partially executed At least one measurement of the test has been executed.
Executed All measurements of the test have been executed.
Note: The test status set in the Oil Analysis test is displayed in the job overview (see 5.2 "Job overview"
on page 44) under Tests. If you do not set the test status to Partially executed or Executed in the Oil
Analysis test, the test status Not executed is displayed in the job overview.
OMICRON 121
CPC 100 PTM User Manual
DANGER
Death or severe injury caused by high voltage or current
The high-voltage cable is double-shielded and therefore safe. However, the last 50 cm (20 inches)
of this cable have no shield.
► Avoid any direct contact of this part of the cable to ground potential or to any objects.
► During a test, consider this part of the cable a live wire and, due to the high-voltage, life-
threatening.
Note: This test name depends on the Profile selected in the Settings (see "Profiles" on page 26):
• IEEE standard: Overall PF & CAP
• IEC standard: Winding DF & CAP
• Custom Profile: for example Overall Tanδ & CAP or Winding Tanδ & CAP
In this section, the terms Power factor (PF) and Overall PF & CAP and will be used.
122 OMICRON
Transformer tests
The following table describes the Overall PF & CAP test settings.
Table 12-7: Overall PF & CAP test settings
Setting Description
Measurement settings
Test frequency ► Set the output frequency for the test.
Sweep settings
Frequency sweep Sweep profile:
None, OMICRON expertise (recommended), or CPC template
• None: no frequency sweep
• OMICRON expertise: sweep frequencies dynamically distributed
within the CPC 100 frequency range for optimum results
• CPC template: sweep frequencies specified by the CPC 100 test
templates
Voltage sweep (tip-up) Sweep profile:
None or OMICRON expertise
• None: no voltage sweep
• OMICRON expertise: sweep voltages dynamically distributed within
the asset-dependent voltage range for optimum results
► Click the pen button to create a frequency or voltage sweep
Sweep profiles
profile.
► Add up to 30 measurement points with individual output voltages or
frequencies. Double-click a value to change it.
► Mark a favorite to use it as the default sweep profile for future
tests.
Note: The predefined profiles None, OMICRON expertise and
CPC template cannot be edited or deleted.
The default sweep profiles for this test are:
• Frequency sweep: OMICRON expertise
• Voltage sweep: None
OMICRON 123
CPC 100 PTM User Manual
Setting Description
Device settings
Tan delta device ► Select your CP TD device:
• CP TD1
• CP TD12
• CP TD15
Enable shield check ► Select the check box if you want that the CPC 100 checks whether
the shield of the high-voltage cable is connected.
Use beeper ► Select the check box if you want to use the CP TD’s beeper during
the measurement.
Test conditions
Custom test conditions ► Select the check box to set test conditions differing from the global
test conditions.
Top oil temperature Oil temperature at the top of the transformer’s tank
Bottom oil temperature Oil temperature at the bottom of the transformer’s tank
Average oil temperature Calculated average oil temperature of the transformer’s tank
Winding temperature Temperature of the transformer’s winding
Ambient temperature Ambient temperature on site
Humidity Relative ambient humidity
Weather Weather during the test
Correction factors
Temperature correction ► Select the check box to activate temperature correction.
Correction temp. ► In the Custom list, click the correction temperature, or type the
correction temperature beneath.
Correction factor ► Click the Calculate button to calculate the correction factor
automatically or type the correction factor beneath.
Use reference voltage ► Select the check box to extrapolate the I out and Watt losses results
for the specified reference voltage.
124 OMICRON
Transformer tests
NOTICE
Equipment damage or loss of data possible
Before making a decision based on the Primary Test Manager automatic assessment, read the
disclaimer 1.3.1 "Automatic assessment" on page 10.
Primary Test Manager supports automatic assessment for the following insulation media if you activate
the temperature correction by selecting the Temperature correction check box:
• Natural ester
• Mineral oil
• Silicone
The following table describes the automatic assessment parameters of the Overall PF & CAP test.
Table 12-8: Overall PF & CAP automatic assessment parameters
Parameter Description
Limit schema Limit schema underlying the assessment
Global assessment criteria
Min. Iout @10 kV Threshold value of the automatic assessment.
If Iout < Min. Iout @10 kV, the automatic assessment status is
Investigate.
To save the selected limit schema as the default setting for all future jobs, click Set as default.
Under Visible limits:
► Click only limits that are used for this measurement to display only limits for the selected
transformer’s insulation type.
► Click all limits to display limits for all supported transformer’s insulation types.
The following tables describe the automatic assessment limits of the Overall PF & CAP test.
Table 12-9: Overall PF & CAP automatic assessment limits based on Power factor1
OMICRON 125
CPC 100 PTM User Manual
If you selected the Use reference voltage check box Primary Test Manager performs additionally a
cross check by using the following additional limits.
Table 12-10: Overall PF & CAP cross check limits based on Power Factor1
Table 12-11: Overall PF & CAP cross check limits based on capacitance
126 OMICRON
Transformer tests
The following table describes the Overall PF & CAP measurement data.
Table 12-12: Overall PF & CAP measurement data
Data Description
No. Number of the measurement
Measurement Arrangement of the measurement
Test mode Test mode according to the IEEE Std 62-1995
Sweep Swept variable: frequency, voltage, or none
V test Test voltage
Freq. Test frequency
V out Measured output voltage
I out Measured output current
Watt losses Measured losses
OMICRON 127
CPC 100 PTM User Manual
DANGER
Death or severe injury caused by high voltage or current
The high-voltage cable is double-shielded and therefore safe. However, the last 50 cm (20 inches)
of this cable have no shield.
► Avoid any direct contact of this part of the cable to ground potential or to any objects.
► During a test, consider this part of the cable a live wire and, due to the high-voltage, life-
threatening.
Note: This test name depends on the Profile selected in the Settings (see "Profiles" on page 26):
• IEEE standard: Bushing H/X PF & CAP – C1
• IEC standard: Bushing Prim/Sec DF & CAP – C1
• Custom Profile: for example Bushing H/X Tanδ & CAP – C1 or Bushing Prim/Sec Tanδ & CAP –
C1
In this section, the terms Power factor (PF) and Bushing PF & CAP – C1 will be used.
128 OMICRON
Transformer tests
The following table describes the Bushing PF & CAP – C1 test settings.
Table 12-13: Bushing PF & CAP – C1 test settings
Setting Description
Measurement settings
Test frequency ► Set the output frequency for the test.
Sweep settings
Frequency sweep Sweep profile:
None, OMICRON expertise (recommended), or CPC template
• None: no frequency sweep
• OMICRON expertise: sweep frequencies dynamically distributed
within the CPC 100 frequency range for optimum results
• CPC template: sweep frequencies specified by the CPC 100 test
templates
Voltage sweep (tip-up) Sweep profile:
None or OMICRON expertise
• None: no voltage sweep
• OMICRON expertise: sweep voltages dynamically distributed
within the asset-dependent voltage range for optimum results
► Click the pen button to create a frequency or voltage sweep
Sweep profiles
profile.
► Add up to 30 measurement points with individual output voltages or
frequencies. Double-click a value to change it.
► Mark a favorite to use it as the default sweep profile for future
tests.
Note: The predefined profiles None, OMICRON expertise and
CPC template cannot be edited or deleted.
The default sweep profiles for this test are:
• Frequency sweep: OMICRON expertise
• Voltage sweep: OMICRON expertise
OMICRON 129
CPC 100 PTM User Manual
Setting Description
Device settings
Tan delta device ► Select your CP TD device:
• CP TD1
• CP TD12
• CP TD15
Enable shield check ► Select the check box if you want that the CPC 100 checks whether
the shield of the high-voltage cable is connected.
Use beeper ► Select the check box if you want to use the CP TD’s beeper during
the measurement.
Test conditions
Custom test conditions ► Select the check box to set test conditions differing from the global
test conditions.
Top oil temperature Oil temperature at the top of the transformer’s tank
Ambient temperature Ambient temperature on site
Humidity Relative ambient humidity
Weather Weather during the test
Correction factors
Temperature correction ► Select the check box to activate temperature correction.
NOTICE
Equipment damage or loss of data possible
Before making a decision based on the Primary Test Manager automatic assessment, read the
disclaimer 1.3.1 "Automatic assessment" on page 10.
Primary Test Manager supports automatic assessment for the following bushing’s insulation types:
• Oil-impregnated paper
• Resin-bonded paper
• Resin-impregnated paper
For another type of the bushing’s insulation the automatic assessment cannot be done.
Note: The assessment can be done only if all bushings have the same insulation type.
130 OMICRON
Transformer tests
The following table describes the automatic assessment parameters of the Bushing PF & CAP – C1 test.
Table 12-14: Bushing PF & CAP – C1 automatic assessment parameters
Parameter Description
Limit schema Limit schema underlying the assessment
Global assessment criteria
Min. Iout @10 kV Threshold value of the automatic assessment.
If Iout < Min. Iout @10 kV, the automatic assessment status is
Investigate.
To save the selected limit schema as the default setting for all future jobs, click Set as default.
Under Visible limits:
► Click only limits that are used for this measurement to display only limits for the selected
bushing’s insulation type and the PF (C1)1 nameplate values (if entered) or the absolute limits.
► Click all limits to display limits for all supported bushing insulation types and the PF (C1)1 nameplate
values (if entered) or the absolute limits.
The following tables describe the automatic assessment limits of the Bushing PF & CAP – C1 test.
Table 12-15: Bushing PF & CAP – C1 automatic assessment limits based on Power Factor1
Table 12-16: Bushing PF & CAP – C1 automatic assessment limits based on capacitance
1. Term depends on the Profile selected in the Settings (see "Profiles" on page 26).
OMICRON 131
CPC 100 PTM User Manual
The following table shows the Primary Test Manager assessment logic.
Table 12-17: Assessment logic
Data Description
No. Number of the measurement
Measurement Terminal name of the bushing under test
Test mode Test mode according to the IEEE Std 62-1995
Sweep Swept variable: frequency, voltage, or none
V test Test voltage
132 OMICRON
Transformer tests
Data Description
Freq. Test frequency
V out Measured output voltage
I out Measured output current
Watt losses Measured losses
DANGER
Death or severe injury caused by high voltage or current
The high-voltage cable is double-shielded and therefore safe. However, the last 50 cm (20 inches)
of this cable have no shield.
► Avoid any direct contact of this part of the cable to ground potential or to any objects.
► During a test, consider this part of the cable a live wire and, due to the high-voltage, life-
threatening.
Note: This test name depends on the Profile selected in the Settings (see "Profiles" on page 26):
• IEEE standard: Bushing H/X PF & CAP – C2
• IEC standard: Bushing Prim/Sec DF & CAP – C2
• Custom Profile: for example Bushing H/X Tanδ & CAP – C2 or Bushing Prim/Sec Tanδ & CAP –
C2
In this section, the terms Power factor (PF) and Bushing PF & CAP – C2 will be used.
OMICRON 133
CPC 100 PTM User Manual
The following table describes the Bushing PF & CAP – C2 test settings.
Table 12-19: Bushing PF & CAP – C2 test settings
Setting Description
Measurement settings
Test frequency ► Set the output frequency for the test.
Enable shield check ► Select the check box if you want that the CPC 100 checks whether
the shield of the high-voltage cable is connected.
Use beeper ► Select the check box if you want to use the CP TD’s beeper during
the measurement.
Test conditions
Custom test conditions ► Select the check box to set test conditions differing from the global
test conditions.
Top oil temperature Oil temperature at the top of the transformer’s tank
Ambient temperature Ambient temperature on site
Humidity Relative ambient humidity
Weather Weather during the test
134 OMICRON
Transformer tests
Setting Description
Correction factors
Temperature correction ► Select the check box to activate temperature correction.
NOTICE
Equipment damage or loss of data possible
Before making a decision based on the Primary Test Manager automatic assessment, read the
disclaimer 1.3.1 "Automatic assessment" on page 10.
Primary Test Manager supports automatic assessment for the following bushing’s insulation types:
• Oil-impregnated paper
• Resin-bonded paper
• Resin-impregnated paper
For another type of the bushing’s insulation the automatic assessment cannot be done.
Note: The assessment can be done only if all bushings have the same insulation type.
The following table describes the automatic assessment parameters of the Bushing PF & CAP – C2 test.
Table 12-20: Bushing PF & CAP – C2 automatic assessment parameters
Parameter Description
Limit schema Limit schema underlying the assessment
Global assessment criteria
Min. Iout @10 kV Threshold value of the automatic assessment.
If Iout < Min. Iout @10 kV, the automatic assessment status is
Investigate.
