Chapter 2. X-Ray Photoelectron and Auger Electron Spectros
Chapter 2. X-Ray Photoelectron and Auger Electron Spectros
2.1 Introduction
In X-ray photoelectron spectroscopy (XPS) and Auger electron spectroscopy
(AES), electrons emitted after the interaction between primary X-rays or
electrons and a sample are detected. The interaction is illustrated in Fig. 2.1
for AES. The amount of electrons having escaped from the sample without
energy loss is typically measured in the range of 20 to 2000 eV. The data is
represented as a graph of intensity versus electron energy. Due to the impact
of the primary beam, the atoms in the sample are ionised, and electrons are
liberated from the surface, either as a result of the photoemission process
(XPS), or of the radiationless de-excitation in the Auger electron emission
process (AES).
Characteristic X-rays
Volum e of
1-3 µm Prim ary
Excitation
Fig. 2.1: Schematic representation of the primary beam – sample interaction in the case of an
AES analysis.
1
Although XPS and AES are comparable methods in the sense that both are
based on the use of a spectrometer to measure electrons of relatively low
energy, the main difference between the two methods consists in the source of
the primary radiation, which is necessary to provoke ionisation of the atoms.
AES makes use of an electron gun while XPS relies on soft X-rays. As a
consequence of that, one of the main differences is the lateral resolution of the
two methods. Since there is a continuous evolution in the design of the
equipment and the performance of the techniques, it is difficult to express an
absolute value for it, but for AES the lateral resolution typically is situated in
the 10 to 100 nm range, while by means of XPS only a lateral resolution of a
few to 100 µm can be reached. In both methods low energy electrons are
measured, giving rise to comparable depth and sensitivity values, which are
respectively in the order of nanometers (see Fig. 2.1) and of about 0.1 %
atomic concentration.
In this chapter, we try to give practical information on XPS and AES, so that
the reader can understand the basic principles, the characteristics and the
potential of the methods and their instrumentation. More detailed
information may be found in dedicated books. Strongly recommended are
e.g. the books of D. Briggs and M.P. Seah [1], T.A. Carlson [2], J.F. Watts [3], B.
Agius et al. [4], M. Thompson et al. [5], and V.I. Nefedov [6]. More practical
information, e.g. spectra and data, is given in [7-9]. We can also refer to the
2
proceedings of important international conferences such as ECASIA
"European Conference for Application of Surface and Interface Analysis" and
QSA "Quantitative Surface Analysis". Many papers on XPS and AES can be
found in journals. “Surface Science” and “Applied Surface Science”, for
example, mostly communicate about the fundamental physical background of
the XPS and AES analysis, while e.g. “Surface and Interface Analysis” and
“Journal of the Vacuum Science and Technology” report more on the
applications.
At the end of this chapter, a paragraph is dedicated to a literature overview of
the use of XPS and AES for the analysis of cultural heritage materials.
Photoelectron
Efermi level
2p
2s
1s
3
The experimental quantity that is measured is the kinetic energy of the
electron, which depends on the energy hν of the primary X-ray source. The
characteristic parameter for the electron is its binding energy. The relation
between these parameters is given by Eq. (2.1) :
EB = hν – EK – W (2.1)
where EB and EK are respectively the binding and the kinetic energy of the
emitted photoelectron, hν is the photon energy, and W is the spectrometer
work function. In a first approximation, the work function is the difference
between the energy of the Fermi level EF and the energy of the vacuum level
EV, which is the zero point of the electron energy scale :
W = EF – E V (2.2)
This quantity is to be determined by calibration for the spectrometer used.
From Eq. (2.1) it is clear that only binding energies lower than the exciting
radiation (1486.6 eV for Al-Kα and 1253.6 eV for Mg-Kα) are probed.
Each element has a characteristic electronic structure and thus a characteristic
XPS spectrum. Fig. 2.3 shows the XPS spectrum of Ag.
x 10 4
3.5
Ag3d5/2
3
-
Ag3d3/2
2.5
-
c/s
-AgMNN
1.5
Ag3p3/2
-
Ag3p1/2
-AgMNN
1
-Ag3s
0.5
-Ag4p
-Ag4d
-Ag4s
0
1200 1000 800 600 400 200 0
Binding Energy (eV)
Fig. 2.3 : XPS spectrum of Ag. Conditions : Al-Kα at 350 W, pass energy = 58.7 eV.
4
background originates from photoelectrons which undergo energy changes
between photoemission from the atom and detection in the spectrometer. The
observed peaks can be grouped into three types : peaks originating from
photoemission (i) from core levels, (ii) from valence levels at low binding
energies (0 to 20 eV) and (iii) from X-ray excited Auger emission (between
1100 and 1200 eV).
