OFV-5000 Xtra Laser Vibrometer: Xtra Sensitivity and Versatility

The OFV-5000 Xtra Laser Vibrometer is a non-contact sensor that uses laser vibrometry to measure vibration, displacement, and acceleration of surfaces. It has high optical sensitivity allowing it to gather data from dark, biological, rotating, or moving objects. It comes with interchangeable lenses and decoders to provide versatile measurement of objects from micrometers to over 100 mm in size over a wide frequency bandwidth up to 24 MHz.

Uploaded by

vnetawz
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
171 views12 pages

OFV-5000 Xtra Laser Vibrometer: Xtra Sensitivity and Versatility

The OFV-5000 Xtra Laser Vibrometer is a non-contact sensor that uses laser vibrometry to measure vibration, displacement, and acceleration of surfaces. It has high optical sensitivity allowing it to gather data from dark, biological, rotating, or moving objects. It comes with interchangeable lenses and decoders to provide versatile measurement of objects from micrometers to over 100 mm in size over a wide frequency bandwidth up to 24 MHz.

Uploaded by

vnetawz
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 12

OFV-5000 Xtra Laser Vibrometer

The OFV-5000 Xtra Vibrometer Controller in combi- An alternative exchangeable fiber lens as an add-
nation with the new MLV-I-120 Sensor Head is designed on to the MLV-I-120 head enables a dual purpose
to flexibly measure vibrational velocity, displacement use for small objects and facilitates overhead
and acceleration. The Xtra optical sensitivity gathers measurements. The compactness of this robust
high-fidelity data from all surfaces – even on dark, sensor with different lens options leads to a versatile
biological, rotating or moving sample objects. use and an easy integration into test setups.

This non-contact approach solves challenging appli-


cations like NDT, biomedical, “long distance” displace-
ment measurements, quasi-static displacement mea-
surement and shaker control.

Highlights
■■ Non-contact vibration sensor

with Xtra sensitivity


■■ High-fidelity data from all

surfaces – even on dark,


biological or moving objects
■■ From µm-sized to large, distant

objects
■■ High dynamic range with wide

bandwidth up to 24 MHz
■■ Remote operation keeps laser

precisely focused

OFV-5000 Xtra Laser Vibrometer


Xtra sensitivity and versatility
Datasheet
Technical data
Models
Version Max. Max. Best velocity Max. Best Acceleration Decoders
frequency velocity resolution1 displacement displacement output
resolution1 O: Option;
–: Not available
M 2.5 MHz 25 m/s 0.04 µm s /√Hz
-1
– – O VX-09

O VX-09, VX-08
1
Best value of the
R 2.5 MHz 25 m/s 0.004 µm s /√Hz –
-1
– typical resolution
of the included
V 24 MHz 25 m/s 0.04 µm s-1/√Hz ±188 nm 0.2 pm/√Hz – VX-09, DX-300 decoders

MD 2.5 MHz 25 m/s 0.04 µm s /√Hz


-1 ±125 mm 0.75 pm/√Hz O VX-09, DX-900

RD 2.5 MHz 25 m/s 0.004 µm s /√Hz ±125 mm


-1
0.75 pm/√Hz O VX-09, VX-08,
DX-900

VD 24 MHz 25 m/s 0.04 µm s-1/√Hz ±125 mm 0.2 pm/√Hz – VX-09, DX-900,


DX-300

2
Depends on decod-
Metrological specifications er configuration.
3
Displacement or
Analog signal outputs BNC, ±10 V2: acceleration signal,
Velocity signal depends on decod-
er configuration (see
Displacement signal table "Models").
AUX output3 4
Notice: frequency
range, max. velocity
Digital signal output S/PDIF, 24 bit, 48/96 kSa/s (available on request, only for VX-08 decoder) and measurement
range depend on
Frequency range4 DC to 24 MHz decoder configura-
tion (see below).
Max. velocity4 ±25 m/s