To save the selected limit schema as the default setting for all future jobs, click Set as default.
Under Visible limits:
► Click only limits that are used for this measurement to display only limits for the selected
bushing’s insulation type and the PF (C2)1 nameplate values (if entered) or the absolute limits.
► Click all limits to display limits for all supported bushing insulation types and the PF (C2)1 nameplate
values (if entered) or the absolute limits.
1. Term depends on the Profile selected in the Settings (see "Profiles" on page 26).
OMICRON 135
CPC 100 PTM User Manual
The following tables describe the automatic assessment limits of the Bushing PF & CAP – C2 test.
Table 12-21: Bushing PF & CAP – C2 automatic assessment limits based on Power Factor1
Table 12-22: Bushing PF & CAP – C2 automatic assessment limits based on capacitance
The following table shows the Primary Test Manager assessment logic.
Table 12-23: Assessment logic
136 OMICRON
Transformer tests
• During measurement
To assess the test while the measurements are running, select the Assess during measurements
check box.
• Manually after all measurements have been finished
To assess the test manually, click Assess measurements.
After a measurement has been assessed, the test settings and the assessment parameters are locked.
If you change nameplate data relevant for the assessment (i.e. insulation type, PF or CAP) after a
measurement has been assessed and you reopen the test the following message appears:
Data Description
No. Number of the measurement
Measurement Terminal name of the bushing under test
Test mode Test mode according to the IEEE Std 62-1995
V test Test voltage
Freq. Test frequency
V out Measured output voltage
I out Measured output current
Watt losses Measured losses
OMICRON 137
CPC 100 PTM User Manual
DANGER
Death or severe injury caused by high voltage or current
The high-voltage cable is double-shielded and therefore safe. However, the last 50 cm (20 inches)
of this cable have no shield.
► Avoid any direct contact of this part of the cable to ground potential or to any objects.
► During a test, consider this part of the cable a live wire and, due to the high-voltage, life-
threatening.
Note: This test name depends on the Profile selected in the Settings (see "Profiles" on page 26):
• IEEE standard: Bushing H/X – Energized Collar
• IEC standard: Bushing Prim/Sec – Energized Collar
In this section, the term Bushing – Energized Collar will be used.
The following table describes the Bushing – Energized Collar test settings.
Table 12-25: Bushing – Energized Collar test settings
Setting Description
Measurement settings
Test frequency ► Set the output frequency for the test.
138 OMICRON
Transformer tests
Setting Description
Device settings
Tan delta device ► Select your CP TD device:
• CP TD1
• CP TD12
• CP TD15
Enable shield check ► Select the check box if you want that the CPC 100 checks whether
the shield of the high-voltage cable is connected.
Use beeper ► Select the check box if you want to use the CP TD’s beeper during
the measurement.
Test conditions
Custom test conditions ► Select the Custom test conditions check box to set test conditions
differing from the global test conditions.
Top oil temperature Oil temperature at the top of the transformer’s tank
Ambient temperature Ambient temperature on site
Humidity Relative ambient humidity
Weather Weather during the test
Correction factors
Use reference voltage ► Select the check box to extrapolate the I out and Watt losses results
for the specified reference voltage.
Data Description
No. Number of the measurement
Measurement Terminal name of the bushing under test
Test mode Test mode according to the IEEE Std 62-1995
V test Test voltage
Freq. Test frequency
OMICRON 139
CPC 100 PTM User Manual
Data Description
V out Measured output voltage
I out Measured output current
Watt losses Measured losses
Assessment Measurement assessment
DANGER
Death or severe injury caused by high voltage or current
The high-voltage cable is double-shielded and therefore safe. However, the last 50 cm (20 inches)
of this cable have no shield.
► Avoid any direct contact of this part of the cable to ground potential or to any objects.
► During a test, consider this part of the cable a live wire and, due to the high-voltage, life-
threatening.
The following table describes the Exciting Current test settings.
Table 12-27: Exciting Current test settings
Setting Description
Tap changer settings
Type of tap changer Tap changer (OLTC/DETC) whose taps are changed during the test
Number of taps Number of taps of the tap changer
Start tap Start tap position of the test
Stop tap Stop tap position of the test
DETC position/ Tap position of the DETC or OLTC
OLTC position1
Measurement settings
Test voltage Test voltage
Test mode Test mode according to the IEEE Std 62-1995
Measurement Phase of the winding under test
140 OMICRON
Transformer tests
Setting Description
Noise suppression settings
Averaging (no. points) Number of averaged measurements
Device settings
Tan delta device ► Select your CP TD device:
• CP TD1
• CP TD12
• CP TD15
Enable shield check ► Select the Enable shield check check box if you want that the
CPC 100 checks whether the shield of the high-voltage cable is
connected.
Use beeper ► Select the Use beeper check box if you want to use the CP TD’s
beeper during the measurement.
Correction factors
Use reference voltage ► Select the Use reference voltage check box to extrapolate the I out
and Watt losses results for the specified reference voltage.
Data Description
Tap Tap changer position
V out Measured output voltage
I out Measured output current
I phase Measured primary current per phase
Watt losses Measured losses
Reactance Measured reactance
Assessment Measurement assessment per tap position
OMICRON 141
CPC 100 PTM User Manual
DANGER
Death or severe injury caused by high voltage or current
The high-voltage cable is double-shielded and therefore safe. However, the last 50 cm (20 inches)
of this cable have no shield.
► Avoid any direct contact of this part of the cable to ground potential or to any objects.
► During a test, consider this part of the cable a live wire and, due to the high-voltage, life-
threatening.
Note: This test name depends on the Profile selected in the Settings (see "Profiles" on page 26):
• IEEE standard: Insulating Fluids PF & CAP
• IEC standard: Insulating Fluids DF & CAP
• Custom Profile: for example Insulating Fluids Tanδ & CAP
In this section, the terms Power factor (PF) and Insulating Fluids PF & CAP will be used.
The following table describes the Insulating Fluids PF & CAP test settings.
Table 12-29: Insulating Fluids PF & CAP test settings
Setting Description
Measurement settings
Test frequency ► Set the output frequency for the test.
142 OMICRON
Transformer tests
Setting Description
Device settings
Tan delta device ► Select your CP TD device:
• CP TD1
• CP TD12
• CP TD15
Enable shield check ► Select the check box if you want that the CPC 100 checks whether
the shield of the high-voltage cable is connected.
Use beeper ► Select the check box if you want to use the CP TD’s beeper during
the measurement.
Test conditions
Oil temperature Temperature of the oil
Custom test conditions ► Select the check box to set test conditions differing from the global
test conditions.
Data Description
No. Number of the measurement
Specimen Oil sample under test
Test mode Test mode according to the IEEE Std 62-1995
V test Test voltage
Freq. Test frequency
V out Measured output voltage
I out Measured output current
Watt losses Measured losses
OMICRON 143
CPC 100 PTM User Manual
Data Description
144 OMICRON
Transformer tests
DANGER
Death or severe injury caused by high voltage or current
The high-voltage cable is double-shielded and therefore safe. However, the last 50 cm (20 inches)
of this cable have no shield.
► Avoid any direct contact of this part of the cable to ground potential or to any objects.
► During a test, consider this part of the cable a live wire and, due to the high-voltage, life-
threatening.
Note: This test name depends on the Profile selected in the Settings (see "Profiles" on page 26):
• IEEE standard: Surge Arrester Watt Losses H/X
• IEC standard: Surge Arrester Watt Losses Prim/Sec
In this section, the term Surge Arrester Watt Losses will be used.
The following table describes the Surge Arrester Watt Losses test settings.
Table 12-31: Surge Arrester Watt Losses test settings
Setting Description
Measurement settings
Test frequency ► Set the output frequency for the test.
OMICRON 145
CPC 100 PTM User Manual
Setting Description
Device settings
Tan delta device ► Select your CP TD device:
• CP TD1
• CP TD12
• CP TD15
Enable shield check ► Select the check box if you want that the CPC 100 checks whether
the shield of the high-voltage cable is connected.
Use beeper ► Select the check box if you want to use the CP TD’s beeper during
the measurement.
Test Conditions
Custom test conditions ► Select the Custom test conditions check box to set test conditions
differing from the global test conditions.
Top oil temperature Oil temperature at the top of the transformer’s tank
Ambient temperature Ambient temperature on site
Humidity Relative ambient humidity
Weather Weather during the test
Correction factors
Use reference voltage ► Select the Use reference voltage check box to extrapolate the I out
and Watt losses results for the specified reference voltage.
Data Description
Measurement Terminal name of the surge arrester under test
Position Surge arrester unit under test
Test mode Test mode according to the IEEE Std 62-1995
V test Test voltage
Freq. Test frequency
146 OMICRON
Transformer tests
Data Description
V out Measured output voltage
I out Measured output current
Watt losses Measured losses
Assessment Measurement assessment
Setting Description
Measurement settings
Test current Test current
OLTC position Tap position of the OLTC
DETC position Tap position of the DETC
3Ph equiv. test
Perform test ► Select the Perform test check box if you want to perform the
3Ph equiv. test.
Show FRSL results ► Select the Show FRSL results check box to show the FRSL results
of the 3Ph equiv. test.
Per-phase test
Perform test ► Select the Perform test check box if you want to perform the Per-
phase test.
Show FRSL results ► Select the Show FRSL results check box to show the FRSL results
of the Per-phase test.
Test conditions
Custom test conditions ► Activate the Custom test conditions check box to set test
conditions differing from the global test conditions.
OMICRON 147
CPC 100 PTM User Manual
Setting Description
Winding temperature Temperature of the transformer’s winding
Temperature correction ► Select the Temperature correction check box to activate
temperature correction.
Single-phase transformers
The following table describes the automatic assessment limits for the single-phase transformer.
Table 12-34: Leakage Reactance automatic assessment limits for single-phase transformer
148 OMICRON
Transformer tests
1. Term depends on the Profile selected in the Settings (see "Profiles" on page 26).
During the test, Primary Test Manager calculates the Zk meas (%) value1. For the automatic
assessment, Primary Test Manager compares Zk meas (%) with the nameplate value Zk leakage
reactance (%). If the deviation is less than the default limit, the assessment is set to Pass, otherwise to
Fail.
Three-phase transformers
The following table describes the automatic assessment limits for the three-phase transformer.
Table 12-35: Leakage Reactance automatic assessment limits for three-phase transformer
3Ph equiv. Zk (%)1 relative to nameplate Low limit (fail) Default limit
Per-phase equiv. Zk (%)1 relative to average of all Low limit (fail) Default limit
phases
1. Term depends on the Profile selected in the Settings (see "Profiles" on page 26).
During the 3Ph equiv. measurement, Primary Test Manager calculates one Zk meas (%) value for all
phases. For the automatic assessment, Primary Test Manager compares Zk meas (%) with the
nameplate value Zk ref (%). If the deviation is less than the default limit the assessment status is set to
Pass, otherwise to Fail.
During the Per-phase measurement, Primary Test Manager calculates the Zk meas (%) values for each
phase. For the automatic assessment, Primary Test Manager compares Zk meas (%) with the average
of Zk meas (%) across all phases. If the deviation is less than the default limit, the assessment status is
set to Pass, otherwise to Fail.
You can assess the test at two points in the process:
• Assessment during the measurement
To assess the test while the measurements are running, activate the
Assess during measurements check box. After this, the test is locked, and you can only change
assessment-relevant data, for example, the nominal values.
• Manual assessment after all measurements have been finished
To assess the test manually, click Assess measurements. After that, the test is locked, and you can
only change assessment-relevant data, for example, the nominal values.
To set new data in the test or to change the test settings, click Clear all assessments.
You can overwrite the Primary Test Manager automatic assessment manually. To do so, click the arrow
next to the automatic assessment, and then select a manual assessment from the list.
Note: The automatic assessment is stored in the Assessment box for documentation purposes.
1. In the Transformer area of the Settings dialog box, you can change the Short-circuit impedance (Leakage reactance)
abbreviation (see "Profiles: Transformer" on page 27).
OMICRON 149
CPC 100 PTM User Manual
Data Description
Phase Phase under test
I AC Measured current
V1 AC Measured voltage
V1 AC phase Phase angle of the voltage V1 AC
Watt losses Measured losses
Zk Measured leakage reactance
Rk Real part of the measured Zk
Xk Imaginary part of the measured Zk
(leakage reactance)
Lk Measured leakage inductance
Data Description
Phase Phase under test
Zk dev (%)1 Deviation of the Zk meas (%) from the Zk ref (%) in percent
Data Description
Rk avg Average value of the per phase Rk
150 OMICRON
Transformer tests
Data Description
Rk dev (%) phase A Deviation of the measured Rk of the phase A from the average value
Rk dev (%) phase B Deviation of the measured Rk of the phase B from the average value
Rk dev (%) phase C Deviation of the measured Rk of the phase C from the average value
Assessment Measurement assessment
Measurements: per-phase
The following table describes the leakage reactance results (Zk).