Although valence level spectra have their analytical value (e.g. in the study of
the electronic structure of materials), they contribute little to the analysis of
cultural heritage materials, and will not be discussed here.
The main information comes from the core level peaks and the Auger peaks.
As can be seen in Fig. 2.3, the notation for XPS and AES peaks is different.
The nomenclature employed to describe XPS and AES features is based on the
momenta associated with the orbiting paths of electrons around atomic
nuclei, indicated by the quantum numbers n, l, j. And yet the translation into
the notation is different for both techniques. XPS uses the spectroscopic
notation : first the principal quantum number (n = 1, 2, 3, …), then l = 0, 1,
2, … indicated as s, p, d, …respectively, and finally the j value given as a
suffix (1/2, 3/2, 5/2, …). The AES nomenclature on the contrary usually
follows the X-ray notation : the states with n = 1, 2, 3, … are designated K, L,
M, …, while the combinations of l and j are given the suffixes 1, 2, 3, …. Both
notations are listed in Table 2.1.
In the case of Ag, the Al-Kα radiation is only energetic enough to probe the
core levels up to the 3s shell. The non s-level peaks are doublets. This is
related to the fact that when l > 0, two possible states characterised by j arise
(see Table 2.1). More details on the difference in energy between the two
states and on their relative intensities can be found in [1]. It is also noted that
the core level peaks have different intensities and widths. The relative
intensities are governed by the ionization efficiencies of the different core
shells, designated by ionization cross section. The line width, defined as the
5
full width at half-maximum intensity (FWHM), is a convolution of several
contributions : the natural width of the core level, the width of the X-ray line
and the resolution of the analyser.
Table 2.1 : XPS and X-ray notation, respectively used for XPS and AES peaks.
The Auger peaks in the spectrum, MNN peaks in the case of Ag, originate
from the decay of the ionized atoms to their ground state. The principle of the
Auger emission is discussed in the next paragraph.
6
(a) (b)
Auger Electron
Incident
X-ray Beam
Photon
Fig. 2.4 : Schematic representation of the X-ray fluorescence (a) and Auger electron emission
(b) processes.
For the example shown here, a hole is created on the K level in the initial
ionization step. This requires a primary energy greater than the binding
energy of the electron in that shell. For the ionization to be efficient, a
primary energy of about 5 times the binding energy is taken. In practice,
typical primary energies are 5 and 10 keV. The hole can be produced by
either the primary beam, or the backscattered secondary electrons The atom
relaxes by filling the hole with an electron coming from an outer level, in the
example shown as L1. As a result, the energy difference EK – EL1 becomes
available as excess energy, which can be used in two ways. The emission of
an X-ray at that energy may occur or the energy may be given to another
electron, either in the same level or in a more shallow one, as is the case in the
example, to be ejected. The first of the two competing processes is X-ray
fluorescence, the second Auger emission. Fortunately, the probability for
Auger emission is much higher for core levels with binding energies below
about 2 keV, as illustrated in Fig. 2.5.
7
Fig. 2.5 : Relative probabilities of X-ray fluorescence and Auger emission.
8
the periodic table have a unique spectrum. Fig. 2.6 shows, as an example, the
Auger spectrum of Ag.
(b)
(a)
5 4
x 10 x 10
6 6
5.5 4
5
2
4.5
0
c/s
c/s
-2
3.5
-4
3
-6
2.5
2 -8
220 240 260 280 300 320 340 360 380 400 220 240 260 280 300 320 340 360 380 400
Kinetic Energy (eV) Kinetic Energy (eV)
Fig. 2.6 : AES spectrum of Ag, (a) direct spectrum, (b) differentiated spectrum. Conditions :
Ep = 5 keV, ∆E/E = 0.25%.
9
(a) (b)
Fig. 2.7 : Influence of back-scattering on intensity and spatial distribution [11]. Ip is peak
intensity due to primary ionization, Ib to back-scattered induced ionization. Primary beam at
(a) 5 keV, (b) 20 keV and 40° angle of incidence.
The back-scattering factors depend both on the energy and the angle of
incidence of the primary beam, and they influence the intensities as well as
the spatial distribution of the detected Auger electrons as illustrated in Fig.
2.7. The sample consists of a 40 nm thick Al layer on Au. The Al KLL peak is
shown.
In general, AES peaks are broader than XPS peaks. This is related to the
complex multiplet splitting due to the number of final states after the
transition that are permitted (see Table 2.1). The KLL series, for example,
consists of five components, i.e. KL1L1 (1 transitions), KL1L2,3 (2 transitions)
and KL2,3L2,3 (3 transitions, of which one is forbidden). Two elements
contribute to the peak broadening in solid species : peak overlap in the
multiplet structure, and solid state peak broadening.
