Data Logger Endress and Hausser
Data Logger Endress and Hausser
Operating Instructions
Memograph M, RSG45
Advanced Data Manager
Memograph M, RSG45 Table of contents
Table of contents
Endress+Hauser 3
Table of contents Memograph M, RSG45
9.2 Switching on the measuring device . . . . . . . . . 39 11.10.2 Functions relating to the SD card or
9.3 Setting the operating language . . . . . . . . . . . . 39 USB stick . . . . . . . . . . . . . . . . . . . . . 57
9.4 Configuring the measuring device (Setup 11.10.3 Notes on e-mail encryption . . . . . . . . 60
menu) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40 11.10.4 Notes on WebDAV encryption . . . . . . 60
9.4.1 Step-by-step: to the first measured 11.10.5 SSL certificates . . . . . . . . . . . . . . . . . 61
value . . . . . . . . . . . . . . . . . . . . . . . . 40 11.11 Showing measured values history . . . . . . . . . . 62
9.4.2 Step-by-step: set or delete the limit 11.11.1 Historical data: changing a group . . . . 62
values . . . . . . . . . . . . . . . . . . . . . . . . 41 11.11.2 Historical data: Scroll speed . . . . . . . . 62
9.4.3 Step-by-step: read HART® values 11.11.3 Historical data: Time scaling . . . . . . . 62
(option) . . . . . . . . . . . . . . . . . . . . . . 41 11.11.4 Historical data: Time range
9.4.4 Step-by-step: HART® communication displayed . . . . . . . . . . . . . . . . . . . . . 62
between an FDT Frame application 11.11.5 Historical data: Screenshot . . . . . . . . . 62
(FieldCare) and a HART® device/ 11.11.6 Historical data: Change the display
sensor (optional) . . . . . . . . . . . . . . . . 41 mode . . . . . . . . . . . . . . . . . . . . . . . . 63
9.4.5 Setup directly at the device . . . . . . . . 42 11.11.7 Historical data: Store text . . . . . . . . . . 63
9.4.6 Setup via SD card or USB stick . . . . . . 42 11.12 Signal analysis . . . . . . . . . . . . . . . . . . . . . . . . 63
9.4.7 Setup via Web server . . . . . . . . . . . . . 42 11.13 Search in trace . . . . . . . . . . . . . . . . . . . . . . . . 63
9.4.8 Setup via FieldCare/DeviceCare 11.14 Changing the display mode . . . . . . . . . . . . . . 63
configuration software (included in 11.15 Store text . . . . . . . . . . . . . . . . . . . . . . . . . . . 64
the delivery) . . . . . . . . . . . . . . . . . . . 43 11.16 Printout . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64
9.5 Advanced settings (Expert menu) . . . . . . . . . . 44 11.17 Adjusting the brightness of the display . . . . . . 64
9.6 Configuration management . . . . . . . . . . . . . . 45 11.18 Limit values . . . . . . . . . . . . . . . . . . . . . . . . . . 64
9.7 Simulation . . . . . . . . . . . . . . . . . . . . . . . . . . . 45 11.19 WebDAV Client . . . . . . . . . . . . . . . . . . . . . . . 65
9.8 Protecting settings from unauthorized 11.19.1 Access to the WebDAV server via
access . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45 HTTP (HTML) . . . . . . . . . . . . . . . . . . 65
4 Endress+Hauser
Memograph M, RSG45 Table of contents
14 Repairs . . . . . . . . . . . . . . . . . . . . . . . . . . . 76
14.1 General notes . . . . . . . . . . . . . . . . . . . . . . . . 76
14.2 Spare parts . . . . . . . . . . . . . . . . . . . . . . . . . . 76
14.3 Return . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 77
14.4 Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . 78
15 Accessories . . . . . . . . . . . . . . . . . . . . . . . 79
15.1 Device-specific accessories . . . . . . . . . . . . . . . 79
16 Technical data . . . . . . . . . . . . . . . . . . . . 81
16.1 Function and system design . . . . . . . . . . . . . . 81
16.2 Input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 84
16.3 Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 88
16.4 Power supply . . . . . . . . . . . . . . . . . . . . . . . . 91
16.5 Performance characteristics . . . . . . . . . . . . . 100
16.6 Installation . . . . . . . . . . . . . . . . . . . . . . . . . 100
16.7 Environment . . . . . . . . . . . . . . . . . . . . . . . . 102
16.8 Mechanical construction . . . . . . . . . . . . . . . 102
16.9 Display and operating elements . . . . . . . . . . 103
16.10 Certificates and approvals . . . . . . . . . . . . . . 106
16.11 Ordering information . . . . . . . . . . . . . . . . . . 107
17 Appendix . . . . . . . . . . . . . . . . . . . . . . . . 109
17.1 Operating items in the "Expert" menu . . . . . . 109
17.1.1 "System" submenu . . . . . . . . . . . . . . 109
17.1.2 "Inputs" submenu . . . . . . . . . . . . . . . 131
17.1.3 "Outputs" submenu . . . . . . . . . . . . . . 162
17.1.4 "Communication" submenu . . . . . . . . 168
17.1.5 "Application" submenu . . . . . . . . . . . 191
17.1.6 "Diagnostics" submenu . . . . . . . . . . . 248
Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 253
Endress+Hauser 5
Document information Memograph M, RSG45
1 Document information
Symbol Meaning
DANGER!
DANGER This symbol alerts you to a dangerous situation. Failure to avoid this situation will result in
serious or fatal injury.
WARNING!
WARNING This symbol alerts you to a dangerous situation. Failure to avoid this situation can result in
serious or fatal injury.
CAUTION!
CAUTION This symbol alerts you to a dangerous situation. Failure to avoid this situation can result in
minor or medium injury.
NOTE!
NOTICE This symbol contains information on procedures and other facts which do not result in
personal injury.
6 Endress+Hauser
Memograph M, RSG45 Document information
Symbol Meaning
Permitted
Procedures, processes or actions that are permitted.
Preferred
Procedures, processes or actions that are preferred.
Forbidden
Procedures, processes or actions that are forbidden.
Tip
Indicates additional information.
Reference to documentation
Reference to page
Reference to graphic
, , … Series of steps
Result of a step
Visual inspection
Symbol Meaning
A, B, C, ... Views
Flow direction
A0013441
Hazardous area
- Indicates a hazardous area.
A0011187
1.3 Terminology
To improve clarity, abbreviations or synonyms are used in these instructions for the
following terms:
• Endress+Hauser:
Term used in these instructions: "Manufacturer" or "Supplier"
• Memograph M RSG45:
Term used in these instructions: "Device" or "Measuring device"
Endress+Hauser 7
Basic safety instructions Memograph M, RSG45
PROFIBUS®
Registered trademark of the PROFIBUS User Organization, Karlsruhe, Germany
PROFINET®
Registered trademark of the PROFIBUS & PROFINET International User Organization e.V.,
Karlsruhe, Germany
Modbus®
Registered trademark of SCHNEIDER AUTOMATION, INC.
EtherNet/IPTM
Registered trademark of ODVA, INC.
Internet Explorer®, ExcelTM
Registered trademarks of the Microsoft Corporation
Mozilla Firefox®
Registered trademark of the Mozilla Foundation
Opera®
Registered trademark of Opera Software ASA.
Google ChromeTM
Registered trademark of Google INC.
8 Endress+Hauser
Memograph M, RSG45 Basic safety instructions
• The manufacturer accepts no liability for damages resulting from incorrect use or use
other than that designated. It is not permitted to convert or modify the device in any
way.
• The device is designed for installation in a panel and must only be operated in an
installed state.
Endress+Hauser 9
Product description Memograph M, RSG45
2.7 IT security
The manufacturer only provides a warranty if the device is installed and used as described
in the Operating Instructions. The device is equipped with security mechanisms to protect
it against any inadvertent changes to the device settings.
IT security measures in line with operators' security standards and designed to provide
additional protection for the device and device data transfer must be implemented by the
operators themselves.
3 Product description
10 Endress+Hauser
Memograph M, RSG45 Installation
4.3.1 Nameplate
Compare the nameplate with the following diagram:
1 Memograph M
Ord. cd.: RSG45-XX/XXX
2 Ser. no.: XXXXXXXXXXX
Front: IP65 Rear: IP20 8
Ext. ord. cd.: RSG45-XXXXXXXXX XXX 12ATEXxx xx X
3 100-230 V AC (±10%) ~50/60 Hz 40 VA II2G Ex px IIC Gb 9
4 -10°C (14°F) < Ta < +50°C (122°F) II2D Ex pD IIIC Db
FW: XX.XX.XX Inst. per XA01362R/09/A3/xx.xx
5 MAC: XX-XX-XX-XX-XX-XX
6 MAC 1: XX-XX-XX-XX-XX-XX
0044
TAG1-XXXXXXXXXXX
7 10
TAG2-XXXXXXXXXXX
A0025806
5 Installation
Endress+Hauser 11
Installation Memograph M, RSG45
12 Endress+Hauser
Memograph M, RSG45 Installation
158.5 (6.24)
195.2 (7.69) 141.2 (5.56) 27(1.06)
1
(5 38 +1
.43 +0
149.2 (5.87)
.04
)
A
37.1
(1.46)
158.5 (6.24)
196 (7.72) 141.2 (5.56) 31.4 (1.24)
1
(5. 38 +1
43 +0.0
150 (5.91)
4
)
B
208 (8.19)
70 138
(2.76) (5.43)
138 (5.43)
162 (6.38)
C
24 (0.94)
A0024610
B B
D D
A A
C C
A0026672
3 Panel mounting
Endress+Hauser 13
Electrical connection Memograph M, RSG45
1. From the rear of the device, push the sealing rubber (B) (supplied) as far as the front
frame of the device (A).
2. Slide the device (A) through the panel cutout from the front (C). To avoid the buildup
of heat, maintain a distance of >12 mm (>0.47 in) from walls and other devices.
3. Hold the device (A) level and hang the fastening clips (D) in the openings (1 x left, 1
x right).
4. Evenly tighten the screws on the fasting clips (D) using a screwdriver to guarantee a
secure seal to the control panel (torque 100 Ncm).
6 Electrical connection
14 Endress+Hauser
Memograph M, RSG45 Electrical connection
Endress+Hauser 15
Electrical connection Memograph M, RSG45
individual grounding points, only one point of the shielding is connected directly with the
reference ground.
If the shielding of the cable is grounded at more than one point in systems without
potential matching, mains frequency equalizing currents can occur. These can damage
the signal cable or significantly impact signal transmission. In such cases the shielding
of the signal cable is to be grounded on one side only, i.e. it may not be connected to
the ground terminal of the housing. The shield that is not connected should be
insulated!
R12
R13
N/-
PE
R21
R22
R31
R32
R41
R42
R51
R52
R61
R62
R11
24V
Out
GND1
D21
D31
D41
D51
D61
D11
+
-
6
GND2
GND2
DC1
DD1
O15
O16
O25
O26
DA1
DB1
DE1
D71
D81
D91
5
RG
RC
RD
RH
RA
RB
RE
RK
RF
RL
RJ
RI
Ch15
Ch16
Ch13
Ch14
G1
G6
D1
D6
F1
F6
E1
E6
…
…
…
.
.
.
.
.
.
.
.
Ch12
Ch11
Ch10
Ch9
C1
3
C6
A1
A6
91
96
B1
B6
…
…
.
.
.
.
.
.
.
.
Ch7
Ch8
Ch6
Ch5
2
71
76
81
86
61
66
51
56
…
…
…
…
.
.
.
.
.
.
.
.
Ch4
Ch3
Ch1
Ch2
1
41
46
31
36
21
26
16
11
…
…
…
…
…
…
.
.
.
.
.
.
.
.
A0024605
Circuit diagram
16 Endress+Hauser
Memograph M, RSG45 Electrical connection
8
Termination
RL R= resistor
9 (typical 120 Ω) RS 485
RxD/TxD(+)
RxD/TxD(-)
Further units
12-24 VDC
>5 mA Digital in (D)
Analog +
_
24 V -
Power supply outputs (O) GNDx GNDx
100-230 VAC 24 V AC/DC (Selectable in unit setup)
(±10%) (-10%; +15%) _
Dx1
1 2
GNDx
50 / 60 Hz 50 / 60 Hz Pulse + - + - Dx1 Dx1
L+ N- PE 12-24 VDC
10k +
1 2
0...20 mA; + - + - -+ 24 V +
Dx1
GNDx
4...20 mA
>20 ms
O15
O16
O25
O26
Rel. max. 250 V / 3 A
L/+
N/-
PE
24V 24V Out: max. 250 mA
R41
R42
R51
R52
R61
R62
R12
R13
R21
R22
R31
R32
R11
Out
GND1
D21
D31
D41
D51
D61
D11
-
+
Slot 6: Power supply slot (standard)
GND2
GND2
DC1
DD1
O15
O16
O25
O26
DA1
DB1
DE1
D71
D81
D91
Slot 5: analog input / HART 17-20 or digital-I/O (option)
RG
RC
RD
RH
RA
RB
RE
RK
RF
RL
RJ
RI
RS 232
Ch15
Ch16
Ch13
Ch14
D6
F1
E6
G1
G6
F6
8 9
…
…
…
…
…
…
…
…
…
…
…
…
…
…
…
…
Ch10
Ch12
Ch11
Ch9
C6
91
C1
B1
B6
96
A1
A6
…
…
…
…
…
…
…
…
…
…
…
…
…
…
…
…
1 5
Ch7
Ch6
Ch5
Ch8
2 3 5 8 9
Slot 2: analog input / HART 5-8 (option)
81
71
51
61
86
56
76
66
…
…
…
…
…
…
…
…
…
…
…
…
…
…
…
…
Ch3
Ch1
Ch4
Ch2
31
41
36
26
16
46
…
…
…
…
…
…
…
…
…
…
…
…
…
…
…
…
5 1
9 6
To PC: Cable with 6 9 Slot 0: CPU (standard)
9 pol. Sub-D plug Ethernet USB RS 232 / RS 485 Bus Interface
1 5
RxD - 3 GND - 5 LED LED
23 5 8 9 green yellow
TxD - 2
Analog inputs
x1 x2 x3 x4 x5 x6 x1 x2 x3 x4 x5 x6 x1 x2 x3 x4 x5 x6
+
TC A, B, C, D, J, K, L, N, T, R, S
-
x1 x2 x3 x4 x5 x6
24 V >40 µs
R=1.2 k
+
+
A0026669-EN
5 For connection examples of the HART® inputs (optional), see the Operating Instructions → 23
Endress+Hauser 17
Electrical connection Memograph M, RSG45
A0019103
100-230 VAC L+ N- PE
24 V AC/DC L+ N- PE
A0019103
The open or close function (= activation or deactivation of the relay coil) in a limit
event can be configured in the setup: "Setup -> Advanced setup -> Outputs -> Relay ->
Relay x". However, in the event of a power failure, the relay adopts its quiescent switch
state regardless of the setting programmed.
Type Terminal
A0019103
18 Endress+Hauser
Memograph M, RSG45 Electrical connection
Type Terminal
A0019103
If the auxiliary voltage is to be used for the digital inputs, the 24 V out - terminal of
the auxiliary voltage output must be connected with the GND1 terminal.
Type Terminal
Chx
x1
x2
x3
x4
x5
x6
A0019303
x1 x2 x3 x4 x5 x6
1) If a universal input is used as a frequency or pulse input, a series resistor must be used in series connection
with the voltage source. Example: 1.2 kΩ series resistor at 24 V
Endress+Hauser 19
Electrical connection Memograph M, RSG45
Type Terminal
Modem
Chx
250Ω 10Ω
x1
x2
x3
x4
x5
x6
x1
x2
x3
x4
x5
x6
A0024862
x1 x2 x3 x4 x5 x6
GND2
GND2
DC1
DD1
DA1
DB1
DE1
O15
O16
O25
O26
D71
D81
D91
RG
RC
RD
RH
RA
RB
RE
RK
RF
RL
RJ
RI
A0024736
Relay 7, 8 RA RB RC RD
Relay 9, 10 RE RF RG RH
Relay 11, 12 RI RJ RK RL
1) NO)
2) NO)
The open or close function (= activation or deactivation of the relay coil) in a limit
event can be configured in the setup: "Setup -> Advanced setup -> Outputs -> Relay ->
Relay x". However, in the event of a power failure, the relay adopts its quiescent switch
state regardless of the setting programmed.
Type Terminal
GND2
GND2
DC1
DD1
DA1
DB1
DE1
O15
O16
O25
O26
D71
D81
D91
RG
RC
RD
RH
RA
RB
RE
RK
RF
RL
RJ
RI
A0024736
20 Endress+Hauser
Memograph M, RSG45 Electrical connection
Type Terminal
GND2
GND2
DC1
DD1
DA1
DB1
DE1
O15
O16
O25
O26
D71
D81
D91
RG
RC
RD
RH
RA
RB
RE
RK
RF
RL
RJ
RI
A0024736
If the auxiliary voltage is to be used for the digital inputs, the 24 V out - terminal of
the auxiliary voltage output (power unit, slot 6) must be connected with the GND2
terminal.
+
- +
3
Y
_
1
_
A0024729
6 Connecting the auxiliary voltage output when using as a transmitter power supply for 2-wire sensors in
the current measuring range
1 Sensor 1 (e.g. Cerabar from Endress+Hauser)
2 Sensor 2
3 External indicator (optional) (e.g. RIA16 from Endress+Hauser)
Endress+Hauser 21
Electrical connection Memograph M, RSG45
_ + +
-
Y
3
+ +
- _
Y
1
A0024730
7 Connecting the auxiliary voltage output when using as a transmitter power supply for 4-wire sensors in
the current measuring range
1 Sensor 1 (e.g. temperature switch TTR31 from Endress+Hauser)
2 Sensor 2
3 External indicator (optional) (e.g. RIA16 from Endress+Hauser)
22 Endress+Hauser
Memograph M, RSG45 Electrical connection
1 3 5 5 6
Y Y Y
2
I I I
+ - + - - + - + + -
4
- + - + 6
Chx
Chx
Chx
Chx
x1
x2
x3
x4
x5
x6
x1
x2
x3
x4
x5
x6
x1
x2
x3
x4
x5
x6
x1
x2
x3
x4
x5
x6
A0024864
The internal auxiliary voltage (24 V OUT) can also be used as the transmitter power
supply.
3
Y
2
I
1 Y
I
Y
I
- +
A0024860
The internal auxiliary voltage (24 V OUT) can also be used as the transmitter power
supply.
Endress+Hauser 23
Electrical connection Memograph M, RSG45
Cable
resistance
Further units
RxD - 3 GND - 5
TxD - 2
A0024732-EN
1 2 3 4 5 6 7 8 9
Only one interface can be used at any one time (RS232 or RS485).
24 Endress+Hauser
Memograph M, RSG45 Electrical connection
Endress+Hauser 25
Electrical connection Memograph M, RSG45
Reference list: HP Color LaserJet CP1515n, HP Color LaserJet Pro CP1525n, Kyocera FS-
C5015N.
GDI printers are not supported!
PROFIBUS–DP slave:
The device can be integrated into a fieldbus system as per the PROFIBUS–DP standard by
means of the PROFIBUS-DP interface. Up to 40 analog inputs and 20 (14 real + 6 virtual)
digital inputs can be transmitted via PROFIBUS-DP and stored in the device. For
bidirectional communication in cyclic data transfer. Connection via Sub-D socket.
Baud rate: maximum 12 Mbit/s.
6.3.10 USB connection, type A (host) (CPU card, rear of device, slot 0)
Two USB 2.0 ports are available on shielded USB A sockets at the rear of the device. A USB
stick as a memory medium, for example, can be connected to these ports. An external
keyboard/mouse for device operation, a USB hub, a barcode reader or a printer (PCL5c or
higher) may also be connected.
26 Endress+Hauser
Memograph M, RSG45 Electrical connection
4 3 2
A0024737
Endress+Hauser 27
Electrical connection Memograph M, RSG45
28 Endress+Hauser
Memograph M, RSG45 Electrical connection
Reference list: HP Color LaserJet CP1515n, HP Color LaserJet Pro CP1525n, Kyocera FS-
C5015N.
GDI printers are not supported!
Are the power supply and signal cables correctly connected? See connection diagram and
rear of device.
Endress+Hauser 29
Operation options Memograph M, RSG45
7 Operation options
Security
External memory
Messages
Screen saver
Barcode reader
Device options,...
Digital inputs
Relay 1-12
Communication Ethernet
HART
Serial interface
Modbus
Profibus,...
Application Maths
Signal analysis
Diagnostics Diagnostics list
Limits
Event logbook
Signal groups
Device information
E-mail
Measured values
Printer
Outputs
Softkeys
Simulation
Texts,...
HART,...
A0024770-EN
30 Endress+Hauser
Memograph M, RSG45 Operation options
+Clear memory
+Expert functions
+Expert functions
+Expert functions
+Expert functions
+Expert functions
Event logbook
Device information
Simulation
A0019596-EN
Operator Tasks during operation: "Operation" Contains all the parameters that are required in
• Configuration of the display. ongoing operation: configuration of the measured value
• Reading measured values. display (displayed values, display format, etc.).
Endress+Hauser 31
Operation options Memograph M, RSG45
Fault elimination: "Diagnostics" Contains all parameters for detecting and analyzing
• Diagnosing and eliminating process errors. errors:
• Interpretation of device error messages and • Diagnosis list
correcting associated errors. All diagnosis messages are listed in chronological
order.
• Event logbook
Events such as limit value violations and power
failures are listed in chronological order.
• Device information
Displays important device information (e.g. serial
number, firmware version, device options for
hardware and software, memory information, etc.).
• Measured values
Display of current measured values of device.
• Outputs
Displays the current status of the outputs e.g. switch
status of relay outputs.
• Simulation
Various functions/signals can be simulated for test
purposes here.
Note: In Simulation mode, normal recording of the
measured values is interrupted and the intervention
is logged in the event log.
• HART®
Displays the exact device information of a selected
HART® device and the HART® communication signal
quality.
• Initialize modem
Initializes the modem connected to the serial
interface (for automatic call answering).
Expert Tasks that require detailed knowledge of the "Expert" Contains all parameters of the device (including those
function of the device: that are already in one of the other menus). The expert
• Commissioning measurements under difficult menu is protected by a code. Factory setting: 0000. This
conditions. menu is structured according to the function blocks of
• Optimal adaptation of the measurement to the device:
difficult conditions. • "System" submenu
• Detailed configuration of the communication Contains all higher-order device parameters that do
interface. not concern the measurement or measured value
• Error diagnostics in difficult cases. communication.
• "Inputs" submenu
Contains all parameters for configuring the analog
and digital inputs.
• "Output" submenu
Contains all parameters for configuring the outputs
(e.g. relays).
• "Communication" submenu
Contains all parameters for configuring the
communication interfaces.
• "Application" submenu
Contains all parameters for configuring application-
specific settings (e.g. group settings, limit values
etc.).
• "Diagnostics" submenu
Contains all parameters needed to detect and
analyze operational errors.
32 Endress+Hauser
Memograph M, RSG45 Operation options
c b a
5 6 7 8 5 6 7 8
9 9
4 10 4 10
3 1
2
A0024709
11 Device front (left: version with navigator and front interfaces; right: version with stainless steel front and
touchscreen)
c USB A socket "Host" e.g. for USB memory stick, external keyboard, barcode reader or printer
d LED at SD slot. Yellow LED lit or flashing when the device writes to the SD card or reads it.