Filters High pass filter: 100 Hz, off


Low pass filter: 5 kHz, 20 kHz, 100 kHz, off

Tracking filter2 3 settings: slow, fast, off

Signal level Bargraph on touchscreen and on sensor head


Output as DC voltage signal (BNC, 0 ... 5 V)

PC interface RS-232, remote control of the instrument settings

Velocity decoders
Decoder Description No. of ranges Typical Max. Frequency
resolution5, 6 velocity range
VX-08 Digital high-resolution 8 0.004 µm s-1/√Hz 1.25 m/s 0 Hz - 25 kHz
velocity decoder
VX-09 Digital velocity decoder 14 0.04 µm s-1/√Hz 25 m/s 0 Hz - 2.5 MHz

5
Noise-limited reso-
Displacement decoders lution in the smallest
measurement range.
Decoder Description No. of ranges Resolution5 Max. Frequency The noise-limited
displacement range resolution is defined
as the signal ampli-
DX-300 Analog displacement decoder 1 0.2 pm/√Hz ±188 nm 30 kHz - 24 MHz tude (RMS) at which
the signal-to-noise
for ultrasonic applications ratio is 0 dB with
1 Hz spectral reso­
DX-900 Broadband digital displacement 16 0.75 pm/√Hz ±125 mm 0 Hz - 2.5 MHz lution, measured on
decoder a mirror.

6
The attainable res-
Acceleration decoder olution is frequency-
dependent. The
Decoder No. of ranges Max. Frequency typical value refers
acceleration range to the center of the
operating frequency
AX-100 8 250,000 m/s2 0.5 Hz - 20 kHz range.

2
General specifications
OFV-5000 Xtra Controller MLV-I-120 Sensor Head Xtra

Interface/display 7" color touchscreen with


interactive menu guidance

Dimensions 19" rack mounting, W x H x L: 135 x 100 x 383 mm


W x D x H: 450 x 360 x 150 mm
(19", 84 HP/3U) without angle brackets

Weight 10 kg 4.55 kg

Protection class IP20 IP40

Operating temperature +5 °C … +40 °C (41 °F ... 104 °F)

Storage temperature -10 °C … +65 °C (14 °F ... 149 °F)

Relative humidity max. 80%, non-condensing

Power supply 100...240 VAC ±10%, 50/60 Hz

Power consumption max. 100 VA

Optical specifications: MLV-I-120 Sensor Head Xtra

Laser type Measurement laser: invisible (IR), wavelength 1550 nm, output power <10 mW
Targeting laser: visible (green), wavelength 510 - 530 nm, effective output power
< 1 mW

Laser class Class 2, eye-safe, with both lasers in operation

Focus Auto focus, remote focus, manual focus

Maximum stand-off distance Up to 100 m (with MLV-O-LRI long range front lens, surface dependent)

1
Included with MLV-
Working distance and laser spot size O-FMI-02 Fiber Lens
(IR).
Front lenses Fiber heads for MLV-O-FMI-02 2
Optional available
for MLV-O-FMI-02
MLV-O-SRI MLV-O-LRI MLV-O-100 1
MLV-O-110 2
Fiber Lens (IR).
short range long range Mini Fiber Head Micro Spot Fiber Head

Focal length [mm] 29 70 16 35

Min. stand-off distance [mm] 25 380 60 56±2

Exit beam diameter (1/e2) [mm] 2...4.5 11...12.4 3.3...4.3 14


Typical spot size in μm at
25 mm 48 – – –
50 mm 77 – – –
56 mm 81 – 8
60 mm 84 – 28 –
75 mm 91 – 37 –
100 mm 97 – 53 –
300 mm 150 – 180 –
380 mm 184 60 224 –
500 mm 236 81 295 –
1,000 mm 448 171 608 –
2,000 mm 906 349 – –
5,000 mm distance 2,766 898 – –

Each additional meter add [μm] – +183 – –

3
Options and accessories
Tripods

VIB-A-T02 Standard Tripod Easy targeting on the object under test

VIB-A-T05 Tripod For precise pointing of the sensor head. The geared
with Geared Pan/Tilt Head pan/tilt head allows quick coarse adjustment and fine
adjustment in 3 axes