In what follows, the analytical possibilities and the strengths and weaknesses
of XPS and AES are discussed. This is however impossible without
considering first the instrumentation
10
2.3 XPS and AES instruments
Fig. 2.8: Picture of a full set-up: 5800 Multitechnique system of Physical Electronics combining
XPS and AES.
11
X-ray Source Energy Analyzer
Hemispherical
sector analyzer
Quartz Crystal
Electron Source Monochromator
Al kα x-rays
Input Lens
15 kV electrons Multi-channel
Detector
Rowland
Circle Photoelectrons
Fig. 2.9: Schematic representation of an XPS set-up with an hemisperical sector analyser.
Electron Source
Multi-Channel
Detector
Cylindrical
Mirror Analyzer
Ion Gun
Sample
Fig. 2.10: Schematic representation of a cylindrical mirror analyser used in AES.
12
XPS and AES apparatus is in a continuous evolution, but with both
techniques the most radical progress has been the improvement of the lateral
resolution. By the introduction of field Emission AES, the level of 10 nm can
now be reached. The lateral resolution of XPS currently reaches the level of a
few micrometers, and XPS mapping facilities are becoming more frequently
available than in the past.
13
but is generally based on diffusion, sputter ion and turbo molecular pumps,
combined with auxiliary tools such as Ti sublimation pumps. The high
vacuum level imposes high requirements on the used materials. Stainless
steel is often selected for the fabrication of the analysis chamber and the joints
are usually made of Cu rings. The trajectory of the electrons is strongly
influenced by the earth's magnetic and electric fields and consequently a
screening is placed around the system. All UHV systems need occasionally to
be baked out to remove contaminants from the chamber walls, the stage and
other contact points.
14
and refocussed on the sample surface. Other benefits are the removal of
contributions to the XPS spectrum of satellite peaks and of the
Bremsstrahlung continuum coming from the X-ray spectrum of the anode.
The drawback of the use of a monochromator is a severe loss in intensity of
the primary X-rays, e.g. up to 40 % for Al-Kα radiation, and thus of the
resulting XPS peaks. Acquisition times are drastically increased, and detecting
low elemental concentrations becomes impossible. As a result, in practice a
monochromator is only used in cases where a high energy resolution,
typically 0.4 eV, is required (see paragraph 2.5.4).
Focusing of X-rays is much more complicated compared to electron beams,
and it is only recently that focussed X-ray systems enabling to reach a lateral
resolution of a few micrometers, are introduced in XPS. Depending on the
constructor of the equipment, two trends may be noted. The first possibility
is to apply an aperture system in order to select only a fraction of the emitted
photoelectrons for detection. The newest development is the so-called
microfocus monochromator, where focusing of the X-rays is realized by
intermediate of an electron gun.
15
and energy dispersion of the primary electrons are less important than for
XPS.
The main characteristic of the electron gun is its brightness. It is defined as
the number of electrons emitted per unit area and unit solid angle. It
determines the number of primary electrons impinging on the surface of the
analysed material and therefore directly determines the data acquisition time
and sensitivity of the AES system.
The simplest form of a thermionic system is a W wire in the shape of a
hairpin. A small electric current provokes heating in the wire, so that the
electrons achieve energies higher than the work function of about 4.5 eV and
are able to escape from the wire. Yet, most AES systems with a thermionic
source use a LaB6 filament because its brightness is higher compared to of W
filament sources.
A real breakthrough on the level of brightness was the introduction of field
emission sources [12]. Field emission is achieved by applying a strong
electrostatic field between a filament, in the shape of a needle with a tip
radius of about 50 nm, and an extraction electrode. The filament is usually a
wire of a W single crystal fashioned into a sharp point. The significance of the
small tip radius is that the electric field can be concentrated to extreme levels,
up to 109 V/m or more. Applying a high field results in narrowing the barrier
as well in width as in height, allowing the electrons to tunnel directly
through the barrier and omitting the requirement of thermal energy. The
small area of emission from the tip into a small solid angle provides a high
brightness compared to thermionic sources. Moreover, a more chromatic (i.e.,
mono-energetic) beam is obtained, reducing for example chromatic
aberrations in the electron lenses, so that smaller spot sizes can be achieved.
Cleanliness of the W single crystal is extremely important, since adsorption of
impurities will increase the work function. Therefore the gun compartment of
the instrument is differentially pumped.
