Ib Modulo 9
Ib Modulo 9
4 Resolution_D_
A. The minimum and maximum
5 Reproducibility_E_ operating limits that the sensor is
expected to measure accurately.
3 Precision_C_ B. Difference between the true value
1 Range_A_ and the actual value measured by
the sensor
2 Accuracy_B_ C. Degree of measurement
reproducibility
D. The smallest distinguishable input
change that can be detected with
certainty
E. How close the measurements are
when the same input is measured
repeatedly over time
Doppler is a noninvasive test that calculates blood flow in blood vessels by bouncing
high-frequency sound waves (ultrasound) off circulating red blood cells.
The heart monitor operates with the use of electrodes connected to the
patient that transform the ionic current of the human heart into low amplitude electronic
current (100uV to 5mV), in turn it is amplified to be digitized.