Do not remove the SD card if the LED is lit or flashing! Risk of data loss!
Endress+Hauser 33
Operation options Memograph M, RSG45
8 In display mode: alternating display indicating the percentage space on the SD card or USB stick that
has already been used.
Status symbols are also displayed in alternation with the memory information (e.g. simulation mode,
data storage active, operation locked, batch active)
In setup mode: the current "direct access" operating code is displayed
9 In display mode: window for measured value display (e.g. curve display).
Display of current measured values and the status in the event of an error/alarm condition. In the case
of counters, the type of counter is displayed as a symbol.
inIf aredmeasuring point has limit value status, the corresponding channel identifier is highlighted
(quick detection of limit value violations). During a limit value violation and device
operation, the acquisition of measured values continues uninterrupted.
10 In display mode: alternating status display (e.g. set zoom range) of the analog or digital inputs in the
appropriate color of the channel.
In setup mode: different information is displayed here depending on the display type.
åD Daily analysis
åW Weekly analysis
åM Monthly analysis
åY Annual analysis
å Totalizer
9 Channel-related symbols:
"Out of specification"
e.g. input signal too high/low
"Maintenance required"
Maintenance is required. The measured value is still valid.
"Device locked"
Setup is locked via a control input. Disable setup lock via a control input.
"Out of specification"
The device is being operated outside its technical specifications (e.g. during
startup or cleaning).
"Function check"
The device is in Service mode.
34 Endress+Hauser
Memograph M, RSG45 Operation options
"Maintenance required"
Maintenance is required. The measured value is still valid.
"External communication"
The device is communicating externally (e.g. via Modbus).
SIM "Simulation"
Simulation is active.
4 "Historical data"
Historical data are currently shown on screen.
Back
Use the "Back" function, which can be found at the bottom of each menu/submenu, to move up
a level in the menu structure.
devices
To quit the menu immediately, press and hold "Back" (>3 sec.) in the Navigator. The
switches to display mode.
Setup changes
Power on
Power off
Limit value on
Service
User administration
Acknowledging message
Back
Continue searching
Endress+Hauser 35
Operation options Memograph M, RSG45
←x Delete backwards.
If this symbol is selected, the character to the left of the cursor position is deleted.
x→ Delete forwards.
If this symbol is selected, the character to the right of the cursor position is deleted.
Delete all.
if this symbol is selected, the entire entry is deleted.
Reject entry.
If this symbol is selected, the entry is rejected and you quit editing mode. The previously set text
remains.
Accept entry.
If this symbol is selected, the entry is applied at the position specified by the user, and you quit
editing mode.
For details, see the Operating Instructions on the analysis software DVD provided.
36 Endress+Hauser
Memograph M, RSG45 Operation options
Function scope
The configuration software is an FDT/DTM-based system asset management tool. It can
configure all smart field devices in a system and helps you manage them. By using the
status information, it is also a simple but effective way of checking their status and
condition. Access is via USB or Ethernet interface (network).
Typical functions:
• Device configuration
• Loading and saving device data (upload/download)
• Documentation of the measuring point
For details, see the Operating Instructions on the configuration software DVD
provided
Endress+Hauser 37
System integration Memograph M, RSG45
8 System integration
8.1.2 Ethernet
Setup → Advanced setup → Communication → Ethernet
The IP address can be entered manually (fixed IP address) or assigned automatically using
DHCP.
The port for data communication is preset to 8000. The port can be changed in the Expert
→ Communication → Ethernet menu.
The following functions are implemented:
• Data communication with PC software (analysis software, configuration software, OPC
server)
• Web server
The following connections are possible at the same time:
• 1x Port 8000 (configuration software, OPC server or analysis software)
• 1x Port 8002 (OPC server only)
• 1x Port 5094 (HART® IP)
• 4x Modbus slave TCP
• 5x Web server
Ports can be changed!
As soon as the maximum number of connections has been reached, new connection
attempts are blocked until an existing connection has been terminated.
38 Endress+Hauser
Memograph M, RSG45 Commissioning
Parity: Even/Odd/None -
Port: - 502
Transfer of values
The actual Modbus TCP protocol is located between layer 5 to 6 in the ISO/OSI model.
To transfer a value, 3 registers of 2 bytes each (2-byte status + 4-byte float) or 5 registers
of 2 bytes each (2-byte status + 8-byte double) are used.
More detailed information about Modbus is provided in the supplementary
documentation.
9 Commissioning
Version with stainless steel front and touchscreen, or when operating with the
external USB mouse:
Calling the main menu, configuring the operating language:
1. Press or click the "Menu" soft key at the bottom edge of the screen
2. The main menu appears on the display with the "Sprache/Language" option
Endress+Hauser 39
Commissioning Memograph M, RSG45
3. To change the default language setting: press or click "Sprache/Language" and select
the desired language from the drop-down menu
4. Use "Back" or "ESC" to quit the main menu
The operating language has been changed.
Use the "Back" function, which can be found at the bottom of each menu/submenu,
to move up a level in the menu structure.
To quit the menu immediately and return to the measured value display, press and
hold "Back" (>3 sec.). The changes made are accepted and saved.
40 Endress+Hauser
Memograph M, RSG45 Commissioning
The device is in the measured value display mode and displays the measured values.
Endress+Hauser 41
Commissioning Memograph M, RSG45
Change date/time UTC time zone Change the date and time
dd.mm.yyyy hh:mm:ss
Advanced setup Advanced settings for the device e.g. system settings,
inputs, outputs, communication, application etc.
A detailed overview of all the operating parameters is provided in the appendix at the
end of the Operating Instructions. → 109
42 Endress+Hauser
Memograph M, RSG45 Commissioning
Please observe the information and communication settings for Ethernet and the Web
server under .→ 38
To configure the device via a Web server, you must have Administrator or Service
access. ID and password administration is performed in the main menu under "Setup
-> Advanced setup -> Communication -> Ethernet -> Configuration Web server ->
Authentication".
Default value for ID: admin; Password: admin
Note: The password should be changed during commissioning!
If security settings are according to "FDA 21 CFR Part 11", you must have
Administrator rights to configure the device via a Web server.
Continue with device configuration in accordance with the Operating Instructions for the
device. The complete Setup menu i.e. all of the parameters listed in the Operating
Instructions, can also be found on the Web server. After configuration, accept the setup
with "Save settings".
Procedure to establish a direct connection via Ethernet (point-to-point connection):
→ 51
NOTICE
Undefined switching of outputs and relays
‣ During configuration using a Web server, the device may assume undefined statuses!
This may result in the undefined switching of outputs and relays.
Endress+Hauser 43
Commissioning Memograph M, RSG45
NOTICE
Undefined switching of outputs and relays
‣ During configuration using the configuration software, the device may assume
undefined statuses! This may result in the undefined switching of outputs and relays.
A detailed overview of all the operating parameters is provided in the appendix at the
end of the Operating Instructions. → 109
44 Endress+Hauser
Memograph M, RSG45 Commissioning
LCAUTION
If the SD card or USB stick are removed directly:
Risk of data loss on SD card or USB stick.
‣ To remove the SD card or the USB stick, always select "Operation -> SD card (or USB
stick) -> Remove safely" in the main menu!
The function for saving the setup data must be enabled at the device for the Web
server under "Setup -> Advanced setup -> Communication-> Ethernet -> Web
server settings; Setup -> Yes".
9.7 Simulation
Various functions/signals can be simulated for test purposes here.
NOTICE
Selecting simulation: Simulation of the relays and the WebDAV client can be found in
the main menu under "Diagnostics -> Simulation". The simulation of the measured
values can be found in the main menu under "Expert -> Diagnostics -> Simulation".
Only the simulated values are recorded during simulation. The simulation is recorded in
the event logbook.
‣ Do not start simulation if measured value recording must not be interrupted!
Endress+Hauser 45
Commissioning Memograph M, RSG45
Setup lock via control input: The settings for the control input can be found in the main
menu under "Setup -> Advanced setup -> Inputs -> Digital inputs -> Digital input X ->
Function: Control input; Action: Lock setup".
It is preferable to lock the setup using a control input.
Setting up an access code: The settings for the access code can be found in the main
menu under "Setup -> Advanced setup -> System -> Security -> Protected by -> Access
code". Factory setting: "open access", i.e. modifications can always be made.
Make a note of the code and store in a safe place.
Setting up user roles: The settings for the user roles (operator, admin and service) are
provided in the main menu under "Setup -> Advanced setup -> System -> Security ->
Protected by -> User roles" . Factory setting: "open access", i.e. modifications can always
be made.
The passwords should be changed during commissioning.
Make a note of the code and store in a safe place.
Setting up user administration according to "FDA 21 CFR Part 11": The settings for user
administration can be found in the main menu under "Setup -> Advanced setup ->
System -> Security -> Protected by -> FDA 21 CFR Part 11". Factory setting: "open
access", i.e. modifications can always be made.
46 Endress+Hauser
Memograph M, RSG45 Fulfilling requirements in accordance with "FDA 21 CFR Part 11"
Endress+Hauser 47
Fulfilling requirements in accordance with "FDA 21 CFR Part 11" Memograph M, RSG45
NOTICE
The user is responsible for observing the following instructions for ensuring
compliance with FDA 21 CFR Part 11:
‣ The scope of application for this FDA 21 CFR Part 11 evaluation refers to the
Memograph M RSG45 and the associated Field Data Manager (FDM) PC software,
which have been produced and tested by the manufacturer. The overall system must be
subjected to a final evaluation by the end user.
‣ Recordings from the device and the associated PC software in accordance with "FDA
Guidance for Industry: Computerized Systems Used in Clinical Investigations: 2007"
cannot be used for clinical studies.
‣ In order to satisfy the requirements of FDA 21 CFR Part 11 with regard to the
electronic signature, the device may only be read out with the associated Field Data
Manager (FDM) PC software.
‣ The device and the associated PC software do not take into account the requirements
of FDA 21 CFR Part 11, §11.30 for "Open systems". Operation within the context of an
open system must be excluded.
‣ The operator must take suitable management precautions to prevent the loss of
password access data/authentication data.
‣ The printer must be connected correctly and its long-term function guaranteed, even in
the event of local changes (e.g. network printer).
‣ Due to the aging of components, the device must be calibrated at regular intervals.
‣ Regular maintenance must also be carried out.
‣ Measures must be taken to prevent the device from being moved (e.g. physical
protection).
‣ No data will be recorded in the event of a power failure.
‣ The battery buffer for the backup clock is checked and, where necessary, a warning
issued.
‣ Wireless transmission may not be used between the device and the PC software. In the
case of wired transmission, operation is only permitted within the closed company
network. This must be checked during the validation of the complete system.
NOTICE
The user is responsible for observing the following instructions for ensuring
compliance with FDA 21 CFR Part 11:
Requirements for connected hardware components:
‣ Associated hardware components (e.g. a PC connected to the device) are part of the
system but must be provided by the user. Backup measures designed to prevent hard
disk defects or a memory overrun are the responsibility of the user.
‣ Wired peripherals, such as a USB keyboard for the device or a keyboard for the PC, can
be scanned in using "Keylogger". It is the user's responsibility to prevent the use of so-
called "sniffers".
48 Endress+Hauser
Memograph M, RSG45 Fulfilling requirements in accordance with "FDA 21 CFR Part 11"
NOTICE
The user is responsible for observing the following instructions for ensuring
compliance with FDA 21 CFR Part 11:
Requirements for the device:
‣ The device must be regarded as a closed system.
‣ Administrator rights are required to change the configuration. The device must be
locked via hardware protection (terminal cover) and a digital input assigned. Changes
are documented in the event logbook / audit trail. The event logbook can be opened
from the main menu under "Diagnostics -> Event logbook".
‣ User administration with ID and password must be active.
‣ All changes must always be documented with the user's name in the event logbook. It is
essential that user administration is active for this. All users must be created
accordingly under User administration.
‣ All settings relating to password complexity and validity period are the responsibility of
the administrator.
‣ During commissioning, the special requirements for the admin role must be taken into
account (stricter password rules). User administration can only be disabled by the
administrator. Under certain circumstances the administrator may lock users out of the
system or sabotage user administration.
‣ The initial password set by the administrator must be changed on the first login
(configurable).
‣ If the wrong password is entered repeatedly (max. number of attempts is adjustable),
access is blocked and an alarm can be sent by e-mail.
‣ After setting up/using the device, the user must log out. Automatic logout after a
configurable time must be activated.
‣ Communication interfaces can be blocked by the administrator. The interface for using
the modem must be switched off.
‣ Hardware locking must be performed for the device update. This ensures that only the
administrator can perform a firmware update.
‣ Any manipulation of the time synchronization is the responsibility of the user.
‣ In order to prevent a loss of data, the data must be read out regularly from the memory.
‣ Administrator rights are required in order to delete the internal memory. A hardware
lock can also be used to prevent the memory from being deleted.
‣ The device produces a diagnosis list in which faults are documented. The diagnosis list
must be checked regularly; this can be done in the main menu under "Diagnostics ->
Diagnosis list ".
‣ Proof testing of the error log is recommended; this is done from the main menu under
"Diagnostics -> Device information -> Error log".
Endress+Hauser 49
Fulfilling requirements in accordance with "FDA 21 CFR Part 11" Memograph M, RSG45
NOTICE
The user is responsible for observing the following instructions for ensuring
compliance with FDA 21 CFR Part 11:
Requirements for the external memory:
‣ The external storage medium (SD card/USB stick) must only be removed using the
function "Operation -> SD card/USB stick -> Remove safely".
‣ In contrast to the protected data format, which features CRC16 protective measures,
the stored data can be manipulated when using the open format (*.csv).
‣ In order to prevent unauthorized access (e.g. through theft), the external memory can
also be disabled or removed.
NOTICE
In order to satisfy the requirements according to FDA 21 CFR Part 11, certain
functions are required in the associated PC software. In particular, some user
administration settings must be made.
‣ Use only operating systems with user administration (e.g. MS Windows® 2000/XP/
Vista/7/8).
‣ Create an administrator: Create an administrator under "Extras -> User
administration".
‣ Activate user administration and password protection according to FDA 21 CFR Part
11: Under "Extras -> Settings -> User administration" select "Activate user
management" and "Password protection in compliance with FDA 21 CFR Part 11".
The software applies the settings following a restart.
‣ Additional users can now be created under "Extras -> User administration" with the
corresponding access roles.
‣ All changes made to the software are recorded in the "Audit trail". This can be opened
via "Extras -> Audit trail".
‣ Where appropriate, use the high-performance automatic functions of the PC software
(e.g. automatic read out, automatic backup function, automatic e-mail alert).
‣ Note:Detailed instructions can be found on the DVD provided with the PC software.
NOTICE
The user is responsible for observing the following instructions for ensuring
compliance with FDA 21 CFR Part 11:
‣ Modem operation cannot be implemented as a closed system.
‣ User administration must be active in order to record the user names in the audit trail.
‣ The password must be changed on the first login.
‣ If the administrator forgets the password, a master password, which is only valid for a
limited period, may be used. Access is granted to the user on request and following
authorization.
‣ If a user does not log out, an automatic lockout (screensaver) is activated via the
Windows® system setting.
‣ It is possible to use the Field Data Manager PC software without identification or
password protection, but this is not compliant with FDA 21 CFR Part 11.
‣ The ending of the automatic system service must be controlled via Windows® user
rights.
‣ Since the PC date/time is used for the audit trail, it is theoretically possible to
manipulate the time stamp. This must be prevented by the user by means of Windows®
user rights.
‣ Regular backups and preventive maintenance (e.g. by means of software updates) must
be carried out.
‣ It is important to ensure that no data can be intercepted or manipulated during
communication between the device and the database.
50 Endress+Hauser
Memograph M, RSG45 Operation
11 Operation
The "Operation" menu is geared towards the tasks and activities of the operator. It contains
all the parameters that are needed in ongoing operation. Historical values and analyses,
for example, can be displayed in the "Operation" menu and display settings can be made.
Any settings made for the onsite display have no effect on the measurement section or the
configured device parameters.
The unit's simple control system and the integrated help function enables you to perform
operation for many applications without the need for hard copy operating instructions.
Endress+Hauser 51
Operation Memograph M, RSG45
c b a
5 6 7 8 5 6 7 8
9 9
4 10 4 10
3 1
2
A0024709
12 Device front (left: Version with navigator and front interfaces; right: Version with stainless steel front and
touchscreen)
a Slot for SD card
b USB B socket "Function" e.g. to connect to PC or laptop
c USB A socket "Host" e.g. for USB memory stick, external keyboard, barcode reader or printer
d Yellow LED for read/write access to the SD card
1 Navigator: press briefly to open the main menu and confirm messages (=Enter); press for longer to open the
online help
2 Green LED (top) lit: Power supply present. Red LED (bottom) flashing: Maintenance required
3 Variable "Softkey" keys 1-4 (from left to right)
4 Function indicator of "Softkey" keys
5 Header: group name, analysis type
6 Header: current date/time
7 User ID (when function is active)
8 Header: alternating display indicating the percentage space on the SD card or USB stick that has already been
used. Status symbols are also displayed in alternation with the memory information
9 Area for measured value display (e.g. curve display). Display of current measured values and the status in the
event of an error/alarm condition. In the case of counters, the type of counter is displayed as a symbol
10 Status bar
An overview of all the symbols and icons is provided in the operating options section.
→ 34
If a measuring point has limit value status, the corresponding channel identifier is
highlighted in red (quick detection of limit value violations). During a limit value
violation and device operation, the acquisition of measured values continues
uninterrupted.
Information on how to rectify a problem in the event of an error is provided in the
"Troubleshooting" section. → 66
52 Endress+Hauser
Memograph M, RSG45 Operation
Endress+Hauser 53
Operation Memograph M, RSG45
The structure of the channel values for the XML file is explained as follows:
Day Description
MAN Manufacturer
54 Endress+Hauser
Memograph M, RSG45 Operation
H4000504428 0000000279 Contains all the measured values of the group from the ANSI
0000000185 2013-11-07 starting time indicated in the file name. An individual CSV
11-18-00 GROUP01.csv file is created for each group.
H4000504428 0000000279 Contains the signal analyses of the active channels from ANSI
0000000185 2013-11-07 the starting time indicated in the file name. An individual
11-30-00 ANALYSIS01.csv CSV file is created for every analysis (01 - 04).
H4000504428 0000000279 Contains the event log from the starting time indicated in Unicode UTF-8
2013-11-07 11-18-34 the file name. (see the notes in
EVENTS.csv the following
section)
Meaning of the values under "Status" and "Limit" when analyzing a group:
Status of the channel:
0: OK
1: Open circuit
6: Error value, i.e. not the calculated value (for mathematics if an input variable is invalid)
7: Sensor/input error
General status:
Endress+Hauser 55
Operation Memograph M, RSG45
Importing CSV data from the event logbook ("Events") into MS ExcelTM (version 2007
and higher):
1. Select the menu "Data -> Get external data - From text"
2. Select the CSV file
3. Follow the instructions in the wizard
4. Select file origin "Unicode UTF-8"
56 Endress+Hauser
Memograph M, RSG45 Operation
Endress+Hauser 57
Operation Memograph M, RSG45
Start the process picture generator and create the first project:
1. Copy the entire "ProcessPictureGenerator" directory from the DVD to a local directory
on the PC
2. Alternatively, unzip the downloaded ZIP file in a local directory on the PC
3. Double-click on the "FieldDiagramer.exe" application it contains
58 Endress+Hauser
Memograph M, RSG45 Operation
Parameters Parameter/description
"Edit" submenu Here you can edit the currently displayed process screen (position, font size, etc.).
"Channel 1-8" Here you can edit the selected channel of the process screen (activate, change position/
submenu font size, etc.).
Channel ident.
Specify whether the channel identifier should be displayed in addition to the measured
value. The channel identifier is displayed above the measured value.
Options: No, Yes; Factory setting: No.
Alignment
Specify how the measured value should be aligned.
IfIf "Align
"Align left" is selected, the x position is the top, left corner of the measured value.
right" is selected, the x position is the top, right corner of the measured
value.
Options: Left-aligned, right-aligned.
Factory setting: Right-aligned.
Endress+Hauser 59
Operation Memograph M, RSG45
Parameters Parameter/description
x-position
Select the x-position at which the measured value should be displayed.
Input: 0 to 799 pixels; factory setting: 10.
the
The point of origin (x-/y-zero point) is the top left. When the x-value is increased
text moves to the right.
y-position
Select the y-position at which the measured value should be displayed.
The point of origin (x-/y-zero point) is the top left. When the y-value is increased
the text moves down.
Input: 0 to 450 pixels; factory setting: 50 (channel 1) to 260 (channel 8).
Font size
Select the font size in which the measured value should be displayed.
Options: Small, Medium, Large, Very large; factory setting: Large.
Accept changes Saves the changes to the process screen in the device memory.
Delete Deletes the selected process screen from the device memory. The device then changes to
the curve display.
60 Endress+Hauser
Memograph M, RSG45 Operation
Advanced setup -> Application -> WebDAV Client -> Enable -> Yes (SSL)" or under
"Expert -> Application -> WebDAV Client -> Enable -> Yes (SSL)".
Only TLS V1.0 (= SSL 3.1) or TLS V1.1 is supported. Older standards are not supported.
The encryption method is automatically agreed with the counterparty.
A certificate must be installed in order to be able to send encrypted data. These certificates
can be obtained from your WebDAV server service provider. The following file formats are
supported:
• *.CER: DER- or Base64-encoded certificate
• *.CRT: DER- or Base64-encoded certificate
• *.DER: DER-encoded certificate
The file name of the certificate may only contain the following characters: a..z, A..Z,
0..9, +, -, _, #, (, ), !
To establish an SSL connection, the device automatically selects the certificate that best
suits the counterparty from all the installed certificates. An error message is displayed if
the device does not have any of the required certificates.
When WebDAV Client encryption is enabled, no data can be sent if a valid certificate
is not available or if the certificate has expired.
Endress+Hauser 61
Operation Memograph M, RSG45
62 Endress+Hauser
Memograph M, RSG45 Operation
Endress+Hauser 63
Operation Memograph M, RSG45
The following display modes are possible: Curve, Curve in ranges, Waterfall, Waterfall in
ranges, Bargraph, Digital display, Instrument display, Circular chart, and Process screen.
The various display modes have no influence on the signal recording.
11.16 Printout
The device settings, user administration data, event log, current measured values or a
screenshot of the measured value display can be printed out in the main menu under
"Operation -> Printout".
NOTICE
Due to technical restrictions, the printout from the device is always in English when
using the following operating languages: Polish, Russian, Swedish, Czech, Japanese,
and Chinese.
Printout via PC:
‣ The device settings or user administration can be saved in the set language in the form
of an RTF file and printed out from the PC.
NOTICE
Due to technical restrictions, the event log cannot be printed out from the device
when using the following operating languages: Polish, Russian, Swedish, Czech,
Japanese, and Chinese.
Printout via PC:
‣ The event log can be printed out via the Field Data Manager (FDM) software.
Alternatively the event log can be saved as a CSV file and printed out from the PC.