Table 12-39: Leakage Reactance results (Zk)
Data Description
Phase Phase under test
I AC Measured current
V1 AC Measured voltage
V1 AC phase Phase angle of the voltage V1 AC
Watt losses Measured losses
Zk Measured leakage reactance
Rk Real part of the measured Zk
Xk Imaginary part of the measured Zk
(leakage reactance)
Lk Measured leakage inductance
Data Description
Phase Phase under test
Zk dev (%)1 Deviation of the measured impedance from the reference impedance in
percent
Xk meas (%) Measured reactance in percent
OMICRON 151
CPC 100 PTM User Manual
Data Description
Xk ref (%) Reference reactance in percent
Xk dev (%) Deviation of the Xk meas (%) from the Xk ref (%) in percent
Dominance order Order of the Zk dev (%) values:
1 for the highest value
2 for the middle value
3 for the lowest value
Assessment Measurement assessment
1. Term depends on the Profile selected in the Settings (see "Profiles" on page 26).
Data Description
Rk avg Average value of the per phase Rk
Rk dev (%) phase A Deviation of the measured Rk of the phase A from the average value
Rk dev (%) phase B Deviation of the measured Rk of the phase B from the average value
Rk dev (%) phase C Deviation of the measured Rk of the phase C from the average value
Assessment Measurement assessment
152 OMICRON
Transformer tests
The following table describes the Zero Sequence Impedance test settings.
Setting Description
Measurement settings
Test current Current of the test
Test conditions
Custom test conditions ► Activate the Custom test conditions check box to set test
conditions differing from the global test conditions.
NOTICE
Equipment damage or loss of data possible
Before making a decision based on the Primary Test Manager automatic assessment, read the
disclaimer 1.3.1 "Automatic assessment" on page 10.
OMICRON 153
CPC 100 PTM User Manual
2. In the Edit automatic assessment limits dialog box, enter the deviation of the Z0 (%) from the
Z0 ref (%) in percent.
The following table describes the Zero Sequence Impedance measurement data.
Table 12-43: Zero Sequence Impedance Measurement Data
Data Description
Tap Tap changer position
V1 AC Measured voltage
V1 AC phase Phase angle of the voltage V1 AC
I AC Measured current
Z meas Measured impedance
Z corr Temperature corrected impedance
Z0 Measured zero sequence impedance
Z0 (%) Measured zero sequence impedance in percent
Z0 ref (%) Reference zero sequence impedance in percent
Z0 dev (%) Deviation of the Z0 (%) from the Z0 ref (%) in percent
Assessment Measurement assessment
154 OMICRON
Transformer tests
Note: This test name depends on the Profile selected in the Settings (see "Profiles" on page 26):
• IEEE standard: TTR H–X
• IEC standard: Turns Ratio Prim–Sec
In this section, the term TTR will be used.
You can perform the TTR test also for the vector groups not supported by Primary Test Manager
according to the IEC 61378-1 standard. To perform the test according to the IEC 61378-1 standard,
select the IEC 61378-1 check box in the Settings and conditions pane.
The following table describes the TTR test settings.
Table 12-44: TTR test settings
Setting Description
Tap changer settings
Type of tap changer Tap changer (OLTC/DETC) whose taps are changed during the test
Number of taps Number of taps of the tap changer
Start tap Start tap position of the test
Stop tap Stop tap position of the test
DETC position/ Tap position of the DETC or OLTC
OLTC position1
CP SB1 settings
Use CP SB1 ► Select the CP SB1 check box to perform the test by using the
CP SB1 switch box.
OMICRON 155
CPC 100 PTM User Manual
Setting Description
Measurement settings
Test voltage Test voltage
Measurement Phase of the winding under test
Ratio Choose between TTR (transformer turns ratio) and VTR (voltage ratio)
to be displayed in the Measurements table.
Assessment settings
Manual ► Click Manual to select the manual assessment settings.
2. In the Edit automatic assessment limits dialog box, enter the deviation of the measured
transformer ratio from the nominal transformer ratio in percent.
156 OMICRON
Transformer tests
Data Description
Tap Tap changer position
Nominal ratio Nominal transformer ratio
V prim Measured primary voltage
OMICRON 157
CPC 100 PTM User Manual
DANGER
Death or severe injury caused by high voltage or current
Injecting direct current into test objects with inductive characteristics will charge the winding of the
test object.
► Follow instructions below.
► Refer to the Safety instructions in the CPC 100 User and Reference manuals.
DANGER
Death or severe injury caused by high voltage or current
► Never open the measuring circuit while current flows.
► After a measurement, wait until the test device has discharged completely.
► Ground all terminals of the test object before touching the test setup.
► Short-circuit the terminals before disconnecting the test leads.
► Disconnect cables not used for testing both from the test object and the test device.
158 OMICRON
Transformer tests
Setting Description
Tap changer settings
Type of tap changer Tap changer (OLTC/DETC) whose taps are changed during the test
Number of taps Number of taps of the tap changer
Start tap Start tap position of the test
Stop tap Stop tap position of the test
DETC Position/ Tap position of the DETC or OLTC
OLTC Position1
CP SB1 settings
Use CP SB1 ► Select the CP SB1 check box to perform the test by using the
CP SB1 switch box.
Measurement settings
CPC output DC current output of the CPC 100
Test current Current of the test
Measurement Phase of the winding under test
Automatic result ► Select the Automatic result check box to stop the measurement
automatically depending on the tolerance R dev and the settling
time.
OMICRON 159
CPC 100 PTM User Manual
Setting Description
Settling time (∆t) Time within which the deviation of consecutive measurement results is
evaluated. If the deviation is less than the tolerance R dev, the
measurement stops.
Show TRTapCheck results ► Select the Show TRTapCheck results check box to display
information about the condition (ripple and slope) of the tap changer
mechanisms
Test conditions
Custom test conditions ► Activate the Custom test conditions check box to set test
conditions differing from the global test conditions.
NOTICE
Equipment damage or loss of data possible
Before making a decision based on the Primary Test Manager automatic assessment, read the
disclaimer 1.3.1 "Automatic assessment" on page 10.
Primary Test Manager supports automatic assessment of the DC Winding Resistance test.
The following table describes the automatic assessment parameters of the DC Winding Resistance test.
Table 12-47: DC Winding Resistance automatic assessment parameters
Parameter Description
Limit schema Limit schema underlying the assessment
To save the selected limit schema as the default setting for all future jobs, click Set as default.
The following table describes the automatic assessment limits of the DC Winding Resistance test.
Table 12-48: DC Winding Resistance automatic assessment limits
During the automatic assessment, Primary Test Manager compares the R corr values of all phases to
each another. If a deviation is greater than the default limit the assessment status is set to Fail.
You can assess the test:
160 OMICRON
Transformer tests
• During measurement
To assess the test while the measurements are running, select the Assess during measurements
check box.
• Manually after all measurements have been finished
To assess the test manually, click Assess measurements.
After a measurement has been assessed, the test settings and the assessment parameters are locked.
To change the test data, the test settings or the assessment parameters, click Clear all assessments.
You can overwrite the Primary Test Manager automatic assessment manually. To do so, click the arrow
next to the automatic assessment, and then select a manual assessment from the list.
Note: The automatic assessment is stored in the Assessment box for documentation purposes.
The following table describes the DC Winding Resistance measurement data.
Table 12-49: DC Winding Resistance measurement data
Data Description
Tap Tap changer position
I DC Measured current
V DC Measured voltage
Time Time between the start and stop of a measurement
R meas Measured resistance
R dev Deviation of two successive measurement results at the time the test was
stopped
R corr Corrected measured resistance
Slope1 Highest measured falling edge slope of the test current during the tap
changer operation
Assessment Measurement assessment per tap position
1. Only available if an OLTC is specified and the Show TRTapCheck results check box is selected
Note: If you did not select the Automatic result check box to stop the measurement automatically, you
can stop the measurement by clicking the Keep results button.
OMICRON 161
CPC 100 PTM User Manual
Use the Dynamic OLTC-Scan (DRM) test card to visualize the on-load tap changer’s transient switching
behavior and assess its condition.
A Dynamic OLTC-Scan (DRM) can only be performed on resistive OLTCs and with the following settings
adjusted:
► Define the transformer’s winding configuration in the Transformer tab of the Asset view
(see "Setting the vector group" on page 55).
► Enter the No. of taps in the Tap changers tab of the Asset view. The No. of taps must be greater
than 1 (see 11.1.2 "Tap changer" on page 101).
DANGER
Death or severe injury caused by high voltage or current
Injecting direct current into test objects with inductive characteristics will charge the winding of the
test object.
► Follow instructions below.
► Refer to the Safety instructions in the CPC 100 User and Reference manuals.
DANGER
Death or severe injury caused by high voltage or current
► Never open the measuring circuit while current flows.
► After a measurement, wait until the test device has discharged completely.
► Ground all terminals of the test object before touching the test setup.
► Short-circuit the terminals before disconnecting the test leads.
► Disconnect cables not used for testing both from the test object and the test device.
Setting Description
Tap changer settings
Number of taps Number of taps of the tap changer
Start tap Start tap position of the test
Stop tap Stop tap position of the test
162 OMICRON
Transformer tests
Setting Description
CP SB1 settings
Mode Switching mode for taps
Manual switching of taps Taps are switched manually.
Auto switching of taps Taps are switched automatically.
Tap time Time for the automatic tap change between two positions
Impulse time Duration of the impulse for the automatic tap change
Measurement settings
CPC output DC current output of the CPC 100
Test current Current of the test
Recording time The time of the switching cycle that shall be recorded
Tolerance R dev Tolerance of the deviation of consecutive measurement results within
the settling time
Settling time (∆t) Time within which the deviation of consecutive measurement results is
evaluated. If the deviation is less than the tolerance R dev, the
measurement stops.
Noise suppression ► Select the Noise suppression check box to activate an interference
filter.
Measurements pane
The Measurements pane in the workspace displays the measurement results in table and graph format.
► Highlight a line to indicate the corresponding graph and vice versa.
► Use the Assessment drop-down box to manually assess the test results.
► To view graphical diagrams of the measurement results, click the Plot tab.
Filters
The following table describes the filters available for the Dynamic OLTC-Scan (DRM):
Table 12-51: Available filters for the Dynamic OLTC-Scan (DRM)
Filter Description
Phase Only results from the currently selected phase are displayed.
Parity Only results from taps with even/odd numbers are displayed.
Direction Only results from one direction of the OLTC’s movement are displayed.
OMICRON 163
CPC 100 PTM User Manual
Note: The Dynamic OLTC-Scan (DRM) is restricted to 50 taps at once. For OLTCs with more than 50
taps, enter the appropriate Start tap and Stop tap to define which set of 50 taps you would like to
examine.
Notification Description
! Data corruption: A disturbance has been detected. This result is invalid.
Repeat the measurement.
!! Time out: No switching operation has been detected during the Tap time.
164 OMICRON
Transformer tests
DANGER
Death or severe injury caused by high voltage or current
► Never open the measuring circuit while current flows.
► After a measurement, wait until the test device has discharged completely.
► Ground all terminals of the test object before touching the test setup.
► Short-circuit the terminals before disconnecting the test leads.
► Disconnect cables not used for testing both from the test object and the test device.
The following table describes the Demagnetization test settings.
Table 12-53: Demagnetization test settings
Setting Description
Measurement settings
CPC output DC current output of the CPC 100
Test current Current of the test
Saturation criteria
Saturation level Saturation level of the transformer
OMICRON 165
CPC 100 PTM User Manual
Data Description
No. Number of the measurement
Measurement Name of the measurement
Stage Test status:
• Wiring check
• Saturating
• Discharging
• Canceled
• Demagnetizing
• Test stopped
• Core is demagnetized1
I DC Measured current
Min. neg. remanence Maximum remanence in negative direction of the hysteresis curve
Max. pos. remanence Maximum remanence in positive direction of the hysteresis curve
Remanence Measured remanence
Initial remanence Measured remanence at start of the test
Assessment Measurement assessment
1. If the measured remanence < 1%.
1. Test name depends on the Profile selected in Settings – Profiles (see "Profiles" on page 26).
166 OMICRON
Transformer tests
OMICRON 167
CPC 100 PTM User Manual
Primary Test Manager supports the following tests of autotransformers without tertiary winding:
• Oil Analysis
• Overall PF & CAP test1
• Bushing PF & CAP – C1 test1
• Bushing PF & CAP – C2 test1
• Bushing – Energized collar test1
• Exciting Current test
• Insulating Fluids PF & CAP test1
• Surge Arrester Watt Losses test1
• Leakage Reactance test1
• Zero Sequence Impedance test1
• TTR test1
• DC Winding Resistance test1
• Dynamic OLTC-Scan (DRM) test1
• Demagnetization test
Note: The tests of the autotransformers without tertiary winding are identical to the respective tests of
the two-winding transformers. For more information, see 12.1 "Two-winding transformers" on page 115.