Optical accessories

MLV-O-SRI SR Front Lens (IR) Short Range front lens for measuring at short working
distances (highest depth of focus)

MLV-O-LRI LR Front Lens (IR) Long Range front lens for measuring at long working
distances

MLV-O-FMI-02 Fiber Lens (IR) 2m Flexible measurements with 2 m fiber cable on


small objects or where space is restricted. Includes
MLV-O-100 Mini Fiber Head and VIB-A-CAS08
Transportation Case

MLV-O-100 Mini Fiber Head Small fiber head (10 mm diameter) with a laser
spot size down to 28 µm for MLV-O-FMI-02 Fiber
Lens (IR) 2 m

MLV-O-110 Micro Spot Fiber Head Small fiber head (24 mm diameter) with a laser spot
size of 8 µm for MLV-O-FMI-02 Fiber Lens (IR) 2 m

4
Positioning stages
VIB-A-P35 Precision 4-Axes Stage XY-traverse stage featuring 18 mm travel with +/- 5°
tip/tilt function for positioning a single 10 mm outer
diameter Mini Fiber Head.

OFV-036 Tip/Tilt Precision Stage For positioning a single 10 mm outer diameter Mini Fiber
Head. Travel range ±5°.

Transportation cases
VIB-A-CAS07 Transportation Case Robust transportation case for the sensor head
for MLV-I-120 Sensor Head Xtra (included with sensor head)

VIB-A-CAS02 Transportation Case Robust transportation case for the OFV-5000 Xtra
OFV-5xxx Controller

Please contact Polytec’s application and sales engineers who will help to choose
the appropriate accessories like VibSoft data acquisition and analysis software.

5
Velocity decoders
VX-09 digital velocity decoder
Measurement Full scale output Typical Frequency Max. Max.
range1 (peak)2 resolution3 range acceleration linearity error

mm/s/V m/s µm s-1/√Hz MHz g %


12.5 0.125 0.04 0 - 0.1 8,000 0.5
25 0.25 0.1 0 - 0.25 40,000 0.5
50 0.5 0.35 0-1 320,000 0.5
125 1.25 0.55 0 - 1.5 1,200,000 0.5
250 2.5 0.55 0 - 1.5 2,400,000 0.5
500 5 1.1 0 - 2.5 8,000,000 0.5
1,250 12.5 1.4 0 - 2.5 20,000,000 0.5
2,500 25 2 0 - 1.5 24,000,000 0.5
Measurement ranges 50 ... 2,500 mm/s/V each feature an additional low pass range with 250 kHz frequency range, higher resolution but reduced max. acceleration.
1

The full scale values correspond to the maximum output voltage of 10 Vpeak.
2

The noise-limited resolution is defined as the signal amplitude (rms) at which the signal-to-noise ratio is 0 dB with 1 Hz spectral resolution, for frequencies above
3

1 kHz measured on a mirror. The attainable resolution is frequency-dependent. The typical value refers to the center of the operating frequency range.

V peak Range diagram


100 m/s
Max. Detectable
Acceleration: 24,000,000 g
25 m/s
2.5 m/s/V
12.5 m/s
10 m/s 1.25 m/s/V
■■ Multi-purpose high frequency 5 m/s
0.5 m/s/V
bandwidth digital decoder 2.5 m/s
250 mm/s/V
1.25 m/s
with high precision and high 1 m/s 125 mm/s/V
500 mm/s
resolution 250 mm/s
50 mm/s/V
25 mm/s/V
125 mm/s
■■ True DC capability 100 mm/s 12.5 mm/s/V
Ranges
■■ Well balanced properties
regarding bandwidth, 10 mm/s

resolution and velocity


limits make it the decoder of 1 mm/s Max. Vibrational Frequency: 2.5 MHz
choice for most measurement
Lower Frequency Limit: 0 Hz