The final spot size on the sample surface with a particular FE electron gun is a
function of the lens system, the beam current and the energy (see Fig. 2.11).
16
Beam Diameter (nm)
1000
20 kV
100 10 kV
5 kV
3 kV
10
1 10 50
Beam Current (nA)
Fig. 2.11: Relation between spot size, beam current and energy for a FE gun.
Fig. 2.12: FEAES analysis of small Cu particles. (a) SEM image, (b) AES image based on the
Cu mapping.
Note however that the obtained lateral resolution depends also on the
characteristics of the sample under investigation.
17
2.3.5 Detection of electron energy
Mainly two types of detectors are used in AES and XPS systems: the
cylindrical mirror analyser (CMA) and the hemispherical sector analyser
(HSA). In the past, the CMA was preferred for AES systems merely for
geometrical reasons and for its high analyser transmission function, and the
HSA analyser for XPS for its superior energy resolution. More and more,
however, depending on the constructor and on the requirements of the
customer, the HSA detector is also used in AES systems. A multifunctional
AES-XPS system is evidently operated by only one electron detector, and
mostly a HSA type is chosen. Yet, it is important to mention that not all of the
requirements imposed by the type of analysis coincide for XPS and AES .
18
noise ratio is improved by scanning the energy over the cylinders several
times so that counts are accumulated. After recording the spectrum in the
direct mode, it can be differentiated numerically as is shown in Fig. 2.6.
The sensitivity of the CMA is related to the analyser transmission function,
giving the effective number of electrons that are measured by the analyser for
a particular energy, and is superior compared to the HSA.
The energy resolution of a CMA is relative to the energy of the peak. Its best
relative energy resolution (about 0.25 % ∆E/E) is clearly inferior compared to
that of the HSA, as will be discussed below. Therefore, the CMA is not often
used for XPS analysis, except maybe in the past where a kind of double CMA
was introduced. Yet this configuration is no longer considered in new
systems.
The major advantage of the CMA is that there are less shadowing effects
when analysing rough surfaces and small particles, which is often the case for
cultural heritage materials (see also paragraph 2.7). This is due to the coaxial
position of the electron gun and the detector, as noted in Fig. 2.10, ensuring
that the collection of the electrons emitted from the surface is done coaxially
around the electron gun. This effect is shown in Fig. 2.13, where the SEM
picture and the AES images (see paragraph 2.5.7) of Ni spheres on an In
substrate are shown for a coaxial geometry. The SEM picture is taken by the
secondary electron detector incorporated in the AES system. The AES image
is obtained by scanning of the primary lens system over the surface. At each
position, the intensity of a number of AES peaks is measured, providing
information on the lateral distribution of elements (see also paragraph 2.5.7).
Operated in this mode, the technique is also called scanning Auger
microscopy (SAM).
A drawback of the use of an electron gun and an analyser operating under the
same angle is that angle resolved depth profiling (see paragraph 2.5.5) is not
possible.
19
(a) (b)
Fig. 2.13: Images of Ni spheres on an In substrate, (a) SEM and (b) coaxial AES imaging based
on the Ni mapping.
20
ramped, and so the electrons are counted as function of energy.
Since Auger electrons have a higher kinetic energy than photoelectrons, they
need to be more retarted. When recording XAES (X-ray induced AES) or AES
spectra, the second mode is applied. In the so-called constant retard ratio
(CRR), also known as fixed retard ratio (FRR), the voltages of the
hemispheres of the analyser are changed with the energy of the spectrum so
that the ratio of the electron kinetic energy to the pass energy is constant. In
this mode, the transmission function is optimized at the expense of the energy
resolution.
To improve the sensitivity of the HSA, the electron detection is done by a
multichannel detector system. Depending on the type of system, the number
of electron multipliers may go up to 16. This parallel electron detection is
especially useful when a monochromator is used due to the loss of intensity of
the primary X-rays.
21
differences in obtained sputter profiles.
The cold static spot gun has usually a large beam size of 5 to 10 mm and is
therefore only used for large spot XPS depth profiling or for precleaning of
the sample. The latter operation is often done in a configuration where the
ion gun is mounted in a separate chamber, called preparation vacuum
chamber. The gun is back-filled with an inert gas such as Ar having a
pressure of about 10-6 torr, with a variable potential between 1 and 10 kV. A
discharge to form Ar+ is realised by an external magnet and the positively
ions are accelerated and extracted by a simple focus electrode.
In the electron impact source, the ions are created by electrons emitted by a
heated filament, accelerated into a cylindrical grid, where they collide with Ar
gas atoms. The ion energy is controlled by the potential applied to the grid
and ion energies up to 5 kV can be achieved. This ion source produces an ion
beam with a narrow energy spread and spot sizes from 2 mm to 50 µm. It can
be operated in static mode or it can be rastered over the surface to produce a
larger and more uniform crater.