64 Endress+Hauser
Memograph M, RSG45 Operation
Endress+Hauser 65
Diagnostics and troubleshooting Memograph M, RSG45
No measured value display; no No supply voltage connected Check the supply voltage of the device
LED lit
Supply voltage is applied; device or The power unit or the device must be
power unit is defective replaced
Diagnostic message is The list of diagnostic messages is provided in the following section
displayed
Dead pixels: Dead pixels refer to pixels on LCD and TFT displays that are defective
due to the technology or manufacturing techniques used. The TFT display used can
have up to 10 dead pixels (Class III as per ISO 13406-2). These dead pixels do not
entitle the user to a warranty claim.
12.2 Troubleshooting
The Diagnostics menu is used for the analysis of the device functions and offers
comprehensive assistance during troubleshooting. Always proceed as follows to locate the
cause of the device errors or alarms.
The contact details of your Endress+Hauser representative can be found on the Internet at
www.endress.com/worldwide.
66 Endress+Hauser
Memograph M, RSG45 Diagnostics and troubleshooting
F105 Invalid value! Measured value is invalid (when calculating --> Check connections and process variables
NAN)
F261 Analog card x out of order! Hardware defect detected Contact the Service Department, replace card
F261 HART card out of order! Hardware defect detected Contact the Service Department, replace card
F261 Power supply out of order! Hardware defect detected Contact the Service Department, replace power unit
F261 Digital card out of order! Hardware defect detected Contact the Service Department, replace card
F261 Fieldbus card out of order! Hardware defect detected Check contacts of Anybus card, contact the Service
Department
M284 Firmware update Firmware has been updated No action required. Message can be acknowledged
F301 Error: Cannot load setup Setup defective Switch the device off and then on again, re-configure,
contact the Service Department if necessary
M302 Setup restored from backup Setup has been loaded from backup Check setup
F303 Error: Device data Device data defective Contact the Service Department
M304 Backup: Device data Device data defective. However it was possible Check settings (e.g. serial number)
to continue working with the backup data
F309 Error: Date/time is not set Invalid date/time (e.g. internal battery is empty) Device was switched off too long. The date/time must
be set again. Battery might need to be replaced
(contact the Service Department)
F310 Error: Cannot save setup The setup could not be saved Contact the Service Department
F311 Error: Device data The device data could not be saved Contact the Service Department
F312 Error: Calibration data The calibration data could not be saved Contact the Service Department
defective
Endress+Hauser 67
Diagnostics and troubleshooting Memograph M, RSG45
F312 Analog card x is not Analog card x is not calibrated! The device Contact the Service Department
calibrated! works with default values, i.e. the measured
values could possibly be inaccurate under
certain circumstances
M313 SRAM has been SRAM was defragmented after firmware update No action required. Message can be acknowledged
defragmented
F314 Error: Option code Activation code is no longer correct (incorrect Enter new code
serial number/program name). Option has been
switched off and setup preset has been
performed
M315 No IP address could be No IP address could be obtained from the DHCP Check the network cable
obtained from the DHCP server!
server!
M316 Invalid MAC address! No or incorrect MAC address Contact the Service Department
M317 Battery voltage < 2 V. Battery needs to be replaced (contact the Service
Please replace battery! Department)
F348 Firmware cannot be Firmware update has been aborted because the Contact the Service Department
updated: firmware file is damaged or is not compatible
• Checksum incorrect with this device
• Firmware incompatible!
M350 Measured value acquisition Measured value acquisition was interrupted/ No action required. Message can be acknowledged
interrupted for calibration/ reactivated for service/maintenance purposes.
service work Causes include:
Measured value acquisition • Calibration of inputs/outputs
restarted • Firmware update
M351 The device will restart The device is rebooting. No action required. Message can be acknowledged
Causes include:
• Following a firmware update
• Change of device options
M502 Device is locked! Device is locked! The message appears when an Check lock per digital channel
attempt is made to update the firmware, for
example
F510 Setup was corrected. The device detected that the configuration was Please check the configuration of the device. If
no longer correct. All the parameters affected hardware has been replaced, no other action is needed
have been reset to the factory default settings.
(recommendation: change the operating language so
that the error message no longer appears after the
Possible causes:
• Input cards have been removed or replaced by next restart)
another type
• An input card is no longer working correctly
A firmware update has caused compatibility
problems
Note: This error message appears each time the
device is restarted until at least one change has
been made to the configuration
F510 User administration was The device detected that the user
corrected administration settings were no longer correct.
All the parameters affected have been reset to
the factory default settings
M520 SMTP: Name could not be Problem with name resolution (DNS). Check the corresponding settings
resolved (DNS)! SMTP: e-mail
SNTP: Name could not be SNTP: time synchronization
resolved (DNS)!
68 Endress+Hauser
Memograph M, RSG45 Diagnostics and troubleshooting
F526 • Support points not OK The support points for the specified Check support points
• Support points: x-value linearization table are not plausible
exists more than once
• Upper and lower support
points are identical
M528 Setup is not compatible with An attempt was made to load a setup which is Check whether the correct file has been selected
this firmware! not compatible with this firmware (e.g. another
device type)
M530 Cannot copy setup An error occurred when a setup was loaded Replace the SD card or USB stick
from an SD card or USB stick Setup file defective?
An error occurred when a setup was saved to an
SD card or USB stick
S901 Input signal too low Input signal too low Check Connections and parameters. Check connected
sensor/transmitter
S902 Input signal too high Input signal too high Check Connections and parameters. Check connected
sensor/transmitter
M905 Set point x Set point x has been violated Note: Error number only occurs if e-mails are sent
M906 End limit value x Set point x no longer violated Note: Error number only occurs if e-mails are sent
F910 This software is not enabled The current firmware is not enabled for this Contact the Service Department
for this device hardware
M920 Too many messages that There are too many messages that need to be Acknowledge messages
need to be acknowledged! acknowledged. Another message could not be
added
M922 No cyclic measurement The instantaneous values were not read out for
transfer a set time
M922 No cyclic transfer The device was not read out via fieldbus for a Check the communication of the fieldbus. Check PLC
configurable time
M923 Error in printing Various problems while printing, e.g.: Check the printer status on the printer
• Printer spooler is full
• Printer out of paper
• Printer not ready
• Please replace the toner/ink
M924 Error accessing SD card! Impossible to access the removable data Check/replace removable data medium
Error accessing USB stick! medium. Causes include:
SD card is not or wrong Memory is larger than 32 GB
formatted! Invalid format (only FAT or FAT32 are
USB stick is not or wrong permitted)
formatted!
M927 Insufficient space free on An attempt was made to save to the SD card or Use other SD card/USB stick. Delete files that are no
data storage medium! USB stick (setup, screenshot, etc.), but not longer needed from the SD card/USB stick
enough free memory space is available
M927 Insufficient space free on An attempt was made to save the data onto the Use other WebDAV server. Delete files no longer
data storage medium! WebDAV server, but there was not sufficient required from the WebDAV server
memory available
Endress+Hauser 69
Diagnostics and troubleshooting Memograph M, RSG45
F929 File is damaged! The file that should be loaded is damaged/ Create file again, use other storage medium
invalid (e.g. wrong checksum).
This message can occur in connection with the
following actions, for instance:
Loading setup from SD card/USB stick
Firmware update
Loading process-related graphics
M940 E-mail could not be sent! (x) E-mail could not be sent! Check settings/network connection
Optional: Error code (x) from server: e.g.: • 451: Try again
• 451: Requested action aborted: local • 554: Use other e-mail provider
processing error
• 554: Transaction failed. Possible reason: e-
mail was not sent as suspected of being
SPAM
• 1: No free buffer
• 2: No receiver specified
M941 No connection to the e-mail A connection to the e-mail server could not be Check settings/network connection
server! established because:
• The entered connection data are incorrect
• The connection has been disconnected
M942 SMTP: fault occurred (x). An error occurred when sending an e-mail. Check settings/network connection
x= error code:
0: SMTP was switched off when the mail was
being sent
3: TCP/IP connection was denied
4: TCP/IP connection error
5: SMTP server denied
6: Error during authentication
7: Connection unexpectedly lost
8: Server responded with error code
9: Timeout
10: Internal protocol error
M945 SNTP: Time was not Time could not be synchronized via SNTP. • Check settings
synchronized! • Check whether the error occurs often. If it does,
Possible reasons:
choose another time server
• SNTP server temporarily unavailable
• Settings not correct
M945 SNTP server 1 not Time could not be synchronized via SNTP. • Check settings
responding. Try server 2 • Check whether the error occurs often. If it does,
Possible reasons:
choose another time server
• SNTP server temporarily unavailable
• Settings not correct
M946 Screenshot could not be Screenshot could not be created. Check/replace the SD card or USB stick
saved (x)! Possible causes (x):
0: Error when writing
1: Insufficient free space
2: Bitmap could not be created
3: SD card/USB stick not available or ready
M947 Modem could not be The connected modem could not be initialized Please check the cable and modem
initialized! Please check the by the device
cable and modem
M950 Unable to load SSL Unable to load SSL certificate. Cause: • Use a certificate with a valid file format
certificate • Import certificate to device again
• Invalid file format
• File is damaged
F951 SSL certificate '…' has Certificates have an expiry date, i.e. they must Install a new certificate
expired! be renewed from time to time
M952 SSL certificate '…' expires The device warns the user shortly before the Install a new certificate
on ...! certificate expires
70 Endress+Hauser
Memograph M, RSG45 Diagnostics and troubleshooting
M953 x certificates have already The device can manage max. 3 X.509 Delete a certificate that is already installed and no
been installed. First of all, certificates longer required
please delete certificates
that are no longer required
M954 SSL certificate not found: Unable to establish SSL connection as a suitable Install a suitable certificate
key ID = … certificate is not installed
M956 Incorrect password. Your Incorrect password. User account has been Please contact your administrator to reactivate your
user account has been deactivated account
deactivated
M956 Incorrect password. Your An incorrect password was entered and the Wait until the temporary block has been disabled or
user account has been account has been temporarily blocked contact your administrator
deactivated for 10 minutes
M957 Wet steam alarm Warning for wet steam alarm Check the application (pressure, temperature inputs)
M965 SMS could not be sent SMS could not be sent because: Check connections and communication settings
• The entered connection data are incorrect
• There is no connection to the service provider
M971 No channels assigned to The batch functionality has been activated but Check group settings
batch x no channels have been assigned to the batch
M980 No connection to the A connection to the WebDAV server could not Check settings/network connection
WebDAV server be established because the entered connection
data are incorrect or the connection has been
disconnected
M982 WebDAV: Directory or file Set directory path does not exist Create a directory manually in the WebDAV server
could not be created
M490 Channel x: Max. 5 devices Max. 5 HART® devices may be connected to the Use other channels
may be connected per input
channel in Multidrop mode.
M960 Value uncertain/ In the case of fieldbuses: The status of the value
communication faulty is uncertain
In the case of HART®: The current value is used
instead of the digital value
Endress+Hauser 71
Diagnostics and troubleshooting Memograph M, RSG45
12.10 Simulation
Various functions/signals can be simulated for test purposes here.
NOTICE
Selecting simulation: Simulation of the relays can be found in the main menu under
"Diagnostics -> Simulation". The simulation of the measured values can be found in
the main menu under "Expert -> Diagnostics -> Simulation".
Only the simulated values are recorded during simulation. The simulation is recorded in
the event logbook.
‣ Do not start simulation if measured value recording must not be interrupted!
72 Endress+Hauser
Memograph M, RSG45 Diagnostics and troubleshooting
Endress+Hauser 73
Diagnostics and troubleshooting Memograph M, RSG45
74 Endress+Hauser
Memograph M, RSG45 Maintenance
13 Maintenance
No special maintenance work is required for the device.
13.3 Cleaning
The front of the housing can be cleaned with a clean dry or damp cloth.
Endress+Hauser 75
Repairs Memograph M, RSG45
14 Repairs
1 2
3
4
6
5
A0025589
3 Digital board extension (8x digital inputs + 6x relays + 2x analog outputs) XPR0011-A7
for slot 5 (rear panel must also be ordered with the extension)
RS232/RS485 adapter set, DIN rail, 230 V AC, galvanic isolation + RSG40A-S6
interface cable for PC/modem
76 Endress+Hauser
Memograph M, RSG45 Repairs
RS232/RS485 adapter set, DIN rail, 150 V AC, galvanic isolation + RSG40A-S7
interface cable for PC/modem
3 Terminals:
Terminal plug 3 pole for mains connection "N L PE" RM5.08 – orange color 71123475
at slot 6
Terminal plug-in 4 pole FMC1,5/4-ST-3,5 for digital board extension, slot 71037350
5 (analog outputs)
Terminal plug-in 6 pole FMC1,5/6-ST-3,5 for analog input at slots 1-5 51009211
Software:
Without software application XPR0012-0 _ _ _
Mathematics XPR0012-B _ _ _
Tele-Alarm + mathematics XPR0012-C _ _ _
Batch + mathematics XPR0012-D _ _ _
Wastewater + stormwater overflow tank + Tele-Alarm + mathematics XPR0012-E _ _ _
Energy software, water + steam + mathematics XPR0012-F _ _ _
Tele-Alarm + energy software + water + steam + mathematics XPR0012-G _ _ _
Option:
Standard XPR0012-_ _ _ A
The software option can be enabled directly on the device. After ordering you will
receive instructions and a code that you must enter.
14.3 Return
The measuring device must be returned if it is need of repair or a factory calibration, or if
the wrong measuring device has been delivered or ordered. Legal specifications require
Endress+Hauser, as an ISO-certified company, to follow certain procedures when handling
products that are in contact with the medium.
To ensure safe, swift and professional device returns, please refer to the procedure and
conditions for returning devices provided on the Endress+Hauser website at
http://www.endress.com/support/return-material
Endress+Hauser 77
Repairs Memograph M, RSG45
14.4 Disposal
The device contains electronic components and must therefore be disposed of as electronic
waste. Please pay particular attention to the national disposal regulations in your country.
78 Endress+Hauser
Memograph M, RSG45 Accessories
15 Accessories
Always quote the serial number of the device when ordering accessories! Installation
instructions are provided with the accessory!
Various accessories, which can be ordered with the device or subsequently from Endress
+Hauser, are available for the device. Detailed information on the order code in question is
available from your local Endress+Hauser sales center or on the product page of the
Endress+Hauser website: www.endress.com.
Field Data Manager analysis software with SQL database support (1 x workstation MS20-A1
license, Professional version)
Designation:
Cable set RS232 for connection to PC or modem RXU10-B _
USB - RS232 converter RXU10-E _
Cable USB-A - USB-B, 1.8 m (5.9 ft) RXU10-F _
Configuration software "FieldCare Device Setup" + USB cable RXU10-G _
320 (12.6)
320 (12.6)
254 (10)
A0024766
14 Dimensions in mm (in)
Endress+Hauser 79
Accessories Memograph M, RSG45
293.4 (11.6)
173 (6.81)
188 (7.4)
236 (9.29) 184 (7.24)
212.6 (8.37)
293.4 (11.6)
173 (6.81)
236 (9.29)
A0024767
15 Dimensions in mm (in)
Version:
Standard RXU10- _ 1
Neutral RXU10- _ 2
A0029023
80 Endress+Hauser
Memograph M, RSG45 Technical data
16 Technical data
Measuring principle Electronic acquisition, display, recording, analysis, remote transmission and archiving of
analog and digital input signals as well as calculated values.
The device is intended for installation in a panel or cabinet door. There is also the option of
operating it in a desktop housing or field housing.
Measuring system Multichannel data recording system with multicolor TFT display (178 mm / 7" screen size),
internal memory, external memory (SD card and USB stick), galvanically isolated universal
inputs (U, I, TC, RTD, pulse, frequency), HART® inputs, digital inputs, transmitter power
supply, limit relays, digital and analog outputs, communication interfaces (USB, Ethernet,
RS232/485), optionally available with Modbus, Profibus DP or PROFINET I/O or
EtherNet/IP.
An Essential Version of the Field Data Manager (FDM) software is included for SQL-
supported data analysis at the PC.
The number of inputs available in the basic device can be individually increased using
a maximum of 5 plug-in cards. The device supplies power directly to connected two-
wire transmitters. The device is configured and operated via the navigator (jog/shuttle
dial) or by touchscreen (optional), using the integrated Web server and a PC, or via an
external USB keyboard or mouse or with the FieldCare / DeviceCare configuration
software. Online help supports the user during local operation.
Ex version:
• The hazardous area version (Ex version) is only available in conjunction with the
stainless steel front and touch control.
• In this version, the SD card is integrated in the device and cannot be removed. The
card can be read out using the Field Data Manager (FDM) software supplied via USB
or Ethernet, or by WebDAV.
Application packages / In the standard version, the Advanced Data Manager has a variety of functions, including
software options an end-to-end safety concept to meet the requirements of FDA 21 CFR Part 11. The
following application packages are available to help users meet the requirements of their
applications and save time:
• Mathematics
• Telealarm
• Batch management
• Wastewater + RSB (rain spillway basin)
• Energy calculation
The application packages contain the standard functions and the specific package
functions. The individual packages can be largely combined as the user requires. The
application packages can also be activated retroactively by entering the activation code.
Standard functions
• Signal analysis: external, 1 min to 12 h, day, week, month, year
• Web server
• User administration compliant with FDA 21 CFR Part 11
• Event log/audit trail
• Process screen
• Operation time counter
• Text entry/comments
• Change language
Endress+Hauser 81
Technical data Memograph M, RSG45
• Time synchronization
• Linearization
• Access protection through release code
• E-mail notification in event of alarms and limit violation
• Encrypted e-mail transmission via SSL (TLS)
• Operation via external USB keyboard and mouse
• External USB or network printer
Mathematics
With the mathematics package, measured values of the inputs or the results of other math
channels can be linked mathematically. A formula with up to 200 characters can be
created using a formula editor. Once entered, the user can then check the plausibility of
the formula.
Functions:
• 12 math channels
• Mathematics functions via formula editor
• Basic arithmetic operations, relational operators, logic operations and functions
Telealarm software
The Telealarm software facilitates user mobility, allowing users to respond to events while
they are on the road. E-mails or SMS messages triggered by process alarms or other
important process events can be sent to several recipients simultaneously or automatically
forwarded to a recipient/destination. Messages can be confirmed, relays controlled
remotely and current values queried by cellular phone. The Advanced Data Manager with
GSM (GPRS) or Ethernet is ideal both for environmental applications to monitor unstaffed
outstations, and for tank monitoring applications.
The Telealarm software contains the mathematics package.
Functions:
• Advanced SMS/e-mail notification in the event of an alarm
• Instantaneous values queried by cellular phone
• Remote relay switching
• Alarm confirmation by SMS
Batch software
Batch management allows users to reliably record and visualize discontinuous processes.
User-definable or externally controlled analysis intervals are possible for up to four
batches simultaneously. Batches are assigned batch-specific values and the measured data,
the start, end and duration of every batch, along with the current batch status, are
displayed on the device and in the Field Data Manager software. At the end of the batch, a
batch print-out is automatically started directly at the device (USB or network printer) or is
printed out at a PC with the Field Data Manager software.
The batch software contains the mathematics package.
Functions:
• Batch report for 4 batches simultaneously
• USB barcode reader
• Automatic batch printout
• Preset counter
82 Endress+Hauser
Memograph M, RSG45 Technical data
quantity channel. Infiltration water recording and the monitoring of rain spillway basins
for reservoir and overflow events are also functions of this software option.
The water/wastewater software contains the mathematics package and the telealarm
software.
Functions:
• Rain spillway basin (reservoir/overflow)
• Highest and lowest values for quantities
• Highest and lowest values from ¼-hourly averages
• Determination of infiltration water
Additional functions:
• 12 math channels
(Channels 1-8: energy-specific formulas and formula editor, channels 9-12: formula
editor)
• Heat quantity + mass calculation for water and steam applications
• Efficiency calculation
Dependability Reliability
Depending on the device version, the mean time between failures (MTBF) is between 52
years and 16 years (calculated based on SN29500 standard at 40°C)
Maintainability
Battery-backed time and data memory. It is advisable to have the backup battery replaced
by a service technician after 10 years.
Endress+Hauser 83
Technical data Memograph M, RSG45
Safety
Recorded data are saved in a tamper-proof format and can be exported and archived with
manipulation protection using the Field Data Manager software.
16.2 Input
HART® inputs
Standard version without HART® inputs. Optional HART® input cards (slot 1-5) with 4
inputs (4/8/12/16/20) each.
Both the digital HART® values and the 4 to 20 mA signal can be evaluated at every input.
The 4 HART® values (PV, SV, TV, QV) of a sensor can be evaluated and the analog HART®
value (PV) can be measured via the digital HART® signal. Up to 40 digital HART® values
can be recorded in total. It is possible to access the HART® sensor in the field from a PC
tool (e.g. FieldCare). In this way, the sensor can be configured from the control room and
the status information of the sensor can be analyzed/displayed. The Memograph M acts as
a HART® Gateway.
Access to the connected sensors is only possible if the device is connected by Ethernet.
Port 5094 must be open in the firewall.
Digital inputs
Standard version: 6 digital inputs
Optional digital card (slot 5): 8 additional digital inputs, 6 additional relays and 2 analog
outputs
84 Endress+Hauser
Memograph M, RSG45 Technical data
Math channels
12 math channels (optional). Mathematics functions can be freely edited via a formula
editor.
Integration of calculated values e.g. for totalization.
Limit values
60 limit values (individual channel assignment)
Calculated values
The values of the universal and HART® inputs can be used to perform calculations in the
math channels.
The results of the math channels can also be used for calculations in other math channels.
Measuring range According to IEC 60873-1: An additional display error of ±1 digit is permitted for every
measured value.