Primary Test Manager supports additionally the TanDelta – PF/DF/Tanδ & CAP manual test. For more
information, see 19.2 "Manual Tan Delta" on page 251.
1. Test name depends on the Profile selected in Settings – Profiles (see "Profiles" on page 26).
168 OMICRON
Transformer tests
Primary Test Manager supports the following tests of autotransformers with tertiary winding:
• Oil Analysis
• Overall PF & CAP test1
• Bushing PF & CAP – C1 test1
• Bushing PF & CAP – C2 test1
• Bushing – Energized collar test1
• Exciting Current test
• Insulating Fluids PF & CAP test1
• Surge Arrester Watt Losses test1
• Leakage Reactance test1
• Zero Sequence Impedance test1
• TTR test1
• DC Winding Resistance test1
• Dynamic OLTC-Scan (DRM) test1
• Demagnetization test
Note: The tests of the autotransformers with tertiary winding are identical to the respective tests of the
two-winding transformers. For more information, see 12.1 "Two-winding transformers" on page 115. The
available tests depend on the accessibility of the tertiary winding.
Primary Test Manager supports additionally the TanDelta – PF/DF/Tanδ & CAP manual test. For more
information, see 19.2 "Manual Tan Delta" on page 251.
1. Test name depends on the Profile selected in Settings – Profiles (see "Profiles" on page 26).
OMICRON 169
CPC 100 PTM User Manual
1. Test name depends on the Profile selected in Settings – Profiles (see "Profiles" on page 26).
170 OMICRON
Spare Bushing tests
1. Test name depends on the Profile selected in Settings (see "Profiles" on page 26).
OMICRON 171
CPC 100 PTM User Manual
DANGER
Death or severe injury caused by high voltage or current
The high-voltage cable is double-shielded and therefore safe. However, the last 50 cm (20 inches)
of this cable have no shield.
► Avoid any direct contact of this part of the cable to ground potential or to any objects.
► During a test, consider this part of the cable a live wire and, due to the high-voltage, life-
threatening.
Note: This test name depends on the Profile selected in the Settings (see "Profiles" on page 26):
• IEEE standard: Spare Bushing PF & CAP – Overall
• IEC standard: Spare Bushing DF & CAP – Overall
• Custom profile: for example Spare Bushing Tanδ & CAP – Overall
In this section, the terms Spare Bushing PF & CAP – Overall and Power Factor (PF) will be used.
The following table describes the Spare Bushing PF & CAP – Overall test settings.
Table 13-1: Spare Bushing PF & CAP – overall test settings
Setting Description
Measurement settings
Test frequency Test frequency
Sweep settings
Frequency sweep Sweep profile:
None, OMICRON expertise (recommended), or CPC template
• None: no frequency sweep
• OMICRON expertise: sweep frequencies dynamically distributed
within the CPC 100 frequency range for optimum results
• CPC template: sweep frequencies specified by the CPC 100 test
templates
Voltage sweep (tip-up) Sweep profile:
None or OMICRON expertise
• None: no voltage sweep
• OMICRON expertise: sweep voltages dynamically distributed within
the asset-dependent voltage range for optimum results
172 OMICRON
Spare Bushing tests
Table 13-1: Spare Bushing PF & CAP – overall test settings (continued)
Setting Description
► Click the pen button to create a frequency or voltage sweep
Sweep profiles
profile.
► Add up to 30 measurement points with individual output voltages or
frequencies. Double-click a value to change it.
► Mark a favorite to use it as the default sweep profile for future
tests.
Note: The predefined profiles None, OMICRON expertise and
CPC template cannot be edited or deleted.
The default sweep profiles for this test are:
• Frequency sweep: OMICRON expertise
• Voltage sweep: none
Enable shield check ► Select the Enable shield check check box if you want that the
CPC 100 checks whether the shield of the high-voltage cable is
connected.
Use beeper ► Select the Use beeper check box if you want to use the CP TD’s
beeper during the measurement.
Test conditions
Custom test conditions ► Activate the Custom test conditions check box to set test
conditions differing from the global test conditions.
OMICRON 173
CPC 100 PTM User Manual
Table 13-1: Spare Bushing PF & CAP – overall test settings (continued)
Setting Description
Correction factors
Temperature correction ► Select the Temperature correction check box to activate
temperature correction.
Data Description
No. Number of the measurement
Measurement Arrangement of the measurement
Test mode Test mode according to the IEEE Std 62-1995
Sweep Swept variable: frequency, voltage, or none
V test Test voltage
Freq. Test frequency
V out Measured output voltage
I out Measured output current
Watt losses Measured losses
174 OMICRON
Spare Bushing tests
DANGER
Death or severe injury caused by high voltage or current
The high-voltage cable is double-shielded and therefore safe. However, the last 50 cm (20 inches)
of this cable have no shield.
► Avoid any direct contact of this part of the cable to ground potential or to any objects.
► During a test, consider this part of the cable a live wire and, due to the high-voltage, life-
threatening.
Note: This test name depends on the Profile selected in the Settings (see "Profiles" on page 26):
• IEEE standard: Spare Bushing PF & CAP – C1
• IEC standard: Spare Bushing DF & CAP – C1
• Custom profile: for example Spare Bushing Tanδ & CAP – C1
In this section, the terms Power Factor (PF) and Spare Bushing PF & CAP – C1 will be used.
The following table describes the Spare Bushing PF & CAP – C1 test settings.
Table 13-3: Spare Bushing PF & CAP – C1 test settings
Setting Description
Measurement settings
Test frequency Test frequency
Perform inverted C1 test ► Select the Perform inverted C1 test check box to include the
inverted C1 measurement in the test.
Sweep settings
Frequency sweep Sweep frequency profile:
None, OMICRON expertise (recommended), or CPC template.
• None: no frequency sweep
• OMICRON expertise: sweep frequencies dynamically distributed
within the CPC 100 frequency range for optimum results
• CPC template: sweep frequencies specified by the CPC 100 test
templates
OMICRON 175
CPC 100 PTM User Manual
Setting Description
Voltage sweep (tip-up) Sweep profile:
None or OMICRON expertise
• None: no voltage sweep
• OMICRON expertise: sweep voltages dynamically distributed within
the asset-dependent voltage range for optimum results
► Click the pen button to create a frequency or voltage sweep
Sweep profiles
profile.
► Add up to 30 measurement points with individual output voltages or
frequencies. Double-click a value to change it.
► Mark a favorite to use it as the default sweep profile for future
tests.
Note: The predefined profiles None, OMICRON expertise and
CPC template cannot be edited or deleted.
The default sweep profiles for this test are:
• Frequency sweep: OMICRON expertise
• Voltage sweep: none
Enable shield check ► Select the Enable shield check check box if you want that the
CPC 100 checks whether the shield of the high-voltage cable is
connected.
Use beeper ► Select the Use beeper check box if you want to use the CP TD’s
beeper during the measurement.
Test conditions
Custom test conditions ► Activate the Custom test conditions check box to set test
conditions differing from the global test conditions.
176 OMICRON
Spare Bushing tests
Setting Description
Correction factors
Temperature correction ► Select the Temperature correction check box to activate
temperature correction.
The following table describes the Spare Bushing PF & CAP – C1 measurement data.
Table 13-4: Spare Bushing PF & CAP – C1 measurement data
Data Description
No. Number of the measurement
Measurement Arrangement of the measurement
Test mode Test mode according to the IEEE Std 62-1995
Sweep Swept variable: frequency, voltage, or none
V test Test voltage
Freq. Test frequency
V out Measured output voltage
I out Measured output current
Watt losses Measured losses
OMICRON 177
CPC 100 PTM User Manual
DANGER
Death or severe injury caused by high voltage or current
The high-voltage cable is double-shielded and therefore safe. However, the last 50 cm (20 inches)
of this cable have no shield.
► Avoid any direct contact of this part of the cable to ground potential or to any objects.
► During a test, consider this part of the cable a live wire and, due to the high-voltage, life-
threatening.
Note: This test name depends on the Profile selected in the Settings (see "Profiles" on page 26):
• IEEE standard: Spare Bushing PF & CAP – C2
• IEC standard: Spare Bushing DF & CAP – C2
• Custom profile: for example Spare Bushing Tanδ & CAP – C2
In this section, the terms Power Factor (PF) and Spare Bushing PF & CAP – C2 will be used.
The following table describes the Spare Bushing PF & CAP – C2 test settings.
Table 13-5: Spare Bushing PF & CAP – C2 test settings
Setting Description
Measurement settings
Test frequency Test frequency
Noise suppression settings
Averaging (no. points) Number of averaged measurements
Bandwidth CPC 100 filter bandwidth
Device settings
Tan delta device ► Select your CP TD device:
• CP TD1
• CP TD12
• CP TD15
Enable shield check ► Select the Enable shield check check box if you want that the
CPC 100 checks whether the shield of the high-voltage cable is
connected.
178 OMICRON
Spare Bushing tests
Setting Description
Use beeper ► Select the Use beeper check box if you want to use the CP TD’s
beeper during the measurement.
Test conditions
Custom test conditions ► Activate the Custom test conditions check box to set test
conditions differing from the global test conditions.
The following table describes the Spare Bushing PF & CAP – C2 measurement data.
Table 13-6: Spare Bushing PF & CAP – C2 measurement data
Data Description
No. Number of the measurement
Measurement Arrangement of the measurement
Test mode Test mode according to the IEEE Std 62-1995
V test Test voltage
Freq. Test frequency
V out Measured output voltage
I out Measured output current
Watt losses Measured losses
OMICRON 179
CPC 100 PTM User Manual
Data Description
180 OMICRON
Spare Bushing tests
DANGER
Death or severe injury caused by high voltage or current
The high-voltage cable is double-shielded and therefore safe. However, the last 50 cm (20 inches)
of this cable have no shield.
► Avoid any direct contact of this part of the cable to ground potential or to any objects.
► During a test, consider this part of the cable a live wire and, due to the high-voltage, life-
threatening.
Note: This test name depends on the Profile selected in the Settings (see "Profiles" on page 26):
• IEEE standard: Spare Bushing PF & CAP – C1+C2
• IEC standard: Spare Bushing DF & CAP – C1+C2
• Custom profile: for example Spare Bushing Tanδ & CAP – C1+C2
In this section, the terms Power Factor (PF) and Spare Bushing PF & CAP – C1+C2 will be used.
The following table describes the Spare Bushing PF & CAP – C1+C2 test settings.
Table 13-7: Spare Bushing PF & CAP – C1+C2 test settings
Setting Description
Measurement settings
Test frequency Test frequency
Noise suppression settings
Averaging (no. points) Number of averaged measurements
Bandwidth CPC 100 filter bandwidth
Device settings
Tan delta device ► Select your CP TD device:
• CP TD1
• CP TD12
• CP TD15
Enable shield check ► Select the Enable shield check check box if you want that the
CPC 100 checks whether the shield of the high-voltage cable is
connected.
OMICRON 181
CPC 100 PTM User Manual
Setting Description
Use beeper ► Select the Use beeper check box if you want to use the CP TD’s
beeper during the measurement.
Test conditions
Custom test conditions ► Activate the Custom test conditions check box to set test
conditions differing from the global test conditions.
The following table describes the Spare Bushing PF & CAP – C1+C2 measurement data.
Table 13-8: Spare Bushing PF & CAP – C1+C2 measurement data
Data Description
No. Number of the measurement
Measurement Arrangement of the measurement
Test mode Test mode according to the IEEE Std 62-1995
V test Test voltage
Freq. Test frequency
V out Measured output voltage
I out Measured output current
Watt losses Measured losses
182 OMICRON
Spare Bushing tests
DANGER
Death or severe injury caused by high voltage or current
The high-voltage cable is double-shielded and therefore safe. However, the last 50 cm (20 inches)
of this cable have no shield.
► Avoid any direct contact of this part of the cable to ground potential or to any objects.