For the rapid removal of material and etching of large areas, the
duoplasmatron design of ion source is sometimes preferred. A magnetically
constricted arc is used to produce a dense plasma from which the ion beam is
extracted, focused and rastered across the specimen by a set of deflector
plates. This type of gun provides intense ion beams with a narrow energy
spread, very suitable for small spot focusing. However, due to its high price,
it is not so much used in AES or XPS systems in contrast to secondary ion
microscopy (SIMS) instruments.
22
The sample holder is mounted on a sample stage which allows for high
resolution positioning in the x,y,z and θ directions. To an increasing extent,
especially in new systems (and highly appreciated in industrial laboratories)
remote control stages are introduced. In this manner, in an automated
fashion, different sample areas may be analysed, and if a parking stage is
available, an automatic exchange of samples is possible. Triggered by the
growing interest of the semiconductor industry, the vacuum systems are
made accessible for large samples, allowing to analyse semi-conductor wafers
without preceding sample preparation. In most research instruments in use
in university laboratories, the sample size is limited in area to a few cm2 and
in thickness to a few mm.
In some systems additional sample manipulation facilities are present such as
e.g. cooling or heating of the sample holder in situ during the analysis. A
valuable tool in materials characterisation is a separate preparation chamber
connected to the analysis system, enabling sample treatments such as heating
or tensile testing. After pretreatment, the sample is introduced into the
vacuum chamber for analysis through a lock gate without coming in contact
with the air.
The angle between the ion beam, the primary beam and the detected electron
beam may induce shadowing and redeposition effects when sputtering rough
surfaces. To optimize the sputtering conditions, a sample stage with tilt and
rotation capabilities is useful. In [13] the Zalar rotation is applied to this
purpose.
23
C or Au layer as is commonly done in SEM-EDX analysis since the escape
depth of the Auger electrons is only a few nm. Thus, AES is especially
appreciated for the analysis of metals and passive films on metals, but there is
a growing tendency to analyse semi conducting and ceramic materials with it.
Our own experience shows that e.g. Al covered with a non conducting oxide
film of about 0.1 µm is measurable [13]. In other cases, special attributes, such
as a conductive grid to be placed on the sample, are necessary to limit
charging effects. Another possibility is to adapt the working conditions. This
can be done by varying the angle of incidence of the primary electrons, or,
more complicated, by performing the analysis under conditions, i.e. low
primary electron energy and low current, where the impinging primary
current is balanced by escaping electrons.
In principle the requirement of a conductive sample is less stringent with a
primary X-ray beam. However, practice shows that also in XPS charging
effects may occur when analysing non-conducting samples, especially when
using a monochromator. Therefore the constructors of XPS equipment add in
their newest configurations a neutralisation system which is able to
compensate the charge automatically.
Both for XPS and AES, the samples should be stable in a UHV system of 10-10
torr. Polymers with low vapour pressure, wet samples and porous materials
need to be handled with care. Many cultural heritage materials belong to this
category, as will be shown in paragraph 2.7.
Most operators prefer flat samples, firstly because the interpretation of the
data is more straightforward, and secondly because roughness deteriorates
the depth resolution in sputtering mode (see paragraph 2.5.5).
24
scattered. The characteristic depth d from which photoelectrons and Auger
electrons are emitted, called the escape depth, is given by:
d = λ(E) cos θ (2.5)
where θ is the angle of emission from the surface normal. The attenuation
length varies according to the element which is emitting the electron and to
the matrix, and depends on the energy of the emission. Typical values of λ
are, as shown in Fig. 2.14, in the range of 1 to 10 atom layers [14]. This is the
basis for the surface sensitivity of XPS and AES spectroscopy.
Fig. 2.14 : Dependence of attenuation length λ on the emitted electron energy [14].
25
4
(a) (b)
x 10 x 10 4
1 2
-Cr2p3/2
1.8
0.5
1.6
1.4
0
-O KLL
1.2
Cr
-Cr LMM
Cr
c/s
c/s
-0.5 1
-Cr2s
O1
0.8
-1
0.6
-Cr3p
0.4
Cr1
-1.5
-O1s
Cr2
0.2
-2 0
400 450 500 550 600 1200 1000 800 600 400 200 0
Kinetic Energy (eV) Binding Energy (eV)
Fig. 2.15 : Complementary information in the AES (a), and XPS (b) spectra of an oxidized Cr
sample. Conditions : Ep = 5 keV for AES, Al-Kα at 350 W, pass energy = 58.7 eV for XPS.