Resistance Pt100: -200 to 850 °C (-328 to 1562 °F) (IEC 60751:2008, α=0.00385) 4-wire: ±0.1% oMR
thermometer Pt100: -200 to 510 °C (-328 to 950 °F) (JIS C 1604:1984, α=0.003916) 3-wire: ±(0.1% oMR + 0.8 K)
(RTD) Pt100: -200 to 850 °C (-328 to 1562 °F) (GOST 6651-94, α=0.00391) 2-wire: ±(0.1% oMR + 1.5 K)
Pt500: -200 to 850 °C (-328 to 1562 °F) (IEC 60751:2008, α=0.00385) Temperature drift: ±0.01%/K oMR
Pt500: -200 to 510 °C (-328 to 950 °F) (JIS C 1604:1984, α=0.003916)
Pt1000: -200 to 600 °C (-328 to 1112 °F) (IEC 60751:2008, α=0.00385)
Pt1000: -200 to 510 °C (-328 to 950 °F) (JIS C 1604:1984, α=0.003916)
Cu50: -50 to 200 °C (-58 to 392 °F) (GOST 6651-94, α=4260) 4-wire: ±0.2% oMR
Cu50: -200 to 200 °C (-328 to 392 °F) (GOST 6651-94, α=4280) 3-wire: ±(0.2% oMR + 0.8 K)
Pt50: -200 to 1100 °C (-328 to 2012 °F) (GOST 6651-94, α=0.00391) 2-wire: ±(0.2% oMR + 1.5 K)
Cu100: -200 to 200 °C (-328 to 392 °F) (GOST 6651-94, α=4280) Temperature drift: ±0.02%/K oMR
Pt46: -200 to 1100 °C (-328 to 2012 °F) (GOST 6651-94, α=0.00391) 4-wire: ±0.3% oMR
Cu53: -200 to 200 °C (-328 to 392 °F) (GOST 6651-94, α=4280) 3-wire: ±(0.3% oMR + 0.8 K)
2-wire: ±(0.3% oMR + 1.5 K)
Temperature drift: ±0.02%/K oMR
Thermocoupl Type J (Fe-CuNi): -210 to 1200 °C (-346 to 2192 °F) (IEC 60584:2013) ±0.1% oMR from -100 °C (-148 °F) ≥1 MΩ
es (TC) Type K (NiCr-Ni): -270 to 1300 °C (-454 to 2372 °F) (IEC 60584:2013) ±0.1% oMR from -130 °C (-202 °F)
Type L (NiCr-CuNi): -200 to 800 °C (-328 to 1472 °F) (GOST R8.585:2001) ±0.1% oMR from -100 °C (-148 °F)
Type L (Fe-CuNi): -200 to 900 °C (-328 to 1652 °F) (DIN 43710-1985) ±0.1% oMR from -100 °C (-148 °F)
Type N (NiCrSi-NiSi): -270 to 1300 °C (-454 to 2372 °F) (IEC 60584:2013) ±0.1% oMR from -100 °C (-148 °F)
Type T (Cu-CuNi): -270 to 400 °C (-454 to 752 °F) (IEC 60584:2013) ±0.1% oMR from -200 °C (-328 °F)
Temperature drift: ±0.01%/K oMR
Endress+Hauser 85
Technical data Memograph M, RSG45
Type A (W5Re-W20Re): 0 to 2500 °C (32 to 4532 °F) (ASTME 988-96) ±0.15% oMR from 500 °C (932 °F) ≥1 MΩ
Type B (Pt30Rh-Pt6Rh): 42 to 1820 °C (107.6 to 3308 °F) (IEC 60584:2013) ±0.15% oMR from 600 °C (1112 °F)
Type C (W5Re-W26Re): 0 to 2315 °C (32 to 4199 °F) (ASTME 988-96) ±0.15% oMR from 500 °C (932 °F)
Type D (W3Re-W25Re): 0 to 2315 °C (32 to 4199 °F) (ASTME 988-96) ±0.15% oMR from 500 °C (932 °F)
Type R (Pt13Rh-Pt): -50 to 1768 °C (-58 to 3214 °F) (IEC 60584:2013) ±0.15% oMR from 100 °C (212 °F)
Type S (Pt10Rh-Pt): -50 to 1768 °C (-58 to 3214 °F) (IEC 60584:2013) ±0.15% oMR from 100 °C (212 °F)
Temperature drift: ±0.01%/K oMR
Pulse input Min. Pulse length 40 μs, max. 12.5 kHz; 0 to 7 mA = LOW; 13 to 20 mA = HIGH Load: 50 Ω
(I) 1) ±1 Ω
Frequency 0 to 10 kHz, overrange: up to 12.5 kHz; 0 to 7 mA = LOW; 13 to 20 mA = HIGH ±0.02% @ f <100 Hz of reading
input (I) 1) ±0.01% @ f ≥100 Hz of reading
Temperature drift: 0.01% of
measured value over the entire
temperature range
1) If a universal input is used as a frequency or pulse input, a series resistor must be used in series connection with the voltage source. Example: 1.2
kΩ series resistor at 24 V
Limit values for input voltage and current as well as cable open circuit detection/line influence/temperature compensation:
Measured variable Limit values (steady-state, without Cable open circuit detection/line influence/temperature compensation
destroying input)
Current (I) Maximum permitted input voltage: 2.5 V 4 to 20 mA range with disengageable cable open circuit monitoring to NAMUR
Maximum permitted input current: 50 mA NE43. The following error ranges apply when NAMUR NE43 monitoring is
switched on:
≤3.8 mA: underrange
≥20.5 mA: overrange
≤ 3.6 mA or ≥ 21.0 mA: open circuit (display shows: – – – –)
Pulse, frequency (I) Maximum permitted input voltage: 2.5 V No cable open circuit monitoring
Maximum permitted input current: 50 mA
Voltage (U) >1 V Maximum permitted input voltage: 35 V 1 to 5 V range with disengageable cable open circuit monitoring:
<0.8 V or >5.2 V: cable open circuit (display shows: - - - -)
Thermocouples (TC) Maximum permitted input voltage: 24 V Influence of line resistance: <0.001%/Ω
Error, internal temperature compensation: ≤ 2 K
86 Endress+Hauser
Memograph M, RSG45 Technical data
Limit values for input voltage and current as well as cable open circuit detection:
Measured variable Limit values (steady-state, without Cable open circuit detection
destroying input)
Current (I) Maximum permitted input voltage: 0.5 V 4 to 20 mA range with disengageable cable open circuit monitoring to NAMUR
Maximum permitted input current: 50 mA NE43. The following error ranges apply when NAMUR NE43 monitoring is
switched on:
≤3.8 mA: underrange
≥20.5 mA: overrange
≤ 3.6 mA or ≥ 21.0 mA: open circuit (display shows: – – – –)
Scan rate
Current/voltage/pulse/frequency input: 100 ms per channel
Thermocouples and resistance temperature detector: 1 s per channel
Endress+Hauser 87
Technical data Memograph M, RSG45
The available storage capacity of the internal and external memory can be displayed
in the main menu under "Diagnostics → Device information → Memory
information". The storage capacity depends on the specific device set-up.
Converter resolution
24 bit
Totalization
The interim value, daily value, weekly value, monthly value, annual value and overall value
can be determined (13-digit, 64 bit).
Analysis
Recording of quantity/operating time (standard function), also a min/max/median
analysis within the set time frame.
Digital inputs Input level Logical "0" (corresponds to –3 to +5 V), activation with logical "1" (corresponds to
+12 to +30 V)
Selectable functions
• Functions of the digital input: control input, ON/OFF message, pulse counter (13-digit,
64-bit), operating time, message+operating time, quantity from time, Profibus DP,
EtherNet/IP, PROFINET.
• Functions of the control input: start recording, screensaver on, lock setup, time
synchronization, change group, limit value monitoring on/off, individual LV on/off,
block keyboard/navigator, start/stop analysis.
Additionally for the batch software option: reset batch number, batch limit values on/
off.
16.3 Output
Auxiliary voltage output The auxiliary voltage output can be used for loop power supply or to control the digital
inputs. The auxiliary voltage is short-circuit proof and galvanically isolated.
88 Endress+Hauser
Memograph M, RSG45 Technical data
Load: ≥1 kΩ
Relay outputs A mix of low voltage (230 V) and safety extra low voltage (SELV circuits) is not permitted
at the connections of the relay contacts.
Alarm relay
1 alarm relay with changeover contact.
Standard relay
5 relays with NO contact, e.g. for limit value messages (can be configured as NC contact).
Optional relays
Optional digital card (slot 5): 6 additional relays with NO contact e.g. for limit value alarms
(can be configured as an NC contact).
Switching cycles
>105
Endress+Hauser 89
Technical data Memograph M, RSG45
Galvanic isolation All inputs and outputs are galvanically isolated from each other and designed for the
following test voltages:
Relay Digital Analog Analog Ethern RS232/ USB Auxiliary voltage output
in in/ out et RS485
HART®
Digital in 2 kVDC 500 500 500 500 500 VDC 500 VDC 500 VDC
VDC VDC VDC VDC
but: 1)
Analog 2 kVDC 500 500 500 500 500 VDC 500 VDC 500 VDC
in/HART® VDC VDC VDC VDC
Analog 2 kVDC 500 500 500 500 500 VDC 500 VDC 500 VDC
out VDC VDC VDC VDC
Ethernet 2 kVDC 500 500 500 - 500 VDC 500 VDC 500 VDC
VDC VDC VDC
RS232/ 2 kVDC 500 500 500 500 - 500 VDC 500 VDC
RS485 VDC VDC VDC VDC
USB 2 kVDC 500 500 500 500 500 VDC Galvanicall 500 VDC
VDC VDC VDC VDC y
connected
Auxiliary 2 kVDC 500 500 500 500 500 VDC 500 VDC -
voltage VDC VDC VDC VDC
output
1) Test voltage applies between inputs on power unit (terminals D11 to D61) to inputs on optional digital
card (terminals D71 to DE1). The inputs are galvanically connected on the same plug-in connector.
90 Endress+Hauser
Memograph M, RSG45 Technical data
at the device). Appropriate internal device wiring measures must be taken to allow
unrestricted operation when EMC interference is present. These measures have been
taken into account for this device. Operation in the event of disturbance variables as per
NAMUR NE21 is thus guaranteed.
Where applicable, national installation regulations and guidelines must be observed
during the installation! Where there are large differences in potential between the
individual grounding points, only one point of the shielding is connected directly with the
reference ground.
If the shielding of the cable is grounded at more than one point in systems without
potential matching, mains frequency equalizing currents can occur. These can damage
the signal cable or significantly impact signal transmission. In such cases the shielding
of the signal cable is to be grounded on one side only, i.e. it may not be connected to
the ground terminal of the housing. The shield that is not connected should be
insulated!
R12
R13
N/-
PE
R21
R22
R31
R32
R41
R42
R51
R52
R61
R62
R11
24V
Out
GND1
D21
D31
D41
D51
D61
D11
+
-
6
GND2
GND2
DC1
DD1
O15
O16
O25
O26
DA1
DB1
DE1
D71
D81
D91
5
RG
RC
RD
RH
RA
RB
RE
RK
RF
RL
RJ
RI
Ch15
Ch16
Ch13
Ch14
G1
4
G6
D1
D6
F1
F6
E1
E6
…
…
…
.
.
.
.
.
.
.
.
Ch12
Ch11
Ch10
Ch9
3
C1
C6
A1
A6
91
96
B1
B6
…
…
.
.
.
.
.
.
.
.
Ch7
Ch8
Ch6
Ch5
2
71
76
81
86
61
66
51
56
…
…
…
…
.
.
.
.
.
.
.
.
Ch4
Ch3
Ch1
Ch2
1
41
46
31
36
21
26
16
11
…
…
…
…
…
…
.
.
.
.
.
.
.
.
A0024605
Supply voltage • Extra-low voltage power supply unit ±24 V AC/DC (-10% / +15%) 50/60Hz
• Low voltage power supply unit 100 to 230 V AC (±10%) 50/60Hz
Overcurrent protection (rated current ≤ 10 A) must be installed for the power cable.
Endress+Hauser 91
Technical data Memograph M, RSG45
The power actually consumed depends on the individual operating state and the device
version (LPS, USB, brightness of screen, number of channels, etc). The active power here is
approx. 3 W to 25 W.
Power supply failure Battery-backed time and data memory. The device starts automatically following a power
failure.
92 Endress+Hauser
Memograph M, RSG45 Technical data
8
Termination
RL R= resistor
9 (typical 120 Ω) RS 485
RxD/TxD(+)
RxD/TxD(-)
Further units
12-24 VDC
>5 mA Digital in (D)
Analog +
_
24 V -
Power supply outputs (O) GNDx GNDx
100-230 VAC 24 V AC/DC (Selectable in unit setup)
(±10%) (-10%; +15%) _
Dx1
1 2
GNDx
50 / 60 Hz 50 / 60 Hz Pulse + - + - Dx1 Dx1
L+ N- PE 12-24 VDC
10k +
1 2
0...20 mA; + - + - -+ 24 V +
Dx1
GNDx
4...20 mA
>20 ms
O15
O16
O25
O26
Rel. max. 250 V / 3 A
L/+
N/-
PE
24V 24V Out: max. 250 mA
R41
R42
R51
R52
R61
R62
R12
R13
R21
R22
R31
R32
R11
Out
GND1
D21
D31
D41
D51
D61
D11
-
+
Slot 6: Power supply slot (standard)
GND2
GND2
DC1
DD1
O15
O16
O25
O26
DA1
DB1
DE1
D71
D81
D91
Slot 5: analog input / HART 17-20 or digital-I/O (option)
RG
RC
RD
RH
RA
RB
RE
RK
RF
RL
RJ
RI
RS 232
Ch15
Ch16
Ch13
Ch14
D6
F1
E6
G1
G6
F6
8 9
…
…
…
…
…
…
…
…
…
…
…
…
…
…
…
…
Ch10
Ch12
Ch11
Ch9
C6
91
C1
B1
B6
96
A1
A6
…
…
…
…
…
…
…
…
…
…
…
…
…
…
…
…
1 5
Ch7
Ch6
Ch5
Ch8
2 3 5 8 9
Slot 2: analog input / HART 5-8 (option)
81
71
51
61
86
56
76
66
…
…
…
…
…
…
…
…
…
…
…
…
…
…
…
…
Ch3
Ch1
Ch4
Ch2
31
41
36
26
16
46
…
…
…
…
…
…
…
…
…
…
…
…
…
…
…
…
5 1
9 6
To PC: Cable with 6 9 Slot 0: CPU (standard)
9 pol. Sub-D plug Ethernet USB RS 232 / RS 485 Bus Interface
1 5
RxD - 3 GND - 5 LED LED
23 5 8 9 green yellow
TxD - 2
Analog inputs
x1 x2 x3 x4 x5 x6 x1 x2 x3 x4 x5 x6 x1 x2 x3 x4 x5 x6
+
TC A, B, C, D, J, K, L, N, T, R, S
-
x1 x2 x3 x4 x5 x6
24 V >40 µs
R=1.2 k
+
+
A0026669-EN
17 For connection examples of the HART® inputs (optional), see the Operating Instructions → 23
Endress+Hauser 93
Technical data Memograph M, RSG45
A0019103
100-230 VAC L+ N- PE
24 V AC/DC L+ N- PE
A0019103
The open or close function (= activation or deactivation of the relay coil) in a limit
event can be configured in the setup: "Setup -> Advanced setup -> Outputs -> Relay ->
Relay x". However, in the event of a power failure, the relay adopts its quiescent switch
state regardless of the setting programmed.
Type Terminal
A0019103
94 Endress+Hauser
Memograph M, RSG45 Technical data
Type Terminal
A0019103
If the auxiliary voltage is to be used for the digital inputs, the 24 V out - terminal of
the auxiliary voltage output must be connected with the GND1 terminal.
Type Terminal
Chx
x1
x2
x3
x4
x5
x6
A0019303
x1 x2 x3 x4 x5 x6
1) If a universal input is used as a frequency or pulse input, a series resistor must be used in series connection
with the voltage source. Example: 1.2 kΩ series resistor at 24 V
Endress+Hauser 95
Technical data Memograph M, RSG45
Type Terminal
Modem
Chx
250Ω 10Ω
x1
x2
x3
x4
x5
x6
x1
x2
x3
x4
x5
x6
A0024862
x1 x2 x3 x4 x5 x6
GND2
GND2
DC1
DD1
DA1
DB1
DE1
O15
O16
O25
O26
D71
D81
D91
RG
RC
RD
RH
RA
RB
RE
RK
RF
RL
RJ
RI
A0024736
Relay 7, 8 RA RB RC RD
Relay 9, 10 RE RF RG RH
Relay 11, 12 RI RJ RK RL
1) NO)
2) NO)
The open or close function (= activation or deactivation of the relay coil) in a limit
event can be configured in the setup: "Setup -> Advanced setup -> Outputs -> Relay ->
Relay x". However, in the event of a power failure, the relay adopts its quiescent switch
state regardless of the setting programmed.
Type Terminal
GND2
GND2
DC1
DD1
DA1
DB1
DE1
O15
O16
O25
O26
D71
D81
D91
RG
RC
RD
RH
RA
RB
RE
RK
RF
RL
RJ
RI
A0024736
96 Endress+Hauser
Memograph M, RSG45 Technical data
Type Terminal
GND2
GND2
DC1
DD1
DA1
DB1
DE1
O15
O16
O25
O26
D71
D81
D91
RG
RC
RD
RH
RA
RB
RE
RK
RF
RL
RJ
RI
A0024736
If the auxiliary voltage is to be used for the digital inputs, the 24 V out - terminal of
the auxiliary voltage output (power unit, slot 6) must be connected with the GND2
terminal.
Connector • Panel-mounted device: connected to mains via plug-in screw terminals with reverse
polarity protection
• Desktop version (option): connected to mains via IEC connector
Overvoltage protection To avoid high-energy transients on long signal cables, connect a suitable surge arrester
upstream (e.g. E+H HAW562) in series upstream.
1 x USB port type B (function) on the front of device (only for version with navigator and
front interfaces)
A USB 2.0 port is available on a shielded USB B socket at the front of the device. This can
be used to connect the device for communication with a laptop, for example.
Endress+Hauser 97
Technical data Memograph M, RSG45
98 Endress+Hauser
Memograph M, RSG45 Technical data
• The following baud rates are supported: 9600, 19200, 38400, 57600, 115200
• Max. cable length with a shielded cable: 2 m (6.6 ft) (RS232), or 1000 m (3281 ft)
(RS485)
Only one interface can be used at any one time (RS232 or RS485).
Endress+Hauser 99
Technical data Memograph M, RSG45
Cable open circuit detection, current input Relays, OC, analog output ≤ 1150
1) If internal measuring point temperature compensation is used, otherwise values as for voltage
Long-term drift As per IEC 61298-2: max. ±0.1%/year (of measuring range)
16.6 Installation
100 Endress+Hauser
Memograph M, RSG45 Technical data
158.5 (6.24)
195.2 (7.69) 141.2 (5.56) 27(1.06)
1
(5 38 +1
.43 +0
149.2 (5.87)
.04
)
A
37.1
(1.46)
158.5 (6.24)
196 (7.72) 141.2 (5.56) 31.4 (1.24)
1
(5. 38 +1
43 +0.0
150 (5.91)
4
)
B
208 (8.19)
70 138
(2.76) (5.43)
138 (5.43)
162 (6.38)
C
24 (0.94)
A0024610
Installation dimensions
• Installation depth (excluding terminal cover): approx. 159 mm (6.26 in) for device incl.
terminals and fastening clips.
• Installation depth including terminal cover (option): approx.198 mm (7.8 in)
• Panel cutout: 138 to 139 mm (5.43 to 5.47 in) x 138 to 139 mm (5.43 to 5.47 in)
• Panel thickness: 2 to 40 mm (0.08 to 1.58 in)
• viewing angle range: 50˚ in all directions from the display central axis
• A minimum distance of 12 mm (0.47 in) between the devices must be observed if
aligning the devices vertically above one another or horizontally beside one another.
• The grid dimension of the panel cutouts for multiple devices must be at least
208 mm (8.19 in) horizontally and at least 162 mm (6.38 in) vertically (tolerance not
considered).
• Securing to DIN 43 834
Field housing assembly and As an option, the device can be ordered ready-mounted in a field housing with IP65.
design (optional) Dimensions (B x H x D) approx.: 320 mm (12.6 in) x 320 mm (12.6 in) x 254 mm (10 in)
Endress+Hauser 101
Technical data Memograph M, RSG45
Desktop housing assembly As an option, the device can be ordered ready-mounted in a desktop housing.
and design (optional) Dimensions (B x H x D) approx.: 293 mm (11.5 in) x 188 mm (7.4 in) x 213 mm (8.39 in)
(dimensions with bracket, feet and installed device)
16.7 Environment
Rear IP20
Electromagnetic EMC to all relevant requirements of the IEC/EN 61326- series and NAMUR NE21. For
compatibility details see declaration of conformity.
• Interference immunity: as per IEC/EN 61326 series (industrial environment) / NAMUR
NE21
Maximum measured error <1% of measuring range
• Interference emissions: as per IEC 61326-1 Class A
Weight • Panel-mounted device with navigator and front interfaces (with maximum
configuration): approx. 2.7 kg (5.9 lbs)
• Panel-mounted device with stainless steel front and touchscreen (with maximum
configuration): approx. 3.2 kg (7 lbs)
• Desktop housing (excluding device): approx. 2.3 kg (5 lbs)
• Field housing (excluding device): approx. 4 kg (8.8 lbs)
102 Endress+Hauser
Memograph M, RSG45 Technical data
Seal towards panel wall; seal in flap; seal towards Rubber EPDM 70 Shore A
navigator
Window seal between front frame and glass Rubber EPDM 60 Shore A
Operating concept The device can be operated directly onsite, or via remote configuration with the PC via
interfaces and operating tools (Web server, configuration software).
Web server
A Web server is integrated into the device. The Web server offers the following range of
functions:
• Easy configuration without additional installed software
• Instantaneous value display and diagnostics information
• Display of current measured value curves via Web browser (remote control)
• Loading/saving of device configuration
• Device firmware update
• Printout of device configuration
Endress+Hauser 103
Technical data Memograph M, RSG45
Resolution
Wide VGA 384,000 pixels (800 x 480 pixels)
Backlight
50,000 h half value time (= half brightness)
Number of colors
262,000 viewable colors, 256 colors used
Viewing angle
Max. viewing angle range: 50˚ in all directions from the display central axis
Screen display
• Users can choose between black or white for the background color
• Active channels can be assigned to up to 10 groups. To clearly identify these groups,
they can be assigned a name, such as "temp. boiler 1" or "daily averages".
• Scales linear or logarithmic
• Measured value history: quick retrieval of historic data with zoom function
• Preformatted display formats, such as horizontal or vertical curves, instrument display,
circular chart, process screen, bargraph or digital display.
104 Endress+Hauser
Memograph M, RSG45 Technical data
c b a
5 6 7 8 5 6 7 8
9 9
4 10 4 10
3 1
2
A0024709
19 Device front (left: version with navigator and front interfaces; right: version with stainless steel front and
touchscreen)
c USB A socket "Host" e.g. for USB memory stick, external keyboard, barcode reader or printer
d LED at SD slot. Yellow LED lit or flashing when the device writes to the SD card or reads it.
Do not remove the SD card if the LED is lit or flashing! Risk of data loss!
Endress+Hauser 105
Technical data Memograph M, RSG45
8 In display mode: alternating display indicating the percentage space on the SD card or USB stick that
has already been used.
Status symbols are also displayed in alternation with the memory information (e.g. simulation mode,
data storage active, operation locked, batch active)
In setup mode: the current "direct access" operating code is displayed
9 In display mode: window for measured value display (e.g. curve display).
Display of current measured values and the status in the event of an error/alarm condition. In the case
of counters, the type of counter is displayed as a symbol.
inIf aredmeasuring point has limit value status, the corresponding channel identifier is highlighted
(quick detection of limit value violations). During a limit value violation and device
operation, the acquisition of measured values continues uninterrupted.
10 In display mode: alternating status display (e.g. set zoom range) of the analog or digital inputs in the
appropriate color of the channel.
In setup mode: different information is displayed here depending on the display type.
Languages The following languages can be selected in the operating menu: German, English, Spanish,
French, Italian, Dutch, Swedish, Polish, Portuguese, Czech, Russian, Japanese, Chinese
(Traditional), Chinese (Simplified)
"Field Data Manager • Export of saved data (measured values, analyses, event log) RS232/RS485, USB,
(FDM)" analysis • Visualization and processing of saved data (measured Ethernet
software, SQL values, analyses, event log)
database support • Safe archiving of exported data in a SQL database
(included in the
delivery)
Web server • Display of current and historical data and measured value Ethernet
(integrated into the curves via the web browser
device; access via • Easy configuration without additional installed software
browser) • Remote access to device and diagnostic information
OPC server The following momentary values can be provided: RS232/RS485, USB,
(optional) • Analog channels Ethernet
• Digital channels
• Mathematics
• Totalizer
CE mark The measuring system meets the legal requirements of the applicable EC guidelines. These
are listed in the corresponding EC Declaration of Conformity together with the standards
applied. Endress+Hauser confirms successful testing of the device by affixing to it the CE
mark.