► During a test, consider this part of the cable a live wire and, due to the high-voltage, life-
threatening.
The following table describes the Spare Bushing – Energized Collar test settings.
Table 13-9: Spare Bushing – Energized Collar test settings
Setting Description
Measurement settings
Test frequency Test frequency
Noise suppression settings
Averaging (no. points) Number of averaged measurements
Bandwidth CPC 100 filter bandwidth
Device settings
Tan delta device ► Select your CP TD device:
• CP TD1
• CP TD12
• CP TD15
Enable shield check ► Select the Enable shield check check box if you want that the
CPC 100 checks whether the shield of the high-voltage cable is
connected.
Use beeper ► Select the Use beeper check box if you want to use the CP TD’s
beeper during the measurement.
Test conditions
Custom test conditions ► Activate the Custom test conditions check box to set test
conditions differing from the global test conditions.
OMICRON 183
CPC 100 PTM User Manual
Setting Description
Weather Weather during the test
Correction factors
Use reference voltage ► Select the Use reference voltage check box to extrapolate the I out
and Watt losses results for the specified reference voltage.
The following table describes the Spare Bushing – Energized Collar measurement data.
Table 13-10: Spare Bushing – Energized Collar measurement data
Data Description
No. Number of the measurement
Measurement Arrangement of the measurement
Test mode Test mode according to the IEEE Std 62-1995
V test Test voltage
Freq. Test frequency
V out Measured output voltage
I out Measured output current
Watt losses Measured losses
Assessment Measurement assessment
184 OMICRON
Circuit breaker tests
1. Test name depends on the Profile selected in Settings (see "Profiles" on page 26).
OMICRON 185
CPC 100 PTM User Manual
DANGER
Death or severe injury caused by high voltage or current
The high-voltage cable is double-shielded and therefore safe. However, the last 50 cm (20 inches)
of this cable have no shield.
► Avoid any direct contact of this part of the cable to ground potential or to any objects.
► During a test, consider this part of the cable a live wire and, due to the high-voltage, life-
threatening.
Note: This test name depends on the Profile selected in the Settings (see "Profiles" on page 26):
• IEEE standard: Circuit Breaker PF & CAP
• IEC standard: Circuit Breaker DF & CAP
• Custom profile: for example Circuit Breaker Tanδ & CAP
In this section, the terms Power Factor (PF) and Circuit Breaker PF & CAP will be used.
The following table describes the Circuit Breaker PF & CAP test settings.
Table 14-1: Circuit Breaker PF & CAP test settings
Setting Description
Measurement settings
Test frequency Frequency of the test
Sweep settings
Frequency sweep Sweep profile:
None, OMICRON expertise (recommended), or CPC template
• None: no frequency sweep
• OMICRON expertise: sweep frequencies dynamically distributed
within the CPC 100 frequency range for optimum results
• CPC template: sweep frequencies specified by the CPC 100 test
templates
Voltage sweep (tip-up) Sweep profile:
None or OMICRON expertise
• None: no voltage sweep
• OMICRON expertise: sweep voltages dynamically distributed within
the asset-dependent voltage range for optimum results
186 OMICRON
Circuit breaker tests
Setting Description
► Click the pen button to create a frequency or voltage sweep
Sweep profiles
profile.
► Add up to 30 measurement points with individual output voltages or
frequencies. Double-click a value to change it.
► Mark a favorite to use it as the default sweep profile for future
tests.
Note: The predefined profiles None, OMICRON expertise and
CPC template cannot be edited or deleted.
The default sweep profiles for this test are:
• Frequency sweep: None
• Voltage sweep: None
Enable shield check ► Select the Enable shield check check box if you want that the
CPC 100 checks whether the shield of the high-voltage cable is
connected.
Use beeper ► Select the Use beeper check box if you want to use the CP TD’s
beeper during the measurement.
Test conditions
Custom test conditions ► Select the Custom test conditions check box to set the test
conditions different from the global test conditions.
OMICRON 187
CPC 100 PTM User Manual
Setting Description
Correction factors
Temperature correction ► Select the Temperature correction check box to activate
temperature correction.
Data Description
No. Number of the measurement
Measurement Tested insulation
Position Contact position of the breaker
Test mode Test mode according to the IEEE Std 62-1995
Sweep Swept variable: frequency, voltage, or none
V test Test voltage
Freq. Test frequency
V out Measured output voltage
I out Measured output current
Watt losses Measured losses
Cap. meas Measured capacitance
188 OMICRON
Circuit breaker tests
DANGER
Death or severe injury caused by high voltage or current
The high-voltage cable is double-shielded and therefore safe. However, the last 50 cm (20 inches)
of this cable have no shield.
► Avoid any direct contact of this part of the cable to ground potential or to any objects.
► During a test, consider this part of the cable a live wire and, due to the high-voltage, life-
threatening.
Note: This test name depends on the Profile selected in the Settings (see "Profiles" on page 26):
• IEEE standard: Bushing PF & CAP – C1
• IEC standard: Bushing DF & CAP – C1
• Custom profile: for example Circuit Breaker Tanδ & CAP – C1
In this section, the terms Power Factor (PF) and Bushing PF & CAP – C1 will be used.
The following table describes the Bushing PF & CAP – C1 test settings.
Table 14-3: Bushing PF & CAP – C1 test settings
Setting Description
Measurement settings
Test frequency Test frequency
Sweep settings
Frequency sweep Sweep profile:
None, OMICRON expertise (recommended), or CPC template
• None: no frequency sweep
• OMICRON expertise: sweep frequencies dynamically distributed
within the CPC 100 frequency range for optimum results
• CPC template: sweep frequencies specified by the CPC 100 test
templates
Voltage sweep (tip-up) Sweep profile:
None or OMICRON expertise
• None: no voltage sweep
• OMICRON expertise: sweep voltages dynamically distributed within
the asset-dependent voltage range for optimum results
OMICRON 189
CPC 100 PTM User Manual
Setting Description
► Click the pen button to create a frequency or voltage sweep
Sweep profiles
profile.
► Add up to 30 measurement points with individual output voltages or
frequencies. Double-click a value to change it.
► Mark a favorite to use it as the default sweep profile for future
tests.
Note: The predefined profiles None, OMICRON expertise and
CPC template cannot be edited or deleted.
The default sweep profiles for this test are:
• Frequency sweep: None
• Voltage sweep: None
Enable shield check ► Select the Enable shield check check box if you want that the
CPC 100 checks whether the shield of the high-voltage cable is
connected.
Use beeper ► Select the Use beeper check box if you want to use the CP TD’s
beeper during the measurement.
Test conditions
Custom test conditions ► Select the Custom test conditions check box if you want to set test
conditions different from the job conditions.
190 OMICRON
Circuit breaker tests
Setting Description
Temperature correction ► Select the Temperature correction check box to activate
temperature correction.
Data Description
No. Number of the measurement
Measurement Terminal name of the bushing under test
Test mode Test mode according to the IEEE Std 62-1995
Sweep Swept variable: frequency, voltage, or none
V test Test voltage
Freq. Test frequency
V out Measured output voltage
I out Measured output current
Watt losses Measured losses
Cap. meas Measured capacitance
OMICRON 191
CPC 100 PTM User Manual
DANGER
Death or severe injury caused by high voltage or current
The high-voltage cable is double-shielded and therefore safe. However, the last 50 cm (20 inches)
of this cable have no shield.
► Avoid any direct contact of this part of the cable to ground potential or to any objects.
► During a test, consider this part of the cable a live wire and, due to the high-voltage, life-
threatening.
Note: This test name depends on the Profile selected in the Settings (see "Profiles" on page 26):
• IEEE standard: Bushing PF & CAP – C2
• IEC standard: Bushing DF & CAP – C2
• Custom profile: for example Bushing Tanδ & CAP – C2
In this section, the terms Power Factor (PF) and Bushing PF & CAP – C2 will be used.
The following table describes the Bushing PF & CAP – C2 test settings.
Table 14-5: Bushing PF & CAP – C2 test settings
Setting Description
Measurement settings
Test frequency Test frequency
Test voltage Test voltage
Noise suppression settings
Averaging (no. points) Number of averaged measurements
Bandwidth CPC 100 filter bandwidth
Device settings
Tan delta device ► Select your CP TD device:
• CP TD1
• CP TD12
• CP TD15
192 OMICRON
Circuit breaker tests
Setting Description
Enable shield check ► Select the Enable shield check check box if you want that the
CPC 100 checks whether the shield of the high-voltage cable is
connected.
Use beeper ► Select the Use beeper check box if you want to use the CP TD’s
beeper during the measurement.
Test conditions
Custom test conditions ► Select the Custom test conditions check box if you want to set test
conditions different from the job conditions.
Data Description
No. Number of the measurement
Measurement Terminal name of the bushing under test
Test mode Test mode according to the IEEE Std 62-1995
V test Test voltage
Freq. Test frequency
V out Measured output voltage
I out Measured output current
Watt losses Measured losses
OMICRON 193
CPC 100 PTM User Manual
Data Description
Cap. meas Measured capacitance
194 OMICRON
Circuit breaker tests
DANGER
Death or severe injury caused by high voltage or current
The high-voltage cable is double-shielded and therefore safe. However, the last 50 cm (20 inches)
of this cable have no shield.
► Avoid any direct contact of this part of the cable to ground potential or to any objects.
► During a test, consider this part of the cable a live wire and, due to the high-voltage, life-
threatening.
The following table describes the Bushing – Energized Collar test settings.
Table 14-7: Bushing – Energized Collar test settings
Setting Description
Measurement settings
Test frequency Test frequency
Test voltage Test voltage
Noise suppression settings
Averaging (no. points) Number of averaged measurements
Bandwidth CPC 100 filter bandwidth
Device settings
Tan delta device ► Select your CP TD device:
• CP TD1
• CP TD12
• CP TD15
Enable shield check ► Select the Enable shield check check box if you want that the
CPC 100 checks whether the shield of the high-voltage cable is
connected.
Use beeper ► Select the Use beeper check box if you want to use the CP TD’s
beeper during the measurement.
Test conditions
Custom test conditions ► Select the Custom test conditions check box if you want to set test
conditions different from the job conditions.
OMICRON 195
CPC 100 PTM User Manual
Setting Description
Humidity Relative ambient humidity
Weather Weather during the test
Correction factors
Use reference voltage ► Select the Use reference voltage check box to extrapolate the I out
and Watt losses results for the specified reference voltage.
Data Description
No. Number of the measurement
Measurement Terminal name of the bushing under test
Test mode Test mode according to the IEEE Std 62-1995
V test Test voltage
Freq. Test frequency
V out Measured output voltage
I out Measured output current
Watt losses Measured losses
Assessment Measurement assessment
196 OMICRON
Circuit breaker tests
DANGER
Death or severe injury caused by high voltage or current
The high-voltage cable is double-shielded and therefore safe. However, the last 50 cm (20 inches)
of this cable have no shield.
► Avoid any direct contact of this part of the cable to ground potential or to any objects.
► During a test, consider this part of the cable a live wire and, due to the high-voltage, life-
threatening.
Note: This test name depends on the Profile selected in the Settings (see "Profiles" on page 26):
• IEEE standard: Insulating Fluids PF & CAP
• IEC standard: Insulating Fluids DF & CAP
• Custom profile: for example Insulating Fluids Tanδ & CAP
In this section, the terms Power Factor (PF) and Insulating Fluids PF & CAP will be used.
The following table describes the Insulating Fluids PF & CAP test settings.
Table 14-9: Insulating Fluids PF & CAP test settings
Setting Description
Measurement settings
Test frequency Test frequency
Test voltage Test voltage
Noise suppression settings
Averaging (no. points) Number of averaged measurements
Bandwidth CPC 100 filter bandwidth
Device settings
Tan delta device ► Select your CP TD device:
• CP TD1
• CP TD12
• CP TD15
Enable shield check ► Select the Enable shield check check box if you want that the
CPC 100 checks whether the shield of the high-voltage cable is
connected.
OMICRON 197
CPC 100 PTM User Manual
Setting Description
Use beeper ► Select the Use beeper check box if you want to use the CP TD’s
beeper during the measurement.
Test conditions
Oil temperature Temperature of the oil
Custom test conditions ► Activate the Custom test conditions check box to set test
conditions differing from the global test conditions.
Data Description
No. Number of the measurement
Specimen Oil sample under test
Test mode Test mode according to the IEEE Std 62-1995
V Test Test voltage
Freq. Test frequency
V out Measured output voltage
I out Measured output current
Watt losses Measured losses
Cap. meas Measured capacitance
198 OMICRON
Circuit breaker tests
WARNING
Death or severe injury caused by high voltage or current possible
► Do not use external power sources for the circuit breaker’s main contacts.