26
intensity of the primary beam, the ionization cross section, on the probability
of Auger transition in the case of AES, on the attenuation length, on the
spectrometer transmission efficiency and on the detector efficiency.
A more practical approach consists in incorporating those parameters into
sensitivity factors. The intensity of a signal from an element A, IA, in a solid
is proportional to its molar fraction xA:
xA = IA / IA0 (2.6)
where IA0 is the intensity from a pure A sample, and may be considered as a
sensitivity factor. In general what is used in the commercial data treatment
software is a set of relative sensitivity factors, normalized to a reference
element. For this reference, in most cases Cu or Ag, the sensitivity factor is set
to unity. The molar fraction xA in a homogeneous specimen composed of i
elements, is then given by :
27
ionization.
An additional difficulty encountered in quantitative analysis is the
determination of the peak areas Ii (see Eq. (2.7)). XPS peaks, and even more
pronounced AES peaks, appear on a background. In some cases, the
background correction is not straightforward. In most commercial data
handling systems different possibilities exist. The simplest one is the straight
line between two suitably chosen points. More complex methods are the
Shirley and the Tougaard background correction procedures [1].
The core level peaks in XPS show a clear shift in binding energy, the so-called
chemical shift, related with differences in the chemical environment of the
emitting element. This is shown in Fig. 2.16 for the C peak of a polyethylene
terephtelate (PET) sample.
The capability to distinguish between different chemical states is the main
characteristic of XPS. Due to this, another acronym is in use for this
technique: ESCA, which stands for ‘Electron Spectroscopy for Chemical
Analysis’.
28
1800
Peak FWHM (eV)
CH C 1s
CH 0.99
C-O 0.98
O=C-O 0.76
c/s
O=C-O
C-O
0
298 288 278
Binding Energy (eV)
Fig. 2.16 : Chemical shift of the C 1s peak as a function of its bonding with O.
The shifts are typically a few to 10 eV or more, and therefore detectors with a
high energy resolution are used (see paragraph 2.3.5.2). To subtract chemical
information , it is imperative to determine peak positions as accurately as
possible. The line of interest is preferentially evoked by means of a
monochromatic X-ray source, and recorded with the highest possible energy
resolution. When dealing with small chemical shifts, overlapping peaks may
anyway occur in the spectra. Peak deconvolution and peak fitting tools are
available in the commercial data handling systems. An illustration is seen in
Fig. 2.16, where the C 1s peak is deconvoluted in its components.
In AES, the situation is less promising. Firstly, most AES lines are by nature
broader than XPS lines. Secondly, the Auger transition involves three
electrons and the overall chemical shift is influenced by the three energy
levels concerned. In general, similar shifts of a few eV as for XPS are to be
expected, especially when core electrons are involved in the transition. Since
in AES-specific equipment, detectors with lower energy resolution are used
(see paragraph 2.3.5.1) it is clear that chemical shifts are not often measured
29
accurately. Moreover, peaks corresponding to transitions with valence
electrons are broad and poorly defined, which makes the assignment of
chemical shifts nearly impossible. In that case, the identification of the
chemical state may be done by a peak shape analysis. In fact the shape of a
CVV or CCV peak, C being a core level, is related to the density of states
(DOS) in the valence band. Since the DOS varies from one chemical
environment to another, a variation in peak shape may be observed. This
effect is commonly noticed in Auger spectra of non-metallic elements such as
C, S, O, N. Fig. 2.17 shows a few examples of chemical effect in differentiated
AES spectra.
Al LMM Al KLL
Elemental Al
dE(E)/EdN
Elemental Al
Al Oxide
Al Oxide
C KLL
40 50 60 70 80 90 1280 1300 1320 1340 1360 1380 1400 1420
Kinetic Energy (eV) Kinetic Energy (eV)
Graphitic C
Si LMM Si KLL
Elemental Si
Elemental Si
W Carbide /
50 60 70 80 90 100 110 1500 1520 1540 1560 1580 1600 1620 1640
Kinetic Energy (eV) Kinetic Energy (eV)
Fig. 2.17 : Chemical shifts and line shape effects in AES spectra of Al, Si and C.
30
AES where a an electron beam is used as primary radiation. However, also in
XPS the phenomenon is encountered. Although an internal standard, e.g. the
C 1s for XPS or the C KVV line for AES, may be used to estimate the value of
the charging shift, it is not very accurate. In order to solve this problem, the
Auger parameter was introduced [16].