106 Endress+Hauser
Memograph M, RSG45 Technical data
Ex approval Information about currently available Ex versions (ATEX, FM, CSA, etc.) can be supplied by
your E+H Sales Center on request. All explosion protection data are given in separate
documentation which is available upon request.
Ordering information Detailed ordering information is available from the following sources:
• In the Product Configurator on the Endress+Hauser website: www.endress.com -> Click
"Corporate" -> Select your country -> Click "Products" -> Select the product using the
filters and search field -> Open product page -> The "Configure" button to the right of the
product image opens the Product Configurator.
• From your Endress+Hauser Sales Center: www.addresses.endress.com
Product Configurator - the tool for individual product configuration
• Up-to-the-minute configuration data
• Depending on the device: Direct input of measuring point-specific information such
as measuring range or operating language
• Automatic verification of exclusion criteria
• Automatic creation of the order code and its breakdown in PDF or Excel output
format
• Ability to order directly in the Endress+Hauser Online Shop
Endress+Hauser 107
Technical data Memograph M, RSG45
• "Field Data Manager (FDM)" analysis software on DVD (Essential, Demo or Professional
version, depending on order)
• "FieldCare Device Setup / DeviceCare" configuration software on DVD
• Delivery note
• Multilanguage Brief Operating Instructions, hard copy
• Ex Safety Instructions, hard copy (optional)
108 Endress+Hauser
Memograph M, RSG45 Appendix
17 Appendix
Direct access
Description Direct access to active operating items (rapid access). Entering the direct access code takes
you directly to the desired operating parameter. The direct access code is displayed in the
Setup menu on the top right of the display (e.g. 00000-000).
Language
Options German, English, Spanish, French, Italian, Dutch, Polish, Portuguese, Russian, Swedish,
Czech, Japanese, Chinese (Simplified), Chinese (Traditional)
Device tag
Endress+Hauser 109
Appendix Memograph M, RSG45
Temp. unit
Factory setting °C
Decimal separator
Fault switching
Description If the device detects a system error (e.g. hardware defect) or a fault (e.g. cable open
circuit), the selected output switches.
Keyboard layout
110 Endress+Hauser
Memograph M, RSG45 Appendix
Description Please select the keyboard layout. Only relevant if external keyboard is used.
Factory setting No
Paper size
Lock operation
Description Local operation is locked in cases of inactivity once the set time has elapsed to prevent
inadvertent operation (e.g. when cleaning the device). The device is unlocked by pressing
the navigator or the OK operating key for 3 s. When using an external keyboard, the
device is unlocked with the key combination "Ctrl-Alt-Del".
Endress+Hauser 111
Appendix Memograph M, RSG45
LED mode
Description NAMUR NE 44: Green LED -> Power supply OK. Red LED -> Measuring signal failure. Red
LED flashing -> Maintenance required.
NAMUR NE 44+: As per NAMUR NE 44, with red LED in the event of a limit value
violation.
PRESET
Clear memory
112 Endress+Hauser
Memograph M, RSG45 Appendix
Date format
Time format
"Date/time" submenu
(Online configuration)
Description Display of the current UTC time zone is on (UTC = universal time coordinated).
Current date/time
(Online configuration)
Endress+Hauser 113
Appendix Memograph M, RSG45
Navigation Expert → System → Date/time setup → Date/time → Change date/time → UTC time
zone
Direct access code: 120010-000
Description Set your UTC time zone (UTC = universal time coordinated).
Options -12:00, -11:00: Samoa, -10:00: Hawaii, -09:30: Marquesas, -09:00: Alaska, -08:00: LA,
-07:00: Denver, -06:00: Chicago, -05:00: New York, -04:00: Caracas, -03:30: St.John’s,
-03:00: Brasilia, -02:00: Atlantic, -01:00: Azores, +00:00: London, +01:00: Berlin,
+02:00: Cairo, +03:00: Moscow, +03:30: Tehran, +04:00: Abu Dhabi, +04:30: Kabul,
+05:00: Islamabad, +05:30: New Delhi, +05:45: Kathmandu, +06:00: Dhaka, +06:30:
Pyinmana, +07:00: Bangkok, +08:00: Peking, +08:45, +09:00: Tokyo, +09:30: Adelaide,
+10:00: Canberra, +10:30: Lord-Howe, +11:00:Solom.Isl., +11:30: Norfolk, +12:00:
Auckland, +12:45: Chatham, +13:00, +14:00
Date/time
(Online configuration)
Description Set the current date and time for the unit here.
NT/ST changeover
114 Endress+Hauser
Memograph M, RSG45 Appendix
NT/ST region
Occurrence
Description Day, when in the spring a change from normal to summer time occurs.
Visible for NT/ST changeover = Automatic or Manual. Only editable if NT/ST changeover =
Manual.
Day
Endress+Hauser 115
Appendix Memograph M, RSG45
Description Day, when in the spring a change from normal to summer time occurs.
Visible for NT/ST changeover = Automatic or Manual. Only editable if NT/ST changeover =
Manual.
Month
Description Month, when in the spring a change from normal to summer time occurs.
Visible for NT/ST changeover = Automatic or Manual. Only editable if NT/ST changeover =
Manual.
Options January, February, March, April, May, June, July, August, September, October, November,
December
Date
Description Date next spring when a change from normal to summer time occurs.
Only visible if NT/ST changeover = Automatic or Manual. Cannot be edited.
Time
Description Point of time when the time is moved forward by 1 hour on the day of the changeover
from normal time to summer time (in the set time format).
Visible for NT/ST changeover = Automatic or Manual. Only editable if NT/ST changeover =
Manual.
116 Endress+Hauser
Memograph M, RSG45 Appendix
Occurrence
Description Day, when in the autumn a change from summer to normal time occurs.
Visible for NT/ST changeover = Automatic or Manual. Only editable if NT/ST changeover =
Manual.
Day
Description Day, when in the autumn a change from summer to normal time occurs.
Visible for NT/ST changeover = Automatic or Manual. Only editable if NT/ST changeover =
Manual.
Month
Description Month, when in the autumn a change from summer to normal time occurs.
Visible for NT/ST changeover = Automatic or Manual. Only editable if NT/ST changeover =
Manual.
Options January, February, March, April, May, June, July, August, September, October, November,
December
Endress+Hauser 117
Appendix Memograph M, RSG45
Date
Description Date next autumn when summer time changes back to normal time.
Only visible if NT/ST changeover = Automatic or Manual. Cannot be edited.
Time
Description Point of time when the time is moved back by 1 hour on the day of the changeover from
summer time to normal time (in the set time format).
Visible for NT/ST changeover = Automatic or Manual. Only editable if NT/ST changeover =
Manual.
"SNTP" submenu
Description Contains settings for time synchronization using the Simple Network Time Protocol
(SNTP).
SNTP
Description If switched on, time synchronization is carried out via SNTP once a day.
Note: Only possible via Ethernet.
Port 123 must be open in the firewall. The user/network administrator is responsible for
the accuracy of the time server.
Factory setting No
118 Endress+Hauser
Memograph M, RSG45 Appendix
SNTP server 1
Description Please specify the address of the time server (or the IP address).
Note: The DNS server must be configured (see Communication/Ethernet).
Your administrator can provide the address where necessary.
SNTP server 2
Description Shows the IP address of the time server if it was automatically allocated via DHCP. Non-
editable display text.
An attempt is always made to synchronize the time via SNTP server 1 first (provided
it is configured).
DHCP must be switched on (see Communication/Ethernet).
DHCP server: Option 42.
"Security" submenu
Description Contains settings that protect the unit against unauthorized operation and configuration.
Protected by
Options Open access, access code, FDA 21 CFR Part 11, user roles
Access code
Endress+Hauser 119
Appendix Memograph M, RSG45
Description Using this code, setup access can be protected from unauthorized persons. In order to
change any parameter the correct code must be entered. Factory setting: "0", i.e. changes
can be made at any time.
Hint: Make a note of the code and store in a safe place.
Only visible if "Protected by" = "Access code".
Factory setting 0
Description If the unit is protected by an access code a set point code can also be defined. The user can
change the set points once the set point code is entered. All other operating positions
remain locked, however.
Only visible if an access code has been defined.
Factory default: "0" means that alarm set points can only be changed by entering the access
code.
Alarm set point code and access code should not be identical!
Factory setting 0
IP address
Description Enter the IP address or DNS name of the PC on which user administration is running.
Please contact your network administrator if necessary.
Port
120 Endress+Hauser
Memograph M, RSG45 Appendix
Lock hardware
Description Device functions/interfaces that are not used can be switched off for security reasons.
Fieldbuses may also be affected in the case of Ethernet or the serial interface.
Please follow the operating instructions.
Options Ethernet (all ports/services), USB A socket front, USB A socket back, USB B socket front,
Serial interface, SD card
"Authentication" submenu
Description Define the passwords for the various user roles to allow device access.
Visible only if "Protected by" = "User roles".
Operator
ID: operator
Password
Endress+Hauser 121
Appendix Memograph M, RSG45
Administrator
ID: admin
Password
Service
ID: service
Password
Description Contains settings for the external data carrier, amongst other things which data are to be
stored in which format on the external data carrier.
Save as
Description "Protected format": All data are stored in a manipulation protected encrypted format. They
can be interpreted only by the PC analysis software provided.
"Open format": data are stored in a CSV format, this can be opened by a number of different
programs (e.g. MS Excel) (Attention: no manipulation security).
122 Endress+Hauser
Memograph M, RSG45 Appendix
SD card
Memory build-up
Description "Stack memory": no more data can be stored once the data carrier is full.
"Ring memory": once the data carrier is full the oldest data are deleted so that new data can
be stored (First in first out (FIFO).
The "Ring memory" setting refers only to the automatic storage of measured values.
Manual save functions ("Operation -> SD card -> Update/Save measured values") are
not affected.
Warning at
Factory setting 90
Switches relay
Endress+Hauser 123
Appendix Memograph M, RSG45
Description When the warning "Data carrier full" is displayed a relay can also be switched on.
CSV settings
Description Configure which separator is used by your application (e.g. in Excel = semicolon).
Date/time
Description Please select if the date and time should be stored in one column or separate columns
when data are saved in CSV format files.
Operating time
124 Endress+Hauser
Memograph M, RSG45 Appendix
"Messages" submenu
Acknowledging messages
Description The time the message is acknowledged can be saved in the events list.
This setting cannot be altered if the user administration system is activated (FDA 21
CFR Part 11).
Switches relay
Description A relay can be switched as soon as a message that has to be confirmed is displayed (e.g.
on/off message, device errors etc.).
The relay assumes the initial status as soon as all messages have been confirmed.
Description To increase the life span of the LCDs the backlight can be switched off (= screen saver).
Endress+Hauser 125
Appendix Memograph M, RSG45
Screen saver
Options Switched off, On after 10 min, On after 30 min, On after 60 min, Switched daily, Control
input
ON daily from
Description Set time (hh:mm) as of when the screensaver should be switched on (e.g. shift end time).
The screen saver is switched off as soon as the device is operated via onsite
operation. After 1 min of inactivity it switches back on automatically.
Only visible if screen saver = switched daily
Description Set time (hh:mm) as of when the screensaver should be switched off (e.g. shift start time).
Only visible if screen saver = switched daily.
Alarm response
126 Endress+Hauser
Memograph M, RSG45 Appendix
Description "Off on alarm": If limit value violations occur or status signal "Maintenance required (Mxxx)"
or "Check function (Cxxx)" is active, the screensaver will be automatically deactivated.
"Always on": If limit value violations occur or status signal "Maintenance required (Mxxx)"
or "Check function (Cxxx)" is active, the screensaver will not be deactivated.
Active messages that require acknowledgment or an active status signal "Failure
(Fxxx)" or "Out of specification (Sxxx)" deactivate the screensaver always.
Description Settings for barcode reader (only relevant if a USB barcode reader is connected to the
device).
Notes on operating a barcode reader: it must behave like a HID keyboard; texts must
be terminated with a carriage return.
Character set
Save as event
Description The device can save texts read in using a barcode reader in the events list.
Texts are stored as events if one of the following conditions is NOT met:
• A command sequence was read in
• The dialog where batch information is input is active
• The dialog for testing the barcode reader is active
• The "Save text" function is executed
Endress+Hauser 127
Appendix Memograph M, RSG45
Factory setting No
Timeout sequences
Description Specify the number of seconds after which a command sequence is canceled if the
necessary data are not read in.
Factory setting 30
Activation code
(Online configuration)
Description Here, you can enter a code to enable the device options.
Options that can be upgraded can be found under "Spare parts" → 77.
Note: When an activation code is entered, the device is restarted in order to enable the
new option.
• The activation code entered is not displayed, i.e. this parameter is always empty
following a restart.
• Pay attention to case-sensitivity.
Slot 1
(Online configuration)
128 Endress+Hauser
Memograph M, RSG45 Appendix
Slot 2
(Online configuration)
Slot 3
(Online configuration)
Slot 4
(Online configuration)
Endress+Hauser 129
Appendix Memograph M, RSG45
Slot 5
(Online configuration)
Communication
(Online configuration)
Fieldbus
(Online configuration)
Modbus Master
(Online configuration)
130 Endress+Hauser
Memograph M, RSG45 Appendix
Application
(Online configuration)
Front of housing
(Online configuration)
Add input
Endress+Hauser 131
Appendix Memograph M, RSG45
Description Addition of an input that must be switched on and configured according to the input
signal.
Factory setting No
Delete input
Factory setting No
Signal
Description Select the type of signal connected (current, voltage, etc.). The channel is switched off if no
signal type is selected (factory default).
Options Switched off, Current, Voltage, Resistance thermometer (RTD), Thermocouple, Pulse
counter, Frequency input, Profibus DP (option), Modbus Slave (option), Modbus Master
(option), HART (option), EtherNet/IP (option), PROFINET (option)
Range
132 Endress+Hauser
Memograph M, RSG45 Appendix
Description Select the input range or which resistance thermometer/thermocouple is connected. The
respective terminal layout can be found in the Operating Instructions or on the rear of the
unit.
Only visible if signal ≠ switched off.
Value
Description Specify the value, which was read out digitally via HART, that should be recorded/
processed.
Only visible if signal = HART.
Endress+Hauser 133
Appendix Memograph M, RSG45
Navigation Expert → Inputs → Universal inputs → Universal input x → Measured value type
Direct access code: 220022-0xx
Examples: Universal input 1: 220022-000; Universal input 12: 220022-011
Connection
Transmission protocol
Options Modbus TCP, Modbus TCP with slave address, Modbus RTU over TCP
IP address
134 Endress+Hauser
Memograph M, RSG45 Appendix
Port
Slave address
Factory setting 1
Readout function
Description Modbus function with which the registers are to be read out.
Only visible if signal = Modbus Master.
Endress+Hauser 135
Appendix Memograph M, RSG45
Register address
Factory setting 1
Data type
Description Describes the data type of the value received and its sequence of bytes.
Only visible if signal = Modbus Master.
Channel ident.
Plot type
136 Endress+Hauser
Memograph M, RSG45 Appendix
Description The analog inputs are scanned in a 100ms cycle. Depending on the memory cycle, the
selected data are determined, saved, and displayed based on the scanned values.
Options Instantaneous value, Average, Minimum value, Maximum value, Minimum + Maximum,
Counter, Current value + Counter
Time base
Description A current value can be determined from the counter reading with the aid of the time base
e.g. input liters, time base = second → current value = liters/second.
Only visible if signal = "Pulse counter" and plot type = "Current value + counter".
Engineering unit
Description Specify the technical (physical) unit for the measuring point connected to this input.
Only visible if signal ≠ switched off.
Unit/dimension counter
Endress+Hauser 137
Appendix Memograph M, RSG45
Pulse counter
Description Please check if the pulse counter used is a fast or slow (up to a max. 25 Hz) counter. For
example if the number of state changes from a relay are to be monitored then you must
set up "up to 25Hz".
Only visible if signal = pulse counter.
Pulse value
Description Factor, that when multiplied by the input signal results in the required physical value.
Example: 1 pulse equals 5 m3-> enter "5".
Only visible if signal = pulse counter.
Factory setting 1
Calc. factor
Description Factor for converting the counter (e.g. the transmitter returns m³/100 -> desired unit is
m³ --> enter 0.01 as the factor)
Only visible if signal = Modbus Master.
138 Endress+Hauser
Memograph M, RSG45 Appendix
Decimal point
Options None, One (X.Y), Two (X.YY), Three (X.YYY), Four (X.YYYY), Five (X.YYYYY)
Navigation Expert → Inputs → Universal inputs → Universal input x → Start value range
Direct access code: 220046-0xx
Examples: Universal input 1: 220046-000; Universal input 12: 220046-011
Factory setting 0
Navigation Expert → Inputs → Universal inputs → Universal input x → End value range
Direct access code: 220047-0xx
Examples: Universal input 1: 220047-000; Universal input 12: 220047-011
Lower frequency
Endress+Hauser 139
Appendix Memograph M, RSG45
Description Configure the lower frequency that corresponds to the start of the measuring range.
Only visible if signal = frequency input.
Range start
Description Transmitters convert the physical measured variable to standardized signals. Enter the
start of the measuring range here.
• The start and end of the measuring range may not be identical.
• The start of the measuring range can also be larger than the end (e.g. for deep
wells).
• The parameter can be defined independently of the number of decimal places
configured for the measured value as these are only taken into consideration for
the display.
Upper frequency
Description Configure the upper frequency that corresponds to the end of the measuring range.
Only visible if signal = frequency input.
140 Endress+Hauser
Memograph M, RSG45 Appendix
Navigation Expert → Inputs → Universal inputs → Universal input x → Meas. range end
Direct access code: 220007-0xx
Examples: Universal input 1: 220007-000; Universal input 12: 220007-011
Description Transmitters convert the physical measured variable to standardized signals. Enter the end
of the measuring range here.
• The start and end of the measuring range may not be identical.
• The end of the measuring range can also be smaller than the start (e.g. for deep
wells).
• The parameter can be defined independently of the number of decimal places
configured for the measured value as these are only taken into consideration for
the display.
Zoom start
Description If the whole value range is not used, you can configure the lower value of the required
section here. Zooming does not affect the save function.
• The zoom can also be set outside the measuring range. The only restriction is that
the start and end of the zoom may not be identical.
• If the signal or the range is changed, the zoom is corrected if it no longer fits the
measuring range.
• The zoom start can also be larger than the zoom end. The device will automatically
rotate the values on the display.
Zoom end
Description Like "Zoom start". However, enter the upper value of the required range here.
• The zoom can also be set outside the measuring range. The only restriction is that
the start and end of the zoom may not be identical.
• If the signal or the range is changed, the zoom is corrected if it no longer fits the
measuring range.
• The zoom end can also be smaller than the zoom start. The device will
automatically rotate the values on the display.
Endress+Hauser 141
Appendix Memograph M, RSG45
Damping
Description The more unwanted interference there is on the measurement signal the higher the value
that should be entered here. Result: Fast changes will be damped/suppressed.
Only visible if signal = current, voltage, resistance thermometer or thermocouple.
Comparison point
Description Internal: Compensation of the voltage error by measuring the terminal temperature.
External: Compensation of the voltage error by using an external controlled comparison
measurement point.
Only visible if signal = thermocouple.
Comparison temp.
Description Settings for the external comparison temperature (only when connecting thermocouples).
Only visible if comparison point = external.
142 Endress+Hauser
Memograph M, RSG45 Appendix
Totalizer
(Online configuration)
Description Initial setting for the totalizer. Useful when continuing measurements recorded to date
with an (electro)-mechanical counter.
Only visible if signal = pulse counter or counter for Modbus Master
Factory setting 0
"Linearization" submenu
Linearization
Factory setting No
Number of Points
Endress+Hauser 143
Appendix Memograph M, RSG45
Description Specify how many support points your linearization table has.
Note: The first and last point always has to correspond to the start and end of the
measuring range respectively.
User entry 2 to 32
Factory setting 2
Zoom start
Description If the total transmitter range is not used, you can enter the lower value of the required
section here (higher resolution).
Example: Transmitter 0-14 pH, required section: 5-9 pH. Set "5" here. Zooming does not
affect the save function.
Factory setting 0
Zoom end
Description Like "Zoom start". However, enter the upper value of the required range here.
Example: Transmitter 0-14 pH, required section: 5-9 pH. Input here: "9".
144 Endress+Hauser
Memograph M, RSG45 Appendix
Points
Sort table
Factory setting No
Check table
Description Here you can check whether the linearization table has been entered correctly.
Factory setting No
x-value (1-32)
Description x-value for the linearization (value from the device input). e.g. 10 cm corresponds to 20
liters --> enter 10.
Endress+Hauser 145
Appendix Memograph M, RSG45
Factory setting 0
y-value (1-32)
Description Enter the y-value that the measured x-value corresponds to e.g. 10 cm corresponds to 20
liters --> enter 20.
Factory setting 0
Navigation Expert → Inputs → Universal inputs → Universal input x → Meas. val. corrct
Offset
Navigation Expert → Inputs → Universal inputs → Universal input x → Meas. val. corrct→ Offset
Direct access code: 220050-0xx
Examples: Universal input 1: 220050-000; Universal input 12: 220050-011
Description This offset is only effective on the analog input signal (no maths/bus channels).
Only visible if signal = resistance thermometer or thermocouple.
Factory setting 0
Correction RPT
146 Endress+Hauser
Memograph M, RSG45 Appendix
Navigation Expert → Inputs → Universal inputs → Universal input x → Meas. val. corrct →
Correction RPT
Direct access code: 220057-0xx
Examples: Universal input 1: 220057-000; Universal input 12: 220057-011
Description Rear panel temperature correction value for this analog input (only required for
thermocouples).
Only visible/editable if the service code has been entered.
Range start
Target value
Navigation Expert → Inputs → Universal inputs → Universal input x → Meas. val. corrct → Target
value
Direct access code: 220052-0xx
Examples: Universal input 1: 220052-000; Universal input 12: 220052-011
Description Enter the lower set point here (e.g. measuring range 0°C to 100°C: 0°C).
Only visible if signal = current or voltage.
Factory setting 0
Actual value
Navigation Expert → Inputs → Universal inputs → Universal input x → Meas. val. corrct → Actual
value
Direct access code: 220053-0xx
Examples: Universal input 1: 220053-000; Universal input 12: 220053-011
Description Enter the lower value actually measured here (e.g. measuring range 0°C to 100°C:
measured value 0.5°C).
Only visible if signal = current or voltage.
Factory setting 0
Endress+Hauser 147
Appendix Memograph M, RSG45
Target value
Navigation Expert → Inputs → Universal inputs → Universal input x → Meas. val. corrct → Target
value
Direct access code: 220055-0xx
Examples: Universal input 1: 220055-000; Universal input 12: 220055-011
Description Enter the upper set point here (e.g. measuring range 0°C to 100°C: 100°C).
Only visible if signal = current or voltage.
Actual value
Navigation Expert → Inputs → Universal inputs → Universal input x → Meas. val. corrct → Actual
value
Direct access code: 220056-0xx
Examples: Universal input 1: 220056-000; Universal input 12: 220056-011
Description Enter the upper value actually measured here (e.g. measuring range 0°C to 100°C:
measured value 100.5°C).
Only visible if signal = current or voltage.
"Totalization" submenu
Totalization
148 Endress+Hauser
Memograph M, RSG45 Appendix
Description By totalizing the analog signal (e.g. flow rate in m³/h) quantities (in m³) can be calculated.