► During the test, supply the circuit breaker’s main contacts only with the CPC 100.
OMICRON 199
CPC 100 PTM User Manual
Setting Description
Measurement settings
Test current Test current
Manual input of V DC ► Select the Manual input of V DC check box to enter the V DC
manually instead of measuring it.
Automatic result ► Select the Automatic result check box to stop the measurement
automatically depending on the tolerance R dev and the settling
time.
NOTICE
Equipment damage or loss of data possible
Before making a decision based on the Primary Test Manager automatic assessment, read the
disclaimer 1.3.1 "Automatic assessment" on page 10.
200 OMICRON
Circuit breaker tests
2. In the Edit automatic assessment limits dialog box, enter the absolute or relative limits.
The following table describes the Contact Resistance measurement data.
Table 14-12: Contact Resistance measurement data
Data Description
Main contact Main contact of the circuit breaker under test
I DC Measured current
V DC Measured voltage
R meas Measured resistance
R min Minimum assessment limit
R max Maximum assessment limit
Assessment Measurement assessment
Note: If you did not select the Automatic result check box to stop the measurement automatically, you
can stop the measurement by clicking the Keep results button.
OMICRON 201
CPC 100 PTM User Manual
1. Test name depends on the Profile selected in the Settings (see "Profiles" on page 26).
2. Only available for IEC 60044 and IEEE C57.13
202 OMICRON
Current transformer tests
DANGER
Death or severe injury caused by high voltage or current
The high-voltage cable is double-shielded and therefore safe. However, the last 50 cm (20 inches)
of this cable have no shield.
► Avoid any direct contact of this part of the cable to ground potential or to any objects.
► During a test, consider this part of the cable a live wire and, due to the high-voltage, life-
threatening.
Note: This test name depends on the Profile selected in the Settings (see "Profiles" on page 26):
• IEEE standard: CT PF & CAP
• IEC standard: CT DF & CAP
• Custom profile: for example CT Tanδ & CAP
In this section, the terms Power factor (PF) and CT PF & CAP will be used.
Setting Description
Measurement settings
Test frequency Frequency of the output signal
HV-shield connectable ► Select the HV-shield connectable check box if the HV screen or
shield of the current transformer is accessible.
Sweep settings
Frequency sweep Sweep profile:
None, OMICRON expertise (recommended), or CPC template
• None: no frequency sweep
• OMICRON expertise: sweep frequencies dynamically distributed
within the CPC 100 frequency range for optimum results
• CPC template: sweep frequencies specified by the CPC 100 test
templates
OMICRON 203
CPC 100 PTM User Manual
Setting Description
Voltage sweep (tip-up) Sweep profile:
None or OMICRON expertise
• None: no voltage sweep
• OMICRON expertise: sweep voltages dynamically distributed within
the asset-dependent voltage range for optimum results
► Click the pen button to create a frequency or voltage sweep
Sweep profiles
profile.
► Add up to 30 measurement points with individual output voltages or
frequencies. Double-click a value to change it.
► Mark a favorite to use it as the default sweep profile for future
tests.
Note: The predefined profiles None, OMICRON expertise and
CPC template cannot be edited or deleted.
The default sweep profiles for this test are:
• Frequency sweep: OMICRON
• Voltage sweep: None
Sweep settings
Noise suppression settings
Averaging (no. points) Number of averaged measurements
Bandwidth CPC 100 filter bandwidth
If this setting is active, the measurement will not be performed at the
Test frequency set in the Measurement section. Primary Test
Manager will instead measure two values at frequencies below and
above the entered Test frequency and calculate the median of those
Avoid test frequency
two values.
The Avoid test frequency setting is predefined for the selected test.
► Only change the default setting for special applications.
Device settings
Tan delta device ► Select your CP TD device:
• CP TD1
• CP TD12
• CP TD15
Enable shield check ► Select the Enable shield check check box if you want that the
CPC 100 checks whether the shield of the high-voltage cable is
connected.
Use beeper ► Select the Use beeper check box if you want to use the CP TD’s
beeper during the measurement.
204 OMICRON
Current transformer tests
Setting Description
Test conditions
Custom test conditions ► Activate the Custom test conditions check box to set test
conditions differing from the global test conditions.
Data Description
Measurement Arrangement of the measurement
Test mode Test mode according to the IEEE Std 62-1995
Sweep Swept variable: frequency, voltage, or none
V test Test voltage
Freq. Test frequency
V out Measured output voltage
I out Measured output current
Watt losses Measured losses
Cap. meas Measured capacitance
OMICRON 205
CPC 100 PTM User Manual
15.2 CT Ratio
The following table describes the CT Ratio test settings.
DANGER
Death or severe injury caused by high voltage or current
► Make sure that no secondary CT windings are open.
Setting Description
Measurement settings
CPC output Selected output range for the current transformer test
Manual measurement I sec ► Select the Manual measurement I sec check box to enter the I sec
manually, for example, if I sec was measured with an external
current clamp not connected to the CPC 100.
Use current clamp ► Select the Use current clamp check box if you want to use current
clamp.
NOTICE
Equipment damage or loss of data possible
Before making a decision based on the Primary Test Manager automatic assessment, read the
disclaimer 1.3.1 "Automatic assessment" on page 10.
206 OMICRON
Current transformer tests
2. In the Edit automatic assessment limits dialog box, enter the assessment limits.
The following table describes the CT Ratio measurement data.
Table 15-4: CT Ratio measurement data
Data Description
Tap Identification of the tap under test
Auto Select the Auto check box to stop the measurement automatically.
Otherwise, the result is saved when you press Keep results.
I test Test current at the primary side of the current transformer
Frequency Frequency of the output signal
I prim Measured current at the primary side of the current transformer
I sec Measured secondary current of the current transformer
Ratio Measured ratio of the current transformer
Ratio dev Deviation of the measured ratio from the rated ratio
Phase Measured phase displacement between the primary and secondary
current
Polarity Polarity status
OMICRON 207
CPC 100 PTM User Manual
Data Description
Measure burden Select the Measure burden check box to measure the connected
secondary burden in VA.
Manual input V sec Select the Manual input V sec check box to enter the measured
secondary voltage at the burden manually, for example, if it is measured
with an external device.
V sec Measured secondary voltage on the burden
Burden Measured burden in VA related to the rated secondary current
Burden nom. Rated power of the current transformer
Phase burden Phase angle of the complex secondary burden
cos φ Phase angle of the complex rated secondary burden
In use Tap on which the CT is operated
Assessment Measurement assessment
1. According to IEC 60444
2. According to IEC 61869 and IEEE C57.13
208 OMICRON
Current transformer tests
DANGER
Death or severe injury caused by high voltage or current
► Never open the measuring circuit while current flows.
► Make sure that no other secondary windings are open.
► After a measurement, wait until the CPC 100 has discharged completely.
► Before disconnecting from the CPC 100, connect the device under test on both ends to
protective earth.
► Short-circuit the winding under test before disconnecting the test leads.
Setting Description
Temperature compensation for Cu
Winding temperature Temperature of the transformer’s winding
Reference temperature Reference temperature to be used for temperature correction
Note: The transformer’s core is magnetized after the measurement. Demagnetization can be done using
the CTExcitation test (see 15.4 "CT Excitation" on page 211).
Setting Description
Tap Identification of the tap under test
CPC output Selected output range for the current transformer test
I test Test current
I DC Measured current
V DC Measured voltage
OMICRON 209
CPC 100 PTM User Manual
Setting Description
R meas Measured resistance
R corr Corrected measured resistance
In use Tap on which the CT is operated
Assessment Measurement assessment
1. According to IEC 60444
2. According to IEC 61869 and IEEE C57.13
210 OMICRON
Current transformer tests
15.4 CT Excitation
The following table describes the CT Excitation test settings.
DANGER
Death or severe injury caused by high voltage or current
Feeding test voltage to a tap of a CT can cause life-threatening voltages on other taps and
windings.
► Make sure that no secondary CT windings are open.
Setting Description
Measurement settings
Test frequency Frequency of the output signal
Noise suppression ► Select the Noise suppression check box to activate an interference
filter.
Setting Description
Tap Identification of the tap under test
Ipn Rated primary current of the current transformer
Isn Rated secondary current of the current transformer
I max Maximum test current
V max Maximum test voltage
I knee Knee point current
V knee Knee point voltage
Knee point calculation Method used for the knee point calculation
In use Tap on which the CT is operated
Assessment Measurement assessment
OMICRON 211
CPC 100 PTM User Manual
15.5 CT Ratio V
The following table describes the CT Ratio V test settings.
DANGER
Death or severe injury caused by high voltage or current
Feeding test voltage to a tap of a CT can cause life-threatening voltages on other taps and
windings.
► Do not touch tapped windings.
► Make sure that no secondary CT windings are open.
Setting Description
Measurement settings
Test frequency Frequency of the output signal
Setting Description
Tap Identification of the tap under test
Note: If the transformer’s knee point voltage is approximated or exceeded due to the transformer’s
saturation, the measurement results are not correct anymore. If the knee point is significantly exceeded,
the transformer can even be damaged. Therefore, the knee point voltage should be known or measured
212 OMICRON
Current transformer tests
beforehand. Generally, it is advisable to set the test voltage to 75 % of the knee-point voltage. The
frequency should be set between 15 and 20 Hz off the mains frequency to avoid interferences by life
systems in the neighborhood.
15.6 CT Burden
The following table describes the CT Burden test settings.
DANGER
Death or severe injury caused by high voltage or current
► Make sure that no secondary CT windings are open.
NOTICE
Disconnect burden from the CT to avoid measurement errors.
Setting Description
Measurement settings
Test frequency Frequency of the output signal
Setting Description
Tap Identification of the tap under test
OMICRON 213
CPC 100 PTM User Manual
DANGER
Death or severe injury caused by high voltage or current
► Never open the measuring circuit while current flows.
► Make sure that no other secondary windings are open.
► After a measurement, wait until the CPC 100 has discharged completely.
► Before disconnecting from the CPC 100, connect the device under test on both ends to
protective earth.
► Short-circuit the winding under test before disconnecting the test leads.
DANGER
Death or severe injury caused by high voltage or current
Feeding test voltage to a tap of a CT can cause life-threatening voltages on other taps with higher
ratios.
► Do not touch tapped windings.
► Make sure that no other secondary windings are open.
The CT Composite Error test is only available for protection CTs:
• classes C, K or T according to IEEE C57.13
• classes P or PR according to IEC 60044/IEC 61869
Note: The transformer’s core is magnetized after the Winding Resistance measurement. Since the
core is demagnetized in the course of the CT Excitation test, we recommend performing the Winding
Resistance measurement first (see 15.4 "CT Excitation" on page 211).
Setting Description
CT Excitation Measurement Settings
Test frequency Frequency of the output signal
Noise suppression ► Select the check box to activate an interference filter.
214 OMICRON
Current transformer tests
Setting Description
Burden settings2
Use rated burden ► Choose whether the rated burden or the operating burden be used
for the composite error calculation. These values are entered in the
Use operating burden asset.
Assessment settings
Manual ► Select for manual assessment of the measurement results
Setting Description
Tap Identification of the tap under test
OMICRON 215
CPC 100 PTM User Manual
Setting Description
V DC Measured voltage
R meas Measured resistance
R corr Corrected measured resistance
In use Indicates if the tap under test is designated as In use in the asset
Assessment Measurement assessment
1. For IEEE C57.13: Only available if Perform winding resistance measurement check box is activated
2. According to IEC 60444
3. According to IEC 61869 and IEEE C57.13
Setting Description
Tap Identification of the tap under test
Setting Description
Tap Identification of the tap under test
216 OMICRON
Current transformer tests
Setting Description
Composite error @ ALF x Isr4 Composite error at rated accuracy limit current
OMICRON 217
CPC 100 PTM User Manual
DANGER
Death or severe injury caused by high voltage or current
The high-voltage cable is double-shielded and therefore safe. However, the last 50 cm (20 inches)
of this cable have no shield.
► Avoid any direct contact of this part of the cable to ground potential or to any objects.
► During a test, consider this part of the cable a live wire and, due to the high-voltage, life-
threatening.