applications
100 µm/s
■■ Universal decoder, included
in all models
10 µm/s

1 µm/s

Min. Detectable Velocity: 0.04 µm/s


1 MHz

10 MHz

Frequency
1.5 MHz
2.5 MHz
1 Hz

10 Hz

100 Hz
0.1 Hz

1 kHz

10 kHz

100 kHz
250 kHz

6
VX-08 digital high-resolution velocity decoder
Measurement Full scale output Typical Frequency Max. Max.
range (peak)1 resolution2 range acceleration linearity error

mm/s/V m/s µm s-1/√Hz kHz g %


0.5 0.005 0.004 0-5 16 0.1
1.25 0.0125 0.004 0-5 40 0.1
2.5 0.025 0.004 0 - 10 160 0.1
5 0.05 0.01 0 - 20 640 0.1
12.5 0.125 0.015 0 - 20 1,600 0.1
25 0.25 0.03 0 - 20 3,200 0.1
50 0.5 0.06 0 - 25 8,000 0.1
125 1.25 0.15 0 - 25 20,000 0.1
1
The full scale values correspond to the maximum output voltage of 10 Vpeak.
2
The noise-limited resolution is defined as the signal amplitude (rms) at which the signal-to-noise ratio is 0 dB with 1 Hz spectral resolution, for frequencies
above 1 kHz measured on a mirror. The attainable resolution is frequency-dependent. The typical value refers to the center of the operating frequency range.

Range diagram
V peak
Max. Detectable
Acceleration: 20,000 g

10 m/s
■■ Digital velocity decoder dedi-
1.25 m/s
125 mm/s/V cated to low frequencies up to
1 m/s
25 kHz with highest resolution
500 mm/s
50 mm/s/V
250 mm/s
25 mm/s/V ■■ Highest optical sensitivity for
125 mm/s
100 mm/s 12.5 mm/s/V measurement on dark surfaces
50 mm/s
5 mm/s/V
25 mm/s
2.5 mm/s/V
■■ True DC capability
12.5 mm/s
10 mm/s 1.25 mm/s/V
5 mm/s
■■ Outstanding linearity and
0.5 mm/s/V
Ranges frequency response
1 mm/s
■■ Included in models R and RD
Max. Vibrational Frequency: 25 kHz
Lower Frequency Limit: 0 Hz

100 µm/s

10 µm/s

1 µm/s

Min. Detectable Velocity: 0.004 µm/s


1 MHz

Frequency
1 Hz

10 Hz

100 Hz
0.1 Hz

5 kHz
1 kHz

20 kHz
10 kHz

25 kHz

100 kHz

7
Displacement decoders
DX-900 broadband digital displacement decoder
Measurement Full scale output Resolution2 Frequency Max.
range (peak-to-peak)1 range3 velocity
µm/V µm nm kHz m/s
0.125 2.5 0.04 0 - 2,500 25
0.25 5 0.08 0 - 2,500 25
0.5 10 0.15 0 - 2,500 25
1.25 25 0.38 0 - 2,500 25
2.5 50 0.75 0 - 2,500 25
5 100 1.5 0 - 2,500 25
12.5 250 3.8 0 - 2,500 25
25 500 7.5 0 - 2,500 25
50 1,000 15 0 - 2,500 25
125 2,500 38 0 - 2,500 25
250 5,000 75 0 - 2,500 25
500 10,000 150 0 - 2,500 25
1,250 25,000 380 0 - 2,500 25
2,500 50,000 750 0 - 2,500 25
5,000 100,000 1500 0 - 2,500 25
12,500 250,000 3800 0 - 2,500 25
1
The full scale values correspond to ±10 V (peak-to-peak) maximum output voltage.
2
The resolution corresponds to the quantization step of approx. 0.3 mV at the analog output.
差不多是16bits AD转换器的量化值(一位符号)
Noise-limited resolution: < 0.75 pm/√Hz in the smallest measurement range. The noise-limited resolution is defined as the signal amplitude (rms) at which the
signal-to-noise ratio is 0 dB with 1 Hz spectral resolution, for frequenciey above 1 kHz measured on a mirror.
3 When a suitable measurement range has been selected for the digital velocity decoder VX-09, which is required for using the DX-900.