Factory setting No
Totalization base
Description Select the required time base. Example: ml/s -> time base seconds (s); m³/h -> time base
hours (h).
Only visible if totalization = yes.
Unit
Description Enter the unit for the calculated quantity (e.g. "m³").
Only visible if totalization = yes.
Navigation Expert → Inputs → Universal inputs → Universal input x → Totalization → Low flow
cut off
Direct access code: 220033-0xx
Examples: Universal input 1: 220033-000; Universal input 12: 220033-011
Description If the volume flow recorded is below the set value, these quantities are not added to the
counter.
If the input is scaled from 0 to y, or if the pulse input is used, all values that are smaller
than the set value are not recorded.
If the input is scaled from -x to +y, all values around the zero point (e.g. also negative
values) are not recorded.
Only visible if totalization = yes.
Endress+Hauser 149
Appendix Memograph M, RSG45
Factory setting 0
Calc. factor
Description Factor for calculating the integrated value (e.g. the transmitter delivers l/s -> totalization
base = second -> engineering unit required is m³ -> enter factor 0.001)
Only visible if totalization = yes.
Totalizer
(Online configuration)
Description Initial setting for the totalizer. Useful when continuing measurements recorded to date
with an (electro)-mechanical counter.
Only visible if totalization = yes.
Factory setting 0
Description Contains settings that define how this channel is to react under fault conditions (e.g. cable
open circuit, over range).
NAMUR NE 43
150 Endress+Hauser
Memograph M, RSG45 Appendix
Navigation Expert → Inputs → Universal inputs → Universal input x → Fault mode → NAMUR
NE 43
Direct access code: 220060-0xx
Examples: Universal input 1: 220060-000; Universal input 12: 220060-011
Description Activate/deactivate the 4-20mA loop monitoring as per NAMUR recommendation NE 43.
The following error ranges apply when NAMUR NE43 is switched on:
≤ 3.8 mA: under range
≥ 20.5 mA: over range
≤ 3.6 mA or ≥ 21.0 mA: sensor error
≤ 2 mA: cable open circuit
Only visible if signal = "Current" and range = "4-20 mA" or "4-20 mA squared".
Options Off, On
Factory setting On
Navigation Expert → Inputs → Universal inputs → Universal input x → Fault mode → Cable open
circ.
Direct access code: 220060-0xx
Examples: Universal input 1: 220060-000; Universal input 12: 220060-011
Options Off, On
Factory setting On
Navigation Expert → Inputs → Universal inputs → Universal input x → Fault mode → Lower error
value
Direct access code: 220065-0xx
Examples: Universal input 1: 220065-000; Universal input 12: 220065-011
Description When NE43 is switched off, defines the value that must be undershot for the device to
output an error.
Only visible if signal = "Current", range = "4-20 mA", and NAMUR NE 43 = "Off".
Endress+Hauser 151
Appendix Memograph M, RSG45
Navigation Expert → Inputs → Universal inputs → Universal input x → Fault mode → Upper error
value
Direct access code: 220066-0xx
Examples: Universal input 1: 220066-000; Universal input 12: 220066-011
Description When NE43 is switched off, defines the value that must be exceeded for the device to
output an error.
Only visible if signal = "Current", range = "4-20 mA", and NAMUR NE 43 = "Off".
Time delay
Navigation Expert → Inputs → Universal inputs → Universal input x → Fault mode → Time delay
Direct access code: 220064-0xx
Examples: Universal input 1: 220064-000; Universal input 12: 220064-011
Description The device only reacts (e.g. Relay switch) when this condition, open circuit/under range/
over range, has been operational for the preset time.
Only visible if NAMUR NE 43 = on.
User entry 0 to 99 s
Factory setting 0s
On error
Navigation Expert → Inputs → Universal inputs → Universal input x → Fault mode → On error
Direct access code: 220061-0xx
Examples: Universal input 1: 220061-000; Universal input 12: 220061-011
Description Configure what value the device should continue working with (for calculations) if the
measured value is not valid (e.g. cable open circuit).
In the event of an error value, all the dependent calculations are flagged accordingly
as "error value". Counters are not flagged, however!
Error value
152 Endress+Hauser
Memograph M, RSG45 Appendix
Navigation Expert → Inputs → Universal inputs → Universal input x → Fault mode → Error value
Direct access code: 220062-0xx
Examples: Universal input 1: 220062-000; Universal input 12: 220062-011
Description The device continues calculating with this value in the event of an error.
Only visible if On error = Error value.
Factory setting 0
Save event
Navigation Expert → Inputs → Universal inputs → Universal input x → Fault mode → Save event
Direct access code: 220063-0xx
Examples: Universal input 1: 220063-000; Universal input 12: 220063-011
Factory setting No
Copy settings
Description Setting up only required if the digital inputs (e.g. events) are to be used.
x = place holder for selected digital input.
Add input
Endress+Hauser 153
Appendix Memograph M, RSG45
Description Addition of a digital input that must be configured according to the function.
Factory setting No
Delete input
Factory setting No
Function
Description Select the required function: Digital inputs are High active; this means the described effect
is achieved by a high input.
Low = -3 to +5V
High = +12 to +30V.
Options Switched off, control input, on/off event, pulse counter, operational time, event+operation
time, quantity from time, Profibus DP (option), Modbus slave (option), EtherNet/IP
(option), PROFINET (option)
Function
154 Endress+Hauser
Memograph M, RSG45 Appendix
Options Switched off, Control input, On/off event, Pulse counter, Operational time, Event
+operation time, Quantity from time
Channel ident.
Description Measurement point name (e.g. "Pump") or description of the function of this input (e.g.
"Fault message").
Only visible if function ≠ switched off.
Engineering unit
Decimal point
Options None, One (X.Y), Two (X.YY), Three (X.YYY), Four (X.YYYY), Five (X.YYYYY)
Endress+Hauser 155
Appendix Memograph M, RSG45
Input factor in
Pulse value
Description Factor, that when multiplied by the input signal results in the required physical value.
Examples:
1 pulse equals 5 m3 -> enter "5".
Only visible if function = pulse counter.
Factory setting 1
Description Factor, that when multiplied by the operating time results in the required physical value.
Examples:
1 second equals 8 l -> enter "8".
Only visible if function = quantity from time.
Factory setting 1
Time delay
156 Endress+Hauser
Memograph M, RSG45 Appendix
Description The high signal must be active for at least the preset time before the device changes the
channel from low to high.
The change from high to low is always immediate.
Only visible if function = control input, on/off event, event+operation time.
Factory setting 0
Action
Action Description
Start/stop recording The device only saves data as long as a high signal is present.
Lock setup The user can only change the setup if a low signal is present.
Time synchronization If a high signal is applied, the device rounds the system time up or down (only for
low→high change) to the nearest minute: 0 to 29 → Round down; 30 to 59 →
round up.
Change group When switching low→high the display switches to the next active group.
Set point monitoring The entire set point monitoring function of the device can be switched on (for
on/off "high") or switched off (for "low").
Individual LV on/off The monitoring of a selected limit value can be switched on (high) or off (low).
Block keyboard/navigator The device can only be operated if a low signal is present. Otherwise all key
activation and navigator actions are discarded.
Start/stop analysis 1-4 Starts/ends one of the max. 4 external analyses (the analysis runs only as long as
the signal is high). Measured value acquisition for the graphic display continues.
Batches are also started/ended with this function.
Note: This function is not available in the case of batch and control input via a
maths channel.
Reset batch number x Resets the automatically generated batch number (1..x) to 0 (in the event of a
(option) LowHigh change).
Options Switched off, Start/stop recording, Screen saver on, Lock setup, Time synchronization,
Change group, Set point monitoring on/off, Individual LV on/off, Block keyboard/
navigator, Start/stop analysis x, Reset batch no. x, Batch x limits on/off
Endress+Hauser 157
Appendix Memograph M, RSG45
Group
Description Select which group should be displayed in the event of a Low->High flank change.
Alternatively the next active group can be displayed.
Only visible if function = control input and action = change group.
Set point
Description Select the set point which should be switched on or off by means of this control input.
Only visible if function = control input and action = individual LV on/off.
Options Switched off, Universal input xx, Digital input xx, Math xx, Set point xx, Relay xx
Switches relay
Description Switches the respective relay when the digital input is low or high. Please take note of the
connection hints in the Operating Instructions!
Only visible if function = control input, on/off event, event+operation time.
158 Endress+Hauser
Memograph M, RSG45 Appendix
Description 'H'
Description Condition description when the digital input is active. This text is both shown in the
display and saved to memory.
Only visible if function = control input, on/off event, event+operation time.
Factory setting On
Description 'L'
Description Condition description when the digital input is not active. This text is both shown in the
display and saved to memory.
Only visible if function = control input, on/off event, event+operation time.
Save event
Description Determines whether the condition change from low to high or high to low is stored in the
event log.
Requires higher memory capacity.
Endress+Hauser 159
Appendix Memograph M, RSG45
Event message
Description "Do not acknowledge": No message is shown if the digital input switches.
"Acknowledge": A message window is shown on the screen which has to be acknowledged
by operating a push button.
Only visible if function = control input, on/off event, event+operation time.
Navigation Expert → Inputs → Digital inputs → Digital input x → Event text L->H
Direct access code: 250010-00x
Examples: Digital input 1: 250010-000; Digital input 6: 250010-005
Description Description of condition change from low to high. Event text is stored (e.g. Start filling).
If no event text is set, the device generates an automatic event text (factory setting),
e.g. digital 1 L->H.
Only visible if function = control input, on/off event, event+operation time.
Navigation Expert → Inputs → Digital inputs → Digital input x → Event text H->L
Direct access code: 250011-00x
Examples: Digital input 1: 250011-000; Digital input 6: 250011-005
Description Description of condition change from high to low. Event text is stored (e.g. Stop filling).
If no event text is set, the device generates an automatic event text (factory setting),
e.g. digital 1 H->L.
Only visible if function = control input, on/off event, event+operation time.
Record duration
160 Endress+Hauser
Memograph M, RSG45 Appendix
Description The duration between "On" and "Off" can be recorded. The duration is appended to the "Off"
event text (<hhhh>h<mm>:<ss>).
Power failure times do not affect the duration. If the digital channel was "on" before the
power failure and is still "on" after the power failure, the duration continues.
Only visible if function = control input, on/off event, event+operation time.
Factory setting No
Totalizer
(Online configuration)
Description Initial setting for the totalizer. Useful when continuing measurements recorded to date
with an (electro)-mechanical counter.
Only visible if function = pulse counter, operational time, event+operation time or quantity
from time.
Factory setting 0
Copy settings
Factory setting No
Endress+Hauser 161
Appendix Memograph M, RSG45
Description Settings for the universal output selected (current or pulse output).
Signal
Reference channel
Options Switched off, Universal input x, Digital input x, Maths x, Set point x, Relay x
All active inputs are available for selection.
Start value
162 Endress+Hauser
Memograph M, RSG45 Appendix
Factory setting 0
Damping/filter
Description Time constant of the first order low pass for the output signal. This is used to prevent
strong fluctuations of the output signal.
Visible only if signal = 4-20 mA or 0-20 mA.
Pulse value
Description The pulse value specifies what quantity an output pulse corresponds to (e.g. 1 pulse = 5
liters).
Only visible if signal = pulse output.
Factory setting 1
Endress+Hauser 163
Appendix Memograph M, RSG45
Pulse width
Description The pulse width limits the maximum possible output frequency of the pulse output. Define
a fixed or dynamic pulse width.
Only visible if signal = pulse output.
Pulse width
Description You can set the pulse width in the range of 0.5 to 1000 ms here.
Only visible if signal = pulse output.
Description Here, you can correct the output current value (necessary only if the device that carries out
the further processing cannot compensate for any measurement section tolerances).
Proceed as follows:
1. On the connected device, read out the displayed value in both the upper and lower
measuring range.
2. Enter the lower and upper target and actual value.
Visible only if signal = 4-20 mA or 0-20 mA.
Target value
164 Endress+Hauser
Memograph M, RSG45 Appendix
Factory setting 0
Actual value
Description Here, enter the lower actual value which is displayed at the connected device.
Visible only if signal = 4-20 mA or 0-20 mA.
Factory setting 0
Target value
Actual value
Endress+Hauser 165
Appendix Memograph M, RSG45
Description Here, enter the upper actual value which is displayed at the connected device.
Visible only if signal = 4-20 mA or 0-20 mA.
Description Configure how the analog output should behave in the event of an error (e.g. if the input
channel has a cable open circuit).
Visible only if signal = 4-20 mA or 0-20 mA.
NAMUR NE 43
Description Activate/deactivate the 4-20 mA loop output as per NAMUR recommendation NE 43. The
following error ranges apply when NAMUR NE43 is switched on:
≤3.8 mA: underrange
≥20.5 mA: overrange
≤ 3.6 mA or ≥ 21.0 mA: Cable open circuit
Visible only if signal = 4-20 mA or 0-20 mA.
Options Off, On
Factory setting On
On error
Description What value should the output assume in the event of an error (e.g. cable open circuit or
calculated value invalid)?
Visible only if signal = 4-20 mA or 0-20 mA.
Error value
166 Endress+Hauser
Memograph M, RSG45 Appendix
Description This value is output in the event of an error. Note: Must be between 0 and 22 mA.
Visible only if signal = 4-20 mA or 0-20 mA.
User entry 0 to 22 mA
Factory setting 0 mA
Operating mode
Identifier
Remote controlled
Endress+Hauser 167
Appendix Memograph M, RSG45
Description Configure whether the relay may be controlled remotely (e.g. PC or SMS).
Only visible with the option "Tele-Alarm".
Factory setting No
Timeout
Description Timeout is modifiable between 1 and 99 seconds. 0 seconds means that the functionality
is inactivated.
Factory setting 0s
Switches
Description After the given timeout the dedicated relay/OC is active while no readout of actual
measured values are in process.
Timeout fieldbus
168 Endress+Hauser
Memograph M, RSG45 Appendix
Description Time within which measured values must be received via fieldbus (otherwise an error will
be set). Not relevant if only measured values are read out.
Factory setting 10 s
"Ethernet" submenu
Description Contains the set-up required if you are using the Ethernet interface of the unit.
MAC-address
(Online configuration)
DHCP
Description The device can get its Ethernet settings through DHCP.
Caution: The settings determined are not displayed until after setup acceptance!
Note: The unit always gets the same IP address if the leasing time is set long enough
on the DHCP server. The PC software needs the IP address determined to establish a
connection!
IP address
Endress+Hauser 169
Appendix Memograph M, RSG45
Description Please enter the IP address (given from your network administrator). Please talk to your
network administrator about this.
Only editable if DHCP = no.
Subnet mask
Description Please enter the subnet mask (given from your network administrator).
Only editable if DHCP = no.
Gateway
Description Please enter the Gateway (given from your network administrator).
Only editable if DHCP = no.
Description Please enter the IP-address of the DNS server (you can get this from your network
administrator).
Is needed if you wish to send e-mails and want to use the e-mail server name instead of
the IP address (e.g. smtp.example.org).
Only editable if DHCP = no.
170 Endress+Hauser
Memograph M, RSG45 Appendix
Disable port
Port
Description The system communicates with the PC software through this communication port.
If your network is protected by a firewall, this port may have to be enabled. Please
contact your network administrator if this is the case.
OPC port
Description Values van be read via OPC server using this communication port.
If your network is protected by a firewall, this port may have to be enabled. Please
contact your network administrator if this is the case.
HART IP port
Endress+Hauser 171
Appendix Memograph M, RSG45
Description Connected HART devices can be accessed via this communication port using the
communication device type manager (DTM).
Only visible if a HART card is present.
Note: If your network is protected by a firewall, this port will have to be enabled.
Please contact your network administrator if this is the case.
The "RSG45 HART CommDTM" is required in order to access connected HART devices via
the RSG45. This establishes the connection between an FDT Frame application and a
HART device. The DTM for the connected device must also be installed in the FDT Frame
application. The "RSG45 HART CommDTM" is available from www.endress.com/rsg45.
Further information → 37
Web server
Description Switch the Web server functionality on/off. The instantaneous values can only be
displayed using an Internet browser when the web browser is activated.
A connection to the Web server can only be established via the Ethernet interface.
Description Configure the Web server or specify which functionality should be possible via Web server.
Only visible if Web server = yes.
Instantaneous value display is always possible once the Web server is switched on.
Port
172 Endress+Hauser
Memograph M, RSG45 Appendix
Factory setting 80
Setup
Firmware update
Factory setting No
Remote control
Factory setting No
Endress+Hauser 173
Appendix Memograph M, RSG45
WebDAV server
Factory setting No
Batch (option)
Factory setting No
Factory setting No
"Authentication" submenu
174 Endress+Hauser
Memograph M, RSG45 Appendix
Description Set the passwords for the various users with which the device can be accessed via Web
server.
Only relevant if the device is not protected by user administration.
Operator
ID
Password
Administrator
Endress+Hauser 175
Appendix Memograph M, RSG45
ID
Password
Service
ID
Password
176 Endress+Hauser
Memograph M, RSG45 Appendix
"Timeouts" submenu
Description Timeouts for the Web server. Settings should only be modified if slow network connections
are causing transmission problems.
The settings are only adopted if the browser has been restarted or a new tab opened.
Caution: Settings should only be changed by experts.
Connection quality
Description Configuration of typical timeout values for the Web server connection.
The default values can be changed if necessary.
Options Please select, Local network (LAN/WLAN), Wireless/mobile (fast connect.), Wireless/
mobile (slow connect)
Get timeout
Description Maximum time to load a new page before the browser ends the connection.
Factory setting 25
Endress+Hauser 177
Appendix Memograph M, RSG45
Set timeout
Description Maximum time to write a value or execute an action before the browser ends the
connection.
Factory setting 5
Put timeout
Description Maximum time to transmit files to or from the device before the browser ends the
connection.
Ping interval
Factory setting 10
Ping timeout
178 Endress+Hauser
Memograph M, RSG45 Appendix
Description Time in which the device must respond before the browser ends the connection.
Factory setting 15
Ping retry
User entry 0 to 5
Factory setting 0
Poll timeout
Factory setting 5
"HART" submenu
Description Specify the values that should be read out via HART.
Master type
Endress+Hauser 179
Appendix Memograph M, RSG45
Description Select the HART master type – usually "Primary". Select "Secondary" if another device (e.g. a
PLC) is already operating as the primary master.
Attempts on error
User entry 0 to 99
Factory setting 3
Failure mode
Description If HART communication is disrupted, the primary process variable (PV) can be calculated if
a valid current is present.
This function is not possible in Multidrop mode.
Start/end of measuring range must be correctly configured.
Add value
Factory setting No
180 Endress+Hauser
Memograph M, RSG45 Appendix
Delete value
Factory setting No
Description Specify the value that should be read out via HART.
Note: This value must then be assigned to a channel for universal inputs.
Connection
Description Select the physical channel to which the HART device is connected and from which you
wish to query the value.
Unit address
User entry 0 to 62
Factory setting 0
Endress+Hauser 181
Appendix Memograph M, RSG45
Process variable
Options Primary process variable (PV), Secondary process variable (SV), Third process variable
(TV), Fourth process variable (QV)
Channel ident.
Description Contains the set-up required if you are using the RS232 or RS485 unit interface.
Type
Description Configure how the serial interface is used. Pay attention to the connections.
Protocol
182 Endress+Hauser
Memograph M, RSG45 Appendix
Options PC software, Printer, Modbus Slave (only if type = RS485), Modbus Master (only if type =
RS485)
Baud rate
Description Transmission speed ("Baud rate") - must be the same as the settings for the PC software.
Parity
Description Parity
Only visible if protocol ≠ PC software.
Stop bits
Options 1, 2
Factory setting 1
Endress+Hauser 183
Appendix Memograph M, RSG45
Unit address
Description Every unit operating using RS232/RS485 must have an individual address (00-30).
Only visible if type = RS485.
User entry 0 to 30
Factory setting 0
Modbus
Unit address
Description Enter the device address where it should be possible to reach this device in the bus.
Only visible if Modbus = RS485.
Factory setting 1
184 Endress+Hauser
Memograph M, RSG45 Appendix
Port
Baud rate
Description Transmission speed ("Baud rate") - must be the same as the settings for the PC software.
Only visible if Modbus = RS485.
Parity
Description Parity
Only visible if Modbus = RS485.
Endress+Hauser 185
Appendix Memograph M, RSG45
Stop bits
Description Parity
Only visible if Modbus = RS485 and parity = none.
Options 1, 2
Factory setting 1
Modbus
Scan cycle
Factory setting 1s
186 Endress+Hauser
Memograph M, RSG45 Appendix
Response timeout
Description Time within which the system must receive a response from the end device.
Only visible if Modbus = RS485.
Factory setting 1s
Description Maximum number of registers that can be read out per command.
Factory setting 20
Connection attempts
User entry 1 to 10
Factory setting 1
Command distribution
Endress+Hauser 187
Appendix Memograph M, RSG45
Description Distributed over the scan cycle: The commands are uniformly distributed over the scan
cycle.
At the start of the scan cycle: The commands are sent at the start of the scan cycle with a
pause between commands. A new scan commences when the scan cycle ends.
Continuously: The commands are sent continuously with only a pause between commands.
The scan cycle is not taken into account.
Only visible if Modbus = RS485.
Options Distributed over the scan cycle, At the start of the scan cycle, Continuously
Description Time system waits between a response and a new command to be sent.
Only visible if Modbus = RS485.
Factory setting 10 ms
Description Contains the settings required if you are using the RS485 interface of the device.
Baud rate
Description Transmission speed ("Baud rate") - must be the same as the settings for the PC software.
Only visible if Modbus = RS485.
Parity
188 Endress+Hauser
Memograph M, RSG45 Appendix
Description Parity
Only visible if Modbus = RS485.
Stop bits
Description Parity
Only visible if Modbus = RS485 and parity = none.
Options 1, 2
Factory setting 1
Slave address
Description Enter the device address where it should be possible to reach this device in the bus.
Factory setting 1
Show status
Endress+Hauser 189
Appendix Memograph M, RSG45
Description The status is shown on the display in addition to the measured value. Status changes are
saved in the event log.
Factory setting No
Master In/Out
Options Not used, 1 AI-PA: 5 Byte, 2 AI-PA: 10 Byte, 3 AI-PA: 15 Byte, 4 AI-PA: 10 Word, 8 DI: 2
Byte, 1 AO-PA: 5 Byte, 2 AO-PA: 10 Byte, 3 AO-PA: 15 Byte, 4 AO-PA: 10 Word, 8 DO: 2
Byte
Byte x...y
Description Please select the value that should be used within the module from this address offset.
Options Switched off, Universal input x, Digital input x, Maths x, Set point x, Relay x
Note: All active inputs are available for selection.
190 Endress+Hauser
Memograph M, RSG45 Appendix
-->
Options Not used, Instantaneous value, State, Totalizer, Total operational time
Bit 0.0-0.7
Description Please select the value that should be used within the module from this address offset.
Options Switched off, Universal input x, Digital input x, Maths x, Set point x, Relay x
All active inputs are available for selection.
Endress+Hauser 191
Appendix Memograph M, RSG45
Function
Channel ident.
Description Measurement point name (e.g. "Pump") or description of the function of this input (e.g.
"Fault message").
