Quality Control Pathologists Shewhart's Individuals Control Chart
Quality Control Pathologists Shewhart's Individuals Control Chart
Quality Control
Intralab QC
Interlab QC
- It involves the proficiency testing programs that periodically provide samples of unknown
concentrations to participating clinical laboratories.
- It is important in maintaining long-term accuracy of the clinical methods
- It is also used to determine state-of-the art interlaboratory performance.
Levey-Jennings Chart
QC procedure(s) to be implemented
Each of the two control materials will be analyzed once per run, providing a total of two control
measurements per run. Control status will be judged by either the 1 2s or 13s rule. These rules are
defined as follows:
12s refers to the control rule that is commonly used with a Levey-Jennings chart when the
control limits are set as the mean plus/minus 2s. In many laboratories, this rule is used to reject a
run when a single control measurement exceeds a 2s control limit.
13s corresponds to a Levey-Jennings chart having control limits set as the mean plus/minus
3s. An analytical run is rejected when a single control measurement exceeds a 3s control limit.
The 12s rule is very commonly used today, and while it provides high error detection, the use of 2s
control limits gives an expected high level of false rejections. The 1 3s rule provides an alternative QC
procedure that has lower false rejections, but also lower error detection. In this exercise, you will see
how to apply both QC procedures and also get a feel for the difference in their performance.
Two sets of control limits will be needed to implement the rules described above. The first set uses
2s control limits (for implementation of the 12s rule) calculated as the mean plus or minus 2 times the
standard deviation. The second set uses 3s control limits (for implementation of the 1 3s rule)
calculated as the mean plus or minus 3 times the standard deviation.
For this example, Control 1 has a mean of 200 and a standard deviation of 4 mg/dL.
The upper control limit would be:
200 + 2*4, which is 208 mg/dL.
The lower control limit would be:
200 - 2*4, or 192 mg/dL.
- A cumulative sum (CUSUM) chart is a type of control chart used to monitor small
shifts in the process mean. It uses the cumulative sum of deviations from a target.
- The CUSUM chart plots the cumulative sum of deviations from the target for
individual measurements or subgroup means.
When designing a CUSUM chart it is necessary to consider the average run length and shift
to be detected. Extensive guidance is available on suitable parameters (NIST 2012,
Montgomery 2012).
It is possible to modify the CUSUM, so it responds more quickly to detect a process that is
out-of-control at start-up. This modification is done using an initial CUSUM equal to half of
the h parameter.
Like the EWMA, CUSUM is sensitive to small shifts in the process mean but does not match
the ability of a Shewhart chart to detect larger shifts. For this reason, it is sometimes used
together with a Shewhart chart (Montgomery 2012).
-
Trend Shift
Specificity
- The specificity of a test is its ability to designate an individual who does not have a
disease as negative.
- "Analytical specificity" refers to the ability of an assay to measure on particular
organism or substance, rather than others, in a sample.
Sensitivity
Accuracy
Precision
Random Error
- Random errors occur because of random and inherently unpredictable events in the
measurement process.
Systematic Error
- -Systematic errors occur when there is a problem in the measurement system that
affects all measurements in the same way.