Range diagram
Amplitude
(peak)

125 mm Max. Displacement: 125 mm


■■ Broadband displacement 100 mm
50 mm
decoder up to 2.5 MHz 25 mm
12.5 mm
■■ Supplements VX-09 velocity 10 mm
5 mm
decoder for high precision 2.5 mm
displacement measurements 1.25 mm
1 mm
2. 25 m /s/ /s/VV

500 µm
1. 5 m m /s/

with resolutions down to


5
m /s/ V
0. 0 m m /s/V

/s/ V

250 µm
25 5 m m /s/V V
12

40 pm
V

125 µm
505 m m/

100 µm
2 5m
12

m m s/

50 µm
16 measurement ranges
.

■■
25 µm
■■ Included in models MD, RD 12.5 µm
10 µm
and VD 5 µm
2.5 µm
1.25 µm
Lower Frequency Limit: 0 Hz

1 µm
Max. Frequency: 2.5 MHz

100 nm

10 nm
1 MHz

10 MHz

Frequency
1.5 MHz
2.5 MHz
1 Hz

10 Hz

100 Hz

1 kHz

10 kHz

100 kHz
250 kHz

8
DX-300 analog displacement decoder for ultrasonic applications
Measurement Full scale output Resolution2 Frequency
range (peak)1 range
nm/V nm pm/√Hz kHz
125 ±188 < 0.2 30 - 24,000 (−3dB)
The full scale value corresponds to the output voltage swing of ±1.5 V at load resistance 50 Ω.
1

The noise-limited resolution is defined as the signal amplitude (rms) at which the signal-to-noise ratio is 0 dB with 1 Hz spectral resolution for frequencies above 30 kHz
2

measured on a mirror.

Range diagram
Amplitude
(peak -peak)

1 µm ■■ Displacement decoder for mea-


Max. Displacement: 375 nm
375 nm
suring smallest motions up to
± 188 nm with a large frequency
100 nm
bandwidth from 30 kHz up to
24 MHz
10 nm ■■ For measurements on ultra-
sonic transducers, detection of
ultrasonic pulses and to observe
1 nm
fast transient motions of MEMS
devices
100 pm ■■ Special high-pass filter suppress-
Lower Frequency Limit: 30 kHz

es low frequency (acoustic)


Max. Frequency: 24 MHz

vibrations and allows the detec-


10 pm
tion of smallest high-frequency
vibrations even under noisy
1 pm
ambient conditions
■■ Additional 2 MHz low-pass filter
< 0.2 pm improves resolution perfor-
mance in lower frequency range
Frequency
1 MHz

24 MHz
10 MHz

100 MHz
10 kHz

30 kHz
1 kHz

100 kHz

■■ Included in models V and VD

9
Acceleration decoder
AX-100 acceleration decoder
Measurement range Measurement range Full scale accelaration Signal frequency range 2
velocity accelaration output (peak) 1 (switchable)

mm/s/V m/s²/V m/s² Hz


12.5 125 1,250 0.5 ... 5,000 / 20,000
25 250 2,500 0.5 ... 5,000 / 20,000
50 500 5,000 0.5 ... 5,000 / 20,000
125 1,250 12,500 0.5 ... 5,000 / 20,000
250 2,500 25,000 0.5 ... 5,000 / 20,000
500 5,000 50,000 0.5 ... 5,000 / 20,000
1,250 12,500 125,000 0.5 ... 5,000 / 20,000
2,500 25,000 250,000 0.5 ... 5,000 / 20,000
1
he full scale values correspond to the maximum output voltage of 10 Vpeak.
T
2
Value is valid, as long as the corresponding velocity measurement range is not saturated.