Formula
192 Endress+Hauser
Memograph M, RSG45 Appendix
Application
Options Water heat quantity, Water heat difference, Steam heat quantity, Steam heat difference,
Water/glycol heat difference, Water DP flow, Steam DP flow, Liquids DP flow, Gas DP flow
Factory setting Water heat quantity or Water DP flow (depends on selected function)
Device type
Options Orifice (corner), Orifice (D/D2), Orifice (flange), Nozzle (ISA1932), Nozzle (l. radius),
Venturi nozzle, Venturi tbe, cast, Venturi tbe, bear., Venturi tbe, steel, V-cone, Pitot tube,
Gilflo
Flow
Engineering unit
Endress+Hauser 193
Appendix Memograph M, RSG45
Description Please select the unit you would like to use for scaling the selected flow input.
Only visible with energy package (option) and selected flow input.
Options m³/h, l/h, ft³/m, ft³/h, gpm, gal/h, kg/h, t/h, ton/h, lb/h
Description Please specify where the flow sensor is installed. This is important so that the correct
temperature is used for density calculation.
Only visible with energy package (option) and active flow input.
Pressure
Description Please select a pressure input. If you select "deactivated" the temperature input is used to
calculate the saturated steam conditions.
Only visible with energy package (option) and selected steam application.
Engineering unit
194 Endress+Hauser
Memograph M, RSG45 Appendix
Description Please select the unit used for scaling the selected pressure input.
Only visible with energy package (option) and selected steam application.
Only visible with energy packet (option) and active pressure input.
Options bar (a), psi (a), MPa (a), inH20 (a), bar (g), psi (g), MPa (g), inH20 (g)
Temperature (water/steam/warm)
Description Please select the temperature input used to measure the warm side (or steam line). In
steam applications, if you select "deactivated" the pressure is used to calculate the saturated
steam conditions.
Only visible with energy package (option) and selected energy function.
Temperature (steam/cold)
Description Please select the temperature input used to measure the cold side (or condensate line).
Only visible with energy package (option) and selected heat difference measurement.
Engineering unit
Description Please select the unit used for scaling the selected temperature sensors.
Only visible with energy package (option) and selected energy function.
Endress+Hauser 195
Appendix Memograph M, RSG45
Factory setting °C
Medium
Description Select a cooling medium. If the medium used is not in list, select ethylene glycol or
propylene glycol.
Only visible with energy package (option) and application = Water/Glycol heat difference.
Water/Glycol concentrat.
User entry 0 to 60 %
Factory setting 20 %
The result is
196 Endress+Hauser
Memograph M, RSG45 Appendix
Description Configure which data type the calculation returns. This setting affects how the channel
saves and is displayed.
If you add 2 analog channels, for example, the result is a "current value".
If you logically link 2 channels, for example, (digital 1 AND digital 2), the result is a "state"
(on/off).
Instantaneous value: If, for example, 2 analog channels are added (AI(1;1)+AI(1;2)), the
result is an instantaneous value.
State: The state/status of an individual analog input can be output as the result. A relay
can also be actuated as a result.
Counter: If, for example, 2 counters from digital inputs are added (DI(3;1)+DI(3;5)), the
result is a counter.
Operating time from status: The status (logical "1" or "0") of one or more digital inputs that
are connected by addition can be analyzed. If the result of the calculation is not equal to 0,
the counter for the operating time starts. The time is increased by 0.1 s every 100 ms.
Operating time from total: If several digital inputs that are configured as "operational time"
are added together, the result is the total of all the individual operating times.
Control input: The function corresponds to a digital input that has been configured as a
control input.
Options Instantaneous value, State, Counter, Operating time from status, Operating time from
total, Control input, Efficiency
Plot type
Options Instantaneous value, Average, Minimum value, Maximum value, Minimum + Maximum,
Counter, Current value + Counter
Engineering unit
Endress+Hauser 197
Appendix Memograph M, RSG45
Engineering unit
Options kW, MW, GJ/h, kBtu/m, kBtu/h, MBtu/h, ther/m, ther/h, ton, RT, kg/h, t/h, lbs/h, ton/h,
kg/m³, lb/ft³, kJ/kg, Btu/lb
Decimal point
Options None, One (X.Y), Two (X.YY), Three (X.YYY), Four (X.YYYY), Five (X.YYYYY)
Action
Action Description
Start/stop recording The device only saves data as long as a high signal is present.
Lock setup The user can only change the setup if a low signal is present.
Time synchronization If a high signal is applied, the device rounds the system time up or down (only for
low→high change) to the nearest minute: 0 to 29 → Round down; 30 to 59 →
round up.
Change group When switching low→high the display switches to the next active group.
198 Endress+Hauser
Memograph M, RSG45 Appendix
Action Description
Set point monitoring The entire set point monitoring function of the device can be switched on (for
on/off "high") or switched off (for "low").
Individual LV on/off The monitoring of a selected limit value can be switched on (high) or off (low).
Block keyboard/navigator The device can only be operated if a low signal is present. Otherwise all key
activation and navigator actions are discarded.
Start/stop analysis 1-4 Starts/ends one of the max. 4 external analyses (the analysis runs only as long as
the signal is high). Measured value acquisition for the graphic display continues.
Batches are also started/ended with this function.
Note: This function is not available in the case of batch and control input via a
maths channel.
Reset batch number x Resets the automatically generated batch number (1..x) to 0 (in the event of a
(option) LowHigh change).
Options Switched off, Start/stop recording, Screen saver on, Lock setup, Time synchronization,
Change group, Set point monitoring on/off, Individual LV on/off, Block keyboard/
navigator, Start/stop analysis x, Reset batch no. x, Batch x limits on/off
Set point
Description Select the set point which should be switched on or off by means of this control input.
Only visible if Action = Individual LV on/off.
Switches relay
Description Switches the respective relay when the digital input is low or high.
Only visible if the result is = control input or state.
Endress+Hauser 199
Appendix Memograph M, RSG45
Description 'H'
Description Condition description when the digital input is active. This text is both shown in the
display and saved to memory.
Only visible if the result is = control input or state.
Factory setting On
Description 'L'
Description Condition description when the digital input is not active. This text is both shown in the
display and saved to memory.
Only visible if the result is = control input or state.
Save event
Description Determines whether the condition change from low to high or high to low is stored in the
event log.
Requires higher memory capacity.
200 Endress+Hauser
Memograph M, RSG45 Appendix
Event message
Description "Do not acknowledge": No message is shown if the state of the mathematics channel
changes.
"Acknowledge": A message window is shown on the screen which has to be acknowledged
by operating a push button.
Only visible if the result is = control input or state.
Description Description of condition change from low to high. Event text is stored (e.g. Start filling).
Only visible if the result is = control input or state.
Description Description of condition change from high to low. Event text is stored (e.g. Stop filling).
Only visible if the result is = control input or state.
Record duration
Endress+Hauser 201
Appendix Memograph M, RSG45
Description The duration between "On" and "Off" can be recorded. The duration is appended to the "Off"
event text (<hhhh>h<mm>:<ss>).
Power failure times do not affect the duration. If the digital channel was "on" before the
power failure and is still "on" after the power failure, the duration continues.
Only visible if the result is = control input or state.
Factory setting No
Zoom start
Description If the whole value range is not used, you can configure the lower value of the required
section here. Zooming does not affect the save function.
Only visible if the result is = instantaneous value.
Factory setting 0
Zoom end
Description Like "Zoom start". However, enter the upper value of the required range here.
Only visible if the result is = instantaneous value.
Totalizer
(Online configuration)
202 Endress+Hauser
Memograph M, RSG45 Appendix
Description Initial setting for the totalizer. Useful when continuing measurements recorded to date
with an (electro)-mechanical counter.
Only visible if the result is = counter, operating time from status or operating time from
total.
Factory setting 0
Formula editor
Enter the desired calculation formula.
The formula can be any combination of arithmetic calculations and logical operations.
Analog, digital or already active mathematics channels can be used.
Formula editor
A text field with the formula currently used appears. If the field is empty a formula
has not yet been defined for the mathematics channel.
Description Individual channels can be mathematically linked and calculated with functions. The
mathematics channels calculated in this way are treated as "real" channels, regardless of
whether they are connected conventionally or via fieldbus. Enter the desired calculation
formula.
The formula can be any combination of arithmetic calculations and logical operations.
Analog and digital channels can be used, as can mathematics channels that are already
active.
A formula with up to 200 characters can be created using this editor. If the formula is
finished, click OK to close the editor and accept the formula entered. The common entry
and arithmetic operators and inputs are described in detail in the following sections.
Inputs
Inputs are described in the formula using the following syntax:
Input type (signal type;channel number)
AI Analog inputs
DI Digital inputs
MI Mathematics inputs
2 State
3 Counter/operational time
Endress+Hauser 203
Appendix Memograph M, RSG45
5 Validity:
The validity of an analog or maths channel is returned.
The relayed value of the function is 0 in the event of:
• Open circuit
• Invalid measured value
• Sensor error
• Input signal too high/low
• Error value
The relayed value of the function is 1 in the event of:
Measured value OK, even if the limit value is breached
6 Delta count
11 Totalizer
12 Duration
Not all signal types are available for each input type. These depend on the respective
device options.
Channel number:
Analog channel 1 = 1, analog channel 2 = 2, digital channel 1 = 1, …
Examples:
DI(2;4) State of digital channel 4
'
Status of a limit value:
LMT (type, limit number)
Type Description
Examples:
LMT (1;1) Instantaneous value of limit value 1
204 Endress+Hauser
Memograph M, RSG45 Appendix
Operators
Arithmetic operators:
Operator Function
+ Addition
* Multiplication
/ Division
^ x high y
Relational operators:
Operator Function
= Equal to
Logical operators:
Functi Syntax Description Example
on
&& Value 1 && Logical "and" (see also function "and") DI(2;1) && DI(2;2)
Value 2
Functions
Standard functions:
Functi Syntax Description Example
on
In the ln(number) Returns the natural logarithm of a number. Natural ln(86) = 4.454347
logarithms are based on a constant e
(2.71828182845904). For values ≤ 0 the result is
undefined. The device continues with 0
log log(number) Calculates the logarithm of the argument for base 10. For log(10) = 1
values ≤ 0 the result is undefined. The device continues
with 0.
exp exp(number) Exponentiates the base e with the number specified as the exp(2.00) =
argument. The constant e is the basis for the natural 7.389056
logarithm and has the value 2.71828182845904.
abs abs(number) Returns the absolute value of a number. The absolute value abs(-1.23) = 1.23
of a number is the number without its sign.
sqrt sqrt(number) sqrt calculates the positive square root of the argument sqrt(4) = 2
"Number". In the case of negative values the result is
undefined. The device continues with 0.
Endress+Hauser 205
Appendix Memograph M, RSG45
mod mod(number; Returns the remainder of a division. The result has the mod(5;2) = 1
divisor) same sign as the divisor. If the divisor has the value 0, the
result is undefined. The device continues with 0.
rnd rnd(number; Rounds a number to a certain number of decimal places. rnd(2.15;1) = 2.2
number_digits) "Number" is the number to which you wish to round up/ rnd(2.149;1) = 2.1
down. rnd(-1.475;2) = -1.48
"Number_digits" indicates to how many decimal places you rnd(-1.473;2) = -1.47
wish to round the number up/down. rnd(21.5;-1) = 20
rnd(5.5;-2) = 10
Notes:
• If "Number_digits" is greater than 0 (zero), the rnd(5.5;-3) = 0
number is rounded to the specified number of
decimal places.
• If "Number_digits" is equal to 0, the number is
rounded to the next whole number.
• If "Number_digits" is less than 0, the part of the
number to the left of the decimal separator is
rounded.
Trigonometric functions:
Functi Syntax Description Example
on
The following functions expect an angle in the radian as the argument. If the angle lies in
the degree, it must be converted to the radian by multiplying it with pi()/180.
Alternatively the function "rad" can be used:
206 Endress+Hauser
Memograph M, RSG45 Appendix
With the following functions the returned angle is specified in radian with a value
between –pi/2 and +pi/2. If the result is to be expressed in degrees, the respective result
must be multiplied by 180/pi() or the "Degrees" function used:
asin asin(number) Returns the arc sine or inverse sine of a number (inverse arcsin(-0.5) = -
function). The arc sine expects a real argument in the range 0.5236
from -1 to +1. In the case of values outside of this range the arcsin(-0.5)*180/
device continues with 0. pi() = -30°
acos acos(number) Returns the arc cosine or inverse cosine of a number arccos(-0.5) =
(inverse function). Arc cosine expects a real argument in 2.094395
the range from -1 to +1. In the case of values outside of this
range the device continues with 0.
atan atan(number) Returns the arc tangent or inverse tangent of a number atan(1) = 0.785398
(inverse function).
Logical functions:
Functi Syntax Description Example
on
if if(check; "Check" is any value or expression; the result can be TRUE or if(x>10;1;0)
then_value; FALSE. This argument can use any comparative calculation If value x is greater
otherwise_value operator. "Then_value" is the value returned if "Check" is than 10 the function
) TRUE. "Otherwise_value" is the value returned if "Check" is returns 1, otherwise 0
FALSE.
or or(true1;true2) Returns TRUE if an argument is TRUE. Returns FALSE if all or(2>1;3>2) = true
arguments are FALSE. or(2<1;3>2) = true
Note: See also operator "||" or(2<1;3<2) = false
and and(true1;true2 Returns TRUE if both arguments are TRUE. If one of the and(2>1;3>2) = true
) arguments is FALSE, this function returns the value FALSE and(2<1;3<2) = false
Note: See also operator "&&"
not not(truth value) Inverts the value of an argument. NOT can be used if a value not(false) = true
does not match a certain other value.
The XX in the following functions stands for one of the inputs described under → 203.
Range functions can only ever be executed with one input type.
Range functions:
Functi Syntax Description Example
on
sumX sumXX(Type;Fro Totals the values for the specified input signal range. sumXX (1;2;5) = Sum
X m;To) "Type": Signal type (see → 203)) of all instantaneous
"From": Channel number from which the values should be values from channel 2
totaled (1 = channel 1) to 5
"To": Channel number to which the values should be totaled
(1 = channel 1)
avgXX avgXX(Type;Fro Calculates the average for the specified input signal range. avgXX(1;1;6)
m;To)
minXX minXX(Type;Fro Returns the lowest value for the specified input signal minXX(1;1;6)
m;To) range.
maxX maxXX(Type;Fro Returns the highest value for the specified input signal maxXX (1;1;6)
X m;To) range.
Endress+Hauser 207
Appendix Memograph M, RSG45
Date/time functions:
Functi Syntax Description Example
on
dow dow() Returns the current day of the week as a number between 1 Sunday = 1
and 7. Monday = 2
Tuesday = 3
Wednesday = 4
Thursday = 5
Friday = 6
Saturday = 7
Decimal separator
Both the decimal point and the decimal comma can be used in the formula editor.
Thousand separators are not supported.
Differential pressure
208 Endress+Hauser
Memograph M, RSG45 Appendix
DP unit
Diameter unit
Factory setting mm
D at 20 °C
Description Internal pipe diameter (D) under design conditions at 20 °C/68 °F.
d at 20 °C
Endress+Hauser 209
Appendix Memograph M, RSG45
Description Internal pipe diameter of the throttle (d) under design conditions at 20 °C/68 °F.
K-factor
Description Set the K-factor (blockage factor) for the pitot tube (see nameplate on the probe).
Only visible if Device type= Pitot tube.
Pipe material
Density
Description Please select density input or mathematics channel, in which the density is calculated.
Only visible if application = Liquids DP flow or Gas DP flow.
210 Endress+Hauser
Memograph M, RSG45 Appendix
Density unit
Design density
Isentropic exponent
Description Input of the kappa isentropic exponent. (Required in order to calculate the expansion
number).
Only visible if application = gas DP flow.
Endress+Hauser 211
Appendix Memograph M, RSG45
"Viscosity" submenu
Description Input of the viscosity at two support points. (Needed to calculate the Reynolds number and
flow coefficient).
Only visible if application = Liquids DP flow or Gas DP flow.
Support point 1
Temperature
Description Temperature
Factory setting 0
Viscosity
Factory setting 1 cp
Support point 2
Temperature
212 Endress+Hauser
Memograph M, RSG45 Appendix
Description Temperature
Viscosity
"Totalization" submenu
Description Settings only needed if the calculated value - e.g. for quantity calculation - should be
integrated. Analysis time frames, see "Signal analysis".
Totalization
Description By totalizing the analog signal (e.g. flow rate in m³/h) quantities (in m³) can be calculated.
Factory setting No
Totalization base
Endress+Hauser 213
Appendix Memograph M, RSG45
Description Select the required time base. Example: ml/s -> time base seconds (s); m³/h -> time base
hours (h).
Only visible if totalization = yes.
Unit
Description Enter the unit for the calculated quantity (e.g. "m³").
Only visible if totalization = yes.
Description Select the unit of the quantity determined by means of totalization here.
Only visible if function = energy or mass calculation and totalization = yes.
Options kWh, MWh, MJ, GJ, kBtu, MBtu, tonh, therm, kg, t, lbs, ton
Navigation Expert → Application → Maths → Maths x → Totalization → Low flow cut off
Direct access code: 400053-00x
Examples: Maths 1: 400053-000; Maths 4: 400053-003
214 Endress+Hauser
Memograph M, RSG45 Appendix
Description If the volume flow recorded is below the set value, these quantities are not added to the
counter.
If the input is scaled from 0 to y, or if the pulse input is used, all values that are smaller
than the set value are not recorded.
If the input is scaled from -x to +y, all values around the zero point (e.g. also negative
values) are not recorded.
Only visible if totalization = yes.
Factory setting 0
Calc. factor
Description Factor for calculating the integrated value (e.g. the transmitter delivers l/s -> totalization
base = second -> engineering unit required is m³ -> enter factor 0.001)
Only visible if totalization = yes.
Totalizer
(Online configuration)
Description Initial setting for the totalizer. Useful when continuing measurements recorded to date
with an (electro)-mechanical counter.
Only visible if totalization = yes.
Factory setting 0
"Linearization" submenu
Endress+Hauser 215
Appendix Memograph M, RSG45
Linearization
Factory setting No
Number of Points
Description Specify how many support points your linearization table has.
Note: The first and last point always has to correspond to the start and end of the
measuring range respectively.
User entry 2 to 32
Factory setting 2
Zoom start
216 Endress+Hauser
Memograph M, RSG45 Appendix
Description If the total transmitter range is not used, you can enter the lower value of the required
section here (higher resolution).
Example: Transmitter 0-14 pH, required section: 5-9 pH. Set "5" here. Zooming does not
affect the save function.
Factory setting 0
Zoom end
Description Like "Zoom start". However, enter the upper value of the required range here.
Example: Transmitter 0-14 pH, required section: 5-9 pH. Input here: "9".
"Points" submenu
Check table
Description Here you can check whether the linearization table has been entered correctly.
Factory setting No
Sort table
Endress+Hauser 217
Appendix Memograph M, RSG45
Factory setting No
x-value (1-32)
Description x-value for the linearization (value from the device input). e.g. 10 cm corresponds to 20
liters --> enter 10.
Factory setting 0
y-value (1-32)
Description Enter the y-value that the measured x-value corresponds to e.g. 10 cm corresponds to 20
liters --> enter 20.
Factory setting 0
Description Contains settings that specify how this channel is to behave in the event of an error (e.g. if
an input channel has a cable open circuit or there is division by 0).
218 Endress+Hauser
Memograph M, RSG45 Appendix
Navigation Expert → Application → Maths → Maths x → Fault mode → Wet steam alarm
Direct access code: 400113-00x
Examples: Maths 1: 400113-000; Maths 4: 400113-003
On error
Description Configure what value the device should continue working with (for calculations) if the
measured value is not valid (e.g. cable open circuit).
Error value
Description The device continues calculating with this value in the event of an error.
Only visible if On error = Error value.
Factory setting 0
Copy settings
Endress+Hauser 219
Appendix Memograph M, RSG45
Factory setting No
Analysis x
Description For the set time frame, determines the minimum, maximum and average value or
quantities and operating times.
If the "Externally controlled" option is to be used, a digital input or a maths channel
must be set to "Function = Control input" and "Action = Start/stop analysis x".
Options Switched off, Externally controlled, 1min, 2min, 3min, 4min, 5min, 10min, 15min, 30min,
1h, 2h, 3h, 4h, 6h, 8h, 12h
Daily analysis, Weekly analysis, Monthly analysis, Annual analysis
Synchron. time
220 Endress+Hauser
Memograph M, RSG45 Appendix
Week starting on
Description The following data can be determined via the signal analysis cycles (e.g. daily analysis):
• How often was the set point violated
• How long was the set point violated
Factory setting No
Description Set how often the weekly- monthly- or yearly analysis are to be calculated.
"No": each individual alarm is counted.
"Yes": The number of days within the analysis cycle in which at least one limit value
violation took place (e.g. required for the number of overflows in a storm overflow tank).
Factory setting No
Reset to zero
(Online configuration)
Endress+Hauser 221
Appendix Memograph M, RSG45
Reset channel
(Online configuration)
Options Please select, Universal input x, Digital input x, Maths x, Set point x, Relay x
All active inputs are available for selection.
Description Specify whether an automatic print-out should follow at the end of an evaluation.
The printout is only made if a USB printer is connected to the device or a network
printer is available.
For supported printers please refer to the Operating Instructions.
If the Batch option is selected, the printout is configured in the Batch mode/Print
menu.
Analysis x
Description Specify whether an automatic print-out should follow at the end of an evaluation.
The print-out is only made if a USB printer is connected to the device!
For supported printers please refer to the Operating Instructions.
If the Batch option is selected, the printout is configured in the Batch mode/Print
menu.
222 Endress+Hauser
Memograph M, RSG45 Appendix
Factory setting No
"Limits" submenu
Description Limit values can monitor the measured values. A relay, for example, can be switched if a
limit value is violated.
Factory setting No
Factory setting No
Description Specify where the limit values can be changed. If you select "Outside of setup also", you can
change limit values in the "Operation" menu as well as in "Setup". This allows you to adjust
your limit values to the process even if setup is locked.
Note: This function can/should be protected by the limit value code.
Endress+Hauser 223
Appendix Memograph M, RSG45
Description View or change the setup for the selected alarm set point.
x = place holder for selected limit value
Channel/value
Description Select which input/calculated value the limit value refers to.
Options Switched off, Universal input x, Digital input x, Maths x, Set point x
Type
Options Switched off, Upper set point, Lower set point, Analysis x, Gradient dy/dt, Analysis x
frequency, Analysis x duration, Inband, Outband
224 Endress+Hauser
Memograph M, RSG45 Appendix
Hysteresis For every set point, the switch point can be controlled via a hysteresis.
The hysteresis is set as an absolute value (positive values only) in the unit of the respective channel (e.g. upper
set point = 100 m, hysteresis = 1 m: Set point on = 100 m, set point off = 99 m).
Upper set point The limit value is active if the value exceeds the configured value. The limit value is switched off if the limit
value, including hysteresis, is undershot.
Measured value
Setpoint
Hysteresis
Threshold
“off”
t
Setpoint “on”
Setpoint “off”
A0010187-EN
Lower set point The limit is active if the value drops below the configured value. The limit value is switched off if the limit value,
including hysteresis, is exceeded.
Measured value
Threshold
“off” Hysteresis
Setpoint
t
Setpoint “on”
Setpoint “off”
A0010186-EN
Endress+Hauser 225
Appendix Memograph M, RSG45
Gradient dy/dt The "Gradient" operating mode is used to monitor the temporal change of the input signal. The alarm is
triggered if the measured value reaches or exceeds the preset value.