Note: This test name depends on the Profile selected in the Settings (see "Profiles" on page 26):
• IEEE standard: VT PF & CAP
1. Test name depends on the Profile selected in the Settings (see "Profiles" on page 26).
2. Only available for inductive voltage transformers
218 OMICRON
Voltage transformer tests
Data Description
Upr Rated primary voltage of the voltage transformer
Usr Rated secondary voltage of the voltage transformer
V test Test voltage at the primary side of the voltage transformer
Test frequency Test frequency of the output signal
V prim Measured voltage at the primary side of the voltage transformer
V sec Measured secondary voltage of the voltage transformer
Ratio Measured ratio of the voltage transformer
Ratio dev Deviation of the measured ratio from the rated ratio
Phase Measured phase displacement between the primary and secondary
voltage
Polarity Polarity status
Assessment Measurement assessment
OMICRON 219
CPC 100 PTM User Manual
Setting Description
Measurement settings
Test frequency Test frequency
Setting Description
Upr Rated primary voltage of the voltage transformer
Usr Rated secondary current of the voltage transformer
V test Test voltage
V meas Measured burden voltage
I meas Measured burden current
Phase Measured phase displacement
Burden (@ Vtest) Burden based on Vtest
Burden (@ Usn) Burden based on Usn
cosφ Cos φ of phase angle
Assessment Measurement assessment
220 OMICRON
Rotating machine tests
Primary Test Manager supports the following tests for rotating machines:
• DC Winding Resistance Stator/Rotor
• Stator Winding PF & CAP1
• Partial Discharge HV Source
1. Test name depends on the Profile selected in the Settings (see "Profiles" on page 26).
OMICRON 221
CPC 100 PTM User Manual
DANGER
Death or severe injury caused by high voltage or current
Injecting direct current into test objects with inductive characteristics will charge the winding of the
test object.
► Follow instructions below.
► Refer to the Safety instructions in the CPC 100 User and Reference manuals.
DANGER
Death or severe injury caused by high voltage or current
► Never open the measuring circuit while current flows.
► After a measurement, wait until the test device has discharged completely.
► Ground all terminals of the test object before touching the test setup.
► Short-circuit the terminals before disconnecting the test leads.
► Disconnect cables not used for testing both from the test object and the test device.
Setting Description
Measurement settings
CPC output DC current output of the CPC 100
Test current Current of the test
Result settings
Automatic result ► Select the Automatic result check box to stop the measurement
automatically depending on the tolerance R dev and the settling
time.
222 OMICRON
Rotating machine tests
Setting Description
Corr. factor Temperature correction factor
Assessment settings • Manual
• Automatic
Note: The Assessment settings are only available for a three-phase
asset in the DC Winding Resistance Stator test.
Assessment
Note: This section is only visible if Automatic is selected under Assessment settings.
Primary Test Manager supports automatic assessment of the DC Winding Resistance Stator test.
Parameter Description
Limit schema Customer-specific limits for automatic assessment
► To save the selected limit schema as the default setting for all future jobs, click Set as default.
During the automatic assessment, Primary Test Manager compares the R corr values of all phases to
each another. If a deviation is greater than the default limit the assessment status is set to Fail.
You can assess the test:
• During measurement
To assess the test while the measurements are running, select the Assess during measurements
check box.
• Manually after all measurements have been finished
To assess the test manually, click Assess measurements.
After a measurement has been assessed, the test settings and the assessment parameters are locked.
To change the test data, the test settings or the assessment parameters, click Clear all assessments.
You can overwrite the Primary Test Manager automatic assessment manually. To do so, click the arrow
next to the automatic assessment, and then select a manual assessment from the list.
The automatic assessment is stored in the Assessment box for documentation purposes.
OMICRON 223
CPC 100 PTM User Manual
Data Description
Phase Measured phase
R meas Measured resistance
R dev Deviation of two successive measurement results at the time the test was
stopped
R corr Corrected measured resistance
Time Time between the start and stop of a measurement
I DC Measured current
V DC Measured voltage
Assessment Measurement assessment per tap position
Note: If you did not select the Automatic result check box to stop the measurement automatically, you
can stop the measurement by clicking the Keep results button.
224 OMICRON
Rotating machine tests
DANGER
Death or severe injury caused by high voltage or current
The high-voltage cable is double-shielded and therefore safe. However, the last 50 cm (20 inches)
of this cable have no shield.
► Avoid any direct contact of this part of the cable to ground potential or to any objects.
► During a test, consider this part of the cable a live wire and, due to the high-voltage, life-
threatening.
Note: This test name depends on the Profile selected in the Settings (see "Profiles" on page 26):
• IEEE standard: Stator Winding PF & CAP
• IEC standard: Stator Winding DF & CAP
• Custom profile: for example Stator Winding Tanδ & CAP
In this section, the term Stator Winding PF & CAP will be used.
OMICRON 225
CPC 100 PTM User Manual
Capacitance and power factor (PF) measurements are conducted to identify defects that affect the
condition of the machine insulation.
For a suitable test object compensation using the CP CR, the capacitance value is required and can be
measured or entered in the Hardware configuration section of the test.
Setting Description
Detect capacitance ► Click Yes if you do not know the test objects capacitance and want
to measure it before testing.
This will add the Capacitance tab to the Measurements section. In
this tab you can perform a measurement to detect the capacitance
before performing the Stator Winding test.
Setting Description
Measurement settings
Test frequency ► Enter the test frequency.
This value is also used for the calculation of the necessary number of
CP CR coils.
Sweep settings
Voltage sweep (tip-up) Sweep profiles:
• None: no voltage sweep
• Umax = L - Gnd (the former OMICRON expertise profile)
• Umax = L - L
• Umax = 1.2*(L - L)
• Umax = 1.2*(L - Gnd)
226 OMICRON
Rotating machine tests
Table 17-6: Stator Winding PF & CAP – Settings and conditions (continued)
Setting Description
► Click the pen button to create a voltage sweep profile.
Sweep profiles
► Add up to 30 measurement points with individual output voltages.
Double-click a value to change it.
► Mark a favorite to use it as the default sweep profile for future
tests.
Note: The predefined profiles cannot be edited or deleted.
The default voltage sweep profile is Umax = L - Gnd.
Enable shield check ► Activate the check box if you want the CPC 100 to check whether the
shield of the high-voltage cable is connected.
Use beeper ► Select the Use beeper check box if you want to use the CP TD’s
beeper during the measurement.
Test conditions
Custom test conditions ► Activate the check box to set test conditions differing from the global
test conditions.
Setting Description
Capacitance1
Measurement Measurement procedure
Test mode Test mode according to IEEE Std 62-1995
Sweep Disabled for the capacitance measurement
V test Test voltage
Freq. Test frequency
OMICRON 227
CPC 100 PTM User Manual
Setting Description
V out Measured output voltage
I tot Measured current in the parallel resonance circuit
Watt losses Measured losses
Measurements
Measurement Measurement procedure
Test mode Test mode according to IEEE Std 62-1995
Sweep Setting selected under Sweep settings
V test Test voltage
I out Measured output current
Watt losses Measured losses
228 OMICRON
Rotating machine tests
DANGER
Death or severe injury caused by high voltage or current
The high-voltage cable is double-shielded and therefore safe. However, the last 50 cm (20 inches)
of this cable have no shield.
► Avoid any direct contact of this part of the cable to ground potential or to any objects.
► During a test, consider this part of the cable a live wire and, due to the high-voltage, life-
threatening.
The CPC 100 is used as a high-voltage sweep source for parallel Partial Discharge (PD) measurements
on a rotating machine with the MPD.
Table 17-8: Partial Discharge HV Source – Wiring diagram
Setting Description
Wiring diagram ► Select a wiring diagram via the respective radio button:
• HV source wiring
• Measurement wiring
Note: The measurement wiring changes according to the quadripole
setting under Coupling in the measurement settings.
There is no wiring diagram shown for calibration.
OMICRON 229
CPC 100 PTM User Manual
For a suitable test object compensation using the CP CR, the capacitance value is required and can be
measured or entered in the Hardware configuration section of the test.
Setting Description
Detect capacitance ► Click Yes if you do not know the test objects capacitance and want
to measure it before testing.
This will add the Capacitance tab to the Measurements section. In
this tab you can perform a measurement to detect the capacitance
before performing the Partial Discharge HV Source test.
230 OMICRON
Rotating machine tests
Setting Description
Measurement settings
Test frequency ► Enter the test frequency.
This value is also used for the calculation of the necessary number of
CP CR coils.
Sweep settings
Voltage sweep (tip-up) Sweep profiles:
• None: no voltage sweep
• Umax = L - Gnd (the former OMICRON expertise profile)
• Umax = L - L
• Umax = 1.2*(L - L)
• Umax = 1.2*(L - Gnd)
OMICRON 231
CPC 100 PTM User Manual
Setting Description
► Click the pen button to create a voltage sweep profile.
Sweep profiles
► Add up to 30 measurement points with individual output voltages.
Double-click a value to change it.
► Mark a favorite to use it as the default sweep profile for future
tests.
Note: The predefined profiles cannot be edited or deleted.
The default voltage sweep profile is Umax = L - Gnd.
Enable shield check ► Activate the check box if you want the CPC 100 to check whether the
shield of the high-voltage cable is connected.
Use beeper ► Select the Use beeper check box if you want to use the CP TD’s
beeper during the measurement.
Test conditions
Custom test conditions ► Activate the check box to set test conditions differing from the global
test conditions.
Setting Description
V-Cal for MPD
Calibration voltage Voltage used for MPD voltage calibration
Start ► Click the Start button to start the voltage output used for MPD
calibration.
Capacitance1
Measurement Measurement procedure
Test mode Test mode according to IEEE Std 62-1995
232 OMICRON
Rotating machine tests
Setting Description
Sweep Disabled for the capacitance measurement
V test Test voltage
Freq. Test frequency
Step duration Step duration of the voltage sweep
V out Measured output voltage
Measurements
Measurement Measurement procedure
Test mode Test mode according to IEEE Std 62-1995
Sweep Setting selected under Sweep settings
V test Test voltage
Freq. Test frequency
Step duration Step duration of the voltage sweep
V out Measured output voltage
Clear results ► Click the Clear results button to clear all results in the measurement
tab.
OMICRON 233
CPC 100 PTM User Manual
Setting Description
V(t), Q(t) Voltage and Q values over time diagram for each measurement
channel
Q(V) QIEC value over the voltage
Advanced statistics
Test voltage Injected test voltage
Channel Measurement channel
QWTD Apparent charge of all PDs over a given time (also known as QIEC)
QPeak Largest absolute charge of any PD event seen during the evaluation
interval
QAvg Average QIEC (or QWTD) value for the evaluation interval
234 OMICRON
Rotating machine tests
Setting Description
Stator
Rated voltage L-L Rated voltage of the stator
Rated frequency Rated frequency of the stator
Calculation based on Whether calculations are based on the winding scheme or dimensions
of the stator
Winding Scheme
Bars per slot The number of bars per slot in the machine
Turns per coil The number of turns per coil according to the winding scheme
Parallel circuits per phase The number of parallel circuits per phase according to the winding
scheme
Dimensions
No. of ventilation ducts The number of ventilation ducts in the machine
Width of ventilation duct The width of the ventilations ducts in the machine
Setting Description
Measurement settings
Stator excitation The desired stator excitation in percent of the rated flux
OMICRON 235
CPC 100 PTM User Manual
Setting Description
Test frequency The frequency used for stator excitation
Cable accessory Select whether an 8-wire or 2-wire cable is being used
Number of turns The number of physical turns of the cable accessory around the stator
Output voltage Output voltage is calculated automatically based on the entered values
Setting Description
Magnitude calibration
Start calibration Prepares the CPC 100 current output to inject a known current to the
measurement calibration slot
Set ratio manually If the calibration factor is already known, input the value manually
Measurement settings
Measurement Select whether the test is being performed automatically (with rail) or
manually (with handle)
Recording mode Select whether to:
► Record new slot;
or if an existing slot is selected:
► Record to selected slot
► Insert new slot below selected
Slot settings
Start with slot Enter slot number for first measurement
Counting direction Select whether slots will be measured in ascending or descending
order
Recording settings1
Rail position1 Offset distance between the beginning of the stator core and the zero
point of the ruler on the rail unit
Start recording at1 Position on the ruler where the recording of measurement values
should start
Position2 Position in the stator to which the currently measured result in manual
mode refers to.
Increment2 Increment (in mm) at which the position is increased after each
recorded measurement
Start Start the test
Record slot Record the measurement data based on the selected recording mode
236 OMICRON
Rotating machine tests
Setting Description
Keep result Keep manually measured slot data
New slot Start manual measurement of a new slot
Delete selected slot Deletes the currently selected slot
Show all Show all slot data on the graph
Hide all Hide all slot data from the graph
Combine charts Combine Quad and InPhase graphs
Stack charts Separate Quad and InPhase graphs
Default zoom Begin graph scale at 0mm and end at 1760mm
Zoom to fit Zoom graph view to display only data range
Fit X-axis Zoom graph view to display only data range along the x-axis
Fit Y-axis Zoom graph view to display only data range along the y-axis
Heatmap
Current threshold The upper current limit as indicated on the heatmap
Tolerance Define which percentage of the set threshold the heatmap displays the
middle color of the gradient
Color scheme Select preferred color scheme
1.Only available in automatic measurement mode.