Acceleration Range diagram


1000000 m/s 2

250000 m/s 2

125000 m/s 2

■■ 0.5Hz – 20kHz bandwidth 100000 m/s 2

50000 m/s 2

■■ 250,000 m/s² maximum 25000 m/s 2

acceleration 12500 m/s 2

10000 m/s 2

5000 m/s 2
/s
■■ 8 measurement ranges m m/s
2500 m/s 25 .5
2

t y 2 s
(coupled to VX-09 velocity 1250 m/s 2
ci y 1 m/ /s s
1000 m/s 2
V elo ocit y 5 .5 m m/ /s
l
. e c y .25 mm /s
i t 2
decoder ranges) ax V o it
M ax. V el loc ity 1 00 mm m/s
M ax. V e loc ty 5 250 5 m
■■ Available as an option in 100 m/s 2 M ax. V e loci ity 12
M ax. V e loc ity
models M, MD, R and RD M ax. V e loc
M ax. . V e
M ax
M
10 m/s 2
Max. Frequency: 20 kHz

1 m/s 2
Min. Frequency Limit: 0.5 Hz

100 mm/s 2

10 mm/s 2
5 mm/s 2

1 mm/s 2
Frequency
1 Hz

10 Hz

100 Hz
0.1 Hz
0.5 Hz

1 kHz

10 kHz

20 kHz

100 kHz
1.6 kHz

10
Compliance with standards

Laser safety IEC 60825-1

Electrical safety IEC 61010-1

EMC IEC 61326-1


Emission: Limit class B
IEC/EN 61000-3-2 and 61000-3-3
Immunity: IEC/EN 61000-4-2 to 61000-4-6 and IEC/EN 61000-4-11

85 174.3
Dimensions 231.8
A

Ø 68
1
B
M6 6
135

Ø 4F8 5
1/4-20 UNC 6.4
+0.028
67.5

67.5
Ø 4F8 +0.010 5
M6 6

60 186.8

365.3
383.2

B
100

53

Manufactured by:
GmbH
D-76337 Waldbronn, Germany
Model No.: MLV
Serial No.: x xx xxxx
Mfg.-Date: mm.yyyy

All dimensions in mm A Stand-off distance B Beam


if not marked otherwise

11
Measuring vibration non-contact and highly precise with Polytec laser Doppler vibrometers. The optional VibSoft data acquisition
hardware and analysis software allows quick and intuitive analysis and visualization of your vibration measurement data.

Polytec GmbH Polytec, Inc. (USA) Polytec Ltd. Polytec Japan Polytec China Ltd.
(Germany) North American (Great Britain) Arena Tower, 13th floor Room 402, Tower B
Polytec-Platz 1-7 Headquarters Lambda House 3-1-9, Shinyokohama Minmetals Plaza
76337 Waldbronn 16400 Bake Parkway Batford Mill Kohoku-ku, Yokohama-shi No. 5 Chaoyang North Ave
2018/05 - Technical specifications are subject to change without notice.

Tel. +49 7243 604-0 Suites 150 & 200 Harpenden, Herts AL5 5BZ Kanagawa 222-0033 Dongcheng District
[email protected] Irvine, CA 92618 Tel. +44 1582 711670 Tel. +81 45 478-6980 100010 Beijing
Tel. +1 949 943-3033 [email protected] [email protected] Tel. +86 10 65682591
Polytec GmbH [email protected] [email protected]
(Germany)
Vertriebs- und Central Office Polytec France S.A.S. Polytec South-East Asia
Beratungsbüro 1046 Baker Road Technosud II Pte. Ltd.
Schwarzschildstraße 1 Dexter, MI 48130 Bâtiment A Blk 4010 Ang Mo Kio Ave 10
OM_DS_OFV-5000_Xtra_E_42498

12489 Berlin Tel. +1 734 253-9428 99, Rue Pierre Semard #06-06 TechPlace I
Tel. +49 30 6392-5140 92320 Châtillon Singapore 569626
East Coast Office Tel. +33 1 496569-00 Tel. +65 64510886
1 Cabot Road [email protected] [email protected] www.polytec.com
Suites 101 & 102
Hudson, MA 01749
Tel. +1 508 417-1040

You might also like