If a positive value has been set, the limit value is monitored on increasing gradients. In the case of negative
values the decreasing gradient is monitored.
The alarm is canceled when the gradient drops below the preset value. A hysteresis is not possible in the
Gradient operating mode. The alarm can be suppressed for the set time delay (unit: seconds s) in order to
decrease the sensitivity.
Measured value
t
Tm
T0
Setpoint “on”
Setpoint“off”
A0010188-EN
Inband The limit value is violated as soon as the measured value to be checked exceeds or drops below a preset
maximum or minimum respectively. The hysteresis must be monitored on the inside of the band. For the limit
value to no longer be violated, the value must lie within the hysteresis range.
Measured value
Setpoint
Hysteresis
Hysteresis
Setpoint t
Setpoint
active
Setpoint
not active
A0010192-EN
226 Endress+Hauser
Memograph M, RSG45 Appendix
Outband The limit value is violated as soon as the measured value to be checked lies within a preset band between
minimum and maximum. The hysteresis must be monitored on the outside of the band. For the limit value to
no longer be violated, the value must lie outside of the hysteresis range.
Measured value
Hysteresis
Setpoint
Setpoint Hysteresis
t
Setpoint
active
Setpoint
not active
A0010189-EN
Special case: Hysteresis and delay In the special case that hysteresis and limit value delay are activated, one limit value is switched according to
for one limit value the following principle. If hysteresis and limit value delay are activated, the delay becomes active when a limit
value is exceeded and measures the time from which the value is exceeded. If the measured value falls below
the limit value, the delay is reset. This also occurs if the measured value falls below the limit value, but
continues to be higher than the set hysteresis value. When the limit value is exceeded again, the time delay
once more becomes active and starts measuring from 0.
Measured
Delay Delay
value
Setpoint
Hysteresis
Delay time
is reset
t
Setpoint
active
Setpoint
not active
A0010193-EN
Identifier
Endress+Hauser 227
Appendix Memograph M, RSG45
Set point
Description Limit value in the set process unit, e.g. in °C, m³/h.
Factory setting 0
Set point 2
Factory setting 0
Time span dt
Description Time span within which the signal must change by the specified value before it is
recognized as a set point.
Note: max. 60 seconds.
Only visible if type = gradient dy/dt.
User entry 0 to 60 s
Factory setting 60 s
Hysteresis (abs.)
228 Endress+Hauser
Memograph M, RSG45 Appendix
Description The alarm condition is only canceled when the signal has changed into the normal
operation range by the preset value.
Factory setting 0
Time delay
Description In order to be interpreted as an alarm the signal must exceed or undercut the preset value
by at least the time set up.
Factory setting 0s
Switches
LV messages
Description "Do not acknowledge": Alarm condition is signaled by highlighting the tag name in red (no
message is output).
"Acknowledge": In the event of an alarm, a message is also displayed. This message then
has to be acknowledged.
Endress+Hauser 229
Appendix Memograph M, RSG45
Save event
Event text LV on
Description This text (including date and time) is shown on the display and/or stored in the event log.
Only available if "LV messages" is set to "Acknowledge" or "Save message" is set to "Yes".
If no text is entered, the device generates its own text (e.g. Analog 1 > 100%).
Description The same as "Event text LV on", but on return from alarm to normal condition.
Record duration of LV on
230 Endress+Hauser
Memograph M, RSG45 Appendix
Description The duration of a set point violation can be recorded. The duration is appended to the
"limit value off" event text (format: <hhhh>h<mm>:<ss>).
Power failure times do not affect the duration. If the set point was violated before the
power failure and is still violated after the power failure, the duration continues.
Factory setting No
Reset relay
Description If LV no longer violated: the relay is switched as long as the set point is violated.
After acknowledging message: even if the set point is no longer violated, the relay remains
switched until the message has been acknowledged. If the set point is still violated when
the message is acknowledged, the relay stays switched until the set point is no longer
violated.
Up to message acknowledgment: the relay remains active until the message is
acknowledged or the set point is no longer active.
Save cycle
Endress+Hauser 231
Appendix Memograph M, RSG45
Description The user can configure whether this set point should be displayed in the graphic as a help
line (in the color of the channel).
Note: 4 lines can be shown per channel in a single group.
Factory setting No
Copy settings
Factory setting No
Description Group the analog, digital and/or mathematics channels such that you can call up
important information at the press of a button during operation (e.g. temperatures, signals
in plant unit 1).
• Maximum 8 channels per group!
• High speed storage (100ms) is only available in group 1.
232 Endress+Hauser
Memograph M, RSG45 Appendix
Identifier
Save cycle
Description Configure the save cycle with which this group should be saved in normal conditions (see
also set point/save cycle).
The save cycle is independent of the measured value display (see Operating
Instructions).
Options Off, 100ms (only for group 1), 1s, 2s, 3s, 4s, 5s, 10s, 15s, 20s, 30s, 1min, 2min, 3min,
4min, 5min, 10min, 15min, 30min, 1h
Alarm cycle
Description Configure the save cycle with which this group should be saved in an alarm condition (set
point violation).
Attention: Requires higher memory capacity.
Endress+Hauser 233
Appendix Memograph M, RSG45
Options Off, 100ms (only for group 1), 1s, 2s, 3s, 4s, 5s, 10s, 15s, 20s, 30s, 1min, 2min, 3min,
4min, 5min, 10min, 15min, 30min, 1h
Display blue
Display
Description Please select which data from the selected channel should be displayed.
If the "Everything" option is selected, the device switches cyclically between the
various values of the channel (instantaneous value, analysis 1 etc.)
Display black
234 Endress+Hauser
Memograph M, RSG45 Appendix
Display
Description Please select which data from the selected channel should be displayed.
Display red
Display
Description Please select which data from the selected channel should be displayed.
Display green
Endress+Hauser 235
Appendix Memograph M, RSG45
Display
Description Please select which data from the selected channel should be displayed.
Display violet
Display
Description Please select which data from the selected channel should be displayed.
Display orange
236 Endress+Hauser
Memograph M, RSG45 Appendix
Display
Description Please select which data from the selected channel should be displayed.
Display cyan
Display
Description Please select which data from the selected channel should be displayed.
Endress+Hauser 237
Appendix Memograph M, RSG45
Display brown
Display
Description Please select which data from the selected channel should be displayed.
Grid divisions
Description Indicates the number of lines ("amplitude grid") that should be displayed. Example: display
of 0 to 100%: select 10 divisions, display 0 to 14pH: select 14 divisions.
Options Logarithmic, 1, 2, 3, 4, 5, 6, 7, 8, 9, 10, 11, 12, 13, 14, 15, 16, 17, 18, 19, 20
Factory setting 10
Min. decade
238 Endress+Hauser
Memograph M, RSG45 Appendix
Description Set the decade from which the display should be split.
Factory setting 1
Max. decade
Curve display
Description The instantaneous values are displayed as standard for the measured value curves.
Alternatively, this current value display can also be disabled which means that more data
can be shown on the display as a result.
Curve display
Endress+Hauser 239
Appendix Memograph M, RSG45
Zoom
Description Defines the zoom that is shown in "Curves" or "Waterfall" display mode. This setting does
not affect other display modes (e.g. Curves in range, Bar graph etc.).
Options Do not display, Scroll display, Display blue, Display black, Display red, Display green,
Display violet, Display orange, Display cyan, Display brown
Bargraph
Description Configure the direction in which the bar graphs should be drawn.
240 Endress+Hauser
Memograph M, RSG45 Appendix
Description The group will always be saved or only when the allocated batch is active.
1 revolution =
Description Configure how long it takes for the circular chart to be written to once completely (one
complete revolution).
Note: The device only ever shows 1/4 of the circular chart.
"E-mail" submenu
SMTP host
Description Enter your SMTP host here. If necessary, contact your network administrator or e-mail
provider.
Endress+Hauser 241
Appendix Memograph M, RSG45
Description Specify whether the e-mail server requires a secure connection (SSL).
STARTTLS: Runs on the same TCP port as unencrypted SMTP (port 25 or 587).
SMTPS: Completely encrypted with own TCP port (465).
If necessary, contact your network administrator or e-mail provider.
Factory setting No
Port
Description Enter your SMTP port here. If necessary, contact your network administrator or e-mail
provider.
Factory setting 25
Sender
Description Enter the e-mail address of the device here (this text appears as the sender of the e-mail).
If necessary, contact your network administrator or e-mail provider.
If a valid e-mail address is not configured this might cause e-mail transmission
problems, depending on the particular provider.
User name
242 Endress+Hauser
Memograph M, RSG45 Appendix
Description Configure the user name of the e-mail account here. If necessary, contact your network
administrator or e-mail provider.
Password
Description Enter the password for authentication here. If necessary, contact your network
administrator or e-mail provider.
Description Here, enter all the e-mail addresses messages should be sent to in the event of an alarm.
Assignment to the alarms is performed later on.
E-mail address x
Endress+Hauser 243
Appendix Memograph M, RSG45
Description Specify who should receive e-mails when limit value violations occur (both on and off
messages).
Only for limit values where "Save message" is set to "Yes".
Recipient x
Description Specify who should receive e-mails when "on"/"off" messages occur (for digital inputs or
maths channels).
Only for inputs where "Save message" is set to "Yes".
Recipient x
Description Specify who should receive e-mails when errors occur (Fxxx and Sxxx messages).
Recipient x
244 Endress+Hauser
Memograph M, RSG45 Appendix
Description Specify who should receive e-mails when maintenance is required (Mxxx messages).
Recipient x
"Printer" submenu
Printer Please select the printer you wish to use. For supported printers please 540000-000
refer to the Operating Instructions.
IP address Enter the IP address of the network printer here. If you do not know the 540001-000
IP address, contact your network administrator.
Note: A DNS name may also be used.
Port Please enter the port of your network printer (available from your 540002-000
network administrator). Note: Port 9100 is usually used.
Color printer Please specify whether you are using a black and white or color printer. 540003-000
Paper size Please select the paper size of your printer. 540004-000
Print Select the print direction as per the properties of the printer used. 540006-000
direction
Endress+Hauser 245
Appendix Memograph M, RSG45
Characters/ Enter the maximum number of characters per line here. 540007-000
line
Blank rows Enter the number of blank lines required at the end of the printout to 540008-000
at the end make it easier to tear off.
Fault You can switch a relay if an error occurs during the printout. The relay 540005-000
switches remains switched until the printer is ready again or the device is
restarted.
"Serial Settings required if you are using the RS232 or RS485 interface of the 150101-000
interface" device. 150103-000
submenu
"Softkeys" submenu
Description You can configure the functions that are assigned to the softkeys of the device.
Softkey 1-3
Options Not assigned, Safe SD Card removal, Remove USB stick safely, Printout, Enter batch info,
Event log/audit trail, Historic measured values, Log on to the device (login), Log out of the
device (logout), Screenshot, Search in trace, Show analyses, Change set point, Next group,
Operation
"Texts" submenu
Description Settings only needed if you want to save text for subsequent reporting. Here, configure the
text that can be saved in the event log during operation.
Text 1-30
246 Endress+Hauser
Memograph M, RSG45 Appendix
Description Contains settings for using the device in the wastewater sector.
Detailed descriptions of this device option can be found in the associated
documentation.
Description Contains settings for alerts via a modem connected to the device or by e-mail.
Detailed descriptions of this device option can be found in the associated
documentation.
Description All recorded data are transmitted to an external WebDAV server (e.g. NAS). The format
can be specified or selected via "Setup → Advanced setup → System → External memory
-> Save as".
Enable Switch the WebDAV Client functionality on/off. When active, the device 472000-000
copies the saved measured values automatically to the configured server.
Endress+Hauser 247
Appendix Memograph M, RSG45
Port This communication port is used to communicate with the WebDAV 472002-000
Server.
enabled.
If your network is protected by a firewall, this port may have to be
Please contact your network administrator if this is the
case.
Entry: Numbers (max. 5 digits)
Factory setting: 80
User name Input of the user name that can access the WebDAV server. 472004-000
Entry: Text (max. 20 characters)
Directory Enter the directory in which the data should be saved. 472005-000
Entry: Text (max. 120 characters)
Save as "Protected format": All data are stored in a manipulation protected 472010-000
encrypted format. They can be interpreted only by the PC analysis
software provided.
"Open format": data are stored in a CSV format, this can be opened by a
number of different programs (e.g. MS Excel) (Attention: no
manipulation security).
Selection: protected format, open format (*.csv)
Factory setting: protected format
Actual diagnostics
(Online configuration)
Last diagnostics
(Online configuration)
248 Endress+Hauser
Memograph M, RSG45 Appendix
Last restart
(Online configuration)
Description Information as to when the device was last restarted (e.g. due to a power failure).
Description Events such as alarm set point infringement and power failure are listed in the correct
time sequence.
Device tag
Serial number
(Online configuration)
Description Individual serial number of the device. Please provide these details when ordering spare
parts or asking any questions about the unit.
Endress+Hauser 249
Appendix Memograph M, RSG45
Order code
(Online configuration)
Firmware Version
(Online configuration)
Description Displays the installed firmware version of the device. Please send these details with any
questions about the unit.
ENP version
(Online configuration)
Description Displays the version of the electronic nameplate. Please send these details with any
questions about the unit.
Description Displays the ENP device name (electronic name plate). Please send these details with any
questions about the unit.
250 Endress+Hauser
Memograph M, RSG45 Appendix
Device name
(Online configuration)
Description Displays the device name. Please send these details with any questions about the unit.
Manufacturer ID
(Online configuration)
Description Displays the manufacturer ID. Please send these details with any questions about the unit.
Manufacturer name
(Online configuration)
Description Displays the manufacturer name. Please send these details with any questions about the
unit.
Firmware
(Online configuration)
Description Displays the installed firmware of the device. Please send these details with any questions
about the unit.
"Simulation" submenu
Endress+Hauser 251
Appendix Memograph M, RSG45
Operating mode
Description Normal operation: Unit plots the signals from the connected measurement points.
Simulation: Instead of operating with the real measurement points the signals are
simulated (using the actual settings).
252 Endress+Hauser
Memograph M, RSG45 Index
Index
Symbols Communication (parameter) . . . . . . . . . . . . . . . . . . 130
--> Profibus DP (parameter) . . . . . . . . . . . . . . . . . . 191 Communication (submenu) . . . . . . . . . . . . . . . . . . . 168
Comparison point (parameter) . . . . . . . . . . . . . . . . . 142
0…9 Comparison temp. (parameter) . . . . . . . . . . . . . . . . 142
1 hour= (parameter) . . . . . . . . . . . . . . . . . . . . . . . . 156 Configuration softwareFieldCare . . . . . . . . . . . . . . . . 43
1 revolution = (parameter) . . . . . . . . . . . . . . . . . . . 241 Configuration Web server (submenu) . . . . . . . . . . . . 172
1 second= (parameter) . . . . . . . . . . . . . . . . . . . . . . 156 Connection quality (parameter) . . . . . . . . . . . . . . . . 177
Connection type (parameter) . . . . . . . . . . . . . . . . . . 134
A Control relays (parameter) . . . . . . . . . . . . . . . . . . . . 174
Access code (parameter) . . . . . . . . . . . . . . . . . . . . . 119 Copy settings (parameter) . . . . . . . . 153, 161, 219, 232
Acknowledging messages (parameter) . . . . . . . . . . . 125 Correction RPT (parameter) . . . . . . . . . . . . . . . . . . . 146
Action (parameter) . . . . . . . . . . . . . . . . . . . . . . . . . 157 Counter engineering unit (parameter) . . . . . . . . . . . 137
Activation code (parameter) . . . . . . . . . . . . . . . . . . 128 CSV settings (parameter) . . . . . . . . . . . . . . . . . . . . . 124
Actual diagnostics (parameter) . . . . . . . . . . . . . . . . 248 Current date/time (parameter) . . . . . . . . . . . . . . . . 113
Actual value (parameter) . . . . . . . . . . . . . 147, 148, 165 Curve display (parameter) . . . . . . . . . . . . . . . . . . . . 239
Add HART value (parameter) . . . . . . . . . . . . . . . . . . 180
Add input (parameter) . . . . . . . . . . . . . . . . . . . 131, 153 D
Add limit value (parameter) . . . . . . . . . . . . . . . . . . . 223 d at 20 °C (parameter) . . . . . . . . . . . . . . . . . . . . . . . 209
Admin ID (parameter) . . . . . . . . . . . . . . . . . . . . . . . 176 D at 20 °C (parameter) . . . . . . . . . . . . . . . . . . . . . . . 209
Admin password (parameter) . . . . . . . . . . . . . . . . . 176 Damping (parameter) . . . . . . . . . . . . . . . . . . . . . . . 142
Administrator (parameter) . . . . . . . . . . . . . . . . . . . 175 Damping/filter (parameter) . . . . . . . . . . . . . . . . . . . 163
Administrator, ID, password (parameter) . . . . . . . . . 122 Data type (parameter) . . . . . . . . . . . . . . . . . . . . . . . 136
Alarm cycle (parameter) . . . . . . . . . . . . . . . . . . . . . 233 Date (parameter) . . . . . . . . . . . . . . . . . . . . . . . 116, 118
Alarm response (parameter) . . . . . . . . . . . . . . . . . . 126 Date format (parameter) . . . . . . . . . . . . . . . . . . . . . 113
Alarm statistics (parameter) . . . . . . . . . . . . . . . . . . 221 Date/time (parameter) . . . . . . . . . . . . . . . . . . 114, 124
Analysis x (parameter) . . . . . . . . . . . . . . . . . . . . . . 220 Date/time (submenu) . . . . . . . . . . . . . . . . . . . . . . . 113
Analysis x for automatic printout (parameter) . . . . . 222 Date/time setup (submenu) . . . . . . . . . . . . . . . . . . . 112
Application (parameter) . . . . . . . . . . . . . . . . . 131, 193 Day (parameter) . . . . . . . . . . . . . . . . . . . . . . . 115, 117
Application (submenu) . . . . . . . . . . . . . . . . . . . . . . 191 Decimal point (parameter) . . . . . . . . . . . . . . . . 139, 155
Authentication (submenu) . . . . . . . . . . . . . . . . . . . . 121 Decimal separator (parameter) . . . . . . . . . . . . . . . . 110
Automatic printout (submenu) . . . . . . . . . . . . . . . . 222 Declaration of Conformity . . . . . . . . . . . . . . . . . . . . . . 9
Delete HART value (parameter) . . . . . . . . . . . . . . . . 181
B Delete input (parameter) . . . . . . . . . . . . . . . . . 132, 154
Barcode reader (submenu) . . . . . . . . . . . . . . . . . . . . 127 Delete limit value (parameter) . . . . . . . . . . . . . . . . . 223
Bargraph (parameter) . . . . . . . . . . . . . . . . . . . . . . . 240 Density (parameter) . . . . . . . . . . . . . . . . . . . . . . . . 210
Batch (parameter) . . . . . . . . . . . . . . . . . . . . . . . . . . 174 Density unit (parameter) . . . . . . . . . . . . . . . . . . . . . 211
Batch assignment (parameter) . . . . . . . . . . . . . . . . . 240 Description 'H' (parameter) . . . . . . . . . . . . . . . . 159, 200
Batch mode (parameter) . . . . . . . . . . . . . . . . . . . . . 232 Description 'L' (parameter) . . . . . . . . . . . . . . . . 159, 200
Baud rate (parameter) . . . . . . . . . . . . . . . . . . . . . . . 183 Design density (parameter) . . . . . . . . . . . . . . . . . . . 211
Begin summer time (parameter) . . . . . . . . . . . . . . . 115 Device info (submenu) . . . . . . . . . . . . . . . . . . . . . . . 249
Device name . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 251
C Device options (submenu) . . . . . . . . . . . . . . . . . . . . 128
Cable open circuit detection (parameter) . . . . . . . . . 151 Device tag . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 249
Calc. factor (parameter) . . . . . . . . . . . . . . 138, 150, 215 Device tag (parameter) . . . . . . . . . . . . . . . . . . . . . . 109
CE mark (declaration of conformity) . . . . . . . . . . . . . . . 9 Device type (parameter) . . . . . . . . . . . . . . . . . . . . . 193
CE-mark . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 106 DHCP (parameter) . . . . . . . . . . . . . . . . . . . . . . . . . . 169
Change date/time (submenu) . . . . . . . . . . . . . . . . . 114 Diagnosis list (submenu) . . . . . . . . . . . . . . . . . . . . . 249
Change set points (parameter) . . . . . . . . . . . . . . . . . 223 Diagnostic messages . . . . . . . . . . . . . . . . . . . . . . . . . 67
Channel ident. (parameter) . . . . . . . . . . . . . . . 136, 155 Diagnostics (submenu) . . . . . . . . . . . . . . . . . . . . . . 248
Channel/value (parameter) . . . . . . . . . . . . . . . . . . . 224 Diameter unit (parameter) . . . . . . . . . . . . . . . . . . . . 209
Character set (parameter) . . . . . . . . . . . . . . . . . . . . 127 Differential pressure (parameter) . . . . . . . . . . . . . . . 208
Check table (parameter) . . . . . . . . . . . . . . . . . . . . . 145 Digital inputs (submenu) . . . . . . . . . . . . . . . . . . . . . 153
Check table for linearization (parameter) . . . . . . . . . 217 Dim. linearized value (parameter) . . . . . . . . . . 144, 216
Circular chart (submenu) . . . . . . . . . . . . . . . . . . . . . 241 Direct access (parameter) . . . . . . . . . . . . . . . . . . . . 109
Clear memory (parameter) . . . . . . . . . . . . . . . . . . . 112 Disable port (parameter) . . . . . . . . . . . . . . . . . . . . . 171
Communication Display (parameter) . . . . . . . . . 234, 235, 236, 237, 238
Ethernet TCP/IP . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Endress+Hauser 253
Index Memograph M, RSG45
254 Endress+Hauser
Memograph M, RSG45 Index
Endress+Hauser 255
Index Memograph M, RSG45
256 Endress+Hauser
Memograph M, RSG45 Index
Web server . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
Function scope . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
Web server (parameter) . . . . . . . . . . . . . . . . . . . . . 172
Web server authentication (submenu) . . . . . . . . . . . 174
WebDAV Client (submenu) . . . . . . . . . . . . . . . . . . . 247
WebDAV server (parameter) . . . . . . . . . . . . . . . . . . 174
Week starting on (parameter) . . . . . . . . . . . . . . . . . 221
Wet steam alarm (parameter) . . . . . . . . . . . . . . . . . 219
Workplace safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
X
x-value (1-32) (parameter) . . . . . . . . . . . . . . . . . . . 145
x-value (1-32) for linearization (parameter) . . . . . . . 218
Y
y-value (1-32) (parameter) . . . . . . . . . . . . . . . . . . . 146
y-value (1-32) for linearization (parameter) . . . . . . . 218
Z
Zoom (parameter) . . . . . . . . . . . . . . . . . . . . . . . . . . 240
Zoom end (parameter) . . . . . . . . . . . . . . . 141, 144, 202
Zoom end for linearization (parameter) . . . . . . . . . . 217
Zoom start (parameter) . . . . . . . . . . . . . . 141, 144, 202
Zoom start for linearization (parameter) . . . . . . . . . 216
Endress+Hauser 257
www.addresses.endress.com