The Auger parameter α is the difference between the kinetic energies of a
photoelectron line and an Auger line in the XPS spectrum . Taking Eq. (2.1)
into account, the expression for α becomes :
α = EK + EB – hν (2.8)
where EK is the kinetic energy of the sharpest Auger line and EB the binding
energy of the most intense photoelectron line. The modified Auger parameter
α’ is introduced to ensure a positive value :
α’ = α + hν = EK + EB (2.9)
Since α and α’ are related to a difference in energy measured on the same
sample in the same spectrometer, any static charge effects cancel out. The
Auger parameter has a unique value for each chemical state and can be used
as a fingerprint. Tabulations of the Auger parameter can be found in
handbooks (see, e.g., [7]).
31
High take-off angle Low take-off angle
Analyzer
e- e-
X-rays
α
α
d = λ sin α 500eV
d Ar+ Ion Beam
32
energetic ions (mostly Ar+ ions with an energy of 1 to 5 keV) , the surface
atoms are sputtered away and the residual surface is analysed. By this
technique, layers up to 1 to 2 µm are accessible.
AES or XPS spectra are recorded, either discontinuously after the subsequent
sputter steps or simultaneously with the sputtering. The original data consist
of signal intensities of the detected elements, mostly peak-to-peak heights in
AES and peak areas in XPS, as a function of sputtering time. A typical
example is shown in Fig. 2.19.
Fig. 2.19 : XPS sputter profile of a Ti oxide layer on top of a TiNb alloy. Sputter conditions : 4
kV Ar+ on a 2mm x 2mm area. Analysis conditions : Al-Kα at 350W.
33
composition, sputtering induced effects, e.g. where a component is
preferentially sputtered, may cause non-linearities in the sputter time and
depth relationship.
In the second step, peak intensities are translated into concentrations of
elements taking into account their respective sensitivity coefficients. The
obtained concentration profile deviates from the real one because differences
in escape depth of Auger or photoelectrons originating from different
elements are neglected.
34
multidimensional data sets with overlapping spectral features. They are
generated either in chemical analysis problems where the same spectral
region is studied on different samples (e.g. metal oxide, hydroxide, sulphide,
…), or either in depth profiling studies where the same sample is measured
under different conditions (e.g., a paint layer on top of a substrate). The two
most commonly applied multivariate techniques are LLSF (linear least
squares fitting) and FA (factor analysis). Both assume that each spectrum can
be represented by a linear combination of component spectra :
[D] = [R] . [C] (2.11)
where [D] is the data matrix of recorded XPS or AES spectra, [R] is the matrix
of spectra of independent components and [C] is a matrix of weighting factors
or concentrations. In LLSF, [C] is determined by a linear least squares
minimalisation. Data analysis by means of FA is more complex, and several
techniques exist, but the basic idea is first to identify the number of
independent components and then to determine the true component spectra
together with the concentrations. Illustrations of data analysis can e.g. be
found in [17,18].
2.5.7 Imaging
Distributions of elements and chemical states over the surface are measurable
using rastering techniques. Obviously, for this application AES is in favour
compared to XPS in view of its smaller analysed area (see paragraph 2.3).
Imaging in AES is also named SAM, Scanning Auger Microscopy. A well
focused incident electron beam is rastered over the surface, and the relevant
spectra are collected. Peak intensities are translated into gray scale values.
Examples are shown in Fig. 2.12 and 2.13. The smaller the spot size, the
smaller the influence of the back-scattering [1]. For spot sizes larger than 100
nm, the back-scattering phenomena may modify the images in a complex
manner. With the developments in small spot XPS, imaging in XPS is
nowadays also feasible, as illustrated in Fig. 2.20.
35
2.6 Comparison of XPS, AES and other surface analytical techniques
XPS and AES are often combined in one experimental set-up because they are
complementary. Sometimes, the system is also equipped with other sources
and detectors, so that other techniques can be applied simultaneously. A few
examples are: energy dispersive X-ray analysis (EDX), secondary ion mass
spectroscopy (SIMS), scanning tunneling microscopy (STM) or ellipsometry.
Making a list of all the available surface analytical methods is even for the
most experienced surface scientist an extremely difficult job as the number of
techniques continuously increases, especially when newer techniques such as
scanning probe or light reflection methods are considered. Each of the many
techniques has its advantages, and it is beyond the scope of this chapter to
compare them all. The main common characteristic of XPS and AES is their
high surface sensitivity. Another technique often used for the same reason is
SIMS. In Table 2.2 these techniques are compared. The characteristic values
that are listed here are only indicative since all equipment is in continuous
evolution. Moreover, some differences in performance exist between the
commercially available systems.