OMICRON 237
CPC 100 PTM User Manual
238 OMICRON
Grounding system tests
WARNING
Death or severe injury caused by high voltage or current possible
The auxiliary current probe carries-life threatening voltages during the test.
In case of an error, unexpected high voltages can occur at any time. The step voltage around the
auxiliary current probe can also be quite high.
► Always press the emergency stop button before working with the auxiliary current probe or the
associated wiring cables.
► Mark an area of 10 m (30 ft) around the electrode as dangerous zone and position a guard
outside this area to keep people from entering the dangerous zone.
NOTICE
If the desired current cannot be reached or an overload occurs, the contact resistance of the auxiliary
current probe to the soil may be too high. Place several electrodes at a distance of a few meters and
connect them all together to keep the resistance to the soil low. This also reduces the hazard due to
high voltages around the electrode.
Setting Description
Measurement settings
Rated frequency ► Select the rated frequency of the substation
• 60 Hz
• 50 Hz
• 16.7 Hz
Input source ► Select the source which is used for the injection of the test current.
• CPC 100 and 6A
• CPC 100 and CP CU1
Assessment settings
OMICRON 239
CPC 100 PTM User Manual
Setting Description
Reduction factor Ratio between current contributing to the ground potential rise and total
injected current
The reduction factor can be determined with a measurement.
Injected currents
Frequency Frequency of the test current from the input signal
Current Test current from the input source
240 OMICRON
Grounding system tests
Setting Description
Add ► Click Add to create a measurement point
Clear all ► Click Clear all to clear all measurement results in the overview
Setting Description
Previous / Next ► Click Previous or Next to switch to the previous or next
measurement respectively
Note: You can also select measurements directly by clicking on the
respective measurement under Overview.
Settings
Distance Distance to the reference point
► Select the Enter distance manually check box to enter the distance
manually.
► Select the Show on map check box to show the measurement on
the map.
Results
Frequency Measured frequency
Voltage Measured voltage
Impedance Impedance at nominal frequency
OMICRON 241
CPC 100 PTM User Manual
Setting Description
Voltage Voltage calculated in relation to the maximum current to earth
Map
Map Map showing the measurement traces
Drag the map by holding left mouse button and moving the mouse
Zoom in or out with the mouse wheel
Draw injection line ► Click Draw injection line to draw a new injection line on the map.
► Click Save injection line to save the injection line.
or
► Click Delete injection line to delete the injection line.
Plot
FFT The FFT of the selected measurement
Impedance Impedance over distance in relation to the distance to the reference
point
Step voltage Step voltage for all valid measurements
Voltage Voltage over distance in relation to the distance to the reference point
242 OMICRON
Grounding system tests
WARNING
Death or severe injury caused by high voltage or current possible
The auxiliary current probe carries life threatening voltages during the test. The step voltage
around the auxiliary current probe can also be quite high.
► Always press the emergency stop button before working with the auxiliary current probe or the
associated wiring cables.
► Mark an area of 10 m (30 ft) around the electrode as dangerous zone and position a guard
outside this area to keep people from entering the dangerous zone.
NOTICE
If the desired current cannot be reached or an overload occurs, the contact resistance of the auxiliary
current probe to the soil might be too high. Place several electrodes at a distance of a few meters and
connect them all together to keep the resistance to the soil low. This also reduces the hazard due to
high voltages around the electrode.
OMICRON 243
CPC 100 PTM User Manual
Table 18-4 Step & Touch voltage measurement – Settings and condtions
Setting Description
Measurement settings
Rated frequency ► Select the rated frequency of the substation
• 60 Hz
• 50 Hz
• 16.7 Hz
Input source ► Select the source which is used for the injection of the test current.
• CPC 100 and 6A AC
• CPC 100 and CP CU1
• COMPANO 100 and V OUT
• COMPANO 100 and I OUT
244 OMICRON
Grounding system tests
Table 18-4 Step & Touch voltage measurement – Settings and condtions
Setting Description
Perm. body current
Perm. touch voltage
Perm. hand-to-hand
voltage1
1. Only for Standard IEEE 80/81
Table 18-5: Step & Touch voltage measurement – Overview
Setting Description
Add ► Click Add to create a measurement point
Clear all ► Click Clear all to clear all measurement results in the overview
OMICRON 245
CPC 100 PTM User Manual
Setting Description
Previous / Next ► Click Previous or Next to switch to the previous or next
measurement respectively
Note: You can also select measurements directly by clicking on the
respective measurement under Overview.
Settings
Location ID Location ID for reporting
Location Description field for the measurement location
Input impedance Input impedance on the HGT1
• Touch Voltage
Scenario1
• Step Voltage
• Hand-to-hand Voltage
Results
Frequency Measured frequency
Voltage Measured voltage
Voltage Voltage at nominal frequency
Assessment settings
Standard ► Select the standard applicable for the measurement
• EN 50522
• IEEE 80/81
Add. considered resistance Additional resistance considered for the fault scenario, depending on
the applicable standard
Body current limit acc. ► Select the body current limit according to:
• EN 50522
• EN 50522
• Custom
• IEEE 80/81
• Dalziel 50 kg
• Dalziel 70 kg
• Biegelmeier
• Custom
246 OMICRON
Grounding system tests
Setting Description
Map Map showing the measurement locations
Drag the map by holding left mouse button and moving the mouse
Zoom in or out with the mouse wheel
Plot
FFT The FFT of the selected measurement
Trend Trending diagram for all measurements
1. Only available when selecting Standard IEEE 80/81 under Assessment Settings
2. Only for Standard EN 50522
OMICRON 247
CPC 100 PTM User Manual
248 OMICRON
Optional manual tests
Setting Description
Test conditions
Test object temperature Temperature of the test object
Custom test conditions ► Activate the Custom test conditions check box to set test conditions
differing from the global test conditions.
Correction factors
Temperature correction ► Select the Temperature correction check box to activate
temperature correction.
OMICRON 249
CPC 100 PTM User Manual
Setting Description
Test data To import a file containing test data:
► Click the Add button to browse your PC and add data from a file.
To directly import data from a measurement file:
► Open the file on your computer.
► In the file press CTRL+A to mark all content, then press CTRL+C to
copy.
► In Primary Test Manager press Paste from clipboard.
The results may take a few seconds to load.
Measurement Name or number of the measurement
PI Polarization index
DAR Dielectric absorption ratio
Time Time at which the given values were recorded
Voltage
Voltage and current values recorded at the Time specified in the first
V DC
column
I DC
250 OMICRON
Optional manual tests
DANGER
Death or severe injury caused by high voltage or current
► Never use the CP TD without a solid connection to ground with at least 6 mm².
► Use a ground point as close as possible to the test object.
► Make sure to position the test object in the high-voltage area.
DANGER
Death or severe injury caused by high voltage or current
The high-voltage cable is double-shielded and therefore safe. However, the last 50 cm (20 inches)
of this cable have no shield.
► Avoid any direct contact of this part of the cable to ground potential or to any objects.
► During a test, consider this part of the cable a live wire and, due to the high-voltage, life-
threatening.
The following table describes the test settings of the Manual Tan Delta test.
Table 19-3: Manual Tan Delta test – Test settings
Setting Description
Measurement settings
Default frequency Default test frequency
Show results ► Select the results you want to display from the list.
All measurement results are stored and can be displayed by selecting
them from the list.
OMICRON 251
CPC 100 PTM User Manual
Setting Description
Device settings
Tan delta device ► Select your CP TD device:
• CP TD1
• CP TD12
• CP TD15
Enable shield check ► Select the Enable shield check check box if you want the CPC 100
to check whether the high-voltage cable shield is connected.
Use beeper ► Select the Use beeper check box if you want to use the CP TD’s
beeper during the measurement.
Test conditions
Ambient temperature Ambient temperature on site
Test object temperature Temperature of the test object
Humidity Relative ambient humidity
Weather Weather during the test
Correction factors
Temperature correction ► Select the Temperature correction check box to activate
temperature correction.
The following tables describe the measurement data of the Manual Tan Delta test.
Table 19-4: Manual Tan Delta test – Common measurement data
Data Description
252 OMICRON
Optional manual tests
Table 19-4: Manual Tan Delta test – Common measurement data (continued)
Data Description
Freq. Test frequency
V out Measured output voltage
I out Measured output current
Assessment Measurement assessment
1. A value is assigned after a measurement has been added.
2. The data is by default empty. You can type your own measurement name.
Note: To add a measurement point to a measurement, click Add meas. point. To add a measurement
to the Manual Tan Delta test, click Add measurement.
Table 19-5: Manual Tan Delta test – Common measurement data (PF, Cap., watt losses)
Data Description
Watt losses Measured losses
Data Description
Z phase Phase of the measured impedance
Z abs Magnitude of the measured impedance
OMICRON 253
CPC 100 PTM User Manual
Data Description
Reactive power Measured reactive power Q
Apparent power Measured apparent power S
Table 19-8: Manual Tan Delta test – Measurement data (Cp, Rp)
Data Description
Cp Measured parallel capacitance
Rp Measured parallel resistance
Table 19-9: Manual Tan Delta test – measurement data (Cp, quality factor)
Data Description
Cp Measured parallel capacitance
Quality factor Measured quality factor
Table 19-10: Manual Tan Delta test – measurement data (Ls, Rs)
Data Description
Ls Measured series inductance
Rs Measured series resistance
Table 19-11: Manual Tan Delta test – measurement data (Ls, quality factor)
Data Description
Ls Measured series inductance
Quality factor Measured quality factor
254 OMICRON
Optional manual tests
19.3 Quick
Quick is the standard test with which all outputs and inputs of the CPC 100 can be freely configured. This
makes it possible to perform a manual test configuration.
Quick only supports the CPC 100 test set.
DANGER
Death or severe injury caused by high voltage or current
Together with the test object’s capacitance, the leakage inductance of the CPC 100’s internal
output transformer forms a series resonant circuit. Especially at frequencies > 50 / 60 Hz this may
result in voltage superelevation.
► When testing capacitive test objects using voltages ≥ 500 V, make sure that the test object’s
capacitance does not exceed 25 nF.
DANGER
Death or severe injury caused by high voltage or current
► Never use Quick in combination with a DC output on test objects with highly capacitive
characteristics.
► Mind the danger of test object’s charged capacitance. Before connecting or disconnecting any
leads, use a grounding/discharging rod
► to discharge all terminals of the test object.
► to connect all terminals of the test object to ground and short-circuit all capacitances.
DANGER
Death or severe injury caused by high voltage or current
► Never use Quick to measure the resistance of windings with highly inductive characteristics.
Turning off the DC source results in life-threatening voltage levels.
► For this kind of measurement only use the special winding resistance tests.
Setting Description
Outputs
Output mode Select the CPC 100 output.
Voltage/Current Start value of the CPC 100 output
Note: Output value is directly applied after pressing the I/0 button on the
CPC 100.
Max. Voltage/Current Maximum output voltage or current value selectable during the test
OMICRON 255
CPC 100 PTM User Manual
Setting Description
Measurements
Signal mode ► Select AC or DC as input signal mode (for 10A I AC/DC only).
Setting Description
Trigger ► Select the Trigger check box to activate follow-up actions in defined
event occurrences.
256 OMICRON
Optional manual tests
Setting Description
Action Define the action in the occurrence of a trigger event
• Continue: additional measurement point is created after the
occurrence of a trigger event
• Stop: trigger event stops the measurement
Noise suppression
Frequency selective Applies a digital filter to the measurement input channels that attenuates
measurement signal components above and below test frequency
Avoid test frequency Performs two measurements with frequencies above and below the
configured test frequency. The result at the test frequency is then
calculated by interpolation.
Note: Cannot be combined with triggers.
Setting Description
Reading 1 Define reading channel 1 for measured quantities
Reading 2 Define reading channel 2 for measured quantities
OMICRON 257
PTMate Getting Started
Support
When you are working with our products we want to provide you with the greatest
possible benefits. If you need any support, we are here to assist you!
OMICRON electronics GmbH, Oberes Ried 1, 6833 Klaus, Austria, +43 59495
254 OMICRON
ENU 1036 06 11