36
(a)
Polyethylene Mapping Area
Substrate
Adhesion Layer
Base Coat
Clear Coat
695 x 320µm
1072 x 812µm
(b)
C O Cl S
Fig. 2.20 : (a) cross section image and (b) C, O, Cl and Si XPS images of a painted substrate.
37
Table 2.2 : Comparison of the main characteristics of XPS, AES and SIMS.
characteristics XPS AES SIMS
primary beam X-rays electrons ions
analyzed beam electrons (energy) electrons (energy) ions (mass)
type of sample all, charging possible conductive all, charging possible
area of analysis 10 µm 10 nm 100 nm
surface selectivity 1 to 5 nm 1 to 5 nm 0.1 to 1 nm
elemental all except H, He all except H, He all
identification
sensitivity 0.1 % 0.1 % < 1 ppm (dynamic)
100 ppm (static)
quantification requiring standards requiring standards requiring
close standards
molecular not possible not possible mostly possible
identification
nature of chemical shift, straightforward shift and shape not possible
bonding requiring data
analysis
depth profiling elemental, chemical elemental, chemical elemental
destructive nature none if not sputtered none if not sputtered always
38
REFERENCES
[1] D. Briggs and M.P. Seah, ‘Practical Surface Analysis. Volume 1 – Auger and X-ray
Photoelectron Spectroscopy’, Second Edition, John Wiley and Sons, Chichester, 1990
[2] T.A. Carlson, ‘Photoelectron and Auger Spectroscopy’, Plenum Press, New York, 1975
[3] J.F. Watts, ‘An Introduction to Surface Analysis by Electron Spectroscopy’, Oxford
University Press, Oxford, 1990
[4] G. Hollinger, P. Pertosa, Chapter 3 in ‘Surfaces interfaces et films minces – Observation et
analyse’, Edited by B. Agius et al., Dunod, Paris, 1990
[5] M. Thompson, M.D. Baker, A. Christie, J.F. Tyson, ‘Auger Electron Spectroscopy’, John
Wiley and Sons, New York, 1985
[6] V.I. Nefedov, ‘X-ray Photoelectron Spectroscopy of Solid Surfaces’, English Edition, VSP
BV, Utrecht, 1988
[7] J.F. Moulder, W.F. Stickle, P.E. Sobol, K.D. Bomben, ‘Handbook of X-ray Photoelectron
Spectroscopy’, Edited by J. Chastain, R.C. King Jr., Physical Electronics, USA, 1995
[8] K.D. Childs, B.A. Carlson, L.A. LaVanier, J.F. Moulder, D.F. Paul, W.F. Stickle, D.G.
Watson, ‘Handbook of Auger Electron Spectroscopy’, Third Edition by C.L. Hedberg,
Physical Electronics, USA, 1995
[9] G. Beamson, D. Briggs, ‘High Resolution XPS of Organic Polymers.The Scienta ESCA300
Database’, John Wiley and Sons, Chichester, 1982
[10] M.F. Chung, L.H. Jenkins, Surf. Sci., 21 (1970) 253
[11] N.M. glezos, A.G. Nassiopoulow, Surf. Sci., 254 (1991) 314
[12] J.I. Goldstein, D.E. Newbury, P.Echlin, D.C. Joy, A.D. Romig Jr., C.E. Lyman, C. Fiori, E.
Lifshin, Chapter 2 in ‘Scanning Electron Microscopy and X-ray Microanalysis’, Second
Edition, Plenum Press, New York, 1992
[13] J. De Laet, K. De Boeck, H. Terryn, J. Vereecken, Surf. Interface Anal., 22 (1994) 175
[14] M.P. Seah, W.A. Dench, Surf. Interface Anal., 1 (1979) 2
[15] M. Detroye, F. Reniers, C. Buess-Herman, J. Vereecken, Proceedings ECASIA 97, 7th
European Conference on Applications of Surface and Interface Analysis, 16-20/06/97,
Göteborg, Sweden, Edited by I. Olefjord, L. Nyborg, D. Briggs, John Wiley, 1997, 995
[16] C.D. Wagner, L.H. Gale, R.H. Raymond, Anal. Chem., 51 (1979) 466 and references cited
herein [17] F. Reniers, A. Hubin, H. Terryn, J. Vereecken, Surf. Interface Anal., 21 (1994) 483
[18] G. Treiger, I. Bondarenko, P. Van Espen, G. Goeminne, N. Roose, H. Terryn, Proceedings
ECASIA 95, 6th European Conference on Applications of Surface and Interface Analysis, 9-
13/10/95, Montreux, Switzerland, Edited by H.J. Mathieu, B. Reihl, D. Briggs, John Wiley,
1996, 775
39
40